Abstract
Cascaded linear systems based modeling techniques have been used in the past to predict important system parameters that have a direct impact on image quality. Such models are also useful in optimizing system parameters to improve image quality. In this work, detailed analysis of a solid-state fluoroscopic imaging system intended for high-resolution angiography is presented with the use of such a model. The imaging system analyzed through this model uses four 8×8 cm three-side buttable interlined charge-coupled devices (CCDs) specifically designed for high-resolution angiography and tiled in a seamless fashion to achieve a field of view (FOV) of 16×16 cm. Larger FOVs can be achieved by tiling more CCDs in a similar manner. The system employs a CsI:Tl scintillator coupled to the CCDs by straight (nontapering) fiberoptics and can potentially be operated in 78, 156, or 234 μm pixel pitch modes. The system parameters analyzed through this model include presampling modulation transfer function, noise power spectrum, and detective quantum efficiency (DQE). The results of the simulations performed indicate that DQE(0) in excess of 0.6 is achievable, with the imager operating at 156 μm pixel pitch, 30 frames/s, and employing a 450-μm-thick CsI:Tl scintillator, even at a low fluoroscopic exposure rate of 1 μR/frame. Further, at a nominal fluoroscopic exposure rate of 2.5 μR/frame there was no noticeable degradation of the DQE even at the 78 μm pixel pitch mode suggesting that it is feasible to perform high-resolution angiography hitherto unattainable in clinical practice.
Keywords: fluoroscopy, angiography, charge-coupled devices, cascaded linear systems analysis, detective quantum efficiency (DQE)
I. INTRODUCTION
A cascaded linear-systems-based model was used to analyze the processes that govern the output image quality of the solid-state high-resolution fluoroscopic system. With the assumptions that the proposed system is linear and shift-invariant, the model was used to describe the output image quality in terms of the objective measure, detective quantum efficiency (DQE). Cascaded linear-systems-based modeling techniques have been used to predict imaging performance of systems developed for x-ray imaging.1–5 Such models have been used to investigate key objective parameters of image quality such as the Wiener spectrum or noise power spectrum (NPS),6–8 noise equivalent quanta,7,9 and detective quantum efficiency (DQE).7,9,10 Over the past several years, many aspects detailing the development of the theory, modeling techniques and its application have been described.11–17 Theoretical calculations based on the cascaded model for evaluating charge-coupled device (CCD)-based imagers with slightly varying design and application have been published.10,18,19
The imaging chain is represented as a cascade of amplifying and scattering mechanisms,13 with the assumption that the system is linear and shift invariant.7,12 While the assumption of linearity can be largely supported, as CCD-based imaging systems exhibit linear signal dependence with incident exposure over most of their dynamic range, the assumption that the system is spatially stationary can be made only in a wide sense as detector nonuniformities exist. For fluoroscopic applications, image lag also limits the validity of such an assumption in the temporal domain.20 Hence, the description of image noise in terms of the spatiotemporal NPS, which incorporates the image lag, has been sought.20 Cunningham et al.20 have shown experimentally and theoretically that “the spatial component of the spatiotemporal DQE of a system operating in the fluoroscopic mode is the same as the conventional DQE of the same system operating in the radiographic mode under quantum-noise-limited conditions.” Analysis of the system parameters in the fluoroscopic mode was performed using this finding so that a single fluoroscopic frame is considered as essentially a radiographic acquisition with an exposure level corresponding to that typically used in fluoroscopy.
II. METHODS AND MATERIAL
The system analyzed through the cascaded linear systems-based approach consists of four, three-side buttable, 8×8 cm large-area interlined CCDs specifically designed for high-resolution fluoroscopy, coupled to a structured CsI:Tl scintillator by a straight (nontapering) fiberoptic plate. Each CCD has a 2048×2048 pixel matrix with a fundamental pixel pitch of 39 μm. The CCDs can potentially be operated in three different pixel pitch modes of 78, 156 and 234 μm, resulting in Nyquist sampling limits of 6.4, 3.2, and 2.1 cycles/mm, respectively. The variable pixel pitch is achieved by grouping (binning) 2×2, 4×4, and 6×6 adjacent pixels, respectively. This CCD with a unique architecture using interline channel (hereafter referred to as the interline CCD) was specifically designed for high-resolution fluoroscopic imaging. The interline channel is opaque to light resulting in degradation of the fill factor. The width of the interline channel is 19 μm and traverses the length of the pixel, resulting in an active area of 20×39 μm for each fundamental pixel. This results in a fill factor of ~51%, independent of the pixel pitch mode. Thallium-doped cesium iodide (CsI:Tl) has been selected as the scintillator of choice, as it has the capability to maintain high spatial resolution due to its structured columnar arrangement. CsI:Tl scintillator thickness of 300–525 μm were considered in this analysis. Optical coupling between the CCD and the CsI:Tl scintillator is achieved by using a 1 in. (2.54 cm) thick nontapering (1:1) fiberoptic plate.
The system was considered as a cascade of discrete stages, which can be represented by one of the following processes: quantum gain, stochastic blurring, or deterministic blurring. For any given stage i, the input and output signal are represented as and in the spatial coordinates of (x,y), the average input and output signal are represented as qi–1 and qi, and the input and output signal are represented as qi–1(u,v) and qi(u,v) in its orthogonal spatial frequency coordinates of (u,v). Similarly, for any given stage i, the input and output Wiener spectrum (NPS) are represented in the spatial frequency coordinates as Wi–1(u,v) and Wi(u,v). For any quantum gain stage i, the average quantum gain of that stage is represented as and its gain variance represented as . The gain variance can also be expressed in terms of the Poisson excess,15 εgi, or in terms of the Swank factor,21 AS. For any blurring stage i, the blur is characterized by its modulation transfer function (MTF) in spatial frequency coordinates and represented as Ti(u,v). The additive noise imparted by any given stage i, is represented in the spatial frequency coordinates as W addi(u,v). Signal and noise transfer relationships from the input to the output for gain and blurring stages have been provided in the past.2,13,15,17
The system was analyzed by modeling the imaging chain into the following elementary stages:
Incident image quanta
Attenuation of x rays by the CsI:Tl scintillator
Generation and emission of optical quanta by the CsI:Tl scintillator
Optical blurring by the CsI:Tl scintillator
Coupling of the optical quanta by straight (nontapering) fiberoptics
Absorption of optical quanta by the CCD
Pixel presampling MTF and fill factor
NPS aliasing
Additive noise
Thus the model provides the presampling signal and the aliased NPS. DQE measurements performed on digital imaging systems are based on the presampling signal (MTF) and the aliased NPS. Hence, the effects of noise aliasing have also been included. This is of particular importance to this imager as the system can be operated in any of the three pixel pitch modes and hence, their impact on the aliased NPS need to be addressed. Analysis of system performance was performed for the three pixel pitch modes of 78, 156, and 234 μm and for 300-, 375-, 450-, and 525-μm-thick CsI:Tl scintillators.
Absorption of K-fluorescent x rays cause a spatial blur and result in spatial correlation between the primary interaction site and the reabsorption site.16 This results in parallel pathways for signal and noise transfer within the scintillator medium. Cunningham et al.16 used the term “parallel cascade” to describe this process in his model. Cunningham16,22 and Yao23 have provided detailed descriptions of the modeling technique using parallel pathways. Zhao et al.24 have studied the effect of characteristic x rays as applied to an amorphous selenium based imager using parallel pathways. Recently, Ganguly et al.19 have analyzed the performance of a CCD-based microangiographic imager designed for neurovascular imaging using the parallel cascade approach. In this work, the parallel-cascaded linear systems-based model16,19,22–24 was applied to the fluoroscopic imager using a clinically representative polyenergetic x-ray spectrum with an aim of predicting system performance and with the assumption that the statistical nature of light follows the Poisson process.
Stage 0: Incident x-ray quanta
The system was analyzed using a polyenergetic 72 kVp x-ray beam filtered by 2.54 cm (1 in.) of Al from a 12° tungsten (W) target, with a first half-value layer (HVL) of 7.2 mm of Al, and is in the range of beam quality used by several investigators.25–31 A plot of this spectrum normalized to unit area is shown in Fig. 1. The fluence per μR of exposure ( ) for this beam was calculated to be 293 x-ray photons/(mm2 μR) based on the definition of Roentgen provided by Johns and Cunningham32 and the technique described by Siewerdsen et al.2
Fig. 1.

Plot of the 72 kVp x-ray spectrum (7.2 mm of Al HVL) incident on the imager and used in the model. The spectrum is normalized to unit area.
Stage 1: Attenuation of x rays by the scintillator
This is a stochastic gain stage that follows the binomial process, where the probability that an incident x-ray photon is attenuated by the CsI:Tl scintillator is given by the average quantum efficiency, . The quantum efficiency, g1(E), of the CsI:Tl scintillator was computed as
| (1) |
The energy-dependent mass attenuation coefficient of CsI:Tl represented as [(μ/ρ)(E)] CsI:Tl was obtained from the Physical Reference Data of the National Institute of Standards and Technology.33 The material density of CsI:Tl was assumed to be 4.51 g/cm3. The surface density (coverage), ρs, for each CsI:Tl scintillator thickness provided by the manufacturer (Hamamatsu Corp., Bridgewater, NJ), and shown in Table I, was used. The energy-dependent quantum efficiency was weighted by the incident spectrum to provide the average quantum efficiency, , as shown in the following:
| (2) |
where, qrel(E) is the incident spectrum normalized to unit area. The calculated for the four CsI:Tl thickness is provided in Table I.
Table I.
Summary of parameters used in the model or estimated from the study.
| Scintillator thickness (μm)
|
||||||
|---|---|---|---|---|---|---|
| Parameter | Notation | Pixel pitch (μm) | 300 | 375 | 450 | 525 |
| Fluence [photons/(mm2 μR)] | q0/X | 293.1 | ||||
| CsI:Tl coverage [mg/cm2] | ρs | 103 | 126 | 148 | 175 | |
| Quantum efficiency | 0.663 | 0.730 | 0.779 | 0.827 | ||
| Detector enclosure (cover) transmission factor | tf | 0.944 | ||||
| Probability of K-shell interactiona | ς | 0.85 | ||||
| Fluorescence yielda | ω | 0.89 | ||||
| K-absorption probability | fK | 0.583 | 0.630 | 0.666 | 0.701 | |
| Mean gain path A | 1608.3 | 1475.8 | 1358 | 1250.1 | ||
| Mean gain path B | 592.2 | 549.3 | 510 | 473.8 | ||
| Mean gain path C | 1023.5 | 933.5 | 854.5 | 782.5 | ||
| Mean quantum gain | 1290.8 | 1219.6 | 1146.9 | 1077.8 | ||
| Fiberoptic coupling efficiency | 0.495 | |||||
| CCD quantum efficiency | 0.53 | |||||
| Fill factor | Ff | 0.513 | ||||
| 78 | 24.3 | |||||
| Additive noise [e−rms] | σT | 156 | 34.1 | |||
| 234 | 45.9 | |||||
| 78 | 193.1 | 201.5 | 202.4 | 201.8 | ||
| Sensitivity [e−/μR] | Γ | 156 | 775.7 | 806.1 | 809.7 | 807.3 |
| 234 | 1745.2 | 1813.8 | 1821.8 | 1816.5 | ||
Reference 22.
Stage 2: Generation and emission of optical quanta by scintillator
Following the work of Cunningham16,22 and Yao,23 paths A, B, and C correspond to local light emission when a K-fluorescent x ray is not produced, local light emission when a K-fluorescent x ray is produced, and remote (K-fluorescent x-ray absorption site) light emission when a K-fluorescent x ray is produced as shown in Fig. 2. With the assumption that the statistical nature of light follows the Poisson process, the NPS at the output of stage 2 along paths A, B, and C can be determined using the transfer relationship13 as
Fig. 2.

Illustration of the parallel pathways (Refs. 16, 22, and 23) in optical quanta generation and blurring within the CsI:Tl scintillator.
| (3) |
where tf is the x-ray transmission factor of the cover plate (detector enclosure), ς is the probability of K-shell photoelectric interaction, ω is the fluorescence yield, fK is the fraction of K-fluorescent x rays reabsorbed in the scintillator, and , and are the mean gains along paths A, B, and C, respectively. If the statistical nature of light is ignored,16,22 then the NPS at the output of stage 2 along paths A, B, and C can be determined using the transfer relationship13 as
| (4) |
It is relevant to note that for large values of , and , the terms , and approximate , and , respectively, which indicate that the statistical nature of light can be ignored. However for completeness, equations for the case that include the statistical nature of light in the theoretical model are presented. The measured tf of the cover plate (detector enclosure) for the specified spectrum was 0.944 and the values for ςand ω were obtained as 0.85 and 0.89, respectively, from Cunningham et al.22
In order to determine the mean gains along each of the three paths, the scintillator was considered to be made of fractional layers of thickness Δt as shown in Fig. 3. Given a layer at a distance z from the output surface of the phosphor (surface of phosphor coupled to the fiberoptic plate), then the incident x rays are attenuated by a factor e−μ(t−z) before it reaches the fractional layer of interest. Within this layer of thickness Δt, a fraction corresponding to (1−e−μΔt) is absorbed and converted to optical quanta. The linear attenuation coefficients were determined from the mass attenuation coefficients by taking into account the density of CsI:Tl and packing fraction. Holl et al.34 measured 50 000 optical quanta per MeV and more recently as high as 64 000 optical quanta per MeV35 have been reported. We chose to use the average of 58 000 optical quanta per MeV (58 optical quanta per keV) for m0. Thus the absorbed fraction within the layer of thickness Δt generates (1−e−μΔt)Em0 optical quanta. However, only a fraction of the generated optical quanta from a fractional layer located at a distance z from the output surface escape the scintillator in the direction of the CCD and this fraction is represented as ηesc(z). The escape efficiency ηesc(z), was determined by fitting the result of Hillen et al.36 with a first-order exponential decay function of the form ηesc(z) = y0+a1e−z/t1, with the assumption that the maximum escape efficiency occurs at the output surface of the scintillator. Hence, the mean gain along paths A, B, and C, represented as , and were determined as
Fig. 3.

Schematic of the model used for determining optical quanta generation and emission from the CsI:Tl scintillator.
| (5) |
where EK is the K edge of CsI:Tl and approximated as 33 keV. The K-fluorescent x-ray absorption factor, fK, was determined using the technique of Chan and Doi37 with the assumption that the x-ray beam is normally incident. The average energy of the characteristic x rays was determined by weighting the Kα1, Kα2, and Kβ1 emission lines of cesium and iodine with their relative intensities.38 The photoelectric and total mass attenuation coefficients corresponding to the average energy (30 keV) were obtained from NIST database.39 The calculated fK for the four CsI:Tl thicknesses are shown in Table I. The frequency-dependent signal along path C can be determined as
| (6) |
where TK(u,v) is the characteristic transfer function (stochastic blur of K-fluorescent x rays). TK(u,v) was determined using the technique of Metz and Vyborny.40 The point spread function of K-fluorescent x-ray absorption (characteristic spread function) was determined using Monte Carlo simulation41 and with absorption and scattering coefficients of CsI:Tl determined at 30 keV.
Since paths B and C are correlated and all quanta are real, the cross-spectral density can be written as16,23,24
| (7) |
The summed mean signal at the output of stage 2 is given by
| (8) |
where the mean gain of stage 2, and is identical to m̄ stated in Ref. 20. The NPS at the output of stage 2 can be written as
| (9) |
Stage 3: Optical blurring by the scintillator
The stochastic blur due to the random displacement of optical quanta, represented as T3(u,v), was approximated from measurement of the scintillator MTF (obtained by deconvolving the pixel presampling MTF from the measured presampling MTF) at energy levels below the K edge of CsI:Tl. Specifically, the measurement was performed with a mammographic x-ray source at 32 kVp using a rhodium (Rh) target and rhodium (Rh) filtration. Alternatively, T3(u,v) can be obtained by Monte Carlo modeling42 or by weighting of the measured presampling MTF as performed by Cunningham et al.22 Based on the transfer relationship,13 the NPS at the output of stage 3 can be written as
| (10) |
Stage 4: Coupling of optical quanta by fiberoptics
This is a stochastic gain stage that follows the binomial process, where the probability that an incident quantum is coupled to the CCD is given by the average fiberoptic coupling efficiency, . The fiberoptic transmission efficiency, calculated using the technique of Hejazi and Trauernicht,18 was determined to be 0.495. Based on the transfer relationship,13 the NPS at the output of stage 4 can be written as
| (11) |
where .
Stage 5: Absorption of optical quanta by the CCD
This is a stochastic gain stage that follows the binomial process, where the probability that the CCD absorbs an incident quantum is given by the average CCD quantum efficiency ( ). Quantum efficiency of 0.53 provided by the manufacturer of the CCD (Fairchild Imaging, Inc., Milpitas, CA) was used in the calculations. Based on the transfer relationship,13 the NPS at the output of stage 5 can be written as
| (12) |
Stage 6: Pixel presampling MTF and fill factor
The pixel presampling MTF represented as T6(u,v) was approximated with a sinc function, and can be stated as7
| (13) |
where ax and ay represent the dimensions of the pixel that is sensitive to light (active dimension) in the x and y directions, respectively. The fill factor, Ff, of 0.513 was used. Based on the transfer relationship,2 the mean signal at the output of stage 6 can be written as
| (14) |
The frequency-dependent signal at the output of stage 6 can be written as
| (15) |
where .
Based on the transfer relationship,2 the NPS at the output of stage 6 for the case where the statistical nature of light is considered can be written as
| (16) |
If the statistical nature of light is ignored,16,22 then the NPS along the u axis at the output of stage 6 can be shown as
| (17) |
where . Equation (17) is identical to Eq. (24) of Ref. 22 for the case tf is unity, is unity, and is expressed as sinc2(πau).
Stage 7: NPS Aliasing
The signal and NPS represented in the above equations are the presampling signal and the presampling NPS. However, DQE measurements reported in the literature5,43–48 for digital imaging (sampled) systems are based either on the presampling MTF or on the expectation MTF computed from the presampling MTF, and the aliased NPS. The aliased NPS represented as W7(u,v) is expressed as49,50
| (18) |
where III(u,v) is the Fourier transform of a rectangular array of δ functions representing the pixel matrix.49,50
Stage 8: Additive noise
The total additive electronic noise (σT) associated with a CCD-based system was calculated using the technique of Hejazi and Trauernicht18 by considering a typical dark current of 15 pA/cm2, measured read noise of 20 electrons rms, measured quantization step-size of 2.2 electrons/digital unit (DU) and with the imager operating at 30 fps. Additive noise levels (σT) thus determined for the three pixel pitch modes and included in Table I were used in the simulation. For the case where the electronic noise is “white” (independent of spatial frequency), the amplitude of the “white” noise power, Wadd8 can be estimated as49
| (19) |
where σT is the estimated electronic noise, UNyq and VNyq are the Nyquist sampling limits along the two orthogonal directions. Hence, the NPS at the output of stage 8 can be stated as
| (20) |
The detective quantum efficiency, which is defined as the ratio of the square of the output signal-to-noise ratio to the square of the input signal-to-noise ratio can be calculated from the above-given equations as
| (21) |
where (q0 /X) is the photon fluence per μR of incident exposure and X is the incident exposure in μR. The presampling MTF represented as MTFpre along the two orthogonal axes were determined as
| (22) |
The sensitivity (Γ) of the imaging system was calculated as .
For fluoroscopic applications, in addition to the spatial characteristics, the temporal characteristics of the imager also need to be addressed. Of particular importance is image lag, which is a result of a fraction of the generated signal from a particular frame being released into subsequent frames. Primary sources that contribute to image lag in pulsed fluoroscopic systems include the decay characteristics of the scintillator, decay characteristics of the x-ray source primarily due to the capacitance of the high-tension cables, and charge traps within the CCD. Measurements of the CsI:Tl scintillator decay characteristics at room temperature by Valentine et al.51 have found two primary decay time constants of 679±10 ns and 3.34±0.14 μs, which contribute to 63.7% and 36.1% of the emission. The system is designed to include a delay of 2 ms after the termination of the x-ray pulse, which is sufficient to allow for almost complete integration of the emitted optical quanta within a particular frame. Also, scientific-grade CCDs are routinely used for fast-framing applications. Hence, image lag is not expected to be significant with this system. The companion paper48 confirms this theoretical expectation, where first-frame image lag of 0.9% was measured with the system operating at 30 fps.
The model was implemented using IDL (Version 5.5, Research Systems, Inc., Boulder, CO) and all simulations were performed with the imager operating at 30 fps. The effect of CsI:Tl thickness on the presampling MTF[MTFpre(u)], NPS after aliasing [W8(u,0) ], and frequency-dependent DQE[DQE(u)], were studied for all the three pixel pitch modes at a nominal fluoroscopic exposure rate of 2.5 μR/ frame. The effect of pixel size on the DQE(u) of the system employing a 450 μm CsI:Tl scintillator was performed at an exposure rate of 2.5 μR/frame. The effect of incident exposure on the frequency-dependent DQE was studied for all pixel pitch modes and for various thickness of the CsI:Tl scintillator by varying the incident exposure rate from 1 to 10.5 μR/frame.
All computations of the frequency-dependent DQE addressed so far are along the u axis in the spatial frequency domain, which corresponds to the active dimension that is unaffected by the fill factor in the spatial domain. However, along the v axis there is degradation of the active dimension due to interline channel. Hence, it is pertinent to compare the presampling MTF, aliased NPS and DQE performance along the two orthogonal axes. Simulations of the presampling MTF, aliased NPS and DQE along the u and v axes, were performed with the imager using a 450 μm CsI:Tl and operating at 30 fps, 156 μm pixel-pitch mode and an exposure rate of 2.5 μR/frame.
III. RESULTS AND DISCUSSION
Table I provides a summary of the parameters used or estimated from the model. Since the values of , and , were large, the terms , and may be approximated to , and , respectively, which indicate that the statistical nature of light generation can be ignored with this imager. However, for completeness all calculations were performed without the approximation. The experimentally measured sensitivity of a prototype imager addressed in the companion paper was 360 DU/μR for 156 μm pixel pitch mode48 and when scaled by quantization step size of 2.2 electrons/DU, yields 792 electrons/μR, which is in good agreement with the theoretical estimate of 810 electrons/μR for 156 μm pixel pitch mode.
Figure 4 shows the presampling MTF along the u axis computed for each thickness of the CsI:Tl scintillator considered with the imager operating at the 78, 156, and 234 μm pixel pitch modes. The plots indicate that limiting MTF (10% of presampling MTF) in excess of 5 cycles/mm can be achieved in the high-resolution 78 μm pixel pitch mode with the imager employing a 450 μm CsI:Tl scintillator. For the 156 and 234 μm pixel pitch modes, presampling MTF of 0.21 and 0.36 were predicted at Nyquist sampling limits of 3.21 and 2.14 cycles/mm for the system employing a 450 μm CsI:Tl scintillator. Such high-spatial resolution may be beneficial in many fluoroscopic procedures such as cardiovascular imaging and interventional neuroangiography.
Fig. 4.

Presampling MTF along the u axis computed for each thickness of the CsI:Tl scintillator considered with the imager operating at the (a) 78, (b) 156, and (c) 234 μm pixel pitch modes. Also, the sinc of pixel pitch for the 78, 156, and 234 μm pixel pitch modes has been included.
Figure 5 shows the aliased NPS along the u axis computed for each thickness of the CsI:Tl scintillator considered with the imager operating at the 78, 156, and 234 μm pixel pitch modes and at a nominal fluoroscopic exposure rate of 2.5 μR/frame. The faster roll-off of the NPS with increasing thickness of CsI:Tl scintillator is due to the combination of increased optical blur [degraded T0(u,v)] and increased K-fluorescent x-ray absorption fraction ( fK) with increasing CsI:Tl thickness.
Fig. 5.

Aliased NPS along the u axis computed for each thickness of the CsI:Tl scintillator considered at an exposure rate of 2.5 μR/frame with the imager operating at the (a) 78, (b) 156, and (c) 234 μm pixel pitch modes.
Figure 6 shows the DQE along the u axis computed for each thickness of the CsI:Tl scintillator considered with the imager operating at the 78, 156, and 234 μm pixel pitch modes and at a nominal fluoroscopic exposure rate of 2.5 μR/frame. The plots indicate that DQE(0) in excess of 0.6 can be achieved with a 450 μm CsI:Tl scintillator even at a high-resolution pixel pitch mode of 78 μm. In the 78 μm pixel pitch mode, while increasing the CsI:Tl scintillator thickness improved the DQE(0), at higher spatial frequencies there was degradation of the DQE, with the transition occurring at ~2.7 cycles/mm. With increasing CsI:Tl thickness, the faster roll-off of the square of the presampling MTF relative to the aliased NPS resulted in this transition. Such a trend has been observed with an a-Si based imager.25 However for the 156 and 234 μm pixel pitch modes, increasing the CsI:Tl scintillator thickness resulted in improved DQE up to their corresponding Nyquist sampling limit. Extrapolation of the DQE plots for the 156 and 234 μm pixel pitch modes beyond their corresponding Nyquist sampling limits indicate that such a transition would have occurred, however the DQE is not defined beyond the Nyquist sampling limit.
Fig. 6.

Spatial frequency-dependent DQE along the u axis computed for each thickness of the CsI:Tl scintillator considered at an exposure rate of 2.5 μR/frame with the imager operating at the (a) 78, (b) 156, and (c) 234 μm pixel pitch modes.
Figure 7 shows the DQE along the u axis computed for the system using a 450 μm CsI:Tl scintillator and at an exposure rate of 2.5 μR/frame for the three pixel pitch modes. This plot indicates that there is a minimal degradation of the DQE(0) with the 78 μm pixel pitch mode compared to that with the 156 μm pixel pitch mode due to relatively larger contribution from the electronic noise to the total NPS with the 78 μm pixel pitch mode. It is also seen that at 156 and 234 μm pixel pitch modes, faster roll-off of the DQE occurs due to the combined effects of reduced presampling MTF and increased noise aliasing. Figure 8 shows the exposure dependence of DQE(u) for each thickness of CsI:Tl scintillator and with the imager operating at the high-resolution 78 μm pixel pitch mode. The plots indicate that for exposure rates of 2.5 μR/frame and above, there is minimal degradation of the DQE at low spatial frequencies. Further the plots indicate that DQE(0) of ~0.7 is achievable with a 525 μm CsI:Tl scintillator at exposure rates of 2.5 μR/frame and above.
Fig. 7.

The effect of pixel pitch on the DQE performance is illustrated by the plot of DQE(u) computed for the system using a 450 μm CsI:Tl scintillator and at an exposure rate of 2.5-μR/frame for the three pixel pitch modes.
Fig. 8.

Exposure dependence of DQE(u) with the imager operating at the high-resolution 78 μm pixel pitch mode for (a) 300 μm CsI:Tl, (b) 375 μm CsI:Tl, (c) 450 μm CsI:Tl, (d) 525 μm CsI:Tl scintillator. The exposure levels indicated in the legend are exposure per frame with the imager operating at 30 fps. The plots are in the same order as in the legend.
Figure 9 shows the exposure dependence of DQE(u) for each thickness of CsI:Tl scintillator and with the imager operating at the 156 μm pixel pitch mode. The plots indicate that the DQE performance had minimal dependence with exposure rate within the range considered and DQE(0) of ~0.64 is achievable with a 450 μm CsI:Tl scintillator even at a low fluoroscopic exposure rate of 1 μR/frame. To facilitate direct comparison with the experimentally determined DQE addressed in a companion paper,48 DQE(u) of the system using the 450-μm-thick CsI:Tl and operating in the 156 μm pixel pitch mode at exposure rates of 1 and 10.5 μR/ frame are also presented. Experimental data are provided only for this configuration, as this version of the prototype imager operates at a fixed 4×4 pixel binning resulting in a 156 μm pixel pitch and uses a 450-μm-thick CsI:Tl. The model shows excellent agreement with experimentally determined DQE. It is important to note that the several of the parameters used in the model related to the scintillator are either approximated or averaged, such as the K edge of CsI:Tl, fluorescence yield of CsI:Tl, characteristic emission energy of CsI:Tl, optical blur of CsI:Tl, and the mean number of optical quanta generated per absorbed 1 keV photon. More accurate information and appropriate treatment of these parameters may yield better accuracy in the theoretical estimates of performance. Further, the model does not include Lubberts effect,52 thus limiting its applicability to large CsI:Tl thickness. Simulations of the exposure dependency for each thickness of CsI:Tl scintillator considered with the system operating at the 234 μm pixel pitch mode, yielded similar trends to that observed with the system operating at the 156 μm pixel pitch mode, and hence are not reported.
Fig. 9.

Exposure dependence of DQE(u) with the imager operating at the 156 μm pixel pitch mode for (a) 300 μm CsI:Tl, (b) 375 μm CsI:Tl, (c) 450 μm CsI:Tl, (d) 525 μm CsI:Tl scintillator. The exposure levels indicated in the legend are exposure per frame with the imager operating at 30 fps. The plots are in the same order as in the legend. To facilitate direct comparison with experimental results (Ref. 48), DQE(u) at 1 (boxes) and 10.5 μR/ frame (triangles) are presented in (c).
All computations addressed so far are along the u axis, which corresponds to the active dimension that is unaffected by the fill factor in the spatial domain. However, along the v axis there is degradation of the active dimension due to interline channel. Figure 10 shows the comparison of the pre-sampling MTF, aliased NPS, and DQE along the u and v axes performed with the imager using a 450 μm CsI:Tl and operating at 156 μm pixel-pitch mode and an exposure rate of 2.5 μR/frame. While noticeable differences between the u and v axes are observed in the presampling MTF and the aliased NPS, the DQE performance remains relatively unchanged. To facilitate direct comparison with the experimentally determined DQE addressed in a companion paper,48 DQE of the system using the 450-μm-thick CsI:Tl and operating in the 156 μm pixel pitch mode at exposure rates of 2.5 μR/frame along the u and v axes are also presented.
Fig. 10.

Comparison of the (a) presampling MTF, (b) aliased NPS, and (c) DQE along the u and v axes performed with the imager using a 450 μm CsI:Tl and operating at 156 μm pixel-pitch mode and an exposure rate of 2.5 μR/frame. The exposure levels indicated in the legend are exposure per frame with the imager operating at 30 fps. To facilitate direct comparison with experimental results (Ref. 48), MTF, NPS, and DQE along u and v axes are also presented.
IV. CONCLUSIONS
The physical characteristics of a solid-state fluoroscopic imaging system developed for high-resolution angiography were analyzed with a parallel-cascaded linear systems-based model16,19,22–24 that takes into account the polyenergetic nature of the incident x-ray beam and the stochastic nature of the generated optical quanta. Good agreement between the model and the experimental results indicate that this modeling technique can serve as a valuable tool for predicting and optimizing system performance. For the fluoroscopic exposure rates considered and with the imager operating at 156 μm pixel pitch mode, minimal or no dependence with exposure rate was observed suggesting quantum noise-limited mode of operation. Results indicate that it is feasible to obtain DQE(0) in excess of 0.6, while attaining limiting spatial resolution (10% presampling MTF) in excess of 5 cycles/ mm, even at a nominal fluoroscopic exposure rate of 2.5 μR/frame.
Results from the DQE calculations and the high spatial resolution capabilities combined with its unique feature for a CCD-based imager to provide multiple resolution modes suited to the imaging requirement suggest that this type of technology or its variants can be a good candidate for high-resolution neuro-interventional imaging including region-of-interest fluoroscopy,19,53–55 cardiac and pediatric angiography.
Acknowledgments
The authors would like to thank Hetal Ved for technical assistance. The authors would also like to thank Fairchild Imaging, Inc., for providing technical data related to the solid-state imaging device. This work was supported in part by Grant No. R01 HL65551 from the National Heart, Lung and Blood Institute (NHLBI) and in part by Grant No. R01 EB002270 from the National Institute of Biomedical Imaging and Bioengineering (NIBIB), National Institutes of Health (NIH). The contents of this paper are solely the responsibility of the authors and do not necessarily represent the official views of NHLBI, NIBIB, or NIH.
References
- 1.Blume H, Colditz J, Eckenbach W, Hoen P, Meijer J, Snoeren RM, Spaak WE, Spekowius G. RSNA categorical course in Physics. 1995. Image-intensifier and x-ray exposure control systems; pp. 87–103. [Google Scholar]
- 2.Siewerdsen JH, Antonuk LE, El-Mohri Y, Yorkston J, Huang W, Boudry JM, Cunningham IA. Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology. Med Phys. 1997;24:71–89. doi: 10.1118/1.597919. [DOI] [PubMed] [Google Scholar]
- 3.Bissonette JP, Cunningham IA, Jaffray DA, Fenster A, Munro P. A quantum accounting and detective quantum efficiency analysis for video-based portal imaging. Med Phys. 1997;24:815–826. doi: 10.1118/1.598009. [DOI] [PubMed] [Google Scholar]
- 4.Zhao W, Rowlands JA. Digital radiology using active matrix readout of amorphous selenium: Theoretical analysis of detective quantum efficiency. Med Phys. 1997;24:1819–1833. doi: 10.1118/1.598097. [DOI] [PubMed] [Google Scholar]
- 5.Siewerdsen JH, Antonuk LE, El-Mohri Y, Yorkston J, Huang W, Cunningham IA. Signal, noise power spectrum and detective quantum efficiency of indirect-detection flat-panel imagers for diagnostic radiology. Med Phys. 1998;25:614–628. doi: 10.1118/1.598243. [DOI] [PubMed] [Google Scholar]
- 6.Jenkins GM, Watts DG. Spectral Analysis and its Applications. Holden-Day; San Francisco: 1968. [Google Scholar]
- 7.Dainty JC, Shaw R. Image Science. Academic; New York: 1974. [Google Scholar]
- 8.Papoulis A. Probability, Random Variables, and Stochastic Processes. 3. McGraw–Hill; New York: 1991. [Google Scholar]
- 9.Tapiovaara MJ, Wagner RF. SNR and DQE analysis of broad spectrum x-ray imaging. Phys Med Biol. 1985;30:519–529. [Google Scholar]
- 10.Williams MB, Simoni PU, Smilowitz L, Stanton M, Phillips W, Stewart A. Analysis of the detective quantum efficiency of a developmental detector for digital mammography. Med Phys. 1999;26:2273–2285. doi: 10.1118/1.598741. [DOI] [PubMed] [Google Scholar]
- 11.Metz CE, Doi K. Transfer function analysis of radiographic imaging systems. Phys Med Biol. 1979;24:1079–1106. doi: 10.1088/0031-9155/24/6/001. [DOI] [PubMed] [Google Scholar]
- 12.Barrett HH, Swindell W. Radiological Imaging—The Theory of Image Formation, Detection and Processing. revised. Academic; New York: 1981. [Google Scholar]
- 13.Rabbani M, Shaw R, Van Metter R. Detective quantum efficiency of imaging systems with amplifying and scattering mechanisms. J Opt Soc Am A. 1987;4:895–901. doi: 10.1364/josaa.4.000895. [DOI] [PubMed] [Google Scholar]
- 14.Rabbani M, Van Metter R. Analysis of signal and noise propagation for several imaging mechanisms. J Opt Soc Am A. 1989;6:1156–1164. [Google Scholar]
- 15.Cunningham IA, Westmore MS, Fenster A. A spatial frequency-dependent quantum accounting diagram and detective quantum efficiency model of signal and noise propagation in cascaded imaging systems. Med Phys. 1994;21:417–427. doi: 10.1118/1.597401. [DOI] [PubMed] [Google Scholar]
- 16.Cunningham IA. Linear-systems modeling of parallel-cascaded stochastic processes: The NPS of radiographic screens with reabsorption of characteristic x radiation. Proc SPIE. 1998;3336:220–230. [Google Scholar]
- 17.Cunningham IA. Applied linear-systems theory, Physics and Psychophysics. In: Beutel J, Kundel HL, Van Metter R, editors. Handbook of Medical Imaging. Vol. 1. SPIE; Bellingham: 2000. pp. 79–160. [Google Scholar]
- 18.Hejazi S, Trauernicht DP. System considerations in CCD-based x-ray imaging for digital chest radiography and digital mammography. Med Phys. 1997;24:287–297. doi: 10.1118/1.598072. [DOI] [PubMed] [Google Scholar]
- 19.Ganguly A, Rudin S, Bednarek DR, Hoffman KR. Micro-angiography for neurovascular imaging. II. Cascade model analysis. Med Phys. 2003;30:3029–3039. doi: 10.1118/1.1617550. [DOI] [PubMed] [Google Scholar]
- 20.Cunningham IA, Moschandreou T, Subotic V. The detective quantum efficiency of fluoroscopic systems: The case for a spatial-temporal approach (or, does the ideal observer have infinite patience?) Proc SPIE. 2001;4320:479–488. [Google Scholar]
- 21.Swank RK. Absorption and noise in x-ray phosphors. J Appl Phys. 1973;44:4199–4203. [Google Scholar]
- 22.Cunningham IA, Yao J, Subotic V. Cascaded models and the DQE of flat-panel imagers: Noise aliasing, secondary quantum noise and reabsorption. Proc SPIE. 2002;4682:61–72. [Google Scholar]
- 23.Yao J, Cunningham IA. Parallel Cascades: New ways to describe noise transfer in medical imaging systems. Med Phys. 2001;28:2020–2038. doi: 10.1118/1.1405842. [DOI] [PubMed] [Google Scholar]
- 24.Zhao W, Ji WG, Rowlands JA. Effects of characteristic x-rays on the noise power spectra and detective quantum efficiency of photoconductive x-ray detectors. Med Phys. 2001;28:2039–2049. doi: 10.1118/1.1405845. [DOI] [PubMed] [Google Scholar]
- 25.Granfors PR. Performance characteristics of an amorphous silicon flat panel x-ray imaging detector. Proc SPIE. 1999;3659:480–490. [Google Scholar]
- 26.Flynn MJ, Samei E. Experimental comparison of noise and resolution for 2k and 4k storage phosphor radiography systems. Med Phys. 1999;26:1612–1623. doi: 10.1118/1.598656. [DOI] [PubMed] [Google Scholar]
- 27.Stierstorfer K, Spahn M. Self-normalizing method to measure the detective quantum efficiency of a wide range of x-ray detectors. Med Phys. 1999;26:1312–1319. doi: 10.1118/1.598626. [DOI] [PubMed] [Google Scholar]
- 28.Samei E, Flynn MJ. An experimental comparison of detector performance for computed radiography systems. Med Phys. 2002;29:447–459. doi: 10.1118/1.1449873. [DOI] [PubMed] [Google Scholar]
- 29.Samei E, Flynn MJ. An experimental comparison of detector performance for direct and indirect digital radiography systems. Med Phys. 2003;30:608–622. doi: 10.1118/1.1561285. [DOI] [PubMed] [Google Scholar]
- 30.Samei E. Image quality in two phosphor-based flat panel digital radiographic detector. Med Phys. 2003;30:1747–1757. doi: 10.1118/1.1578772. [DOI] [PubMed] [Google Scholar]
- 31.Granfors PR, Aufrichtig R, Possin GE, Giambattista BW, Huang ZS, Liu J, Ma B. Performance of a 41×41 cm2 amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications. Med Phys. 2003;30:2715–2726. doi: 10.1118/1.1609151. [DOI] [PubMed] [Google Scholar]
- 32.Johns HE, Cunningham JR. The Physics of Radiology. 4. Thomas; Springfield, IL: 1983. [Google Scholar]
- 33.Hubbell JH, Seltzer SM. National Institute of Standards and Technology; Gaithersburg, MD: 1997. Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients (version 1.03) [Online] Available: http://physics.nist.gov/xaamdi. [Google Scholar]
- 34.Holl I, Lorenz E, Mageras G. A measurement of light yield of common inorganic scintillators. IEEE Trans Nucl Sci. 1988;35:105–109. [Google Scholar]
- 35.Jing T, Goodman CA, Cho G, Drewery J, Hong WS, Lee H, Kaplan SN, Mireshghi A, Perez-Mendez V, Wildermuth D. Amorphous silicon pixel layers with cesium iodide converters for medical radiography. IEEE Nuclear Science Symposium; San Francisco, CA. 2–5 November, 1993. [Google Scholar]
- 36.Hillen W, Eckenbach W, Quadfleig P, Zaengel T. Signal-to-noise performance in cesium iodide x-ray fluorescent screens. Proc SPIE. 1991;1443:120–131. [Google Scholar]
- 37.Chan HP, Doi K. Energy and angular dependence of x-ray absorption and its effect on radiographic response in screen-film systems. Phys Med Biol. 1983;28:565–579. doi: 10.1088/0031-9155/28/5/009. [DOI] [PubMed] [Google Scholar]
- 38.Kortright JB, Thompson AC. X-ray Data Booklet. Section 1.2 X-ray Emission Energies. Lawrence Berkeley National Laboratory; Berkeley, CA: 2000. [Online] http://xdb.lbl.gov/Section1/Sec_1-2.html. [Google Scholar]
- 39.Chantler CT, Olsen K, Dragoset RA, Kishore AR, Kotochigova SA, Zucker DS. X-Ray Form Factor, Attenuation and Scattering Tables (version 2.0) NIST; Gaithersburg, MD: 2003. [Online] Available: http://physics.nist.gov/ffast. [Google Scholar]
- 40.Metz CE, Vyborny CJV. Wiener spectral effects of spatial correlation between the sites of characteristic x-ray emission and reab-sorption in radiographic screen-film systems. Phys Med Biol. 1983;28:547–564. doi: 10.1088/0031-9155/28/5/008. [DOI] [PubMed] [Google Scholar]
- 41.Jacques SL. Light distribution from points, line and plane sources for photochemical reactions and fluorescence in turbid biological tissues. Photochem Photobiol. 1998;67:23–32. [Online] Source code Available: http://omlc.ogi.edu/classroom/ece532/class4/ssmc/index.html. [PubMed] [Google Scholar]
- 42.Badano A, Leimbach R. Depth-dependent phosphor blur in indirect x-ray imaging sensors. Proc SPIE. 2002;4682:94–106. [Google Scholar]
- 43.Zhao W, Blevis I, Germann S, Rowlands JA, Waechter D, Huang Z. Digital radiology using active matrix readout of amorphous selenium: Construction and evaluation of a prototype real-time detector. Med Phys. 1997;24:1834–1843. doi: 10.1118/1.598098. [DOI] [PubMed] [Google Scholar]
- 44.Vedantham S, Karellas A, Suryanarayanan S, Levis I, Sayag M, Kleehammer R, Heidsieck R, D’Orsi CJ. Mammographic imaging with a small format CCD-based digital cassette: Physical characteristics of a clinical system. Med Phys. 2000;27:1832–1840. doi: 10.1118/1.1286720. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Vedantham S, Karellas A, Suryanarayanan S, Albagli D, Han S, Tkaczyk EJ, Landberg CE, Opsahl-Ong B, Granfors PR, Levis I, D’Orsi CJ, Hendrick RE. Full breast digital mammography with an amorphous silicon-based flat panel detector: Physical characteristics of a clinical prototype. Med Phys. 2000;27:558–567. doi: 10.1118/1.598895. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Granfors PR, Aufrichtig R. Performance of a 41×41-cm2 amorphous silicon flat-panel x-ray detector for radiographic imaging applications. Med Phys. 2000;27:1324–1331. doi: 10.1118/1.599010. [DOI] [PubMed] [Google Scholar]
- 47.Dobbins JT, Ergun DL, Rutz L, Hinshaw DA, Blume H, Clark DC. DQE(f) of four generation of computed radiography acquisition devices. Med Phys. 1995;22:1581–1593. doi: 10.1118/1.597627. [DOI] [PubMed] [Google Scholar]
- 48.Vedantham S, Karellas A, Suryanarayanan S, Onishi SK. Solid-state fluoroscopic imager for high-resolution angiography: Physical characteristics of an 8 cm × 8 cm experimental prototype. Med Phys. doi: 10.1118/1.1750992. accepted. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Siewerdsen JH. Ph. D. dissertation. University of Michingan; Ann Arbor, MI: 1998. Signal, noise, and detective quantum efficiency of a-Si:H flat-panel imagers. [Google Scholar]
- 50.Antonuk LE, Jee KW, El-Mohri Y, Maolinbay M, Nassif S, Rong X, Zhao Q, Siewerdsen JH, Street RA, Shah KS. Strategies to improve the signal and noise performance of active matrix, flat panel imagers for diagnostic x-ray applications. Med Phys. 2000;27:289–306. doi: 10.1118/1.598831. [DOI] [PubMed] [Google Scholar]
- 51.Valentine JD, Moses WW, Derenzo SE, Wehe DK, Knoll GF. Temperature dependence of CsI(Tl) gamma-ray excited scintillation characterisitics. Nucl Instrum Methods Phys Res A. 1993;325:147–157. [Google Scholar]
- 52.Lubberts G. Random noise produced by x-ray fluorescent screens. J Opt Soc Am. 1968;58:1475–1483. [Google Scholar]
- 53.Rudin S, Wu Y, Kyprianou IS, Ionita CN, Wang Z, Ganguly A, Bednarek DR. Micro-angiographic detector with fluoroscopic capability. Proc SPIE. 2002;4682:344–354. [Google Scholar]
- 54.Massoumzadeh P, Rudin S, Bednarek DR. Quantitative evaluation of amorphous selenium (a-Se) for region-of-interest fluoroscopy. Proc SPIE. 2000;3977:658–669. [Google Scholar]
- 55.Ganguly A, Rudin S, Bednarek DR, Hoffman KR, Kyprianou IS. Micro-angiography for neurovascular imaging. I. Experimental evaluation and feasibility. Med Phys. 2003;30:3018–3028. doi: 10.1118/1.1617549. [DOI] [PubMed] [Google Scholar]
