INFLUENCE OF LIGHT ENERGY DENSITY ON HEAT GENERATION DURING PHOTOACTIVATION OF DENTAL COMPOSITES WITH DIFFERENT DENTIN AND COMPOSITE THICKNESS

Ricardo Danil Guiraldo 1, Simonides Consani 2,, Aline Simprine de Souza 3, Rafael Leonardo Xediek Consani 4, Mario Alexandre Coelho Sinhoreti 2, Lourenço Correr Sobrinho 2
PMCID: PMC4327644  PMID: 19668987

Abstract

Objective:

The aim of this study was to determine the influence of different energy densities on the heat generated during photoactivation of Filtek Z250 (3M/ESPE) and Z100 (3M/ESPE) composite resins with different dentin and composite thickness.

Material and Methods:

The temperature increase was registered with a type-K thermocouple connected to a digital thermometer (Iopetherm 46). A chemically polymerized acrylic resin base was prepared to serve as a guide for the thermocouple and as a support for 0.5-, 1.0-, and 1.5-mm-thick bovine dentin discs. Circular elastomer molds (1.0 mm-height x 3.0-mm diameter or 2.0-mm height x 3.0-mm diameter) were adapted on the acrylic resin base to standardize the composite resin thickness. A conventional halogen light-curing unit (XL 2500, 3M/ESPE) was used with light intensity of 700 mW/cm2. Energy density was calculated by the light intensity applied during a certain time with values of 28 J/cm2 for Z100 and 14 J/cm2 for Filtek Z250. The temperature change data were subjected to three-way ANOVA and Tukey's test at 5% level.

Results:

The higher energy density (Z100) promoted greater temperature increase (p<0.05) than the lower energy density (Filtek Z250). For both composites and all composite thicknesses, the lowest dentin thickness (0.5 mm) yielded significantly higher (p<0.05) temperature increase than the other two dentin thicknesses. The 1-mm-thick composite resin layer yielded significantly higher (p<0.05) temperature changes for both composites and all dentin thicknesses.

Conclusions:

Temperature increase was influenced by higher energy density and dentin/composite thickness.

Keywords: Energy density, Dentin thickness, Composite thickness, Production of heat, Temperature increase

INTRODUCTION

A dental composite can be defined as a three-dimensional combination of at least two chemically different materials with a different interface separating the components14. Basically they are composed of an organic matrix, load particles (glass, quartz and/or melted silica) and a bonding agent, usually an organic silane, with a dual characteristic making possible the chemical bonding with the load particle and the co-polymerization with the monomers of the organic matrix14. The resin matrix is a flowing component that becomes rigid when polymerized, due to the formation of free radicals that end up inducing covalent bonds among the organic molecules, generating macro-molecules denominated polymers3. For such purpose, it is necessary that an activator-initiator system generates enough energy to break the amina molecule and form free radicals7,14.

Photoactivation is performed with visible light belonging to the blue region of the electromagnetic spectrum to excite camphorquinone, which is the most common photoinitiator present in composite resins and has an absorption spectrum in the 400 and 500 nm range. The most efficient wavelength range for polymerization would be between 468 and 470 nm13, which excites the camphorquinone and induces the rupture of amina16, followed by an addition reaction polymerization.

Among the light-curing units available in the market, the most traditional ones are those that use halogen lamps as light source. However, the main irradiation produced by those lamps is in the infrared spectral region, which is absorbed by the composite and results in great molecular vibration and generation of heat19. Thus, light sources that use halogen lamps need absorbent filters that reduce the passage of infrared energy to the tooth16. The efficiency of these filters varies according to the manufacturer and, as a result, the non-absorbed energy can produce heat.

Increase in light intensity can lead to temperature rise during polymerization due to larger radiation energy supplied by the light-curing unit6. Polymerization of composites using visible light generates a temperature increase due to both the exothermal reaction process, which is typical of addition polymerization, and the energy absorbed during irradiation with the halogen light source12,18. For this reason, when the thickness of the residual dentin is small in cavities without lining and the light intensity is high, the curing time should be minimum18.

It has been suggested that visible light photoactivation can also contribute to increase the temperature inside the pulp chamber, causing pulpal damages9,11,12. However, the findings of a recent study investigating the thermal variations in the pulp chamber associated with composite insertion techniques and light-curing methods showed that when the bulk-increment technique was used there was no significant difference in temperature increase among the photoactivation methods (Continuous, Intermittent light, or Soft-Start)4, as well as between different shades of the same composite1. Pulp damage can occur when a source of external heat is applied on the dental structure because the increase of the temperature inside the pulp chamber can result in an irreversible inflammatory pulpal reaction8,20.

Based on these considerations, it would be important to verify the effect of different energy densities on the heat generated during the photoactivation of Filtek Z250 and Z100 composite resins with different dentin (0.5, 1.0, and 1.5 mm) and composite (1.0 and 2.0 mm) thicknesses. The null hypothesis tested was that different energy densities influence the heat produced during composite photoactivation with different dentin and composite thicknesses.

MATERIAL AND METHODS

Two restorative composite resins were used in this study: Filtek Z250 and Z100 (Figure 1). A conventional halogen light-curing unit (XL 2500, 3M/ESPE, St. Paul, MN, USA) was used with light intensity of 700 mW/cm2 checked with a curing radiometer (Model 100; Demetron Research Corporation, Danbury, CT, USA). The energy density was calculated as follows: energy density (J/cm2) is the light intensity (mW/cm2) applied during a certain time (s) divided by 1,000 (Table 1). The spectral distributions were obtained using a spectromer (USB 2000, Ocean Optics, Dunedin, FL, USA) (Figure 2).

FIGURE 1. Tested materials.

FIGURE 1

TABLE 1. Characteristics of the experimental groups.

Groups (n=10) Composite resin Dentin thickness Composite thickness Light intensity (mW/cm2) Exposure (s) Energy density (J/cm2)
Z 250 0.5 mm 1.0 mm 700 20 14
II Z250 0.5 mm 2.0 mm 700 20 14
III Z250 1.0 mm 1.0 mm 700 20 14
IV Z250 1.0 mm 2.0 mm 700 20 14
V Z250 1.5 mm 1.0 mm 700 20 14
VI Z250 1.5 mm 2.0 mm 700 20 14
VII Z100 0.5 mm 1.0 mm 700 40 28
VIII Z100 0.5 mm 2.0 mm 700 40 28
IX Z100 1.0 mm 1.0 mm 700 40 28
X Z100 1.0 mm 2.0 mm 700 40 28
XI Z100 1.5 mm 1.0 mm 700 40 28
XII Z100 1.5 mm 2.0 mm 700 40 28

FIGURE 2. Spectral distributions of XL 2500.

FIGURE 2

Temperature changes were recorded using a type-K thermocouple connected to a digital thermometer (Iopetherm 46; IOPE, São Paulo, SP, Brazil) accurate to 0.1°C. An apparatus for temperature measurement similar to that developed by Schneider, et al.17 (2005) was used. A chemically polymerized acrylic resin base (JET; Artigos Odontológicos Clássico, São Paulo, SP, Brazil) was prepared to serve as guide for the thermocouple and as support for 0.5-, 1.0-, and 1.5-mm-thick bovine dentin discs. The discs were positioned between the acrylic resin base and an elastomer mold to simulate different dentin thicknesses remaining after cavity preparation. Circular elastomer molds measuring either 1.0 mm-height and 3.0-mm diameter or 2.0-mm height and 3.0-mm diameter were adapted on the acrylic resin base to standardize the composite resin thickness (Figure 3).

FIGURE 3. Apparatus used for measurement of temperature increase.

FIGURE 3

The composite was inserted in the circular elastomer mold, covered with a Mylar strip and digitally pressed. For photoactivation, the tip of the light guide of the curing unit was placed in contact with the Mylar strip. The curing time was 20 s for Filtek Z250 and 40 s for Z100 according to manufacturer's instructions, which produced different energy densities. Twelve experimental groups (n=10) were established from the interactions among 2 energy densities, 3 dentin thicknesses and 2 composite thicknesses (Table 1).

All measurements were taken in a temperature/humidity-controlled room, with a constant temperature of 20°C ± 2°C and 30% relative humidity. For temperature measurements, the initial temperature was recorded following temperature stabilization. The composite was then light-cured and the temperature peak was registered. The initial temperature was deducted from the final temperature in order to determine the temperature increase.

The temperature change data were submitted to threeway ANOVA and the means were compared by Tukey's test at 5% significance level.

RESULTS

There was statistically significant difference (p<0.05) for temperature increase (Table 2) between the larger energy density (Z100) and the lower energy density (Z250).

TABLE 2. Mean temperature increase (°C) for Z250 and Z100 composite resins, regardless of the dentin and composite thicknesses.

Composite Mean temperature increase
Z 100 3.18 (1.12) a
Filtek Z250 2.13 (0.92) b

Different letters indicate statistically significant difference (Tukey's test, p<0.05). Standard deviations are given in parentheses.

For Z250 and Z100 composites with 1.0 mm thickness, the 0.5 mm-thick dentin differed significantly from the 1.0 mm-thick dentin, which in turn was significantly different (p<0.05) from the 1.5 mm-thick dentin. For both composites with 2.0 mm in thickness, the 0.5 mm-thick dentin differed significantly (p<0.05) from the other two dentin thicknesses, which however were no significantly different (p>.05) from each other. For both composites and all dentin thicknesses, the 1.0 mm-thick composite layer presented significantly greater temperature changes (p<0.05) than the 2.0 mm-thick composite increment (Tables 3 and 4).

TABLE 3. Mean temperature increase (°C) for Z250 composite resin for each dentin and composite thickness.

Composite Dentin thickness Composite thickness
1.0 mm 2.0 mm
Filtek Z250 0.5 mm 3.87 (0.38) a,A 1.97 (0.20) a,B
1.0 mm 2.56 (0.39) b,A 1.34 (0.17) b,B
1.5 mm 1.81 (0.15) c,A 1.26 (0.09) b,B

Different lowercase letters in columns and uppercase letters in rows indicate statistically significant difference (Tukey's test, p<0.05). Standard deviations are given in parentheses.

TABLE 4. Mean temperature increase (°C) for the composite Filtek Z100 for each dentin and composite thickness.

Composite Dentin thickness Composite thickness
1.0 mm 2.0 mm
Z100 0.5 mm 5.01 (0.19) a,A 2.95 (0.30) a,B
1.0 mm 4.24 (0.12) b,A 2.10 (0.21) b,B
1.5 mm 2.85 (0.19) c,A 1.97 (0.18) b,B

Different lowercase letters in columns and uppercase letters in rows indicate statistically significant difference (Tukey's test, p<0.05). Standard deviations are given in parentheses.

DISCUSSION

The null hypothesis that different energy densities influence the heat produced during composite photoactivation with different dentin and composite thicknesses was accepted.

External heat applied to the tooth can increase the temperature in the pulp chamber resulting in irreversible damage to the pulp tissue8,20. Trauma can be induced by the cavity preparation, exothermal reaction of cements, restorative materials or heat generated by light-curing units18,19. Photoactivation of composites using visible light can also contribute to the increase in temperature inside the pulp chamber, with potential thermal damages to the pulp11,12.

The increase of the temperature caused by photoactivation would result from the energy density emitted by the light-curing units5,16. Table 2 shows that the temperature increase produced by higher energy density (Z100) was significantly higher when compared to the lower energy density (Filtek Z250). It is likely the reason why in this study Z100 with energy density of 28 J/cm2 promoted higher temperature increase than Z250 with energy density of 14J/cm2. As the same light-curing unit (700 mW/cm2) was used for both materials, the difference in the exposure time s (40 s for Z100 and 20 s for Filtek Z250, as per manufacturer's instructions) seems to be the main responsible for the differences in heat generation by altering the energy density. Z100 and Z250 composites were used in this study because they possess similar organic matrixes and different filler volumes, which did not influence the increase of temperature. In this situation, the results suggest that energy density was the major factor accounting for temperature rise.

The heat that reaches the pulp can be minimized by the dentin thickness remaining after cavity preparation. The thicker the dentin layer the greater the minimization temperature that reaches the pulp because the dentin has low thermal conductivity10. This situation was observed in this study as smaller thickness of residual dentin yielded higher temperature increase values with statistically significant difference for both Filtek Z250 (Table 3) and Z100 (Table 4), except for the 1.0- and 1.5-mm-thick dentin discs with 2 mm-thick composite layer. This exception suggests that the resulting heat can be lessened by using a thickest composite layer. The light passing through the polymerized composite is absorbed and reflected, thus loosing intensity7,15, which can generate similar temperature increase. Similarly, for both Filtek Z250 (Table 3) and Z100 (Table 4) composites and for all dentin thicknesses (0.5, 1.0 and 1.5 mm), the 1.0-mm-thick composite layer always resulted in significantly higher temperature increase than the thickness of 2.0 mm. A previous study showed that the heat produced in an irradiated empty cavity was always higher when compared to the polymerization of a 2.0-mm-thick composite increment. This is due to the extenuating capacity of the material to exceed the exothermal effect of the polymerization18.

Within the limitations of this study, the results seem to indicate that the temperature increase resulting from the photoactivation of composites would be well tolerable by the dental pulp, even considering the extreme condition observed with Z100 composite, in which a higher energy density (28 J/cm2) was used with 0.5 mm-thick dentin and 1.0 mm-thick composite increment. It has been demonstrated that temperature changes over 5.5°C would be necessary to induce pulp alterations in monkey's teeth20.

CONCLUSIONS

Based on the analyzed and discussed results it may be concluded that:

  1. The higher energy density (Z100) promoted higher temperature increase than the lower energy density (Z250).

  2. For both composites and all composite thicknesses, the lowest dentin thickness yielded significantly higher temperature increase than the other two dentin thicknesses.

  3. For both composites and all dentin thicknesses, the 1.0 mm-thick composite layer presented significantly greater temperature changes than the 2.0 mm-thick composite increment.

  4. Temperature increase was influenced by higher energy density and dentin/composite thickness

ACKNOWLEDGMENTS

The authors wish to thank CNPq/PIBIC/UNICAMP for the financial support.

REFERENCES

  • 1.Consani S, Farina EDP, Guiraldo RD, Sinhoreti MAC, Correr L., Sobrinho Influence of shade and composition in the generation of heat during the dental composite photoactivation. Braz J Oral Sci. 2006;5(19):1213–1216. [Google Scholar]
  • 2.Goodis HE, White JM, Andrews J, Watanabe LG. Measurement of temperature gererated by visible-light-cure lamps in an in vitro mode. Dent Mater. 1989;5(4):230–234. doi: 10.1016/0109-5641(89)90066-3. [DOI] [PubMed] [Google Scholar]
  • 3.Guggenberger R, Weinmann W. Exploring beyond methacrylates. Am J Dent. 2000;13(sp. issue):82D–84D. [PubMed] [Google Scholar]
  • 4.Guiraldo RD, Consani S, Sinhoreti MAC, Correr L, Sobrinho, Schneider LFJ. Thermal variations in the pulp chamber associated with composite insertion techniques and light-curing methods. J Contemp Dent Pract. 2009;10(1):17–24. [PubMed] [Google Scholar]
  • 5.Guiraldo RD, Consani S, Lympius T, Schneider LFJ, Sinhoreti MAC, Correr L., Sobrinho Influence of the light-curing unit and thickness of residual dentin on generation of heat during composite photoactivation. J Oral Sci. 2008;50(2):137–142. doi: 10.2334/josnusd.50.137. [DOI] [PubMed] [Google Scholar]
  • 6.Hansen EK, Asmussen E. Correlation between depth of cure and temperature rise of a light-activated resin. Scand J Dent Res. 1993;101(3):176–179. doi: 10.1111/j.1600-0722.1993.tb01659.x. [DOI] [PubMed] [Google Scholar]
  • 7.Knezevic A, Tarle Z, Meniga A, Sutalo J, Pichler G, Ristic M. Degree of conversion and temperature rise during polymerization of composite resin samples with blue diodes. J Oral Rehabil. 2001;28(6):586–591. doi: 10.1046/j.1365-2842.2001.00709.x. [DOI] [PubMed] [Google Scholar]
  • 8.Lisanti VF, Zander HA. Thermal injury to normal dog teeth: in vivo measurements to pulp temperature increases and their effect on the pulp tissue. J Dent Res. 1952;31(4):548–558. doi: 10.1177/00220345520310040501. [DOI] [PubMed] [Google Scholar]
  • 9.Lloyd CH, Joshi A, McGlynn E. Temperature rises produced by light sources and composites during curing. Dent Mater. 1986;2(4):170–174. doi: 10.1016/s0109-5641(86)80030-6. [DOI] [PubMed] [Google Scholar]
  • 10.Loney RW, Price RBT. Temperature transmission of High-output light-curing units through dentin. Oper Dent. 2001;26(5):516–520. [PubMed] [Google Scholar]
  • 11.Masutani S, Setcos JC, Schinell RJ, Philips RW. Temperature rise during polymerization of visible light-activated composite resins. Dent Mater. 1998;4(4):174–178. doi: 10.1016/s0109-5641(88)80059-9. [DOI] [PubMed] [Google Scholar]
  • 12.McCabe JF. Cure performance of light-activated-composites by differential thermal analysis (DTA) Dent Mater. 1985;1(6):231–234. doi: 10.1016/S0109-5641(85)80048-8. [DOI] [PubMed] [Google Scholar]
  • 13.Nomoto R. Effect of light wavelength on polymerization of light-cured resins. Dent Mater J. 1997;16(1):60–73. doi: 10.4012/dmj.16.60. [DOI] [PubMed] [Google Scholar]
  • 14.Peutzfeldt A. Resin composites in dentistry: the monomer systems. Eur J Oral Sci. 1997;105(2):97–116. doi: 10.1111/j.1600-0722.1997.tb00188.x. [DOI] [PubMed] [Google Scholar]
  • 15.Rueggeberg FA, Caughman WF, Curtis JW., Jr Effect of light intensity and exposure duration on cure of resin composite. Oper Dent. 1994;19(1):26–32. [PubMed] [Google Scholar]
  • 16.Rueggeberg FA. Contemporary issues in photocuring. Compend Contin Educ Dent. 1999;(Suppl.)(25):S4–S15. [PubMed] [Google Scholar]
  • 17.Schneider LFJ, Consani S, Sinhoreti MAC, Correr L, Sobrinho, Milan FM. Temperature change and hardness with different resin composites and photo-activation methods. Oper Dent. 2005;30(4):516–521. [PubMed] [Google Scholar]
  • 18.Shortall AC, Harrington E. Temperature rise during polymerization of light-activated resin composites. J Oral Rehabil. 1998;25(12):908–913. doi: 10.1046/j.1365-2842.1998.00336.x. [DOI] [PubMed] [Google Scholar]
  • 19.Uhl A, Mills RW, Jandt KD. Polymerization and light-induced heat of dental composites cured with LED and halogen technology. Biomaterials. 2003;24(10):1809–1820. doi: 10.1016/s0142-9612(02)00585-9. [DOI] [PubMed] [Google Scholar]
  • 20.Zach L, Cohen G. Pulp response to externally applied heat. Oral Surg Oral Med Oral Pathol. 1965;19(4):515–530. doi: 10.1016/0030-4220(65)90015-0. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Applied Oral Science are provided here courtesy of Bauru School of Dentistry