Skip to main content
Journal of Diabetes Science and Technology logoLink to Journal of Diabetes Science and Technology
. 2014 Jan;8(1):142–149. doi: 10.1177/1932296813511725

Recommendations for Insulin Dose Calculator Risk Management

Christen Rees 1,
PMCID: PMC4454096  PMID: 24876550

Abstract

Several studies have shown the usefulness of an automated insulin dose bolus advisor (BA) in achieving improved glycemic control for insulin-using diabetes patients. Although regulatory agencies have approved several BAs over the past decades, these devices are not standardized in their approach to dosage calculation and include many features that may introduce risk to patients. Moreover, there is no single standard of care for diabetes worldwide and no guidance documents for BAs, specifically. Given the emerging and more stringent regulations on software used in medical devices, the approval process is becoming more difficult for manufacturers to navigate, with some manufacturers opting to remove BAs from their products altogether. A comprehensive literature search was performed, including publications discussing: diabetes BA use and benefit, infusion pump safety and regulation, regulatory submissions, novel BAs, and recommendations for regulation and risk management of BAs. Also included were country-specific and international guidance documents for medical device, infusion pump, medical software, and mobile medical application risk management and regulation. No definitive worldwide guidance exists regarding risk management requirements for BAs, specifically. However, local and international guidance documents for medical devices, infusion pumps, and medical device software offer guidance that can be applied to this technology. In addition, risk management exercises that are algorithm-specific can help prepare manufacturers for regulatory submissions. This article discusses key issues relevant to BA use and safety, and recommends risk management activities incorporating current research and guidance.

Keywords: bolus calculator, diabetes, insulin, medical device regulation, risk management, dose calculator, insulin pump


Automated insulin dose bolus advisors (BAs) are used by insulin-treated diabetic patients to calculate insulin dosages for correction of out-of-target blood glucose and/or intake of carbohydrates. BAs provide a useful and simplified alternative to manual calculations of insulin dosing. To calculate appropriate bolus doses, a BA requires accurate settings for the insulin-to-carbohydrate ratio, glucose-correction factor, duration-of-insulin action, and correction target.1 The BA then utilizes a proprietary algorithm to calculate the dose of insulin recommended for the user based on their inputs, and in some insulin dose advisors, historical data input by the user. Current literature has shown that integration of BAs into insulin pump software may help individuals more accurately meet prandial insulin dosage requirements, improve postprandial glycemic excursions, and achieve optimal glycemic control.2-5 However, evidence is sparse regarding the specifics of risk management for algorithms such as those used in BAs. In addition, there is no worldwide standard of care for diabetes, which complicates the development of software and/or devices intended to optimize glycemic control. Not surprisingly, BAs currently on the market around the world differ widely in their approaches to dosage calculation. More specifically, inputs required for calculation, handling of insulin-on-board calculations, and outputs shown to the user all vary widely from manufacturer to manufacturer.

Medical device safety is a risk management process that must encompass the life span of medical devices from inception to disposal.6 These activities should be conducted in a consistent fashion, globally, to ensure the health and safety of diabetic patients in all countries. There are many local and global organizations with the charter of ensuring adequate regulation of medical devices and publishing and communicating regulations and standards to their audience. This article discusses key issues relevant to BA use and safety and provides recommendations for risk management activities incorporating current research and guidance. The purpose of this article is to present the current landscape and provide prompt further discussion on the topic of BA risk management.

Current Guidance for Development of BA Algorithms

Literature Review

Published articles, regulatory submissions, and books, primarily published within the past 10 years, with the majority being published within the last 3 years, were reviewed. International and local guidance documents were reviewed irrespective of year published, only with regard to topic and latest revision. A total of 142 sources were reviewed (see the appendix), with the following breakdown: insulin BA use/benefit (28%), standards of care/guidance documents (18%), insulin BA regulation/safety (25%), novel insulin dose algorithm (25%), and other (2%). Sources were included if they contained relevant information on drug BAs or insulin BAs, medical device regulation, and medical product risk management. Literature was excluded from the review if it did not include discussion regarding risks of drug BA use, regulation of medical devices or software, BA standards of care, development of insulin dose algorithms, or insulin BA use.

Current Guidance Documents

While numerous studies have shown that use an insulin BA is beneficial to most diabetic patients on continuous subcutaneous insulin infusion (CSII) therapy,2-5 the standards of care published by medical organizations vary in their recommendations for therapy goals, glycemic targets, and starting parameters for insulin BAs and CSII.7,8 For example, the 2011 American Association of Clinical Endocrinologists (AACE) Diabetes Care Plan Guidelines recommend fasting plasma glucose levels of less than 110 mg/dL and 2-hour postprandial glucose concentrations of less than 140 mg/dL. However, the 2012 American Diabetes Association (ADA) Standards of Diabetes Care recommends preprandial capillary plasma glucose of 70-130 mg/dL and peak postprandial capillary plasma glucose of less than 180 mg/dL. Neither guideline describes use or setting up of BAs for diabetes care.7,8 Moreover, many of the publications regarding medical device risk management are case studies and implementation guides, which can make interpretation and utilization of this information difficult for manufacturers.

Local and international regulatory bodies currently provide no guidance documents published specifically pertaining to insulin BAs. However, representatives from the regulatory agencies have shared expectations regarding BA functionality and safety at various meetings and conferences in recent years. As well, both the UK Medicines and Healthcare Products Regulatory Agency (MHRA) in the United Kingdom and the US Food and Drug Administration (FDA) have recently released documents pertaining to infusion pumps specifically, which often incorporate BA technology. Both of these documents describe basic infusion pump safety considerations for their respective countries.

The MHRA document (Device Bulletin, Infusion Systems) was a device bulletin released in 2010 and is intended for use by physicians and medical staff using infusion pumps.9 It provides general infusion pump safety considerations for use and training of the devices.

The FDA document (Guidance for Industry and FDA Staff—Total Product Life Cycle: Infusion Pump—Premarket Notification) was released in 2010 as draft guidance and has not been finalized as of the date of this research.10 This document provides recommendations for infusion pump regulatory submission as well as specific information on insulin dosing algorithms; the MHRA document does not provide these recommendations. The document also states that for infusion pumps that contain algorithms intended to provide dosing recommendations, it is recommended that manufacturers include the following information in their premarket notification: each dosing algorithm used, each algorithm in symbolic form, all parameters in each algorithm with identification of which parameters can be modified by the user, clinical data or other justification for why it is believed that the algorithm is appropriate for the target population, proof of verification and validation activities showing that calculations made by the BA are correct, and data demonstrating that the algorithm calculates insulin doses correctly.10

In presentations at various meetings and conferences, the FDA has identified decision-making software such as insulin BAs as medical devices and actively regulates these devices.11 Additional guidance released by the FDA’s Center for Devices and Radiological Health (CDRH) in 2005 (Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) describes the expectations regarding software development, level of concern, and scope of the guidance.12 The FDA considers insulin BAs to fall under this guidance11 and recognizes 4 primary review areas for insulin BAs: (1) intended use, (2) algorithm, (3) clinical evidence, and (4) documentation.

Risk management activities are also described in documents from the American National Standards Institute (ANSI), Association for the Advancement of Medical Instrumentation (AAMI), and the International Organization for Standardization (ISO). Medical Devices—Application of Risk Management to Medical Devices 13 and Medical Device Software Risk Management 14 recommend the following software-specific risk management activities: (1) analysis of the clinical or other intended use(s) of the device to identify clinical hazards and related software functionality providing clinical information or control commands related to safety and effectiveness; (2) identification of potential software causes that could contribute to a hazardous situation; (3) identification of software components that implement or could affect functionality related to safety or the implementation of risk control measures and identification of hardware, software, and/or labeling risk control measures to prevent software errors from resulting in a hazardous situation or to reduce the probability of such errors; (4) evaluation of residual risk after risk control measures are identified and determination of whether the risk is acceptable given the intended use and target population of the device; and (5) verification of risk control measures and software-related software components.14

Use of an In Vitro Simulator

While some rules regarding BA algorithm assessment are supported by current literature and standards of care, other rules require additional support to demonstrate safety. Exercising the algorithm with an in vitro simulator can provide medical evidence of safety.

In vitro simulation allows the full functionality of the algorithm to be tested, especially the least likely scenarios that may not occur in a clinical trial. However, for product risk management activities, it is important that the simulator used be standardized and approved by regulatory agencies.

In 2011, the FDA approved the University of Virginia (UV) simulator for in silico testing of artificial pancreas control systems in the United States. This simulator is a computer-based diabetes simulation tool developed by UV and has been accepted by the FDA as an alternative to animal testing of type 1 diabetes control strategies.15 The simulator uses a software algorithm to model the human metabolic system. There are now hundreds of theoretical, FDA-accepted patient models that researchers can use to test control algorithms.16

Based on patient data from 300 children, adolescents, and adults with type 1 diabetes, the UV algorithm uses 26 different parameters to mimic human metabolism at the individual level, through several distinct patient profiles. Within these individual profiles, variables such as diet, exercise behavior, and insulin intake can be manipulated to test the accuracy or effectiveness of a new product under varying conditions—or to compare it to existing products.15 This simulator, or a similar approach, would be ideal for exercising the capabilities of insulin dose algorithms to demonstrate safety.

Recommendations

The availability of published guidance documents can help alleviate some of the challenges that manufacturers face and can help ensure that products brought to market are safe and effective around the world. This is especially important when considering the safety concerns of algorithms, which do not specifically fall under the guidance of medical devices or medical software. However, if the guidance is not harmonized internationally, risk management activities and, ultimately, product safety may be critically inconsistent. My opinion, after performing the extensive research (appendix) and case study activities described in this report, is as follows: because of their unique functionality and risks posed to users, insulin dose calculating algorithms should have specifically dedicated risk management activities assigned. The following recommendations may be helpful in developing standardized, international guidance for software designers, manufacturers, and regulatory agencies (Table 1).

Table 1.

Recommendations for Insulin Dose Calculator Risk Management.

Recommendation 1 Require in vitro testing of insulin dosing algorithms utilizing an in vitro simulator, such as the previously described University of Virginia simulator.
Recommendation 2 Require risk management activities by global regulatory agencies in accordance with Medical Device Software Risk Management and Guidance for Industry and FDA Staff—Total Product Life Cycle: Infusion Pump—Premarket Notification; Draft Guidance, Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices or (similar updated documents) as an international standard.
Recommendation 3 Require standardized algorithm documentation for submission of stand-alone insulin dosing calculators, as well as those that are a constituent of a medical device.
Recommendation 4 Develop a worldwide database of registered products to improve regulatory transparency around the world, as well as to monitor and track device safety throughout the device life cycle.
Recommendation 5 Avoid requiring standardization of insulin dosing algorithms to preclude the stifling of scientific and technologic advances in the area.
  • Recommendation 1: Require clinical study and/or in vitro testing of insulin dosing algorithms utilizing an in vitro simulator, such as the previously described UV simulator.

In vitro testing of insulin dosing algorithms with the educational version of the simulator would provide a standard platform for algorithm testing. If accepted as an international standard, all algorithms developed in other countries would be held to the same standard of proof. This would promote consistency and safety of insulin dosing algorithms, and would assist regulatory agencies in reviewing in vitro testing results.

  • Recommendation 2: Require risk management activities by global regulatory agencies in accordance with Medical Device Software Risk Management and Guidance for Industry and FDA Staff—Total Product Life Cycle: Infusion Pump—Premarket Notification,10 Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices,12 or similar updated documents as an international standard in addition to ISO 14971.

Risk management activities in accordance with these recommendations would ensure the consistency of risk management activities for all insulin dosing algorithms. Once provided as an international standard, this document would provide the much-needed guidance for manufacturers to follow and ensure the safety of their algorithms, as well as for submission of their algorithm-containing software or device to regulatory agencies. For the regulatory bodies, having an approved guidance would ensure clear directives for submissions to follow, thus making the approval process more straightforward with clear expectations.

  • Recommendation 3: Require standardized algorithm documentation for submission of stand-alone insulin dosing calculators, as well as those that are a constituent of a medical device.

This type of exploration and documentation would be extremely helpful for manufacturers and regulatory agencies as it clearly demonstrates the medical evidence associated with each component of the insulin BA algorithm. In addition, appropriately modified information of this type may even be beneficial to consumers. The exploration of literature and justification of algorithm rules would identify gaps in the literature support of the algorithm. The areas of the algorithm not supported by current literature or standards of care could then be justified in another manner, such as in vitro simulation or clinical studies, to meet regulatory standards for safety. This would ensure that all aspects of the algorithm are thoroughly explored and supported to ensure safety.

  • Recommendation 4: Develop a worldwide database of registered products to improve regulatory transparency around the world, as well as to monitor and track device safety throughout the device life cycle.

The benefit of these databases is well known; however, their benefit is limited to the geographic area(s) which they support. If a product is sold in an area of the world where it is not registered, then that country will have no record of the product or any postmarket reporting on that product. This is especially problematic in the EU, where registers are maintained by competent authorities and a manufacturer needs register with only one to market its product throughout the EU.

  • Recommendation 5: Avoid requiring standardization of insulin dosing algorithms to preclude the stifling of scientific and technologic advances in the area.

There are many different insulin dosing algorithms currently available globally, and novel algorithms continue to be researched and developed. Many of these algorithms behave differently from one another in their calculation method and logic, but generate similar recommendations. Fine-tuning the algorithm, allowing varying inputs, creating “smart” algorithms, and integrating blood glucose (BG) and continuous glucose monitoring (CGM) data all enhance the functioning of insulin BAs over time. To avoid hindering innovation, it is essential that a standardized algorithm not be mandated by regulatory agencies. Acceptance of a policy such as this would stifle the evolution of these algorithms and prevent potential breakthroughs in diabetes care.

Conclusions

There is a general consensus that insulin BAs are useful and beneficial.2-5 Current literature also shows that additional novel insulin dosing algorithms are being researched and developed.17-20 After an extensive literature research (see the appendix) and case study exploration, my opinion on this topic is as follows: clearer guidance is needed to assist both the manufacturers creating these algorithms and the regulatory agencies that will ultimately review them.

Specific risk management activities for BA algorithms, including simulator testing, flowchart diagramming, and literature support search, have been recommended in this article. Risk management activities, as described in a few key guidance documents related to medical device software and infusion pumps, have also been identified.

Manufacturers and developers of insulin BAs would benefit from standardized guidance because risk management expectations would be clearly defined and well organized, thus facilitating greater adherence. It would also provide consistency for regulatory agencies reviewing submissions, to ensure the testing and results were consistent and more easily reviewed.

However, while required standardization of algorithm may ease regulatory burden regarding insulin BAs, this would likely stifle innovation in the field of diabetes. Mandating a standard method for calculating insulin dosage may hinder the development of algorithmic advances that could potentially provide breakthroughs in the field of diabetes care around the world.

In summary, international acceptance and adoption of the recommendations put forth in this article for insulin BA risk management and guidance may greatly benefit manufacturers, regulatory agencies, and, most important, people with diabetes worldwide. This article provides a short but solid survey of the problem and will hopefully prompt further attention and publication regarding this topic.

Acknowledgments

Christen Rees penned this article while a student in the Health Product Risk Management program at Virginia Polytechnic Institute and State College. This article is a modification of the final project and report for her master’s degree. The author would like to acknowledge Robin Wagner, Patsy Allen, Ann Buskirk, George Flick, Stan Mastrangelo, Al Dolan, and Lori Marsh for their inspiration and guidance during her work.

Appendix

Appendix.

Comprehensive Literature Review List

Group First author Subject Year
Insulin dose algorithm/use
Barnard Use of an automated bolus calculator in MDI patients 2012
Bernasko Intensive insulin therapy 2007
Blazik The education of patients 2010
Bode Diabetes management in the new millennium 2002
Cavan Automated bolus advisor control and usability study (ABACUS): protocol 2012
Clavel Treatment by external insulin pump 2010
Colin Glucose meters with built-in automated bolus calculator 2013
Colquitt Clinical and cost-effectiveness 2004
Cukiermann-Yaffe Key elements for successful 2011
Gross A bolus calculator 2003
Hirsch Practical pearls 2010
King Basal bolus dosing 2005
King A prospective evaluation 2007
Klupa Clinical usefulness of a bolus calculator 2008
Mastrototaro The integrated minimed 2009
Michon Insulin pump and intensive 2002
Murphy Reducing hyperglycemia 2009
Pollom Optimizing inpatient glycemic control 2010
Renard Insulin pump use in Europe 2010
Roemer Tips and tools for insulin pump 2009
Romatowska The effect of using the insulin pump bolus calculator compared to standard insulin dosage calculations in patients with type 1 diabetes mellitus—systematic review 2013
Scheiner Insulin pump therapy 2009
Schmidt Use of an automated bolus calculator in MDI-treated type 1 diabetes: the BolusCal Study 2012
Shalitin Benefits of a bolus calculator 2010
Shashaj Benefits of a bolus calculator 2008
Spellman Management of diabetes 2003
Tamborlane Insulin pump therapy in childhood 2003
Unger Optimizing insulin therapy 2008
Walsh Guidelines for insulin dosing 2010
Walsh Guidelines for optimal bolus 2011
Walsh Pumping insulin 2003
Walsh Pumping insulin 2006
Wei Basal-bolus insulin protocols 2011
Weissberg Insulin pump therapy 2003
White Insulin pump therapy 2007
White When and how to implement 2009
Wilson Intensive insulin therapy 2007
Wolpert Hands-on demonstration 2005
Zhang The current situation review 2010
Zhang Second insulin pump safety 2010
Zisser Bolus calculator: a review 2008
Zisser Clinical performance of 3 bolus 2010
Zisser Clinical update on optimal prandial insulin 2009
Standards of care/guidance documents
AACE Diabetes care plan guidelines 2011
AADE Consensus insulin pump therapy 2008
AAMI Medical device software risk management 2005
ADA Practical insulin 2011
ADA Standards of medical care 2011
EC Directive 98/79/EC 1998
EN Biological evaluation 2009
EN Clinical investigation of medical devices 2009
EN Medical device software 2006
EN Medical devices—application of risk management 2009
EN Medical devices—application of usability 2008
EN Medical devices—quality management 2009
EN Medical electrical equipment part 1-10 2008
EN Medical electrical equipment part 2-24 1998
FDA Draft guidance 2011
FDA Guidance for industry 2010
ISO Guide to the development 2012
ISO Risk management 2009
ISO Safety aspects 1999
Malerbi Brazilian diabetes society 2006
TIR Technical information report 2009
WHO Medical device regulations 2003
aami.org
fda.gov
iso.org
Medical device regulation/safety
Babamir Constructing a model-based 2010
Bartoo Risk management 2003
Benet A risk driven approach 2011
Burton Improving software risk management 2008
Burton A risk management capability model 2008
Burton A software risk management capability model 2008
Coverity Governance, risk, and compliance 2011
Devices4Limited Regulation of health apps 2012
Donawa Effective risk management 2004
Goodman Medical device error 2002
Grimes Medical device security 2004
Hedge Case study—risk management 2011
Israelski Human factors risk management 2004
Jardine Risk management frameworks 2003
Jones Risk management in the design 2002
Klonoff Insulin pump safety meeting 2009
McCaffery Improving software risk 2009
McCaffery Risk management capability model 2009
Mingziao A quantitative approach 2011
Narvaez Prospective risk management 2010
Paul A review of the security 2011
Pietzsch Review of US medical device 2007
Rakitin Coping with defective software 2006
Reviewers Guidance for industry and FDA 2000
Robson Risk management and biomedical 2005
Schultz Medical device safety 2007
Shah Current status of the regulation 2008
Sherwin-Smith Medical applications—the future of regulation 2012
Small Medical device-associated safety 2004
Wang Discussions on the regulatory 2009
Wood Software risk management 2008
Xie Risk management for medical devices 2007
Xu Discussion about the application 2006
Zimlicki Bolus calculator regulation 2011
Zuckerman Medical device recalls 2011
Novel algorithm
Anderson Multiple daily injections 2009
Boiroux Optimal insulin administration 2009
Boiroux Strategies for glucose control 2010
Bondia Coordinated basal-bolus 2009
Campbell Calculating insulin on board for extended bolus
Charpentier The Diabeo software 2011
Cook The intelligent dosing system 2005
Davidson Analysis of guidelines 2008
Edelman Insulin pump therapy 2010
FDA Endotool drug dose calculator 2006
FDA Glucommander plus 2006
FDA Glytec, LLC 2010
FDA Intelligent dosing system 2008
Gonzalez-Rodriguez Should the amount of fat and protein 2010
Grandinetti Web-based prediction 2011
Hedbrandt The Sarimner diabetes simulator 2007
Juneja Safety and effectiveness 2008
King How much do I give? 2010
Kordonouri Supplementary fat plus protein 2010
Laha Glucose control in critical illness 2008
Lippaiova Safety of new algorithms 2008
Marling Case-based decision support 2008
Mathews A new model to estimate bolus 2009
Olinder Postprandial glucose levels 2009
Palerm Prandial insulin dosing using 2007
Pankowska Bolus calculator with nutritional 2010
Pankowska A review of the “bolus guide” 2010
Pelzer Preliminary application of a new bolus 2011
Revert Calculation of the best basal-bolus therapy 2011
Revert Combining basal-bolus 2010
Rossi Diabetes interactive diary 2010
Rossi Interactive diary for diabetes 2009
Sahota Computerized clinical decision 2011
Schwartz Use of case-based reasoning 2008
Shapira Bolus guide: a novel insulin 2010
Other
ADA Workgroup on Hypoglycemia 2005
Estabrook Commercially available UV simulator 2011
Kravitz From insight to implementation 2005

Footnotes

Abbreviations: AACE, American Association of Clinical Endocrinologists; AAMI, Association for the Advancement of Medical Instrumentation; ADA, American Diabetes Association; BA, bolus advisor; BG, blood glucose; CDRH, Center for Devices and Radiological Health; CE, Conformité Européenne; CGM, continuous glucose monitoring; CSII, continuous subcutaneous insulin infusion; FDA, Food and Drug Administration; GHTF, Global Harmonization Task Force; HbA1c, glycosylated hemoglobin; IDDM, insulin-dependent diabetes mellitus; ISO, International Organization for Standardization; MDD, Medical Devices Directive; MDI, multiple daily injections; MHRA, Medicines and Healthcare Products Regulatory Agency; TIR, Technical Information Report; UN, United Nations; WHO, World Health Organization.

Declaration of Conflicting Interests: The author(s) declared the following potential conflict of interest with respect to the research, authorship, and/or publication of this article: Christen Rees is employed at Roche Diagnostics Operations, Indianapolis, IN, USA.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

References

  • 1. Walsh J. Pumping Insulin. 4th ed. San Diego, CA: Torrey Pines Press; 2006. [Google Scholar]
  • 2. Gross TM, Kayne D, King A, Rother C, Juth S. A bolus calculator is an effective means of controlling pump therapy. Diabetes Technol Ther. 2003;5(3):365-369. [DOI] [PubMed] [Google Scholar]
  • 3. Hirsch IB. Practical pearls in insulin pump therapy. Diabetes Technol Ther. 2010;12(suppl 1):S23-S27. [DOI] [PubMed] [Google Scholar]
  • 4. Klupa T, Benbenek-Klupa T, Malecki M, Szalecki M, Sieradzki J. Clinical usefulness of a bolus calculator in maintaining normoglycaemia in active professional patients with type 1 diabetes treated with continuous subcutaneous insulin infusion. J Int Med Res. 2008;36(5):1112-1116. [DOI] [PubMed] [Google Scholar]
  • 5. Shashaj B, Busetto E, Sulli N. Benefits of a bolus calculator in pre- and postprandial glycaemic control and meal flexibility of paediatric patients using continuous subcutaneous insulin infusion (CSII). Diabet Med. 2008;25(9):1036-1042. [DOI] [PubMed] [Google Scholar]
  • 6. World Health Organization. Medical device regulations: global overview and guiding principles. Journal of healthcare information management. J Healthcare Inf Manage. 2003;24:36-40. [Google Scholar]
  • 7. American Association of Clinical Endocrinologists. American Association of Clinical Endocrinologists medical guidelines for clinical practice for developing a diabetes mellitus comprehensive care plan. Endocr Pract. 2011;17(suppl 2):287-302. [DOI] [PubMed] [Google Scholar]
  • 8. American Diabetes Association. Standards of medical care in diabetes—2012. Diabetes Care. 2012;35(suppl 1):S11-S63. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Medicines and Healthcare Products Regulatory Agency. Device Bulletin Infusion Systems. DB2003(02) v2.0; November 2010.
  • 10. US Food and Drug Administration. Guidance for Industry and FDA Staff—Total Product Life Cycle: Infusion Pump—Premarket Notification [510(k)] Submissions Draft Guidance. April 23, 2010.
  • 11. Zimlicki C. FDA’s perspectives for insulin dose calculators. Paper presented at: Diabetes Technology Meeting; October 27-30, 2001; San Francisco, CA. [Google Scholar]
  • 12. US Food and Drug Administration. Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. May 11, 2005.
  • 13. ANSI/AAMI/ISO 14971:2007/(R)2010. Medical Devices—Application of Risk Management to Medical Devices. October 6, 2010.
  • 14. AAMI TIR32:2004. Medical Device Software Risk Management. December 23, 2004.
  • 15. Estabrook M. Commercially available U.Va. diabetes simulator accelerating the pace of research and development. UVaToday. July 6, 2011.
  • 16. http://www.fda.gov/MedicalDevices/ProductsandMedicalProcedures/HomeHealthandConsumer/ConsumerProducts/ArtificialPancreas/ucm259561.htm.
  • 17. Charpentier G, Benhamou PY, Dardari D, et al. The Diabeo software enabling individualized insulin dose adjustments combined with telemedicine support improves HbA1c in poorly controlled type 1 diabetic patients: a 6-month, randomized, open-label, parallel-group, multicenter trial (TeleDiab 1 Study). Diabetes Care. 2011;34(3):533-539. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Cook CB, McMichael JP, Lieberman R, et al. The intelligent dosing system: application for insulin therapy and diabetes management. Diabetes Technol Ther. 2005;7(1):58-71. [DOI] [PubMed] [Google Scholar]
  • 19. Dimensional Dosing Systems, Inc. Intelligent dosing system for insulin—my insulin doser. 510(K) Number K082512. November 7, 2008.
  • 20. Glytec LLC. G+ system. 510(K) Number K101344. June 28, 2010.

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES