Extreme hemodynamic fluctuations: Importance of understanding the principles of syringe pump function

Chakrabarti Dhritiman 1, Byrappa Vinay 1,
PMCID: PMC4397639  PMID: 25878440

Sir,

Infusion devices remain high-risk devices capable of delivering drugs unpredictably leading to mortality and morbidity. The problem of performance, degradation, and quality was the cause for 20% of reported incidences relating to the infusion pump in UK between 1990 and 2000.[1] Thus, it is important for the clinician to be familiar with the features and function of the infusion device they are using.

A 55-year-old lady with middle cerebral artery aneurysm underwent microsurgical clipping of the aneurysm under general anesthesia uneventfully. After clipping, with the target of raising systolic blood pressure (SBP) to 160 mm Hg, noradrenalin infusion was started at 1.6 μg/min. The solution was prepared in a 50 ml syringe with 8 mg of noradrenaline bitartrate diluted in 40 mL normal saline and infused at 2 mL/h. The solution was infused using Orchestra® DPS + (Fresenius Kabi) infusion pump. (Ref 082594/21126459) La Grand Chamin-38590 Brezins- France. Initially, there was a surge in SBP up to 240 mm Hg. No intervention was done as the surge was expected to settle down. However, after the blood pressure decreased to 110 mm Hg, there was another surge in SBP up to 200 mm Hg after approximately 5 min. This was followed by 4 similar cyclical surges at 10 min intervals [Figure 1]. We changed the syringe but again the fluctuations in blood pressure persisted.

Figure 1.

Figure 1

Snapshot showing graphical representation of fluctuation in invasive blood pressure

Pulsatile flow of drug at low dose rates in syringe pump was then thought to be the cause and to smoothen the fluctuations, the drug in the syringe solution was double diluted (i.e. 4 mg in 40 mL) and the dose rate was also doubled (4 mL/h). Following this intervention, the surges flattened out considerably, and the SBP fluctuated between 170 and 150 mm Hg thereafter [Figure 1].

The Orchestra® DPS + pump uses stepper motors with the lead screw, and the flow rate accuracy of the motor is documented to be ± 1% on drive mechanism and ±2% on syringes.[2] However, at low flow rates, the flow may be discontinuous and pulsatile. Stepper motors are known to produce flow in a series of discrete fixed volume pulses. The volume and rate of pulses at fixed flow rates depend on the number of steps per revolution of the lead screw. If the time duration between the pulses is relatively large compared to the half-life of the drug, the result may be wide variation in effect of the drug.[3] Another reason may be stiction (i.e., the friction which tends to prevent stationary surfaces from being set in motion) between syringe plunger and barrel.[4] A similar problem was investigated by Capes et al. and they found Terumo syringes to be significantly associated with the noncontinuous flow.[5]

In our case, the syringe change did not affect the fluctuations while increasing the flow rates with the same syringe caused smoothening of fluctuations. Thus, it appears that the motor mechanism to be at fault at low flow rates. This effect though occasionally observed is of serious concern and the technique described here may be easily employed to mitigate the problem.

References

  • 1.Diba A. Infusion equipment and intravenous anaesthesia. In: Davey JA, Diba A, editors. Ward's Anaesthetic Equipment. 5th ed. Philadelphia, USA: Elsevier Saunders; 2005. pp. 409–26. [Google Scholar]
  • 2. [Last accessed on 2014 Oct 01];Orchestra® DPS+Data Sheet (Fresenius Kabi, Bad Homburg, Germany) [Google Scholar]
  • 3. [Last accessed on 2014 Jul 17]. Available from: http://www.hospira.com/Images/P11_3241_Infusion_Pump_Therapy_Pocket_Guide_FINAL_32-90336_1.pdf .
  • 4.Ferrari R, Beech DR. Infusion pumps: Guidelines and pitfalls. Aust Prescriber. 1995;18:49–51. [Google Scholar]
  • 5.Capes DF, Dunster KR, Sunderland VB, McMillan D, Colditz PB, McDonald C. Fluctuations in syringe-pump infusions: Association with blood pressure variations in infants. Am J Health Syst Pharm. 1995;52:1646–53. doi: 10.1093/ajhp/52.15.1646. [DOI] [PubMed] [Google Scholar]