Clinical Validation of the Lyra Direct HSV 1+2/VZV Assay for Simultaneous Detection and Differentiation of Three Herpesviruses in Cutaneous and Mucocutaneous Lesions

Feinan Fan a,b, Jeffrey Stiles b, Albina Mikhlina b, Xuedong Lu a, N Esther Babady b, Yi-Wei Tang b,c,
Editor: M J Loeffelholz
PMCID: PMC4187752  PMID: 25078915

Abstract

We evaluated the Lyra Direct HSV 1+2/VZV multiplex real-time PCR assay for the detection and differentiation of herpes simplex virus 1 (HSV-1), HSV-2, and varicella-zoster virus (VZV) on 695 consecutive cutaneous and mucocutaneous lesion specimens. The intra-assay and interassay coefficient of variation values for the Lyra assay were 0.29 to 1.30% and 2.33 to 2.61%, respectively. The sensitivities, specificities, and positive and negative predictive values were 93.4 to 95.0%, 96.1 to 96.8%, 78.0 to 80.3%, and 99.0 to 99.1%, respectively, in comparison to those of viral culture. The values were further improved when a resolution analysis was performed with a laboratory-developed PCR assay.

TEXT

Herpes simplex virus 1 and 2 (HSV-1 and HSV-2, respectively) and varicella-zoster virus (VZV) cause serious cutaneous and mucocutaneous lesions at every stage in immunocompromised patients (1). Due to the similarities in the clinical presentations of infections caused by these three herpesviruses, a clinical diagnosis needs to be confirmed by laboratory testing to prevent misdiagnosis (2). The need for accurate and specific laboratory diagnostics is especially important in immunocompromised patients who may present with atypical lesions confounding the clinical diagnosis and delaying the institution of appropriate antiviral therapy (1, 3, 4). Consequently, etiologic diagnosis and differentiation of HSV-1, HSV-2, and VZV infections is critical for both patient care and infection control (3, 4).

Traditional laboratory methods for the diagnosis of cutaneous and mucocutaneous HSV-1, HSV-2, and VZV infections include the Tzanck smear, direct fluorescent assay (DFA), and cell culture (including shell vial culture) (3). Several reports have been published on the increased sensitivity of real-time PCR assays for detecting HSV/VZV over the traditional methods described above (57). Most published molecular assays utilize separate reactions for either HSV or VZV, and limited multiplex formats for the simultaneous detection and differentiation of HSV-1, HSV-2, and VZV have been reported (79). Currently, four molecular assays have been cleared by the Food and Drug Administration (FDA) for the detection and/or typing of HSV-1 and HSV-2 in genital and/or oral lesions: the MultiCode HSV-1&2 assay (Luminex Corporation, Austin, TX), the ProbeTec HSV QX amplified DNA assay (BD Diagnostics, Sparks, MD), the IsoAmp HSV assay (BioHelix Corporation, Beverly, MA), and the AmpliVue HSV 1+2 assay (Quidel Corporation, San Diego, CA) (10).

(This study was presented in part at the 114th Annual Meeting of the American Society for Microbiology, Boston, MA, 17 to 20 May 2014.)

In May 2014, the FDA cleared the Lyra Direct HSV 1+2/VZV assay (Quidel Corporation, San Diego, CA). This is the first FDA-cleared commercial multiplex real-time PCR assay that simultaneously detects and differentiates HSV-1/2 and VZV DNA in both cutaneous and mucocutaneous lesion specimens. The objectives of this preclinical study were to validate the performance of the Lyra assay on cutaneous and mucosal specimens, specifically in an immunocompromised oncology patient population.

Clinical samples.

Cutaneous or mucocutaneous lesions from symptomatic patients were sampled using a FLOQSwabs (Copan Diagnostics, Inc., Murrieta, CA) and submitted to the laboratory in Remel M4 transport medium (M4RT; Thermo Fisher Scientific, Lenexa, KS) for HSV and/or VZV culture. Six hundred ninety-five consecutive swab specimens were collected between 14 May 2012 and 25 November 2013. Following the completion of the culture setup, the leftover M4RT medium specimens were collected, aliquoted, and stored at −80°C until testing with the Lyra assay and laboratory-developed tests (LDT). The study was granted a waiver of the Health Insurance Portability and Accountability Act (HIPAA) authorization and informed consent (WA0346-13) by the Memorial Sloan-Kettering Cancer Center institutional review board committee.

Tube and shell vial culture.

Cell culture was performed by adding 250 μl of M4RT medium to an MRC-5 cell culture tube and an A549 shell vial (Diagnostic Hybrids, Athens, OH), followed by incubation at 37°C for a maximum of 14 days (5, 7, 9). The culture tubes were read every day to detect cytopathic effects. For the shell vial culture, identification was confirmed using fluorescence-labeled monoclonal antibodies against HSV (Bartels; Trinity Biotech, Wicklow, Ireland) and VZV (Merifluor; Meridian Diagnostic, Inc., Cincinnati, OH) at 24 and 48 h after incubation. The slides were read using an Olympus BX40 microscope with a BX-FLA reflected fluorescent light attachment.

Lyra Direct HSV 1+2/VZV assay.

The Lyra Direct HSV 1+2/VZV assay uses multiplex real-time PCR to detect and differentiate HSV-1, HSV-2, and VZV DNA in a given specimen. The multiplex real-time PCR is carried out under optimized conditions in a single tube, generating amplicons for HSV-1, HSV-2, VZV, and the process control (PRC). The identification and differentiation of HSV-1, HSV-2, and VZV occurs by the use of genotype-specific hydrolysis probes. The Lyra assay was performed on the QuantStudio Dx instrument (Life Technologies, Foster City, CA), according to the manufacturer's investigational use only kit's instructions (Quidel Corp., San Diego, CA). For sample preparation, 100 μl of M4RT medium specimen was placed in a 60°C heat block for 5 min. The heated specimen was mixed with 25 μl process buffer, and 5 μl of the mixture was transferred to a microtiter plate well, to which 15 μl of master mix was added. The plate was then sealed and placed into the QuantStudio Dx instrument for amplification and detection using primers and TaqMan-based probes specific for HSV-1, HSV-2, and VZV.

Laboratory-developed tests.

Discordant samples between culture and the Lyra assay were processed and amplified by the Cepheid HSV analyte-specific reagent (ASR) (Sunnyvale, CA) for HSV detection and typing (11) or by the Argene VZV R-gene ASR (bioMérieux, Durham, NC) for VZV detection (12). For HSV, 5 μl of heated specimen was added to 20 μl of the Cepheid HSV master mix. For VZV, 10 μl of the heated specimen was added to 15 μl of the VZV master mix. The tubes were closed and amplified on the SmartCycler II instrument (Cepheid). All positive amplicons were subsequently cloned and confirmed by sequencing (13).

A total of 695 specimens, of which 57 (8.2%) were genital/anal specimens, were included in the study. They were collected from 340 males (48.9%) and 355 females (51.1%), with an average ± standard deviation age of 52.8 ± 19.6 years. The tube and/or shell vial culture method detected 99 (14.2%) HSV-1/2-positive and 76 (10.9%) VZV-positive swabs. The Lyra Direct assay detected 117 (16.9%) HSV-1/2 and 91 (13.1%) VZV specimens and produced one invalid result, for an invalid result rate of 0.14%. Among the 93 specimens subjected to randomized testing three times on different days, 92 specimens had concordant results, with an overall agreement of 98.9%. One specimen that tested negative in the first two runs was positive for VZV at the third run. The intra-assay % coefficient of variation (%CV) values for the Lyra Direct assay ranged from 0.50 to 0.90% for HSV-1, 0.41 to 1.30% for HSV-2, and 0.29 to 0.75% for VZV (Table 1). The interassay %CV values of the Lyra assay were 2.33% for HSV-1, 2.36% for HSV-2, and 2.61% for VZV (Table 1). In comparison to cell culture, the sensitivity, specificity, and positive and negative predictive values of the Lyra Direct HSV 1+2/VZV assay were 95.0%, 96.1%, 80.3%, and 99.1%, respectively, for HSV-1/2 and 93.4%, 96.8%, 78.0%, and 99.2%, respectively, for VZV (Table 2).

TABLE 1.

Intra- and interassay variations of the Lyra Direct HSV-1+2/VZV assay

Virus by variation Specimen IDa Mean CTb SD CT CV (%)c
Intra-assay
    HSV-1 1110 25.89 0.19 0.74
1141 32.90 0.30 0.90
1108 29.65 0.15 0.50
    HSV-2 1124 18.82 0.08 0.41
1151 32.21 0.42 1.30
1120 29.93 0.13 0.44
    VZV 1066 17.60 0.10 0.57
1025 30.82 0.09 0.29
1028 26.16 0.20 0.75
Interassay
    HSV-1 PCd 27.20 0.63 2.33
    HSV-2 PC 25.47 0.58 2.26
    VZV PC 22.42 0.58 2.61
a

Specimens used for the intra-assay variations are clinical specimens enrolled in this study. The specimens used for interassay variations were positive controls provided in the kit. ID, identification.

b

CT, threshold cycle.

c

CV, coefficient of variation.

d

PC, positive control.

TABLE 2.

Sensitivity, specificity, PPV, and NPV of the Lyra Direct HSV-1+2/VZV assay (n = 694)a

Viruses by resolution analysis time No. of specimens by resultb
Test performance (95% CI) (%)c
R+ L+ R− L+ R+ L− R− L− Sensitivity Specificity PPV NPV
Pre-resolution analysis
    HSV-1/2 94d 23 5 573 95.0 (88.6–98.3) 96.1 (94.2–97.5) 80.3 (72.0–87.1) 99.1 (98.0–99.7)
    VZV 71 20 5 598 93.4 (87.8–99.0) 96.8 (95.4–98.2) 78.0 (69.5–86.5) 99.2 (98.4–99.9)
Post-resolution analysis
    HSV-1 68 2 1 623 98.6 (92.3–99.8) 99.7 (98.9–100.0) 97.1 (90.0–99.6) 99.8 (99.1–100.0)
    HSV-2 45 2 0 647 100.0 (92.1–100.0) 99.7 (98.9–100.0) 95.7 (85.4–99.4) 100.0 (99.4–100.0)
    VZV 72 19 0 603 100.0 (95.0–100.0) 96.9 (95.3–98.2) 79.1 (69.3–86.9) 100.0 (99.4–100.0)
a

PPV, positive predictive value; NPV, negative predictive value.

b

R, reference; L, Lyra. In the pre-resolution analysis, the tube or shell vial culture results were used as references, and in post-resolution analysis, the majority-matched results among culture, Lyra, and LDT were used as references.

c

CI, confidence interval.

d

One specimen was positive for both HSV-1 and HSV-2 by the Lyra assay.

There were 53 specimens with discordant results between culture and the Lyra assay. They were tested with two independent real-time PCR LDT. Among them, 10 specimens were culture positive (5 HSV and 5 VZV)/Lyra assay negative, and 43 specimens were culture negative/Lyra assay positive (23 HSV and 20 VZV). Among the 10 culture-positive/Lyra negative specimens, 9 specimens were negative by both LDT PCR assays (4 HSV and 5 VZV). For the 43 Lyra-positive/culture-negative samples, only one of 20 VZV-positive specimens by the Lyra assay was positive by the LDT-PCR. In contrast, among the 23 specimens that were HSV positive by Lyra assay, 19 (82.6%) were positive by the LDT-PCR. In comparison to a combined standard based on the majority results of the viral culture, Lyra, and LDT-PCR assays, the resolved sensitivity, specificity, and positive and negative predictive values of the Lyra assay were 98.6%, 99.7%, 97.1%, and 99.8%, respectively, for HSV-1, 100%, 99.7%, 95.7%, and 100%, respectively, for HSV-2, and 100%, 96.9%, 79.1%, and 100%, respectively, for VZV (Table 2).

The extraction of nucleic acids from specimen is not always necessary to perform before nucleic acid amplification and detection (14). However, this step increases the hands-on time and the turnaround time to results, raises expenses, and potentially introduces contamination. Sakai et al. (14) reported quick detection of HSV-1, HSV-2, and VZV from skin vesicles and exudates without nucleic acid extraction (14). Furthermore, several reports pointed out that omission of the nucleic acid extraction step prior to real-time PCR resulted in a more rapid and cost-effective assay, with little impact upon the sensitivity of detection (9, 1416). The Lyra device does not require the time-consuming nucleic acid extraction step and simply treats diluted clinical specimens in a 60°C heat block for 5 min. The equipment needed for the Lyra assay reaction is a heat block that furnishes a constant temperature of 60°C and a molecular real-time PCR platform, like the QuantStudio Dx instrument. The Lyra Direct assay consists of simple dilution-only sample preparation and automatic amplification that can bring accurate and rapid HSV-1/2 and VZV diagnostics to a laboratory that lacks cell culture expertise.

Among the 53 specimens with discordant results between culture and the Lyra assay, the Lyra assay detected more HSV (23 samples) and VZV (20 samples) than did culture (5 HSV and 5 VZV samples). When an LDT PCR was used to resolve the discordant results, among the 43 Lyra-positive/culture-negative specimens, 19 Lyra VZV-positive specimens were negative by the LDT-PCR, and only one of the 20 Lyra VZV-positive specimens was positive by the LDT-PCR. In contrast, among the 23 Lyra HSV-positive specimens, 82.6% were positive by the LDT-PCR (8 HSV-1 and 11 HSV-2). Similar to other molecular assays, our data indicate that the Lyra assay is more sensitive than culture for detecting HSV and VZV in lesion specimens from immunocompromised cancer patients. Previous studies have indicated that molecular methods have a significantly higher sensitivity for VZV detection in a variety of clinical specimen types than that of culture (6, 9, 17). After resolution analysis, the Lyra assay had 19 VZV-positive results (2.7%) that were not confirmed by either culture or LDT PCR. Whether these additional VZV-positive findings are related to carryover contamination that occurred during operation merits further investigation (6).

The Lyra Direct HSV 1+2/VZV assay is precise, reproducible, and sensitive for simultaneously detecting and differentiating HSV-1, HSV-2, and VZV in cutaneous and mucocutaneous lesions. Requiring no time-consuming nucleic acid extraction step, the Lyra assay is able to cover three viruses in a single tube with a total sample-to-result time of <60 min.

ACKNOWLEDGMENTS

We thank the Clinical Microbiology Service staff of the Memorial Sloan-Kettering Cancer Center for helping collect the clinical specimens.

This work was supported in part by a research contract between the Memorial Sloan-Kettering Cancer Center and the Quidel Corporation (SK2013-0910).

Footnotes

Published ahead of print 30 July 2014

REFERENCES

  • 1.Tan HH, Goh CL. 2006. Viral infections affecting the skin in organ transplant recipients: epidemiology and current management strategies. Am. J. Clin. Dermatol. 7:13–29. 10.2165/00128071-200607010-00003 [DOI] [PubMed] [Google Scholar]
  • 2.Kalman CM, Laskin OL. 1986. Herpes zoster and zosteriform herpes simplex virus infections in immunocompetent adults. Am. J. Med. 81:775–778. 10.1016/0002-9343(86)90343-8 [DOI] [PubMed] [Google Scholar]
  • 3.Frisch S, Guo AM. 2013. Diagnostic methods and management strategies of herpes simplex and herpes zoster infections. Clin. Geriatr. Med. 29:501–526. 10.1016/j.cger.2013.01.003 [DOI] [PubMed] [Google Scholar]
  • 4.Styczynski J, Reusser P, Einsele H, de la Camara R, Cordonnier C, Ward KN, Ljungman P, Engelhard D. 2009. Management of HSV, VZV and EBV infections in patients with hematological malignancies and after SCT: guidelines from the Second European Conference on Infections in Leukemia Bone Marrow Transpl. 43:757–770. 10.1038/bmt.2008.386 [DOI] [PubMed] [Google Scholar]
  • 5.Druce J, Catton M, Chibo D, Minerds K, Tyssen D, Kostecki R, Maskill B, Leong-Shaw W, Gerrard M, Birch C. 2002. Utility of a multiplex PCR assay for detecting herpesvirus DNA in clinical samples. J. Clin. Microbiol. 40:1728–1732. 10.1128/JCM.40.5.1728-1732.2002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Espy MJ, Uhl JR, Sloan LM, Buckwalter SP, Jones MF, Vetter EA, Yao JD, Wengenack NL, Rosenblatt JE, Cockerill FR, III, Smith TF. 2006. Real-time PCR in clinical microbiology: applications for routine laboratory testing. Clin. Microbiol. Rev. 19:165–256. 10.1128/CMR.19.1.165-256.2006 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Tan TY, Zou H, Ong DC, Ker KJ, Chio MT, Teo RY, Koh MJ. 2013. Development and clinical validation of a multiplex real-time PCR assay for herpes simplex and varicella zoster virus. Diagn. Mol. Pathol. 22:245–248. 10.1097/PDM.0b013e3182914291 [DOI] [PubMed] [Google Scholar]
  • 8.Bennett S, Carman WF, Gunson RN. 2013. The development of a multiplex real-time PCR for the detection of herpes simplex virus 1 and 2, varizella [sic] zoster virus, adenovirus and Chlamydia trachomatis from eye swabs. J. Virol. Methods 189:143–147. 10.1016/j.jviromet.2012.08.020 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Buelow DR, Bankowski MJ, Fofana D, Gu Z, Pounds S, Hayden RT. 2013. Comparison of two multiplexed PCR assays for the detection of HSV-1, HSV-2, and VZV with extracted and unextracted cutaneous and mucosal specimens. J. Clin. Virol. 58:84–88. 10.1016/j.jcv.2013.05.008 [DOI] [PubMed] [Google Scholar]
  • 10.Fan F, Day S, Lu X, Tang YW. 2014. Laboratory diagnosis of herpes simplex virus and varicella-zoster virus infections. Future Virol., in press [Google Scholar]
  • 11.Podzorski RP. 2006. Evaluation of the Cepheid herpes simplex virus typing real-time PCR assay using dermal and genital specimens. Diagn. Microbiol. Infect. Dis. 56:173–177. 10.1016/j.diagmicrobio.2006.03.011 [DOI] [PubMed] [Google Scholar]
  • 12.Dekeyser S, El Nady M, Diaz E, Cassagnou M, Descamps D. 2009. Diagnosis of herpetic esophagitis in the immunocompetent subject by PCR (Herpès Consensus Générique-Argène). Report of six cases. Pathol. Biol. (Paris) 57:101–106 (In French.) 10.1016/j.patbio.2008.07.005 [DOI] [PubMed] [Google Scholar]
  • 13.Tang YW, Espy MJ, Persing DH, Smith TF. 1997. Molecular evidence and clinical significance of herpesvirus coinfection in the central nervous system. J. Clin. Microbiol. 35:2869–2872 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Sakai K, Wakasugi S, Muchemwa FC, Ihn H. 2008. Quick detection of herpes viruses from skin vesicles and exudates without nucleic acid extraction using multiplex PCR. Biosci. Trends 2:164–168 [PubMed] [Google Scholar]
  • 15.Pandori MW, Lei J, Wong EH, Klausner J, Liska S. 2006. Real-Time PCR for detection of herpes simplex virus without nucleic acid extraction. BMC Infect. Dis. 6:104. 10.1186/1471-2334-6-104 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Binkhamis K, Al-Siyabi T, Heinstein C, Hatchette TF, LeBlanc JJ. 2014. Molecular detection of varicella zoster virus while keeping an eye on the budget. J. Virol. Methods 202:24–27. 10.1016/j.jviromet.2014.02.009 [DOI] [PubMed] [Google Scholar]
  • 17.Mehta SK, Tyring SK, Cohrs RJ, Gilden D, Feiveson AH, Lechler KJ, Pierson DL. 2013. Rapid and sensitive detection of varicella zoster virus in saliva of patients with herpes zoster. J. Virol. Methods 193:128–130. 10.1016/j.jviromet.2013.05.019 [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)