Abstract
Sepsis caused by Staphylococcus aureus is a major health problem worldwide. Better outcomes are achieved when rapid diagnosis and determination of methicillin susceptibility enable early optimization of antimicrobial therapy. Eight large clinical laboratories, seven from the United States and one from Scotland, evaluated the combination of the Staphylococcus QuickFISH BC and the new mecA XpressFISH assay (both AdvanDx, Woburn, MA, USA) for the detection of methicillin-resistant S. aureus in positive blood cultures. Blood cultures flagged as positive by automated blood culture instruments and demonstrating only Gram-positive cocci in clusters on Gram stain were tested by QuickFISH, a 20-min assay. If only S. aureus was detected, mecA XpressFISH testing followed. The recovered S. aureus isolates were tested by cefoxitin disk diffusion as the reference method. The QuickFISH assay results were concordant with the routine phenotypic testing methods of the testing laboratories in 1,211/1,221 (99.1%) samples and detected 488/491 S. aureus organisms (sensitivity, 99.4%; specificity, 99.6%). Approximately 60% of the samples (730) contained coagulase-negative staphylococci or nonstaphylococci as assessed by the QuickFISH assay and were not tested further. The 458 compliant samples positive exclusively for S. aureus by the QuickFISH assay were tested by the mecA XpressFISH assay, which detected 209 of 211 methicillin-resistant S. aureus organisms (sensitivity, 99.1%; specificity, 99.6%). The mecA XpressFISH assay also showed high reproducibility, with 534/540 tests performed by 6 operators over 5 days achieving reproducible results (98.9% agreement). The combination of the Staphylococcus QuickFISH BC and mecA XpressFISH assays is sensitive, specific, and reproducible for the detection of methicillin-resistant S. aureus and yields complete results in 2 h after the blood culture turns positive.
INTRODUCTION
Members of the genus Staphylococcus are the most common organisms isolated from blood cultures, with Staphylococcus aureus often being the most common cause of bloodstream infections (BSI) (1). The annual incidence of infection in the United States is approximately 50 per 100,000 population. Methicillin-resistant S. aureus (MRSA) leads to significant increases in morbidity and mortality compared to methicillin-susceptible S. aureus (MSSA) (2). MRSA infections may be health care associated or community acquired (3). The treatment of MSSA bacteremia with vancomycin may lead to delays in the clearance of the organism (4). Since MRSA represents an epidemiologic concern, patients infected with S. aureus may be placed in isolation until the susceptibility to methicillin is ascertained (http://www.cdc.gov/mrsa/healthcare/clinicians/precautions.html). Isolation represents a significant medical expense and may adversely affect patient care (5). The rapid identification of the etiologic agent of bloodstream infection leads to early optimization of antibiotic therapy and improved patient outcomes, including decreased mortality, shorter stays in the intensive care unit, shorter hospital stays, and significant reductions in hospital costs (4, 6–8). The FDA has approved/cleared several nucleic acid-based techniques that expedite the identification of S. aureus directly from positive blood culture bottles (http://www.fda.gov/MedicalDevices/ProductsandMedicalProcedures/InVitroDiagnostics/ucm330711.htm#microbial). In addition to AdvanDx, the manufacturers of these assays include BioFire (Salt Lake City, UT), BD GeneOhm (Franklin Lakes, NJ), Nanosphere (Northbrook, IL), and Cepheid (Sunnyvale, CA).
A peptide nucleic acid fluorescence in situ hybridization (PNA FISH) kit for the rapid detection of S. aureus and coagulase-negative staphylococci (CoNS) was launched in 2003 by AdvanDx. Since then, S. aureus/CoNS PNA FISH tests have been used in many clinical microbiology laboratories for the rapid identification of Staphylococcus organisms and to distinguish S. aureus from CoNS in positive blood cultures. PNA FISH probes consist of peptide backbones to which the nucleic acid bases are attached. Such probes lack the net negative charge common to nucleic acid probes, which facilitates both entry into bacterial cells and tighter binding to the RNA sequences that serve as targets for the assays (8–10). Recently, the FDA approved the S. aureus QuickFISH BC (QFISH) (AdvanDx) assay, which is similar to its predecessor but takes about 20 min (10). Both the original and QFISH assays accurately distinguish S. aureus from CoNS (8, 9). However, neither distinguishes MSSA, which can often be treated with oxacillin and some cephalosporins, from MRSA, which responds to a more limited spectrum of antibiotics, such as vancomycin, daptomycin, ceftaroline, and linezolid, and which may be toxic, expensive, or both. Thus, patients whose cultures exhibit Gram-positive cocci in clusters, suggestive of S. aureus, previously had to be treated with an antimicrobial suitable for MRSA until phenotypic susceptibility results became available after about 2 days.
The MRSA phenotype is due to the mecA gene, which encodes the altered penicillin-binding protein 2a (PBP-2a) (11). The proportion of MRSA varies with the patient setting but is currently around 50% with both health care-associated and community-acquired disease (3) (http://www.cdc.gov/mrsa/healthcare/clinicians/precautions.html). Phenotypic identification and susceptibility testing require a minimum of 2 days after the culture bottle is flagged as positive. AdvanDx recently designed a PNA-FISH assay for the detection of mRNA encoded by the S. aureus mecA gene called mecA XpressFISH (mecAX). Part of the data presented here was submitted to the FDA. Eight sites participated in the original study. However, the data from the site that used the Trek system were not submitted to the FDA due to the limited number of samples tested at that site.
(This study was presented as a poster at the 114th General Meeting of the American Society for Microbiology, 17 to 20 May 2014, Boston, MA.)
MATERIALS AND METHODS
Laboratories and study design.
Seven geographically diverse U.S. microbiology laboratories and one in the United Kingdom participated in this study. Blood cultures were collected and incubated according to each laboratory's standard operating procedure. When a bottle was flagged as positive by the automated blood culture instrument, a Gram stain was performed on the fluid, and any organisms present were identified by the standard automated phenotypic procedures of the laboratory. If the Gram stain showed only Gram-positive cocci in clusters (GPCCL) and the specimen was eligible for enrollment (see below), the residual specimen was deidentified, given a unique study number, and tested by the QFISH assay. Those specimens containing only S. aureus by the QFISH assay were next tested by mecAX. The human subjects research committee of each institution waived the requirement for informed consent. Prior to initiating the study, the coordinators and technical personnel were trained at each site. Training samples were tested at each site, and a reproducibility study was performed at three sites.
Enrollment criteria and routine testing procedures.
A single set of blood cultures was enrolled per patient. Both bottles of that individual set were eligible for enrollment, provided the organisms in each bottle were worked up separately for organism identification. One of three continuously monitored blood culture systems (BacT/Alert, bioMérieux, Durham, NC; Bactec, Becton Dickinson, Franklin Lakes, NJ; or VersaTREK, Thermo Fisher Scientific, Oakwood Village, OH) was used at each of the eight sites that participated in this study. Of the eight sites, five used Bactec, two used BacT/Alert, and one site used VersaTREK automated blood culture systems. Bottles containing charcoal to adsorb antibiotics and the VersaTREK REDOX2 blood culture bottles were not used, as they are not compatible with the QFISH assay, according to the manufacturer's instructions (10).
The sites performed their standard bacterial identification and susceptibility testing according to the manufacturer's instructions for the instrumentation and in accordance with CLSI guidelines for any other testing (12). The automated instrumentation included MicroScan (Siemens Medical Solutions, Deerfield, IL) (2 sites), Phoenix (BD Diagnostics, Sparks, MD) (2 sites), and Vitek (bioMérieux, Durham, NC) (4 sites). Blood culture bottles that were flagged as positive on an automated blood culture device and exhibited only GPCCL by Gram stain were enrolled in the first part of the study. The testing personnel were given a deidentified bottle and the Gram stain results. In addition, they were blinded to any organism identification or antibiotic susceptibility test results that might have already been available. The residual fluid from the bottle was first subjected to testing with the QFISH assay for the rapid identification of S. aureus and CoNS, according to the manufacturer's instructions. Briefly, fluorescein-labeled S. aureus and Texas Red-labeled CoNS probes are added to a fixed smear prepared from liquid from a positive blood culture bottle. After hybridization and a stringent wash, the slide is coverslipped and observed under a fluorescence microscope (10).
The samples positive for S. aureus only by the QFISH assay were then tested with the mecAX assay. Polymicrobial specimens were excluded. The samples were stored at room temperature (20 to 23°C) between bottle positivity and testing. Testing was required to be completed within 60 h. Most were completed within 24 h, and only 13 (3.8%) exceeded 48 h. Repeat mecAX testing of specimens with discrepant mecAX assay results took place within 1 week of bottle positivity.
mecA XpressFISH BC assay.
Positive blood culture bottles that contained only S. aureus as assessed by the QFISH assay were tested for the presence of mecA by the mecAX assay, according to the manufacturer's instructions, which are summarized briefly below. The fluids from the positive bottles were filtered if they contained resin beads (anaerobic bottles). Approximately 10 drops of sample were transferred to a proprietary filter vial, and the plunger was depressed to push the fluid into the vial while retaining the beads. The filtrates and fluids obtained directly from the culture bottles were subsequently treated identically. Because the mecAX assay detects mecA mRNA, transcription of the gene is first induced (13) by incubating the organism in the presence of cefoxitin (FOX). The mecAX kit includes induction tubes containing 0.5 ml of Trypticase soy broth and swabs containing FOX. Prior to induction, a FOX swab is inserted into the broth in the induction tube, swirled to release the antibiotic, and discarded. The induction tubes are inoculated with 250 μl of filtrate or fluid from blood culture bottles positive for S. aureus (FOX final concentration, 6.3 μg/ml), mixed, and incubated at 33 to 35°C for 40 to 50 min. mecAX testing is carried out immediately.
Proprietary slides containing a sample well and positive- and negative-control wells are heated to 55 ± 1°C. Fluid (10 μl) from the induction tube is added to the center of the sample well and spread. One drop of fixing solution is added and spread evenly throughout the sample well. After the smear becomes visibly dry, 1 drop of mecAX probe solution is added to the sample and negative-control wells. The sample well is covered with one coverslip, and the negative- and positive-control wells are covered with another. The slides, still on the heating block, are incubated for 10 to 20 min. The slides are then transferred to a stringent wash solution preheated to and maintained at 57 ± 1°C. The coverslips fall off spontaneously or are removed with forceps. After incubation for 10 to 20 min, the slides are air-dried. One drop of mounting fluid is added to the sample well and another to the negative-control well. Separate coverslips are applied to the sample well and the control wells. The slides are examined with a fluorescence microscope using a filter provided by the manufacturer. S. aureus carrying mecA fluoresce bright green against a yellowish-brown background, while negative organisms should not stain.
Reference antimicrobial susceptibility testing method.
The comparator assay for the mecAX assay was FOX disc susceptibility. This was performed at all sites using 30-μg discs, according to Clinical and Laboratory Standards Institute M100-S23 guidelines (12), regardless of the standard procedure generally employed in the laboratory. A zone size of ≥22 mm was interpreted as methicillin susceptible, while ≤21 mm was defined as resistant to methicillin. Discordant FOX and mecAX test results were resolved by repeating the mecAX testing. The unresolved discrepant sample was subjected to analysis by the FilmArray blood culture identification system (BioFire, Salt Lake City, UT), according to the manufacturer's instructions.
mecAX reproducibility analysis.
A mecAX assay reproducibility study was performed by two operators at each of three sites on 5 separate testing days. Three organisms were included (S. aureus strains CT-178 [formerly NRS674], [BEI Resources, National Institute of Allergy and Infectious Diseases {NIAID}, NIH], a homogeneous MRSA; ATCC 4330 [ATCC, Gaithersburg, MD], a heterogeneous MRSA; and ATCC 29213, an MSSA). The specimens were prepared at AdvanDx, encoded, and shipped on ice to the testing sites. All contained approximately 1 × 107 to 6 × 107 organisms/ml. Each sample was tested in triplicate by each operator on each testing day.
Data analysis.
Sensitivity and specificity were calculated from routine 2 by 2 result tables. The 95% confidence intervals were calculated by the method of Clopper and Pearson (14) using the online calculator at http://statpages.org/confint.html.
RESULTS
S. aureus QuickFISH BC assay.
After a Gram stain that revealed only GPCCL, 1,221 residual clinical specimens from 1,082 patients were enrolled in the QFISH assay study. The 1,221 specimens were from Bactec (918 blood culture bottles), BacT/Alert (266 bottles), and VersaTREK (37 bottles). There were 139 paired aerobic and anaerobic blood culture bottles from the same patient included in the final analysis. All the others were single bottles from unique patients. Six cultures were positive for both CoNS and S. aureus by the QFISH assay and were included in the analysis of the sensitivity and specificity of the QFISH assay. Of the QFISH results, 1,211 of the 1,221 (99.2%) agreed with the results of standard phenotypic testing (Table 1), with 488/491 S. aureus (green fluorescence), 682/689 CoNS (red fluorescence), and 41/41 other GPCCL (no fluorescence). Three false-negative results were obtained for S. aureus and 7 for CoNS. Four of the seven CoNS results remained negative on retesting, including one Staphylococcus simulans. which is a known limitation of the QFISH assay (data not shown). The remaining three were not retested. The CoNS and “other” categories were combined to calculate a specificity for S. aureus, while the S. aureus and “other” categories were combined to calculate a specificity for CoNS (Table 1).
TABLE 1.
Routine identification | No. identified by Staphylococcus
QuickFISH BC assay |
Performance (% [no. detected/total no.] [95% CI]) of Staphylococcus
QuickFISH BC assay |
||||
---|---|---|---|---|---|---|
S. aureus | CoNS | Other GPCCL | Total | Sensitivity | Specificitya | |
S. aureus | 488 | 3 | 0 | 491 | 99.4 (488/491) (98.2–99.4) | 99.6 (730/733) (98.8–99.9) |
CoNS | 0 | 682 | 7 | 689 | 99.0 (682/689) (97.9–99.5) | 98.1 (529/539) (96.6–99.1) |
Other GPCCL | 0 | 0 | 41 | 41 | 100 (41/41) (91.4–100) |
For calculation of the specificity and 95% confidence intervals, see Materials and Methods.
mecAX testing.
Of the 488 specimens identified by QFISH as containing S. aureus, 458 provided evaluable results for the mecAX assay. Six specimens fluoresced both red and green, indicating the presence of both S. aureus and CoNS. According to protocol, these were not subjected to the mecAX evaluation. In addition, 24 specimens were not included in the mecAX evaluation due to procedural errors or protocol violations: 2 samples were accidentally skipped for mecAX testing, 6 were eliminated due to the mecA mRNA induction time being too short or too long, and the first 16 samples from one site were excluded due to an incorrect stringent wash temperature (the same as the hybridization temperature).
The results of the 458 evaluable S. aureus specimens that were subjected to both FOX disk diffusion testing and mecAX testing are shown in Table 2. FOX disk diffusion revealed 211/458 or 46.1% of the S. aureus specimens to be MRSA; of these, 209/211 were mecA XpressFISH positive. Of the 247 MSSA by FOX disk diffusion, 246 were mecAX negative. The overall sensitivity and specificity of mecAX were 99.1% and 99.6%, respectively.
TABLE 2.
mecA XpressFISH result | No. in cefoxitin disk diffusion test found to beb: |
|
---|---|---|
Methicillin resistant (≤21 mm) | Methicillin susceptible (≥22 mm) | |
Positive | 209 | 1c |
Negative | 2d | 246 |
The sensitivity of the assay was 99.1% (209/211) (95% CI, 96.6 to 99.7%), and the specificity was 99.6% (246/247) (95% CI, 97.8 to 99.99).
n = 458.
The single false positive (FOX, 28 mm) originally exhibited atypical weakly positive green fluorescence and was negative on repeat testing.
One of the two false negatives (FOX, 19 mm) was positive on repeat mecA XpressFISH testing. The other (FOX, 11 mm) remained negative on repeat mecA XpressFISH testing. That specimen was subjected to PCR for mecA, which was positive.
To resolve the three discrepant results, the mecAX testing was repeated. The single false-positive specimen originally exhibited an atypical weak green mecAX signal and was negative on repeat testing. One false-negative mecAX result (FOX disc, 19 mm) was positive on repeat testing, while the other remained mecAX negative (FOX disc, 11 mm). The persistently false-negative specimen was subjected to mecA PCR using an FDA-cleared mecA PCR assay from a different manufacturer and was found to be mecA positive.
mecAX reproducibility study.
The tests were performed in triplicate by two operators per site at three sites over 5 days and showed agreement for 534/540 (360/360 mecAX-positive and 174/180 mecAX-negative) samples (98.9%). The six nonreproducible false-positive results (triplicate determinations of 2 negative samples) were obtained at one site on the same day by the same operator. This is presumed to represent a systematic error, but its nature was not determined.
DISCUSSION
This multicenter study is the first report of the clinical performance of the AdvanDx mecAX test. The combination of the S. aureus QFISH and mecAX tests for the identification of MRSA in blood cultures is rapid, robust, reliable, sensitive, and specific. Our data also revealed high intra- and interlaboratory reproducibility. Our results with the QFISH assay parallel those of the multicenter study by Deck et al. (9). In that study, the authors reported sensitivities for detecting S. aureus and CoNS of 99.5% and 98.8%, respectively. They also reported that the Staphylococcus QFISH assay turnaround time was <30 min and the hands-on time was <5 min. Likewise, Carretto et al. (15) reported 100% sensitivity and specificity for S. aureus detection in their single-center evaluation, with minimal impact with respect to modifying laboratory workflow to accommodate testing.
Compared to FOX testing, the mecAX test had two false-negative and one atypical, weak, green, and false-positive fluorescent result (sensitivity and specificity, 99.1 and 99.6%, respectively; Table 2). One false-negative and one false-positive result resolved on retesting. The third repeatedly mecAX false-negative result was positive for mecA by the FilmArray test. We did not investigate possible sequence differences that might have accounted for the mecAX assay result that remained falsely negative on repeat testing.
The combination of the Staphylococcus QFISH and mecAX assays might significantly speed up the diagnosis of MRSA BSI. After the 20-min QFISH assay indicates the presence of S. aureus, labs using the mecAX assay can determine its susceptibility to methicillin and notify the physician in <2 h. Compared with other rapid tests, the combination of the QFISH and mecAX assays has low barriers to implementation. The training is straightforward, and test performance does not require any molecular skills beyond those expected of a competent clinical laboratory technologist. In total, the hands-on time is approximately 15 min. As Carretto et al. (15) point out, the cost of PNA-FISH is two-thirds the cost of some reverse transcription-PCR (RT-PCR) assays. To perform the QFISH and mecAX assays, only a slide warmer, water bath, and fluorescence microscope equipped with a dual-band filter are needed. In addition, variable costs for the mecAX assay, including fixative, probes, wash solutions, slides, and coverslips, are minimal. This compares very favorably to matrix-assisted laser desorption ionization–time of flight mass spectrometry and nucleic acid amplification, both of which require expensive instruments for test performance. Nucleic acid amplification may also require additional equipment for nucleic acid extraction and perhaps a laboratory designed specifically for DNA amplification with unidirectional workflow and clean rooms.
Many of our laboratories have been successfully using the various generations of S. aureus PNA FISH testing for years, and the addition of the mecAX assay to the test menu would require little additional effort.
It has already been shown that the implementation of PNA FISH and the more rapid QFISH technologies lead to improved patient outcomes and antibiotic utilization (8, 9). After utilizing QFISH technology to identify S. aureus, the detection of methicillin resistance by phenotypic techniques still requires 2 days. Thus, performing the mecAX assay immediately after the QFISH test might enhance patient outcomes significantly, providing there is a rapid clinical response to the results. The implementation of an additional test, such as the mecAX, into the laboratory workflow depends on clinical considerations, convenience, and cost-effectiveness (including laboratory staffing levels, clinician availability, and communication support). This may include the availability of pharmacy representatives to reinforce rapid and appropriate antibiotic changes. As shown by Holtzman et al. (16), without active antimicrobial stewardship intervention, the implementation of a dual-probe staphylococcal PNA FISH test did not lead to shorter hospital stays or decreased vancomycin use after the identification of CoNS. Whether the implementation of the QFISH and mecAX assays will be cost-effective in individual hospitals, especially those that batch their PNA-FISH tests and perform them once per shift (for example), will require cost-benefit analyses and outcome studies.
ACKNOWLEDGMENTS
This study was supported by research grants from AdvanDx, Inc., Woburn, MA.
Aside from research support, we have no financial interests in AdvanDx.
Footnotes
Published ahead of print 27 August 2014
REFERENCES
- 1.Styers D, Sheehan DJ, Hogan P, Sahm DF. 2006. Laboratory-based surveillance of current antimicrobial resistance patterns and trends among Staphylococcus aureus: 2005 status in the United States. Ann. Clin. Microbiol. Antimicrob. 5:2. 10.1186/1476-0711-5-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.van Hal SJ, Jensen SO, Vaska VL, Espedido BA, Paterson DL, Gosbell IB. 2012. Predictors of mortality in Staphylococcus aureus bacteremia. Clin. Microbiol. Rev. 25:362–386. 10.1128/CMR.05022-11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Hadler JL, Petit S, Mandour M, Cartter ML. 2012. Trends in invasive infection with methicillin-resistant Staphylococcus aureus, Connecticut, USA, 2001–2011. Emerg. Infect. Dis. 18:917–924. 10.3201/eid1806.120182. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Khatib R, Saeed S, Sharma M, Riederer K, Fakih MG, Johnson LB. 2006. Impact of initial antibiotic choice and delayed appropriate treatment on the outcome of Staphylococcus aureus bacteremia. Eur. J. Clin. Microbiol. Infect. Dis. 25:181–185. 10.1007/s10096-006-0096-0. [DOI] [PubMed] [Google Scholar]
- 5.Stelfox HT, Bates DW, Redelmeier DA. 2003. Safety of patients isolated for infection control. JAMA 290:1899–1905. 10.1001/jama.290.14.1899. [DOI] [PubMed] [Google Scholar]
- 6.Kumar A, Ellis P, Arabi Y, Roberts D, Light B, Parrillo JE, Dodek P, Wood G, Kumar A, Simon D, Peters C, Ahsan M, Chateau D, Cooperative Antimicrobial Therapy of Septic Shock Database Research Group 2009. Initiation of inappropriate antimicrobial therapy results in a fivefold reduction of survival in human septic shock. Chest 135:1237–1248. 10.1378/chest.09-0087. [DOI] [PubMed] [Google Scholar]
- 7.Paul M, Kariv G, Goldberg E, Raskin M, Shaked H, Hazzan R, Samra Z, Paghis D, Bishara J, Liebovici L. 2012. Importance of appropriate empirical antibiotic therapy for methicillin-resistant Staphylococcus aureus bacteraemia. J. Antimicrob. Chemother. 65:2658–2665. 10.1093/jac/dkq373. [DOI] [PubMed] [Google Scholar]
- 8.Forrest GN. 2007. PNA FISH: present and future impact on patient management. Expert Rev. Mol. Diagn. 7:231–236. 10.1586/14737159.7.3.231. [DOI] [PubMed] [Google Scholar]
- 9.Deck MK, Anderson ES, Buckner RJ, Cofasaante G, Couli JM, Crystal B, Della Latta P, Fuchs M, Fuller D, Harris W, Hazen K, Kilmas LL, Lindao D, Meltzer MC, Morgan M, Shephard J, Stevens S, Wu F, Fiandaca MJ. 2012. Multicenter evaluation of the Staphylococcus QuickFISH method for simultaneous identification of Staphylococcus aureus and coagulase-negative staphylococci directly from blood culture bottles in less than 30 minutes. J. Clin. Microbiol. 50:1994–1998. 10.1128/JCM.00225-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.AdvanDx, Inc. Staphylococcus QuickFISH BC Staphylococcus aureus/coagulase-negative staphylococci culture identification kit package insert. AdvanDx, Inc., Woburn, MA: http://www.advandx.com/AdvanDX/media/Documents/Package%20Inserts/PN1878D---Staphylococcus-QuickFISH-BC-Kit.pdf. [Google Scholar]
- 11.Ubukata K, Nonoguchi R, Matsuhashi M, Konno M. 1989. Expression and inducibility in Staphylococcus aureus of the mecA gene, which encodes a methicillin-resistant S. aureus-specific penicillin-binding protein. J. Bacteriol. 171:2882–2885. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Clinical and Laboratory Standards Institute. 2013. Performance standards for antimicrobial susceptibility testing; 23rd informational supplement. CLSI document M100-S23 Clinical and Laboratory Standards Institute, Wayne, PA. [Google Scholar]
- 13.de Lencastre H, de Jonge BL, Matthews PR, Tomasz A. 1994. Molecular aspects of methicillin resistance in Staphylococcus aureus. J. Antimicrob. Chemother. 33:7–24. 10.1093/jac/33.1.7. [DOI] [PubMed] [Google Scholar]
- 14.Clopper CJ, Pearson ES. 1934. The use of confidence or fiducial limits illustrated in the case of the binomial. Biometrika 26:404–413. [Google Scholar]
- 15.Carretto E, Bardaro M, Russello G, Mirra M, Zuelli C, Barbarini 2013. Comparison of the Staphylococcus QuickFISH BC test with the tube coagulase test performed on positive blood cultures for evaluation and application in a clinical routine setting. J. Clin. Microbiol. 51:131–135. 10.1128/JCM.02103-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Holtzman C, Whitney D, Barlam T, Miller NS. 2011. Assessment of impact of peptide nucleic acid fluorescence in situ hybridization for rapid identification of coagulase-negative staphylococci in the absence of antimicrobial stewardship intervention. J. Clin. Microbiol. 49:1581–1582, 2011. 10.1128/JCM.02461-10. [DOI] [PMC free article] [PubMed] [Google Scholar]