Abstract
An optical characterization method is presented based on the use of the impulse response to characterize the damping imparted by the shell of an air-filled ultrasound contrast agent (UCA). The interfacial shell viscosity was estimated based on the unforced decaying response of individual echogenic liposomes (ELIP) exposed to a broadband acoustic impulse excitation. Radius versus time response was measured optically based on recordings acquired using an ultra-high-speed camera. The method provided an efficient approach that enabled statistical measurements on 106 individual ELIP. A decrease in shell viscosity, from 2.1 × 10−8 to 2.5 × 10−9 kg/s, was observed with increasing dilatation rate, from 0.5 × 106 to 1 × 107 s−1. This nonlinear behavior has been reported in other studies of lipid-shelled UCAs and is consistent with rheological shear-thinning. The measured shell viscosity for the ELIP formulation used in this study [κs = (2.1 ± 1.0) × 10−8 kg/s] was in quantitative agreement with previously reported values on a population of ELIP and is consistent with other lipid-shelled UCAs. The acoustic response of ELIP therefore is similar to other lipid-shelled UCAs despite loading with air instead of perfluorocarbon gas. The methods described here can provide an accurate estimate of the shell viscosity and damping for individual UCA microbubbles.
I. INTRODUCTION
Physical models for encapsulated ultrasound contrast agent (UCA) microbubbles have been developed and improved over the past two decades. The models give accurate predictions of the nonlinear radial dynamics of individual UCA microbubbles, particularly under forced, or acoustically driven, conditions.1 Such models are largely based on Rayleigh–Plesset-type equations, which describe the dynamics of a gas bubble in response to pressure variations. The effects of the shell encapsulation on the microbubble motion are incorporated by adding additional terms that describe the viscoelastic behavior of the shell.2 The presence of the shell increases the resonance frequency from that of a free bubble and damps the oscillations in response to the acoustic forcing.3 The shell also affects the nonlinear response of microbubbles, a key property of UCAs that is often exploited in diagnostic imaging modes for enhanced discrimination between the contrast agent and surrounding tissue.4
In addition to traditional diagnostic ultrasound imaging, there has been recent interest in advancing the applications of microbubbles for molecular imaging and therapeutic applications. These applications exploit the nonlinear response for selective detection of microbubbles in a given volume for diagnosis and specific targeting of disease. Molecular imaging techniques with targeted UCAs are being used increasingly for noninvasive diagnosis of inflammation, thrombus, and neovascularization.5 Targeted microbubble agents are also being developed for controlled drug-delivery applications.6 Clinical application of these UCAs necessitates accurate prediction of the frequency-dependent response of single isolated microbubbles. A better understanding of the shell properties that can be expected for individual UCA microbubbles within a population is also critical. The acoustic response of microbubbles can depend on the morphology and viscoelastic properties of the encapsulating shell,7 particularly if lipids are not homogeneously distributed throughout the shell.8 Therefore single microbubble characterization techniques and physical models that accurately describe the dynamics of an encapsulated microbubble are increasingly important.
The introduction of optical methods to resolve microbubble oscillations has provided new insights into the dynamic response of single microbubbles. Optical methods based on direct measurement of the bubble radius versus time can be used due to availability of ultra-high-speed imaging systems9,10 such as the Brandaris 128 fast-framing (0.5–25 Mfps) camera.11 This system enables optical characterization studies by imaging single microbubble dynamical phenomena occurring at multiple time scales.12 Acoustic measurements of the scattered echo from an isolated microbubble have been simultaneously recorded and compared to the optically measured radial dynamics to verify the effectiveness of this approach.13
Optical studies of the radial dynamics of UCAs typically rely on measuring the steady-state forced response of a microbubble to a narrow-band burst excitation. For example, the amplitude response of a microbubble exposed to multiple-cycle, low amplitude ultrasound bursts at various frequencies can be analyzed to build up a resonance curve and fit to a linearized model to derive the viscoelastic shell parameters.14 This microbubble spectroscopy technique14–17 requires multiple recordings to characterize the frequency-dependent response of an individual microbubble. In addition, the shell morphology can be altered or destroyed due to multiple burst excitations.18 A method using a single excitation would improve the efficiency of UCA characterization techniques and would minimize the effect of alterations to the microbubble shell under successive excitations.
In this study, we investigated the use of the impulse response to characterize the damping imparted by the encapsulation of microbubbles stabilized by a lipid shell, echogenic liposomes (ELIP).19 The shell viscosity is the dominant mechanism affecting the damping of microbubble response to acoustic excitation and most directly influences the onset of nonlinear oscillations. An understanding of this physical property of the shell encapsulation can be exploited for optimization of contrast-enhanced imaging and therapeutic applications. Through this technique, we aim to acquire a better understanding of the transient dynamics of ELIP in response to short-pulse excitations for both diagnostic and therapeutic applications. This method is also of particular interest to improve the efficiency of UCA characterization techniques in general as it requires only a single acoustic excitation when the impulse response is recorded using an ultra-high-speed camera.
ELIP are under development as theragnostic ultrasound contrast agents and differ from other commercially available UCAs primarily in shell material and gas content.4 ELIP formulations include a small amount of cholesterol to increase membrane rigidity and are echogenic due to the presence of air, which is entrapped and stabilized by the lipid during the rehydration process.19 The exact location of the entrapped air pockets, morphology of the encapsulation, and amount of air in each carrier has not been fully ascertained.19–21 The objective of the present study was to measure the damping due to the encapsulation and to estimate the shell viscosity based on the unforced response of individual ELIP exposed to a broadband acoustic impulse excitation.
An ultra-high-speed camera was used to capture the radius versus time response. During each recording, an isolated ELIP was excited using two ultrasound pulses with a phase difference of 180°. Following each pulse, the bubble oscillated at its natural frequency with diminishing amplitude determined by the damping, and eventually returned to its resting radius. A generic model was developed to estimate the shell viscosity based on the amplitude decay time constant, which was determined from the experimentally measured radius versus time curve for an individual ELIP. Gas diffusion was neglected in our model, and this approach was verified experimentally by comparing the resting radius before and after the impulse excitation. The shell viscosity of individual ELIP obtained using the optical method presented here is compared to our previously reported value obtained for a population of ELIP using a broadband acoustic attenuation spectroscopy technique.21
II. MATERIALS AND METHODS
A. Experimental setup
Freeze-dried ELIP dispersions consisting of L-α-phosphatidylcholine,1,2-dipalmitoyl-sn-glycero-3-phosphocholine (DPPC),1,2-dipalmitoyl-sn-glycero-3-phosphoethanolamine (DPPE),1,2-dipalmitoyl-sn-glycero-3-phospho-(1′-rac-glycerol) (DPPG),cholesterol (27:42:8:8:15, mol. %) [all lipids were purchased from Avanti Polar Lipids (Alabaster, AL)] were prepared as previously described.22 ELIP suspensions were prepared by reconstituting 10 mg/ml lyophilized lipid powder using air-saturated, filtered (type I) water at room temperature as described by Huang.19 The suspension was diluted (∼100×) into air-saturated phosphate-buffered saline (PBS) mixed with 0.5% (wt./vol.) bovine serum albumin (BSA) solution (Sigma Chemical Co., St. Louis, MO), loaded into an OptiCell® (Nunc/Thermo Scientific, Wiesbaden, Germany), and placed on an optical stage in a 37 °C water bath. A microscope with a 60× water-immersion objective and 2× magnifier was used to image the bubble dynamics at approximately 17 Mfps using the Brandaris 128 fast-framing camera.11,12 A diagram of the ultra-high-speed optical imaging setup is presented in Fig. 1.
Figure 2(a) shows an optical image of a dilute suspension of ELIP suspended in an OptiCell® using a 40× objective to obtain a larger field of view. A single frame from a Brandaris 128 recording using the 60× objective and 2× magnifier is shown in Fig. 2(b). Figures 2(c) and 2(d) show super-resolution fluorescence microscope images of fluorescently labeled ELIP obtained using the methods outlined by Kooiman et al.23 Briefly, a 4Pi confocal laser-scanning microscope (Leica TCS 4Pi) with two opposing objective lenses (100×, HCX PL APO, 1.35 NA) was used to generate super-resolution three-dimensional image stacks. For this supporting study, 2 mol. % fluorescently labeled 1,2-dipalmitoyl-sn-glycero-3-phosphoethanolamine-N-(lissamine rhodamine B sulfonyl) (rhodamine-DPPE) (Avanti Polar Lipids) was substituted into the ELIP formulation and the reconstituted liposomes were suspended in 87% (vol./vol.) glycerol between quartz cover slips at room temperature. The images show the distribution of the lipid molecules in the shell encapsulation of the ELIP. The size distribution of the microbubbles deduced from optical images presented in Figs. 2(a) and 2(c) is consistent with Coulter counter measurements of the particle size distribution for this ELIP formulation, which indicated a volume-weighted mean diameter of 2 μm.21 Previously published transmission electron microscopy images have shown liposomes on the order of 2 μm diameter24 as well as vesicles in the nanometer size range that could not be resolved using the optical methods described here.25,26 Atomic force microscopy20 and fluorescence microscopy studies26 also indicate a broad size distribution, with particle sizes ranging over three orders of magnitude from tens of nanometers to microns.
B. Acoustic excitation pulse
A focused, broadband PVDF transducer (23 mm diameter, 25 mm focal distance, PA275; Precision Acoustics, Dorchester, UK) was positioned in the water bath at a 45° angle below the sample and the acoustic focus (0.5 mm full-width at half-maximum pressure) aligned with the optical field of view. To detect potential nonlinear bubble dynamics, a pulse inversion pair was utilized to excite individual ELIP. For each optical recording, a sequence of two phase-inverted acoustic pulses, temporally separated by 3 μs, was used to excite each ELIP impulsively. The pulses had a phase difference of 180° but were otherwise identical with center frequency of 4 MHz and pulse duration of ∼0.33 μs (1.5 cycles). Pulse excitations with peak pressure amplitude of 125, 250, or 500 kPa (210, 420, or 840 kPa peak-to-peak acoustic pressure, respectively) were generated by a programmable arbitrary waveform generator (8026, Tabor Electronics Ltd., Tel Hanan, Israel) and amplified using a wideband RF amplifier (0.3–35 MHz, A-500; Electronic Navigation Industries, Rochester, NY). During calibration, a 0.2 -μm needle-type hydrophone (Precision Acoustics) was positioned approximately 2 mm from the sample holder membrane (a modified OptiCell® with one membrane removed), such that the hydrophone location corresponded to the location of the ELIP during the optical measurement. The transmitted acoustic pressure amplitude in situ was attenuated by a factor of 3 dB relative to the free-field pressure due to the presence of the membrane and the 45° angle of incidence of the acoustic wave. The in situ pressure waveform and power spectrum for a 250 kPa peak pressure amplitude excitation pulse are shown in Fig. 3.
C. Estimation of the shell viscosity
Linearization of a Rayleigh–Plesset-type bubble dynamics equation yields the equation of motion for a damped harmonic oscillator. In the absence of forcing this equation can be written,1
(1) |
The eigenfrequency, ω0, is given by the Minnaert relation for a free gas bubble,17 and for an encapsulated bubble, this expression can be modified to include an additional term that depends on the elasticity of the shell.27 The damped natural frequency for unforced oscillations, ωd, is related to the eigenfrequency of the system by the damping coefficient, , where δ represents the total damping of the system.28,29 If the system is modeled as a linearized, single degree-of-freedom system which is underdamped (δ < 2), the solution can be written in terms of an oscillatory signal with a decaying exponential envelope characterized by the time constant, .
Analysis of the experimental data consists of calculating the envelope signal, A(t), from the measured radius versus time curves for each individual pulse excitation using the Hilbert transform. The time constant can be readily determined by fitting a line to the natural logarithm of the envelope signal,30 with a slope given by
(2) |
In general, the total damping consists of contributions from all of the dissipative loss mechanisms, including: Acoustic radiation (δrad), thermal diffusion and conduction (δth), viscosity of the surrounding liquid (δvis), and viscosity of the shell (δsh), such that δ = δrad + δth + δvis + δsh.28 For an encapsulated bubble, the last term represents the damping imparted by the shell,3 and for an unencapsulated bubble, δsh = 0. Assuming a Newtonian rheology, the expression for δsh in terms of the total non-dimensional damping coefficient can be written,31
(3) |
where δ0 = δrad + δth + δvis represents the total damping coefficient evaluated at resonance for an unencapsulated gas bubble with the same radius and eigenfrequency as the encapsulated bubble. The shell viscosity in dimensional units, κs, can be estimated as a function of the time constants, τ and τ0,
(4) |
where τ and τ0 are the time constants for an encapsulated bubble (measured) and unencapsulated bubble with the same initial radius (calculated), respectively.
The expression given in Eq. (2) for the interfacial shell viscosity is consistent with the Rayleigh–Plesset-type model for phospholipid-encapsulated bubbles developed by Marmottant et al.32 In this model, the effective surface tension is described by three regimes of shell behavior: Buckled, elastic, or ruptured. In the elastic regime, the shell contributes an additional restoring force, characterized by the elasticity χ, which shifts the resonance for small amplitude oscillations of an encapsulated bubble to higher frequencies. Oscillations in the purely elastic regime occur only at exceedingly low vibrational amplitudes [e.g., R/R0 < 1.01 for χ = 2.5 N/m (Ref. 17)]. Above a threshold radius, the shell is in the ruptured regime and the interfacial dynamics are also controlled by the free air-water surface tension. We estimate the rupture radius for ELIP to be R/R0 = 1.02 based on the elasticity determined previously [χ = 1.55 N/m (Ref. 21)]. Overvelde et al.17 demonstrated that for oscillations exceeding this amplitude, the bubble is no longer oscillating in the purely elastic regime and the natural frequency for a encapsulated bubble approaches a value that is only slightly greater (∼10%) than would be predicted for an unencapsulated bubble. For simplicity, we consider ω0 to be the same for both encapsulated and unencapsulated bubbles and use the eigenfrequency of an unencapsulated bubble [e.g., Eq. (5) in Ref. 26] to calculate τ0 in Eq. (4).
Each ELIP was excited by both peak-negative and peak-positive pulses, which were analyzed separately. A subset of ELIP were excited several times using 125, 250, or 500 kPa peak pressure amplitude pulses. An example of the damping analysis for an individual 0° pulse excitation is shown in Figs. 4(a) and 4(b). The natural frequency of unforced oscillation was estimated from the Fourier spectrum of the radius versus time response, shown in Fig. 4(c).
D. Simulations
To evaluate the validity of the shell viscosity estimates obtained for individual ELIP, radius versus time curves were simulated using the dynamical model for lipid-encapsulated microbubbles developed by Marmottant et al.32 The measured acoustic pressure waveforms (see Fig. 3) were used as the forcing function for the bubble dynamics simulations that were compared with the experimental data. The values of the physical parameters used for the simulations were: Ambient pressure P0 = 100 kPa, density of the liquid, ρ = 1007 kg/m3, viscosity of the liquid (at 37 °C), μ = 0.76 × 10–3 kg/(m s), surface tension of the free air-water interface, σwater = 0.072 N/m, and speed of sound, c = 1536 m/s. These parameters correspond to air-filled microbubbles suspended in PBS with 0.5% BSA solution at 37 °C.21 The effective polytropic exponent was calculated from the ratio of specific heats using the expression given by Hoff et al.33 following other researchers who have used similar expressions.34 The parameters used in the Marmottant model for describing the interfacial rheological properties of the shell are the elasticity, χ = 0.1–1.55 N/m, and the shell surface viscosity, κs, which was estimated based on experimental measurements as described in Sec. II C. Marmottant et al.32 and others17 have demonstrated that a phospholipid shell stabilizes a quiescent gas bubble by counteracting the Laplace pressure, therefore most microbubbles have an initial surface tension much lower than the surface tension of the free air-water interface, σwater.35 In this study, we assume the shell encapsulation is initially in the pre-buckled state (i.e., the transition between the elastic and buckled regions) and consequently the initial surface tension was taken as σ0 = 0 for all numerical simulations.
E. Excitation of ELIP with a tone burst
A subset of ELIP was also excited using a 3 MHz tone burst to record the radius versus time response to acoustic forcing. Burst excitations were generated using the same arbitrary waveform generator and transducer but consisted of a 10-cycle, 3-MHz sinusoidal wave with a 2-cycle Gaussian ramp and taper. The in situ acoustic pressure amplitude for the tone burst excitation was 180 kPa, calibrated using the same methods as described in the preceding text. An impulse response excitation was recorded prior to the tone burst excitation to obtain an estimate of the shell viscosity using the methods described in the preceding text. This data acquisition scheme enabled a post hoc comparison of the measured forced response to simulations using the estimated value of shell viscosity for a particular liposome.
III. RESULTS
A total of 476 radius versus time curves for 106 individual ELIP were analyzed. The equilibrium radii of the individual ELIP ranged between 0.9 and 3.4 μm. Only trials for which the microbubble did not show any dissolution were analyzed in this study (|Δr| < 0.1 μm, where Δr is the difference between the bubble resting radius before and after the acoustic excitation). A small number of recordings (12 in total, <3%) with |Δr| > 0.1 were therefore not included in the analysis.
Figure 5 shows the estimated natural frequency obtained from the Fourier spectrum of the individual radius versus time responses. The median value of the measured resonance frequency was 16% higher than the eigenfrequency of an unencapsulated bubble, given by 2πω0.
Results of the measured shell viscosity as function of size are shown in Fig. 6(a). Due to the large number of individual data points, the mean value and standard deviation of the points within radius bins spaced by 0.2 μm are also shown for clarity. There is a considerable spread of the shell viscosity estimates among microbubbles of the same size range (indicated by vertical error bars). However, the estimated shell viscosity increases with increasing bubble size, as has been shown in other investigations of lipid-shelled agents.14,36,37 The shell viscosity estimates range from 2.1 × 10–9 kg/s for microbubbles of 0.9–1.4 μm radius to 2.3 × 10–8 kg/s for microbubbles of 2.5–3.4 μm radius. These values are the same order of magnitude as values obtained in other studies based on population estimates of commercially available lipid-shelled microbubbles such as SonoVue® (κs = 5.4 × 10–9 kg/s),38 Sonazoid® (κs = 1.2 × 10–8 kg/s),39 and Definity® (κs = 3 × 10–9 kg/s).40
The dependence of the shell viscosity on the maximum dilatation rate is shown in Fig. 6(b). The maximum dilatation rate ()max was determined directly from the experimentally measured radius versus time curves by calculating the derivative with respect to time. Despite the dispersion in the individual shell viscosity estimates, this plot demonstrates a decrease of the shell viscosity with increasing dilatation rate. By fitting the data to a power law relation of the form,41 , the 95% confidence interval for the power-law index n is between 0.1 and 0.25. A power-law index less than one is consistent with rheological shear-thinning behavior and has been reported earlier for lipid-shelled bubbles.14 For low dilatation rates (<0.50 × 106 s–1), the mean value and standard deviation of the shell viscosity estimates in this study was κs = (2.1 ± 1.0) × 10–8 kg/s (n = 23).
Figure 7 shows three examples of numerically simulated and experimentally measured radius versus time curves for individual ELIP with equilibrium radii of 1.5, 2.3, and 3.0 μm. The excitation peak pressure amplitude for all cases was 250 kPa. Simulations were carried out using the experimentally derived shell viscosity (κs) for each individual ELIP (i.e., the average of the estimates derived from both 0° and 180° impulses). The shell elasticity (χ) was varied over one order of magnitude with the upper limit corresponding to the value previously determined for ELIP in the linear regime, χ = 0.1–1.55 N/m.21 The numerical simulations show good agreement with the experimentally measured transient impulse response dynamics for each case.
Figure 8 shows an example of simulated and experimentally measured radius versus time curves for a single ELIP exposed to three different peak pressure amplitude pulses: 125, 250, and 500 kPa. As expected, the shell viscosity values obtained from each independent excitation are similar. The estimates of κs for each pressure are 5.7, 5.6, and 5.6 × 10–9 kg/s, respectively.
IV. DISCUSSION
The goal of this study was to improve the understanding of the transient oscillation dynamics and shell properties of ELIP. We estimated the damping and shell viscosity from the unforced response of individual ELIP excited by a broadband acoustic impulse excitation measured using ultra-high-speed imaging. This technique can achieve the measurement of individual ELIP in a single run using the Brandaris 128 ultra-high-speed framing camera and provides an efficient alternative approach to the microbubble spectroscopy technique as reported previously.14–16 The results derived from estimates for individual ELIP were in good agreement with our previously obtained values based on bulk acoustic attenuation measurements. Overall the shell viscosity estimates for ELIP in this study are consistent with values reported for commercially available lipid-shelled UCAs. Furthermore, we found that the Marmottant model accurately describes the transient dynamics of an individual ELIP when compared with radius versus time measurements obtained from ultra-high-speed recordings. Thus despite the encapsulation of air instead of perfluorocarbon gas, the viscoelastic shell properties of ELIP are similar to other lipid-shelled agents such as SonoVue® and Definity®.
A. Natural frequency of unforced oscillations
The eigenfrequency of an unencapsulated bubble is used to calculate the time constant, τ0, in Eq. (4). To evaluate the validity of the assumption that the surface tension dominates the bubble dynamics, we obtained estimates of the natural frequency from the Fourier transform of the radius versus time curves (Fig. 5). The median natural frequency measured was 16% higher than the eigenfrequency for an unencapsulated bubble, given by 2πω0. A 16% error in the natural frequency translates to a ∼4% change in τ0 (and therefore in our estimates of κs) for the size range of bubbles studied.
B. Contributions to damping effects
The total damping coefficient is the summation of the shell damping (δsh) and three other damping terms: The radiation damping (δrad), the thermal damping (δth), and the viscous damping due to the surrounding liquid (δvis). Theoretical expressions for predicting the damping factors of a spherically oscillating gas bubble are well-established,28 and the latter three terms can be readily calculated based on the known equilibrium size and eigenfrequency of the microbubble. Figure 9 shows the resonance damping coefficients versus radius for an air bubble under the experimental conditions used in this study. Viscous damping (δvis or δsh) dominates, and the radiation damping is approximately an order of magnitude smaller for bubbles in the size range considered in this study. However, thermal damping can be appreciable and is included in our model for air-filled microbubbles at resonance.
Thermal damping is not accounted for explicitly in the Rayleigh–Plesset formulation and is often neglected in linearized models of UCAs as well.1 However, thermal damping can be significant for resonant bubbles larger than about 1 μm and is dominant for resonant bubbles larger than ∼10 μm.3 Figure 9 shows that for a 1.3 μm radius bubble, δth is approximately equal to δrad and only amounts to 15% of the viscous term δvis. For a bubble of about 3.5 μm radius, δth is approximately equal to the viscous damping coefficient. In this study, we estimated the thermal damping and polytropic exponent based on the resting radius of the microbubble using the expressions given by Hoff et al.33 It should be noted that the values of δth shown in Fig. 9 were calculated for an unencapsulated air bubble at resonance and include the effects of surface tension on the equilibrium gas pressure. However, δth may be less significant for an encapsulated bubble due to a lower equilibrium gas pressure resulting from the reduced initial surface tension for a bubble with a stabilizing shell (for example, σ0 = 0 for a pre-buckled bubble as was assumed in this study).42 Therefore we may have a slight underestimation of the shell damping effect.
C. Shell viscosity
Overall our estimates of the shell viscosity of individual ELIP based on the optical method developed in this study are consistent with our previous measurements obtained using a broadband attenuation spectroscopy technique.21,25 For low dilatation rates (< 0.5 × 10–6 s–1), the average value of the shell viscosity in this study [κs = (2.1 ± 1.0)× 10–8 kg/s, n = 23] is in agreement with our previously reported bulk acoustic measurements in the linear regime [κs = (2.0 ± 0.14) × 10–8 kg/s].21 Moreover the shell viscosity estimates derived from damping measurements on 106 individual ELIP in this study indicate an apparent increase of the shell viscosity with increasing radius and a decrease with increasing dilatation rate, indicative of shear-thinning behavior. This observation is consistent with previous measurements on commercially available lipid-shelled microbubbles such as SonoVue® (Ref. 14) and Definity® (Ref. 37) but is not explained by existing shell models where the viscosity is assumed to be constant over all radii. However, there have recently been attempts to account for this behavior using ad-hoc rheological models.43,44
The shell viscosity estimates based on the optical approach also showed considerable variations among ELIP of the same size range (see Fig. 6). Variations in shell properties and acoustic response have also been reported in previous studies on lipid-shelled microbubbles. Kwan and Borden35 and Kooiman et al.23 present convincing microscopic images of heterogeneous lipid domains and ascribe this phenomenon to phase separation of the different lipid species on the bubble surface. Similar variations in the local lipid concentration on the shell have been found to affect the nonlinear behavior of DSPC microbubbles at high frequencies,45 although no clear relation between the nonlinear response and shell microstructure was evident. Hosny et al.7 quantified the spatial distribution of viscosity in the microbubble shell using a fluorescence lifetime imaging technique and also found a large variation in viscosity that was correlated to differences in the ultrasound response of microbubbles of similar size. Based on the chemical composition and preparation method of ELIP, we expect similar inhomogeneities in local lipid concentration as well as variation between individual ELIP particles within a population. Moreover, lipid diffusion and the phase transition influenced by the surrounding temperature can be other key factors affecting the heterogeneity of the ELIP shell.8,22
Nanostructural surface morphology that could not be resolved using the optical techniques in this study (e.g., Fig. 2) may nevertheless play a large role in the dynamic behavior of ELIP. For lipid-encapsulated bubbles, the shell is usually assumed to be a monolayer due to thermodynamic considerations.35 However, the shell thickness has not been directly measured for microscopic encapsulated gas bubbles and estimates based on the dynamic response of the microbubble only give information about the interfacial rheological properties of the microbubble (i.e., zero-thickness interface model).20 Therefore the large variation in viscosity estimates could be due to the number of lipid layers surrounding the encapsulated microbubbles.7 The resolution of the fluorescence images presented in Fig. 2 is limited by the focal spot size, which is about 220 nm in the lateral dimension. Thus we are unable to conclude whether the shell is a lipid monolayer or if it is multilamellar. Multilamellar vesicles, in which the lipid monolayer immediately adjacent to the gas bubble is surrounded by one or more bilayers, have been observed in transmission electron microscopy images of ELIP.25 The spread of the shell viscosity estimates for ELIP of similar size may indicate the presence of multilamellar vesicles, where the apparent viscosity depends on the number of lamallae.
Another factor affecting the interpretation of the variation in the shell viscosity estimates is the uncertainty, which can be approximated as the random error associated with separate measurements on the same ELIP. We calculated the random error by considering the relative difference between the values obtained from separate analysis of the two phase-inverted excitation pulses (0° and 180°), i.e., the absolute difference between the two shell viscosity estimates normalized by the average of the two estimates. The median random error was 40% which is approximately the same as the standard deviation shown in Figs. 6(a) and 6(b).
D. Impulse response measurements and analysis
Evaluation of the damping coefficient based on the impulse response is a new and efficient approach applied to characterization of lipid-shelled ELIP in this study. A broadband pulse was applied to acquire the radius versus time impulse response of an individual ELIP, requiring only a single exposure of a few microseconds using an ultra-high-speed camera. Using the Brandaris 128, we were able to record the unforced, exponentially decaying amplitude response from two acoustic impulse excitations, a pulse inversion pair, during each high speed recording (128 frames). This method provides an efficient approach that enabled statistical measurements on a large number of individual ELIP. Additionally, data analysis was straightforward, and the simulated transient response of the lipid-encapsulated microbubble accurately tracked the measured radius versus time curves (Figs. 7 and 8).
We also assessed the validity of this technique by comparing the forced response measured using a 10-cycle narrow-band burst excitation with numerical simulations. An example is shown in Fig. 10 in which simulations using two different shell viscosity values (derived from the optical method described here or the bulk acoustic method reported in Raymond et al.,21 respectively) are compared with experimental measurements. Excellent agreement was found between the simulated and measured radius versus time curves when the effect of the shell was taken into account using the viscosity estimate obtained in this study. For the example shown in Fig. 10(a), the shell viscosity estimate (κs = 3.0 × 10–9 kg/s) was obtained during a previous impulse response exposure of the same particular liposome. Alternatively, using the population-averaged shell viscosity obtained in Raymond et al.21 (κs = 2.0 × 10–8 kg/s) results in an under-prediction of the amplitude of the response to narrow-band forcing [Fig. 10(b)]. Use of the estimated value obtained for a specific ELIP in this study allows one to obtain much more precise agreement between the experimental theoretical data using the Marmottant model.
E. Implications for contrast-enhanced imaging
The size range of ELIP is known to be polydisperse with particle sizes ranging from tens of nanometers to several microns.21,24,25 Differences in the characteristics of nanoscale versus microscale ELIP vesicles may play a role in the scattering properties of individual ELIP. This study found that smaller ELIP are characterized by a substantially lower shell viscosity, which may contribute to increased nonlinear scattering, especially at higher frequencies. Such nonlinear behavior of the smaller populations of ELIP could be exploited for diagnostic ultrasound applications that utilize higher frequencies, such as intravascular ultrasound. For example, the nonlinear acoustic signatures of ELIP could be utilized for improved imaging of pathology in the vasa vasorum, the proliferative small vessels that play a role in atheroma progression in the cardiovascular system.
Goertz et al.46 demonstrated that sub-populations of the lipid-shelled clinical UCA Definity® exhibited different frequency-dependent scattering properties as a function of size. These authors hypothesized that differences in the encapsulation microstructure for small and large bubbles may be a key factor influencing the nonlinear scattering at high frequencies. A subsequent study by Helfield et al.47 demonstrated that smaller sub-populations of lipid-shelled microbubbles were not only resonant at higher frequencies but were also characterized by a substantially lower shell viscosity, suggesting a possible frequency dependence of the lipid shell properties. It is still not clear whether this effect is due solely to the size or if frequency-dependent considerations need to be taken into account in the dynamical models. Doinikov et al.44 point out that existing shell models may not capture the observed radius or frequency dependence of the shell material properties and that full description of the rheological properties of the shell may require the use of more complex models in order to describe the shear-thinning behavior. A more detailed exploration of alternative rheological models could be explored in future studies.
F. Limitations
Inherent limitations of the optical system resolution result in a bias of the size range of ELIP selected for analysis in this study. The optical resolution of the ultra-high-speed imaging system was 0.4 μm (Ref. 11), and the smallest individual ELIP that was measured in this study was several times larger than the optical resolution limit. The size range of ELIP considered here focused on measurable particles larger than 0.9 μm in radius only and smaller ELIP were not considered for investigation.
The proximity of the microbubble to the OptiCellTM membrane may affect the measured response of ELIP to acoustic excitation. Previous studies have found that the presence of a membrane near an oscillating microbubble can affect the resonance frequency and oscillation amplitude in response to acoustic forcing.48,49 Functionalized UCAs such as ELIP are being developed for targeted imaging and drug-delivery applications, and therefore similar effects are expected to play a role in vivo. The acoustic response of adherent targeted bubbles may be influenced by the ligand distribution and bubble-wall interaction.23 Therefore future work to investigate the damping for adherent targeted agents using this approach is recommended.
The simulations presented in Figs. 7, 8, and 10 are not unique solutions because there is more than one free parameter in the Marmottant model, namely the shell elasticity, χ, and initial surface tension, σ0. We have chosen to assume the phospholipid shell stabilizes a gas bubble by counteracting the Laplace pressure, so that σ0 = 0, throughout this paper. Using empirical fits, the values for the data presented in Fig. 7 are: χ = 0.1, 0.2, and 1.55 N/m for R0 = 1.5, 2.3, and 3.0 μm, respectively. A radius-dependent shell elasticity parameter has also been observed and modeled by Chetty et al.9 and Doinikov et al.44 The measured oscillation amplitudes were not always in the “elastic” regime range of the Marmottant model (R/R0 < 1.02). However, the 3.0 μm bubble did behave near this regime (R/R0 ∼ 1.07) and the shell elasticity determined from the optical measurements agreed with previous measurements.21
V. CONCLUSIONS
In this study, a model based on time domain analysis was developed to estimate the shell viscosity for individual ELIP using optically measured radius versus time curves. The decay time constant was measured from the unforced response of the ELIP following excitation by a short acoustic pulse. At low dilatation rates, the shell viscosity is in quantitative agreement with our previously determined values for ELIP. The viscoelastic shell of ELIP was found to behave similarly to other lipid-shelled agents. We demonstrated that qualitative agreement between the measured and simulated radius versus time curves under transient and steady-state acoustic forcing is highly dependent on the shell viscosity. The methods described here can provide accurate estimates of the shell viscosity and damping for individual UCA microbubbles. In contrast to acoustic measurements of a microbubble population with a relatively wide size distribution, this method provides some insight on the variation of properties among single particles within a population. This method also enables efficient measurements on a large number or individual UCA microbubbles because it is based on the response to a single acoustic impulse excitation.
ACKNOWLEDGMENTS
The authors would like to thank Robert Beurskens and Frits Mastik from the Department of Biomedical Engineering, Erasmus MC, for technical assistance with Brandaris experiments. The authors would also like to thank the Erasmus Optical Imaging Centre of Erasmus MC for use of their facilities and Wiggert A. van Cappellen and Alex Nigg from the Erasmus Optical Imaging Centre, Department of Pathology, Erasmus MC, for technical assistance. J.L.R. was supported by a fellowship from the Whitaker International Program administered by the Institute of International Education. This work was supported in part by the U.S. Department of Health and Human Services, National Institutes of Health (Grant Nos. NIH R01HL74002 and NIH R01HL059586) and NanoNextNL, a micro and nanotechnology consortium of the Government of the Netherlands and 130 partners. This work was performed while the first author was a visiting fellow at the Department of Biomedical Engineering, Thoraxcenter, Erasmus MC, Rotterdam, Netherlands.
Portions of this work were presented in “Impulse response of echogenic liposomes,” Proceedings of the 19th European Symposium on Ultrasound Contrast Imaging, Rotterdam, Netherlands, January 2014.
References
- 1.Faez T., Emmer M., Kooiman K., Versluis M., Van Der Steen A., and De Jong N., “ 20 years of ultrasound contrast agent modeling,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control 60, 7–20 (2013) 10.1109/TUFFC.2013.2533. [DOI] [PubMed] [Google Scholar]
- 2.Doinikov A. A. and Bouakaz A., “ Review of shell models for contrast agent microbubbles,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control 58, 981–993 (2011). 10.1109/TUFFC.2011.1899 [DOI] [PubMed] [Google Scholar]
- 3.Jong N. de, Bouakaz A., and Frinking P., “ Basic acoustic properties of microbubbles,” Echocardiography 19, 229–240 (2002). 10.1046/j.1540-8175.2002.00229.x [DOI] [PubMed] [Google Scholar]
- 4.Stride E., “ Physical principles of microbubbles for ultrasound imaging and therapy,” Cerebrovasc. Dis. 27 Suppl. 2, 1–13 (2009). 10.1159/000203122 [DOI] [PubMed] [Google Scholar]
- 5.Wen Q., Wan S., Liu Z., Xu S., Wang H., and Yang B., “ Ultrasound contrast agents and ultrasound molecular imaging,” J. Nanosci. Nanotechnol. 14, 190–209 (2014). 10.1166/jnn.2014.9114 [DOI] [PubMed] [Google Scholar]
- 6.Castle J., Butts M., Healey A., Kent K., Marino M., and Feinstein S. B., “ Ultrasound-mediated targeted drug delivery: Recent success and remaining challenges,” Am. J. Physiol. Heart Circ. Physiol. 304, H350–H357 (2013). 10.1152/ajpheart.00265.2012 [DOI] [PubMed] [Google Scholar]
- 7.Hosny N. A., Mohamedi G., Rademeyer P., Owen J., Wu Y., Tang M.-X., Eckersley R. J., Stride E., and Kuimova M. K., “ Mapping microbubble viscosity using fluorescence lifetime imaging of molecular rotors,” Proc. Natl. Acad. Sci. U.S.A. 110, 9225–9230 (2013). 10.1073/pnas.1301479110 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Borden M., “ Nanostructural features on stable microbubbles,” Soft Matter 5, 716–720 (2009). 10.1039/B815506B [DOI] [Google Scholar]
- 9.Chetty K., Stride E., Sennoga C., Hajnal J., and Eckersley R., “ High-speed optical observations and simulation results of SonoVue microbubbles at low-pressure insonation,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control 55, 1333–1342 (2008). 10.1109/TUFFC.2008.796 [DOI] [PubMed] [Google Scholar]
- 10.Chen X., Wang J., Versluis M., De Jong N., and Villanueva F. S., “ Ultra-fast bright field and fluorescence imaging of the dynamics of micrometer-sized objects,” Rev. Sci. Instrum. 84, 063701 (2013). 10.1063/1.4809168 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Chin C. T., Lancée C., Borsboom J., Mastik F., Frijlink M. E., De Jong N., Versluis M., and Lohse D., “ Brandaris 128: A digital 25 million frames per second camera with 128 highly sensitive frames,” Rev. Sci. Instrum. 74, 5026–5034 (2003). 10.1063/1.1626013 [DOI] [Google Scholar]
- 12.Gelderblom E. C., Vos H. J., Mastik F., Faez T., Luan Y., Kokhuis T. J. A., Van Der Steen A. F. W., Lohse D., De Jong N., and Versluis M., “ Brandaris 128 ultra-high-speed imaging facility: 10 years of operation, updates, and enhanced features,” Rev. Sci. Instrum. 83, 103706 (2012). 10.1063/1.4758783 [DOI] [PubMed] [Google Scholar]
- 13.Sijl J., Vos H. J., Rozendal T., De Jong N., Lohse D., and Versluis M., “ Combined optical and acoustical detection of single microbubble dynamics,” J. Acoust. Soc. Am. 130, 3271–3281 (2011). 10.1121/1.3626155 [DOI] [PubMed] [Google Scholar]
- 14.van der Meer S. M., Dollet B., Voormolen M. M., Chin C. T., Bouakaz A., Jong N. de, Versluis M., and Lohse D., “ Microbubble spectroscopy of ultrasound contrast agents,” J. Acoust. Soc. Am. 121, 648–656 (2007). 10.1121/1.2390673 [DOI] [PubMed] [Google Scholar]
- 15.Faez T., Skachkov I., Versluis M., Kooiman K., and Jong N. de, “ In vivo characterization of ultrasound contrast agents: Microbubble spectroscopy in a chicken embryo,” Ultrasound Med. Biol. 38, 1608–1617 (2012). 10.1016/j.ultrasmedbio.2012.05.014 [DOI] [PubMed] [Google Scholar]
- 16.Luan Y., Faez T., Gelderblom E., Skachkov I., Geers B., Lentacker I., van der Steen T., Versluis M., and Jong N. de, “ Acoustical properties of individual liposome-loaded microbubbles,” Ultrasound Med. Biol. 38, 2174–2185 (2012). 10.1016/j.ultrasmedbio.2012.07.023 [DOI] [PubMed] [Google Scholar]
- 17.Overvelde M., Garbin V., Sijl J., Dollet B., Jong N. de, Lohse D., and Versluis M., “ Nonlinear shell behavior of phospholipid-coated microbubbles,” Ultrasound Med. Biol. 36, 2080–2092 (2010). 10.1016/j.ultrasmedbio.2010.08.015 [DOI] [PubMed] [Google Scholar]
- 18.Thomas D. H., Butler M., Anderson T., Emmer M., Vos H., Borden M., Stride E., De Jong N., and Sboros V., “ The quasi-stable lipid shelled microbubble in response to consecutive ultrasound pulses,” Appl. Phys. Lett. 101, 071601 (2012). 10.1063/1.4746258 [DOI] [Google Scholar]
- 19.Huang S.-L., “ Ultrasound-responsive liposomes,” in Liposomes: Methods and Protocols, Methods in Molecular Biology, edited by Weissig V. ( Humana, New York, 2010), Vol. 605, pp. 113–128. [DOI] [PubMed] [Google Scholar]
- 20.Paul S., Nahire R., Mallik S., and Sarkar K., “ Encapsulated microbubbles and echogenic liposomes for contrast ultrasound imaging and targeted drug delivery,” Comput. Mech. 53, 413–435 (2014). 10.1007/s00466-013-0962-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Raymond J. L., Haworth K. J., Bader K. B., Radhakrishnan K., Griffin J. K., Huang S.-L., McPherson D. D., and Holland C. K., “ Broadband attenuation measurements of phospholipid-shelled ultrasound contrast agents,” Ultrasound Med. Biol. 40, 410–421 (2014). 10.1016/j.ultrasmedbio.2013.09.018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Buchanan K. D., Huang S., Kim H., MacDonald R. C., and McPherson D. D., “ Echogenic liposome compositions for increased retention of ultrasound reflectivity at physiologic temperature,” J. Pharm. Sci. 97, 2242–2249 (2008). 10.1002/jps.21173 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Kooiman K., Kokhuis T. J. A., van Rooij T., Skachkov I., Nigg A., Bosch J. G., van der Steen A. F. W., van Cappellen W. A., and de Jong N., “ DSPC or DPPC as main shell component influences ligand distribution and binding area of lipid-coated targeted microbubbles,” Eur. J. Lipid Sci. Technol. 116, 1217–1227 (2014). 10.1002/ejlt.201300434 [DOI] [Google Scholar]
- 24.Paul S., Russakow D., Nahire R., Nandy T., Ambre A. H., Katti K., Mallik S., and Sarkar K., “ In vitro measurement of attenuation and nonlinear scattering from echogenic liposomes,” Ultrasonics 52, 962–969 (2012). 10.1016/j.ultras.2012.03.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Kopechek J. A., Haworth K. J., Raymond J. L., Douglas Mast T., S. R. Perrin, Jr. , Klegerman M. E., Huang S., Porter T. M., McPherson D. D., and Holland C. K., “ Acoustic characterization of echogenic liposomes: Frequency-dependent attenuation and backscatter,” J. Acoust. Soc. Am. 130, 3472–3481 (2011). 10.1121/1.3626124 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Nahire R., Paul S., Scott M. D., Singh R. K., Muhonen W. W., Shabb J., Gange K. N., Srivastava D. K., Sarkar K., and Mallik S., “ Ultrasound enhanced matrix metalloproteinase-9 triggered release of contents from echogenic liposomes,” Mol. Pharm. 9, 2554–2564 (2012). 10.1021/mp300165s [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Katiyar A. and Sarkar K., “ Excitation threshold for subharmonic generation from contrast microbubbles,” J. Acoust. Soc. Am. 130, 3137–3147 (2011). 10.1121/1.3641455 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Ainslie M. A. and Leighton T. G., “ Review of scattering and extinction cross-sections, damping factors, and resonance frequencies of a spherical gas bubble,” J. Acoust. Soc. Am. 130, 3184–3208 (2011). 10.1121/1.3628321 [DOI] [PubMed] [Google Scholar]
- 29.Leander J. L., “ On the eigenfrequency of a gas bubble in a liquid,” J. Acoust. Soc. Am. 102, 1900–1903 (1997). 10.1121/1.420113 [DOI] [Google Scholar]
- 30.Agneni A. and Balis-Crema L., “ Damping measurements from truncated signals via Hilbert transform,” Mech. Syst. Signal Process. 3, 1–13 (1989). 10.1016/0888-3270(89)90019-8 [DOI] [Google Scholar]
- 31.Chatterjee D. and Sarkar K., “ A Newtonian rheological model for the interface of microbubble contrast agents,” Ultrasound Med. Biol. 29, 1749–1757 (2003). 10.1016/S0301-5629(03)01051-2 [DOI] [PubMed] [Google Scholar]
- 32.Marmottant P., Van Der Meer S., Emmer M., Versluis M., De Jong N., Hilgenfeldt S., and Lohse D., “ A model for large amplitude oscillations of coated bubbles accounting for buckling and rupture,” J. Acoust. Soc. Am. 118, 3499–3505 (2005). 10.1121/1.2109427 [DOI] [Google Scholar]
- 33.Hoff L., Sontum P. C., and Hovem J. M., “ Oscillations of polymeric microbubbles: Effect of the encapsulating shell,” J. Acoust. Soc. Am. 107, 2272–2280 (2000). 10.1121/1.428557 [DOI] [PubMed] [Google Scholar]
- 34.Crum L. A., “ Polytropic exponent of a gas contained within air bubbles pulsating in a liquid,” J. Acoust. Soc. Am. 73, 116–120 (1983). 10.1121/1.388844 [DOI] [Google Scholar]
- 35.Kwan J. J. and Borden M. A., “ Lipid monolayer collapse and microbubble stability,” Adv. Colloid Interface Sci. 183–184, 82–99 (2012). 10.1016/j.cis.2012.08.005 [DOI] [PubMed] [Google Scholar]
- 36.Morgan K. E., Allen J. S., Dayton P., Chomas J. E., Klibanov A. L., and Ferrara K. W., “ Experimental and theoretical evaluation of microbubble behavior: Effect of transmitted phase and bubble size,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control 47, 1494–1509 (2000). 10.1109/58.883539 [DOI] [PubMed] [Google Scholar]
- 37.Tu J., Swalwell J. E., Giraud D., Cui W., Chen W., and Matula T. J., “ Microbubble sizing and shell characterization using flow cytometry,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control 58, 955–963 (2011). 10.1109/TUFFC.2011.1896 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Gorce J.-M., Arditi M., and Schneider M., “ Influence of bubble size distribution on the echogenicity of ultrasound contrast agents: A study of SonoVueTM,” Invest. Radiol. 35, 661–671 (2000). 10.1097/00004424-200011000-00003 [DOI] [PubMed] [Google Scholar]
- 39.Sarkar K., Shi W. T., Chatterjee D., and Forsberg F., “ Characterization of ultrasound contrast microbubbles using in vitro experiments and viscous and viscoelastic interface models for encapsulation,” J. Acoust. Soc. Am. 118, 539–550 (2005). 10.1121/1.1923367 [DOI] [PubMed] [Google Scholar]
- 40.Faez T., Goertz D., and De Jong N., “ Characterization of DefinityTM ultrasound contrast agent at frequency range of 5-15 MHz,” Ultrasound Med. Biol. 37, 338–342 (2011). 10.1016/j.ultrasmedbio.2010.11.014 [DOI] [PubMed] [Google Scholar]
- 41.Malkin A. I., Rheology Fundamentals ( ChemTec, Toronto-Scarborough, Ontario, Canada, 1994). [Google Scholar]
- 42.Church C. C., “ The effects of an elastic solid surface layer on the radial pulsations of gas bubbles,” J. Acoust. Soc. Am. 97, 1510–1521 (1995). 10.1121/1.412091 [DOI] [Google Scholar]
- 43.Li Q., Matula T. J., Tu J., Guo X., and Zhang D., “ Modeling complicated rheological behaviors in encapsulating shells of lipid-coated microbubbles accounting for nonlinear changes of both shell viscosity and elasticity,” Phys. Med. Biol. 58, 985–998 (2013). 10.1088/0031-9155/58/4/985 [DOI] [PubMed] [Google Scholar]
- 44.Doinikov A. A., Haac J. F., and Dayton P. A., “ Modeling of nonlinear viscous stress in encapsulating shells of lipid-coated contrast agent microbubbles,” Ultrasonics 49, 269–275 (2009). 10.1016/j.ultras.2008.09.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Helfield B. L., Cherin E., Foster F. S., and Goertz D. E., “ Investigating the subharmonic response of individual phospholipid encapsulated microbubbles at high frequencies: A comparative study of five agents,” Ultrasound Med. Biol. 38, 846–863 (2012). 10.1016/j.ultrasmedbio.2012.01.011 [DOI] [PubMed] [Google Scholar]
- 46.Goertz D. E., Jong N. de, and van der Steen A. F., “ Attenuation and size distribution measurements of Definity and manipulated Definity populations,” Ultrasound Med. Biol. 33, 1376–1388 (2007). 10.1016/j.ultrasmedbio.2007.03.009 [DOI] [PubMed] [Google Scholar]
- 47.Helfield B. L., Leung B. Y. C., Huo X., and Goertz D. E., “ Scaling of the viscoelastic shell properties of phospholipid encapsulated microbubbles with ultrasound frequency,” Ultrasonics 54, 1419–1424 (2014). 10.1016/j.ultras.2014.03.014 [DOI] [PubMed] [Google Scholar]
- 48.Garbin V., Cojoc D., Ferrari E., Di Fabrizio E., Overvelde M. L. J., Van Der Meer S. M., De Jong N., Lohse D., and Versluis M., “ Changes in microbubble dynamics near a boundary revealed by combined optical micromanipulation and high-speed imaging,” Appl. Phys. Lett. 90, 114103 (2007). 10.1063/1.2713164 [DOI] [Google Scholar]
- 49.Helfield B. L., Leung B. Y. C., and Goertz D. E., “ The effect of boundary proximity on the response of individual ultrasound contrast agent microbubbles,” Phys. Med. Biol. 59, 1721–1745 (2014). 10.1088/0031-9155/59/7/1721 [DOI] [PubMed] [Google Scholar]