Abstract
We present a novel analysis of arterial pulse wave propagation that combines traditional wave intensity analysis with identification of Windkessel pressures to account for the effect on the pressure waveform of peripheral wave reflections. Using haemodynamic data measured in vivo in the rabbit or generated numerically in models of human compliant vessels, we show that traditional wave intensity analysis identifies the timing, direction and magnitude of the predominant waves that shape aortic pressure and flow waveforms in systole, but fails to identify the effect of peripheral reflections. These reflections persist for several cardiac cycles and make up most of the pressure waveform, especially in diastole and early systole. Ignoring peripheral reflections leads to an erroneous indication of a reflection-free period in early systole and additional error in the estimates of (i) pulse wave velocity at the ascending aorta given by the PU–loop method (9.5% error) and (ii) transit time to a dominant reflection site calculated from the wave intensity profile (27% error). These errors decreased to 1.3% and 10%, respectively, when accounting for peripheral reflections. Using our new analysis, we investigate the effect of vessel compliance and peripheral resistance on wave intensity, peripheral reflections and reflections originating in previous cardiac cycles.
Keywords: haemodynamics, pulse wave propagation, wave intensity analysis, one-dimensional modelling, Windkessel effect, PU–loop method, systemic circulation
1 INTRODUCTION
Blood pressure and flow waveforms in systemic arteries carry valuable information for the diagnosis and treatment of cardiovascular disease and play a significant role in clinical conditions such as hypertension. The waveforms result from a complex ventricular-vascular interaction involving cardiac contraction, impedance of large and medium-sized distensible arteries and resistance of smaller arteries and arterioles. Blood behaves as an incompressible fluid in arteries, which distend to accommodate the sudden increase in blood volume delivered by cardiac contraction. When elastic energy stored in the distended arterial walls is released, arteries contract. The regular expansion and contraction of arteries (the pulse) that follows cardiac contraction propagates in the form of pulse waves. These produce continuous changes in blood pressure and flow that can be studied as pressure and flow wavefronts (infinitesimal changes in pressure and flow)‡ running forwards and backwards (away from and towards the heart, respectively), with backward wavefronts originating from reflected forward wavefronts at sites of vascular impedance mismatch.
Figure 1 shows typical blood pressure waveforms measured in vivo along human (left) and rabbit (right) aortas, from the root to the aorto-iliac bifurcation, under normal conditions. The slope of the line joining the feet of these waveforms shows clearly that the pressure wavefront originated at the start of cardiac contraction propagates away from the heart; the measured space-averaged speed is 6.9 m s − 1 in the human and 6.1 m s − 1 in the rabbit. Thus, during a typical cardiac cycle, which takes about 1 s in the human and 0.25 s in the rabbit, a pulse wave has sufficient time to travel from the heart to the arterial vasculature and back multiple times.
Several studies have used wave intensity analysis (WIA) to investigate the role of wave reflections in shaping in vivo pressure and flow waveforms in systemic arteries 2–11, including the coronary circulation 12–14. Given simultaneous measurements of blood pressure and flow velocity with time at an arbitrary location in the arterial network, we can calculate the local pulse wave velocity (PWV) and apply WIA to quantify the timing, direction and magnitude of the predominant waves that shape the pressure and velocity waveforms 15,16. Accurate estimation of PWV is not only important for WIA but is also clinically relevant, because PWV is an important predictor of cardiovascular events 17.
Numerical modelling has been used to assess the following: (i) the ability of WIA to quantify reflection coefficients 18; (ii) haemodynamic information provided by WIA in a model of aortic coarctation 19 and the fetal circulation 20; (iii) a modified WIA based on the reservoir-wave separation 21; and (iv) the performance of several methods for PWV calculation 22–25. Numerically generated pressure and flow waveforms are free of measurement errors, and the theoretical values of haemodynamic properties that affect waveforms (e.g. PWV, location of reflection sites, and magnitude of reflected waves produced) are available for comparison with corresponding estimates given by WIA and methods of calculating PWV.
In the present work, we used pressure and flow velocity waveforms measured in vivo in the rabbit or generated numerically in several models of human compliant vessels to (i) show the inability of traditional WIA to identify the important role of peripheral reflections in shaping the pressure waveform; (ii) test the accuracy of the modified PU–loop method of calculating PWV proposed by Mynard et al. 24, which accounts for peripheral reflections originating in previous cardiac cycles; and (iii) propose a new analysis of arterial pulse wave propagation to study the predominant waves that shape pressure and flow waveforms during systole and the contribution to the pressure waveform, over the whole cardiac cycle, of wave reflections originating in previous cardiac cycles, vessel compliances, peripheral resistances, outflow pressures and the flow at the root. We used our new analysis to study the effects of vessel stiffness and peripheral resistance on numerically-generated aortic pressure and flow waveforms.
We generated all numerical data using the nonlinear one-dimensional (1-D) formulation of blood flow in compliant vessels, because WIA is derived from this formulation and, hence, 1-D model pressure and velocity waveforms provide an ideal mathematical framework for our study. Several comparisons against in vivo 26–29, in vitro 30–34 and 3-D numerical 35 data have shown the ability of the 1-D formulation to capture the main features of pressure and flow waveforms in large human arteries. The nomenclature and abbreviations used in this paper are listed in the supplementary material.
2 METHODS
We first describe the in vivo (Section 2.1) and numerical (Section 2.2) pressure and flow waveforms used in our work. Next, we summarise the mathematical formulation of traditional WIA (Section 2.3) and show how we quantified the magnitude and timing of wave reflections (Section 2.4). We then introduce the Windkessel model to study arterial haemodynamics during diastole and account for peripheral reflections in WIA and the PU–loop method (Section 2.5).
2.1 Rabbit in vivo pressure and flow waveforms
Ten New Zealand white male rabbits (Harlan UK) with the properties shown in Table I were maintained on a standard laboratory diet and housed at 18°C on a 12-h light cycle. The rabbits were pre-medicated with Hypnorm (0.1 ml kg − 1) intramuscularly and anaesthetised with sodium pentobarbitone (35 mg kg − 1), administered intravenously via the marginal vein of the right ear. They were then placed in the supine position and artificially ventilated with a Harvard Small Animal Ventilator (with a respiratory rate of 50 breaths min − 1; inflation pressure set for trough 25 cmH 2O and peak 50 cmH 2O). Body temperature was maintained near 39 °C (monitored using a rectal thermistor probe) by placing the rabbits on a heating blanket with a temperature controller (CWE Inc. TC1000). The animal procedures complied with the Animals (Scientific Procedures) Act (1986) and were approved by the Imperial College London local ethical review process.
Table I.
Rabbit | Age | W | C-R | HR | CO | Pm | PP | Pout | cff | cPU | ||
---|---|---|---|---|---|---|---|---|---|---|---|---|
No. | (days) | (kg) | (cm) | (beat/s) | (ml/s) | (kPa) | (kPa) | (kPa) | (s) | (m/s) | (m/s) | (m/s) |
1 | 100 | 3.22 | 43 | 3.7 | 3.1 | 8.1 | 2.9 | 3.0 | 0.32 | 4.3 | 3.9 | 4.3 |
2 | 114 | 3.20 | 41 | 2.9 | 2.3 | 5.2 | 2.8 | 2.2 | 0.37 | 4.5 | 4.2 | 4.4 |
3 | 93 | 2.55 | 40 | 4.0 | 2.4 | 6.9 | 2.8 | 2.3 | 0.28 | 4.7 | 4.2 | 4.6 |
4 | 72 | 2.29 | 40 | 4.5 | 2.6 | 6.2 | 2.2 | 2.2 | 0.28 | 5.1 | 4.6 | 5.1 |
5 | 92 | 2.67 | 40 | 4.7 | 4.2 | 7.7 | 3.4 | 2.8 | 0.18 | 4.5 | 4.0 | 4.5 |
6 | 136 | 2.54 | 42 | 3.1 | 4.1 | 6.0 | 2.4 | 1.3 | 0.28 | 3.5 | 3.0 | 3.4 |
7 | 143 | 3.00 | 41 | 3.4 | 4.9 | 8.6 | 3.1 | 2.3 | 0.30 | 4.5 | 4.2 | 4.6 |
8 | 114 | 3.26 | 43 | 3.6 | 4.0 | 10.5 | 2.8 | 4.5 | 0.32 | 4.4 | 4.0 | 4.4 |
9 | 159 | 3.48 | 45 | 4.7 | 4.4 | 8.1 | 2.6 | 1.7 | 0.34 | 4.3 | 3.9 | 4.3 |
10 | 133 | 3.13 | 41 | 5.2 | 5.4 | 8.5 | 3.0 | 3.1 | 0.26 | 3.9 | 3.5 | 3.8 |
Mean | 116 | 2.93 | 41.6 | 4.0 | 3.7 | 7.6 | 2.8 | 2.5 | 0.29 | 4.37 | 3.96 | 4.34 |
SEM | 9 | 0.12 | 0.5 | 0.2 | 0.3 | 0.5 | 0.1 | 0.3 | 0.02 | 0.13 | 0.14 | 0.14 |
We calculated Pout and by fitting an exponential function of the form given by Equation (19) to the decline in pressure during diastole, with T0 the time at the beginning of the exponential fit. The last two rows are mean and standard error of the mean (SEM) for all ten rabbits.
For each rabbit, a midline thoracotomy was made and the rib cage was retracted to expose the ascending aorta and place a perivascular blood flow probe (6 mm diameter, type MA6PSB, Transonic Systems Inc.) around it, near the aortic valve. The thoracotomy was closed using clamps. The right femoral artery was exposed at the level between the knee and groin for insertion of pressure measurement wires. Two pressure waveforms were simultaneously measured with the aortic flow waveform, from the aortic root to the iliac artery in 1 cm increments (Figure 1(b)) using a dual sensor Millar Mikro-tip catheter transducer (model SPC-721, size 2.5F, sensors 5 cm apart). The aortic pressure waveforms at the root and 5 cm distally (Figure 2(b)) allowed us to calculate the PWV using the ‘foot-to-foot’ method (cff) 25, which we used as the gold standard PWV at the aortic root (Table I). Brief asystole (cessation of heart contraction) were induced by gently tapping the left ventricle (LV) (Figure 2(a,b)). All data were acquired at a sampling rate of 1 kHz using the NOTOCORD-hem acquisition software (NOTOCORD Systems, France).
Prior to euthanasia, heparin (2,000 units) was administered intravenously for anticoagulation of the blood in preparation for production of a resin cast of the arterial system (Figure 2(e)). The thorax was reopened and the LV was cannulated with a polythene tube (filled with saline solution: 9 g/L NaCl) via an apical stab wound. The cannula was flushed through with approximately 5 ml saline. Casting resin (Batson's no. 17, Plastic Replica and Corrosion kits, Polyscience, USA, prepared according to the manufacturer's instructions) was infused into the arterial system via the cannula, at a pressure equivalent to the mean arterial pressure that was recorded in vivo in the thoracic aorta. Infusion was continued until the casting resin ceased to flow because of setting of the resin; the final perfusion volume was approximately 100 ml of resin. The cast was allowed to cure for at least 12 h; then the carcass was completely submerged in an aqueous solution of potassium hydroxide (25% w/v). The alkaline corrosion process was allowed to continue for 14 days. Then the corrosion solution was removed to reveal the arterial cast, which was submerged in a strong warm solution of detergent (Decon 90, Decon Laboratories Ltd., East Sussex, UK) and left for 24 h. The cast was rinsed gently and thoroughly in water and allowed to dry.
We used the cast to locate accurately the sites of pressure measurements and determine the luminal cross-sectional area of the aortic root, which is required to convert measured flow rate to flow velocity; the latter, combined with simultaneous pressure at the root, allowed WIA.
2.2 Human numerical pressure and flow waveforms
To generate the numerical pressure and flow data for this study, we solved the nonlinear 1-D equations of blood flow in compliant vessels in a single-vessel model of the human thoracic aorta 35 (Figure 3(a,b,c), Table II) and also in a model of the 55 larger systemic arteries in the human 1 (Figure 3(d,e,f,g,h), Table III), using a DG scheme with a spectral/hp spatial discretisation 32. These equations can be derived by applying conservation of mass and momentum to a differential 1-D control volume of the vessel 36,37,
1 |
where t is the time, x is the distance along the vessel, A(x,t) is the area of the luminal cross section of the vessel, U(x,t) is the axial blood flow velocity averaged over the cross section, P(x,t) is the blood pressure averaged over the cross-section, ρ is the density of blood (assumed to be constant), and f(x,t) = − 22πμU is the frictional force per unit length, with μ = 4 mPa s the viscosity of blood.
Table II.
Parameter | Value |
---|---|
Length, l | 24.137 cm |
Radius at diastolic pressure, rd | 1.2 cm |
Wall thickness, h | 1.2 mm |
Blood density, ρ | 1,060 Kg m − 3 |
Young's modulus, E | 400.0 kPa |
Mean flow rate, | 6.170 l min − 1 |
Windkessel resistance, R1 | 11.75 Pa s cm − 3 |
Windkessel compliance, C | 10.16 mm 3 Pa − 1 |
Windkessel resistance, R2 | 111.67 Pa s cm − 3 |
The resulting wave speed at mean pressure is 5.2 m s − 1.
Table III.
Mean | Mean | Peripheral | Peripheral | Wall | ||||
---|---|---|---|---|---|---|---|---|
Arterial segment | Length | pressure | flow | resistance | compliance | viscosity | ||
name | (cm) | (mm) | (m s − 1) | (kPa) | (ml s − 1) | (GPa s m − 3) | (m 3 GPa − 1) | (kPa s) |
1. Ascending aorta | 5.8 | 15.4 → 15.4 | 4.0 → 4.0 | 13.34 | 102.8 | — | — | 0.5 |
2. Aortic arch A | 2.3 | 13.2 → 12.6 | 4.2 → 4.2 | 13.33 | 89.3 | — | — | 0.5 |
3. Brachiocephalic | 3.9 | 10.6 → 9.4 | 4.5 → 4.6 | 13.32 | 13.5 | — | — | 1.0 |
4. R. subclavian | 3.9 | 6.0 → 4.7 | 5.3 → 5.7 | 13.32 | 7.0 | — | — | 1.0 |
5. R. common carotid | 10.8 | 5.7 → 2.9 | 5.3 → 6.5 | 13.29 | 6.5 | — | — | 6.0 |
6. R. vertebral | 17.1 | 1.9 → 1.4 | 8.1 → 8.7 | 11.98 | 2.3 | 4.51 | 0.090 | 6.0 |
7. R. brachial | 48.5 | 4.2 → 2.4 | 6.4 → 7.5 | 12.68 | 4.7 | — | — | 2.5 |
8. R. radial | 27.0 | 1.9 → 1.6 | 8.0 → 8.4 | 10.82 | 2.4 | 3.96 | 0.099 | 6.0 |
9. R. ulnar A | 7.7 | 1.9 → 1.7 | 8.0 → 8.2 | 12.25 | 2.3 | — | — | 6.0 |
10. R. interosseous | 9.1 | 1.1 → 0.9 | 9.5 → 10.0 | 11.89 | 0.2 | 63.22 | 0.032 | 6.0 |
11. R. ulnar B | 19.7 | 1.6 → 1.4 | 8.4 → 8.7 | 10.08 | 2.2 | 3.96 | 0.077 | 6.0 |
12. R. internal carotid | 20.5 | 2.9 → 2.2 | 7.1 → 7.7 | 12.29 | 5.8 | 1.88 | 0.259 | 6.0 |
13. R. external carotid | 18.7 | 1.3 → 0.8 | 9.1 → 10.4 | 9.53 | 0.7 | 10.42 | 0.193 | 6.0 |
14. Aortic arch B | 4.5 | 11.2 → 10.9 | 4.4 → 4.4 | 13.28 | 83.8 | — | — | 0.5 |
15. L. common carotid | 16.0 | 5.1 → 2.5 | 5.5 → 6.8 | 13.18 | 5.5 | — | — | 6.0 |
16. L. internal carotid | 20.5 | 2.2 → 1.7 | 7.7 → 8.2 | 11.1 | 5.1 | 1.88 | 0.189 | 6.0 |
17. L. external carotid | 18.7 | 1.0 → 0.6 | 9.8 → 11.1 | 7.21 | 0.5 | 10.42 | 0.173 | 6.0 |
18. Thoracic aorta A | 6.0 | 10.4 → 9.9 | 4.5 → 4.6 | 13.25 | 76.5 | — | — | 0.5 |
19. L. subclavian | 3.9 | 5.7 → 4.4 | 5.3 → 5.8 | 13.26 | 7.2 | — | — | 1.0 |
20. L. vertebral | 17.0 | 1.9 → 1.4 | 8.1 → 8.7 | 11.98 | 2.3 | 4.51 | 0.090 | 6.0 |
21. L. brachial | 48.5 | 4.2 → 2.4 | 6.4 → 7.5 | 12.64 | 4.9 | — | — | 2.5 |
22. L. radial | 27.0 | 1.8 → 1.4 | 8.2 → 8.7 | 10.25 | 2.2 | 3.96 | 0.085 | 6.0 |
23. L. ulnar A | 7.7 | 2.2 → 2.2 | 7.7 → 7.7 | 12.40 | 2.7 | — | — | 6.0 |
24. L. interosseous | 9.1 | 0.9 → 0.9 | 10.0 → 10.0 | 11.94 | 0.2 | 63.22 | 0.028 | 6.0 |
25. L. ulnar B | 19.7 | 2.1 → 1.9 | 7.8 → 8.0 | 11.54 | 2.6 | 3.96 | 0.130 | 6.0 |
26. Intercostals | 9.2 | 6.6 → 4.9 | 5.1 → 5.6 | 13.20 | 2.0 | 6.00 | 0.104 | 0.5 |
27. Thoracic aorta B | 12.0 | 8.6 → 6.7 | 4.7 → 5.1 | 13.05 | 74.5 | — | — | 0.5 |
28. Abdominal aorta A | 6.1 | 6.3 → 6.3 | 5.2 → 5.2 | 13.00 | 61.3 | — | — | 0.5 |
29. Celiac A | 2.3 | 4.1 → 3.6 | 5.9 → 6.1 | 13.05 | 13.2 | — | — | 0.5 |
30. Celiac B | 2.3 | 2.7 → 2.5 | 6.7 → 6.8 | 13.04 | 8.9 | — | — | 0.5 |
31. Hepatic | 7.6 | 2.8 → 2.3 | 7.2 → 7.6 | 12.97 | 4.3 | 2.72 | 0.205 | 2.5 |
32. Gastric | 8.2 | 1.6 → 1.5 | 8.4 → 8.5 | 12.22 | 2.6 | 4.06 | 0.082 | 6.0 |
33. Splenic | 7.2 | 2.2 → 2.0 | 7.7 → 7.9 | 12.45 | 6.3 | 1.74 | 0.140 | 6.0 |
34. Superior mesenteric | 6.8 | 4.1 → 3.7 | 5.9 → 6.1 | 13.00 | 16.7 | 0.70 | 0.481 | 1.0 |
35. Abdominal aorta B | 2.3 | 6.0 → 5.9 | 5.3 → 5.3 | 13.00 | 44.6 | — | — | 0.5 |
36. L. renal | 3.7 | 2.7 → 2.7 | 6.7 → 6.7 | 12.77 | 13.4 | 0.85 | 0.231 | 2.5 |
37. Abdominal aorta C | 2.3 | 6.1 → 6.1 | 5.2 → 5.2 | 12.98 | 31.2 | — | — | 0.5 |
38. R. renal | 3.7 | 2.7 → 2.7 | 6.7 → 6.7 | 12.72 | 13.4 | 0.85 | 0.231 | 2.5 |
39. Abdominal aorta D | 12.2 | 6.0 → 5.7 | 5.3 → 5.3 | 12.96 | 17.8 | — | — | 0.5 |
40. Inferior mesenteric | 5.8 | 2.4 → 1.6 | 7.5 → 8.4 | 12.92 | 2.2 | 5.16 | 0.133 | 2.5 |
41. Abdominal aorta E | 2.3 | 5.6 → 5.4 | 5.4 → 5.4 | 12.94 | 15.6 | — | — | 0.5 |
42. L. common iliac | 6.8 | 4.1 → 3.6 | 5.9 → 6.1 | 12.92 | 7.8 | — | — | 1.0 |
43. R. common iliac | 6.8 | 4.1 → 3.6 | 5.9 → 6.1 | 12.92 | 7.8 | — | — | 1.0 |
44. L. external iliac | 16.6 | 3.3 → 3.1 | 6.3 → 6.3 | 12.72 | 5.9 | — | — | 2.5 |
45. L. internal iliac | 5.8 | 2.1 → 2.1 | 7.9 → 7.9 | 12.78 | 1.9 | 5.96 | 0.137 | 6.0 |
46. L. femoral | 50.9 | 2.7 → 1.9 | 7.3 → 7.9 | 11.32 | 2.9 | — | — | 6.0 |
47. L. deep femoral | 14.5 | 2.1 → 1.9 | 7.9 → 8.0 | 12.02 | 3.0 | 3.58 | 0.127 | 6.0 |
48. L. posterior tibial | 36.9 | 1.6 → 1.4 | 8.4 → 8.6 | 7.77 | 1.8 | 3.58 | 0.074 | 6.0 |
49. L. anterior tibial | 39.8 | 1.3 → 1.1 | 8.9 → 9.1 | 6.08 | 1.1 | 4.19 | 0.051 | 6.0 |
50. R. external iliac | 16.6 | 3.3 → 3.1 | 6.3 → 6.3 | 12.72 | 5.9 | — | — | 2.5 |
51. R. internal iliac | 5.8 | 2.1 → 2.1 | 7.9 → 7.9 | 12.78 | 1.9 | 5.96 | 0.137 | 6.0 |
52. R. femoral | 50.9 | 2.7 → 1.9 | 7.3 → 7.9 | 11.32 | 2.9 | — | — | 6.0 |
53. R. deep femoral | 14.5 | 2.1 → 1.9 | 7.9 → 8.0 | 12.02 | 3.0 | 3.58 | 0.127 | 6.0 |
54. R. posterior tibial | 36.9 | 1.6 → 1.4 | 8.4 → 8.6 | 7.77 | 1.8 | 3.58 | 0.074 | 6.0 |
55. R. anterior tibial | 39.8 | 1.3 → 1.1 | 8.9 → 9.1 | 6.08 | 1.1 | 4.19 | 0.051 | 6.0 |
Mean pressures and flows calculated in the midpoint of the segment. The outflow pressure ( Pout) is 1.33 kPa at each terminal branch and the blood density is ρ = 1,050 Kg m − 3. R., right; L., left.
Both models are based on data from young and healthy humans. The single-vessel model is uniform and contains a single peripheral reflection site, which simplifies our initial assessment of WIA. The 55-artery model allows us to assess WIA in the presence of tapering and multiple outflows and reflection sites; all these properties have been shown relevant for analysis of pulse wave propagation phenomena 38–41.
To account for the fluid-structure interaction of the problem and close the system of equations (1), we obtained the following explicit algebraic relationship between P and A (or tube law). Assuming the arterial wall to be a thin, incompressible, homogeneous, isotropic, Voigt-type visco-elastic membrane, which deforms axisymmetrically, each cross-section independently of the others, we can relate blood pressure (P) to luminal cross-sectional area (A) through 32
2 |
with
3 |
where Pe is the elastic component of pressure, h(x) is the wall thickness, E(x) is the Young's modulus, ϕ(x) is the wall viscosity, and A0(x) is the reference area at P = 0 and . For some simulations, we considered a purely elastic tube law (i.e. P = Pe) as it is assumed in WIA.
We implemented all the boundary conditions of our simulations and solved matching conditions at bifurcations by taking into account the correct propagation of the characteristic information and neglecting energy losses and visco-elastic effects (see 1 for a detailed description).
Both models exhibit the following characteristic features of the pressure and flow waveforms that are observed in vivo under normal conditions. The foot of the pressure and flow waveforms in early systole propagates away from the heart (compare the simulated pressures in Figure 3(a,d) with the in vivo pressures in Figure 1(a,b)). The pulse pressure (the difference between the maximum, systolic, and minimum, diastolic, pressures) increases in the aorta with increasing distance from the heart (Figures 3(a,d) and 1(a,b)), whereas mean pressure gradually decreases§. In the ascending aorta, pressure features a ‘shoulder’ or point of inflection (Figures 3(a,d) and 1(a,b)). From the ascending aorta to the upper thoracic aorta, a small pressure peak is observed at the start of diastole, which forms the dicrotic notch. This vanishes in the lower thoracic region (Figures 3(d) and 1(a,b)), but it is observed in other proximal arteries such as the common carotid arteries (Figure 3(f)) 43. Moreover, the initial pressure increase becomes steeper and narrower in time in more peripheral locations (wave steepening) (Figures 3(a,d) and 1(a,b)), and a wide pressure peak appears in diastole from the abdominal aorta to the leg arteries 44,45 and in the brachial artery 46 (Figure 3(d,f)). With increasing distance from the heart, the aortic flow waveform (Figure 3(c,g)) becomes characterised by an increase in width, a reduction in the amount of reverse flow and a decrease in amplitude and mean value¶ 44,45. Reversed flow is absent in the suprarenal region of the aorta 29,47, renal arteries 44,48 and carotid arteries 29,49,50. There is, however, a region of reverse flow in early diastole in the infrarenal region of the aorta and leg arteries 47,50.
2.3 Traditional wave intensity analysis
Wave intensity analysis is derived from the system of equations (1). Velocity and pressure waveforms are decomposed into successive wavefronts, with dP and dU as changes in pressure and velocity, respectively, across a wavefront. Fluid viscous losses are assumed to be negligible locally, and A is assumed to depend only upon P through a purely elastic tube law, with uniform and constant properties. Under these conditions, Riemann's method of characteristics applied to the system of equations (1) shows (Figure 4(a)) that for any point (X,T) in the (x,t) space, there are two characteristic paths, Cf and Cb, defined by , on which Wf and Wb are constant 15. The quantities Wf and Wb are generally known as the characteristic variables or Riemann invariants and satisfy
4 |
where the pressure-dependent c is the PWV, that is the speed at which pulse wavefronts travel in the absence of convective velocity (U) (further details are given in Appendix A). For the purely elastic tube law given by Equation (3),
5 |
Under physiological flow conditions, c is much greater than the maximum U, so that U + c > 0 and U − c < 0 (i.e. the flow is subcritical). Therefore, changes in P and U propagate in the forward and backward directions (we define the forward direction as the direction of mean blood flow, in which x increases) with speeds of U + c and U − c, respectively.
In vivo measurements of P and U with time are typically taken at a fixed point x = X, rather than along a characteristic line. Solving the two equations in (4) at x = X for dP and dU yields
6 |
Wave intensity ( dI) is defined as 15,16
7 |
which is the flux of energy per unit area carried by the wavefront as it propagates and has dimensions of power/unit area and SI units W m − 2. Note that dI is calculated with time at a fixed point x = X. According to Equation (7), dI is positive if and negative if . Therefore, dI ‘measures’ the importance with time of changes in P and U in the forward and backward directions at x = X. Whenever dI > 0, forward changes in P and U dominate over backward changes; the flow is accelerated if dP > 0 and decelerated if dP < 0. Whenever dI < 0, backward changes in P and U dominate over forward changes; the flow is accelerated if dP < 0 and decelerated if dP > 0.
2.3.1 Application to measured data
Given simultaneous measurements of P(t) and U(t) at an arterial site of the numerical models or rabbit, we used a Savitzky–Golay filter to smooth dP(t) = P(t + dt) − P(t) and dU(t) = U(t + dt) − U(t), where dt is the time between two adjacent sampling points of P or U. This filter is commonly used for WIA because it preserves peaks in dP and dU (and hence dI) 16,51,52 ,p. 650. We normalised the value of dI given by Equation (7) by dt2 to make the magnitude of dI independent of the sampling frequency. We used customised matlab software (The MathWorks, Inc., MA, USA) for our data analysis.
2.3.2 Forward and backward waveforms
The measured waveforms P(t) and U(t) can be separated into forward-travelling (Pf(t), Uf(t)) and backward-travelling (Pb(t), Ub(t)) components, that is and . Separating dP and dU into changes across the forward ( dPf, dUf) and backward ( dPb, dUb) wavefronts, that is and , and using the water hammer equations,
8 |
yield 15
9 |
A derivation of Equations (8) from the system of equations (1) using the method of characteristics is given in 15. In Appendix A, we provide an alternative derivation of Equations (8) by directly applying conservation of mass and momentum to a control volume moving with the forward or backward pulse wavefronts.
If the PWV (c) is known, Equation (9) allows us to obtain Pf,b(t) and Uf,b(t) from the measured P(t) and U(t) by adding the differences dPf,b(t) and dUf,b(t), that is and . We considered the integration constants P0 and U0 to be half the pressure and velocity, respectively, at the end of diastole. Thus, wave intensity ( dI) at a fixed point x = X can be separated into forward ( dIf > 0) and backward ( dIb < 0) components,
10 |
We can calculate c from simultaneous measurements of P and U using the PU–loop 15,53 method (see Appendix B for more details) or from two measurements of P (or U) at two different sites using foot-to-foot, least squared difference or cross-correlation techniques 25.
2.4 Wave reflections
At sites of impedance mismatch, pulse waves are partly reflected and partly transmitted. Using a linearised version of the 1-D equations (1), we can relate the changes in pressure and flow velocity across the reflected wavefront ( dPref, dUref) in the parent and two daughter vessels of an arterial bifurcation to the corresponding changes in pressure and velocity in the incident wavefront ( dPinc, dUinc) through
11 |
The superscripts p, d1 and d2 refer to the parent and first and second daughter vessels, respectively. The reflection coefficients for wavefronts propagating in the parent and daughter ( and ) vessels can be expressed as a function of the characteristic admittance, , j = p,d1,d2, where c0 is the PWV at zero pressure (see Appendix C for a detailed derivation of Equations (11) and (12)),
12 |
At the outlet of each terminal branch coupled to a single resistance (R1) we have 42
13 |
where Pout is the outflow pressure, Rf = (R1 − Z0)/(R1 + Z0) is the local reflection coefficient and Z0 = 1/Y 0 is the characteristic impedance of the terminal branch. Note that R1 = Z0 yields Rf = 0, in which case dPinc and dUinc are completely absorbed by the outflow model.
From Equation (11) to Equation (13) with Pout = 0, we can relate the wave intensity of the reflected wavefront () to the wave intensity of the incident wavefront () through
14 |
Given a wave intensity profile separated into dIf and dIb, we defined the initial positive and negative regions as the incident and reflected waves, respectively (Figure 5). Note that these waves are made of many wavefronts across which wave intensity ( dI) changes. We then calculated the time of arrival ( and ) and magnitude (Iinc and Iref) of the incident and reflected waves using either peak values or area-average values; they are indicated by empty squares or stars, respectively, in Figure 5. Given the wavefronts that make up the incident, dIinc, or reflected, dIref, wave, we defined the area-average values as
15 |
where the sum is taken from the start to the end of the wave, and tini and tend are, respectively, the times at the start and end of the wave. These area-average values account for all the wavefronts that make up the incident and reflected waves and not only the wavefronts that define their peaks. Finally, we calculated the transit time (TT) between the incident and reflected waves and the apparent reflection coefficient () based on Equation (14) as
16 |
Thus, we obtained a pair of TT and using peak values and another pair using area-average values. We also calculated TT using two foot values, which we defined as the point when dIinc and dIref are greater than 1% of the peak Iinc and Iref, respectively (they are indicated by empty circles in Figure 5).
2.5 Haemodynamics during diastole
We can describe pressure and flow during diastole using a zero-dimensional Windkessel model. At any point in a distributed 1-D model, pressure becomes increasingly well described with increasing time in diastole by a space-independent Windkessel pressure, pw(t) (Figure 3(a,d,f)), given by 39,54
17 |
18 |
where N is the number of arterial segments, with i = 1 the aortic root and terminal segments coupled to resistance-compliance-resistance (RCR) Windkessel models (Figure 3(b)), M − 1 < N is the number of terminal branches, Qin(t) is the flow waveform at the aortic root, RT is the net peripheral resistance of the arterial network, CT is the total systemic compliance, Cc is the total conduit compliance, is the compliance of segment i (with luminal area , length li and PWV ci), Cp is the total peripheral compliance, pw(T0) is the pressure pw at the reference time t = T0 (T0 = 0 in Figure 3(a,d,f)) and is the outflow in the terminal segment j. The parameters of the RCR Windkessel models are R1 = Z0, C, R2 and Pout (Figure 3(b)). During diastole, it is reasonable to assume Qin = 0 and , , in normal conditions, which reduces pw to
19 |
Equations (17) and (19) neglect nonlinearities, flow inertia and flow viscous dissipation within the 1-D model arterial segments and assume that wall compliance and fluid peripheral resistance are the dominant mechanisms of blood flow. These are reasonable assumptions towards the end of diastole, when Equation (19) provides accurate predictions of blood pressure, as shown by in vivo studies in dogs 55 and numerical solutions of the three-dimensional Navier–Stokes equations in compliant vessels 35.
The flow rate (qw) driven by during diastole is linearly dependent on x in each arterial segment 56,
20 |
where is the flow rate at the inlet of the segment. The wavefronts associated with pw and during diastole at a fixed point x in Segment i, , are
21 |
respectively, with dpw(t) uniform in space. The wave intensity given by dpw and is
22 |
where is the wave intensity at the inlet of Segment i and we have used .
At the ascending aorta, we have , since we assumed zero flow at the root during diastole. Assuming the aorta to be a uniform vessel without branches, we have during diastole. Note that (dpw)2 is exponential with a time constant (see Equation (19)). During the systolic ejection, we can relate the velocity ( dUs) and pressure ( dPs) wavefronts using 16,53, which yield an early-systole wave intensity . Finally, dIw and dIs are related through
23 |
which shows that local (c) and global (RT and CT) properties of the vasculature are responsible for the change of wave intensity from early systole to late diastole.
3 RESULTS AND DISCUSSION
We first study the predominant waves identified by traditional WIA in the aorta of our numerical models and rabbits and discuss their role in shaping the pressure and flow waveforms (Section 3.1 to 3.5). In particular, we show that the wave intensity profile is misleading by indicating a reflection-free period in early systole, which introduces additional error in the estimate of PWV at the aortic root by the traditional PU–loop method (Section 3.2). We then show that WIA does not identify peripheral reflections (Section 3.6) and assess the ability of the Windkessel model to describe haemodynamics during diastole (Section 3.7) and the modified PU–loop method for PWV calculation suggested in 24 (Section 3.8). We describe our new WIA (Section 3.9), which accounts for peripheral reflections using the Windkessel model, and use it to study the effect of vessel compliance and peripheral resistance on blood pressure and flow waveforms (Section 3.10). Lastly, we analyse the sensitivity of our new WIA to sampling frequency and errors in the estimate of PWV (Section 3.11).
3.1 Predominant wave intensity waves in the aorta
At any point in the single-vessel aortic model coupled to a matched three-element Windkessel outlet (Figure 3(b)), we identified four predominant waves in the cardiac cycle from the forward ( dIf) and backward ( dIb) wave intensity profiles as follows (Figure 6(c) shows the profiles at the inlet): (i) an initial forward compression wave (i.e. travelling towards the outlet, because dI > 0, and with dP > 0) produced by the increase in blood flow at the inlet (inflow) in early systole; (ii) a backward compression wave indicating reflection of the initial wave at the outlet; (iii) a forward decompression ( dP < 0) wave (a ‘suction’ wave) caused by the decrease in inflow towards the end of systole; and (iv) a forward compression wave due to the short increase in inflow at the end of systole. The flow is accelerated by the first and fourth waves and decelerated by the second and third. These four waves were also observed in the ascending aorta of the 55-artery model (Figure 3(e), results not shown).
Furthermore, the four waves were observed in the ascending aorta of the rabbit (Figure 6(d)). Their timing and direction of propagation suggest that the first wave is caused by the contraction of the LV, the second is due to reflections of the initial contraction wave at downstream sites of impedance mismatch, the third is produced by the relaxation of the LV and the fourth results from the closure of the aortic valve. The first three waves have been extensively reported for the human aorta 15,16. The fourth wave has been observed in human coronary arteries 12,14; the aorta of the dog 53,57, sheep 58 and swine 59; and large arteries of the fetal lamb 20.
3.2 Left ventricular-contraction wave – an apparent reflection-free period in early systole
Both numerical and in vivo results in Figure 6(c,d) show that there is a period in early systole during which dIb is almost zero. This indicates that changes in pressure and velocity are generated by the contraction of the LV only and not affected by wavefronts reflected in the vasculature. Indeed, the increase in pressure (Figure 6(a,b), orange dashed lines) is made up entirely of forward pressure wavefronts ( dPf) during this apparently reflection-free period (Figure 6(e,f), orange dashed lines). According to Equation (10), if dIb = 0 in early systole, then dP = ρc dU; that is pressure (P) and flow velocity (U) are proportional and make the linear part of the PU–loop (Figure 6(g,h), orange dashed lines).
However, Figure 6(e,f) shows clearly that the backward pressure (Pb) and velocity (Ub) waveforms are not constant during the apparently reflection-free period in early systole: Pb is made up of backward pressure wavefronts ( dPb) that decrease P, and Ub is made up of backward velocity wavefronts ( dUb) that increase U. According to Equation (9), non-zero dPb and dUb indicate that dP = ρc dU is not satisfied. Indeed, from the slope of the PU–loops (Equation (B.1)) in Figure 6(g,h) we have cPU = 4.7 m s − 1 in the aortic model and cPU = 4.0 m s − 1 in the rabbit, which differ respectively from c = 5.2 m s − 1 given by Equation (5) (with A the mean area) and cff = 4.4 m s − 1 calculated using the foot-to-foot method. At the aortic root of the 55-artery model, cPU = 3.4 m s − 1, which differs from c = 4.0 m s − 1 using Equation (5) with mean area. The mean error of cPU in our ten rabbits relative to cff is 9.5% (Table I), which is similar to the error reported in 24 using numerical data only. Therefore, the wave intensity profile is misleading by indicating a reflection-free period in early systole, which leads to an error in the estimate of c by the PU–loop method.
3.3 Reflection of the left ventricular-contraction wave – pressure augmentation
Reflections of the forward wavefronts that make up the LV-contraction (compression) wave yield a reflected wave that is made up of negative velocity wavefronts ( dUref < 0) and positive pressure wavefronts ( dPref > 0), the wave labelled ‘reflection’ in Figure 6(c,d). This wave produces a Ub that decelerates the net forward flow and a Pb that augments pressure (Figure 6(e,f)), thereby generating the ‘shoulder’ or point of inflexion (Figure 6(a,b)) that defines the pressure augmentation index 60.
In the single-vessel aortic model, reflected wavefronts originate from the outlet, which is coupled to a matched three-element Windkessel model (Figure 3(b)). In the more realistic 1-D distributed model, however, the origin of the reflected wave is more complex. Several studies 38–41 have identified multiple reflection sites that reflect wavefronts towards the aorta, where they arrive at different times with a magnitude that decreases exponentially with time 39. The net effect of multiple reflection sites on the TT and given by Equation (16) along the aorta is an apparent reflection site that appears to move away as the measurement location approaches it 40.
3.4 Left ventricular-relaxation wave – flow deceleration
The LV-relaxation (forward) wave drops pressure and flow during the deceleration phase towards the end of systole (Figure 6(a,b)). Flow velocity is reduced by both forward (Uf) and backward (Ub) velocities and becomes negative at the end of systole, whereas pressure (P) is reduced by only the forward pressure (Pf); Pb continues increasing P (Figure 6(e,f)). Thus, during the LV-relaxation phase, wavefronts originating at the aortic root decrease both pressure and velocity, whereas wavefronts reflected in peripheral locations increase pressure and decelerate the flow.
Two mechanisms have been suggested to describe the origin of the LV-relaxation wave, active myocardial relaxation and flow inertia, which are discussed in 61 ,p. 95.
3.5 Valve wave – dicrotic notch
The closure of the aortic valve produces a forward compression wave that augments pressure, creating the dicrotic notch, and accelerates the flow, from negative to zero at the root (Figure 6(a,b,c,d)). The dicrotic notch is shaped by the forward pressure component (Pf) (Figure 6(e,f)), that is by wavefronts travelling from the aortic root, and progressively vanishes in the aorta with increasing distance from the heart (Figure 1(a,b)). In the thoracic aorta of the 55-artery model (Figure 3(e)), the dicrotic notch is absent if wall viscosity is modelled and is present if wall viscosity is neglected (Figure 7(a)). Therefore, wall viscosity dissipates the dicrotic notch. Wall viscosity also dissipates all predominant waves in the wave intensity profile as they travel towards distal locations, almost abolishing the valve wave (Figure 7(b)). In vivo measurements in humans show that the valve wave is small or absent in the brachial, radial and femoral arteries 7,11.
3.6 Why does wave intensity vanish during diastole? The importance of peripheral wave reflections from previous cardiac cycles
Numerical and in vivo wave intensity profiles in the aorta during diastole do not show any reflection of the four predominant waves discussed previously (Figure 6(c,d)). Here, we show that WIA fails to identify the important contribution to the pressure waveform of peripheral reflections, using our single-vessel (Section 3.6.1) and distributed (Section 3.6.2) models.
3.6.1 Single-vessel aortic model
According to this model, wave intensity vanishes during diastole because reflected wavefronts are spread in time by peripheral compliance. To illustrate this, at the inlet of the vessel we prescribed a narrow Gaussian-shaped flow (Figure 8(a)) that approximates a compression wavefront immediately followed by a decompression wavefront. At the outlet, we first coupled a single-resistance with a zero outflow pressure, so that peripheral compliance is absent. This model generates multiple Gaussian-shaped pulse waves (Figure 8(b)). As predicted by Equation (13), both outlet and inlet change the direction of the blood flow; the outlet yields reflected waves travelling towards the inlet with 83% (Rf = (R1 − Z0)/(R1 + Z0) = 0.83) of the pressure and flow amplitudes of the incident wave, and the inlet behaves as a closed end (Rf = 1, because Qin = 0 by the time the first pulse wave is reflected) producing reflected waves with the same amplitude as the incident wave.
The wave intensity profile is able to identify the timing, direction and magnitude of the multiple waves reflected in the model (Figure 8(d)). Equation (16) applied to the wave intensity profile in the midpoint yields errors in the estimates smaller than 2% using peak values and 4% using area-average values (Equation (15)), both relative to the theoretical Rf = 0.83. Furthermore, the errors in the estimates of TT in this model are smaller than 1% using foot, peak and area-average values, which shows the ability of Equation (16) to estimate accurately TT and Rf in the presence of a single peripheral reflection site without peripheral compliance.
Adding peripheral compliance using, for example, a matched RCR Windkessel outflow model (i.e. with R1 = Z0 to minimise wave reflections 42, Figure 3(b)) has the effect of smoothing reflected waves (Figure 8(c)), which is necessary for obtaining physiological-like waveforms (e.g. those in Figure 6(a,c)). Although reflected waves alter the pressure and flow waveforms after t = 0.1 s (Figure 8(c)), the wave intensity profile does not reveal them (Figure 8(e)), because the initial pulse wave has a wave energy several orders of magnitude greater than its reflections.
3.6.2 Distributed 1-D model
We can study the role of peripheral reflections in shaping the pressure waveform using the following methodology. Neglecting nonlinear effects, we can separate the pressure and flow waveforms at an arbitrary point in a given distributed model into a conduit (or arterial) waveform, which is made up of pulse wavefronts propagating from the aortic root and being reflected at the arterial junctions, aortic root and tapered vessels, and a peripheral waveform, which is made up of wavefronts originating from reflections at terminal branches. As detailed in 39, the conduit waveform is obtained by running the simulation with each terminal branch coupled to a single resistance equal to the characteristic impedance of the branch, so that any wavefront leaving the vessel is completely absorbed by the boundary condition; that is so that in Equation (13), . The peripheral waveform is the difference between the total and conduit waveforms.
In the aorta and main branches, most of the pressure waveform at the start of systolic ejection is made up of peripheral reflections, as Figure 9(a) shows for the left common carotid artery. These reflections originated in previous cardiac cycles, because they are present before the start of cardiac ejection in the current cardiac cycle. The maximum contribution of peripheral reflections to the pressure waveform occurs in early diastole, when these reflections produce a concave shape in the proximal arteries of our 55-artery model (Figures 3(d) and 9(a)). This feature of the pressure waveform is sometimes present in vivo in the aorta of the human 3,60,62, rabbit (e.g. Figure 6(b)) and dog 5,55 and in the human carotid, brachial and radial arteries 6–9,11. Furthermore, in the carotid artery, peripheral reflections provide a pressure gradient that drives less than 50% of the mean blood flow in systole, but more than 90% of the mean flow in diastole (Figure 9(b)). All wavefronts that did not originate as peripheral reflections and which make up the conduit waveform are responsible for the features of the pressure and flow waveforms in systole (e.g. pulse pressure and dicrotic notch, Figure 9(a); flow amplitude, Figure 9(b)). In diastole, these wavefronts produce a conduit waveform that decreases exponentially and contributes little to the pressure and flow waveforms in the next cardiac cycle.
Although the diastolic pressure and flow waveforms in our distributed 1-D model are mainly made up of peripheral reflections, WIA fails to identify them. The wave intensity profile only reveals the waves that shape the conduit waveform, as Figure 9(c) shows: the conduit pressure and flow waveforms produce a profile similar in shape and magnitude to the ‘total’ profile, whereas the ‘peripheral’ wave intensity profile has a considerably smaller magnitude than the ‘total’ profile, because changes in pressure and velocity across the wavefronts that make the peripheral waveform are smaller than those that make the conduit waveform. As a result, the apparent reflection coefficient given by Equation (16) accounts little for peripheral reflections. Indeed, calculation of reflection coefficients using forward and backward pressures produces smaller errors than using wave intensity profiles 18.
We can use Equation (23) to quantify the reduction in wave intensity in the aorta from systole ( dIs) to diastole ( dIw) because of net peripheral resistance (RT) and total compliance (CT) acting on the blood flow through the Windkessel effect. Taking x = 20 cm, c = 5 m s − 1 and = 1.8 s as representative values for the thoracic aorta and , then dIw is 2% of dIs. However, because the magnitude of pulse wavefronts decreases as they get reflected and attenuated in the arterial network. Therefore, dIw is several orders of magnitude smaller than dIs (Figure 7(b)).
Next, we discuss the validity of the Windkessel model to study haemodynamics during diastole.
3.7 Using the Windkessel model to study haemodynamics during diastole
Our rabbit in vivo data show that the space-independent pressure pw(t) given by Equation (19) is able to describe pressure towards the end of diastole. Fitting Equation (19) to the decline in pressure during asystole measured along the aorta and iliac artery at 5 cm increments (at least five measurements were taken for each rabbit) produced a small variability of the time constant () and asymptotic pressure (Pout); for example, ms and Pout = 4.5 ± 0.2 kPa in rabbit 8 (Figure 2(c,d) and Table I).
Despite the rabbit having about four times the human heart rate, human and rabbit pressure waveforms are similar in shape when normalised in time (compare Figures 1(a,b)). Therefore, pressure in diastole falls at a faster rate with time in the rabbit, as indicated by the average time constant in the rabbit ( 0.29 ± 0.02 s; Table I) being one order of magnitude smaller than in the human ( 1.79 ± 0.13 s 46). This is due to a smaller total systemic compliance (CT) in the rabbit ( 0.21 ± 0.05 mm 3 Pa − 1) than in the human ( 17.0 ± 1.0 mm 3 Pa − 1 46); vessels are less compliant (i.e. C0D = A0l/(ρc2) is smaller) for the shorter and thinner rabbit vessels, despite having similar PWV to the human vessels.
In agreement with Equation (20), the flow in the last part of diastole increases linearly from the inlet to the outlet of the single-vessel aortic model (Figure 3(c)) and from the aortic root to the thoracic aorta of the 55-artery model (Figure 3(g)). For a given point x in the aorta, the flow towards the end of diastole decays exponentially (Figure 3(c,g,h)). Furthermore, the PU–loop is approximately linear in the last part of diastole (Figure 6(g,h), green dashed lines), which is in agreement with pw and being proportional in the last part of diastole, according to Equation (20).
These results provide further evidence to support the conclusions in 35,55 that the Windkessel model has application for study of haemodynamics in diastole. Application of this model enables us to account for wave reflections originating from previous cardiac cycles and thereby modify the PU–loop method (Section 3.8) and WIA (Section 3.9).
3.8 Modified –loop method
Figure 10(c,d) shows pressure cleared of contributions of waves reflected in previous cardiac cycles versus flow velocity (U) at the root of the single-vessel aortic model and in vivo rabbit aorta. In the numerical model, these contributions were obtained by prolonging the decaying pw given by Equation (19) from the previous cardiac cycle into the current cycle, with the time constant () and asymptotic pressure (Pout) given by the theoretical parameters of the model. In the rabbit, and Pout were calculated by fitting an exponential function of the form given by Equation (19) to the decline in pressure during several asystole generated along the aorta and iliac artery at 5 cm increments (Figure 2(b,e)). The values of and Pout were taken to be the corresponding mean values for each rabbit (Figure 2(c,d)). Figure 10(a,b) shows pw prolonged from the previous cardiac cycle and the pressure waveform , which is cleared of contributions of reflected waves originating from previous cardiac cycles, for the numerical and in vivo data. We did not clear U of contributions of reflected waves originating from previous cycles, because U approximates zero at the aortic root. In more distal locations, these contributions could be obtained by prolonging the decaying U from the previous cardiac cycle into the current cycle. Theoretically, the decaying U should be approximated by given by Equation (20), with Ad the diastolic area.
The backward pressure and velocity wavefronts calculated by applying Equation (9) to and U are zero in early systole (i.e. and are constant, as shown in Figure 10(e,f), orange dashed lines) and, hence, is satisfied in early systole (unlike dP = ρc dU). As a result, the slope of versus U in Figure 10(c,d) (orange dashed lines) yields m s − 1 in the aortic model, in agreement with the theoretical value calculated using Equation (5) with A the mean area, and m s − 1 in the rabbit aorta, in agreement with the value calculated using the foot-to-foot method (Table I). In all ten rabbits, the modified –loop method provided smaller errors than the traditional PU–loop method in the estimates of PWV in the ascending aorta, relative to the foot-to-foot estimates (three last columns of Table I). The mean relative errors for the ten rabbits were 1.3% using the modified –loop method and 9.5% using the traditional PU–loop method, which are comparable to the errors reported in 24 obtained from numerical data.
Our results support the use of the modified –loop method when an exponential fit to the decline in pressure during diastole is possible. An exponential fit may be challenging when using in vivo data without asystole and when the pressure decay does not develop into an exponential (e.g. see Figure 6(b)) 63. Tapering and energy losses are other potential sources of error when using the PU–loop, because they are not accounted for in the derivation of Equation (B.1) (see Appendices A and B).
Lastly, we note that it is not necessary to eliminate the pressure contribution from previous cardiac cycles in order to estimate accurately the local PWV from simultaneous P and luminal area (A) measurements at an arbitrary location in the arterial network. This is because pressure and luminal area wavefronts are reflected with the same reflection coefficient (Rf), whereas velocity wavefronts are reflected with − Rf and, hence, reflections from previous cardiac cycles have a similar effect on P and A, unlike on P and U (see Appendix C). Applying Equation (B.3) from Appendix B to P and A at any point of the visco-elastic 55-artery model (Figure 3(e)) yields errors in the estimates of PWV smaller than 2%, relative to the theoretical PWV obtained using Equation (6) with A the mean area over the cardiac cycle. Such a small error, however, remains to be verified using in vivo P and A data.
3.9 Novel wave intensity analysis
We propose a modified WIA that consists of (i) the new wave intensity calculated using and U to study haemodynamics during systole (Section 3.9.1); (ii) the Windkessel model given by Equation (19) to study haemodynamics during diastole (as we showed in Section 3.7) and quantify the contribution to the pressure waveform of wave reflections originating from previous cardiac cycles (Section 3.9.2); and (iii) the Windkessel model given by Equation (17) (with , , when analysing in vivo data) to analyse the effect on the pressure waveform during both systole and diastole of vessel compliances, peripheral resistances, outflow pressures and the flow at the root (Qin(t)) (Section 3.10).
3.9.1 Haemodynamics during systole
The modified profiles are not influenced by wave reflections originating from previous cardiac cycles and, hence, allow us to do WIA starting from a ‘true’ reflection-free period in early systole. In normal conditions, they are very similar to the traditional dIf,b profiles (e.g. compare Figure 6(c,d) with Figure 10(g,h), and ‘visco-elastic’ in Figure 7(b) with ‘M1’ in Figure 11(d)) and contain the four predominant waves described in Sections 3.1 – 3.5. However, using foot values of and decreases the error of the TT estimate provided by Equation (16) in the single-vessel aortic model coupled to a matched RCR Windkessel model (Figure 3(b)). From the profiles dIf and dIb in the midpoint of the vessel, calculated with c at diastolic pressure, we obtained TT = 67 ms using peak values, TT = 102 ms using area-average values (Equation (15)) and TT = 65 ms using foot values. Their errors are, respectively, 31%, 100% and 27%, relative to the theoretical 51 ms calculated as , according to the characteristics analysis (see Figure 4(b)), where l = 24.1 cm is the vessel length, c = 5.0 m s − 1 is the theoretical PWV at diastolic pressure given by Equation (5), Uinc = 0.04 m s − 1 is the average flow velocity during the propagation towards the outlet of the wavefront that forms the foot of the incident wave, and Uref = 0.5 m s − 1 is the average flow velocity during the propagation towards the inlet of the wavefront that forms the foot of the reflected wave. Similar relative errors were obtained at any point in the vessel, suggesting that foot values should be used for the calculation of TT. The relative error given by foot values of and was reduced to 10%.
Conceptually, the new wave intensity profile should be used to study haemodynamics during systole. In practice, however, the new and traditional wave intensity profiles may be very similar due to measurement errors.
3.9.2 Contributions of wave reflections from previous cardiac cycles
The contribution of wave reflections from several previously occurring cardiac cycles to the pressure waveform (P) can be studied by prolonging the decay of pw given by Equation (19) at the end of each previous cycle. Figure 10(a,b) shows the contribution to P at the aortic root in the numerical model and rabbit in vivo of the second (II), third (III), fourth (IV) and fifth (V) previous cycles, and Table IV quantifies them. In the rabbit, the pressure at which flow to the microcirculation ceases (Pout) contributes 4.5 kPa to P. In the numerical model, Pout was set to zero and does not play a role in shaping P. In both the numerical model and the rabbit, the current cycle contributes to P more than any previous cycle, providing the systolic features of pressure that can be studied using , and, in the rabbit, the diastolic concave shape. Pressure contributions from previous cycles decrease exponentially, with earlier cycles generating a smaller percentage of P than later cycles.
Table IV.
Cardiac cycle | Single-vessel aorta | Rabbit in vivo |
---|---|---|
Current | 41.9 | 27.0 |
I | 25.7 | 17.4 |
II | 14.3 | 7.3 |
III | 8.4 | 3.0 |
IV | 4.0 | 1.3 |
V | 2.5 | 0.5 |
Pout | 0 | 43.0 |
Each contribution is quantified using pw, as described in Section 3.9.2, and is expressed in % relative to the area under P.
Because conduit pressures vanish almost completely by the end of diastole (Figure 9(a)), pressure contributions from previous cardiac cycles are mostly made up of reflected waves originating from peripheral reflection sites. These reflected waves persist for several cardiac cycles, because they become trapped within the arterial network between the aortic valve and peripheral vessels 39. In diastole, the valve is closed and so reflects backward wavefronts with a reflection coefficient close to one 39,56. As a result, forward and backward pressures (Pf,b and ) are equal and forward and backward velocities (Uf,b and ) are opposite, as Figures 6(e,f) and 10(e,f) show.
3.10 Effect of vessel compliance and peripheral resistance
We use our modified WIA to study the effects of changes in vessel compliance and peripheral resistance on the pressure and flow waveforms in the thoracic aorta of the visco-elastic 55-artery model (Figure 3(e)), with the parameter values described in Table III (hereinafter referred to as Model M1, normal young) and the following three variations:
Model M2, normal old: The elastic modulus E is increased three-fold in all the arterial segments (except for the terminal branches) to decrease the total conduit compliance (Cc) and model the increase in the stiffness of the arterial wall with ageing 64.
Model M3, hypertensive young: The total resistance R1 + R2 at the outflow of each 1-D model terminal branch (Figure 3(b)) is increased by 40% to raise blood pressure.
Model M4, hypertensive old: The changes introduced in M2 and M3 are combined.
Using the Windkessel pressure (pw) as a zero-order approximation to the pressure waveform (P) provides relevant information on the effects of vessel compliance and resistance. First, we note that the total compliance (CT) and peripheral resistance (RT) appear in the term of the integrand of Equation (17). This term increases the contribution to P of the decaying flow out of the heart (Qin; Figure 3(g), ‘Root’) in the last part of systole; the greater CT and RT are, the more they spread the shape of Qin (and hence pw) in time. Furthermore, the magnitude of P produced by Qin in systole decreases with increasing CT through the term multiplying the integral. Thus, the pulse pressure decreases with increasing CT. According to Equation (18), CT increases with the increasing compliance of each segment in the arterial network (C0D), which is greater in young models (M1 and M3) than in old models (M2 and M4). As a result, the pulse pressure of both pw and P at the thoracic aorta is less in young than old models (Figure 1(a,c)). In vivo measurements in humans show an increase in pulse pressure with age 62 and disease (such as atherosclerosis and diabetes) 65, which can be explained by a decrease in vessel compliance. Lastly, both P and pw decay in diastole at a greater rate in old than in young models (Figure 1(a,c)), because the time constant () is less in old than young models, in agreement with Equation (19).
Using pw to quantify the portion of P at the thoracic aorta originating from previous cardiac cycles reveals similar contributions to P from the first previous cycle in the four models studied (Table V). However, in old models a smaller portion of P is made up of reflected waves originating from earlier cycles because pw falls at a faster rate with decreasing CT, and in hypertensive models a greater portion of P comes from earlier cycles (compare M1 with M3 and M2 with M4 in Table V) because pw falls at a slower rate with increasing RT.
Table V.
Cardiac cycle | M1 | M2 | M3 | M4 |
---|---|---|---|---|
Current | 38.6 | 58.9 | 30.8 | 48.7 |
I | 24.6 | 21.4 | 23.2 | 24.5 |
II | 12.8 | 6.6 | 14.5 | 10.7 |
III | 6.7 | 2.0 | 9.0 | 4.7 |
IV | 3.5 | 0.6 | 5.6 | 2.1 |
V | 1.8 | 0.2 | 3.5 | 0.9 |
Pout | 10.2 | 10.2 | 7.7 | 7.7 |
We can study how CT and RT affect systolic haemodynamics using the wave intensity profile . The peak values of are greater with decreasing CT and, hence, the energy carried by pulse wavefronts in old models (M2 and M4) is greater than in young models (M1 and M3) (Figure 11(d)). This result suggests that the LV must produce more energy to propel the same amount of blood flow throughout the vasculature of old models. On the other hand, normal and hypertensive models of the same age (i.e. M1 and M3 or M2 and M4) have similar contours of ; that is, changes in RT have little effect on . Therefore, peripheral reflections have a minor effect on the flux of energy in the thoracic aorta in systole. This result provides further evidence to support our result in Section 3.6 on the inability of the wave intensity profile to identify peripheral reflections. Furthermore, this result suggests that changes in RT have little effect on aortic augmentation index, in agreement with 41.
The foot of the incident wave in the profile occurs earlier in old models (Figure 11(d)), which feature greater PWV in proximal vessels, but the TT between the incident and reflected waves calculated using Equation (16) with foot values changes by less than 2 ms in all four models. Furthermore, the apparent reflection coefficient is not affected by changes in RT, but it increases with decreasing CT; is 7% greater in M2 than in M1. Different results are obtained, however, in the single-vessel aortic model; TT decreases with decreasing CT and increasing RT, in agreement with theoretical predictions, and increases with increasing CT and RT (see supplementary material for further details). Therefore, varying CT and RT has a different effect on TT and in the presence of one or multiple peripheral reflection sites.
We can further investigate the effects of CT and RT on the pulse waveform by calculating conduit and peripheral waveforms as described in Section 3.6.2. Decreasing CT increases the amplitude of both the conduit and peripheral pressure waveforms, but increasing RT only increases the magnitude (and not the shape) of the peripheral waveform (Figure 1(e,g)). These results indicate that vessel compliance has a similar effect on reflections originating from internal junctions, aortic root and tapered vessels to those originating from the periphery, whereas peripheral resistances only affect reflections originating from the periphery. Thus, ‘conduit’ and ‘peripheral’ mechanisms underlie changes in pulse pressure, whereas only ‘peripheral’ mechanisms underlie changes in mean pressure. Changes in CT affect both the conduit and peripheral flow waveforms (Figure 1(f,h)). In particular, the amplitude of the conduit waveform increases in old models, which leads to greater peak flows and less flow damping (Figure 1(b)). Changes in RT have little effect on total, conduit and peripheral flow waveforms (Figure 1(b,f,h)).
We cannot measure conduit and peripheral pressures in vivo, but we can approximate them through the reservoir and excess pressures, respectively 56, which can be calculated from a pressure waveform measured at an arbitrary location in arteries 66. This approximation improves with the reflection coefficients for forward-travelling waves at junctions given by Equation (12) approaching zero 56. Thus, reservoir and excess pressures in normal conditions and under the effect of pharmacological drugs affecting vessel compliance and resistance could be used for an in vivo verification of the numerical results presented in this section.
3.11 Sensitivity of the modified wave intensity profile to sampling frequency and errors in pulse wave velocity estimate
We analyse the sensitivity of the modified forward and backward wave intensity profiles (), transit time (TT) and apparent reflection coefficient () in the midpoint of the single-vessel aortic model to decreasing sampling frequency (Section 3.11.1) and increasing errors in the estimate of PWV (Section 3.11.2).
3.11.1 Sampling frequency
The forward and backward wave intensity profiles change with decreasing sampling frequency of pressure and flow velocity (Figure 12(a)). From 1 to 0.2 kHz, the peak values of and change by less than 15%, the errors in the TT estimates given by foot values remain below 10% relative to the theoretical TT = 51 ms (Figure 12(b)), and the calculated using peak and area-average values change by less than 6% relative to the corresponding values at 1 kHz (Figure 12(c)). We observed greater differences in and at 0.1 kHz (Figure 12(a)) and below, which lead to greater relative errors in TT and (Figure 12(b,c)).
Furthermore, we observed a drop in the peak of the valve wave (described in Section 3.5) with decreasing sampling frequency; this wave is small at 0.2 kHz and absent at 0.1 kHz (Figure 2(a)). This result is in agreement with aortic in vivo data showing that the valve wave is small or absent in the traditional wave intensity profile for sampling frequencies below 0.2 kHz 2,4,15 and present for greater frequencies 21,57–59.
Our results suggest a small sensitivity of the wave intensity profile to pressure and velocity data sampled at 0.2 kHz or above, assuming data free of any other error, and support the use of foot values of and to estimate the TT to a dominant reflection site. However, sampling frequencies greater than 0.2 kHz may be necessary at higher heart rates and under strong myocardial contractions.
3.11.2 Pulse wave velocity error
Introducing an error in the PWV given by Equation (5) at diastolic area when calculating results in a backward compression wave developing at the same time as the (incident) LV-contraction wave (Figure 12(d)). The amplitude of this fictitious wave increases with increasing PWV error, which leads to increasingly greater changes in the estimated TT (Figure 12(e)) and (Figure 12(f)) with respect to their values free of PWV error. A PWV error of 20% changes the estimates of TT and by less than 7% using peak values and 16% using area-average values, relative to the corresponding values free of error. Greater PWV errors yield dramatic changes in the calculated TT and using peak values, because the peak value of is attained in the fictitious wave instead of the genuine reflected wave (Figure 12(d)). Lastly, using the local minimum between the fictitious and genuine reflected waves to determine the foot of the reflected wave, the TT estimates given by foot values are more sensitive to PWV errors within ± 20% than those given by peak values (Figure 12(e)).
Our results suggest a small variability of in the aorta using estimates of PWV with errors up to ± 20%, in agreement with results reported for traditional dIf,b using in vivo measurements in human coronary arteries 67. In Section 3.8, we showed that estimates of local PWV at the ascending aorta with errors smaller than 20% are feasible from simultaneous pressure and flow velocity measurements using the modified –loop method.
4 CONCLUSIONS
We have presented a novel analysis of arterial pulse wave propagation that combines traditional WIA with identification of Windkessel pressures to account for the effect on the pressure waveform of peripheral wave reflections. Our modified wave intensity profiles allow us to study the predominant waves that shape pressure and flow waveforms during systole, starting from a ‘true’ reflection-free period in early systole. The Windkessel pressure allows us to (i) quantify the contribution to the pressure waveform over the whole cardiac cycle of reflected waves originating in previous cardiac cycles and (ii) study the buffering effect of the vasculature, which depends on vessel compliances, peripheral resistances, outflow pressures and the flow at the root (Section 3.9).
We have shown that traditional WIA identifies the timing, direction and magnitude of the predominant waves that shape aortic pressure and flow waveforms in systole (Section 3.1 to 3.5) but fails to identify the important contribution to the pressure waveform of peripheral reflections. These reflections persist for several cardiac cycles and make up most of the pressure waveform, especially in diastole and at the start of cardiac contraction (Section 3.6). Ignoring the contribution of peripheral reflections to the pressure waveform leads to an erroneous indication of a reflection-free period in early systole and additional error in the estimates of (i) PWV at the ascending aorta given by the PU–loop method (9.5% error using rabbit in vivo data) and (ii) TT to a dominant reflection site calculated from the wave intensity profile (27% error using numerical data). These errors decreased to 1.3% and 10%, respectively, when peripheral reflections were considered in the calculations using the Windkessel pressure (Section 3.8 and 3.9.1).
We have used our new analysis of wave propagation to study the effects of vessel compliance and peripheral resistance on numerically-generated aortic pressure (P) and flow (Q) waveforms (Section 3.10). With decreasing compliance, the pulse pressure increases, a smaller portion of P is made up of reflected waves originating from earlier cycles, there is less damping of Q, and pulse wave energy increases, suggesting that the LV must produce more energy to propel the same amount of blood flow throughout a stiffer vasculature. With increasing resistance, the mean pressure raises (but not the pulse pressure), a greater portion of P is made up of reflected waves originating from earlier cycles, and there is little change in Q and wave energy. We have also shown that vessel compliance has a similar effect on reflected waves originating from internal junctions, aortic root and tapered vessels to those originating from the periphery, whereas peripheral resistances only affect reflected waves originating from the periphery.
Lastly, our results suggest a small sensitivity of the forward and backward wave intensity profiles to pressure and velocity data sampled at 0.2 kHz or above or to errors in the estimate of PWV within ± 20%, assuming data free of any other error.
It is important to note that our modified WIA differs from the reservoir-wave hypothesis 55,56,66, which has been shown not to be beneficial for WIA 21. In this hypothesis, all the Windkessel pressure is separated from the measured pressure waveform, whereas our new method separates only the Windkessel pressure from previous cardiac cycles. The latter allows us to do WIA on all components of the pressure waveform generated in the current cardiac cycle, starting from a ‘true’ reflection-free period.
Acknowledgments
The authors would like to thank Mr Peter Sowinski, Dr Katherine Fraser and Dr Emma Bailey (Bioengineering, Imperial College) for their assistance with the in vivo experiments, Dr Sukhjinder Nijjer (Hammersmith Hospital, Imperial College Healthcare NHS Trust) for providing the Volcano ComboWires and the reviewers for their constructive comments. The authors gratefully acknowledge the support of a British Heart Foundation Intermediate Basic Science Research Fellowship (FS/09/030/27812) and Programme Grant. Jordi Alastruey was also supported by the Centre of Excellence in Medical Engineering (funded by the Wellcome Trust and EPSRC under grant number WT 088641/Z/09/Z) and the National Institute for Health Research (NIHR) Biomedical Research Centre at Guy's and St Thomas' National Health Service (NHS) Foundation Trust and King's College London. The views expressed are those of the authors and not necessarily those of the NHS, the NIHR or the Department of Health.
APPENDIX A: DERIVATION OF THE WATER HAMMER EQUATIONS
Following 15, we consider the area (A), velocity (U) and pressure (P) waveforms to be made of successive wavefronts, which travel through the arterial network propagating changes in A, U and P. Across an individual wavefront, we define the changes in area, velocity and pressure as dA, dU and dP, respectively. We take the linear assumption that these changes can be separated into changes across the forward-travelling ( dAf, dUf, dPf) and backward-travelling ( dAb, dUb, dPb) wavefronts (Figure A.1), that is
A.1 |
We define the forward direction as the direction of mean blood flow, in which x increases.
Applying conservation of mass in the control volume for the frame moving with the forward-travelling wavefront (Figure 13(c)) yields , where ρ is the density of blood assumed to be constant and c is the local PWV. This expression reduces to
A.2 |
ignoring the term . Applying conservation of momentum to the same moving frame yields , which reduces to
A.3 |
ignoring terms with products of dAf, dUf and dPf. Replacing c dAf in (A.3) with A dUf from (A.2) yields the water hammer equation . Replacing A dUf in (A.3) with c dAf from (A.2) yields
A.4 |
Similarly, conservation of mass and momentum in the control volume for the frame moving with the backward-travelling wavefront (Figure 13(d)) yields the other water hammer equation ,
A.5 |
A.6 |
Combining Equations (A.4) and (A.6), using and , yields
A.7 |
APPENDIX B: CALCULATION OF THE PULSE WAVE VELOCITY
Equations (8), (A.2) and (A.7) allow us to calculate the local PWV (c) at an arbitrary location in the arteries from simultaneous measurements of P and U, A and U, and P and A, respectively, using the methods described in the following two subsections. In all these methods, c is assumed to be constant; that is c = c(x), and dA, dU and dP are calculated as the change in the measured A, U and P, respectively, in a sampling period dt (e.g. dA(t) = A(t + dt) − A(t)) using a Savitzky–Golay filter 16,51,52 ,p. 650.
From simultaneous measurements of P (or A) and U using ‘loop’ methods
The ‘loop’ methods rely on the existence of a period within the cardiac cycle in which the plots of P versus U (PU–loop) and ln A versus U (lnAU–loop) are approximately linear. In normal physiological conditions, this linearity has been observed in early systole in the PU–loop of the human aorta and main pulmonary artery 16,53 and lnAU–loop of the human carotid artery 68, which suggests that dA, dU and dP in early systole are predominantly made of wavefronts propagating towards distal locations. Thus, dA = dAf, dU = dUf and dP = dPf are reasonable assumptions and Equations (8) and (A.2) yield, respectively,
B.1 |
B.2 |
They allow us to calculate c from the slope of the (approximately) linear part in the PU– and lnAU–loops; this slope is ρc and 1/c, respectively. In the case of the PU–loop, we need to know the density of blood (ρ).
In Section 3.8, we showed that elimination of the Windkessel pressure generated in previous cardiac cycles from P improves the accuracy of the value of c given by the PU–loop method in the ascending aorta.
From simultaneous measurements of P and A using a new analytical expression
According to Equation (A.7), (dP)2 = ρ2c4 (d(ln A))2. Adding the terms on the left and right of this equation over a cardiac cycle, assuming ρ and c to be constant, yields
B.3 |
This equation is similar in form to the so-called ‘single-point’ method 16,69 for simultaneous P and U measurements, though Equation (B.3) follows directly from conservation of mass and momentum and does not require minimising wave energy over the cardiac cycle as does the ‘single-point’ method.
APPENDIX C: WAVEFRONT REFLECTIONS AT JUNCTIONS
Consider three arterial segments Ωj, j = p,d1,d2, joining in a splitting or merging flow junction (Figure C.1). Three perturbations of the initial states , j = p,d1,d2, of luminal cross-sectional area, elastic component of pressure and flow rate, respectively, propagating towards the junction (along each corresponding segment Ωj) will produce a new wave in each segment, denoted by , j = p,d1,d2, which will propagate away from the junction. This appendix shows how these three new waves can be calculated using the linearised 1-D equations,
C.1 |
where a, pe and q are the perturbation variables for area, elastic component of pressure and flow rate, respectively, that is , and
C.2 |
are the elastic wall compliance, fluid inertia and viscous fluid resistance, respectively, per unit length of vessel. Equations (C.1) follow from linearising (1) and (2) about the reference state (A,P,Pe,Q) = (A0,0,0,0), with Γ = 0 and β and A0 constant along x 39,54.
Conservation of mass and continuity of the total pressure at the junction yield
C.3 |
C.4 |
In addition, the linear characteristic variables‖, , propagating towards the junction remain unchanged after the reflection. Thus, for a splitting flow junction (Figure 14, left) we have
C.5 |
where , j = p,d1,d2, is the characteristic impedance of segment Ωj and is the pulse wave velocity at zero blood pressure. Moreover,
C.6 |
because the initial characteristic variables moving away from the junction are all zero. Combination of Equations (C.5) and (C.6) yields
C.7 |
Substitution of these expressions into Equation (C.3) and using Equation (C.4) leads to
C.8 |
where , j = p,d1,d2, is the characteristic admittance of segment Ωj.
If we perturb one segment at a time with , j = p,d1,d2, so that for i ≠ j, then from Equation (C.8) we have
C.9 |
Defining the reflection coefficient , j = p,d1,d2, as the ratio of the change of pressure across the reflected wave to the change of pressure in the incident wave; that is , j = p,d1,d2, and using Equation (C.9) to write as a function of yields Equation (12).
For a merging flow junction (Figure 14, right), Equations (C.3) and (C.4) are also valid, whereas Equations (C.5) and (C.6), respectively, become
C.10 |
C.11 |
Combining Equations (C.3), (C.4), (C.10) and (C.11) as described previously for the splitting flow case also yields Equation (12).
Note that from in (C.1), we have and , j = p,d1,d2. Thus, , j = p,d1,d2, is satisfied, which shows that the reflection coefficients also give the ratio of the change of area across the reflected wave to the change of area in the incident wave. However, in terms of changes in the flow rates, we have , j = p,d1,d2, which follows from transforming δqj and Δqj into pressures using Equations (C.6) and (C.7) for the splitting flow case and Equations (C.10) and (C.11) for the merging flow case. Defining the changes in flow velocity as and , we have , j = p,d1,d2.
Footnotes
In this article, the term ‘wave’ refers to a change in blood pressure, flow and luminal area with a finite duration; ‘wavefront’ refers to infinitesimal changes in these properties; and ‘waveform’ refers to the shape of these properties over the cardiac cycle. Wavefronts are the elemental waves in wave intensity analysis (WIA).
This is difficult to observe in Figures 1(a,b) and 3(a,d) because the decrease in mean pressure is small compared to pulse pressure.
The mean flow does not decrease in the single-vessel aortic model because the outlet is the only outflow of this model.
The expressions given for wf,b follow from applying the method of characteristics to the system of equations (C.1).
Supporting Information
Additional supporting information may be found in the online version of this article at the publisher's web site.
REFERENCES
- Alastruey J, Parker KH, Sherwin SJ. Arterial pulse wave haemodynamics. In: Anderson S, editor. 11th International Conference on Pressure Surges, Lisbon, Portugal, 24th – 26th October. Virtual PiE Led t/a BHR Group; 2012. pp. 401–442. (ed.)., ISBN: 9781855981331. [Google Scholar]
- Jones C, Parker K, Hughes R, Sheridan D. Nonlinearity of human arterial pulse wave transmission. Transactions of the ASME Journal of Biomechanical Engineering. 1992;114:10–14. doi: 10.1115/1.2895433. [DOI] [PubMed] [Google Scholar]
- Jones C, Sugawara M. “Wavefronts” in the aorta – implications for the mechanisms of left ventricular ejection and aortic valve closure. Cardiovascular Research. 1993;27:1902–1905. doi: 10.1093/cvr/27.11.1902. [DOI] [PubMed] [Google Scholar]
- Khir A, Henein M, Koh T, Das S, Parker K, Gibson D. Arterial waves in humans during peripheral vascular surgery. Clinical Science. 2001;101:749–757. [PubMed] [Google Scholar]
- Jones C, Sugawara M, Kondoh Y, Uchida K, Parker K. Compression and expansion wavefront travel in canine ascending aortic flow: wave intensity analysis. Heart Vessels. 2002;16:91–98. doi: 10.1007/s003800200002. [DOI] [PubMed] [Google Scholar]
- Ohte N, Narita H, Sugawara M, Niki K, Okada T, Harada A, Hayano J, Kimura G. Clinical usefulness of carotid arterial wave intensity in assessing left ventricular systolic and early diastolic performance. Heart Vessels. 2003;18:107–111. doi: 10.1007/s00380-003-0700-5. [DOI] [PubMed] [Google Scholar]
- Zambanini A, Cunningham S, Parker K, Khir AW, Thom S, Hughes A. Wave-energy patterns in carotid, brachial, and radial arteries: a noninvasive approach using wave-intensity analysis. American Journal of Physiology - Heart and Circulatory Physiology. 2005;289:H270–H276. doi: 10.1152/ajpheart.00636.2003. [DOI] [PubMed] [Google Scholar]
- Avgeropoulou C, Illmann A, Schumm-Draeger PM, Kallikazaros J, von Bibra H. Assessment of arterio-ventricular coupling by tissue Doppler and wave intensity in type 2 diabetes. British Journal of Diabetes & Vascular Disease. 2006;6:271–278. [Google Scholar]
- Sugawara M, Niki K, Ohte N, Okada T, Harada A. Clinical usefulness of wave intensity analysis. Medical and Biological Engineering and Computing. 2009;47:197–206. doi: 10.1007/s11517-008-0388-x. [DOI] [PubMed] [Google Scholar]
- Kolyva C, Pantalos G, Giridharan G, Pepper J, Khir A. Discerning aortic waves during intra-aortic balloon pumping and their relation to benefits of counterpulsation in humans. Journal of Applied Physiology. 2009;107:1497–1503. doi: 10.1152/japplphysiol.00413.2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu J, Yuan LJ, Zhang ZM, Duan YY, Xue JH, Yang YL, Guo Q, Cao TS. Effects of acute cold exposure on carotid and femoral wave intensity indexes: evidence for reflection coefficient as a measure of distal vascular resistance. Journal of Applied Physiology. 2011;110:738–745. doi: 10.1152/japplphysiol.00863.2010. [DOI] [PubMed] [Google Scholar]
- Sun Y, Anderson T, Parker K, Tyberg J. Wave-intensity analysis: a new approach to coronary hemodynamics. Journal of Applied Physiology. 2000;89:1636–1644. doi: 10.1152/jappl.2000.89.4.1636. [DOI] [PubMed] [Google Scholar]
- Sun Y, Anderson T, Parker K, Tyberg J. Effects of left ventricular contractility and coronary vascular resistance on coronary dynamics. American Journal of Physiology - Heart and Circulatory Physiology. 2004;286:H1590–H1595. doi: 10.1152/ajpheart.01100.2001. [DOI] [PubMed] [Google Scholar]
- Davies J, Whinnett Z, Francis D, Manisty C, Aguado-Sierra J, Willson K, Foale R, Malik I, Hughes A, Parker K, et al. Evidence of a dominant backward-propagating “suction” wave responsible for diastolic coronary filling in humans, attenuated in left ventricular hypertrophy. Circulation. 2006;113:1768–1778. doi: 10.1161/CIRCULATIONAHA.105.603050. [DOI] [PubMed] [Google Scholar]
- Parker K, Jones C. Forward and backward running waves in the arteries: analysis using the method of characteristics. Journal of Biomechanical Engineering. 1990;112:322–326. doi: 10.1115/1.2891191. [DOI] [PubMed] [Google Scholar]
- Parker K. An introduction to wave intensity analysis. Medical and Biological Engineering and Computing. 2009;47:175–188. doi: 10.1007/s11517-009-0439-y. [DOI] [PubMed] [Google Scholar]
- Vlachopoulos C, Aznaouridis K, Stefanadis C. Prediction of cardiovascular events and all-cause mortality with arterial stiffness: a systematic review and meta-analysis. Journal of the American College of Cardiology. 2010;55:1318–1327. doi: 10.1016/j.jacc.2009.10.061. [DOI] [PubMed] [Google Scholar]
- Mynard J, Penny D, Smolich J. Wave intensity amplification and attenuation in non-linear flow: implications for the calculation of local reflection coefficients. Journal of Biomechanics. 2008;41:3314–3321. doi: 10.1016/j.jbiomech.2008.10.002. [DOI] [PubMed] [Google Scholar]
- van den Wijngaard J, Siebes M, Westerhof B. Comparison of arterial waves derived by classical wave separation and wave intensity analysis in a model of aortic coarctation. Medical and Biological Engineering and Computing. 2009;47:211–220. doi: 10.1007/s11517-008-0387-y. [DOI] [PubMed] [Google Scholar]
- Smolich J, Mynard J, Penny D. Pulmonary trunk, ductus arteriosus, and pulmonary arterial phasic blood flow interactions during systole and diastole in the fetus. Journal of Applied Physiology. 2011;110:1362–1373. doi: 10.1152/japplphysiol.00038.2011. [DOI] [PubMed] [Google Scholar]
- Mynard J, Penny D, Davidson M, Smolich J. The reservoir-wave paradigm introduces error into arterial wave analysis: a computer modelling and in-vivo study. Journal of Hypertension. 2012;20:734–743. doi: 10.1097/HJH.0b013e32834f9793. [DOI] [PubMed] [Google Scholar]
- Trachet B, Reymond P, Kips J, Swillens A, Buyzere MD, Suys B, Stergiopulos N, Segers P. Numerical validation of a new method to assess aortic pulse wave velocity from a single recording of a brachial artery waveform with an occluding cuff. Annals of Biomedical Engineering. 2010;38:876–888. doi: 10.1007/s10439-010-9945-1. [DOI] [PubMed] [Google Scholar]
- Alastruey J. Numerical assessment of time-domain methods for the estimation of local arterial pulse wave speed. Journal of Biomechanics. 2011;44:885–891. doi: 10.1016/j.jbiomech.2010.12.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mynard J, Davidson M, Penny D, Smolich J. Proceedings of the 33rd Annual International Conference of the IEEE EMBS. 2011. pp. 6446–6449. Robustness of the P-U and lnD-U loop wave speed estimation methods: effects of the diastolic pressure decay and vessel wall non-linearities, Boston, MA, USA, [DOI] [PubMed]
- Gaddum N, Alastruey J, Beerbaum P, Chowienczyk P, Schaeffter T. A technical assessment of pulse wave velocity algorithms applied to non-invasive arterial waveforms. Annals of Biomedical Engineering. 2013:1–13. doi: 10.1007/s10439-013-0854-y. , DOI: 10.1007/s10439-013-0854-y. [DOI] [PubMed] [Google Scholar]
- Stettler J, Niederer P, Anliker M. Theoretical analysis of arterial hemodynamics including the influence of bifurcations. Part II: critical evaluation of theoretical model and comparison with noninvasive measurements of flow patterns in normal and pathological cases. Annals of Biomedical Engineering. 1981;9:165–175. doi: 10.1007/BF02363534. [DOI] [PubMed] [Google Scholar]
- Olufsen M, Peskin C, Kim W, Pedersen E, Nadim A, Larsen J. Numerical simulation and experimental validation of blood flow in arteries with structured-tree outflow conditions. Annals of Biomedical Engineering. 2000;28:1281–1299. doi: 10.1114/1.1326031. [DOI] [PubMed] [Google Scholar]
- Steele B, Wan J, Ku J, Hughes T, Taylor C. In vivo validation of a one-dimensional finite-element method for predicting blood flow in cardiovascular bypass grafts. IEEE Transactions on Biomedical Engineering. 2003;50:649–656. doi: 10.1109/TBME.2003.812201. [DOI] [PubMed] [Google Scholar]
- Reymond P, Merenda F, Perren F, Rüfenacht D, Stergiopulos N. Validation of a one-dimensional model of the systemic arterial tree. American Journal of Physiology - Heart and Circulatory Physiology. 2009;297:H208–H222. doi: 10.1152/ajpheart.00037.2009. [DOI] [PubMed] [Google Scholar]
- Segers P, Dubois F, Wachter DD, Verdonck P. Role and relevancy of a cardiovascular simulator. Cardiovascular Engineering. 1998;3:48–56. [Google Scholar]
- Bessems D, Giannopapa C, Rutten M, van de Vosse F. Experimental validation of a time-domain-based wave propagation model of blood flow in viscoelastic vessels. Journal of Biomechanics. 2008;41:284–291. doi: 10.1016/j.jbiomech.2007.09.014. [DOI] [PubMed] [Google Scholar]
- Alastruey J, Khir A, Matthys K, Segers P, Sherwin S, Verdonck P, Parker K, Peiró J. Pulse wave propagation in a model human arterial network: assessment of 1-D visco-elastic simulations against in vitro measurements. Journal of Biomechanics. 2011;44:2250–2258. doi: 10.1016/j.jbiomech.2011.05.041. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Saito M, Ikenaga Y, Matsukawa M, Watanabe Y, Asada T, Lagrée PY. One-dimensional model for propagation of a pressure wave in a model of the human arterial network: comparison of theoretical and experimental results. Journal of Biomechanical Engineering. 2011;133:121005. doi: 10.1115/1.4005472. [DOI] [PubMed] [Google Scholar]
- Huberts W, Canneyt KV, Segers P, Eloot S, Tordoir J, Verdonck P, van de Vosse F, Bosboom E. Experimental validation of a pulse wave propagation model for predicting hemodynamics after vascular access surgery. Journal of Biomechanics. 2012;45:1684–1691. doi: 10.1016/j.jbiomech.2012.03.028. [DOI] [PubMed] [Google Scholar]
- Xiao N, Alastruey J, Figueroa C. A systematic comparison between 1-D and 3-D hemodynamics in compliant arterial models. International Journal for Numerical Methods in Biomedical Engineering. 2013 doi: 10.1002/cnm.2598. , DOI: 10.1002/cnm.2598. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sherwin S, Franke V, Peiró J, Parker K. One-dimensional modelling of a vascular network in space-time variables. Journal of Engineering Mathematics. 2003;47:217–250. [Google Scholar]
- Peiró J, Veneziani A. Reduced models of the cardiovascular system. In: Formaggia L, Quarteroni A, Veneziani A, editors. Cardiovascular Mathematics. Modeling and Simulation of the Circulatory System. Milano: Springer-Verlag; 2009. pp. 347–394. (eds.). [Google Scholar]
- Segers P, Verdonck P. Role of tapering in aortic wave reflection: hydraulic and mathematical model study. Journal of Biomechanics. 2000;33:299–306. doi: 10.1016/s0021-9290(99)00180-3. [DOI] [PubMed] [Google Scholar]
- Alastruey J, Parker K, Peiró J, Sherwin S. Analysing the pattern of pulse waves in arterial networks: a time-domain study. Journal of Engineering Mathematics. 2009;64:331–351. [Google Scholar]
- Davies J, Alastruey J, Francis D, Hadjiloizou N, Whinnett Z, Manisty C, Aguado-Sierra J, Willson K, Foale R, Malik I, et al. Attenuation of wave reflection by wave entrapment creates a ‘horizon effect’ in the human aorta. Hypertension. 2012;60:778–785. doi: 10.1161/HYPERTENSIONAHA.111.180604. [DOI] [PubMed] [Google Scholar]
- Westerhof B, Westerhof N. Magnitude and return time of the reflected wave: the effects of large artery stiffness and aortic geometry. Journal of Hypertension. 2012;30:932–939. doi: 10.1097/HJH.0b013e3283524932. [DOI] [PubMed] [Google Scholar]
- Alastruey J, Parker K, Peiró J, Sherwin S. Lumped parameter outflow models for 1-D blood flow simulations: effect on pulse waves and parameter estimation. Communications in Computational Physics. 2008;4:317–336. [Google Scholar]
- Valdez-Jasso D, Bia D, Zócalo Y, Armentano R, Haider M, Olufsen M. Linear and nonlinear viscoelastic modeling of aorta and carotid pressure–area dynamics under in vivo and ex vivo conditions. Annals of Biomedical Engineering. 2011;39:1438–1456. doi: 10.1007/s10439-010-0236-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- McDonalds D. Blood Flow in Arteries. London: Edward Arnold; 1974. [Google Scholar]
- Caro C, Pedley T, Schroter R, Seed W. The Mechanics of the Circulation. Cambridge: Cambridge University Press; 2011. , (2nd edn), [Google Scholar]
- Simon A, Safar M, Levenson J, London G, Levy B, Chau N. An evaluation of large arteries compliance in man. American Journal of Physiology. 1979;237:H550–H554. doi: 10.1152/ajpheart.1979.237.5.H550. [DOI] [PubMed] [Google Scholar]
- Cheng C, Herfkens R, Taylor C. Abdominal aortic hemodynamic conditions in healthy subjects aged 50–70 at rest and during lower limb exercise: in vivo quantification using MRI. Atherosclerosis. 2003;168:323–331. doi: 10.1016/s0021-9150(03)00099-6. [DOI] [PubMed] [Google Scholar]
- Yamamoto T, Ogasawara Y, Kimura A, Tanaka H, Hiramatsu O, Tsujioka K, Lever J, Parker K, Jones C, Caro C, et al. Blood velocity profiles in the human renal artery by Doppler ultrasound and their relationship to atherosclerosis. Arteriosclerosis, Thrombosis, and Vascular Biology. 1996;16:172–177. doi: 10.1161/01.atv.16.1.172. [DOI] [PubMed] [Google Scholar]
- Holdsworth D, Norley C, Frayne R, Steinman D, Rutt B. Characterization of common carotid artery blood-flow waveforms in normal human subjects. Physiological Measurement. 1999;20:219–240. doi: 10.1088/0967-3334/20/3/301. [DOI] [PubMed] [Google Scholar]
- Oates C. Cardiovascular Haemodynamics and Doppler Waveforms Explained, Appendix C. Cambridge: Greenwich Medical Media LTD; 2001. [Google Scholar]
- Savitzky A, Golay M. Smoothing and differentiation of data by simplified least squares procedures. Analytical Chemistry. 1964;36:1627–1639. [Google Scholar]
- Press W, Teukolsky S, Vetterling W, Flannery B. Numerical Recipes in C: The Art of Scientific Computing. New York: Cambridge University Press; 1992. , (2nd edn), [Google Scholar]
- Khir A, O'Brien A, Gibbs J, Parker K. Determination of wave speed and wave separation in the arteries. Journal of Biomechanics. 2001;34:1145–1155. doi: 10.1016/s0021-9290(01)00076-8. [DOI] [PubMed] [Google Scholar]
- Alastruey J, Passerini T, Formaggia L, Peiró J. Physical determining factors of the arterial pulse waveform: theoretical analysis and estimation using the 1-D formulation. Journal of Engineering Mathematics. 2012;77:19–37. [Google Scholar]
- Wang JJ, O'Brien A, Shrive N, Parker K, Tyberg J. Time-domain representation of ventricular-arterial coupling as a windkessel and wave system. American Journal of Physiology - Heart and Circulatory Physiology. 2003;284:H1358–H1368. doi: 10.1152/ajpheart.00175.2002. [DOI] [PubMed] [Google Scholar]
- Alastruey J. On the mechanics underlying the reservoir–excess separation in systemic arteries and their implications for pulse wave analysis. Cardiovascular Engineering. 2010;10:176–189. doi: 10.1007/s10558-010-9109-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Khir A, Parker K. Wave intensity in the ascending aorta: effects of arterial occlusion. Journal of Biomechanics. 2004;38:657–655. doi: 10.1016/j.jbiomech.2004.05.039. [DOI] [PubMed] [Google Scholar]
- Penny D, Mynard J, Smolich J. Aortic wave intensity analysis of ventricular-vascular interaction during incremental dobutamine infusion in adult sheep. American Journal of Physiology - Heart and Circulatory Physiology. 2008;294:H481–H489. doi: 10.1152/ajpheart.00962.2006. [DOI] [PubMed] [Google Scholar]
- van Houwelingen M, Merkus D, Hekkert M, van Dijk G, Hoeks A, Duncker D. Initiation of ventricular contraction as reflected in the aortic pressure waveform. Physiological Measurement. 2012;33:557–569. doi: 10.1088/0967-3334/33/4/557. [DOI] [PubMed] [Google Scholar]
- Kelly R, Hayward C, Avolio A, O'Rourke M. Noninvasive determination of age-related changes in the human arterial pulse. Circulation. 1989;80:1652–1659. doi: 10.1161/01.cir.80.6.1652. [DOI] [PubMed] [Google Scholar]
- Mynard J. Ph.D. Thesis. 2011. Computer Modelling and Wave Intensity Analysis of Perinatal Cardiovascular Function and Dysfunction, University of Melbourne, Australia,
- Murgo J, Westerhof N, Giolma J, Altobelli S. Aortic input impedance in normal man: relationship to pressure wave forms. Circulation. 1980;62:105–116. doi: 10.1161/01.cir.62.1.105. [DOI] [PubMed] [Google Scholar]
- Vermeersch S, Rietzschel E, Buyzere MD, Bortel LV, Gillebert T, Verdonck P, Segers P. The reservoir pressure concept: the 3-element windkessel model revisited? Application to the Asklepios population study. Journal of Engineering Mathematics. 2009;64:417–428. [Google Scholar]
- Avolio A, Chen S, Wang R, Zhang C, Li M, O'Rourke M. Effects of aging on changing arterial compliance and left ventricular load in a northern Chinese urban community. Circulation. 1983;68:50–58. doi: 10.1161/01.cir.68.1.50. [DOI] [PubMed] [Google Scholar]
- Safar M, Blacher J, Jankowski P. Arterial stiffness, pulse pressure, and cardiovascular disease – Is it possible to break the vicious circle? Atherosclerosis. 2011;218:263–271. doi: 10.1016/j.atherosclerosis.2011.04.039. [DOI] [PubMed] [Google Scholar]
- Aguado-Sierra J, Alastruey J, Wang JJ, Hadjiloizou N, Davies J, Parker K. Separation of the reservoir and wave pressure and velocity from measurements at an arbitrary location in arteries. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2008;222:403–416. doi: 10.1243/09544119JEIM315. [DOI] [PubMed] [Google Scholar]
- Siebes M, Kolyva C, Verhoeff BJ, Piek JJ, Spaan JA. Potential and limitations of wave intensity analysis in coronary arteries. Medical and Biological Engineering and Computing. 2009;47:233–239. doi: 10.1007/s11517-009-0448-x. [DOI] [PubMed] [Google Scholar]
- Feng J, Khir A. Determination of wave speed and wave separation in the arteries using diameter and velocity. Journal of Biomechanics. 2010;43:455–462. doi: 10.1016/j.jbiomech.2009.09.046. [DOI] [PubMed] [Google Scholar]
- Davies J, Whinnett Z, Francis D, Willson K, Foale R, Malik I, Hughes A, Parker K, Mayet J. Use of simultaneous pressure and velocity measurements to estimate arterial wave speed at a single site in humans. American Journal of Physiology - Heart and Circulatory Physiology. 2006;290:H878–H885. doi: 10.1152/ajpheart.00751.2005. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.