Abstract
Self-illuminating fluorescence imaging without autofluorescence background interference has recently aroused more research interests in molecular imaging. Currently, only a few self-illuminating probes were developed, based mainly on toxic quantum dots such as CdSe, CdTe. Herein, we report a novel design of nontoxic self-illuminating gold nanocluster (64Cu-doped AuNCs) for dual-modality positron emission tomography (PET) and near-infrared (NIR) fluorescence imaging based on Cerenkov resonance energy transfer (CRET). PET radionuclide 64Cu was introduced by a chelator-free doping method, which played dual roles as the energy donor and the PET imaging source. Meanwhile, AuNCs acted as the energy acceptor for NIR fluorescence imaging. 64Cu-doped AuNCs exhibited efficient CRET-NIR and PET imaging both in vitro and in vivo. In a U87MG glioblastoma xenograft model, 64Cu-doped AuNCs showed high tumor uptake (14.9%ID/g at 18 h) and produced satisfactory tumor self-illuminating NIR images in the absence of external excitation. This self-illuminating nanocluster with non-toxicity and good biocompatibility can be employed as a novel imaging contrast agent for biomedical applications, especially for molecular imaging.
1. Introduction
Self-illuminating fluorescence imaging is a newly developed imaging technique, overcoming several shortcomings of traditional fluorescence imaging. It has many distinguished imaging advantages, such as: i) no need of excitation from external illumination sources to fluoresce, resulting in no interference of strong background autofluorenscence [1, 2], and ii) no issue of excitation attenuation caused by photons absorption/scattering in bio-tissue transmission [3], which serves extremely important for in vivo imaging, as most imaging targets are non-superficial. So far, only a few self-illuminating fluorescence probes are designed and constructed based on three different energy transfer mechanisms such as bioluminescence resonance energy transfer (BRET), chemiluminescence resonance energy transfer (CLRET) and Cerenkov resonance energy transfer (CRET). Quantum dots (QDs), due to their high quantum yield, tunable emission peaks, long fluorescence lifetimes, and negligible photobleaching, have been employed as the energy transfer acceptor. For example, So et al. described self-illuminating QDs based on BRET, in which the bioluminescence energy of the luciferase-catalyzed oxidation of coelenterazine was transferred to the CdSe/ZnS QDs [2]. Zhang et al. reported another CLRET-CdSe/ZnS QDs methodology for in vivo myeloperoxidase (MPO) [4]. Recently, we constructed CRET-based self-illuminating QDs. In this system, Cerenkov radiation of 64Cu was transferred to excite the CdSe/ZnS QDs [5]. However, self-illuminating QDs are limited for further biomedical applications because most QDs contain heavy metal elements (such as Cd2+, Pb2+, etc.) and tend to self-aggregate inside living cells [6-8]. The cytotoxicity of the released heavy metal ions in biological systems and potential environmental hazard of these ions always hinder their practical applications [9, 10]. Therefore, it is highly desirable to develop novel self-illuminating systems with non-toxicity and good biocompatibility for further biomedical applications, especially in molecular imaging and therapy.
Recently, noble metal nanoclusters (such as gold nanoclusters (AuNCs)), due to their high fluorescence, good photostability, non-toxicity, excellent biocompatibility and water solubility, have aroused great interest in the development of new types of luminescent probes [11-18]. More importantly, these nanoclusters are usually prepared by a biomimetic synthesis method using naturally functional macromolecules (proteins, peptides, and enzymes etc.) as templates, which provides the nanoclusters with good colloidal stability and solubility in biological systems and avoid them from undesired or nonspecific binding [11, 17, 19]. Meanwhile, the macromolecules provide bioactive functional groups, such as −NH2 and −COOH, which can be used for further surface functionalization [20-22]. Therefore, metal nanoclusters are a good choice as energy transfer acceptors for developing novel self-illuminating fluorescent probes.
Herein, we strategically designed and prepared self-illuminating 64Cu-doped gold nanoclusters (64Cu-doped AuNCs) for dual-modality positron emission tomography (PET) and near infrared (NIR) imaging. The radioactive 64Cu plays dual roles as both the energy donor to excite AuNCs and the positron-emitting radionuclide (tracer) for PET imaging, while AuNCs act as the energy acceptor for self-illuminating NIR fluorescence imaging (Scheme 1). A novel chelator-free 64Cu doping method was developed for rapid radiolabeling of the AuNCs (within 1 h). The as-prepared 64Cu-doped AuNCs exhibited good stability, water dispersibility and solubility in vitro, and noncytotoxicity and good biocompatibility in vivo. In a U87MG glioblastoma xenograft model, 64Cu-doped AuNCs displayed satisfactory synergistic dual-modality PET and self-illuminating NIR tumor imaging.
2. Material and methods
2.1. Chemicals
All chemicals were purchased from Sigma-Aldrich Chemical Co. and used without further purification. Water was purified with Milli-Q plus system (Millipore Co., Bedford, MA), and the resistivity was kept to be over 18 MΩ cm.
2.2. Preparation of the HSA-AuNCs and 64Cu-doped AuNCs
The procedure for synthesizing HSA-AuNC synthesis was similar to that for BSA-AuNCs. Briefly, the aqueous HAuCl4 solution (5 mL, 10 mM) was added to HSA solution (5 mL, 50 mg/mL) under vigorous stirring. Five min later, NaOH solution (0.5 mL, 1 M) was introduced, and the pH changed to ~12. The reaction was allowed to proceed under vigorous stirring at 37 °C for 12 h. The reaction mixture was kept in the refrigerator (4 °C) in the dark for further use.
For 64Cu doping, all the solutions used were de-gassed for 40 min before use, and the AuNC solution was washed to pH 8~9 by PBS. 5-6 mCi 64CuCl2 was added to 1 mL of AuNCs solution under stirring for 5 min. Then, 10 μL of hydrazine hydrate was added and stirring continued for another 40 min at room temperature. Purification was done in a PD-10 column washed with 0.1% Na-ascorbate in PBS. Radiochemical purity was measured by instant thin layer chromatography (ITLC) plates using 0.1 M citric acid as eluent. 1-10% (molar ratio) Cu-doped AuNCs were prepared with the same procedure.
2.3. Characterization of 64Cu-doped AuNCs
Transmission electron microscope (TEM) images were acquired using a JEM-2010 (JEOL Ltd., Japan) operated at an accelerating voltage of 200 kV. UV-Vis-NIR spectra were measured on a Genesys 10S UV-Vis spectrophotometer (Thermo Scientific, Waltham, MA) using quartz cuvettes with an optical path of 1 cm. Fluorescence emission spectra were recorded on an F-7000 fluorescence spectrophotometer (Hitachi, Tokyo, Japan). CD spectra (190–300 nm) were recorded on a Jasco J-815 spectropolarimeter.
2.4. Cell culture and animal model
U87MG glioblastoma cell line was bought from ATCC and cultured in EMEM containing 10% fetal bovine serum (Gibco, Grand Island, NY) and incubated at 37 °C in a humidified 5% CO2 incubator. All animal work was performed following a protocol approved by the National Institutes of Health Clinical Center Animal Care and Use Committee (NIH CC/ACUC). Athymic nude mice purchased from Harlan (Indianapolis, IN) were subcutaneously implanted with 5 × 106 U87MG cells in the right front flank. In vivo imaging was performed 3 weeks after the inoculation when the tumor volume reached around 75 mm3.
2.5. CKK8 safety assay
5 × 103 cells per well were seeded in 96-well plate and incubated with AuNCs or decayed 64Cu-doped AuNCs for 24 h at different concentrations (100, 50, 25, 12.5 and 6.25 μM). Then the CKK8 agent (Dojindo Laboratories, Japan) was added to each well and incubated for 2 h, the absorbance at 450 nm was measured. Cell viability was normalized by control group without any treatment.
2.6. Cell uptake and FACS
50 μM AuNCs or decayed 64Cu-doped AuNCs were co-incubated with U87MG cells at 37 °C in a humidified 5% CO2 incubator. At different time points (0.5, 1, 2, 4, and 18 h), the uptake of AuNCs was analyzed by an Acurri C6 Flow Cytometer (BD Biosciences). The untreated cells were served as a control. The microscopic observation of internalization was carried out at 6 h after incubation. The nucleus was counter stained with DAPI. The images were captured by an inverted fluorescence microscope (Olympus IX81, Japan).
2.7. MicroPET imaging and Biodistribution study
The U87MG tumor-bearing mice were anesthetized with isoflurane and were injected with 100 μL 7.4 MBq (200 μCi) 64Cu-doped AuNCs intravenously. All PET scans were performed on an Inveon small-animal PET scanner (Siemens, Erlangen, Germany) at indicated time point post injection. The images were collected for 10 min. For each PET scan, 3-dimensional volumes of interest (VOIs) were drawn over the tumor and muscle on decay-corrected whole-body coronal images and analyzed by Inveon Research Workplace (Siemens). At the endpoint of experiment, the mice were sacrificed and interested organs were harvested, weighted and the radioactivity was measured in a Beckman 8000 gamma counter (Beckman, Brea, CA). Standards were prepared and the organ uptake was calculated as percent of injected dose/gram of tissue (%ID/g).
2.8. In vitro and in vivo self-illuminating fluorescence imaging
For in vitro self-illuminating fluorescence imaging, 11.1 MBq (300 μCi) of 64CuCl2 and 64Cu-doped AuNCs were imaged by an IVIS Lumina II small animal imaging system (Caliper Life Sciences, Hopkinton, MA). The images were taken under different emission filter sets (no filter, <510 nm, >590 nm, 515-575 nm, 575-650 nm, 695-770 nm and 810-885 nm), with the exposure time of 5 min, f/top = 1, binning = 4. For in vivo imaging, after the tumor-bearing mice were transferred to the light-tight chamber of IVIS imaging system, the images were taken with three different emission filter sets (no filter, <510 nm and >590 nm).The acquisition conditions were: exposition time 10 min, f/top = 1, binning = 4. All the images were analyzed by the Living Image 3.0 software (Caliper Life Science) and the signal was represented as photons per second per centimeter square per steradian (p/s/cm2/sr).
3. Results and discussion
3.1. Characterization of Cu-doped AuNCs
Fluorescent AuNCs were prepared by a human serum albumin (HSA)-directed, solution-phase, biomimetic synthetic method. As shown in Figure 1a and b, both AuNCs and Cu-doped AuNCs (1% Cu) are spherical. After Cu-doping, the size of AuNCs is increased from 0.93 ± 0.25 to 2.56 ± 0.50 nm (Figure 1c and d). UV-vis absorbance and fluorescence emission spectra of Cu-doped AuNCs (1% Cu) are shown in Figure 1e. The Cu-doped AuNCs (1% Cu) has one characteristic absorption peak at 280 nm, which is assigned to the π→π* transition of the aromatic amino acid residues of HSA. Under the excitation at 514 nm, Cu-doped AuNCs (1% Cu) exhibits maximum emission at ~667 nm. In addition, the structure conformation of pure HSA and Cu-doped AuNCs (1% Cu) is also investigated using a circular dichroism (CD) spectropolarimeter (Figure 1f). Pure HSA shows a positive absorption band at 190 nm and two negative absorption bands at 209 and 222 nm. With the formation of Cu-doped AuNCs (1% Cu), the positive peak at 190 nm ascends and the two negative peaks display slight blue shifts to 205 and 221 nm, respectively.
3.2. The effects of different doping concentration of Cu
The fluorescence intensity of Cu-doped AuNCs gradually deceases with the increase of Cu-doping ratio (Figure 2a and Supplementary Figure S1a). Cu-doped AuNCs display good stability in water, PBS and serum (Figure 2b, c and Supplementary Figure S1b). More importantly, both AuNCs and Cu-doped AuNCs (1% Cu) exhibit higher photostability than conventional fluorescent dyes (Figure 2d).
3.3. Dual imaging potentials of 64Cu-doped AuNCs
Next, we prepared the Cu-doped AuNCs with “hot” copper using the same procedure. After the implementation of hydrazine hydrate reduction, almost all the 64Cu was reduced and doped onto the AuNCs (Supplementary Figure S2). To analyze the 64Cu doping location, we used HSA as control. The instant thin layer chromatography (ITLC) results showed that HSA alone cannot be labeled by 64Cu with the same doping method (Supplementary Figure S3), suggesting that 64Cu was doped on the AuNCs, rather than on HSA. We then examined the NIR and PET imaging capability of 64Cu-doped AuNCs. 5.5 MBq (150 μCi, 250 μL, 5 mM) 64Cu-doped AuNCs were imaged by PET and fluorescence imaging, respectively. The 64Cu-doped AuNCs exhibited high PET signal and displayed comparable NIR imaging performance to the corresponding AuNCs under the same external excitation (Figure 3a). This result demonstrates that 64Cu doping provides AuNCs with PET imaging ability and the trace amount of doped 64Cu had no obvious impairment to the NIR fluorescence of AuNCs. As Cerenkov radiation is most intense at high frequencies (ultraviolet/blue area) [23], which overlaps exactly with the absorption spectrum of the AuNCs, the CRET would happen within 64Cu-doped AuNCs. In order to reveal CRET and evaluate subsequent self-illumination, multi-spectra imaging analysis was carried out with IVIS system. 11.1 MBq (300 μCi) 64Cu-doped AuNCs and 64CuCl2 were imaged with serial filter sets (no filter, < 515 nm, > 590 nm, 515-575 nm, 575-650 nm, 595-770 nm and 810-875 nm). Each filter only allows certain wavelength of light to pass (Figure 3b). In the no filter group, which collected all the signals, the 64Cu-doped AuNCs showed an intensity of 1.6- that of 64CuCl2. A reasonable explanation for this excessive intensity is that AuNCs yield additional signals under the excitation of Cerenkov radiation from 64Cu. Indeed, with the filter of > 590 nm, which covers most of the emission spectrum of 64Cu-doped AuNCs, we recorded an intensity 2.7 times that of 64CuCl2 (Figure 3c). Moreover, with filter sets of 575-650 nm and 695-770 nm, which contain parts of the emission spectrum of AuNCs, the intensities of 64Cu-doped AuNCs were 1.5-fold and 4.3-fold that of 64CuCl2 (Figure 3d). In contrast, with the filter < 510 nm, which is believed to be the main emission area of Cerenkov radiation, the 64Cu-doped AuNCs showed lower intensity than that of 64CuCl2. This is reasonable, as a portion of the Cerenkov radiation was transferred from 64Cu to AuNCs (Figure 3d).
3.4. In vivo CRET imaging of 64Cu-doped AuNCs
Cerenkov radiation itself is not suitable for in vivo studies due to its short wavelength [5, 23]. However by CRET, the energy from Cerenkov radiation can be transferred to AuNCs, which have a longer emission wavelength and deeper tissue penetration. To identify the imaging potential of 64Cu-doped AuNCs in vivo, we chose to utilize the U87MG glioblastoma xenograft model, as cell uptake and flow cytometry assay identified high uptake of Cu-doped AuNCs on this cell line, CKK8 viability assay disclosed no obvious cytotoxicity to the U87MG cells (Figure 4a-c). U87 Tumor-bearing mice were given 7.4 MBq (200 μCi) of 64Cu-doped AuNCs intravenously and then sequentially scanned by IVIS with 3 different filter sets (no filter, > 590 nm and < 510 nm). The whole-body IVIS images showed an enhanced tumor uptake of 64Cu-doped AuNCs after injection, reaching maximum at 8 h post injection followed by a decrease in intensity due to isotope decay (Figure 5a). Importantly, the CRET-NIR imaging (measured by > 590 nm filter) showed satisfactory tumor uptake compared to Cerenkov imaging (measured by filter < 510 nm), in which tumor uptake can barely be detected under the same intensity scale bar (Supplementary Figure S4). To compare the intensities between the 3 filters, regions of interests (ROIs) were drawn over the tumors. The intensities of CRET-NIR were slightly lower than those of the no filter group, accounting for around 4/5 of the total signal intensities. This is because the lower wavelength of Cerenkov radiation was deducted from the total intensity. On the contrary, only about 1/7 of the total signal was coming from Cerenkov radiation (Figure 5b). These encouraging results further confirmed the in vivo imaging potentials of self-illuminating 64Cu-doped AuNCs and demonstrated that CRET-NIR was the main luminescent contributor to self-illumination, compared with Cerenkov radiation alone.
3.5. In vivo PET imaging of 64Cu-doped AuNCs
The simultaneous PET imaging is another advantage of self-illuminating 64Cu-doped AuNC. Complementary to NIR imaging, PET has infinite penetration and provides more quantitative information [24, 25]. The PET imaging was taken immediately after the IVIS scan. The decay-corrected whole-body PET images and 3D reconstruction showed high tumor to normal contrast (Figure 6a and Supplementary Movie S1). Quantitative three-dimensional volumes of interest (VOIs) analysis identified about 6.9 percent of injected dose/gram of tissue (%ID/g) tumor uptake of 64Cu-doped AuNCs at 1 h post injection, which increased to 14.9 %ID/g and 15.2 %ID/g at 18 h and 24 h, respectively (Figure 6b). The tumor to muscle ratios (T/NT) also increased from 5.8 at 1 h post injection, to a peak of 28.4 at 18 h post injection (Figure 6c). The imaging discrepancy between PET and previous self-illuminating imaging may be attributed to the isotope decay. The %ID/g value is a decay-corrected value in PET imaging. Therefore, we recorded an increased %ID/g value of tumor, while the absolute radioactivity decreased because of isotope decay. In self-illuminating imaging, 64Cu was the energy donor. The luminescent intensity is thus positively correlated with its radioactivity. This is why we recorded maximum tumor uptake at 18 h time point in PET, but at 8 h time point in self-illuminating imaging. Actually, if we calculate and plot the images without decay correction, the pattern of PET would be consistent with IVIS imaging results (Supplementary Figure S5).
3.6. The correlation between CRET and PET signals
To further understand the correlation among PET, CRET-NIR (> 590 nm) and Cerenkov radiation (< 510 nm), we compared their signal intensities. At 24 h post injection, the tumor and five other organs were extracted, and acquired by IVIS and PET imaging respectively (Figure 7a, b, and Supplementary Figure S6). The ROI measurement displayed good linear correlation between PET and total signal (no filter, r2 = 0.9340, Figure 7c), and PET and CRET-NIR signal (> 590 nm; r2 = 0.9687, Figure 7d). In contrast, a weaker linear correlation was found between PET signals and Cerenkov radiation (< 510 nm; r2 = 0.7616, Figure 7e). As we know Cerenkov radiation is linearly correlated with the PET activity [26, 27], these results demonstrate that a portion of the energy of Cerenkov radiation serves to excite AuNCs in 64Cu-doped AuNCs. Furthermore, high linear correlation suggests that the self-illumination intensity of 64Cu-doped AuNCs can be applied as an alternative indicator of PET signal. In addition, organ distribution results indicate that the liver and kidneys may be responsible for the route of clearance of 64Cu-doped AuNCs (Figure 7f). Histological examination showed no obvious organ damage at 1 day and 7 days after injection of 64Cu-doped AuNCs (Supplementary Figure S7)
4. Conclusion
In summary, we developed a novel self-illuminating gold nanocluster, 64Cu-doped AuNCs, for dual-modality PET and CRET-NIR fluorescence imaging. To the best of our knowledge, it is the first time that noble metal nanoclusters have been reported as energy transfer acceptors for multimodality imaging. The 64Cu-doped AuNCs we developed have the following features: 1) both PET and self-illuminating NIR imaging capabilities; 2) in vivo self-illumination, converting short wavelengths of Cerenkov radiation into longer wavelengths; 3) small size, high solubility and stability in aqueous media, and 4) non-toxicity and good biocompatibility. These self-illuminating nanoclusters provide novel biomedical research tools, especially for molecular imaging and cancer therapy.
Supplementary Material
5. Acknowledgment
This work was supported, in part, by the Intramural Research Program (IRP) of the NIBIB, NIH, the National Key Basic Research Program (973 Project) (2010CB933901, 2013CB733802, 2014CB744503) and the National Natural Science Foundation of China (No.81090270, No.81090273 and No.81370585). We also thank Dr. Jie Lu, Dale Kiesewetter, and Andrea Balbo for their technical support.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- [1].Frangioni JV. Self-illuminating quantum dots light the way. Nat Biotechnol. 2006;24:326–8. doi: 10.1038/nbt0306-326. [DOI] [PubMed] [Google Scholar]
- [2].So MK, Xu C, Loening AM, Gambhir SS, Rao J. Self-illuminating quantum dot conjugates for in vivo imaging. Nat Biotechnol. 2006;24:339–43. doi: 10.1038/nbt1188. [DOI] [PubMed] [Google Scholar]
- [3].Wu Q, Chu M. Self-illuminating quantum dots for highly sensitive in vivo real-time luminescent mapping of sentinel lymph nodes. Int J Nanomed. 2012;7:4531. doi: 10.2147/IJN.S30709. vol 7, pg 3443, 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [4].Zhang N, Francis KP, Prakash A, Ansaldi D. Enhanced detection of myeloperoxidase activity in deep tissues through luminescent excitation of near-infrared nanoparticles. Nat Med. 2013;19:500–5. doi: 10.1038/nm.3110. [DOI] [PubMed] [Google Scholar]
- [5].Sun X, Huang X, Guo J, Zhu W, Ding Y, Niu G, et al. Self-illuminating 64Cu-doped CdSe/ZnS nanocrystals for in vivo tumor imaging. J Am Chem Soc. 2014;136:1706–9. doi: 10.1021/ja410438n. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Weng J, Ren J. Luminescent quantum dots: a very attractive and promising tool in biomedicine. Curr Med Chem. 2006;13:897–909. doi: 10.2174/092986706776361076. [DOI] [PubMed] [Google Scholar]
- [7].Huang P, Lin J, Wang X, Wang Z, Zhang C, He M, et al. Light-triggered theranostics based on photosensitizer conjugated carbon dots for simultaneous enhanced fluorescence imaging and photodynamic therapy. Adv Mater. 2012;24:5104–10. doi: 10.1002/adma.201200650. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Huang P, Li Z, Lin J, Yang D, Gao G, Xu C, et al. Photosensitizer-conjugated magnetic nanoparticles for in vivo simultaneous magnetofluorescent imaging and targeting therapy. Biomaterials. 2011;32:3447–58. doi: 10.1016/j.biomaterials.2011.01.032. [DOI] [PubMed] [Google Scholar]
- [9].Hardman R. A toxicologic review of quantum dots: Toxicity depends on physicochemical and environmental factors. Environ Health Persp. 2006;114:165–72. doi: 10.1289/ehp.8284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [10].Huang P, Bao L, Yang D, Gao G, Lin J, Li Z, et al. Protein-directed solution-phase green synthesis of BSA-conjugated MxSey (M= Ag, Cd, Pb, Cu) nanomaterials. Chem Asian J. 2011;6:1156–62. doi: 10.1002/asia.201000818. [DOI] [PubMed] [Google Scholar]
- [11].Luo Z, Zheng K, Xie J. Engineering ultrasmall water-soluble gold and silver nanoclusters for biomedical applications. Chem Commun (Camb) 2014;50:5143–55. doi: 10.1039/c3cc47512c. [DOI] [PubMed] [Google Scholar]
- [12].Xie J, Zheng Y, Ying JY. Protein-directed synthesis of highly fluorescent gold nanoclusters. J Am Chem Soc. 2009;131:888–9. doi: 10.1021/ja806804u. [DOI] [PubMed] [Google Scholar]
- [13].Li M, Yang DP, Wang X, Lu J, Cui D. Mixed protein-templated luminescent metal clusters (Au and Pt) for H2O2 sensing. Nanoscale Res Lett. 2013;8:182. doi: 10.1186/1556-276X-8-182. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [14].Zheng J, Nicovich PR, Dickson RM. Highly fluorescent noble-metal quantum dots. Annu Rev Phys Chem. 2007;58:409–31. doi: 10.1146/annurev.physchem.58.032806.104546. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [15].Lin J, Zhou Z, Li Z, Zhang C, Wang X, Wang K, et al. Biomimetic one-pot synthesis of gold nanoclusters/nanoparticles for targeted tumor cellular dual-modality imaging. Nanoscale Res Lett. 2013;8:1–7. doi: 10.1186/1556-276X-8-170. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [16].Huang P, Lin J, Wang S, Zhou Z, Li Z, Wang Z, et al. Photosensitizer-conjugated silica-coated gold nanoclusters for fluorescence imaging-guided photodynamic therapy. Biomaterials. 2013;34:4643–54. doi: 10.1016/j.biomaterials.2013.02.063. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [17].Zhang C, Zhou Z, Qian Q, Gao G, Li C, Feng L, et al. Glutathione-capped fluorescent gold nanoclusters for dual-modal fluorescence/X-ray computed tomography imaging. J Mater Chem B. 2013;1:5045–53. doi: 10.1039/c3tb20784f. [DOI] [PubMed] [Google Scholar]
- [18].Zhang XD, Chen J, Luo Z, Wu D, Shen X, Song SS, et al. Enhanced tumor accumulation of sub-2 nm gold nanoclusters for cancer radiation therapy. Adv Healthc Mater. 2014;3:133–41. doi: 10.1002/adhm.201300189. [DOI] [PubMed] [Google Scholar]
- [19].Xavier PL, Chaudhari K, Baksi A, Pradeep T. Protein-protected luminescent noble metal quantum clusters: an emerging trend in atomic cluster nanoscience. Nano Rev. 2012;3 doi: 10.3402/nano.v3i0.14767. doi: 10.3402/nano.v3i0.14767. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [20].Wang YL, Chen JJ, Irudayaraj J. Nuclear Targeting Dynamics of Gold Nanoclusters for Enhanced Therapy of HER2(+) Breast Cancer. Acs Nano. 2011;5:9718–25. doi: 10.1021/nn2032177. [DOI] [PubMed] [Google Scholar]
- [21].Retnakumari A, Setua S, Menon D, Ravindran P, Muhammed H, Pradeep T, et al. Molecular-receptor-specific, non-toxic, near-infrared-emitting Au cluster-protein nanoconjugates for targeted cancer imaging. Nanotechnology. 2010;21 doi: 10.1088/0957-4484/21/5/055103. [DOI] [PubMed] [Google Scholar]
- [22].Retnakumari A, Setua S, Menon D, Ravindran P, Muhammed H, Pradeep T, et al. Molecular-receptor-specific, non-toxic, near-infrared-emitting Au cluster-protein nanoconjugates for targeted cancer imaging. Nanotechnology. 2010;21:055103. doi: 10.1088/0957-4484/21/5/055103. [DOI] [PubMed] [Google Scholar]
- [23].Mitchell GS, Gill RK, Boucher DL, Li C, Cherry SR. In vivo Cerenkov luminescence imaging: a new tool for molecular imaging. Philos Trans A Math Phys Eng Sci. 2011;369:4605–19. doi: 10.1098/rsta.2011.0271. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [24].Zhao Y, Sultan D, Detering L, Cho S, Sun G, Pierce R, et al. Copper-64-alloyed gold nanoparticles for cancer imaging: improved radiolabel stability and diagnostic accuracy. Angew Chem Int Ed Engl. 2014;53:156–9. doi: 10.1002/anie.201308494. [DOI] [PubMed] [Google Scholar]
- [25].Chen F, Ellison PA, Lewis CM, Hong H, Zhang Y, Shi S, et al. Chelator-free synthesis of a dual-modality PET/MRI agent. Angew Chem Int Ed Engl. 2013;52:13319–23. doi: 10.1002/anie.201306306. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [26].Xu YD, Liu HG, Cheng Z. Harnessing the Power of Radionuclides for Optical Imaging: Cerenkov Luminescence Imaging. J Nucl Med. 2011;52:2009–18. doi: 10.2967/jnumed.111.092965. [DOI] [PubMed] [Google Scholar]
- [27].Ruggiero A, Holland JP, Lewis JS, Grimm J. Cerenkov luminescence imaging of medical isotopes. J Nucl Med. 2010;51:1123–30. doi: 10.2967/jnumed.110.076521. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.