Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2016 Jan 31.
Published in final edited form as: IEEE Trans Med Imaging. 2014 Oct 8;34(2):541–550. doi: 10.1109/TMI.2014.2361810

Tissue Electrical Property Mapping from Zero Echo-Time Magnetic Resonance Imaging

Seung-Kyun Lee 1, Selaka Bulumulla 2, Florian Wiesinger 3, Laura Sacolick 4, Wei Sun 5, Ileana Hancu 6
PMCID: PMC4455904  NIHMSID: NIHMS695097  PMID: 25312919

Abstract

The capability of magnetic resonance imaging (MRI) to produce spatially resolved estimation of tissue electrical properties (EPs) in vivo has been a subject of much recent interest. In this work we introduce a method to map tissue EPs from low-flip-angle, zero-echo-time (ZTE) imaging. It is based on a new theoretical formalism that allows calculation of EPs from the product of transmit and receive radio-frequency (RF) field maps. Compared to conventional methods requiring separation of the transmit RF field (B1+) from acquired MR images, the proposed method has such advantages as: (i) reduced theoretical error, (ii) higher acquisition speed, and (iii) flexibility in choice of different transmit and receive RF coils. The method is demonstrated in electrical conductivity and relative permittivity mapping in a salt water phantom, as well as in-vivo measurement of brain conductivity in healthy volunteers. The phantom results show the validity and scan-time efficiency of the proposed method applied to a piece-wise homogeneous object. Quality of in-vivo EP results was limited by reconstruction errors near tissue boundaries, which highlights need for image segmentation in EP mapping in a heterogeneous medium. Our results show the feasibility of rapid EP mapping from MRI without B1+ mapping.

Index Terms: magnetic resonance imaging, electrical properties, MREPT, ZTE

I. INTRODUCTION

In-Vivo imaging of tissue permittivity and electrical conductivity can prove useful for a number of applications. For example, specific absorption rate (SAR) modeling relies on the availability of spatially resolved maps of tissue electrical properties (EPs) as input [1]. Accurate classification of tumor tissue as benign or malignant can also benefit from the additional information provided by EP mapping; while some disagreement exists in literature [2], prompting for further, properly controlled experimental verification, it is generally agreed upon that cancers tend to exhibit higher conductivity and permittivity than normal tissues [35].

MRI was recognized more than two decades ago as having the capability of producing conductivity and permittivity maps [6]. However, progress in implementing tissue EP mapping in the clinic has been slow. In most common implementations, two acquisitions are required for EP reconstruction: the first one maps the magnitude of the transmit radio-frequency (RF) field (B1+), using one of the many B1+ mapping techniques [79]; the second one maps the transceive RF phase, which is then used to approximate the phase of B1+ [10]. The complex B1+ maps, thus obtained with a phase approximation, are subsequently processed using the integral or differential forms of Maxwell's equations to retrieve permittivity and conductivity maps [1113]. This general approach has been known as MR-based electrical property tomography (MREPT). Unfortunately, B1+ mapping is typically a relatively slow process, seldom used clinically for diagnostic purposes. Its acquisition adds burden to the total scan time for a patient and B1+ maps usually have lower SNR compared to standard, clinical gradient echo (GRE) or spin echo (SE) images. Since processing in Maxwell's equations involves computation of the derivatives of B1+, an operation very sensitive to noise, low SNR in the input B1+ can severely degrade the quality of the EP maps. Reconstruction of EPs out of more SNR-efficient acquisitions would therefore be beneficial, enabling one to determine tissue EPs faster, and with higher accuracy.

Recently, fast conductivity mapping methods have been demonstrated that are based on a single acquisition of a steady-state free precession (SSFP)[14] or an ultra-short echo time (UTE)[15] scan. These methods do not require B1+ magnitude mapping as they employ a "phase-only" approximation [16] of the conventional, complex B1+-based method. These fast methods, however, intrinsically do not allow reconstruction of permittivity. Furthermore, the speed afforded by the phase-only approximation comes at a cost of additional systematic error in conductivity estimation (at the level of 10%) [17, 18]. Marques et al. have reported a novel, single-acquisition mapping of EPs using relative sensitivity maps of multiple receive coils [19]. This method enabled both permittivity and conductivity mapping without B1+, and was compatible with a wide range of imaging sequences. However, its dependence on the 3rd order derivative of the coil sensitivity data made it more sensitive to input image noise.

In this work, we show that by rearranging Maxwell's equations, conductivity and permittivity mapping is possible from a single, low flip-angle GRE acquisition. In particular, we propose to use a zero-echo-time (ZTE) sequence, due to its immunity to eddy current and static magnetic field (B0) inhomogeneity-induced phase changes, as well as its speed and SNR efficiency. By utilizing both the magnitude and phase information of the acquired images, our method achieves an acquisition speed higher than in the UTE method [15] while allowing both conductivity and permittivity reconstruction. In our approach, numerical processing of the image data can share the same routines developed for the conventional method [17], and therefore does not add more numerical complexity or processing time.

After theoretical derivation of the method, we show results obtained with the new formalism in phantom experiments. EPs reconstructed from human in-vivo brain images using the new formalism are also presented. The advantages and disadvantages of the method are discussed in the Discussion section.

II. THEORY

A. Background and motivation for a new formalism

In a medium of constant electrical conductivity (σ) and relative permittivity (εr), each of the circularly polarized components of the RF magnetic field satisfies the wave equation:

2B1++k2B1+=0 (1)
2B1+k2B1=0 (2)

with the complex wave vector given by

k2=µεrε0ω2iµσω. (3)

Here B1+(B1) is the complex amplitude of the transmit (receive) RF magnetic field inside the medium, µ is the magnetic permeability, ε0 = 8.854 × 10−12 F/m is the permittivity of vacuum, and ω = 2πf = γB0 is the RF frequency; γ is proton's gyromagnetic ratio. In biological tissue, µ can be replaced by the permeability in vacuum, µ = µ0 = 4π × 10−7 H/m.

A challenge in applying Eqs. (13) to MREPT is that in MRI, the absolute phase of either B1+orB1 is not directly accessible. A much used approximation is to replace the B1+ phase by one half of the transceive phase [12],

B1+12(B1++B1++B1). (4)

Eqs. (1,3,4) lead to the conventional MREPT equations,

εr1µε0ω2Re(2|B1+|ei(B1++B1)/2|B1+|ei(B1++B1)/2) (5)
σ1µωIm(2|B1+|ei(B1++B1)/2|B1+|ei(B1++B1)/2). (6)

Note that in this as well as in the new approach below, the z component of the RF magnetic field, B1z, is not needed.

Eq. (4) transmit-receive coil in vacuumor when the imaged object possesses certain symmetry [20]. In general, interaction of the tissue with the RF field renders B1+andB1 different from each other, with the difference growing at higher frequencies [21, 22]. Since tissue-RF interaction is precisely what makes MREPT possible, this difference is unavoidable. In Appendix 1, we show that in EP reconstruction based on B1+ magnitude and Eq. (4), the error in EPs resulting from the B1+,B1 difference is a first order effect with respect to the difference. In human MRI at 3 T and higher, such error can be significant. Thus, an important motivation of the present work is to demonstrate a theoretical formula which involves approximation that is less sensitive to the difference between B1+andB1.

B. Derivation

By multiplying Eq. (1) with B1 and Eq. (2) with B1+, and adding them together, we obtain the following:

2(B1+B1)+2k2B1+B12B1+·B1=0. (7)

Replacing B1+B1with(B1+,B1)2, we can show that (see Appendix 2 for details)

2B1+B1B1+B1+k2+kerror2=0 (8)

in which

kerror214lnB1B1+·lnB1B1+. (9)

Here the square root of a complex quantity B1+B1 can be defined unambiguously by defining the phase of B1+B1 as 1/2 times the phase of B1+B1; as long as the latter is properly phase-unwrapped to avoid phase jump in the region of interest (ROI), so is B1+B1 free from phase jump.

Ignoring kerror2 and substituting Eq. (3) for k2 in Eq. (8) lead to Eqs. (1011). These define the proposed theoretical method of image-based EP mapping.

εr1µε0ω2Re(2B1+B1B1+B1) (10)
σ1µωIm(2B1+B1B1+B1). (11)

C. Error estimation and analytical examples

In the limit where B1+=B1 (up to a constant factor, which we assume to be 1 without loss of generality), the error term kerror2 is zero. When the ratio B1/B1+ is spatially varying but is close to unity, kerror2 can be estimated as

kerror214(B1B1+B1+)·(B1B1+B1+). (12)

Here we used the leading order Taylor expansion ln(1 + z) ≈ z for small 𝓏. Appendix 2 further breaks down kerror2 into the real and imaginary parts and expresses each of them in terms of the magnitude and phase differences of B1+andB1 (Eqs. (A19, A20)). The analysis shows that the error in the EPs is a second order effect. This is in contrast to the first order effect in the conventional method (Appendix 1). This provides a support for an argument that, when B1+,B1 are close to each other, our method can be advantageous over the conventional one in terms of the theoretical accuracy of EP mapping.

We have previously demonstrated theoretical advantage of the new method through an RF simulation study [23]. Here we illustrate the benefit of the new method using known analytical solutions of the RF field inside an elliptical cylinder. Figure 1 shows the EPs of a homogeneous, elliptical cross-section cylinder reconstructed from analytical RF field solutions [24]. The cylinder, infinitely long, is axially located inside a circularly polarized RF transmit-receive coil (at 128 MHz) and has the following parameters: major axis = 20 cm, minor axis = 15 cm, εr = 80, σ = 0.8 S/m. The B1+andB1 fields differ from each other by up to 37%, which did not make significant errors in εr and σ reconstructed by our method. On the other hand, in the εr and σ maps obtained by the conventional method, errors are very noticeable in regions with different B1+andB1. Similar errors in an object with broken cylindrical symmetry was reported in [18].

Fig. 1.

Fig. 1

EPs of a homogeneous elliptical cylinder reconstructed from analytical RF solutions. (a) Magnitudes of B1+andB1 in a circularly polarized transmit-receive coil. (b) Magnitude of the fractional difference. (c) True EPs. (d,e) Reconstructed EPs from Eqs. (1011) (d), and Eqs. (56) (e).

Next, we examine the magnitude of kerror2 in comparison with the corresponding quantity in the conventional method for a range of model EPs. Figure 2(a,b) shows the magnitude and phase difference, respectively, between analytically calculated B1+andB1 for different combinations of the model EPs of the cylinder, ranging from εr = 10 to 90, and σ = 0.1 to 0.9 S/m. Figure 2(c) shows the map of the relative error magnitude |kerror2/k2|, whereas Fig. 2(d) shows the corresponding quantity for the conventional method. Both methods show small errors when the B1+,B1 difference is small. As the latter grows, however, the error in the new method grows much more slowly than the conventional method. For example, for the geometry considered the new method is nearly error-free at σ> 0.5 S/m or εr < 50.

Fig. 2.

Fig. 2

Magnitude (a) and phase (b) difference between analytically calculated B1+andB1 in an elliptical cylinder (at 128 MHz), and theoretical error in k2 in the proposed (c) and the conventional (d) method. Color scales are identical for all images. Each ellipse represents an axial cross section of an infinitely long elliptical cylinder (inset).

D. Phase-only method

When the magnitude of B1+B1 varies more slowly in space than its phase, Eq. (11) can be further approximated as

σ1µω2ϕ (13)

where ϕ is the phase of B1+B1. Here we used the fact that, after ignoring the magnitude variation, the imaginary part of the fraction (∇2e)/e equals ∇2ϕ. This represents a previously known, phase-only conductivity mapping method [15, 16, 25] derived in a different way.

E. Image-based tissue electrical property mapping

A practical advantage of Eqs. (10,11) is that the magnitude and phase of the product B1+B1 are more readily available from standard MR images than individual complex quantities B1+andB1. For example the complex image from a low-flip-angle spoiled GRE scan can be expressed as

I=I0B1+B1. (14)

Here I0 is the "true" MR image determined by the proton density and other tissue contrasts. Separating out the RF term B1+B1 from the as-acquired image I has been a subject of intense research [26] for the purpose of image shading removal in high-field MRI. Whereas some of those methods could be adopted for our purpose, here we pay attention to the fact that MREPT typically targets localized reconstruction of piece-wise constant electrical properties. That is, because Eqs. (1,2) are only valid in regions of constant electrical properties, EP mapping methods derived from these equations are considered valid only in regions of constant EPs, such as regions of a single tissue type. If we therefore restrict ourselves to ROIs each of which containing a single tissue type, it is permissible to assume that I0 is nearly constant in each region. We can then substitute Eq. (14) for B1+B1 in Eqs. (10,11), where I0 factors out of the Laplacian operator and cancels out in the fraction. This results in a desired, image-based EP mapping method without B1+ mapping. We reiterate that the assumption of constant I0 implies that the method best holds for individual segments in a tissue-segmented image, as was considered, for example, in [27, 28].

In practice, a GRE image can contain B0 inhomogeneity-induced, spatially varying phase even inside a single tissue ROI, complicating isolation of the B1+B1 phase. While a separate SE scan can help this, in this work we propose using a low-flip-angle ZTE image for simultaneous acquisition of the magnitude and phase of B1+B1. A ZTE sequence has negligible time for B0 induced phase change, and in our implementation its limited gradient switching also minimizes eddy current concerns. In what follows, we calculated EPs by replacing B1+B1 of Eqs. (10,11) by a complex ZTE image IFGH.

F. Use of a receiver array

Suppose that a low-flip-angle ZTE image was obtained with an N-channel receiver array. The complex combined image is proportional to

IZTEB1+·(c1B1(1)+c2B1(2)+cNB1(N)) (15)

where c1, c2,⋯, cN are the complex weights. If we denote the quantity in parenthesis in Eq. (15) as B1(tot), such quantity satisfies the wave equation Eq. (2), as does each individual receiver field map. Therefore, the derivation leading to Eqs. (10,11) is valid as we replace B1byB1(tot) in Eqs. (711), and B1+B1(tot) can be replaced by IZTE, provided, as before, the true (unshaded) image varies slowly in an ROI. Here, the error in EP calculation (Eq. (12)) is determined by the fractional difference between the transmit B1+ field (usually from a body coil) and the combined receiver field B1(tot). This suggests that the complex weights should be adjusted to better match the two RF field maps. When one of the RF field maps (typically B1+) is approximately homogeneous over an ROI, matching the two field maps roughly corresponds to homogenizing the magnitude of the image IZTE itself. Such optimization can be done separately for each ROI in which EP mapping is desired. Since homogenizing |IZTE| aims at matching B1(tot)with/B1+, rather than with B1+, it will be only effective in improving EP mapping results when B1+ itself is substantially homogeneous. We note that complex coil combination to improve the accuracy of MREPT, assuming homogeneity of body-coil B1+, was demonstrated in [25] in the context of phase-only conductivity mapping.

III. EXPERIMENTAL METHODS

A. Image acquisition

All images were acquired with a Discovery 3.0 T MR750 scanner (GE Healthcare, Waukesha, WI) with a standard GE head birdcage coil and a GE 8-channel phased array coil.

ZTE imaging

ZTE imaging was implemented in a way similar to the RUFIS method [29]. This pulse sequence consists of a nonselective hard pulse excitation followed by three-dimensional (3D) center-out radial sampling. Image encoding starts immediately, leading to a nominal echo time TE = 0; consequently, the acquired images have little sensitivity to chemical shift and B0 inhomogeneities. Note that the three sources of off-resonance phase accumulation in UTE imaging [30], i.e., finite excitation pulse width, ramp sampling, and readout interval, are minimized or eliminated through use of a short hard pulse, lack of ramp sampling (Fig. (3)), and high-bandwidth readout. The effective off-resonance dephasing time estimated from the theories of ref. [30] was on the order of 20 µs.

Fig. 3.

Fig. 3

Pulse sequence for zero TE imaging. Four “spokes” in a 3D radial k-space scan are shown.

The pulse sequence was grouped into segments, each containing a number of spokes; a diagram of one of the segments is shown in Fig. (3). For typical B1+ amplitudes of clinical whole-body MR scanners (~15 µT), flip angles of only a few degrees are possible with the hard pulse, satisfying the low-flip-angle requirement. The 3D radial spokes were uniformly distributed in all directions and sequentially ordered along a spiral path [31]. This limits the gradient switching between repetitions to incremental directional updates only, resulting in reduced eddy currents. In our case the amplitude of the encoding gradient was 4.9 mT/m, with each update involving gradient switching of amplitude of only 0.13 mT/m.

The following parameters were used for both the phantom and in-vivo ZTE scans: Number of spokes = 1282, TE/TR = 0/1 ms, bandwidth = ±31 kHz, flip angle = 3°, FOV = 28~30 cm, resolution = 128 µ 128 µ 128, slice thickness = 2.2~2.3 mm. Each 3D scan lasted for 27 seconds.

Phantom scan

Three cylinders with diameters 7.7 cm, 7.7 cm, 9.4 cm were filled with water with varying salt concentrations. The cylinders were placed with their axes parallel to the main magnetic field. ZTE imaging on the phantom was done with both a head transmit/receive (T/R) coil and an 8 channel brain array. To verify the measured electrical properties, the same phantom was also scanned for conventional MREPT using B1+ map and SE phase [32]. The conventional scan data were acquired with the head T/R coil on 15 slices with slice thickness = 2 mm, FOV = 30 cm, in-plane resolution = 128 × 128, and SAR-limited scan time of 11 min (B1+ map) and 3 min 44 sec (SE). Finally, the conductivity of liquids in all cylinders at 128 MHz was measured with an impedance analyzer (E4991A) coupled to a dielectric probe (E85070) (Agilent Technologies, CA, USA).

In-vivo scan

All in-vivo ZTE scans were carried out with the head T/R coil. Informed consent was obtained from the two volunteers enrolled for the study according to the Institutional Review Board of the lead author’s institution.

B. Data processing and EP reconstruction

EPs on a particular slice were calculated from the Laplacian of a 3D complex map of IZTE centered around the slice of interest. For the 8 channel ZTE data, the individual coil’s complex images were combined with complex weights determined by:

{cn}=argmin{cn}1/min(c1I1+c2I2++c8I82), (16)

where |c1| = ⋯ = |c8| = 1. Here I1,⋯, I8 are individual complex ZTE images. One set of weights cn was determined for each segment (cylinder) of the phantom. The optimization procedure found relative phases of different channels that maximized the minimum intensity of the combined image. Such max-min optimization was previously used in transmit RF shimming [33]. Phase-only optimization was adopted here because it was found to produce more smoothly-varying image phase than full complex optimization.

The Laplacian was numerically calculated by quadratic fitting in the three Cartesian directions [27]; depending on the proximity to the image mask boundary, typically 4–6 pixels on each side of the pixel of interest were used for fitting in one direction. To quantify the accuracy and precision of the method, the mean and pixel standard deviation of the EPs of each cylinder were calculated on an axial slice.

In-vivo brain EPs were reconstructed pixel by pixel without tissue segmentation. No post-processing filtering was applied to the EP maps. In the display of the conductivity results below, pixels with diverging conductivities near the skull (coming from reconstruction across tissue boundaries) were excluded by truncating the conductivity map with a mask obtained by 2D region growing from the center of the brain. For this a Matlab (Mathworks, Natick, MA, USA) code was written which let a user to graphically select a seed pixel in the raw conductivity map, from which a region was grown by adding neighboring pixels that satisfied a homogeneity criterion.

IV. RESULTS

Figure 4(a) shows the measured |B1+| map and one half of the spin echo phase (ϕSE) map of the phantom on an axial slice, as well as the EP maps calculated from Eqs. (5,6). Figure 4(b,c) shows the magnitude and phase of the square root of IZTE and the EP maps obtained by the new method. Comparison of Fig. 4(a–c) shows that the new method, with both the head T/R coil (Fig. 4b) and the array coil (Fig. 4c), produced EPs that are in good agreement with the conventionally obtained map as well as the true map (Fig. 4e). For the 8 channel ZTE data, the individual coils' magnitude images and coil combination phases ϕc are shown in Fig. 4(d). The ϕc plot emphasizes the fact that the combination coefficients were determined for each cylinder separately.

Fig. 4.

Fig. 4

Phantom data. (a) |B1+| map (normalized) and one-half of the SE phase on an axial slice, and EPs calculated from them using Eqs. (5,6). (b) Magnitude and phase of IZTE, and EPs calculated with the new method. Head T/R birdcage coil was used. (c) Same as (b) but with body coil transmission and an 8-channel array reception, after coil combination. (d) 8-channel IZTE magnitudes and coil combination phases optimized per segment. (e) True EPs.

Table 1 shows the mean and the pixel standard deviation of the EPs in each cylinder. The statistics on the phantom EPs obtained by the ZTE method are comparable to those obtained conventionally using B1+ and ϕSE. Given a significant scan time advantage of the ZTE method compare to B1+ map-based methods, our results suggest that the proposed method can be useful for fast estimation of both permittivity and conductivity in piece-wise homogeneous objects.

Table 1.

Measured (mean ± standard deviation) and true EPs in each cylinder (segment) of the phantom. True values are taken from the literature (permittivity) and dielectric probe measurement (conductivity).

Method Segment εr σ [S/m]
ZTE head T/R 1 83.7 ± 15.7 0.11 ± 0.09
2 88.6 ± 15.1 0.44 ± 0.11
3 82.0 ± 14.9 0.90 ± 0.10
8 ch array 1 89.1 ± 16.0 0.08 ± 0.08
2 90.3 ± 12.6 0.45 ± 0.11
3 84.6 ± 12.7 0.94 ± 0.09
B1+ and SE with head T/R 1 79.1 ± 12.1 0.10 ± 0.09
2 80.9 ± 18.3 0.43 ± 0.11
3 78.3 ± 11.1 0.95 ± 0.11
True values 1 80 0.08
2 80 0.46
3 80 0.92

Figure 4(a,b) shows high degree of similarity between the (normalized) magnitude of IZTE and the |B1+| map of the phantom. This is expected since for axially located cylindrical objects in a T/R birdcage coil, |B1+|=|B1|and|IZTE| is proportional to their product (Eq. (14)).

Figure 5(a,b) shows orthogonal 3-plane in-vivo ZTE images and conductivity maps for two volunteers. The conductivity maps in the brain are shown overlaid on the amplitude images of the head. The average conductivity in the brain for the masked pixels (obtained by numerical averaging of all color pixel values in the three planes shown) was 0.55 S/m and 0.56 S/m for volunteers 1 and 2, respectively. The quality of the conductivity maps compares favorably with previously reported results obtained with an ultra-short TE sequence [15]. Whereas no attempt was made to segment the brain into different tissue types, the observed average conductivity is compatible with the brain tissue conductivity values in literature: white matter (0.34 S/m), grey matter (0.59 S/m), and cerebrospinal fluid (2.14 S/m) [34, 35].

Fig. 5.

Fig. 5

(a–b) Three-plane in-vivo brain ZTE images (amplitude, left column; phase, middle column) and conductivity maps (right column) overlaid on the amplitude image. Average conductivity values were 0.55 S/m (volunteer 1) and 0.56 S/m (volunteer 2). (c–d) Comparison of conductivity maps reconstructed from IZTE phase only vs complex IZTE. Black arrows indicate noise that is stronger in phase-only reconstruction.

Significant noise and artifacts in in-vivo EP maps can come from reconstruction errors near tissue boundaries [36]. Whereas this is the case for any method based on Eqs. (1,2), an image-based method can be particularly sensitive to boundary effects, since in addition to EP discontinuities, image contrast at tissue boundaries can cause large errors in Laplacian calculation. In-vivo permittivity maps of Fig. 6(a,b) demonstrate this difficulty. Many pixels with unphysical permittivity values (negative or greater than 100) appear near tissue boundaries in the corresponding |IZTE| images.

Fig. 6.

Fig. 6

Normalized IZTE image amplitudes on a zoomed color scale (left) and reconstructed permittivity maps (right) on an axial slice of head. Black arrows indicate tissue boundaries around which unrealistic permittivity values were obtained (white arrows).

Thanks to the phase dominance in conductivity reconstruction [16], in-vivo conductivity maps were less affected by the |IZTE| boundary effect. In order to check this, we recalculated conductivity maps with the IZTE phase only, emulating the method of [15]. Figure 5(c,d) shows the comparison between conductivity maps obtained from IZTE phase only vs complex IZTE. Qualitatively, including IZTE amplitude did not appear to increase the noise beyond what was already present in phase-only reconstruction (arrows). On the other hand, the difference map shows, consistently for the two volunteers, that phase-only reconstruction estimates conductivity about 10% higher than full IZTE reconstruction. This observation is interestingly similar to the previously reported amount of overestimation of conductivity when only phase is used for conductivity reconstruction [17]. This suggests that full IZTE reconstruction is likely more accurate than phase-only reconstruction of conductivity; i.e., including |IZTE| in our method has merit.

V. DISCUSSION

A. Comparison with B1+ mapping-based methods

Quantitative comparison between different EP mapping methods must take into account different theoretical assumptions, image acquisition methods, and reconstruction algorithms. A recent review [17] provided a brief comparison between some of the existing, B1+ map-based MREPT methods. Whereas a comprehensive comparative analysis is outside the scope of this work, here we provide a qualitative explanation of the differences between the new, ZTE-based method and the B1+ map-based ones.

Scan time

The scan time advantage of the new method comes mainly from the elimination of the B1+ mapping step. Phase acquisition in both the conventional and the new methods can be quick; for example, a UTE [15] and a steady-state precession [37] sequence have been used for rapid conductivity mapping using phase acquisition alone. However, in order to map permittivity, or to improve the accuracy in conductivity, conventional methods require an additional B1+ mapping scan. The speed and accuracy of different B1+ mapping methods have been compared in a recent review [38]. It showed that the Bloch-Siegert phase shift method [8] provided good accuracy (a few % error) over a large flip angle range which is desirable for MREPT. In the same study, a representative scan time for in-vivo B1+ mapping at 3T with matrix size 128 × 50 × 48 was 4 min 48 seconds, limited by SAR. We estimate therefore that a 3D B1+ mapping scan with resolution comparable to our 128 × 128 × 128 ZTE scan would be at least 10 minutes. This compares with 27 seconds used for the ZTE scan. Note that SAR limitation in B1+ mapping worsens at higher magnetic fields due to increased induced electric fields at higher RF frequencies. Overall, we estimate that the proposed method has scan time advantage of at least an order of magnitude compared to the conventional methods for a complete 3D EP mapping.

Accuracy

Main sources of errors in most common MREPT methods include: (i) systematic errors due to applying Eqs. (1,2) to EP-heterogeneous medium, (ii) systematic errors associated with replacing exact equations with approximate ones expressed in measurable quantities (such as replacing B1+ phase with one half of the transceiver phase), and (iii) random noise, especially amplified by derivative operation. Our method does not necessarily address the first and the third sources better than the existing methods, but rather improves EP accuracy by reducing the second type of error (when, of course, the other sources are not the limiting factors), through utilizing approximations (Eqs. (10, 11)) that are less sensitive to experimental conditions. Specifically, methods relying on transceiver phase approximation lose accuracy quickly as the object becomes asymmetric and the B1+, B1 difference grows. The proposed method, on the other hand, is less sensitive to B1+, B1 difference as was demonstrated here and previously [23] through numerical models.

Artifacts

Since our method operates directly on ZTE images, any image intensity artifact will more directly affect the EP results than in conventional methods. For example, radial k-space under-sampling is known to produce streaking artifacts [39]. In our case the number of radial spokes (1282) was a factor of π smaller than what is required for Nyquist sampling. Although this did not cause significant artifacts in the images themselves, a low level artifact can be amplified to a high level through Laplacian operation. Similarly, Gibbs ringing from k-space truncation, commonly present near sharp edges of a phantom, can be more problematic in the Laplacian of the image. Our ZTE protocol is relatively insensitive to off-resonance and motion-related artifacts due to a high readout bandwidth and lack of ramp sampling. However, the protocol is more demanding of precise timing of RF and gradient pulses. In this work we have not investigated imperfections in the pulse sequence realization which can contribute to the observed EP noise.

In comparison, B1+ map-based methods can be sensitive to B1+ map inaccuracies caused by conditions such as relaxation time variation and resonance offset [38, 40]. Also, artifacts in images used to calculate B1+ can translate into EP artifacts, although such link is expected to be weak for phase-based B1+ mapping.

B. Comparison with phase-only conductivity mapping

B1+ phase-based conductivity mapping [16] shares two advantages with our method. First it allows fast, B1+ amplitude-free conductivity reconstruction [14, 15]. Second, separate transmit and receive coils can be used, including a receiver array [25, 40]. However, the phase-only method is different from our complex ZTE image-based method in the following ways. First, it is not applicable to permittivity, even in a homogeneous medium. Second, for conductivity, the phase-only method is known to introduce additional errors compared to when B1+ amplitude is included in reconstruction. In ref. [17] such error was estimated to be about 10%, whereas our earlier analytical model [23] showed that the error can be much bigger near the edge of a high conductivity medium. What we demonstrated in this work is that, by utilizing the magnitude as well as the phase of transceive B1 (measurable through IZTE), one can have a fast and RF-coil-flexible EP mapping method that includes permittivity, and estimates conductivity more accurately.

C. Prospect for in-vivo permittivity mapping

A challenge of the proposed method is that permittivity reconstruction is more severely affected by image intensity artifacts and tissue contrast, as compared to B1+ map-based methods. We have shown that including |IZTE| in brain conductivity mapping did not increase the noise compared to phase-only reconstruction. For permittivity, however, we found that in-vivo permittivity maps calculated from IZTE were of poorer quality than typical published permittivity maps in brain based on B1+ maps. This is likely because permittivity is more affected by IZTE magnitude than its phase, and IZTE magnitude has more pronounced image discontinuity at tissue boundaries than IZTE phase or B1+. In order for the proposed method to reliably produce permittivity values in vivotissue of interest (e.g. tumor) should be segmented out first [27, 28] so that EP reconstruction is done within a segment of homogeneous tissue properties as much as possible. Further validation of the ZTE method for in-vivo permittivity mapping remains as a future work. Its success will likely depend on the quality and reliability of tissue segmentation, as well as the number of pixels within each segment.

VI. CONCLUSION

In conclusion, we have demonstrated a fast EP mapping method based on ZTE imaging. Compared to the conventional method, the proposed method is more tolerant to differences in spatial distributions of B1+ and B1 that arise from tissue-RF interaction. It also allows EP reconstruction from images acquired with receiver arrays. Its present limitation lies in greater boundary artifacts due to image intensity discontinuity at tissue contrast boundaries. For in-vivo applications, therefore, tissue segmentation to minimize the boundary effects is key to taking full advantage of the new method. Application of the method to EP mapping in anatomies where receiver arrays are typically used for imaging, such as breast, is one of the future research directions.

Acknowledgments

This work was supported by the Grant R01CA154433 from the National Cancer Institute.

APPENDIX 1: ERROR TERM IN CONVENTIONAL METHOD

The following shows how the error involved in the approximation Eq. (4) propagates into EP calculation.

The quantity which approximates the true complex B1+ in the conventional method is

|B1+|exp(iB1++B12)=|B1+|exp(iB1+)exp(iB1B1+2)=B1+exp(iδp/2) (A1)

where

δpB1B1+. (A2)

The Laplacian of the right-hand side of Eq. (A1) is

2(B1+eiδp/2)=eiδp/2{2B1++B1+(i2δp/2δp·δp/4)+iB1+·δp}. (A3)

If we divide this equation by B1+eiδp/2, and use the fact that (2B1+)/B1+=k2, we obtain

2(B1+eiδp/2)B1+eiδp/2+k2+kerror,conv2=0 (A4)

where

kerror,conv214δp·δpi22δpiB1+B1+·δp. (A5)

Substituting B1+=|B1+|eiB1+ in the above equation, we can separate the real and imaginary parts of the error term as

Re(kerror,conv2)=14δp·δp+B1+·δp, (A6)
Im(kerror,conv2)=122δp|B1+||B1+|·δp. (A7)

Dividing Eqs. (A6, A7) by µε0ω2 and −µω, respectively, gives the errors in εr and σ involved in the conventional method. We note that both Eqs. (A6, A7) contain a term (second term on the right-hand side) that is a product of ∇δp and a quantity that is not necessarily as small as ∇δp. As δp departs from zero, therefore, we can say that the electrical property errors in the conventional method grow as a first order effect with respect to δp. Note that while "first order effect" refers to an order-of-magnitude estimate for kerror,conv2, it does not necessarily mean that the actual spatial variation of errors in k2 is proportional to δp, since Eqs. (A6, A7) depend on derivatives of δp as well as B1+.

APPENDIX 2: DERIVATION OF EQS. (8,9) AND SMALLNESS OF kerror2

Let us define aB1+B1andbB1B1+. Then Eq. (7) can be rewritten as

2a2+2k2a22(ab)·(ab)=0. (A8)

Using the identity ∇2a2 = 2a2a + 2∇a · ∇a, we can rewrite this as

2a2a+2a·a+2k2a22(ab)·(ab)=0. (A9)

Dividing Eq. (A9) by 2a2, we get

2aa+k2+kerror2=0 (A10)

in which

kerror21a2(a·a(ab)·(ab))=1a2(a·a(abab2b)·(ab+ba))=1b2b.b=14lnb2·lnb2. (A11)

Finally, substituting the definitions of a and b in Eqs. (A10, A11) gives Eqs. (8,9), respectively.

Now we estimate the size of the real and imaginary parts of kerror2, to be compared with Eqs. (A6, A7). Recall that we are interested in kerror2 when B1+andB1 are close to each other. The closeness can be formally defined in terms of the magnitude of the complex fractional difference δB,

δB|B1B1+B1+|1. (A12)

Under this condition, we consider two additional small quantities; the fractional magnitude difference δm,

δm|B1||B1+||B1+|. (A13)

and the phase difference δp of Eq. (A2).

Figure (A1) graphically demonstrates the inequalities:

|δm|δB (A14)
|δp|arcsinδBδB. (A15)

graphic file with name nihms695097f7.jpg

Fig. A1. Relationship among magnitude, phase, and complex differences between B1+andB1. In the complex plane, B1 lies on a circle centered at B1+ with radius |B1B1+|. Among all possibilities, |B1| is the most different from |B1+|whenB1 is at an intersection between the circle and a line passing B1+ and the origin (O). The maximum magnitude difference is |B1B1+|. This proves Eq. (A14). The maximum phase difference occurs when the line OB1 is tangential to the circle. The phase difference δp,max in this case satisfies sin δp,max = δB, proving Eq. (A15).

The last approximation follows from δB ≪ 1. Eqs. (A14, A15) say that δm, δp are always both at most as small as δB. This allows us to use δm, δp as well as δB as a small parameter in the error analysis.

We now express the real and imaginary parts of (B1B1+)/(B1+) in terms of δm and δp.

Re(B1B1+B1+)=|B1||B1+|cosδp1=(1+δm)(112δp2+14!δp4)1=δm12δp212δmδp2+=δm+O(δB2). (A16)

Here O(δB2) means that the rest of the terms are at most on the order of δB2. In the second line of Eq. (A16) we used the Taylor expansion of the cosine function.

Similarly, using the Taylor expansion of sin x we get

lm(B1B1+B1+)=|B1||B1+|sinδp=(1+δm)(δp13!δp3+)=δp+O(δB2). (A17)

Eqs. (A16, A17) show that, to the leading order, the real and imaginary parts of (B1B1+)/(B1+) equal δm and δp, respectively.

Using these relations, Eq. (12) can be expressed as

kerror214(δm+iδp)·(δm+iδp) (A18)

from which we get

Re(kerror2)14δp·δp+14δm·δm, (A19)
lm(kerror2)12δm·δp. (A20)

All terms on the right-hand side contain a product of a pair of small quantities. In this sense, the theoretical error in k2 involved in our method is a second order effect with respect to the B1+,B1 difference.

Comparing Eq. (A19) with Eq. (A6), we find that Re(kerror,conv2) will be larger than Re(kerror2) in magnitude if B1+·δp is much larger than ∇δm · ∇δm/4 in magnitude. Although this may not be true in all cases, it is still expected to hold in a majority of cases given the smallness of δp, δm. Similarly, comparing Eq. (A20) with Eq. (A7), we find that although |Im(kerror,conv2)|>|Im(kerror2)| may not be always true (depending on the specifics of B1+,B1), it may hold in most cases due to the second order nature of Eq. (A20). Lack of rigorous proofs calls for experimental tests for practical validation. Numerical examples in the text provide a support for the qualitative arguments made here.

Contributor Information

Seung-Kyun Lee, Email: lsk@ge.com, GE Global Research, Niskayuna, NY, USA.

Selaka Bulumulla, Email: bulumull@ge.com, GE Global Research, Niskayuna, NY, USA.

Florian Wiesinger, Email: florian.wiesinger@research.ge.com, GE Global Research, Munich, Germany.

Laura Sacolick, Email: sacolick@ge.com, GE Healthcare, Munich, Germany.

Wei Sun, Email: Wei.Sun@geahk.ge.com, GE Healthcare, Waukesha, WI, USA.

Ileana Hancu, Email: hancu@ge.com, GE Global Research, Niskayuna, NY, USA.

REFERENCES

  • 1.Yeo DT, Wang Z, Loew W, Vogel MW, Hancu I. Local specific absorption rate in high-pass birdcage and transverse electromagnetic body coils for multiple human body models in clinical landmark positions at 3T. J Magn Reson Imaging. 2011 May;33:1209–1217. doi: 10.1002/jmri.22544. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Lazebnik M, Popovic D, McCartney L, Watkins CB, Lindstrom MJ, Harter J, Sewall S, Ogilvie T, Magliocco A, Breslin TM, Temple W, Mew D, Booske JH, Okoniewski M, Hagness SC. A large-scale study of the ultrawideband microwave dielectric properties of normal, benign and malignant breast tissues obtained from cancer surgeries. Phys Med Biol. 2007;52:6093–6115. doi: 10.1088/0031-9155/52/20/002. [DOI] [PubMed] [Google Scholar]
  • 3.Chaudhary SS, Mishra RK, Swarup A, Thomas JM. Dielectric properties of normal & malignant human breast tissues at radiowave & microwave frequencies. Indian J Biochem Biophys. 1984 Feb;21:76–79. [PubMed] [Google Scholar]
  • 4.Joines WT, Zhang Y, Li C, Jirtle RL. The measured electrical properties of normal and malignant human tissues from 50 to 900 MHz. Med Phys. 1994 Apr;21:547–550. doi: 10.1118/1.597312. [DOI] [PubMed] [Google Scholar]
  • 5.Surowiec AJ, Stuchly SS, Barr JB, Swarup A. Dielectric properties of breast carcinoma and the surrounding tissues. IEEE Trans Biomed Eng. 1988 Apr;35:257–263. doi: 10.1109/10.1374. [DOI] [PubMed] [Google Scholar]
  • 6.Haacke E, Petropoulos L, Nilges E, Wu D. Extraction of conductivity and permittivity using magnetic resonance imaging. Phys Med Biol. 1991;36:723–734. [Google Scholar]
  • 7.Nehrke K, Bornert P. DREAM--a novel approach for robust, ultrafast, multislice B(1) mapping. Magn Reson Med. 2012 Nov;68:1517–1526. doi: 10.1002/mrm.24158. [DOI] [PubMed] [Google Scholar]
  • 8.Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B1 mapping by Bloch-Siegert shift. Magn Reson Med. 2010 May;63:1315–1322. doi: 10.1002/mrm.22357. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med. 2007 Jan;57:192–200. doi: 10.1002/mrm.21120. [DOI] [PubMed] [Google Scholar]
  • 10.van Lier ALHMW, Brunner DO, Pruessmann KP, Klomp DWJ, Luijten PR, Lagendijk JJW, van den Berg CAT. B1+ Phase mapping at 7 T and its application for in vivo electrical conductivity mapping. Magn Reson Med. 2012;67:552–561. doi: 10.1002/mrm.22995. [DOI] [PubMed] [Google Scholar]
  • 11.Wiesinger F. Prospects of absolute B1 calibration. Unsolved Problems and Unmet Needs in Magnetic Resonance. Proc 14th Annual Meeting of ISMRM; Seattle, WA, USA. 2006. [Google Scholar]
  • 12.Katscher U, Voigt T, Findeklee C, Vernickel P, Nehrke K, Dossel O. Determination of electric conductivity and local SAR via B1 mapping. IEEE Trans Med Imaging. 2009 Sep;28:1365–1374. doi: 10.1109/TMI.2009.2015757. [DOI] [PubMed] [Google Scholar]
  • 13.Bulumulla SB, Yeo DT, Zhu Y. Direct calculation of tissue electrical parameters from B1 maps. Proc 17th Annual Meeting of ISMRM; Honolulu, HI, USA. 2009. p. 3043. [Google Scholar]
  • 14.Stehning C, Voigt TR, Katscher U. Real-Time Conductivity Mapping using Balanced SSFP and Phase-Based Reconstruction. Proc 19th Annual Meeting of ISMRM; Montreal, Canada. 2011. p. 128. [Google Scholar]
  • 15.Schweser F, Huang K-H, Herrmann L, Kramer M, Deistung A, Reichenbach R. Conductivity mapping using Ultrashort Echo Time (UTE) imaging. Proc 21st Annual Meeting of ISMRM; Salt Lake City, UT, USA. 2013. p. 4190. [Google Scholar]
  • 16.Voigt T, Katscher U, Doessel O. Quantitative conductivity and permittivity imaging of the human brain using electric properties tomography. Magn Reson Med. 2011 Aug;66:456–466. doi: 10.1002/mrm.22832. [DOI] [PubMed] [Google Scholar]
  • 17.Katscher U, Kim DH, Seo JK. Recent progress and future challenges in MR electric properties tomography. Comput Math Methods Med. 2013:546562. doi: 10.1155/2013/546562. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.van Lier ALHWW, Raaijmakers A, Voigt T, Lagendijk JJW, Luijten PR, Katscher U, van den Berg CAT. Electrical properties tomography in the human brain at 1.5, 3, and 7T: a comparison study. Magn Reson Med. 2014;71:354–363. doi: 10.1002/mrm.24637. [DOI] [PubMed] [Google Scholar]
  • 19.Marques JP, Sodickson DK, Ipek O, Collins CM, Gruetter R. Single acquisition electrical property mapping based on relative coil sensitivities: A proof-of-concept demonstration. Magn Reson Med. 2014 Aug;5 doi: 10.1002/mrm.25399. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Lee S, Bullumula S, Dixon W, Yeo D. Workshop on MR based impedance imaging. Seoul, Korea: 2010. B1+ phase mapping for MR based electrical property measurement of a symmetric phantom. [Google Scholar]
  • 21.Watanabe H. Investigation of the asymmetric distributions of RF transmission and reception fields at high static field. Magn Reson Med Sci. 2012;11:129–135. doi: 10.2463/mrms.11.129. [DOI] [PubMed] [Google Scholar]
  • 22.Seo JK, Kim DH, Lee J, Kwon OI, Sajib SZK, Woo EJ. Electrical tissue property imaging using MRI at dc and Larmor frequency. Inverse Problems. 2012;28:1–26. [Google Scholar]
  • 23.Bulumulla SB, Lee S-K, Hancu I. Comparison of data approximation methods used in MR-based tissue electrical property mapping – a simulation study. Proc 21st Annual Meeting of ISMRM; Salt Lake City, UT, USA. 2013. p. 4171. [Google Scholar]
  • 24.Sled JG, Pike GB. Standing-wave and RF penetration artifacts caused by elliptic geometry: an electrodynamic analysis of MRI. IEEE Trans Med Imaging. 1998 Aug;17:653–662. doi: 10.1109/42.730409. [DOI] [PubMed] [Google Scholar]
  • 25.Shin J, Kim MJ, Lee J, Kim M, Choi N, Nam Y, Kim D-H. Coil combine for conductivity mapping of breast cancer. Proc 21st Annual Meeting of ISMRM; Salt Lake City, UT, USA. 2013. p. 4180. [Google Scholar]
  • 26.Vovk U, Pernus F, Likar B. A review of methods for correction of intensity inhomogeneity in MRI. IEEE Trans Med Imaging. 2007 Mar;26:405–421. doi: 10.1109/TMI.2006.891486. [DOI] [PubMed] [Google Scholar]
  • 27.Katscher U, Djamshidi K, Voigt T, Ivancevic M, Abe H, Newstead G, Keupp J. Estimation of breast tumor conductivity using parabolic phase fitting. Proc 20th Annual Meeting of ISMRM; Melbourne, Australia. 2012. p. 3482. [Google Scholar]
  • 28.Huang L, Schweser F, Herrmann K-H, Kramer M, Deistung A, Reichenbach JR. A Monte Carlo method for overcoming the edge artifacts in MRI-based electrical conductivity mapping. Proc 22nd Annual Meeting of ISMRM; Milan, Italy. p. 3190. [Google Scholar]
  • 29.Madio DP, Lowe IJ. Ultra-fast imaging using low flip angles and FIDs. Magn Reson Med. 1995 Oct;34:525–529. doi: 10.1002/mrm.1910340407. [DOI] [PubMed] [Google Scholar]
  • 30.Carl M, Chiang JT. Investigations of the origin of phase differences seen with ultrashort TE imaging of short T2 meniscal tissue. Magn Reson Med. 2012 Apr;67:991–1003. doi: 10.1002/mrm.23075. [DOI] [PubMed] [Google Scholar]
  • 31.Wong ST, Roos MS. A strategy for sampling on a sphere applied to 3D selective RF pulse design. Magn Reson Med. 1994 Dec;32:778–784. doi: 10.1002/mrm.1910320614. [DOI] [PubMed] [Google Scholar]
  • 32.Bulumulla SB, Lee SK, Yeo DT. Conductivity and permittivity imaging at 3.0T. Concepts Magn Reson Part B Magn Reson Eng. 2012 Feb 1;41B:13–21. doi: 10.1002/cmr.b.21204. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Setsompop K, Wald LI, AE Reduced-Voltage RF Shimming for Adiabatic Pulse Design in Parallel Transmission. Proc 15th Annual Meeting of ISMRM; Berlin, Germany. 2007. p. 1687. [Google Scholar]
  • 34.Zhang X, Zhu S, He B. Imaging electric properties of biological tissues by RF field mapping in MRI. IEEE Trans Med Imaging. 2010 Feb;29:474–481. doi: 10.1109/TMI.2009.2036843. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissues. Phys Med Biol. 1996 Nov;41:2271–2293. doi: 10.1088/0031-9155/41/11/003. [DOI] [PubMed] [Google Scholar]
  • 36.Seo JK, Kim MO, Lee J, Choi N, Woo EJ, Kim HJ, Kwon OI, Kim DH. Error analysis of nonconstant admittivity for MR-based electric property imaging. IEEE Trans Med Imaging. 2012 Feb;31:430–437. doi: 10.1109/TMI.2011.2171000. [DOI] [PubMed] [Google Scholar]
  • 37.Pohmann R, Scheffler K. A theoretical and experimental comparison of different techniques for B(1) mapping at very high fields. NMR Biomed. 2013 Mar;26:265–275. doi: 10.1002/nbm.2844. [DOI] [PubMed] [Google Scholar]
  • 38.Morrell GR, Schabel MC. An analysis of the accuracy of magnetic resonance flip angle measurement methods. Phys Med Biol. 2010;55:6157–6174. doi: 10.1088/0031-9155/55/20/008. [DOI] [PubMed] [Google Scholar]
  • 39.Bernstein MA, King KF, Zhou ZJ. Handbook of MRI pulse sequences. Amsterdam ; Boston: Academic Press; 2004. [Google Scholar]
  • 40.Voigt T, Homann H, Katscher U, Doessel O. Patient-individual local SAR determination: in vivo measurements and numerical validation. Magn Reson Med. 2012 Oct;68:1117–1126. doi: 10.1002/mrm.23322. [DOI] [PubMed] [Google Scholar]

RESOURCES