Optimized Saturation Pulse Train for Human First-Pass Myocardial Perfusion Imaging at 7T

Yuehui Tao 1,*, Aaron T Hess 1, Graeme A Keith 1, Christopher T Rodgers 1, Alexander Liu 1, Jane M Francis 1, Stefan Neubauer 1, Matthew D Robson 1
PMCID: PMC4377098  PMID: 24753130

Abstract

Purpose

To investigate whether saturation using existing methods developed for 3T imaging is feasible for clinical perfusion imaging at 7T, and to propose a new design of saturation pulse train for first-pass myocardial perfusion imaging at 7T.

Methods

The new design of saturation pulse train consists of four hyperbolic-secant (HS8) radiofrequency pulses, whose peak amplitudes are optimized for a target range of static and transmit field variations and radiofrequency power deposition restrictions measured in the myocardium at 7T. The proposed method and existing methods were compared in simulation, phantom, and in vivo experiments.

Results

In healthy volunteer experiments without contrast agent, average saturation efficiency with the proposed method was 97.8%. This is superior to results from the three previously published methods at 86/95/90.8%. The first series of human first-pass myocardial perfusion images at 7T have been successfully acquired with the proposed method.

Conclusion

Existing saturation methods developed for 3T imaging are not optimal for perfusion imaging at 7T. The proposed new design of saturation pulse train can saturate effectively, and with this method first-pass myocardial perfusion imaging is feasible in humans at 7T. Magn Reson Med 73:1450–1456, 2015. © 2014 The Authors. Magnetic Resonance in Medicine Published by Wiley Periodicals, Inc. on behalf of International Society of Medicine in Resonance.

Keywords: 7T, cardiac magnetic resonance imaging, myocardial perfusion, saturation, radiofrequency pulse, B1 inhomogeneity

INTRODUCTION

First-pass myocardial perfusion imaging (1) is one of the most important applications of cardiac magnetic resonance (MR). It has been reported that 3T cardiac MR perfusion imaging is superior to 1.5T for prediction of significant single vessel and multivessel coronary disease (2). Further improvement can be expected at 7T due to higher signal-to-noise ratio and longer T1 of the myocardium (3) but no clinical cardiac perfusion at 7T has yet been reported.

The major challenges at 7T are large static field (B0) and transmit radiofrequency (RF) field (B1) variations across the heart and limited RF power deposition allowed for a complete and uniform saturation. Incomplete saturation in the perfusion sequence can result in unintended high signal in the myocardium in perfusion images, which may make it difficult to distinguish a perfusion defect from healthy myocardium. This problem was already observed when transitioning from 1.5T to 3T. Perfusion methods developed at 1.5T were found to operate inadequately at 3T owing to large B0 and B1 variation due to limited capability of commercially available shimming and calibration methods for the heart. Several methods have been proposed to overcome these challenges. Composite adiabatic pulses known as B1-insensitive rotation (BIR4) (4,5) have been demonstrated capable of effective saturation at 3T (6). A train of two BIR4 pulses has also been proposed for saturation at 3T (7,8). The disadvantage of the BIR4 pulse is that it has an excessive specific absorption rate (SAR), which is problematic even at 3T (6).

Another group of saturation methods use a train of RF pulses. Each RF subpulse is interleaved between gradient spoilers to remove residual transverse magnetization. Perhaps the simplest variation uses a pulse train of three nonselective 90° rectangular pulses (9) (hereafter referred to as the standard pulse train). This method is adopted on commercially available 1.5T and 3T systems but has been shown to be susceptible to B1 inhomogeneity at 3T (8,10). Alternatively, the tailored pulse train (10) consists of rectangular pulses of different flip angles. The flip angles are optimized for a range of B0 and B1 variation measured a priori over the heart. The hybrid pulse train (8) is an extension of the tailored pulse train, combining two rectangular pulses and an adiabatic half-passage pulse. Both the tailored and hybrid pulse trains produced better saturation than the standard pulse train at 3T (8,10), but their performance depends on the predetermined range of B0 and B1 variation that the pulse trains were optimized for. The B1-insensitive train to obliterate signal BISTRO (11), developed for water suppression and outer-volume suppression when using local transmit coils, uses a large number of amplitude- and frequency-modulated RF pulses, each of which is offset-independent adiabatic and produces a uniform saturation inside the limits of a frequency sweep range. To compensate for large B1 variation, the amplitudes of the RF pulses in the BISTRO pulse train are varied. A sufficiently large number of RF pulses are required to ensure that all regions experience a saturating flip angle at least once. The limitation of the BISTRO method is that the RF power must exceed a threshold level and the required number of RF pulses can be large [32 in (11)].

According to initial experience with our 7T whole body system, the range of B0 and B1 variation is much larger than at 3T, and the peak B1 level that is available is much lower. Further, the SAR limit becomes much more challenging at 7T owing to the higher frequency of the RF. Consequently, uniform saturation presents many challenges.

In this study, we first determine the range of B0 and B1 and the effect of the SAR limit from scans of healthy volunteers on our 7T system. Then, we investigate whether saturation using existing methods developed at 3T is feasible for perfusion imaging at 7T. Further, we propose a new design of saturation pulse train for first-pass myocardial perfusion imaging at 7T. The new method is compared to existing techniques in simulation, phantom, and in vivo experiments. We also present the first series of human first-pass myocardial perfusion images at 7T.

METHODS

All experiments used a Magnetom whole-body 7T MRI scanner (Siemens Healthcare, Erlangen, Germany). The coil is an eight-channel strip-line transverse electromagnetic transceiver array (12). Maximum gradient strength is 70 mT/m, and maximum slew rate is 200 T/m/s.

A fast low-angle shot sequence was modified to be able to use different saturation pulse trains prior to image acquisition. All images were acquired in the midventricular short-axis slice. One image was collected in each heartbeat with the perfusion sequence. The Siemens vector electrocardiogram was used for triggering, with a trigger delay of 300 ms. Unless otherwise stated, the imaging parameters were: field of view 380 × 285 mm2; acquisition matrix 192 × 144; slice thickness 8 mm; in-plane resolution 2 × 2 mm2; nominal flip angle 10°; echo time 1.18 ms; repetition time 2.82 ms; receiver bandwidth 1002 Hz/pixel. Generalized autocalibrating partially parallel acquisitions (13) was applied with an acceleration factor of 2. The acquisition of each image consisted of 84 phase-encoding steps. Time between the end of the saturation pulse train and the central k-space readout was 125 ms.

Eight healthy volunteers (age 31 ± 7 years, weight 77 ± 8 kg) were recruited. Ethics approval was granted for all study procedures and informed consent was obtained from all subjects.

B0 and B1 Variation

Manufacturer provided software was used for B0 shimming over an adjustment volume covering the whole left ventricle (LV). B0 offset was measured with a double-echo gradient echo sequence.

Local B1 shimming (14) was performed to maximize the minimum B1 over a region of interest (ROI) manually selected in the midventricle short-axis slice covering the LV. The B1 map was determined from images acquired with a B0-independent saturation preparation method (15).

B0 and B1 maps in the midventricle short-axis slice were acquired from eight volunteers. LV myocardium was manually segmented in the B0 and B1 maps. The observed number of pixels as a function of B0 and B1 was used to fit the joint probability distribution for B0 and B1 in the LV myocardium. The distribution was approximated as a Gaussian mixture distribution (16) in the fitting as in the design of the tailored pulse train (10). A range of B0 offset (−250 to 250 Hz) and B1 (150–400 Hz, 1 µT of B1 corresponds to 42.58 Hz) was then manually selected to cover the aforementioned distribution with a probability greater than 99%.

Energy Limit for the Saturation Pulse Train

The online RF-power monitoring system of the scanner was used to determine the maximum energy available for the saturation pulse train. The system monitors RF power averaged in real-time over durations of 0.1/1/10/360 s, and compares it with a set of preset RF power limits. A scan will be stopped automatically when any of the preset RF power limits is exceeded. Each transmit element of the eight-channel coil is connected to an independently controlled RF amplifier and dedicated power monitor. Conservative safety limits are imposed based on SAR simulations for the coil in Semcad X (Schmid and Partner Engineering AG, Switzerland). These worst-case limits assume constructive interference of E-fields everywhere in space. The preset safety limit that is most likely breached is the 1 s limit. For perfusion scans saturating once and collecting one image per heartbeat, the total energy of all RF pulses within 1 s also depends on the heart rate. For a typical heart rate between 50 and 100 bpm, the design of the saturation pulse train needs to allow both saturation and image acquisition to be applied twice within 1 s.

The limit of relative energy for a saturation pulse train was obtained using a maximum-amplitude rectangular pulse as the saturation pulse for the perfusion pulse sequence and gradually increasing the duration of the rectangular pulse, in steps of 0.1 ms, to challenge the RF-power monitoring system. The relative energy is defined as the total energy of a set of RF pulses divided by the energy of a 1-ms rectangular pulse of the maximum amplitude.

According to the limits measured from eight volunteers and assuming a normal distribution, the limit of relative energy for a saturation pulse train was set to 6. With the limit at this level, the probability of triggering an automatic stop of scan due to excessive energy deposition by the saturation RF pulses is less than 1% at a heart rate between 50 and 100 bpm. This limit would be breached if an 8-ms BIR4 pulse at maximum amplitude was used for saturation.

Optimization of Saturation Pulse Train for 7T

The development of the saturation pulse train for perfusion at 7T used the same approach adopted for the tailored pulse train (10). The flip angles (peak amplitudes) of the RF pulses in the pulse train were optimized over the range of B0 and B1 variation determined above. Large B0 variation requires RF pulses of broad bandwidth (hence a short duration high voltage pulse), while regions of low peak B1 also increase the demands on the coil voltage. It is difficult to fulfill both requirements with a rectangular pulse of fixed amplitude because its flip angle increases with its duration, but its bandwidth decreases proportionally. No good solution was found in the search for a tailored rectangular pulse train under the conditions at 7T.

Full-passage hyperbolic-secant (HS8) pulses (17) are more suited to this saturation problem. When the duration of a HS8 pulse is fixed at 5 ms, and the frequency sweep ±500 Hz, the effective bandwidth is not sensitive to the change of peak amplitude at low levels of peak B1 comparable to the B1 range at 7T.

To minimize the relative energy [Inline graphic] of the HS8 pulse train, an exhaustive search of the peak amplitudes [Inline graphic] of the four HS8 pulses (duration 5 ms, frequency sweep ±500 Hz) was carried out. The peak amplitudes ranged from 0 to 100% of the maximum amplitude, in steps of 1%. Numerical Bloch simulations were performed over the range (Inline graphic) of B0 offset (±250 Hz) and B1 (150–400 Hz) previously determined from volunteer measurements. T1 and T2 relaxation was ignored. Ideal spoiling was assumed in the simulation, so the transverse magnetization was zero before each RF pulse. Solutions were excluded when the normalized residual longitudinal magnetizationInline graphic was outside the range −6% to 0, or when the relative energy of the four HS8 pulses was greater than 6.

graphic file with name mrm0073-1450-m5.jpg

The peak amplitudes determined from the search are 29/47/47/84%, respectively, of the maximum amplitude. The relative energy of the HS8 pulse train is 4.9. The probability of a stop of scan triggered by the HS8 pulse train due to excessive RF energy is virtually zero, according to the distribution of the energy limits previously measured from eight volunteers. To complete the HS8 pulse train, each HS8 pulse is interleaved between gradient spoilers (duration 1/8/6/4/1 ms including 0.5 ms for each linear ramp, peak amplitude 50 mT/m), as shown in Figure 1. The durations and amplitudes of the gradient spoilers were determined empirically to ensure adequate spoiling.

FIG 1.

FIG 1

Pulse sequence diagram of the HS8 saturation pulse train.

Simulation, Phantom, and In Vivo Comparisons

The saturation performance of the HS8 pulse train and four existing methods (the standard/tailored/hybrid pulse train and the BIR4 pulse) was evaluated in numerical Bloch simulation of the saturation of a single isochromat in the B1 range of 150–400 Hz and B0 offset range of ±250 Hz. The parameters of the five methods are listed in Table 1. T1 and T2 relaxation was ignored and ideal spoiling was assumed.

Table 1.

Parameters of the Saturation Pulse Trains

Pulse train RF pulse

Relative energya Durationb (ms) Duration (ms) Peak amplitudec
Standard (9,10) 1.5 6.5 0.5 1
0.5 1
0.5 1
BIR4 (4,10) 6.4 8 8 (BIR4) 1
Tailored (10) 2.648 7.648 0.548 1
1.296 1
0.804 1
Hybrid (8) 2.1 7.84 0.8 0.73
0.5 0.75
1.54 (AHP) 1
HS8 (17) 4.9 38 5 (HS8) 0.29
5 (HS8) 0.47
5 (HS8) 0.47
5 (HS8) 0.84
a

Relative energy is defined as the total energy of the pulse train divided by the energy of a 1-ms rectangular RF pulse at the maximum amplitude.

b

Pulse train duration excludes leading and trailing gradient spoilers. The spoiler durations for the standard/tailored/hybrid pulse trains are 1, 3, 2, and 3 ms. The spoiler durations for the HS8 pulse train are 1, 8, 6, 4, and 1 ms, including 0.5 ms for each linear ramp.

c

Peak amplitude is relative to the maximum amplitude.

The HS8 pulse train and three existing methods (the standard/tailored/hybrid pulse train) were also tested in scans of a 10-cm spherical oil phantom (T1 1420 ms). To simulate the B0 and B1 variation, the amplitude of the saturation RF pulses were scaled from 20 to 100% of the maximum, in steps of 10% of the maximum amplitude, and a frequency offset between ±320 Hz, in steps of 20 Hz, was added to all saturation RF pulses. All phantom images were acquired in the same axial plane at isocenter. In each scan, the test saturated image was collected 8 s after an unsaturated image of otherwise identical imaging parameters.

Each test saturated image was divided pixel-wise by its corresponding unsaturated image to calculate the normalized absolute residual longitudinal magnetization, which was then corrected for T1 relaxation effects. The correction calculated the normalized residual longitudinal magnetization at the end of the saturation pulse train from signal measured 125 ms after, assuming simple T1 relaxation (T1 of oil 1420 ms). Six 2 × 2 (4 pixels) ROIs were selected at locations of different B1 levels in each image. The normalized absolute residual longitudinal magnetization was then averaged in each ROI. The B1 field over the imaged slice was not homogeneous, so the six ROIs in each image could be selected at locations of desired B1 levels. With the scaling of the amplitudes of the saturation RF pulses, 26 B1 levels, in steps of 10 Hz, in the range of 150–400 Hz were measured.

Eight volunteers were scanned without contrast using the HS8 pulse train and three existing methods (the standard/tailored/hybrid pulse train). As in the phantom scans, two short axis images were acquired in each breath hold, the first unsaturated and the second saturated. The delay between the two images was 8 s. The LV myocardium was manually segmented into six segments (18). Each segment was treated as an independent ROI. The residual signal was normalized by calculating a pixel-wise ratio image between each saturated image and its corresponding unsaturated image. Mean and maximum values of the normalized residual signal were calculated over each ROI. Mean ± (standard deviation) of the mean and maximum values were then calculated over all 48 ROIs.

Rest Perfusion Scan

Three male volunteers were scanned during the first pass of gadolinium-chelate contrast agent (Dotarem, Guerbet, France). The HS8 pulse train was used for saturation. 60 images were collected, one each heartbeat, over an end-expiratory breath hold. The volunteers were asked to hold breath as long as possible, and then breathe smoothly. The peripheral injection of a contrast bolus (0.05 mmol/kg, flow rate 6 mL/s) with a power injector (Accutron MR, MEDTRON, Germany) was started between the third and fourth heartbeat. The first image was not saturated.

The LV myocardium was manually segmented into six segments. The signal intensities were averaged in each segment. The average intensities from all images except the first one were divided by the corresponding averages from the first image to calculate the normalized signal intensity. The normalized signal intensity was also calculated for the LV blood pool.

RESULTS

Figure 2 shows an example of B0 and B1 variation from the same volunteer. In the inferolateral segment the B0 variation is visibly larger than other regions. This is also observed in images from all other volunteers and is consistent with the literature (19). Figure 3 shows the B0 and B1 variations measured from all volunteers with the range selected for the optimization of the saturation pulse train.

FIG 2.

FIG 2

B0 offset and B1 maps over the LV myocardium measured from one volunteer in the midventricular short-axis slice.

FIG 3.

FIG 3

Histogram of B0 offset and B1 measurements over the LV myocardium from eight volunteers.

Figure 4 shows the worst case results obtained with different methods in simulation. All methods except the HS8 pulse train left residual magnetization over 10% when B1 was below 200 Hz. This is also shown in Figure 5 with results from phantom experiments, which agree well with the simulation results. The effective bandwidth of the tailored train is broader than that of the hybrid train, leading to a lower maximum absolute normalized residual longitudinal magnetization than that from the hybrid train over the whole range of B1 from 150 to 400 Hz.

FIG 4.

FIG 4

The maximum (worst case) absolute normalized residual longitudinal magnetization (Inline graphic) over the B0 offset range of ±250 Hz as a function of B1, from numerical Bloch simulations.

FIG 5.

FIG 5

The absolute normalized residual longitudinal magnetization produced with the standard, tailored, hybrid, and HS8 pulse trains in simulation (top row) and phantom scans (bottom row).

Figure 6 shows the images saturated with different methods, and statistics of mean and maximum residual signal intensity over the LV myocardium from eight volunteers are listed in Table 2. Given that the relaxation time between the end of saturation pulse train and the central k-space readout was 125 ms, the normalized residual myocardium signal immediately after the saturation was calculated assuming T1 in the myocardium was 1950 ms (3). The average residual from the HS8 train was 2.2% (average saturation efficiency 97.8%), less than the average residual from the standard/tailored/hybrid trains at 14.0/5.0/9.2% (average saturation efficiency 86/95/90.8%). The maximum residual from the HS8 train was 4.7%, less than the maximum residual from the standard/tailored/hybrid trains at 25.2/13.3/19.2%.

FIG 6.

FIG 6

Representative images from one volunteer: (a) nonsaturated, saturated with the (b) HS8, (c) standard, (d) tailored, (e) hybrid pulse train. All saturated images (b–e) are displayed in the same grayscale.

Table 2.

Mean (Inline graphic) and Maximum (Inline graphic) Residual Signal Intensity Over the Left Ventricle Myocardium Measured from All Volunteers before T1 Correction

Saturation pulse train Smean (%) Smax (%)
HS8 8.3 ± 0.9 10.6 ± 1.1
Standard 19.3 ± 7.1 29.8 ± 12.5
Tailored 10.9 ± 4.2 18.7 ± 6.5
Hybrid 14.8 ± 2.2 24.2 ± 7.1

Figures 7 and 8 shows first-pass perfusion results from a male volunteer (age 32 years, weight 80 kg). The normalized residual signals from all segments are close to each other, suggesting a uniform saturation in this healthy volunteer. These images demonstrate that first-pass perfusion imaging in humans at 7T is feasible.

FIG 7.

FIG 7

First-pass perfusion images with the HS8 pulse train for saturation: (a) nonsaturated, (b) before arrival of contrast agent, (c) right ventricle blood enhancement, (d) LV blood enhancement, (e) LV wall enhancement. All saturated images (b–e) are displayed in the same grayscale.

FIG 8.

FIG 8

The time course of the normalized signal intensity in LV blood pool and six midventricular short-axis segments (18) during the first pass of the contrast agent.

DISCUSSION

With this work, we have developed a new design of saturation pulse train for cardiac perfusion imaging at 7T, and successfully acquired the first series of human first-pass myocardial perfusion images at 7T with the proposed method. In simulation, phantom, and in vivo experiments, we have demonstrated that existing saturation methods developed for 3T imaging are not optimal for cardiac perfusion imaging at 7T. This can be explained by differences in the ranges of B0 and B1 variation targeted by these saturation methods. Previously proposed methods were designed for a higher B1 level and narrower B0 bandwidth. These were appropriate for body coil on 3T scanners but not suitable for the transmitting surface coil on our 7T system. Lower B1 levels require more power for the saturation, however, the RF power deposition allowed at 7T is very limited. This excluded the BIR4 pulse, and limited the number of pulses available for use in all pulse trains. The HS8 pulse and the adiabatic half-passage pulse in the hybrid pulse train are both operating outside adiabatic range at low B1 levels in the targeted range of B1 at 7T, which may explain the relatively high residual normalized signal after the saturation with the hybrid pulse train.

The range of B0 offset (±250 Hz) and B1 (150–400 Hz) determined from in vivo measurements had an important role in the design of the HS8 pulse train. No good solution was found in the search for a tailored rectangular pulse train for this range. The duration and frequency sweep of the HS8 pulses were specifically chosen for the above range. The results of the search were sensitive to changes to the range, especially changes to the lower limit of the B1 range. With the lower B1 limit at 140 Hz, no solution was found. At 200 Hz, the required relative energy was reduced to 2.7. Above 235 Hz, the required relative energy was less than 2, enabling coverage of three slices. Reducing the higher limit of the B1 range did not save much energy. With the higher limit at 320 Hz, the saturation still required 4.8 in relative energy. Because the effective saturation bandwidth of the chosen HS8 pulse is mainly determined by its duration and frequency sweep, the results of the search were not sensitive to reduction of the B0 offset bandwidth. With the bandwidth reduced to ±150 Hz, the required relative energy was 4.8. If the range was redefined as an elliptical region spanning ±250 Hz in B0 offset and 150–400 Hz in B1, the required relative energy would be 4.4. The HS8 pulse train proposed in this work (29/47/47/84%) would not saturate effectively if B0 offset and B1 were out of the predetermined range. According to numerical Bloch simulation ignoring relaxation, the maximum normalized longitudinal magnetization immediately after saturation would be 5.6% at B1 of 135 Hz and 0.2% at 415 Hz. If the B0 offset bandwidth was ±300 Hz, the maximum normalized longitudinal magnetization would be 5.4%. Within the range selected in this work, the performance of the HS8 pulse train is not sensitive to small deviations in amplitude of any single HS8 pulse in the pulse train. If the peak amplitude of any single RF pulse in the HS8 pulse train changes by 1% of the maximum amplitude, the maximum change in minimum or maximum normalized longitudinal magnetization will be less than 0.01.

The HS8 pulse train bears an appearance similar to the BISTRO method, both using amplitude- and frequency-modulated RF pulses interleaved between gradient spoilers. The RF pulses used in the BISTRO method have a wide frequency sweep (8 KHz), and they are offset-independent adiabatic. The frequency sweep of the HS8 pulses in the HS8 pulse train in this work is only 1000 Hz. In the low B1 field observed in this study, these 5-ms HS8 pulses are not offset-independent adiabatic. The peak amplitudes of the RF pulses of the BISTRO method are not optimized for a target range of B0 and B1 variation, and the BISTRO method requires a sufficiently large number of RF pulses to ensure that all regions experience a flip angle near 90° at least once, which was not required in the design of the HS8 pulse train. In regions of low B1, each single HS8 pulse alone in the HS8 pulse train will not produce a flip angle of 90°, but the combined effect of four HS8 pulses will produce a saturating flip angle.

Other types of numerically optimized RF pulses can also be used to build the saturation pulse train. However, it is difficult to cover the whole universe of RF pulses. We did evaluate a selection of different pulses and found HS8 to be the best. In addition to a pulse train of all HS8 pulses, we also observed in numerical Bloch simulations that a hybrid train of rectangular and HS8 pulses could also saturate effectively in the target B0 and B1 range but that it deposits slightly greater power into the subject. At low flip angles (short pulse durations), a hard pulse may have similar saturation effects as an HS8 pulse of slightly less power, when the effective bandwidth of the hard pulse is wide enough to cover the target range of B0. An optimized hybrid train of HS8 and hard pulses may be built by replacing low-flip-angle HS8 pulses in the optimized HS8 pulse train with hard pulses of slightly greater power. However, the high-flip-angle HS8 pulses cannot be replaced in the same way, because hard pulses of high flip angles (long durations) will not saturate effectively over the target range of B0. This may explain why no good solution was found in the search for a tailored rectangular pulse train under the conditions at 7T.

It should be noted that the imaging protocol used for the rest perfusion scan was not optimized specifically for 7T. The image quality in these first acquisitions is already comparable to that at 3T, supporting the great potential for perfusion imaging at 7T.

Multislice cardiac perfusion imaging at 7T which saturates multiple times in each heart beat remains challenging. According to the relative energy limit allowed for the saturation pulse measured in this study, the relative energy of the saturation pulse train needs to be reduced to 2 if the heart is to be saturated three times each heartbeat. One option may be to improve the B1 shimming performance. If the lower limit of B1 can be increased to 200 Hz and the upper limit unchanged, the RF energy of the saturation pulse train may be reduced by 43% by simply scaling down the amplitudes of the RF pulses. Alternatively improved B0 shimming would allow the optimization to be performed over a narrower B0 range which would be a second means of reducing RF power deposition. Finally, the RF SAR calculations that we have used are by necessity cautious and future improvements in RF coils and/or RF monitoring technology should allow us to deliver more power into the subject in complete safety. These are all areas of on-going development.

CONCLUSIONS

We demonstrated that the proposed HS8 pulse train saturated effectively within the target range of B0 and B1 variation and the energy limit at 7T, while the existing methods designed for 3T imaging were not as effective. We also demonstrated that single-slice first-pass myocardial perfusion imaging is feasible for human at 7T.

Acknowledgments

This work was supported by the Medical Research Council (UK). MDR also acknowledges support from the Oxford NIHR Biomedical Research Centre.

REFERENCES

  • 1.Gerber BL, Raman SV, Nayak K, Epstein FH, Ferreira P, Axel L, Kraitchman DL. Myocardial first-pass perfusion cardiovascular magnetic resonance: history, theory, and current state of the art. J Cardiovasc Magn Reson. 2008;10:18. doi: 10.1186/1532-429X-10-18. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Cheng AS, Pegg TJ, Karamitsos TD, Searle N, Jerosch-Herold M, Choudhury RP, Banning AP, Neubauer S, Robson MD, Selvanayagam JB. Cardiovascular magnetic resonance perfusion imaging at 3-tesla for the detection of coronary artery disease: a comparison with 1.5-tesla. J Am Coll Cardiol. 2007;49:2440–2449. doi: 10.1016/j.jacc.2007.03.028. [DOI] [PubMed] [Google Scholar]
  • 3.Rodgers CT, Piechnik SK, Delabarre LJ, Van de Moortele PF, Snyder CJ, Neubauer S, Robson MD, Vaughan JT. Inversion recovery at 7 T in the human myocardium: measurement of T1, inversion efficiency and B1+ Magn Reson Med. 2013;70:1038–1046. doi: 10.1002/mrm.24548. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Staewen RS, Johnson AJ, Ross BD, Parrish T, Merkle H, Garwood M. 3-D FLASH imaging using a single surface coil and a new adiabatic pulse, BIR-4. Invest Radiol. 1990;25:559–567. doi: 10.1097/00004424-199005000-00015. [DOI] [PubMed] [Google Scholar]
  • 5.Kim D, Cernicanu A, Axel L. B0 and B1-insensitive uniform T1-weighting for quantitative, first-pass myocardial perfusion magnetic resonance imaging. Magn Reson Med. 2005;54:1423–1429. doi: 10.1002/mrm.20704. [DOI] [PubMed] [Google Scholar]
  • 6.Kim D, Gonen O, Oesingmann N, Axel L. Comparison of the effectiveness of saturation pulses in the heart at 3 T. Magn Reson Med. 2008;59:209–215. doi: 10.1002/mrm.21423. [DOI] [PubMed] [Google Scholar]
  • 7.Cunningham CH, Pauly JM, Nayak KS. Saturated double-angle method for rapid B1mapping. Magn Reson Med. 2006;55:1326–1333. doi: 10.1002/mrm.20896. [DOI] [PubMed] [Google Scholar]
  • 8.Kim D, Oesingmann N, McGorty K. Hybrid adiabatic-rectangular pulse train for effective saturation of magnetization within the whole heart at 3 T. Magn Reson Med. 2009;62:1368–1378. doi: 10.1002/mrm.22140. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Oesingmann N, Zhang Q, Simonetti O. Improved saturation RF pulse design for myocardial first-pass perfusion at 3T. J Cardiovasc Magn Reson. 2004;6:373–374. [Google Scholar]
  • 10.Sung K, Nayak KS. Design and use of tailored hard-pulse trains for uniform saturation of myocardium at 3 Tesla. Magn Reson Med. 2008;60:997–1002. doi: 10.1002/mrm.21765. [DOI] [PubMed] [Google Scholar]
  • 11.Luo Y, de Graaf RA, DelaBarre L, Tannus A, Garwood M. BISTRO: an outer-volume suppression method that tolerates RF field inhomogeneity. Magn Reson Med. 2001;45:1095–1102. doi: 10.1002/mrm.1144. [DOI] [PubMed] [Google Scholar]
  • 12.Snyder CJ, DelaBarre L, Moeller S, Tian J, Akgun C, van de Moortele PF, Bolan PJ, Ugurbil K, Vaughan JT, Metzger GJ. Comparison between eight- and sixteen-channel TEM transceive arrays for body imaging at 7 T. Magn Reson Med. 2012;67:954–964. doi: 10.1002/mrm.23070. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A. Generalized autocalibrating partially parallel acquisitions (GRAPPA) Magn Reson Med. 2002;47:1202–1210. doi: 10.1002/mrm.10171. [DOI] [PubMed] [Google Scholar]
  • 14.Metzger G, Van de Moortele P, Vaughan JT, Snyder C, Akgun C. Local B1+ shimming for prostate imaging with transceiver arrays at 7 tesla based on subject dependent phase measurements. Magn Reson Med. 2008;59:396–409. doi: 10.1002/mrm.21476. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Tao Y, Hess AT, Keith GA, Rodgers CT, Robson MD. 2013. p. 2596. B0-independent quantitative measurement of low B1 field for human cardiac MRI at 7T. In Proceedings of the 21st Annual Meeting of ISMRM, Salt Lake City, USA.
  • 16.McLachlan G, Peel D. Finite mixture models. New York: Wiley; 2000. p. 456. [Google Scholar]
  • 17.Tannus A, Garwood M. Adiabatic pulses. NMR Biomed. 1997;10:423–434. doi: 10.1002/(sici)1099-1492(199712)10:8<423::aid-nbm488>3.0.co;2-x. [DOI] [PubMed] [Google Scholar]
  • 18.Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS. American heart association writing group on myocardial S, registration for cardiac I. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: a statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002;105:539–542. doi: 10.1161/hc0402.102975. [DOI] [PubMed] [Google Scholar]
  • 19.Atalay MK, Poncelet BP, Kantor HL, Brady TJ, Weisskoff RM. Cardiac susceptibility artifacts arising from the heart-lung interface. Magn Reson Med. 2001;45:341–345. doi: 10.1002/1522-2594(200102)45:2<341::aid-mrm1043>3.0.co;2-q. [DOI] [PubMed] [Google Scholar]

Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley

RESOURCES