Redox Cycling Without Reference Electrode

Sahana Sarkar a, Klaus Mathwig a, Shuo Kang a, Ab F Nieuwenhuis a, Serge G Lemay a
PMCID: PMC4465345  NIHMSID: NIHMS632806  PMID: 25271709

Abstract

The reference electrode is a key component in electrochemical measurements, yet it remains a challenge to implement a reliable reference electrode in miniaturized electrochemical sensors. Here we explore experimentally and theoretically an alternative approach based on redox cycling which eliminates the reference electrode altogether. We show that shifts in the solution potential caused by the lack of reference can be understood quantitatively, and determine the requirements for accurate measurements in miniaturized systems in the absence of a reference electrode.

Introduction

Motivated by the wide applications of electrochemical sensors in the biomedical domain, there is continuously rising interest in miniaturised assays suitable for integration. Electrochemical sensors are particularly well suited for cost effective realization and integration with fluidics due to their inherent compatibility with microfabrication111. Irrespective of the particular measurement method employed (amperometry, potentiometry, impedance spectroscopy, etc.), the reference electrode is a key component of electrochemical sensors. In miniaturized systems, however, the volume-to-surface area ratio of reference electrodes decreases, thereby reducing their lifetime and stability1216. For example, due to downscaling, dissolution of the electrode material, leakage of the reference solution and contamination at the junction of salt bridges become difficult to avoid, rendering many sensors short-lived or unreliable. Realizing lab-on-a-chip devices without the use of conventional reference electrodes12, 1417 could therefore significantly improve the performance and usability of miniaturised electrochemical sensors.

One potential route for bypassing the reference electrode is offered by systems based on redox cycling, in which a current is generated by successively reducing and oxidizing an analyte at two independently biased electrodes. While the primary purpose of redox cycling is typically to increase the magnitude of the electrochemical current or to improve selectivity1820, it also provides a pathway for electrochemically generated currents to flow through the analyte solution without the need for a reference or auxiliary electrode. In this work we explore the underlying principles of operation of redox cycling systems without a reference electrode. As a prototypical system we focus on fluidic nanogap sensors, sketched in Fig. 1a, consisting of two closely spaced parallel electrodes separated by a few tens of nanometers and forming the floor and roof of a fluid-filled nanochannel. Electrochemically active molecules undergo successive redox cycling by diffusing back and forth between these electrodes, generating a current that is proportional to the analyte concentration. The cross sectional diagram and top views of such a device are shown in Fig.1b and 1c, respectively. An example application of a variant of this device in which operation without a reference electrode would be necessary is a sealable chamber allowing locally generated analyte molecules (e.g., by a catalytic reaction) to be collected and detected in a fL-level volume (Fig. 1d). Another example is the incorporation of a nanogap directly in (or, more practically, in a side channel of) a capillary flow detection system (Fig. 1e). Other configurations where the same general principles apply include interdigitated arrays19, 2124 and ring-disk electrodes.25, 26

Figure 1.

Figure 1

a) Nanogap sensor in the conventional 3-electrode configuration that includes a reference electrode. Redox molecules undergo successive oxidation and reduction reactions at the two planar electrodes while diffusing within the nanochannel. An ionic current also flows to the reference electrode. b) Schematic diagram of the cross section of the device prior to etching (not drawn to scale). La and z represent the active (or overlapping) length of the top and bottom electrodes and the height of the nanogap, respectively. c) Optical microscopic image of the top view of the device before the sacrificial layer is etched. d) Nanogap sensor with an integrated microfluidic valve (PDMS). In the closed state the reference electrode (not shown) gets disconnected, as represented by the dashed lines. e) Capillary based fluid flow system consisting of two electrodes imbedded in a flow through (nano) channel.

We emphasize that this work does not in any way diminish the relevance of previous efforts at realizing miniaturized reference electrodes, but instead provides an alternative, complementary approach.

Materials and Methods

A nanogap sensor with two parallel planar electrodes, as shown in Fig. 1a–c, was used in our experiments. The method for micro fabricating the devices was similar to that reported previously27. In brief, the device consisted of a sacrificial chromium layer (thickness z = 60±5 nm) deposited between two platinum electrodes. The areas of the top and bottom electrodes were 60 µm2 and 102 µm2, respectively, whereas the overlapping area between the two electrodes where redox cycling took place was 30 µm2. Prior to the measurements, chromium etchant (BASF, Chromium Etch Selectipur) was filled through the inlets in order to etch the sacrificial layer and create the nanochannel. The electrodes were cleaned before conducting the measurements by cycling their potential in 5 mM H2SO4. A polydimethylsiloxane (PDMS) reservoir filled with analyte solution was positioned above the device and a standard Ag/AgCl electrode (BASi, MF 2079, RE-5B) inserted in this reservoir was used as reference electrode. The reference electrode was thus located outside the nanochannel in the reservoir. No auxiliary electrode was necessary as the current through the reference electrode remained well below 1 nA, as shown in detail below (Fig. 2). All solutions were prepared in 18 MΩ cm Milli-Q water (Millipore). Ferrocenedimethanol (Fc(MeOH)2, Sigma-Aldrich, cat. no- 372625, diffusion coefficient D = 6.7 × 10−10 m2/s) was chosen as a simple prototypical redox species and was prepared as a 1 mM solution reduced Fc(MeOH)2 with 0.1 M KCl (Sigma-Aldrich, cat. no- P3911) added as supporting electrolyte.

Figure 2.

Figure 2

Cyclic voltammograms for 1 mM Fc(MeOH)2 in 0.1 M KCl at a scan rate of 5 mV/s. The black, red and green lines represent the currents measured at the top, bottom and reference electrodes, respectively. a) The bottom electrode (Eb) was swept while the top electrode was left floating; the rest potential was Vrp = 0.21 V. b) The top electrode (Et) was swept while the bottom electrode was left floating; again Vrp = 0.21 V. c) Eb was swept while Et was biased at 0 V giving rise to a redox cycling current of 30 nA. d) The current through the reference electrode during redox cycling, Iref, was essentially identical to that observed in cases (a) and (b). e) Et was swept while Eb was biased at 0 V. f) Iref measured simultaneously with the voltammogram in panel (e). The current is nearly constant, in sharp contrast with panel (d).

In order to investigate the role of the reference electrode in redox cycling measurements, experiments were conducted with two complementary electrode configurations. In the first configuration the Ag/AgCl electrode was held at a fixed bias, thus serving as a proper reference electrode, while the current flowing through it was monitored using a transimpedance amplifier (Femto, DDPCA-300). In the second scheme, there was no true reference electrode present. The standard Ag/AgCl reference electrode instead served as a non-invasive probe to the solution potential, Esol. A Keithley 617 electrometer was used for these potentiometric measurements. The chassis ground and common (COM) terminals of all the instruments were connected together throughout the measurements; we refer to this potential as the circuit reference potential (or ground for short). Scan rates were kept sufficiently low such that steady-state mass transport conditions were achieved.

Experimental Results

We first concentrate on measurements conducted with an active reference electrode. Fig. 2a shows the cyclic voltammogram obtained at the bottom electrode when its potential, Eb, was varied from 0 to 0.5 V while the potential of the top electrode, Et, was disconnected (black curve). A sigmoidal cyclic voltammogram was observed with a wave height (difference between the plateau currents at Et = 0 V and Et = 0.5 V) of ilim = 305 pA. The cyclic voltammogram simultaneously obtained at the reference electrode is also shown (green curve); this current has the same magnitude but an opposite sign to that measured at the bottom electrode, demonstrating that all current from the working electrode flows to the reference electrode and therefore that there are no significant parasitic current leakage pathways in the system. The voltammogram (obtained at the bottom electrode) is shifted downward by 52 pA compared to that expected for the original solution of Fc(MeOH)2 in the reduced form. This indicates that this solution became partly oxidized after exposure to the reservoir, consistent with previous observations28 and as verified explicitly by re-extracting the solution and analysing it ex-situ (see Supporting Information).

Fig. 2b shows that sweeping Et while keeping Eb floating gave rise to an essentially identical cyclic voltammogram. While this may appear counterintuitive since the top electrode is located further from the access holes, and thus one might expect smaller currents, this effect can be understood by noting that the bottom electrode must necessarily adjust its potential to that of the top electrode because of the strong coupling introduced by redox cycling, as discussed earlier28. Similarly, measurements where both Et and Eb were swept simultaneously (data not shown) yielded once again the same voltammogram. In all the above voltammograms, the potential at which the obtained current crosses the zero-current axis – that is, the rest potential, Vrp – was found to be 0.21 V.

Fig. 2c illustrates a typical voltammogram obtained during redox cycling with a reference electrode and Eb swept between 0 and 0.5 V while Et was kept biased at 0 V. An oxidising current at the bottom electrode giving ilim of 30 nA was observed. An equal and opposite current was observed at the top (reducing) electrode, as expected for redox cycling. The amplification factor for redox cycling can be defined as the ratio of ilim at the electrode during redox cycling to that of the current at the reference electrode (Fig.2a and 2c), yielding an amplification factor of ~98. It is however crucial to note that a finite current still flows through the reference electrode during redox cycling, as explicitly shown in Fig. 2d. This current originates from molecules from the bulk that diffused through the access holes into the channel, reacted at the bottom electrode and returned once again to the bulk reservoir. Correspondingly, the values of Iref observed in Fig. 2d are essentially equal to those obtained in the case of Fig 2a. Fig. 2e shows that when Et was varied while Eb was biased at 0 V, oxidising and reducing currents were now observed at the top and bottom electrodes, respectively, as expected since the role of the electrodes was reversed. In this case, however, the current through the reference electrode (Fig. 2f) was roughly constant (apart from slight drift and hysteresis) with a value close to the Eb = 0 V values of 50 pA observed in Fig.2a and 2c. This reflects the fact that mass transport to the reference electrode is much more efficient from the bottom than from the top electrode due to the channel geometry: the bottom electrode is directly exposed to the access holes, whereas molecules leaving the top electrode interact with near-certainty with the bottom electrode on their way to the bulk. As a result, the potential of the top electrode hardly influences Iref.

We now turn to measurements done in the second configuration, in which no active reference electrode was used and the potential of the solution was allowed to float. Fig. 3a shows a cyclic voltammogram wherein Et was biased at 0 V while Eb was cycled between −0.5 and 0.5 V with respect to the circuit reference potential. These curves are qualitatively different from those observed when the reference was connected (Fig. 2c), with both electrodes exhibiting reducing and oxidizing currents depending on the polarity of Eb. Thereafter, we introduced an Ag/AgCl electrode to monitor the potential of the floating bulk electrolyte with respect to the circuit reference potential, as shown in Fig. 3b

Figure 3.

Figure 3

Cyclic voltammograms for 1 mM Fc(MeOH)2 in 0.1 M KCl at a scan rate of 5 mV/s in the absence of the reference electrode. a) Currents obtained at the top (black) and bottom (red) electrodes when the bottom electrode (Eb) was swept while the top electrode (Et) was biased at 0 V with respect to the circuit reference potential. b) Schematic diagram of the setup configuration; the Ag/AgCl electrode now serves as a probe of the solution potential, Esol. All potentials are referred to the external circuit potential, here labelled as Ground. c) Eb was varied (black) while Et was grounded (red) and Esol was measured (orange). Esol closely follows the potential of the sweeping electrode, Eb. d) Et was varied while Eb was grounded. Esol experiences only minor variations compared to the sweeping potential, Et.

Fig. 3c shows the corresponding measured potential of the solution, Esol (orange curve) while cycling Eb (black curve) and keeping Et equal to 0 V (red curve). Esol follows the applied potential quite closely. Similarly, Fig. 3d shows the complementary case where Et was cycled. In this case, sweeping the potential had a negligible effect on the bulk solution. These results illustrate qualitatively that the more effective mass transport from the bottom electrode to the bulk solution observed in Fig. 2 directly translate into a stronger influence of the bottom electrode on the solution potential.

Analysis

Using the measured potential of the solution, Esol, and the known applied potentials, Et and Eb (all referred to the external circuit potential), we can reconstruct the potential differences across the solid-liquid interfaces at both electrodes, Et,sol and Eb,sol. These are the preferred variables for describing electrochemical processes since it is these potential differences that drive electrochemical reactions. In terms of the directly measurable quantities defined above, these potentials are simply given by

Et/b,sol=Et/bEsol (1)

The potentials of Fig.3c and 3d re-expressed in terms of Et,sol and Eb,sol are shown in Fig.4a and 4b, respectively. In Fig. 4a, it is observed that although Eb was swept over a broad range of potentials, Eb,sol remained relatively constant near the Vrp at 0.21 V. On the contrary, Et,sol changed from 0.6 V to −0.27 V, thereby showing that the top electrode was effectively being swept during the measurement while the bottom electrode held a nearly constant potential with respect to the solution. Fig. 4b exhibits exactly the same behaviour except that the potential curves are reflected through the Et/b = 0 V axis. While this may appear to contrast with the large differences observed between Fig.3c and 3d, this mirror symmetry must necessarily be present since, in the case where the solution potential is allowed to float, only the potential difference, EtEb, can be relevant in establishing the interfacial potentials. Both electrode configurations thus lead to the same situation wherein the bottom electrode remains at a nearly constant potential while the top electrode experiences an effective potential sweep that is applied as a potential difference between the two electrodes.

Figure 4.

Figure 4

a) The potentials at the electrodes with respect to the solution, Eb,sol and Et,sol, while sweeping Eb. Eb, sol remained relatively constant near a value of Vrp = 0.21 V while Et,sol varied from 0.6 V to −0.27V. b) Same as a) except that ET is swept. c) The expected current was calculated using Eq. (2) and the potentials of panel a). The calculated and measured currents (represented as black and green lines, respectively) are in excellent agreement. d) Same as c) for the case where Et is swept (panel b)). The calculated and measured currents are represented as red and green lines, respectively.

In order to further test the validity of the values of Et/b,sol extracted above, we attempted to reconstitute the cyclic voltammograms of Fig. 3a using the potentials determined in Fig.4a and 4b. The shape of redox-cycling voltammograms in nanogap devices has previously been shown to be well described by the Butler-Volmer formalism28,

Itrc=Ibrc=ilim11+efηt11+efηb1+Dk0z(eαfηt1+efηt+eαfηb1+efηb) (2)

Here z is the height of the nanochannel, f = F/RT, F is the Faraday constant, R is the gas constant, T is the absolute temperature and ηt,b are the overpotentials applied to the top and bottom electrodes, respectively. That is, ηt/b = Et/b,sol − Ef, where Ef is the formal potential of the reaction. Ef was measured directly using the same solution with a platinum UME (BASi, MF 2005) and the same reference electrode as used during the other measurements and was found to be 0.26 V. ilim obtained during the redox cycling (30 nA, Fig. 2c,e) was directly substituted in Eq. (2). The calculated value of ilim is marginally higher (~32 nA) and this difference can be attributed to the buckling of the electrodes due to stress (resulting from the passivation) that leads to an increase in the height of the channel29. The remaining unknown parameters, i.e., the transfer coefficient, α, and the heterogeneous rate constant, k0, were found to be 0.50 and 0.06 m/s, respectively, consistent with previous measurements28

Substituting the measured interfacial potentials from Fig.4a and 4b into Eq. (2) yields the calculated cyclic voltammograms (green curves) plotted in Fig.4c and 4d, respectively. These are in excellent agreement with the measured currents at the bottom (black curve) and top (red curve) electrodes obtained when Eb and Et were swept, respectively.

Discussion

We first briefly summarize the situation for the classical system consisting of cyclic voltammetry at a UME. Within the operating window of the solvent, the detailed shape of the voltammogram depends on the solution composition and generically includes both a target analyte signal as well as parasitic currents resulting from reactions involving dissolved oxygen, hydronium ions and/or contaminants. Under typical steady-state mass transport conditions the current varies monotonically with electrode potential and vanishes at the rest potential, Vrp, as discussed before (Fig. 2a,b). A floating working electrode starting at an arbitrary potential will reduce or oxidize molecules in solution and accumulate charge until its potential drifts to Vrp, and this potential can be measured with a high-impedance voltmeter (Fig. 3). This description was further extended by Xiong and White to the case of a cell comprising of two polarisable electrodes in the absence of a reference electrode30.

The same principle applies to the redox-cycling case: in the absence of a reference electrode, the solution potential is determined by the condition such that no net current can be injected into the solution at steady state,

It+Ib=0 (3)

Here It and Ib are the currents at the top and bottom electrodes, respectively. That is, the solution potential will drift to the value at which Eq. (3) is satisfied. In redox-cycling systems, the steady state current at each electrode can further be broken into two components: a redox-cycling current, It/brc, and an excess current, It/b0. The latter includes all other current components such as parasitic reactions, irreversible reactions and reactions occurring in regions of the electrodes not implicated in redox cycling. This can be written as

It/b=It/b0+It/brc (4)

As was seen in Fig. 2, the magnitude of It/brc is far larger than It/b0 and dominates the overall current levels during redox cycling. Noting that the redox-cycling current obtained at the two electrodes is, by definition, equal and opposite (Itrc=Ibrc), however, equation (1) and (2) reduce to the more specific condition

It0+Ib0=0 (5)

That is, in the absence of a reference electrode, the potential of the electrolyte solution is fully determined by the excess current, It/b0, even though it represents only a small fraction of the total currents flowing through the electrodes. This situation is reminiscent of positive-feedback scanning electrochemistry (SECM) measurements on a floating substrate3136, but here it is the solution itself that adjusts its potential with respect to the electrodes.

Conclusions

Redox cycling measurements without a well-defined reference electrode are a priori difficult to interpret because it is not known how a potential difference applied between the two working electrodes is divided between them. We have argued that determining this potential is further complicated by the fact that it is fully determined by “excess” background currents rather than by the (usually much larger) redox cycling current. On the other hand, we have shown that the geometry of the redox cycling device can be exploited to alleviate this difficulty to a significant extent. In situations where one of the redox cycling electrodes has a much more significant exposure to the sample solution, this electrode can readily pin the solution potential to a nearly constant value, allowing the potential of the second electrode with respect to the solution to be controlled. While we have concentrated here on nanogap electrodes embedded in nanochannels, the same principles can be applied to other asymmetric redox cycling geometries such as recessed ring-disk electrodes or even interdigitated electrodes if one of the electrodes is recessed with respect to the other.

Supplementary Material

ESI

Acknowledgements

The authors acknowledge E. D. Goluch for useful discussions and preliminary experiments. We also gratefully acknowledge financial support from The Netherlands Organization for Scientific Research (NWO) and the European Research Council (ERC). This publication was further made possible by Grant Number 1R01HG006882-01 from National Institutes of Health (NIH); its contents are solely the responsibility of the authors and do not necessarily represent the official views of NIH.

Footnotes

† Electronic Supplementary Information (ESI) available: micro fabrication methods and characterization of solution after exposure to device. See DOI: 10.1039/b000000x/

Notes and references

  • 1.Dutse SW, Yusof NA. Sensors-Basel. 2011;11:5754–5768. doi: 10.3390/s110605754. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Lee TMH, Carles MC, Hsing IM. Lab Chip. 2003;3:100–105. doi: 10.1039/b300799e. [DOI] [PubMed] [Google Scholar]
  • 3.Webster TA, Goluch ED. Lab Chip. 2012;12:5195–5201. doi: 10.1039/c2lc40650k. [DOI] [PubMed] [Google Scholar]
  • 4.Fragoso A, Latta D, Laboria N, von Germar F, Hansen-Hagge TE, Kemmner W, Gartner C, Klemm R, Drese KS, O'Sullivan CK. Lab Chip. 2011;11:625–631. doi: 10.1039/c0lc00398k. [DOI] [PubMed] [Google Scholar]
  • 5.Goral VN, Zaytseva NV, Baeumner AJ. Lab Chip. 2006;6:414–421. doi: 10.1039/b513239h. [DOI] [PubMed] [Google Scholar]
  • 6.Koudelka-Hep M, van der Wal PD. Electrochim Acta. 2000;45:2437–2441. [Google Scholar]
  • 7.Kraly JR, Holcomb RE, Guan Q, Henry CS. Anal Chim Acta. 2009;653:23–35. doi: 10.1016/j.aca.2009.08.037. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Lin YH, Matson DW, Bennett WD, Thrall KD, Timchalk C. Microreaction Technology: Industrial Prospects. 2000:588–596. [Google Scholar]
  • 9.Wang J. Electroanalysis. 2005;17:1133–1140. [Google Scholar]
  • 10.Zou ZW, Kai JH, Rust MJ, Han J, Ahn CH. Sensors and Actuators a-Physical. 2007;136:518–526. [Google Scholar]
  • 11.Grayson ACR, Shawgo RS, Li YW, Cima MJ. Adv Drug Deliver Rev. 2004;56:173–184. doi: 10.1016/j.addr.2003.07.012. [DOI] [PubMed] [Google Scholar]
  • 12.Shinwari MW, Zhitomirsky D, Deen IA, Selvaganapathy PR, Deen MJ, Landheer D. Sensors-Basel. 2010;10:1679–1715. doi: 10.3390/s100301679. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Suzuki H. Electroanalysis. 2000;12:703–715. [Google Scholar]
  • 14.Schwarz MA, Galliker B, Fluri K, Kappes T, Hauser PC. Analyst. 2001;126:147–151. doi: 10.1039/b007383k. [DOI] [PubMed] [Google Scholar]
  • 15.Simonis A, Luth H, Wang J, Schoning MJ. Sensor Actuat B-Chem. 2004;103:429–435. [Google Scholar]
  • 16.Franklin RK, Johnson MD, Scott KA, Shim JH, Nam H, Kipke DR, Brown RB. Ieee Sensor. 2005:1400–1403. [Google Scholar]
  • 17.Suzuki H, Shiroishi H, Sasaki S, Karube I. Analytical chemistry. 1999;71:5069–5075. doi: 10.1021/ac9811468. [DOI] [PubMed] [Google Scholar]
  • 18.Wolfrum B, Zevenbergen M, Lemay S. Analytical chemistry. 2008;80:972–977. doi: 10.1021/ac7016647. [DOI] [PubMed] [Google Scholar]
  • 19.Goluch ED, Wolfrum B, Singh PS, Zevenbergen MAG, Lemay SG. Analytical and Bioanalytical Chemistry. 2009;394:447–456. doi: 10.1007/s00216-008-2575-x. [DOI] [PubMed] [Google Scholar]
  • 20.Straver MG, Odijk M, Olthuis W, van den Berg A. Lab on a chip. 2012;12:1548–1553. doi: 10.1039/c2lc21233a. [DOI] [PubMed] [Google Scholar]
  • 21.Dam VAT, Olthuis W, van den Berg A. Analyst. 2007;132:365–370. doi: 10.1039/b616667a. [DOI] [PubMed] [Google Scholar]
  • 22.Heo JI, Shim DS, Teixidor GT, Oh S, Madou MJ, Shin H. J Electrochem Soc. 2011;158:J76–J80. [Google Scholar]
  • 23.Niwa O, Xu Y, Halsall HB, Heineman WR. Analytical chemistry. 1993;65:1559–1563. doi: 10.1021/ac00059a013. [DOI] [PubMed] [Google Scholar]
  • 24.Sanderson DG, Anderson LB. Analytical chemistry. 1985;57:2388–2393. [Google Scholar]
  • 25.Ma CX, Contento NM, Gibson LR, Bohn PW. Acs Nano. 2013;7:5483–5490. doi: 10.1021/nn401542x. [DOI] [PubMed] [Google Scholar]
  • 26.Zhao G, Giolando DM, Kirchhoff JR. Analytical chemistry. 1995;67:1491–1495. doi: 10.1021/ac00104a031. [DOI] [PubMed] [Google Scholar]
  • 27.Kang S, Mathwig K, Lemay SG. Lab Chip. 2012;12:1262–1267. doi: 10.1039/c2lc21104a. [DOI] [PubMed] [Google Scholar]
  • 28.Zevenbergen MAG, Wolfrum BL, Goluch ED, Singh PS, Lemay SG. Journal of the American Chemical Society. 2009;131:11471–11477. doi: 10.1021/ja902331u. [DOI] [PubMed] [Google Scholar]
  • 29.Zevenbergen MAG, Krapf D, Zuiddam MR, Lemay SG. Nano letters. 2007;7:384–388. doi: 10.1021/nl062571g. [DOI] [PubMed] [Google Scholar]
  • 30.Xiong JW, White HS. Journal of electroanalytical chemistry. 2013;688:354–359. [Google Scholar]
  • 31.Wipf DO, Bard AJ. J Electrochem Soc. 1991;138:469–474. [Google Scholar]
  • 32.Macpherson JV, Slevin CJ, Unwin PR. J Chem Soc Faraday T. 1996;92:3799–3805. [Google Scholar]
  • 33.Selzer Y, Turyan I, Mandler D. J Phys Chem B. 1999;103:1509–1517. [Google Scholar]
  • 34.Zoski CG, Simjee N, Guenat O, Koudelka-Hep M. Anal Chem. 2004;76:62–72. [Google Scholar]
  • 35.Xiong H, Guo JD, Amemiya S. Anal Chem. 2007;79:2735–2744. doi: 10.1021/ac062089i. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Oleinick AI, Battistel D, Daniele S, Svir I, Amatore C. Anal Chem. 2011;83:4887–4893. doi: 10.1021/ac2006075. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

ESI

RESOURCES