Abstract
Background Context
The optimal animal model for in vivo testing of spinal implants, particularly total or partial disc replacement devices, has not yet been determined. Mechanical and morphological similarities of calf and human spines have been reported; however, limitations of the calf model include open growth plates and oversized vertebrae with growth. Mature runt cows (Corrientes breed) may avoid these limitations.
Purpose
This study compared vertebral morphology and biomechanical properties of human and runt cow lumbar motion segments.
Study Design
In vivo disc pressure measurements were obtained in six mature runt cows at L4–5. In vitro evaluation was performed on these same segments and repeated on twelve human motion segments.
Methods
Disc pressures were measured in vivo in runt cow (Corrientes breed) L45 discs using a percutaneous transducer with the animal performing various activities. These motion segments were then harvested and morphologic and biomechanical evaluations (disc pressure in compression, flexibility tests to 7.5 Nm) were performed on both cow and male human L23 and L45 segments.
Results
The transverse lumbar disc dimensions were slightly smaller for (mixed gender) cow vs. (male) humans, but were within the range of reported (mixed gender) human values. The mean ± SD disc height was smaller for runt cow (7 ± 1 mm) vs. human discs (13 ± 2 mm, p<0.001). The vertebral bodies of the cow were approximately twice as tall as the human. In vitro testing revealed significantly greater disc pressure response to applied axial loading in the runt cow vs. humans (1.27 ± 0.18 kPa/N vs. 0.84 ± 0.15 kPa/N respectively) but similar overall stiffness (2.15 ± 0.71 kN/mm vs. 1.91 ± 0.94 kN/mm, respectively). Runt cow and human segments flexibility curves were similar with the following exceptions: runt cow stiffness was ~40% greater in torsion (p<0.05), runt cow segment lateral bending motion was greater vs. humans (range of motion by 30%, neutral zone by 100%; both p<0.05) and flexion range of motion tended to be smaller in runt cow vs. human specimens (by ~40%, p=NS). In vivo, the standing disc pressure in the runt cow was 0.80 ± 0.24 MPa.
Conclusions
Although no animal replicates the human motion segment, the runt cow lumbar spine had a number of biomechanical and morphological measurements within the range of human values. The closed physes and temporally stable morphology of the mature runt cow may make this model more suitable vs. standard calf models for human intradiscal implant studies.
Introduction
Interest in lumbar spine arthroplasty has increased because spinal fusion for end-stage degenerative disc disease may result in decreased range of motion and accelerated deterioration of the remaining adjacent discs, particularly when multiple disc levels are fused[1–3]. Disc arthroplasty devices must be durable and able to continue to function for the duration of the patients’ lives of which most would have implantation during their middle ages. The durability requirements for spinal disc arthroplasty thus far exceeds that of fusion devices, which only need to survive until fusion is obtained, and joint arthroplasty of the appendicular skeleton, which usually applies to the elderly patient with limited lifespan. Animal models may be useful in the design and testing of spinal disc replacements. Optimal in vivo testing should use a mature mammal with lumbar disc morphology and in vivo loading that approximate those of the human spine.
Domestic mammal spines have often been used for in vitro testing of lumbar spinal fusion implants because motion segment load-displacement relationships are somewhat similar to human data[4–9]. But most of these animals, such as the dog, sheep, goat and pig, have vertebra and disc anatomy which are substantially smaller than that of humans[10–15]. Lumbar calf spines are of similar size to human spines, but often fail at relatively low testing loads because the growth plates are still cartilaginous[7;16–18].
There is also controversy as to whether the appropriate animal model for in vivo testing should have an upright posture[19]. Some believe that spinal loading depends on upright posture and have recommended the use of kangaroos or nonhuman primates. However, there are difficulties with these “erect” animal models; as most are scarce, expensive, difficult to work with and are in reality primarily quadruped [20]. Because of their small disc height and diameter, prior studies using nonhuman primates have necessitated routine removal of substantial portions of the endplates in order to fit even miniaturized disc implants[20;21]. Furthermore, similarities in vertebral anatomy and trabecular orientation suggest similarities in the predominant axial forces in bipeds and quadrupeds[19].
In a pilot study, the authors found that the mature runt (“mini-”) cow and male pig (boar) lumbar spines had a number of anatomical size similarities to the human lumbar spine particularly with respect to the intervertebral disc[22]. Surgical access to the lumbar spine of the runt cow is simple relative to that in the boar, and may therefore be a more suitable animal model for the study of human disc prostheses[23].
The purpose of this study is to compare and contrast runt cow and human lumbar anatomy, in vitro and in vivo intradiscal pressures, and multi-axial load-displacement properties. One hypothesis of this study is that the disc of the mature runt cow, which is similar to human size, has intradiscal pressures similar to human (e.g., 0.5 to 1.2 MPa range while standing) [24–28]. The combined in vitro and in vivo results would validate that the runt cow may be a suitable in vivo animal model for study and treatment of human lumbar intervertebral conditions.
Methods
In vivo cow intradiscal pressure testing
After approval by the institutional animal care and use committee, six adult runt cows (Corrientes breed, 3 female and 3 male, size and age in Table 1) were obtained for study. Each cow was premedicated with xylazine (0.02 to 0.05 mg/kg, intramuscularly [IM]), and anesthetized with ketamine (2–4 mg/kg, intravenous [IV]) and Pentothal (3–4 mg/kg, IV, to effect.) Buprenorphine (0.005–0.01 mg/kg, IM) was administered to provide preemptive analgesia and then each animal was intubated and maintained in anesthesia with isoflurane or sevoflurane inhalant anesthetic.
Table 1.
Vertebral body and disc space measurements and DEXA score for both human and runt cow specimens. See Figure 1 for corresponding anatomy.
| Measure (Drawing Abbreviation) | Runt Cow | Human (male) | Significant Difference | Prior Human Studies, Mean (References) |
|---|---|---|---|---|
| Mean (SD) | Mean (SD) | p | ||
|
|
||||
| Age (years) | 1.9 (0.2) | 46.5 (5.7) | ||
| Weight (Kg) | 264 (22) | 88.7 (26.9) | ||
| Disc Measurements | ||||
| Disc height L45 (DH, mm) | 7.2 (0.8) | 13.3 (2.1) | <0.001 | 8–14(1,2,10,13,15,17,19) |
| Ratio of nucleus to total disc area | 36% (6%) | 41% (4%) | NS | 30–50%(18,19) |
| Disc area (cm2) | 13.4 (3.6) | 22.1 (4.4) | <0.01 | 9–26(3,4,14,15,16,17,19) |
| L45 disc wedging (⊖L, degrees lordosis) | 1.8 (1.6) | 8.3 (3.7) | <0.01 | 15–17(2,5,6) |
| Disc degeneration (Thompson grade I-V,) | all gr I | 4x gr II, 6x gr III, 2x gr IV | ||
| Endplate diameter (mm) | ||||
| L4 Coronal (ED-C) | 45.5 (2.6) | 58.0 (6.8) | <0.01 | 47–54(4,7,8,) |
| L4 Sagittal (ED-S) | 36.0 (3.8) | 39.8 (5.5) | NS | 33–40(1,4,7,8,10,13) |
| L5 Coronal (ED-C) | 46.8 (2.6) | 59.8 (6.2) | 0.01 | 50–53(4,7,8,11,13) |
| L5 Sagittal (ED-S) | 36.2 (4.0) | 38.7 (5.0) | NS | 33–40(1,4,7,8,10,11,13) |
| Vertebral body diameter (waist, mm) | ||||
| Coronal L4 (VBD-C) | 41.7 (3.8) | 47.2 (6.5) | NS | 41–45(7,9) |
| Sagittal L4 (VBD-S) | 30.0 (3.7) | 37.0 (3.9) | <0.01 | 33–35(7,9) |
| Coronal L5 (VBD-C) | 45.2 (6.2) | 51.0 (6.6) | NS | 46–48(7,9) |
| Sagittal L5 (VBD-S) | 30.2 (3.5) | 36.2 (4.3) | 0.02 | 32–36(7,9) |
| Other (mm) | ||||
| L4 Vertebral body height (VBH) | 52.5 (3.6) | 27.8 (2.1) | <0.001 | 24–28(4,7,10,11) |
| L5 Vertebral body height (VBH) | 53.8 (3.3) | 28.3 (1.6) | <0.001 | 23–28(4,7,10,11) |
| L4 Transverse process length (TPL) | 97.0 (16.7) | 22.0 (4.5) | <0.001 | 16.5(4) |
| L5 Transverse process length (TPL) | 95.3 (14.2) | 20.4 (3.3) | <0.001 | 23(4) |
| DEXA | ||||
| L4 and L5 Vertebral body bone density (anteroposterior, gm/cm2) | 1.529 (0.230) | 1.194 (0.225) | NS | 1.233(12) |
| L4 and L5 Vertebral body bone density (lateral, gm/cm2) | 1.704 (0.220) | 0.855 (0.265) | <0.01 | |
Henry 2004
Schultz 1979
SD = standard deviation
NS = not significant
The animal was placed in a semi-sternal recumbent position and its L4–5 disc was identified under fluoroscopy. After application of local anesthetic, a 1.5 mm diameter trocar was percutaneously placed from a dorsolateral approach into the disc. A cannulated guide tube was placed over the trocar and secured in the annulus. The trocar was removed, and through the guide tube a calibrated intradiscal pressure transducer (1.3 mm diameter × 20 cm length, <5% error, Medical Measurements Incorporated, Hackensack, New Jersey) was placed into the L4–5 disc using a technique previously described[28], but modified to allow wireless data acquisition. Biplanar fluoroscopy was used to confirm transducer placement within the central region of the disc.
Opening intradiscal pressure values were obtained with the animal at rest and the changes related to ventilation (0.02 to 0.05 MegaPascal, MPa) and spontaneous breathing (mean 0.05 MPa, range 0.01 to 0.16 MPa) were recorded to determine their artifact. The animals were transferred to a 3 × 5 m pen and allowed to recover (typically 30 to 90 minutes) after which intradiscal pressure measurements were again obtained with the animal lying holding its head up, standing, and walking. After satisfactory data acquisition, animals were sedated with xylazine (0.2 mg/kg IM) and ketamine (2–4 mg/Kg), and then euthanized with an overdose of phenobarbitol.
Morphometric data
After in vivo testing, the lumbar cow spines were harvested and stored at −20° C. Male human L2–3 (n=6) and L4–5 (n=6) motion segments were obtained from an ongoing study[29]. Radiographs and bone mineral densities (DEXA, GE Lunar Medical Systems, Madison, WI) were obtained for the runt cow and human spines. Morphometric data for both runt cow and human specimens were obtained from radiographs and direct measurements using calipers. This included vertebral endplate diameters in coronal and sagittal planes, L45 intervertebral lordosis (Cobb technique), disc and vertebral body height, and transverse process length (Figure 1). The error between radiographic and direct measurements was 0.7 mm on average and ranged from 0 to 3 mm. After flexibility testing, the motion segments were disarticulated and the discs cross-sections digitally photographed with a standardized distance marker. The nucleus, annulus, and total disc cross-sectional area were measured using image-analysis software (Scion Corporation, Frederick, MD). Disc degeneration was categorized over five grades using the Thompson classification (I = normal to V = severe degeneration)[30]).
Figure 1.
Radiographic tracings of the runt cow L4–5 motion segment vertebrae. See Table 1 for dimensions.
In vitro cow and human intradiscal pressure testing and calibration
Explanted L4–5 bovine motion segments were isolated (longitudinal and posterior ligaments, disc and facet capsules were left intact) and the ends potted for in vitro testing. Intradiscal pressure measurements were conducted as described above unloaded and under applied axial loads. Axial loads were applied in compression using a servopneumatic load frame (Enduratec, Minneapolis, MN) with all rotations and off-axis translations fully constrained. A peak compression load of 1500 N was applied in a sinusoidal fashion at a mean rate of 80 N/s. Three load cycles were applied with a 10 second dwell time between cycles; the third cycle was used for data analysis.
In three animals the cannulas were still well fixed in the annulus from the in vivo testing, which was conducted using posterolateral access, and were tested with the sensing needle passed through this cannula. The other three motion segments had the needles inserted using the posterolateral approach simulating that in vivo. All motion segments were also tested with the needle placed anteriorly, and with the needle sensing surface oriented both vertically and horizontally; artifact due to anterior vs. posterolateral access and vertical vs. horizontal needle orientation was less than 4%. Manual bending of the transducer cannula produced artifactual readings ranging from 0.005 to 0.02 MPa. The 12 human cadaver motion segments were tested in vitro using identical methods but, because needle access direction (anterior vs. posterolateral) appeared to have little influence, specimens were tested only using the simpler anterior needle access.
The applied load vs. intradiscal pressure relationship was always linear in both the cow and human motion segments (R2 = .99). Accordingly, a linear load vs. pressure curve was fit for each animal. The slope and offset from this curve were recorded for each animal and were used to convert the bovine in vivo pressures to estimated in vivo axial loads.
In vitro cow and human flexibility testing
Flexibility testing (sagittal and lateral bending and axial torsion) was conducted in an identical fashion for both cow and human motion segments. Unconstrained moments were applied sinusoidally to 7.5 Newton-meters (Nm) at 0.03 Hz, with an additional 100 N axial compressive load applied coincident with the long axis of the distal vertebrae. Three preconditioning cycles were used for each test and the fourth cycle was used for data analysis. A 10 second pause held at 0 Nm occurred at after each moment cycle to allow for viscoelastic recovery. Rotation data were recorded using a video motion measurement system (VICON, Oxford Metrics, Oxford, UK) by tracking three reflective markers mounted on each vertebra. Marker translations were converted to anatomic rotations using Euler angle formulations. Range of motion and neutral zone were calculated as previously described[31]. Initial stiffness was calculated as the slope of the flexibility curves in the low-stiffness neutral zone, and the elastic zone stiffness was calculated in the higher stiffness elastic zone region. Rotation data were sub-sampled at 1.0 Nm increments, and the rotation at each increment was averaged to yield a mean ± standard deviation for plotting moment-rotation curves (Figures 2–4).
Figure 2.

Flexion-extension flexibility curves (mean +/− standard deviation) for runt cow and human lumbar spine.
Figure 4.

Axial torsion flexibility curves (mean +/− standard deviation) for runt cow and human lumbar spine.
Statistical Analysis
Statistical analysis was performed to compare cow and human vertebrae in terms of range of motion, neutral zone, stiffness and dimensional measurements. Comparisons were made using the student’s t-tests with significance considered at p<0.05. The error bars in the graphs represent plus or minus one standard deviation (SD).
Results
All the young adult runt cow discs were well hydrated whereas the middle-aged human discs varied in their degree of degeneration (Table 1). Measurements of runt cow endplate diameter were within approximately 20% of the human values measured in this study. The superior endplates of the cows were dome shaped, whereas human endplates were comparably more planar. The transverse dimensions of the cow endplates were slightly smaller than those for human segments. Within this small group of runt cows the weight was 5% greater and the endplates were 4% larger for the males relative to the females. The runt cows had significantly larger vertebral body axial height, longer transverse processes, smaller disc height and less lordosis relative to the human motion segments.
Overall, the runt cow flexibility profiles were similar in shape to those observed in the human motion segment, but tended to differ in magnitude (Figures 2–4). For the number of animals tested, the combined flexion/extension motion segment properties were not statistically different between humans and runt cows in terms of neutral zone, initial stiffness and elastic zone stiffness (Figures 5–6). The mean range of motion of the human specimens tended to be greater than that of the runt cow in both flexion (5.7 ± 3.2° vs. 3.3° ± 0.8°) and extension (2.9 ± 2.0° vs. 2.1° ± 0.6°), but these differences did not reach a level of statistical significance with this sample size (p ≥ 0.174). In lateral bending, the runt cow had a significantly greater range-of-motion (p = 0.022). This difference was primarily due to the runt cow’s large neutral zone, which was significantly greater than that of the human motion segment (p = 0.001). In axial torsion, there tended to be more motion in the human vs. runt cow motion segments; this difference was marginally significant in terms of elastic zone stiffness only (p = 0.046).
Figure 5.

Range-of-motion and neutral zone for runt cow and human lumbar spine in 3 axis of rotation (mean +/− standard deviation).
Figure 6.
Initial and elastic zone stiffness for runt cow and human lumbar spine (mean +/−standard deviation).
Over the entire axial compression loading curve the mean stiffness was 2.15 ± 0.71 kN/mm and 1.91 ± 0.94 kN/mm for the cow and human motion segments respectively (Figure 7a). From the mean load-displacement curves, the stiffness in the high load region (>500 N applied) 2.27 kN/m and 2.24 kN/m for the cow and human motion segments respectively. The intradiscal pressure response was significantly greater in the cow (1.27 ± 0.18 kPa/N) relative to the human (0.84 ± 0.15 kPa/N, p<0.001, Figure 7b). Human pressure response under axial loading was significantly lower in degenerated specimens (disc grade ≥ 3, 0.67 kPa/N) than those in normal motion segments (disc grade ≤ 2, 0.93 kPa/N, p=0.02), whereas between levels (L23 versus L45) there was no significant difference (p=0.5).
Figure 7.
(a, b). (a) Axial load-displacement characteristics and (b) intradiscal pressure of runt cow L4–5 lumbar discs, and human L2–3 and L4–5 lumbar discs (mean +/− standard deviation).
In vivo cow L4–5 intradiscal pressures for various activities are shown in Figure 8. Intradiscal pressures were lowest in sedentary positions, greater in static activities that actively used back muscles, such as holding the head up, and greatest during dynamic activities such as walking or pushing off with the legs. In vivo loads as estimated from the individual specimens’ in vitro load vs. pressure “calibration” curves averaged 167 N (range 88 – 292 N) when sedated and lying prone, 256 N (75 – 409 N) lying sedated in the pen, 437 N (385 – 958 N) when lifting head, 632 N (385 – 955 N) when standing, and reached a mean peak load of 2585 N (1313 – 4375 N) during ambulation.
Figure 8.
In vivo runt cow L4–5 intradiscal pressures for various activities. Values represent the mean (+1 SD) values with the exception of Wilke [27], where the median (maximum) of static measurements from each category in Table 1 in that study. The “holding weights” values represent static values for sitting (Polga[28]; Nachemson[24]) or standing (Wilke[27]).
Discussion
This comprehensive comparison of cow and human lumbar segment morphology and biomechanics has two unique features: first, this model evaluated a mature animal with closed physes and stable morphology; second, in vivo disc pressures were evaluated, which has not been previously published for quadrupeds. In this discussion, the authors also validate the human results by comparison to prior human in vitro studies.
Anatomic measurements found the anterior disc height of intact human lumbar discs of the present study to be similar to those previously reported[32–36]. Many anatomical human measurements of the present study of males, Table 1, found a mean size slightly greater than prior reports, which typically reported measurements from combined male and female spines[33;37;38].
Disc height values were less for the cow, but were comparable to those seen in humans with disc degeneration (Table 1). Lordosis between adjacent L4 and L5 endplates was significantly less in the cows than for the humans of this study and that reported in prior human studies (p < 0.01)[35;39;40]. Morphometric data pertinent to disc implant studies found that the runt cow and human both had oval shaped endplates. The cow endplate dimensions were smaller than the male humans of the present study; however, they were of similar size in diameter and cross sectional area to prior mixed gender human reports [32;33;36–38;41–44]. The ratio of disc nucleus to annulus in cows was slightly less than that for the human specimens in this study, but is close to previously reported human values[45]. Although the endplates were oval for cow and human, the waists of the bovine vertebral bodies were slightly triangular in cross-section. The transverse process lengths of the cows were 4 to 5 times as long as for the humans, which suggest that the cows may experience greater bending moments due to the processes acting as long levers. This idea is supported by the high transient pressures seen in vivo in this study.
The bone mineral density of the bovine vertebrae were within the range predicted for mature animals of the size tested in the present study[46]. Anteroposterior DEXA scans found 25% greater bone density in cows compared to humans (p = NS). Lateral scans found a greater difference in apparent density; this difference may be underestimated by the smaller coronal plane diameter in the bovine vertebra. This may be due in part to the greater cortical thickness of the bovine vertebral body which the authors noted qualitatively, and has been described previously [16]. Discrepancies in the DEXA score are less likely to be due to trabecular bone density since a previous study using quantitative computed tomography relating compressive mechanical testing found calf trabecular bone characteristics to be similar to that of young human specimens[47].
Axial compression loading with intradiscal pressure monitoring of the human specimens found stiffness to be in the range (2 – 6 kN/mm) of prior reports[34;44;45;48–50]. In vitro disc pressure response to applied axial loading in the human motion segments was in agreement with previous human studies[24;51;52]. Decreased disc pressure for degenerated discs, as found in the present study, has been noted previously[53–59].
Human flexibility results of the present study were similar to prior reports. The range of motion and neutral zone were well within the range of previous studies[31;45;49;60–63;63–65]. Bending stiffness of the human specimens was also similar to previous studies which have typically found initial stiffness in the range of 0.5 to 1 Nm/degree and elastic zone stiffness of 2 to 4 Nm/degree[31;45;49;60;61;63;63;64].
This report is the first to study the morphology and biomechanics of the lumbar spine of the Corrientes breed mature runt cow. The intradiscal pressure response to applied loading for the cow was slightly greater than for the human spines but similar to a prior animal (pig) study[66]. The axial stiffness of the runt cow motion segments was slightly greater than for the human specimens (2.00 versus 1.85 MN/m overall, 3.3 MN/m for both at loads > 500 N). The difference in cow versus human, and the difference due to degree of degeneration within the human discs is most likely due to nucleus hydration, greater integrity of the annulus in the animal and greater osmotic pressure in the cow nucleus due to the type and size of proteoglycans, which also may reflect the age of the disc[67;68]
Our mature cow range of motion results were similar to one prior calf spine study[6] but slightly less in axial rotation compared to another study[7;9]. Compared to humans, the runt cows had slightly less range-of-motion in flexion/extension (especially flexion), significantly greater range-of-motion and neutral zone in lateral bending, and significantly greater axial torsion stiffness. Limitations of the present study are the small number of animals tested and the varying degree of disc degeneration of the human specimens. Although none of the human specimens had advanced disc degeneration, even moderate degeneration in the human specimens may have contributed to their flexibility variability; this resultant variance in human specimen range of motion may have masked actual population differences. Post-hoc analysis of the data based on degeneration level (“minimal to mild” vs. “moderate” dehydration) revealed slightly increased flexion-extension mean range of motion for the “minimal to mild” discs. This difference was not statistically significant (p=0.3). A sample size analysis using the standard deviation of nearly 50% of the mean observed in this study, ~45 samples would be required to detect a significant range of motion difference of 30% between minimal and moderate dehydration groups.
In vivo cow intradiscal pressures for a number of activities were determined in the present study and values for standing, 0.80 ± 0.24 MPa, were found to be comparable to human studies with reported pressures ranging from 0.55 to 1.2 MPa[24–28]. The mean peak in vivo dynamic pressures during walking were greater in the runt cow (2.3 MPa) compared to the human (0.53 MPa - 0.85 MPa[27]). Compressive loads during various cow activities were estimated using the in vitro intradiscal pressure vs. axial load experiment as a motion segment specific calibration curve. Average runt cow estimated loads were 170 N to 260 N lying at rest, 440 N while holding up the head, 630 N while standing, and 2590 N (peak) during ambulation. The dynamic loads for the cow were greater than predicted human peak dynamic loads, such as estimated human peak ambulation (1060 N – 1590 N)[27;69]. However, static loads were comparable to previous human estimates for lying at rest (144 to 240 N) and standing (500 to 800N)[26;27;53;69]. Animal studies in baboon and sheep, which have smaller discs, have estimated in vivo loads of 200 N to 800 N for a variety of activities[20;70]. These findings and others[19;71] suggest that an upright posture is not required for axial spinal loading.
In vivo loading in articular joints has been measured in both domestic mammals and humans[72–80]. Based on these studies, in vivo joint contact pressures are estimated to generally range from 1.0 to 2.5 MPa (and up to 10 MPa for peak loading). The findings in the current and previous studies that both fibrous joints (e.g., the disc) and articular joints generally exist in a 0.5 to 2.5 MPa dynamic stress environment may not be coincidental because chondrocytes are metabolically most active in this load range[81–86]. A stress environment that is static, too low, or too high results in decreased chondrocyte proteoglycan production, increased protein degradation, and chondrocyte cell death[85;87–91]. One can hypothesize that the size of a mammalian joint or disc is therefore determined by the stress environment in order to maintain a healthy chondrocyte population.
In conclusion, while no animal replicates the human lumbar motion segment, the runt cow lumbar motion segment is similar to that of the human in terms of transverse disc size, flexibility characteristics in several loading directions, and in vitro and in vivo intradiscal pressures representing static positions and were within the range of those reported previously. However, several differences also exist, including slightly smaller disc sizes, greater vertebral body heights and higher dynamic pressures in the runt cow. These differences and similarities should be considered when choosing an animal model for study of human spinal conditions and treatments. The mature Corrientes runt cow lumbar spine may be useful for the in vivo animal testing of human implants, and particularly intradiscal instrumentation, and provide the advantage of stable morphology in survival studies.
Figure 3.

Lateral bending flexibility curves (mean +/− standard deviation) for runt cow and human lumbar spine.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
Contributor Information
Dr. Glenn Robin Buttermann, Midwest Spine Institute Stillwater, MN.
Brian P Beaubien, Gustillo Education Center.
Louis C Saeger, Midwest Spine Institute.
References
- 1.Ghiselli G, Wang JC, Bhatia NN, et al. Adjacent segment degeneration in the lumbar spine. J Bone Joint Surg Am. 2004;86-A:1497–503. doi: 10.2106/00004623-200407000-00020. [DOI] [PubMed] [Google Scholar]
- 2.Gillet P. The fate of the adjacent motion segments after lumbar fusion. J Spinal Disord Tech. 2003;16:338–45. doi: 10.1097/00024720-200308000-00005. [DOI] [PubMed] [Google Scholar]
- 3.Park P, Garton HJ, Gala VC, et al. Adjacent segment disease after lumbar or lumbosacral fusion: review of the literature. Spine. 2004;29:1938–44. doi: 10.1097/01.brs.0000137069.88904.03. [DOI] [PubMed] [Google Scholar]
- 4.Asazuma T, Stokes IA, Moreland MS, et al. Intersegmental spinal flexibility with lumbosacral instrumentation. An in vitro biomechanical investigation. Spine. 1990;15:1153–8. doi: 10.1097/00007632-199011010-00013. [DOI] [PubMed] [Google Scholar]
- 5.Lysack JT, Yen D, Dumas GA. In vitro flexibility of an experimental pedicle screw and plate instrumentation system on the porcine lumbar spine. Med Eng Phys. 2000;22:461–8. doi: 10.1016/s1350-4533(00)00059-x. [DOI] [PubMed] [Google Scholar]
- 6.Riley LH, III, Eck JC, Yoshida H, et al. A biomechanical comparison of calf versus cadaver lumbar spine models. Spine. 2004;29:E217–E220. doi: 10.1097/00007632-200406010-00021. [DOI] [PubMed] [Google Scholar]
- 7.Wilke HJ, Krischak S, Claes L. Biomechanical comparison of calf and human spines. J Orthop Res. 1996;14:500–3. doi: 10.1002/jor.1100140321. [DOI] [PubMed] [Google Scholar]
- 8.Wilke HJ, Kettler A, Claes LE. Are sheep spines a valid biomechanical model for human spines? Spine. 1997;22:2365–74. doi: 10.1097/00007632-199710150-00009. [DOI] [PubMed] [Google Scholar]
- 9.Wilke HJ, Krischak ST, Wenger KH, et al. Load-displacement properties of the thoracolumbar calf spine: experimental results and comparison to known human data. Eur Spine J. 1997;6:129–37. doi: 10.1007/BF01358746. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.McLain RF, Yerby SA, Moseley TA. Comparative morphometry of L4 vertebrae: comparison of large animal models for the human lumbar spine. Spine. 2002;27:E200–E206. doi: 10.1097/00007632-200204150-00005. [DOI] [PubMed] [Google Scholar]
- 11.Gotfried Y, Bradford DS, Oegema TR., Jr Facet joint changes after chemonucleolysis-induced disc space narrowing. Spine. 1986;11:944–50. doi: 10.1097/00007632-198611000-00016. [DOI] [PubMed] [Google Scholar]
- 12.Yingling VR, Callaghan JP, McGill SM. The porcine cervical spine as a model of the human lumbar spine: an anatomical, geometric, and functional comparison. J Spinal Disord. 1999;12:415–23. [PubMed] [Google Scholar]
- 13.Sasaki M, Takahashi T, Miyahara K, et al. Effects of chondroitinase ABC on intradiscal pressure in sheep: an in vivo study. Spine. 2001;26:463–8. doi: 10.1097/00007632-200103010-00008. [DOI] [PubMed] [Google Scholar]
- 14.Wilke HJ, Kettler A, Wenger KH, et al. Anatomy of the sheep spine and its comparison to the human spine. Anat Rec. 1997;247:542–55. doi: 10.1002/(SICI)1097-0185(199704)247:4<542::AID-AR13>3.0.CO;2-P. [DOI] [PubMed] [Google Scholar]
- 15.O’connell GD, Vresilovic EJ, Elliot DM. Comparative intervertebral disc anatomy across several animal species. Chicago, IL. 52nd Annual Meeting of the Orthopaedic Research Society; 3-19-2006.2006. [Google Scholar]
- 16.Cotterill PC, Kostuik JP, D’Angelo G, et al. An anatomical comparison of the human and bovine thoracolumbar spine. J Orthop Res. 1986;4:298–303. doi: 10.1002/jor.1100040306. [DOI] [PubMed] [Google Scholar]
- 17.Allan DG, Russell GG, Moreau MJ, et al. Vertebral end-plate failure in porcine and bovine models of spinal fracture instrumentation. J Orthop Res. 1990;8:154–6. doi: 10.1002/jor.1100080121. [DOI] [PubMed] [Google Scholar]
- 18.Buttermann GR, Beaubien BP. Stiffness of prosthetic nucleus determines stiffness of reconstructed lumbar calf disc. Spine J. 2004;4:265–74. doi: 10.1016/j.spinee.2003.11.003. [DOI] [PubMed] [Google Scholar]
- 19.Smit TH. The use of a quadruped as an in vivo model for the study of the spine -biomechanical considerations. Eur Spine J. 2002;11:137–44. doi: 10.1007/s005860100346. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ledet EH, Tymeson MP, DiRisio DJ, et al. Direct real-time measurement of in vivo forces in the lumbar spine. Spine J. 2005;5:85–94. doi: 10.1016/j.spinee.2004.06.017. [DOI] [PubMed] [Google Scholar]
- 21.Cunningham BW, Dmitriev AE, Hu N, et al. General principles of total disc replacement arthroplasty: seventeen cases in a nonhuman primate model. Spine. 2003;28:S118–S124. doi: 10.1097/00007632-200310151-00005. [DOI] [PubMed] [Google Scholar]
- 22.Buttermann GR, Beaubien BP, Saeger LC. In vivo and in vitro lumbar disc loading in domestic quadrupeds. 12th International Meeting on Advanced Spine Techniques; July 7–9, 2005; Banff, Canada. [Google Scholar]
- 23.Mendenhall HV. The Small Cow as an Animal Model for Lumbar Intervertebral Disc. 12th International Meeting on Advanced Spine Techniques; Banff, Canada. 7-7-2005. [Google Scholar]
- 24.Nachemson AL, Elfstrom G. Intravital dynamic pressure measurements in lumbar discs. A study of common movements, maneuvers and exercises. Scand J Rehabil Med Suppl. 1970;1:1–40. [PubMed] [Google Scholar]
- 25.Nachemson AL. Disc pressure measurements. Spine. 1981;6:93–7. doi: 10.1097/00007632-198101000-00020. [DOI] [PubMed] [Google Scholar]
- 26.Schultz A, Andersson G, Ortengren R, et al. Loads on the lumbar spine. Validation of a biomechanical analysis by measurements of intradiscal pressures and myoelectric signals. J Bone Joint Surg Am. 1982;64:713–20. [PubMed] [Google Scholar]
- 27.Wilke HJ, Neef P, Caimi M, et al. New in vivo measurements of pressures in the intervertebral disc in daily life. Spine. 1999;24:755–62. doi: 10.1097/00007632-199904150-00005. [DOI] [PubMed] [Google Scholar]
- 28.Polga DJ, Beaubien BP, Kallemeier PM, et al. Measurement of in vivo intradiscal pressure in healthy thoracic intervertebral discs. Spine. 2004;29:1320–4. doi: 10.1097/01.brs.0000127179.13271.78. [DOI] [PubMed] [Google Scholar]
- 29.Buttermann GR, Beaubien BP. Spinal load sharing of a compressible annulus sparing disc prosthesis. 19th Annual Meeting of the North American Spine Society; Chicago, IL. 2004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Thompson JP, Pearce RH, Schechter MT, et al. Preliminary evaluation of a scheme for grading the gross morphology of the human intervertebral disc. Spine. 1990;15:411–5. doi: 10.1097/00007632-199005000-00012. [DOI] [PubMed] [Google Scholar]
- 31.Beaubien BP, Mehbod AA, Kallemeier PM, et al. Posterior augmentation of an anterior lumbar interbody fusion: minimally invasive fixation versus pedicle screws in vitro. Spine. 2004;29:E406–E412. doi: 10.1097/01.brs.0000141187.53366.9b. [DOI] [PubMed] [Google Scholar]
- 32.Brinckmann P, Grootenboer H. Change of disc height, radial disc bulge, and intradiscal pressure from discectomy. An in vitro investigation on human lumbar discs. Spine. 1991;16:641–6. doi: 10.1097/00007632-199106000-00008. [DOI] [PubMed] [Google Scholar]
- 33.Hall LT, Esses SI, Noble PC, et al. Morphology of the lumbar vertebral endplates. Spine. 1998;23:1517–22. doi: 10.1097/00007632-199807150-00002. [DOI] [PubMed] [Google Scholar]
- 34.Koeller W, Muehlhaus S, Meier W, et al. Biomechanical properties of human intervertebral discs subjected to axial dynamic compression--influence of age and degeneration. J Biomech. 1986;19:807–16. doi: 10.1016/0021-9290(86)90131-4. [DOI] [PubMed] [Google Scholar]
- 35.Shao Z, Rompe G, Schiltenwolf M. Radiographic changes in the lumbar intervertebral discs and lumbar vertebrae with age. Spine. 2002;27:263–8. doi: 10.1097/00007632-200202010-00013. [DOI] [PubMed] [Google Scholar]
- 36.Tibrewal SB, Pearcy MJ. Lumbar intervertebral disc heights in normal subjects and patients with disc herniation. Spine. 1985;10:452–4. doi: 10.1097/00007632-198506000-00009. [DOI] [PubMed] [Google Scholar]
- 37.Berry JL, Moran JM, Berg WS, et al. A morphometric study of human lumbar and selected thoracic vertebrae. Spine. 1987;12:362–7. doi: 10.1097/00007632-198705000-00010. [DOI] [PubMed] [Google Scholar]
- 38.Panjabi MM, Goel V, Oxland T, et al. Human lumbar vertebrae. Quantitative three-dimensional anatomy. Spine. 1992;17:299–306. doi: 10.1097/00007632-199203000-00010. [DOI] [PubMed] [Google Scholar]
- 39.Stagnara P, De Mauroy JC, Dran G, et al. Reciprocal angulation of vertebral bodies in a sagittal plane: approach to references for the evaluation of kyphosis and lordosis. Spine. 1982;7:335–42. doi: 10.1097/00007632-198207000-00003. [DOI] [PubMed] [Google Scholar]
- 40.Jackson RP, McManus AC. Radiographic analysis of sagittal plane alignment and balance in standing volunteers and patients with low back pain matched for age, sex, and size. A prospective controlled clinical study. Spine. 1994;19:1611–8. doi: 10.1097/00007632-199407001-00010. [DOI] [PubMed] [Google Scholar]
- 41.Gilad I, Nissan M. A study of vertebra and disc geometric relations of the human cervical and lumbar spine. Spine. 1986;11:154–7. doi: 10.1097/00007632-198603000-00010. [DOI] [PubMed] [Google Scholar]
- 42.Scoles PV, Linton AE, Latimer B, et al. Vertebral body and posterior element morphology: the normal spine in middle life. Spine. 1988;13:1082–6. doi: 10.1097/00007632-198810000-00002. [DOI] [PubMed] [Google Scholar]
- 43.van Schaik JJ, Verbiest H, van Schaik FD. Morphometry of lower lumbar vertebrae as seen on CT scans: newly recognized characteristics. AJR Am J Roentgenol. 1985;145:327–35. doi: 10.2214/ajr.145.2.327. [DOI] [PubMed] [Google Scholar]
- 44.Edwards WT, Ordway NR, Zheng Y, et al. Peak stresses observed in the posterior lateral anulus. Spine. 2001;26:1753–9. doi: 10.1097/00007632-200108150-00005. [DOI] [PubMed] [Google Scholar]
- 45.White AA, III, Panjabi MM. Clinical Biomechanics of the Spine. 2. Philadelphia, PA: J.B. Lippincott Company; 1990. Chapter 1 - Physical properties and functional biomechanics of the spine; pp. 1–83. [Google Scholar]
- 46.Mitchell AD, Scholz AM, Pursel VG. Total body and regional measurements of bone mineral content and bone mineral density in pigs by dual energy X-ray absorptiometry. J Anim Sci. 2001;79:2594–604. doi: 10.2527/2001.79102594x. [DOI] [PubMed] [Google Scholar]
- 47.Swartz DE, Wittenberg RH, Shea M, et al. Physical and mechanical properties of calf lumbosacral trabecular bone. J Biomech. 1991;24:1059–68. doi: 10.1016/0021-9290(91)90022-f. [DOI] [PubMed] [Google Scholar]
- 48.Markolf KL, Morris JM. The structural components of the intervertebral disc. A study of their contributions to the ability of the disc to withstand compressive forces. J Bone Joint Surg Am. 1974;56:675–87. [PubMed] [Google Scholar]
- 49.Markolf KL. Deformation of the thoracolumbar intervertebral joints in response to external loads: a biomechanical study using autopsy material. J Bone Joint Surg Am. 1972;54:511–33. [PubMed] [Google Scholar]
- 50.Smeathers JE, Joanes DN. Dynamic compressive properties of human lumbar intervertebral joints: a comparison between fresh and thawed specimens. J Biomech. 1988;21:425–33. doi: 10.1016/0021-9290(88)90148-0. [DOI] [PubMed] [Google Scholar]
- 51.Nachemson AL. Lumbar intradiscal pressure. Experimental studies on post-mortem material. Acta Orthop Scand. 1960;(Suppl 43):1–104. doi: 10.3109/ort.1960.31.suppl-43.01. [DOI] [PubMed] [Google Scholar]
- 52.Frei H, Oxland TR, Nolte LP. Thoracolumbar spine mechanics contrasted under compression and shear loading. J Orthop Res. 2002;20:1333–8. doi: 10.1016/S0736-0266(02)00058-X. [DOI] [PubMed] [Google Scholar]
- 53.Sato K, Kikuchi S, Yonezawa T. In vivo intradiscal pressure measurement in healthy individuals and in patients with ongoing back problems. Spine. 1999;24:2468–74. doi: 10.1097/00007632-199912010-00008. [DOI] [PubMed] [Google Scholar]
- 54.Horst M, Brinckmann P. 1980 Volvo award in biomechanics. Measurement of the distribution of axial stress on the end-plate of the vertebral body. Spine. 1981;6:217–32. doi: 10.1097/00007632-198105000-00004. [DOI] [PubMed] [Google Scholar]
- 55.Shirazi-Adl SA, Shrivastava SC, Ahmed AM. Stress analysis of the lumbar disc-body unit in compression. A three-dimensional nonlinear finite element study. Spine. 1984;9:120–34. doi: 10.1097/00007632-198403000-00003. [DOI] [PubMed] [Google Scholar]
- 56.McNally DS, Shackleford IM, Goodship AE, et al. In vivo stress measurement can predict pain on discography. Spine. 1996;21:2580–7. doi: 10.1097/00007632-199611150-00007. [DOI] [PubMed] [Google Scholar]
- 57.Pollintine P, Przybyla AS, Dolan P, et al. Neural arch load-bearing in old and degenerated spines. J Biomech. 2004;37:197–204. doi: 10.1016/s0021-9290(03)00308-7. [DOI] [PubMed] [Google Scholar]
- 58.Adams MA, McNally DS, Dolan P. ‘Stress’ distributions inside intervertebral discs. The effects of age and degeneration. J Bone Joint Surg Br. 1996;78:965–72. doi: 10.1302/0301-620x78b6.1287. [DOI] [PubMed] [Google Scholar]
- 59.Adams MA, Dolan P. Spine biomechanics. J Biomech. 2005;38:1972–83. doi: 10.1016/j.jbiomech.2005.03.028. [DOI] [PubMed] [Google Scholar]
- 60.Volkman T, Horton WC, Hutton WC. Transfacet screws with lumbar interbody reconstruction: biomechanical study of motion segment stiffness. J Spinal Disord. 1996;9:425–32. [PubMed] [Google Scholar]
- 61.Brodke DS, Dick JC, Kunz DN, et al. Posterior lumbar interbody fusion. A biomechanical comparison, including a new threaded cage. Spine. 1997;22:26–31. doi: 10.1097/00007632-199701010-00005. [DOI] [PubMed] [Google Scholar]
- 62.Lindsey DP, Swanson KE, Fuchs P, et al. The effects of an interspinous implant on the kinematics of the instrumented and adjacent levels in the lumbar spine. Spine. 2003;28:2192–7. doi: 10.1097/01.BRS.0000084877.88192.8E. [DOI] [PubMed] [Google Scholar]
- 63.Nachemson AL, Schultz AB, Berkson MH. Mechanical properties of human lumbar spine motion segments. Influence of age, sex, disc level, and degeneration. Spine. 1979;4:1–8. doi: 10.1097/00007632-197901000-00001. [DOI] [PubMed] [Google Scholar]
- 64.Panjabi MM, Oxland TR, Yamamoto I, et al. Mechanical behavior of the human lumbar and lumbosacral spine as shown by three-dimensional load-displacement curves. J Bone Joint Surg Am. 1994;76:413–24. doi: 10.2106/00004623-199403000-00012. [DOI] [PubMed] [Google Scholar]
- 65.Pearcy MJ, Tibrewal SB. Axial rotation and lateral bending in the normal lumbar spine measured by three-dimensional radiography. Spine. 1984;9:582–7. doi: 10.1097/00007632-198409000-00008. [DOI] [PubMed] [Google Scholar]
- 66.Ekstrom L, Holm S, Holm AK, et al. In vivo porcine intradiscal pressure as a function of external loading. J Spinal Disord Tech. 2004;17:312–6. doi: 10.1097/01.bsd.0000092068.78152.00. [DOI] [PubMed] [Google Scholar]
- 67.Urban JP, McMullin JF. Swelling pressure of the lumbar intervertebral discs: influence of age, spinal level, composition, and degeneration. Spine. 1988;13:179–87. doi: 10.1097/00007632-198802000-00009. [DOI] [PubMed] [Google Scholar]
- 68.Panjabi M, Brown M, Lindahl S, et al. Intrinsic disc pressure as a measure of integrity of the lumbar spine. Spine. 1988;13:913–7. doi: 10.1097/00007632-198808000-00008. [DOI] [PubMed] [Google Scholar]
- 69.Cappozzo A. Compressive loads in the lumbar vertebral column during normal level walking. J Orthop Res. 1984;1:292–301. doi: 10.1002/jor.1100010309. [DOI] [PubMed] [Google Scholar]
- 70.Hauerstock D, Reindl R, Steffen T. Telemetric measurement of compressive loads in the sheep lumbar spine. 47th annual meeting of the Orthopaedic Research Society; San Francisco, CA. 1-1-2001. [Google Scholar]
- 71.Lanyon LE. In vivo bone strain recorded from thoracic vertebrae of sheep. J Biomech. 1972;5:277–81. doi: 10.1016/0021-9290(72)90044-9. [DOI] [PubMed] [Google Scholar]
- 72.Afoke NY, Byers PD, Hutton WC. Contact pressures in the human hip joint. J Bone Joint Surg Br. 1987;69:536–41. doi: 10.1302/0301-620X.69B4.3611154. [DOI] [PubMed] [Google Scholar]
- 73.Bergmann G, Deuretzbacher G, Heller M, et al. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34:859–71. doi: 10.1016/s0021-9290(01)00040-9. [DOI] [PubMed] [Google Scholar]
- 74.Bergmann G, Graichen F, Rohlmann A. Hip joint forces in sheep. J Biomech. 1999;32:769–77. doi: 10.1016/s0021-9290(99)00068-8. [DOI] [PubMed] [Google Scholar]
- 75.Buttermann GR, Schendel MJ, Kahmann RD, et al. In vivo facet joint loading of the canine lumbar spine. Spine. 1992;17:81–92. doi: 10.1097/00007632-199201000-00013. [DOI] [PubMed] [Google Scholar]
- 76.Davy DT, Kotzar GM, Brown RH, et al. Telemetric force measurements across the hip after total arthroplasty. J Bone Joint Surg Am. 1988;70:45–50. [PubMed] [Google Scholar]
- 77.Hodge WA, Carlson KL, Fijan RS, et al. Contact pressures from an instrumented hip endoprosthesis. J Bone Joint Surg Am. 1989;71:1378–86. [PubMed] [Google Scholar]
- 78.Lanyon LE, Paul IL, Rubin CT, et al. In vivo strain measurements from bone and prosthesis following total hip replacement. An experimental study in sheep. J Bone Joint Surg Am. 1981;63:989–1001. [PubMed] [Google Scholar]
- 79.Paul JP. Forces transmitted by joints in the human body. Proc Instru Mech Engrs. 1966;181:8–15. [Google Scholar]
- 80.Rydell NW. Forces acting on the femoral head-prosthesis. A study on strain gauge supplied prostheses in living persons. Acta Orthop Scand. 1966;37(S):1–132. doi: 10.3109/ort.1966.37.suppl-88.01. [DOI] [PubMed] [Google Scholar]
- 81.Sah RL, Kim YJ, Doong JY, et al. Biosynthetic response of cartilage explants to dynamic compression. J Orthop Res. 1989;7:619–36. doi: 10.1002/jor.1100070502. [DOI] [PubMed] [Google Scholar]
- 82.Handa T, Ishihara H, Ohshima H, et al. Effects of hydrostatic pressure on matrix synthesis and matrix metalloproteinase production in the human lumbar intervertebral disc. Spine. 1997;22:1085–91. doi: 10.1097/00007632-199705150-00006. [DOI] [PubMed] [Google Scholar]
- 83.Li KW, Williamson AK, Wang AS, et al. Growth responses of cartilage to static and dynamic compression. Clin Orthop Relat Res. 2001:S34–S48. doi: 10.1097/00003086-200110001-00005. [DOI] [PubMed] [Google Scholar]
- 84.Kasra M, Goel V, Martin J, et al. Effect of dynamic hydrostatic pressure on rabbit intervertebral disc cells. J Orthop Res. 2003;21:597–603. doi: 10.1016/S0736-0266(03)00027-5. [DOI] [PubMed] [Google Scholar]
- 85.Walsh AJ, Lotz JC. Biological response of the intervertebral disc to dynamic loading. J Biomech. 2004;37:329–37. doi: 10.1016/s0021-9290(03)00290-2. [DOI] [PubMed] [Google Scholar]
- 86.Wenger KH, Woods JA, Holecek A, et al. Matrix remodeling expression in anulus cells subjected to increased compressive load. Spine. 2005;30:1122–6. doi: 10.1097/01.brs.0000162395.56424.53. [DOI] [PubMed] [Google Scholar]
- 87.Lotz J, Hsieh A, Palmer E, et al. Role of annular tension in disc repair. 3rd Annual Meeting of the Spine Arthroplasty Society; Scottsdale, AZ. 5-1-2003. [Google Scholar]
- 88.Ariga K, Yonenobu K, Nakase T, et al. Mechanical stress-induced apoptosis of endplate chondrocytes in organ-cultured mouse intervertebral discs: an ex vivo study. Spine. 2003;28:1528–33. [PubMed] [Google Scholar]
- 89.Hall AC, Urban JP, Gehl KA. The effects of hydrostatic pressure on matrix synthesis in articular cartilage. J Orthop Res. 1991;9:1–10. doi: 10.1002/jor.1100090102. [DOI] [PubMed] [Google Scholar]
- 90.Hsieh AH, Lotz JC. Prolonged spinal loading induces matrix metalloproteinase-2 activation in intervertebral discs. Spine. 2003;28:1781–8. doi: 10.1097/01.BRS.0000083282.82244.F3. [DOI] [PubMed] [Google Scholar]
- 91.Ishihara H, McNally DS, Urban JP, et al. Effects of hydrostatic pressure on matrix synthesis in different regions of the intervertebral disk. J Appl Physiol. 1996;80:839–46. doi: 10.1152/jappl.1996.80.3.839. [DOI] [PubMed] [Google Scholar]





