Improved Visualization of Intracranial Vessels with Intraoperative Coregistration of Rotational Digital Subtraction Angiography and Intraoperative 3D Ultrasound

Dino Podlesek 1, Tobias Meyer 2, Ute Morgenstern 2, Gabriele Schackert 1, Matthias Kirsch 1,*
Editor: Nima Etminan3
PMCID: PMC4372211  PMID: 25803318

Abstract

Introduction

Ultrasound can visualize and update the vessel status in real time during cerebral vascular surgery. We studied the depiction of parent vessels and aneurysms with a high-resolution 3D intraoperative ultrasound imaging system during aneurysm clipping using rotational digital subtraction angiography as a reference.

Methods

We analyzed 3D intraoperative ultrasound in 39 patients with cerebral aneurysms to visualize the aneurysm intraoperatively and the nearby vascular tree before and after clipping. Simultaneous coregistration of preoperative subtraction angiography data with 3D intraoperative ultrasound was performed to verify the anatomical assignment.

Results

Intraoperative ultrasound detected 35 of 43 aneurysms (81%) in 39 patients. Thirty-nine intraoperative ultrasound measurements were matched with rotational digital subtraction angiography and were successfully reconstructed during the procedure. In 7 patients, the aneurysm was partially visualized by 3D-ioUS or was not in field of view. Post-clipping intraoperative ultrasound was obtained in 26 and successfully reconstructed in 18 patients (69%) despite clip related artefacts. The overlap between 3D-ioUS aneurysm volume and preoperative rDSA aneurysm volume resulted in a mean accuracy of 0.71 (Dice coefficient).

Conclusions

Intraoperative coregistration of 3D intraoperative ultrasound data with preoperative rotational digital subtraction angiography is possible with high accuracy. It allows the immediate visualization of vessels beyond the microscopic field, as well as parallel assessment of blood velocity, aneurysm and vascular tree configuration. Although spatial resolution is lower than for standard angiography, the method provides an excellent vascular overview, advantageous interpretation of 3D-ioUS and immediate intraoperative feedback of the vascular status. A prerequisite for understanding vascular intraoperative ultrasound is image quality and a successful match with preoperative rotational digital subtraction angiography.

Introduction

Preoperative investigations of intracranial aneurysms are: Computed tomography angiography (CTA) magnetic resonance imaging angiography (MRA) and digital subtraction angiography (DSA). According to our experience, the interpretation of preoperative diagnostic imaging does not always match the real intraoperative situation in patients.

Complete clipping of intracranial aneurysms, preservation of the vascular tree, and identification of vessels beyond the visual field are the primary goals during aneurysm surgery. Le Roux et al. detected aneurysm remnants in 5.7% of 637 patients treated by clipping upon postoperative angiography [1]. Mere use of a microscope for aneurysm surgery is limited to the visual field. The intraoperative assessment of aneurysm clip placement has been evaluated by intravenous fluorescence angiography and intraoperative near-infrared indocyanin green video angiography [24]. As these techniques are limited to the visual field, possible entrapment and affection of distal vessels, e.g. by improper clip placement, may be missed. Intraoperative ultrasound (3D-ioUS) might improve this uncertainty since it is not limited to the visual field and allows for online imaging and real time update of images, respectively. Previous 3D-ioUS reports have shown that 3D-ioUS identified up to 90% of cerebral aneurysms using 2-dimensional transcranial color-coded ultrasonography [5,6]. In addition, advanced 3D-ioUS can render reconstructions of the vascular tree within seconds, therefore offering an additional measure to improve the diagnostic reconstruction of the vessel-tree and, thus, intraoperative safety [7].

The purpose of our study was to evaluate the usability, the accuracy and the intraoperative applicability of volume-based registration of 3D-ioUS in comparison with preoperative rDSA

Patients and Methods

Patients

Thirty-nine patients (13 men, 26 women, median age 57 years, range 35–80 years) with 59 saccular aneurysms were included in this study. Forty-three of 59 aneurysms were chosen for neurosurgical clipping and intraoperative ultrasound imaging. The rest of the aneurysms were treated with coiling. Twenty-one patients underwent elective procedures and 18 were treated within 24 hours after suffering from subarachnoid hemorrhage (SAH; Hunt & Hess grades II-V).

Data acquisition

Our study was performed according to the Declaration of Helsinki on Biomedical Research involving Human Subjects. It was approved by the Ethics Committee of the Faculty of Medicine Carl Gustav Carus at the Technische Universität Dresden, Germany (EK63042005; EK348092013). The participants or their relatives provided informed consent either written or verbally to participate in this study. In six patients, written consent was not obtained but approval was given by their relatives and documented in the patients file.

Pre-operative imaging studies

Thirty-nine patients underwent DSA with 3D. Selective pan-angiography DSA scans with multiple planes of each carotid artery and 4 planes of vertebrobasilar tree were routinely performed. In 6 patients, preoperative computed tomographic angiography (CTA) was obtained for the acute diagnostic approach.

Clipping procedure

The head was fixed in a Sugita head frame. Pterional craniotomy or modified frontotemporal approach were chosen for clipping the aneurysms located on anterior cerebral circuit. Before aneurysm dissection, the ultrasound volume was obtained. This was accomplished by holding the ultrasound probe manually or by immobilizing the probe with a pneumatic-flexible arm (Unitrack Aesculap, Tuttlingen, Germany). Setup and completion time were less than 10 minutes, even if several volumes were acquired. The time limiting factor was most frequently a technical issue, human error, or limitation due to a poor clinical status of the patient.

Consecutively, 3D-ioUS data sets were imported to a laptop with a stereoscopic display and coregistered with preoperative rDSA resulting in a rapid 3D display of the coregistered vascular tree. Routinely, intraoperative micro-Doppler measurements were performed before and after clip placement using a 16 MHz conventional micro-Doppler probe (diameter 1 mm, DWL, Singen Germany) by applying the ultrasound-probe directly on the visible vessels. In selected patients, near-infrared indocyanine green (ICG) videoangiography was performed as well [3].

Intraoperative 3D-ultrasound

Intraoperative high-resolution power Doppler 3D imaging (Voluson 730, GE Healthcare, formerly Kretz, Brussels) was used in 39 patients with anterior, middle cerebral and internal carotid artery aneurysms to visualize the aneurysm and the vascular tree prior to and after clipping. For this purpose, a curved linear array transducer (4D micro-convex endocavitary transducer, 5–9 MHz ultrasound probe, RIC5–9 curved linear array volumetric real time 3D probe, GE Healthcare, formerly Kretz, Brussels) was used. The ultrasound probe contains the mechanism that pivots the ultrasound crystals during the acquisition of volume data. The probe was attached to a pneumatic arm (Unitrack Aesculap, Tuttlingen, Germany) to minimize motion artefacts. Both power Doppler and conventional B-mode volume data sets were captured in varying B-image angles of 40°-130° and panning angles of 20°-95°. Before the coregistration procedure and data reconstruction of the intracranial vessels, analysis of 3D-ioUS MIP (maximum intensity projection; MIP) images was performed. Whenever possible, series of 3D power Doppler and color-coded duplex measurements were performed. These data were transferred to a laptop with a 3D display within eye-sight of the neurosurgeon.

Hardware and software visualization tools

A notebook with integrated autostereoscopic display (Sharp AL3DU; Sharp Corporation, Osaka, Japan) enabled immediate 3D impressions of the vascular morphology without using additional viewing aids such as shutter glasses.

A 3D mouse (Spaceball 5000, 3DConnexion, Fremont, California, USA) supported navigating and maneuvering objects in 3D space. Digital 3D planning and editing was performed using the commercial visualization software package AMIRA (Visage Imaging, Berlin, Germany, formerly Mercury Computer Systems, TGS Inc., San Diego, USA). Beyond basic 3D software tools, AMIRA approves development and integration of customized 3D visualization and processing algorithms. Hence, specific software modules were implemented for neurosurgical-oriented demand. The 3D-ioUS steps of processing and visualization have been automated and the interactive steps have been reduced to a minimum during the process of coregistration and visualization. 3D-reconstruction, visualization and coregistration of rDSA, as well as 3D-ioUS were immediately performed after ultrasound imaging. US data sets were post processed by digital elimination of artefacts (isolated voxel), visual optimization (smoothing with Gaussian filter), and updating those for intraoperative rDSA fusion. An automated rigid (translation, rotation) registration algorithm (standard AMIRA module AffineRegistration) registered the 3D-ioUS data set to the reference data set (rDSA). The algorithm progressed stepwise for multiple resolutions: Registration is initially performed with a coarse resolution of the data sets. Then, a registration with finer resolutions used the transformation parameters of the prior registration. Due to the different modalities of the image data, the normalized mutual information metric was used to estimate the similarity between rDSA and 3D-ioUS for automatic registration. In case of insufficient alignment, the registration was manually conducted using the 3D navigation device.

For a more accurate comparison of 3D-ioUS data with preoperative rDSA, local misalignment was compensated. These misalignments originated from deformation of the brain because of cranial opening and the different imaging geometry of rDSA and 3D-ioUS. A non-rigid registration algorithm can compensate for local deformations of the 3D-ioUS data after rigid registration. The module was programmed using the B-Spline deformable transform from the Insight Segmentation and Registration Toolkit (ITK) [8]. An evenly spaced grid was laid over the 3D-ioUS data set. The grid nodes can move in each dimension independently. Parameterized cubic curves (B-Splines) interpolated the deformation between the single grid nodes. The normalized mutual information was used as similarity metric. The grid nodes moved in respect to the optimization of the similarity between the data sets. After coregistration, the rDSA and the 3D-ioUS were visualized. A surface rendering algorithm (standard AMIRA module IsoSurface) visualized the rDSA data sets. The surface of the vessels were reconstructed as a tetrahedral grid and visualized as a shaded surface. 3D-volume rendering visualized the 3D-ioUS data set. Transparent volume rendering made simultaneous perceptibility of both data sets possible.

All data processing steps described above were integrated into a single AMIRA-based program which was executed on demand. For practicability and ease of 3D navigation, a 3D mouse encased into a sterile wrapping was available on demand for the neurosurgeon. All patients with ultrasound data sets were evaluated retrospectively as well.

Post-operative imaging

All patients had a postoperative CT scan on the following day. For detection of vasospasm, Transcranial Doppler US of intracranial vessels (DWL Doppler; Compumedics GmbH, 2MHz probe, Singen, Germany) was performed daily to bidaily. Within one week, postoperative conventional digital subtraction angiography was performed in all patients. In 6 patients, postoperative rDSA was performed because the patients had a second surgery for an additional aneurysm or unclear remnants in conventional DSA. To validate intraoperative findings, post-operative control rDSA results were matched with the intraoperatively acquired 3D-ioUS volume.

Assessment of 3D-ioUS quality

We determined the intraoperative aneurysm carrier morphology, aneurysm volume, aneurysm neck width, presence of vasospasm, and compared those to preoperative rDSA. 3D-ioUS data were evaluated by an ultrasound experienced vascular neurosurgeon. Postoperatively, those were re-evaluated independently by two neurosurgeons with regard to aneurysm visualization and anatomy, detection of parent vessel, aneurysm neck depiction, and successful coregistration of 3D-ioUS and rDSA. Post-processing and intraoperative adaption of US data sets were time consuming, which improved over time, resulting in a data acquisition and processing time of less than 15 min.

Similarity of aneurysm volume visualization of rDSA and 3D-ioUS

We compared the volume (voxel count) overlap between the rDSA (Vvessel, rDSA, reference data set) and the 3D-ioUS (Vvessel, ioUS, transformed data set) to evaluate the similarity after registration (Fig. 1). The volume of interest (VOI) was centered on the aneurysm in both data sets and cropped to fit the aneurysm including the aneurysm neck. The true positive (TP), the false positive (FP), and the false negative (FN) counts of the overlap were calculated for each registration. The TP value describes the intersection between the rDSA and the transformed 3D-ioUS. FN and FP represent the non-overlapping volumes of the DSA and the 3D-ioUS, respectively.

Fig 1. Pattern of the evaluation of registration results by overlapping the rDSA and the 3D-ioUS (rDSA rotational digital subtraction angiography, 3D-ioUS: intraoperative 3D-ultrasound, TN: true negative, FN: false negative, TP: true positive, FP: false positive).

Fig 1

Evaluation of registration; a.: preoperative rDSA; b.: preoperative 3D-ioUS; c.: intraoperative coregistration of a. and b.

These data were used to calculate both the Jaccard and the Dice coefficients [9]. The Jaccard- index describes the size of the intersection of the volume of the two data sets relative to the size of their union (J(A,B) = (A and B)/(A or B)). Dice-coefficient displays more robust estimation of spatial overlap than the Jaccard-index, using the following equation: D(A,B) = 2 (A and B)/(A + B) where A and B are the vessels in the rDSA data set and the 3D- ioUS data sets, respectively. This formula represents the size of the union of 2 sets divided by the average size of the both sets. It can be expressed as: D(A,B) = 2 TP /((FP+TP)+(TP+FN)). It equals 1 if the volumes overlap is complete and equals 0 if there is no overlap (Fig. 1).

Results

Patient characteristics

Before the aneurysm dissection and clipping, we obtained 3D-ioUS volumes in 39 patients and repeated the measurement in 26 patients after clipping. Thirty-nine patients harbored 43 aneurysms and underwent both 3D-ioUS and rDSA. Duplex sonography of associated intracranial vessels was obtained in 24 patients.

As a result of improper insonation in 5 out of 39 patients (13%), aneurysms were outside of the field of view, yet they provided the additional information on ipsilateral vascular tree beyond the visual field. We visualized 35 of 43 aneurysms (81%) with 3D-ioUS and reconstructed those successfully with AMIRA software. Due to inaccurate insonation angle and aneurysm wall calcification in patient 13 and patient 32 only the aneurysm neck and the outgoing vessels were visualized and successfully matched.

Simultaneous coregistration of preoperative rDSA data with intraoperative pre-clipping 3D-ioUS was performed sufficiently in 37 of 39 (95%) patients. The coregistration was insufficient in 8 of 26 patients in post clipping 3D-ioUS due to clip artefacts.

The result of coregistration was demonstrated to two ultrasound-experienced vascular neurosurgeons who subjectively evaluated the coregistration results intraoperatively. Their assessment was excellent with coregistered identification of the vascular structures (Table 1).

Table 1. Evaluation of 3D-ioUS.

Patient Location Aneurysm volume (cm3) 3D-ioUS availability 3D-ioUS sufficiency Aneurysm Carrier vessel Aneurysm neck Outgoing vessels Coregistration of ioUS/rDSA Comments
Pre / post Clip
1 MCA 0.133 +/− +++/− +++/− +++/− +++/− +++/− +/−
2 MCA 0.298 +/+ +++/+++ +++/+++ +++/+++ +++/++ +++/++ +/+
3 MCA 0.08 +/− +++/− +++/− +++/− +++/− +++/− +/−
4 MCA 0.044 +/+ +++/+++ +++/++ +++/++ +++/+++ +++/+++ +/+
5 MCA 1.44 +/− +++/− +++/− +++/− +++/− ++/− +/−
6 AcommA 1.4 +/− +/− −/− −/− −/− +/− −/− Aneurysm not in FOV
7 MCA/ +ICA-T 0.03 +/+ +++/+++ +++/+++ +++/+++ ++/++ +++/+ +/+
8 AcommA. 0.22 +/− ++/− −/− −/− −/− −/− +/− Aneurysm not in FOV
9 AcommA. 0.16 +/− +++/− −/− −/− −/− −/− +/− Aneurysm not in FOV
10 MCA 0.76 +/− +++/− ++/− ++/− ++/− ++/− +/−
11 AcommA. 0.98 +/− ++/− −/− −/− −/− −/− +/− Aneurysm not in FOV
12 AcommA. 1.17 +/+ +++/+ +++/− +++/− +++/++ ++/− +/+
13 AcommA. 0.64 +/+ +++/+ −/− +++/+ ++/− +++/− +/− Thrombosed aneurysm
14 MCA 1.32 +/+ +++/+ +++/− +++/− +++/− ++/− +/−
15 MCA 0.084 +/+ +++/+++ +++/++ +++/++ +++/+++ +++/++ +/+
16 MCA 0.21 +/− +++/− ++/− +++/− ++/− +++/− +/−
17 MCA 0.81 +/+ +++/+ +++/− ++/+ ++/− +++/+ +/−
18 MCA 0.13 +/+ ++/+ ++/− +/− +/− ++/++ +/+
19 MCA/MCA 0.036 +/+ +++/+++ ++/++ +++/+++ +++/+++ +++/+++ +/+
20 MCA 0.36 +/+ +++/+++ +++/++ +++/++ +++/++ ++/++ +/+
21 MCA/MCA 0.165 +/+ +++/++ +++/+ +++/+ +++/+++ +++/++ +/+
22 MCA/MCA/MCA 0.364 +/+ +/+ +/+ +++/++ ++/++ ++/++ +/+
23 MCA 0.01 +/− +++/− ++/− ++/− ++/− ++/− +/−
24 MCA/MCA 0.7 +/+ +++/+ +++/− +++/+ ++/− ++/− +/−
25 MCA 0.247 +/+ +++/++ ++/++ +++/++ ++/+++ +++/++ +/+
26 AcommA./MCA 0.04 +/+ +/+ −/− +/− +/− +/− −/− Aneurysm not in FOV
27 MCA 0.56 +/+ +++/+++ +++/+++ +++/+++ +++/+++ +++/+++ +/+
28 AcommA./MCA 0.193 +/+ +++/++ ++/++ +++/+++ +++/+++ ++/++ +/+
29 MCA 2.02 +/+ ++/+ ++/− ++/− ++/− ++/− +/−
30 MCA/MCA/MCA 0.59 +/− +++/− +++/− +++/− +++/− +++/− +/−
31 AcommA 0.77 +/+ +++/+ +++/− +++/− +++/− ++/− +/−
32 ICA-T 1.1 +/− ++/− ++/− ++/− ++/− ++/− +/− Aneurysm partially thrombosed
33 MCA 2.5 +/− +++/− +++/− +++/− +++/− ++/− +/−
34 MCA 0.068 +/− +++/− +++/− +++/− +++/+++ −/− +/−
35 ICA-T 1.81 +/− ++/− ++/− ++/− ++/− ++/− +/−
36 MCA 0.81 +/+ +++/− +++/− +++/− +++/− +++/+ +/−
37 AcommA/MCA/MCA 0.14 +/− +++/− +++/− +++/− ++/− ++/− +/−
38 AcommA 0.1 +/+ +/− +/− −/− −/− −/− −/−
39 MCA 0.035 +/+ +++/++ +++/++ ++/++ ++/++ ++/+++ +/+

Aneurysm volumes were calculated from the maximum intensity projection (MIP) volumes of ioUS. As indicated, ioUS measurements were performed pre- and post-clipping. Two independent investigators evaluated the data which is depicted on a scale from – (not available or insufficient images) to +, ++, or +++. The overlap presents the volumetric ratio of the aneurysms.

Accuracy of rDSA—3D-ioUS coregistration

We used the Dice-coefficient and Jaccard-index to quantify the intersection overlap between 3D-ioUS aneurysm volume and preoperative rDSA aneurysm volume (Table 2). The segmented volumes of 3D-ioUS and preoperative rDSA data sets were very similar with 94% accuracy. Although the aneurysm volumes were approximately the same size in rDSA and 3D-ioUS, they had slightly different 3D conformations resulting in a Jaccard-index of 0.56. The brain shift could be one of the factors influencing those data. Therefore, the Dice-coefficient was employed to provide a more robust estimate [9] of the spatial similarity of different imaging modalities resulting in a good overlap of 0.71. This should provide a clinically meaningful index to assess the comparability of the imaging modalities.

Table 2. Results of Dice-coefficient, Jaccard-Index and volume ratios for rDSA and 3D-ioUS volumes. (Due to incomplete aneurysm visualization or absent aneurysm in the field of view patients 6, 8, 9, 11, 13, 26 and 32 were excluded from calculations).

Patient Aneurysm Jaccard-index Dice-coefficient OverlapV3D-ioUS: VrDSA
1 MCA 0.76 0.87 0.93
2 MCA 0.58 0.74 0.77
3 MCA 0.59 0.74 0.68
4 MCA 0.47 0.64 0.76
5 MCA 0.45 0.62 0.86
6 AcommA - - -
7 MCA 0.27 0.43 0.89
8 AcommA - - -
9 AcommA - - -
10 MCA 0.74 0.85 0.96
11 AcommA - - -
12 AcommA. 0.43 0.60 0.93
13 AcommA - - -
14 MCA 0.22 0.36 0.80
15 MCA 0.61 0.76 1.20
16 MCA 0.59 0.74 0.99
17 MCA 0.52 0.68 0.97
18 MCA 0.51 0.68 0.89
19 MCA 0.56 0.72 0.84
20 MCA 0.46 0.63 0.89
21 MCA 0.66 0.80 0.97
22 MCA 0.61 0.76 0.97
23 MCA 0.65 0.79 0.83
24 MCA 0.52 0.69 0.90
25 MCA 0.71 0.83 1.01
26 AcommA./MCA - - -
27 MCA 0.70 0.82 1.06
28 AcommA. 0.61 0.76 1.08
29 MCA 0.63 0.78 1.15
30 MCA 0.54 0.70 1.13
31 AcommA 0.68 0.81 1.18
32 ICA-T - - -
33 MCA 0.56 0.72 1.04
34 MCA 0.56 0.72 0.85
35 ICA-T 0.52 0.69 0.94
36 MCA 0.66 0.79 1.01
37 AcommA 0.50 0.67 1.02
38 AcommA 0.40 0.58 0.88
39 MCA 0.60 0.75 1.05
Mean 0.56 0.71 0.94

The index equals 1 if the volumes overlap completely and 0 if there is no overlap.

Coregistration of preoperative 3D-ioUS and rDSA

Illustrative cases

In patient 13, the aneurysm was only partially visualized in 3D-ioUS compared to rDSA, suggesting either a partially thrombosed aneurysm or a thickened and calcified aneurysm wall.

In patient 17, a complex MCA bifurcation aneurysm was clipped. 3D-ioUS obtained a good match with the pre-operative rDSA, however, it showed a vasospasm of the temporal M2 branch distal to the clip. This finding was confirmed in postoperative conventional DSA and in the corresponding MIP images of 3D-ioUS, although this was initially interpreted as an occlusion of M2-segment in 3D-ioUS (Fig. 2).

Fig 2. Patient 17: The black arrow indicates M2 segment (temporal branch of MCA) of the medial cerebral artery, vasospasm of M2 segment visualized in post-clipping 3D-ioUS (E) which led to repositioning of the clip; (rDSA (red), 3D-ioUS (blue, orange); A: pre-clipping conventional DSA; B: pre-clipping rDSA; C: pre-clipping 3D-ioUS; D: post-clipping conventional DSA; E: post-clipping 3D-ioUS; F: coregistration of C+E.

Fig 2

In patients 5, 29, and 30, the combination of 3D-ioUS with rDSA revealed details not apparent on the single modality imaging. The combination with ultrasound flow measurements (duplex) demonstrated a degree of detail that was not visible in rDSA alone. This is potentially an effect of high flow velocity in the parent vessel. In patient 5, 3D-ioUS provided additional information of M1 branch anatomy (Fig. 3).

Fig 3. Patient 5: The yellow arrow indicates the aneurysm of the medial cerebral artery bifurcation; the black Arrow indicates the aneurysm origin/neck; improved visualization of the aneurysm neck in 3D-ioUS (B) with narrow M1 segment of the medial cerebral artery due to high blood flow velocity in bifurcation marked with a white arrow (D, E).; rDSA (red), 3D-ioUS (blue, glow-orange), A: pre-clipping rDSA; B: pre-clipping 3D-ioUS; C: coregistration of A+B; D: pre-clipping duplex; E: MIP 3D-ioUS; F: coregistration of A+E.

Fig 3

In particular, the aneurysm neck was clearly separated from the surrounding artery. In patient 30, 3D-ioUS demonstrated two separate aneurysm domes in contrast to a fused dome in pre-operative rDSA (Fig. 4).

Fig 4. Patient 30: The white arrows indicate the aneurysms of the medial cerebral artery bifurcation.

Fig 4

Contrary to rDSA (G; 2 aneurysms detected), 3 aneurysms were displayed in 3D-ioUS (H, J, K; 2 on MCA bifurcation and the 3rd aneurysm on M2 branch) as confirmed by the intraoperative view; (rDSA (red), 3D-ioUS (blue, glow- orange)); G: pre-clipping rDSA; H: pre-clipping 3D-ioUS; I: coregistration of G+H; J: MIP 3D-ioUS; K: MIP of 3D-ioUS; L: coregistration of K+H.

Similarly, patient 29 revealed a better separation of individual M2 branches compared to rDSA (Fig. 5).

Fig 5. Patient 29: Aneurysm of the medial cerebral artery bifurcation; the black and white arrows indicate the M2 segments of the medial cerebral artery; uncertainty in preoperative rDSA interpretation of M2 segment (single broad vessel vs. 2 nearby vessels); detection of 2 M2 branches (N, P and R) in 3D-ioUS as confirmed intraoperatively; rDSA (red), 3D-ioUS (blue, glow-orange); M: pre-clipping rDSA; N: pre-clipping 3D-ioUS; O: coregistration of M+N; P,Q: MIP of 3D-ioUS; R: coregistration of R+M.

Fig 5

3D-ioUS had its limitations due to calcification and clip artefacts. It gave us an assumption of aneurysm calcifications by producing acoustic shadowing. In some patients, acoustic shadowing hindered the interpretation of 3D-ioUS. Several 3D-ioUS volumes of the region of interest were necessary to interprete the vascular anatomy (patients 2, 4, 13, 17, 24, 27, 30, 35). Pre-existing clip placement generated artefacts in patient 24, however, imaging at different angles made the aneurysm visualization possible.

By obtaining several 3D-ioUS volumes from different angles and fixing the probe to a pneumatic arm, clip and motion artefacts decreased to a minimum. The parent vessel and the aneurysm neck were visualized with high precision in absence of motion artefacts (Fig. 6).

Fig 6. Patient 4: Aneurysm of the medial cerebral artery bifurcation indicated by the white arrow.

Fig 6

Precise aneurysm neck visualization in 3D-ioUS and clip segmentation for virtual planning of clipping procedure. A: pre-clipping rDSA (red); B: post- clipping rDSA (green); C: pre-clipping 3D-ioUS (blue); D: post-clipping 3D-ioUS (pink); E: post- clipping rDSA with segmented clips (red); F: post- clipping 3D-ioUS with segmented clips (red); G—H: coregistration of C+D; I: coregistration of A+C; J: coregistration of B+G.

Discussion

Intraoperative semi-automated coregistration of 3D intraoperative ultrasound with preoperative rDSA combines the advantages of ultrasound, namely speed, non-invasiveness and visualization beyond the visible field, with an improved anatomical interpretation of the coregistered 3D-US images. Using standard visualization tools, it is possible to simulate and evaluate the intraoperative strategy when the anatomical structures were afflicted by brain shifting and other changes concerning the anatomical discrepancy to the preoperative data.

Successful detection and interpretation of ioUS requires a learning curve in particular for small intracranial aneurysms and small vessels. Despite the inability to visualize small neck remnants and perforating vessels after clipping, intraoperative 3D-ioUS coregistered with preoperative rDSA was positively evaluated by independent neurosurgeons.

Since the first report of transcranial ultrasound application, Stendel et al. has shown the reliability, practicability and the necessity of intraoperative microvascular Doppler ultrasonography in cerebral aneurysm surgery [10,11]. The aneurysm detection by transcranial power Doppler depends on aneurysm location and its morphology. In 3D transcranial power sonography, the differentiation of artefacts and real time vascular changes is superior to 2-dimensional transcranial color- coded sonography [5,12]. A diameter smaller than 5–6 mm, poor bone window, and adverse aneurysm location were the cause of insufficient aneurysm visualization using transcranial color-coded duplex sonography [12,13]. Turner et al. showed that the sensitivity and specificity increase with the aneurysm diameter and are 100% in aneurysms with a 12mm diameter. The application of the intravenous contrast agent increases its sensitivity and specificity [14]. Klötzsch et al. reported a higher rate of undetected aneurysms using nonenhanced 3D-TCCS than contrast enhanced ultrasound [12]. Other studies employing neurosonography integrated in neuronavigation systems have been introduced and have clearly shown intraoperative advantages [1522]. Lindseth et al. have shown that blood-flow imaging as an independent ultrasound modality and integrated with 3D navigation technology has a high potential for the visualization of flow dynamics of intracranial vascular pathologies [13].

In the retrospective analysis of 180 patients with 205 aneurysm Brinjikji et al. showed the differences in measurements of dome-to-neck ratios between 2D and 3D-rDSA [23]. There is a tendency to overestimate the neck size due to volume averaging in 3D imaging techniques. As we have shown for patient 17 the evaluation of intraoperative 2D-ioUS revealed a severe M2 vasospasm previously interpreted as an occlusion in 3D-ioUS. For accurate imaging interpretation and, therefore, assessment of vessel anatomy, 2D imaging is still crucial. The noninvasiveness and possibility of repeated ioUS measurements from different angles are advantageous for improved imaging interpretation.

As an alternative to the established vascular imaging modalities, such as ICG angiography, vascular US and conventional DSA, a selective catheterization of the exposed ipsilateral superficial temporal artery has been proposed as a rapid intraoperative angiography [24]. Although the procedure is simple and can be repeated during the procedure once the catheter is in place, the preparation of the STA is time-consuming compared to US, its usability for DSA not obvious beforehand, it sacrifices the superficial temporal artery, and radiation exposure is inevitable [24]. The major advantage is its superior resolution and comparability with standard DSA imaging.

Sufficient clipping can be accessed by ICG application. Good visualization depends on the aneurysm location relative to the parent vessel and its morphology. Sharma et al. advocated the intraoperative ICG application as a complementary technique to intraoperative digital subtraction angiography [25]. ICG has a superior resolution to visualize small and perforating arteries as long as they are in the visual field. Even though the technique has a low risk profile and is useful for adequacy of aneurysmal obliteration and patency of perforating vessels, the evaluation of perfused vessels is limited to the visual field and is flow dependent [3,2633]. Only DSA and US are able to visualize blood flow in the non-visible compartment.

Although the intraoperative DSA has been introduced by several studies as a “gold standard” technique for the detection of aneurysm remnants, the intraoperative risk, the invasive procedure and the cost effectiveness are the main disadvantages. The visualization of the perfusion of the perforating vessels is still not sufficient due to imaging resolution of the intraoperative DSA [3437].

In our study none of the aneurysm remnants (patient 24, 27, and 33) were detected with ioUS due to the immediate proximity to the surgical clip. However, regrown aneurysms were detectable. Pre-clipping, aneurysms were visualized down to 3 mm in diameter. After the clipping procedure, the aneurysm and the parent vessels were better visualized in 3D-ioUS compared to 2D MIP images as seen in several patients. The Dice-coefficient for automatic comparison between two different imaging modalities proved to be more robust than the Jaccard-index. In fact, the aneurysm volumes were very similar upon 3D-ioUS and rDSA imaging. The interpretation of 3D-ioUS is improved when comparing (coregistration) those with preoperatively obtained rDSA data.

The spatial resolution of vascular ultrasound is inferior to standard 3D-DSA, and within the operative field, inferior to ICG angiography. In particular, very small arteries, such as perforating arteries are difficult to discern. However, the diagnostic accuracy continuously evolves with the application of intraoperative ultrasound contrast enhancing agents and newer technologies [13].

Calcified plaques and the clip-material, producing an acoustic shadow, are the limiting factors for intracranial ultrasound imaging. In such cases, we showed that several ioUS-volumes from different angles were required for successful vascular reconstruction. Image reconstruction and coregistration is able to decrease clip artefacts. Even better results might be possible using elastic fusion algorithms [3840], or the use of adaptive coregistration algorithms for approximation of the resolution of two different image modalities, as well as vascular ultrasound contrast agents.

Despite its limitations, the combined use of the diagnostic information of DSA, ICG and vascular ultrasound imaging fills most intraoperative diagnostic gaps as complementary tools.

Further developments should combine power 3D-ioUS integrated into a neuronavigation system for identification and reduced misinterpretation of ioUS data due to brain-shift, in perspective allowing for transcranial 3D color coded Doppler and the possibility for blood flow velocity quantification in 3D. In addition, mathematical algorithms have been developed to improve the image quality by reducing the artefacts [41,42]. The immediate intraoperative coregistration of 3D-ioUS and rDSA of cerebral aneurysms conveys one of the many benefits of US in the neurosurgical operating theatre and shows that although the ultrasound data volumes were difficult to interpret on their own, the coregistration to rDSA images is important for interpretation.

Acknowledgments

We would like to thank Prof. Dr. Rüdiger von Kummer, Dept. of Neuroradiology for providing the preoperative diagnostic imaging.

Data Availability

All relevant data are within the paper.

Funding Statement

The authors have no support or funding to report.

References

  • 1. Le Roux PD, Elliott JP, Eskridge JM, Cohen W, Winn HR (1998) Risks and benefits of diagnostic angiography after aneurysm surgery: a retrospective analysis of 597 studies. Neurosurgery 42: 1248–1254; discussion 1254–1245. [DOI] [PubMed] [Google Scholar]
  • 2. de Oliveira JG, Beck J, Seifert V, Teixeira MJ, Raabe A (2007) Assessment of flow in perforating arteries during intracranial aneurysm surgery using intraoperative near-infrared indocyanine green videoangiography. Neurosurgery 61: 63–72; discussion 72–63. [DOI] [PubMed] [Google Scholar]
  • 3. Raabe A, Nakaji P, Beck J, Kim LJ, Hsu FP, et al. (2005) Prospective evaluation of surgical microscope-integrated intraoperative near-infrared indocyanine green videoangiography during aneurysm surgery. Journal of neurosurgery 103: 982–989. [DOI] [PubMed] [Google Scholar]
  • 4. Wrobel CJ, Meltzer H, Lamond R, Alksne JF (1994) Intraoperative assessment of aneurysm clip placement by intravenous fluorescein angiography. Neurosurgery 35: 970–973; discussion 973. [DOI] [PubMed] [Google Scholar]
  • 5. Klotzsch C, Nahser HC, Fischer B, Henkes H, Kuhne D, et al. (1996) Visualisation of intracranial aneurysms by transcranial duplex sonography. Neuroradiology 38: 555–559. [DOI] [PubMed] [Google Scholar]
  • 6. Wardlaw JM, Cannon JC (1996) Color transcranial "power" Doppler ultrasound of intracranial aneurysms. Journal of neurosurgery 84: 459–461. [DOI] [PubMed] [Google Scholar]
  • 7. Yasuda K, Nakajima S, Wakayama A, Oshino S, Kubo S, et al. (2003) Intraoperative three-dimensional reconstruction of power Doppler vascular images. Minimally invasive neurosurgery: MIN 46: 323–326. [DOI] [PubMed] [Google Scholar]
  • 8. Yoo TS, Ackerman MJ, Lorensen WE, Schroeder W, Chalana V, et al. (2002) Engineering and algorithm design for an image processing Api: a technical report on ITK—the Insight Toolkit. Studies in health technology and informatics 85: 586–592. [PubMed] [Google Scholar]
  • 9. Zou KH, Warfield SK, Bharatha A, Tempany CM, Kaus MR, et al. (2004) Statistical validation of image segmentation quality based on a spatial overlap index. Academic radiology 11: 178–189. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Aaslid R, Markwalder TM, Nornes H (1982) Noninvasive transcranial Doppler ultrasound recording of flow velocity in basal cerebral arteries. Journal of neurosurgery 57: 769–774. [DOI] [PubMed] [Google Scholar]
  • 11. Stendel R, Pietila T, Al Hassan AA, Schilling A, Brock M (2000) Intraoperative microvascular Doppler ultrasonography in cerebral aneurysm surgery. Journal of neurology, neurosurgery, and psychiatry 68: 29–35. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Klotzsch C, Bozzato A, Lammers G, Mull M, Lennartz B, et al. (1999) Three-dimensional transcranial color-coded sonography of cerebral aneurysms. Stroke; a journal of cerebral circulation 30: 2285–2290. [DOI] [PubMed] [Google Scholar]
  • 13. Lindseth F, Lovstakken L, Rygh OM, Tangen GA, Torp H, et al. (2009) Blood flow imaging: an angle-independent ultrasound modality for intraoperative assessment of flow dynamics in neurovascular surgery. Neurosurgery 65: 149–157; discussion 157. 10.1227/01.NEU.0000345945.92559.C5 [DOI] [PubMed] [Google Scholar]
  • 14. Turner CL, Kirkpatrick PJ (2000) Detection of intracranial aneurysms with unenhanced and echo contrast enhanced transcranial power Doppler. Journal of neurology, neurosurgery, and psychiatry 68: 489–495. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Berntsen EM, Gulati S, Solheim O, Kvistad KA, Lindseth F, et al. (2009) Integrated pre- and intraoperative imaging in a patient with an arteriovenous malformation located in eloquent cortex. Minimally invasive neurosurgery: MIN 52: 83–85. 10.1055/s-0028-1124104 [DOI] [PubMed] [Google Scholar]
  • 16. Letteboer MM, Willems PW, Viergever MA, Niessen WJ (2005) Brain shift estimation in image-guided neurosurgery using 3-D ultrasound. IEEE transactions on bio-medical engineering 52: 268–276. [DOI] [PubMed] [Google Scholar]
  • 17. Lindseth F, Kaspersen JH, Ommedal S, Lango T, Bang J, et al. (2003) Multimodal image fusion in ultrasound-based neuronavigation: improving overview and interpretation by integrating preoperative MRI with intraoperative 3D ultrasound. Computer aided surgery: official journal of the International Society for Computer Aided Surgery 8: 49–69. [DOI] [PubMed] [Google Scholar]
  • 18. Lindseth F, Lango T, Bang J, Nagelhus Hernes TA (2002) Accuracy evaluation of a 3D ultrasound-based neuronavigation system. Computer aided surgery: official journal of the International Society for Computer Aided Surgery 7: 197–222. [DOI] [PubMed] [Google Scholar]
  • 19. Muns A, Meixensberger J, Arnold S, Schmitgen A, Arlt F, et al. (2011) Integration of a 3D ultrasound probe into neuronavigation. Acta neurochirurgica 153: 1529–1533. 10.1007/s00701-011-0994-5 [DOI] [PubMed] [Google Scholar]
  • 20. Rasmussen IA Jr, Lindseth F, Rygh OM, Berntsen EM, Selbekk T, et al. (2007) Functional neuronavigation combined with intra-operative 3D ultrasound: initial experiences during surgical resections close to eloquent brain areas and future directions in automatic brain shift compensation of preoperative data. Acta neurochirurgica 149: 365–378. [DOI] [PubMed] [Google Scholar]
  • 21. Sergeeva O, Uhlemann F, Schackert G, Hergeth C, Morgenstern U, et al. (2006) Integration of intraoperative 3D-ultrasound in a commercial navigation system. Zentralblatt fur Neurochirurgie 67: 197–203. [DOI] [PubMed] [Google Scholar]
  • 22. Solheim O, Selbekk T, Lindseth F, Unsgard G (2009) Navigated resection of giant intracranial meningiomas based on intraoperative 3D ultrasound. Acta neurochirurgica 151: 1143–1151. 10.1007/s00701-009-0395-1 [DOI] [PubMed] [Google Scholar]
  • 23. Brinjikji W, Cloft H, Lanzino G, Kallmes DF (2009) Comparison of 2D digital subtraction angiography and 3D rotational angiography in the evaluation of dome-to-neck ratio. AJNR American journal of neuroradiology 30: 831–834. 10.3174/ajnr.A1444 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Lee MC, Macdonald RL (2003) Intraoperative cerebral angiography: superficial temporal artery method and results. Neurosurgery 53: 1067–1074; discussion 1074–1065. [DOI] [PubMed] [Google Scholar]
  • 25. Sharma M, Ambekar S, Ahmed O, Nixon M, Sharma A, et al. (2014) The Utility and Limitations of Intraoperative Near-Infrared Indocyanine Green Videoangiography in Aneurysm Surgery. World neurosurgery 82: e607–e613. 10.1016/j.wneu.2014.05.033 [DOI] [PubMed] [Google Scholar]
  • 26. Hanggi D, Etminan N, Steiger HJ (2010) The impact of microscope-integrated intraoperative near-infrared indocyanine green videoangiography on surgery of arteriovenous malformations and dural arteriovenous fistulae. Neurosurgery 67: 1094–1103; discussion 1103–1094. 10.1227/NEU.0b013e3181eb5049 [DOI] [PubMed] [Google Scholar]
  • 27. Raabe A, Beck J, Seifert V (2005) Technique and image quality of intraoperative indocyanine green angiography during aneurysm surgery using surgical microscope integrated near-infrared video technology. Zentralblatt fur Neurochirurgie 66: 1–6; discussion 7–8. [DOI] [PubMed] [Google Scholar]
  • 28. Ozgiray E, Akture E, Patel N, Baggott C, Bozkurt M, et al. (2013) How reliable and accurate is indocyanine green video angiography in the evaluation of aneurysm obliteration? Clinical neurology and neurosurgery 115: 870–878. 10.1016/j.clineuro.2012.08.027 [DOI] [PubMed] [Google Scholar]
  • 29. Murai Y, Adachi K, Takagi R, Koketsu K, Matano F, et al. (2011) Intraoperative Matas test using microscope-integrated intraoperative indocyanine green videoangiography with temporary unilateral occlusion of the A1 segment of the anterior cerebral artery. World neurosurgery 76: 477 e477–477 e410. [DOI] [PubMed] [Google Scholar]
  • 30. Fischer G, Stadie A, Oertel JM (2010) Near-infrared indocyanine green videoangiography versus microvascular Doppler sonography in aneurysm surgery. Acta neurochirurgica 152: 1519–1525. 10.1007/s00701-010-0723-5 [DOI] [PubMed] [Google Scholar]
  • 31. Abla AA, Lawton MT (2014) Indocyanine green angiography for cerebral aneurysm surgery: advantages, limitations, and neurosurgeon intuition. World neurosurgery 82: e585–586. 10.1016/j.wneu.2014.08.019 [DOI] [PubMed] [Google Scholar]
  • 32. Dashti R, Hernesniemi J (2014) Intraoperative assessment of a quality of microneurosurgical clipping: role of indocyanine green videoangiography. World neurosurgery 82: e589–590. 10.1016/j.wneu.2014.08.060 [DOI] [PubMed] [Google Scholar]
  • 33. Schnell O, Morhard D, Holtmannspotter M, Reiser M, Tonn JC, et al. (2012) Near-infrared indocyanine green videoangiography (ICGVA) and intraoperative computed tomography (iCT): are they complementary or competitive imaging techniques in aneurysm surgery? Acta neurochirurgica 154: 1861–1868. 10.1007/s00701-012-1386-1 [DOI] [PubMed] [Google Scholar]
  • 34. Sindou M, Acevedo JC, Turjman F (1998) Aneurysmal remnants after microsurgical clipping: classification and results from a prospective angiographic study (in a consecutive series of 305 operated intracranial aneurysms). Acta neurochirurgica 140: 1153–1159. [DOI] [PubMed] [Google Scholar]
  • 35. Klopfenstein JD, Spetzler RF, Kim LJ, Feiz-Erfan I, Han PP, et al. (2004) Comparison of routine and selective use of intraoperative angiography during aneurysm surgery: a prospective assessment. Journal of neurosurgery 100: 230–235. [DOI] [PubMed] [Google Scholar]
  • 36. Derdeyn CP, Moran CJ, Cross DT, Grubb RL Jr., Dacey RG Jr (1995) Intraoperative digital subtraction angiography: a review of 112 consecutive examinations. AJNR American journal of neuroradiology 16: 307–318. [PMC free article] [PubMed] [Google Scholar]
  • 37. Chiang VL, Gailloud P, Murphy KJ, Rigamonti D, Tamargo RJ (2002) Routine intraoperative angiography during aneurysm surgery. Journal of neurosurgery 96: 988–992. [DOI] [PubMed] [Google Scholar]
  • 38. Meyer T, Kuss J, Uhlemann F, Wagner S, Kirsch M, et al. (2013) Autostereoscopic 3D visualization and image processing system for neurosurgery. Biomedizinische Technik Biomedical engineering 58: 281–291. 10.1515/bmt-2012-0079 [DOI] [PubMed] [Google Scholar]
  • 39. Ferrant M, Nabavi A, Macq B, Black PM, Jolesz FA, et al. (2002) Serial registration of intraoperative MR images of the brain. Medical image analysis 6: 337–359. [DOI] [PubMed] [Google Scholar]
  • 40. Ferrant M, Nabavi A, Macq B, Jolesz FA, Kikinis R, et al. (2001) Registration of 3-D intraoperative MR images of the brain using a finite-element biomechanical model. IEEE transactions on medical imaging 20: 1384–1397. [DOI] [PubMed] [Google Scholar]
  • 41. Munteanu C, Morales FC, Ruiz-Alzola J (2008) Speckle reduction through interactive evolution of a general order statistics filter for clinical ultrasound imaging. IEEE transactions on bio-medical engineering 55: 365–369. 10.1109/TBME.2007.897833 [DOI] [PubMed] [Google Scholar]
  • 42. Tay PC, Acton ST, Hossack JA (2011) A wavelet thresholding method to reduce ultrasound artifacts. Computerized medical imaging and graphics: the official journal of the Computerized Medical Imaging Society 35: 42–50. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

All relevant data are within the paper.