Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2015 Jan 18.
Published in final edited form as: Int J Cardiovasc Imaging. 2014 Sep 13;31(1):95–103. doi: 10.1007/s10554-014-0532-7

Carotid magnetic resonance imaging for monitoring atherosclerotic plaque progression: a multicenter reproducibility study

Jie Sun 1,, Xue-Qiao Zhao 2, Niranjan Balu 3, Daniel S Hippe 4, Thomas S Hatsukami 5, Daniel A Isquith 6, Kiyofumi Yamada 7, Moni B Neradilek 8, Gádor Cantón 9, Yunjing Xue 10, Jerome L Fleg 11, Patrice Desvigne-Nickens 12, Michael T Klimas 13, Robert J Padley 14, Maria T Vassileva 15, Bradley T Wyman 16, Chun Yuan 17,
PMCID: PMC4297722  NIHMSID: NIHMS642597  PMID: 25216871

Abstract

This study sought to determine the multicenter reproducibility of magnetic resonance imaging (MRI) and the compatibility of different scanner platforms in assessing carotid plaque morphology and composition. A standardized multi-contrast MRI protocol was implemented at 16 imaging sites (GE: 8; Philips: 8). Sixty-eight subjects (61 ± 8 years; 52 males) were dispersedly recruited and scanned twice within 2 weeks on the same magnet. Images were reviewed centrally using a streamlined semiautomatic approach. Quantitative volumetric measurements on plaque morphology (lumen, wall, and outer wall) and plaque tissue composition [lipid-rich necrotic core (LRNC), calcification, and fibrous tissue] were obtained. Inter-scan reproducibility was summarized using the within-subject standard deviation, coefficient of variation (CV) and intraclass correlation coefficient (ICC). Good to excellent reproducibility was observed for both morphological (ICC range 0.98–0.99) and compositional (ICC range 0.88–0.96) measurements. Measurement precision was related to the size of structures (CV range 2.5–4.9 % for morphology, 36–44 % for LRNC and calcification). Comparable measurement variability was found between the two platforms on both plaque morphology and tissue composition. In conclusion, good to excellent inter-scan reproducibility of carotid MRI can be achieved in multicenter settings with comparable measurement precision between platforms, which may facilitate future multicenter endeavors that use serial MRI to monitor atherosclerotic plaque progression.

Keywords: Magnetic resonance imaging, Carotid artery, Atherosclerosis, Reproducibility, Multicenter study

Introduction

The fine delineation of vessel wall morphology by magnetic resonance imaging (MR, MRI) has made cardiovascular MR increasingly popular in longitudinal studies for monitoring atherosclerosis progression [110]. Additionally, the combination of contrast weightings enables accurate segmentation of plaques into areas of lipid-rich necrotic core (LRNC), calcification and fibrous tissue [1113]. Understanding morphological and compositional changes of atherosclerotic plaques under various conditions may facilitate the pursuit of optimal management strategies.

For imaging biomarkers to be used in longitudinal studies, scan-rescan reproducibility provides critical information for study planning. Carotid MRI reproducibility has been primarily explored in single-center studies [1416]. However, there is a pressing need for contemporary clinical studies to adopt multicenter designs, either to ensure large-scale recruitment or to study clinical conditions with low prevalence. Unique challenges exist for multicenter MRI studies considering that substantial variations in technique instrumentation are typically present among imaging sites. In brain MRI, pooling data from multiple sites have been shown to introduce variations and biases related to scanner platforms and acquisition protocols [17]. In carotid MRI, even with carefully matched imaging parameters, Saam et al. [18] noted systemic measurement differences between platforms at 1.5 T.

In this prospective study, we sought to determine the multicenter reproducibility of MRI and the compatibility of different scanner platforms in assessing carotid plaque morphology and composition. To provide informative data for future studies, subjects with various extent of carotid atherosclerosis were recruited to encompass the full spectrum of plaque severity; standardized MRI protocol and quality assurance procedures were implemented across sites; and the same scanner was consistently used for individual subjects.

Materials and methods

Study population

A multicenter MRI study (NCT00880178 and NCT01178320; http://clinicaltrials.gov) was carried out in parallel to the Atherothrombosis Intervention in Metabolic syndrome with low HDL/high triglycerides: Impact on Global Health outcomes (AIM-HIGH) trial [19]. AIM-HIGH trial participants, who were men and women aged 45 and older, with established atherosclerosis in coronary, cerebrovascular/carotid or peripheral arteries and dyslipidemia [19], were invited to participate in the MRI sub-study at ten imaging centers if they had no contraindications for MRI examination (e.g. metal implants, claustrophobia) or contrast injection (e.g. renal insufficiency). Similar subjects with identical inclusion criteria were also recruited at six non-AIM-HIGH imaging centers to increase the number of centers. Additionally, each center was restricted to enroll no more than six subjects to ensure an adequate sample distribution across centers.

A total of 68 subjects were recruited at 16 imaging centers in the US (n = 10), Canada (n = 3) and China (n = 3) (Supplemental Table 1). Each subject was scheduled for two carotid MRI scans at the same imaging center within 2 weeks. There was no change in medications or clinical status during the time interval; measurement differences were thus considered to primarily reflect scan-rescan variability of MRI examination and image analysis. Institutional review board approval was obtained at all participating sites. Enrolled subjects provided written informed consent.

Carotid MRI

The 16 imaging sites involved two platforms (Supplemental Table 1). All scans were performed at 3.0 T using commercially available carotid phased-array coils (GE: 6-channel, Neocoil LLC, Pewaukee, WI, USA; Philips: 8-channel, Shanghai Chenguang Medical Technologies, Shanghai, China) [20]. A standardized MRI protocol was implemented on both platforms with carefully matched sequences and parameters, which acquires multi-contrast, cross-sectional images around the carotid bifurcation (Supplemental Table 2). Post-contrast T1-weighted images were acquired about 5 min after contrast injection (Magnevist, Bayer Healthcare, Wayne, NJ, USA). Spatial resolution was 0.625 × 0.625 × 2 mm3 before interpolation. Total acquisition time was approximately 45 min.

Standardized quality assurance procedures were followed in image acquisition (Supplemental Fig. 1). All personnel who performed MRI scans for the study were trained by a core lab (University of Washington, Seattle, WA, USA). Phantom and volunteer scans were performed prior to real subject scans. During the study, images were assessed by the core lab within 48 h of acquisition, and the site was notified if the scan needed to be repeated because of inadequate image quality. In subjects with bilateral carotid plaques, the side with larger lesion (visual inspection) was designated as the index side for follow-up imaging.

Image review

Images were transferred and analyzed centrally using a custom-designed image analysis software package (CASCADE, University of Washington, Seattle, WA, USA) that facilitates multiple-series registration and semiautomatic plaque analysis [21]. Analysis was limited to the index side assigned during image quality check to ensure independent sampling. First and repeat scans were aligned using carotid bifurcation as a landmark, which were then separated, randomized, and analyzed independently using a streamlined multi-step approach.

One reader (J.S.) outlined lumen and outer wall boundaries on cross-sectional slices using a computer-assisted contouring tool. A second reader (K.Y.) performed peer-review which involves contour modification and discussion with the first reader as necessary. Lastly, an integrated plaque segmentation tool automatically classified wall area into LRNC, calcification and fibrous tissue, following a previously described algorithm [22]. After finishing all cases, slice-based area measurements were exported and aggregated for volumetric measurements (morphology: lumen, wall and outer wall volumes; composition: LRNC, calcification and fibrous tissue volumes), which indicate plaque progression patterns and are frequently used in previous studies [110].

Statistical analysis

Data was presented as mean ± standard deviation (SD) or median (inter-quartile range) for continuous variables as appropriate, and count (percentage) for categorical variables. Measurement reproducibility was summarized using the within-subject SD, the coefficient of variation (CV) and the intraclass correlation coefficient (ICC). Linear mixed models were used to decompose the total variance into within-subject, between-subject and between-site variances to compute reproducibility parameters. CV was defined as 100 % × within-subject SD/mean. ICC was defined as (total variance – within-subject variance)/total variance. Presence or absence of LRNC and calcification for each scan was defined as whether or not the component volume was greater than 0 mm3. Inter-scan agreement on component presence/absence was then quantified using Cohen's kappa [23]. Relationships between measurement value (within-subject mean) and precision (within-subject SD and within-subject CV) were tested using the Spearman correlation coefficient. Platform-specific within-subject SDs were computed and the ratio of these SDs were tested against 1 (no difference in measurement precision between platforms). These ratios were also adjusted for the magnitude of measurement, as the precision tended to be related to the magnitude. Specifically, the within-subject SD was modeled as a power function of the within-subject mean, so the adjusted SD ratio can be interpreted as the ratio of platform-specific SDs for subjects with the same mean value [24]. Confidence intervals (CIs) and corresponding p-values were computed using bootstrap technique, where sites were re-sampled as clusters. Data analyses were performed using R 2.14.1 (R Foundation for Statistical Computing, Vienna, Austria). Statistical significance was defined as p < 0.05 (two-tailed).

Results

All 68 subjects completed two scans with a mean time interval between scans of 12.1 ± 6.8 days (Fig. 1). The average age was 61 ± 8 years and 52 (76 %) were men (Table 1). Despite motion artifacts in some cases, all scans were deemed to have diagnostic image quality during image quality check. Therefore, there were no repeated scans or excluded subjects. The mean scan coverage was 28.8 ± 4.5 mm. Based on the first-scan data, 36 (53 %) and 23 (54 %) had detectable LRNC and calcification, respectively.

Fig. 1.

Fig. 1

First and repeat scans on different scanner platforms. a A plaque with calcification (short arrows) and LRNC (long arrows) was scanned twice on GE platform. b A plaque with ulceration (short arrows) and LRNC (long arrows) was scanned twice on Philips platform. Primary weightings are presented to highlight specific features. The last image in each panel illustrates contour-based measurements (yellow contours indicate LRNC). CE-T1W contrast-enhanced T1-weighted, LRNC lipid-rich necrotic core, T1W T1-weighted, T2W T2-weighted, TOF time-of-flight

Table 1. Clinical and imaging characteristics of study population.

Mean ± SD, median (IQR) or n (%) Range
Clinical characteristics
 Age, years 61 ± 8 45–80
 Male sex 52 (76)
 Current smokera 17 (25)
 Hypertensiona 56 (84)
 Diabetes mellitusa 17 (25)
 LDL-C, mg/dla 81 ± 34 29–205
 HDL-C, mg/dla 35 ± 6 20–56
 Triglycerides, mg/dla 169 ± 64 80–388
Imaging characteristicsb
 Lumen volume, mm3 1,239 ± 540 489–3,361
 Wall volume, mm3 849 ± 264 426–1,694
 Total vessel volume, mm3 2,088 ± 746 973–4,707
 LRNC volume, mm3 1.2 (0, 27) 0–358
 Calcification volume, mm3 0 (0, 3.5) 0–84
 Fibrous tissue volume, mm3 751 ± 197 412–1,288

IQR inter-quartile range, LRNC lipid-rich necrotic core, SD standard deviation

a

Excluding 1 subject missing clinical information

b

Based on the first scan

Lumen, wall and outer wall volumes together provide a comprehensive characterization of vessel wall dimensions, all of which demonstrated small measurement errors (CV ranged from 2.5 to 4.9 %) and excellent reproducibility (ICC ranged from 0.98 to 0.99) with repeat scanning (Table 2). Weak but statistically significant correlations (r = 0.24–0.25, p = 0.04–0.05) were noted between within-subject mean and within-subject SD (absolute precision) (Fig. 2). No apparent relationships were noted for within-subject CV (relative precision) with within-subject mean.

Table 2. Reproducibility of MRI measurements on plaque morphology and tissue composition.

N Mean (mm3) Median (mm3) SD (mm3) Within-subject SD (mm3) CV (95 % CI) (%) ICC (95 % CI)
Morphological measurements
 Lumen volume 68 1,221 1,092 545 43 3.5 (2.9, 4.2) 0.99 (0.99, 1.00)
 Wall volume 68 849 805 263 41 4.9 (4.1, 5.6) 0.98 (0.96, 0.98)
 Total vessel volume 68 2,066 1,903 749 52 2.5 (2.1, 2.9) 0.99 (0.99, 1.00)
Compositional measurements
 LRNC volume 46 42 22 74 15 36 (25, 47) 0.96 (0.89, 0.98)
 Calcification volume 24 18 12 23 7.9 44 (28, 61) 0.88 (0.59, 0.96)
 Fibrous tissue volume 68 750 729 195 37 5.0 (4.3, 5.5) 0.96 (0.94, 0.98)

CV coefficient of variation, ICC intra-class correlation coefficient, LRNC lipid-rich necrotic core, SD standard deviation

Fig. 2.

Fig. 2

Bland–Altman plots on plaque morphology. Bland–Altman plots are shown for lumen volume (a), wall volume (b) and total vessel volume (c). A logarithmic scale is used for the x-axis to aid visualization as the measurements spanned a wide range. Note that the actual values shown are on the original scale. Dashed lines indicate the mean difference and dotted lines indicate 95 % limits of agreement

The numbers of subjects where LRNC was detected in both scans, only one scan and neither scan were 32 (47 %), 14 (21 %) and 22 (32 %), respectively (kappa: 0.58; 95 % CI [0.39, 0.78]). For calcification, the numbers were 21 (31 %), 3 (4 %) and 44 (65 %), respectively (kappa: 0.90; 95 % CI [0.79, 1.00]). LRNCs that were inconsistently detected were found to be smaller, with a median volume of 2 (1, 3) mm3, compared to 34 (5, 69) mm3 in those that were consistently detected on both scans. Furthermore, kappa tended to improve when subjects with LRNC volumes <1, 2, or 4 mm3 were excluded (kappa: 0.62, 0.69, and 0.80, respectively).

Variability on quantifying plaque tissue composition was estimated using only subjects that exhibited corresponding components in at least one scan (values were assigned zero if not detected on the other scan). Overall good to excellent reproducibility was observed (ICC ranged from 0.88 to 0.96). Compared to morphological measurements, quantification of compositional structures showed smaller within-subject SD but larger CV (LRNC: 36 %; calcification: 44 %) (Table 2). Notably, within-subject SDs of LRNC and calcification volumes positively correlated with within-subject means (r = 0.77 and 0.74, respectively; Fig. 3), whereas within-subject CVs showed negative correlations (r = −0.76 and −0.65, respectively).

Fig. 3.

Fig. 3

Bland–Altman plots on plaque composition. Bland–Altman plots are shown for LRNC volume (a), CA volume (b) and fibrous tissue volume (c). A logarithmic scale is used for the x-axis to aid visualization as the measurements spanned a wide range. Note that the actual values shown are on the original scale. Dashed lines indicate the mean difference and dotted lines indicate 95 % limits of agreement

Pooled inter-scan variability was calculated for each platform separately. Within-subject SD, indicative of measurement precision, was not significantly different between the two platforms studied (Table 3). Furthermore, the SD ratios (See “Statistical analysis” for details) tended to get closer to 1 after adjustment to control for the positive relationship between within-subject SD and mean.

Table 3. Comparison of measurement errors between scanner platforms.

GE Philips SD ratioa



N Within-subject SD (mm3) N Within-subject SD (mm3) Crude (95 % CI) p value Adjusted (95 % CI) p value
Morphological measurements
 Lumen volume 35 32 33 51 1.6 (0.8, 2.3) 0.158 1.2 (0.8, 2.0) 0.257
 Wall volume 35 36 33 47 1.3 (0.8, 1.9) 0.233 1.1 (0.7, 1.7) 0.560
 Total vessel volume 35 49 33 52 1.1 (0.7, 1.5) 0.856 1.0 (0.7, 1.4) 0.884
Compositional measurements
 LRNC volume 22 13 24 15 1.2 (0.3, 2.9) 0.555 0.9 (0.6, 1.3) 0.500
 Calcification volume 10 9.7 14 5.7 0.6 (0.3, 2.9) 0.396 0.9 (0.6, 1.3) 0.515
 Fibrous tissue volume 35 35 33 41 1.2 (0.9, 1.6) 0.198 1.0 (0.7, 1.5) 0.900

Abbreviations as in Table 2

a

The crude SD ratios were computed as the ratio of platform-specific within-subject SDs. The adjusted ratios involved controlling for the within-subject mean of each individual, which tended to be positively associated with within-subject SD (Figs. 2, 3)

Discussion

With a rigorous multicenter design, we prospectively evaluated the reproducibility of MRI and the compatibility of scanner platforms in assessing carotid plaque morphology and composition. Although previous studies have shown that MRI-derived plaque morphological and compositional measurements are promising biomarkers for expanding our knowledge on atherosclerosis progression, fulfilling their potential lies in large-scale multicenter studies that hitherto have been scarce. This systematic examination of multicenter reproducibility of frequently-used MRI biomarkers therefore may provide helpful information for the planning of future clinical studies, particularly regarding sample size determination, quality control, and image analysis.

Multicenter MRI studies usually entail the involvement of multiple scanner platforms. Quantitative measurements are known to be influenced by platform-specific characteristics including hardware, pulse sequence design and image reconstruction algorithms [17, 18, 25]. Although these factors are typically beyond control of clinical investigators, our data show that it is feasible for multicenter studies to employ different platforms and pool serial data acquired at multiple sites if individuals are followed on the same scanner across the study course. Implementation of a standardized protocol with carefully matched pulse sequences, sequence parameters and flow-suppression techniques could be one key factor in obtaining comparable measurement precision between platforms. Moreover, participating sites may have diverse levels of experience in plaque imaging. Quality assurance procedures are recommended throughout the study. This not only includes initial training and case-specific instructions on subject/coil positioning and identifying motion sources, but to a larger extent refers to timely feedback from experienced readers on image quality and directions for improvement even if completed scans are acceptable.

A number of studies have examined inter-scan reproducibility of plaque morphological measurements by carotid MRI [14, 16, 2629]. The present study, being the largest one and the first multicenter clinical trial, supports the excellent reproducibility of carotid MRI in characterizing plaque morphology. It was demonstrated that measurement precision in multicenter studies can match that reported in single-center studies [14, 16, 2628], when strict quality assurance procedures are implemented and central review is performed.

Few studies have examined inter-scan reproducibility of plaque compositional measurements [2729]. Because of the small size of plaque components and complex features that may co-localize with them, measurement variability of plaque components such as the LRNC has been consistently noted to be larger than morphological measurements. Saam et al. [27] summarized measurement errors from four serial scans of subjects treated with placebo. Multiple scans of the same subject were reviewed simultaneously but mixed with those receiving the active treatment, blinding readers from knowing which cases would be used for summarizing reproducibility. CV of LRNC and calcification volumes were found to be 11.1 % (within-subject SD 16 mm3) and 30.8 % (within-subject SD 9 mm3), respectively. Li et al. [28] used a 3T protocol and showed CV of 31.7 % (within-subject SD 15 mm3) and 22.0 % (within-subject SD 9 mm3) for LRNC and calcification volumes. In another study, Wasserman et al. [29] reported a CV of 86 % (within-subject SD 60 mm3) for LRNC volume and identified reader variability as the main contributing factor. In the latter two studies as well as the present one, all images were acquired for measuring reproducibility and therefore paired scans were analyzed separately in order to blind readers.

While our findings suggest the feasibility of multicenter studies, the larger variability of LRNC compared to morphology has additional implications for future studies. Image acquisition (scanning) and image analysis (reading) both contribute to the variability of contour-based plaque measurements. Reader variability is not trivial for small structures, as suggested by Wasserman et al. [29]. Indeed, much smaller measurement variability has been shown by Saam et al. [27] when serial scans were analyzed side-by-side. A sensitive and unbiased reading approach may be that serial scans are analyzed in the same session while readers are blinded to time sequence. With such a design, reader variability in drawing contours is reduced, and artifacts from unsuppressed flow or partial volume are more consistently interpreted. Consequently, biological differences on plaque composition between serial images can be detected more readily [68]. Additionally, our data indicated that smaller LRNCs were associated with lower kappa and larger CV, which was not shown in previous studies of smaller sample size. It's conceivable that factors limiting area measurement including the spatial resolution will become more influential in the setting of small LRNCs. Therefore, studies with a particular interest in LRNC progression may benefit from examining relatively large ones. Although this requires a screening step, either utilizing the baseline scan for screening purpose or a separate scan with simplified protocol, the sample size needed to test hypotheses may be substantially reduced.

This study used a standard two-dimensional protocol, whose wide availability, especially in university hospitals, forms a solid basis for planning multicenter studies. Yet the small coverage (32 mm) means that technicians need to carefully position scanning slabs to enable following the same segment during the study. Experience from this study suggests that the chance to exclude subjects due to insufficient coverage is low. As carotid atherosclerosis almost unexceptionally affects the carotid bifurcation, the two-dimensional protocol represents a reasonable solution in planning multicenter longitudinal studies on plaque progression, given previous validation studies and demonstrated successes in single-center reports. On the other hand, lesions outside the current coverage are possible, and the slice thickness (2 mm) is a limiting factor in the reproducibility of volumetric measurements. The development of a multi-contrast protocol based on emerging large-coverage, isotropic 3-dimensional (3D) sequences may afford further advantages. As scan-rescan contrast variations and their impact on plaque measurements in multicenter settings were the main focus of this study, the use of a validated automatic plaque classifier has enabled an unbiased comparison between scanner platforms, and the results are expected to hold independent of readers. In contrast to plaque morphology where excellent reader reproducibility is consistently reported [1416], previous literature shows that manual segmentation of plaque components suffers from variable reader reproducibility [13, 2931]. Automatic plaque classifiers, which consistently produce the same results with the same image set, are less subject to reader variability. In terms of accuracy, the automatic plaque classifier has demonstrated good agreement with histology that is non-inferior to expert human reading [21, 22]. Indeed, previous studies have used it to understand the contrast properties of MR sequences as well as changes in plaque composition [5, 32, 33].

This study has several noteworthy limitations. The most important one may be that only two platforms were studied. The feasibility of plaque imaging on a Siemens platform with coils of similar design has been shown [18]. Further efforts are needed to standardize protocols that can be applied across all major MRI manufacturer platforms with comparable performance. Another limitation is that intra-reader and inter-reader reproducibility were not assessed. With a focus on multicenter reproducibility and compatibility of different platforms, we used a standardized workflow including image analysis, which can be readily adopted for future clinical studies but precluded the assessment of reader reproducibility. However, reader reproducibility has been well documented in previous reports [1416]. Finally, as individual subjects were not scanned on different scanners, we could not estimate inter-scanner variation. Platform changes or upgrades during longitudinal studies might undermine study design and likely reduce reproducibility.

Conclusions

This study demonstrates the feasibility of multicenter MRI studies employing different scanner platforms for monitoring carotid atherosclerotic plaque progression, which can maintain good to excellent inter-scan reproducibility with comparable measurement precision between platforms. Our findings may facilitate future multicenter endeavors that use serial MR plaque imaging to seek optimal clinical management as well as improved understanding of the biology of atherosclerosis.

Supplementary Material

Acknowledgements

Supplemental Fig. 1. Image acquisition workflow including quality assurance procedures for (a) setting up imaging sites (b) and monitoring image quality

Supplemental Table 1. Subject distribution across imaging centers

Supplemental Table 2. Carotid MRI protocol

Acknowledgments

This study was supported by a Grant from the Foundation for the National Institutes of Health Biomarkers Consortium made possible by funds from Merck, Pfizer, and Abbott, and by R01HL088214. Carotid coils were provided by GE Healthcare and Philips Healthcare. This manuscript represents the views of the authors and not necessarily those of the National Institutes of Health or the United States government.

Footnotes

Electronic supplementary material: The online version of this article (doi:10.1007/s10554-014-0532-7) contains supplementary material, which is available to authorized users.

Conflict of interest: Xue-Qiao Zhao reported research grants from Abbvie, Kowa, Merck, and Pfizer. Daniel S. Hippe reported grant support for analysis of unrelated data from GE Healthcare, Philips Healthcare, Society of Interventional Radiology, and RSNA Research and Education Foundation. Thomas S. Hatsukami reported research grants from Philips Healthcare. Michael T. Klimas is an employee of Merck. Robert J. Padley is an employee of Abbvie and reported stocks of Abbvie. Bradley T. Wyman was a former employee of Pfizer and reported stocks of Pfizer. Chun Yuan reported research grants from NIH, VP Diagnostics, Philips Healthcare, and consulting fees from Bristol Myers Squibb Medical Imaging and Philips Healthcare. The remaining authors reported no conflicts of interest.

Contributor Information

Jie Sun, Email: sunjie@u.washington.edu, Department of Radiology, University of Washington, 850 Republican St Brotman 127, Seattle, WA 98109, USA.

Xue-Qiao Zhao, Division of Cardiology, Department of Medicine, University of Washington, Seattle, WA, USA.

Niranjan Balu, Department of Radiology, University of Washington, 850 Republican St Brotman 127, Seattle, WA 98109, USA.

Daniel S. Hippe, Department of Radiology, University of Washington, 850 Republican St Brotman 127, Seattle, WA 98109, USA

Thomas S. Hatsukami, Department of Surgery, University of Washington, Seattle, WA, USA

Daniel A. Isquith, Division of Cardiology, Department of Medicine, University of Washington, Seattle, WA, USA

Kiyofumi Yamada, Department of Radiology, University of Washington, 850 Republican St Brotman 127, Seattle, WA 98109, USA.

Moni B. Neradilek, Mountain-Whisper-Light Statistics, Seattle, WA, USA

Gádor Cantón, Department of Mechanical Engineering, University of Washington, Seattle, WA, USA.

Yunjing Xue, Department of Radiology, University of Washington, 850 Republican St Brotman 127, Seattle, WA 98109, USA.

Jerome L. Fleg, National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, MD, USA

Patrice Desvigne-Nickens, National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, MD, USA.

Michael T. Klimas, Merck, Whitehouse Station, NJ, USA

Robert J. Padley, Abbvie, North Chicago, IL, USA

Maria T. Vassileva, Foundation for the National Institutes of Health, Bethesda, MD, USA

Bradley T. Wyman, Pfizer, New London, CT, USA

Chun Yuan, Email: cyuan@u.washington.edu, Department of Radiology, University of Washington, 850 Republican St Brotman 127, Seattle, WA 98109, USA.

References

  • 1.Corti R, Fayad ZA, Fuster V, Worthley SG, Helft G, Chesebro J, Mercuri M, Badimon JJ. Effects of lipid-lowering by simvastatin on human atherosclerotic lesions: a longitudinal study by high-resolution, noninvasive magnetic resonance imaging. Circulation. 2001;104:249–252. doi: 10.1161/01.cir.104.3.249. [DOI] [PubMed] [Google Scholar]
  • 2.Lima JA, Desai MY, Steen H, Warren WP, Gautam S, Lai S. Statin-induced cholesterol lowering and plaque regression after 6 months of magnetic resonance imaging-monitored therapy. Circulation. 2004;110:2336–2341. doi: 10.1161/01.CIR.0000145170.22652.51. [DOI] [PubMed] [Google Scholar]
  • 3.Yonemura A, Momiyama Y, Fayad ZA, Ayaori M, Ohmori R, Higashi K, Kihara T, Sawada S, Iwamoto N, Ogura M, Taniguchi H, Kusuhara M, Nagata M, Nakamura H, Tamai S, Ohsuzu F. Effect of lipid-lowering therapy with atorvastatin on atherosclerotic aortic plaques detected by noninvasive magnetic resonance imaging. J Am Coll Cardiol. 2005;45:733–742. doi: 10.1016/j.jacc.2004.11.039. [DOI] [PubMed] [Google Scholar]
  • 4.Lee JM, Wiesmann F, Shirodaria C, Leeson P, Petersen SE, Francis JM, Jackson CE, Robson MD, Neubauer S, Channon KM, Choudhury RP. Early changes in arterial structure and function following statin initiation: quantification by magnetic resonance imaging. Atherosclerosis. 2008;197:951–958. doi: 10.1016/j.atherosclerosis.2007.09.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Underhill HR, Yuan C, Zhao XQ, Kraiss LW, Parker DL, Saam T, Chu B, Takaya N, Liu F, Polissar NL, Neradilek B, Raichlen JS, Cain VA, Waterton JC, Hamar W, Hatsukami TS. Effect of rosuvastatin therapy on carotid plaque morphology and composition in moderately hypercholesterolemic patients: a high-resolution magnetic resonance imaging trial. Am Heart J. 2008;155:581–584. doi: 10.1016/j.ahj.2007.11.018. [DOI] [PubMed] [Google Scholar]
  • 6.Boussel L, Arora S, Rapp J, Rutt B, Huston J, Parker D, Yuan C, Bassiouny H, Saloner D. Atherosclerotic plaque progression in carotid arteries: monitoring with high-spatial-resolution MR imaging–multicenter trial. Radiology. 2009;252:789–796. doi: 10.1148/radiol.2523081798. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Zhao XQ, Dong L, Hatsukami T, Phan BA, Chu B, Moore A, Lane T, Neradilek MB, Polissar N, Monick D, Lee C, Underhill H, Yuan C. MR imaging of carotid plaque composition during lipid-lowering therapy: a prospective assessment of effect and time course. J Am Coll Cardiovasc Imaging. 2011;4:977–986. doi: 10.1016/j.jcmg.2011.06.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Sun J, Balu N, Hippe DS, Xue Y, Dong L, Zhao X, Li F, Xu D, Hatsukami TS, Yuan C. Subclinical carotid atherosclerosis: short-term natural history of lipid-rich necrotic core–a multicenter study with MR imaging. Radiology. 2013;268:61–68. doi: 10.1148/radiol.13121702. [DOI] [PubMed] [Google Scholar]
  • 9.Fayad ZA, Mani V, Woodward M, Kallend D, Abt M, Burgess T, Fuster V, Ballantyne CM, Stein EA, Tardif JC, Rudd JH, Farkouh ME, Tawakol A. Safety and efficacy of dalcetrapib on atherosclerotic disease using novel non-invasive multimodality imaging (dal-PLAQUE): a randomised clinical trial. Lancet. 2011;378:1547–1559. doi: 10.1016/S0140-6736(11)61383-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Kawahara T, Nishikawa M, Kawahara C, Inazu T, Sakai K, Suzuki G. Atorvastatin, etidronate, or both in patients at high risk for atherosclerotic aortic plaques: a randomized, controlled trial. Circulation. 2013;127:2327–2335. doi: 10.1161/CIRCULATIONAHA.113.001534. [DOI] [PubMed] [Google Scholar]
  • 11.Cai JM, Hatsukami TS, Ferguson MS, Kerwin WS, Saam T, Chu BC, Takaya N, Polissar NL, Yuan C. In vivo quantitative measurement of intact fibrous cap and lipid-rich necrotic core size in atherosclerotic carotid plaque: comparison of high-resolution, contrast-enhanced magnetic resonance imaging and histology. Circulation. 2005;112:3437–3444. doi: 10.1161/CIRCULATIONAHA.104.528174. [DOI] [PubMed] [Google Scholar]
  • 12.Trivedi RA, U-King-Im JM, Graves MJ, Horsley J, Goddard M, Kirkpatrick PJ, Gillard JH. MRI-derived measurements of fibrous-cap and lipid-core thickness: the potential for identifying vulnerable carotid plaques in vivo. Neuroradiology. 2004;46:738–743. doi: 10.1007/s00234-004-1247-6. [DOI] [PubMed] [Google Scholar]
  • 13.Cappendijk VC, Heeneman S, Kessels AG, Cleutjens KB, Schurink GW, Welten RJ, Mess WH, van Suylen RJ, Leiner T, Daemen MJ, van Engelshoven JM, Kooi ME. Comparison of single-sequence T1w TFE MRI with multisequence MRI for the quantification of lipid-rich necrotic core in atherosclerotic plaque. J Magn Reson Imaging. 2008;27:1347–1355. doi: 10.1002/jmri.21360. [DOI] [PubMed] [Google Scholar]
  • 14.Alizadeh DR, Doornbos J, Tamsma JT, Stuber M, Putter H, van der Geest RJ, Lamb HJ, de Roos A. Assessment of the carotid artery by MRI at 3T: a study on reproducibility. J Magn Reson Imaging. 2007;25:1035–1043. doi: 10.1002/jmri.20904. [DOI] [PubMed] [Google Scholar]
  • 15.Syed MA, Oshinski JN, Kitchen C, Ali A, Charnigo RJ, Quyyumi AA. Variability of carotid artery measurements on 3-Tesla MRI and its impact on sample size calculation for clinical research. Int J Cardiovasc Imaging. 2009;25:581–589. doi: 10.1007/s10554-009-9468-8. [DOI] [PubMed] [Google Scholar]
  • 16.Duivenvoorden R, de Groot E, Elsen BM, Lameris JS, van der Geest RJ, Stroes ES, Kastelein JJ, Nederveen AJ. In vivo quantification of carotid artery wall dimensions: 3.0-Tesla MRI versus B-mode ultrasound imaging. Circ Cardiovasc Imaging. 2009;2:235–242. doi: 10.1161/CIRCIMAGING.108.788059. [DOI] [PubMed] [Google Scholar]
  • 17.Reig S, Sánchez-González J, Arango C, Castro J, González-Pinto A, Ortuño F, Crespo-Facorro B, Bargalló N, Desco M. Assessment of the increase in variability when combining volumetric data from different scanners. Hum Brain Mapp. 2009;30:355–368. doi: 10.1002/hbm.20511. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Saam T, Hatsukami TS, Yarnykh VL, Hayes CE, Underhill H, Chu BC, Takaya N, Cai JM, Kerwin WS, Xu DX, Polissar NL, Neradilek B, Hamar WK, Maki J, Shaw DW, Buck RJ, Wyman B, Yuan C. Reader and platform reproducibility for quantitative assessment of carotid atherosclerotic plaque using 1.5T Siemens, Philips, and General Electric scanners. J Magn Reson Imaging. 2007;26:344–352. doi: 10.1002/jmri.21004. [DOI] [PubMed] [Google Scholar]
  • 19.The AIM-HIGH Investigators. The role of niacin in raising high-density lipoprotein cholesterol to reduce cardiovascular events in patients with atherosclerotic cardiovascular disease and optimally treated low-density lipoprotein cholesterol Rationale and study design. The Atherothrombosis Intervention in Metabolic syndrome with low HDL/high triglycerides: impact on Global Health outcomes (AIM-HIGH) Am Heart J. 2011;161:471–477. doi: 10.1016/j.ahj.2010.11.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Balu N, Yarnykh VL, Scholnick J, Chu BC, Yuan C, Hayes C. Improvements in carotid plaque imaging using a new eight-element phased array coil at 3T. J Magn Reson Imaging. 2009;30:1209–1214. doi: 10.1002/jmri.21890. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Kerwin WS, Xu D, Liu F, Saam T, Underhill HR, Takaya N, Chu BC, Hatsukami TS, Yuan C. Magnetic resonance imaging of carotid atherosclerosis: plaque analysis. Top Magn Reson Imaging. 2007;18:371–378. doi: 10.1097/rmr.0b013e3181598d9d. [DOI] [PubMed] [Google Scholar]
  • 22.Liu F, Xu DX, Ferguson MS, Chu BC, Saam T, Takaya N, Hatsukami TS, Yuan C, Kerwin WS. Automated in vivo segmentation of carotid plaque MRI with morphology-enhanced probability maps. Magn Reson Med. 2006;55:659–668. doi: 10.1002/mrm.20814. [DOI] [PubMed] [Google Scholar]
  • 23.Cohen J. A coefficient of agreement for nominal scales. Educ Psychol Meas. 1960;20:37–46. [Google Scholar]
  • 24.Pinheiro JC, Bates DM. Mixed-effects models in S and S-PLUS. Springer; New York, NY: 2000. [Google Scholar]
  • 25.Friedman L, Glover GH. Report on a multicenter fMRI quality assurance protocol. J Magn Reson Imaging. 2006;23:827–839. doi: 10.1002/jmri.20583. [DOI] [PubMed] [Google Scholar]
  • 26.Vidal A, Bureau Y, Wade T, Spence JD, Rutt BK, Fenster A, Parraga G. Scan-rescan and intra-observer variability of magnetic resonance imaging of carotid atherosclerosis at 1.5 T and 3.0 T. Phys Med Biol. 2008;53:6821–6835. doi: 10.1088/0031-9155/53/23/011. [DOI] [PubMed] [Google Scholar]
  • 27.Saam T, Kerwin WS, Chu BC, Cai JM, Kampschulte A, Hatsukami TS, Zhao XQ, Polissar NL, Neradilek B, Yarnykh VL, Flemming K, Huston J, Insull W, Morrisett JD, Rand SD, Demarco KJ, Yuan C. Sample size calculation for clinical trials using magnetic resonance imaging for the quantitative assessment of carotid atherosclerosis. J Cardiovasc Magn Reson. 2005;7:799–808. doi: 10.1080/10976640500287703. [DOI] [PubMed] [Google Scholar]
  • 28.Li F, Yarnykh VL, Hatsukami TS, Chu B, Balu N, Wang J, Underhill HR, Zhao X, Smith R, Yuan C. Scan-rescan reproducibility of carotid atherosclerotic plaque morphology and tissue composition measurements using multicontrast MRI at 3T. J Magn Reson Imaging. 2010;31:168–176. doi: 10.1002/jmri.22014. [DOI] [PubMed] [Google Scholar]
  • 29.Wasserman BA, Astor BC, Sharrett AR, Swingen C, Catellier D. MRI measurements of carotid plaque in the atherosclerosis risk in communities (ARIC) study: methods, reliability and descriptive statistics. J Magn Reson Imaging. 2010;31:406–415. doi: 10.1002/jmri.22043. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Takaya N, Cai JM, Ferguson MS, Yarnykh VL, Chu BC, Saam T, Polissar NL, Sherwood J, Cury RC, Anders RJ, Broschat KO, Hinton D, Furie KL, Hatsukami TS, Yuan C. Intra- and interreader reproducibility of magnetic resonance imaging for quantifying the lipid-rich necrotic core is improved with gadolinium contrast enhancement. J Magn Reson Imaging. 2006;24:203–210. doi: 10.1002/jmri.20599. [DOI] [PubMed] [Google Scholar]
  • 31.Touze E, Toussaint JF, Coste J, Schmitt E, Bonneville F, Vandermarcq P, Gauvrit JY, Douvrin F, Meder JF, Mas JL, Oppenheim C. Reproducibility of high-resolution MRI for the identification and the quantification of carotid atherosclerotic plaque components: consequences for prognosis studies and therapeutic trials. Stroke. 2007;38:1812–1819. doi: 10.1161/STROKEAHA.106.479139. [DOI] [PubMed] [Google Scholar]
  • 32.Kerwin WS, Liu F, Yarnykh V, Underhill H, Oikawa M, Yu W, Hatsukami TS, Yuan C. Signal features of the atherosclerotic plaque at 3.0 Tesla versus 1.5 Tesla: Impact on automatic classification. J Magn Reson Imaging. 2008;28:987–995. doi: 10.1002/jmri.21529. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Liu W, Balu N, Sun J, Zhao X, Chen H, Yuan C, Zhao H, Xu J, Wang G, Kerwin WS. Segmentation of carotid plaque using multicontrast 3D gradient echo MRI. J Magn Reson Imaging. 2012;35:812–819. doi: 10.1002/jmri.22886. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Acknowledgements

Supplemental Fig. 1. Image acquisition workflow including quality assurance procedures for (a) setting up imaging sites (b) and monitoring image quality

Supplemental Table 1. Subject distribution across imaging centers

Supplemental Table 2. Carotid MRI protocol

RESOURCES