Ultrasensitive and label-free molecular level detection enabled by light phase control in magnetoplasmonic nanoantennas

Nicolò Maccaferri 1, Keith Gregorczyk 1, Thales V A G de Oliveira 1, Mikko Kataja 2, Sebastiaan van Dijken 2, Zhaleh Pirzadeh 3, Alexandre Dmitriev 3, Johan Åkerman 4, Mato Knez 1,5, Paolo Vavassori 1,5,*
PMCID: PMC4340560  EMSID: EMS61555  PMID: 25639190

Abstract

Systems allowing label-free molecular detection are expected to have enormous impact on biochemical sciences. Research focuses on materials and technologies based on exploiting localized surface plasmon resonances in metallic nanostructures. The reason for this focused attention is their suitability for single molecule sensing, arising from intrinsically nanoscopic sensing volume, and the high sensitivity to the local environment. Here we propose an alternative route, which enables radically improved sensitivity compared torecently reported plasmon-based sensors. Such high sensitivity is achieved by exploiting the control of the phase of light in magnetoplasmonic nanoantennas. We demonstrate a manifold improvement of refractometric sensing figure-of-merit. Most remarkably, we show a raw surface sensitivity (i.e., without applying fitting procedures) of two orders of magnitude higher than the current values reported for nanoplasmonic sensors. Such sensitivity corresponds to a mass of ~0.8 ag per nanoantenna of polyamide-6.6 (n=1.51), which is representative for a large variety of polymers, peptides and proteins.

INTRODUCTION

The most prominent routes for high-sensitivity and label-free detection in a compact device setting presently rely on localized surface plasmon resonance (LPR) based technologies. The greatly enhanced electromagnetic fields at the surface of a resonant plasmonic nanostructure1,2 allow for probing minute changes in the surrounding environment. Due to the evanescent nature of the fields, the sensing volume of plasmonic nanostructures is only marginally larger than the structures themselves, making them ideal probes for localized changes in a medium. For these reasons, the utilization of LPRs for label-free molecular detection is under very active investigation for biochemical and biomedical applications3-12. They feature optical interrogation schemes, small footprint, high sensitivity of refractometric detection, potentially down to single-molecule level due to the electromagnetic field enhancement at the nanoscale3,4, and easy integration with a wide range of fluidic systems for analyte delivery.

To quantify the sensing performance of a LPR-based sensor, the bulk refractive index sensitivity SRIλ*/Δn is often considered, where Δλ* is the shift of the LPR peak position λ* in nanometers measured in the extinction spectra over the change in the environment refractive index Δn. For detection at the molecular level, sensitivity to local variation of the index of refraction is most relevant. In this case sensor performances are quantified in terms of surface sensitivity SSurf = Δλ*/Δt at a given refractive index n, where t is the thickness of an assembled thin layer of the material with refractive index n being sensed on top of the active surface. Since the final accuracy of the peak tracking depends both on the magnitude of the peak shift and on the resonance line-width, the most crucial performance-defining parameter is the figure-of-merit (FoM), obtained by dividing either SRI or SSurf by the full width at half maximum (FWHM) of the resonance. Current LPR based sensors have a surface FoM lower or at most comparable to that of propagating surface plasmon resonances (SPRs) based sensors13-15, which are the core and reference systems for label-free optical detection. However, SPR based sensors are not an alternative for single-molecule level detection due to the lack of local sensitivity to the index of refraction. Therefore, tremendous research efforts aimed at improving the FoM of LPR based sensors have been conducted9,10,12,16-18.

Here we unveil a sensing modality that utilizes the unique optical properties of nanostructured magnetoplasmonic nanoantennas to combine and enhance all mentioned features of plasmonic sensing. Most importantly, we demonstrate that our approach delivers a manifold improvement of surface sensitivity and orders of magnitude higher surface FoM with respect to recent values reported for plasmonic-based detectors, including SPR-based sensors14,15,18. The proposed approach relies on magnetoplasmonic ferromagnetic (FM) nanoantennas deposited on a transparent substrate. FM nanoantennas are known to support LPRs19 and, once activated by an external magnetic field, they acquire an intrinsic magneto-optical (MO) activity. The key point is the selection of the FM material and nanoantennas’ design in order to produce exact phase compensation in the electric field components of the otherwise elliptically polarized transmitted light at a specific wavelength λε. Under this condition, a vanishing ellipticity ε (ε null-point, i.e. full linear polarization) is produced at λε. Our underlying strategy relies on the fact that light polarization changes can be measured precisely especially near null-conditions. The determination of λε provides a phase sensitive identification of the nanoantennas’ LPR position and enables tracking of the spectral shift caused by local refractive index variations with unprecedented precision.

RESULTS

Theoretical background

In ferromagnetic bulk materials and continuous films, the magneto-optical activity is governed by the spin-orbit interaction, which is a property intrinsic to a given material. Conversely, the magneto-optical response of a magnetic nanoantenna is governed also by localized surface plasmon resonances (LPRs)19-24. As depicted in Fig. 1, light interacting with a magnetic field-activated FM nanoantenna excites two coupled LPRs, one directly driven by the electric field of the incident light and the second induced orthogonally by the inherent MO activity. Desired light polarization behavior in transmission (and/or reflection) is achievable by tuning the relative phase of these two excited LPRs by designing the nanostructures. In detail, is the LPR polarized perpendicular to the driving electric field and light propagation direction that is governing the magneto-optical response of the nanoantenna. This fundamental physical effect can be visualized as follows; the incident electric field Ex0 induces an electric dipole along the x-axis described as px=αxxEx0, where αxx is the diagonal element of the polarizability tensor. The corresponding oscillation of the conduction electrons along the x-axis is spin-orbit coupled with the magnetization M within the nanoantenna, which is aligned along the z-axis by the applied magnetic field in the experimental geometry adopted here. This coupling induces an additional oscillation motion of conduction electrons, i.e., a second electric dipole py, along the in-plane transverse y-axis direction. In general, the expression for the spin-orbit induced transverse dipole can be written conveniently in terms of the off-diagonal polarizability tensor elements αij = εijαiiαjj/(ε0 - εm)2, where εij and ε0, are the off-diagonal and diagonal elements of the dielectric tensor of the given constituent material, and εm is the dielectric constant of the embedding medium. This expression shows that the off–diagonal elements of the polarizability tensor are proportional to the product between the diagonal components of the tensor along the two mixed directions. In the present case, the spin-orbit transversally induced dipole is given by px=αyxEx0=[εyxαyyαxx(ε0εm)2]Ex0 (considering the circular shape of our nanoantennas, αyy = αxx, i.e., the two LPRs resonate at the same wavelength, and the expression above can be simplified to px=αyxEx0=[εyxαxx2(ε0εm)2]Ex0).

Figure 1. LPR phase-sensitivity in the transmitted light polarization.

Figure 1

a. When an incident light beam hits a ferromagnetic nanoantenna, the conduction electrons inside the nanostructure oscillate driven by the electric field Ei. These Ei-driven oscillations can be modeled as a damped spring-mass harmonic oscillator. A LPR is induced at a specific photon wavelength λ*, yielding a peak in the extinction spectrum (I0It)/I0 = 1-(Et/Ei)2, displayed in the top panel. b. If the nanoantenna is magnetized perpendicularly to the surface plane, a MO-activity is turned on inducing a second MO-coupled LPR (MO-LPR) orthogonal to that directly driven by Ei. In a circular nanoantenna the MO-LPR resonates at the same λ*. The simultaneous excitation of LPR and MO-LPR induces an elliptical polarization ε of the transmitted field Et20,21. The null condition ε = 0 is generated at a desired λε (in general λελ*) simply through engineering of the size of the circular nanoantenna21,34,35. Measurement of λε provides a precise phase sensitive detection of the LPR position. The top panel displays typical Δε spectrum (red-line), as well as the 1/|Δε| spectrum (blue-line) and its resonance at λε. The close-up view of the 1/|Δε| spectrum around λε shown in the inset features a very narrow FWHM (1.7 nm). c. Similarly to the case described in (b), the concerted action of the simultaneously excitated LPR and MO-LPR can be exploited to actively manipulate the reflected light’s polarization inducing the condition ε = 0 at a desired λ′ε. In general λ′ελε since in this case also the additional phase introduced by the substrate reflectivity contributes to the polarization of the reflected field Er. As in transmission geometry, the detection of λ′ε provides precise phase sensitive detection of the LPR position. The top panel displays typical Δε spectrum (red-line), as well as the 1/|Δε| spectrum (blue-line) and its resonance at λ′ε. The close-up view of the 1/|Δε| spectrum around λ′ε shown in the inset features a very narrow FWHM (< 1.7 nm). Both in transmission and reflection, the sensitivity increases further by measuring the magnetic field- induced variation Δε as ε reverses its sign upon inverting H (see Supplementary Fig. 2).

Notably, the transversally induced electric dipole py is of second order in terms of polarizability as it depends on the product of two polarizabilites, αxx and αyy, while the directly induced dipole px depends only on αxx. This difference is crucial for the present application and can be understood considering that the transverse oscillation py is not driven directly by Ex0 but by the induced dipole px with the mediation of the spin-orbit coupling. The polarization state of the far-field radiated in the z-direction, either transmitted or reflected (Faraday or Kerr geometry), can be represented by the ratio of these two in-plane and mutually orthogonal oscillating electric dipoles, namely by the complex MO polarization angle Θ = py / px = εyx αxx/(ε0 - εm)2. This expression reveals that the polarization of the transmitted and reflected light is governed by both the intrinsic properties of the constituent material [αM = εyx /(ε0 - εm)2] and by the in-plane LPR in the nanoantenna (αyy, which coincides with αxx for circular nanoantennas). In detail, the transmitted and reflected light will have an elliptical polarization described through measurable rotation and ellipticity angles given by θ = Re(Θ) and ε =Im(Θ), respectively. Physically, both θ and ε at any light wavelength are determined by the relative phase ΔΦ of the two scattering electric dipoles px and py, since the complex MO angle Θ ∝ eiΔΦ, where ΔΦ =Φ [αM] + Φ [αyy]. For a given constituent material Φ [αM] is fixed. A null condition ε = 0 (ΔΦ = 0, π, 2π…) can be generated at a desired λε simply through engineering of the size of the circular nanoantenna, which controls both the spectral position and phase Φ [αyy] of the LPR21. As a result, measuring λε provides a precise and phase sensitive detection of any shift of the LPR position (λε and λ* are not bound to be identical) induced by modifications of the dielectric properties of the near-field region local environment.

Bulk refractive index sensitivity

The magnetoplasmonic nanostructures investigated here are bottom-up short-range ordered nickel (Ni) cylindrical nanoantennas on glass (Fig. 2a)25. For determining SRI, we measured the extinction spectra by immersing the samples in solutions with different indices of refraction n (Fig. 2b and Supplementary Fig. 1). The observed shift of λ* leads to SRI of 180 and 230 nm per refraction index unit (nm RIU−1) for 30 nm thick Ni nanoantennas with diameters of 100 and 160 nm, respectively (upper panels of Fig. 2b and Supplementary Fig. 1). In parallel, we determined the null-point wavelength λε by measuring the spectral dependence of the polarization ellipticity variation Δε of the transmitted light, induced by applying and reversing the MO activation field H (Methods and Supplementary Fig. 2). The wavelength λε is more precisely visualized by plotting the quantity 1/|Δε|, which resonates at λε, as shown in the top-right panel of Fig. 1. Plotting the data in this manner makes it much easier to visualize the enormous gain in precision (limit of detection) and sensitivity of our approach. In addition, and for the sake of direct comparison to the FoM, which is conventionally utilized for defining the performances of a plasmonic detector, the plot of 1/|Δε| provides the most sensible way to perform this comparison. We make it clear that comparison between sensitivity (both bulk and surface) and limit of detection performances of our approach with respect to plasmonic based sensors are instead derived from direct measurements of Δε spectra and the precision (noise level) with which λε is determined from such measurements.

Figure 2. Refractive index sensitivity of Ni magnetoplasmonic nanoantennas.

Figure 2

a 3D AFM profile of Ni magnetoplasmonic nanoantennas on glass, with lateral dimensions of 103 ± 5 nm (diameter) and 30 ± 0.5 nm (thickness). b (Top panel) extinction spectra of Ni cylindrical nanoantennas for different values of the embedding refractive index (clean n=1, water n=1.33, 50% Vol. glycerol n=1.41, and glycerol n=1.47; the inset shows a zoom of the resonance peaks in the spectral region 450-600 nm); (bottom panel) plot of the inverse of transmitted light ellipticity variation 1/|Δε| for the same values of the embedding refractive index as above. c Comparison between the bulk sensitivities of Au26λ*/Δn) (red-dashed line) and Ni (both Δλ*/Δn and Δλεn) (blue and green markers, respectively) cylindrical nanoantennas on glass. d Comparison between the bulk figure-of-merit of Au26 [(Δλ*/Δn)/FWHM] (purple dashed line) and Ni [(Δλεn)/FWHM] (blue markers are experimental data, blue dotted line is a guide for eyes) cylindrical nanoantennas, and Au surface plasmon resonance14 (red dashed line), in the spectral range 420-750 nm.

The spectra of 1/|Δε| for the two studied samples and for different values of n are shown in the lower panels of Fig. 2b and Supplementary Fig. 1. We observe that λε undergoes a shift equal to that of λ*, confirming the SRI values determined above. It is clear that monitoring λε instead of the extinction peak λ* enables tracking of the resonance shift with an exceptionally higher precision.

In principle, our approach is characterized by a virtually unlimited value of the FoM since 1/|Δε| is diverging at the resonant wavelength λε. Practically, however, we can estimate a FWHM by accounting for experimental errors. Although polarization parameters can be determined with sub-micro-radiant resolution (down to 10 nrad), here we assume an experimental resolution of 5 micro-radians in the determination of Δε, which is easily achievable without the utilization of advanced tools and/or fitting procedures, and consequently truncate the 1/|Δε| spectra at 2×105 rad−1 (see the inset of the top-right panel of Fig. 1). We obtain a FWHM ranging from 1.5 to 1.8 nm for the resonance of the 1/|Δε| spectra (Fig. 2b and Supplementary Fig. 2). Taking the average FWHM of ~1.7 nm means a FoM of more than 100 RIU−1 and approaching 150 RIU−1 for Ni nanoantennas with diameters of 100 and 160 nm, respectively. Such large FoM values greatly exceed even those for SPR in the same spectral range, which are often considered as theoretical limit references10. Fig. 2c,d show a comparison of the sensitivities of Au and Ni cylindrical nanoantennas on glass as well as their FoM, including that of SPR based sensors.

The outstanding sensing performance of our magnetoplasmonic nanoantennas is better appreciated if compared directly with recently developed nanoplasmonic sensors. Improvements in FoM have been achieved by boosting the SRI and/or reducing the FWHM of the LPR. SRI enhancement has been realized by lifting metal nanostructures above substrates with dielectric pillars in order to expose more efficiently the surrounding environment to the LPR enhanced electric field26. More efforts have been devoted to a reduction of the FWHM of LPRs caused by inherent losses in metallic nanostructures. The most effective approach relies on resonant coupling of LPRs with modes that possess a smaller FWHM. Such modes could, for example, be SPRs in an optically coupled thin metallic continuous layer16, diffractive coupling among the nanostructures arranged in periodic arrays15, and coherent coupling of different localized plasmon modes in nanostructures with complex shapes9,17. FoM values towards ~60 have been reported upon resonant coupling of LPRs and SPRs16. More recently, a FoM value of up to 108 has been reported from periodic arrays of gold “nano-mushrooms” that combine the two aforementioned approaches10. It is worth noting that all pathways towards enhancement of the FoM rely on a substantial increase in the complexity of the plasmonic nanostructures and arrangements design. In addition, the performance boost applies to a specific and narrow wavelength range where resonant coupling occurs. By exploiting the built-in phase sensitivity of the individual magnetoplasmonic nanoantennas, we circumvent the limitation due to their inherent losses, while keeping the nanostructuring process extremely simple and maintaining the outstanding performance over a broad spectral range.

Surface sensitivity

Applications to molecular detection rely on high sensitivity and FoM to the local variation of the index of refraction n. A reliable and precise assessment of the detection performance is an experimental challenge, as control of Δn on the molecular level is needed. The method of choice is the controlled deposition of extremely thin and removable films using molecular layer deposition (MLD), an organic variant of atomic layer deposition27,28. The cyclic and self-terminating growth mechanisms of this method allow molecular scale control of the polymer film growth and surface-dependent nucleation characteristics can be used for area selective growth through matching of substrate and precursor chemistries. More importantly, MLD deposits polymer films from vaporized pure molecular fragments avoiding potentially negative impact of solvents (see also Methods).

For our experiments we used MLD of polyamide 6.6 (PA-6.6) with n=1.5128. This polymer was chosen for convenience as the MLD process is well established and also, being a polyamide, it is representative for a large variety of polymers, peptides, and proteins. In a first experiment, PA-6.6 was deposited 20 cycles at a time. The film thickness was measured by AFM. The corresponding change of the local index of refraction was monitored by tracking λε. Since the surface was modified only locally, we conducted polarization measurements also in reflection geometry as an alternative way to track λε (see Fig. 1c). The experimental spectra and the resulting shift of λε are shown in Fig. 3a,b. AFM images and their quantitative analysis show that PA-6.6 nucleated selectively on the Ni nanodisks and linearly grew up to ~100 cycles (Fig. 3c,d). Combining the two plots in Fig. 3b,d, we find SSurf values of ~3 and ~5.3 in transmission and reflection geometry, respectively. The remarkable enhancement by a factor 1.7 of the SSurf in reflection geometry, confirmed also by calculations, is consequence of a larger shift of λε caused by the additional phase contribution of uncovered substrate reflectivity (details in Supplementary Fig. 3). Such SSurf value of ~5.3 is 3.5 times the surface sensitivity reported for Au based SPR detectors in the same spectral region for the same refractive index13. Even more remarkable is the surface FoM (SSurf /FWHM) of our magnetoplasmonic nanoantennas approach, whose value slightly larger than 3 is two orders of magnitude higher than the best surface FoM values achieved with plasmonic systems14.

Figure 3. Surface sensitivity assessment combining polarimetry and AFM measurements.

Figure 3

a Plot of the inverse of the transmitted (left panel) and reflected (right panel) light ellipticity 1/|Δε| spectra as a function of molecular layer deposition (MLD) cycles (at steps of 20 cycles). b 1/|Δε| resonance wavelength λε as a function of the MLD cycles for the two measurements geometries. In both cases, a linear dependence of λε vs. number of MLD cycles is observed (the black dashed lines are guide for eyes). The shift of λε saturates for a number of MLD cycles equal to 120, corresponding to a PA-6.6 thickness of ~35 nm, as shown in Supplementary Fig. 7 for transmission geometry case. Such PA-6.6 thickness agrees well with the near-field spatial extension (see Supplementary Fig. 4). c AFM images taken from the same sample region (total area imaged 2.7 × 2.7 μm2) after different numbers of MLD cycles. The AFM images show that PA-6.6 grows only on top of the nanoantennas. The colors of the frames refer to the corresponding colored polarimetry (b) and thickness (d) data points. The length of the white scale bars in the images corresponds to 1 μm. d PA-6.6 average thickness as function of the MLD cycles after AFM topography image analysis. Surface sensitivities (spectral variation of λε divided by the average nylon thickness) of ~3 (transmission) and ~5.3 (reflection) are found combining plots (b) and (d). The error bars indicate the standard deviation from the average thicknesses measured analyzing the AFM images.

In order to demonstrate that very small quantities of material can be detected, PA-6.6 deposition, polarization ellipticity measurements and AFM characterization were performed sequentially for the very first cycles of MLD. Fig. 4 shows that already after 2 MLD cycles an average increase in t of ~1.6 nm is determined by AFM. The resulting Δλε for the two measurement geometries confirm the surface sensitivities obtained in the previous experiment. Based on our very conservative error bar estimate in the measurement of Δε, we can detect λε with ~0.5 nm precision without the application of any fitting procedure (raw limit of detection).

Figure 4. Surface sensitivity in the first few cycles of PA-6.6 MLD.

Figure 4

a Schematic of one cycle of the MLD process for PA-6.6. A substrate with –OH surface groups is exposed to a pulse of adipoyl chloride (AC). The AC reacts with these–OH groups creating the by-product HCl, which is purged away along with any unreacted AC. Next a pulse of 1.6-hexamethylenediamine (HD) is introduced to the reaction chamber and reacts with the available –Cl groups. Again the byproduct is HCl, which is purged away along with any un-reacted HD. This process is repeated until the desired thickness is achieved. Nominally the process has a growth rate of ~0.8 nm/cycle29. b AFM images taken from the same sample region (total area imaged 1.2 ×1.2 μm2) before and after PA-6.6 MLD. The colors of the frames refer to the corresponding colored thickness (c) and polarimetry (d) data points. c PA-6.6 average thickness as function of the MLD cycles after AFM topography image analysis. The error bars indicate the standard deviation from the average thicknesses shown in the inset, which shows the line profiles of all the disks included in the images in (b). The line profiles are taken along two orthogonal directions, which are shown as white dashed lines only in the AFM image of the clean sample in (b). d Plot of the inverse of transmitted and reflected light ellipticity λε as a function of MLD cycles (black dashed lines are guide for eyes). Surface sensitivities of ~3.1 (transmission) and ~5.4 (reflection) are found combining plots (c) and (d), in excellent agreement with the results presented in Fig. 3. The horizontal error bars indicate the standard deviation from the average thicknesses shown in the inset in (c). The vertical error bars indicate the experimental error in the magneto-optical measurements. The insets show the corresponding 1/|Δε| spectra for the two measurement geometries (reflection – top-left inset, and transmission – bottom-right inset).

Given the surface sensitivity (SSurf) values ~3 and ~5.3 of our approach (transmission and reflection geometry, respectively), detection of sub-nanometer thick PA-6.6 coverage, viz., a discontinuous monolayer (ML) of PA-6.6 on individual cylindrical Ni nanoantennas is achievable (~1.7 Å and ~1.0 Å for the two measurement geometries). A ML of PA-6.6 is ~8.5Å thick29, the precision above corresponds, in the case of transmission geometry, to a coverage of less than 0.2 ML, namely a ML covering less than 20% of the nanoantenna exposed surface. The minimum detectable coverage reduces to 0.1 ML in reflection geometry. Such sub-ML coverage values are equivalent to PA-6.6 volumes of ~1200 and 800 nm3, i.e. a remarkable mass sensitivity of ~1.2 ag and 0.8 ag per disk corresponding to about 3300 and 2200 molecules of PA-6.6 per disk (density of amorphous PA-6.6 is 1.05 g/cm3 and its molecular weight is 226 g/mol). Such minimum detectable coverage is based on extrapolation and assumes that the electric near field is uniformly distributed in the vicinity of the Ni nanoantenna surface exposed to the environment, i.e., the surface sensitivity is not, or weakly, space-dependent. This condition is fulfilled to a good extent for our nanodisks with a diameter of 100 nm as we verified by the simulations of the near field produced by the excitation of a LPR performed using Lumerical (see Supplementary Fig. 4). A raw limit of detection of a few zg per nanoantenna can be achieved considering the sub-micro-radiant resolution of advanced polarimetry tools.

These sensing performances could be further improved by applying the same strategies as employed for noble metal nanoplasmonic systems for label-free detection, such as lifting of the nanostructures from the substrate and exploiting resonant coupling between LPRs to modes with narrow FWHM.

Also reducing the diameter of the cylindrical Ni nanoantennas would improve the limit of detection, since the surface sensitivity is weakly dependent on the diameter and, consequently, also the ability to detect a ML coverage of 20% and 10% of the nanoantenna exposed surface, which for smaller nanoantennas corresponds to proportionally smaller volumes of material.

So far, the limit of detection of our approach was derived based on the signal-to-noise ratio of the measurements, without any mathematical fit of the data, and compared with those of similar “raw” estimates of sensitivity based on absorption spectrum measurement. This comparison demonstrates the radically improved sensitivity enabled by our nano-magnetoplasmonic approach with respect to plasmon based sensors.

Higher sensitivity and limit of detection values are reported in literature for plasmon based sensors, which are achieved by application of fitting procedures30,31. We mention here that the application of fitting procedures of our data, confirm the higher sensitivity of our approach as shown in the Supplementary Fig. 5. Indeed, a mass sensitivity in the sub-zg per nanoantenna, down to a few yg per nanoantenna, can be achieved through the application of fitting procedures opening a pathway to mass sensitivity corresponding to ~10 molecules of PA-6.6 per disk (or, equivalently, of any material having n~1.5 and a density of ~1 gcm−3, which is the case of many polymers and bio-molecules).

DISCUSSION

Magnetoplasmonic nanoantennas that support magneto-optically induced localized plasmon resonances have been synthesized in order to induce a null condition of the transmitted/reflected light polarization ellipticity at desired wavelengths. The obtained null condition allows for an easily measurable and extremely precise phase sensitive detection of localized plasmon resonances. Such magnetoplasmonic nanoantennas can be used for optical sensing of local refractive index variations with enhanced sensitivity and unrivalled values for the figure-of-merit, even exceeding the theoretically predicted upper limit for sensing based on propagating surface plasmon resonances. Our approach requires extremely simple and scalable nanostructuring processes and offers a remarkably improved sensitivity performance in a large spectral range. To conclude we would like to emphasize that the ultrasensitive sensing capabilities of our magnetoplasmonic nanoantennas can be used also in nanoplasmonic biosensing (for instance cancer serum detection). Ni surfaces are covered by an ultrathin layer of oxidized Ni. Such surfaces can be functionalized with either silanes or even better with Histidin tags (His6)32 on par with Au surface functionalization with thiolate chemistry, thus informing our selected Ni as the material for our nanoantennas. As an alternative, one could even deposit a thin layer of Au, which will not affect the plasmonic behavior (even improve it, according to literature), and use conventional thiolate chemistry. In addition to biosensing, there are also many other potential civil and/or military applications that do not require surface functionalization and would enormously benefit from our approach like: chemical sensing of toxic materials, explosives, and ultra-precise thickness monitoring applications.

METHODS

Optical and magneto-optical measurements

The extinction spectra (I0It)/I0, where I0 and It are the intensities of the incident and transmitted light, respectively, were taken in the wavelength range 420-900 nm. The intensity of the light passing through the substrate without nanostructures on top was taken as the reference I0 signal.

The wavelength dependence of the magneto-optically induced ellipticity change Δε of the transmitted or reflected light was measured using magneto-optical Faraday (incidence angle 0°) and Kerr (incidence angle 2.5°) effect spectrometers working in polar geometry in the wavelength range 420-950 nm. The incident light beam was linearly polarized with either p- or s-polarization. Δε was measured by switching the polarity of a magnetic field H = 4 kOe applied normal to the sample plane to activate the magneto-optical coupling in the nanoantennas’ constituent material. As shown schematically in Supplementary Fig. 2, in the polar-Kerr geometry Δε was measured at each wavelength with the transmitted beam passing through a photoelastic phase modulator and a polarizer before detection. A lock-in amplifier was used to filter the signal at the modulation frequency in order to retrieve Δε (ref. 33). Our experimental setup allows measuring polarization parameters, like ε and Δε, with sub-micro-radiant precision, although in our analysis of the measured spectra we assumed a 5 micro-radians error bar, for the sake of demonstrating that our approach guarantees a high sensitivity even using extremely simple measurement setups as that shown in the Supplementary Fig. 6. It is worth mentioning that Δε can be equivalently measured in reflection geometry with the light impinging on either magnetoplasmonic sample surface side (Fig. 1c). Measurement of Δε with the light beam impinging from the glass bottom surface can be particularly appealing for implementation of our approach in practical devices (Supplementary Fig. 6).

Sensing experiments

Bulk sensitivity experiments were performed in a microfluidic cell where different liquids, with different refractive index, were injected every time. We used liquids and mixtures whose refractive indexes are well known, viz. water (n=1.33), glycerol (n=1.47) and 50% glycerol volume in water (n=1.41).

For surface sensitivity assessment, molecular layer deposition of polyamide 6.6 (PA-6.6) was done in a commercially available Cambridge Nanotech/Ultratech Savannah system using alternating pulses of adipoyl chloride (AC) (Sigma Aldrich) and 1,6-hexamethylenediamine (HD) (Sigma Aldrich) with the following pulse/purge/pulse/purge parameters; 0.5s/30s/0.5s/30s. The AC and HD were heated to 70°C and 80°C respectively. The reaction chamber was maintained at 85°C throughout the reaction. Before deposition the samples were treated using 5 min cycles of acetone and isopropanol ultrasonication, and then exposed to Ar plasma at 90 W for 60 sec to eliminate residues of organic materials. Coating and cleaning procedures were repeated several times to ensure the reproducibility of the experiment (both polarization and thickness of PA-6.6 deposits). The thickness of the PA-6.6 deposits was monitored via AFM. The AFM measurements were done in the same region after each MLD cycles and MOKE measurement step, using an Agilent 5500 AFM microscope with a Si tip operated in tapping mode. The images were analyzed using the Gwyddion® software package in order to extract the average thickness values of PA-6.6 deposits. Both bulk and surface sensitivity experiments were repeated 5 times each, giving the same results within the experimental errors reported above.

Sample fabrication

Bottom-up hole mask colloidal lithography was used to pattern the Ni nanodisks25. The following process steps were applied in all presented examples. The substrates used are 10×10×1 mm3 pieces of microscope slide glass (VWR International). The glass substrates were first cleaned through 5 min cycles of acetone, isopropanol and water ultrasonication. A poly(methyl methacrylate), PMMA, film (2 wt % PMMA diluted in anisole, MW = 950000) was spin coated onto a clean surface and followed bysoft baking (170 °C, 10 min on a hot plate). Reactive oxygen plasma treatment (50 W, 5 s, 250 mTorr, Plasma Therm Batchtop RIE 95m) was applied in order to decrease the polymer film hydrophobicity and avoid spontaneous de-wetting of the surface during subsequent polyelectrolyte and particle deposition steps, which would introduce inhomogenities in the particle distribution. Providing a net charge to the PMMA surface by pipetting a solution containing a positively charged polyelectrolyte onto the film (polydiallyldimethylammonium (PDDA) MW 200000–350000, Sigma Aldrich, 0.2 wt % in Milli-Q water, Millipore), followed by careful rinsing with de-ionized water in order to remove excess PDDA and blow-drying with a N2 stream. Deposition of a water suspension containing negatively charged polystyrene particles (sulfate latex, Invitrogen, 0.2 wt % in Mili-Q water) and N2 drying in a similar fashion as described above, leaving the PMMA surface covered with uniformly distributed PS-spheres. Evaporation of an oxygen plasma resistant thin film of Au. Removing the PS-spheres using tape stripping (SWT-10 tape, Nitto Scandinavia AB), resulting in a mask with holes arranged in a pattern determined by the self-assembled colloidal particles. Transfer of the hole-mask pattern into the sacrificial layer via an oxygen plasma treatment (50 W, 250 mTorr, Plasma Therm Batchtop RIE 95m), which effectively removes all PMMA situated underneath the holes in the film, leaving the surface covered with a thin film mask supported on a perforated, undercut polymer film. Ni deposition is done by e-beam-assisted evaporation (AVACHVC600). Lift off was done using acetone at room temperature or 50 °C for 5–10 min. The filling factor (surface covered by Ni nanoantennas normalized to area of the sample surface) can be varied between 5% to 30% in a controlled manner. The filling factor of the samples utilized in our experiments can be estimated to be around 15%.

Supplementary Material

1

Acknowledgements

P.V., N. M., K. G., and M.Kn. acknowledge financial support from the Basque Government (Programs No. PI2012-47, PI2013-56) and the Spanish Ministry of Economy and Competitiveness (Projects No. MAT2012-36844 and MAT2012-38161). N. M. acknowledges support from the Doctoral Program of the Department of Education, Linguistic Policy, and Culture of the Basque Government (Grant No. PRE_2013_1_975). P.V. and N.M. also thank Paolo Biagioni for support on simulations using Lumerical FDTD Solutions Software. T. V. A. G. d. O. acknowledges support from the European Research Council (Grant No. 257654-SPINTROS). Z. P., A. D., and J. Å. acknowledge the Swedish Research Council (VR) and the Swedish Foundation for Strategic Research, SSF (Z. P. and A. D.: Framework program Functional Electro-magnetic Metamaterials, Project No. RMA08; J. Å.: Successful Research Leader Program). M. K. and S. v. D. acknowledge support from the National Doctoral Programme in Nanoscience and the Academy of Finland (Grant No. 263510).

Footnotes

Competing Financial Interests statement

The authors declare no competing financial interests.

References

  • [1].Sandtke M, Kuipers L. Slow guided surface plasmons at telecom frequencies. Nature Photon. 2007;1:573–576. [Google Scholar]
  • [2].Ferry VE, Sweatlock LA, Pacifici D, Atwater HA. Plasmonic nanostructure design for efficient light coupling into solar cells. Nano Lett. 2008;8:4391–4397. doi: 10.1021/nl8022548. [DOI] [PubMed] [Google Scholar]
  • [3].Cui Y, Wei QQ, Park HK, Lieber CM. Nanowire nanosensors for highly sensitive and selective detection of biological and chemical species. Science. 2001;293:1289–1292. doi: 10.1126/science.1062711. [DOI] [PubMed] [Google Scholar]
  • [4].Anker JN, et al. Biosensing with plasmonic nanosensors. Nature Mater. 2008;7:442–453. doi: 10.1038/nmat2162. [DOI] [PubMed] [Google Scholar]
  • [5].Stewart ME, et al. Nanostructured plasmonic sensors. Chem. Rev. 2008;108:494–521. doi: 10.1021/cr068126n. [DOI] [PubMed] [Google Scholar]
  • [6].Larsson EM, Syrenova S, Langhammer C. Nanoplasmonic sensing for nanomaterials science. Nanophotonics. 2012;1:249–266. [Google Scholar]
  • [7].Mayer KM, Hafner JH. Localized surface plasmon resonance sensors. Chem. Rev. 2011;111:3828–3857. doi: 10.1021/cr100313v. [DOI] [PubMed] [Google Scholar]
  • [8].Chung T, Lee SY, Song EY, Chun H, Lee B. Plasmonic nanostructures for nano-scale bio-sensing. Sensors. 2011;11:10907–10929. doi: 10.3390/s111110907. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Kravets VG, et al. Singular phase nano-optics in plasmonic metamaterials for label-free single-molecule detection. Nature Mater. 2013;12:304–309. doi: 10.1038/nmat3537. [DOI] [PubMed] [Google Scholar]
  • [10].Shen Y, et al. Plasmonic gold mashroom arrays with refractive index sensing figures of merit approaching the theoretical limit. Nature Comm. 2013;4:2381. doi: 10.1038/ncomms3381. [DOI] [PubMed] [Google Scholar]
  • [11].Aćimović SS, et al. LSPR chip for parallel, rapid, and sensitive detection of cancer markers in Serum. Nano Lett. 2014;14:2636–2641. doi: 10.1021/nl500574n. [DOI] [PubMed] [Google Scholar]
  • [12].Kabashin AV, et al. Plasmonic nanorod metamaterials for biosensing. Nature Mater. 2009;8:867–871. doi: 10.1038/nmat2546. [DOI] [PubMed] [Google Scholar]
  • [13].Svedendahl M, Chen S, Dmitriev A, Käll M. Refractometric sensing using propagating versus localized surface plasmons: a direct comparison. Nano Lett. 2009;9:4428–4433. doi: 10.1021/nl902721z. [DOI] [PubMed] [Google Scholar]
  • [14].Otte MA, et al. Identification of the optimal spectral region for plasmonic and nanoplasmonic sensing. ACS Nano. 2010;4:349–357. doi: 10.1021/nn901024e. [DOI] [PubMed] [Google Scholar]
  • [15].Offermans P, et al. Universal scaling of the figure of merit of plasmonic sensors. ACS Nano. 2011;5:5151–5157. doi: 10.1021/nn201227b. [DOI] [PubMed] [Google Scholar]
  • [16].Lodewijks K, et al. Tuning the Fano resonance between localized and propagating surface plasmon resonances for refractive index sensing applications. Plasmonics. 2013;8:1379–1385. [Google Scholar]
  • [17].Verellen N, et al. Plasmon line shaping using nanocrosses for high sensitivity localized surface Plasmon resonance sensing. Nano Lett. 2011;11:391–397. doi: 10.1021/nl102991v. [DOI] [PubMed] [Google Scholar]
  • [18].Lodewijks K, Van Roy W, Borghs G, Lagae L, Van Dorpe P. Boosting the Figure-Of-Merit of LSPR-based refractive index sensing by phase-sensitive measurements. Nano Lett. 2012;12:1655–1659. doi: 10.1021/nl300044a. [DOI] [PubMed] [Google Scholar]
  • [19].Chen J, et al. Plasmonic nickel nanoantennas. Small. 2011;7:2341–2347. doi: 10.1002/smll.201100640. [DOI] [PubMed] [Google Scholar]
  • [20].Bonanni V, et al. Designer magnetoplasmonics with nickel nanoferromagnets. Nano Lett. 2011;11:5333–5338. doi: 10.1021/nl2028443. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Maccaferri N, et al. Tuning the magneto-optical response of nanosize ferromagnetic Ni disks using the phase of localized plasmons. Phys. Rev. Lett. 2013;111:167401. doi: 10.1103/PhysRevLett.111.167401. [DOI] [PubMed] [Google Scholar]
  • [22].Belotelov VI, et al. Enhanced magneto-optical effects in magnetoplasmonic crystals. Nature Nanotech. 2011;6:370–376. doi: 10.1038/nnano.2011.54. [DOI] [PubMed] [Google Scholar]
  • [23].Armelles G, Cebollada A, García-Martín A, González MU. Magnetoplasmonics: combining magnetic and plasmonic functionalities. Adv. Opt. Mater. 2013;1:10–35. [Google Scholar]
  • [24].Lodewijks K, Maccaferri N, Pakizeh T, Dumas Randy K., Zubritskaya I, Åkerman J, Vavassori P, Dmitriev A. Magnetoplasmonic design rules for active magneto-optics. Nano Lett. 2014;14:7207–7214. doi: 10.1021/nl504166n. [DOI] [PubMed] [Google Scholar]
  • [25].Fredriksson H, et al. Hole-Mask Colloidal Lithography. Adv. Mater. 2007;19:4297–4302. [Google Scholar]
  • [26].Dmitriev A, et al. Enhanced nanoplasmonic optical sensors with reduced substrate effect. Nano Lett. 2008;8:3893–3898. doi: 10.1021/nl8023142. [DOI] [PubMed] [Google Scholar]
  • [27].Knez M, Nielsch K, Niinistö L. Synthesis and surface engineering of complex nanostructures by atomic layer deposition. Adv. Mater. 2007;19:3425–3438. [Google Scholar]
  • [28].George SM. Atomic layer deposition: An overview. Chem. Rev. 2010;110:111–131. doi: 10.1021/cr900056b. [DOI] [PubMed] [Google Scholar]
  • [29].Du Y, George SM. Molecular layer deposition of nylon 66 films examined using in situ FTIR spectroscopy. J. Phys. Chem. C. 2007;111:8509–8517. [Google Scholar]
  • [30].Dahlin AB, Tegenfeldt JO, Höök F. Improving the instrumental resolution of sensors based on localized surface plasmon resonance. Anal. Chem. 2006;78:4416–4423. doi: 10.1021/ac0601967. [DOI] [PubMed] [Google Scholar]
  • [31].Chen S, Svedendahl M, Käll M, Gunnarsson L, Dmitriev A. Ultrahigh sensitivity made simple: nanoplasmonic label-free biosensing with an extremely low limit-of-detection for bacterial and cancer diagnostics. Nanotechnology. 2009;20:434015. doi: 10.1088/0957-4484/20/43/434015. [DOI] [PubMed] [Google Scholar]
  • [32].Hochuli E, Bannwarth W, Döbeli H, Gentz R, Stüber D. Genetic Approach to Facilitate Purification of Recombinant Proteins with a Novel Metal Chelate Adsorbent. Nat. Biotechnol. 1988;6:1321–1325. [Google Scholar]
  • [33].Vavassori P. Polarization modulation technique for magneto-optical quantitative vector magnetometry. Appl. Phys. Lett. 2000;77:1605–1607. [Google Scholar]
  • [34].Maccaferri N, et al. Polarizability and magnetoplasmonic properties of magnetic general nanoellipsoids. Opt. Express. 2013;21:9875–9889. doi: 10.1364/OE.21.009875. [DOI] [PubMed] [Google Scholar]
  • [35].Maccaferri N, et al. Effects of a non-absorbing substrate on the magneto-optical Kerr response of plasmonic ferromagnetic nanodisks. Phys. Status Solidi A. 2014;211:1067–1075. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1

RESOURCES