. 2014 Dec 18;7(6):930–934. doi: 10.3980/j.issn.2222-3959.2014.06.03

Reliability of Tonolab measurements in rats

Li-Fang Liu 1, Chu-Kai Huang 1, Ming-Zhi Zhang 1
PMCID: PMC4270983  PMID: 25540741

Abstract

AIM

To assess the repeatability and reproducibility of Tonolab tonometer in rats with high intraocular pressure (IOP) and evaluate its ability to detect IOP changes in rats with general anaesthesia.

METHODS

Left eyes of adult Fischer rats (F344) were photocoagulated by 532 nm diode laser to induce high IOP. Hypertensive eyes of 30 conscious rats were randomly chosen to measure IOP on a single occasion. Two observers independently and alternately undertook IOP measurements consecutively for three times using the same Tonolab tonometer blind to the other observer's IOP measurements. The within subject standard deviation (Sw), coefficient of variation (CVw) (100×Sw/overall mean), and intraclass correlation coefficient (ICC) were calculated to evaluate intra-observer repeatability. Inter-observer difference was analysed by using 95% limits of agreement described by Bland-Altman and paired sample t-test. Also, another 13 normal F344 rats were intraperitoneally administrated with ketamine/xylazine or chloral hydrate, and IOPs of both eyes were measured by a single operator once every 5min until animals came to conscious. IOPs at various time points were compared by using one-way ANOVAs.

RESULTS

Mean IOP was 35.58 mm Hg (range 17.33 to 65.33 mm Hg). For intraobserver repeatability, the Sw, CVw and ICC of high IOP for two observers were 5.20 mm Hg/3.41 mm Hg, 9.98%/8.08% and 0.820/0.928 respectively. The inter-observer difference was 14.76%±19.76% of the mean IOP of two observers, with a 95% limits of agreement -23.97% to 53.50%, and the difference between mean IOP of these two observers was statistically significant (P=0.001). IOPs dropped slightly during the first 15min post-aneathesia, with a IOP change between 0.17 and 1.17 mm Hg. IOPs changed from basline of 11.75±2.05 mm Hg (n=12) to 8.75±1.06 mm Hg 20min post-anesthesia (P=0.001), and this hypotensive condition persisted until 80min post-anesthesia.

CONCLUSION

In this sample of hypertensive rats, Tonolab measurements demonstrated high levels of intraobserver repeatability, however, its interobserver reproducibility was poor. Longitudinal changes of IOP caused by genral anaesthesia can be sensitively detected by Tonolab. So we suggested that measurements of IOP using Tonolab are best measured by a single observer, and it could be included in experimental glaucoma.

Keywords: Tonolab, repeatability, reproducibility, rats

INTRODUCTION

Rodent models for mimicking glaucoma and evaluating the intraocular pressure (IOP)-lowering mechanism of drugs has been widely used[1]. To ascertain the association between IOP and optic nerve damage, accurate and reproducible IOP measurements are indispensable. However, it is difficult to measure IOP in rodents because of their small eyes. Direct measurements with microneedle cannulation into anterior chamber can record IOP directly and accurately[2], but this procedure is invasive and needs to be performed under general anaesthesia, so it cannot be repeatedly used during a short period of time, furthermore, it involves corneal perforation and carries a risk of infection. To overcome these disadvantages, some noninvasive methods of measuring IOP in rodents have been reported[3], [4]. TonoPen XL electronic tonometer is the most popular one[5]. Although there are many reports of IOP measurements using TonoPen in rodents, it was originally designed for use in humans, suggesting it may be less accurate in small animals[5][8].

Since year 2001, the induction-impact rebound tonometer, the first tonometer designed for use in rodents, has become available[9]. This is a instrument that is suitable for repeated use, and its lightweight probe makes it possible to measure IOP without anaesthesia. Though the ability of this noninvasive tonometer to reflect true IOP has been reported in several studies, its repeatability, reproducibility and sensitivity to measure IOP in rodents has not frequently been reported, so in this study, we measure IOP in ocular hypertensive and general anaesthetic rats to determine its accuracy and ability to detect changes of IOP[2],[10][14].

MATERIALS AND METHODS

Animals

Adult Fischer344 (F344) rats (12-16 weeks old) were used in this study (Vital River, Beijing, China). Animals were adapted to an alternating 12h periods of light and dark light schedule (12L:12D, light on at 06:00 a.m.), with free access to food and water. All studies were in compliance with the Association for Research in Vision and Ophthalmology (ARVO) Resolution on the Use of Animals in Research.

Study of Repeatability and Reproducibility

To produce increased IOP we used a diode laser (Vitra, Quantel Medical, France) at 532 nm. Animals were intraperitoneally anaesthetized with a 1:1 mixture (1.5 mL/kg) of ketamine (100 mg/mL), xylazine (20 mg/mL) and topical proparacaine 1% eye drops (Alcaine, Alcon Laboratories, Belgium). Laser treatment was administered unilaterally by 55-70 applications directed towards trabecular meshwork. Laser beam was applied with a power of 440 mW for 0.7s, producing a spot size of 100 µm[15]. One week later, the same procedure were repeated if IOP difference between the treated eye and the fellow eye was less than 8 mm Hg[16]. Topical 5% erythromycin was applied at the end of each procedure.

IOP was measured with TonoLab (Tiolat, Helsinki, Finland) under the manufacturer's instructions in conscious rats. A new disposable probe was used for each animal. Choose the “rat” measurement model. Keep the probe horizontal, start probe-cornea distance about 3-5 mm and angel of 25° limit relative to the visual axis at cornea apex[13]. Only measurements that are judged by data analytical system to be within its parameters are marked as valid: if the measurement is successful, there is a short beep, otherwise, the tonometer will beep twice and show an error message. After effectively measuring for six times, average is calculated and displayed automatically. This machine-generated average is used as IOP for each eye in this study.

Hypertensive eyes of 30 rats (mean IOP 35.58 mm Hg, range 17.33 to 65.33 mm Hg) were randomly chosen to measure IOP with the same Tonolab by two observers in a randomised order, each observer measured for three groups of IOP consecutively. All measurements were completed within 5min, and results were not revealed until all data had been collected. Measurements from a single observer were used to evaluate intra-observer repeatability, and the mean of these three readings was used to assess inter-observer reproducibility.

Anaesthetic Study

Baseline IOPs of 13 normal F344 rats were measured immediately before induction of anesthesia. Then, rats were intraperitoneally administrated with 1:1 mixture (1.5 mL/kg) of ketamine (100 mg/mL) and xylazine (20 mg/mL, n=6) or chloral hydrate (3 mL/kg, n=7). The point when animals reached to a deep anesthesia (determined by absence of pain and tail-pinch reflex) was definded as “time 0”, and post-anesthesia measurements were obtained at 5min intervals up to 80min or 60min when animals came to conscious[17]. IOPs in awake and anesthetized animals were measured in both eyes by the same observer. Care was taken to avoid restraint or eyelid manipulation during all measurements.

Statistical Analysis

Statistical analyses were performed using SPSS version 16.0 (SPSS Inc., Chicago, Illinois, USA) and Excel (Excel 2003, Microsoft, Redmond, WA, USA).

According to formula: n=1.962/[2×0.22×(m-1)], where n represents the number of subjects and m is the number of observers[18], it was estimated that a minimal sample size would be twenty-five. Within subject standard deviation (Sw) was calculated as the square root of the within-subject mean square of error[19]. Statistical comparison of intraclass correlation coefficient (ICC) was performed by Wilcoxon signed ranks test[20], and ICC values larger than 0.75 may be regarded as acceptable[21].The coefficient of variation (CVw) were also computed, and a CV<10% was indicative of good repeatability[20]. Bland–Altman method shows the mean difference and 95% limit of agreement between two sets of measurement, and we used this method to evaluate reproducibility of IOP measurements between two observers[22]. The mean IOP of these two observers was also compared by paired sample t-tests.

In anaesthetic study, average IOPs of each measurement time point in both eyes were determined, and statistical analyses were performed by using one-way ANOVAs followed by Bonferroni's multiple comparison. P<0.05 was considered statistically significant.

RESULTS

The mean of high IOP was 35.58±12.00 mm Hg (range 17.33 to 65.33 mm Hg). Intraobserver repeatability was high, with CVw<10%, Sw between 3.41 and 5.20 mm Hg, ICC between 0.81 and 0.93 (Table 1).

Table 1. The overall mean (x±s), Sw, ICC and CVw of measurements in ocular hypertensive rats.

Observer Mean (mm Hg) Sw (95% CI) ICC (95% CI) CVw
Observer 1 37.80±3.84 5.20 (2.51-6.91) 0.81 (0.69-0.90) 9.98
Observer 2 33.30±2.61 3.41 (1.74-4.50) 0.93 (0.88-0.96) 8.08

95% CI: 95% confidence interval.

(n=30)

Interobserver reproducibility, however, was low. Mean±SD was 4.43±6.49 mm Hg (95% limit of agreement: -8.28 to 17.15 mm Hg). If transformed the bias into percentage, it was 14.76%±19.76% of the mean IOP of two observers, with a 95% limit of agreement -23.97% to 53.50% (Figure 1). Paired sample t-test showed that the difference between mean IOP of these two observers was statistically significant (P=0.001).

Figure 1. Bland–Altman plot of data from the interobserver reproducibility study.

Figure 1

Mean IOP for two observers ( mm Hg) was plotted against the difference in IOP of these two observers (%). Solid line represents the mean bias and dotted lines represent the 95% limits of agreement. All but two of the data points (n=30) lie at or within 1.96 SDs of the mean.

Awake IOPs in 26 eyes of normal F344 rat were 13.28±2.56 mm Hg(95%CI: 12.36-14.20 mm Hg, n=13), there was not statistically difference between IOPs of left and right eyes (P=0.916). For anaesthetics of chloral hydrate (n=7), the difference between postanesthetic IOPs (from time point 10min to 1h) and baseline IOPs was not statistically significant in both eyes (P=0.072, P=0.676; Figure 2). IOPs between left and right eyes was also not statistically different except for 35min postanesthesia (P=0.025). For anaesthetics of ketamine/xylazine, IOP dropped slightly but not statistically during the first 15min post-aneasthesia, with a change between 0.17 and 1.17 mm Hg. However, from 20 to 80min, IOP reduced to a point (7 to 8 mm Hg) that was statistically different from the baseline IOP (P=0.000; Figure 3).

Figure 2. IOP danymics after chloral hydrate anaesthesia in 14 normal rat eyes.

Figure 2

Figure 3. IOP danymics after ketamine/xylazine anaesthesia in 12 normal rat eyes.

Figure 3

DISCUSSION

The TonoLab tonometer is designed by the induction-impact mechanism, it pushes a magnetic probe onto the cornea and detects its return-bounce motion with a sensing coil. The motion parameters of the probe vary according to eye pressure and can be used to automatically transform into IOP. Potential advantages of this tonometer include easy portability, longitudinal measurements without anesthesia, and adaptability to small eyes of rodents.

Despite advantages of Tonolab described above, it is necessary to assess its accuracy and sensitivity before usefulness of this tonometer can be confirmed. Repeatability is the variability of the measurements obtained by one person while measuring the same item repeatedly, it represents inherent precision of the equipment. While reproducibility is the variability of the measurement caused by differences in observers' behaviors. So inaccuracy can arise from instruments or systematic differences between observers.

In this study, individual observer appeared to be internally reliable when obtaining repeated measurements (Table 1), indicating that intraobserver repeatability was high. However, when IOPs of two observers were compared, differences in measurement appeared. The average measurement difference (4.43±6.49 mm Hg) was moderate, with a 95% limits of agreement -8.28 to 17.15 mm Hg, if transformed into percentage, this means a 14.76%±19.76% bias, with a 95% limit of agreement -23.97% to 53.50%. This large discrepancy could not be accepted in experimental glaucoma. It was also inconsistent with data reported by Lee et al[2] and Goldblum et al[24] (Table 2). Lee et al [2] evaluated the accuracy of the TonoLab in eight cannulated rat eyes by manometrically changing intraocular pressure between 20 to 100 mm Hg, and IOPs were measured with a rebound tonometer three times by a single observer. When compared data recorded by Tonolab with the cannulation “golden standard” method by using Bland and Altman method, they got a mean bias of 7.41%±7.87%, which was only 50% of ours. Likewise, Goldblum et al[24] found that the mean difference between induction-impact and cannulation method was -3.85±3.01 mm Hg, and mean bias was ±4.6% of mean (range ±1.9% to ±7.3%), which was also much smaller than data in this study. Although we did not include the cannulation method as a comparison in this study because of technical reasons, we can still get our conclusion that the interobserver reproducibility in our study was rather poor. This suggested that although one observer may consistently use the same points for measurement in repeated examinations, a different observer may choose different points when performing a particular measurement.

Table 2. Papers about rebound tonometer in rodents searched from PubMed.

Reference Subjects Accuracy
Morrison et al[13] Hypertensive Brown Norway rats Cannulated eyes: R2=0.99
Ohashi et al [5] Sprague Dawley rats R2=0.985; circadian change: 2.1±1.1 mm Hg; latanoprost induced change: 3.0±2.1 mm Hg
Lee et al [2] Cannulated rat eyes R2=0.963; mean bias 7.41%±7.87%; limits of agreement -8.02 to 22.84 mm Hg
Pease et al [14] Normal rats; normal mice; hypertensive rats Normal eyes within 1 mm Hg; glaucomatous eyes: R2=0.98; normal mouse eyes: R2=0.98
Wang et al[23] Cannulated Wistar rats; different strains of anaesthetic mice R2>0.94; IOP changes: 5.1 to 7.2 mm Hg
Goldblum et al[24] Wistar rats R2=0.6603; mean difference: –3.85±3.01 mm Hg;sStandard deviations: ±4.6% of mean (range: ±1.9% to ±7.3%)
Kontiola et al[9] Wistar rats ex vivo R2=0.95; in vivo R2=0.67

R2: Square of the coefficient of multiple correlation, it ranges from 0 to 1, with 0 denoting that model does not explain any variation and 1 denoting that it perfectly explains the observed variation.

To further prove the accuracy of Tonolab, we assessed its ability to detect IOP changes caused by ketamine/xylazine or chloral hydrate anaesthesia in normal rats. Our study demonstrated that during the first 15min of ketamine/xylazine anaesthesia, Tonolab could detect a minimal IOP change between 0.17 and 1.17 mm Hg (Table 3). If we analysed all data during the 80min study period (data not shown), the range of IOP change was 0.34 to 4.83 mm Hg. As for chloral hydrate anaesthesia, IOP change was smaller, with range from 0.14 to 2.71 mm Hg (Table 4). This means Tonolab was sensitive enough to detect a minimal change of IOP, which was much smaller than previous reported ones (Table 2). Ohashi et al[5] evaluated the accuracy of Tonolab by studying nocturnal and Latanoprost induced IOP changes, and found a 2 to 3 mm Hg dynamics. Another study by Wang et al[23] also reported an anesthesia induced IOP change of 5.1 to 7.2 mm Hg. So, we consider Tonolab could be used in experimental research due to its ability to detect minimal IOP changes. Furthermore, our study also demonstrated that IOP measured under anesthesia may not reflect true IOP. According to our data, IOP dropped significantly after 15min of ketamine/xylazine anaesthesia. This may due to the mechanism of xylazine, which is a α2-adrenergic receptor agonist, so by directly depressing cardiac function and lowering heart rate, it also has a indirect effect on IOP[25]. As for chloral hydrate anesthesia, IOP did not change significantly, but fluctuation existed. During the first 10min, IOP increased by 2-3 mm Hg, then it tended to decline to the baseline level. This can be explained by the pharmacological properties of chloral hydrate, which plays its role via enhancing the GABA receptor complex and has similar action to barbiturates, so the effect to IOP may be very small[26]. To confirm our presumption, we find evidence from other experiments, and result showed that chloral hydrate did not alter IOP levels in both normal and glaucomatous eyes. From all above, we suggest IOP measurement either under ketamine/xylazine or chloral hydrate anaesthesia, the time window should be chosen when IOP is stabilized[27].

Table 3. IOP history during the first 20min of Ketamine/Xylazine anaesthesia.

Laterality Conscious After anaesthesia (min)
5 10 15 20
OD 11.17±1.60 10.83±2.14 9.83±1.47 9.33±0.52 8.33±1.21
OS 12.33±2.42 11.17±1.47 10.00±1.90 9.83±1.33 9.17±0.75

Table 4. IOP history during 60min of chloral hydrate anaesthesia.

Laterality Conscious After anaesthesia (min)
10 20 30 40 50 60
OD 14.14±2.48 16.43±3.21 15.00±3.70 14.57±2.52 14.29±1.80 15.29±2.87 14.00±3.21
OS 13.29±2.81 16.00±2.16 14.86±4.10 15.14±3.89 12.14±2.41 13.14±3.13 12.71±2.69

In summary, based on our studies, we suggest that before and after measurements of IOP are best measured at the present time by a single observer, and Tonolab could be included in experimental glaucoma study.

Acknowledgments

Foundation: Supported by National Natural Science Foundation of China (No.30672282).

Conflicts of Interest: Liu LF, None; Huang CK, None; Zhang MZ, None.

REFERENCES

  • 1.Johnson TV, Tomarev SI. Rodent models of glaucoma. Brain Res Bull. 2010;81(2-3):349–358. doi: 10.1016/j.brainresbull.2009.04.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Lee EJ, Park KH, Kim DM, Yoo YC, Kang SH, Kim YJ. Assessing intraocular pressure by rebound tonometer in rats with an air-filled anterior chamber. Jpn J Ophthalmol. 2008;52(6):500–503. doi: 10.1007/s10384-008-0591-1. [DOI] [PubMed] [Google Scholar]
  • 3.Nissirios N, Goldblum D, Rohrer K, Mittag T, Danias J. Noninvasive determination of intraocular pressure (IOP) in nonsedated mice of 5 different inbred strains. J Glaucoma. 2007;16(1):57–61. doi: 10.1097/IJG.0b013e31802b3547. [DOI] [PubMed] [Google Scholar]
  • 4.Filippopoulos T, Matsubara A, Danias J, Huang W, Dobberfuhl A, Ren L, Mittag T, Miller JW, Grosskreutz CL. Predictability and limitations of non-invasive murine tonometry: comparison of two devices. Exp Eye Res. 2006;83(1):194–201. doi: 10.1016/j.exer.2005.12.005. [DOI] [PubMed] [Google Scholar]
  • 5.Ohashi M, Aihara M, Saeki T, Araie M. Efficacy of TonoLab in detecting physiological and pharmacological changes in rat intraocular pressure: comparison of TonoPen and microneedle manometry. Jpn J Ophthalmol. 2008;52(5):399–403. doi: 10.1007/s10384-008-0532-z. [DOI] [PubMed] [Google Scholar]
  • 6.Reitsamer HA, Kiel JW, Harrison JM, Ransom NL, McKinnon SJ. Tonopen measurement of intraocular pressure in mice. Exp Eye Res. 2004;78(4):799–804. doi: 10.1016/j.exer.2003.11.018. [DOI] [PubMed] [Google Scholar]
  • 7.Lamparter J, Hoffmann EM. Measuring intraocular pressure by different methods. Ophthalmologe. 2009;106(8):676–682. doi: 10.1007/s00347-009-1971-8. [DOI] [PubMed] [Google Scholar]
  • 8.Saeki T, Aihara M, Ohashi M, Araie M. The efficacy of TonoLab in detecting physiological and pharmacological changes of mouse intraocular pressure--comparison with TonoPen and microneedle manometery. Curr Eye Res. 2008;33(3):247–252. doi: 10.1080/02713680801919716. [DOI] [PubMed] [Google Scholar]
  • 9.Kontiola AI, Goldblum D, Mittag T, Danias J. The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp Eye Res. 2001;73(6):781–785. doi: 10.1006/exer.2001.1088. [DOI] [PubMed] [Google Scholar]
  • 10.Pease ME, Cone FE, Gelman S, Son JL, Quigley HA. Calibration of the TonoLab tonometer in mice with spontaneous or experimental glaucoma. Invest Ophthalmol Vis Sci. 2011;52(2):858–864. doi: 10.1167/iovs.10-5556. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Kim CY, Kuehn MH, Anderson MG, Kwon YH. Intraocular pressure measurement in mice: a comparison between Goldmann and rebound tonometry. Eye (Lond) 2007;21(9):1202–1209. doi: 10.1038/sj.eye.6702576. [DOI] [PubMed] [Google Scholar]
  • 12.Morris CA, Crowston JG, Lindsey JD, Danias J, Weinreb RN. Comparison of invasive and non-invasive tonometry in the mouse. Exp Eye Res. 2006;82(6):1094–1099. doi: 10.1016/j.exer.2005.12.001. [DOI] [PubMed] [Google Scholar]
  • 13.Morrison JC, Jia L, Cepurna W, Guo Y, Johnson E. Reliability and sensitivity of the TonoLab rebound tonometer in awake Brown Norway rats. Invest Ophthalmol Vis Sci. 2009;50(6):2802–2808. doi: 10.1167/iovs.08-2465. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Pease ME, Hammond JC, Quigley HA. Manometric calibration and comparison of TonoLab and TonoPen tonometers in rats with experimental glaucoma and in normal mice. J Glaucoma. 2006;15(6):512–519. doi: 10.1097/01.ijg.0000212276.57853.19. [DOI] [PubMed] [Google Scholar]
  • 15.Kwong JM, Vo N, Quan A, Nam M, Kyung H, Yu F, Piri N, Caprioli J. The dark phase intraocular pressure elevation and retinal ganglion cell degeneration in a rat model of experimental glaucoma. Exp Eye Res. 2013;112:21–28. doi: 10.1016/j.exer.2013.04.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Huang C, Cen LP, Liu L, Leaver SG, Harvey AR, Cui Q, Pang CP, Zhang M. Adeno-associated virus-mediated expression of growth-associated protein-43 aggravates retinal ganglion cell death in experimental chronic glaucomatous injury. Mol Vis. 2013;27;19:1422–1432. [PMC free article] [PubMed] [Google Scholar]
  • 17.Burnside WM, Flecknell PA, Cameron AI, Thomas AA. A comparison of medetomidine and its active enantiomer dexmedetomidine when administered with ketamine in mice. BMC Vet Res. 2013;9:48. doi: 10.1186/1746-6148-9-48. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Bland JM, Altman DG. Measurement error. BMJ. 1996;313(7059):744. doi: 10.1136/bmj.313.7059.744. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Bayhan HA, Aslan Bayhan S, Muhafiz E, Can I. Repeatability of aberrometric measurements in normal and keratoconus eyes using a new Scheimpflug-Placido topographer. J Cataract Refract Surg. 2014;40(2):269–275. doi: 10.1016/j.jcrs.2013.07.046. [DOI] [PubMed] [Google Scholar]
  • 20.Fujimura F, Kamiya K, Fujiwara K, Shoji N, Shimizu K. Repeatability and reproducibility of measurements using a NT-530P noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure. Biomed Res Int. 2013;2013:370592. doi: 10.1155/2013/370592. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Kim JH, Kim KB, Kim WC, Kim JH, Kim HY. Accuracy and precision of polyurethane dental arch models fabricated using a three-dimensional subtractive rapid prototyping method with an intraoral scanning technique. Korean J Orthod. 2014;44(2):69–76. doi: 10.4041/kjod.2014.44.2.69. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Petersen AK, Maribo T, Hjortdal VE, Laustsen S. Intertester reliability of the talk test in a cardiac rehabilitation population. J Cardiopulm Rehabil Prev. 2014;34(1):49–53. doi: 10.1097/HCR.0000000000000029. [DOI] [PubMed] [Google Scholar]
  • 23.Wang WH, Millar JC, Pang IH, Wax MB, Clark AF. Noninvasive measurement of rodent intraocular pressure with a rebound tonometer. Invest Ophthalmol Vis Sci. 2005;46(12):4617–4621. doi: 10.1167/iovs.05-0781. [DOI] [PubMed] [Google Scholar]
  • 24.Goldblum D, Kontiola AI, Mittag T, Chen B, Danias J. Non-invasive determination of intraocular pressure in the rat eye. Comparison of an electronic tonometer (TonoPen), and a rebound (impact probe) tonometer. Graefes Arch Clin Exp Ophthalmol. 2002;240(11):942–946. doi: 10.1007/s00417-002-0571-y. [DOI] [PubMed] [Google Scholar]
  • 25.Xu Q, Ming Z, Dart AM, Du XJ. Optimizing dosage of Ketamine and Xylazine in murine echocardiography. Clin Exp Pharmacol Physiol. 2007;34(5-6):499–507. doi: 10.1111/j.1440-1681.2007.04601.x. [DOI] [PubMed] [Google Scholar]
  • 26.Lu J, Greco MA. Sleep circuitry and the hypnotic mechanism of GABAA drugs. J Clin Sleep Med. 2006;15;2(2):S19–26. [PubMed] [Google Scholar]
  • 27.Jaafar MS, Kazi GA. Effect of oral chloral hydrate sedation on the intraocular pressure measurement. J Pediatr Ophthalmol Strabismus. 1993;30(6):372–376. doi: 10.3928/0191-3913-19931101-07. [DOI] [PubMed] [Google Scholar]

Articles from International Journal of Ophthalmology are provided here courtesy of Press of International Journal of Ophthalmology