Abstract
Background and aims. Early diagnosis of external root resorption is important for accurate treatment. The purpose of this study was to compare the efficacy of a charge-coupled device (CCD) sensor and a photostimulable phosphor (PSP) plate receptor in the diagnosis of artificial external root resorption.
Materials and methods. In this diagnostic in-vitro study, 40 maxillary incisors were mounted in a segment of dry bone and preliminary radiographs were obtained using CCD and PSP sensors. Artificial resorption cavities were produced on the middle-third in half of the samples and on the cervical-third in the other half on the buccal root surfaces. Radiographs were repeated and images were evaluated. Data were statistically analyzed using chi-square and diagnostic tests.
Results. There were no significant differences between the two sensors in the sensitivity (p=0.08 and 0.06) and specificity (p=0.13) for the diagnosis of resorption in both root areas. The overall accuracy of CCD was higher than PSP sensor; however, the difference was not statistically significance (p>0.05).
Conclusion. CCD and PSP sensors chosen for the present study produced similar results in diagnosing simulated external root resorption.
Keywords: Dental radiography, digital radiography, root resorption
Introduction
External root resorption results in dissolution of cementum, dentin and sometimes extends toward the pulp. It has an unknown etiology in most cases; however, inflammation, tumors, and excessive mechanical and occlusal forces might contribute to the situation. One of the most common regions for external root resorption is the cervical area of the root, although it happens less probably in middle area too, and the highest incidence rate has been reported in anterior teeth.1 Root resorption was first described by Bell in 1830.2 In most cases, no clinical symptoms are present and it is usually detected during routine radiography. When such lesions are found, they are at an advanced level and treatment is almost impossible. In the early stages, diagnosis is based on the radiographic evidence of a resorption defect; therefore, detection of the lesion on a radiograph is very important for a dental practitioner. On the other hand, due to superimpositions, radiographs might fail to show lesions below certain diameters or depths.3
Recently, digital detectors have replaced radiographic films. Charge-coupled device (CCD) sensors have a silicon wafer as a base for image recording and the photo-stimulable phosphor (PSP) plates consist of polyester base, coated with a crystalline halide composed of europium-activated barium fluorohalide compounds.4 The quality of the image produced by a CCD detector depends on factors like chip pixel dimension, type of the scintillation layer, and acquisition and display software, while in PSP plates, the quality of the image may also depend on scanning procedure as well as electronics.5
Digital systems reduce radiation dose. They are time saving and offer image enhancement software, with easy communication and storage.6 Since the radiological diagnosis of external root resorption is important and considering the potential difference in diagnostic performance of these two digital detectors including their resolutions, this study was undertaken to compare the efficacy of these sensors in detecting external root resorption.
Materials and Methods
In this diagnostic in-vitro study, 40 extracted maxillary incisors were selected. Teeth with root fractures, caries, external or internal root resorption, or anatomic variations on preliminary radiographs were excluded from the study. The teeth were immersed in 0.5% sodium hypochlorite solutions for 24 hours for disinfection. In order to simulate alveolar bone covering the tooth, a dry sheep alveolar block was provided and fixed on a firm flat stand perpendicular to the surface using a special jig. Teeth were numbered and consecutively placed within empty sockets. Radiographs were provided with both sensors under the same condition using a sensor holder (Kerr Sensor Holder, Hawa SA, Switzerland; Figure 1).
Figure 1.
Dry alveolar block is fixed on a stand for an orthoradial radiograph.
The focus-receptor distance was set at 20 cm and radiographs of the teeth were taken in an orthoradial direction. All images were exposed with a Minray radiographic machine (SOREDEX, Tusula, Finland) operated at 70kVp, 8 mA and exposure times of 0.03 and 0.12 seconds for PSP and CCD sensors, respectively, based on a pilot study for optimal image quality. Two digital sensors were used: Dr. Suni CCD detector (Suni Medical Imaging, San Joe, USA) size 2, offering 2 megapixels and pixel size of 22 μm with a resolution of 23 Lp/mm; and PSP detector (Digora®Optime; SOREDEX, Tusula, Finland) size 2 and scanner 40 μm (pixel), offering 12.5 Lp/mm spatial resolution. The exposed phosphor plates were immediately scanned.
The preliminary radiographs were saved in two separate files. Then the teeth were divided into two groups of 20 each. In group 1, artificial resorption defects (ISO 0.6 mm depth/0.6 mm diameter holes), were created on the buccal surfaces of the middle-third of the roots using a round 006 bur (Diatec, Switzerland).2,3,7 In group 2, the same procedures were carried out in the cervical-thirds of the roots. The size of defects were chosen based on the pilot study. The teeth were once again placed back to the sockets and radiographs were taken using the same sensors.
For each of the imaging systems, 80 radiographs (with and without resorption and with defects created in different locations) were obtained making a total of 160 radiographs. Before starting the observations, a precise definition of the radiographic appearance of the lesion was achieved among observers.
Images were coded and evaluated in a random order separately by two observers. Examiners evaluated each image without previous knowledge of the presence or location of the root surface cavities.
The results of the observations including the presence/absence and the location of resorptions were recorded for each tooth. The observation conditions were the same for all the images and they were displayed in a darkened room on a 15.6 inch laptop monitor (Dell, Inspiron N65, China) with a screen resolution set at 768×1366 pixels and color set to a 64-bit depth. Digital images were evaluated in their own software and the contrast and brightness tool of each system was used if necessary (Figure 2). Observations were made in 4 sessions with 2-week intervals. Diagnostic indices including the sensitivity, specificity, and accuracies of both sensors at different root areas, Kappa values for inter-observer agreement and Kappa values between sensors were calculated. Data were statistically analyzed using chi-square and diagnostic test at the significance level of 0.05.
Figure 2.
Example of images of premolar teeth. Arrows show the resorption cavities. A, B: Images of a tooth with and without cavity obtained by PSP sensor. C, D: Images of different teeth with and without cavity obtained by CCD sensor.
Results
The Kappa value (inter-observer agreement) between the observers for the CCD sensor was 89%. Also, the Kappa value between the observers for the PSP sensor was 90%.
A total of 160 digital images of teeth with and without resorption defects in the middle and cervical thirds with the CCD and PSP sensors were evaluated.
The observers of the CCD images found 15 (18.75%) of the total detected resorption defects to be in the middle thirds, 17 (21.25%) in the cervical thirds and none in the remaining 48 (60%), while 25% of them were actually found to have defects in the middle thirds, 25% in the cervical thirds and 50% remaining were found to have no actual resorption cavities.
The observers of the PSP images found 14 (17.5%) of the total detected resorption defects to be in the middle thirds, 19 (23.75%) in the cervical thirds and none in the remaining 47 (58.75%), while 25% of them were actually found to have defects in the middle thirds, 25% in the cervical thirds and 50% remaining were found to have no actual resorption cavities.
The diagnostic sensitivities in the middle-thirds, were 70% and 55% for CCD and PSP sensors, respectively; in the cervical-thirds, they were 75% and 60% for CCD and PSP sensors, respectively. There were no significant differences in the diagnostic sensitivities between the two sensors and the two root areas (P = 0.08 and P = 0.06, respectively).
The diagnostic specificities were 95% and 85% for CCD and PSP sensors, respectively. According to chi-square test, there were no significant differences in the diagnostic specificities between the two sensors (P = 0.13; Tables1 & 2).
Table 1. Distribution of readings (percent) with CCD sensor according to the presence or absence and location of the defect and standard method .
Gold standard | ||||
CCD reading | Middle third | Cervical third | No defect | Total |
Middle third | 14(17.5%) | 1(1.25%) | 0 | 15(18.75%) |
Cervical third | 0 | 15(18.75%) | 2(2.5%) | 17(21.25%) |
No defect | 6(7.5%) | 4(5%) | 38(47.5%) | 48(60%) |
Total | 20(25%) | 20(25%) | 40(50%) | 80(100%) |
Table 2. Distribution of readings (percent) with PSP sensor according to the presence or absence and location of the defect and standard method .
Gold standard | ||||
PSP reading | Middle third | Cervical third | No defect | Total |
Middle third | 11(13.75%) | 2(2.5%) | 1(1.25%) | 14(17.5%) |
Cervical third | 2(2.5%) | 12(15%) | 5(6.25%) | 19(23.75%) |
No defect | 7(8.75%) | 6(7.5%) | 34(42.5%) | 47(58.75%) |
Total | 20(25%) | 20(25%) | 40(50%) | 80(100%) |
The diagnostic accuracies were 83.75% and 71.25% with CCD and PSP sensors, respectively. This difference was not statistically significant (P > 0.05).
Discussion
Dentists are faced with the challenge of timely diagnosis with regards to external root resorption, which can lead to a loss of tooth structure and a lower chance to preserve the tooth. Resorptive lesions are first diagnosed by intraoral radiographs, and therefore, the use of reliable radiographic techniques to diagnose such lesions is a necessity.1,8,9
In clinical situations, factors such as observation conditions, tissue superimposition and the position of the tooth involved in the dental arch influence the diagnosis of resorptive lesions. However, with in-vitro studies and elimination of confounding factors, the ability of detectors are studied more accurately.
In this study, teeth with artificial resorptive cavities were considered as the gold standard. The use of gold standard enabled the exact calculation of the percentage of positive and negative readings. Although borders of artificially created external root resorption cavities are relatively distinctive—and it seems that detecting such cavities are easier than that of natural cavities which have more diffuse borders, we could not employ a better method for creating artificial external root cavitations. In addition, bone marrow cavities imitated root defects, which possibly reduced the observers’ detection ability.
Since the results showed that the sensitivity and specificity for the diagnosis of the lesions were the same for both sensors in the two root segments studied, it seems that the observers’ ability to identify the defect was not affected by its location.
Although the CCD sensor revealed higher correct percentage of readings and the overall accuracy of CCD was higher than PSP sensor, differences were not statistically significant. Despite the higher resolution of the CCD sensor, the obtained results might have been due to software limitations.
Several studies have compared the digital systems and conventional films in detecting resorption defects;2,3,7,10-12 however, there are limited studies comparing CCD and PSP detectors in this respect. In a study by Kamburoğlu et al,2 conventional and digital radiography were compared in the diagnosis of external root resorption. Cavities were produced in the apical, middle and cervical-thirds on the two buccal and proximal root surfaces, which measured 0.5 and 0.8 mm in depth and cavities on the cervical and proximal areas were diagnosed much better.2 Also in a study by Dalili et al,12 cavities created using a bur in the middle-thirds were detected better than those in the cervical thirds, which is not in line with the findings of the present study.
Although in our study the cavities were created only on the buccal surfaces and only in the middle and cervical segments, the diagnostic efficacy of the lesions in the middle and cervical thirds was the same, which is similar to the results of the study by Kamburoğlu et al,2 where the CCD sensor performed better than the PSP sensor, probably due to low resolution of the PSP system.
In a study by Borg et al10 with the aim of comparing the efficacy of PSP and CCD sensors and conventional radiography in the diagnosis of external root resorption, both sensors yielded identical results, which is in accordance with our findings. In the latter study, cavities with 0.6 and 0.9 mm in diameter were created, and similar to the study carried out by Kamburoğlu et al,2 deeper lesions were diagnosed much better. Size is effective in detecting root defects particularly on intraoral radiographs.7,10,13 In the present study, all the lesions were 0.6 mm in depth, which is an appropriate size in the evaluation of the accuracy of the diagnosis of artificial resorption defects. In another study by Kamburoğlu et al7 carried out to diagnose internal resorption with the use of conventional radiography and various digital sensors, the results showed the number of correct diagnoses in the PSP images were less than other techniques, which is not in agreement with the results of the present study. In the latter study, the exposure times were 0.16 and 0.32 seconds with CCD and PSP sensors, respectively.7 PSP produces favorable images at higher x-ray doses, too, due to a higher dynamic range, which is considered a disadvantage for these systems.14
Given the disadvantages of conventional radiographs and the advances in digital radiography with the production of new sensors, the present study was undertaken to determine the superior digital system in the diagnosis of external root resorption.
Having similar sensitivities and specificities for both sensors in the diagnosis of resorptive lesions is clinically important. Generally, various factors can influence radiographic diagnoses. These include the imaging system (conventional vs. digital radiography), manipulation of images, observation conditions,14,15 and finally the type, dimension and location of the lesion. It appears that the use of technology including monitors, sensors, and appropriate software to assist in accurate diagnosis of various lesions can be very effective. However, another important consideration in addition to the familiarity of observers with hardware is their experience in working with digital images, especially operating the relevant computer programs. This has a significant role in the accuracy and correctness of the diagnosis.16-20Three-dimensional CBCT assessment has also been demonstrated for diagnosis and treatment planning of root resorption.21-23However, the use of digital intraoral sensors should be continued in this regard as doses of currently used CBCT units are still high.24,25
Conclusion
Considering the methodology and the results of the present study, the chosen CCD and the PSP sensors produced similar results in diagnosing simulated external root resorption lesions.
Acknowledgement
The authors would like to thank the Departments of Oral and Maxillofacial Radiology and Endodontics at Islamic Azad University of Tehran, Dental Branch for their support.
References
- 1.Cohen S, Hargreaves K. Pathways of the Pulp, 9th ed. St. Luis: Mosby; 2006. 4-47, 704-27 [Google Scholar]
- 2.Kamburoğlu K, Tesis I, Kfir A, Kaffe I. Diagnosis of artificially induced external root resorption using conventional intraoral film radiography, CCD, and PSP: an ex vivo study. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;106:885–91. doi: 10.1016/j.tripleo.2008.01.005. [DOI] [PubMed] [Google Scholar]
- 3.Westphalen VPD, Gomes de Moraes, Westphalen FH, Martins WD, Couto Sousa PH. Conventional and digital radiographic methods in detection of simulated external root resorptions: a comparative study. Dentomaxillofac Radiol. 2004;33:233–35. doi: 10.1259/dmfr/65487937. [DOI] [PubMed] [Google Scholar]
- 4.Ludlow JB, Mol A. Digital imaging. In: White SC, Pharoah MJ. Oral Radiology, Principles and Interpretation, 6th ed. St. Louis: Mosby; 2008. 78-99 [Google Scholar]
- 5.Farman AG, Farman TT. A comparison of 18 different x-ray detectors currently used in dentistry. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005;99:485–89. doi: 10.1016/j.tripleo.2004.04.002. [DOI] [PubMed] [Google Scholar]
- 6.Kitagawa H, Scheetz JP, Farman AG. Comparison of complementary metal-oxide semiconductor and charge-coupled device intraoral x-ray detectors using subjective image quality. Dentomaxillofac Radiol. 2003;32:408–11. doi: 10.1259/dmfr/19990417. [DOI] [PubMed] [Google Scholar]
- 7.Kamburoğlu K, Barenboim SF, Kaffe I. Comparison of conventional film with different digital and digitally filtered image in detection of simulated internal root resorption cavities: an ex vivo study in human cadaver jaws. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;105:790–97. doi: 10.1016/j.tripleo.2007.05.030. [DOI] [PubMed] [Google Scholar]
- 8.Seltzer S, Krasner P. Endodontology: Biologic Considerations in Endodontic Procedures, 2nd ed. Philadelphia: Lea and Sibiger; 1998. 170-190, 527-43 [Google Scholar]
- 9.Torabinejad M, Walton RE. Endodontics: Principles and Practice. 4th ed. St. Louis: Saunders; 2009. 21-38 [Google Scholar]
- 10.Borg E, Källqvist A, Gröndahl k, Gröndahl HG. Film and digital radiography for detection of simulated root resorption cavities. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1998;86:110–4. doi: 10.1016/s1079-2104(98)90159-7. [DOI] [PubMed] [Google Scholar]
- 11.Levander E, Bajka R, Malmgren O. Early radiographic diagnosis of apical root resorption during orthodontic treatment: a study of maxillary incisors. Eur J Orthod. 1998;20:57–63. doi: 10.1093/ejo/20.1.57. [DOI] [PubMed] [Google Scholar]
- 12.Dalili Z, Taromsari M, Mohtavipoor ST, Ojhand I. Comparison between CMOS digital detector and film in detection of simulated external root resorption: an in vitro study. J Mash Dent Sch. 2011;35:73–84. [Google Scholar]
- 13.Bernardes RA, de Paulo RS, Pereira LO, Hungaro Duarte MA, Ordinola-Zapata R, Ribamar de Azevedo. Comparative study of cone beam computed tomography and intraoral periapical radiographs in diagnosis of lingual-simulated external root resorptions. Dental Traumatol. 2012;28:268–72. doi: 10.1111/j.1600-9657.2011.01113.x. [DOI] [PubMed] [Google Scholar]
- 14.Wenzel A, Møystad A. Work flow with digital intraoral radiography: a systematic review. Acta Odontol Scand. 2010;68:106–14. doi: 10.3109/00016350903514426. [DOI] [PubMed] [Google Scholar]
- 15.White SC. Caries detection with xeroradiographs: the influence of observer experience. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1987;64:118–22. doi: 10.1016/0030-4220(87)90126-5. [DOI] [PubMed] [Google Scholar]
- 16.Ramanathan GP. Diagnosis of dental caries: a comparison of three radiograph viewing techniques. Clin Pediatr Dent. 1999;23:235–45. [PubMed] [Google Scholar]
- 17.Wenzel A. Effect of image enhancement for delectability of bone lesion in digitized intraoral radiography. Scand J Dent Res. 1988;96:149–60. doi: 10.1111/j.1600-0722.1988.tb01422.x. [DOI] [PubMed] [Google Scholar]
- 18.Araujo EA, Castilho JC, Medici Filho E, de Moraes ME. A comparison of direct digital and conventional imaging with Ekta speed plus and insight films for the detection of approximal caries. Am J Dent. 2005;18:241–44. [PubMed] [Google Scholar]
- 19.Uprichard KK, Potter BJ, Russell CM, Schafer TE, Adair S, Weller RN. Comparison of direct digital and conventional radiography for detection of the proximal surface caries in the mixed dentition. Pediatr Dent. 2000;22:9–15. [PubMed] [Google Scholar]
- 20.Syriopoulos K, Sanderink GC, Velders XL, van der Stelt. Radiographic detection of approximal caries: a comparison of dental films and digital imaging system. Dentomaxillofac Radiol. 2000;29:312–18. doi: 10.1038/sj.dmfr.4600553. [DOI] [PubMed] [Google Scholar]
- 21.Patel S, Dawood A, Wilson R, Horner K, Mannocci F. The detection and management of root resorption lesions using intraoral radiography and cone beam computed tomography – an in vivo investigation. Int Endod J. 2009;42:831–8. doi: 10.1111/j.1365-2591.2009.01592.x. [DOI] [PubMed] [Google Scholar]
- 22.da Silveira HLD, Silveira HED, Liedke GS, Lermen CA, dos Santos RB, de Figueire do. Diagnostic ability of computed tomography to evaluate external root resorption in vitro. Dentomaxillofac Radiol. 2007;36:393–6. doi: 10.1259/dmfr/13347073. [DOI] [PubMed] [Google Scholar]
- 23.Durack C, Patel S, Davies J, Wilson R, Mannocci F. Diagnostic accuracy of small volume cone beam computed tomography and intraoral periapical radiography for the detection of simulated external inflammatory root resorption. Int Endod J. 2011;44:136–47. doi: 10.1111/j.1365-2591.2010.01819.x. [DOI] [PubMed] [Google Scholar]
- 24.Grünheid TH, Kolbeck Schieck JR, Pliska BT, Ahmad M, Larson BE. Dosimetry of a cone-beam computed tomography machine compared with a digital x-ray machine in orthodontic imaging. Am J Dentofacial Orthop. 2012;141:436–43. doi: 10.1016/j.ajodo.2011.10.024. [DOI] [PubMed] [Google Scholar]
- 25.Sezgin OS, Kayipmaz S, Yasar D, Yilmaz AB, Ozturk MH. Comparative dosimetry of dental cone beam computed tomography, panoramic radiography, and multislice computed tomography. Oral Radiol. 2012;28:32–37. doi: 10.1007/s11282-011-0078-5. [DOI] [Google Scholar]