Abstract
The Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) maintains the Australian primary standard of absorbed dose. Until recently, the standard was used to calibrate ionisation chambers only in 60Co gamma rays. These chambers are then used by radiotherapy clinics to determine linac output, using a correction factor (kQ) to take into account the different spectra of 60Co and the linac. Over the period 2010–2013, ARPANSA adapted the primary standard to work in megavoltage linac beams, and has developed a calibration service at three photon beams (6, 10 and 18 MV) from an Elekta Synergy linac. We describe the details of the new calibration service, the method validation and the use of the new calibration factors with the International Atomic Energy Agency’s TRS-398 dosimetry Code of Practice. The expected changes in absorbed dose measurements in the clinic when shifting from 60Co to the direct calibration are determined. For a Farmer chamber (model 2571), the measured chamber calibration coefficient is expected to be reduced by 0.4, 1.0 and 1.1 % respectively for these three beams when compared to the factor derived from 60Co. These results are in overall agreement with international absorbed dose standards and calculations by Muir and Rogers in 2010 of kQ factors using Monte Carlo techniques. The reasons for and against moving to the new service are discussed in the light of the requirements of clinical dosimetry.
Keywords: Radiotherapy, Dosimetry, Absorbed dose, Protocol, TRS-398, Megavoltage calibration
Introduction
Dosimetry for external beam radiotherapy in Australia is traceable to the primary standard of absorbed dose, a graphite calorimeter of the Domen type [1,2], operated in a beam of 60Co radiation [3], This standard is maintained at the Melbourne laboratories of the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) under an authorisation from the National Measurement Institute (Australia). Radiotherapy facilities use the calibration service in combination with a calculated correction factor, kQ, tabled in the IAEA TRS-398 Code of Practice [4], to measure linac output. In Australia and New Zealand, secondary standard laboratories can also provide a 60Co calibration which can be used with TRS-398. In this way, all patient treatments delivered using external beams in this region are traceable to the calorimeter operated in 60Co.
ARPANSA has recently developed a calibration service to allow most therapy ionisation chamber types to be calibrated directly at three megavoltage photon energies: 6, 10 and 18 MV. The new service is more accurate than using a calculated kQ since the chambers are calibrated in beams very similar to those in radiotherapy clinics. The accuracy of a calculated kQ factor depends on the precision of the chamber geometry (including any deviations from the specified geometry) and the accuracy of the calculation. To use a measured kQ only requires that the chamber response be reasonably insensitive to photon energy, so that users can interpolate with the beam quality index. ARPANSA provide a quadratic fit to the calibration coefficients as a function of the beam quality TPR20,10 in the calibration report for purposes of interpolation. The new service is based on the same primary standard graphite calorimeter adapted to work with linac beams. Adaptations were required to account for variations in dose rate during irradiation and between irradiations, and for the higher dose rate compared to 60Co. Additional graphite buildup was also employed [2, 3].
The purpose of this article is to explain the fundamental differences between these two methods of traceability for absorbed dose in linac photon beams (via 60Co or via “direct” calibration). We present the validation of the new method, and compare the uncertainties in both cases. We present the change in absorbed dose expected in the clinic when shifting between the methods. We include a detailed discussion of the advantages and disadvantages of these methods from the radiotherapy users’ point of view. In particular, we discuss the two services in the light of the expected shift in dose, the origin of dose prescriptions, the improvement in dose accuracy which comes from the new service, and the expected shift in patient doses in Australasia compared to the rest of the world.
The new service is available outside Australia and could be used by New Zealand clinics. The standards laboratory of New Zealand, the National Centre for Radiation Science, has taken part in tests of the service. The pros and cons of this article are relevant, with the additional drawback of having to ship dosemeters overseas.
ARPANSA plans to extend the new method to mega-voltage electron beams. However at present electron dosimetry remains traceable to the primary standard realised on 60Co—either through cross-calibrations performed in the clinic, or by using the new cross-calibration service at ARPANSA. In both cases a parallel plate chamber is calibrated on a high energy electron beam against a thimble chamber which has been calibrated in 60Co.
Review of calibration services outside Australia
The primary standards laboratory of the UK, the National Physical Laboratory (NPL), introduced a calibration service for megavoltage photon beams in 1989 and for electron beams in 2001. These calibrations are used with the UK Codes of Practice [5, 6] as the basis for reference dosimetry in UK clinics.
In North America, the primary standards laboratories of the US and Canada, the National Institute of Standards and Technology, NIST, and National Research Council Canada, NRCC, respectively, maintain linear accelerators and are able to calibrate ionisation chambers directly in terms of absorbed dose at linac energies. They do not routinely provide calibrations to all radiotherapy facilities, however, as the large number of linacs in North America (more than 5,000) requires the existence of a network of secondary standards laboratories. These laboratories (Accredited Dosimetry Calibration Laboratories, ADCLs), do not have easy access to linacs, and presumably a business case weighing the relatively small improvements in accuracy brought about by using linacs is difficult to make when the reliability and lower cost of 60Co is taken into account.
The NRCC tests representative chambers to ensure that those calibrated via 60Co do not result in clinical dose errors. For example, in 2010 NRCC published results of measurements of a wide range of chambers indicating which chambers were suitable for reference dosimetry [7]. Megavoltage external beams in North America are calibrated according to the AAPM protocol TG-51 [8] which only allows 60Co as the reference quality. A review in 2014 continued this recommendation [9].
Across Europe, most ionisation chamber calibrations are performed in 60Co. In France and Germany, the primary standards laboratories Laboratoire National Henri Becquerel (LNHB) and Physikalisch-Technische Bundesanstalt (PTB), respectively, maintain primary standards on both 60Co and medical linacs. The majority of clinical dose measurements are traceable to chambers calibrated in 60Co [10]. In Belgian and Dutch hospitals [11], chambers are calibrated in 60Co and used with a local protocol NCS-18 [12] which includes kQ factors derived from measurements made in a representative set of chambers at several hospitals using portable water calorimeters.
In Japan, Korea and Taiwan, dosimetry is currently based on 60Co, but the standards laboratories have installed linacs and are in the process of setting up calibration services. Whether the clinics use these services directly, or whether the services are used to check 60Co based dosimetry, remains to be seen.
In summary, nearly all external beam radiotherapy around the world is traceable to primary standards of absorbed dose realised on 60Co and occasionally MV linacs, using a variety of protocols.
Cobalt-60 versus linac calibration service
In Australia, nearly all external beam radiotherapy is delivered using linacs with accelerating potentials in the range 4–18 MV [13]. Beam output dose measurements are performed with either a Farmer chamber (model 2571) or a chamber of similar design (“Farmer-type”). The PTW 30013 and IBA FC65-G are the most common chambers in addition to the 2571. As mentioned, these chambers are calibrated for absorbed dose to water in 60Co but AR-PANSA can now calibrate them directly in linac beams. Here we consider the differences in the two calibration chains and in the results.
Firstly, 60Co gamma rays are continuous, have relatively low dose rate in radiotherapy, and have a lower average energy than medical linac beams. These differences are summarised in Table 1.
Table 1.
Differences between 60Co gamma radiation and MV linac X-rays (nominal values only)
| Quantity | 60Co | MV linac X-rays |
|---|---|---|
| Spectrum | Two gammas plus scatter | Bremsstrahlung plus scatter |
| Nominal energy | 1.3 MeV | 1–25 MeV |
| Nominal Dw rate | 1–10 mGy/s | 60 mGy/s |
| Peak Dw rate | 10 mGy/s | 60,000 mGy/s |
| Type | Continuous | Pulsed |
| Output over a few minutes | Constant | Ramp up then usually constant to within about 3 % |
Using a calculated factor to take into account the different response in a linac beam is therefore an extrapolation which involves assumptions about the dosemeter. The ionisation chamber geometry is assumed to be the same as that used for the calculation of kQ. The response of the electrometer to a pulsed current is assumed to be the same as for a constant current. These assumptions have been shown to hold nearly all of the time for modern radiotherapy dosemeters [7]. Nevertheless, there is always the risk that a chamber or electrometer does not conform to these assumptions, either because it is damaged, has been manufacturer differently, or the design is flawed.
In a direct calibration, the response of the dosemeter is determined by interpolation of kQ for different TPR20,10 values. The beam qualities, dose rates and pulse behaviour are qualitatively similar for all medical linacs. No knowledge of the chamber construction or behaviour is required, and almost any chamber type can be calibrated.
The second, related point to consider is the uncertainty in clinical dose measurements derived from each method. Here we intend ‘clinical doses’ to refer to all doses delivered by the clinic, i.e. patient doses and doses delivered under reference conditions. Using 60Co with factors from TRS-398, for example, the uncertainty in the calibration coefficient of a Farmer chamber at megavoltage photons is around 1.1 % (standard uncertainty k = 1). Using the direct calibration method this can be reduced to around 0.6 %. Clinical doses determined via the direct calibration route should therefore be more accurate (although we note that the use of more accurate Monte Carlo calculated kQ factors such as those recently included in TG-51 [9] would reduce this improvement).
While the absolute accuracy of dose measurements will increase with the direct calibration service, it is less clear that the overall consistency in clinical doses will also increase. That is, random contributions to calibration factors on a linac are likely to be worse than for 60Co where the beam is more reproducible. The agreement between standards laboratories of different countries is slightly worse for linacs than it is for 60Co, as can be seen in the results from the Key Comparison Database [14] (accessed in February 2014) (Table 2). This database records the results of comparisons between standards laboratories around the world and the central primary standards laboratory, the Bureau International des Poids et Mesures (BIPM). The standard deviation of the ratios of absorbed dose to water for the BIPM.RI(I)-K6 comparison (linacs) for four laboratories and three photon beams is 0.0038 (0.38 %). The same ratio for 60Co absorbed dose to water in BIPM.RI(I)-K4 (for 60Co) is 0.0020 (0.20 %) for thirteen laboratories (Table 2).
Table 2.
Results for key comparisons of absorbed dose to water in linac beams (three energy ranges and four laboratories) and in 60Co (13 laboratories)
| Lab, i | Country | TPR20,10 | Ratio of Dw (Lab, i/BIPM) xi |
Standard uncertainty in ratio ui |
Date of measurement |
|---|---|---|---|---|---|
| NRC | Canada | 0.681 | 0.9973 | 0.0055 | 2009 |
| PTB | Germany | 0.683 | 1.0013 | 0.0052 | 2010 |
| NIST | US | 0.674 | 1.0035 | 0.0057 | 2010 |
| LNE-LNHB | France | 0.675 | 0.9952 | 0.0044 | 2012 |
| NRC | Canada | 0.731 | 1.0008 | 0.0055 | 2009 |
| PTB | Germany | 0.733 | 1.0034 | 0.0057 | 2010 |
| LNE-LNHB | France | 0.749 | 0.9948 | 0.0047 | 2012 |
| NRC | Canada | 0.800 | 0.9942 | 0.0055 | 2009 |
| PTB | Germany | 0.798 | 1.0018 | 0.0064 | 2010 |
| NIST | US | 0.783 | 0.9958 | 0.0059 | 2010 |
| Standard deviation of linac comparison ratios xi = 0.37 % | |||||
| NIST | US | Co-60 | 0.9984 | 0.0051 | 1997 |
| METAS | Switzerland | Co-60 | 0.9999 | 0.0054 | 2000 |
| MKEH | Hungary | Co-60 | 0.9983 | 0.0048 | 2001 |
| PTB | Germany | Co-60 | 0.9961 | 0.0037 | 2005 |
| VSL | Netherlands | Co-60 | 0.9926 | 0.0049 | 2005 |
| ENEA | Italy | Co-60 | 0.9999 | 0.0044 | 2007 |
| NPL | UK | Co-60 | 0.9980 | 0.0064 | 2007 |
| BEV | Austria | Co-60 | 0.9996 | 0.0044 | 2009 |
| VNIIFTRI | Russia | Co-60 | 0.9976 | 0.0043 | 2009 |
| NRC | Canada | Co-60 | 0.9980 | 0.0052 | 2009 |
| NMIJ | Japan | Co-60 | 0.9960 | 0.0046 | 2009 |
| ARPANSA | Australia | Co-60 | 0.9973 | 0.0053 | 2010 |
| LNE-LNHB | France | Co-60 | 0.9971 | 0.0039 | 2013 |
| Standard deviation of 60Co comparison ratios xi = 0.20 % | |||||
Compiled from the online key comparison database [14], February 2014
Other differences between 60Co and linac based calibrations are practical. A 60Co unit costs much less than a linac. The linac has higher maintenance costs and presumably greater down-time, but also a potentially longer life, as 60Co can only be maintained for around 10 years before the source decays to below a usable level. The linac also requires more detailed quality assurance measurements to make sure the beams are symmetric and the energy is constant, while 60Co requires physical security measures and presents difficulties in transport and disposal at its end of life. We estimate the cost of the linac to around ten times that of 60Co. Nevertheless, the linac brings many capabilities beyond calibrating ionisation chambers (research, development of radiotherapy dosimetry auditing procedures, liaison with the medical physics community) which should be considered in any cost-benefit analysis.
In 60Co, the calibration procedure uses the source output measured by the primary standard and corrected for source decay. An ARPANSA ionisation chamber is used to check this rate during the calibration. The direct calibration service uses a secondary standard chamber, against which the user chamber is cross-calibrated at three energies. The total amount of time for each calibration (60Co or 3 linac beams) is of the same order.
Validation of the new service
Before introducing the new calibration service, ARPANSA undertook three international comparisons and a literature survey of measured kQ values. ARPANSA also conducted a field trial to test the service, seek feedback, and to allow enough time to establish baselines for quality control.
Comparisons with other measurements of absorbed dose in linac beams
In September 2012, a team from the BIPM visited ARPANSA with a primary standard graphite calorimeter to perform absorbed dose measurements on the ARPANSA linac. These measurements were compared with ARPANSA’s measurements and the resulting comparison was published recently [15]. ARPANSA also took part in two indirect bilateral comparisons where ionisation chambers were measured at ARPANSA and in another laboratory. These comparisons are less direct because the chambers are calibrated on different linacs and some interpolation of the results is required. The results showing the ratio of the ARPANSA dose to water calibration coefficient to the other laboratory is shown in Table 3 and Fig. 1, for all three comparisons. Most of the ratios fall within one standard uncertainty of unity, indicating reasonable agreement. The agreement is worst for the highest energy beam, but still within two standard uncertainties.
Table 3.
Results of three international comparisons of absorbed dose to water (1) the Key Comparison BIPM.RI(I)-K6 ARPANSA and BIPM [15], (2) ARPANSA and NRCC (Canada) and (3) ARPANSA and National Measurement Institute (Japan) NMTJ [16]. R is the ratio of the absorbed dose to water realised by ARPANSA to the value realised by the other laboratory
| Nominal accelerating voltage (MV) |
Measured TPR20,10 |
Ratio R ARP/BIPM |
uc(R)/R | Ratio R ARP/NRCC |
uc(R)/R | Ratio R ARP/NMIJ |
uc(R)/R |
|---|---|---|---|---|---|---|---|
| 6 | 0.673 | 0.9965 | 0.0055 | 0.9954 | 0.006 | 1.0000 | 0.007 |
| 10 | 0.734 | 0.9924 | 0.0060 | 0.9928 | 0.006 | 0.9972 | 0.007 |
| 18 | 0.777 | 0.9932 | 0.0059 | 0.9920 | 0.006 | 0.9930 | 0.007 |
Fig. 1.
Results of international comparisons of absorbed dose to water (Table 3) between ARPANSA and three other laboratories. The results for NRC and NMTJ have been interpolated to the ARPANSA beam qualities shown on this graph. A small horizontal offset was used to separate the data
Comparison with measured and calculated 2571 kQ values
As a second form of method validation, we compare the kQ factors for the 2571 chamber type obtained by ARPANSA with published values. This factor is known to vary by only a small amount from chamber to chamber [7]. Measured kQ’s are plotted in Fig. 2 for the Netherlands [12], Canada [7], UK [12], France [12] and ARPANSA. Also shown are the calculated values from TRS-398 [4] and the more recent Monte Carlo calculations of Muir and Rogers [17]. The results are consistent with the comparisons results given in the previous section: ARPANSA is lower than the average but agrees within the spread of international values.
Fig. 2.
Correction factors kQ for the 2571 chamber type measured by different laboratories on medical linacs (The Netherlands [12], Canada [7], UK [12], France [12] and Australia). Values from TRS-398 [4] and Monte Carlo calculations of Muir and Rogers [17] are also shown. Standard uncertainties are shown for ARPANSA and TRS-398, the other points have uncertainties in the range 0.5–1.0 %
Uncertainties
The uncertainty in the user chamber calibration coefficient is given in Table 4 (based on a calibration in 60Co and use of TRS-398) and Table 5 (based on calibration in the three ARPANSA reference linac beams and interpolation with TPR20,10). The uncertainty in the calibration coefficient is significantly smaller for the direct calibration method.
Table 4.
Uncertainty in the calibration coefficient of an ionisation chamber at the user quality Q when based on a 60Co calibration
| Source of uncertainty | u (%) |
|---|---|
| ARPANSA calibration coefficient ND,w,Co-60 | 0.4 |
| TRS-398 kQ | 1.0 |
| Combined relative standard uncertainty (k = 1) | 1.1 |
Table 5.
Uncertainty in the calibration coefficient of an ionisation chamber at the user quality Q when based on calibration in the ARPANSA reference qualities Q0
| Source of uncertainty |
u (%) |
||
|---|---|---|---|
| 6 MV | 10 MV | 18 MV | |
| ARPANSA calibration coefficient ND,w,Q0 | 0.44 | 0.49 | 0.49 |
| Interpolation to user qualities | 0.1 | 0.1 | 0.1 |
| Spectral differences between calibration and clinical beama | 0.40 | 0.40 | 0.40 |
| Combined relative standard uncertainty (k= 1) | 0.62 | 0.65 | 0.65 |
Based on modelled differences in ionisation chamber response for beams of the same TPR20,10 but different spectra [18]
Use of direct linac calibration service with TRS-398
The TRS-398 dosimetry protocol recommends the use of direct calibrations where available, over 60Co, for both megavoltage photons and electrons. Our uncertainty calculation indicates that the resulting clinical doses will have lower uncertainties than those based on 60Co.
Section 6.5.2 of TRS-398 concerns the use of a measured kQ. It recommends that the user interpolate to the user energies. A method of interpolation is not specified, but ARPANSA provides a simple quadratic fit. It also recommends the use of reference quality Q0 and normalising other calibration coefficients to this value to produce a set of kQ,Qo correction factors. ARPANSA has adopted this recommendation and uses 60Co for Qo.
Shift in absorbed dose when adopting the new service
Of particular interest to radiotherapy clinics is the shift in patient dosimetry that would accompany adoption of the new service. While clinics will welcome the reduction in uncertainty that comes with the new service, they will also be mindful of time lost checking discontinuities in quality control data and the need to re-establish some baseline measurements.
Figure 3 shows the ratio of the factor kQ measured by ARPANSA, to that tabulated in TRS-398, for the three most common chamber types in use in Australia and for three representative beam qualities (nominally 6, 10 and 18 MV). The measured kQ’s are the average of two different chambers of each type.
Fig. 3.
Ratio of the kQ’s measured by ARPANSA to the values tabulated in TRS-398, for three common chamber types and three beam qualities
The differences shown in Fig. 3 will translate directly into differences in clinical linac output measurements. For example, a clinic using a 2571 chamber should expect a dosimetry shift of nearly 0.5 and 1.2 % at 6 and 18 MV, respectively. The shift at all energies will lower ND,w values compared to those based on 60Co. The consequence of this shift is that linacs will be calibrated to deliver more radiation when based on dosimetry from the new (lower) ND,w coefficients. The magnitude of the shift is within the combined standard uncertainties of the two calibration methods, however the change is not insignificant from a user perspective.
Discussion
Traceability and consistency in radiotherapy dosimetry are important for several reasons. We discuss these issues with regard to the different uncertainty and shift in dosimetry which accompanies the new calibration service.
Dose accuracy requirements
Differences in clinical doses which arise due to differences in traceability are usually not reported. A recent article by Andreo is an exception [19]. Nevertheless, improvements in the consistency and accuracy of ionisation chamber calibrations are considered to be important in the clinic. For example, dose errors of more than 5 % have been shown to have measurable effects on patient outcome for some treatments [20]. This implies that ionisation chambers need to be calibrated to better than 2.5 % standard uncertainty if all other contributions to the clinical uncertainty are zero, and much more accurately than this if clinical delivery uncertainties are taken into account. Boyer and Schultheiss [21] conclude that 1 % in dose correlates with about a 2 % change in early-stage tumour control. A more recent analysis by Thwaites et al. [22] concludes that 3 % standard uncertainty in the dose delivered to the patient can be taken as the “currently recommended general accuracy requirement”. Further discussion can be found in references [20–23]. These arguments support the need for more accurate absorbed dose standards.
The origin of clinical dose prescriptions
Radiotherapy prescriptions are in most cases derived from clinical trials. A wide variety of dose fractionation schemes are employed for the same clinical scenario [24, 25]. Sometimes the different prescriptions arise from different intent (the level of complications considered acceptable versus the degree of tumour control, for example), other times because they are based on different trials, or different historical practices and experience.
In all of these trials, the traceability of the dosimetry in the trial is difficult to extract, and is considered to be below the threshold that would influence the trial results. Dosimetry based on air kerma standards in 60Co is known to result in linac outputs which differ by up to a percent when compared to dosimetry from 60Co absorbed dose to water standards [26, 27]. Differences of this order between countries are also possible because the primary standards can differ by this amount, even absorbed dose standards. It is therefore not possible to argue for or against the use of 60Co from a prescription point of view.
Absolute dose accuracy and treatment repeatability
The ability to measure absorbed dose to water in absolute terms is required for radiotherapy only because it is the quantity chosen for prescriptions. Prescriptions need to be made in a quantity that (a) can be realised with a small uncertainty, (b) can be measured consistently and internationally, and (c) has a response which is proportional to tissue and tumour response. Absorbed dose to water is used mainly because it fulfils these criteria, but its relationship to tissue response needs to be established by clinical trials. It is possible to conceive of more appropriate quantities (such as the number of double-strand breaks per cell) which might be more proportional to tissue response. It is also possible to conceive of quantities that could be realised more consistently around the world (such as the ionisation in an air filled chamber). We make this comment to note that improving the accuracy of absorbed dose standards is not necessarily what is required by the radiotherapy profession. For example, radiotherapy doses might be more consistent if based on 60Co and agreed values of kQ were adopted and prescribed for use, even if the absorbed doses were incorrect. We believe, however, that the best way to improve consistency in radiotherapy is to improve the measurement of absorbed dose. By basing radiotherapy on a physical quantity, dose prescriptions will always have a physical meaning. This will not only help prevent misunderstandings with existing treatments, but will ensure that new treatment modalities have a framework for dosimetry.
Clinical doses in Australasia versus North America
In addition, we can look at linac output in Australia and New Zealand compared to the rest of the world. The Radiological Physics Centre (RPC) [28] in Houston regularly conducts remote audits of linac output around the world. Their results for photon beams audited in Australasia between 1999 and 2013 are shown in Fig. 4. These audits used TLDs up until 2010 and OSLDs afterwards. The same data for North America have a mean of 0.998 and standard deviation of 0.015. The mean of the Australasian data is 0.989 with standard deviation 0.015 from 448 measurements. The standard deviation of the mean is 0.07 % (less for North America) which indicates that the results are statistically significant: audited Australasian clinical doses are slightly lower than in North America. Differences in dosimetry protocols and primary standards in Australasia and North America account for some of the difference. Other differences, such as reference temperature of 22 °C in North America versus 20 °C in Australia, and potential differences in reporting dose to tissue verses dose to water, are not expected to have an effect. In 2013, a comparison conducted with a single ionisation chamber between the RPC and ARPANSA showed only 0.16% difference at 60Co. Nevertheless, based on these audit data, we can expect that Australian linacs will be closer to those of North America if the output is increased by about 1 %.
Fig. 4.
Ratio of the RPC-measured dose and clinic stated dose for remote output measurements, for all radiotherapy clinics audited by the RPC in Australasia. All photon beam audit results are shown. Repeat measurements and beam energy are not differentiated. Data courtesy of the RPC, Houston, USA
Monte Carlo kQ factors
A final comment on the adoption of the direct calibration service concerns the accuracy of the calculated kQ factors in TRS-398. These are known to differ from recent measurements and calculations by a small amount (within their stated uncertainty) [19], mainly because the chamber stem was not included in the calculation of kQ. When Monte Carlo calculations are used to derive kQ factors and the stem is included, the results are in better agreement with AR-PANSA’s and other laboratories’ measured values. Figure 5 shows the Monte Carlo calculated kQ factors calculated by Muir and Rogers [17] for the most common chambers used in Australia. These values are also included in Fig. 2 for the 2571 chamber. The results indicate that at least some of the difference between TRS-398 used with 60Co and the ARPANSA direct calibration is due to the TRS-398 kQ factors.
Fig. 5.
Ratio of kQ’s calculated using Monte Carlo techniques by Muir and Rogers [17] to the values tabulated in TRS-398, for five common chamber types, over all the TPR20,10 values in TRS-398. Nominal accelerating potential ranges are shown as vertical stripes
Flattening Filter Free (FFF) beams
The use of the TRS-398 Code of Practice with FFF beams has been shown to introduce a small additional error of around 0.5 % [29], because the relationship between TPR20,10 and stopping power ratios is different for FFF beams compared to flattened beams. This difference results in a different chamber response for the same TPR20,10, depending on whether the beam is a normal flattened beam or an FFF beam. A chamber calibrated directly on flattened beams will therefore entail the same error on an FFF beam as using a chamber calibrated at 60Co, when the beam quality index is TPR20,10. Possible solutions include reporting the calibration results in terms of %dd(10)x, which has a better correlation with stopping powers, so that the measured kQ ‘s could be used with TG-51 [8] for FFF beams, or ARPANSA could implement FFF beams on their linac and provide direct calibrations for these beams.
Conclusion
The ARPANSA calibration service for megavoltage linac photons has been described. The results of three comparisons and an indirect comparison of kQ factors for the 2571 chamber indicate that ARPANSA dosimetry is within the estimated uncertainty of other international standards of absorbed dose. The shift in clinical dosimetry when using the service instead of 60Co is expected to be in the range 0.5–1.3 %, with directly measured calibration coefficients being lower and clinical doses being higher, than for 60Co, when using the TRS-398 Code of Practice over the range TPR20,10 = 0.62–0.80. The advantages of the new service have been discussed in the light of the clinical impact of changing the service, with the conclusion that the move is justified based on the improvements in dose accuracy. Radiotherapy facilities can use this information to decide whether to shift to the new service when the additional cost is taken into account.
Acknowledgments
The authors wish to thank staff from the primary standards laboratories who performed comparisons with ARPANSA during the development of this service (BIPM, NMIJ and NRC), and Rebecca Day from the Wellington Blood and Cancer Centre in New Zealand, who first raised our awareness of the possibility of using audit data to produce Fig. 4. The Australian Clinical Dosimetry Service is a joint initiative between the Department of Health and ARPANSA. The Radiological Physics Center is supported by grant CA 10953 from the National Cancer Institute, NIH, DHHS.
Contributor Information
D. J. Butler, Email: Duncan.Butler@arpansa.gov.au, Australian Radiation Protection and Nuclear Safety Agency, 619 Lower Plenty Road, Yallambie, VIC 3085, Australia.
G. Ramanathan, Australian Radiation Protection and Nuclear Safety Agency, 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
C. Oliver, Australian Radiation Protection and Nuclear Safety Agency, 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
A. Cole, Australian Radiation Protection and Nuclear Safety Agency, 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
J. Lye, Australian Radiation Protection and Nuclear Safety Agency, 619 Lower Plenty Road, Yallambie, VIC 3085, Australia Australian Clinical Dosimetry Service, 619 Lower Plenty Road, Yallambie, VIC 3085, Australia.
P. D. Harty, Australian Radiation Protection and Nuclear Safety Agency, 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
T. Wright, Australian Radiation Protection and Nuclear Safety Agency, 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
D. V. Webb, Australian Radiation Protection and Nuclear Safety Agency, 619 Lower Plenty Road, Yallambie, VIC 3085, Australia
D. S. Followill, Radiological Physics Centre (now the Imaging and Radiation Oncology Core Quality Assurance Centre), M. D. Anderson Cancer Center, 1515 Holcombe Blvd, Houston, TX 77030, USA
References
- 1.Domen SR, Lamperti PJ. A heat-loss-compensated calorimeter: theory, design and performance. J. Res. Nat. Bur. Stand. 1974;78A:595–610. doi: 10.6028/jres.078A.037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Ramanathan G, Harty P, Wright T, Lye J, Butler D, Webb D, Huntley R. The Australian Primary Standard for absorbed dose to water (graphite calorimeter). Technical Report No. 166. Yallambie: Australian Radiation Protection and Nuclear Safety Agency; 2013. [Google Scholar]
- 3.Lye JE, Butler DJ, Franich RD, Harty PD, Oliver CP, Ramanathan G, Webb DV, Wright T. Direct MC conversion of absorbed dose to graphite to absorbed dose to water for 60Co radiation. Radiat Prot Dosimetry. 2013;155:100–109. doi: 10.1093/rpd/ncs276. [DOI] [PubMed] [Google Scholar]
- 4.Andreo P, Burns DT, Hohlfeld K, Huq MS, Kanai T, Laitano F, Smythe VG, Vynckier S. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. IAEA Technical Report Series No. 398. 2000 [Google Scholar]
- 5.IPSM Working Party. Lillicrap SC, Owen B, Williams JR, Williams PC. Code of practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service. Phys Med Biol. 1990;35:1355–1360. [Google Scholar]
- 6.IPEM Working Party. Thwaites DI, DuSautoy AR, Jordan T, McEwen MR, Nisbet A, Nahum AE, Pitchford WG. The IPEM code of practice for electron dosimetry for radiotherapy beams of initial energy from 4 to 25 MeV based on an absorbed dose to water calibration. Phys Med Biol. 2003;48:2929–2970. doi: 10.1088/0031-9155/48/18/301. [DOI] [PubMed] [Google Scholar]
- 7.McEwen MR. Measurement of ionization chamber absorbed dose kQ factors in megavoltage photon beams. Med Phys. 2010;37:2179–2193. doi: 10.1118/1.3375895. [DOI] [PubMed] [Google Scholar]
- 8.Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, Rogers DW. AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys. 1999;26:1847–1870. doi: 10.1118/1.598691. [DOI] [PubMed] [Google Scholar]
- 9.McEwen M, DeWerd L, Ibbott G, Followill DS, Rogers DW, Seltzer S, Seuntjens J. Addendum to the AAPM’s TG-51 protocol for clinical reference dosimetry of high energy photon beams. Med Phys. 2014;41:041501. doi: 10.1118/1.4866223. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.DIN 6800-2. Dosismessverfahren nach der Sondenmethode für Photoncn- und Elektronenstrahlung - Teil 2: Dosimetric hochenergetischer Photonen- und Elektronenstrahlung mit Ionisationskammern. 2008. Mar, [Google Scholar]
- 11.Audit of High-Energy Photon Beams in Belgian and Dutch Radiotherapy Departments. NEDERLANDSE COMMISSIE VOOR STRALINGSDOSIMETRIE. Report 23 of the Netherlands Commission on Radiation Dosimetry. 2013 Dec; 2013. [Google Scholar]
- 12.NCS report #18. Code of practice for the absorbed dose determination in high energy photon and electron beams, Delft, the Netherlands. 2008 www.radiationdosimetry.org.
- 13.Brown RL, Butler DJ. Results of the 2009 dosimetry survey of Australian radiotherapy centres. Australian Phys Eng Sci Med. 2010;33:285–297. doi: 10.1007/s13246-010-0031-7. [DOI] [PubMed] [Google Scholar]
- 14.The BIPM Key Comparison database: KCDB. [Accessed Feb 2014];Results for the BIPM.RI(I)-K6 comparison of absorbed dose to water at megavoltage linac energies and BIPM.RI(I)-K4 comparison of absorbed dose to water at 60Co. 2014 http://kcdb.bipm.org/
- 15.Picard S, Burns DT, Roger P, Harty PD, Ramanathan G, Lye JE, Wright T, Butler DJ, Cole A, Oliver C, Webb DV. Key Comparison BIPM.RI(I)-K6 of the standards for absorbed dose to water of the ARPANSA, Australia and the BIPM in accclerator photon beams. Metrologia 51 Tech. Suppl 06006. 2014 [Google Scholar]
- 16.Shimizu M, Morishita Y, Kato M, Tanaka T, Kurosawa T, Takata N, Saito N, Ramanathan G, Harty PD, Oliver C, Wright T, Butler DJ. Comparison of the NMIJ and ARPANSA standards for absorbed dose to water in high-energy photon beams. Radiat Prot Dosim. 2014 doi: 10.1093/rpd/ncu272. [DOI] [PubMed] [Google Scholar]
- 17.Muir BR, Rogers DWO. Monte Carlo calculations of kQ, the beam quality conversion factor. Med Phys. 2010;37:5939–5950. doi: 10.1118/1.3495537. [DOI] [PubMed] [Google Scholar]
- 18.Lye JE, Butler DJ, Ramanathan G, Franich RD. Spectral differences in 6 MV beams with matched PDDs and the effect on chamber response. Phys Med Biol. 2012;57:7599–7614. doi: 10.1088/0031-9155/57/22/7599. [DOI] [PubMed] [Google Scholar]
- 19.Andreo P, Wulff J, Burns DT, Palmans H. Consistency in reference radiotherapy dosimetry: resolution of an apparent conundrum when (60)Co is the reference quality for charged-particle and photon beams. Phys Med Biol. 2013;58:6593–6621. doi: 10.1088/0031-9155/58/19/6593. [DOI] [PubMed] [Google Scholar]
- 20.ICRU Report 24. Determination of absorbed dose in a patient irradiated by beams of X or gamma rays in radiotherapy procedures. International Commission on Radiation Units and Measurement, Washington, Section 7.1. Accuracy required in clinical dosimetry. 1976 [Google Scholar]
- 21.Boyer AL, Schultheiss T. Effects of dosimetric and clinical uncertainty on complication-free local tumor control. Radiother Oncol. 1988;11:65–71. doi: 10.1016/0167-8140(88)90046-1. [DOI] [PubMed] [Google Scholar]
- 22.Thwaites D. Accuracy required and achievable in radiotherapy dosimetry: have modern technology and techniques changed our views? 7th International conference on 3D radiation dosimetry. J Phys Conf Ser. 2013;444:012006. [Google Scholar]
- 23.Castro P, García-Vicente F, Mínguez C, Floriano A, Sevillano D, Perez L, Torres J. Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration. J Appl Clin Med Phys. 2008;9:70–86. doi: 10.1120/jacmp.v9i1.2676. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Radiotherapy Dose-Fractionation. Board of the Faculty of Clinical Oncology. Royal College of Radiologists; London: 2006. ISBN 1-905034-14-8. [Google Scholar]
- 25.Wu J, Wong R, Johnston M, Bezjak A, Whelan T. Meta-analysis of dose-fractionation radiotherapy trials for the palliation of painful bone metastases. Int J Radiat Oncol Biol Phys. 2003;55(3):594–605. doi: 10.1016/s0360-3016(02)04147-0. [DOI] [PubMed] [Google Scholar]
- 26.Andreo P, Huq MS, Westermark M, Song H, Tilikidis A, DeWerd L, Shortt K. Protocols for the dosimetry of high-energy photon and electron beams: a comparison of the IAEA TRS-398 and previous international codes of practice. Phys Med Biol. 2002;47:3033–3053. doi: 10.1088/0031-9155/47/17/301. [DOI] [PubMed] [Google Scholar]
- 27.Palmans H, Nafaa L, De JJ, Gillis S, Hoornaert MT, Martens C, Piessens M, Thierens H, Van der Plaetsen A, Vynckier S. Absorbed dose to water based dosimetry versus air kerma based dosimetry for high-energy photon beams: an experimental study. Phys Med Biol. 2002;47:421–440. doi: 10.1088/0031-9155/47/3/305. [DOI] [PubMed] [Google Scholar]
- 28.Ibbott GS. QA in radiation therapy: the RPC perspective. J Phys. 2010;250:1–7. [Google Scholar]
- 29.Xionga G, Rogers DWO. Relationship between %dd(10)x and stopping-power ratios for flattening filter free accelerators: a Monte Carlo study. Med Phys. 2008;35:2104–2109. doi: 10.1118/1.2905028. [DOI] [PubMed] [Google Scholar]





