Abstract
Near-infrared fluorescence (NIRF) molecular imaging holds great promise as a new “point-of-care” medical imaging modality that can potentially provide the sensitivity of nuclear medicine techniques, but without the radioactivity that can otherwise place limitations of usage. Recently, NIRF imaging devices of a variety of designs have emerged in the market and in investigational clinical studies using indocyanine green (ICG) as a non-targeting NIRF contrast agent to demark the blood and lymphatic vasculatures both non-invasively and intraoperatively. Approved in the USA since 1956 for intravenous administration, ICG has been more recently used off label in intradermal or subcutaneous administrations for fluorescence imaging of the lymphatic vasculature and lymph nodes. Herein, we summarize the devices of a variety of designs, summarize their performance in lymphatic imaging in a tabular format and comment on necessary efforts to develop standards for device performance to compare and use these emerging devices in future, NIRF molecular imaging studies.
Near-infrared fluorescence (NIRF) imaging is an emerging clinical technology that requires administration of a fluorescence-imaging agent that can be excited at near-infrared (NIR) wavelengths of ≥760 nm. Upon illuminating tissue surfaces with penetrating NIR light to excite the imaging agent within the tissues, the generated fluorescence is collected to form a two-dimensional (2D) image demarking the tissue deposition of the NIRF imaging agent. While far-red and NIR light between the wavelength ranges of 690–900 nm penetrate deeply in tissues, endogenous chromophore fluorescence when excited by light of wavelengths <780 nm, creates a high autofluorescence background for molecularly targeting exogenous imaging agents in tissues. The tissue depth to which the NIRF imaging can detect NIRF imaging agents is dependent upon their brightness and the sensitivity of the device, but has been estimated to be between 3 and 4 cm beneath the tissue surfaces in intensified devices1 and <2 cm in others.2 Three-dimensional tomographic imaging using 2D projection data as well as time-dependent and independent methods has been developed for small animal imaging3,4 but, owing to these limitations in tissue penetration, has not been translated to clinical imaging.
The exciting concept of conjugating a NIR excitable fluorophore to a small molecule, protein or antibody that targets an extracellular disease marker for diagnostic, molecular imaging has been postulated for years by several investigators.5–7 The use of NIRF imaging for molecularly guided surgical resection of cancers could dramatically reduce residual tumour burden as well as surgical morbidity associated with excising sufficient tissues to avoid having positive surgical margins. However, the tissue depth, concentration and dose at which a “first-in-humans” imaging agent can be detected in tissues depends upon several factors but, most importantly, upon the sensitivity of the imaging device. Unlike positron emission tomography, scintigraphy, single-photon emission and the γ probe used to detect radiolabelled molecular targeting agents for diagnostic imaging and intraoperative detection, fluorescence imaging devices do not have phantoms and standards to assess performance metrics, and there are no traceable standards to quantify or compare performance between fluorescence imaging devices. Different fluorescent imaging device designs, summarized for clinical devices in an excellent review by Alander et al,8 likely result in varying performance for detecting a NIRF imaging agent. As a result in the USA, the regulatory strategy for securing market approval of fluorescent imaging agents is to pair the drug approval process to a specific model of imaging device [e.g. approval of Cysview® (Photocure, Oslo, Norway) with Karl Storz blue light cystoscope (PDD system; Karl Storz GmbH and Co., Tuttlingen, Germany)]. This practice could limit the entry and adoption of molecularly targeted NIRF imaging agents into clinical practice.
To date, the only NIR-excited fluorophore used clinically is indocyanine green (ICG). Since 1956, ICG has been approved by the US Food and Drug Administration for intravenous (i.v.) administration at a concentration of 2.5 mg ml−1 with doses of up to 25 mg in adults, 12.5 mg in children and 6.25 mg in infants. ICG has been used in the clinic as a reagent for determining cardiac output, hepatic function and ophthalmic angiography. It has an excellent record of safety, and there is no demonstrable evidence of phototoxicity associated with its use.9 After administration, ICG binds tightly to plasma proteins and has a half-life of several minutes in blood circulation, which allows repeated intraoperative i.v. administration for fluorescence angiography. In the plasma, the absorption and emission peaks of ICG are shifted towards longer wavelengths, to around 807 and 822 nm,10 respectively, but still reside in the “optical window” of the tissues. Compared with other NIR-excited fluorophores, the quantum efficiency (QE) of ICG is low and reported to be 0.02 at 780 nm excitation and 830 nm emission;11 it is comparatively unstable once reconstituted in saline; and it has no functional group for conjugation to compound for molecular imaging. Using ICG as a non-specific blood vascular imaging agent, NIRF angiography has been used intraoperatively in coronary, neurosurgical and vascular surgeries8 as well for non-invasive assessment of superficial perfusion.12
Most recently, ICG has been used in off-label intradermal and subcutaneous administrations at varying doses for evaluating the lymphatic circulation to identify sentinel lymph nodes (SLNs) in surgical oncology,10,13,14 assess lymphovenous anastomoses (LVA) surgery15–17 and non-invasively map the lymphatic vasculature.18–20 Indeed, “ICG lymphography” has been found to be superior to lymphoscintigraphy for diagnostic imaging of early lymphoedema in upper extremities20 and enables early diagnosis of lymphoedema before the onset of symptoms.20–22 Yet, the doses of ICG and the design of devices used in these and other clinical studies vary widely, suggesting variable device performance. The available ICG fluorescence imaging devices on the market include photodynamic eye (PDE; Hamamatsu Photonics Co., Hamamatsu, Japan) and SPY (Novadaq Technologies Inc., Toronto, ON, Canada), with other investigational devices such as frequency-domain photon migration (FDPM) imager (University of Texas Health Science Center at Houston, TX) and mini-FLARE™ (Israel Beth Deaconess Medical Center, Boston, MA) also employed in clinical studies. Although the architecture can be dramatically different among these devices, the core components are (i) the light source for exciting ICG; (ii) optical filters for separating emitted fluorescent signals from strong backscattered excitation light and ambient light signals; and (iii) an area detector for sensing the emitted fluorescent signals. Undoubtedly, the performance of a device is ultimately determined by these core components, requiring different dosages of ICG ranging from micrograms to milligrams per injection for visualizing the lymphatics. In this review, we first compare three core components employed in various ICG fluorescence imaging systems and then review the varying ICG concentrations used in clinical imaging of the lymphatic vasculature and lymph nodes as an indicator of device performance. This review complements those of others10,13,14,23 but summarizes literature results in a tabulated and quantitative format to enable readers to compare performance. Finally, we comment on a proposal for a systematic approach to quantify and report imaging device performance using standardized measurements of certifiable phantoms that could accelerate the future translation of NIRF molecular imaging agents used with these fluorescence imaging devices.
INSTRUMENTATION
In order to understand differences in performance presented in the ICG imaging of lymph nodes for cancer staging and the Lymphatic imaging sections, the differences in imaging devices employed in these studies are first described. As described in past reviews,8,24 there are three core components common to the ICG fluorescence imaging systems, summarized in Table 1. These are the incident light sources used to excite ICG; the optics that allow for collection of ICG fluorescence and rejection of ambient and backscattered incident light; and the area detector used to register the collected light (Figure 1).
Table 1.
Summary of various indocyanine green fluorescence imaging devices from excitation source, fluorescence collection, working distance (WD) and field of view (FOV)
| Device | Excitation source |
Fluorescence collection |
WD (cm) | FOV (cm2) | |||||
|---|---|---|---|---|---|---|---|---|---|
| Source type | Excitation wavelength λe (nm) | Fluence rate (mW cm−2) | Camera type | Dynamic range (bit) | Integration time (ms) | Collection wavelength λc (nm) | |||
| FDPM imager | Laser | 785 | <1.90 | Intensified CCD | 16 | 50–800 | 825–835 | <76.2 | <900 |
| Mini-FLARE™ | LED | 760 | 7.70 | CCD | 12 | 0.1–8000 | 800–848 | 10–32 | 12 × 9 at 32 cm WD |
| SPY | Laser | 806 | 31 (narrow FOV) 4 (wide FOV) | CCD | 8 | Real-time video | Centred at 830 | 30 | 7.6 × 5.0 19.0 × 12.7 |
| Photodynamic eye | LED | 760 | 4.0 | CCD | 8 | NS | >820 | 20 | 10.0 × 6.7 |
| HyperEye Medical System | LED | 760–780 | NS | CCD | NS | NS | 800–850 | NS | NS |
| FLUOBEAM® | Laser | 750 | 5.0 | CCD | 8 | 1–1000 | >800 | 15–25 | 2.2 × 1.5 to 20 × 14 |
| IC-View | Laser | 780 | Incident power NS | CCD | NS | NS | NS | NS | NS |
| Visual Navigator | LED | 740 | Incident power NS | CCD | NS | NS | Centred at 820 | NS | NS |
| The prototype surgical navigation system | LED | 760 | Incident power NS | Electron-multiplying CCD | 16 | Real-time video | 810–870 | 60 | 7.9 × 7.9 to 12.5 × 12.5 |
| Leica FL800 | Xenon | 700–800 | NS | CCD | NS | NS | 820–860 | NS | NS |
| INFRARED™ 800™ | Xenon | 700–780 | NS | CCD | NS | NS | 820–900 | NS | NS |
| FIREFLY™ for robotic surgery (da Vinci®) | Laser | 806 | NS | CCD | NS | NS | NS | NS | NS |
| Laparoscopic near-infrared fluorescence system | Xenon | NS | NS | CCD | NS | NS | NS | NS | NS |
CCD, charge-coupled devices; LED, light-emitting diode; NS, not specified.
The devices mentioned in the table and their manufacturer details in parenthesis are as follows: FDPM imager (University of Texas Health Science Center at Houston, TX), FIREFLY for robotic surgery (da Vinci; Novadaq Technologies Inc., Toronto, ON, Canada), FLUOBEAM (Fluoptics, Grenoble, France), HyperEye (Mizuho Medical Co., Ltd, Tokyo, Japan), IC-View (Pulsion Medical Systems SE, Feldkirchen, Germany), INFRARED 800 (Zeiss™; Karl Zeiss Inc., Jena, Germany), Laparoscopic near-infared fluorescence system (Olympus®, Tokyo, Japan), Leica FL800 (Leica Microsystems Inc., Buffalo Grove, IL) Mini-FLARE (Israel Beth Deaconess Medical Center, Boston, MA), Photodynamic eye (PDE; Hamamatsu Photonics Co., Hamamatsu, Japan), SPY (Novadaq Technologies Inc.), The prototype surgical navigation system (Chinese Academy of Sciences, Beijing, China) and Visual Navigator (SH System, Seoul, Republic of Korea).
Figure 1.
Schematic of indocyanine green (ICG)-based near-infrared fluorescence imaging system consisting of the incident light source, collection optics and area detector. (a) Represents the spectra of 785-nm laser diode, 780-nm light-emitting diode (LED) and xenon lamp, showing the laser diode with narrow bandwidth, the LED with relatively wide bandwidth and xenon with very broad bandwidth, and (b) the focus lens and emission filter. (c) A plot of signal-to-noise ratio (SNR) vs contrast for different charge-coupled device (CCD)-based detectors with integration time of 200 ms. Both front-illuminated intensified CCD (FICCD) and back-illuminated intensified CCD (BICCD) are superior to the other types of CCD cameras (Zhu et al25). BCCD, back-illuminated CCD; EMCCD, electron-multiplying CCD; FCCD, front-illuminated CCD; OD, optical density.
Incident light sources for collection of indocyanine green fluorescence
The commonly used excitation light sources in ICG fluorescence imaging systems in order of increasing spectral bandwidth are (i) laser diodes; (ii) light-emitting diodes (LEDs); and (iii) filtered lamp sources, which have typical spectra illustrated in Figure 1a. Because rejection of backscattered excitation light is performed spectrally through the use of interference filters, laser diodes enable the greatest sensitivity, since the “background” arising from “leakage” of backscattered excitation light is the lowest. By contrast, LEDs generate a broader band of wavelengths with relatively lower power output, requiring tens of LEDs integrated together for milliwatts per square centimetre of incident light. For the filtered lamp sources, the lamp sources are filtered to generate excitation light with a narrow band, but the generated excessive heat needs to be dissipated to extend the lifetime of the filter. In addition, the filtered lamp sources have low efficiencies, making it difficult to couple into an optical fibre. Hence, laser diodes and LEDs are widely adopted in the ICG fluorescence imaging systems used clinically. One might expect that the greater amount of incident excitation light (measured as a “fluence” in milliwatts per square centimetre) would result in more ICG fluorescence and a larger amount of collected fluorescence signal. However, there are maximum permissible exposure (MPE) limits for both eye safety as well as skin safety. The American National Standards Institute's (#13997) MPE for eye exposure of a 700–1050 nm laser beam is 102(λ−0.700) × 10−3 W cm−2 when the duration is between 10 and 30,000 s, and the MPE for skin exposure is 0.2 × 102(λ−0.700) W cm−2 (where the wavelength λ is reported in units of micrometre). For 785 nm, the limit for eye safety is 1.48 mW cm−2 and for skin safety is 296 mW cm−2. As shown in Table 1, all laser-based devices used in clinical studies probably exceed MPE for eye safety but not for skin safety. It should be noted that devices that exceed the MPE require the use of eye protection for the patient, and precautions for the healthcare personnel. In addition, with greater excitation incident power, there is greater backscattered light, which, as discussed in the next section, can contribute to obtain a high background, obscure collection of weak ICG signals and limit device performance.
Collection optics
Without exception, all ICG fluorescence imaging systems have detection sensitivities that are limited by background signals (or noise floor) that arise from one or more of the following: (i) the spectral overlap between the backscattered excitation light and collected ICG fluorescence allowing non-fluorescent signals to be registered as fluorescence; (ii) a “blue shifting” of optical filters that allows passage of non-collimated, backscattered excitation light not normally incident on the filter surface, and (iii) the limited optical density of optical filters that allows passage of a small amount of ambient and excitation light that is still significant compared with the weak, collected ICG fluorescence (Figure 1b).26 If light that does not originate from the fluorescent dye resident in the tissues is collected, then it represents the “noise floor” making it less probable that the device can image small quantities of a NIRF agent in tissues.
It is also important to note that the sensitivity of the fluorescent imaging device can also be impacted in the environment in which it is used. While it is not feasible to work in total darkness, variation in ambient light, whether provided by fluorescent or incandescent room lighting, surgical lamps or white light illumination in endoscopy or laparoscopic devices, can have small but significant spectral contributions within the spectral bandwidths in which ICG fluorescence is collected.
Area detectors
Currently, charge-coupled device (CCD) detectors are primarily used in ICG fluorescence imaging systems and depend upon integrating the collected photons over millisecond time frames. The CCD detectors used can be divided into (i) front- and back-illuminated CCDs (FCCDs and BCCDs), (ii) electron-multiplying CCDs (EMCCDs) and (iv) intensified CCDs (ICCD), including front- and back- illuminated ICCD based on their configurations. Most imaging systems employed in humans use either FCCD or BCCDs (Table 1). FCCDs are constructed in a fashion similar to the human eye by orienting polysilicon gates at the front, wiring in the middle and photodiodes for light collection at the back. This configuration blocks incident light from reaching the photodiodes resulting in relatively low QE. BCCDs contain the same elements, but by rearranging the gate structure to the back of the photosensitive area of the CCDs and by reducing the thickness of the silicon layer via proprietary etching techniques, BCCDs have more than two-fold improvement in QE than do their front-illuminated counterparts. Unlike a conventional CCD, an EMCCD has an additional electron-emitting register installed between the normal serial register and the output amplifier to multiply weak signals. The EM register is split up into several hundred stages, and each stage acts as an avalanche diode to multiply the signal. The degree of multiplication gain can be controlled by varying the clock voltages applied to the EM register.
By contrast, ICCDs have a different multiplication mechanism, in which an image intensifier amplifies the collected image before it is registered by a CCD. The intensifier consists of three main components, a photocathode, an EM microchannel plate (MCP) and a phosphor screen. The photocathode coverts the low level of incoming light into electrons, which are then accelerated and amplified by a high electrical field towards the MCP and, finally, the amplified signal is converted back into a light signal at the phosphor screen ready for the CCD acquisition. In NIR sensitive generation III intensifiers, a thin film of sintered aluminium oxide is attached to the MCP to protect the photocathode from spurious backscattered ions. Generation II intensifiers are more common but lack the sensitivity to NIR light of generation III intensifiers.
The potential noise sources in these CCD devices consist of the signal inherent shot noise, dark current noise and readout noise.27 Shot noise results from the inherent statistical fluctuations in the number of photons incident on the CCD and represents the detection limit of a detector. For the ICCDs and EMCCDs, shot noise is a function of the mean incident photon flux, multiplication gain, QEs, integration time and a noise factor that characterizes the noise introduced by the gain process. Dark current noise arises from statistical variation in the number of thermally generated electrons that accumulate in the pixels of all CCDs. The rate of dark current generation highly depends on the temperature of the CCD and is especially important for EMCCDs, since it is multiplied by the on-chip multiplication gain. For the ICCDs, additional dark current noise is generated by spontaneous electron output from the intensifier photocathode and is magnified by the multiplication gain of the intensifier. Cooling the CCD can reduce the dark current noise to a negligible level over a typical exposure interval in high-performance CCD cameras. Readout noise is a combination of system noise components inherent to the process of converting CCD charge carriers into an analogue voltage signal for quantification, and the subsequent analogue-to-digital (A/D) converter that converts the analogue voltage signal into a digital representation. For EMCCDs, the readout noise consists of two components: the charge transfer noise arising from the multiplication effect of gain register and all other non-multiplied readout noises. Readout noise becomes dominant in conventional CCDs operating at high frame readout frequency. Our studies showed that the imaging performance of generation III ICCD systems is superior and better than that of unintensified CCD systems in both signal-to-noise ratio and target-to-background ratio when the integration time of CCD camera is <200 ms,25 as shown in Figure 1c. This performance may be expected owing to the optimal amplification of the collected NIRF signal by the intensifier and the subsequent integration that reduces the amplified noise. In general, increased integration times result in maximum signal and reduced noise levels, but at long integration times, motion artefacts and camera saturation owing to limited dynamic range can result. Dynamic range of CCD-based measurements is limited by the A/D register and is typically 8, 10, 12 or the maximum dynamic range of 16-bits.
Investigational and marketed near-infrared fluorescence imaging devices
Several investigational and market NIRF imaging devices have been used clinically (Table 1) and can be divided into the following classifications: (i) hand-held [including HyperEye Medical System (Mizuho Medical), FLUOBEAM® (Fluoptics, Grenoble, France), PDE (Hammatsu Photonics Co., Hamamatsu, Japan), IC-View (Pulsion Medical Systems SE, Feldkirchen, Germany) and Visual Navigator (SH System, Seoul, Republic of Korea)], (ii) cart-based (including FDPM imager, mini-FLARE, SPY and the Prototype Surgical Navigation System) and (iii) incorporated into existing devices, such as the da Vinci® Robotics System (Novadaq Technologies Inc.), by the incorporation of NIR illumination and detection capability (including INFRARED™ 800™, FIREFLY™ and laparoscopic NIRF system). Figure 2 are the photos of typical investigational (FDPM imager) and marketed (PDE and SPY) NIRF imaging devices. The fluence rate of fluorescence excitation light source varies among these systems with the range of <1.9–31 mW cm−2. The fluorescence excitation wavelengths are also different in these systems. For example, FLUOBEAM, FDPM imager, SPY and FIREFLY employ laser didoes operating at 750, 785 and 806 nm, respectively. mini-FLARE, PDE, HyperEye Medical System and Visual Navigator employ LEDs with a centred wavelength located at 760 or 740 nm. INFRARED 800 and FIREFLY utilize the filtered xenon with wavelengths ranging from 700 to 800 nm. The CCD cameras in these systems have varying dynamic ranges and integration times that vary from 0.1 to 1000.0 ms. Collection wavelengths in these systems are mainly centred at 830 nm, which is close to the emission peak of ICG fluorophore and at which CCD cameras have moderately high sensitivity. Some systems such as SPY and PDE have fixed working distance (WD) of 30 and 20 cm, respectively, and fixed fields of view (FOVs) of 19 × 12.7 and 10 × 6.7 cm2, respectively. In these systems, the proper WD is found by positioning the camera head until there is a superposition of two points of LED illumination on the tissue surface to be imaged. Other systems such as the FDPM imager, mini-FLARE and FLUOBEAM have adjustable WDs and FOVs. To accommodate various surgical applications, some systems (including FDPM imager and mini-FLARE) are designed with adjustable WD and FOV accomplished through manual or automatic focus lens along with a moveable articulating arm. During intraoperative or non-invasive applications, it is important to correlate the collected fluorescent image to the actual FOV either by (1) collecting light from a single view in one optical path and spectrally separating the “white-light” and “fluorescent” images for collection by different detectors for direct overlay on display or by (2) collecting two different “white-light” and fluorescent views and co-registering the separate views for combined display.
Figure 2.
Photos of typical investigational and marketed near-infrared fluorescence imaging devices: (a) FDPM imager (University of Texas Health Science Center at Houston, TX), (b) Photodynamic eye (PDE; Hamamatsu Photonics Co., Hamamatsu, Japan) and (c) SPY (Novadaq Technologies Inc., Toronto, ON, Canada).
As depicted in Table 1, the WD, FOV, wavelength specifications for excitation illumination and fluorescence collection, and, finally, detector type varies widely among the devices used in clinical studies. Because the variation in instrument design could result in variable performance, one might expect that varying amounts of ICG employed in these clinical studies could be used to infer performance in lieu of the missing standardized measurements and specification of performance as discussed above. Secondly, NIRF imaging provides an unmet clinical need to assess the lymphatic vasculature and its function.18,19 For these two reasons, we focus upon the clinical application of these devices in which the lymphatic vasculature is interrogated with varying amounts of ICG, whether for identification of SLNs for cancer staging or for assessment of the lymphatic vasculature for diagnosis of various conditions or diseases.
INDOCYANINE GREEN IMAGING OF LYMPH NODES FOR CANCER STAGING
Lymph node mapping for identification of tumour draining lymph nodes for resection and subsequent pathological application in many cancers is commonly performed intraoperatively using blue dyes and/or transcutaneously and intraoperatively with non-specific technetium-99m (99mTc) radiocolloid. While emerging technologies employ i.v. administration of ultrasmall superparamagnetic iron oxide particles28 and intradermally administered radiolabelled mannose sugar, tilmanocept29, intradermal and subcutaneous administration of ICG to locate draining lymph nodes represents an emerging area in nodal staging. The primary advantages of using a radiolabelled fluorescent molecule or NIR active compound is the direct relationship between the regions of fluorescence emitted from tissue surfaces within the surgical FOV. Lymph node detection is often achieved intraoperatively using a gamma probe to provide an audible signal when placed transcutaneously or over a small tissue region containing radioactivity within the surgical FOV. The gamma probe enables localization but does not provide sufficient boundary discrimination to finely guide surgical resection. The advantage for using an NIR fluorescent agent such as ICG over a blue intravital dye is the opportunity to detect subsurface lymph nodes or tissues that may not otherwise be detected intraoperatively within the surgeon's FOV. While ICG fluorescence remains limited in tissue penetration, diagnostic nuclear imaging remains the hallmark for assessing tumour-draining deep lymph nodes in a pre-operative fashion. As a result, in the majority of studies using ICG fluorescence navigation for localizing tumour draining lymph nodes, it is used intraoperatively, rather than non-invasively, in a pre-operative setting as reviewed by Vahrmeijer et al.23 It is important to note that because ICG is not cancer specific, it cannot be used for NIRF detection of cancer-positive lymph nodes, which is the goal of emerging NIRF molecular imaging agents.30–32 A recent review of NIRF molecular imaging agents for identification of tumour margins during cancer surgery is not contained herein but rather in Vahrmeijer et al.23 The following sections detail the performance of the various devices used intraoperatively for SLN detection.
Sentinel lymph node mapping in breast cancer
The majority of studies using ICG fluorescence imaging for lymph node detection are performed in breast cancer as summarized in Table 2. Kitai et al33 first utilized the fluorescence properties of ICG for SLN detection in patients with early breast cancer using the PDE prototype device using 25 mg of ICG in 5 ml of water. Intraoperatively, sentinel nodes were found by visually following the green-stained lymphatic vessels to the first-draining SLNs that were fluorescent. Since then, comparative studies have been performed to assess performance of ICG, blue dye and radioisotope alone or in combination using a range of devices and ICG doses.34–38,40,43–50,52–56 In general, reports are mixed but show that fluorescence imaging provides higher detection rate and lower false-negative rate than do blue dye and radioisotope methods,34,35,37,50,53 and the combination of fluorescence imaging with blue dye visualization was found to be superior to the blue dye method alone.39,47 Moreover, the combination of fluorescence imaging with blue dye visualization had the highest sensitivity method, avoiding the need of the radioisotope method,46 and in another study, no benefit was found in using the blue dye when ICG was combined with radioisotope method.45 Investigators who used the mini-FLARE device often combined ICG with human serum albumin (HSA) to improve quantum yield and SLN retention of ICG fluorophore.36,41,42 Addition of HSA to ICG was not found to significantly alter SLN detection rate when the PDE device was used for detection.42 In another study of 134 patients with breast cancer, detection of SLNs from ICG was not found to be superior to the radiotracer method using the PDE device.55 Combination of ICG and radiotracer has also been attempted for NIRF and gamma detection of SLNs in breast cancer,51 enabling pre-operative localization from nuclear imaging and intraoperative guidance from NIRF imaging, but earlier studies showed that ICG was especially unstable and lost its fluorescent properties when mixed with the acidic solutions used to suspend and administer 99mTc radiocolloid.1,57 To date, various total doses of ICG, ranging from as little as 10 µg by Sevick-Muraca et al,1 to as much as 25 mg used by Kitai et al,33 have been successfully injected for non-invasive and intraoperative detection of visualization of lymph drainage pathways and SLNs. Mieog et al41 determined that the optimal total dose of ICG detected using the mini-FLARE device lies between 0.062 and 1.240 mg of ICG in 1.6 ml of total injection volume. The injection sites of ICG fluorophore have been intradermal periareolar, subcutaneous periareolar, subareolar and deep tumour or a combination thereof, but there is no evidence showing that the SLN detection rate from ICG is affected by intradermal vs subcutaneous route of administration. However, because ICG is non-specific, there is a chance that ICG may drain from the injection sites and receiving lymph node basins prior to detection, resulting in false-negative SLN detection. A typical ICG fluorescence-guided SLN harvesting in a patient with breast cancer using the mini-FLARE device is shown in Figure 3. The number of detected SLNs reported in different clinical studies using ICG fluorescence imaging is variable but, generally, is reported to have higher detection rates than do conventional methods.
Table 2.
Summary of sentinel lymph node (SLN) detection in breast cancer patients
| Study | Number of subjects | Mean age of subjects (years) | Total dose of ICG | Injection site | Device | Operative time (min) | Averaged number of SLNs | Detection rate (%) | Comments | |
|---|---|---|---|---|---|---|---|---|---|---|
| Kitai et al33 | 18 | 56.9 | 25 mg in 5 ml | Subcutaneous periareolar | Prototype of PDE |
<30 | 2.8 (1–6) | 17/18 (94) | The use of prototype PDE for SLN detection | |
| Tagaya et al34 | 25 | – | 5 mg in 1 ml | Subdermal periareolar | PDE |
15 (8–25) | 5.5 (3–10) | 100 | ICG is superior to blue dye with a larger number of SLNs detected | |
| Sevick-Muraca1 | 24 | 57.7 (30–85) | 0.00031–0.1 mg in 0.1–3.0 ml | Periareolar and deep peritumour | FDPM imager |
Immediately | 1.7 (dose >10 µg) | 87 (dose >10 µg) | Determine the minimum dose needed for visualization of lymph drainage pathways | |
| Murawa et al35 | 30 | 56 (27–84) | 5 mg in 1 ml, 10 mg in 2 ml, 15 mg in 3 ml | Intradermal periareolar | IC-View |
5–10 | 1.75 | 29/30 (97) | Visualization of lymphatic vessels depends on the different doses of ICG. ICG is superior to radioisotope | |
| Troyan et al36 | 6 | 61 (51–65) | 0.0124 mg in 1.6 ml | Peritumour | Mini-FLARE™ |
5 | 1.5 (1–4) | 9/10 | The use of ICG:HSA detected by mini-FLARE | |
| Sugie et al37 | 411 | 57.9 (30–91) | 5 mg in 1 ml | Periareolar | PDE |
– | 2.3 (1–9) | 408/411 (99) | ICG is superior to blue dye | |
| Hirche et al38 | 43 | 58.4 (27–83) | 11 mg in 2.2 ml | Subareolar | IC-View |
5–15 | 2.0 (1–3) | 42/43 (98) | – | |
| Hojo et al39 | 141 | 57.6 (34–83) | 10 mg in 2 ml | Intradermal peritumour and subareolar | PDE |
>5 | 3.8 | 140/141 (99) | Comparison of ICG, patent blue and radioisotope. The combination of ICG and blue dye is a highly sensitive method | |
| Tagaya et al40 | 50 | – | 1.25 mg in 0.5 ml | Subdermal areola | PDE |
10.3 (6–18) | 3.7 (1–7) | 100 | With additional imaging overly navigation system | |
| Mieog et al41 | 24 | 59.5 (33–81) | 0.062–1.24 mg in 1.6 ml | Peritumour or periareolar | Mini-FLARE |
16 ± 3 | 1.43 (1–3) | 100 | Optimal injection dose of ICG:HSA ranged between 0.062 and 1.240 mg per 1.6 ml−1 validated on mini-FLARE | |
| Hutteman et al42 | 18 | 59.5 (38–72) and 57.5 (40–73) | 0.62 mg in 1.6 ml | Four intradermal four periareolar | Mini-FLARE |
15.6 ± 2.2 | 1.4 ± 0.5 | 100 | Comparison of ICG:HSA with ICG alone, showing no difference in SLN identification rate | |
| Abe et al43 | 128 | 53 (29–73) | 0.75 mg in 0.15 ml | Intradermal | PDE |
<15 | 3.1 (1–6) | 100 | ICG fluorescence demonstrated higher sensitivity compared with that of blue dye | |
| Aoyama et al44 | 312 | 57.4 (29–85) | 0.625 mg in 5 ml | Subareolar periareolar | PDE |
A few | 3.41 (1–12) | 100 | – | |
| van der Vorst et al45 | 24 | 59 (39–75) | 0.62 mg in 1.6 ml | Four intradermal four periareolar | Mini-FLARE |
15.2 ± 3.0 | 1.5 ± 0.8 (1–4) | 23/24 (96) | No benefit of using patent blue when combining ICG with radioisotope | |
| Wishart et al46 | 100 | 60 (34–81) | 10 mg in 2 ml | Intradermal and subcutaneousperiareolar | PDE |
– | 1.93 (1–5) | 100 | Comparison of ICG, blue dye and radioisotope. The combination of ICG and blue dye had the highest nodal sensitivity method, avoiding the need of the radioisotope method | |
| Hirano et al47 | 108 | 60.5 (31–85) | 5 mg in 2.5 ml | Subareolar | PDE |
– | 2.2 (1–5) | 100 | The combination of blue dye with ICG is superior to blue dye alone | |
| Takeuchi et al48 | 145 | 57.2 (30–83) | 5 mg in 1 ml | Subcutaneous periareolar | PDE |
– | – | 144/145 (99) | – | |
| Polom et al49 | 49 | 54.4 (31–71) | 0.465–10 mg in 1 ml | Intradermal | PDE |
– | – | 47/49 (96) | – | |
| Sugie et al50 | 99 | 60 (29–75) | 2.5–5.0 mg in 0.5–1.0 ml | Subareolar | PDE |
– | 3.4 (1–7) | 98/99 (99) | ICG is superior to blue dye | |
| Schaafsma et al51 | 32 | 56 (34–82) | 0.05 mg in 0.2 ml | Periareolar | Mini-FLARE |
24 h (19–29) | 1 (1–2) | 100 | Integration of ICG and radioactive labels in a single tracer | |
| Jung et al52 | 43 | 48.2 ± 9.6 | 0.18 mg in 0.3 ml | Subareolar | Visual Navigator |
6.5 ± 5.2 | 3.4 ± 1.4 | 100 | A mixture of ICG, radioisotope and blue dye compared with radioisotope alone, high number SLNs detected by the mixture method | |
| Guo et al53 | 36 | 55.2 (30–81) | 5 mg in 1 ml | Subareolar | PDE | 5–10 |
3.6 | 35/36 (97) | ICG is superior to blue dye | |
| Chi et al54 | 22 | 49 (32–68) | 1–2 mg in 0.2–0.4 ml | Subcutaneous areolar | The prototype surgical navigation system | 5–10 |
2.7 (1–6) | 100 | Use of a prototype surgical navigation system for SLN detection | |
| Ballardini et al55 | 134 | – | 5 mg in 1 ml | Subdermally, peritumoural | PDE | – |
– | 99.6 | Concordance between ICG and radiotracer methods. ICG and NIRF imaging can be used with equal success of the radiotracer method to detect SLNs | |
| Verbeek et al56 | 95 | 57 (30–75) | 0.62 mg in 1.6 ml | Intradermal periareolar or peritumour | Mini-FLARE | 19 ± 7.1 | 1.9 (1–5) | 94/95 (99) | The comparison of ICG, radioisotope and blue dye; call for a standard method |
HSA, human serum albumin; ICG, indocyanine green; NIRF, near-infrared fluorescence; PDE, photodynamic eye.
The devices mentioned in the table and their manufacturer details in parenthesis are as follows: FDPM imager (University of Texas Health Science Center at Houston, TX), IC-View (Pulsion Medical Systems SE, Feldkirchen, Germany), Mini-FLARE (Israel Beth Deaconess Medical Center, Boston, MA), PDE (Hamamatsu Photonics Co., Hamamatsu, Japan) and Visual Navigator (SH System, Seoul, Republic of Korea).
Figure 3.
Single sentinel lymph node (SLN) identified and resected for a patient with breast cancer after injection (Inj.) of indocyanine green. Colour video images are shown in the left, near-infrared (NIR) fluorescence imaging is shown in the middle and a pseudocoloured merge of the two are shown in the right, where arrows point to the positions of the SLN. LC, lymphatic channel. Reproduced from Troyan et al36 with permission from Springer Science + Business Media.
Skin cancer
Studies using ICG fluorescence imaging for SLN detection in skin cancer are summarized in Table 3. Fujiwara et al58 first reported the use of fluorescence method to identify the SLN successfully in melanoma and squamous cell carcinoma of the skin. Moreover, they found that use of fluorescence method avoids the “shine-through” phenomenon seen in lymphoscintigraphy, thereby providing high detection rate of the cervical SLN.64 In general, 0.5 mg in 0.1 ml dose of ICG was injected intradermally at various sites around the tumour, with a total dose volume ranging from 0.4 to 1.0 ml. To improve fluorescence signal, ICG:HSA was also used in melanoma SLN detection,69 and the ICG:HSA dose escalation clinical trial was also performed.71 In addition, the effectiveness and limitations of fluorescence detection of lymph nodes were evaluated with transcutaneous imaging, and imaging within the surgical FOV and of the excised tissues. Studies showed a lower detection of lymph nodes from transcutaneous as opposed to from within the surgical FOV or from resected tissues containing lymph nodes.70 It was observed that transcutaneous SLN detection rate could be improved by manual manipulation of overlying tissues or by increasing the dose of ICG.70 Other opportunities for improving detection could result from improving device sensitivity.
Table 3.
Summary of sentinel lymph node (SLN) detection in patients with skin cancer
| Study | Number of subjects | Mean age of subjects (years) | Total dose of ICG | Injection site | Device | Operative time (min) | Averaged number of SLNs | Detection rate (%) | Comments |
|---|---|---|---|---|---|---|---|---|---|
| Fujiwara et al58 | 10 | 68 (34–80) | 3–5 mg in 0.6–1.0 ml | Intradermal peritumour | PDE | <15 | – | – | The detection of the SLN in melanoma and squamous cell carcinoma of the skin: the first report |
| Tsujino et al59 | 2 | 75, 88 | 5 mg in 1 ml | Intradermal | PDE | 2–3 | – | – | – |
| Tanaka et al60 | 6 | – | 5–8 mg in 1 ml | Intradermal | PDE | – | – | 100 | ICG fluorescence is superior to the blue dye method |
| Mizukami et al61 | 24 | – | 3–10 mg in 0.6–2.0 ml | Intradermal | PDE | – | – | 23/24 (96) | – |
| Namikawa et al62 | 49 | 62 (18–85) | 5 mg in 1 ml | Intradermal peritumour | PDE | Approximately minutes | 4.0 | 61.1, 100 | A combination of blue dye, radioisotope and fluorescence method for improving SLN detection rate |
| Fujisawa et al63 | 16 | – | 1.5 mg in 0.3 ml to 3.0 mg in 0.6 ml | Intradermal peritumour | FNS | 10 | 1.5 per basin (1–3) | 100 | A custom-made low-cost intraoperative system developed and tested |
| Hayashi et al64 | 1 | 47 | 2 mg in 0.4 ml | Intradermal peritumour | PDE | – | – | 100 | Fluorescence method overcomes the shine-through phenomenon in head and neck/lip mucosa melanomas |
| Stoffels et al65 | 22 | 51.63 | 0.2 mg in 1 ml | Peritumour | PDE | 5–10 | 2.8 | 100 | Fluorescence method is an attractive option with radioisotope method |
| Fujisawa et al67 | 34 | – | 2.0 mg in 0.4 ml to 4.0 mg in 0.8 ml | Intradermal peritumour | – | <15 | 2.18 | 100 | Fluorescence method is superior to the conventional methods |
| Rasmussen et al68 | 4 | – | 0.4 mg in 1.6 ml | Peritumour, intradermal | FDPM imager | – | – | – | Study assessed tumour draining lymphatics longitudinally in patients with cancer but did not assess surgically guided resection of SLN |
| Polom et al66 | 10 | 56.4 (26–79) | 0.2–0.93 mg in 2 ml | Peritumoural, intradermal | PDE | – | 1.9 | – | Radioisotope and ICG methods yield the same results |
| van der Vorst et al69 | 15 | 55 (21–70) | 0.75, 1.0, 1.2 and 1.5 mg in 1.6 ml | Peritumour | Mini-FLARE™ | 12.6 ± 3.6 | 2.0 ± 1.6 | 93 | An ICG:HSA concentration of 0.744 mg per 1.6 ml appears optimal with mini-FLARE |
| Namikawa et al70 | 86 | 63 (18–87) | 2.0 mg in 0.4 ml to 5.0 mg in 1.0 ml | Intradermal peritumour | PDE | – | – | 63.4, 98.9, 100 | The evaluation of the SLN detection rate at the “before skin incision”, “after skin incision” and “ex vivo” surgical stages |
| Gilmore et al71 | 25 | 47 (25–79) | 77.5, 193.8 or 387.5 µg in 1.0 ml | Intradermal peritumour | Mini-FLARE | >5 | 2.6 | 24/25 (96) | The dose escalation clinical trial |
| Nakamura et al72 | 12 | 66 (32–85) | 3.0 mg in 0.6 ml | Intradermal peritumour | – | – | 3.2 | 95 | The improvement in the detection rate of cervical sentinel nodes in head and neck skin cancer using fluorescence method in combination with the standard technique |
FNS, fluorescence navigation system; HSA, human serum albumin; ICG, indocyanine green; PDE, photodynamic eye.
The devices mentioned in the table and their manufacturer details in parenthesis are as follows: FNS (University of Tsukuba, Tsukuba, Japan), PDE (Hamamatsu Photonics Co., Hamamatsu, Japan), FDPM imager (University of Texas Health Science Center at Houston, TX), Mini-FLARE (Israel Beth Deaconess Medical Center, Boston, MA).
Other cancers
Studies using ICG fluorescence imaging for SLN detection in other cancers are summarized in Table 4. Miyashiro et al73 used the fluorescence method to detect the SLN of gastric cancer in open gastrectomy, and ICG was injected endoscopically around the tumour 1 day before surgery. Later studies showed that pre-operative, endoscopic ICG injection provides a larger averaged number of SLNs, a higher accuracy rate and a lower false-negative rate than does intraoperative ICG injection in cT1-stage gastric cancer.74 Some investigators showed the possibility of SLN mapping guided by ICG fluorescence imaging during laparoscopy-assisted gastrectomy.75 The adequate dose of ICG injected submucosally on the day before operation has been reported to be four doses of 0.5 ml containing 50 µg in 1 ml ICG for SLN mapping using HyperEye Medical System in the gastric cancer surgery.77 The NIRF method has also been applied for SLN mapping in rectal,79 anal,80 colon78,81 and colorectal82 cancers with acceptable detection rates. Moreover, the SLN could be detected not only in T1 disease but also in T2 and T3 disease.79 In these studies, the total dose of ICG ranged from 14 mg in 2.8 ml to 25 mg in 5 ml injected around the tumours. The da Vinci, integrated with FIREFLY, has been used for SNL mapping in bladder and prostate cancer.83,84 Recently, the application of the NIRF method for SLN detection was attempted in cervical cancer,85,86 showing a similar detection rate as that using the blue dye and radioisotope method. As in breast and skin cancers, 1.6 ml of ICG:HSA was injected around the tumour for SLN mapping in patients with cervical cancer.87 Further studies demonstrated that there is no advantage of ICG:HSA over ICG alone for SLN detection in early stage cervical cancer, thus avoiding the added cost and complexity in using ICG:HSA.89 SLN mapping has also been demonstrated using robotically assisted endoscopic NIR imaging after cervical injection of appropriate dose of ICG around the tumour.88 The NIRF method for SLN mapping has been introduced in other types of cancers, such as vulvar and endometrial cancer,90–94 head and neck cancer95–97 and non-small-cell lung cancer98,99 with high detection rate.
Table 4.
Summary of sentinel lymph node (SLN) detection in other patients with cancer
| Study | Type of cancer | Number of subjects | Mean age of subjects (years) | Total dose of ICG | Injection site | Device | Operative time (min) | Averaged number of SLNs | Detection rate (%) | Comments |
|---|---|---|---|---|---|---|---|---|---|---|
| Miyashiro et al73 | Gastric cancer | 3 | 66 (60, 66, 73) | 10–20 mg in 2–4 ml | Peritumour | PDE | A few and 1 day | 3 (4, 1, 4) | 100 | ICG injected 1 day before surgery |
| Tajima et al74 | 56 | 68.4 ± 10.1 | 10 mg in 2 ml | Submucosa peritumour or into subserosa | PDE | A few or 1–3 days | 7.2 ± 7.0 (1–30) | 54/56 (96) | Pre-operative injection of ICG is preferable with cT1 stage cancer | |
| Tajima et al75 | 77 | 67.2 ± 10.6 | 10 mg in 2 ml | Submucosa peritumour or subserosa | PDE-2 | 1–3 days | 7.9 (LAG) 7.2 (OG) | 94.7, 94.9 | SLN mapping possible, even during LAG | |
| Miyashiro et al76 | 10 | 68 ± 8 (54–79) | 1–10 mg in 2–4 ml | Peritumour | A prototype laparoscopy | Approximately 5 | 3.1 ± 1.5 (1–6) | 100 | Development of a prototype laparoscopic system for SLN detection in gastric cancer | |
| Yoshida et al77 | 12 | 58.4 ± 14.3 | 0.05–0.2 mg in 2 ml | Submucosal peritumour | HyperEye Medical System | 1 day | 3.6 ± 2.1 | 100 | Determination of ICG dose and harvesting time | |
| Watanabe et al78 | Rectal, anal, colon, colorectal bladder cancer | 10 | 62.5 (43–80) | 5 mg in 2 ml | Submucosa | PDE | – | – | 100 | Detection of tumour |
| Noura et al79 | 25 | 58.4 ± 11.6 (33–74) | 5 mg in 1 ml | Peritumour | PDE | <30 | 2.1 ± 0.8 (1–4) | 23/25 (92) | Detection of lateral SLN not only in T1 and T1 diseases, but also in T3 diseases | |
| Hirche et al80 | 12 | 58 (21–90) | 25 mg in 5 ml | Subdermal peritumour | IC-View | 10–15 | 1.6 | 10/12 (83) | Application in anal cancer | |
| Hirche et al81 | 26 | 67 (46–87) | 5–20 mg in 1–4 ml | Peritumour | IC-View | 3–10 | 1.7 (0–5) | 25/26 (96) | The application in colon cancer | |
| Cahill et al82 | 18 | 66 (48–80) | 10–15 mg in 1–3 ml | Submucosal peritumour | Laparoscopic NIRF system | 1 day (1 case), 10 | 12 (1 day) and 3.6 (1–5) | 100 | Use of NIR laparoscopy for SLN detection in early stage colorectal neoplasia | |
| Manny and Hemal83 | Prostate cancer | 10 | 71 (54–77) | 5 mg in 2 ml | Peritumour | FIREFLY™ (da Vinci®) | 30 | 16 (12–25) | 100 | Using the combined cystoscopic and intravenous injection of ICG in bladder cancer |
| Manny et al84 | 50 | 66 (51–73) | 1 mg in 0.4 ml | The lobe of the prostate | FIREFLY (da Vinci) | 30 | – | 100 | Application in prostate cancer | |
| Furukawa et al85 | Cervical, vulvar, endometrial cancer | 12 | 58 (36–68) | 4 mg in 0.8 ml | In the cervix | PDE | 30 (20–40) | 7 (3–10) | 10/12 (83) | Fluorescence method comparable with the conventional method |
| Crane et al86 | 20 | 51.3 (39–74) | 0.5 mg in 1 ml | In the cervix | Custom | 0.5 | 1.5 | 6/9 (67) | – | |
| van der Vorst et al87 | 9 | 40 (29–77) | 0.62, 0.93 or 1.24 mg in 1.6 ml | Transvaginal submucosal peritumour | Mini-FLARE™ | 51 ± 18 | 3.4 ± 1.2 (1–5) | 100 | The optimal concentration of ICG:HSA is 0.62 mg in1.6 ml with Mini-FLARE | |
| Rossi et al88 | 20 | 61 (28–82) | 1 mg in 2 ml | 1 cm into the cervical stroma | SPY | 37.3 (8–60) | 4.5 (0–9) | 85 | Appropriate dose required with robotic assistance | |
| Schaafsma et al89 | 18 | 40 (28–67) | 0.62 mg in 1.6 ml | Submucosal peritumour | Mini-FLARE | 43 ± 14 | 2.9 (ICG:HSA) 2.7 (ICG) | 14/18 (78) | No advantage of ICG:HSA over ICG alone for the SLN detection in early stage cervical cancer | |
| Crane et al90 | 10 | 67.5 (46–82) | 0.5 mg in 1 ml | Peritumour | Custom | Immediately | 2.6 (1–6) | 100 | Transcutanous SLN mapping possible in 50% of all patients | |
| Hutteman et al91 | 9 | 50 (30–72) | 0.62–1.24 mg in 1.6 ml | Peritumoural | Mini-FLARE | 19 ± 4 | 1.6 (1–4) | 100 | ICG:HSA solution used | |
| Holloway et al92 | 35 | 63.4 | 2.5 mg in 2 ml | Injected into cervix | FIREFLY (da Vinci) | 10 | – | 100 | Robotic-assisted lymphadenectomy | |
| Rossi et al93 | 29 | 62 63.5 | 0.5–1.0 mg | Cervical (n = 17) and hysteroscopic injections (n = 12) | FIREFLY (da Vinci) or SPY | – | 5 (1–9) 2.5 (1–3) | 82% for cervical injection 33% for hysteroscopic | A higher SLN detection rate by cervical injection compared with hysteroscopic endometrial injection | |
| Jewell et al94 | 227 | 60 (28–90) | 5 mg in 4 ml | Into cervix, stroma and submucosa | FIREFLY (da Vinci) | – | 3 (1–23) | 216/227 (95) | Blue dye appears unnecessary | |
| Bredell95 | Head and neck cancer | 8 | – | 10 mg in 1 ml | Peritumour | PDE | 5 | 3 (1–5) | 100 | – |
| van den Berg et al96 | 14 | 65.5 (51–84) | 0.05 mg in 0.4 ml | Peritumour | PDE | 10 | 3.4 (1–7) | 100 | ICG can detect SLNs located in close proximity to the primary tumour | |
| van der Vorst et al 97 | 10 | 59.5 (33–73) | 0.62 mg in 1.6 ml | Peritumour | Mini-FLARE | 5–30 | 1.7 ± 0.8 | 100 | – | |
| Yamashita et al98 | Non-small-cell lung cancer | 31 | 63 (54–83) | 10 mg in 2 ml | Peritumour | Infrared Camera Systems | 10 | 1.3–1.4 | 25/31 (81) | – |
| Gilmore et al99 | 38 | 64.9 ± 10.3 | 0.0038–2.5 mg | Peritumour | Mini-FLARE/and Novadaq® NIR camera | 5 | 1.7 | 90 at the dose of ICG >1 mg | Identification of optimized dose of ICG |
HSA, human serum albumin; ICG, indocyanine green; LAG, laparoscopy-assisted gastrectomy; NIRF, near-infrared fluorescence; OG, open gastrectomy; PDE, photodynamic eye.
The devices mentioned in the table and their manufacturer details in parenthesis are as follows: FIREFLY (da Vinci; Novadaq Technologies Inc., Toronto, ON, Canada), HyperEye Medical Systems (Mizuho Medical Co., Ltd, Tokyo, Japan), IC-View (Pulsion Medical Systems SE, Feldkirchen, Germany), Infrared Camera Systems (Olympus, Tokyo, Japan), Mini-FLARE (Israel Beth Deaconess Medical Center, Boston, MA), Novadaq NIR camera (Novadaq Technologies Inc.), PDE (Hamamatsu Photonics Co., Hamamatsu, Japan), SPY scope (Novadaq Technologies Inc.).
In the Western world, the incurable condition of lymphoedema is generally caused after both SLN and full lymph node resections, presenting without warning weeks to years after surgery. Although lymphoscintigraphy has been performed routinely to image the lymphatics, the long integration times and relatively low spatial resolution of this technique limit its application for real-time imaging of lymphatic contractile function and for visualizing fine lymphatic architecture. Recently, ICG fluorescence imaging has been adopted for human lymphatic imaging, as described in the following section.
LYMPHATIC IMAGING
Clinical lymphatic imaging studies using NIRF imaging systems are summarized in Table 5. NIRF imaging has been used to characterize lymphoedema of the upper and lower extremities using both PDE2,16,20,100–104,107–113,116,117,119,120,122 and FDPM devices.1,22,105,106,114,118,121 Interpretation of NIRF images include aberrant lymphatic vasculature as seen comparatively to normal lymphatic vasculature in Figure 4a (Supplementary Video A), impaired transit time measurements (i.e., the time between ICG injection and its appearance in the nearest draining lymph nodes) using the PDE device; as well as measurements of reflux or impaired lymphatic propulsion using the FDPM device (Figure 4b; Supplementary Video B). Using the PDE device, Mihara et al20 showed NIRF imaging to be a superior diagnostic to lymphoscintigraphy when evaluating for lymphoedema. NIRF imaging has also been used to detect lymphatic abnormalities before the onset of swelling symptoms using the FDPM imager,21,22 earlier than lymphoscintigraphy122 and following cancer surgery.68 Using both devices, multiple intradermal injections of single ICG doses were used for lymphatic mapping, typically 0.025 mg in 0.1 ml saline using the FDPM imager and 0.5 mg in 0.1 ml water using the PDE device. Assessment of lymphatic architecture and transport was also performed with NIRF imaging devices in subjects with head and neck lymphoedema113,115 and used in the treatment setting to assess the effect of lymphoedema treatments, i.e. manual lymphatic drainage and pneumatic compression devices, on lymphatic function.105,106,115 The results suggest the ability to use NIRF imaging to direct therapy and to stratify which subjects could benefit by these non-surgical therapies on the basis of stimulated lymphatic pumping or lymphatic vessel recruitment. LVA, a surgical therapy for late-stage and severe lymphoedema, also uses NIRF imaging in a therapeutic setting to stratify patients who would benefit from the treatment,122 to guide the skin incision site,15,17 identify functional subcutaneous and dermal lymphatic vessel LVA16,17,107 and assess patency of LVA after surgery.15,108 In another application, NIRF imaging has been used to perform axillary reverse mapping in patients with breast cancer, to identify those lymph nodes that are drained by the arm and thus could be potentially spared in lymph node dissection to prevent later the onset of upper extremity lymphoedema.101,117,123–125 Lymphatic mapping using NIRF imaging has also been used to provide a lymphatic phenotype for genotyping;118 to evaluate the lymphatics in infants and neonates with lymphatic pleural effusion and ascites;120 and to direct surgical management of post-operative chylothorax in infants.121
Table 5.
Summary of lymphatic imaging using indocyanine green (ICG) fluorophore
| Study | Number of subjects | Total dose of ICG | Injection site | Device | Comments |
|---|---|---|---|---|---|
| Unno et al100 | 22 | 2 mg in 0.4 ml | Subcutaneously at the dorsum of the foot | PDE | The first published report of ICG fluorescence imaging of human lymphatics in lymphoedema and normal subjects |
| Ogata et al16 | 5 | 2 mg in 0.4 ml | Intracutaneously into the dorsal aspect of foot (first web space) | PDE | Guiding lymphaticovenular anastomoses |
| Unno et al2 | 10 | 3 mg in 0.6 ml | Subcutaneously at the dorsum of the foot | PDE | Quantitation of lymph function from transit time measurements |
| Ogasawara et al101 | 37 | 25 mg in 5 ml | Subdermal subareolar (3 ml); subdermal peritumour (2 ml) | PDE | Evaluation of the number of lymphatic pathways in patients with breast cancer |
| Sevick-Muraca et al1 | 24 | 0.31–100 µg in 0.1–0.3 and 1–3 ml | Intradermal subcutaneous | FDPM imager | Determination of the minimum microdose of ICG for transcutaneous lymph tracking to sentinel lymph nodes in patients with breast cancer and showed active lymph propulsion |
| Suzuki et al102 | 54 | 1.5–3 mg in 0.3 ml | Subcutaneously in dorsum of foot | PDE | Assessed lymphatic recovery after vein stripping |
| Kamiya et al103 | 1 | 7.5 mg in 1.5 ml | Subcutaneous in bilateral inguinal region | PDE | Located lymphatic fistula causing thorax following oesophagectomy |
| Rasmussen et al22 | 44 | 0.4 mg in 1.6 ml | Intradermal | FDPM imager | Assessment of lymphatic architecture and propulsive transport in arms and legs |
| Unno et al104 | 65 | 3 mg in 0.6 ml | Subcutaneous injection in dorsum of foot | PDE | Measured lymphatic pressure with inflatable cuff |
| Adams et al105 | 9 | 0.3 mg in 1.2 ml | Intradermal injection | FDPM imager | Assessed movement of ICG before and after pneumatic compression therapy in arms |
| Tan et al106 | 20 | 0.4 mg in 1.6 ml | Intradermal | FDPM imager | Assessment of lymphatic contractile function after manual lymphatic drainage in arms and legs |
| Furukawa et al107 | 9 | (Not specified) in 0.2 ml | Intracutaneously into the dorsal aspect of each second web space | PDE | Targeting dermal lymphatic backflow during microsurgical lymphaticovenous implantation |
| Mukenge et al108 | 11 | (Not specified) in 0.2–1.0 ml | Subcutaneous | PDE | Assessment of the patency of LVA after surgery |
| Mihara et al17 | 6 | (Not specified) in 0.2 ml | Intracutaneously into the dorsal aspect of the foot (first web space) | PDE | Identifying subcutaneous lymph vessels before and during scarless LVA |
| Yamamoto et al109 | 20 | 0.5 mg in 0.2 ml | Subcutaneous in arms and the second web space of the hand | PDE | Proposed classification system for arm lymphoedema |
| Yamamoto et al110 | 45 | 1 mg in 0.4 ml | Subcutaneous in leg and first web space of the foot | PDE | Proposed classification system for leg lymphoedema |
| Yamamoto et al111 | 62 | 0.5 mg in 0.2 ml | Subcutaneous in leg and first web space of the foot | PDE | Assessed ICG imaging as a means to diagnose subclinical lymphoedema |
| Unno et al112 | 399 | 3 mg in 0.6 ml | Subcutaneous | PDE | Assessed the effect of age and gender on lymphatic function |
| Yamamoto et al113 | 5 | 0.04 mg in 0.3 ml | Subcutaneous | PDE | Imaging of lymphatics in normal controls and patients with head and neck lymphoedema |
| Aldrich et al114 | 24 | 0.3 mg in 1.2 ml | Intradermal | FDPM imager | Assessed systemic progression of lymphoedema in breast cancer survivors |
| Maus et al115 | 1 | 0.225 mg in 0.9 ml | Intradermal | FDPM imager | Imaging of lymphatics in head and neck lymphoedema used to redirect manual lymphatic drainage |
| Akita et al116 | 134 | (Not specified) in 0.6 ml | Web spaces of feet | PDE | Showed that ICG lymphography could detect earlier stages of lymphoedema than does lymphoscintigraphy |
| Akita et al122 | 100 | (Not specified) in 0.6 ml | Subcutaneously at the first web space of the foot | PDE | Determination of early stage lymphoedema after lymph node dissection in gynaecological cancer based on the detected patterns |
| Sakurai et al117 | 327 | (Not specified) in 0.15 ml | Subcutaneously into the interdigital area | PDE | Identifying the risk group of post-operative lymphoedema through axillary reverse mapping |
| Burrows et al118 | 1 | 0.3 mg in 1.2 ml | Intradermal | FDPM imager | Imaging of lymphatic abnormalities associated with RASA1 gene mutations |
| Liu et al15 | 20 | 2–4 mg in 0.2–0.4 ml | Subdermal | Leica FL800 | Guiding the skin incision site of LVA and assessment of the patency of the anastomosis |
| Mihara et al119 | 72 | 2 mg in 0.4 ml | Subcutaneously | PDE | Diagnosed and classified lymphoedema |
| Shibasaki et al120 | 10 | 0.25 mg in 0.1 ml (single dose of multiple injections) | Subcutaneously on hands and feet | PDE | Evaluation of the severity of lymphatic dysfunction in infants and neonates with congenital lymphatic pleural effusion and ascites |
| Tan et al121 | 1 | 0.050 mg in 0.2 ml | Intradermally on hands and feet | FDPM imager | Imaging of abnormal lymphatic drainage in an infant with post-operative chylothorax to guide surgical management |
LVA, lymphovenous anastomoses; PDE, photodynamic eye; RASA1, gene encoding RAS p21 protein activator (GTPase activating protein).
The devices mentioned in the table and their manufacturer details in parenthesis are as follows: Leica FL800 (Leica Microsystems Inc., Buffalo Grove, IL), FDPM imager (University of Texas Health Science Center at Houston, TX), PDE (Hamamatsu Photonics Co., Hamamatsu, Japan).
Figure 4.
(a) Static near-infrared fluorescence (NIRF) imaging of indocyanine green (ICG)-laden lymphatics in the arm of a normal subject. Supplementary Video A shows increased lymphatic contractile transport in response to manual lymphatic drainage applied by a trained therapist. (b) Static NIRF imaging of ICG-laden lymphatics in the arm of a breast cancer survivor with lymphoedema in the right extremity as depicted in (c). Supplementary Video B shows the comparative lack of lymphatic propulsion compared with that shown in the arm of a normal subject shown in (a). Reproduced from Tan et al,106 with permission from Elsevier.
PHANTOMS FOR ASSESSING CLINICAL PERFORMANCE
Currently, the success of NIRF imaging is based upon the ability of the device to detect ICG in humans, as outlined in Tables 2–5, with the expectation that “first-in-human” molecularly targeted imaging agents can likewise be detected by these devices already employed to detect ICG. ICG is comparatively dim compared with other emerging NIR fluorophores, such as IRDye800® (Li-Cor Biosciences, Lincoln, NE) and others,57 which promise brighter fluorescent signals and the ability to detect targeted tissues deeper and with smaller concentrations than is used for haemo- and lymphovascular imaging of ICG. Device performance is critical and depends upon optimal excitation of NIRF imaging agents, rejection of backscattered excitation and ambient light, and selective collection of fluorescence emanating from the fluorophore. While instrument design differs widely (Table 1), and NIRF molecularly targeted imaging agents can be expected to have broad spectral characteristics that make them amendable to different devices, there remains no systematic measurement procedure and test to enable prediction that a given molecular imaging agent can be detected in humans by a given NIRF imaging device. Furthermore, as technologies evolve and as NIRF imaging device components change, there remains no standardized means that could track device improvements over time and establish clinical performance without involving clinical trials that are often costly. There is also no standardized measurement and approach to qualify the consistent, equivalent operational status of a single device as a function of clinical usage time. Hence, there is an urgent need to develop working standards that are stable over time and enable qualification of imaging device performance and that decouple the device and drug regulatory processes. For example, if a “first-in-humans” imaging agent fails to demark a diseased tissue, how is one to determine whether it was the imaging agent that failed to target the diseased tissue or whether the imaging device was not sensitive enough for its detection? The ICG dose escalation studies by Sevick-Muraca et al1 and Mieog et al41 provide important information for setting the dosing and detection levels of imaging agents to begin safety and toxicity studies for eventual translation but inevitably are specific to the particular imaging device used.
Many groups have embarked on the process to construct solid phantoms for quantifying the performance of complete NIRF imaging systems, as well as for comparing systems.25 While past work has focused upon adding fluorophore or imaging agent in milk or intralipid solutions, these liquid standards have a finite shelf life, are not often replicable and require comprehensive characterization for each use. Table 6 summarizes the developmental efforts found in the literature that seek to establish solid phantoms in which a fluorophore or luminescent material is embedded in a solid medium that mimics or exaggerates tissue-scattering properties.25,27,126–129 These phantoms are essential to sensitively and accurately detect changes in device performance during manufacture, installation and operation in the clinic. Because ambient light can change with time in the environment, the use of the standards at the point of care can define the expected performance of the device. Hence, if the ICG is not detected during lymphatic mapping in a specific site, then one can safety assume that there are no functional lymphatics to be visualized. Likewise, if a molecular imaging agent is not detected, one can conclude that the molecular imaging agent was not present in or subsurface to the FOV. Long-term stability (>1 year) of these phantoms remains elusive, and while promising candidates are under development, as shown in Table 6, there needs to be an adoption of an industry standard to promote further technological evolution of NIRF devices and their use with emerging NIRF imaging agents.
Table 6.
Summary of near-infrared and far-red fluorescent solid phantoms
| Phantom | Base material | Scatter | Absorber | Fluorescent dye | Excitation peak (nm) | Emission peak (nm) | Fluorescence stability | Comments |
|---|---|---|---|---|---|---|---|---|
| Boehm et al126 | Gelatin | Milk powder | None | NIR96010 | 755 | 790 | NS | To determine the concentration-dependent contrast resolution of NIR mammography |
| Grand et al127 | Gelatin | Intralipid | Haemoglobin | ICG; Pam78 | NS | NS | Long-term stability for both ICG and Pam78 at some wavelengths for a period of 4 weeks (measured weekly) | Assessing NIRF imaging system and training surgeon |
| Baeten et al128 | Silicone | White silicone pigments | Black silicone pigments | IR-676 iodide; IR-780 iodide | 676; 780 | 700; 799 | Long-term stability during (measured every 2 h) and after the curing process (measured periodically) for IR-676 iodide and IR-780 after curing (approximately 2 months) | The hydrophobic fluorophores (IR-676 and IR-780) embedded within silicone are suitable for diffuse fluorescence tomography standards |
| Polyester resin | TiO2 | India ink | Cy5.5; AF 750 | 675; 749 | 694; 775 | Unstable in the first 24 h (measured every 2 h) and then fairly steady degradation after curing (measured periodically) for Cy5.5 over a period of 2 months; unstable during the curing process for AF 750 dye | ||
| Pleijhuis et al129 | Gelatin | Intralipid | Haemoglobin | ICG | 780 | 820 | NS | Assessing intraoperative techniques |
| Zhu et al27 | Polyurethane | None | None | QDots 800 | None | 800 | Long-term stability over a 3-month period after curing (measured periodically) | High reflectance to quantify excitation light leakage |
| Zhu et al25 | Polyurethane | TiO2 | None | QDots 800 | None | 800 | Long-term stability over a 3-month period after curing (measured periodically) | Different concentrations of QDots 800 to assess the performance of NIRF imaging systems |
| XFM-2 Fluorescent Phantom | Polyurethane | NS | None | AF 680; AF750; QDots 800 | 640; 745; none | 700; 800; 800 | Up to 60% reduction in fluorescent activity over a 4-month period | To test fluorescent imaging, spectral unmixing or fluorescent tomography |
AF 680, Alexa Fluor 680 (Invitrogen, Carlsbad, CA); AF 750, Alexa Fluor 750 (Invitrogen); Cy5.5, cyanine dye 5.5 (Amershan Biosciences, Piscataway, NJ); Gelatin (Sofertgelatine; RUF Lebensmittelwork, Quakenbrück, Germany; ICG, indocyanine green; NIRF, near-infrared fluorescence; NS, not specified; QDots 800, Quantum Dots 800 (Qdot® 800 ITkTM, Q21 771MP, Invitrogen); TiO2, titania or titanium dioxide.
IR-676 and IR-780 iodide (Sigma-Aldrich Chemical Company, Inc., Milwaukee, WI), NIR96010 (Institut for Diagnostikforshuing, Schering AG, Berlin, Germany) XFM-2 Fluorescent Phantom (Caliper Life Science, Hopkinton, MA).
SUMMARY
NIRF imaging is rapidly emerging for clinical characterization of the lymphatic vasculature, whether for SLN mapping in cancer surgeries or for assessing lymphatic vessel architecture and function in health and a variety of diseases. Herein, we summarized the performance variety of NIRF imaging devices used in the clinic to detect ICG, the only NIR excitable dye currently used in humans. The performance of these devices, while variable, offers promise for the emerging NIRF molecular imaging agents that promise profound impact in surgical and non-surgical applications. Now that there is a plethora of NIRF devices in the clinic and a range of ICG doses used as summarized herein, a critical evaluation of the efficient approaches to evolve to the next step of using “first-in-humans,” molecularly targeted imaging agents is needed.
FUNDING
The work was compiled under grant support National Institutes of Health U54 CA136404 and R01 HL09232 (EMS).
Acknowledgments
ACKNOWLEDGMENTS
We thank Dr John C Rasmussen for reviewing the contribution.
REFERENCES
- 1.Sevick-Muraca EM, Sharma R, Rasmussen JC, Marshall MV, Wendt JA, Pham HQ, et al. Imaging of lymph flow in breast cancer patients after microdose administration of a near-infrared fluorophore: feasibility study. Radiology 2008; 246: 734–41. doi: 10.1148/radiol.2463070962 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Unno N, Nishiyama M, Suzuki M, Yamamoto N, Inuzuka K, Sagara D, et al. Quantitative lymph imaging for assessment of lymph function using indocyanine green fluorescence lymphography. Eur J Vasc Endovascular Surg 2008; 36: 230–6. doi: 10.1016/j.ejvs.2008.04.013 [DOI] [PubMed] [Google Scholar]
- 3.Darne C, Lu Y, Sevick-Muraca EM. Small animal fluorescence and bioluminescence tomography: a review of approaches, algorithms and technology update. Phys Med Biol 2014; 59: R1–64. doi: 10.1088/0031-9155/59/1/R1 [DOI] [PubMed] [Google Scholar]
- 4.Leblond F, Davis SC, Valdés PA, Pogue BW. Pre-clinical whole-body fluorescence imaging: Review of instruments, methods and applications. J Photochem Photobiol B: Biol 2010; 98: 77–94. doi: 10.1016/j.jphotobiol.2009.11.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Weissleder R, Ntziachristos V. Shedding light onto live molecular targets. Nat Med 2003; 9: 123–8. [DOI] [PubMed] [Google Scholar]
- 6.Gurfinkel M, Ke S, Wen X, Li C, Sevick-Muraca EM. Near-infrared fluorescence optical imaging and tomography. Dis Markers 2004; 19: 107–21. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Contag CH, Fraser S, Weissleder R. Strategies in in vivo molecular imaging. NeoReviews 2000; 1: e225–32. [Google Scholar]
- 8.Alander JT, Kaartinen I, Laakso A, Pätilä T, Spillmann T, Tuchin VV, et al. A review of indocyanine green fluorescent imaging in surgery. Int J Biomed Imaging 2012; 2012: 94085. doi: 10.1155/2012/940585 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Giraudeau C, Moussaron A, Stallivieri A, Mordon S, Frochot C. Indocyanine green: photosensitizer or chromophore? Still a debate. Curr Med Chem 2014; 21: 1871–97. [DOI] [PubMed] [Google Scholar]
- 10.Schaafsma BE, Mieog JSD, Hutteman M, Van der Vorst JR, Kuppen PJK, Löwik CW, et al. The clinical use of indocyanine green as a near-infrared fluorescent contrast agent for image-guided oncologic surgery. J Surg Oncol 2011; 104: 323–32. doi: 10.1002/jso.21943 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Sevick-Muraca EM, Lopez G, Reynolds JS, Troy TL, Hutchinson CL. Fluorescence and absorption contrast mechanisms for biomedical optical imaging using frequency-domain techniques. Photochem Photobiol 1997; 66: 55–64. [DOI] [PubMed] [Google Scholar]
- 12.Kaiser M, Yafi A, Cinat M, Choi B, Durkin AJ. Noninvasive assessment of burn wound severity using optical technology: a review of current and future modalities. Burns 2011; 37: 377–86. doi: 10.1016/j.burns.2010.11.012 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Polom K, Murawa D, Rho YS, Nowaczyk P, Hünerbein M, Murawa P. Current trends and emerging future of indocyanine green usage in surgery and oncology. Cancer 2011; 117: 4812–22. doi: 10.1002/cncr.26087 [DOI] [PubMed] [Google Scholar]
- 14.Marshall MV, Rasmussen JC, Tan IC, Aldrich MB, Adams KE, Wang X, et al. Near-infrared fluorescence imaging in humans with indocyanine green: a review and update. Open Surg Oncol J 2010; 2: 12–25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Liu HL, Pang SY, Chan YW. The use of a microscope with near-infrared imaging function in indocyanine green lymphography and lymphaticovenous anastomosis. J Plast Reconstr Aesthet Surg 2014; 67: 231–6. doi: 10.1016/j.bjps.2013.10.039 [DOI] [PubMed] [Google Scholar]
- 16.Ogata F, Narushima M, Mihara M, Azuma R, Morimoto Y, Koshima I. Intraoperative lymphography using indocyanine green dye for near-infrared fluorescence labeling in lymphedema. Ann Plast Surg 2007; 59: 180–4. [DOI] [PubMed] [Google Scholar]
- 17.Mihara M, Hara H, Kikuchi K, Yamamoto T, Iida T, Narushima M, et al. Scarless lymphatic venous anastomosis for latent and early-stage lymphoedema using indocyanine green lymphography and non-invasive instruments for visualising subcutaneous vein. J Plast Reconstr Aesthet Surg 2012; 65: 1551–8. doi: 10.1016/j.bjps.2012.05.026 [DOI] [PubMed] [Google Scholar]
- 18.Sevick-Muraca EM, Kwon S, Rasmussen JC. Emerging lymphatic imaging technologies for mouse and man. J Clin Invest 2014; 124: 905–14. doi: 10.1172/JCI71612 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Sevick-Muraca EM. Translation of near-infrared fluorescence imaging technologies: emerging clinical applications. Annu Rev Med 2012; 63: 217–31. doi: 10.1146/annurev-med-070910-083323 [DOI] [PubMed] [Google Scholar]
- 20.Mihara M, Hara H, Araki J, Kikuchi K, Narushima M, Yamamoto T, et al. Indocyanine green (ICG) lymphography is superior to lymphoscintigraphy for diagnostic imaging of early lymphedema of the upper limbs. PLoS One 2012; 7: e38182. doi: 10.1371/journal.pone.0038182 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Rasmussen JC, Tan IC, Marshall MV, Fife CE, Sevick-Muraca EM. Lymphatic imaging in humans with near-infrared fluorescence. Curr Opin Biotechnol 2009; 20: 74–82. doi: 10.1016/j.copbio.2009.01.009 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Rasmussen JC, Tan IC, Marshall MV, Adams KE, Kwon S, Fife CE, et al. Human lymphatic architecture and dynamic transport imaged using near-infrared fluorescence. Transl Oncol 2010; 3: 362–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Vahrmeijer AL, Hutteman M, van der Vorst JR, van de Velde CJH, Frangioni JV. Image-guided cancer surgery using near-infrared fluorescence. Nat Rev Clin Oncol 2013; 10: 507–18. doi: 10.1038/nrclinonc.2013 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Sevick-Muraca EM, Rasmussen JC. Molecular imaging with optics: primer and case for near-infrared fluorescence techniques in personalized medicine. J Biomed Opt 2008; 13: 041303. doi: 10.1117/1.2953185 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Zhu B, Rasmussen JC, Sevick-Muraca EM. A matter of collection and detection for intraoperative and noninvasive near-infrared fluorescence molecular imaging: to see or not to see? Med Phys 2014; 41: 022105. doi: 10.1118/1.4862514 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Zhu B, Rasmussen JC, Lu Y, Sevick-Muraca EM. Reduction of excitation light leakage to improve near-infrared fluorescence imaging for tissue surface and deep tissue imaging. Med Phys 2010; 37: 5961–70. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Zhu B, Tan IC, Rasmussen JC, Sevick-Muraca EM. Validating the sensitivity and performance of near-infrared fluorescence imaging and tomography devices using a novel solid phantom and measurement approach. Technol Cancer Res Treat 2012; 11: 95–104. [DOI] [PubMed] [Google Scholar]
- 28.Anzai Y, Blackwell KE, Hirschowitz SL, Rogers JW, Sato Y, Yuh WT, et al. Initial clinical experience with dextran-coated superparamagnetic iron oxide for detection of lymph node metastases in patients with head and neck cancer. Radiology 1994; 192: 709–15. [DOI] [PubMed] [Google Scholar]
- 29.Wallace AM, Han LK, Povoski SP, Deck K, Schneebaum S, Hall NC, et al. Comparative evaluation of [(99m)tc] tilmanocept for sentinel lymph node mapping in breast cancer patients: results of two phase 3 trials. Ann Surg Oncol 2013; 20: 2590–9. doi: 10.1245/s10434-013-2887-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Hall MA, Pinkston KL, Wilganowski N, Robinson H, Ghosh P, Azhdarinia A, et al. Comparison of mAbs targeting epithelial cell adhesion molecule for the detection of prostate cancer lymph node metastases with multimodal contrast agents: quantitative small-animal PET/CT and NIRF. J Nucl Med 2012; 53: 1427–37. doi: 10.2967/jnumed.112.106302 [DOI] [PubMed] [Google Scholar]
- 31.Heath CH, Deep NL, Sweeny L, Zinn KR, Rosenthal EL. Use of panitumumab-IRDye800 to image microscopic head and neck cancer in an orthotopic surgical model. Ann Surg Oncol 2012; 19: 3879–87. doi: 10.1245/s10434-012-2435-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Gao P, Pinkston KL, Wilganowski N, Robinson H, Zhu B, Sevick-Muraca EM, et al. Deglycosylation of mAb by EndoS for improved molecular imaging sensitivity. Mol Imaging Biol; in press. Epub ahead of print August 2014. doi: 10.1007/s11307-014-0781-9 [DOI] [PubMed]
- 33.Kitai T, Inomoto T, Miwa M, Shikayama T. Fluorescence navigation with indocyanine green for detecting sentinel lymph nodes in breast cancer. Breast Cancer 2005; 12: 211–15. [DOI] [PubMed] [Google Scholar]
- 34.Tagaya N, Yamazaki R, Nakagawa A, Abe A, Hamada K, Kubota K, et al. Intraoperative identification of sentinel lymph nodes by near-infrared fluorescence imaging in patients with breast cancer. Am J Surg 2008; 195: 850–3. doi: 10.1016/j.amjsurg.2007.02.032 [DOI] [PubMed] [Google Scholar]
- 35.Murawa D, Hirche C, Dresel S, Hünerbein M. Sentinel lymph node biopsy in breast cancer guided by indocyanine green fluorescence. Br J Surg 2009; 96: 1289–94. doi: 10.1002/bjs.6721 [DOI] [PubMed] [Google Scholar]
- 36.Troyan SL, Kianzad V, Gibbs-Strauss SL, Gioux S, Matsui A, Oketokoun R, et al. The FLARE intraoperative near-infrared fluorescence imaging system: a first-in-human clinical trial in breast cancer sentinel lymph node mapping. Ann Surg Oncol 2009; 16: 2943–52. doi: 10.1245/s10434-009-0594-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Sugie T, Kassim KA, Takeuchi M, Hashimoto T, Yamagami K, Masai Y, et al. A novel method for sentinel lymph node biopsy by indocyanine green fluorescence technique in breast cancer. Cancers 2010; 2: 713–20. doi: 10.3390/cancers2020713 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Hirche C, Murawa D, Mohr Z, Kneif S, Hünerbein M. ICG fluorescence-guided sentinel node biopsy for axillary nodal staging in breast cancer. Breast Cancer Res Treat 2010; 121: 373–8. doi: 10.1007/s10549-010-0760-z [DOI] [PubMed] [Google Scholar]
- 39.Hojo T, Nagao T, Kikuyama M, Akashi S, Kinoshita T. Evaluation of sentinel node biopsy by combined fluorescent and dye method and lymph flow for breast cancer. Breast 2010; 19: 210–13. doi: 10.1016/j.breast.2010.01.014 [DOI] [PubMed] [Google Scholar]
- 40.Tagaya N, Aoyagi H, Nakagawa A, Abe A, Iwasaki Y, Tachibana M, et al. A novel approach for sentinel lymph node identification using fluorescence imaging and image overlay navigation surgery in patients with breast cancer. World J Surg 2011; 35: 154–8. doi: 10.1007/s00268-010-0811-y [DOI] [PubMed] [Google Scholar]
- 41.Mieog JSD, Troyan SL, Hutteman M, Donohoe KJ, van der Vorst JR, Stockdale A, et al. Toward optimization of imaging system and lymphatic tracer for near-infrared fluorescent sentinel lymph node mapping in breast cancer. Ann Surg Oncol 2011; 18: 2483–91. doi: 10.1245/s10434-011-1566-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Hutteman M, Mieog JS, van der Vorst JR, Liefers GJ, Putter H, Löwik CW, et al. Randomized, double-blind comparison of indocyanine green with or without albumin premixing for near-infrared fluorescence imaging of sentinel lymph nodes in breast cancer patients. Breast Cancer Res Treat 2011; 127: 163–70. doi: 10.1007/s10549-011-1419-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Abe H, Mori T, Umeda T, Tanaka M, Kawai Y, Shimizu T, et al. Indocyanine green fluorescence imaging system for sentinel lymph node biopsies in early breast cancer patients. Surg Today 2011; 41: 197–202. [DOI] [PubMed] [Google Scholar]
- 44.Aoyama K, Kamio T, Ohchi T, Nishizawa M, Kameoka S. Sentinel lymph node biopsy for breast cancer patients using fluorescence navigation with indocyanine green. World J Surg Oncol 2011; 9: 157. doi: 10.1186/1477-7819-9-157 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.van der Vorst JR, Schaafsma BE, Verbeek FP, Hutteman M, Mieog JS, Lowik CW, et al. Randomized comparison of near-infrared fluorescence imaging using indocyanine green and 99m technetium with or without patent blue for the sentinel lymph node procedure in breast cancer patients. Ann Surg Oncol 2012; 19: 4104–11. doi: 10.1245/s10434-012-2466-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Wishart GC, Loh SW, Jones L, Benson JR. A feasibility study (ICG-10) of indocyanine green (ICG) fluorescence mapping for sentinel lymph node detection in early breast cancer. Eur J Surg Oncol 2012; 38: 651–6. doi: 10.1016/j.ejso.2012.05.007 [DOI] [PubMed] [Google Scholar]
- 47.Hirano A, Kamimura M, Ogura K, Kim N, Hattori A, Setoguchi Y, et al. A comparison of indocyanine green fluorescence imaging plus blue dye and blue dye alone for sentinel node navigation surgery in breast cancer patients. Ann Surg Oncol 2012; 19: 4112–16. doi: 10.1245/s10434-012-2478-0 [DOI] [PubMed] [Google Scholar]
- 48.Takeuchi M, Sugie T, Abdelazeem K, Kato H, Shinkura N, Takada M, et al. Lymphatic mapping with fluorescence navigation using indocyanine green and axillary surgery in patients with primary breast cancer. Breast J 2012; 18: 535–41. doi: 10.1111/tbj.12004 [DOI] [PubMed] [Google Scholar]
- 49.Polom K, Murawa D, Nowaczyk P, Rho YS, Murawa P. Breast cancer sentinel lymph node mapping using near infrared guided indocyanine green and indocyanine green–human serum albumin in comparison with gamma emitting radioactive colloid tracer. Eur J Surg Oncol 2012; 38: 137–42. doi: 10.1016/j.ejso.2011.11.004 [DOI] [PubMed] [Google Scholar]
- 50.Sugie T, Sawada T, Tagaya N, Kinoshita T, Yamagami K, Suwa H, et al. Comparison of the indocyanine green fluorescence and blue dye methods in detection of sentinel lymph nodes in early-stage breast cancer. Ann Surg Oncol 2013; 20: 2213–18. doi: 10.1245/s10434-013-2890-0 [DOI] [PubMed] [Google Scholar]
- 51.Schaafsma BE, Verbeek FP, Rietbergen DD, van der Hiel B, van der Vorst JR, Liefers GJ, et al. Clinical trial of combined radio-and fluorescence-guided sentinel lymph node biopsy in breast cancer. Br J Surg 2013; 100: 1037–44. doi: 10.1002/bjs.9159 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Jung SY, Kim SK, Kim SW, Kwon Y, Lee ES, Kang HS, et al. Comparison of sentinel lymph node biopsy guided by the multimodal method of indocyanine green fluorescence, radioisotope, and blue dye versus the radioisotope method in breast cancer: a randomized controlled trial. Ann Surg Oncol 2013; 21: 1254–9. doi: 10.1245/s10434-013-3437-0 [DOI] [PubMed] [Google Scholar]
- 53.Guo W, Zhang L, Ji J, Gao W, Liu J, Tong M. Breast cancer sentinel lymph node mapping using near-infrared guided indocyanine green in comparison with blue dye. Tumor Biol 2014; 35: 3073–8. doi: 10.1007/s13277-013-1399-2 [DOI] [PubMed] [Google Scholar]
- 54.Chi C, Ye J, Ding H, He D, Huang W, Zhang G-J, et al. Use of indocyanine green for detecting the sentinel lymph node in breast cancer patients: from preclinical evaluation to clinical validation. PLoS One 2013; 8: e83927. doi: 10.1371/journal.pone.0083927 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Ballardini B, Santoro L, Sangalli C, Gentilini O, Renne G, Lissidini G, et al. The indocyanine green method is equivalent to the 99mTc-labeled radiotracer method for identifying the sentinel node in breast cancer: a concordance and validation study. Eur J Surg Oncol 2013; 39: 1332–6. doi: 10.1016/j.ejso.2013.10.004 [DOI] [PubMed] [Google Scholar]
- 56.Verbeek FP, Troyan SL, Mieog JS, Liefers GJ, Moffitt LA, Rosenberg M, et al. Near-infrared fluorescence sentinel lymph node mapping in breast cancer: a multicenter experience. Breast Cancer Res Treat 2014; 143: 333–42. doi: 10.1007/s10549-013-2802-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Azhdarinia A, Ghosh P, Ghosh S, Wilganowski N, Sevick-Muraca EM. Dual-labeling strategies for nuclear and fluorescence molecular imaging: a review and analysis. Mol Imaging Biol 2011; 14: 261–76. doi: 10.1007/s11307-011-0528-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Fujiwara M, Mizukami T, Suzuki A, Fukamizu H. Sentinel lymph node detection in skin cancer patients using real-time fluorescence navigation with indocyanine green: preliminary experience. J Plast Reconstr Aesthet Surg 2009; 62: e373–8. doi: 10.1016/j.bjps.2007.12.074 [DOI] [PubMed] [Google Scholar]
- 59.Tsujino Y, Mizumoto K, Matsuzaka Y, Niihara H, Morita E. Fluorescence navigation with indocyanine green for detecting sentinel nodes in extramammary Paget's disease and squamous cell carcinoma. J Dermatol 2009; 36: 90–4. doi: 10.1111/j.1346-8138.2009.00595.x [DOI] [PubMed] [Google Scholar]
- 60.Tanaka R, Nakashima K, Fujimoto W. Sentinel lymph node detection in skin cancer using fluorescence navigation with indocyanine green. J Dermatol 2009; 36: 468–70. doi: 10.1111/j.1346-8138.2009.00679.x [DOI] [PubMed] [Google Scholar]
- 61.Mizukami T, Fujiwara M, Suzuki A, Nagata T, Fukamizu H. Sentinel lymph node detection by indocyanine green fluorescence imaging in skin cancer patients: technical refinement. Open Surg Oncol J 2010; 2: 57–61. [Google Scholar]
- 62.Namikawa K, Yamazaki N. Sentinel lymph node biopsy guided by indocyanine green fluorescence for cutaneous melanoma. Eur J Dermatol 2011; 21: 184–90. doi: 10.1684/ejd.2010.1237 [DOI] [PubMed] [Google Scholar]
- 63.Fujisawa Y, Nakamura Y, Kawachi Y, Otsuka F. A custom-made, low-cost intraoperative fluorescence navigation system with indocyanine green for sentinel lymph node biopsy in skin cancer. Dermatology 2011; 222: 261–8. doi: 10.1159/000327080 [DOI] [PubMed] [Google Scholar]
- 64.Hayashi T, Furukawa H, Oyama A, Funayama E, Saito A, Yamao T, et al. Sentinel lymph node biopsy using real-time fluorescence navigation with indocyanine green in cutaneous head and neck/lip mucosa melanomas. Head Neck 2012; 34: 758–61. doi: 10.1002/hed.21651 [DOI] [PubMed] [Google Scholar]
- 65.Stoffels I, von der Stück H, Boy C, Pöppel T, Körber N, Weindorf M, et al. Indocyanine green fluorescence-guided sentinel lymph node biopsy in dermato-oncology. J Dtsch Dermatol Ges 2012; 10: 51–7. doi: 10.1111/j.1610-0387.2011.07843.x [DOI] [PubMed] [Google Scholar]
- 66.Polom K, Murawa D, Rho YS, Spychala A, Murawa P. Skin melanoma sentinel lymph node biopsy using real-time fluorescence navigation with indocyanine green and indocyanine green with human serum albumin. Br J Dermatol 2012; 166: 682–3. doi: 10.1111/j.1365-2133.2011.10634.x [DOI] [PubMed] [Google Scholar]
- 67.Fujisawa Y, Nakamura Y, Kawachi Y, Otsuka F. Indocyanine green fluorescence-navigated sentinel node biopsy showed higher sensitivity than the radioisotope or blue dye method, which may help to reduce false-negative cases in skin cancer. J Surg Oncol 2012; 106: 41–5. doi: 10.1002/jso.23045 [DOI] [PubMed] [Google Scholar]
- 68.Rasmussen JC, Kwon S, Sevick-Muraca EM, Cormier JN. The role of lymphatics in cancer as assessed by near-infrared fluorescence imaging. Ann Biomed Eng 2012; 40: 408–21. doi: 10.1007/s10439-011-0476-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.van der Vorst JR, Schaafsma BE, Verbeek FP, Swijnenburg RJ, Hutteman M, Liefers GJ, et al. Dose optimization for near-infrared fluorescence sentinel lymph node mapping in patients with melanoma. Br J Dermatol 2013; 168: 93–8. doi: 10.1111/bjd.12059 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Namikawa K, Tsutsumida A, Tanaka R, Kato J, Yamazaki N. Limitation of indocyanine green fluorescence in identifying sentinel lymph node prior to skin incision in cutaneous melanoma. Int J Clin Oncol 2013; 19: 198–203. doi: 10.1007/s10147-013-0524-y [DOI] [PubMed] [Google Scholar]
- 71.Gilmore DM, Khullar OV, Gioux S, Stockdale A, Frangioni JV, Colson YL, et al. Effective low-dose escalation of indocyanine green for near-infrared fluorescent sentinel lymph node mapping in melanoma. Ann Surg Oncol 2013; 20: 2357–63. doi: 10.1245/s10434-013-2905-x [DOI] [PubMed] [Google Scholar]
- 72.Nakamura Y, Fujisawa Y, Nakamura Y, Maruyama H, Furuta J-i, Kawachi Y, et al. Improvement of the sentinel lymph node detection rate of cervical sentinel lymph node biopsy using real-time fluorescence navigation with indocyanine green in head and neck skin cancer. J Dermatol 2013; 40: 453–7. doi: 10.1111/1346-8138.12158 [DOI] [PubMed] [Google Scholar]
- 73.Miyashiro I, Miyoshi N, Hiratsuka M, Kishi K, Yamada T, Ohue M, et al. Detection of sentinel node in gastric cancer surgery by indocyanine green fluorescence imaging: comparison with infrared imaging. Ann Surg Oncol 2008; 15: 1640–3. doi: 10.1245/s10434-008-9872-7 [DOI] [PubMed] [Google Scholar]
- 74.Tajima Y, Yamazaki K, Masuda Y, Kato M, Yasuda D, Aoki T, et al. Sentinel node mapping guided by indocyanine green fluorescence imaging in gastric cancer. Ann Surg 2009; 249: 58–62. doi: 10.1097/SLA.0b013e3181927267 [DOI] [PubMed] [Google Scholar]
- 75.Tajima Y, Murakami M, Yamazaki K, Masuda Y, Kato M, Sato A, et al. Sentinel node mapping guided by indocyanine green fluorescence imaging during laparoscopic surgery in gastric cancer. Ann Surg Oncol 2010; 17: 1787–93. doi: 10.1245/s10434-010-0944-0 [DOI] [PubMed] [Google Scholar]
- 76.Miyashiro I, Kishi K, Yano M, Tanaka K, Motoori M, Ohue M, et al. Laparoscopic detection of sentinel node in gastric cancer surgery by indocyanine green fluorescence imaging. Surg Endosc 2011; 25: 1672–6. doi: 10.1007/s00464-010-1405-3 [DOI] [PubMed] [Google Scholar]
- 77.Yoshida M, Kubota K, Kuroda J, Ohta K, Nakamura T, Saito J, et al. Indocyanine green injection for detecting sentinel nodes using color fluorescence camera in the laparoscopy-assisted gastrectomy. J Gastroenterol Hepatol 2012; 27: 29–33. doi: 10.1111/j.1440-1746.2012.07067.x [DOI] [PubMed] [Google Scholar]
- 78.Watanabe M, Tsunoda A, Narita K, Kusano M, Miwa M. Colonic tattooing using fluorescence imaging with light-emitting diode-activated indocyanine green: a feasibility study. Surg Today 2009; 39: 214–18. doi: 10.1007/s00595-008-3849-9 [DOI] [PubMed] [Google Scholar]
- 79.Noura S, Ohue M, Seki Y, Tanaka K, Motoori M, Kishi K, et al. Feasibility of a lateral region sentinel node biopsy of lower rectal cancer guided by indocyanine green using a near-infrared camera system. Ann Surg Oncol 2010; 17: 144–51. doi: 10.1245/s10434-009-0711-2 [DOI] [PubMed] [Google Scholar]
- 80.Hirche C, Dresel S, Krempien R, Hünerbein M. Sentinel node biopsy by indocyanine green retention fluorescence detection for inguinal lymph node staging of anal cancer: preliminary experience. Ann Surg Oncol 2010; 17: 2357–62. doi: 10.1245/s10434-010-1010-7 [DOI] [PubMed] [Google Scholar]
- 81.Hirche C, Mohr Z, Kneif S, Doniga S, Murawa D, Strik M, et al. Ultrastaging of colon cancer by sentinel node biopsy using fluorescence navigation with indocyanine green. Int J Colorectal Dis 2012; 27: 319–24. doi: 10.1007/s00384-011-1306-5 [DOI] [PubMed] [Google Scholar]
- 82.Cahill RA, Anderson M, Wang LM, Lindsey I, Cunningham C, Mortensen NJ. Near-infrared (NIR) laparoscopy for intraoperative lymphatic road-mapping and sentinel node identification during definitive surgical resection of early-stage colorectal neoplasia. Surg Endosc 2012; 26: 197–204. doi: 10.1007/s00464-011-1854-3 [DOI] [PubMed] [Google Scholar]
- 83.Manny TB, Hemal AK. Fluorescence-enhanced robotic radical cystectomy using unconjugated indocyanine green for pelvic lymphangiography, tumor marking, and mesenteric angiography: the initial clinical experience. Urology 2014; 83: 824–30. doi: 10.1016/j.urology.2013.11.042 [DOI] [PubMed] [Google Scholar]
- 84.Manny TB, Patel M, Hemal AK. Fluorescence-enhanced robotic radical prostatectomy using real-time lymphangiography and tissue marking with percutaneous injection of unconjugated indocyanine green: the initial clinical experience in 50 patients. Eur Urol 2014; 65: 1162–8. doi: 10.1016/j.eururo.2013.11.017 [DOI] [PubMed] [Google Scholar]
- 85.Furukawa N, Oi H, Yoshida S, Shigetomi H, Kanayama S, Kobayashi H. The usefulness of photodynamic eye for sentinel lymph node identification in patients with cervical cancer. Tumori 2010; 96: 936–40. [PubMed] [Google Scholar]
- 86.Crane LM, Themelis G, Pleijhuis RG, Harlaar NJ, Sarantopoulos A, Arts HJ, et al. Intraoperative multispectral fluorescence imaging for the detection of the sentinel lymph node in cervical cancer: a novel concept. Mol Imaging Biol 2011; 13: 1043–9. doi: 10.1007/s11307-010-0425-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.van der Vorst JR, Hutteman M, Gaarenstroom KN, Peters AA, Mieog JS, Schaafsma BE, et al. Optimization of near-infrared fluorescent sentinel lymph node mapping in cervical cancer patients. Int J Gynecol Cancer 2011; 21: 1472. doi: 10.1097/IGC.0b013e31822b451d [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88.Rossi EC, Ivanova A, Boggess JF. Robotically assisted fluorescence-guided lymph node mapping with ICG for gynecologic malignancies: a feasibility study. Gynecol Oncol 2012; 124: 78–82. doi: 10.1016/j.ygyno.2011.09.025 [DOI] [PubMed] [Google Scholar]
- 89.Schaafsma BE, van der Vorst JR, Gaarenstroom KN, Peters AA, Verbeek FP, de Kroon CD, et al. Randomized comparison of near-infrared fluorescence lymphatic tracers for sentinel lymph node mapping of cervical cancer. Gynecol Oncol 2012; 127: 126–30. doi: 10.1016/j.ygyno.2012.07.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Crane LM, Themelis G, Arts HJ, Buddingh KT, Brouwers AH, Ntziachristos V, et al. Intraoperative near-infrared fluorescence imaging for sentinel lymph node detection in vulvar cancer: first clinical results. Gynecol Oncol 2011; 120: 291–5. doi: 10.1016/j.ygyno.2010.10.009 [DOI] [PubMed] [Google Scholar]
- 91.Hutteman M, Van Der Vorst JR, Gaarenstroom KN, Peters AAW, Mieog JSD, Schaafsma BE, et al. Optimization of near-infrared fluorescent sentinel lymph node mapping for vulvar cancer. Am J Obstetrics Gynecol 2012; 206: 89.e1–5. doi: 10.1016/j.ajog.2011.07.039 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92.Holloway RW, Bravo RA, Rakowski JA, James JA, Jeppson CN, Ingersoll SB, et al. Detection of sentinel lymph nodes in patients with endometrial cancer undergoing robotic-assisted staging: a comparison of colorimetric and fluorescence imaging. Gynecol Oncol 2012; 126: 25–9. doi: 10.1016/j.ygyno.2012.04.009 [DOI] [PubMed] [Google Scholar]
- 93.Rossi EC, Jackson A, Ivanova A, Boggess JF. Detection of sentinel nodes for endometrial cancer with robotic assisted fluorescence imaging: cervical versus hysteroscopic injection. Int J Gynecol Cancer 2013; 23: 1704–11. doi: 10.1097/IGC.0b013e3182a616f6 [DOI] [PubMed] [Google Scholar]
- 94.Jewell EL, Huang JJ, Abu-Rustum NR, Gardner GJ, Brown CL, Sonoda Y, et al. Detection of sentinel lymph nodes in minimally invasive surgery using indocyanine green and near-infrared fluorescence imaging for uterine and cervical malignancies. Gynecol Oncol 2014; 133: 274–7. doi: 10.1016/j.ygyno.2014.02.028 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95.Bredell MG. Sentinel lymph node mapping by indocyanin green fluorescence imaging in oropharyngeal cancer-preliminary experience. Head Neck Oncol 2010; 2: 31. doi: 10.1186/1758-3284-2-31 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.van den Berg NS, Brouwer OR, Klop WM, Karakullukcu B, Zuur CL, Tan IB, et al. Concomitant radio-and fluorescence-guided sentinel lymph node biopsy in squamous cell carcinoma of the oral cavity using ICG-99mTc-nanocolloid. Eur J Nucl Med Mol Imaging 2012; 39: 1128–36. doi: 10.1007/s00259-012-2129-5 [DOI] [PubMed] [Google Scholar]
- 97.van der Vorst JR, Schaafsma BE, Verbeek FP, Keereweer S, Jansen JC, van der Velden LA, et al. Near-infrared fluorescence sentinel lymph node mapping of the oral cavity in head and neck cancer patients. Oral Oncol 2013; 49: 15–19. doi: 10.1016/j.oraloncology.2012.07.017 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Yamashita S, Tokuishi K, Anami K, Miyawaki M, Moroga T, Kamei M, et al. Video-assisted thoracoscopic indocyanine green fluorescence imaging system shows sentinel lymph nodes in non-small-cell lung cancer. J Thorac Cardiovasc Surg 2011; 141: 141–4. doi: 10.1016/j.jtcvs.2010.01.028 [DOI] [PubMed] [Google Scholar]
- 99.Gilmore DM, Khullar OV, Jaklitsch MT, Chirieac LR, Frangioni JV, Colson YL. Identification of metastatic nodal disease in a phase 1 dose-escalation trial of intraoperative sentinel lymph node mapping in non-small cell lung cancer using near-infrared imaging. J Thorac Cardiovasc Surg 2013; 146: 562–70. doi: 10.1016/j.jtcvs.2013.04.010 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100.Unno N, Inuzuka K, Suzuki M, Yamamoto N, Sagara D, Nishiyama M, et al. Preliminary experience with a novel fluorescence lymphography using indocyanine green in patients with secondary lymphedema. J Vasc Surg 2007; 45: 1016–21. [DOI] [PubMed] [Google Scholar]
- 101.Ogasawara Y, Ikeda H, Takahashi M, Kawasaki K, Doihara H. Evaluation of breast lymphatic pathways with indocyanine green fluorescence imaging in patients with breast cancer. World J Surg 2008; 32: 1924–9. doi: 10.1007/s00268-008-9519-7 [DOI] [PubMed] [Google Scholar]
- 102.Suzuki M, Unno N, Yamamoto N, Nishiyama M, Sagara D, Tanaka H, et al. Impaired lymphatic function recovered after great saphenous vein stripping in patients with varicose vein: venodynamic and lymphodynamic results. J Vasc Surg 2009; 50: 1085–91. doi: 10.1016/j.jvs.2009.06.003 [DOI] [PubMed] [Google Scholar]
- 103.Kamiya K, Unno N, Konno H. Intraoperative indocyanine green fluorescence lymphography, a novel imaging technique to detect a chyle fistula after an esophagectomy: report of a case. Surg Today 2009; 39: 421–4. doi: 10.1007/s00595-008-3852-1 [DOI] [PubMed] [Google Scholar]
- 104.Unno N, Nishiyama M, Suzuki M, Tanaka H, Yamamoto N, Sagara D, et al. A novel method of measuring human lymphatic pumping using indocyanine green fluorescence lymphography. J Vasc Surg 2010; 52: 946–52. doi: 10.1016/j.jvs.2010.04.067 [DOI] [PubMed] [Google Scholar]
- 105.Adams KE, Rasmussen JC, Darne C, Tan I, Aldrich MB, Marshall MV, et al. Direct evidence of lymphatic function improvement after advanced pneumatic compression device treatment of lymphedema. Biomed Opt Express 2010; 1: 114–25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Tan I, Maus EA, Rasmussen JC, Marshall MV, Adams KE, Fife CE, et al. Assessment of lymphatic contractile function after manual lymphatic drainage using near-infrared fluorescence imaging. Arch Phys Med Rehabil 2011; 92: 756–64.e1. doi: 10.1016/j.apmr.2010.12.027 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107.Furukawa H, Osawa M, Saito A, Hayashi T, Funayama E, Oyama A, et al. Microsurgical lymphaticovenous implantation targeting dermal lymphatic backflow using indocyanine green fluorescence lymphography in the treatment of postmastectomy lymphedema. Plast Reconstr Surg 2011; 127: 1804–11. doi: 10.1097/PRS.0b013e31820cf2e2 [DOI] [PubMed] [Google Scholar]
- 108.Mukenge SM, Catena M, Negrini D, Ratti F, Moriondo A, Briganti A, et al. Assessment and follow-up of patency after lymphovenous microsurgery for treatment of secondary lymphedema in external male genital organs. Eur Urol 2011; 60: 1114–19. doi: 10.1016/j.eururo.2010.11.020 [DOI] [PubMed] [Google Scholar]
- 109.Yamamoto T, Yamamoto N, Doi K, Oshima A, Yoshimatsu H, Todokoro T, et al. Indocyanine green-enhanced lymphography for upper extremity lymphedema: a novel severity staging system using dermal backflow patterns. Plast Reconstr Surg 2011; 128: 941–7. doi: 10.1097/PRS.0b013e3182268cd9 [DOI] [PubMed] [Google Scholar]
- 110.Yamamoto T, Narushima M, Doi K, Oshima A, Ogata F, Mihara M, et al. Characteristic indocyanine green lymphography findings in lower extremity lymphedema: the generation of a novel lymphedema severity staging system using dermal backflow patterns. Plast Reconstr Surg 2011; 127: 1979–86. doi: 10.1097/PRS.0b013e31820cf5df [DOI] [PubMed] [Google Scholar]
- 111.Yamamoto T, Matsuda N, Doi K, Oshima A, Yoshimatsu H, Todokoro T, et al. The earliest finding of indocyanine green lymphography in asymptomatic limbs of lower extremity lymphedema patients secondary to cancer treatment: the modified dermal backflow stage and concept of subclinical lymphedema. Plast Reconstr Surg 2011; 128: 314e-21. doi: 10.1097/PRS.0b013e3182268da8 [DOI] [PubMed] [Google Scholar]
- 112.Unno N, Tanaka H, Suzuki M, Yamamoto N, Mano Y, Sano M, et al. Influence of age and gender on human lymphatic pumping pressure in the leg. Lymphology 2011; 44: 113–20. [PubMed] [Google Scholar]
- 113.Yamamoto T, Lida T, Matsuda N, Kikuchi K, Yoshimatsu H, Mihara M, et al. Indocyanine green (ICG)-enhanced lymphography for evaluation of facial lymphoedema. J Plast Reconstr Aesthet Surg 2011; 64: 1541–4. doi: 10.1016/j.bjps.2011.05.025 [DOI] [PubMed] [Google Scholar]
- 114.Aldrich MB, Guilliod R, Fife CE, Maus EA, Smith L, Rasmussen JC, et al. Lymphatic abnormalities in the normal contralateral arms of subjects with breast cancer-related lymphedema as assessed by near-infrared fluorescent imaging. Biomed Opt Express 2012; 3: 1256–65. doi: 10.1364/BOE.3.001256 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115.Maus EA, Tan IC, Rasmussen JC, Marshall MV, Fife CE, Smith LA, et al. Near-infrared fluorescence imaging of lymphatics in head and neck lymphedema. Head Neck 2012; 34: 448–53. doi: 10.1002/hed.21538 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 116.Akita S, Mitsukawa N, Kazama T, Kuriyama M, Kubota Y, Omori N, et al. Comparison of lymphoscintigraphy and indocyanine green lymphography for the diagnosis of extremity lymphoedema. J Plast Reconstr Aesthet Surg 2013; 66: 792–8. doi: 10.1016/j.bjps.2013.02.023 [DOI] [PubMed] [Google Scholar]
- 117.Sakurai T, Endo M, Shimizu K, Yoshimizu N, Nakajima K, Nosaka K, et al. Axillary reverse mapping using fluorescence imaging is useful for identifying the risk group of postoperative lymphedema in breast cancer patients undergoing sentinel node biopsies. J Surg Oncol 2013; 109: 612–15. doi: 10.1002/jso.23528 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 118.Burrows PE, Gonzalez-Garay ML, Rasmussen JC, Aldrich MB, Guilliod R, Maus EA, et al. Lymphatic abnormalities are associated with RASA1 gene mutations in mouse and man. Proc Natl Acad Sci U S A 2013; 110: 8621–6. doi: 10.1073/pnas.1222722110 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 119.Mihara M, Seki Y, Hara H, Iida T, Oka A, Kikuchi K, et al. Predictive lymphatic mapping: a method for mapping lymphatic channels in patients with advanced unilateral lymphedema using indocyanine green lymphography. Ann Plast Surg 2014; 72: 706–10. doi: 10.1097/SAP.0b013e31826a18b1 [DOI] [PubMed] [Google Scholar]
- 120.Shibasaki J, Hara H, Mihara M, Adachi S, Uchida Y, Itani Y. Evaluation of lymphatic dysplasia in patients with congenital pleural effusion and ascites using indocyanine green lymphography. J Pediatr 2014; 164: 1116–20.e1. doi: 10.1016/j.jpeds.2013.12.052 [DOI] [PubMed] [Google Scholar]
- 121.Tan IC, Balaguru D, Rasmussen JC, Guilliod R, Bricker JT, Douglas WI, et al. Investigational lymphatic imaging at the bedside in a pediatric postoperative chylothorax patient. Pediatr Cardiol 2014; 35: 1295–300. doi: 10.1007/s00246-014-0946-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 122.Akita S, Mitsukawa N, Rikihisa N, Kubota Y, Omori N, Mitsuhashi A, et al. Early diagnosis and risk factors for lymphedema following lymph node dissection for gynecologic cancer. Plast Reconstr Surg 2013; 131: 283–90. doi: 10.1097/PRS.0b013e318277870f [DOI] [PubMed] [Google Scholar]
- 123.Noguchi M, Noguchi M, Nakano Y, Ohno Y, Kosaka T. Axillary reverse mapping using a fluorescence imaging system in breast cancer. J Surg Oncol 2012; 105: 229–34. doi: 10.1002/jso.22094 [DOI] [PubMed] [Google Scholar]
- 124.Ikeda K, Ogawa Y, Komatsu H, Mori Y, Ishikawa A, Nakajima T, et al. Evaluation of the metastatic status of lymph nodes identified using axillary reverse mapping in breast cancer patients. World J Surg Oncol 2012; 10: 233. doi: 10.1186/1477-7819-10-233 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 125.Meric-Bernstam F, Rasmussen JC, Krishnamurthy S, Tan I, Zhu B, Wagner JL, et al. Toward nodal staging of axillary lymph node basins through intradermal administration of fluorescent imaging agents. Biomed Opt Express 2014; 5: 183–96. doi: 10.1364/BOE.5.000183 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 126.Boehm T, Hochmuth A, Malich A, Reichenbach JR, Fleck M, Kaiser WA. Contrast-enhanced near-infrared laser mammography with a prototype breast scanner: feasibility study with tissue phantoms and preliminary results of imaging experimental tumors. Invest Radiol 2001; 36: 573–81. [DOI] [PubMed] [Google Scholar]
- 127.Grand AM, Lomnes SJ, Ohnishi S, Morgan TG, Gogbashian A, Frangioni JV, et al. Tissue-like phantoms for near-infrared fluorescence imaging system assessment and the training of surgeons. J Biomed Opt 2006; 11: 014007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 128.Baeten J, Niedre M, Dunham J, Ntziachristos V. Development of fluorescent materials for diffuse fluorescence tomography standards and phantoms. Opt Express 2007; 15: 8681–94. [DOI] [PubMed] [Google Scholar]
- 129.Pleijhuis RG, Langhout GC, Helfrich W, Themelis G, Sarantopoulos A, Crane LM, et al. Near-infrared fluorescence (NIRF) imaging in breast-conserving surgery: assessing intraoperative techniques in tissue-simulating breast phantoms. Eur J Surg Oncol 2011; 37: 32–9. doi: 10.1016/j.ejso.2010.10.006 [DOI] [PubMed] [Google Scholar]