Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2014 Nov 17.
Published in final edited form as: Metabolomics. 2013 Sep 12;10(2):312–323. doi: 10.1007/s11306-013-0582-1

Validation of a dual LC-HRMS platform for clinical metabolic diagnosis in serum, bridging quantitative analysis and untargeted metabolomics

Ilya Gertsman 1,*, Jon A Gangoiti 1, Bruce A Barshop 1
PMCID: PMC4234038  NIHMSID: NIHMS639451  PMID: 25411574

Abstract

Mass spectrometry-based metabolomics is a rapidly growing field in both research and diagnosis. Generally, the methodologies and types of instruments used for clinical and other absolute quantification experiments are different from those used for biomarkers discovery and untargeted analysis, as the former requires optimal sensitivity and dynamic range, while the latter requires high resolution and high mass accuracy. We used a Q-TOF mass spectrometer with two different types of pentafluorophenyl (PFP) stationary phases, employing both positive and negative ionization, to develop and validate a hybrid quantification and discovery platform using LC-HRMS. This dual-PFP LC-MS platform quantifies over 50 clinically relevant metabolites in serum (using both MS and MS/MS acquisitions) while simultaneously collecting high resolution and high mass accuracy full scans to monitor all other co-eluting non-targeted analytes. We demonstrate that the linearity, accuracy, and precision results for the quantification of a number of metabolites, including amino acids, organic acids, acylcarnitines and purines/pyrimidines, meets or exceeds normal bioanalytical standards over their respective physiological ranges. The chromatography resolved highly polar as well as hydrophobic analytes under reverse-phase conditions, enabling analysis of a wide range of chemicals, necessary for untargeted metabolomics experiments. Though previous LC-HRMS methods have demonstrated quantification capabilities for various drug and small molecule compounds, the present study provides an HRMS quant/qual platform tailored to metabolic disease; and covers a multitude of different metabolites including compounds normally quantified by a combination of separate instrumentation.

Keywords: untargeted metabolomics, targeted metabolomics, bioanalytical validation, mass spectrometry, Q-TOF, LC-HRMS, comprehensive metabolite profiling

Introduction

Untargeted metabolomics has become a popular tool in metabolite biomarker discovery and in the evaluation of metabolic pathway changes associated with disease (Dunn et al. 2011; Kell 2007; Mamas et al. 2011). Such methods are effective in identifying metabolites that have statistically significant changes among sample cohorts. Recent instrument and chromatographic advances have allowed investigators to measure hundreds to thousands of unique metabolites that are representative of diverse chemical classes, all within one or several runs (Crews et al. 2009; Yanes et al. 2011).

A disadvantage of pure untargeted metabolomic studies is that metabolites are only quantified relative to one another and an actual estimate of concentration often cannot be determined, an aspect that makes the technique better suited for research than actual clinical diagnosis. Certain metabolites have wide reference ranges, and a 5-fold change in relative intensity for example, may in fact have no phenotypic consequence, whereas the same relative change in a metabolite that is tightly regulated may be very clinically relevant. On the other hand, targeted metabolomics methods enable absolute quantification of metabolites of interest using stable isotope dilution, and quality control analysis. In the realm of clinical testing, different regulatory bodies (CAP, CLIA, GLP) require standard procedures for assurance of specificity, precision, accuracy, linearity, sensitivity, recovery and stability in the presence of potentially interfering compounds, and verification of calibration with at least a minimum or zero value, a low, middle, and a high value near the upper limit of the analytical range (Bansal et al. 2007; FDA 2001). Unfortunately, triple-quadrupole mass spectrometers, best suited for accurate quantification, lack the resolution and mass accuracy necessary to identify the other thousands of co-extracted analytes.

In LC-MS/MS targeted platforms, the compounds are generally quantified by selected reaction monitoring analysis (SRM), focusing just on the transition of the precursor to a product ion of a particular analyte (Honour 2011). SRM experiments reduce the background signal of unspecified compounds that have similar mass and elution times, and increase the duty cycle of the instrument, resulting in greater sensitivity and a linear dynamic range of quantification. A new generation of Q-TOF instruments combine the sensitivity and dynamic range of triple-quadrupole instruments with time of flight technology, allowing for high resolution and mass accuracy (Fung et al. 2011). We have used such an instrument to develop assay methods which allow for absolute quantification of a range of metabolites present in serum and plasma, while simultaneously collecting high resolution, accurate mass data for comprehensive untargeted metabolomics, as shown in the workflow in Figure 1. Several LC-HRMS have been recently published that have successfully quantified different compounds using related mass spec techniques, but the previous publications have focused on drug compounds or peptides, which are known to behave well in standard reverse phase conditions using C18 columns (Dillen et al. 2012; Henry et al. 2012; Kaufmann et al. 2011; Zhang et al. 2009). We have developed and validated methods that enable quantification of small polar compounds such as amino acids and organic acids in addition to the more hydrophobic chemical classes, enabling a hybrid quantification/discovery platform with more comprehensive capabilities and a clinically oriented focus.

Figure 1. Targeted/Untargeted Workflow.

Figure 1

The figure describes the different analyses that can be done from a Q-TOF run. (A) Total intensity scan of a single MS run. Each run includes multiple experiments within each 0.9 second scan cycle, including a 0.25sec TOF-MS scan (B) from which specific masses can be extracted using narrow m/z windows. Statistical analyses such as PCA plots (C) and T-tests help distinguish cohorts and implicate ions of interest. (D) A box plot of one such analyte from the current study, identified as inosine by MS/MS data (>5-fold difference between disease [HCC/cirrhotic] and control groups from our t-test analysis with p-values <0.05). From the same TOF-MS scan (50-1000m/z) used in untargeted processing, narrow m/z ranges can be extracted for absolute quantification of labeled and non-labeled standards (E), using standard curve (shown on right) and QC analysis. Within the same cycle, short MS/MS scans of specific masses (F) permit quantification of product ions, shown here for lysine.

Typically, clinical biochemical genetic tests include separate analyses of organic acids (Duez et al. 1996; Hoffmann et al. 1989; Kumps et al. 1999; Sweetman 1991), amino acids (Shapira 1989), and acylcarnitines (Millington et al. 2011; Vreken et al. 1999), using various instruments, but are measured here with a single system. For our purposes, we selected 55 compounds that are measured in the diagnosis of metabolic dysfunction; commonly associated with, but not limited to inherited metabolic disorders (Chace et al. 2005; Wilcken et al. 2008). These include all 20 proteinogenic amino acids, a selection of non-proteinogenic amino acids, organic acids, purines, pyrimidines, and acylcarnitines (Table 1).

Table 1. Accuracy and Precision results of Clinical Validation.

Metabolite Range
(uM)
Stable
Iso-
tope
Actual conc. of
QC’s in μM LOQ, low,
medium, and high
Accuracy
LOQ QC: Inter-day avg.
(Intra-day
averages)
Accuracy
Low QC: Inter-day avg.
(Intra-day
averages)
Accuracy
Medium QC:
Inter-day avg.
(Intra-day
averages)
Accuracy
High QC:
Inter-day avg.
(Intra-day
averages)
Inter-day
%CV:
QC’sLOQ.
low, medium,
high
amino acids
(proteinogenic)
 alanineA 70-700 d4 70.170.313, 578 102(117.89.100) 83(86.81.82) 94 (94.96.93) 104(95,103,112) 17, 9. 8. 11
 leucineA 30-250 d3 30.70.113.206 104(107.105.99) 109(107.114.107) 102(101.101.103) 101(104.101.97) 5. 4. 3. 5
 isoleucineA 10-125 d10 10.25. 56,103 93 (91.96.91) 101(101.103.99) 101(105.98.98) 99(105,95.96) 3, 3. 4. 6
 valine#A 40-350 d8 40.95.158.289 96(108.92.89) 99(101.97.93) 97(102.96.94) 98(99.97.98) 10. 6. 5. 5
 prolineA 40-500 d7 40. 100.215.413 91 (91.93.89) 102(101.105.101) 100(100.102.98) 98(104.94.95) 3. 3. 3. 5
 phenylalanineA 20-200 d5 20.50.90.165 99(100.98.100) 111(108.115.109) 106(104.107.106) 96 (99.96.93) 3. 4. 3. 6
 tryptophanB 1-120 d5 1, 6, 45, 99 ND 104(122.98.91) 103(102.103.103) 95(94.97.96) 14. 2, 3
 tyrosineA 20-150 d4 20, 35, 68, 124 100(100.99.99) 96(95.98.95) 101 (100.103.101) 98(101.95.98) 1. 3. 2. 3
 cystineA 30-200 d4 30.50,90,165 115(116,114,113) 99(96.100.100) 100(106.93.101) 107(110.101.112) 4. 8. 12. 10
 methionineA 5-50 1-13C.
d3
5, 13, 23, 41 100(103.100.96) 103(101.107.102) 100(102.99.100) 100(101.98.100) 4. 3. 2. 2
 threonineA 25-300 1-13C,
d3met
25, 75, 135, 248 93 (84.95.101) 101(110.97.97) 96(89.103.96) 89(88.87.91) 18. 15. 8. 4
 serine#A 30-300 d3 30, 75, 135, 248 109(116,117.93) 103(95.120.95) 96(98,95.94) 103(104.101.105) 8. 20. 8. 8
 asparagine#A 20-150 d5 gln 20, 38, 68, 124 109(111.118.99) 100(93.96.111) 104(106.107.98) 97(97.100.94) 12. 9. 7. 6
 glutamineA 100-800 d5 100, 188, 350, 660 99(102.103,92) 98(92.102.99) 96(101.96.90) 100(102.102.96) 8. 8. 7. 6
 aspartate#A 0.6-50 d3 0.6. 2.1, 18, 42 ND 100(111.99.91) 98(99.98.98) 106(99.112.106) 19, 9, 9
 glutamateA 0.5-150 d3 0.5. 3, 53, 124 ND 89(83.81.105) 92(85.95.94) 97(98.96.97) 33, 13, 8
 lysine#A 21-300 d4 21, 59, 130, 248 127*(120,142,120) 98(93.96.105) 91 (90.88.95) 98(106.90.98) 13. 8, 4. 7
 histidine#A 40-250 d3 40, 70, 113, 206 100(113.100.87) 96(93.96.98) 95(102.87.96) 101(108.97.99) 7. 3. 7. 7
 arginineA 20-150 d4 10, 38, 68, 124 96(100.97.90) 98 (97.103,93) 99(100.96.101) 101 (108,99.95) 5, 5, 3. 7
non-proteinogenic
amino acids/other
 ornithine#A 10-200 d2 10, 35, 88, 165 99(103.94.99) 95(96,87.102) 100(103.98.97) 97(93,99,100) 14, 9, 11, 6
 homocystineA 5-75 d8 5, 15, 34, 62 99(98.110.89) 97(102.104.86) 96(95.96.96) 98(95, 100, 98) 13. 12, 7. 4
 creatineB 5-150 d3 5, 18, 63, 124 93 (97.91.92) 99(103.98.97) 95(96.91.99) 93 (97.90.93) 5. 3. 3. 3
 creatinineB 2-100 d3 0.5, 1.5, 20, 41 N/A 114*(172.96.74) 103(108.92.108) 94(100.90.92) 53, 10, 8
 pipecolicB 1-30 d9 1, 3.5, 13, 25 87(91.88.81) 98(103,97.94) 99(101.96.101) 96(101.94.95) 6, 5. 3. 3
 5-oxoproline#C 10-120 d5 10, 25, 53, 99 110(111.119,102) 103(112.99.97) 101 (113.95.96) 99(106.96.95) 10. 8. 9. 6
 taurine#C 10-100 d4 10, 25, 45, 83 99(88.112.97) 95 (98.95,92) 94 (98.94.91) 102(110.99.97) 11, 4. 6. 9
 citrullineA 2-100 d2 2, 7, 40, 83 101 (93.112.96) 100(111,95,93) 97(115.76,101) 99(110,95.93) 26, 15, 18, 9
purines/pyrimidines
 hypoxanthineA 0.5-50 d4tyr 0.5, 3, 20, 41 79(76,74,871 113(109.116,115) 103(112,94,103) 102(113.94,99) 6, 7, 8. 10
 uracil#B 0.3-3 2-15N 0.3, 0.7, 1.3, 2.5 93(106.98.90) 83(85.73,91} 93(89.90.101) 88(84.80,100) 19, 18, 13, 11
 thymine#A. 0.05-15 d4 0.05, 0.3, 5.3, 12.4 98(129.88,73) 94(98.101,83) 97(104.92.94) 95(99.95,89) 31. 11. 6. 7
 cytidine#B 0.05-15 d2 0.05, 0.3, 5.3, 12.4 92(81,100,95) 96(95.97,96) 97(96,92,103) 95(98,89,96) 12, 4. 7. 5
 uridine#B 0.05-15 d2 0.05, 0.3, 5.3, 12.4 ND 97(117,97,77) 103(101.100.109) 96(98,90,100) 18, 6, 6
 orotic acid#C 0.2-7 2-15N 0.2, 0.7,2,8. 5.8 100(99,112,90) 94(100,94,89) 101 (102,100.100) 100(97,100,103) 12, 7, 3. 6
 guanosineA 0.1-3 d5 0.1,0.35, 1.25, 2.48 91 (83,77.113) 85(92,74,90) 90(92,89,88) 96(105.77,107) 4. 12,7, 17
acylcarnitines
 carnitineA 2-125 d3 2.9.46.103 79(82. 78.77) 116(115.116.117) 104(105.103.104) 101 (105.100.99) 6. 5. 2. 4
 acetyl-carnitineA 0.5-50 d3 0.5, 3,18,41 ND 118(116.118,121) 110(110,108.111) 100(100.101.99) 3,3,4
 propionyl-
 carnitineA
0,05-5 d3 0.05.0.3.1.8.4.1 130*(127,159,106) 89(92.92,85) 85(86.88.81) 96(96.99,93) 23, 4. 4, 4
 isobutyryl-
 carnitineA
0.005-0.5 d7 0.005.0.03.0.18.
0.41
111 (120.104.108) 91 (91,93.88) 87(89.94.79) 94(93.99.90) 10. 9. 8, 11
 isovaleryl-
 carnitineB
0,005-0.5 d3 0.005.0.03.0.18.
0.41
ND 102(105.102.100) 105(106.104.106) 98(100.100.95) 3, 3, 3
 octanoyl-
 carnitineA
0.005-0.5 d3 0.005.0.03.0.18.
0.41
ND 99(96.101.99) 103(104.100.104) 100(103.98.98) 4. 3. 3
 palmitoyl-
 carnitineB
0.005-0.5 d3 0.005.0.03,0.18,
0.41
113*(110,119,NA) 87(95,76,91) 88(97,83,84) 102(103,101,103) 9, 11, 8, 1
organic acids
 2-OH butyric
 acidC
2-100 d3 2,7, 40.83 105(104.120.91) 103(109.103.98) 96(101,94,92) 96(102.92,93) 15, 5. 5. 6
 3-OH-butyric
 acidC
6-300 2-13C 6.26,125. 248 104(117.102,93) 99(105.96.95) 98(98.103,93) 95(102.92.89) 11. 6. 5, 7
 citricC 21-300 d4c 21,59.125, 248 N/A N/A 108(112.111.102) 100(122.90.87) 10,18
 fumaric#C 0.1-5 d2 0.1,0.35, 1.8.4.1 ND 84(92,69,90) 100(105,101.94) 104(121.96,94) 15, 7,14
 lacticC 510-3600 3-13C. 510, 875, 1600,
2970
97(100.101.92) 103(109.100.101) 98(101.98.95) 95(104.91.91) 5, 5, 5, 9
 malic#C 0.3-25 d3 0.3,1.1,8.8,21 101 (99.114,91) 99(112.97,88) 95(100.92,94) 97(105,93,93) 13,11.4,7
 Pyruvic#C 6-300 2-13C 6,21,115, 248 ND 83(91,85,72) 109(117,109,102) 98(105,94,97) 10, 8,6
 succinicC 1-50 4-13C 1.3.5,18,41 108(108.119.97) 99(104.101.92) 96(101,94,92) 97(104.96,91) 11, 7. 4, 6
 octanoicC 5-300 d15 5,18,113, 248 ND 96(118.90,80) 85(98.79,79) 107(119.103.100) 17,10.9
 subericC 0.5-15 d4 0.5,1.5,6.3,12.4 84(91,86.76) 101 (107.95.99) 104(108.103.100) 98(99,98,97) 12, 6. 4. 3
 azelaicC 0,5-70 d14 0.5. 2.7.5.17 ND 114(120.114.108) 108(114.106.103) 98(105.92.98) 7, 5.6
 Alphaketo-
 glutaric#C
0.5-30 4-13C 0.5, 3,13,25 104(119.104,88) 99(101,95,101) 99(105,97,95) 100(104.93,104) 22, 6, 5. 7
 ketoisovalericC 0.5-40 5-13C 0.5, 3.15, 33 ND 94(100.93,89) 99(104.98.96) 99(105.96,96) 21, 5. 4, 5
 2-oxoisocaproicC 1-60 d3 1,3.5,23,50 70* (59.81.70) 98(104.92,98) 102(107.98.100) 100(107.94.98) 11, 5. 4, 6

Note, all compounds used their own stable isotopic versions for quantification, except for where noted.

ND, not determined because the concentration was below the limit of quantification.

N/A, not available, due to a high endogenous level of analyte in the serum that prohibits accurate measurement at a quality control level.

A

denotes positive mode on EPIC-PFP column

B

denotes positive mode on C18-PFP column

C

denotes negative mode on C18-PFP column

#

denotes quantification in MS/MS mode

*

denotes that the measurement did not meet validation requirements for particular QC level.

LOQ, denotes lower limit of quantitation

Methods and Materials

Strategy

The concentration ranges used for the analytical validation of standards were selected to span normal reference ranges previously reported in plasma, but also encompassed some of the reported abnormal range seen in related diseases (Blau 2003; Hoffmann et al. 1993; Psychogios et al. 2011). These ranges extend from one to over two orders of magnitude in concentration, depending on the metabolite (Table 1). To further verify these ranges, we have processed 15 human serum samples and analyzed them along with the validation experiments. The 15 patients included 3 cohorts, representing (respectively) controls (4 patients), hepatocellular carcinoma (HCC) patients (5 patients), and patients with cirrhosis (6 patients). Though many of the compounds measured are key metabolites in inherited metabolic disorders (i.e. MCADD, MSUD, PKU, Tyrosinemia, etc.), it was not necessary to include such samples in the validation. The respective pathological concentrations are more readily detectable than the minor deviations seen in patients without Mendelian metabolic disease, and the analytical ranges we present here demonstrate that we can distinguish a diseased range for those various key metabolites. We instead confirm that our method enables quantification of these metabolites in normal and non-inherited metabolic disease samples, while demonstrating the versatility of the untargeted aspects of the platform. Previous untargeted studies of HCC and liver disease with a much larger sampling provided a good comparison for evaluating this platform, as these studies reported changes in potential biomarkers spanning a wide range of the physiochemical spectrum, including amino acids, organic acids, acylcarnitines, and a variety of bile and fatty acids (Chen et al. 2011; Zhou et al. 2012).

Sample preparation

We used a final concentration of cold 80% methanol for all sample extractions (Bruce et al. 2008). For calibration curve and quality control materials, 90μL of analyte-stripped serum (Biocell, Rancho Dominguez, CA), was supplemented with 10μL of a non-labeled standards mixture, followed by 400μL of methanol containing 10μL of a stable isotope standard mixture. Sera from patients were collected as part of a separate study approved by the UCSD IRB. Those serum samples were collected within 12 months of each other and processed within 4 hours of collection, as reported previously (Cooper et al. 2001), followed by aliquoting and immediate storage in a −80°C freezer. Samples were extracted with cold methanol as described above within several months after all sera were collected.

Stable isotopes used as internal standards were purchased from CDN Isotopes (Quebec, Canada) and Cambridge Isotopes Laboratory (Andover, MA). Several of the isotopes are 13C and 15N labeled. Although the majority are deuterated, all are labeled at carbon atoms positions that are not readily exchangeable with protons in solution. The specific atoms that are labeled for each compound are noted in a table in the supplementary methods section. Two separate stocks of isotopic standard mixes were formulated based on chemical compatibility. These were dissolved in either 0.02N HCl (final concentration) or methanol. Separate concentrations of calibrator mixes were formulated to satisfy the unique physiological concentration range of each metabolite (Suppl. Methods). A calibration curve (blank, zero, and 8 concentration levels) and four different QC levels were evaluated. On each day of validation, each QC level was extracted in triplicate and run together with a calibration curve. All standard curves had coefficient of correlation values greater than 0.99, most of which contained all 8 calibrator points (Suppl. Data). To satisfy linearity, however, several compounds had either 1 or 2 outliers removed.

HPLC and MS methodologies

All quantification was performed with two different types of pentafluorophenyl columns (C18-PFP; 2.1mm × 150mm, 3μm particle, by Mac-Mode Analytical [Chadds Ford, PA], and Epic-PFP; 2.1mm × 150mm,3μm particle, by ES Industries [West Berlin, NJ]) (Godat et al. 2010; Yang et al. 2010). Material from a single extraction was used for both column types. Amino acids and most of the acylcarnitines were quantified in positive mode on the Epic-PFP, while organic acids were quantified in negative ionization mode on the C18-PFP. The Epic-PFP provided better results for polar compounds in positive ion mode (data not shown). Eight compounds exhibited better performance with the C18-PFP in positive ionization mode (tryptophan, creatine, creatinine, cytidine, uridine, uracil, as well as isovalerylcarnitine and palmitoylcarnitine). As described for each compound table 1, three runs were used in total to quantify all 54 compounds, though most of these could be quantified with just two: positive mode on the EPIC-PFP and negative mode on the C18-PFP. Both columns were kept at 30° C for all runs. The flow rate was 300 μL/min (Shimadzu Prominence UFLC) for all runs, using similar reverse phase gradients (ACN/0.1%FA) for both columns, and optimized for comprehensive metabolite profiling.

All mass spec analysis was performed on an AB-Sciex model 5600 Triple-TOF using the Turbo ESI source. Instrument parameters are described in the supplementary methods. Nearly 70% of the samples were quantified in TOF-MS mode (Table 1). This was enabled by sufficient chromatographic selectivity of these compounds as well the ability to extract for narrow mass ranges during peak integration, enabled by the high resolution (>30,000) of the Q-TOF instrument. In all runs, each scan included a 250 millisecond TOF-MS scan followed by 50-100 millisecond individual product ion scans for specific compounds analyzed by MS/MS. Scan times were adjusted so that the total cycle was less than 1 second. This cycle time enabled at least 8-10 scans for each compound, based on their elution profiles. Several of the compounds were quantified in both MS and MS/MS mode for comparison (Table 2). For MS/MS identification of potential biomarkers, we combined the 15 patients’ sera into 3 separate pools and analyzed these on the Q-TOF with a method that maximizes for data dependent acquisitions, acquiring MS/MS scans on up to 12 non-redundant precursor ions within a 1 second total scan cycle. All compounds were quantified using MultiQuant software (AB Sciex), while MarkerView software (AB Sciex) was used for untargeted analysis. Background subtraction, peak identification, and alignment were all performed in MarkerView, as well as t-test and principal component analysis for the identification of statistically significant variations between patient cohorts.

Table 2. MS vs MS/MS quantification.

metabolite QC accuracies and %CV
MS-mode
Inter-day %Acc (Inter-day %CV)
for QC LOQ, low, med, and high
QC accuracies and %CV
MS/MS-mode
Inter-day %Acc (Inter-day %CV) for
QC LOQ, low, med, and high
malic 103(14%), 95(11%), 95(2%), 94(4%) 101(13%), 99(11%), 95(4%). 97(7%)
fumaric 103(53%), 63(11%), 99(10%), 97(10%) 150(38%), 84(15%), 100(7%), 104(14%)
pyruvic 73(3%), 90(31%), 99(21%) 83(10%), 109(8%). 102(6%)
alphaketoglutaric 88(34%), 102(9%), 102(8%). 98(6%) 104(22%). 99(6%). 99(5%). 100(7%)
orotic 74(16%), 95(11%), 101(5%), 96(7%) 100(12%). 94(7%). 101(3%). 100(6%)
uracil 88(26%), 86(14%), 82(14%), 84(14%) 98(19%). 83(18%). 93(13%). 88(11%)
thymine 96(11). 101(5%), 100(6%) 97(31%). 94(11%), 97(6%), 95(7%)
aspartate 110(28%), 111(14%). 96(12%) 100(19%). 98(9%). 106(9%)
guanosine 91(4%). 85(12%). 90(7%). 96(17%) 82(39%), 84(10%), 92(9%), 85(11%)

Comparing accuracies and %CV’s of 9 compounds that were quantified in both MS mode and MS/MS using the 5600 Q-TOF. Bolded values indicate the better mode for quantification.

Comparisons to standard clinical methods

Five of the patients’ sera were also analyzed by standard clinical methods. The majority of the compounds measured in the above Q-TOF methods were also measured using GC-MS analysis for TMS-derivatives of organic acids (Agilent 5973), Amino acid analyzer for amino acids (Biochrom 30), and triple-quadrupole LC-MS/MS for butyryl-esters of acylcarnitines (API 2000). These methods were performed as previously described (Duez et al. 1996; Hoffmann et al. 1989; Kumps et al. 1999; Millington et al. 2011; Shapira 1989; Sweetman 1991).

Results

HPLC column performance

Chromatographic runs showed good repeatability in both intra-day and inter-day assays as shown in Figure 2. Figure 2 also demonstrates the versatility in retention of organic acids and amino acids, while maintaining efficient retention and peak shape of more hydrophobic compounds such as octanoic acid, as well as octanoyl- and palmitoylcarnitine. The ability of both PFP columns to separate isomers including 2- and 3-hydroxybutyrate, isoleucine and leucine, as well as 2-methylbutyryl-, isovaleryl- and valerylcarnitine, is shown in Figure2B-G. Isobutyryl- and butyrylcarnitine were separated on the C18-PFP but not the Epic-PFP (data not shown). Nearly all the compounds maintained relatively sharp Gaussian profiles, though a few showed poorer peak shape. For example, arginine and lysine (Figure 2E) start to elute shortly at 1 minute in a broad peak that appears non-Gaussian, due in part to ionic suppression occurring near the void volume of the chromatographic runs (0.9min). Nevertheless, using stable isotopes of these compounds experiencing the same elution profiles and suppression effects, the quantification statistics were sufficient, as seen in Table 1. Most of the smaller organic acids eluted between 1.5 minutes and 4 minutes while the larger ones eluted later in the gradient. Nearly all of the amino acids also eluted within the first several (Figure 2). The unique feature of the PFP columns is that polar compounds such as organic and amino acids can be analyzed at low physiological levels under reverse phase conditions, while still showing sharp elution profiles of hydrophobic compounds. Figure 2H-I shows the chromatographic extraction from a serum sample run of 6 different hydrophobic compounds (2 bile acids and 4 long chain fatty acids) previously identified as potential biomarkers in a previous metabolomics study of HCC (Zhou et al. 2012). Though the changes in their abundances were not shown to be significant among our small cohorts, we show these traces to demonstrate the comprehensive capabilities of the Epic-PFP and C18-PFP columns in measuring both polar and non-polar/hydrophobic metabolites.

Figure 2. Resolution of quantified analytes on the PFP columns.

Figure 2

(A) 6 overlayed QC traces are shown from 3 different days on the C18-pfp in negative mode. (B+C) Organic acids resolved on C18-pfp. Note the resolution between isomers 2- and 3-hydroxybutyric acid. (D) 6 overlayed QC traces from 3 different days on Epic-PFP in positive mode. (E) Polar and neutral amino acids resolved on Epic-PFP. (F) A Trace of non-polar amino acids as well as (iso)butyrylcarnitine and hypoxanthine (tautomer described in text) are shown resolved on the Epic-PFP. Note resolution between leucine and isoleucine. (G) Resolution of isovalerylcarnitine from 2-methylbutyryl- and valeryl-carnitine standards are shown. Only isovaleryl and 2-methylbutyryl carnitines were significantly present in the patient sera. (H-I) Peak extractions of various fatty acids and bile acids are shown, compounds that were reported in previous studies as potential biomarkers of HCC.

Quantification and bioanalytical validation

Table 1 demonstrates the high accuracy and precision results obtained in the experiments. The results for 54 of the 55 compounds tested are shown, excluding glycine, which could not be detected in the lower half of the standard curve samples. 34 of the remaining 54 compounds satisfied the following standard criteria (Bansal et al. 2007; FDA 2001): (a) back-calculated accuracy deviations of ≤20% (80%-120%) for each day’s replicate measurements of LOQ QC samples and a maximum of a 20% inter-day coefficient of variation (%CV), (b) a maximum of 15% accuracy deviation (85%-115%) of intra-day replicates of low, middle, and high QC samples, and a maximum %CV of 15. Four of the compounds included in this list of 34 only missed this criterion by ≤2 percentage points for just one of the QC level measurements and were therefore included with the other 30. Many of the compounds that met these requirements included amino and organic acids, several acylcarnitines, and several purines/pyrimidines. An additional 5 compounds including lysine, hypoxanthine, citrulline, propionyl- and octanoylcarnitine missed the criteria for the LOQ by several percentage points. Another 15 compounds had accuracy and CV deviations that did not satisfy the requirements for LOQ (indicated in Table 1), but did however meet the guidelines for the other three levels of quality control replicates, enabling their validation for an adjusted range still sufficient for quantification in human samples, as seen in the reported values found in the patient samples in table 3. In those 15 serum samples, only propionylcarnitine and octanoic acid were seen at concentrations lower than the low QC, while all the other measurements fell within the adjusted range for the compounds listed with asterisks in Table 1. Six compounds (uridine, cytidine, thymine, suberic acid, azelaic acid, and octanoic acid) did not have measurable values in the patient samples. For these compounds, nearly all measurements in the patient serums fell below the lower limit of quantification, and therefore their presence will likely only be quantified in patient samples with abnormal levels. Two compounds, citrate and creatinine, had very linear standard curves and had very good signal in the measurements, but had poor accuracy and precision results at the lower levels due to high residual endogenous concentrations of these compounds in the analyte-stripped serum, causing background levels which obscure measurements at the lower analytical ranges. For future quantification of creatinine and citrate in this platform, standard curves and QC’s will be done by varying the isotopic standards of these compounds, while keeping the endogenous form constant. The results show that most of the compounds are accurately and reproducibly quantified in MS mode, though some of the compounds required MS/MS based quantification, due either to overlap with co-eluting analytes including isobaric compounds, or just due to increased sensitivity obtained in product ion scanning mode (Table 2). For example, the isotopic standard for taurine had nearly the identical mass as 5-oxoproline, necessitating specific product scans and quantitation on compound specific product ions. Similarly, an isobaric compound with a similar elution time and precursor mass as valine necessitated product ion quantification.

Table 3. Metabolite concentrations in patient samples.

Ala
70-700
Leu
30-250
lie
10-125
Val
40-350
Pro
40-500
Phe
20-200
Try
6-120
Tyr
20-150
Cystine
30-200
Met
5-50
Thr
25-300
Ser
30-300
Asn
20-150
Gin
100-800
Lys
21-300
His
40-250
Arg
20-150
Asp
2-50
Glu
3-150
Control 180
(78)
115
(57)
60
(25)
176
(65)
235
(65)
64
(8.6)
45
(17)
80
(20)
37
(11)
25
(10)
185
(36)
81
(20)
47
(15)
547
(48)
220
(10)
59
(11)
96
(4.0)
6.8
(3.0)
40
(22)
HOC 241
(126)
96
(44)
48
(22)
140
(68)
201
(88)
108
(43)
57
(27)
142
(76)
43
(13)
83
(92)
268
(128)
153
(118)
72
(46)
606
(211)
226
(121)
60
(46)
96
(30)
16
(10)
41
(25)
Cirr. 198
(33)
79
(32)
39
(9.0)
125
(48)
167
(38)
90
(26)
47
(13)
106
(31)
51
(23)
36
(12)
180
(70)
108
(23)
50
(12)
545
(110)
204
(36)
52
(10)
95
(35)
15
(8.6)
46
(32)
Ornithine
10-200
Homo-
cystine
5-75
Creatine
5-150
Pipecolic
1-30
5-oxo-
proline
10-120
Citrulline
2-100
Hypoxan-
thine
0.5-50
Uracil
0.3-3.0
Thymine
0.05-15
Cytidine
0.05-15
Uridine
0.3-15
Orotic
0.2-7
Guanosine
0.1-3
Carnitine
2-125
Acetyl-
carnitine
3-50
Propionyl-
camitine
0.005-0.5
Control 97
(18)
7.9
(3.8)
24
(15)
4.3
(1.8)
81
(5.9)
36
(21)
5.9
(2.6)
0.07
(0.03)
NA 0.03
(0.02)
4.3
(2.1)
0.20
(0.28)
0.76
(0.76)
39
(8.6)
8.9
(1.9)
0.09
(0.02)
HCC 108
(72)
4.8
(4.2)
20
(14)
6.3
(3.4)
86
(28)
34
(8)
7.1
(3.1)
0.07
(0.02)
NA 0.03
(0.02)
3.5
(0.94)
0.11
(0.04)
0.15
(0.03)
45
(20)
9.2
(3.8)
0.10
(0.03)
Cirr. 101
(24)
4.1
(2.8)
14
(4.9)
10
(9.1)
81
(16)
39
(11)
9.0
(4.1)
0.08
(0.02)
NA 0.06
(0.05)
4.3
(0.80)
0.14
(0.12)
0.16
(0.03)
33
(7.5)
10
(5.2)
0.06
(0.02)
Isobutyryl-
camitine.
0.005-0.5
Isovaleryl-
camitine
0.03-0.5
Octanoyl-
camitine
0.03-0.5
Palmitoyl-
camitine
0.005-0.5
2-OH-
butyrate
2-100
3-OH-
butyrate
6-300
Fumarate
0.3-5
Lactate
510-3600
Malate
0.3-25
Pyruvate
21-300
Succinate
1-50
Octanoic
18-300
Suberic
0.5-15
Azelalc
2-20
Alphaketo-
gluatric.
0.5-30
Ketoiso-
valeric
3-40
2-oxoiso-
caproic.
1-60
Control 323
(151)
131
(65)
108
(47)
81
(23)
22
(10)
52
(31)
3.1
(1.6)
1256
(287)
10
(11)
163
(49)
4.2
(1.8)
NA NA NA 14
(8.6)
10
(3.8)
20
(9.5)
HCC 176
(107)
64
(33)
81
(14)
94
(10)
42
(27)
18
(20)
3.3
(2.7)
2717
(858)
6.6
(14)
175
(95)
3.5
(1.0)
NA NA NA 18
(12)
13
(1.9)
18
(5.9)
Cirr. 141
(45)
68
(37)
125
(39)
121
(19)
37
(18)
62
(47)
3.9
(3.4)
2368
(890)
6.7
(1.6)
168
(60)
4.2
(13)
NA NA NA 19
(8.0)
10
(1.9)
16
(5.7)

All Concentrations are in μM. Standard deviations of average concentrations are shown in parentheses.

Ranges of concentrations measured are shown below each respective metabolite name. Those compounds that did not qualify at the LOQ target concentration (listed in 31) show an adjusted range based on the low QC.

N/A indicates that the concentration of the metabolite was too low to accurately measure in patient sera.

We used stable isotopes of each respective compound for quantification, except for a few cases. Asparagine was quantified with a glutamine standard and threonine was quantified with the stable isotope of methionine, as their respective isotopes were not available to us at the time of the study. Nonetheless the accuracy and precision results show these are adequate substitutes. Additionally, hypoxanthine was quantified using deuterated tyrosine. A hypoxanthine isotope was included but surprisingly eluted several minutes from the non-labeled standard (Figure 2), which we verified by MS/MS. This could be due to the tautomerization of hypoxanthine (Hernandez et al. 1996).

Comparisons to existing clinical methodologies (Cross-validation)

We have done preliminary comparisons between our current methodology with standard clinical methodologies currently used for amino acid, organic acid, and acylcarnitine analysis, as described in the methods. Five patients’ sera samples were used for the comparison of many of the metabolites verified in the Q-TOF methods (Suppl. Table 1). Most of the amino acids and acylcarnitines showed accuracy measurements between 75 and 125% with standard deviations below 20%, while some of the organic acids showed greater discrepancy. This is not surprising considering that the GC-MS method did not use stable isotopes, and several of the organic acids were too low in concentration to be accurately reported by the older method. The preliminary comparisons showed that the Q-TOF methods exhibited increased sensitivity for a number of compounds, especially organic acids. Another advantage of the Q-TOF platform was the ability to better resolve isomers such as isovaleryl-, 2-methylbutyryl- and valerylcarnitine, which co-elute in the original method (Forni et al. 2010).

Patient Samples and Untargeted Analysis

We used untargeted methods to process the control, HCC, and cirrhotic samples from the same runs from which the quantitative data was extracted. Using marker view software, peaks were aligned and integrated followed by both multivariate and ANOVA statistical processing. From t-test analysis of all peak groups, we identified over 30 unique analytes of interest. One such peak was identified by MS/MS database matching (Tautenhahn et al. 2012) as inosine (box plot in Figure 1D), a compound determined in the previous HCC metabolomics study as being the most differentiated biomarker (Chen et al. 2011). Biological conclusions cannot be formed from this data though, due to the small sample number of each cohort, intended primarily for analytical validation.

Discussion

Using the high resolution and mass accuracy of a new Q-TOF instrument, we demonstrate that quantification of clinically relevant metabolites can be done in MS mode instead of MS/MS mode in many cases, allowing each cycle to collect data on all ionized precursor ions, which could later be evaluated by untargeted analysis. Quantifying in MS mode has been considered a challenge because of background ions from the matrix as well as lower signal intensity. We have been able to eliminate the background for most of the targeted analytes by employing recent advances in chromatographic stationary phases, specifically PFP-based columns. For several compounds, including taurine, ornithine, 5-oxoproline acid and valine, background signals from the serum matrix and/or spiked compounds obscured signals in MS mode, necessitating specific product scans (MS/MS) for accurate quantification. In other cases, as demonstrated in Table 2, product ion quantification improved accuracy and precision results, specifically at the lower levels of concentration. The fast scanning rate of the instrument (up to 100Hz) enabled us to collect short MS/MS scans within the same scanning cycle as the precursor ion data collection (TOF-MS scan). Triple-quadrupole instruments, which lack a time-of-flight tube but include an additional quadrupole for daughter ion selection, generally allow for even faster scanning with less ion loss, and could therefore quantify a greater number of compounds by MS/MS. However, a triple-quadrupole instrument lacks the ability to profile co-eluting ions at high resolution and mass accuracy. The results show that the majority of analytes quantified on the Q-TOF, including most of the amino acids, many organic acids, acylcarnitines and purines/pyrimidines, met or exceeded accuracy and precision criteria common to analytical validation of triple-quadrupole methods. Though Q-TOF instruments were previously thought to lack the dynamic range for quantification of such compounds at physiological ranges, improvements in this new generation Q-TOF instrument, in conjunction with the aforementioned columns, allowed this goal to be achieved. The sensitivity levels also met or exceeded those measured by standard clinical methods routinely used for measuring many of these compounds. Organic acids were quantifiable at concentrations as low as 0.3μM (malic) and 5nM for several acylcarnitines. We used fewer replicates per day in the current validation than prescribed in other validation protocols, which often use 6 replicates to test intra-day characteristics (Honour 2011). This was due to length of our runs (40min. each), which limited the number of QC replicates we could analyze so that each batch did not greatly exceed 1 full day of collection, potentially compromising post-preparative stability. Further validation experiments will be performed to further determine long-term reproducibility and ruggedness of the method.

Though most of the compounds that were quantified were polar in nature, we have shown by extracting signals for a variety of fatty acids, bile acids, and other lipid forms from the patient serum samples, that this new Q-TOF/LC-MS strategy is effective in analyzing many classes of non-polar compounds with untargeted analysis, some of which we plan to quantitate in future iterations of this platform. Arguably, polar metabolite profiling (especially of organic acids) has been under-represented in traditional reverse-phase LC-MS methods. The PFP-columns have significantly advanced the ability to quantitate polar metabolites relevant to clinical applications, without the use of ion-pairing agents (which limit analysis to a single ionization mode) or HILIC columns, which may compromise analysis of hydrophobic compounds (Buescher et al. 2010; Ikegami et al. 2008).

A hybrid targeted/untargeted method held to the conventional rigorous validation standards used with triple-quadrupole methodology, permits for comprehensive metabolic diagnosis, without sacrificing all of the other information potentially useful for future untargeted studies. Specifically, such a platform would be effective as both a clinical and research tool for the diagnosis and follow-up of Mendelian metabolic disorders (i.e. where pathognomonic compounds are elevated, such as tyrosine in tyrosinemia, phenylalanine in PKU, and branched chain amino acids in MSUD), additionally enabling discovery of unforeseen metabolic changes, or the effects of drug therapies and consequent drug metabolism. These methods can also be used in a more traditional research approach for the study of non-Mendelian disease states such as liver disease and cancer, where known changes in major metabolic pathways (i.e. amino acid metabolism and cellular/energy pathways (Dejong et al. 2007; Moreno-Sanchez et al. 2007)) can be quantified, while still enabling the search for potential biomarkers. The current methods help to reduce the need for untargeted sample processing to occur within very short time frames, usually done to limit instrument drift effects. Having many internal standards and absolute quantification results of different classes of metabolites can improve the normalization of samples processed that are run many weeks or months apart, something currently presented as a challenge in untargeted studies. Also, collecting full scan, accurate mass data with good chemometric properties allows one to revisit the original dataset if a new, relevant metabolite is later identified.

Supplementary Material

suppl data
suppl table1

Acknowledgements

We thank Julio A. Gutierrez and Robert T. Schooley for providing patient sera. We would like to thank William L. Nyhan for valuable suggestions. We thank Bonnie Holmes, Kasie Auler and Priscilla Burks for running GC-MS, amino acid analyzer, and triple-quadrupole methods.

Abbreviations

CLIA

Clinical laboratory improvement act

CAP

College for American pathologists

GLP

Good laboratory practices

HCC

Hepatocellular carcinoma

Cirr

cirrhotic patient

ND

not determined

TOF

Time of flight

LOQ

(lower) limit of quantification

QC

quality control

CV

coefficient of variation

HRMS

high resolution mass spectrometry

MSUD

Maple syrup urine disease

MCADD

Medium chain acyl-CoA dehydrogenase deficiency

PKU

Phenylketonuria

References

  1. Bansal S, DeStefano A. Key elements of bioanalytical method validation for small molecules. AAPS J. 2007;9:E109–14. doi: 10.1208/aapsj0901011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Blau N, Duran M, Blaskovics ME, Gibson KM. Physician’s Guide to the Laboratory Diagnosis of Metabolic Diseases. Xxxvii. Springer; Berlin: 2003. [Google Scholar]
  3. Bruce SJ, et al. Evaluation of a protocol for metabolic profiling studies on human blood plasma by combined ultra-performance liquid chromatography/mass spectrometry: From extraction to data analysis. Anal Biochem. 2008;372:237–49. doi: 10.1016/j.ab.2007.09.037. [DOI] [PubMed] [Google Scholar]
  4. Buescher JM, Moco S, Sauer U, Zamboni N. Ultrahigh performance liquid chromatography-tandem mass spectrometry method for fast and robust quantification of anionic and aromatic metabolites. Anal Chem. 2010;82:4403–12. doi: 10.1021/ac100101d. [DOI] [PubMed] [Google Scholar]
  5. Chace DH, Kalas TA. A biochemical perspective on the use of tandem mass spectrometry for newborn screening and clinical testing. Clin Biochem. 2005;38:296–309. doi: 10.1016/j.clinbiochem.2005.01.017. [DOI] [PubMed] [Google Scholar]
  6. Chen T, et al. Serum and urine metabolite profiling reveals potential biomarkers of human hepatocellular carcinoma. Mol Cell Proteomics. 2011;10:M110004945. doi: 10.1074/mcp.M110.004945. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Cooper HM, Paterson Y. Production of polyclonal antisera. Curr Protoc Immunol. 2001 doi: 10.1002/0471142735.im0204s13. Chapter 2, Unit 2 4. [DOI] [PubMed] [Google Scholar]
  8. Crews B, et al. Variability analysis of human plasma and cerebral spinal fluid reveals statistical significance of changes in mass spectrometry-based metabolomics data. Anal Chem. 2009;81:8538–44. doi: 10.1021/ac9014947. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Dejong CH, van de Poll MC, Soeters PB, Jalan R, Olde Damink SW. Aromatic amino acid metabolism during liver failure. J Nutr. 2007;137:1579S–1585S. doi: 10.1093/jn/137.6.1579S. [DOI] [PubMed] [Google Scholar]
  10. Dillen L, et al. Comparison of triple quadrupole and high-resolution TOF-MS for quantification of peptides. Bioanalysis. 2012;4:565–79. doi: 10.4155/bio.12.3. [DOI] [PubMed] [Google Scholar]
  11. Duez P, Kumps A, Mardens Y. GC-MS profiling of urinary organic acids evaluated as a quantitative method. Clin Chem. 1996;42:1609–15. [PubMed] [Google Scholar]
  12. Dunn WB, Broadhurst DI, Atherton HJ, Goodacre R, Griffin JL. Systems level studies of mammalian metabolomes: the roles of mass spectrometry and nuclear magnetic resonance spectroscopy. Chem Soc Rev. 2011;40:387–426. doi: 10.1039/b906712b. [DOI] [PubMed] [Google Scholar]
  13. Food and Drug Adminstration. U.S. Department of Health and Human Services. Center for Drug Evaluation and Research (CDER) Center for Veterinary Medicine(CVM) Guidance for Industry: Bioanalytical Method Validation. 2001 [Google Scholar]
  14. Forni S, Fu X, Palmer SE, Sweetman L. Rapid determination of C4-acylcarnitine and C5-acylcarnitine isomers in plasma and dried blood spots by UPLC-MS/MS as a second tier test following flow-injection MS/MS acylcarnitine profile analysis. Mol Genet Metab. 2010;101:25–32. doi: 10.1016/j.ymgme.2010.05.012. [DOI] [PubMed] [Google Scholar]
  15. Fung EN, Xia YQ, Aubry AF, Zeng J, Olah T, Jemal M. Full-scan high resolution accurate mass spectrometry (HRMS) in regulated bioanalysis: LC-HRMS for the quantitation of prednisone and prednisolone in human plasma. J Chromatogr B Analyt Technol Biomed Life Sci. 2011;879:2919–27. doi: 10.1016/j.jchromb.2011.08.025. [DOI] [PubMed] [Google Scholar]
  16. Godat E, Madalinski G, Muller L, Heilier JF, Labarre J, Junot C. Mass spectrometry-based methods for the determination of sulfur and related metabolite concentrations in cell extracts. Methods Enzymol. 2010;473:41–76. doi: 10.1016/S0076-6879(10)73002-0. [DOI] [PubMed] [Google Scholar]
  17. Henry H, Sobhi HR, Scheibner O, Bromirski M, Nimkar SB, Rochat B. Comparison between a high-resolution single-stage Orbitrap and a triple quadrupole mass spectrometer for quantitative analyses of drugs. Rapid Commun Mass Spectrom. 2012;26:499–509. doi: 10.1002/rcm.6121. [DOI] [PubMed] [Google Scholar]
  18. Hernandez B, Orozco M, Luque FJ. Tautomerism of xanthine and alloxanthine: a model for substrate recognition by xanthine oxidase. J Comput Aided Mol Des. 1996;10:535–44. doi: 10.1007/BF00134177. [DOI] [PubMed] [Google Scholar]
  19. Hoffmann G, Aramaki S, Blum-Hoffmann E, Nyhan WL, Sweetman L. Quantitative analysis for organic acids in biological samples: batch isolation followed by gas chromatographic-mass spectrometric analysis. Clin Chem. 1989;35:587–95. [PubMed] [Google Scholar]
  20. Hoffmann GF, Meier-Augenstein W, Stockler S, Surtees R, Rating D, Nyhan WL. Physiology and pathophysiology of organic acids in cerebrospinal fluid. J Inherit Metab Dis. 1993;16:648–69. doi: 10.1007/BF00711898. [DOI] [PubMed] [Google Scholar]
  21. Honour JW. Development and validation of a quantitative assay based on tandem mass spectrometry. Ann Clin Biochem. 2011;48:97–111. doi: 10.1258/acb.2010.010176. [DOI] [PubMed] [Google Scholar]
  22. Human Metabolome Database [Accessed 06/10/12];2012 www.hmdb.org.
  23. Ikegami T, Tomomatsu K, Takubo H, Horie K, Tanaka N. Separation efficiencies in hydrophilic interaction chromatography. J Chromatogr A. 2008;1184:474–503. doi: 10.1016/j.chroma.2008.01.075. [DOI] [PubMed] [Google Scholar]
  24. Kaufmann A, Butcher P, Maden K, Walker S, Widmer M. Quantitative and confirmative performance of liquid chromatography coupled to high-resolution mass spectrometry compared to tandem mass spectrometry. Rapid Commun Mass Spectrom. 2011;25:979–92. doi: 10.1002/rcm.4952. [DOI] [PubMed] [Google Scholar]
  25. Kell DB. Metabolomic biomarkers: search, discovery and validation. Expert Rev Mol Diagn. 2007;7:329–33. doi: 10.1586/14737159.7.4.329. [DOI] [PubMed] [Google Scholar]
  26. Kumps A, Duez P, Genin J, Mardens Y. Gas chromatography-mass spectrometry analysis of organic acids: altered quantitative response for aqueous calibrators and dilute urine specimens. Clin Chem. 1999;45:1297–300. [PubMed] [Google Scholar]
  27. Mamas M, Dunn WB, Neyses L, Goodacre R. The role of metabolites and metabolomics in clinically applicable biomarkers of disease. Arch Toxicol. 2011;85:5–17. doi: 10.1007/s00204-010-0609-6. [DOI] [PubMed] [Google Scholar]
  28. Millington DS, Stevens RD. Acylcarnitines: analysis in plasma and whole blood using tandem mass spectrometry. Methods Mol Biol. 2011;708:55–72. doi: 10.1007/978-1-61737-985-7_3. [DOI] [PubMed] [Google Scholar]
  29. Moreno-Sanchez R, Rodriguez-Enriquez S, Marin-Hernandez A, Saavedra E. Energy metabolism in tumor cells. FEBS J. 2007;274:1393–418. doi: 10.1111/j.1742-4658.2007.05686.x. [DOI] [PubMed] [Google Scholar]
  30. Psychogios N, et al. The human serum metabolome. PLoS One. 2011;6:e16957. doi: 10.1371/journal.pone.0016957. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Shapira E, Blitzer MG, Miller JB, Affrick DK. Biochemical Gentics: A Laboratory Manual. Oxford University; US: 1989. [Google Scholar]
  32. Sweetman L. Organic acid analysis. In: Holmes FA, editor. Techniques in Diagnostic Human Biochemical Genetics: A Laboratory Manual. Wiley-Liss; New York: 1991. pp. 143–176. [Google Scholar]
  33. Tautenhahn R, Cho K, Uritboonthai W, Zhu Z, Patti GJ, Siuzdak G. An accelerated workflow for untargeted metabolomics using the METLIN database. Nat Biotechnol. 2012;30:826–8. doi: 10.1038/nbt.2348. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. Vreken P, van Lint AE, Bootsma AH, Overmars H, Wanders RJ, van Gennip AH. Rapid diagnosis of organic acidemias and fatty-acid oxidation defects by quantitative electrospray tandem-MS acyl-carnitine analysis in plasma. Adv Exp Med Biol. 1999;466:327–37. doi: 10.1007/0-306-46818-2_38. [DOI] [PubMed] [Google Scholar]
  35. Wilcken B, Wiley V. Newborn screening. Pathology. 2008;40:104–15. doi: 10.1080/00313020701813743. [DOI] [PubMed] [Google Scholar]
  36. Yanes O, Tautenhahn R, Patti GJ, Siuzdak G. Expanding coverage of the metabolome for global metabolite profiling. Anal Chem. 2011;83:2152–61. doi: 10.1021/ac102981k. [DOI] [PMC free article] [PubMed] [Google Scholar]
  37. Yang S, Sadilek M, Lidstrom ME. Streamlined pentafluorophenylpropyl column liquid chromatography-tandem quadrupole mass spectrometry and global (13)C-labeled internal standards improve performance for quantitative metabolomics in bacteria. J Chromatogr A. 2010;1217:7401–10. doi: 10.1016/j.chroma.2010.09.055. [DOI] [PMC free article] [PubMed] [Google Scholar]
  38. Zhang NR, Yu S, Tiller P, Yeh S, Mahan E, Emary WB. Quantitation of small molecules using high-resolution accurate mass spectrometers - a different approach for analysis of biological samples. Rapid Commun Mass Spectrom. 2009;23:1085–94. doi: 10.1002/rcm.3975. [DOI] [PubMed] [Google Scholar]
  39. Zhou L, et al. Serum metabolomics reveals the deregulation of fatty acids metabolism in hepatocellular carcinoma and chronic liver diseases. Anal Bioanal Chem. 2012;403:203–13. doi: 10.1007/s00216-012-5782-4. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

suppl data
suppl table1

RESOURCES