Radiation dose in coronary angiography and intervention: initial results from the establishment of a multi-centre diagnostic reference level in Queensland public hospitals

James A Crowhurst 1,2,, Mark Whitby 1,3, David Thiele 3, Toni Halligan 4, Adam Westerink 5, Suzanne Crown 6, Jillian Milne 7
PMCID: PMC4175852  PMID: 26229649

Abstract

Introduction

Radiation dose to patients undergoing invasive coronary angiography (ICA) is relatively high. Guidelines suggest that a local benchmark or diagnostic reference level (DRL) be established for these procedures. This study sought to create a DRL for ICA procedures in Queensland public hospitals.

Methods

Data were collected for all Cardiac Catheter Laboratories in Queensland public hospitals. Data were collected for diagnostic coronary angiography (CA) and single-vessel percutaneous intervention (PCI) procedures. Dose area product (PKA), skin surface entrance dose (KAR), fluoroscopy time (FT), and patient height and weight were collected for 3 months. The DRL was set from the 75th percentile of the PKA.

Results

2590 patients were included in the CA group where the median FT was 3.5 min (inter-quartile range = 2.3–6.1). Median KAR = 581 mGy (374–876). Median PKA = 3908 uGym2 (2489–5865) DRL = 5865 uGym2. 947 patients were included in the PCI group where median FT was 11.2 min (7.7–17.4). Median KAR = 1501 mGy (928–2224). Median PKA = 8736 uGym2 (5449–12,900) DRL = 12,900 uGym2.

Conclusion

This study established a benchmark for radiation dose for diagnostic and interventional coronary angiography in Queensland public facilities.

Keywords: Coronary angiography, coronary intervention, diagnostic reference level, radiation dose

Introduction

Cardiovascular disease (CVD) is the leading cause of death in Australia with coronary artery disease (CAD), the most common form of CVD being the largest single cause of death.1,2 Invasive coronary angiography (ICA) has been utilised in the diagnosis and treatment of CAD for over 30 years and there is consistent growth in the numbers of these procedures being performed each year.3

One disadvantage of ICA is the radiation dose to the patient from the fluoroscopy used during the procedure. At high X-ray exposures, there is a risk of deterministic radiation effects to the skin, such as erythema, permanent epilation and at very high doses, dermal atrophy and ulceration. There is evidence to suggest that these effects are being increasingly reported.4,5 There is also the increased stochastic risk by way of cancer. The risk of cancer from medical imaging procedures is largely unknown but is related to the cumulative effective dose received from imaging procedures.6 A comparison of the effective doses for medical imaging procedures, including coronary angiography and intervention, is demonstrated in Figure 1. Radiation doses delivered during ICA procedures have not changed significantly over the years and remain one of the highest of any X-ray examination in the acute care setting.7,8 By way of ensuring patient safety, there are maximum permissible X-ray outputs for cardiac catheterisation laboratory (CCL) systems in fluoroscopy mode in Queensland. However, there is no such limit for digital acquisitions, where the radiation dose can be up to 15 times that of fluoroscopy for the same beam on time.9 Operator dose from scattered radiation from the patient is also a consideration as procedures become more lengthy and complicated. The long-term effects by way of cancer to operators from long-term exposure to low-energy ionising radiation are being increasingly recognised.1012

Figure 1.

Figure 1

Relative patient effective dose for medical imaging procedures. Typical doses for medical imaging procedures. Adapted from data from Mettler et al.7 mSv, millisievert; CT, computed tomography; PCI, percutaneous coronary intervention.

Unsurprisingly, radiation protection and advisory bodies suggest that the radiation dose for radiological procedures should be monitored closely at a local, regional and national level.13 In order to keep doses low, whilst also maintaining adequate image quality, physicians and their support staff require established, evidence-based data to benchmark against and to date in Queensland and Australia, there is no benchmark. The publication of a benchmark or diagnostic reference level (DRL) has been performed numerous times by radiation regulatory bodies around the world. It is most commonly used in providing a benchmark for diagnostic radiological imaging but can equally be utilised for interventional procedures.13 They are a guide to good practice, but are neither dose limits nor thresholds that define competent performance of the operator or the equipment.14 The DRL provides physicians with a guide for which the median dose of a particular procedure type should fall below and should be used as a tool for optimising patient dose. The DRL is most commonly derived from the dose area product (DAP or PKA) value for fluoroscopy procedures and is a product of the dose output and the area exposed.13

The purpose of this study is to establish Queensland Public Facility (Queensland Health) radiation DRLs for cardiac catheter procedures, for the purpose of providing benchmarks for ongoing quality assurance and audit.

Methods

Site participation

All seven public hospitals in the state of Queensland, Australia, with a CCL were invited to participate in the study.

Procedures included

The study included patients undergoing ICA procedures from January 2013 through to April 2013 inclusive. Procedures were separated into two groups:

  • Diagnostic coronary angiography-only group (CA): adult patients undergoing coronary angiography ± left heart catheterisation and/or left ventriculography.

  • Diagnostic coronary angiography in conjunction with single vessel percutaneous coronary intervention (PCI): adult patients undergoing coronary angiography ± left heart catheterisation and/or left ventriculography plus PCI.

The study included all adult patients that matched the above criteria. All other procedures, including graft studies and complex multi-vessel PCI, were eliminated from the study.

Data collection

Radiation data, automatically stored by each X-ray machine, were prospectively entered into an electronic image and reporting system (Impax CV; Agfa Healthcare, Mortsel, Belgium) at the time of procedure for all cases. Data were extracted from the cardiac catheter laboratory image and reporting system using structured query language and exported into SPSS version 20 (IBM Corporation, Armonk, NY, USA) for analysis. Radiation data collected were as follows:

  • Examination type

  • Patient height/weight/BMI

  • Fluoroscopy time

  • Dose area product (DAP or PKA)

  • System-calculated skin surface entrance dose (KAR).

Data analysis

In line with the ICRP recommendations, the DRL was set from the entire population for each procedure and is determined by the 75th percentile of the PKA.13

Patient size is known to affect dose.4 Therefore, patient height, weight and body mass index (BMI) were additional measures for comparison. Fluoroscopy time (FT) was collected as it demonstrates ‘beam on’ time. KAR was measured because in Queensland, the trigger value for a reportable dose and subsequent patient follow-up is determined by exceeding an entrance surface dose of 5 Gy. Individual centres were de-identified and allocated a number for analysis.

Additional measures

The dose area product meter (DAP meter) housed within the X-ray tube was calibrated on all systems within the 12 months prior to the data collection for each CCL. A variation of ±25% was deemed acceptable and was in line with similar studies.15,16 CCL X-ray equipment in Queensland Health facilities (QH) is serviced by equipment vendors on a regular basis and the X-ray output of the systems is measured annually by QH radiation physicists to ensure compliance and quality assurance. Ethics approval was granted for this study by the Prince Charles Hospital Human Research Ethics Committee, Queensland Health.

Results

All seven QH facilities identified as performing ICA participated in the study – a total of twelve individual X-ray suites. Eleven of the twelve X-ray units were supplied by one manufacturer, with only one from a different manufacturer. All were <10 years old and incorporated flat detector technology.

A total of 3537 procedures fitted the criteria of the two groups and were included in the study. Of those, 2590 were CA and 947 were PCI. Overall data collection rates for the specified fields were 97.95% in the CA group and 98.33% in the PCI group.

The dose results and patient-related data for CA and PCI procedures are demonstrated in Table1. Distribution curves are demonstrated in Figure 2 for the overall populations. How the different facilities/sites contributed to the data is outlined in Table 2. The DRL, as determined by the 75th percentile of the PKA value for the study population, was 5865 uGym2 for CA, and 12,900 uGym2 for single-vessel PCI procedures. How these results compare to the literature is demonstrated in Tables 3 and 4.

Table 1.

Baseline results in this study for the two identified groups, demonstrating the various dose measures collected in this study.

Measure Coronary angoigraphy Percutaneous coronary intervention
Number of patients 2590 947
Median patient age 62.71 (54–72) 61.73 (53–71)
Median patient height (cm) 170 (177–162) 172 (165–178)
Median patient weight (kg) 83 (71.0–96.0) 82 (73–94)
Median patient BMI 28.7 (24.9–32.9) 27.8 (24.8–31.9)
Median fluoro time (min) 3.5 (2.3–6.1) 11.2 (7.7–17.4)
Median PKA (uGym2) 3908 (2489–5865) 8736 (5449–12,900)
Median KAR (mGy) 581 (374–876) 1501 (928–2224)
Calculated DRL (uGym2) 5865 12,900

Numbers in parenthesis indicate the interquartile range. BMI, body mass index; KAR, patient skin surface entrance dose; PKA, dose area product; DRL, diagnostic reference level; CA, coronary angiography; PCI, percutaneous coronary intervention.

Figure 2.

Figure 2

Distribution plots of the two groups included in this study. Distribution charts demonstrating how the distribution of PKA is non-normal in both diagnostic coronary angiography and percutaneous coronary intervention groups. The y-axis demonstrates the fraction of procedures that fell within a particular PKA range. PKA, dose area product.

Table 2.

How each site enrolled in the study contributed to the data for PKA and the DRL.

Site 1 Site 2 Site 3 Site 4 Site 5 Site 6 Site 7 All
Diagnostic coronary angiography (CA) (n = 193) (n = 230) (n = 542) (n = 331) (n = 261) (n = 571) (n = 462) (n = 2590)
Mean 3263 4301 4909 4701 3620 4740 5601 4448
Median 2774 3826 4235 4197 2899 3886 4641 3908
SD 2033 2386 3168 2769 2617 3630 4057 2952
25th % 1798 2594 2893 2827 1955 2531 2605 2489
75th % 4178 5472 6288 5788 4461 5779 7176 5865
Percutaneous coronary intervention (PCI) (n = 76) (n = 127) (n = 186) (n = 119) (n = 68) (n = 215) (n = 156) (n = 947)
Mean 8303 10,023 10,615 10,034 6912 12,350 8655 9566
Median 7071 8445 9283 8710 6528 10,801 7291 8736
Std. Dev. 5111 6177 6986 6401 3483 7580 5690 5918
25th % 4707 5424 5917 5463 4525 6753 4853 5449
75th % 9471 12,403 12,900 11,089 6528 15,973 11,789 12,900

The values given are all in uGym2. DRL, diagnostic reference level; PKA, dose area product; Std. Dev., one standard deviation; 25th %, 25th percentile; 75th %, 75th percentile.

Table 3.

Demonstration of how the results from this study compare to those previously published and the country where the data originates – diagnostic coronary angiography.

Country Author Median PKA (uGym2) 75th percentile
CA group
Switzerland (non-academic centres) Samara et al.15 5800 10,200
USA Miller et al.16 4900 8300
Turkey Bor et al.24 4910
Switzerland (academic centres) Samara et al.15 4500 9000
Italy Neofotistou et al.29 4240
Australia This study 3908 5864
Finland Neofotistou et al.29 3960
Greece Neofotistou et al.29 3800
Ireland Neofotistou et al.29 3330
Ireland D'Helft et al. 4 3100 4200
Italy Neofotistou et al.29 2820
Spain Neofotistou et al.29 2780
United Kingdom Hart et al.17 2350 2900
England Neofotistou et al.29 1910

CA, coronary angiogram; PKA, dose area product.

Table 4.

Demonstration of how the results from this study compare to those previously published and the country where the data originates – percutaneous coronary intervention.

Country Author Median PKA (uGym2) 75th Percentile
PCI group
USA Miller et al.16 11,700 19,300
Turkey Bor et al.24 10,690
Switzerland (academic centres) Samara et al.15 9000 17,000
Australia This study 8736 12,900
Italy Neofotistou et al.29 8200
Switzerland (non-academic centres) Samara et al.15 6700 12,000
Finland Neofotistou et al.29 6690
Ireland Neofotistou et al.29 4850
Italy Neofotistou et al.29 4240
Ireland D'Helft et al. 4 4200 8400
United Kingdom Hart et al.17 3600 5000
Greece Neofotistou et al.29 3900
Spain Neofotistou et al.29 3900
England Neofotistou et al.29 2710

PCI, percutaneous coronary intervention; PKA, dose area product.

Discussion

The investigation and publication of a DRL for coronary angiography has been performed in the UK every 5 years since 1992.17 This has subsequently been followed by studies in Europe18 and the United States.16 Studies in the UK have found an incremental drop in radiation dose since they started investigating and reporting radiation dose.17

Many different factors affect radiation dose in ICA procedures and these factors have been evaluated before.1922 Factors such as X-ray system set-up, operator technique and clinical practice all play a part and it may well be beneficial to locally investigate these factors on a site by site basis in future. The complexity of the procedure has also been demonstrated to significantly affect radiation dose,23,24 but again was not measured as part of this initial study.

It is advantageous to collect the height and weight of patients for these studies. Previous studies have normalised their PKA results to patient weight.25 This is in line with a paper by Chapple et al. to an average of 70 kg.26 Normalising the dose to a particular patient weight is advantageous for a comparison between X-ray systems. Normalising the dose data to a reference patient weight is difficult with a large number of facilities. Most multi-centre studies have not elected to do this16,2729 and this was not performed as part of this study. A higher PKA value than some of the literature may be the result, due to the higher median patient weight of 83 kg seen in this study. Another method utilised to normalise the PKA data is to exclude patients outside a weight range of 80 ± 5 kg, with the same aim.17 Normalising may not be logical in establishing a DRL for practical routine use, as the DRL would only be relevant to a small portion of patients that fall within a certain weight range.

The results obtained from this study are comparable with those from other studies in the literature. Tables 3 and 4 show how this study compares to similar studies. It is evident that there is a great variation between these multi-centre studies. As an example, reference levels for CA range from between 2900 and 10,200 uGym2, with this study sitting just below the average of 6744 uGym2 at 5865 uGym2. Neofotistou et al. attributed higher doses to teaching hospitals in their study,29 as it is known that in their first year of training, operators use higher levels of radiation due to extended fluoroscopy.30 It is noteworthy therefore that all seven facilities involved in this study are teaching hospitals, training registrars and fellows in ICA, and this has the potential to affect the results. Neofotistou et al. also attributed the possible difference in dose between the highest and lowest centres in its study to the dose rate under fluoroscopy.28 Although dose rates of individual X-ray units were not measured in this study, all units are measured annually in terms of fluoroscopy dose, which is governed by state regulations. Sample size may also be important. There are differences in study population size within the literature, with sample sizes for CA examinations ranging from 311 patients15 to 34,236 patients.17 It has been proposed that 50 examinations from each facility will produce sufficient statistical power in these kind of studies.31 With that in mind, the sample sizes of 2590 CA examinations and 947 PCI examinations seen in this study should make one confident that the DRLs calculated here are based on a sufficiently sound sample size.

This study would indicate that Queensland public facilities are delivering appropriate levels of radiation to patients during ICA procedures. However, there is always work that can be done to reduce dose and audits such as this are a good starting point in raising awareness. They are a fundamental foundation for ongoing audits as part of a quality assurance programme for radiation dose.14

In line with recommendations,14 it is planned that doses for cardiac angiography and intervention be measured against benchmarks on an annual basis and for a study such as this to be repeated in 3–5 years. The DRL should be seen a guideline only, as it is well recognised that fluoroscopy procedures are difficult to benchmark and that each individual procedure may deviate from the DRL for many legitimate reasons.13 DRLs are intended to provide guidance on what is achievable with current good practice rather than optimum performance.14 However, to date, there is no DRL for these procedures in Australia and this study could be used as an interim yardstick for other cardiac catheter laboratories in Queensland and Australia, until a larger national study can be performed.

Conclusion

This study allowed for the calculation of a DRL for diagnostic and interventional coronary angiography procedures in Queensland health facilities. The establishment of a benchmark means that these DRLs can be used for ongoing audit in these and new facilities across Queensland and potentially other similar facilities in Australia.

Acknowledgments

Financial support for this project was provided by the Allied Health Professions' Office of Queensland, Queensland Health. The authors thank Mr Ian Smith for his advice and technical support. The authors also thank the State-wide Cardiac Clinical Network for their support and the individual staff of each facility involved in collection of the data. The following individuals were responsible for data collection and access of data at each facility: Steven Coverdale and Kim Duffy (Cairns Base Hospital, Cairns, Qld); Rohan Jayasinghe and Michael Trikilis (Gold Coast Hospital, Southport, Qld); Carmel McErlean and Peter Larsen (Nambour Hospital, Nambour, Qld); Paul Garrahy (Princess Alexandra Hospital, Woolloongabba, Qld); Peter Stewart (Royal Brisbane and Women's Hospital, Herston, Qld); Michael Savage, Christopher Raffel and Darren Walters (The Prince Charles Hospital, Chermside, Qld); Bernard McGuire and Raiban Yadav (The Townsville Hospital, Townsville, Qld).

Conflict of Interest

The authors declare no conflict of interest.

References

  1. Australian Institute of Health and Welfare. 2008. Australia's Health 2008. Australia's health 2008. Cat. no. AUS 99. Canberra: AIHW. (accessed 26 August 2013). Available from: http://www.aihw.gov.au/publication-detail/?id=6442468102.
  2. Reid CM, Ademi Z, Nelson MR, et al. Outcomes from the REACH Registry for Australian general practice patients with or at high risk of atherothrombosis. Med J Aust. 2012;196:193–7. doi: 10.5694/mja11.10731. [DOI] [PubMed] [Google Scholar]
  3. Tsapaki V, Ahmed NA, AlSuwaidi JS, et al. Radiation exposure to patients during interventional procedures in 20 countries: initial IAEA project results. AJR Am J Roentgenol. 2009;193:559–69. doi: 10.2214/AJR.08.2115. [DOI] [PubMed] [Google Scholar]
  4. D'Helft CJ, Brennan PC, McGee AM, et al. Potential Irish dose reference levels for cardiac interventional examinations. Br J Radiol. 2009;82:296–302. doi: 10.1259/bjr/14857366. [DOI] [PubMed] [Google Scholar]
  5. Koenig TR, Wolff D, Mettler FA, et al. Skin injuries from fluoroscopically guided procedures: part 1, characteristics of radiation injury. AJR Am J Roentgenol. 2001;177:3–11. doi: 10.2214/ajr.177.1.1770003. [DOI] [PubMed] [Google Scholar]
  6. Picano E, Vano E, Rehani MM, et al. The appropriate and justified use of medical radiation in cardiovascular imaging: a position document of the ESC Associations of Cardiovascular Imaging, Percutaneous Cardiovascular Interventions and Electrophysiology. European Heart Journal. 2014;35:665–72. doi: 10.1093/eurheartj/eht394. [DOI] [PubMed] [Google Scholar]
  7. Mettler FA, Huda W, Yoshizumi T, et al. Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology. 2008;248:254–63. doi: 10.1148/radiol.2481071451. [DOI] [PubMed] [Google Scholar]
  8. Wagner LK, Archer BR, Cohen AM. Management of patient skin dose in fluoroscopically guided interventional procedures. J Vasc Interv Radiol. 2000;11:25–33. doi: 10.1016/s1051-0443(07)61274-3. [DOI] [PubMed] [Google Scholar]
  9. Hirshfeld JW, Jr, Balter S, Brinker JA, et al. ACCF/AHA/HRS/SCAI clinical competence statement on physician knowledge to optimize patient safety and image quality in fluoroscopically guided invasive cardiovascular procedures: a report of the American College of Cardiology Foundation/American Heart Association/American College of Physicians Task Force on Clinical Competence and Training. Circulation. 2005;111:511–32. doi: 10.1161/01.CIR.0000157946.29224.5D. [DOI] [PubMed] [Google Scholar]
  10. Buchanan GL, Chieffo A, Mehilli J, et al. The occupational effects of interventional cardiology: results from the WIN for safety survey. EuroIntervention. 2012;8:658–63. doi: 10.4244/EIJV8I6A103. [DOI] [PubMed] [Google Scholar]
  11. Roguin A, Goldstein J, Bar O. Brain malignancies and ionising radiation: more cases reported. EuroIntervention. 2012;8:169–70. doi: 10.4244/EIJV8I1A26. [DOI] [PubMed] [Google Scholar]
  12. Roguin A, Goldstein J, Bar O. Brain tumours among interventional cardiologists: a cause for alarm? Report of four new cases from two cities and a review of the literature. EuroIntervention. 2012;7:1081–6. doi: 10.4244/EIJV7I9A172. [DOI] [PubMed] [Google Scholar]
  13. Committee 3 ICRP. 2012. pp. 1–7. Diagnostic reference Levels in medical imaging:Review and additional advice. (accessed 04/08/2013). Available from: http://www.icrp.org/docs/DRL_for_web.pdf.
  14. Cousins C, Miller DL, Bernardi G, et al. ICRP Publication 120: radiological protection in cardiology. Ann ICRP. 2013;42:1–125. doi: 10.1016/j.icrp.2012.09.001. [DOI] [PubMed] [Google Scholar]
  15. Samara ET, Aroua A, De Palma R, et al. An audit of diagnostic reference levels in interventional cardiology and radiology: are there differences between academic and non-academic centres? Radiat Prot Dosimetry. 2012;148:74–82. doi: 10.1093/rpd/ncq600. [DOI] [PubMed] [Google Scholar]
  16. Miller DL, Hilohi CM, Spelic DC. Patient radiation doses in interventional cardiology in the U.S.: advisory data sets and possible initial values for U.S. reference levels. Med Phys. 2012;39:6276–86. doi: 10.1118/1.4754300. [DOI] [PubMed] [Google Scholar]
  17. Hart D, Hillier MC, Wall BF. National reference doses for common radiographic, fluoroscopic and dental X-ray examinations in the UK. Br J Radiol. 2009;82:1–12. doi: 10.1259/bjr/12568539. [DOI] [PubMed] [Google Scholar]
  18. Aroua A, Rickli H, Stauffer JC, et al. How to set up and apply reference levels in fluoroscopy at a national level. Eur Radiol. 2007;17:1621–33. doi: 10.1007/s00330-006-0463-3. [DOI] [PubMed] [Google Scholar]
  19. Larrazet F, Dibie A, Phillipe F, et al. Factors influencing fluoroscopy time and dose-area product values during ad hoc one-vessel percutaneous coronary angioplasty. Br J Radiol. 2003;76:473–7. doi: 10.1259/bjr/21553230. [DOI] [PubMed] [Google Scholar]
  20. Azpiri-lopez J, Assad-Morell J, Gonzalez-Gonzalez J, et al. Effect of physician training on the X-ray dose delivered during coronary angiography. J Invasive Cardiol. 2013;25:109–13. [PubMed] [Google Scholar]
  21. Delewi R, Hoebers LP, Ramunddal T, et al. Clinical and procedural characteristics associated with higher radiation exposure during percutaneous coronary interventions and coronary angiography. Circ Cardiovasc Interv. 2013;6:501–6. doi: 10.1161/CIRCINTERVENTIONS.113.000220. [DOI] [PubMed] [Google Scholar]
  22. Park EY, Shroff AR, Crisco LV, et al. A review of radiation exposures associated with radial cardiac catheterisation. EuroIntervention. 2013;9:745–53. doi: 10.4244/EIJV9I6A119. [DOI] [PubMed] [Google Scholar]
  23. Bernardi G, Padovani R, Morocutti G, et al. Clinical and technical determinants of the complexity of percutaneous transluminal coronary angioplasty procedures: analysis in relation to radiation exposure parameters. Catheter Cardiovasc Interv. 2000;51:1–9. doi: 10.1002/1522-726x(200009)51:1<1::aid-ccd1>3.0.co;2-k. [DOI] [PubMed] [Google Scholar]
  24. Bor D, Olgar T, Toklu T, et al. Patient doses and dosimetric evaluations in interventional cardiology. Physica Med. 2009;25:31–42. doi: 10.1016/j.ejmp.2008.03.002. [DOI] [PubMed] [Google Scholar]
  25. Bleeser F, Hoornaert MT, Smans K, et al. Diagnostic reference levels in angiography and interventional radiology: a Belgian multi-centre study. Radiat Prot Dosimetry. 2008;129:50–5. doi: 10.1093/rpd/ncn028. [DOI] [PubMed] [Google Scholar]
  26. Chapple CL, Broadhead DA, Faulkner K. A phantom based method for deriving typical patient doses from measurements of dose-area product on populations of patients. Br J Radiol. 1995;68:1083–6. doi: 10.1259/0007-1285-68-814-1083. [DOI] [PubMed] [Google Scholar]
  27. D'Helft C, McGee A, Rainford L, et al. Proposed preliminary diagnostic reference levels for three common interventional cardiology procedures in Ireland. Radiat Prot Dosimetry. 2008;129:63–6. doi: 10.1093/rpd/ncn048. [DOI] [PubMed] [Google Scholar]
  28. Padovani R, Vano E, Trianni A, et al. Reference levels at European level for cardiac interventional procedures. Radiat Prot Dosimetry. 2008;129:104–7. doi: 10.1093/rpd/ncn039. [DOI] [PubMed] [Google Scholar]
  29. Neofotistou V, Vano E, Padovani R, et al. Preliminary reference levels in interventional cardiology. Eur Radiol. 2003;13:2259–63. doi: 10.1007/s00330-003-1831-x. [DOI] [PubMed] [Google Scholar]
  30. Watson LE, Riggs MW, Bourland PD. Radiation exposure during cardiology fellowship training. Health Phys. 1997;73:690–3. doi: 10.1097/00004032-199710000-00016. [DOI] [PubMed] [Google Scholar]
  31. Balter S, Miller DL, Vano E, et al. A pilot study exploring the possibility of establishing guidance levels in x-ray directed interventional procedures. Med Phys. 2008;35:673–80. doi: 10.1118/1.2829868. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Medical Radiation Sciences are provided here courtesy of Wiley