. 2008 Sep 1;2(3):130–136. doi: 10.1016/SASJ-2008-0008-NT

The Combined Use of a Posterior Dynamic Transpedicular Stabilization System and a Prosthetic Disc Nucleus Device in Treating Lumbar Degenerative Disc Disease With Disc Herniations

Mehdi Sasani a,, Ahmet Levent Aydin a, Tunc Oktenoglu a, Murat Cosar b, Yaprak Ataker c, Tuncay Kaner d, Ali Fahir Ozer a
PMCID: PMC4365650  PMID: 25802613

Abstract

Background

Prosthetic replacement of spinal discs is emerging as a treatment option for degenerative disc disease. Posterior dynamic transpedicular stabilization (PDTS) and prosthetic disc nucleus (PDN) devices have been used sporadically in spinal surgery.

Methods

This was a prospective study of 13 patients averaging 40.9 years of age with degenerative disc disease who underwent posterior placement of a PDN with a PDTS. The Oswestry low-back pain disability questionnaire and visual analog scale (VAS) for pain were used to assess patient outcomes at the 3rd, 6th, and 12th postoperative months.

Lumbar range of motion was evaluated using a bubble inclinometer preoperatively and at 12 months postoperatively. Radiological parameters including lumbar lordosis angle (LL), segmental lordosis angle (α), disc height at the operated level (DHo), and disc height of the adjacent level (DHu) were evaluated. A typical midline posterior approach for complete discectomy was followed by the simultaneous placement of the PDN with PDTS.

Results

Both the Oswestry and VAS scores showed significant improvement postoperatively (P < .05). There were no significant differences in LL, α, DHo, and DHu parameters. We observed complications in 3 patients including 2 patients who had the PDN device embedded into the adjacent corpus; 1 had massive endplate degeneration, and the other experienced interbody space infection. In 1 patient, the PDN device migrated to one side in the vertebral space.

Conclusion

The use of a PDN in combination with posterior dynamic instrumentation can help to restore the physiologic motion of the anterior and posterior column and could help to establish posterior dynamic instrumentation as an important treatment of degenerative disc disease. Theoretically this concept is superior, but practically we need more advanced technology to replace disc material. Because this study examined the combination of the PDN and stabilization instrumention, the results cannot be compared with those reported in the literature for either PDN alone or dynamic screws alone.

Level of Evidence

Prospective cohort study with good follow-up (level 1b).

Keywords: Dynamic instrumentation, prosthetic disc nucleus, degenerative disc disease, non fusion treatment, lumbar disc herniation

INTRODUCTION

Prosthetic replacement of spinal discs is becoming a treatment option for degenerative disc disease, led by the success of hip and knee replacement. The prosthetic disc nucleus (PDN) device has been in clinical use in spine disc arthroplasty since 1996. This device is composed of a hydrogel core that can absorb fluid and expand when implanted and a polyethylene jacket.1 Similar to disc prostheses, the ideas behind the design of dynamic systems are to prevent devicerelated bone loss by allowing limited motion and to provide physiological loading by maintaining the natural posture in the lumbar spine.2 A dynamic pedicular screw hinge has monoaxial mobility capacity and provides controlled load sharing between implant and bone, reducing the frequency of postoperative damage to the joint segments.3

We evaluated 13 patients with primary (6 patients) and recurrent (7 patients) degenerative lumbar disease who received a PDN combined with posterior dynamic transpedicular stabilization (PDTS). The basic concept of this procedure was to create simultaneous anterior and posterior column stability to control the movement that occurs in the anterior and posterior column using PDTS, while at the same time supporting the height of the anterior column with the PDN.

MATERIALS AND METHODS

Patient Population

Thirteen patients, averaging 40.9 years of age, presented with 6 primary cases and 7 recurrent lumbar disc herniations. They underwent the simultaneous application of a PDN and PDTS (Table 1). Patients’ outcomes were followed for 12 postoperative months using the visual analog scale (VAS) for pain, which ranges from 0 (no pain) to 10 (worst pain imaginable) and the Oswestry Disability Index, a questionnaire comprised of 10 items designed to assess daily living activities that are most likely to be impaired in patients with low-back pain; a high Oswestry Disability Index percentage indicates high disability.4, 5, 6

Table 1.

Patients’ Demographic Data

Patient No. Gender Age Preoperative Neurological Findings Radiologic Findings (X-Ray, CT, MRI) Operation New/Recurrent Disc Herniation Complication Outcome (12 months)
1 F 26 − Right radiating pain
− Right SLR: 30º +
− Right S1 nerve root DH
− L4-5: black disc
− L5-S1: central and right disc extrusion
− Right L5-S1 discectomy + PDN placement
− L4-L5-S1 PDTS
New disc None Improved

2 F 40 − Left radiating pain
− Left SLR: 10º +
− Left gastrocnemius 2/5 MW
− L5-S1 disc sequestrated
− Both endplate hyperintense
− Left L5-S1 discectomy + PDN placement
− L5-S1 PDTS
New disc None Improved

3 M 57 − Right radiating pain
− Right SLR: 45º +
− L5-S1 right disc extrusion,
− Right L5 hemilaminotomy defect
− Right L5-S1 discectomy+ PDN placement
− L5-S1 PDTS
Recurrent disc − Discitis,
− PDN device embedded into the L5 corpus
− Treatment: antibiotic administration, PDN was removed
− LBP to 3 months
− Late improved

4 M 34 − Right radiating pain
− Right SLR: 45º +
− L4-5: bulging and annulus fibrosis rupture
− L5-S1: Right disc protrusion
− Right L5-S1 discectomy + PDN placement
− L4-L5-S1 PDTS
New disc None Improved

5 M 38 − LBP
− Right radiating pain
− L5-S1 right disc protrusion
− Both endplates hyperintense
− Right L5 hemilaminotomy defect
− Right L5-S1 discectomy + PDN placement
− L5-S1 PDTS
Recurrent disc None Improved

6 M 42 − LBP, left radiating pain
− Left SLR: 30º +
− Left L5 nerve root DH
− L4-5: left protrusion
− Both endplates hyperintense
− Left L4 hemilaminotomy defect
− Left L4-L5 discectomy + PDN placement
− L4-L5 PDTS
Recurrent disc None Improved

7 M 44 − Left radiating pain
− Left SLR: 45º +
− Left L5 nerve root DH
− L4-5 left protrusion − Left L4-L5 microdiscectomy +PDN placement
− L4-L5 PDTS
New disc None Improved

8 M 45 − Right radiating pain
− Right SLR: 30º +
− L3-L4: annulus fibrosis rupture
− L4-L5: right L4 hemilaminotomy defect+ right disc extrusion
− Right L4-L5 discectomy + PDN placement
− L3-L4-L5 PDTS
Recurrent disc None Improved

9 F 40 − Left leg radiating pain
− Left SLR: 45º +
− L4-5 left disc extrusion.
− Left L5 hemilaminotomy defect
− Left L4-L5 discectomy + PDN placement
− L4-L5 PDTS
Recurrent disc None Improved

10 M 42 − Left leg radiating pain
− Left SLR: 45º +
− L4-5 left disc extrusion − Left L4-L5 discectomy +PDN placement
− L4-L5 PDTS
Recurrent disc PDN embedded into the L4 corpus and massive endplate degeneration in both endplates − PTR+ conservative treatment to 3 months
− LBP to 3 months
− Late improved

11 M 39 − Right leg radiating pain
− Right SLR: 45º +
− Right gastrocnemius 2/5 MW
− L5-S1 foraminal disc protrusion − Right L5-S1 discectomy +PDN placement
− L5-S1 PDTS
New disc None Improved

12 M 45 − Left leg radiating pain
− Left SLR: 45º +
− L4-5 left disc extrusion − Left L4-L5 discectomy +PDN placement
− L4-L5 PDTS
New disc None Improved

13 M 40 − LBP
− Right radiating pain
− L5-S1 right disc protrusion − Right L5-S1 discectomy +PDN placement
− L5-S1 PDTS
Recurrent disc PDN migrated to right side in the intervertebral space Improved

SLR: Straight Leg Rising

LBP: Lower back pain

DH: Dermatome hypoesthesia

MW: Muscle Weakness

PTR: Physical Treatment and Rehabilitation

PDN: Prosthetic Disc Nucleus

PDTS: Posterior Dynamic Transpedicular Stabilization

Lumbar range of motion was evaluated by the bubble inclinometer technique as described by Mayer et al. in 1984.7

Radiological evaluations prior to and after surgery consisted of anteroposterior (AP) and lateral X-rays, computed tomography (CT), and magnetic resonance (MR) studies. The heights of the operated disc (DHo) and superior disc (DHu), segmental lordotic angles (α), and lumbar lordosis angle (LL) were used to evaluate patients’ outcomes (Figure 1, Table 2).

Figure 1.

Figure 1

A schematic calculation: (A) A ratio of disc height and segmental lordosis angle (α angle) for the upper adjacent segment was calculated pre- and postoperatively. (B) Measurement of lordosis of the lumbar spine L1 to S1 was calculated.

Table 2.

Patients’ Outcomes and Radiological Evaluations Prior To and After Surgery

Patient No. 1 2 3 4 5 6 7 8 9 10 11 12 13
Preop VAS 6 10 5 7 7 8 8 7 8 7 8 7 8
ODI 24 82 38 72 64 56 72 56 76 60 60 72 62
LLº aº 27 4 52 12 37 12 47 12 32 11 38 8 54 10 73 12 51 7 76 34 43 8 31 3 55 12
α º 4 12 12 12 11 8 10 12 7 34 8 3 12
DHu 0.271 0.261 0.275 0.297 0.187 0.333 0.285 0.19 0.287 0.273 0.197 0.148 0.25
DHo 0.203 0.141 0.166 0.231 0.271 0.226 0.22 0.154 0.3 0.257 0.264 0.237 0.211

Early Postop(3rd Day) LLº 26 48 30 52 31 29 45 64 55 65 48 45 40
αº 1 7 7 10 8 6 8 4 13 12 12 1 11
DHu 0.308 0.279 0.333 0.4 0.297 0.256 0.244 0.257 0.329 0.324 0.223 0.25 0.27
DHo 0.3 0.227 0.198 0.297 0.324 0.288 0.255 0.155 0.38 0.261 0.269 0.297 0.244

3 Mos. Postop VAS 2 2 3 1 2 3 2 3 1 2 2 3 3
ODI 20 32 26 26 18 38 32 26 18 28 20 48 38
LLº 37 61 26 58 33 41 49 52 65 68 47 43 51
αº 4 26 10 12 8 8 12 2 14 12 18 4 8
DHu 0.25 0.314 0.275 0.273 0.283 0.303 0.297 0.24 0.317 0.296 0.192 0.227 0.217
DHo 0.267 0.189 0.169 0.319 0.291 0.352 0.237 0.2 0.295 0.12 0.185 0.228 0.267

6 Mos. Postop VAS 2 1 1 1 1 2 2 1 2 1 3 2 2
ODI 20 26 16 18 12 18 20 18 12 12 12 24 24
LLº 35 63 31 58 21 32 49 49 57 64 45 41 45
αº 4 26 9 14 7 7 9 2 12 14 12 4 11
DHu 0.189 0.319 0.223 0.267 0.229 0.295 0.231 0.242 0.296 0.257 0.186 0.251 0.241
DHo 0.291 0.192 0.177 0.32 0.314 0.276 0.261 0.271 0.311 0.2 0.191 0.215 0.261
12 Mos. Postop VAS 2 1 1 0 0 0 2 1 1 1 3 2 2
ODI 18 20 16 8 6 6 12 16 18 12 8 20 26
LLº 34 67 30 58 20 35 47 54 52 52 47 43 50
αº 3 28 10 11 10 9 9 2 11 12 11 3 18
DHu 0.155 0.337 0.217 0.291 0.13 0.292 0.211 0.185 0.281 0.247 0.192 0.148 0.257
DHo 0.32 0.202 0.167 0.317 0.312 0.268 0.257 0.196 0.297 0.298 0.188 0.197 0.125

VAS: Visual Analog Scale for Pain

ODI: Oswestry Disability Index

LL: Lumbar lordosis angle

α: Segmental lordosis angle

DHu: Disc height of the adjacent level

DHo: Disc height at the operated level

Surgical Technique

A typical midline dorsal approach for complete 1-level lumbar discectomy was followed by the simultaneous placement of the PDN with dynamic pedicles screw/rod placement under fluoroscopy (Figure 2). Each patient was operated on by the same experienced spine surgeon. During the operation, patients were under general anesthesia in a prone position and received antibiotic prophylaxis for at least 48 hours. A midline incision was performed, and the affected segments were exposed by subperiosteal muscle dissection. Under an operation microscope, ligamentum flavum excision, dura exposure, and a total nucleotomy were performed on each side. The facet joints were preserved during nucleotomy.

Figure 2.

Figure 2

(A) Prosthetic disc nucleus (PDN-Solo, Raymedica, LLC) devices. (B) Titanium alloy Cosmic (Ulrich Gmbh & Co. KG, Ulm, Germany) dynamic pedicle screw.

Following the discectomy, the prosthetic disc nucleus devices (PDN-Solo, Raymedica, LLC, Minneapolis, Minnesota) were placed under microscope, and their positions were controlled by fluoroscopy. Afterwards, under fluoroscopy, the titanium alloy pedicle screws (Cosmic, Ulrich Gmbh & Co. KG, Ulm, Germany) were inserted. The screws were then connected with titanium alloy rods under optimum compression. Then, a deep drainage catheter was inserted, and the fascia was carefully closed. The wound was closed and all patients were mobilized within the next day of the operation without a brace.

Statistical Methods

The preoperative versus postoperative clinical findings (VAS, Oswestry scale) were compared using the Wilcoxon test.

RESULTS

The averages of obtained data are summarized in Table 3. The average Oswestry (improved 46.78+ points) and VAS (improved 6.15+ points) scores demonstrated improvement over the first postoperative year (P < .05). The averages of DHo and DHu pre- and post-operatively showed an increase (0.02+ points) and decrease (0.03+ points), respectively (P > .05). The average LL and α angles showed decreases (2.08º and 0.62º, respectively) over the first postoperative year (P > .05).

Table 3.

Averages of Preoperative and Postoperative Data Points

VAS ODI LLº αº DHu DHo
Preoperative 7.38 61.08 47.38 11.15 0.25 0.22

Early Postoperative (3rd day) 6.82 ---- 44.46 7.69 0.29 0.27

3 Month Follow-up 2.23 28.46 48.53 10.61 0.26 0.23

6 Month Follow-up 1.61 17.84 45.38 10.07 0.24 0.25

12 Month Follow-up 1.23 14.30 45.30 10.53 0.22 0.24

VAS: Visual analog scale for pain

ODI: Oswestry Disability Index

LL: Lumbar lordosis angle

α: Segmental lordosis angle

DHu: Disc height of the adjacent level

DHo: Disc height at the operated level

The Oswestry score showed significant improvement from the preoperative exam to the third postoperative month (P < .05). This improvement was stable at the 6- and 12-month exams. The VAS showed significant pain reduction after the operation (P < .05), and improved during the following 6 and 12 months. The postoperative radiologic evaluation showed a central position for all PDN devices, except in 1 patient (patient #13), and proper position of the implanted pedicle screws (Figure 3).

Figure 3.

Figure 3

A postoperative lumbar CT scan showing the central position of the PDN device and proper position of the implanted pedicle screws.

The Oswestry and VAS scores showed significant improvements at 3, 6, and 12 months postoperatively. There were no significant differences between any of the 4 measured radiologic parameters within the first postoperative year.

Range of motion was measured in all patients. There were no significant differences between preoperative and 12-month postoperative range of motion values (P > .05) (Table 4).

Table 4.

Preoperative And Postoperative 12-month Range of Motion Values

Patient Preop Flexion 12-month Postop Flexion Preop Extension 12-month Postop Extension
1 48 53 15 18
2 43 47 13 15
3 35 38 11 15
4 62 64 16 23
5 45 50 11 15
6 51 53 15 17
7 45 49 14 18
8 56 58 18 24
9 41 47 10 13
10 45 50 16 22
11 53 58 14 20
12 52 55 16 19
13 45 49 17 19
Mean: 47.76923 51.61538 14.30769 18.30769

We observed postoperative complications in 3 patients. In 2 patients (patients #3 and #10) the PDN device embedded into the adjacent corpus. Patient #3 developed an interbody space infection, and patient #10 had massive endplate degeneration. In a third patient (#13) the PDN device migrated to one side in the intervertebral space (Figure 4).

Figure 4.

Figure 4

Postoperative complications: (A) A plain X-ray shows the device has migrated to lateral interspace vertebra (patient #13). (B) A T1-weighted sagittal MRI demonstrates that the PDN-Solo device has become embedded in the adjacent vertebra with massive degeneration at both endplate vertebral bones (patient #10). (C) A sagittal CT shows the PDN-Solo device has become embedded in the adjacent vertebra in a patient with discitis (patient #3).

None of the patients showed significant intervertebral space height reduction at the operated level (DHo), except in the cases of embedded PDN. In these 2 cases, severe inflammatory reactions occurred with loading disc height in adjacent vertebrae (Figure 4). We can accept these changes as early findings of Modic degeneration. The intervertebral spaces at the upper adjacent levels did not change significantly in any of the patients.

DISCUSSION

Spinal instability caused by degenerative changes in the disc with subsequent spondylolisthesis, ankylosis, and neuroforaminal stenosis may induce back pain.2 From among the causes of spinal instability, degenerative disc disease (with its characteristic clinical syndromes of disc herniation, spondylosis, and radiculopathy) is associated with vascular, biomechanical, and anatomic changes in the disc.

For many decades, stabilization, intended to fix the abnormal movement of the spine, has become the treatment of choice for instability. The goal of fusion was to provide motion stability in arthrodesis and deformity correction, thus controlling and even strengthening the movement of the segment.8 However, good fusion rates are not correlated with the clinical relief of the patient's back pain symptoms. The best results measured by different clinical scales show around 30% of suboptimal healing after instrumented fusion surgery despite a fusion rate of more than 90%.9 Many authors reported adjacent segment degeneration following lumbar and lumbosacral fusion with rigid instrumentation.10 Kumar et al. described degenerative changes in 49% of patients after 5 years of follow-up.11 Because of these problems, less rigid stabilization systems have recently become more popular in spine surgery.

The PDN device has been in clinical application since 1996. It was developed by Charles D. Ray. This device is composed of a hydrogel core that can absorb fluid and expand when implanted.1 The hydrogel core is covered with a polyethylene jacket. The core and the jacket are designed to assume the cushioning function of a healthy disc, while restoring and maintaining disc height and allowing a normal range of motion. The high-tenacity polyethylene jacket allows the device to absorb fluid, expanding only in height. This expansion restores disc height to relieve pressure on the spinal nerves. Biomechanical studies using a human cadaver model showed that PDN can effectively increase the heights of the intervertebral discs.12 However, clinical observations showed us that in some cases the PDN, when used alone, slips to one side of the midline in the disc space.

Another promising system is the dynamic pedicular screwrod system. It is a non-fusion dynamic implant system that controls displacement in rotation and translation. It stabilizes while avoiding fusion. This system allows potential sagittal mobility at the hinge site between the screw head and the shank of the screw. Mobility occurs mechanically between the longitudinally oriented rod and the sagittally placed screw shank. This articulated connection between the rod and the screw is the reason for the lower rate of implant failure, because of a reduction in flexion strain. Part of the load sharing is transferred from the implant to the spine, ie, there is a reduction of stress-shielding effects on the bone.13 This reduced stiffness helps distribute load and minimize or prevent a motion discontinuity at the adjacent segment. Sharing of motion and stress load prevents deterioration of the neighboring superior disc level, slowing the process of degenerative progression at the adjacent levels, and tends to restrict mobility in flexion, extension, lateral bending, and axial rotation while sharing load and permitting movement.14 Also, the capability of a controlled load sharing between the implant and spine could reduce the frequency of postoperative damage to the joint segments.

The indication for PDN application is painful degeneration of the lumbar spinal disc, which has become sufficiently disabling despite 6 months of conservative medical care, rest, exercise, change in lifestyle, etc. The contraindications for PDN device use are spondylolisthesis greater than grade I, fractured or symptomatic degenerated facet joints, an incompetent annulus, or disc height at the affected level of < 5 mm. These criteria are all based largely on the history, examination, MRI, and, perhaps, provocative discography.1 Also, the vertebral endplates must be free of significant defects such as Schmorl's nodules or fractures before PDN application.15 In our patient group, these features were taken into consideration for PDN application.

Dynamic stabilization using a hinged screw allows motion of disc and facet joints because of the hinged line just posterior to the facet joints. In contrast, posterior dynamic stabilization transfers the load through the posterior column and stabilizes the spine almost like a rigid stabilization technique, preserving motion.3 Preservation of motion makes this technique useful for conditions where PDN use might be contraindicated, such as degenerative spondylolisthesis. As a result, the use of these 2 instruments together minimizes their disadvantages and restores both anterior and posterior columns.

In this limited series, we observed slippage of PDN in 2 cases. However, slippage of PDN with Modic degeneration of adjacent vertebrae is consistent with findings in the literature related to the use of a standalone PDN. On the other hand, in 2 other cases we noticed severe inflammation of adjacent vertebrae due to fracture of upper and lower endplates. In these cases, the PDN was embedded into the upper and lower vertebrae. Surprisingly, there were no clinical symptomatic or neurologic findings in these cases.

We assumed that large size of the PDN device due to the excessively stiff high-tenacity polyethylene jacket enclosing the polymeric hydrogel caused the fracture of endplates, which then made contact with cartilage tissue and blood, provoking an immune response, and resulting in massive degeneration of the vertebral bone.16 On the other hand, the hydrolysis time of 3 weeks was too long; therefore a rigid material between the 2 vertebral corpuses carries the risk of fracture of the cartilaginous endplates.

Dynamic stabilization systems without fusion and PDN devices have been used sporadically in spinal surgery. Ideally, semi-rigid implants dedicated to dynamic stabilization should control movement and share the load with the disc. Constraining the movement of the dynamic screw is more for providing adequate stability and mobility of the segment, and has less to do with avoiding the hot spot pressure on the disc.

The reasons for using the dynamic screw system in combination with PDN are to restore and preserve the disc height with the PDN device, while providing dynamic transpedicular stabilization with the screw system.

Although the VAS and Oswestry scores were satisfactory in our limited series of patients, for those with migrated PDN and infection, we thought that we could eliminate the migration complication with the use of posterior dynamic stabilization. However, we could not. The surprising result was that we saw cartilaginous endplate fracture due to the PDN and associated infection findings on MR studies, but the clinical outcome was not negatively affected.

PDN alone is not soft enough to resemble disc tissue. More appropriate material resembling disc tissue should be developed, and a more dynamic posterior system will need to be found. Theoretically, this concept is superior, but practically we need more advanced technology.

Because this study examined the combination of the PDN and stabilization instrumention, the results cannot be compared with those reported in the literature for either PDN alone or dynamic screws alone.

REFERENCES

  • 1.Jin D, Qu D, Zhao L, Chen J, Jiang J. Prosthetic disc nucleus (PDN) replacement for lumbar disc herniation: preliminary report with six months’ follow-up. J Spinal Disord Tech. 2003;16(4):331–337. doi: 10.1097/00024720-200308000-00004. [DOI] [PubMed] [Google Scholar]
  • 2.Inceoglu S. Posterior Dynamic Stabilization of the Lumbar Spine. World Spine J. 2006;1(2):62–67. [Google Scholar]
  • 3.Bozkus H, Senoglu M, Ozer AF, Sonntag VK, Crawford NR. Comparative stabilization properties of rigid and hinged-dynamic pedicle screw fixation techniques; (Unpublished data results) [Google Scholar]
  • 4.Jensen MP, McFarland CA. Increasing the reliability and validity of pain intensity measurement in chronic pain patients. Pain. 1993;55(2):195–203. doi: 10.1016/0304-3959(93)90148-I. [DOI] [PubMed] [Google Scholar]
  • 5.Baker D, Pynsent P, Fairbank J. The Oswestry Disability Index revisited. In: Roland M, Jenner J, editors. Back Pain: New Approaches to Rehabilitation and Education. Manchester, UK: Manchester University Press; 1989. pp. 174–186. [Google Scholar]
  • 6.Fairbank JCT, Pynsent PB. The Oswestry Disability Index. Spine. 2000;25(22):2940–2952. doi: 10.1097/00007632-200011150-00017. [DOI] [PubMed] [Google Scholar]
  • 7.Mayer TG, Tencer AF, Kristoferson S, Mooney V. Use of noninvasive techniques for quantification of spinal range-of-motion in normal subjects and chronic low-back dysfunction patients. Spine. 1984;9(6):588–595. doi: 10.1097/00007632-198409000-00009. [DOI] [PubMed] [Google Scholar]
  • 8.McNally DS. The objectives for the mechanical evaluation of spinal instrumentation have changed. Eur Spine J. 2002;11(2 Suppl):S179–185. doi: 10.1007/s00586-002-0435-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Mulholland RC, Sengupta DK. Rationale, principles and experimental evaluation of the concept of stabilization. Eur Spine J. 2002;11(2 Suppl):S198–205. doi: 10.1007/s00586-002-0422-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Hilibrand AS, Robbins M. Adjacent segment degeneration and adjacent segment disease: the consequences of spinal fusion? Spine J. 2004;4(6 Suppl):190S–194. doi: 10.1016/j.spinee.2004.07.007. [DOI] [PubMed] [Google Scholar]
  • 11.Kumar MN, Baklanov A, Chopin D. Correlation between sagittal plane changes and adjacent segment degeneration following lumbar spine fusion. Eur Spine J. 2001;10(4):314–319. doi: 10.1007/s005860000239. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Eysel P, Rompe JD, Schoenmayr R, Zoellner J. Biomechanical behaviour of a prosthetic lumbar nucleus. Acta Neurochir (Wien) 1999;141(10):1083–1087. doi: 10.1007/s007010050486. [DOI] [PubMed] [Google Scholar]
  • 13.Strempel A, Neekritz A, Muelenaere P, et al. Dynamic versus rigid spinal implants. In: Gunzburg R, Szpalski M, editors. Lumbar Spinal Stenosis. Philadelphia: Lippincott-Williams and Wilkins; 2000. pp. 275–285. [Google Scholar]
  • 14.Putzier M, Schneider SV, Funk JF, Tohtz SW, Perka C. The surgical treatment of the lumbar disc prolapse: nucleotomy with additional transpedicular dynamic stabilization versus nucleotomy alone. Spine. 2005;30(5):E109–114. doi: 10.1097/01.brs.0000154630.79887.ef. [DOI] [PubMed] [Google Scholar]
  • 15.Ray CD. The PDN prosthetic disc-nucleus device. Eur Spine J. 2002;11(2 Suppl):S137–142. doi: 10.1007/s00586-002-0425-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Bogduk N. The innervation of intervertebral discs. In: Ghosh P, editor. The Biology of the Intervertebral Disc. Boca Raton, FL: CRC Press; 1988. pp. 135–149. [Google Scholar]

Articles from SAS Journal are provided here courtesy of International Society for the Advancement of Spine Surgery

RESOURCES