Abstract
The Charité® lumbar intervertebral disc replacement was subjected to wear testing in an electro-mechanical spine simulator. Sinusoidally varying compression (0.6–2 kN, frequency 2 Hz), rotation (±2°, frequency 1 Hz), flexion–extension (6° to −3°, frequency 1 Hz) and lateral bending (±2°, frequency 1 Hz) were applied out of phase to specimens immersed in diluted calf serum at 37 °C. The mass of the ultra-high-molecular weight polyethylene component of the device was measured at intervals of 0.5, 1, 2, 3, 4 and 5 million cycles; its volume was also measured by micro-computed tomography. Total mass and volume losses were 60.3 ± 4.6 mg (mean ± standard deviation) and 64.6 ± 6.0 mm3. Corresponding wear rates were 12.0 ± 1.4 mg per million cycles and 12.8 ± 1.2 mm3 per million cycles; the rate of loss of volume corresponds to a mass loss of 11.9 ± 1.1 mg per million cycles, that is, the two sets of measurements of wear agree closely. Wear rates also agree closely with measurements made in another laboratory using the same protocol but using a conventional mechanical spine simulator.
Keywords: Intervertebral disc replacement, spine simulators, ultra-high-molecular weight polyethylene, wear rates
Introduction
This technical note presents results on the wear of the Charité® intervertebral disc replacement (DePuy Spine Inc., Raynham, MA, USA) obtained using an electro-mechanical simulator. Replacement of an intervertebral disc in the lumbar spine is one possible surgical treatment for patients with chronic low back pain.1 The Charité device consists of two metal plates, fixed to bone above and below the joint, with a socket in each; an ultra-high-molecular weight polyethylene (UHMWPE) spacer separates the plates and protrudes into the sockets, forming a ball-and-socket joint.2 There are several published studies of wear of the UHMWPE spacer in this device, obtained using conventional mechanical simulators;3,4 later studies have focussed on regional wear of the spacer.4,5 The Bose SDWS-1 Spine Simulator (Bose Corporation, Minnesota, MN, USA) has 6 degrees of freedom and is driven by electro-mechanical motors. It has been used to investigate friction6–8 and wear9 in possible candidates for improved disc replacement devices; it has also been used to determine wear in the NuNec® device intended to replace intervertebral discs in the cervical spine10 and the flexural properties of an elastomeric device that functions as a flexible coupling rather than as a ball-and-socket device.11
The main purpose of this technical note is to determine whether results obtained by the electro-mechanical simulator are comparable to those using a conventional mechanical simulator. It also provides further information on the wear of the Charité device. Comparing the results from the two types of simulator is important for two reasons. First, the electro-mechanical simulator has been used in fundamental studies of friction that could not be performed with a conventional wear simulator.6–8 Second, it has been used to measure the behaviour of devices that have not been investigated by conventional mechanical simulators or testing machines. Investigation of the wear of a device that has been investigated using a conventional simulator provides a method for validating the performance of the new electro-mechanical simulator. In a study of dental restorative materials, significantly different results were obtained using different simulators.12 There is no good reason for preferring one type of simulator over the other; however, mechanical simulators have been used in the earlier work on other artificial disc replacements13–16 and have a long history of successful application, especially in determining the wear of artificial hip replacements.17 However, unlike the Bose SDWS-1 Spine Simulator, most conventional simulators test several devices at once that can lead to artefacts.18
Materials and methods
Specimens
Four samples of the Charité device were investigated, which is a similar number used in some previous wear tests.19,20 They were the smallest size available (Size 1) with the thinnest UHMWPE spacer (7.5 mm) and parallel plates (0°). The metal marker rings were removed from the spacer for comparability with published tests.4,5,21 The ring detects the location of the implanted UHMWPE core. It does not affect the device’s mechanical properties. The devices were not washed before testing since they had been supplied sterile for implantation, packed in double-sealed plastic packages. However, an air spray (RS Components Ltd, Northants, UK) was used to remove any dust that could have settled after opening the packs.
Mechanical testing
The metal plates of the device were attached to the testing machine by the fixture, shown in Figure 1, manufactured from grade 316 stainless steel. This fixture consists of two lids and two main bodies. Each plate was inserted snugly into a recess in a body (clearance ± 0.1 mm) and secured in place by a lid, as shown in Figure 1. The bodies were attached to the base and actuator of the simulator with the plates of the device horizontal. UHMWPE spacers were soaked in distilled water at 37 °C for 2 weeks before testing to stabilise fluid uptake.22
Figure 1.

Attachment of a plate of the Charité® device to the spine simulator: (a) the plate (left-hand side) of the device, the main body (centre) of the fixture that is attached to the base or actuator and the lid (right-hand side); (b) the plate fitted into the recess of the body and (c) the plate secured in the recess by the lid.
Tests were performed using a Bose SDWS-1 Spine Simulator (see above); a Bose ElectroForce® 3330 Series II Testing Machine was used as a ‘load soak’ control station. Wear tests were performed in accordance with BS ISO 18192-1,23 except when stated otherwise. Sinusoidally varying compression (0.6–2 kN, frequency 2 Hz), rotation (±2° about the vertical axis, frequency 1 Hz), flexion–extension (corresponding to bending the spine forwards and backwards, +6° to −3°, frequency 1 Hz) and lateral bending (corresponding to sideways bending of the spine, ±2°, frequency 1 Hz) were applied out-of-phased as described by the standard.
During testing, the devices were immersed in diluted calf serum (to a protein concentration of 15 ± 2 g/L) with distilled water; sodium azide (0.3 g/L) was added to reduce bacterial contamination at a temperature of 37 °C. Three devices were tested for 5 million cycles; the fourth was a control sample that was subjected to load soak conditions of sinusoidal compression (0.6–2 kN, frequency 2 Hz). The purpose of this control was to monitor liquid uptake or loss during the test. Specimens were cleaned and measured at intervals of 0.5, 1, 2, 3, 4 and 5 million cycles. Specimens were cleaned as per BS ISO 14242-2:2000.24 and the Standard Operating Protocol for Spine Wear Simulator Studies (SOP01.6) of the Institute of Medical and Biological Engineering, University of Leeds.
Measuring wear
The mass of each spacer was measured six times with a precision of 0.2 mg using a standard laboratory balance (Ohaus GA200D; Ohaus Scales and Balances, Thetford, Norfolk, UK). The volume of each spacer was measured by micro-computed tomography using a SkyScan-1172 high-resolution system (SkyScan, Kontich, Belgium) in the School of Dentistry, University of Birmingham.
Results
Figure 2(a) and (b) shows that there was a linear mass loss for the three tested samples during the 5 million test cycles. The total mass lost was 60.3 ± 4.6 mg (mean ± standard deviation) and the wear rate was 12.0 ± 1.4 mg per million cycles. The control sample appeared to show a total mass increase of 0.5 mg, that is, a mass increase of 0.1 mg every million cycles which was the longest period between measurements. Since measurements of mass increase were less than the precision of the balance (0.2 mg), they were considered to be negligible.
Figure 2.
Wear of the UHMWPE spacer of the Charité® device represented as (a) mass loss for each sample, (b) average mass loss, (c) volume loss for each sample and (d) average volume loss, as a function of the number of cycles. In both average cases, a regression line is fitted through the points (R2 = 0.99, p = 0.00) and error bars represent standard deviations. For the individual sample plots, regression lines are fitted through the points.
Results for the load soak control (Sample 1) are not shown; the behaviour of this sample is described in section ‘Results’.
Figure 2(c) and (d) shows that there was a linear volume loss for the three tested samples during the 5 million test cycles. The total volume loss was 64.6 ± 6.0 mm3 and the wear rate was 12.8 ± 1.2 mm3 per million cycles. Multiplication by the density of UHMWPE (0.931 g/cm3)21 enables the volume loss and wear rate to be converted into mass loss and wear rate; the results are 60.1 ± 5.6 and 11.9 ± 1.1 mg per million cycles, respectively. The Bland–Altman statistical test25 shows no significant difference (p < 0.05) between the results of the two techniques.
Discussion
The linear wear shown in Figure 2 is consistent with the published results on the Charité device.4,5 These published results were obtained with a mechanical spine simulator. Another study of the Charité device showed a period of running-in followed by a period of steady, slower wear.3 However, this study was performed using a hip simulator under different loading and angular displacement conditions as described in an earlier ASTM F2423-0526 standard. It is possible that the difference can be attributed to these different conditions, emphasising the need to compare the results obtained following the same standards when comparing the performance of different devices. An initial running-in period has been observed in intervertebral disc replacements with metal-on-metal articulation8,14,16 and in metal-on-metal hip replacement devices27,28 where the initial phase has been attributed to wear of surface asperities. This initial wear, leading to reduced friction, has been called ‘self-polishing’.28,29 Previous studies of wear of intervertebral disc replacement devices, in which UHMWPE articulates with metal, have reported evidence for abrasive and adhesive wear.5,30,31 Light scratches were just visible on both articulating surfaces during the tests reported here and could be consistent with abrasive wear. In this study, the UHMWPE surfaces developed a glossy appearance that is consistent with adhesive wear.
Reported wear rates for the Charité device were 13.1 ± 1.14 and 12.9 ± 2.5 mm3 per million cycles.5 These results are very close to those obtained here, using an electro-mechanical simulator, of 12.8 ± 1.2 mm3 per million cycles. Furthermore, the wear rate of the UHMWPE component of the Pro-Disc® device has been reported to be 11.6 ± 1.2 mg per million cycles,16 that is, 12.5 ± 1.3 mm3 per million cycles, which is very close to the UHMWPE wear rate reported here (12.0 ± 1.4 mg per million cycles).
The tests in this study were performed to 5 million cycles. Although BS ISO 18192-123 suggests testing to 10 million cycles, it has been shown that the wear rate of total hip replacements reaches steady state after the first few million cycles16,32 and the same was found for this study. Testing to 10 million cycles would not have changed the wear rates under steady-state conditions.
Acknowledgments
We thank Professor Richard Hall for advice and discussion and Dr Michelle Holder for making the micro-computed tomography measurements.
Footnotes
Declaration of conflicting interests: The authors declare that there is no conflict of interest.
Funding: This study was supported by the Engineering and Physical Sciences Research Council (grant no. EP/F14562/1). The ARG-2 rheometer (TA Instruments) used in this research was obtained through Birmingham Science City: Innovative Uses for Advanced Materials in the Modern World (West Midlands Centre for Advanced Materials Project 2), with support from Advantage West Midlands (AWM) and part funded by the European Regional Development Fund (ERDF).
References
- 1. Hughes SPF, Freemont AJ, Hukins DWL, et al. The pathogenesis of the intervertebral disc degeneration and new and emerging therapies in managing low back pain. J Bone Joint Surg Br 2012; 94: 1298–1304. [DOI] [PubMed] [Google Scholar]
- 2. Kurtz SM. Total disc arthroplasty (ed Kurtz SM, Edidin AA.). Amsterdam: Elsevier, 2006, pp.313–351. [Google Scholar]
- 3. Serhan HA, Dooris AP, Parsons ML, et al. In vitro wear assessment of the Charité artificial disc according to ASTM recommendations. Spine 2006; 31: 1900–1910. [DOI] [PubMed] [Google Scholar]
- 4. Vicars R, Prokopovich P, Brown TD, et al. The effect of anterior-posterior shear on the wear of Charité total disc replacement. Spine 2012; 37: E528–E534. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Prokopovich P, Perni S, Fisher J, et al. Spatial variation of wear on Charité lumbar discs. Acta Biomater 2011; 7: 3914–3926. [DOI] [PubMed] [Google Scholar]
- 6. Moghadas P, Mahomed A, Hukins DWL, et al. Friction in metal-on-metal total disc arthroplasty. J Biomech 2012; 45: 504–509. [DOI] [PubMed] [Google Scholar]
- 7. Moghadas P, Shepherd DET, Hukins DWL, et al. Polymer-on-metal or metal-on-polymer total disc arthroplasty: does it make a difference? Spine 2012; 37: 1834–1838. [DOI] [PubMed] [Google Scholar]
- 8. Moghadas P, Mahomed A, Hukins DWL, et al. Effect of lubricants on friction in laboratory tests of total disc arthroplasty. Proc IMechE, Part H: J Engineering in Medicine 2013; 227: 988–993. [DOI] [PubMed] [Google Scholar]
- 9. Moghadas P, Mahomed A, Hukins DWL, et al. Wear in metal-on-metal total disc arthroplasty. Proc IMechE, Part H: J Engineering in Medicine 2013; 227: 356–361. [DOI] [PubMed] [Google Scholar]
- 10. Xin H, Shepherd DET, Dearn KD. A tribological assessment of a PEEK based self-mating total cervical disc replacement. Wear 2013; 303: 473–479. [Google Scholar]
- 11. Mahomed A, Moghadas P, Shepherd DET, et al. Effect of axial load on the flexural properties of an elastomeric total disc replacement. Spine 2012; 37: E908–E912. [DOI] [PubMed] [Google Scholar]
- 12. Heintze SD, Zappini G, Roussin V. Wear of ten dental restorative materials in five wear simulators – results of a round robin test. Dent Mater 2005; 21: 304–317. [DOI] [PubMed] [Google Scholar]
- 13. Anderson PA, Rouleau JP, Toth JM, et al. A comparison of simulator-tested and retrieved cervical disc prostheses. J Neurosurg Spine 2004; 1: 202–210. [DOI] [PubMed] [Google Scholar]
- 14. Paré PE, Chan FW, Powell ML. Wear characterisation of the A-MAV™ anterior motion replacement using a spine wear simulator. Wear 2007; 263: 1055–1059. [Google Scholar]
- 15. Erkan S, Rivera Y, Wu C, et al. Biomechanical comparison of a two-level Maverick disc replacement with a hybrid one-level disc replacement and one-level anterior lumbar interbody fusion. Spine J 2009; 9: 830–835. [DOI] [PubMed] [Google Scholar]
- 16. Vicars R, Hyde PJ, Brown TD, et al. The effect of anterior-posterior shear load on the wear of ProDisc-L TDR. Eur Spine J 2010; 19: 1356–1362. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Affatato S, Spinelli M, Zavalloni M, et al. Tribology and total hip joint replacement: current concepts in mechanical simulation. Med Eng Phys 2008; 30: 1305–1317. [DOI] [PubMed] [Google Scholar]
- 18. Mejia LC, Brierly TJ. A hip wear simulator for the evaluation of biomaterials in hip arthroplasty components. Biomed Mater Eng 1994; 4: 259–271. [PubMed] [Google Scholar]
- 19. Scholes SC, Green SM, Unsworth A. The wear of metal-on-metal total hip prostheses measured in a hip simulator. Proc IMechE, Part H: J Engineering in Medicine 2001; 215: 523–530. [DOI] [PubMed] [Google Scholar]
- 20. Tipper JL, Firkins PJ, Besong AA, et al. Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator. Wear 2001; 250: 120–128. [Google Scholar]
- 21. Vicars R, Fisher J, Hall RM. The accuracy and precision of a micro computer tomography volumetric technique for the analysis of in-vitro tested total disc replacements. Proc IMechE, Part H: J Engineering in Medicine 2009; 223: 383–388. [DOI] [PubMed] [Google Scholar]
- 22. Grupp TM, Meisel HJ, Cotton JA, et al. Alternative bearing materials for intervertebral disc arthroplasty. Biomaterials 2010; 31: 523–531. [DOI] [PubMed] [Google Scholar]
- 23. BS ISO 18192-1:2011: Implants for surgery. Wear of total intervertebral spinal disc prostheses. Loading and displacement parameters for wear testing and corresponding environmental conditions for test. London, British Standards Institution. [Google Scholar]
- 24. BS ISO 14242-2:2000: Implants for Surgery- Wear of Total Hip Joint Prostheses- Part 2: Methods of Measurement. London, British Standard Institution. [Google Scholar]
- 25. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 327: 307–310. [PubMed] [Google Scholar]
- 26. ASTM F2423-05. Standard guide for functional, kinematic and wear assessment of total disc prostheses. Pennsylvania, PA: American Society for Testing and Materials. [Google Scholar]
- 27. Vassilou K, Elfick APD, Scholes SC, et al. The effect of ‘running-in’ on the tribology and surface morphology of metal-on-metal Birmingham hip resurfacing device in simulator studies. Proc IMechE, Part H: J Engineering in Medicine 2006; 220: 269–277. [DOI] [PubMed] [Google Scholar]
- 28. Tipper JL, Firkins PJ, Ingham E, et al. Quantitative analysis of the wear and wear debris from low and high carbon content cobalt chrome alloys using in metal on metal total hip replacements. J Mater Sci Mater Med 1999; 10: 353–362. [DOI] [PubMed] [Google Scholar]
- 29. McKellop H, Park SH, Chiesa R, et al. In vivo wear of three types of metal on metal hip prostheses during two decades of use. Clin Orthop Relat Res 1996; 329: S128–S140. [DOI] [PubMed] [Google Scholar]
- 30. Rieker C, Konrad R, Schoun R. In vitro comparison of two hard-hard articulations for total hip replacements. Proc IMechE, Part H: J Engineering in Medicine 2001; 215: 153–160. [DOI] [PubMed] [Google Scholar]
- 31. Anderson PA, Kurtz SM, Toth JM. Explant analysis of total disc replacement. Semin Spine Surg 2006; 18: 109–116. [Google Scholar]
- 32. Kurtz SM, van Ooij A, Ross R, et al. Polyethylene wear and rim fracture in total disc arthroplasty. Spine J 2007; 7: 12–21. [DOI] [PubMed] [Google Scholar]
- 33. Nechtow W, Hinter M, Bushelow M, et al. IVD replacement mechanical performance depends strongly on input parameters. Transactions of the 52nd Annual Meeting of the Orthoapedic Research Society, Chicago, Illinois, USA, 19–22 March 2006. [Google Scholar]

