Errata: Lateral and axial measurement differences between spectral-domain optical coherence tomography systems

Francisco A Folgar a, Eric L Yuan a, Sina Farsiu a,b, Cynthia A Toth a,b
PMCID: PMC4453585  PMID: 26016449

This article [J. Biomed. Opt. 19(1), 016014 (2014)] was originally published online on 17 January 2014 with errors in the row and column labels of Tables 4 and 5 on p. 5. The corrected tables are reprinted below.

Table 4.

Conversion factors for lateral measurements across platforms.

  Convert from this platform
Lateral scaling Zeiss Cirrus™ Heidelberg Spectralis™ Bioptigen Envisu™ Bioptigen SDOIS
Convert to this platform Zeiss Cirrus™   0.980 0.991 0.994
Heidelberg Spectralis™ 1.020   1.011 1.014
Bioptigen Envisu™ 1.009 0.989   1.003
Bioptigen SDOIS 1.006 0.986 0.997  

Table 5.

Conversion factors for axial measurements across platforms.

  Convert from this platform
Axial scaling Zeiss Cirrus™ Heidelberg Spectralis™ Bioptigen Envisu™ Bioptigen SDOIS
Convert to this platform Zeiss Cirrus™   1.011 1.037 1.174
Heidelberg Spectralis™ 0.990   1.025 1.160
Bioptigen Envisu™ 0.964 0.975   1.132
Bioptigen SDOIS 0.852 0.862 0.884  

This article was corrected online on 18 May 2015.