Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2015 Feb 23.
Published in final edited form as: IEEE Trans Biomed Eng. 2012 Nov 15;60(2):361–368. doi: 10.1109/TBME.2012.2226885

Simultaneously Extracting Multiple Parameters via Fitting One Single Autocorrelation Function Curve in Diffuse Correlation Spectroscopy

Lixin Dong 1, Lian He 1, Yu Lin 1, Yu Shang 1, Guoqiang Yu 1,*
PMCID: PMC4337031  NIHMSID: NIHMS658482  PMID: 23193446

Abstract

Near-infrared diffuse correlation spectroscopy (DCS) has recently been employed for noninvasive acquisition of blood flow information in deep tissues. Based on the established correlation diffusion equation, the light intensity autocorrelation function detected by DCS is determined by a blood flow index αDB, tissue absorption coefficient μa, reduced scattering coefficient μs’, and a coherence factor β. The present study is designed to investigate the possibility of extracting multiple parameters such as μa, μs’, β, and αDB through fitting one single autocorrelation function curve and evaluate the performance of different fitting methods. For this purpose, computer simulations, tissue-like phantom experiments and in-vivo tissue measurements were utilized. The results suggest that it is impractical to simultaneously fit αDB and μa or αDB and μs’ from one single autocorrelation function curve due to the large crosstalk between these paired parameters. However, simultaneously fitting β and αDB is feasible and generates more accurate estimation with smaller standard deviation compared to the conventional two-step fitting method (i.e., first calculating β and then fitting αDB). The outcomes from this study provide a crucial guidance for DCS data analysis.

Keywords: Near-infrared spectroscopy, diffuse correlation spectroscopy, blood flow, noise model, autocorrelation function

I. INTRODUCTION

Near-infrared (NIR) light has recently been employed for noninvasive acquisition of blood flow information in deep tissues (up to several centimeters), which is referred to as NIR diffuse correlation spectroscopy (DCS) [1, 2] or diffusing-wave spectroscopy (DWS) [3-5]. Blood flow variations measured by DCS have been validated in various organs and tissues against other standards, including Doppler ultrasound [6], power Doppler ultrasound [7], laser Doppler [8], Xenon-CT [9], fluorescent microsphere flow measurement [10], and perfusion MRI [11]. DCS delivers continuous-wave (CW) coherent NIR light into tissue wherein photons encounter absorption and, more commonly, scattering events. The probabilities of these events are described by tissue optical properties: absorption coefficient μa and reduced scattering coefficient μs’. Scattered light is detected by a photodetector placed on the tissue surface at a certain distance (e.g., several centimeters) from a light source. Most photons detected experience multiple scattering events and each scattering event is associated with a random scattering phase shift. The superposition of multiple light fields with different phases creates a speckle pattern of interference.

The motions of moving scatterers, primarily red blood cells in biological tissues, cause fluctuations in light intensity, leading to changes in speckle pattern. These fluctuations/changes carry information about the dynamic properties of moving red blood cells. Time dependent light intensity fluctuations can be measured by the photodetector on the tissue surface and quantified by temporal autocorrelation functions. The electric field autocorrelation function is related to the measured light intensity autocorrelation function through the Siegert relation [12]. It has been found that the electric field autocorrelation function is governed by a correlation diffusion equation [1, 2], and blood flow index (BFI) in biological tissues can be calculated by fitting the measured autocorrelation function curve with the solution of correlation diffusion equation.

Based on the correlation diffusion equation and Siegert relation, however, the measured light intensity autocorrelation function is determined by not only blood flow, but also tissue optical properties (i.e., μa and μs’) and a coherence factor β. β relies mainly on light source and detection optics. It is thus desirable to extract as much information as possible (i.e., multiple parameters) from one single autocorrelation function. Some previous studies have chosen to use the values of μa and μs’ from literature respective to the measured tissue type (e.g., brain or muscle) for the calculation of DCS blood flow [13, 14]. These assumptions are susceptible to deviations in tissue optical properties [15]. A few of recent studies have employed hybrid instruments (NIR spectroscopy combining DCS) allowing for concurrent measurements of both μa and μs’ to extract accurate BFI [6, 16-18]. In addition, most previous studies estimated β based on the Siegert relation using the measured autocorrelation function data at the earliest correlation delay time, and then fitted BFI (i.e., two-step fitting method) [7, 19-22]. Although a few recent studies claimed fitting β and BFI simultaneously [18, 23, 24], none of them have compared the performance of the two methods (i.e., simultaneous fitting versus two-step fitting).

The present study is designed to investigate the possibility of simultaneously extracting multiple parameters such as μa, μs’, β, and BFI through fitting one single autocorrelation function curve and evaluate the performance of different fitting methods. For this purpose, computer simulations, tissue-like phantom experiments, and in-vivo tissue measurements were utilized. It is expected that the outcomes from this study will ultimately improve DCS data analysis.

II. METHODS

A. Diffuse Correlation Spectroscopy (DCS) for Flow Measurements

The flow index is quantified by a DCS flowmeter built in our laboratory. Details about DCS for flow measurements can be found elsewhere [8, 21, 22, 25, 26]. Briefly, long-coherence (> 5 meters) NIR CW light emitted from a laser diode (785 nm, ~100 mw, Crystalaser Inc., NV) enters the tissue via a 200 μm diameter multimode source fiber (Thorlabs, Inc., NJ). The transported/scattered light through tissue is collected by a ~5 μm diameter single-mode detector fiber (SM 600, Fibercore Inc., CA) connected to a single-photon-counting avalanche photodiode (APD, PerkinElmer Inc., Canada). The transistor-transistor logic (TTL) pulses are output from the APD and associated with the number of photons detected from a small area covering a single speckle on tissue surface. These signals are fed into a correlator board (correlator.com, NJ) for computing the light intensity [i.e., photon count rate with a unit of kilo counts per second (kcps)] and intensity temporal autocorrelation function [1, 2]:

g2(r,τ)=I(r,t)I(r,t+τ)I2 (1)

Here I(r,t) is the detected light intensity at position r and time t, ⟨…⟩ denotes a time average, and τ is the autocorrelation delay time. The sampling rate for the DCS flowmeter is 1 Hz.

In highly scattering media, such as biological tissues, the electric field temporal autocorrelation function G1(r,τ)=E(r,t)E(r,t+τ) satisfies the correlation diffusion equation [1, 2]:

(D2νμa13νμsk02αΔr2(τ))G1(r,τ)=νS(r) (2)

Here, v is the speed of light in the medium, k0 is the wavenumber of light in the medium, S(r) is the source light distribution, α is defined as the ratio of moving scatterers to total scatterers, μa is medium absorption coefficient, μs’ is reduced scattering coefficient, D = v/3 (μa + μs’) is the photon diffusion coefficient, and <Δr2(τ)> is the mean-square displacement of moving scatterers in time τ. Intuitively, the random flow model might be considered the best model with which to fit DCS data. In practice, however, it has been observed that the diffusion model fits the autocorrelation curves rather well over a broad range of tissue types [6, 7, 10, 13, 15, 16, 19-21, 27-33]. For the case of diffusive motion, <Δr2(τ)> = 6DBτ, where DB is the effective Brownian diffusion coefficient of scatterers. The combined term, αDB, is referred to as blood flow index (BFI) in biological tissues and is commonly used to calculate the relative change of blood flow (rBF), compared to baseline BFI before physiological changes. The unit of BFI (αDB) is cm2/s. Although this unit is different from the classical blood flow unit in biological tissues (ml/min/100ml), percentage changes in αDB have been found to correlate well with the blood flow changes measured by many other established modalities [8-11, 34].

The homogeneous CW solution to Eq. (2) for a semi-infinite geometry is [35]

G1(ρ,τ)=νS04πD(exp(K(τ)r1)r1exp(K(τ)r2)r2) (3)

Here ρ is the source-detector separation, S0 is source intensity, K2(τ)=3μaμs+μs2k02αΔr2(τ)=3μaμs+6μs2k02αDBτ, r1=[ρ2+(zz0)2]12, r2=[ρ2+(z+z0+2zb)2]12, z0=1μs, zb=2(1+Reff)3μs(1Reff), Reff = −1.440n−2 + 0.710n−1 + 0.668 + 0.0636n and n ≈ 1.34. The Reff is the internal reflection coefficient which accounts for the refractive index mismatch between the medium and air, and n is the ratio of reflective indices between them.

The normalized electric field autocorrelation function g1(r,τ)=G1(r,τ)G1(r,0) is related to the measured intensity autocorrelation function g2(r,τ) through the Siegert relation [12]:

g2(r,τ)=1+βg1(r,τ)2 (4)

Here β is a coherence factor and inversely proportional to the number of speckles detected. Although complete understanding of all factors that affect β needs further investigation, it is thought to mainly depend on light source and detection optics. The light source may be influenced by light coherence, laser stability and stray light while the detection optics may be affected by the detector stability and fiber-tissue coupling coefficient. Some of these factors (if not all) may change during the time course of measurements, leading to a variation in β. When a single-mode fiber is used for DCS flow detection, the maximum β value should be ~0.5 considering the two orthogonal polarization modes collected from the fiber [36]. When a polarizer is placed on the detector fiber, a β value of ~1 can be achieved. However, β will decrease when few-mode fibers (instead of the single-mode fiber) are used [37].

B. Noise Model for Simulation of Autocorrelation Functions

In order to simulate autocorrelation functions measured in real media, a proper estimate of measurement noises is needed. Previously, a noise model with single scattering limit in fluorescence correlation spectroscopy [38] has been adopted for use in diffuse correlation experiments wherein photons experience multiple scattering events [28, 37]. The phantom experiments demonstrated that the noise model provided a good estimate of DCS measurement noises in homogeneous media with infinite geometry. Briefly, the measured correlation function [g2(τ) − 1] was assumed to decay approximately exponentially, i.e., g2(τ)−1 = β exp(−Γτ. The experimental configuration was characterized by the correlator bin time interval T, bin index m corresponding to the delay time τ, average number of photons <n> within T [i.e., <n> = I·T, where I was the detected photon intensity], total averaging time t, and coherence factor β. The noise [standard deviation σ(τ)] of the measured correlation function [g2(τ) − 1] at each delay time τ was estimated to be [28, 38]:

σ(τ)=Tt[β2(1+e2ΓT)(1+e2Γτ)+2m(1e2ΓT)e2Γτ(1e2ΓT)+2n1β(1+e2Γτ)+n2(1+βeΓτ)]12 (5)

Accordingly, the signal-noise-ratio (SNR) of DCS measurements at delay time τ was SNR(τ)=g2(τ)1σ(τ), and 1/SNR(τ) can be used to estimate the noise level of DCS measurements.

For the case of diffuse reflectance measurement on medium surface, a semi-infinite geometry should be considered instead and Eq. (3) should be used to calculate [g2(τ) − 1] rather than the assumption of exponential decay function as used in Reference [28]. However, mathematically it is difficult to derive a noise model based on the complex Eq. (3). In the present study, we designed phantom experiments to test the accuracy of this noise model [Eq. (5)] for use in homogeneous media with semi-infinite geometry. Tissue-like liquid phantoms were created with Intralipid for control of scattering (μs’) and particle Brownian motion (αDB), India ink for control of absorption (μa), and distilled water [15]. Although temperature can affect Brownian motions (αDB) of Intralipid particles, the room temperature was controlled constant (~23 °C) in order to obtain stable αDB (flow). Tissue-like liquid phantoms have been commonly used for the calibration of DCS techniques [1, 15, 27]. The phantom was contained in a glass aquarium. A fiber-optic probe with a pair of source and detector fibers at a distance of 2.5 cm was secured in contact with the surface of the liquid phantom solution using a custom-designed holder. We set constant μs’ = 8 cm−1 and vary μa (0.075, 0.100, 0.150 cm−1) by adding ink to test the noise-model under different levels of noise. Higher μa was associated with lower number of photons detected (I), thus leading to higher level of noise [σ(τ)].

After the noise model was verified for semi-infinite geometry, it was used to generate the normalized intensity autocorrelation curve g2 with noise. An autocorrelation curve (g2) was firstly generated using Eq. (3) and Eq. (4). The standard deviation σ(τ) of [g2(τ) − 1] was then calculated using Eq. (5), wherein the β and Γ were obtained concurrently by fitting the g2 curve with the exponential approximation (the concurrent fitting method can be found in Section II-D), i.e., g2(τ)−1 = β exp(−Γτ). Following a Gaussian distribution with zero mean and standard deviation σ(τ), noises that varied at different delay time τ were generated. The simulated noises were then applied on g2(τ) to create an intensity autocorrelation curve with noise. As indicated in Eq. (5), the SNR and noise level (1/SNR) of simulated autocorrelation functions were adjusted by changing the light intensity or photon count rate (I = <n>/T).

C. Extraction of Multiple Parameters through Fitting One Single Autocorrelation Curve

Multiple parameters examined (i.e., αDB, μa, μs’ and β) were extracted by fitting the measured autocorrelation function curve to the analytical solution of correlation diffusion equation [Eq. (3)]. The goal was to minimize the sum of squared differences (SSD) between the measured and calculated autocorrelation functions. The minimization of the objective function SSD = ∑[g2,m(τ) - g2,c(τ)]2 was done by using Nelder-Mead simplex algorithm (fminsearch function) in Matlab (Mathwork, Inc., MA), where g2,m(τ) was the measured intensity autocorrelation function and g2,c(τ) was the analytical model of autocorrelation in the semi-infinite reflection geometry [Eq. (3)]. Initial guesses for these parameters were assigned randomly using the “rand” function in Matlab. The random variation ranges of these parameters were determined based on the dynamic ranges in DCS phantom experiments and in-vivo measurements: μa = 0.05 to 0.4 cm−1, μs’ = 2 to 15 cm−1, αDB = 0.4 to 2 × 10−8 cm2/s, and β = 0.1 to 0.9. We set the termination tolerance for the fitted variables (TolX) at 10−11, which is ~1000 times smaller than the value of αDB (0.4 to 2 × 10−8 cm2/s). This termination criterion is strict enough to obtain precise results for all four parameters.

In order to determine the possibility of fitting μa, μs’ or β along with αDB from one single autocorrelation curve, we examined the SSD change patterns by varying three pairs of the four parameters (i.e., μa and αDB, μs’ and αDB, β and αDB), respectively. For this purpose, a reference light intensity autocorrelation curve without noise, g20, was initially generated using Eq. (3) and Eq. (4) with the given parameters: μa = 0.12 cm−1, μs’ = 8 cm−1, αDB = 10−8 cm2/s, and β = 0.45. The paired parameters were then varied to generate multiple testing autocorrelation curves. The variation ranges of these parameters were the same as those indicated above. The SSDs between the testing autocorrelation curves and the reference autocorrelation curve g20 were calculated and presented in contour plots as functions of these paired parameters, respectively. Different SSD patterns (e.g., convergence or divergence) implied the possibility of extracting multiple parameters from one single autocorrelation curve.

The possibility of extracting multiple parameters was further examined by fitting the paired parameters simultaneously from the reference autocorrelation curve g20 with two different levels of noise (I = 100 and 50 kcps). For each noise level, 1000 simulated curves were generated and fitted to extract the paired parameters simultaneously. The discrepancies between the fitted and given values of the paired parameters are expressed as “percentage errors”.

D. Comparison of Two Methods for Extracting β and αDB

In most previous studies, extracting αDB began with using Eq. (4) to determine β. Using DCS measured g2 data at earliest τ and letting g1 ≈ 1 [i.e., g1(ρ,0)=G1(ρ,0)G1(ρ,0)=1] led to β = g2(ρ,τ ≈ 0)−1. Using g2(ρ,τ), β and Eq. (4), g1(ρ,τ) was calculated for all τ. Equation (3) was then used with the unknown parameter αDB to fit the g1(ρ,τ) derived from DCS measurements. In addition, one could also average more g2 datasets (i.e., several data points instead of one single data point) at early τ to reduce the noise influence for determining β.

During the study of extracting multiple parameters from one single autocorrelation curve (see Section II-C), we found it possible to simultaneously fit both β and αDB (see Sections III-A and III-C). To compare the performance between the two methods (i.e., two-step fitting versus simultaneously fitting) for extracting β and αDB, computer simulations, phantom experiments and in-vivo tissue measurements were utilized. For simulations, the reference autocorrelation curves g20 with noise were generated based on the procedures described in Section II-B, and the levels of noise were changed by varying photon count rate from 20 to 500 kcps. At each noise level, 1000 curves were created and fitted by the two methods to examine if they were able to extract the expected values of β and αDB from the simulated curves.

The simulation results were further verified with the data collected from the phantom experiments (see Section II-B) and in-vivo tissue measurements in forearm flexor muscle. A single-mode detector fiber was placed at a distance of 2.5 cm from the source fiber for the phantom or tissue measurement. The source and detector fibers were confined in their positions by a foam pad to form a fiber-optic probe which was placed on the surface of the measured phantom or tissue. The in-vivo measurement was taken from one healthy volunteer who signed the informed consent approved by the University of Kentucky Institutional Review Board. Before the experiment, absolute values of tissue optical properties (μa and μs’) in forearm flexor muscle were measured by a frequency-domain NIR tissue-oximeter (Imagent, ISS Inc., IL). The measured μa and μs’ were used as input parameters in calculation of β and αDB with the two methods. To reduce the influence of physiological variations, the baseline drift of the in-vivo tissue measurement was removed using a first order high-pass Butterworth filter with a cutoff frequency of 0.05 Hz. For statistical analyses, significances of the difference between the two methods were tested using a paired t-test. The criterion for significance is p < 0.05.

III. RESULTS

A. The Noise Model can be Used in Homogeneous Media with Semi-infinite Geometry

Figure 1 shows the results from phantom experiments to verify the feasibility of applying the noise model in homogeneous media with semi-infinite geometry. During the three titrations of varying μa (0.075, 0.100, 0.150 cm−1) while keeping μs’ constant (8 cm−1), photon counting rates changed correspondingly (66, 41, 20 kcps). 267, 237 and 235 autocorrelation curves were collected sequentially at the three titration steps. The noise [σ(τ)] (Fig. 1a) and SNR (Fig. 1b) of the autocorrelation function at each τ were calculated and plotted (see the dots in Fig. 1). The solid curves represent the calculated noises or SNRs using Eq. (5) with the parameters obtained from the phantom experiments; β and Γ were obtained simultaneously by fitting the experimental data with the exponentially decaying function (see Sections II-C and II-D); the averaging time to obtain one correlation function curve was kept constant (t = 1 s) for all measurements; the photon count rates were recorded by the correlator board; the bin time interval T was 121 ns for the first 32 channels and doubled every 16 channels thereafter. As shown in Fig. 1, the measurement noise decreased as the delay time τ or light intensity increased, whereas the SNR increased as the light intensity increased and changed with the variation of delay time τ. These results are consistent with the predictions from Eq. (5) and suggest that the noise model provides a good estimate for DCS noises measured in homogeneous media with semi-infinite geometry.

Fig. 1.

Fig. 1

Tissue-like phantom experiments to verify the feasibility of applying the noise model in homogeneous media with semi-infinite geometry. DCS measurements were performed in three liquid phantoms with different μa (0.075, 0.100, 0.150 cm−1) to test the noise model under different noise levels. Higher μa was associated with lower number of photons detected (I), thus leading to higher measurement noise [σ(τ)] and lower signal-to-noise ratio [SNR(τ)]. (a) Comparison of the measurement noises between the measured autocorrelation curves from the phantoms (dots) and calculated noises predicted by the noise model (solid curves). The measurement noise decreased as the delay time τ increased. The “steps” were due to the multi-tau arrangement of the correlator. (b) Comparison of the SNRs between the measured autocorrelation curves and model predictions. Although the measurement noise decreased as the delay time τ increased, the SNR of DCS measurement also decreased because the “signal” dropped even faster than the noise as τ increased.

We then used this noise model to generate autocorrelation curves (g2) with three different levels of noise (I = 20, 41, 66 kcps) observed in phantom tests. Figure 2 (a, b, c) shows the results comparing the simulated and measured (from phantoms) g2 curves with the same levels of noise. Again, the results demonstrate that the noise model works well in homogeneous media with semi-infinite geometry.

Fig. 2.

Fig. 2

Comparison of the simulated and measured g2 curves at three different levels of noise: (a) I = 66 kcps; (b) I = 41 kcps; and (c) I = 20 kcps. The input parameters for simulation were acquired from the phantom experiments.

B. SSD Patterns Imply the Possibility of Extracting Multiple Parameters from One Single Autocorrelation Curve

As indicated in Section II-C, a reference g2 curve without noise was generated with the given parameters: μa = 0.12 cm−1, μs’ = 8 cm−1, αDB = 10−8 cm2/s, β = 0.45. The SSDs between the reference curve g20 and the testing curves generated by varying the values of paired parameters were calculated and presented in Fig. 3. The SSD values were marked on the curves.

Fig. 3.

Fig. 3

Contour plots of sum of squared differences (SSD) between a reference autocorrelation curve g20 obtained with the given parameters [αDB = 10−8 cm2/s, μa (785 nm) = 0.12 cm−1, and μs’ (785 nm) = 8 cm−1] and the testing curves generated by varying the values of paired parameters. The true reference values of g20 are marked at the cross. (a) The SSDs obtained by varying μs’ from 2 to 15 cm−1 and αDB from 0.4 to 2 × 10−8 cm2/s. The dashed black curve represents the points at which local minima were achieved with compositions of μs’ and αDB. (b) The SSDs obtained by varying μa from 0.05 to 0.4 cm−1 and αDB from 0.4 to 2 × 10−8 cm2/s. The dashed black curve illustrates the points at which local minima were achieved. (c) The SSDs obtained by varying β from 0.1 to 0.9 and αDB from 0.4 to 2 × 10−8 cm2/s. The minimum value of SSD was reached at the true reference values of β and αDB.

Different patterns of SSD were observed for different pairs of parameters. The curves for the pairs of μs’/αDB (Fig. 3a) and μa/αDB (Fig. 3b) were divergent although the proportional relationships between the two paired parameters were opposite. Large crosstalk existed between αDB and μa or αDB and μs’, even if when the SSDs were close to zero (see the dashed curves in Fig. 3a and Fig. 3b). Due to the fact that the SSD was a highly nonlinear function of μa, μs’, β and αDB, the points on the dashed lines did not have exactly the same SSD value with one another, but the differences among these values were extremely small. Although the cross (+) point (with the true reference values) was the absolute minimum, there were numerous local minima along the dashed line. This made it difficult to obtain the absolute minimum at the cross point because the searching algorithm could get stuck easily in a local minimum. Conversely, the curves for the pairs of β and αDB (Fig. 3c) were convergent. The minimum value of SSD was reached at the true reference values of β and αDB (the cross point). In total, these results suggest that it is possible to fit β and αDB simultaneously and precisely.

C. Simultaneously Fitting μa, μs’ or β along with αDB from One Single Autocorrelation Curve

The results we got from Section III-B were further confirmed by simultaneously fitting the simulated autocorrelation g20 curves with different levels of noise. Figure 4 shows the fitting results for extracting the three pairs of parameters simultaneously. The upper (Fig. 4a and Fig. 4b) and lower (Fig. 4c and Fig. 4d) panels represent the simulation results with two different levels of noise (I = 100 and 50 kcps). As expected, simultaneously fitting μa and αDB may result in estimation errors; underestimating/overestimating μa of −60%/+200% led to flow index errors up to −100%/+50%. Simultaneously fitting μs’ and αDB may generate even larger errors than simultaneously fitting μa and αDB; underestimating/overestimating μs’ from −40%/+200% led to flow index errors up to +120%/−80%. These errors were mainly due to the crosstalk between the paired parameters. By contrast, when fitting β and αDB simultaneously, the output values of fitted parameters (β and αDB) clustered around the true values (Fig. 4b and Fig. 4d). The estimation errors for both β and αDB at the two noise levels were smaller than 15%.

Fig. 4.

Fig. 4

Percentage errors when simultaneously fitting three pairs of four variables respectively from the simulated autocorrelation curves (g20) generated at two different levels of noise. The initial guesses of the fitted variables were assigned randomly when fitting each of the 1000 simulated g20 curves._The discrepancies between the fitted and given values of the paired parameters are expressed as “% errors”. The upper (a and b) and lower (c and d) panels represent simulation results with two different levels of noise (I = 100 and 50 kcps). Higher noise level caused larger evaluation errors. The left panel (a and c) shows the results when simultaneously fitting αDB and μs’ or αDB and μa. Large crosstalk between the paired parameters was apparent, resulting in large estimation errors. The right panel (b and d) shows the results when simultaneously fitting β and αDB. The estimation errors for both β and αDB were much smaller than those shown in the left panel (a and c).

D. Results from the Comparison of Two Methods for Extracting β and αDB

The results shown in Sections III-B and III-C indicate that β and αDB can be simultaneously extracted by fitting one single autocorrelation curve. To compare the performance of the concurrent-fitting method with the conventional two-step fitting method (i.e., first calculating β and then fitting αDB) for extracting β and αDB, computer simulations, phantom experiments, and in-vivo tissue measurements were conducted in the present study.

Figure 5 shows the comparison results from the simulated g20 curves with different levels of noise (i.e., photon count rate varied from 20 to 500 kcps). For the two-step method, we used either the first data point (upper panel) or averaged seven data points (lower panel) of g20 at early delay time τ to estimate β. For both concurrent and two-step fitting methods, the standard deviation of fitted values increased with the increase of noise level. However, simultaneously fitting β and αDB generated more accurate values with significantly smaller standard deviations (error bars) compared to the two-step fitting method (p < 0.001). Inaccurate estimation of β resulted in errors in fitting αDB, which became more remarkable when the noise level increased. As expected, using one point to estimate β resulted in large standard deviations of estimation, which can be reduced by averaging more data points (7 points in this simulation) of g20. Because of the decay of autocorrelation curve with τ (see Fig. 2), however, the averaging led to significant underestimations of β and αDB (p < 0.001).

Fig. 5.

Fig. 5

Comparison of the two methods for extracting β and αDB from the simulated g20 curves under different noise levels (i.e., the photon count rate changes from 500 kcps to 20 kcps). Dashed lines indicate the expected values (β = 0.45 and αDB = 1 × 10−8 cm2/s). For both methods, the standard deviation (error bar) of the fitted values increased with the increase of noise level. Fitting β and αDB simultaneously resulted in unbiased and more accurate estimation of the means and smaller standard deviations compared to the two-step fitting method. When estimating β from the first point (a and b), the standard deviations of β and αDB were large. Estimating β from more points (c and d) reduced the standard deviations of estimation with the cost of estimation biases in β and αDB.

These simulation results were confirmed by the phantom experiments and in-vivo tissue measurements. Figure 6a shows the estimation deviations of β and αDB from the phantom experiments described in Section II-B. Notice that only the data from the second step of titration (μa = 0.10 cm−1) are presented although the results from the other two steps (μa = 0.075 and 0.15 cm−1) were similar. Since the concurrent-fitting method generated accurate estimates for β and αDB (see simulation results above), the mean values of β and αDB obtained by this method were assumed to be ‘true’ values of the measured phantom. The percentage deviations of β and αDB estimated from each autocorrelation curve are presented as error bars in Fig. 6a. Compared to the concurrent-fitting method, inaccurate estimation of β by the two-step fitting method may result in significant estimation errors in αDB (p < 0.001) and lead to larger error bars.

Fig. 6.

Fig. 6

The performance of the two fitting methods evaluated with the (a) phantom test and (b) in-vivo measurement. The mean values obtained by the simultaneous fitting method were assumed as ‘true’ values, and the discrepancies between the values extracted using the simultaneously fitting method or two-step fitting method are presented as percentage errors. The error bars are shown with crosses respectively and the mean values are located at the center of the crosses.

Similarly, Figure 6b shows the estimations of β and αDB from the data collected in in-vivo tissue measurements described in Section II-D. 574 autocorrelation curves were collected from the subject’s forearm using the DCS device. The results shown in Fig. 6b agreed with those of the simulations shown in Fig. 5 and the phantom experiments shown in Fig. 6a.

IV. DISCUSSION AND CONCLUSIONS

This study was designed to investigate the possibility of extracting multiple parameters such as μa, μs’, β, and αDB through fitting one single autocorrelation function curve and compare the performance of different fitting methods. For this purpose, the patterns of the sum of squared differences (SSD) between the reference autocorrelation curve (g20) and the testing autocorrelation curves generated by varying the values of paired parameters were examined; different SSD patterns (e.g., convergence or divergence) implied the possibility of extracting multiple parameters from a single autocorrelation curve. These results were then verified by computer simulations, phantom experiments and in-vivo tissue measurements.

For simulations with our measurement configuration, a noise model for homogeneous media with semi-infinite geometry is needed to generate autocorrelation curves with noise. Previous studies have used a noise model described in Section II-B as an approximation [28, 37], which has never been validated in semi-infinite geometry. In the present study, we designed phantom experiments to test the accuracy of this noise model for use in homogeneous media with semi-infinite geometry. The phantom experimental results agreed with the theoretical predictions from the noise model (Fig. 1a and Fig. 1b) suggesting that it provides a good estimate for DCS noises measured in homogeneous media with semi-infinite geometry. This noise model was thus used to generate autocorrelation curves with different levels of noise, and the simulated curves were compared with the curves collected from phantom measurements (Fig. 2a, Fig. 2b and Fig. 2c). The comparison results confirmed that this noise model worked well in homogeneous media with semi-infinite geometry.

To investigate the possibility of fitting μa, μs’ or β along with αDB from one single autocorrelation curve, we examined the SSD change patterns by varying three pairs of the four parameters. According to the results, a large crosstalk between the αDB and μa or αDB and μs’ (Fig. 3a and Fig. 3b) existed, suggesting that it is impractical to simultaneously extract αDB and μa or αDB and μs’ from one single autocorrelation curve. Conversely, the SSD curves for the pairs of β and αDB were convergent, suggesting a possibility to extract β and αDB simultaneously. These results were then verified by simultaneously fitting the paired parameters from the simulated autocorrelation curves with noise generated by the noise model. Fitting αDB and μa or αDB and μs’ simultaneously caused large estimation errors (Fig. 4a and Fig. 4c) that were majorly due to the large crosstalk between the paired parameters. These simulation results (Fig. 4a and Fig. 4c) agree very well with our previous findings in phantom titration tests (see Fig. 6 in Ref. [15]). By contrast, when fitting β and αDB simultaneously, the estimation errors for both parameters were much smaller (Fig. 4b and Fig. 4d), although they were increased with the increase of noise level.

Upon examination of the K(τ) definition [see Eq. (3)], the crosstalk between αDB and μa or αDB and μs’ is expected as these paired parameters can compensate each other to generate a similar autocorrelation curve. The decay of an autocorrelation curve is determined by K(τ) which can be rewritten as: K(τ)=(3μaμs+6μs2k02αDBτ)12=[3μaμs(1+2μsk02αDBμaτ)]12. It is apparent from this expression that the decay of an autocorrelation curve is influenced by the term of 2μ2k02αDBμaτ and a variation in αDB can be compensated by a variation in μa μs’. By contrast, based on Eq. (3) and Eq. (4), β does not affect the decay of the autocorrelation curve and does not compensate the variation of αDB. Therefore, it is possible to fit β and αDB simultaneously without causing the crosstalk between them.

We then compared the two methods for extracting β and αDB. All the results from the simulations (Fig. 5), phantom experiments (Fig. 6a), and tissue measurements (Fig. 6b) suggested that simultaneously fitting β and αDB from the entire autocorrelation curve resulted in more accurate values with smaller standard deviations compared to the two-step fitting method. For the two-step fitting method, large standard deviations of estimation resulted mainly from the inaccurate β estimated using only several points at early τ of g2 curve; limited datasets may be contaminated by noises. The estimation bias was due to the decay of autocorrelation curve with τ, which led to underestimations of β and αDB.

In conclusion, the possibility of extracting multiple parameters (αDB, μa, μs’ and β) via fitting one single autocorrelation function curve has not previously been investigated for DCS measurements. It is not trivial to get the answer regarding such possibility because the autocorrelation function depends on all four parameters [see Eq. (3) and Eq. (4)] and explicitly expressing the relations among them is difficult. In the present study, for the first time, we comprehensively investigated the possibility of fitting multiple parameters from one single autocorrelation curve and evaluated the performance of the two methods with computer simulations, tissue-like phantom experiments and in-vivo tissue measurements. The results from this study suggest that it is impractical to simultaneously fit αDB and μa or αDB and μs’ from one single autocorrelation function curve due to the large crosstalk between these paired parameters. However, simultaneous fitting of β and αDB is feasible and generates more accurate estimation with smaller standard deviation compared to the conventional two-step fitting method.

The outcomes from this study imply that absolute values of μa and μs’ are needed for extracting accurate β and αDB. Our laboratory has recently developed a hybrid NIR diffuse optical instrument combining a commercial frequency-domain tissue-oximeter and a DCS flowmeter, which allows for simultaneous measurements of μa and μs’ as well as β and αDB [21]. It is expected that the use of this type of hybrid instrument and simultaneous fitting algorithms will provide accurate measurement results.

Acknowledgments

This work was partially supported by the National Institutes of Health (NIH) under grants R01 CA149274 (GY), R21 AR062356 (GY), and UL1RR033173 (GY), and by the American Heart Association (AHA) Great Rivers Affiliate under the grant 11POST7360020 (YS).

REFERENCES

  • [1].Boas DA, Campbell LE, Yodh AG. Scattering and Imaging with Diffusing Temporal Field Correlations. Physical Review Letters. 1995 Aug 28;75:1855–1858. doi: 10.1103/PhysRevLett.75.1855. [DOI] [PubMed] [Google Scholar]
  • [2].Boas DA, Yodh AG. Spatially varying dynamical properties of turbid media probed with diffusing temporal light correlation. Journal of the Optical Society of America a-Optics Image Science and Vision. 1997 Jan;14:192–215. [Google Scholar]
  • [3].Maret G, Wolf PE. Multiple light scattering from disordered media. The effect of brownian motion of scatterers. Z. Phys. B. 1987;65:409–413. [Google Scholar]
  • [4].Pine DJ, Weitz DA, Chaikin PM, Herbolzheimer Diffusing-wave spectroscopy. Phys.Rev.Lett. 1988;60:1134–1137. doi: 10.1103/PhysRevLett.60.1134. [DOI] [PubMed] [Google Scholar]
  • [5].Stephen MJ. Temporal fluctuations in wave propagation in random media. Phys.Rev.B. 1988;37:1–5. doi: 10.1103/physrevb.37.1. [DOI] [PubMed] [Google Scholar]
  • [6].Roche-Labarbe N, Carp SA, Surova A, Patel M, Boas DA, Grant PE, Franceschini MA. Noninvasive optical measures of CBV, StO(2), CBF index, and rCMRO(2) in human premature neonates’ brains in the first six weeks of life. Hum Brain Mapp. 2010 Mar;31:341–52. doi: 10.1002/hbm.20868. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Yu G, Durduran T, Zhou C, Wang HW, Putt ME, Saunders HM, Sehgal CM, Glatstein E, Yodh AG, Busch TM. Noninvasive monitoring of murine tumor blood flow during and after photodynamic therapy provides early assessment of therapeutic efficacy. Clin Cancer Res. 2005 May 1;11:3543–52. doi: 10.1158/1078-0432.CCR-04-2582. [DOI] [PubMed] [Google Scholar]
  • [8].Shang Y, Chen L, Toborek M, Yu G. Diffuse optical monitoring of repeated cerebral ischemia in mice. Opt Express. 2011 Oct 10;19:20301–15. doi: 10.1364/OE.19.020301. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Kim MN, Durduran T, Frangos S, Edlow BL, Buckley EM, Moss HE, Zhou C, Yu G, Choe R, Maloney-Wilensky E, Wolf RL, Grady MS, Greenberg JH, Levine JM, Yodh AG, Detre JA, Kofke WA. Noninvasive measurement of cerebral blood flow and blood oxygenation using near-infrared and diffuse correlation spectroscopies in critically brain-injured adults. Neurocrit Care. 2010 Apr;12:173–80. doi: 10.1007/s12028-009-9305-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [10].Zhou C, Eucker SA, Durduran T, Yu G, Ralston J, Friess SH, Ichord RN, Margulies SS, Yodh AG. Diffuse optical monitoring of hemodynamic changes in piglet brain with closed head injury. J Biomed Opt. 2009 May-Jun;14:034015. doi: 10.1117/1.3146814. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].Yu G, Floyd T, Durduran T, Zhou C, Wang JJ, Detre JA, Yodh AG. Validation of diffuse correlation spectroscopy for muscle blood flow with concurrent arterial spin labeled perfusion MRI. Opt Express. 2007 Feb 5;15:1064–1075. doi: 10.1364/oe.15.001064. [DOI] [PubMed] [Google Scholar]
  • [12].Rice SO. Mathematical analysis of random noise. In: Wax N, editor. Noise and Stochastic Processes. Dover; New York: 1954. p. 133. [Google Scholar]
  • [13].Buckley EM, Cook NM, Durduran T, Kim MN, Zhou C, Choe R, Yu G, Schultz S, Sehgal CM, Licht DJ, Arger PH, Putt ME, Hurt HH, Yodh AG. Cerebral hemodynamics in preterm infants during positional intervention measured with diffuse correlation spectroscopy and transcranial Doppler ultrasound. Opt Express. 2009 Jul 20;17:12571–81. doi: 10.1364/oe.17.012571. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [14].Shang Y, Symons TB, Durduran T, Yodh AG, Yu G. Effects of muscle fiber motion on diffuse correlation spectroscopy blood flow measurements during exercise. Biomed Opt Express. 2010;1:500–511. doi: 10.1364/BOE.1.000500. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [15].Irwin D, Dong L, Shang Y, Cheng R, Kudrimoti M, Stevens SD, Yu G. Influences of tissue absorption and scattering on diffuse correlation spectroscopy blood flow measurements. Biomed Opt Express. 2011 Jul 1;2:1969–85. doi: 10.1364/BOE.2.001969. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Sunar U, Quon H, Durduran T, Zhang J, Du J, Zhou C, Yu G, Choe R, Kilger A, Lustig R, Loevner L, Nioka S, Chance B, Yodh AG. Noninvasive diffuse optical measurement of blood flow and blood oxygenation for monitoring radiation therapy in patients with head and neck tumors: a pilot study. J Biomed Opt. 2006 Nov-Dec;11:064021. doi: 10.1117/1.2397548. [DOI] [PubMed] [Google Scholar]
  • [17].Durduran T, Choe R, Yu G, Zhou C, Tchou JC, Czerniecki BJ, Yodh AG. Diffuse optical measurement of blood flow in breast tumors. Opt Lett. 2005 Nov 1;30:2915–7. doi: 10.1364/ol.30.002915. [DOI] [PubMed] [Google Scholar]
  • [18].Mesquita RC, Skuli N, Kim MN, Liang J, Schenkel S, Majmundar AJ, Simon MC, Yodh AG. Hemodynamic and metabolic diffuse optical monitoring in a mouse model of hindlimb ischemia. Biomed Opt Express. 2010;1:1173–1187. doi: 10.1364/BOE.1.001173. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Durduran T, Yu G, Burnett MG, Detre JA, Greenberg JH, Wang J, Zhou C, Yodh AG. Diffuse optical measurement of blood flow. Opt Lett. 2004 Aug 1;29:1766–8. doi: 10.1364/ol.29.001766. [DOI] [PubMed] [Google Scholar]
  • [20].Yu G, Durduran T, Lech G, Zhou C, Chance B, Mohler ER, 3rd, Yodh AG. Time-dependent blood flow and oxygenation in human skeletal muscles measured with noninvasive near-infrared diffuse optical spectroscopies. J Biomed Opt. 2005 Mar-Apr;10:024027. doi: 10.1117/1.1884603. [DOI] [PubMed] [Google Scholar]
  • [21].Shang Y, Zhao Y, Cheng R, Dong L, Irwin D, Yu G. Portable optical tissue flow oximeter based on diffuse correlation spectroscopy. Opt Lett. 2009 Nov 15;34:3556–8. doi: 10.1364/OL.34.003556. [DOI] [PubMed] [Google Scholar]
  • [22].Shang Y, Cheng R, Dong L, Ryan SJ, Saha SP, Yu G. Cerebral monitoring during carotid endarterectomy using near-infrared diffuse optical spectroscopies and electroencephalogram. Phys. Med. Biol. 2011;56:3015–3032. doi: 10.1088/0031-9155/56/10/008. [DOI] [PubMed] [Google Scholar]
  • [23].Diop M, Verdecchia K, Lee TY, St Lawrence K. Calibration of diffuse correlation spectroscopy with a time-resolved near-infrared technique to yield absolute cerebral blood flow measurements. Biomed Opt Express. 2011 Jul 1;2:2068–81. doi: 10.1364/BOE.2.002068. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Carp SA, Dai GP, Boas DA, Franceschini MA, Kim YR. Validation of diffuse correlation spectroscopy measurements of rodent cerebral blood flow with simultaneous arterial spin labeling MRI; towards MRI-optical continuous cerebral metabolic monitoring. Biomed Opt Express. 2010;1:553–565. doi: 10.1364/BOE.1.000553. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [25].Yu G, Shang Y, Zhao Y, Cheng R, Dong L, Saha SP. Intraoperative evaluation of revascularization effect on ischemic muscle hemodynamics using near-infrared diffuse optical spectroscopies. J Biomed Opt. 2011 Feb;16:027004. doi: 10.1117/1.3533320. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [26].Dong L, Kudrimoti M, Cheng R, Shang Y, Johnson EL, Stevens SD, Shelton BJ, Yu G. Noninvasive diffuse optical monitoring of head and neck tumor blood flow and oxygenation during radiation delivery. Biomed Opt Express. 2012 Feb 1;3:259–72. doi: 10.1364/BOE.3.000259. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [27].Cheung C, Culver JP, Takahashi K, Greenberg JH, Yodh AG. In vivo cerebrovascular measurement combining diffuse near-infrared absorption and correlation spectroscopies. Physics in Medicine and Biology. 2001 Aug;46:2053–2065. doi: 10.1088/0031-9155/46/8/302. [DOI] [PubMed] [Google Scholar]
  • [28].Zhou C, Yu G, Furuya D, Greenberg JH, Yodh AG, Durduran T. Diffuse optical correlation tomography of cerebral blood flow during cortical spreading depression in rat brain. Optics Express. 2006;14:1125–1144. doi: 10.1364/oe.14.001125. [DOI] [PubMed] [Google Scholar]
  • [29].Menon C, Polin GM, Prabakaran I, Hsi A, Cheung C, Culver JP, Pingpank JF, Sehgal CS, Yodh AG, Buerk DG, Fraker DL. An integrated clinically relevant approach to measuring tumor oxygen status using VEGF-transfected human melanoma xenografts as a model. Cancer Research. 2003;63:7232–7240. [PubMed] [Google Scholar]
  • [30].Li J, Dietsche G, Iftime D, Skipetrov SE, Maret G, Elbert T, Rockstroh B, Gisler T. Noninvasive detection of functional brain activity with near-infrared diffusing-wave spectroscopy. Journal of Biomedical Optics. 2005 Jul-Aug;10:044002-1–12. doi: 10.1117/1.2007987. [DOI] [PubMed] [Google Scholar]
  • [31].Durduran T, Zhou C, Edlow BL, Yu G, Choe R, Kim MN, Cucchiara BL, Putt ME, Shah Q, Kasner SE, Greenberg JH, Yodh AG, Detre JA. Transcranial optical monitoring of cerebrovascular hemodynamics in acute stroke patients. Opt Express. 2009 Mar 2;17:3884–902. doi: 10.1364/oe.17.003884. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [32].Li J, Ninck M, Koban L, Elbert T, Kissler J, Gisler T. Transient functional blood flow change in the human brain measured noninvasively by diffusing-wave spectroscopy. Opt Lett. 2008 Oct 1;33:2233–5. doi: 10.1364/ol.33.002233. [DOI] [PubMed] [Google Scholar]
  • [33].Zhou C, Choe R, Shah N, Durduran T, Yu G, Durkin A, Hsiang D, Mehta R, Butler J, Cerussi A, Tromberg BJ, Yodh AG. Diffuse optical monitoring of blood flow and oxygenation in human breast cancer during early stages of neoadjuvant chemotherapy. J Biomed Opt. 2007 Sep-Oct;12:051903. doi: 10.1117/1.2798595. [DOI] [PubMed] [Google Scholar]
  • [34].Durduran T. Ph.D. Dissertation, Physics. University of Pennsylvania; 2004. Non-Invasive Measurements of Tissue Hemodynamics with Hybrid Diffuse Optical Methods. [Google Scholar]
  • [35].Boas D. Ph.D. Dissertation, Physics. University of Pennsylvania, Philadelphia; 1996. Diffuse Photon Probes of Structural and Dynamical Properties of Turbid Media: Theory and Biomedical Applications. [Google Scholar]
  • [36].Zhou C. Ph.D. Dissertation, Physics. University of Pennsylvania; Philadelphia: 2007. In-vivo Optical Imaging and Spectroscopy of Cerebral Hemodynamics. [Google Scholar]
  • [37].Dietsche G, Ninck M, Ortolf C, Li J, Jaillon F, Gisler T. Fiber-based multispeckle detection for time-resolved diffusing-wave spectroscopy: characterization and application to blood flow detection in deep tissue. Appl Opt. 2007 Dec 10;46:8506–14. doi: 10.1364/ao.46.008506. [DOI] [PubMed] [Google Scholar]
  • [38].Koppel DE. Statistical Accuracy in Fluorescence Correlation Spectroscopy. Physical Review A. 1974;10:1938–1945. [Google Scholar]

RESOURCES