Abstract
Parkinson’s disease (PD) is a neurodegenerative disease characterized by progressive loss of dopaminergic neurons in the basal ganglia. Single photon emission computed tomography (SPECT) scans using [99mTc]TRODAT-1 can image dopamine transporters and provide valuable diagnostic information of PD. In this study, we optimized the scanning parameters for [99mTc]TRODAT-1/SPECT using the Taguchi analysis to improve image quality. SPECT scans were performed on forty-five healthy volunteers according to an L 9 orthogonal array. Three parameters were considered, including the injection activity, uptake duration, and acquisition time per projection. The signal-to-noise ratio (SNR) was calculated from the striatum/occipital activity ratio as an image quality index. Ten healthy subjects and fifteen PD patients were used to verify the optimal parameters. The estimated optimal parameters were 962 MBq for [99mTc]TRODAT-1 injection, 260 min for uptake duration, and 60 s/projection for data acquisition. The uptake duration and time per projection were the two dominant factors which had an F-value of 18.638 (38%) and 25.933 (53%), respectively. Strong cross interactions existed between the injection activity/uptake duration and injection activity/time per projection. Therefore, under the consideration of as low as reasonably achievable (ALARA) for radiation protection, we can decrease the injection activity to 740 MBq. The image quality remains almost the same for clinical applications.
Introduction
Parkinson’s disease (PD) is a neurodegenerative disease characterized by progressive loss of dopaminergic neurons in the basal ganglia. Direct measurements of dopamine transporters (DAT) can indicate the severity of neuronal degeneration. Single photon emission computed tomography (SPECT) scans using [99mTc]TRODAT-1 has been used to image DAT concentration and proven to be a valuable method to diagnose the early stage of PD [1] and other diseases [2–4]. The image quality of SPECT scans depends on the imaging parameters, which are determined empirically in the nuclear medicine department. Therefore, designing a dedicated scanning procedure for [99mTc]TRODAT-1/SPECT is essential to promote diagnostic accuracy of dopamine-related diseases.
Various parametric settings existed for [99mTc]TRODAT-1/SPECT scans. For the injection activity, 740 MBq (20 mCi) was given intravenously to predict the clinical response after treatment of the attention deficit hyperactivity disorder [5], while 925 MBq (25 mCi) was injected to make a differential diagnosis of vascular Parkinsonism and PD [6]. For the scanning time of SPECT, a total of 120 projections each with 20 s were acquired for imaging PD patients with social anxiety disorder [7], whereas 120 projections but each with 60 s were used for imaging patients with early-stage corticobasal degeneration [8]. For the uptake duration, 180 min was taken to investigate Japanese Encephalitis [9], whereas 240 min was selected to diagnose clinically unclear parkinsonian syndromes [10] and PD patients with depression [11].
The optimal imaging time for [99mTc]TRODAT-1/SPECT has been investigated [12]. The result recommends that 240 min post-injection for clinical routine can achieve the highest caudate/occipital and putamen/occipital ratios. However, the image quality is influenced by multiple parameters interactively. Finding an optimal combination of the parameters and elucidating the relationships among them require extended investigation. In this study, we investigated the imaging parameters for [99mTc]TRODAT-1/SPECT scans by the Taguchi analysis. Three parameters were considered, including the injection activity, uptake duration, and time per projection. The cross-interactions between these parameters were evaluated. The final goal was to develop a dedicated acquisition procedure for [99mTc]TRODAT-1/SPECT scans.
Materials and Methods
Taguchi analysis and robust design
The Taguchi method is a kind of fractional factorial testing, which allows us to simultaneously examine several key elements in one study. It has become a popular tool for designing a high-quality system and has been applied in various research fields [13–16]. With a properly designed orthogonal array and the use of F-test, the Taguchi method can evaluate the sensitivity and significance of different imaging parameters, providing a more efficient way to establish scanning procedures used in nuclear medicine departments. Recently, it has been used to optimize the image quality of Ga-67-citrate gamma camera scanning [17]. Detailed descriptions of the Taguchi method and robust design can be found in other literatures [18–20].
Orthogonal array
Three design parameters, including the injection activity, uptake duration, and acquisition time per projection, were considered with three various levels of each for [99mTc]TRODAT-1/SPECT scans. A total of 27 (3×3×3) different combinations are shown in Table 1. According to Taguchi’s recommendation [18], an L 9 orthogonal array was used to rearrange the levels of parameters (Table 2).
Table 1. Three design parameters including the injection activity, uptake duration, and time per projection each with three different levels.
| Symbol | Design parameter | Unit | Level 1 | Level 2 | Level 3 |
|---|---|---|---|---|---|
| A | Injection activity | MBq (mCi) | 740 (20) | 851 (23) | 962 (26) |
| B | Uptake duration | min | 180 | 220 | 260 |
| C | Time per projection | s/projection | 40 | 50 | 60 |
Table 2. Parametric layout in an L 9 orthogonal array of nine groups.
| Group | Level A | Level B | Level C |
|---|---|---|---|
| 1 | 1 | 1 | 1 |
| 2 | 1 | 2 | 2 |
| 3 | 1 | 3 | 3 |
| 4 | 2 | 1 | 2 |
| 5 | 2 | 2 | 3 |
| 6 | 2 | 3 | 1 |
| 7 | 3 | 1 | 3 |
| 8 | 3 | 2 | 1 |
| 9 | 3 | 3 | 2 |
The number in each column indicates various levels for the specific parameters.
Human subjects
A total of forty-five healthy volunteers (age of 24–57 years old and mean age of 41.3±12.0 years), who were reviewed by a neurologist to exclude those with psychiatric diagnoses/medications, neurological diseases, and a history of movement disorders [21], were recruited. All protocols were approved by the Institutional Review Board of Show Chwan Memorial Hospital for clinical research, and all volunteers participated in this study gave written informed consent.
SPECT scans
Nine groups were formed in the orthogonal array (Table 2). Each group contained five randomly assigned subjects; thus, a total of 45 normal participants were included, and the corresponding level of imaging parameters was applied according to Table 1. A dual head gamma camera (Sophia DXT-XLi, GE/SMV, Versailles, France) was used and a total of 60 projection angles were acquired over 180° for each detector head. The image was reconstructed into a 128 × 128 matrix with a pixel size of 2.11 × 2.11 mm2 using the filtered back-projection (FBP) algorithm with a Metz filter (order of 3.5 and cutoff of 7 cm−1). The slice thickness was 2.9 mm. Subsequently, photon attenuation correction was performed using the Chang’s first-order method with a broad-beam attenuation coefficient of 0.12 cm−1 [22, 23].
The reconstructed image with the highest signal in the basal ganglia was extracted and summed together with its two adjacent slices as a single composite image. Several regions of interest (ROIs) were drawn by a neuroradiologist in the striatum and occipital areas at each hemisphere on the composite image, based on reference of the corresponding T1-weighted MR images (Fig. 1) [24]. The striatum/occipital (S/O) activity ratio of the subject was taken as the specific to non-specific uptake ratio, which is expected to be the higher the better owing to accumulation of [99mTc]TRODAT-1 in the striatum. The signal-to-noise ratio (SNR) for each group of the orthogonal array was analyzed as follows:
| (1) |
where η is the SNR in dB, y i is the S/O activity ratio of the ith subject, n is the total number of subjects in each group (n = 5). A larger SNR is considered preferable. Namely, the optimal level of parameters is that with the highest η. The average SNRs for different levels of parameters were further calculated to evaluate the sensitivity of each parameter.
Fig 1. ROIs drawn for the striatum (upper) and occipital (lower) areas on the composite image.
Analysis of variance, ANOVA
ANOVA was used to provide a measurement of confidence and to determine which parameters were statistically significant. Variance was decomposed; the sum of squared deviation for a specific parameter (S x) and the sum of squared error (S E) were calculated as follows:
| (2) |
| (3) |
where is the mean S/O ratio of all subjects, is the mean S/O ratio for the kth level of parameter x, n is the number of groups (n = 9), r is the number of trials in each group (r = 5), L is the number of levels for each parameter (L = 3), and S i is the standard deviation for the ith group. Subsequently, the F-test [25] was performed as follows:
| (4) |
where F x is the F-value of the specific parameter x, f x is the degree of freedom, and S x/f x is the variance. The F-test is an auxiliary tool to examine whether the factors are dominant in the system. If the confidence level of the F-value is greater than 99%, the imaging parameter is considered statistically significant. Additionally, the larger the F-value, the more dominant the parameter.
The isopreference curve, defined as the points on the curve with the same image quality, was further analyzed by fixing the level of one parameter and changing the level of the other two parameters. The corresponding SNR was normalized by the average SNR of the level of the fixed parameter to eliminate its effect. The relationship between parameters can be revealed.
Verification
For further verification, SPECT scans were performed on ten additional healthy subjects (age of 45–64 years old with average age of 52.9 years) and fifteen PD patients (age of 45–76 years old with average age of 60.7 years) using the predicted optimal parametric setting found by the Taguchi analysis and the conventional setting previously used in our hospital. For the conventional setting, the injection activity of 851 MBq, uptake duration of 220 min, and scanning time of 50 s/projection were applied. The average S/O ratios of healthy subjects and PD patients were estimated and compared for statistically significant using the unpaired t-test.
Results
Interpretation of SNR
The levels of parameters used for the L 9 orthogonal array are summarized in Table 3. Table 4 lists the estimated S/O ratios for each subject and the corresponding SNRs for each group (see also S1 Table). Group 1 had the lowest SNR of 6.57 dB among all the groups. When the levels of uptake duration and time per projection rose from 1 to 3 (group 3), the SNR increased 53% to a maximum of 10.08 dB. This finding suggests that the uptake duration and scanning time per projection have a positive impact on the image quality.
Table 3. Nine study groups in the L 9 orthogonal array with three different levels of parameters according to the Taguchi’s recommendation.
| Group | Injection activity (MBq) | Uptake duration (min) | Time per projection(s/projection) |
|---|---|---|---|
| 1 | 740 | 180 | 40 |
| 2 | 740 | 220 | 50 |
| 3 | 740 | 260 | 60 |
| 4 | 851 | 180 | 50 |
| 5 | 851 | 220 | 60 |
| 6 | 851 | 260 | 40 |
| 7 | 962 | 180 | 60 |
| 8 | 962 | 220 | 40 |
| 9 | 962 | 260 | 50 |
Table 4. Estimated S/O ratios for the normal subjects and corresponding SNRs for the nine groups.
| Group | y 1 | y 2 | y 3 | y 4 | y 5 | Average S/O ratio | SNR, η (dB) |
|---|---|---|---|---|---|---|---|
| 1 | 2.51 | 2.12 | 2.21 | 2.01 | 1.93 | 2.16 ± 0.22 | 6.57 |
| 2 | 2.41 | 2.47 | 2.36 | 2.25 | 2.47 | 2.39 ± 0.09 | 7.56 |
| 3 | 3.02 | 3.16 | 3.22 | 3.42 | 3.18 | 3.20 ± 0.14 | 10.08 |
| 4 | 2.46 | 2.53 | 2.63 | 2.57 | 2.48 | 2.53 ± 0.07 | 8.07 |
| 5 | 2.86 | 2.67 | 2.77 | 2.89 | 2.98 | 2.83 ± 0.12 | 9.03 |
| 6 | 2.32 | 2.74 | 3.23 | 2.91 | 2.89 | 2.82 ± 0.33 | 8.84 |
| 7 | 2.86 | 2.87 | 2.74 | 2.97 | 3.65 | 3.02 ± 0.36 | 9.47 |
| 8 | 2.34 | 2.65 | 2.63 | 2.29 | 2.45 | 2.47 ± 0.16 | 7.82 |
| 9 | 2.87 | 2.75 | 3.07 | 2.93 | 2.95 | 2.91 ± 0.12 | 9.27 |
The SNR is expected to be the higher the better.
y i is the S/O ratio of the ith subject in each group.
The average SNRs of the three levels for the injection activity, uptake duration, and time per projection are presented in Table 5 (see also S2 Table) and the main effect graph in Fig. 2. The average SNR for the time per projection significantly increased 23% from level 1 to level 3, while it increased approximately 17% for the uptake duration. Relatively, the average SNR for the injection activity had the slowest slope among the three parameters and the improvement was less than 10%, implying that the injection activity is insensitive to SNR.
Table 5. Average SNRs of the three levels for the injection activity, uptake duration, and time per projection.
| Parameter | Level 1 (dB) | Level 2 (dB) | Level 3 (dB) |
|---|---|---|---|
| Injection activity | 8.07 | 8.65 | 8.85 |
| Uptake duration | 8.03 | 8.13 | 9.40 |
| Time per projection | 7.74 | 8.30 | 9.53 |
Fig 2. The main effect graph for different levels of parameters.
The injection activity was less sensitive to levels than the uptake duration and time per projection.
Cross interactions between parameters
As clearly depicted in Fig. 2, the optimal setting for acquiring the highest SNR should be set as 962 MBq (level 3) for the injection activity, 260 min (level 3) for the uptake duration, and 60 s/projection (level 3) for the scanning time. However, to further inspect the cross interactions between parameters as indicated in Fig. 3, the injection activity had strong cross interactions to either the time per projection (Fig. 3A) or uptake duration (Fig. 3C). Therefore, it may only hold the injection activity in the minimum amount to maintain the maximum SNR. Moreover, the cross interaction between the time per projection and uptake duration (Fig. 3B) depicted a weak correlation and agreed well to the calculated results as implied in Fig. 2. Namely, both parameters in level 3 achieve the maximum SNR. Therefore, by concerning the multiple cross interactions between parameters, another possible candidate for parametric setting becomes as the same as indicated in group 3.
Fig 3. Major cross interactions between (A) injection activity and time per projection, (B) time per projection and uptake duration, and (C) uptake duration and injection activity.

ANOVA and F-test
Table 6 shows the F-test results as well as the confidence levels for the three parameters. If the confidence level exceeds 99%, the parameter is considered statistically significant [25]. The uptake duration and time per projection were the two significant factors in which the latter one dominated due to its largest F-value of 25.933. The second priority was the uptake duration which contributed 38.27% of the total F-value. The injection activity, on the other hand, was a minor factor which occupied only approximately 8.48% of the total F-value and was not statistically significant.
Table 6. F-test results and confidence levels for the three parameters.
| Design parameter | S x | f x | V x | F-value | Confidence level |
|---|---|---|---|---|---|
| Injection activity | 0.372 | 2 | 0.186 | 4.132 | 97.28% |
| Uptake duration | 1.678 | 2 | 0.839 | 18.638 | 100.00% |
| Time per projection | 2.335 | 2 | 1.168 | 25.933 | 100.00% |
The uptake duration and time per projection were the two significant factors since their confidence levels were over 99%.
S x is the sum of squared deviation.
f x is the degree of freedom.
V x is the variance.
Verification
Fig. 4 illustrates some [99mTc]TRODAT-1/SPECT images of the subjects using the conventional setting and the optimal setting. Table 7 indicates the average S/O ratios and SNRs for the normal subjects and PD patients for verification (see also S3 Table). The data from the original group 3 in Table 4 are also listed for comparison, since it fulfills the primary Taguchi’s suggestion on the basis of considering cross interactions between parameters. For the healthy subjects, the SNR of the optimal setting was 25% higher than that of the conventional setting, and the difference of the average S/O ratios was statistically significant (P<0.01). The optimal setting had a slightly higher average S/O ratio than the result of the group 3 setting. However, there was no significant difference (P = 0.955). For the PD patients, the differences of the average S/O ratios between the conventional/optimal settings and the conventional/group-3 settings were not statistically significant. Additionally, the average S/O ratios between the healthy subjects and PD patients all showed significant differences (P<0.01) under these three parameter settings. These findings confirm that 962 MBq, 260 min, and 60 s/projection are the best combination for [99mTc]TRODAT-1/SPECT scans. Additionally, the group 3 setting can achieve fine image quality comparable to the optimal setting.
Fig 4. [99mTc]TRODAT-1/SPECT images using the conventional setting for (A) a 34-year-old healthy female and (B) a 48-year-old male with PD, and using the optimal setting for (C) a 56-year-old healthy male and (D) a 45-year-old female with early PD.
Table 7. Comparison of the average S/O ratios and SNRs between the normal subjects and PD patients using the conventional setting, predicted optimal setting, and original group-3 setting.
| Parameter | Conventional setting | Optimal setting | Original group 3 |
|---|---|---|---|
| Injection activity (MBq) | 851 | 962 | 740 |
| Uptake duration (min) | 220 | 260 | 260 |
| Time per projection (s/proj.) | 50 | 60 | 60 |
| Normal subjects | |||
| SNR (dB) | 8.12 | 10.16 | 10.08 |
| Average S/O ratio | 2.55 ± 0.12 | 3.23 ± 0.17* | 3.20 ± 0.14 † |
| PD patients | |||
| SNR (dB) | 1.16 | 1.53 | 1.31 |
| Average S/O ratio | 1.19±0.23 | 1.25±0.24 ‡ | 1.24 ± 0.26 § |
*significant difference from the conventional setting (P<0.01)
†no significant difference from the optimal setting (P = 0.955)
‡no statistical significance from the conventional setting (P = 0.705)
§no statistical significance from the conventional setting (P = 0.719)
By fixing one of the parameters, isopreference curves through the level space defined by the other two parameters are drawn in Fig. 5. Points lying on the curve correspond to the levels achieving the same image quality. The isopreference curves shifted upper right to level 3. With this flexibility, the scanning protocol can be adjusted based on the patient condition, such as the ability to remain still for a longer acquisition time or the reduction of radiation doses by injecting less radioactivity and adjusting the other two variables.
Fig 5. Isopreference curves through the level space for (A) uptake duration and time per projection, (B) injection activity and time per projection, and (C) injection activity and uptake duration.

The number on the curve indicates the percentage of image quality.
Discussion
The scanning time per projection is the most dominant parameter in [99mTc]TRODAT-1/SPECT scans. This is mainly because the statistical variation in counts depends on the number of photons interacting with scintillation detectors. The quantum noise, the major source of image degradation in nuclear medicine, can be decreased by increasing the acquisition time. However, we should not prolong the total scanning time to more than 60 min [26]. Patients may feel uncomfortable and undesired motion blurring could jeopardize the image quality [27]. Motion during data acquisition can also result in an underestimation of DAT concentration, leading to quantitative errors and misdiagnosis [28].
The uptake duration is another dominant factor which can provide a superior and stable target to non-target ratio. The uptake duration applied in numerous studies ranged from 180 to 240 min for [99mTc]TRODAT-1/SPECT scans [29,30]. Based on our results, a longer delay time of 260 min is suggested for clinical practice. This recommendation is slightly longer than the imaging time investigated from other study [12]. Further prolonged uptake duration is not recommended due to additional biological and physical decay of radiotracers in the stratum [31].
The amount of [99mTc]TRODAT-1 injected for SPECT scans usually ranges from 740 to 962 MBq. Our results indicated that 962 MBq can achieve a slightly higher SNR than 740 and 851 MBq, but the differences were not statistically significant. Obviously, increasing the injection activity is straightforward to gain more photon counts. However, dead-time loses of the counting system may become severe and compromise the benefit of increasing activity. Another drawback is that the internal organ dose of patients increases as the injection activity increases. The dose-limiting organ, liver, receives approximately 45.2 mGy (0.047 mGy/MBq) and the effective dose achieves 11.6 mSv (0.012 mSv/MBq) for 962-MBq injection [32]. Comparing with other 99mTc-labeled compounds, the effective dose for [99mTc]TRODAT-1 is generally higher [33]. Therefore, a tradeoff should be made between the additional radiation risk and image quality, since the injection activity is a minor parameter in terms of SNR. We suggest that the minimum amount of 740 MBq, which still holds the Taguchi’s recommendation on the basis of considering cross interactions, should be used according to the concept of as low as reasonably achievable (ALARA) in radiation protection.
In medical research, single factor experiments are often conducted. Although considering only one variable at a time is simple and direct, the optimal results may not be revealed due to that the effect of one variable depends on the values of other variables. These cross interactions frequently occur in real-life situations. The Taguchi analysis is a fractional factorial testing which can significantly reduce sample sizes and examine several key parameters at the same time. Through the system design, parameter design, tolerance design, and verification phases, we can obtain useful information about the problem domain. More importantly, the cross-interaction between parameters should be evaluated carefully before giving suggestions to prevent potential bias of results.
The average age of the normal participants is younger than that of the PD cohort. This is because we recruited younger volunteers for the optimization study to prevent additional factors affecting DAT function with aging. In the verification study, the average age of the normal subjects is 52.9 years (age of 45–64 years old), which is higher than that of the subjects in the parameter optimization. The optimized parameters can still enhance the SNR. In this study, the average age of the PD cohort is 60.7 (age of 45–76 years old). We did not examine the effect of optimized scanning parameters on early onset PD patients, which is the limitation of this study.
In addition to the three major parameters examined in this study, other parameters could affect the image quality as well. Future work will focus on using statistical reconstruction algorithms, such as the Ordered Subset Expectation-Maximization (OSEM) and the Maximum a Posteriori (MAP). The iterative methods can provide outstanding spatial resolution between the caudate and putamen, and may gradually replace the traditional FBP algorithm for quantification [34]. The impact on the optimized scanning procedure should be further investigated.
Conclusion
In this study, we applied the Taguchi analysis to investigate the optimal imaging parameters for [99mTc]TRODAT-1/SPECT scans. Our analysis indicates that 962 MBq for the injection activity, 260 min for the uptake duration, and 60 s/projection for the acquisition are the best choice in terms of image quality. Since the injection activity is a minor factor, 740 MBq should be used to reduce the internal radiation dose according to ALARA. The optimal imaging parameters could be applied to clinical practice to elevate the diagnostic accuracy of PD and other dopamine-related disorders.
Supporting Information
(DOCX)
(DOCX)
(DOCX)
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
The authors have no support or funding to report.
References
- 1. Wu H, Lou C, Huang Z, Shi G (2011) SPECT imaging of dopamine transporters with 99mTc-TRODAT-1 in major depression and Parkinson's disease. J Neuropsychiatry Clin Neurosci 23: 63–67. 10.1176/appi.neuropsych.23.1.63 [DOI] [PubMed] [Google Scholar]
- 2. Braga-Neto P, Felicio AC, Hoexter MQ, Pedroso JL, Dutra LA, et al. (2012) Cognitive and olfactory deficits in Machado–Joseph disease: a dopamine transporter study. Parkinsonism Relat Disord 18: 854–858. 10.1016/j.parkreldis.2012.04.015 [DOI] [PubMed] [Google Scholar]
- 3. Felicio AC, Godeiro-Junior C, Moriyama TS, Shih MC, Hoexter MQ, et al. (2010) Degenerative parkinsonism in patients with psychogenic parkinsonism: a dopamine transporter imaging study. Clin Neurol Neurosurg 112: 282–285. 10.1016/j.clineuro.2009.12.010 [DOI] [PubMed] [Google Scholar]
- 4. Yang YK, Yeh TL, Yao WJ, Lee IH, Chen PS, et al. (2008) Greater availability of dopamine transporters in patients with major depression—a dual-isotope SPECT study. Psychiatry Res 162: 230–235. 10.1016/j.pscychresns.2007.08.008 [DOI] [PubMed] [Google Scholar]
- 5. la Fougère C, Krause J, Krause KH, Gildehaus FJ, Hacker M, et al. (2006) Value of 99mTc-TRODAT-1 SPECT to predict clinical response to methylphenidate treatment in adults with attention deficit hyperactivity disorder. Nucl Med Commun 27: 733–737. [DOI] [PubMed] [Google Scholar]
- 6. Tzen KY, Lu CS, Yen TC, Wey SP, Ting G (2001) Differential diagnosis of Parkinson’s disease and vascular parkinsonism by 99mTc-TRODAT-1. J Nucl Med 42: 408–413. [PubMed] [Google Scholar]
- 7. Moriyama TS, Felicio AC, Chagas MH, Tardelli VS, Ferraz HB, et al. (2011) Increased dopamine transporter density in Parkinson's disease patients with social anxiety disorder. J Neurol Sci 310: 53–57. 10.1016/j.jns.2011.06.056 [DOI] [PubMed] [Google Scholar]
- 8. Lai SC, Weng YH, Yen TC, Tsai CC, Chang HC, et al. (2004) Imaging early-stage corticobasal degeneration with [99mTc] TRODAT-1 SPET. Nucl Med Commun 25: 339–345. [DOI] [PubMed] [Google Scholar]
- 9. Lin CY, Changlia SP, Huang CK, Lin MS, Yeh CH (2007) Positive Tc-99m TRODAT findings in Japanese encephalitis. Clin Nucl Med 32: 484–485. [DOI] [PubMed] [Google Scholar]
- 10. Felicio AC, Godeiro-Junior C, Shih MC, Borges V, Silva S, et al. (2010) Evaluation of patients with clinically unclear Parkinsonian syndromes submitted to brain SPECT imaging using the technetium-99m labeled tracer TRODAT-1. J Neurol Sci 291: 64–68. 10.1016/j.jns.2009.12.024 [DOI] [PubMed] [Google Scholar]
- 11. Felicio AC, Moriyama TS, Godeiro-Junior C, Shih MC, Hoexter MQ, et al. (2010) Higher dopamine transporter density in Parkinson’s disease patients with depression. Psychopharmacology 211: 27–31. 10.1007/s00213-010-1867-y [DOI] [PubMed] [Google Scholar]
- 12. Kao PF, Tzen KY, Yen TC, Lu CS, Weng YH, et al. (2001) The optimal imaging time for [99Tcm] TRODAT-1/SPET in normal subjects and patients with Parkinson's disease. Nucl Med Commun 22: 151–154. [DOI] [PubMed] [Google Scholar]
- 13. Pan LK, Wang CC, Wei SL, Sher HF (2007) Optimizing multiple quality characteristics via Taguchi method-based grey analysis. J Mater Process Tech 182: 107–116. [Google Scholar]
- 14. Lin CL, Chang WJ, Lin YS, Chang YH, Lin YF (2009) Evaluation of the relative contributions of multi-factors in an adhesive MOD restoration using FEA and the Taguchi method. Dent Mater 25: 1073–1081. 10.1016/j.dental.2009.01.105 [DOI] [PubMed] [Google Scholar]
- 15. Chi CT, Chao CG, Liu TF, Wang CC (2008) Relational analysis between parameters and defects for electron beam welding of AZ-series magnesium alloys. Vacuum 82: 1177–1182. [Google Scholar]
- 16. Hsu WH, Chao CK, Hsu HC, Lin J, Hsu CC (2009) Parametric study on the interface pullout strength of the vertebral body replacement cage using FEM-based Taguchi methods. Med Eng Phys 31: 287–294. 10.1016/j.medengphy.2008.07.001 [DOI] [PubMed] [Google Scholar]
- 17. Yeh DM, Chang PJ, Pan LK (2013) The optimum Ga-67-citrate gamma camera imaging quality factors as first calculated and shown by the Taguchi's analysis. Hell J Nucl Med 16: 25–32. 10.1967/s002449910068 [DOI] [PubMed] [Google Scholar]
- 18. Taguchi G (1986) Introduction to quality engineering: designing quality into products and processes ARRB Group Limited. [Google Scholar]
- 19. Nian C, Yang W, Tarng Y (1999) Optimization of turning operations with multiple performance characteristics. J Mater Process Tech 95: 90–96. [Google Scholar]
- 20. Roy RK (2001) Design of experiments using the Taguchi approach: 16 steps to product and process improvement John Wiley & Sons. [Google Scholar]
- 21. Gelb DJ, Oliver E, Gilman S (1999) Diagnostic criteria for Parkinson disease. Arch Neurol 56: 33–39. [DOI] [PubMed] [Google Scholar]
- 22. Chang LT (1978) A method for attenuation correction in radionuclide computed tomography. IEEE Trans Nuclear Sci 25: 638–643. [Google Scholar]
- 23. Acton PD, Newberg A (2006) Artificial neural network classifier for the diagnosis of Parkinson's disease using [99mTc] TRODAT-1 and SPECT. Phys Med Biol 51: 3057–3066. [DOI] [PubMed] [Google Scholar]
- 24. Huang WS, Lee MS, Lin JC, Chen CY, Yang YW, et al. (2004) Usefulness of brain 99mTc-TRODAT-1 SPET for the evaluation of Parkinson’s disease. Eur J Nucl Med Mol Imaging 31: 155–161. [DOI] [PubMed] [Google Scholar]
- 25. Fisher R (1935) The design and analysis of experiments Edinburgh: Oliver and Boyd. [Google Scholar]
- 26. Tatsch K, Asenbaum S, Bartenstein P, Catafau A, Halldin C, et al. (2002) European Association of Nuclear Medicine procedure guidelines for brain neurotransmission SPET using (123)I-labelled dopamine D(2) transporter ligands. Eur J Nucl Med Mol Imaging 29: BP30–35. [PubMed] [Google Scholar]
- 27. Green MV, Seidel J, Stein SD, Tedder TE, Kempner KM, et al. (1994) Head movement in normal subjects during simulated pet brain imaging with and without head restraint. J Nucl Med 35: 1538–1546. [PubMed] [Google Scholar]
- 28. Koch W, Mustafa M, Zach C, Tatsch K (2007) Influence of movement on FP-CIT SPECT quantification: a Monte Carlo based simulation. Nucl Med Commun 28: 603–614. [DOI] [PubMed] [Google Scholar]
- 29. Acton PD, Newberg A, Plössl K, Mozley PD (2006) Comparison of region-of-interest analysis and human observers in the diagnosis of Parkinson's disease using [99mTc] TRODAT-1 and SPECT. Phys Med Biol 51: 575–585. [DOI] [PubMed] [Google Scholar]
- 30. Hwang WJ, Yao WJ, Fu YK, Yang AS (2008) [99mTc] TRODAT-1/[123I] IBZM SPECT studies of the dopaminergic system in Tourette syndrome. Psychiatry Res 162: 159–166. 10.1016/j.pscychresns.2007.04.006 [DOI] [PubMed] [Google Scholar]
- 31. Kushner SA, McElgin WT, Kung MP, Mozley PD, Plössl K, et al. (1999) Kinetic modeling of [99mTc] TRODAT-1: a dopamine transporter imaging agent. J Nucl Med 40: 150–158. [PubMed] [Google Scholar]
- 32. Mozley PD, Stubbs JB, Plossl K, Dresel SH, Barraclough ED, et al. (1998) Biodistribution and dosimetry of TRODAT-1: a technetium-99m tropane for imaging dopamine transporters. J Nucl Med 39: 2069–2076. [PubMed] [Google Scholar]
- 33. Lombardi MH (2012) Radiation safety in nuclear medicine CRC Press. [Google Scholar]
- 34. Koch W, Hamann C, Welsch J, Pöpperl G, Radau PE, et al. (2005) Is iterative reconstruction an alternative to filtered backprojection in routine processing of dopamine transporter SPECT studies? J Nucl Med 46: 1804–1811. [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
(DOCX)
(DOCX)
(DOCX)
Data Availability Statement
All relevant data are within the paper and its Supporting Information files.



