Inconsistencies between navigation data and radiographs in total knee arthroplasty are system-dependent and affect coronal alignment

Alberto Carli *,, Ahmed Aoude *, Avishai Reuven , Bogdan Matache *, John Antoniou , David J Zukor
PMCID: PMC4183676  PMID: 25265103

Abstract

Background

Few studies have compared the effect of different computer navigation systems on postoperative alignment in patients who have had total knee arthroplasty (TKA). We examined 2 computed tomography (CT)–free computer navigation systems by comparing the accuracy of intraoperative measurements to postoperative alignment.

Methods

Patients underwent unilateral TKA performed by a single surgeon using 1 of 2 CT-free navigation systems. We compared final intraoperative tibial and femoral coronal angles and mechanical axis with the same angles measured on standing postoperative radiographs.

Results

Groups of 31 and 50 patients underwent TKA with the 2 systems, respectively. We noted a significant difference in the coronal tibial implant angle (1.29º ± 1.35º) and in the mechanical axis (1.59º ± 2.36º) for one navigation system (both p < 0.001), while only the coronal tibial implant angle showed a significant difference (1.17º ± 1.65º, p < 0.001) for the second system. The number of radiographic outliers also significantly differed. A significantly higher proportion (32%; p < 0.01) of patients in the second cohort exhibited unacceptable malalignment compared with the first cohort (24%).

Conclusion

Navigation systems for TKA continue to increase in sophistication and popularity. Owing to the significant difference in the proportion of alignment outliers in the 2 navigation systems tested in this study, orthopedic surgeons should not consider all TKA navigation systems equivalent. Additional investigations are needed to compare the accuracy of a variety of CT-free and CT-based navigation systems and to confirm our finding that accuracy is system-dependent.


Computer-assisted orthopedic surgery (CAOS) for total knee arthroplasty (TKA) is generally considered to lead to component positioning that is more accurate and reproducible than conventional techniques.1,2 Additional advantages for the use of COAS in TKA include decreased perioperative blood loss owing to the avoidance of intramedullary instrumentation,1,35 fewer postoperative thromboembolic events6 and fewer systemic emboli.79 Hoffart and colleagues10 also demonstrated that CAOS provides better clinical outcomes at 5 years than conventional TKA. Furthermore, recent meta-analyses by Cheng and colleagues11 and Bauwens and colleagues12 demonstrated that using COAS for TKA significantly reduced the relative risk of excessive implant misalignment by 25% compared with conventional TKA. Another recent meta-analysis by Hetaimish and colleagues13 showed that CAOS produced better implant alignment in the coronal plane for both the tibial and femoral components as well as femoral and tibial slope than the conventional technique.

Although these investigations have demonstrated the improvement of component alignment with CAOS, they did not take into account a critical component of the operative procedure: the type of computer navigation system used. The only comparisons of alignment among different computer navigation systems have come from 2 studies that each compared a computed tomography (CT)–based system to a CT-free system.14,15 Given that studies used in the meta-analysis by Cheng and colleagues11 used a total of 6 different CT-free navigation systems and 2 CT-based systems while those used in the analysis by Bauwens and colleagues12 included 5 CT-free systems and 3 CT-based systems, further investigations are required to confirm that computer navigation systems are equivalently accurate in measuring TKA alignment. Such investigations are especially needed in CT-free CAOS systems, whose differences in hardware and software design have been previously identified as a potential source of inaccuracy.16

The current gold standard for assessing the mechanical alignment of the lower extremity following TKA is the standing full lower extremity radiograph. Therefore, the ideal CAOS navigation system, when properly used, should provide intraoperative measurements for femoral and tibial cuts that match the final implant position and limb alignment as measured on postoperative radiographs. Using this logic, it is possible to compare CAOS navigation systems with one another by assessing the difference between intraoperative alignment measurements by the CAOS system to actual postoperative radiographic alignment values. Thus, it can be said that the smaller the difference between intraoperative and postoperative measurements, the more accurate the CAOS navigation system. The more accurate a CAOS navigation system is, the more confident the surgeon can be that measurements taken intraoperatively reflect the actual postoperative alignment. We have shown in previous work that angular measurements attained by a single CT-free CAOS navigation system intraoperatively can differ significantly from postoperative measurements taken from full length lower extremity radiographs.17 The present study compares this initial system to a second CT-free CAOS navigation system in order to determine if one is more accurate overall and produces fewer alignment outliers (≥ 3° of varus/valgus alignment).

We hypothesized that the variation between final intraoperative CAOS bony cut measurements and postoperative implant and extremity alignment would be different between the 2 CT-free navigation systems. We also hypothesized that the number of postoperative mechanical outliers (≥ 3º of varus/valgus) would be decreased for the navigation system that had more similar intraoperative and postoperative measures.

Methods

Patient selection

We retrospectively reviewed the medical records and radiographs for all consecutive patients who underwent CAOS-TKA at a single orthopedic institution in 2 time periods (November 2007–February 2008 and March 2009–December 2010). For the first cohort, all operations were performed with the assistance of the ORTHOsoft Knee 2.1 Universal system. For the second cohort, all operations were performed with the assistance of the Brainlab AG Ci Knee essential 2.1.1 system. All procedures were performed by a single arthroplasty surgeon (D.J.Z.) who received formal training for each navigation system. The specific time periods were selected for study evaluation as they represented the time by which the surgeon had surpassed the accepted threshold for the learning curve (30–50 cases)18,19 of each CAOS-TKA system. Consecutive patient inclusion was performed to minimize selection bias.

Operative procedure and perioperative data collection

All procedures were performed under spinal anesthesia, and the surgeon used a medial parapatellar approach. On exposure of the knee joint, articular landmarks and important bone areas were recorded using a pointer-like device, and external optical tracking devices were placed on the tibia and femur to template bony cuts. The 2 imaging systems used similar static landmarks, including the femoral condylar surfaces, centre of the talus, femoral intercondylar notch, borders of the tibial plateau and tibial spine, as well as dynamic motions to determine the centre of the femoral head. The main difference in landmark identification was the need to trace out (or “paint”) the entirety of the distal femoral and proximal tibial articular surfaces with the Brainlab system compared with several single point landmarks on the ORTHOsoft system. Apart from these differences and slight instrument sizing differences, the 2 navigation systems similarly use optical trackers that provide data regarding knee position to an infrared receiver, which then displays real-time positional information on a digital display. Following navigation-assisted femoral and tibial cuts, trial components and a fitting spacer were placed, and knee range of motion was assessed. Following cement application and component placement, a final alignment assessment could be performed in both the Brainlab and ORTHOsoft systems to allow any further alignment correction due to the cement mantle. Knee range of motion and mechanical axis were then clinically assessed, and the subcutaneous tissue and skin was closed with a Jones bandage.

Data were collected intraoperatively through each CT-free navigation system. For both systems, the following parameters were recorded after insertion of final components: the coronal angle of the tibial cut, the coronal angle of the femoral cut and the final mechanical axis of the lower extremity.

Postoperative data collection

As per standard protocol at our institution, all patients underwent a full weight-bearing long axis radiograph of the lower extremities 6 weeks after surgery. Standing radiographs were obtained using a long film cassette with a radiography tube distance of 305 cm to capture the hips, knees and ankles. Although use of a long cassette is not universal20 and is subject to errors due to knee flexion and lower extremity rotation,2123 its use has been cited as an acceptable reference standard for alignment of the lower limb.2427 Radiographs were then viewed in their native DICOM format using Horizon Rad Station viewing software version 11.0.8.2172 (McKesson). Measured values were obtained by 2 individuals (A.C. and B.M), who did not participate in any of the surgical procedures, were blinded to the CAOS system allocation and who were both experienced in taking radiographic measurements for research purposes. Using the same methodology as described in our previous work,17 the following measurements were obtained from postoperative radiographs: the coronal angle of the tibial component, the coronal angle of the femoral component and the final mechanical axis. Interrater reliability was calculated, and the average of each measurement made by the 2 researchers was then used for comparison. Furthermore, Knee Society scores were recorded for each patient at the 6-week visit.

Statistical analysis

We assessed interrater reliability for radiographic measurements by calculating the intraclass correlation coefficient using SPSS statistical software version 19, using a 2-way random-effects model assuming a single measurement and absolute agreement. An intraclass correlational coefficient of 1 represents perfect reliability, and any value greater than 0.8 indicates excellent reliability.28 We compared intraoperative navigation and postoperative radiographs using paired t tests, with the level of significance set at p < 0.05. The proportions of postoperative complications and radiographic outliers were recorded for each system and compared using a χ2 test. A radiographic outlier was defined as a postoperative measurement of 3º of varus/valgus or greater.

Results

Our study included 81 patients: 31 in the ORTHOsoft group and 50 in the Brainlab group. All patients successfully underwent unilateral TKA with no significant intraoperative complications. Patient demographics were similar in both groups. The mean patient age was 70.12 ± 10.66 years in the ORTHOsoft group and 74.27 ± 7.60 years in the Brainlab group. All patients began weight bearing by the second postoperative day and yielded similar rates of postoperative transfusions (χ2 = 1.547, p = 0.21) and deep vein thrombosis (χ2 = 2.46, p = 0.11). No major complications, including deep wound infection, myocardial infarction and pulmonary embolus, were noted in either group during the first 6 weeks. The mean Knee Society score at 6 weeks was comparable between the groups (mean 67.77 ± 14.15 in the ORTHOsoft group v. 67.09 ± 12.32 in the Brainlab group). Overall, interrater reliability for postoperative radiograph measurements was excellent, with intraclass correlation coefficients of 0.834 for the coronal angle of the tibial component (p < 0.001), 0.842 for the coronal angle of the femoral component (p < 0.001) and 0.962 for the mechanical axis (p < 0.001).

We noted significant differences between intraoperative navigation measurements and postoperative radiograph measurements in each navigation group (Table 1). However, such differences were not observed in the same measurements. In the ORTHOsoft group, we noted a significant difference in the coronal tibial implant angle (1.29º ± 1.35º, p < 0.001) and in the mechanical axis (1.59º ± 2.36º, p < 0.001). In the Brainlab group, only the coronal tibial implant angle showed a significant difference (1.17º ± 1.65º, p < 0.001); no difference was observed in the mechanical axis (0.75º ± 2.67º, p = 0.05). In both groups, the coronal angle of the femur showed no significant change from intraoperative data to postoperative radiographs (0.30 ± 1.74, p = 0.33 in the ORTHOsoft group v. 0.22 ± 1.94, p = 0.49 in the Brainlab group).

Table 1.

Comparison of intraoperative navigation measures and manual radiolographic measurements made 6 weeks after surgery for the Brainlab and ORTHOsoft navigation systems*

Measure Group; mean ± SD Mean difference 95% CI p value
Navigation system values Radiograph values
Brainlab, n = 50
 Coronal plane of femoral implant 0.14 ± 0.95 0.35 ± 1.70 0.21 ± 0.75 −0.33 to 0.77 0.49
 Coronal plane of tibial implant 0.59 ± 0.76 −0.58 ± 1.65 1.17 ± 0.89 −1.64 to −0.70 < 0.001
 HKA angle 0.02 ± 0.70 −0.73 ± 2.69 0.75 ± 1.99 −1.51 to 0.01 0.05
ORTHOsoft, n = 31
 Coronal plane of femoral implant 0.17 ± 0.63 0.48 ± 1.65 0.30 ± 1.74 −0.33 to 0.94 0.33
 Coronal plane of tibial implant −0.66 ± 1.10 0.62 ± 1.78 1.29 ± 1.35 0.79 to 1.78 < 0.001
 HKA angle 0.45 ± 1.20 2.04 ± 2.13 1.59 ± 2.36 0.73 to 2.46 < 0.001

CI = confidence interval; HKA = hip–knee–ankle; SD = standard deviation.

*

Positive measurements represent valgus orientation; negative measurements represent varus orientation. Comparison was made using a paired t test, with significance set at p < 0.05.

The number of radiographic outliers also differed in each navigation system. The overall percentage of patients with at least 1 radiographic parameter more than 3° was 32% (10 of 31) in the ORTHOsoft group and 24% (12 of 50) in the Brainlab group (χ2 = 19.121, p < 0.01). The percentage of outliers was higher in the ORTHOsoft group for both the mechanical axis (29% v. 22%; Fig. 1) and the coronal angle of the femoral component (6.4% v. 6%; Fig. 2). The Brainlab group had a higher percentage of coronal tibial angle outliers (12% v. 9.6%; Fig. 3).

Fig. 1.

Fig. 1

Comparison of intraoperative measurements of the mechanical axis measured by Brainlab and ORTHOsoft navigation systems and postoperative measurements on standing radiographs. Intraoperative navigation measurements are organized from the most valgus to most varus.

Fig. 2.

Fig. 2

Comparison of intraoperative measurements of the coronal orientation of the femoral implant measured by Brainlab and ORTHOsoft navigation systems and postoperative measurements on standing radiographs. Intraoperative navigation measurements are organized from the most valgus to most varus.

Fig. 3.

Fig. 3

Comparison of intraoperative measurements of the coronal orientation of the tibial implant measured by Brainlab and ORTHOsoft navigation systems and postoperative measurements on standing radiographs. Intraoperative navigation measurements are organized from the most valgus to most varus.

Discussion

Interest in comparing computer navigation systems in orthopedics is steadily emerging. Although not relevant to TKA, a study by Honl and colleagues29 compared acetabular orientation in 5 computer-assisted navigation systems for total hip arthroplasty and found variations among the systems. Specific to TKA, Harvie and colleagues30 prospectively followed 40 patients who underwent the procedure using either an imageless full navigation system or an articular surface-mounted navigation system. They found no significant difference between the 2 systems with regard to the postoperative implant position verified. A comparison among imageless navigation systems has not been reported despite the general preference for CT-free over CT-dependent systems.

To our knowledge, the present study is the first to compare deviations between intraoperative navigation data and postoperative radiograph measurements of 2 CT-free navigation systems. It is also, to our knowledge, the first to propose that this form of intraoperative and postoperative comparison be used as a manner to measure accuracy for CAOS systems. Our results confirm our initial hypothesis that the greater the difference between intraoperative navigation system measurements and postoperative radiograph measurements, the greater the incidence of coronal misalignment in patients who have undergone TKA. It should be specifically emphasized that the differences between the 2 systems, although significant, were small and within 2° of each other. However, such small differences were enough to significantly change the overall proportion of radiographic outliers between systems, which is a pertinent finding given that one of the driving purposes of computer navigation in TKA was to eliminate knee malalignment. Since the difference between intraoperative and postoperative measurements was not the same for the 2 navigation systems studied, our results indicate that different CT-free navigation systems are not necessarily equivalent in predicting postoperative alignment during TKA.

Computed tomography–free navigation systems have been previously shown to decrease the prevalence of radiographic outliers to 10%.3134 In our study, neither navigation system met this threshold. The navigation system that provided intraoperative measurements that better matched postoperative alignment (Brainlab) had fewer mechanical outliers (24%) than its less accurate counterpart (ORTHOsoft). This latter system produced an incidence of mechanical outliers that surpassed the 30% rate seen with conventional instrumentation.35,36

Controversy exists in the literature regarding the clinical relevance of the improved coronal alignment attained through CAOS-TKA. Khan and colleagues37 found that patients who underwent computer-assisted TKA with a mechanical axis alignment of 3° or more from neutral had lower Western Ontario and McMaster Universities Arthritis Index scores, indicating increased difficulty with daily activities. Furthermore, Lützner and colleagues38 randomly assigned 80 patients to receive either conventional or navigated TKA and found that patients who underwent navigated TKA were less likely to have mechanical axis malalignment postoperatively and had significantly improved Knee Society scores. However, a follow-up study involving the same cohort revealed no difference between the groups 5 years after surgery.39 A similar result was found at a 2-year follow-up of 71 patients who were also randomized to conventional or navigated TKA.40

In addition to findings of equivalent outcome scores despite improved alignment, concerns for higher revision rates after navigated TKA have been raised. Gotheson and colleagues41 presented data from the Norwegian registry regarding 10 000 TKAs and found significantly higher revisions rates for navigation versus conventional (3.6% v. 2.1%) procedures. However, this finding is offset by findings of equivalent revision rates in a comparison study of 100 navigated and conventional TKAs42 as well as a retrospective evaluation of 1121 TKAs that yielded a significantly lower revision rate for navigated TKAs.43 Although these conflicting findings in the literature raise valid concerns regarding the long-term benefits of navigated TKA, it remains generally accepted that establishing a normal mechanical axis is a fundamental technical objective in TKA and that CAOS-TKA more reliably accomplishes this objective than conventional methods.44 However, findings from the present study indicate that some CAOS-TKA systems may not be as reliable in providing intraoperative alignment information that adequately correlates with the postoperative mechanical axis. Ultimately, additional studies comparing multiple CAOS-TKA systems and how their respective divergences between intraoperative and postoperative alignment variables affect long-term clinical outcome must be performed to evaluate the relevance of the present findings.

Limitations

This study has limitations. Although the navigation data and postoperative radiograph measurements were taken prospectively, patients were not randomized to a specific navigation system. Preoperative alignment and range of motion data were not compared between the 2 groups, a notable limitation given that a previous study reported greater divergence between intraoperative and postoperative alignment indices in patients with more severe preoperative alignment deformities.45 The study also does not provide short-term patient outcome information after the 6-week postoperative mark, making it difficult to determine the clinical importance of the observed discrepancies between intraoperative and postoperative alignment. Although patients in both groups were included only after the learning curve for each CAOS system had been surpassed, we acknowledge that the treating surgeon was more experienced in CAOS-TKA when performing surgeries on the Brainlab cohort and that this may have improved his intraoperative alignment technique for this group. Postoperative radiograph measurements, although recorded by the 2 blinded observers with excellent inter-rater reliability, are inherently susceptible to errors due to patient positioning as well as knee flexion and external rotation.4649 In addition, as remarked by Choi and colleagues,50 intraoperative and postoperative measurements are performed under considerably different circumstances. Intraoperative measurements are taken before complete soft tissue closure in full extension and without weight bearing, whereas postoperative measurements are taken with full weight bearing and cannot control for soft tissue contractures elsewhere in the lower extremity.

Despite such differing circumstances, our inclusion of only a single surgeon performing the surgical procedures should minimize intersurgeon variation in how the intra-operative data were collected, making it less difficult to compare against postoperative radiographs. Furthermore, our choice to use long-cassette radiographs as the gold standard is justified, as this modality is considered an acceptable reference for alignment27 and sufficient for routine postoperative assessment.51 Another consideration is the cement mantle, which if applied unevenly can cause the final implant position to be different from the bony cut. Both systems used in the present study permit intraoperative alignment after cement and components are inserted, permitting the surgeon to apply coronal stresses or impact implants further before cement curing. This feature of final component alignment surveillance is not universal among CAOS-TKA systems and, along with various cementing techniques, could theoretically affect final component alignment. We are interested in evaluating the effect of cementing technique and CAOS final alignment capability, and we plan to perform a prospective evaluation at our centre. The slight difference in the way surface landmarks between the 2 systems were made is also of interest, but the effect this difference would have on alignment is unknown and should also be recorded in future system comparisons. Finally, long-term data were not included, making it difficult to conclude if such variability in discrepancy between navigation data and radiograph measurements impacted clinical outcome.

Conclusion

Computer-assisted orthopedic surgery for TKA continues to increase in sophistication and popularity, with new computerized systems attempting to incorporate aspects of both full navigation and conventional TKA.30 Although a sizeable amount of literature exists advocating the advantages in alignment that accompanies the use of CAOS, results from the present study suggest that orthopedic surgeons should not consider all TKA navigation systems to be the same. Additional investigations are needed to compare the accuracy of a variety of CT-free and CT-based navigation systems and to confirm our finding that accuracy is system-dependent, with greater accuracy leading to fewer mechanical outliers. This finding should be taken into account when reviewing conflicting data obtained from pooled results involving different systems, and should also be considered when switching to different systems in clinical practice.

Footnotes

Competing interests: None declared

Contributors: A. Carli, A. Aoude, A. Reuven, J. Antoniou and D. Zukor designed the study and acquired the data. A. Carli, B. Matache, J. Antoniou and D. Zukor analyzed the data. A. Carli, A. Aoude, A. Reuven, J. Antoniou and D. Zukor wrote the article, which A. Carli, B. Matache, J. Antoniou and D. zukor reviewed. All authors approved the final version for publication.

References

  • 1.Chauhan SK, Scott RG, Breidahl W, et al. Computer-assisted knee arthroplasty versus a conventional jig-based technique. A randomised, prospective trial. J Bone Joint Surg Br. 2004;86:372–7. doi: 10.1302/0301-620x.86b3.14643. [DOI] [PubMed] [Google Scholar]
  • 2.Decking R, Markmann Y, Fuchs J, et al. Leg axis after computer- navigated total knee arthroplasty: a prospective randomized trial comparing computer-navigated and manual implantation. J Arthroplasty. 2005;20:282–8. doi: 10.1016/j.arth.2004.09.047. [DOI] [PubMed] [Google Scholar]
  • 3.Kalairajah Y, Simpson D, Cossey AJ, et al. Blood loss after total knee replacement: effects of computer-assisted surgery. J Bone Joint Surg Br. 2005;87:1480–2. doi: 10.1302/0301-620X.87B11.16474. [DOI] [PubMed] [Google Scholar]
  • 4.Bierbaum BE, Callaghan JJ, Galante JO, et al. An analysis of blood management in patients having a total hip or knee arthroplasty. J Bone Joint Surg Am. 1999;81:2–10. doi: 10.2106/00004623-199901000-00002. [DOI] [PubMed] [Google Scholar]
  • 5.McConnell J, Dillon J, Kinninmonth A, et al. Blood loss following total knee replacement is reduced when using computer-assisted versus standard methods. Acta Orthop Belg. 2012;78:75–9. [PubMed] [Google Scholar]
  • 6.Ooi LH, Lo NN, Yeo SJ, et al. Does computer-assisted surgical navigation total knee arthroplasty reduce venous thromboembolism compared with conventional total knee arthroplasty? Singapore Med J. 2008;49:610–4. [PubMed] [Google Scholar]
  • 7.Kalairajah Y, Cossey AJ, Verrall GM, et al. Are systemic emboli reduced in computer-assisted knee surgery?: A prospective, randomised, clinical trial. J Bone Joint Surg Br. 2006;88:198–202. doi: 10.1302/0301-620X.88B2.16906. [DOI] [PubMed] [Google Scholar]
  • 8.Morawa LG, Manley MT, Edidin AA, et al. Transesophageal echocardiographic monitored events during total knee arthroplasty. Clin Orthop Relat Res. 1996;(331):192–8. doi: 10.1097/00003086-199610000-00027. [DOI] [PubMed] [Google Scholar]
  • 9.Church JS, Scadden JE, Gupta RR, et al. Embolic phenomena during computer-assisted and conventional total knee replacement. J Bone Joint Surg Br. 2007;89:481–5. doi: 10.1302/0301-620X.89B4.18470. [DOI] [PubMed] [Google Scholar]
  • 10.Hoffart HE, Langenstein E, Vasak N. A prospective study comparing the functional outcome of computer-assisted and conventional total knee replacement. J Bone Joint Surg Br. 2012;94:194–9. doi: 10.1302/0301-620X.94B2.27454. [DOI] [PubMed] [Google Scholar]
  • 11.Cheng T, Zhao S, Peng X, et al. Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials. Knee Surg Sports Traumatol Arthrosc. 2012;20:1307–22. doi: 10.1007/s00167-011-1588-8. [DOI] [PubMed] [Google Scholar]
  • 12.Bauwens K, Matthes G, Wich M, et al. Navigated total knee replacement. A meta-analysis. J Bone Joint Surg Am. 2007;89:261–9. doi: 10.2106/JBJS.F.00601. [DOI] [PubMed] [Google Scholar]
  • 13.Hetaimish BM, Khan MM, Simunovic N, et al. PK Z. Meta-analysis of navigation vs conventional total knee arthroplasty. J Arthroplasty. 2012;27:1177–82. doi: 10.1016/j.arth.2011.12.028. [DOI] [PubMed] [Google Scholar]
  • 14.Martin A, von Strempel A. CT-based and CT-free navigation in total knee arthroplasty–a prospective comparative study with respects to clinical and radiological results [article in German] Z Orthop Ihre Grenzgeb. 2005;143:323–8. doi: 10.1055/s-2005-836568. [DOI] [PubMed] [Google Scholar]
  • 15.Bäthis H, Perlick L, Luring C, et al. CT-based and CT-free navigation in knee prosthesis implantation. Results of a prospective study [article in German] Unfallchirurg. 2003;106:935–40. doi: 10.1007/s00113-003-0685-7. [DOI] [PubMed] [Google Scholar]
  • 16.Stulberg SD, Loan P, Sarin V. Computer-assisted navigation in total knee replacement: results of an initial experience in thirty-five patients. J Bone Joint Surg Am. 2002;84-A(Suppl 2):90–8. [PubMed] [Google Scholar]
  • 17.Livshetz I, Brin YS, Holcroft C, et al. Deviations between intra-operative navigation data and post-operative weight-bearing X-rays. Comput Aided Surg. 2010;15:56–62. doi: 10.3109/10929088.2010.480884. [DOI] [PubMed] [Google Scholar]
  • 18.Jenny J-Y, Miehlke RK, Giurea A. Learning curve in navigated total knee replacement. A multi-centre study comparing experienced and beginner centres. Knee. 2008;15:80–4. doi: 10.1016/j.knee.2007.12.004. [DOI] [PubMed] [Google Scholar]
  • 19.Daubresse F, Vajeu C, Loquet J. Total knee arthroplasty with conventional or navigated technique: comparison of the learning curves in a community hospital. Acta Orthop Belg. 2005;71:710. [PubMed] [Google Scholar]
  • 20.Jiang C-C, Insall JN. Effect of rotation on the axial alignment of the femur: pitfalls in the use of femoral intramedullary guides in total knee arthroplasty. Clin Orthop Relat Res. 1989;(248):50–6. [PubMed] [Google Scholar]
  • 21.Krackow KA, Pepe C, Galloway E. A mathematical analysis of the effect of flexion and rotation on apparent varus/valgus alignment at the knee. Orthopedics. 1990;13:861. doi: 10.3928/0147-7447-19900801-09. [DOI] [PubMed] [Google Scholar]
  • 22.Brouwer R, Jakma T, Brouwer K, et al. Pitfalls in determining knee alignment: a radiographic cadaver study. J Knee Surg. 2007;20:210–5. doi: 10.1055/s-0030-1248045. [DOI] [PubMed] [Google Scholar]
  • 23.Hunt MA, Fowler PJ, Birmingham TB, et al. Foot rotational effects on radiographic measures of lower limb alignment. Can J Surg. 2006;49:401. [PMC free article] [PubMed] [Google Scholar]
  • 24.Langenbach MR, Dohle J, Zirngibl H. Determination of the axis after total endoprosthesis of the knee: functional X-ray photography as golden standard [article in German] Z Orthop Ihre Grenzgeb. 2002;140:32–6. doi: 10.1055/s-2002-22088. [DOI] [PubMed] [Google Scholar]
  • 25.Duparc J, Massare C. Radiological measurement of the angular deviation of the knee in the frontal plane [article in French] Ann Radiol (Paris) 1967;10:635–6. [PubMed] [Google Scholar]
  • 26.Reed SC, Gollish J. The accuracy of femoral intramedullary guides in total knee arthroplasty. J Arthroplasty. 1997;12:677–82. doi: 10.1016/s0883-5403(97)90141-8. [DOI] [PubMed] [Google Scholar]
  • 27.Cooke TDV, Sled EA, Scudamore RA. Frontal plane knee alignment: a call for standardized measurement. J Rheumatol. 2007;34:1796. [PubMed] [Google Scholar]
  • 28.Fleiss J. Statistical methods for rates and proportions. 2nd ed. New York (NY): John Wiley and Sons; 1981. [Google Scholar]
  • 29.Honl M, Schwieger K, Salineros M, et al. Orientation of the ace-tabular component: a comparison of five navigation systems with conventional surgical technique. J Bone Joint Surg Br. 2006;88:1401. doi: 10.1302/0301-620X.88B10.17587. [DOI] [PubMed] [Google Scholar]
  • 30.Harvie P, Sloan K, Beaver RJ. Three-dimensional component alignment and functional outcome in computer-navigated total knee arthroplasty a prospective, randomized study comparing two navigation systems. J Arthroplasty. 2011;26:1285–90. doi: 10.1016/j.arth.2010.12.022. [DOI] [PubMed] [Google Scholar]
  • 31.Bäthis H, Perlick L, Tingart M, et al. Alignment in total knee arthroplasty. A comparison of computer-assisted surgery with the conventional technique. J Bone Joint Surg Br. 2004;86:682–7. doi: 10.1302/0301-620x.86b5.14927. [DOI] [PubMed] [Google Scholar]
  • 32.Sparmann M, Wolke B, Czupalla H, et al. Positioning of total knee arthroplasty with and without navigation support. A prospective, randomised study. J Bone Joint Surg Br. 2003;85:830–5. [PubMed] [Google Scholar]
  • 33.Kim SJ, MacDonald M, Hernandez J, et al. Computer assisted navigation in total knee arthroplasty: improved coronal alignment. J Arthroplasty. 2005;20(Suppl 3):123–31. doi: 10.1016/j.arth.2005.05.003. [DOI] [PubMed] [Google Scholar]
  • 34.Stöckl B, Nogler M, Rosiek R, et al. Navigation improves accuracy of rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 2004;(426):180–6. doi: 10.1097/01.blo.0000136835.40566.d9. [DOI] [PubMed] [Google Scholar]
  • 35.Bolognesi M, Hofmann A. Computer navigation versus standard instrumentation for TKA: a single-surgeon experience. Clin Orthop Relat Res. 2005;(440):162–9. doi: 10.1097/01.blo.0000186561.70566.95. [DOI] [PubMed] [Google Scholar]
  • 36.Brys DA, Lombardi AV, Jr, Mallory TH, et al. A comparison of intramedullary and extramedullary alignment systems for tibial component placement in total knee arthroplasty. Clin Orthop Relat Res. 1991;(263):175–9. [PubMed] [Google Scholar]
  • 37.Khan MM, Khan MW, Al-Harbi HH, et al. Assessing short-term functional outcomes and knee alignment of computer-assisted navigated total knee arthroplasty. J Arthroplasty. 2012;27:271–7. doi: 10.1016/j.arth.2011.04.038. [DOI] [PubMed] [Google Scholar]
  • 38.Lützner J, Günther K-P, Kirschner S. Functional outcome after computer-assisted versus conventional total knee arthroplasty: a randomized controlled study. Knee Surg Sports Traumatol Arthrosc. 2010;18:1339–44. doi: 10.1007/s00167-010-1153-x. [DOI] [PubMed] [Google Scholar]
  • 39.Lützner J, Dexel J, Kirschner S. No difference between computer-assisted and conventional total knee arthroplasty: five-year results of a prospective randomised study. Knee Surg Sports Traumatol Arthrosc. 2013;21:2241–7. doi: 10.1007/s00167-013-2608-7. [DOI] [PubMed] [Google Scholar]
  • 40.Spencer JM, Chauhan SK, Sloan K, et al. Computer navigation versus conventional total knee replacement: no difference in functional results at two years. J Bone Joint Surg Br. 2007;89:477–80. doi: 10.1302/0301-620X.89B4.18094. [DOI] [PubMed] [Google Scholar]
  • 41.Gøthesen O, Espehaug B, Havelin L, et al. Short-term outcome of 1,465 computer-navigated primary total knee replacements 2005–2008: A report from the Norwegian Arthroplasty Register. Acta Orthop. 2011;82:293–300. doi: 10.3109/17453674.2011.575743. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Lüring C, Kauper M, Bäthis H, et al. A five to seven year follow-up comparing computer-assisted vs freehand TKR with regard to clinical parameters. Int Orthop. 2012;36:553–8. doi: 10.1007/s00264-011-1297-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Schnurr C, Güdden I, Eysel P, et al. Influence of computer navigation on TKA revision rates. Int Orthop. 2012;36:2255–60. doi: 10.1007/s00264-012-1606-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Mason JB, Fehring TK, Estok R, et al. Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty. 2007;22:1097–106. doi: 10.1016/j.arth.2007.08.001. [DOI] [PubMed] [Google Scholar]
  • 45.Rauh MA, Boyle J, Mihalko WM, et al. Reliability of measuring longstanding lower extremity radiographs. Orthopedics. 2007;30:299–303. doi: 10.3928/01477447-20070401-14. [DOI] [PubMed] [Google Scholar]
  • 46.Lonner JH, Laird MT, Stuchin SA. Effect of rotation and knee flexion on radiographic alignment in total knee arthroplasties. Clin Orthop Relat Res. 1996;(331):102–6. doi: 10.1097/00003086-199610000-00014. [DOI] [PubMed] [Google Scholar]
  • 47.Wong AK, Inglis D, Beattie KA, et al. Reproducibility of computer-assisted joint alignment measurement in OA knee radiographs. Osteoarthritis Cartilage. 2009;17:579–85. doi: 10.1016/j.joca.2008.10.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Schmidt GL, Altman GT, Dougherty JT, et al. Reproducibility and reliability of the anatomic axis of the lower extremity. J Knee Surg. 2004;17:141–3. [PubMed] [Google Scholar]
  • 49.Cooke TD, Sled EA, Scudamore RA. Frontal plane knee alignment: a call for standardized measurement. J Rheumatol. 2007;34:1796–801. [PubMed] [Google Scholar]
  • 50.Choi WC, Lee S, An JH, et al. Plain radiograph fails to reflect the alignment and advantages of navigation in total knee arthroplasty. J Arthroplasty. 2011;26:756–64. doi: 10.1016/j.arth.2010.07.020. [DOI] [PubMed] [Google Scholar]
  • 51.McGrory JE, Trousdale RT, Pagnano MW, et al. Preoperative hip to ankle radiographs in total knee arthroplasty. Clin Orthop Relat Res. 2002;(404):196–202. doi: 10.1097/00003086-200211000-00032. [DOI] [PubMed] [Google Scholar]

Articles from Canadian Journal of Surgery are provided here courtesy of Canadian Medical Association

RESOURCES