Prospects for measuring the 229Th isomer energy using a metallic magnetic microcalorimeter

GA Kazakov a,b,, V Schauer b, J Schwestka b, SP Stellmer b, JH Sterba b, A Fleischmann c, L Gastaldo c, A Pabinger c, C Enss c, T Schumm b
PMCID: PMC4376067  PMID: 25844000

Abstract

The Thorium-229 isotope features a nuclear isomer state with an extremely low energy. The currently most accepted energy value, 7.8±0.5 eV, was obtained from an indirect measurement using a NASA x-ray microcalorimeter with an instrumental resolution 26 eV. We study, how state-of-the-art magnetic metallic microcalorimeters with an energy resolution down to a few eV can be used to measure the isomer energy. In particular, resolving the 29.18 keV doublet in the γ-spectrum following the α-decay of Uranium-233, corresponding to the decay into the ground and isomer state, allows to measure the isomer transition energy without additional theoretical input parameters, and increase the energy accuracy. We study the possibility of resolving the 29.18 keV line as a doublet and the dependence of the attainable precision of the energy measurement on the signal and background count rates and the instrumental resolution.

Keywords: Thorium-229, Isomer energy, Gamma spectroscopy, Design of experiment

1. Introduction

The nuclear level scheme of the Thorium-229 isotope is expected to feature a long-lived isomer state, 229mTh, extremely close to the nuclear ground state. The most recent value for the isomer energy Eis, 7.8±0.5 eV, obtained from indirect measurements with a NASA x-ray microcalorimeter1 [1,2], is within the reach of modern optical laser spectroscopy and could serve as a “nuclear frequency standard” [3]. This standard could reach an uncertainty level of 10−19 [4], and provide a new powerful instrument for testing the stability of fundamental constants [5,6]. It has been shown that an ensemble of Thorium nuclei doped into a transparent crystal may demonstrate superradiance with a non-trivial emission dynamics [7], and may be used for building an ultraviolet (UV) laser [8]. Finally, the frequency shifts and broadenings produced by such a crystal lattice environment might be used in studies of material properties, as is commonly done in Mössbauer spectroscopy [3]. The necessary step towards all of these exciting applications is a direct observation and precise determination of the isomer state energy.

The existence of the low-energy state in the 229Th nucleus was first conjectured by Kroger and Reich based on studies of the γ-ray spectrum following the α-decay of Uranium-233 [9]. They concluded that this nucleus has a Jπ=3/2+ isomer level lying within 100 eV above the Jπ=5/2+ ground state level. The development of high quality germanium detectors (resolution from 300 to 900 eV) allowed Helmer and Reich to measure more precise γ-energies in 1989–1993 and to predict the energy of the nuclear transition to be Eis=3.5±1.0eV, placing it into the range of optical frequencies [10]. The decay pattern and combinations of transitions used by Helmer and Reich are presented in Fig. 1(a).

Fig. 1.

Fig. 1

Partial level schemes of the 229Th nucleus with decay paths and energies (all in keV). Boxes in each panel denote the energy combinations used to derive Eis in the “indirect” methods discussed in the main text. (a) According to Helmer and Reich [10]; (b) according to Beck et al. [1,2], the interband transitions (dashed arrows) are taken into account; (c) approach discussed here using a high-resolution (Δinst39eV) microcalorimeter to resolve the 29.1 keV doublet (first proposed in Ref. [31]); (d) direct detection of “nuclear light” (many unsuccessful attempts [11–14] and new proposal [17]). Schemes (a), (b), and (c) are indirect measurements, involving keV energy transitions whereas scheme (d) is direct, only measuring the isomer transition of a few eV energy.

This unnaturally low value of Eis triggered a multitude of investigations, both theoretical and experimental, trying to determine the transition energy precisely, and to specify other properties of the Jπ=3/2+ excited state (such as lifetime and magnetic moment). However, searches for direct photon emission from the low-lying excited state performed in the late 1990s [11,12] have failed to observe a signal [13,14]. In 2005, Guimãraes-Filho and Helene re-analyzed the data of Helmer and Reich, taking into account new information about the nuclear decay pattern [15]. They derived Eis=5.5±1.0eV.

In 2007, a cryogenic NASA x-ray microcalorimeter with instrumental resolution Δinst from 26 to 30 eV (FWHM) allowed Beck et al. [1] to perform a new indirect measurement of Eis, involving lower energy nuclear states, as depicted in Fig. 1(b). In this measurement, the obtained transition energy (7 eV) was corrected by accounting for the theoretical branching ratios 29.19keVTh229g estimated as 1/13, and 42.43keVTh229m estimated as 2% in Ref. [2]. This correction yields the currently most accepted value Eis=7.8±0.5eV, now placing the transition into the vacuum UV range (160nm).

In the experiments described above [1,9,10] the isomer transition energy Eis is not measured directly but is derived from the spectrum of higher-energy (keV) γ-radiation of a spontaneously decaying 233U source. We will refer to these measurements as indirect passive. Possible alternatives are direct passive and active approaches.

In the direct passive schemes (Fig. 1(d)), the aim is to perform spectroscopy of the ultraviolet radiation emitted from the isomer appearing in the α-decay of 233U (2% of the nuclei decay is expected to lead into the isomer state). This method has two main difficulties: a relatively high false count rate caused by the Uranium sample radioactivity, and a high probability of non-radiative decay (quenching) of the isomer state in neutral Thorium atoms (up to 109 times higher than the radiative decay rate [16]). To overcome these problems, it was proposed in Ref. [17] to extract α-recoil Thorium ions ejected from an Uranium sample, and collect them in a small spot on a MgF2 coated surface to minimize the quenching rate. Vacuum ultraviolet spectroscopy of the emitted fluorescence radiation may then allow to measure the isomer transition energy.

On the contrary, in active approaches, Thorium nuclei (in the ground state) will be illuminated by tunable radiation to excite them to the isomer state. In the solid-state approach a macroscopic (10121018) number of Thorium ions doped into UV transparent crystals can be excited, for example, by synchrotron radiation, and the emerging fluorescence signal can be studied [18–20]. Apparent advantage of this approach is the huge number of simultaneously excited nuclei. At the same time, crystal fluorescence can cause difficulties in identifying the Thorium isomer transition, and various crystal effects can hamper the precise determination of Eis. Another approach is the spectroscopy of trapped Thorium ions. At PTB, Germany, work is under way to excite nuclei of Th+ ions into the isomer state using a two-photon scheme, exploiting the electronic bridge mechanism [21,22]. In Georgia University of Technology, USA, the laser manipulation of Th3+ ions is under investigation [4,23]. Detection of the excitation of the Thorium into the isomer state may be based on a change of the electronic hyperfine structure [3]. Studies of the hyperfine structure of Thorium are also performed at the IGISOL facility in Jyväskylä, Finland, in collaboration with a group of the University of Mainz, Germany [24].

We should also mention a number of studies aimed to measure the lifetime of the isomer state without a determination of Eis. In Ref. [25], the half life of the isomer state for a bare nucleus was derived theoretically based on the calculations of the matrix element of the nuclear magnetic moment and on the experimental data concerning transitions at higher energies. They predict a half-life of T1/2=(10.95h)/(0.025E3) for the isomer transition, where E is given in eV, which yields T1/2=55min for E=7.8 eV. Direct measurements of this lifetime were performed in several groups [26–28]. The obtained results vary from 2 min [26] to 6 h [28]. This discrepancy may be explained either by an incorrect interpretation of the observations [29] or by a difference in chemical composition of the Thorium resulting in different internal conversion rates.

All active approaches and eventually all nuclear spectroscopy applications require irradiation of the sample with some external narrow-band tunable radiation, and study of the emerging fluorescence. The estimation of the error σ=0.5eV on the isomer energy presented in Ref. [1] corresponds to one standard deviation, therefore it is necessary to scan the excitation source over 2 eV (±2σ) to find the transition with 95% probability, or over 3 eV (±3σ) to find the transition with 99.7% probability. Sakharov re-estimated the influence of the uncertainty of the 29.39 keV peak on the isomer energy derivation in Ref. [1] and obtained an error of 1.3–1.5 eV [30]. Moreover, an analysis of more recent experimental data led him to claim that the energy of the isomer state can be anywhere in the range 0–15 eV, if the isomer state exists at all.

We believe it will be technically difficult, if not impossible, to cover such a broad energy range with a tunable narrow-band source of ultraviolet radiation in a reasonable time. We therefore propose to first increase the energy resolution on Eis by an improved indirect measurement compared to Ref. [1]. As we show below, it appears possible to resolve the 29.18 keV doublet [31] presented in Fig. 1(c) with todays state-of-the-art x-ray spectrometers. Resolving this doublet would significantly increase confidence in the existence of the isomer state. Moreover, the isomer energy would be measured without additional theoretical input parameters like branching ratios, etc. The aim of this study is to investigate the possibility of resolving the 29.18 keV line clearly as a doublet over a broad range of values for the isomer energy splitting and the branching ratio, and to analyze the precision that can be obtained on Eis depending on the relevant experimental parameters.

2. Statistical aspects of the envisioned experiment

The operation principle of high-resolution x-ray microcalorimeters is to detect the heat deposited by an x- or γ-ray interacting with an absorber, using a very sensitive thermometer. Interaction with the absorber material mainly proceeds through the photoeffect. The energy of the produced photoelectron as well as Auger electrons together with their thermalization cascade should be effectively deposited within the volume of the absorber [32, 33]. On the other hand, the absorber should have a small heat capacity Ca for good instrumental energy resolution Δinst. Various microcalorimeters differ in geometry, absorber material, sensor, etc., which leads to different energy resolutions, stopping powers, total detector surfaces etc. Many of these parameters are connected and cannot be optimized independently. For example, increasing the size of absorber increases the solid angle and/or stopping power but degrades the instrumental energy resolution. Finally, we note that after a detection event, dissipation of the deposited heat leads to a detector-specific dead time, during which the energy of a successive photon cannot be measured correctly. Therefore it is impossible to improve the precision of the measurements infinitely simply by using a more active sample, or by placing the sample very close to the detector. The total count rate can be reduced using a designed filter which will primarily absorb photons outside the 29.18 keV region of interest.

This study aims to answer two questions: how does the possibility to resolve the 29.18 keV peak as a doublet depend on the experimental parameters, and how to attain the most precise determination of the isomer transition energy Eis. As outlined above, parameters of the experimental setup can be controlled to some extent only. For a proper design of suitable detectors and experimental configurations, we analyze how the key parameters affect these two points.

2.1. Specification of the problem and statistical model

The model employed for the statistical study should not contain too many parameters to make it accessible to a multi-factor analysis. On the other hand, it should be sufficiently comprehensive for a realistic feasibility study. For the sake of convenience, we assume a fixed total measurement time of t=106s, approximately 11 days, which corresponds to the total time of the successful measurement in Ref. [1]. Also we suppose that the background count rate near the 29.18 keV doublet is flat and symmetric, and that the monoenergetic line has Gaussian shape with full width at half maximum equal to Δinst (see Section 4 for a discussion of these approximations).

The considered total energy interval (0–70 eV) is subdivided to a set of 0.4 eV bins (approximately a factor 10 below the expected instrumental energy resolution). The number of counts in the ith energy bin is a Poissonian random number ni with a mean value λi equal to

λi=d·R29·tσ2π[(1b)exp((EiE1)22σ2)+bexp((EiE2)22σ2)]+d·rbg·t, (1)

where R29 is the signal count rate, rbg is a specific rate of background counts per 1 eV energy interval, E1, E2 are the centers of lines of the components of the 29.18 keV doublet, σ=Δinst/22ln(2), d=0.4 eV is one energy bin, Ei is the mean energy of the ith bin, and b is the branching ratio. The set {n1,,nN} of experimental data can be represented as a position vector n in an N-dimensional “sample space” (N=175).

For a given sample n, we perform a nonlinear regression fit by a vector function f with N components

fi=d·ts2π[J1exp((EiE1˜)22s2)+J2exp((EiE1˜E˜is)22σ˜2)]+d·r˜bg·t. (2)

This fit has 6 free parameters {J1,J2,E1˜,E˜is,r˜bg,σ˜}={θ1,,θ6}θ. For the estimation of these parameters, we use the maximum likelihood method. The likelihood function L(n|θ) is the probability for realizing the set n, if true mean values λi are equal to fi(θ). We also introduce the logarithmic likelihood function

(n|θ)=logL(n|θ)=i=1N[nilogfi(θ)log(ni!)]. (3)

2.2. Resolving the 29.18 keV line as a doublet

As it was outlined above, resolving the 29.18 keV line as a doublet would significantly reduce the doubts [30] in the existence of the isomer state in the 229Th nucleus. Our first aim is to discuss the feasibility of such an identification depending on experimental parametes and the (yet unknown) values of the isomer energy splitting Eis=E2E1 and the branching ratio b.

To check whether a given set of experimental data corresponds to a single line or to a doublet, one can apply the likelihood ratio test [34–36]. The essence of this test and the method to estimate the significance level is described in Appendix A. Here we define the significance level as the probability to identify incorrectly the single peak as a doublet or the doublet as a single peak using the likelihood ratio test in the situation when we have either a single peak or a doublet whose parameters are specified. Figs. 2 and 3 show level curves for the signal count rates R29 which are necessary to attain a significance level of 1% for various values of the instrumental resolution Δinst and the R29/rbg ratio, in the (b,Eis) plane. It is interesting to note that the optimal energy resolution is attained when b0.25 and not for equally strong components of the doublet. This is explained by the fact that such a branching ratio leads to a noticeable asymmetry of the peak which facilitates the identification of a second component, increasing b leads to a reduction of the main peak at fixed total signal count rate R29.

Fig. 2.

Fig. 2

Curves of constant levels of signal count rate R29 (in mHz) required to resolve the 29.18 keV line as a doublet at 1% significance level for different values of the detector resolution Δinst and the signal-to-noise ratio R29/rbg at 106 s of measurement time. The red spot corresponds to the area of the branching ratio b and isomer transition energy Eis according to Ref. [2]. (For interpretation of the references to color in this figure caption, the reader is referred to the web version of this paper.)

Fig. 3.

Fig. 3

Curves of constant levels of signal count rate R29 (in mHz) required to resolve the 29.18 keV line as a doublet at 1% significance level for different values of the detector resolution Δinst and the signal-to-noise ratio R29/rbg at 106 s of measurement time. The red spot corresponds to the area of the branching ratio b and isomer transition energy Eis according to Ref. [2]. (For interpretation of the references to color in this figure caption, the reader is referred to the web version of this paper.)

2.3. Precision in the determination of Eis: Monte-Carlo simulations

The aim of the proposed spectroscopy study is not only to resolve the 29.18 keV line in the γ-spectrum of 233U as a doublet but to determine the energy splitting with maximum precision.

We study the standard deviation

δEis=(E^isEis)2 (4)

of the isomer energy as a characteristic measure of precision (in the following we call δEis the uncertainty of Eis). Here and below, angular brackets denote expectation values, Eis and E^is denote “true” and measured values of the isomer transition energy respectively. For the sake of brevity, we suppose that the true values of the energy splitting Eis and of the branching ratio b are equal to 7.8 eV [2] and 1/14 [10] respectively.2 As before, we assume a total measurement time t=106 s. Scaling the results to other values of Eis, b, or t is straightforward.

To investigate the dependence of δEis on the instrumental energy resolution Δinst, the signal count rate R29, and the specific background count rate rbg, we perform a Monte-Carlo study of δEis. For any set of parameters, we simulate the sample n as shown in Fig. 4, and estimate the parameters θ={J1,J2,E1˜,Eis,r˜bg,s} maximizing the sum (3). We repeat this procedure 104 times and calculate δEis according to Eq. (4).

Fig. 4.

Fig. 4

Examples of Monte-Carlo simulated “experimental data” (black dots) and fit (red curves) for Δinst=3eV (a), 6 eV (b) and 9 eV (c). Other parameters are: R29=7.74mHz, rbg=3.9μHz/eV, t=106s. Plots are depicted in a “square-root scale” where the Poissonian noise is mapped onto signal-independent deviations. The origin of the energy axis is chosen arbitrarily. (For interpretation of the references to color in this figure caption, the reader is referred to the web version of this paper.)

In Fig. 5 we present curves of constant level of δEis in the plane (R29, rbg) obtained from the Monte-Carlo simulation. Finally, in Fig. 6 we present the curves of constant δEis in the plane (R29,Δinst) for fixed ratios R29/rbg. One can see that improving the instrumental resolution by 1 eV increases the precision on the determination of δEis by the same amount as doubling the 29.18 keV signal count rate, or doubling the measurement time t.

Fig. 5.

Fig. 5

Curves of constant δEis labeled in eV in the (R29,rbg) plane for different values of Δinst and 106 s of total measurement time.

Fig. 6.

Fig. 6

Curves of constant δEis labeled in eV in the (R29,Δinst) plane for different values of R29/rbg and 106 s of total measurement time.

3. Experimentally attainable count rates and expected precision

In this section we estimate the attainable count rates and the resulting precision in a measurement of the isomer energy Eis that can be achieved with a state-of-the-art high-resolution microcalorimeter. In particular we consider the metallic magnetic microcalorimeter maXs-20 as described in Ref. [33]. The aim is to demonstrate, that valuable results can be obtained with currently available technology. In Section 5 we describe ongoing work towards a more refined, dedicated detector setup.

The maXs-20 microcalorimeter consists of 8 detector elements (pixels) each of which has an absorber plate for incoming radiation (250×250×5μm3 Au plate) connected to a 160×160×1.3μm3 paramagnetic temperature sensor (Er-doped Au) through 24 gold stems (10μm diameter and 5μm height each). Each sensor is connected to a thermal bath, the system is installed in a cryostat operating at a temperature of about 30 mK. Energy deposited into the absorber plate heats the paramagnetic sensor and causes a change of its magnetization in an external magnetic field. Measuring this change in magnetization using SQUIDS, it is possible to determine the amount of absorbed energy and hence the energy of the incoming x- or γ-rays. Note that after the detection of an x- or γ-ray, the individual pixel can only detect again after a certain relaxation time of the order of 100 ms. Therefore the total count rate RT should not be too high.

As a sample, we consider 1 mCi of 233U electrodeposited as a film onto a metal planchet with a radius of R=10 mm. We assume the sample to be situated 40 mm from the detector (outside the cryostat) with a total surface s=0.5 mm2. Also we suppose the presence of additional material related to the cryostat vacuum system (sealing, input window of the cryostat, some other intentionally positioned shielding, etc.), which we refer to as filters.

According to the NuDat 2.6 database [37], each single decay of 233U is accompanied (on average) by one αparticle with an energy from 4.309 to 4.824 MeV, 0.213 conversion electrons with energies from 2.3 to 600 keV (97.7% of the electrons have energies below 50 keV), and 0.0544 photons most of which (0.052 per decay) are L-shell x-rays with a mean energy of 13 keV. As no individual L x-rays are listed in NuDat 2.6 (only average energy and total intensity), we have taken the lacking data from the X-Ray Data Booklet [38].

To estimate the detector count rates, we suppose that all αparticles and electrons emitted from the Uranium sample are stopped by the sample itself or by the filter materials, therefore, we consider only x- and γ-rays. Also it is supposed that all secondary electrons and photons generated in the filters are absorbed in the material locally. This assumption is correct for relatively thick filters made from light materials like Aluminium.

We take into account absorption of the photons within the sample itself, the filters, and the detector. The total count rate RT is

RT=i=1imaxA·Ω4π·Ii1eUaU(Ei)UaU(Ei)×efaf(Ei)·(1eAuaAu(Ei)). (5)

Here the sum is taken over all photon energies Ei, A is the activity of the Uranium sample, Ii is the relative intensity (quantum output) of photons with the energy Ei per single decay event, Au and f are the thicknesses of the gold absorber and filters respectively, Ω is a solid angle. Linear absorption coefficients aκ(Ei) (κ=U,Au,f) were taken from the XCOM Photon Cross-Sections Database [39]. The count rate R29 of signal photons is

R29=i=1,2A·Ω4πIi1eUaU(Ei)UaU(Ei)efaf(Ei)×[1eAuaAu(Ei)ϖXIXPes(Ei,EX,Au)], (6)

where i=1,2 corresponds to the two components of the doublet, ϖ=0.331 is the Au L shell fluorescence yield [40], IX is the probability that an energy of a fluorescence photon emitted by a Au atom is equal to EX, and Pes(Eγ,EX,Au) is a probability that an incoming γ-quant with energy Ei will be absorbed, and an x-ray photon following this absorption leaves the absorber (escape line). Supposing an isotropic spatial distribution of these secondary photons, we obtain

Pes=120aγeaγx(0π/2exp[aX(x)cosθ]sinθdθ+π/2πexp[aXxcosθ]sinθdθ)dx, (7)

where =Au, aγ=aAu(Eγ), aX=aAu(EX). Also we suppose that the x-ray photon is emitted from the L shell, i.e. the deepest shell that is accessible by energy conservation, and the probability IX for emission of the photon is the relative intensity tabulated in Ref. [39] normalized to the sum of relative intensities from the L shell.

Evaluating the expressions (5)–(7) yields a total detector count rate RT=1Hz and a signal count rate of R29=7.74mHz for a 1.3 mm thick Aluminium filter. Without any filter, the count rates for the same parameters are: RT=13.6Hz, and R29=11.78mHz. We see that the Aluminium filter absorbs approximately 92% of all photons emitted from the sample, but only about 34% of the signal photons. We conclude that filtering is an effective method to decrease the total count rate RT, caused mainly by low-energy Thorium L shell x-ray.

The background count rate rbg is caused by the escape of some fraction of the dissipated energy of γ-quanta absorbed in the detector. In Ref. [1], the number of background counts close to the 29.18 keV doublet was about 30–40 events per 3 eV bin whereas the total number of counts in the 29.18 keV peak was about 2.7×104 events. This yields the ratio R29/rbg=2000eV. Assuming that a similar ratio can be realized with the maXs-20 detector, we find that the uncertainty δEis on the measured isomer transition energy Eis will be equal to 0.06 eV for an instrumental resolution of Δinst=3eV, signal count rate R29=7.74mHz and total measurement time t=106s. Therefore the proposed experiment to determine the isomer energy Eis is expected to be almost one order of magnitude more precise than the results obtained in the previous experiment [1]. Reducing the experimental resolution to 6 eV and 9 eV yields δEis=0.19eV and δEis=0.56eV respectively. Increasing the total measurement time to t=2.6×106s, 1 month, we can measure the isomer transition energy with an uncertainty δEis=0.037eV for Δinst=3eV, δEis=0.12eV for Δinst=6eV, or δEis=0.33eV for Δinst=9eV.

4. Further statistical aspects

We are aware of certain simplifications and assumptions in the above analysis. Here we briefly resume some additional issues that could arise, a detailed discussion of these points is beyond the scope of this work.

First, the shape of signal peaks can deviate from Gaussian. For example, a long low energy tail on the spectral lines may lead to the appearance of a noticeable step in the background count rate (see, for example, Fig. 2(a) in Ref. [1]). We believe that in the work of Ref. [1] this effect is caused mainly by the escape of energy from the absorber material, for example in the form of a thermal phonons [33]. The yield of Compton scattering is not sufficient to explain this step, see Appendix B. To take this effect into account correctly, we will have to modify our model of the background. A more difficult situation arises when the escaping energy is relatively small, which would lead to an asymmetry of the line rather than the appearance of a tail. In this case, it would be useful to study an isolated single peak separated from the doublet of interest but intense enough to give good statistics, and/or to perform an independent study with another γsource, e.g. 241Am.

Another issue that may appear is a slow time-dependent fluctuation of the response function caused by an uncontrollable drift of ambient magnetic fields and/or cryostat temperature over the duration of the measurement. We believe we can suppress such drifts below 10 eV by temperature stabilization and mu-metal shielding of the setup. Additionally, we will monitor the position of a series of reference x- and γ-ray lines for a correct tracing of this drift, realizing a time-dependent calibration of the detector. Note that x-ray lines generally have a much broader linewidth than γlines [42] which simplifies the identification. An auxiliary calibration source, for example 241Am, can help to enhance the quality of this calibration.

Also we should mention possible interference of the 29.18 keV doublet with coincidence and escape lines of x- and γ-rays of 223U and other elements present in the sample. We plan to study the composition of the sample using “ordinary” low-precision γ-spectrometry.

5. Planned experimental implementation

5.1. Detector development

We are currently developing a dedicated new detector for the measurement described above, to some extend interpolating between the maXs-20 (0–20 keV) and the maXs-200 (0–200 keV) series [33]. It will feature a linear array of magnetic calorimeters, each with an active area of 250×250μm2. We will increase the absorber thickness by a factor of 2–3 in comparison with 5μm in the maXs-20, leading to a stopping power of about 50% at 30 keV. We will operate the detectors in a dry 3He/4He-dilution refrigerator at about 20 mK. In this situation, the intrinsic energy resolution of the detector caused by thermal noises of all kinds is expected to be below 2 eV (FWHM). According to the calorimetric detection principle of metallic magnetic calorimeters [41], this resolution is independent of energy as long as the total gain (including operational temperature and external magnetic fields) is stable, the dependence of the detector response on the event position in the absorber is negligible, and the photon energy E is still small enough to be within the range of linear detector response Φ(E). Also we expect that the minimal time between two correctly measurable counts in a single detector element (pixel) will be about 100 ms.

So far, we have achieved resolving powers up to about E/Δinst=3700 (corresponding to 1.6 eV (FWHM) at 5.9 keV) with our maXs-20 devices (unpublished), being limited by a combination of both, instabilities of the operating temperature and a position dependence. We believe that we can improve the short-term stability of the total gain and keep the position dependence of absorption events small enough to allow resolving powers beyond 104 in the planned experiment. Also, the response of the present maXs-20 detector to photon energies below 60 keV has a small quadratic deviation from a linear behavior, see Fig. 7. At an energy of 30 keV this deviation is only about 3%, which yields a 6% degradation of the intrinsic energy resolution compared to the low-energy signals, i.e. below 2.12 eV on an absolute scale.

Fig. 7.

Fig. 7

Detector response versus the energy of absorbed photons (a) and deviation from the linear behavior (b) based on the measurement of 3 characteristic lines of an 241Am γ-spectrum (c).

5.2. Sample preparation and characterization

The sample should ideally consist of isotopically pure 233U to avoid a too high count rate not carrying relevant information and possible interference with the 29.18 keV Thorium doublet signal. For this project, we have 560 mg (about 5 mCi activity) raw sample material available (in oxide powder form). The origin and preparation procedure of this material is unknown, from the γ-spectrum we suspect that originally, 232Th has been activated in a high-flux neutron reactor and the 233U has been separated chemically. A mass spectrum of the raw material, produced by an in-house ICP-MS can be seen in Fig. 8. The raw material contains >90% 233U, together with traces of 232U, 234U, 235U, 238U, and the decay product 229Th. Further daughter products of the 233U chain has not been detected.

Fig. 8.

Fig. 8

Left: inductively coupled plasma mass spectrometer (ICP-MS) data of the raw 233U material composition. Mass signals above 240 amu are molecular fragments and can be ignored. Right: photo of the electroplated UO2 test sample (with 238U).

To further purify the sample, we will perform a PUREX Uranium extraction procedure. We have also observed an efficient additional element separation in the electrodeposition process. In-house analysis using γ- and α-spectroscopy, ICP-MS, and neutron activation analysis will allow us to quantify the success of this procedure and finally know the exact composition of the final measurement sample.

The sample will be produced by electroplating 233U from a liquid solution onto a stainless steel or aluminium planchets. The target activity of 1 mCi corresponds to 104 mg of pure 233U or 118 mg of UO2. Producing correspondingly thick films (1520μm) turned out to be difficult in electroplating [43,44]. We have therefore developed a process to deposit up to 20 mg Uranium onto stainless steel or aluminium foils of only 1050μm thickness. These samples can easily be stacked to realize the target activity without the carrier foils significantly reducing the count rate in the 29.18 keV peak.

6. Conclusion

We have analyzed the feasibility of an indirect measurement of the low-energy isomer state in 229Th using a high-resolution magnetic microcalorimeter. We propose to resolve the 29.18 keV doublet in the γ-radiation spectrum following the α-decay of 233U. Such a measurement would provide a strong indication for the existence of the isomer state and improve the accuracy on the energy measurement significantly. The measurement appears feasible with currently available detector technology and samples.

Acknowledgments

We thank C. Streli, P. Wobrauschek, and J. Seres for discussions and calculations concerning γ-ray filters. We thank S. Smolek and M. Gollowitzer for preliminary tests and characterizations of 233U samples. This work was supported by the ERC project 258604-NAC, the FWF Project Y481, the WPI Thematic Program “Tailored Quantum Materials”, and the FWF project M1272-N16 “TheoNAC”.

Footnotes

This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

1

We will refer to these devices as “x-ray” spectrometers, corresponding to their primary field of application. In the measurements described here, they detect both, x-rays and γ-rays.

2

According to Ref. [1], b=1/13 has 8% error, therefore the value b=1/14 can be considered as a consistent but slightly pessimistic (from the point of view of spectral resolution) estimation.

Appendix A. Likelihood ratio test and estimation of the significance level a

Here we describe the essence of likelihood ratio tests for regression models in the simple case of normally distributed observables with known dispersions, and the method which we actually used for the estimation of the significance level attainable in the experiment.

Let us have N experimental observables yi normally distributed around their (unknown) expectation values λi with known dispersions σi. Without loss of generality, we can set σi=1 for all i. Also, we have 2 regression models, one of which (short model) is a particular case of another one (long model). In the long model, it is supposed that the expectation values of observations yi are some known functions fi(θ) of the l-dimensional parameter θ={θ1,,θl}. In the short model, it is supposed also that the parameters θ has some additional restrictions, and the short model has s=lr degrees of freedom. It is supposed that the long hypothesis anyhow is correct. The question we want to answer is whether the short hypothesis is correct? Or, more precisely: how plausible (or unplausible) is it to obtain the set of observables yi, if the short model is correct?

The likelihood ratio test is a powerful method to answer this question. To illustrate the essence of this test, let us represent N observations yi as a point y in the N-dimensional Euclidean sample space. N functions fi(θ1,,θl) form an ł-dimensional surface C corresponding the long hypothesis. In turn, these functions with additional conditions corresponding the short hypothesis form the s-dimensional surface S within C, see Fig. 9. We suppose that these surfaces are sufficiently smooth. In the case of linear regression models, the functions fi(θ) are linear, and the surfaces C and S are just hyperplanes.

Fig. 9.

Fig. 9

Sketch of the sample space: C and S are the long and short hypotheses surfaces respectively. Point y=f(θ) corresponds to the true value of parameters, S is an s-dimensional surface passing through y whose points are equidistant to S. y is an experimental point, L=f(θL), and S=f(θS) corresponds the best fits within the long and short hypotheses respectively. The non-centrality parameter is the square of the distance a between S and S.

For normally distributed observables yi with zero mean and unit dispersion and for some set of parametes θ, the logarithmic likelihood function is

(y|θ)=const12i=1N(fi(θ)yi)2. (8)

Therefore the maximization of (θ) is equivalent to a minimization of the distance between the points f(θ) and y in the sample space. The square of this distance we denote as |fθy|2. Let L=f(θc) and S=f(θs), where θc and θs are the best likelihood estimations of parameters θ within the long and short hypotheses respectively. It is easy to see that |yS|2|yL|2+|LS|2 which yields |LS|2=2((y|θc)(y|θs)). In turn, |LS|2 is a sum of squares of r normally distributed random values κα with unit dispersion and different means μα. Therefore, |LS|2 is a non-central χ2 random value with r degrees of freedom and non-centrality parameter a2=α=1rμα2. For the sake of brevity, we denote this random value as χr2(a2). It is easy to see that a is just a distance between the short hypothesis surface S and the point y=f(θ) corresponding to the true value θ of parameters θ, see Fig. 9. If the short hypothesis is true (it corresponds the situation when the surface S coincided with S in Fig. 9), then a=0 and |LS|2 is just a “usual” χ2r random value with r degrees of freedom.

To test whether the short hypothesis is true or not, one should choose some desirable significance level α, and compare the value

D=|LS|2=2((n|θc)(n|θs)), (9)

with some critical value λα(χr2) such, that the probability

P(χr2>λα(χr2))=α. (10)

If D>λα(χr2), the short hypothesis is rejected on significance level α, otherwise it is accepted. The probability to reject the short hypothesis incorrectly is equal to α. On the other hand, the probability to accept the short hypothesis incorrectly is P(χr2(a2)<λα(χr2)). Therefore if the non-centrality parameter a2 will be larger than ar2(α) such that

P[χr2(ar2(α))<λα(χr2)]=α, (11)

the probability to accept the short hypothesis falsely will be less than α.

In our case, the observables ni are not normal random values with known dispersion but Poissonian random values. To estimate the possibility to identify the 29.18 keV line as a doublet, we consider ni as observables and approximate their distribution function by a normal distribution with mean λi and dispersion 1/4. This approximation is not very precise (the bias is about 20% for λ=1), but it seems to be applicable for a coarse estimation. Then, for specific values of b, Eis t, Δinst, R29, and rbg we calculate the set λi according to Eq. (1), and fit it by a vector function whose components are given by the square root of Eq. (2) with additional restriction J2=0. Then we calculate the non-centrality parameter

a2=4i=1N(fi(θs)λi)2. (12)

Our short hypothesis (single peak fit) has 4 degrees of freedom whereas the long hypothesis (double peak fit) has 6 degrees of freedom which yields r=2. For significance level α=0.01, conditions (10) and (11) are fulfilled at λ0.01(χ22)=2log(100)9.21, and a22(0.01)27.4. The values of R29 necessary to attain a2=27.4 for given values of the other parameters were calculated numerically and presented in Figs. 5 and 6.

Appendix B. The role of Compton scattering in the formation of an asymmetry or low-energy tail in absorption lines

In Ref. [1] the 29.18 keV line shows a low-energy tail, causing a noticeable step of about 10 counts per 3 eV bin (see Fig. 2(a) in Ref. [1]). This step can be caused either by the absorption of photons which had 29.18 keV energy initially, but lost some fraction of it in the source or filter material, or by the escape of some energy fraction from the absorber. The first scenario could be explained by the Compton “almost forward” scattering, when a photon loses tiny parts of its energy, about a few eV. The Klein–Nishina differential cross-section of the Compton scattering into the elementary solid angle is

dσdΩ=re22(EγEγ)2[EγEγ+EγEγsin2θ], (13)

where Eγ and Eγ are the energies of the photon before and after the Compton scattering respectively, re=e2/(mec2) is the classical electronic radius, and θ is the scattering angle. Using a well-known relation between Eγ and θ, it is easy to express the differential cross-section in units of Eγ

dσdEγ=πre2mec2Eγ2(EγEγ)2[EγEγ+EγEγsin2θ]. (14)

Note that Eγ=Eγ20eV corresponds to θ8.8°. For such small θ, Eq. (14) yields approximately

dσdEγ2πEγ2re2mec23×1028cm2/eV. (15)

The differential probability for the 29.18 keV photon to be scattered into the 1 eV range can be coarsely estimated as

dPdEγ=dσdEγ[zUnUU2+zfnff], (16)

where zi, ni, and i are the nuclear charge number, the atomic density, and the thickness of the Uranium (i=U) or filter (i=f) material respectively.

In the experiment [1], the Uranium activity of one 19 mm diameter planchet was about 0.02 mCi which corresponds to U0.3μm. It was covered by a Titanium foil (“filter”) with f=50.8μm. An estimation following Eq. (16) yields dP/dEγ1.9×106eV1. The peak corresponding to the absorption of 29.18 keV photons has Gaussian shape with about 3×103 events per 3 eV bin height in the maximum, and 26 eV FWHM, see Fig. 2(a) in Ref. [1]. This corresponds to a total number of counts forming this peak of about 3×104. Therefore, Compton scattering of 29.18 keV photons in the source or filter material produces only 0.053 events per 1 eV interval, or 0.16 events per 3 eV bin. This value is significantly lower than the observed step of 10 counts per 3 eV bin. A similar estimation for a U=15μm layer of Uranium and f=1.3mm of Aluminium filter (parameters of the planned experiment) leads to dP/dEγ3.16×105eV1.

Note that in Eq. (16) we treat all electrons as free. A more accurate estimation requires the substitution of the incoherent scattering function instead of z but this function does not exceed z, see Ref. [45] for details. Therefore, a more accurate calculations can only decrease the contribution of Compton scattering to the step of the background count rate.

References

  • 1.Beck B.R., Becker J.A., Beiersdorfer P., Brown G.V., Moody K.J., Wilhelmy J.B., Porter F.S., Kilbourne C.A., Kelley R.L. Physical Review Letters. 2007;98:142501. doi: 10.1103/PhysRevLett.98.142501. [DOI] [PubMed] [Google Scholar]
  • 2.B.R. Beck, C.Y. Wu, P. Beiersdorfer, G.V. Brown J.A. Becker, K.J. Moody, J.B. Wilhelmy, F.S. Porter, C.A. Kilbourne, R.L. Kelley, Improved value for the energy splitting of the ground-state doublet in the nucleus 229Th, in: 12th International Conference on Nuclear Reaction Mechanisms, Varenna, Italy, LLNL-PROC-415170, 2009.
  • 3.Peik E., Tamm C. Europhysics Letters. 2003;61:181. [Google Scholar]
  • 4.Campbell C.J., Radnaev A.G., Kuzmich A., Dzuba V.A., Flambaum V.V., Derevianko A. Physical Review Letters. 2012;108:120802. doi: 10.1103/PhysRevLett.108.120802. [DOI] [PubMed] [Google Scholar]
  • 5.Litvinova E., Feldmeier H., Dobaczewski J., Flambaum V. Physical Review C. 2009;79:064303. arXiv:0901.1240. [Google Scholar]
  • 6.Berengut J.C., Dzuba V.A., Flambaum V.V., Porsev S.G. Physical Review Letters. 2009;102:210801. doi: 10.1103/PhysRevLett.102.210801. [DOI] [PubMed] [Google Scholar]
  • 7.Liao W.-T., Das S., Keitel Christoph H., Pálffy A. Physical Review Letters. 2012;109:262502. doi: 10.1103/PhysRevLett.109.262502. [DOI] [PubMed] [Google Scholar]
  • 8.Tkalya E.V. Physical Review Letters. 2011;106:162501. doi: 10.1103/PhysRevLett.106.162501. [DOI] [PubMed] [Google Scholar]
  • 9.Kroger L.A., Reich C.W. Nuclear Physics A. 1976;259:29. [Google Scholar]
  • 10.Helmer R.G., Reich C.W. Physical Review C. 1994;64:1845. doi: 10.1103/physrevc.49.1845. [DOI] [PubMed] [Google Scholar]
  • 11.Irwin G.M., Kim K.H. Physical Review Letters. 1997;79:990. [Google Scholar]
  • 12.Richardson D.S., Benton D.M., Evans D.E., Griffith J.A.R., Tungate G. Physical Review Letters. 1998;80:3206. [Google Scholar]
  • 13.Shaw R.W., Young J.P., Cooper S.P., Webb O.F. Physical Review Letters. 1999;82:1109. [Google Scholar]
  • 14.Utter S.B., Beiersdorfer P., Barnes A., Lougheed R.W., Crespo López-Urrutia J.R., Becker J.A., Weiss M.S. Physical Review Letters. 1999;82:505. [Google Scholar]
  • 15.Guimãraes-Filho Z.O., Helene O. Physical Review C. 2005;71:044303. [Google Scholar]
  • 16.Karpeshin F.F., Trzhaskovskaya M.B. Physical Review C. 2007;76:054313. [Google Scholar]
  • 17.Wense L.v.d., Thirolf P.G., Kalb D., Laatiaoui M. Journal of Instrumentation. 2013;8:P03005. [Google Scholar]
  • 18.Rellergert W.G., DeMille D., Greco R.R., Hehlen M.P., Torgerson J.R., Hudson E.R. Physical Review Letters. 2010;104:200802. doi: 10.1103/PhysRevLett.104.200802. [DOI] [PubMed] [Google Scholar]
  • 19.Kazakov G.A., Litvinov A.N., Romanenko V.I., Yatsenko L.P., Romanenko A.V., Schreitl M., Winkler G., Schumm T. New Journal of Physics. 2012;14:083019. [Google Scholar]
  • 20.Hehlen M.P., Greco R.R., Rellergert W.G., Sullivan S.T., DeMille D., Jackson R.A., Hudson E.R., Torgerson J.R. Journal of Luminescence. 2013;133:91. [Google Scholar]
  • 21.Porsev S.G., Flambaum V.V., Peik E., Tamm C. Physical Review Letters. 2010;105:182501. doi: 10.1103/PhysRevLett.105.182501. [DOI] [PubMed] [Google Scholar]
  • 22.Herrera-Sancho O.A., Okhapkin M.V., Zimmermann K., Tamm C., Peik E. Physical Review A. 2012;85:033402. [Google Scholar]
  • 23.Radnaev A.G., Campbell C.J., Kuzmich A. Physical Review A. 2012;86:060501(R). [Google Scholar]
  • 24.Sonnenschein V., Moore I.D., Raeder S., Hakimi A., Popov A., Wendt K. European Physical Journal A. 2012;48:52. [Google Scholar]
  • 25.Ruchowska E., Płóciennik W.A., Żylicz J., Mach H., Kvasil J., Algora A., Amzal N., Bäck T., Borge M.G., Boutami R., Butler P.A., Cederkäll J., Cederwall B., Fogelberg B., Fraile L.M., Fynbo H.O.U., Hagebø E., Hoff P., Gausemel H., Jungclaus A., Kaczarowski R., Kerek A., Kurcewicz W., Lagergren K., Nacher E., Rubio B., Syntfeld A., Tengblad O., Wasilewski A.A., Weissman L. Physical Review C. 2006;73:044326. [Google Scholar]
  • 26.Inamura T.T., Haba H. Physical Review C. 2009;79:034313. [Google Scholar]
  • 27.Kikunaga H., Kasamutsu Y., Haba H., Mitsugashira T., Hara M., Takamiya K., Ohtsuki T., Yokoyama A., Nakanishi A., Shinohara A. Physical Review C. 2009;80:034315. [Google Scholar]
  • 28.Zhao X., de Escobar Y.N.M., Rundberg R., Bond E.M., Moody A., Vieira D.J. Physical Review Letters. 2012;109:160801. doi: 10.1103/PhysRevLett.109.160801. [DOI] [PubMed] [Google Scholar]
  • 29.Peik E., Zimmerman K. Physical Review Letters. 2013;111:018901. doi: 10.1103/PhysRevLett.111.018901. [DOI] [PubMed] [Google Scholar]
  • 30.Sakharov S.L. Physics of Atomic Nuclei. 2010;73:1. [Google Scholar]
  • 31.B. Beck, Experiments on the elusive 229Th meta-stable. ECT meeting “Atomic Effects in Nuclear Excitation and Decay”, 15–19 June 2009, Trento, Italy.
  • 32.Porter F.S., Brown G.V., Boyce K.R., Kelley R.L., Kilbourne C.A., Beiersdorfer P., Chen H., Terracol S., Kahn S.M., Szymkowiak A.E. Review of Scientific Instruments. 2004;75:3772. [Google Scholar]
  • 33.Pies C., Schäfer S., Heuser S., Kempf S., Pabinger A., Porst J.-P., Ranitsch P., Foerster N., Hengstler D., Kampkötter A., Wolf T., Gastaldo L., Fleischmann A., Enss C. Journal of Low Temperature Physics. 2012;167:269. [Google Scholar]
  • 34.Neyman J., Pearson E.S. Philosophical Transactions of the Royal Society of London A. 1933;231:289. [Google Scholar]
  • 35.Wilks S.S. Annals of Mathematical Statistics. 1938;9:60. [Google Scholar]
  • 36.Wald A. Transactions of the American Mathematical Society. 1943;54:426. [Google Scholar]
  • 37.NuDat 2.6 database, viewed 07 May 2013, 〈http://www.nndc.bnl.gov/nudat2/〉.
  • 38.A.C. Thompson, D.T. Attwood, E.M. Gullikson, M.R. Howells, J.B. Kortright, A.L. Robinson, J.H. Underwood Kwang-Je Kim, J. Kirz, I. Lindau, P. Pianetta, H. Winick, G.P. Williams, J.H. Scofield, X-Ray Data Booklet. Lawrence Berkeley National Laboratory. University of California, Berkeley, CA 94720, 2009.
  • 39.XCOM: Photon Cross-Sections Database, viewed 06 May 2013, 〈http://www.nist.gov/pml/data/xcom/index.cfm〉.
  • 40.Hubbel J.H., Trehan P.N., Singh N., Chand B., Metha D., Garg M.L., Garg R.R., Singh S., Puri S. Journal of Physical and Chemical Reference Data. 1994;23:339. [Google Scholar]
  • 41.Fleischmann A., Enss C., Seidel G.M. Topics in Applied Physics. 2005;99:151. [Google Scholar]
  • 42.Raboud P.A., Dousse J.Cl., Hoszowska J., Savoy I. Physical Review A. 1999;61:012507. [Google Scholar]
  • 43.Hallstadius L. Nuclear Instruments and Methods in Physics Research. 1984;223:266. [Google Scholar]
  • 44.Luna-Zaragoza D., Serrano J.G. Applied Radiation and Isotopes. 1999;51:499. [Google Scholar]
  • 45.Hubbell J.H., Veigele Wm.J., Briggs E.A., Brown R.T., Cromer D.T., Howerton R.J. Journal of Physical and Chemical Reference Data. 1975;4:471. [Google Scholar]

RESOURCES