Abstract
BACKGROUND
According to field effect theory, by detecting microvasculature changes in the early increase of blood supply (EIBS) in the surrounding tissue neoplastic lesions can be identified from a distance.
OBJECTIVE
To determine the feasibility and efficacy of a fiberoptic probe containing novel Polarization Gating Spectroscopy (PGS) technology to identify patients with pancreatic adenocarcinoma (PAC) by field effect theory.
DESIGN
Prospective cohort (pilot) study
SETTING
Outpatient tertiary care center
PATIENTS
Adult (≥18 years) patients undergoing EGD-EUS were screened. Patients with PAC were included in the “cancer” group and patients without PAC were included in the “control” group. We excluded patients with other known malignancies and gastro-duodenal premalignant lesions.
INTERVENTIONS and MAIN OUTCOME MEASURES
Spectroscopic measurements of EIBS variables, such as deoxyhemoglobin concentration (DHb) and mean blood vessel radius (BVR), were obtained from five peri-ampullary locations. The Mann-Whitney rank sum test was used for the statistical analysis (p≤0.05).
RESULTS
Fourteen patients (mean age: 72 years,79% male) in the cancer group and 15 patients (mean age 63 years,60% male) in the control group were included in the final analysis. At the ampullary site, both DHb (p=0.001) and BVR (p=0.03) were higher in PAC patients than in the controls. The DHb alone (92% sensitivity,86% specificity) or in combination with BVR (92% sensitivity,71% specificity) can differentiate PAC from controls with high accuracy.
LIMITATIONS
Small sample size, Unmatched controls
CONCLUSIONS
Spectroscopic measurements of EIBS by fiberoptic probes are feasible. Preliminary evidence suggests that in vivo measurement of normal-appearing duodenal tissue can differentiate PAC patients from a distance with high accuracy.
Keywords: Pancreas adenocarcinoma, Spectroscopy, fiberoptic probe
Background and Introduction
Pancreatic cancer is the fourth leading cause of cancer death in the United States and is associated with a poor prognosis. The mortality rate is approximately 74% at 1 year and 94% at 5 years.1 The average life expectancy after diagnosis is approximately 5 to 8 months.2 At present, successful surgical resection is the only curative therapy that can improve long-term survival. However, it can be achieved only when a tumor is detected at an early stage.3 Unfortunately, due to non-specific symptoms associated with pancreatic cancer, it is commonly detected in the later stages of the disease.4
In the past decade, there has been significant improvement in the quality of imaging studies (e.g. CT scan, MRI, Endoscopic ultrasound) and development of disease specific molecular markers (e.g. CA 19–9). However, their use has failed to significantly improve the mortality rate of pancreatic cancer.5 Unfortunately, fewer than 20% of patients are considered eligible candidates for curative surgical resection at the time of diagnosis.6 This suggests that although modern imaging methods may define the diagnosis of pancreatic cancer with more precision, they have failed to improve the prognosis associated with it. Therefore, identifying patients with early pancreatic cancer and developing screening strategies for high-risk patients is of immense importance.7, 8 So far, there have been no definite biomarkers or imaging techniques proven to be a safe, sensitive, and cost effective strategy for pancreatic cancer screening in the general population. However, recent technological advances have been able to detect early neoplastic changes in the field of tissue surrounding the pancreas and other solid tumors for its early detection.9–12
In this study, we hypothesized that pancreatic adenocarcinoma (PAC) could be detected by measuring the changes in the early increase in blood supply (EIBS)13 found in the surrounding normal-appearing duodenal tissue with Polarization Gating Spectroscopy (PGS) technology.
We aim to evaluate the feasibility and efficacy of PGS measurements in the duodenum during endoscopy procedures and to evaluate EIBS markers, deoxyhemoglobin concentration (DHb), and average blood vessel radius (BVR) in patients with PAC versus controls.
Materials and Methods
Study Design and Sample Size
This pilot study was a single center, open label, prospective cohort study performed at the Mayo Clinic in Jacksonville, Florida in partnership with the Biomedical Engineering Department of Northwestern University in Evanston, Illinois. We planned to recruit a total of 15 patients with pathologically confirmed PAC (cancer group) and 15 patients without PAC (control group) to compare spectroscopic measurements. The sample size was based on both feasibility and the primary aim to gain sufficient pilot data to provide reasonable point estimates for each measurement to plan future studies. The Institutional Review Board of Mayo Clinic approved the study and all patients signed the informed consent documentation.
Study patients
We screened all patients who were pre-scheduled to receive Esophagogastroduodenoscopy (EGD) with upper Endoscopic Ultrasound (EUS). Patients with a known, recent history of PAC (untreated) were included in the cancer group, and patients without a known history of PAC were included in the control group. Patient's eligibility in the cancer group or the control group was determined based on the inclusion and exclusion criteria of our study (see Table 1). All patients in the cancer group were diagnosed with pancreatic adenocarcinoma. Patients with pancreatic neuroendocrine tumors were excluded from the study. As described in the exclusion criteria, patients in the control group had no malignant or premalignant lesions in the pancreas or gastro-duodenal area. Patients with chronic pancreatitis were excluded from the PAC group, but not from the control group. All patients with visible inflammatory conditions in the upper gastrointestinal tract were excluded from both the PAC group and the control group. Patients in the control group received an upper EUS for the indication of abdominal pain. Patients with EGD/upper EUS findings that did not meet the exclusion criteria were considered a “screen failure” and were excluded from the final study analysis.
Table 1.
Inclusion and exclusion criteria of the study
| Inclusion Criteria | Age 18 years or older |
| Able to provide informed written consent | |
| Patient scheduled for previously planned EGD with upper EUS | |
| Patients with known adenocarcinoma of the pancreas included in the cancer group | |
| Patients with abdominal imaging studies (e.g. CT abdomen or MRI abdomen) negative for malignancy in past 5 years included in the control group | |
| Exclusion Criteria | Unable to obtain biopsy specimen or fine-needle aspiration results of the pancreas lesion (e.g. coagulation disorder, inadequate sample) |
| Presence of malignant lesion* in the pancreas or the duodenum (e.g. Neuroendocrine tumor, Gastrointestinal stromal tumor) | |
| Known familial disorder with high risk of pancreas cancer development (e.g. Familial adenomatous polyposis syndrome, Hereditary non-polyposis colorectal cancer syndrome, Juvenile polyposis syndrome) | |
| Significant family history of pancreatic cancer (at least one first degree relative with pancreas cancer) | |
| Presence of premalignant lesions (e.g. Duodenal adenoma, Pancreas intraductal papillary mucinous neoplasm) | |
| Active visible inflammation/ulcer in the stomach or the duodenum. | |
| Patients with known chronic pancreatitis were excluded from the cancer group† | |
| Know pregnancy or sexually active females of childbearing age not receiving an accepted form of birth control method |
Other than pancreas adenocarcinoma
Chronic pancreatitis patients were allowed to be included in the “control” group only
Polarization Gating Spectroscopy (PGS)
PGS is an optical method, which measures the intensity of light scattering with the help of both polarization and wavelength (λ) (see Figure 1). The polarization dependence of the scattered light allows depth-selective interrogation of tissue. The collection of polarized light parallel (I∥) and perpendicular (I⊥) to the incident beam allows the analysis of the I∥, I⊥, and difference between I∥ and I⊥ of the signals.14 PGS light signals interrogate progressively deeper into the tissue at estimated maximum penetration depths of 200, 270, and 400 μm at the ampulla, respectively. Analysis of the wavelength dependence of these signals allows measurements of oxyhemoglobin concentration (OHb), deoxyhemoglobin concentration (DHb), and the mean blood vessel radius (BVR) via a modified Beer-Lambert algorithm (see supplement 1 for details).14–16
Figure 1.

Schematic of the polarization-gated probe (A) and representation of data acquisition graph (B).
The measurement method with a PGS fiber-optic probe has been described previously in detail.16, 17 The PGS measurement unit consists of the following components: (1) a mobile cart with a light source, a spectrometer, a computer processor, monitor, a keyboard, and calibration equipment (figure 2-A) and (2) a fiber-optic probe (figure 2-B). The fiber-optic probe is reusable. It was sterilized and reprocessed with Cidex® solution (Ethicon endo-surgery Inc., Cincinnati, Ohio) in a similar standardized fashion as with other endoscopes.
Figure 2.




(A) Spectroscopy processor, monitor, and keyboard of polarization gating spectroscopy measurement unit. (B) Fiber-optic probe of polarization gating spectroscopic measurement unit. (C) Polarization gating fiber-optic probe fed through the accessory channel of the standard upper endoscope (D) Peri-ampullary locations where spectroscopy measurements were obtained.
Procedural Details
Each patient received EGD with upper EUS as scheduled by an experienced endoscopist participating in the study (M.W., M.R., T.W.). All patients were sedated with monitored anesthesia under the guidance of an anesthesiology provider.
Before obtaining each spectroscopic measurement, calibration of the fiber-optic system was performed using air, water, and 99% diffuse reflectance standards. The air and water standards accounted for stray reflections in the probe, whereas the reflectance standard was used to account for the intensity spectrum of the light source and system. During EGD, the optic probe was inserted inside the accessory channel of the upper endoscope and advanced to the tip of the endoscope (see figure 2-C). The optic probe was gently brought in contact with the peri-ampullary duodenal mucosa at five different desired locations: (1) directly on the ampulla, (2) approximately 5 mm proximal from the ampulla, (3) approximately 5 mm distal from the ampulla, (4) 1 cm proximal from the ampulla, and (5) 1 cm distal from the ampulla. We obtained spectroscopy measurements four times in each of these five peri-ampullary locations (see figure 2-D). The rest of the EGD and upper EUS-FNA endoscopy procedures were then completed as clinically indicated. All results, including FNA results (if obtained), were recorded. During the procedure, all visualized mucosal abnormalities were recorded and photographed.
Patients were monitored and then discharged after the endoscopy procedure per standard protocol. Upon discharge, participants were provided information to contact the study investigator if they developed any new or worsening symptoms. Any reported post-procedural adverse events were recorded.
Data analysis
After obtaining spectroscopy measurements, the data was stored in the spectroscopy processor and transferred to Northwestern University biomedical engineering laboratory for final analysis. Spectra measurements at each desired location with an ratio between 1 and 5, I∥ signal intensity >1% of the reflectance standard intensity, and oxygenation saturation greater than 2% were considered optimal quality. Spectra not meeting the above criteria were excluded from the final analysis. These criteria ensured avoidance of inadequate intensity signals in the analysis. EIBS variables, DHb, and BVR were calculated from each spectra, and their average value at each desired periampullary location was calculated. A given periampullary location had to possess at least 2 satisfactory for the average values to be included in the final analysis. To minimize the bias, the engineer performing the data analysis was blinded to the patient status (cancer versus control).
Demographic information was described using frequency statistics. The DHb and BVR measurements were compared between patients in cancer and the control group by using the Mann-Whitney rank sum test using Minitab version 16.1.1 (Minitab Inc., State College, PA). At various cutoff values on the receiver operating characteristic (ROC) curve, the sensitivity and specificity of DHb and BVR were determined to differentiate patients with PAC from controls. The leave-one-out cross-validation procedure was performed using Stata version 8 (StataCorp, College Station, TX). The results were assessed for any effect of potential confounding factors, such as age, gender, ethnicity, history of smoking or alcohol use, tumor size, or tumor location, using the analysis of covariance (ANCOVA) test. We used a p-value of less than 0.05 to determine the statistical significance of the test results. This manuscript was prepared and reviewed by the contributing authors of this study. Additionally, all authors had access to the study data and had reviewed and approved the final manuscript.
Results
We enrolled a total of 37 patients in the study. Of these, five patients with neuroendocrine tumors and 2 patients with duodenal anatomy limiting the use of the probe were excluded due to “screen failure.” We recruited a total 15 patients with PACs to the cancer group and 15 patients without PACs to the control group. These individuals met all inclusion and exclusion criteria of the study. Among the 15 patients in the cancer group, 1 patient was excluded from the final analysis due to suboptimal measurements. In the final analysis, 14 patients in the cancer group were compared with 15 patients in the control group. The demographics of the recruited patients in the cancer and control groups are shown in Table 2. Among the 15 patients in the control group, 2 patients were found to have chronic pancreatitis based on five or more EUS features. The rest of the patients had normal pancreatic architecture and received the diagnosis of functional abdominal pain. The preparation of the cart and the spectroscopic measurements took an average approximately 10 minutes.
Table 2.
Demographics of patients in “cancer” and “control” groups.
| Demographic Variables | “Cancer” group (n=14) | “Control” group (n=15) |
|---|---|---|
| Mean age (± SD) | 72 years (± 10 years) | 63 years (± 11 years) |
| Male gender | 11 (79%) | 9 (60%) |
| White ethnicity | 13 (93%) | 13 (87%) |
| History of smoking tobacco | 7 (50%) | 9 (60%) |
| History of alcohol use | 6 (43%) | 13 (87%) |
(SD- Standard Deviation)
Ampulla – Optimal Location for Spectroscopy Measurements
Among all five peri-ampullary locations, the EIBS variables, DHb (p = 0.001) and BVR (p = 0.03) measured directly from the ampulla had the least variability and showed maximal ability to differentiate the PAC patients from controls (supplement 2). Below, we will discuss the details of the measurements obtained directly from the ampulla.
Accuracy to Differentiate Patients with PAC
The scatterplot of DHb and BVR measurements from the ampulla between patients in the cancer and control groups is shown in figure 3-A. The scatterplot demonstrates the higher levels of both DHb and BVR among patients in the cancer group compared with patients in the control group. After plotting DHb and BVR measurements on ROC curves (figure 3-B, 3-C), their sensitivity and specificity were determined using variable cutoff values (see table 3). In ROC curve analysis, DHb measurements alone achieved a sensitivity of 92% and specificity of 86% to differentiate PAC from controls. DHb in combination with BVR measurements achieved a sensitivity of 92% and specificity of 71% to differentiate PAC from controls. After performing leave-one-out cross-validation analysis, the DHb retained a sensitivity of 92% and specificity of 71%.
Figure 3.


(A) The scatterplot of ampullary measurements among patients in “cancer” and “control” group. (B) Receiving Operating Curve (ROC) using DHb (C) ROC using BVR to differentiate presence of pancreas cancer compared with control.
Table 3.
Accuracy of blood vessel radius (BVR) and deoxyhemoglobin concentration (DHb) measurements.
| Variables | Sensitivity | Specificity | AUC in ROC |
|---|---|---|---|
| Deoxyhemoglobin (DHb) | 92% | 86% | 0.87 |
| Blood vessel radius (BVR) | 77% | 79% | 0.79 |
| Combine DHb and BVR | 92% | 71% | 0.88 |
After excluding patients with chronic pancreatitis (n=2) from the control group, DHB measurements alone retained a sensitivity of 92% and specificity of 67% to differentiate PAC from controls, and DHB in combination with BVR measurements retained a sensitivity of 92% and specificity of 50%.
Effects of Confounding Factors
We observed no statistically significant effect of age, race, and history of smoking tobacco or alcohol use on the EIBS measurements between the 2 groups (see Table 4). DHb measurements were found to be higher in males than in females (p=0.02), whereas gender did not show a statistically significant effect on BVR measurements (p=0.22).
Table 4.
Effect of various confounding variables on measurements obtained in patients with pancreas cancer compared with controls.
| Confounding Variables | DHb (p-value*) | BVR (p-value*) |
|---|---|---|
| Age | 0.40 | 0.65 |
| Gender | 0.02 | 0.46 |
| White ethnicity | 0.23 | 0.19 |
| History of active smoking | 0.82 | 0.27 |
| History of active alcohol use | 0.94 | 0.93 |
Statistical analysis was performed using ANCOVA test
In terms of tumor characteristics, the average size of the PAC tumor was 3.3 cm (± 1.0 cm standard deviation). The distribution of the PAC tumors was 43% (n=6) in head/neck region and 57% (n=8) in the body/tail region of the pancreas. Tumor size and tumor location in the pancreas did not show any statistically significant effect on the EIBS measurements.
Adverse Events
None of the patients who participated in our study reported any adverse events during or after the endoscopy procedure with this spectroscopic measurement.
Discussion
Our study results suggest that patients with PAC have field effect changes in the normal-appearing duodenal mucosa that is distant from the site of the tumor. PGS, a novel optical biomarker, is able to distinguish PAC from non-cancer controls by detection of EIBS changes in the microvasculature of the adjacent normal-appearing duodenum. We found that measurement of EIBS markers, DHb, and BVR in the duodenal mucosa by PGS fiber-optic probe was simple, feasible, and safe during the endoscopy. In this early study, DHb alone or in combination with BVR can differentiate patients with PAC from controls with high sensitivity and specificity.
Development and progression of neoplastic lesions are known to be associated with EIBS (primarily through angiogenesis) and high oxygen demand.16, 18 Field effect theory proposes that neoplastic changes, such as epigenetic alteration and angiogenesis,can be present in the contiguous normal tissues. Detecting such changes can reveal the presence of a neoplastic lesion from a distance.16, 18 The use of field effect biomarkers have already been reported in colon cancer12, stomach cancer19, lung cancer20, and breast cancer.21
Recently, the development of modern molecular technologies such as 4-dimensional elastic light-scattering spectroscopy (4D-ELF), and low-coherence enhanced backscattering spectroscopy (LEBS) have been replicated in a miniature form called polarization gated spectroscopy (PGS) to explore the molecular abnormalities in tissue micro-architecture that appears histologically normal.11 PGS allows quantitative assessment of the absorption spectra and light scattering in real time based on how the light interacts with the tissue.14, 22–25
According to theory, field effect changes may be present at several locations surrounding the pancreas, but the measurement of optical and biochemical biomarkers from the duodenum is appealing due to its accessibility and anatomic continuation with the pancreatic tissue and the potential for fewer adverse events related to pancreatic duct manipulation. PGS via a fiber-optic probe is a relatively convenient device applied via the accessory channel of the standard upper endoscope. The preparation and measurement processes add only a few minutes to endoscopy procedure time. PGS technology uses the subset of normal white light (specific wavelength between 450 nm – 650 nm), which does not impose any risk of radiation injury. The spectroscopic measurement involves bringing the fiber-optic probe in contact with the tissue surface and therefore, adds very limited risks to the planned endoscopy procedure. Contrary to other advanced imaging techniques (e.g. confocal microscopy, magnification narrow band imaging), PGS technology can objectively quantify field effect changes without the need of complex training in image interpretation or, in the case of confocal endomicroscopy, a contrast agent.26
A prior study of spectroscopy measurements on an ex vivo duodenal biopsy sample showed a 95% sensitivity and 91% specificity in detecting the presence of pancreatic cancer27, which is comparable to our study results of spectroscopic measurements obtained in vivo. Common confounding factors, such as age, race, history of smoking tobacco, history of alcohol use, tumor size, and tumor location, did not show significant impact on spectroscopic measurements in either group.
Limitations
The number of the patients in the cancer and control groups is relatively small to derive to any strong conclusion. However, having approximately 15 patients in each group is a reasonable number of patients for a pilot study designed to understand the feasibility, safety, and rough estimation of the efficacy of this technology to determine the presence of PAC. We performed a leave-one-out cross-validation statistical analysis, which showed comparable accuracy of the results.
The patient demographics showing higher older male patients in the cancer (PAC) group are suggestive of a potential referral bias and a possible representation of similar demographics of the PAC patients in the general population. However, other confounding variables did not show statistically significant effects on the spectroscopic measurements. Similar to prior reports, DHb measurements were also found to be higher in male patients compared with female patients.28 The cause of gender differences in DHb measurement is currently unknown.
Spectroscopy technology has limitations in differentiating EIBS changes from the neoplastic lesion versus tissue inflammation. After excluding patients with visible inflammation in the gastro-duodenal region, the results achieved high accuracy to differentiate the presence of PAC from controls. So far, the effect of chronic pancreatitis on the spectroscopic measurements is not well understood. To increase the validity of the study results, a small number of patients with chronic pancreatitis were included in the control group. However, the study sample size is not large enough to accurately determine the true effect of chronic pancreatitis on the spectroscopic measurements. There was some theoretical limitation of the front viewing endoscope within the duodenum necessitating a tangential view to obtain spectral measurement of the periampullary tissues. However, the standard upper endoscope, instead of the side-viewing ERCP scope, was chosen to increase its potential applicability and reduce the risk of damaging the delicate fiber-optic probe by the elevator. In our experience, we have found that the fiber-optic probe is quite sturdy with no significant issue maneuvering it through the endoscope. Although the intensity of pressure applied to the tissue with the probe is subjective, the equipment provides immediate feedback if there is poor contact, prompting a repeat measurement. It does not however compensate for excessive contact nor artifacts from luminal contents (bile) or mucosal abrasion. Fortunately, these were rare and adjusted for by moving the probe to a nearby location.
Future implications
This PGS technology has been shown to be a safe, feasible, and accurate option to differentiate the presence of PAC during endoscopy. However, its validity and the effect of potential confounding variables have yet to be tested in a large number of patients with PAC.
Supplementary Material
Take-home Message
The detection of the field effect changes such as early increase in blood supply surrounding the malignant lesion can potentially be useful to detect a malignant lesion from a distance. We can detect variables of early increase in the blood supply with a simple through-the-endoscope fiber-optic probe and can predict the presence of pancreas adenocarcinoma with high accuracy.
Such technology may bring additional tools to stratify the malignant potential of the pancreatic lesion.
Grant Acknowledgement
NIH grants R01CA128641, R01 CA156186, R01 EB003682, U01 CA111257, and NSF grant CBET-1240416
Grant Support: Mayo Clinic CR grant
ClinicalTrials.gov number NCT01015820.
Abbreviations
- 4DELF
four-dimensional elastic light-scattering spectroscopy
- AUC
area under the curve
- ANCOVA
analysis of covariance
- BVR
mean blood vessel radius
- DHb
deoxyhemoglobin concentration
- EGD
esophagogastroduodenoscopy
- EIBS
early increase in blood supply
- ERCP
Endoscopic retrograde cholangiopancreatography
- EUS
endoscopic ultrasound
- FNA
fine needle aspiration
- LEBS
low-coherence enhanced backscattering spectroscopy
- OHb
oxygenated hemoglobin
- PAC
pancreas adenocarcinoma
- PGS
Polarization Gating Spectroscopy
- ROC
Receiver operating characteristic
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
Institutions were work was performed: 1. Mayo Clinic, 4500 San Pablo Road, Jacksonville, FL 32224
2. Northwestern University, 633 Clark Street, Evanston, IL 60208
3. Boston University Medical Center, 650 Albany Street, Suite 526, Boston MA 02118
References
- 1.American Cancer Society . Cancer Facts & Figures 2012. American Cancer Society; Atlanta: 2012. [Accessed 06/01/2013]. http://www.cancer.org/acs/groups/content/@epidemiologysurveilance/documents/document/acspc-031941.pdf. [Google Scholar]
- 2.American Cancer Society Cancer Facts & Figures 2011. [Accessed 06/01/2013];SEER database. http://seer.cancer.gov.
- 3.Ariyama J, Suyama M, Satoh K, et al. Imaging of small pancreatic ductal adenocarcinoma. Pancreas. 1998;16:396–401. doi: 10.1097/00006676-199804000-00030. [DOI] [PubMed] [Google Scholar]
- 4.Eguia V, Gonda TA, Saif MW. Early detection of pancreatic cancer. JOP : Journal of the pancreas. 2012;13:131–4. [PubMed] [Google Scholar]
- 5.American Cancer Society . Cancer Facts & Figures 2013. American Cancer Society; Atlanta: 2013. [Accessed 06/01/2013]. www.cancer.org. [Google Scholar]
- 6. [Accessed 06/01/2013];Surveillance, Epidemiology, and End Results (SEER) Program ( www.seer.cancer.gov) SEER*Stat Database: Incidence - SEER 9 Regs Research Data, Nov 2011 Sub (1973–2009)
- 7.Bartosch-Harlid A, Andersson R. Diabetes mellitus in pancreatic cancer and the need for diagnosis of asymptomatic disease. Pancreatology : official journal of the International Association of Pancreatology. 2010;10:423–8. doi: 10.1159/000264676. [DOI] [PubMed] [Google Scholar]
- 8.Verna EC, Hwang C, Stevens PD, et al. Pancreatic cancer screening in a prospective cohort of high-risk patients: a comprehensive strategy of imaging and genetics. Clinical cancer research : an official journal of the American Association for Cancer Research. 2010;16:5028–37. doi: 10.1158/1078-0432.CCR-09-3209. [DOI] [PubMed] [Google Scholar]
- 9.Testoni PA, Mangiavillano B. Optical coherence tomography in detection of dysplasia and cancer of the gastrointestinal tract and bilio-pancreatic ductal system. World journal of gastroenterology : WJG. 2008;14:6444–52. doi: 10.3748/wjg.14.6444. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Slaughter DP, Southwick HW, Smejkal W. Field cancerization in oral stratified squamous epithelium; clinical implications of multicentric origin. Cancer. 1953;6:963–8. doi: 10.1002/1097-0142(195309)6:5<963::aid-cncr2820060515>3.0.co;2-q. [DOI] [PubMed] [Google Scholar]
- 11.Chai H, Brown RE. Field effect in cancer-an update. Annals of Clinical and Laboratory Science. 2009;39:331–7. [PubMed] [Google Scholar]
- 12.Shen L, Kondo Y, Rosner GL, et al. MGMT promoter methylation and field defect in sporadic colorectal cancer. Journal of the National Cancer Institute. 2005;97:1330–8. doi: 10.1093/jnci/dji275. [DOI] [PubMed] [Google Scholar]
- 13.Wali RK, Roy HK, Kim YL, et al. Increased microvascular blood content is an early event in colon carcinogenesis. Gut. 2005;54:654–60. doi: 10.1136/gut.2004.056010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Siegel MP, Kim YL, Roy HK, et al. Assessment of blood supply in superficial tissue by polarization-gated elastic light-scattering spectroscopy. Applied Optics. 2006;45:335–42. doi: 10.1364/ao.45.000335. [DOI] [PubMed] [Google Scholar]
- 15.Fang C, Brokl D, Brand RE, et al. Depth-selective fiber-optic probe for characterization of superficial tissue at a constant physical depth. Biomedical Optics Express. 2011;2:838–49. doi: 10.1364/BOE.2.000838. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Turzhitsky VM, Gomes AJ, Kim YL, et al. Measuring mucosal blood supply in vivo with a polarization-gating probe. Applied Optics. 2008;47:6046–57. doi: 10.1364/ao.47.006046. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Roy HK, Gomes AJ, Ruderman S, et al. Optical measurement of rectal microvasculature as an adjunct to flexible sigmoidosocopy: gender-specific implications. Cancer Prev Res (Phila) 2010;3:844–51. doi: 10.1158/1940-6207.CAPR-09-0254. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Yamato I, Sho M, Shimada K, et al. PCA-1/ALKBH3 contributes to pancreatic cancer by supporting apoptotic resistance and angiogenesis. Cancer Research. 2012;72:4829–39. doi: 10.1158/0008-5472.CAN-12-0328. [DOI] [PubMed] [Google Scholar]
- 19.Endoh M, Tamura G, Honda T, et al. RASSF2, a potential tumour suppressor, is silenced by CpG island hypermethylation in gastric cancer. British Journal of Cancer. 2005;93:1395–9. doi: 10.1038/sj.bjc.6602854. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Franklin WA, Gazdar AF, Haney J, et al. Widely dispersed p53 mutation in respiratory epithelium. A novel mechanism for field carcinogenesis. The Journal of Clinical Investigation. 1997;100:2133–7. doi: 10.1172/JCI119748. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Yan PS, Venkataramu C, Ibrahim A, et al. Mapping geographic zones of cancer risk with epigenetic biomarkers in normal breast tissue. Clinical cancer research : an official journal of the American Association for Cancer Research. 2006;12:6626–36. doi: 10.1158/1078-0432.CCR-06-0467. [DOI] [PubMed] [Google Scholar]
- 22.Roy HK, Liu Y, Wali RK, et al. Four-dimensional elastic light-scattering fingerprints as preneoplastic markers in the rat model of colon carcinogenesis. Gastroenterology. 2004;126:1071–81. doi: 10.1053/j.gastro.2004.01.009. discussion 948. [DOI] [PubMed] [Google Scholar]
- 23.Roy HK, Kim YL, Wali RK, et al. Spectral markers in preneoplastic intestinal mucosa: an accurate predictor of tumor risk in the MIN mouse. Cancer epidemiology, biomarkers & prevention : a publication of the American Association for Cancer Research, cosponsored by the American Society of Preventive Oncology. 2005;14:1639–45. doi: 10.1158/1055-9965.EPI-04-0837. [DOI] [PubMed] [Google Scholar]
- 24.Roy HK, Kim YL, Liu Y, et al. Risk stratification of colon carcinogenesis through enhanced backscattering spectroscopy analysis of the uninvolved colonic mucosa. Clinical cancer research : an official journal of the American Association for Cancer Research. 2006;12:961–8. doi: 10.1158/1078-0432.CCR-05-1605. [DOI] [PubMed] [Google Scholar]
- 25.Roy HK, Backman V. Spectroscopic applications in gastrointestinal endoscopy. Clinical gastroenterology and hepatology : the official clinical practice journal of the American Gastroenterological Association. 2012;10:1335–41. doi: 10.1016/j.cgh.2012.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Kiesslich R, Goetz M, Hoffman A, et al. New imaging techniques and opportunities in endoscopy. Nature reviews. Gastroenterology & Hepatology. 2011;8:547–53. doi: 10.1038/nrgastro.2011.152. [DOI] [PubMed] [Google Scholar]
- 27.Liu Y, Brand RE, Turzhitsky V, et al. Optical markers in duodenal mucosa predict the presence of pancreatic cancer. Clinical cancer research : an official journal of the American Association for Cancer Research. 2007;13:4392–9. doi: 10.1158/1078-0432.CCR-06-1648. [DOI] [PubMed] [Google Scholar]
- 28.Roy HK, Gomes A, Turzhitsky V, et al. Spectroscopic microvascular blood detection from the endoscopically normal colonic mucosa: biomarker for neoplasia risk. Gastroenterology. 2008;135:1069–78. doi: 10.1053/j.gastro.2008.06.046. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Rogers JD, Capoglu IR, Backman V. Nonscalar elastic light scattering from continuous random media in the Born approximation. Opt. Lett. 2009;34:1891–3. doi: 10.1364/ol.34.001891. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Reif R, Amorosino MS, Calabro KW, et al. Analysis of changes in reflectance measurements on biological tissues subjected to different probe pressures. J. Biomed. Opt. 2008;13:010502. doi: 10.1117/1.2870115. [DOI] [PubMed] [Google Scholar]
- 31.van Veen RL, Verkruysse W, Sterenborg HJ. Diffuse-reflectance spectroscopy from 500 to 1060 nm by correction for inhomogeneously distributed absorbers. Opt. Lett. 2002;27:246–8. doi: 10.1364/ol.27.000246. [DOI] [PubMed] [Google Scholar]
- 32.Amelink A, Sterenborg HJCM, Bard MPL, et al. In vivo measurement of the local optical propertiesof tissue by use of differential path-length spectroscopy. Opt. Lett. 2004;29:1087–1089. doi: 10.1364/ol.29.001087. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.