Abstract
Purpose
High-resolution intravascular (IV) MRI is susceptible to degradation from physiological motion and requires high frame-rates for true endoscopy. Traditional cardiac-gating techniques compromise efficiency by reducing the effective scan rate. Here we test whether compressed sensing (CS) reconstruction and ungated motion-compensation employing projection shifting, could provide faster motion-suppressed, IVMRI.
Theory and Methods
CS reconstruction is developed for under-sampled Cartesian and radial imaging using a new IVMRI-specific cost function to effectively increase imaging speed. A new motion correction method is presented wherein individual IVMRI projections are shifted based on the IVMRI detector's intrinsic amplitude and phase properties. The methods are tested at 3T in fruit, human vessel specimens, and a rabbit aorta in vivo. Images are compared using Structural-Similarity and ‘Spokal-Variation’ indices.
Results
Although some residual artifacts persisted, CS acceleration and radial motion compensation strategies reduced motion artefact in vitro and in vivo, allowing effective accelerations of up to eightfold at 200-300μm resolution.
Conclusion
3T IVMRI detectors are well-suited to CS and motion correction strategies based on their intrinsic radially-sparse sensitivity profiles and high signal-to-noise ratios. While benefits of faster free-breathing high-resolution IVMRI and reduced motion sensitivity are realized, there are costs to spatial resolution, and some motion artifacts may persist.
Keywords: Intravascular MRI, Compressed Sensing, motion correction, Projection Reconstruction
Introduction
Unlike conventional magnetic resonance imaging (MRI) with external detector coils, intravascular (IV) MRI receives the signal using tiny internal detectors (1-6). At field strengths of 3T, these can afford high signal-to-noise ratios (SNR) that permit 80-300μm resolution imaging of vessel walls and associated pathology as fast as several frames per second (fps) (7,8). With suitably modified sensitivity profiles, transmit/receive IVMRI probes can also provide high-resolution images from the probe's point-of-view (9), analogous to optical endoscopy, IV Ultrasound, and Optical Coherence Tomography. Nevertheless, while excelling in soft-tissue contrast, 3T IV MRI still lags behind these other modalities in speed.
Factors contributing to IVMRI's slowness include the physics of spatially encoding the MRI data, and the delays required to accommodate spin relaxation. MRI uses linear magnetic field gradients that encode the object in the spatial frequency domain (k-space). The encoding scheme is ordinarily subject to the Nyquist criterion which requires that sufficient k-space be filled to avoid aliasing errors upon reconstruction. Satisfying Nyquist ordinarily contributes directly to the long acquisition times. MRI scan-times can be reduced using recent ‘compressed sensing’ (CS) reconstruction techniques (10,11) which permit image reconstruction with a limited subset of k-space. Successful CS implementation requires ‘sparse’ data with sufficient SNR, and has been used in several conventional MRI settings including neurological, cardiac and dynamic visualization applications (10-14). However, to date CS has not been applied to IVMRI whose intrinsically high local SNR, spatial sparsity and need for speed, make it an ideal application.
Slow scan speeds are endemic to ultra-high resolution MRI, and also render IVMRI susceptible to the effects of rigid and non-rigid physiological motion—blood flow and respiration, as well as motion during probe advancement. Conventional cardiac- and breath-gating techniques reduce motion-sensitivity, but typically restrict speed and scan-time, depending on the period or incidence of the corresponding motion. While ungated acquisitions could allow frame-rates limited only by the SNR (or spatial resolution), their implementation would require other strategies for motion compensation. In past, several strategies for correcting conventional projection-reconstruction (PR) images for motion have employed data consistency criteria based on the zeroth and/or higher-order moments of each image projection (15-17) or projections of internal loop detectors (18). These methods are easily affected by intensity variations and can fail in high-resolution IVMRI where differences of the order of a millimeter can translate to ten or more pixels. Motion correction has also been reported using tiny external (19) or internal coils (20,21) deployed as tracking markers for conventional imaging. However high-resolution IVMRI requires high-resolution tracking information, and the use of external micro-coils to sense motion may offer limited value when they are remote from the microscopic field-of-view (FOV) of an internal imaging probe. Adding internal tracking coils may also make the IVMRI probe larger and more complex, and a minimally-invasive procedure less so.
In the present work we first address the speed limitation by developing CS reconstruction techniques for under-sampled IVMRI. We use these to test effective frame-rate acceleration factors of up to eight-fold. Second, we present a motion-correction method for ungated PR IVMRI based on the intrinsic amplitude and phase properties of the IVMRI detector's sensitivity profile. Third, the motion-correction method is applied to radially under-sampled data sets to test its efficacy in generating faster motion-corrected IV images. The methods are investigated in fruit, human vessel specimens, and a rabbit aorta in vivo.
Theory
A. Accelerating image acquisition
In two-dimensional (2D) Cartesian MRI, data are acquired to fill a 2D rectilinear grid in k-space and the images are reconstructed by 2D Fourier Transformation (FT). On the other hand, the PR method fills k-space along radial spokes, and the FT of each radial spoke in k-space corresponds to a projection in image space through the object at an angle perpendicular to the direction of the spoke. This direct relationship permits image reconstruction in either domain, by regridding the radial k-space data onto a rectangular grid followed by 2D FT using a Non-Uniform fast FT (NUFFT) algorithm, or using filtered back-projection techniques in image space, as in Computed Tomography. The two methods are equivalent, so we limit the description to the former approach, implemented here with an open-source NUFFT algorithm (22).
For alias-free reconstruction, both Cartesian and PR acquisition schemes ordinarily require sufficient k-space samples to satisfy the Nyquist criterion. Under-sampling in Cartesian MRI results in ‘ghost’ artefacts that appear as replicates of the original image displaced in the under-sampled dimension. Under-sampling artefacts in PR acquisitions are manifest as hyper-intense radial lines or ‘streaking’ (23). CS reconstruction ameliorates these artefacts by tailoring the under-sampling scheme so that the artefacts appear ‘noise-like’, and uses an iterative reconstruction that incorporates penalties to suppress the noise in the final image. In Cartesian acquisitions, the central k-space is fully-sampled but outer k-space is randomly sampled (11). For PR, k-space is under-sampled with a uniform angular distribution (10). All unsampled k-space data points are zero-filled.
For CS, image reconstruction is set up as a constrained minimization problem, where a functional, Φ(x), is minimized:
[1] |
Here y is the measured k-space complex data from all (PR) spokes or (Cartesian) lines stacked to form a vector, x is the image vector to be estimated, Aevaluates the image's FT in under-sampled k-space, the ℓ2 vector norm is denoted , and there are n penalty functions R(x) with weighting factors λ to constrain the resultant image based on a priori knowledge.
Suitable penalty functions that minimize the coefficients of the underlying image in a sparse transform space such as the Wavelet Transform (WT) or spatial finite-differences (minimizing Total Variation, TV), have been described earlier (10,11). The WT is a multi-scale representation of the image with coarse- and fine-scale wavelet coefficients representing low- and high-resolution image components, respectively. The TV constraint assumes that the underlying image consists of areas with constant (or mildly varying) intensity. The penalty functions using these transforms are
[2] |
[3] |
where Ψ is the WT and the ℓ1 vector norm ‖z‖1 = Σi|zi|, Dx, DY denote the derivatives in X and Y direction, respectively.
Here, an additional penalty function was designed for PR-IVMRI based on the field profile of the probes. For a loopless antenna oriented parallel to the main field (z-axis), the trans-axial RF-field, B1,Φ in cylindrical co-ordinates, was approximated by that of a long conductor with current I (24):
[4] |
where ϕ is the azimuthal angle, r is the radial distance from the probe, and μ is the permeability of the medium. Even when the probe is not parallel to the z-axis, Bϕ ∝ 1/r still holds (because MRI is only sensitive to the transverse B1 component) causing hyper-intensity at the probe location (Figs. 1 a, b). Under-sampling leads to streaking artefacts with spokes emanating from the probe location (Fig. 1c). The streaking is manifest in variations of the azimuthal amplitude with the probe as centre. The new penalty function smears the variation between the spokes (thereby reducing ‘spokal variation’, SV) for a more uniform image. We divided the image pixels into concentric annular sectors, computed complex pixel-sums, determined differences between azimuthally adjacent sectors, and summed the absolute differences over all sectors. (Fig. 1d):
[5] |
[6] |
where x(u, v) denotes the pixels in the uth annulus and vth sector.
With the penalty functions of Eqs. [2], [3] and [6], the functional in Eq. [1] becomes
[7] |
This functional was minimized using iterative CS algorithms as noted below {IRGNTV (10,25), GRASP (26), SPARSE-MRI (11)}. These algorithms involved searching for minima using Gauss-Newton and/or conjugate gradient methods by sequentially varying each of the three λ terms in Eq. [7]. Optimum values of λ were selected by comparing sample datasets from the under-sampled and the fully-sampled datasets using: (a) the ratio of SV values (Eq. 6), SVratio = SVA/SVB, wherein a smaller ratio indicates less streaking in image A with respect to image B or; (b) the Structural Similarity index {SSIM (27)},
[8] |
where A and B are the images being compared, σAB is their cross-correlation, μ and σ are the mean and standard deviation of the pixel intensities (if A=B, SSIM = 1). Reconstruction artefacts were analyzed using the point-spread-function (PSF) of the under-sampled trajectory. The width at which the Cartesian PSF drops to 2/π (≈ 64%, the Rayleigh criterion for distinguishing objects), and the location/amplitude of the radial PSF side-lobe were used as metrics (28).
B. Motion Correction
In-plane motion
Since the IVMRI probe is embedded within the body, in-plane body motion could cause the location of the probe to vary within the imaging plane. Assuming only rigid-body motion, if the probe location could be determined in each projection of a PR-MRI acquisition, then shifting or aligning each projection at the probe location should ameliorate the effect of the motion. To detect the probe in each projection, it can be seen from Eq. [4] and Figs. 1(a, b) that: (i) the field strength falls rapidly with distance r from the probe; and (ii) the detection phase varies azimuthally around the probe and reverses direction at the probe in every projection. In addition, (iii) the probe itself is metallic and does not contribute any MRI signal. Thus, each projection in image-space has a volcano-like signal intensity maximum close to the probe, a dip or ‘crater’ in intensity at the probe location (Fig. 2a, top), and a phase reversal at the exact location of the probe (Fig. 2a, bottom). A combination of both amplitude and phase detection provided the most robust means of detecting the probe. The center-of-mass (COM) of each projection served as the starting point. Note that the COM often differs from the probe location due to skewed intensity profiles (Fig. 2a, red diamond). Given the known physical dimensions of the probe and the image resolution, the width in pixels of the crater was determined and then used as a priori knowledge in a peak-fitting algorithm to find the crater. The probe's location was then refined using the phase information. The zero-crossing point of the phase-reversal was found using a level detection algorithm and/or maximum slew-rate detection. Once the probe's location was detected, each projection was shifted to the center of the image FOV, and all projections were aligned based on the location of the probe (Figs. 2b, c). The image was then reconstructed with the probe at its center.
A further refinement to this method was implemented at high-resolution, where the shifting process was observed to be sensitive to intra-pixel jitter, which contributed to residual radial streaking. This involved comparing the detected probe location from one projection to its location in previously acquired projections, and determining the pixel-shift in image-space. If large positional deviations from the preceding, or an average of both preceding and succeeding projections were detected, a corresponding phase-shift in k-space was computed from the inverse FT of the pixel-shift (i.e. if then ). This ‘corrected’ phase-shift was then applied to the motion-corrupted projection in k-space, prior to reconstruction. In this way, only those few projections that were corrupted by motion were processed and subject to potential jitter introduced by the detection algorithm. Images reconstructed using this refinement do not necessarily have the probe at the center of the FOV, but because the probe was already located in preceding steps, a simple image translation was applied to shift it to the FOV center post-reconstruction as described previously (8).
Through-plane motion
Through-plane motion, especially from respiration, is a source of artefacts in vivo, in general. Conventionally, respiration artefacts are overcome by navigator echoes placed on the lung-diaphragm interface (29). In the absence of external navigators, we observed that when stacked, radial k-space projections exhibit a pattern of aberrant jumps (Fig. 2d) that correlate with breathing (every ∼2s for a rabbit, or approximately every 10th projection with repetition time TR = 200ms). These projections served as ‘internal navigators’ that signified motion-corrupted projections that could be discarded, as in traditional navigator-gating. Alternatively, because in PR-MRI successive projections are highly correlated, the aberrant projections were replaced by an average of the preceding and succeeding projections (Fig. 2e). This process corrected aberrant projections from other periodic or sporadic sources as well.
C. Under-sampled motion-corrected reconstruction
Since the motion-correction methods outlined above act on each projection, they can also be applied to radially under-sampled data sets, as illustrated in Fig. 3. The motion-corrected under-sampled data were iteratively reconstructed to produce an effectively faster motion-corrected image. These were compared both visually and using the SSIM (Eq. 8) in vitro, and using the SVratio in vivo where motion-free images were unavailable for comparison.
Methods
First, CS methods were applied to data acquired employing Cartesian encoding and a 5-turn transmit/receive loop IVMRI endoscope (8,9). The loop antenna was 2.3 mm in diameter at its widest, tuned with a 91pF micro-capacitor, and connected to an 0.8mm nitinol cable. The cable was connected to a single channel transmit/receive interface with a switchable PIN diode to decouple the receiver during conventional MRI, as described previously (9).
CS and motion-corrected radial PR-MRI were performed in a Philips 3T Achieva whole-body MRI scanner using internal loopless antennae (1) for signal reception and the scanner's body-coil for excitation (7). The loopless-antenna was either a 2.75mm outer-diameter (OD) 400mm-long semi-rigid copper coaxial cable with the inner conductor extended 42 mm to form the whip, or a 0.8mm OD biocompatible super-elastic nitinol coaxial cable with a 42mm whip suitable for IVMRI applications (7).
Data are presented from fruit, human vessel specimens in vitro, and rabbits in vivo. The first study was performed with the semi-rigid loopless antenna in an orange that was manually shaken (∼± 3mm in-plane, randomly) during MRI. For comparison, projection-shifting was compared with the first-order moment correction method implemented as described elsewhere (15). The in vitro blood vessel study was performed with the semi-rigid loopless antenna in human iliac artery specimens immersed in saline. To mimic physiologically relevant motion, the container was placed on the abdomen of a free-breathing volunteer during MRI: in-plane motion due to respiration was ∼± 4mm. In vivo studies approved by our Institutional Animal Care and Use Committee were performed on 3 healthy New Zealand white rabbits. The rabbits were sedated with intramuscular acepromazine (1 mg/kg) and ketamine (40 mg/kg), induced with intravenous sodium thiopental, and intubated to maintain an open airway. The loopless antenna was advanced into the descending aorta to the renal bifurcation via a femoral incision, as confirmed by contrast-enhanced X-ray C-arm CT prior to transfer to MRI. The loop antenna was inserted in the descending aorta via a surgical cut-down just inferior to the renal arteries. Probes were fixed in place with surgical ties.
Scout MRI was performed to locate the probe, prescribe the imaging plane, and to ‘volume shim’ the vicinity of the internal probe. IVMRI was performed with and without cardiac gating using radial and/or Cartesian gradient echoes (spoiled or balanced). For radial acquisitions, each readout was symmetric about the echo with alternating direction reversals (‘even/odd flyback’), and azimuthal angles were incremented sequentially to span the entire circle. All processing was performed off-line on ‘raw’ k-space data using scripts written in MATLAB® (Mathworks, Nattick USA). IV images were weighted by the receiver sensitivity using an r-1 image intensity filter centered on the probe location (Fig. 5a, b) (1,7), except where noted. Since the probe location was already determined by the motion correction algorithm, this step was relatively trivial.
All reconstruction was performed on a standard DELL XPS L502X laptop (Intel Core i7-2760QM 2.40GHz processor; 8 GB RAM; Windows 7 Professional 64-bit operating system). Freely available CS implementations {IRGNTV (25), GRASP (26), SPARSE-MRI (11)} were adapted for reconstruction. Even and odd k-space trajectories were handled separately for motion correction and recombined for reconstruction, except where noted. Motion correction was done using either projection shifting alone (for phantom and in vitro data), or with the averaging of aberrant k-space projections (in vivo data). For calculating SV in Eq. [5], the annular rings were confined to 1.5-3.5cm from the probe. The scan parameters, CS software and reconstruction penalties are summarized in Table 1. SSIM and SVratio indices are reported in the figure captions for the corresponding images.
Table 1.
Fig. | Image details |
---|---|
1b | 2D radial GRE; 200 spokes; 250μm in-plane resolution; TR/TE=15/6 ms. Tacq = 3s. |
1c | Data from 1(b) undersampled to 50 spokes. NUFFT reconstruction, Trec ∼ 100 ms. |
4a | Cardiac gated 3D Cartesian GRE; TR/TE=250/12 ms; in-plane resolution 80μm; Tacq = 3.1 min/5 contiguous slices. Center slice shown. |
4b | Data from 4(a) undersampled three-fold. CS reconstruction using SPARSE-MRI. λWT = 0.09; λTV = 0.01; λSV = 0. Trec = 2 mins. |
4c | Data from 4(b). Regular FFT reconstruction, Trec ∼ 10 ms. |
5a | Same as 1(b). |
5b | Data from 1(b) with r-1 intensity filter and zoomed. |
5c | Data from 1(c). CS reconstruction using IRGNTV. λWT = 0; λTV = 2, 0.2, 0.02, 0.05 … (reduced every iteration); λSV = 0. Trec = 4 mins. |
5d | Data from 1(c) with r-1 intensity filter and zoomed. |
5e | Acquisition parameters same as 1(b) but with motion. |
5f | Data from 5(e), motion-corrected using COM. NUFFT reconstruction, Trec ∼ 100 ms. |
5g | Data from 5(e), motion-corrected using projection shift. NUFFT reconstruction, Trec ∼ 100 ms. |
5h | Data from 5(g) undersampled to 50 spokes. CS reconstruction using IRGNTV, λ same as 5(c). Trec = 17mins. |
7a | 2D radial GRE; 0.3×0.3×2mm3 voxel; TR/TE=150/6 ms, 1000 spokes, 20mm FOV shown. |
7b | Acquisition parameters same as 7(a) above, but with motion. |
7c | Data from 7(b), motion-corrected. NUFFT reconstruction, Trec ∼ 100 ms. |
7d | Data from 7(c) undersampled to 250 spokes. CS reconstruction using IRGNTV, λ same as 5(c). Trec = 18mins. |
7e | Data from 7(d). NUFFT reconstruction, Trec ∼ 100 ms. |
8a | Ungated, 0.2×0.2×5mm3 voxel, 2D Cartesian Balanced GRE, FA 90°, TR/TE 700/16ms, 25mm FOV. Tacq = 1.5 mins. |
8b | Ungated, 0.2×0.2×5mm3 voxel, 200 spokes (odd) balanced radial 2D GRE, FA 90°, TR/TE = 700/6ms, 50mm FOV shown. Effective Tacq = 2.5 mins. |
8c | Data from 8(b), motion-corrected. NUFFT reconstruction, Trec ∼ 100 ms. |
8d | Cardiac gated, 0.2×0.2×5mm3 voxel, 400 spokes spoiled 2D radial GRE, FA 40°, TE 6ms, 40mm FOV. Tacq = 7 mins. |
8e | Data from 8(c) undersampled to 50 spokes. CS reconstruction using GRASP. λWT = 0; λTV = 0.1 ; λSV = 0.0005. Trec = 7 mins. |
8f | Data from 8(e). NUFFT reconstruction, Trec ∼ 100 ms. |
9a | Ungated, 0.2×0.2×5mm3 voxel, 400 spokes (even+odd) balanced radial 2D GRE, FA 90°, TR/TE = 700/6ms, 15mm FOV shown. Tacq = 5 mins. |
9b | Data from 9(a) undersampled to 100 spokes. CS reconstruction using IRGNTV, λ same as 5(c). Trec = 10 mins. |
9c | Data from 9(a), but using only odd spokes (200 total) and motion-corrected. NUFFT reconstruction, Trec ∼ 500 ms. |
9d | Data from 9(c) undersampled to 50 spokes. CS reconstruction using IRGNTV, λ same as 5(c). Trec = 25 mins. |
9e | Ungated, 0.3×0.3×5mm3 voxel, Cartesian Balanced 2D GRE, FA 90°, TR/TE 9/4ms, 15mm FOV. Tacq = 0.5 secs. |
9f | Data from 9(e) undersampled two-fold. CS reconstruction using SPARSE-MRI. λWT = 0.09; λTV = 0.01; λSV = 0. Trec = 3 mins. |
Results
The effect of CS reconstruction of three-fold undersampled in vivo Cartesian data acquired from a loop-coil IV MRI endoscope with conventional cardiac-gating is shown in Fig. 4b. Traditional NUFFT reconstruction using the same k-space data as in Fig. 4(b) shows under-sampling artefacts especially in the high-SNR regions surrounding the probe (Fig. 4c). The effective speed-up factor in Figs. 4(b, c) vs Fig. 4(a) is 3-fold; the effects on the PSF and resolution are discussed later.
Structures revealed in an IVMRI of a stationary orange acquired with a loopless antenna (Fig. 5b) are retained in a four-fold under-sampled radial CS reconstruction (Fig. 5c). Some blurring of the septa between the fruit segments is evident, but not the streaking artefacts seen in a conventional under-sampled NUFFT reconstruction (Fig. 5d). Motion of the shaken orange obliterated virtually all structure in the PR image (Fig. 5e). Motion-correction of the same data set as (e) using the COM algorithm (15,16) substantially restores gross structure but shows some streaking (Fig. 5f; arrow). The projection-shifting algorithm offers further improvement, as evident in the fruit's central void and septal regions (Fig. 5g) as compared to Fig. 5(f) as well as to the source data, Fig. 5(e). This is reflected in a higher SSIM (see caption). Application of the combined algorithm comprising motion-correction and 4-fold under-sampling (Fig. 5h), reveals morphology albeit with some deterioration in resolution of the septa, as compared to Fig. 5(g). Both these images offer a dramatic improvement over the original motion-corrupted source image data in Fig. 5(e).
The effect of the undersampling in Figs. 4 and 5 is illustrated by plots of the PSF in Fig. 6. The 3-fold Cartesian under-sampling in Fig. 4(b) results in a near doubling of the PSF main-lobe width, as compared to the fully-sampled case (∼2 pixels vs. 1 pixel, see Fig. 6a). For the 4-fold under-sampled radial images in Fig. 5(c), the PSF side-lobe is about four times larger and four times closer to the main lobe, as compared to the fully sampled case (see Fig. 6b).
The effect of physiological motion of an iliac vessel specimen in a saline tank placed on the abdomen of a free-breathing volunteer is debilitating, with most structure lost in this example (Fig. 7b vs. 7a). Motion correction of the corrupted source data in Fig. 7(b) removes most of the radial streaking to reveal the underlying vessel wall structure in Fig. 7c (vs. Fig. 7b). Remaining vertical artefacts are attributable to residual motion. Vessel structure is also largely retained following a four-fold under-sampled CS reconstruction of the same image, shown in Fig. 7(d). Furthermore, the CS image is less noisy than a NUFFT reconstruction of the undersampled and motion-corrected data shown in Fig. 7(e).
Without cardiac or any other gating, in vivo images acquired with the loopless antenna at 200μm resolution in the rabbit aorta can exhibit intense ghost and radial streak artefacts in Cartesian-and radially-encoded acquisitions respectively (Figs. 8a, b). Motion correction by projection-shifting and k-space averaging of the ungated radial data shows a decrease in radial streak artefact intensity in Fig. 8(c), as compared to Fig. 8(b). A cardiac-gated image acquired separately from approximately the same location but requiring a longer scan time, is shown in Fig. 8(d). CS reconstruction applied to 1/4th of the motion-corrected data, although somewhat patchy, retains overall structure (Fig. 8e), whereas conventional NUFFT reconstruction of the same under-sampled data is overwhelmed by streaking artefacts (Fig. 8f). Zooming in on the aorta, we see that radial compressed sensing with only 1/4th of the original data (Fig.9b) provides comparable rendition vs the original (Fig. 9a), with streaking artefacts reduced by the motion correction (Fig. 9c). CS reconstruction applied to 1/8th of the motion-corrected data, provides an effectively eight-fold faster motion-suppressed in vivo image (Fig. 9d). Ungated acquisition affords imaging speeds of up to 2 fps (Fig.9e) and CS reconstruction using 1/2 the data (effectively 4 fps) faithfully renders the aorta but with some loss of detail in low-SNR regions (Fig. 9f).
CS reconstruction times depend significantly on the image resolution, FOV and iterative reconstruction parameters, ranging from 2 to 25 minutes, whereas the corresponding FFT and NUFFT based reconstruction was of the order of milliseconds (Table 1). The SSIM of all the corrected and/or undersampled CS reconstructed images was ≥ 0.4 and SVratio of corrected images was ≤ 0.95 (see figure captions).
Discussion
This paper presented strategies for accelerating image acquisition and suppressing motion artefacts in IVMRI, and demonstrated results in fruit for testing motion effects, vessel specimens in vitro (mounted on a volunteer to simulate physiological motion), and in rabbit aorta in vivo. CS reconstruction has been widely studied for MRI applications in the past, and our implementation was based on freely available open-source CS implementations. We extended previously used penalties with a new IVMRI-specific constraint (SV). As far as we are aware, this is the first application of CS reconstruction to IVMRI applications (30). Additional smoothing functions may be incorporated (10) to avoid some of the CS associated ‘patchiness’. The TV and SV constraints within the CS reconstruction favor smooth areas, blurring some areas with low contrast-to-noise ratios, while better preserving sharper edge morphology. As in any under-sampling scheme, there is some SNR loss. A possible remedy is 7T IVMRI which potentially offers a quadratic SNR improvement with field strength (31). In quantifying reconstruction artefacts, the simple PSF analysis presented here closely approximates the more sophisticated analysis of nonlinear CS reconstruction reported elsewhere (28). CS reconstruction takes significantly longer than conventional FT reconstruction (minutes vs. milliseconds), so that further software and hardware optimization, for example employing graphics processing units (32), would likely be required for practical use. Finally, the benefits of CS reconstruction are likely to grow as the dimensionality of the data is increased from the present 2D case (11).
Motion correction strategies based on PR have been used in the past in conventional MRI settings (15-17). Moment-based methods can yield erroneous results when applied to IVMRI projections wherein intensity profiles are highly skewed towards the coil (Fig. 2a, Fig. 5f). Our new method used explicit probe localization based on the amplitude and phase profiles of IVMRI detectors. Moreover, in high-resolution IVMRI applications, there are dynamic changes in projections such that corrections based on a sampling of central k-space or assumptions of a constant spin-density (17) were ineffective in our studies. All of our radial acquisitions employed gradient echoes, which was the standard radial sequence provided by the scanner manufacturer. Use of radial spin-echo sequences could offer improved immunity to field-inhomogeneity (T2*) effects and suppress signals from flowing blood for vessel wall imaging. Experimentally compensating for view-dependent gradient timing errors (33) may reduce artefacts as well. On the other hand, the reader should also be advised that in vivo IVMRI can often exhibit little motion artefact (e.g., Fig 9e, and the cine rabbit data in Ref. 8), lest the focus here on the effects of motion and its correction unduly create the impression of an overwhelming problem.
We showed that motion correction using projection shifting and averaging of aberrant k-space projections was useful in correcting rigid-body motion in vitro, and was moderately effective in dealing with respiration and sporadic motion that occurred in relatively short periods (1-2 TRs) in vivo. When the motion lasts for a significant fraction of the imaging time, traditional cardiac and respiratory gating and/or repeat imaging may be necessary. All these correction methods have limitations in compensating for the following sources of motion. First, probe motion with respect to the body due to natural physiological motion could cause position-dependent amplitude or phase variations, although here, the tensile properties of the in vivo nitinol probe tended to hold it against the vessel wall. In any case, this motion should have affected the cardiac-gated images (Fig. 8d) which were relatively free of such artefacts. Second, object rotation can cause a mismatch between the nominal acquisition angle and the actual projection. To correct this in high-resolution images, the motion could potentially be modeled based on fast low-resolution acquisitions. Third, pulsatile blood flow causes the vessel to expand and contract during image acquisition, affecting the surrounding tissue to a lesser extent. We did explore a correction method wherein each projection was dilated/compressed to account for the radial tissue expansion/contraction, but found it difficult to reliably model the non-linear changes.
Our motion correction algorithm is applied at the TR of each projection prior to reconstruction, and is therefore amenable to real-time application to the data stream, including MRI endoscopy (9). Applying the algorithm in real-time would permit a sliding window reconstruction strategy wherein the latest projection replaces the oldest, resulting in an image-stream that updates at the TR rate. The intent is to make fast lower-resolution (200-300μm) images in real-time followed by slower high-resolution (≤100μm) images at sites where pathology is suspected (7,8) or therapy is to be delivered. When the azimuthal angle increment in PR is small, the cross-correlation between successive projections can be used for probe detection, but can introduce a delay between acquisition and motion-correction. This will be larger when non-sequential angle increments are employed. Using the amplitude and phase profiles alone, automated detection was achieved for phantom and in vitro data, but some manual processing was required for in vivo data when spurious phase wraps occurred. Transmit/receive loop probes as in MR endoscopy (9) do not exhibit phase variations so that probe detection would have to rely solely on the near-field radial sensitivity profile.
We conclude that 3T IVMRI detectors are amenable to CS reconstruction due to their sparsely-localized sensitivity profiles and intense, yet localized SNR. For radial IVMRI, projection-shifting ameliorates motion artifacts based on the probe's intrinsic radial sensitivity profile. While there is a cost to using CS both in terms of spatial resolution and reconstruction time, and some motion artifacts are likely to persist, these techniques should nevertheless offer tangible benefits for faster high-resolution IVMRI in vivo.
Acknowledgments
We'd like to thank Michael Lustig, Kai Tobias Block, Ergin Atalar and Emi Murano for helpful discussions and suggestions; and Laurie Pipitone, Juls Myers and Arcan Erturk for the in vivo and in vitro studies.
Grant support: NIH R01 EB007829.
References
- 1.Ocali O, Atalar E. Intravascular magnetic resonance imaging using a loopless catheter antenna. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 1997;37(1):112–118. doi: 10.1002/mrm.1910370116. [DOI] [PubMed] [Google Scholar]
- 2.Atalar E, Bottomley PA, Ocali O, Correia LC, Kelemen MD, Lima JA, Zerhouni EA. High resolution intravascular MRI and MRS by using a catheter receiver coil. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 1996;36(4):596–605. doi: 10.1002/mrm.1910360415. [DOI] [PubMed] [Google Scholar]
- 3.Martin AJ, McLoughlin RF, Chu KC, Barberi EA, Rutt BK. An expandable intravenous RF coil for arterial wall imaging. Journal of Magnetic Resonance Imaging. 1998;8(1):226–234. doi: 10.1002/jmri.1880080138. [DOI] [PubMed] [Google Scholar]
- 4.Quick HH, Ladd ME, Nanz D, Mikolajczyk KP, Debatin JF. Vascular stents as RF antennas for intravascular MR guidance and imaging. Magnetic resonance in medicine. 1999;42(4):738–745. doi: 10.1002/(sici)1522-2594(199910)42:4<738::aid-mrm16>3.0.co;2-y. [DOI] [PubMed] [Google Scholar]
- 5.Hillenbrand CM, Elgort DR, Wong EY, Reykowski A, Wacker FK, Lewin JS, Duerk JL. Active device tracking and high-resolution intravascular MRI using a novel catheter-based, opposed-solenoid phased array coil. Magnetic resonance in medicine. 2004;51(4):668–675. doi: 10.1002/mrm.20050. [DOI] [PubMed] [Google Scholar]
- 6.Lederman RJ, Guttman MA, Peters DC, Thompson RB, Sorger JM, Dick AJ, Raman VK, McVeigh ER. Catheter-Based Endomyocardial Injection With Real-Time Magnetic Resonance Imaging. Circulation. 2002;105(11):1282–1284. doi: 10.1161/01.CIR.0000012425.71261.FC. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.El-Sharkawy AM, Qian D, Bottomley PA. The performance of interventional loopless MRI antennae at higher magnetic field strengths. Medical physics. 2008;35(5):1995–2006. doi: 10.1118/1.2905027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Sathyanarayana S, Schar M, Kraitchman DL, Bottomley PA. Towards real-time intravascular endoscopic magnetic resonance imaging. JACCCardiovascular imaging. 2010;3(11):1158–1165. doi: 10.1016/j.jcmg.2010.08.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Sathyanarayana S, Bottomley PA. MRI endoscopy using intrinsically localized probes. Medical physics. 2009;36(3):908–919. doi: 10.1118/1.3077125. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Block KT, Uecker M, Frahm J. Undersampled radial MRI with multiple coils. Iterative image reconstruction using a total variation constraint. Magnetic Resonance in Medicine. 2007;57(6):1086–1098. doi: 10.1002/mrm.21236. [DOI] [PubMed] [Google Scholar]
- 11.Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magnetic Resonance in Medicine. 2007;58(6):1182–1195. doi: 10.1002/mrm.21391. [DOI] [PubMed] [Google Scholar]
- 12.Gamper U, Boesiger P, Kozerke S. Compressed sensing in dynamic MRI. Magnetic Resonance in Medicine. 2008;59(2):365–373. doi: 10.1002/mrm.21477. [DOI] [PubMed] [Google Scholar]
- 13.Usman M, Atkinson D, Odille F, Kolbitsch C, Vaillant G, Schaeffter T, Batchelor PG, Prieto C. Motion corrected compressed sensing for free-breathing dynamic cardiac MRI. Magnetic Resonance in Medicine. 2013;70(2):504–516. doi: 10.1002/mrm.24463. [DOI] [PubMed] [Google Scholar]
- 14.Schirra C, Weiss S, Krueger S, Pedersen S, Razavi R, Schaeffter T, Kozerke S. Toward true 3D visualization of active catheters using compressed sensing. Magnetic Resonance in Medicine. 2009;62(2):341–347. doi: 10.1002/mrm.22001. [DOI] [PubMed] [Google Scholar]
- 15.Glover GH, Noll DC. Consistent projection reconstruction (CPR) techniques for MRI. Magnetic Resonance in Medicine. 1993;29(3):345–351. doi: 10.1002/mrm.1910290310. [DOI] [PubMed] [Google Scholar]
- 16.Glover GH, Pauly JM. Projection reconstruction techniques for reduction of motion effects in MRI. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 1992;28(2):275–289. doi: 10.1002/mrm.1910280209. [DOI] [PubMed] [Google Scholar]
- 17.Shankaranarayanan A, Wendt M, Lewin JS, Duerk JL. Two-step navigatorless correction algorithm for radial k-space MRI acquisitions. Magnetic Resonance in Medicine. 2001;45(2):277–288. doi: 10.1002/1522-2594(200102)45:2<277::aid-mrm1037>3.0.co;2-1. [DOI] [PubMed] [Google Scholar]
- 18.Homagk AK, Umathum R, Korn M, Weber MA, Hallscheidt P, Semmler W, Bock M. An expandable catheter loop coil for intravascular MRI in larger blood vessels. Magnetic Resonance in Medicine. 2010;63(2):517–523. doi: 10.1002/mrm.22228. [DOI] [PubMed] [Google Scholar]
- 19.Ooi MB, Aksoy M, Maclaren J, Watkins RD, Bammer R. Prospective motion correction using inductively coupled wireless RF coils. Magnetic Resonance in Medicine. 2013;70(3):639–647. doi: 10.1002/mrm.24845. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Qin L, Schmidt EJ, Tse ZTH, Santos J, Hoge WS, Tempany-Afdhal C, Butts-Pauly K, Dumoulin CL. Prospective motion correction using tracking coils. Magnetic Resonance in Medicine. 2013;69(3):749–759. doi: 10.1002/mrm.24310. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Rasche V, Holz D, Köhler J, Proksa R, Röschmann P. Catheter tracking using continuous radial MRI. Magnetic resonance in medicine. 1997;37(6):963–968. doi: 10.1002/mrm.1910370623. [DOI] [PubMed] [Google Scholar]
- 22.Fessler JA, Sutton BP. Nonuniform fast Fourier transforms using min-max interpolation. Signal Processing, IEEE Transactions on. 2003;51(2):560–574. [Google Scholar]
- 23.Peters DC, Lederman RJ, Dick AJ, Raman VK, Guttman MA, Derbyshire JA, McVeigh ER. Undersampled projection reconstruction for active catheter imaging with adaptable temporal resolution and catheter-only views. Magnetic Resonance in Medicine. 2003;49(2):216–222. doi: 10.1002/mrm.10390. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Ramo S, Whinnery JR, Van Duzer T. Fields and Waves in Communication Electronics. John Wiley and Sons; 2003. p. 78. [Google Scholar]
- 25.Knoll F, Clason C, Bredies K, Uecker M, Stollberger R. Parallel imaging with nonlinear reconstruction using variational penalties. Magnetic Resonance in Medicine. 2012;67(1):34–41. doi: 10.1002/mrm.22964. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Feng L, Chandarana H, Xu J, Block T, Sodickson DK, Otazo R. k-t radial SPARSE-SENSE: Combination of compressed sensing and parallel imaging with golden angle radial sampling for highly accelerated volumetric dynamic MRI; 20th Annual ISMRM Meeting; Melbourne, Australia. 2012; 2012. p. 81. [Google Scholar]
- 27.Zhou W, Bovik AC, Sheikh HR, Simoncelli EP. Image quality assessment: from error visibility to structural similarity. Image Processing, IEEE Transactions on. 2004;13(4):600–612. doi: 10.1109/tip.2003.819861. [DOI] [PubMed] [Google Scholar]
- 28.Wech T, Stäb D, Budich JC, Fischer A, Tran-Gia J, Hahn D, Köstler H. Resolution evaluation of MR images reconstructed by iterative thresholding algorithms for compressed sensing. Medical Physics. 2012;39(7):4328–4338. doi: 10.1118/1.4728223. [DOI] [PubMed] [Google Scholar]
- 29.Ehman RL, Felmlee JP. Adaptive technique for high-definition MR imaging of moving structures. Radiology. 1989;173(1):255–263. doi: 10.1148/radiology.173.1.2781017. [DOI] [PubMed] [Google Scholar]
- 30.Hegde SS, Zhang Y, Bottomley PA. Accelerated, motion-corrected high-resolution intravascular MRI at 3T. Proceedings of the International Society of Magnetic Resonance in Medicine. 2013:473. [PMC free article] [PubMed] [Google Scholar]
- 31.Erturk MA, El-Sharkawy AMM, Moore J, Bottomley PA. 7 Tesla MRI with a transmit/receive loopless antenna and B1-insensitive selective excitation. Magnetic Resonance in Medicine. 2013 doi: 10.1002/mrm.24910. Early View. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Uecker M, Zhang S, Voit D, Karaus A, Merboldt KD, Frahm J. Real-time MRI at a resolution of 20 ms. NMR in biomedicine. 2010;23(8):986–994. doi: 10.1002/nbm.1585. [DOI] [PubMed] [Google Scholar]
- 33.Peters DC, Derbyshire JA, McVeigh ER. Centering the projection reconstruction trajectory: Reducing gradient delay errors. Magnetic Resonance in Medicine. 2003;50(1):1–6. doi: 10.1002/mrm.10501. [DOI] [PMC free article] [PubMed] [Google Scholar]