Skip to main content
Biomedical Optics Express logoLink to Biomedical Optics Express
. 2015 Apr 14;6(5):1707–1725. doi: 10.1364/BOE.6.001707

In vivo photothermal optical coherence tomography for non-invasive imaging of endogenous absorption agents

Shuichi Makita 1, Yoshiaki Yasuno 1,*
PMCID: PMC4467701  PMID: 26137374

Abstract

In vivo photothermal optical coherence tomography (OCT) is demonstrated for cross-sectional imaging of endogenous absorption agents. In order to compromise the sensitivity, imaging speed, and sample motion immunity, a new photothermal detection scheme and phase processing method are developed. Phase-resolved swept-source OCT and fiber-pigtailed laser diode (providing excitation at 406 nm) are combined to construct a high-sensitivity photothermal OCT system. OCT probe and excitation beam coaxially illuminate and are focused on tissues. The photothermal excitation and detection procedure is designed to obtain high efficiency of photothermal effect measurement. The principle and method of depth-resolved cross-sectional imaging of absorption agents with photothermal OCT has been derived. The phase-resolved thermal expansion detection algorithm without motion artifact enables in vivo detection of photothermal effect. Phantom imaging with a blood phantom and in vivo human skin imaging are conducted. A phantom with guinea-pig blood as absorber has been scanned by the photothermal OCT system to prove the concept of cross-sectional absorption agent imaging. An in vivo human skin measurement is also performed with endogenous absorption agents.

OCIS codes: (110.4500) Optical coherence tomography, (170.6935) Tissue characterization, (350.5340) Photothermal effects, (120.5050) Phase measurement

1. Introduction

Micro-scale non-invasive cross-sectional imaging of endogenous agents in vivo may attain further level of investigation and diagnosis of diseases. Optical coherence tomography (OCT) enables non-invasive in vivo biopsy of biological samples [1] by imaging the micro-scale inner structure of tissue. In addition to obtaining structural information from the backscattering intensity, the measurement of optical absorption has been attempted; i.e., spectroscopic [2,3], pump-probe [4], and photothermal OCT [59]. Spectroscopic OCT classifies tissue according to the spectrum attenuation, which is the combination of absorption and scattering [2]. To extract the absorption of a target, an appropriate wavelength range should be adopted for the absorption peak of the target; e.g., visible light for chromophores such as hemoglobin [3]. Pump-probe OCT uses the transient absorption to contrast absorption materials according to change in absorption of a probe beam due to absorption of a modulated pump beam. Photothermal OCT contrasts absorbers through measurement of the optical path length (OPL) change due to the photothermal effect and has been applied to the in vivo measurement of oxygen saturation [10].

Photothermal OCT has advantages over other OCT methods in that there is no restriction on the selection of the wavelength of the probe; i.e., the probe beam of OCT and the excitation beam for the photothermal effect are totally decoupled. Hence, there is no need to alter the core part of OCT while photothermal OCT can be adapted for different absorbers by changing the excitation wavelength. This allows the use of excitation light that maximizes the photothermal effect and a probe of near-infrared light to optimize imaging penetration. However, in vivo cross-sectional imaging with endogenous agents has not been demonstrated.

Exogenous agents, such as metal nanoparticles [69,1116], carbon nanotubes [17], and biocompatible nanoparticles [18], can be designed to fulfill the requirements of applications, and the absorption coefficients of exogenous agents are usually larger than those of endogenous absorption agents. In addition, in vivo imaging employing photothermal OCT is severely affected by the bulk motion of samples. Namely, the OPL change due to the photothermal effect is measured by the phase-resolved processing of OCT and is disturbed by the bulk tissue motion. To obtain a cross-sectional depth-resolving absorption contrast, the accumulation of OPL change due to the photothermal effect should be resolved [5, 9, 19, 20]. These hurdles make the in vivo photothermal OCT imaging of endogenous agents challenging.

In this study, we demonstrate photothermal OCT for in vivo absorption contrast imaging. A theoretical model relating the photothermal effect and the measurement of the local OPL change by phase-resolved OCT is investigated. A new phase processing scheme that achieves depth-resolved absorption-agent imaging and eliminates bulk-motion artifacts is developed. Spatio-temporal modulation and demodulation of thermal excitation are used to extract and enhance photothermal signals. The signal with a tissue phantom is evaluated. The performance of the depth-resolving absorption-agent imaging is evaluated with a phantom incorporated guinea-pig blood. In vivo human skin imaging of endogenous absorption agents is then demonstrated.

2. Principle of photothermal OCT imaging

Photothermal OCT locates the distribution of absorption materials within tissue by measuring local OPL change induced by light absorption. To measure a tiny local OPL change, a phase-sensitive nanoscale OPL change measurement method that uses complex OCT signals, such like Doppler OCT processing, is used.

By describing the temperature distribution upon laser excitation, the measurement of local OPL change by phase-resolved OCT, and the relationship between the phase measurement of OCT and the temperature distribution, the concept of the localization of absorbers is developed.

2.1. Model of temperature change due to laser excitation

The three-dimensional temperature distribution due to the absorption of light is modeled by the heat conduction equation. By assume some specific conditions, the solutions in analytical forms are obtained. In the presented method (Section 3), we cannot assume that absorption coefficient is homogeneous [6,21], heat transfer due to laser excitation is steady periodic condition [7,22], or excitation laser beam size is significantly large [19]. Hence, here we describe from the basic heat conduction solution.

When the material is infinite and thermal properties are homogeneous, the temperature distribution can be expressed with a Green’s function as [23]

ΔT(r,t)=0texp[|rr|24κ(tt)][4πκ(tt)]3/2q˙(r,t)ρcVdt dx dy dz, (1)

where ΔT(r, t) is the distribution of the temperature change from the initial constant temperature: T(r, t) = T0 + ΔT(r, t), is the volumetric heat generation rate, and spatial coordinate r = (x, y, z). κ, ρ, and cV are the thermal diffusivity, density, and specific heat capacity, respectively.

Endogenous absorbers of a tissue become the heat sources by illuminating an excitation laser. Consider that an excitation laser profile on a sample surface is a spatial Gaussian distribution and the distribution of the excitation laser intensity at the sub-surface obeys the Lambert–Beer’s law, the rate of heat generation can be then described as [24]

q˙(r,t)=YH2P(t)πwE2e2(xxE)2+(yyE)2wE2μα(r)e0zμ(x,y,l)dl, (2)

where μα(r) and μ(r) are the three-dimensional distributions of the absorption and attenuation coefficients in the tissue, respectively. YH is the heat yield by a unit optical energy, z0 is the depth location of the sample surface, (xE, yE) is the illumination point of the excitation laser, wE is the radius of the excitation beam and P(t) is the temporal profile of the excitation laser power.

By substituting Eq. (2) into Eq. (1), the distribution of the temperature change is described as

ΔT(r,t)=YHρcV2πwE20tP(t)exp[|rr|24κ(tt)][4πκ(tt)]3/2e2(xxE)2+(yyE)2wE2×μα(r)ez0zμ(x,y,l)dldt dx dy dz. (3)

When a point absorber in the tissue is assumed, μα(r) = μαδ(|rra|). Then Eq. (3) is rewritten as

ΔTδ(r,t)=YHρcV2πwE2e2(xaxE)2+(yayE)2wE2μαez0zaμ(xa,ya,l)dl0tP(t)exp[|rra|24κ(tt)][4πκ(tt)]3/2dt, (4)

where ra = (xa, ya, za) is the location of the point absorber. Equations (3) and (4) suggest that peaks of temperature change are located at the absorbers in the region illuminated by the excitation laser; i.e., (xE, yE). The temporal integration term of Eq. (4) represents the temporal change of the temperature and thermal diffusion from the point absorber.

2.2. Absorption contrast obtained by phase-resolved photothermal OCT

In photothermal OCT, the temperature change is obtained from a phase-resolved OCT measurement. In detail, the temperature change induces a tissue displacement and a refractive index change. Then, they are measured as OPL change by using phase-resolved OCT. This subsection describes the principle of phase-resolved OCT and then formulates the relationship between the distribution of the temperature change and the displacement of tissue.

2.2.1. Local optical path length change measurement by OCT

The complex OCT signal detected by Fourier-domain OCT, Γ, is expressed as

Γ(x,y,z,t)=η|Eref||Es(x,y,z,t)|eiϕ(x,y,z,t), (5)

where Es(x, y, z, t) is the electric field of probe beam backscattered from the point (x, y, z) at time t, Eref is the electric field of the reference beam, ϕ is the phase difference between the backscattered and reference light, and η is the factor of the conversion from the light energy to signal.

A temporal phase shift Δtϕ(z, t; Δt) = ϕ(z, t + Δt) − ϕ(z, t) is proportional to a temporal OPL change as [19, 25]

Δtϕ(z,t;Δt)2kc=[z0(t+Δt)z(t+Δt)ns(l,t+Δt)dl+noz0(t+Δt)][z0(t)z(t)ns(l,t)dl+noz0(t)], (6)

where z0(t) is the axial position of the surface of the tissue at time t, Δt is the time separation for temporal phase shift calculation, ns and no are the refractive indexes of the tissue and the surrounding media, respectively, and kc is the central optical wavenumber of the light source in vacuum. Δt indicate the temporal variation in the quantity to its right. In this and the following equations, the transversal positions are omitted for simplicity, without losing generality.

To resolve the accumulated OPL change along depth, axial differentiation should be applied to resolve a local OPL change at a local tissue [5,19,20,26]. The axial differentiation of Eq. (6) with axial separation Δz becomes as

ΔzΔtϕ(z,t;Δz,Δt)2kc=za(t+Δt)zb(t+Δt)ns(l,t+Δt)dlza(t)zb(t)ns(l,t)dl, (7)

where za and zb are above and bottom surface positions of a local tissue region where Δz = zb(0) − za(0) and z=za(0)+zb(0)2. Because of axial phase differentiation, the surface movement effect [25] has been removed.

By assuming that the axial separation Δz is significantly larger than the temporal thickness change due to thermal expansion of the local region, then, zb(t) − za(t) ≈ Δz. Then, Eq. (7) is expanded as

ΔzΔtϕ(z,t;Δz,Δt)2kcΔz=ns¯(z,t)Δtw(zb,t;Δt)Δtw(za,t;Δt)Δz+ns¯(z,t+Δt)ns¯(z,t), (8)

where w is the distribution of the axial displacement of the tissue, ns¯(z,t)=1Δzzazbns¯(l,t)dl is the mean refractive index of the the local region.

2.2.2. Thermal effects and local optical path length change

Photothermal effect alter the OPL due to thermal expansion and refractive index change [19]. The relationship between the temperature distribution resulting from laser excitation and tissue displacement depends on the mechanical properties of a tissue. By assuming the tissue to be a linear elastic body, the strain–stress relationship (with thermal stress generated by the temperature distribution) can be expressed using Duhamel–Neumann relations [27] as εij = ∑k,l Sijklσkl + δijαiΔT, where Sijkl, εij, and σkl are entries of the compliance, strain, and stress tensors, respectively. ΔT is the temperature change, α is a linear expansion coefficient, and δij is the Kronecker delta. The subscripts i, j, k, l are integers and denote the orthogonal directions of Cartesian coordinates (1, 2, 3 are corresponding to x-, y-, and z-directions). In the case of an isotropic material,

εij=1+νEσij+δij(νEσkk+αΔT), (9)

where E and ν are Young’s modulus and Poisson’s ratio, respectively.

According to the definition of the axial strain εz=wz, where εz is an alias of ε33, the first term of the left part of Eq. (8) can be approximate measurement of the change of axial strain as

Δtw(zb,t;Δt)Δtw(za,t;Δt)ΔzΔtεz(z,t). (10)

The thermal expansion also causes the lateral thermal strain as shown in Eq. (9). This will cause lateral displacement of tissue and decorrelation phase noise could be occurred [28, 29]. However, it will cause only nanometer scale displacement; the effect of such small lateral displacement can be ignored.

According to Eq. (9), the axial strain εz = [σzν(σx + σy)]/E + αΔT, where σx = σ11 and σy = σ22. The first term of the left part of Eq. (8) can be replaced with axial strain change [Eq. (10)] and the refractive index change is considered due to temperature change as

ΔzΔtϕ(z,t;Δz,Δt)2kcΔz=[αns(z,t)+dnsdT]ΔtΔT(z,t;Δt)+ns(z,t)Δtσzν(Δtσx+Δtσy)E, (11)

where dnsdT is the thermal coefficient of the refractive index, Δtσx,y,z are the changes in stresses during the time lag of Δt at time t.

Because the temperature distribution represents the distribution of absorbing tissue [Eq. (3)], Eq. (11) represents the absorber’s distribution when stresses after Δt are close to the initial condition (Δσx = Δσy = Δσz = 0). Note that the tissue properties except optical properties are assumed to be homogeneous for simplicity. The real biological tissues are perhaps mechanically inhomogeneous and not infinite. Then, stress changes Δtσx,y,z are not zeros. Real axial strain should be obtained by solving the constitutive equation. As discussed before, dnsdT will be negative value by assuming the main content of tissues is water [6]. Moreover, dnsdT is also depending on the density of the material [24]. Hence, the state of the thermal expansion also affect the refractive index change because the expansion alter the density. When the refractive index change is dominant comparing to thermal expansion, Eq. (11) will have negative peak at the location of absorber.

3. Methods

3.1. Phase-sensitive swept-source photothermal OCT system

Phase-sensitive swept-source OCT is developed to measure photothermal OPL change. A schematic diagram of the experimental setup is shown in Fig. 1. The light source is an MEMS-based wavelength-wept laser (AXSUN Technologies Inc., Billerica, Massachusetts) with a center wavelength of 1310 nm, scanning rate of 47 kHz and scanning range of 100 nm. Photocurrent from the balanced photoreceiver (Model 1817-FC, New Focus, Newport corporation, Irvine, California) is filtered by an RF low-pass filter (LP62CH3-0S, R&K Co., Ltd., Shizuoka, Japan, 62-MHz cut-off frequency) and sampled by a high-speed digitizer (ATS9350, Alazar Technologies Inc., Pointe-Claire, Quebec, Canada). The k-clock output from the swept laser source is used for the external sampling clock of the digitizer to linearly sample spectral interference signals in terms of the optical wavenumber. A fiber Bragg grating (FBG, reflection at 1266 ± 0.2 nm) is used to generate a trigger that is used to recognize the start of wavelength sweeping. The trigger is detected by the photodetector (PDA10CF, Thorlabs Inc.) and the electric trigger signal is sent to the digitizer. The incident power of the probe on the sample is 29 mW. Sensitivity of 110 dB was achieved.

Fig. 1.

Fig. 1

Schematic diagram of the photothermal OCT system. (a) The schema of the OCT engine and (b) the configuration of probe. Light from the source split and coupled by 90/10 fiber coupler where 90% portion directed to probe arm. FBG: fiber-Bragg grating, BPR: balanced photoreceiver, ND: neutral density filter, PD: photodetector, PC: polarization controller, DM: shortpass dichroic mirror, and LPF: lowpass filter.

In the probe arm [Fig. 1(b)], a fiber-pigtailed laser diode with a wavelength of 406 nm (LP406-SF20, Thorlabs Inc., Newton, New Jersey) is used as a photothermal excitation laser. The injection current and temperature of the laser diode are controlled through the laser diode mount with thermo-electric coolers (TCLDM9, Thorlabs Inc.) and laser diode controllers (LDC220C and TED200C, Thorlabs Inc.). The excitation laser is coaxially combined with the OCT probe using a short pass dichroic mirror (DMSP805R, Thorlabs Inc.). The galvanometric scanner and excitation laser modulation are controlled and synchronized using a function generator board (PCI-6713, National Instruments, Austin, Texas).

To minimize sampling jitter, the arrival timing of photocurrent from the balanced receiver, triggers generated by the FBG, and k-clock output to the digitizer are synchronized by adjusting optical fiber lengths and introducing an electronic delay (DB64, Stanford Research Systems, Sunnyvale, California) in the k-clock line. High phase stability close to noise-limited sensitivity ( 1/SNR) [30] was achieved.

To maximize the signal-to-noise ratio of photothermal imaging, the probe and excitation beams are precisely aligned to illuminate the tissue coaxially. Off-axis displacement results in ineffective photothermal imaging as suggested by Eq. (3). The beams focus into the tissue with e−2 spot diameters of approximately 18 μm (probe) and 12 μm (excitation).

The maximum permissible exposure (MPE) in a period of 10 s for skin is 10,000 W m−2 for the probe beam (MPEP) from IEC 60825-1 [31]. The current injected to the laser modulates the laser intensity and the peak power is approximately 5 mW. For the excitation beam, the continuous illumination with averaging power of 5 mW is the most restricted condition. Here we use the peak power of 5 mW as the average power for safety. The MPE of this condition is 2,000 W m−2 for the excitation beam (MPEE). The accessible emissions in the case of skin are calculated as 3,014 W m−2 and 519 W m−2 for the probe and excitation beams, respectively. The wavelengths of the probe and the excitation beams are shown as additive in Table 2 of Ref. [31], hence, the exposure of the current configuration is below the safety limit as 3,014W m2MPEP+519W m2MPEE=0.561<1.

The correspondence between the sign of phase shift and the direction of the axial movement of a sample has been confirmed by using manually translated scattering sample before the experiments.

3.2. Excitation and detection protocol

The axial distribution of the temporal alteration of temperature change is obtained from Eq. (11). A large temporal temperature change using high-intensity excitation and/or long time lag providing sufficient time for heating tissue is preferred for high-sensitivity measurement. However, the excitation laser intensity is limited by the safe exposure limit. Additionally, the temporal increase in temperature gradually reduces as the temperature change ΔT increases because the temperature distribution is going to be approaching the steady state temperature distribution (See Section 4.1). Moreover, a long time lag between OCT measurements Δt can collapse the phase-sensitive measurement of OCT. In the phase-sensitive method, OCT signals should be measured at the same location in tissues to preserve their correlation between different time points [28,29,32]. A time lag that is too long will generate huge displacement owing to sample motion and reduce the correlation. The following protocol is designed to address these issues.

The excitation, scanning, and detection timing is schematically shown in Fig. 2. For better sensitivity and a weaker effect of sample motion, it is necessary to measure just after turning on or turning off the excitation and to measure again after a time interval that is as short as possible. Hence, the excitation laser is sequentially switched on and off as moving to the next measurement location on the sample. Excitation and cooling times that are sufficient to warm and cool a region having a size similar to or larger than that of the OCT probe beam are required to detect photothermal signals efficiently. Suitable times may be comparable to the time constant, ts=wP24κ [33]. The radius of the current OCT probe beam is wP = 9 μm and the thermal diffusivity of water is κwater = 0.143 × 10−6 m2/s, resulting in ts = 142 μs. Several tens of sets of OPL change measurements are acquired at each location (M-scan). Hence, the duration of the acquisition at a single location is approximately 1 ms. The modulation frequency of the excitation laser is approximately 200 Hz to 300 Hz. The probe beam is translated by the galvanometric scanning mirror during around 400 μs temporal margin. This margin is required to stabilize the galvanometric scanning mirror. Afterward, the next excitation state is realized and the M-scan acquisition begins.

Fig. 2.

Fig. 2

Timing of the excitation and detection. GATE: the gate signal turns on and off acquisition, TRIG: A-scan trigger, GALVO: the waveform that controls the fast scanning galvanometer mirror, EXCT.: the waveform to modulate the photothermal excitation diode laser.

3.3. Signal processing

To extract the photothermal signal, two-dimensional (2D) phase differences for approximating mixed second-order partial derivatives (axial and temporal) are calculated for M-scan data. The flow of the processing is described in Fig. 3. The Hermitian product between axial profiles with a lag of Δn lines is calculated as

Ht(xl,zm,tn)j=ssΓ*(xl,zm+j,tn)Γ(xl,zm+j,tn+Δnδt), (12)

where δt is temporal interval between consecutive lines. Here, axial averaging with a kernel size of 3 (s = 1) is applied to enhance the phase sensitivity. The Hermitian product with axial pixel separation of Δm is then obtained as

Ht,z(xl,zm,tn)Ht*(xl,zmΔm/2,tn)Ht(xl,zm+Δm/2,tn). (13)

The phase of Eq. (13) is a local OPL change [Fig. 3(c)]. The Hermitian product is averaged over time within the M-scan as

Ht,z(xl,zm)n=0NΔn1M(xl,zm,tn)Ht,z(xl,zm,tn), (14)

where N is the total number of axial scans at the same location and M is a mask defined as

M={1|Ht,z|>θ0otherwise.

The mask M with a threshold level θ discards low–SNR data.

Fig. 3.

Fig. 3

The flow of the photothermal OCT processing. (a) A schematic diagram of scanning and excitation waveforms. (b) A raw 2D phase map of one M-scan. The signal processing described in Section 3.3 is applied ((c)–(g)) to obtain photothermal image.

The phase of the product represents the 2D phase difference [Fig. 3(d)] as

ΔzΔtϕ(xl,zm;Δmδz,Δnδt)arg[Ht,z(xl,zm)], (15)

where δz is axial interval between consecutive pixels. This 2D phase difference ΔzΔtϕ is associated with the temperature change and represents the distribution of the absorption agent as expressed in Section 2.2.2.

Because the excitation of the laser is modulated (Section 3.2), the photothermal signal Δ2ϕt,z is also modulated along the spatio-temporal direction. When the excitation laser is turned on, temperature change increases and tissue around light absorbers expands. In contrast, the temperature change is going to be flat due to thermal diffusion and expanded tissues constrict when the excitation laser is turned off. The photothermal signal that is synchronized with excitation modulation is extracted by spatio-temporal frequency filtering and de-modulated. The 2D phase difference is Fourier transformed along the spatio-temporal direction, x(t), as

ΔzΔtϕ(xl,zm)ΔzΔtΦ(νL,zm).

The Fourier spectrum ΔzΔt Φ is multiplied with a Tukey window w(νL) centered at the modulation frequency (i.e., the Nyquist frequency) which is a high pass filter to pass through frequency components over 20 % of the Nyquist frequency [Fig. 3(e)] and then demodulated as

ΔzΔtΦ˜(νL,zm){w(νLfm)ΔzΔtΦ(νLfm,zm)0<νLw(νL+fm)ΔzΔtΦ(νL+fm,zm)νL0, (16)

where fm is the modulation frequency of the excitation. The demodulated spectrum is then inversely Fourier transformed as

ΔzΔtΦ˜(νL,zm)1ΔzΔtϕ˜(xl,zm).

The real part of the filtered and demodulated phase difference normalized by temporal and axial intervals represents the distribution of photothermal response [Fig. 3(f)] as

Iph(xl,zm)Re[ΔzΔtϕ˜(xl,zm)]ΔmδzΔnδt. (17)

The sign of Iph represents the response is in-phase or out-of-phase to the excitation modulation. As discussed in Section 2.2.2, they are according to whether thermal expansion or refractive index change is dominant. Then, the photothermal OCT image, i.e. the magnitude of [Eq. (17)] reveals the distribution of absorbers [Fig. 3(g)].

Because modulated and non-modulated terms are separated in the spatio-temporal frequency domain [Eq. (16)], only the half of the detection bandwidth can be used for each term. It means that this method requires two times dense spatial sampling comparing to full band detection. This condition can be expressed as Δx/R < π/8 [34], where Δx is the spatial sampling step and R=2κτ is the thermal diffusion length and τ is the duration time of heating. In the current setup, τ equals to half the period of the modulation, 1 to 2 ms, then the lateral separation between successive M-scans should be less than 9.4 μm.

3.3.1. Artifact removal

The real phase of the OCT signal consists of not only photothermal OPL change but also the Doppler phase shift due to bulk motion of the sample and blood flow, and the texture of the tissue. However, these factors are canceled out in the above processing as explained as follows. The phase of the OCT signal can thus be generally expressed as

ϕ=ϕ(ph)(x,z,t)+ϕ(b)(x,t)+ϕ(tx)(x,z)+ϕ(bf)(x,z,t), (18)

where ϕ(ph), ϕ(b), ϕ(tx), and ϕ(bf) are phase components of the photothermal effect, bulk phase shift, structural texture, and blood flow, respectively. By applying temporal and axial derivatives such as those described in Eq. (15), the structural texture and the phase shift due to bulk sample motion are removed as

2ϕ(z,t)zt=2ϕ(ph)(z,t)zt+2ϕ(bf)(z,t)zt. (19)

The blood flow term is modulated by blood pulsation with frequency usually less than 2 Hz. Hence, if the excitation frequency is higher than 2 Hz [in this paper, approximately 200 Hz to 300 Hz (Section 3.2)], this term can also be excluded. After temporal averaging ϕt,z¯(x,z)=2ϕ[x,z,t]¯zt, spatio-temporal filtering is applied to extract the component at the excitation modulation frequency, fm, and the low-frequency component such as pulsation is filtered out. Only the photothermal effect, then, remains as

ϕt,z¯[x(t),z]|f=fm=ϕt,z(ph)¯[x(t),z]|f=fm. (20)

All phase components except that of the photothermal effect are removed as shown above. However, decorrelation due to displacement increases the phase noise [28, 29, 32]. This term manifests as artifacts in photothermal imaging. To mitigate the term, temporal averaging divided by the standard deviation of the 2D difference is replaced by the expression in Eq. (15) as

ΔzΔtϕ(xl,zm)=arg[Ht,z(xl,zm)]Var[arg[Ht,z(xl,zm,tn)]M(xl,zm,tn)]. (21)

Spatio-temporal filtering and demodulation are then applied.

4. Results

4.1. Photothermal signal verification with porcine liver

A porcine liver has been scanned by the photothermal OCT. One hundred axial lines were detected at a single location and 256 transversal positions (1 mm) were measured. The corresponding excitation modulation frequency was 235 Hz. The 2D phase difference was calculated with a time lag of 213 μs and an axial separation of 34.8 μm (refractive index of 1.38). The total acquisition time was approximately 0.68 seconds.

Figure 4 shows the photothermal signals of a porcine liver. Temporal and axial distributions of local OPL change in consecutive M-scans when heating stage [Fig. 4(a)] and cooling stage [Fig. 4(b)] are shown. The profile of the local OPL change at a certain depth (arrow in Fig. 4) presents the temporal slope of local OPL change [Fig. 4(c)]. Within one period of modulation of excitation, we assume the the laser power P(t) is constant at the former half period and turned off at the later half.

P(t)={P(tτ)0(t>τ) (22)

Temporal profile of photothermal signals in a M-scan [Eq. (11)] are related to the temporal derivative of Eq. (4). By introduce Eq. (22),

ΔT(r,t)t{P(4πκt)3/2exp(|r|24κt)(tτ)P(4πκ)3/2[1t3/2exp(|r|24κt)1(tτ)3/2exp(|r|24κ(tτ))](t>τ) (23)

Eq. (23) is fitted to the experimental data. As shown in Fig. 4(c), they are in good agreement.

Fig. 4.

Fig. 4

Photothermal signals in M-mode scan. Phase derivatives of two consecutive M-scans (a) when the excitation laser is on and (b) after the laser is turned off. (c) Temporal profile of photothermal signal. Scale bars are 500 μm in tissue. Each profile obtained at a certain depth of the porcine liver (black arrow).

Figure 5 shows cross-sectional OCT and photothermal images with/without excitation laser illumination. For quantitative comparison, the normalization process [Eq. (21)] was skipped. With photothermal excitation, the photothermal signal appeared inside of the liver tissue. At the surface of the tissue, strong negative signal is obtained. The reason may be the shrinkage of the superficial layer by dehydration due to photothermal excitation because the photothermal signals before frequency filtering are close to zero at locations when the excitation laser off (data is not shown). The positive and negative values are appeared inside of the tissue. This might be present the magnitude of the thermal expansion and refractive index change are comparable. For absorption contrast, the magnitude of photothermal signal is obtained and mapped. The signal-to-noise ratio is measured as the ratio of averaged photothermal signal magnitude with excitation to that of without excitation. The averaged signals are calculated within the white boxes shown in Fig. 5. The SNR of ∼ 7.2 was obtained. For a comparison, Fourier transform method is applied for phase derivative ΔzΔtϕ. Because the probe beam is scanned along lateral direction during the modulation, not peak magnitude detection but bandpass filtering around the modulation frequency and inverse Fourier transform were applied. They are shown in Fig. 5 Each image show the case of without and with excitation. With this Fourier transform method, SNR is measured as ∼ 4.8. The proposed method might have the advantage in sensitivity over the previous phase modulation detection technique at least 1.5 times.

Fig. 5.

Fig. 5

The images are compared. 213 us separation, 34.9 um separation, 235 Hz modulation. For quantitative comparison, normalization [Eq. (21)] was not applied. Scale bars are 500 μm in tissue.

4.2. Phantom imaging

A phantom was imaged to validate the depth-resolving performance of the developed photothermal OCT in locating absorbers. A scattering phantom was made by fixing 1 % soybean oil lipid emulsion (Intralipos®20%, Otsuka Pharmaceutical Factory Inc., Japan) with 10 % porcine gelatin (G2500, Sigma-Aldrich Corp., St. Louis, MO) and injecting guinea-pig blood. The size of a red blood cell of a guinea pig is close to that of a human. The diameter of the red blood cell was ∼ 7 μm and the mean corpuscular volume was ∼ 80 μm3 [35], whereas the values for a human are ∼ 8 μm and ∼ 90 μm3, respectively. The blood was diluted with Alsever’s solution to 50 % for preservation.

The phantom with guinea-pig preserved blood was scanned by photothermal OCT. The excitation modulation period was 4.26 ms, corresponding to a frequency of 235 Hz. The 2D phase difference was calculated with a time lag of 213 μs and an axial separation of 34.8 μm (refractive index of 1.38). One hundred axial lines were detected at a single location and 256 transversal positions were measured. The total acquisition time was approximately 0.68 seconds.

The beam scanning was aimed at the location of injected guinea-pig blood. Photothermal signals generated within the phantom with the excitation laser [Fig. 6(d)] were not present without excitation [Fig. 6(c)]. This contrast was not obtained in conventional structural OCT intensity images [Figs. 6(a) and 6(b)]. These signals probably represent the location of injected blood. The presented photothermal OCT thus provides depth-resolved contrast of absorption materials.

Fig. 6.

Fig. 6

OCT (a, b) and photothermal OCT (c, d) images of a phantom made of gelatin and soybean oil emulsion with guinea-pig blood. The left (a, c) and right (b, d) columns show images without and with excitation, respectively. Scale bars are 500 μm in tissue.

4.3. In vivo human skin imaging

To demonstrate the feasibility of the developed photothermal OCT, in vivo imaging of human skin was conducted. The imaging was performed in two regions; i.e., around a mole of the outer forearm and in a non-mole region. Fifty axial lines were detected at a single location and 256 × 128 transversal positions (0.5 × 0.5 mm2) were measured. A settling time of 425 μs (20 axial lines) was set after the transition of the galvanometric scanning mirror. The excitation modulation period was then 2.98 ms; i.e., the modulation frequency was 335.7 Hz. In addition, the excitation laser was turned on and off along the slow scanning direction (1.3 Hz) to obtain the photothermal OCT image with and without excitation modulation sequentially. The phase difference was calculated with a time lag of 213 μs (10 axial lines) and axial separation of 34.8 μm (refractive index of 1.38). The acquisition time of one B-scan was around 0.38 seconds. The total volumetric acquisition time was 49 seconds.

Two volumes were scanned on human skin. The images are shown in Fig. 7 [7(a)–7(h): around a mole, 7(i)–7(p): normal skin]. To create en-face images of photothermal OCT, the magnitudes of photothermal signals were integrated along the depth. For reference, the phase-resolved Doppler OCT algorithm [36] with bulk motion correction [37] was applied for MB-scans without excitation to contrast the blood flow distribution. The photothermal OCT images of the mole with excitation modulation clearly show strong signals close to the skin surface [Figs. 7(c) and 7(d), orange arrows]. There is a clear difference in the comparison with the photothermal image obtained without excitation [Figs. 7(e) and 7(f), orange arrows]. The strong signal might correspond to melanin in the mole. In addition to the skin’s melanin, a strong photothermal signal of hair is seen in photothermal OCT images obtained with excitation (green arrows).

Fig. 7.

Fig. 7

Volumes of OCT (first row) and photothermal OCT (second and third rows) images of human skin. Images in the the fourth row are images of Doppler blood flow for comparison. Two volumes around a mole (a–h) and of normal skin (i–p) are obtained. En-face projections (left) indicate the locations of the cross-sectional images (left).

Although the intensity is weak, there is structure in the photothermal image obtained without excitation [Fig. 7(e)]. It is clear that the signature in the en-face projection image well matches blood flow imaging using a Doppler phase shift [Fig. 7(g)]. Decorrelation phase shift noise due to blood flow might generate these artifacts.

In the case of normal inner forearm skin, a layered signal, definitely weaker than signal of the mole, is seen in the superficial skin with excitation [Figs. 7(k)–7(n), orange arrows]. This supports the idea that the photothermal signal of melanocyte at the epidermis is detected. A strong photothermal signal and blood flow signal are observed in the slightly deep tissue [Figs. 7(l) and 7(p), green arrows], which is not clear in the photothermal image obtained without excitation [Fig. 7(n), green arrow]. The appearance of this signal in the cross section and en-face image is in good agreement with the structure of capillaries in skin [38] and skin angiographies obtained in other studies [39, 40]. The signal might therefore correspond to a capillary of the skin.

5. Discussions

To increase the speed of photothermal OCT imaging, increasing the excitation modulation frequency required. However, a higher modulation frequency reduces the photothermal signals [6] since the maximum modulation frequency which retains the photothermal signal is limited by the size of the probe beam [33]. Optical coherence microscopy with other detection scheme is required to use high modulation frequency without loss of photothermal signal strengths [41]. We have been applied half period of modulation for each M-scan to achieve high imaging speed and low modulation frequency. In this configuration, Fourier transform method can not applied directly. The photothermal signals are estimated at each M-scan and modulation is applied with transversal beam scanning. Frequency filtering in spatio-temporal frequency domain. This algorithm is not only adapted for the presented detection protocol, but efficient to detect non-sinusoidal modulation as shown in Section 4.1.

For depth-resolved photothermal OCT imaging, the accumulation of OPL change due to the photothermal effect is resolved by differentiating phase information along axial direction after temporal differentiation. The temporal differentiation removes the texture noise (Section 3.3.1) and prevents the phase signal from the decorrelation noise due to axial separation. The direct phase differentiation between two depths [5, 19] should be affected by the decorrelation noise due to large separation between two depths compared to the axial resolution [42, 43]. On the other hand, the differentiation between adjacent pixels [9] makes OPL change very small because the magnitude of OPL change is proportional to a measured tissue thickness, i.e., the separation of two measurement depths. The other approach is calculating the slope of OPL change [20]. This is derived from the temperature change inside of a homogeneous absorption tissue. It assumes that the cumulative OPL change is monotonically increase or decrease. However, this assumption is controversial in inhomogeneous tissue. The photothermal effect on OPL consists of the thermal expansion and the refractive index change, and these effects probably cancel to each other [6, 12]. Although the magnitude of the coefficient of thermal expansion is roughly 2 times of that of the refractive index change [6, 12], real volume expansion and axial displacement are depends on mechanical confinement [26]. For example, if a stiff layer exists at the tissue surface, internal tissue cannot expand axially but its lateral expansion reduces the density of the tissue and then refractive index. In this case, the refractive index change play a main role of the OPL change. Hence, in biological tissues, it is hard to conclude that the thermal expansion is the dominant effect in the OPL change.

The contrast of blood in vivo requires further investigation. The photothermal signal of blood was probably generated by not only thermal expansion and refractive index change but also complex movement of blood cells due to the convection. The current theoretical model does not consider fluid such as blood. Thermal dynamics of fluid in a tube should thus be taken into account. Modeling of the photothermal effect with blood is required to optimize and understand the photothermal OCT measurement of blood. However, as shown in the phantom measurement, an advantage of photothermal OCT is that the technique can detect static blood due to such as hemorrhage. This may be difficult to obtain by employing Doppler or other motion contrast methods that are used to contrast blood flow [44].

The mismatch of the wavelengths between the excitation and probe beams cause a chromatic aberration issues. If longitudinal focal shift occurs, it will alter the laser fluence rate. However, in our configuration, the longitudinal focal shift was around 150 μm according to ray-tracing (OpticStudio, ZEMAX, LLC, Redmond, Washington). It was only around half the Rayleigh range of 280 μm. On the other hand, the transversal focal shift induces the offset between heated location and the OCT probe beam. This is affect to the temporal temperature change profile at the location of OCT probe beam. Hence, if the characteristic length is long enough to cover the offset plus beam spot size of the probe beam, there is no significant signal loss. However, too large offset results in the image mismatch between OCT and photothermal OCT images. It is expected around 5 μm in maximum when 0.5 mm transversal scan is applied in the current setup. Since the spot sizes of the excitation and probe beams are 12 μm and 18 μm, the effect of transversal focal shift would be negligible.

The limitation of the imaging depth is the combination of the limited penetration depth of OCT imaging and the limited fluence depth of the excitation laser. In this study, a single-mode fiber-pigtailed diode laser was selected because it provides a small illumination beam spot, high response of temporal output modulation, and stable operation at low cost. Some studies have expected the penetration depth of a blue (∼ 400 nm) wavelength to be less than 1 mm. A red (∼ 630 nm) wavelength is an alternative that provides deep penetration of tissue; additionally, consumer products of the laser diode are available for this wavelength. However, the imaging depth of photothermal OCT does not rely only the light penetration depth but fluence depth, i.e., light intensity times absorption coefficient [Eq. (2)]. Because absorption coefficients of endogenous agents, such as hemoglobin and melanin, at blue wavelength range are stronger than other visible lights, there is a certain depth where the blue wavelength exhibits higher fluence rate comparing to other visible lights and it might be deeper than the light penetration depth.

The photothermal fluence rate fPhTh is defined as

fPhTh(z)μa(agent)I0exp[(μs(skin)+μa(skin))z], (24)

where μa(agent) is the absorption coefficient of an absorption agent, μs(skin) and μa(skin) are scattering and absorption coefficients of skin, respectively, and z is the depth from the surface.

Figure 8 plots the photothermal fluence rate profiles of hemoglobin in skin for blue (405 nm) and red (630 nm) wavelengths. The plot indicates the strength of the photothermal signal against the depth of blood within skin. The absorption and scattering coefficients of hemoglobin ( μα(hemoglobin)(405 nm)=1,400 cm1, μα(hemoglobin)(630 nm)=26 cm1) [45] and the human skin ( μs(skin)(405 nm)=70 cm1, μs(skin)(630 nm)=20 cm1, μα(skin)(405 nm)=3.8 cm1, μα(skin)(630 nm)=0.75 cm1) [46] are used in this numerical simulation. The transition point is located around 1 mm depth. It means that photothermal signal with blue excitation laser will be higher than that of red wavelength till depth of around 1 mm. Because the absorption coefficients of absorbers and attenuation of tissues are variable, the optimum excitation wavelength varies on each target. To determine the imaging depth of photothermal OCT, the SNR-dependent noise of phase-resolved OCT and the conversion factor from the laser fluence to photothermal signal magnitude should be accounted.

Fig. 8.

Fig. 8

Simulated axial fluence profile of the photothermal effect of hemoglobin in skin for 400 nm (blue line) and 630 nm (red line) wavelengths.

6. Conclusion

We presented a new photothermal OCT detection scheme for the imaging of endogenous absorption agents. A photothermal excitation protocol was designed and a phase processing method developed to sense thermal expansion in in vivo photothermal imaging. A phantom experiment demonstrated that the method has the ability to locate absorbers in a cross section. Non-invasive in vivo imaging of the human skin demonstrated the possibility of photothermal OCT applications.

References and links

  • 1.Fujimoto J. G., Brezinski M. E., Tearney G. J., Boppart S. A., Bouma B., Hee M. R., Southern J. F., Swanson E. A., “Optical biopsy and imaging using optical coherence tomography,” Nat. Med. 1, 970–972 (1995). 10.1038/nm0995-970 [DOI] [PubMed] [Google Scholar]
  • 2.Fleming C. P., Eckert J., Halpern E. F., Gardecki J. A., Tearney G. J., “Depth resolved detection of lipid using spectroscopic optical coherence tomography,” Biomed. Opt. Express 4, 1269–1284 (2013). 10.1364/BOE.4.001269 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Yi J., Wei Q., Liu W., Backman V., Zhang H. F., “Visible-light optical coherence tomography for retinal oximetry,” Opt. Lett. 38, 1796–1798 (2013). 10.1364/OL.38.001796 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Jacob D., Shelton R. L., Applegate B. E., “Fourier domain pump-probe optical coherence tomography imaging of Melanin,” Opt. Express 18, 12399–12410 (2010). 10.1364/OE.18.012399 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Akkin T., Davé D. P., Youn J.-I., Telenkov S. A., Rylander H. G., Milner T. E., “Imaging tissue response to electrical and photothermal stimulation with nanometer sensitivity,” Lasers Surg. Med. 33, 219–225 (2003). 10.1002/lsm.10221 [DOI] [PubMed] [Google Scholar]
  • 6.Adler D. C., Huang S.-W., Huber R., Fujimoto J. G., “Photothermal detection of gold nanoparticles using phase-sensitive optical coherence tomography,” Opt. Express 16, 4376–4393 (2008). 10.1364/OE.16.004376 [DOI] [PubMed] [Google Scholar]
  • 7.Skala M. C., Crow M. J., Wax A., Izatt J. A., “Photothermal optical coherence tomography of epidermal growth factor receptor in live cells using immunotargeted gold nanospheres,” Nano Lett. 8, 3461–3467 (2008). 10.1021/nl802351p [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Jung Y., Reif R., Zeng Y., Wang R. K., “Three-Dimensional High-Resolution Imaging of Gold Nanorods Uptake in Sentinel Lymph Nodes,” Nano Lett. 11, 2938–2943 (2011). 10.1021/nl2014394 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Chi T.-T., Tu Y.-C., Li M.-J., Chu C.-K., Chang Y.-W., Yu C.-K., Kiang Y.-W., Yang C. C., “Photothermal optical coherence tomography based on the localized surface plasmon resonance of Au nanoring,” Opt. Express 22, 11754–11769 (2014). 10.1364/OE.22.011754 [DOI] [PubMed] [Google Scholar]
  • 10.Kuranov R. V., Kazmi S., McElroy A. B., Kiel J. W., Dunn A. K., Milner T. E., Duong T. Q., “In vivo depth-resolved oxygen saturation by dual-wavelength photothermal (DWP) OCT,” Opt. Express 19, 23831 (2011). 10.1364/OE.19.023831 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Oh J., Feldman M. D., Kim J., Sanghi P., Do D., Mancuso J. J., Kemp N., Cilingiroglu M., Milner T. E., “Detection of macrophages in atherosclerotic tissue using magnetic nanoparticles and differential phase optical coherence tomography,” J. Biomed. Opt. 13, 054006 (2008). 10.1117/1.2985762 [DOI] [PubMed] [Google Scholar]
  • 12.Kim J., Oh J., Kang H. W., Feldman M. D., Milner T. E., “Photothermal response of superparamagnetic iron oxide nanoparticles,” Lasers Surg. Med. 40, 415–421 (2008). 10.1002/lsm.20650 [DOI] [PubMed] [Google Scholar]
  • 13.Zhou C., Tsai T.-H., Adler D. C., Lee H.-C., Cohen D. W., Mondelblatt A., Wang Y., Connolly J. L., Fujimoto J. G., “Photothermal optical coherence tomography in ex vivo human breast tissues using gold nanoshells,” Opt. Lett. 35, 700–702 (2010). 10.1364/OL.35.000700 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Paranjape A. S., Kuranov R., Baranov S., Ma L. L., Villard J. W., Wang T., Sokolov K. V., Feldman M. D., Johnston K. P., Milner T. E., “Depth resolved photothermal OCT detection of macrophages in tissue using nanorose,” Biomed. Opt. Express 1, 2–16 (2010). 10.1364/BOE.1.000002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Tucker-Schwartz J. M., Meyer T. A., Patil C. A., Duvall C. L., Skala M. C., “In vivo photothermal optical coherence tomography of gold nanorod contrast agents,” Biomed. Opt. Express 3, 2881–2895 (2012). 10.1364/BOE.3.002881 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Xiao P., Li Q., Joo Y., Nam J., Hwang S., Song J., Kim S., Joo C., Kim K. H., “Detection of pH-induced aggregation of “smart” gold nanoparticles with photothermal optical coherence tomography,” Opt. Lett. 38, 4429–4432 (2013). 10.1364/OL.38.004429 [DOI] [PubMed] [Google Scholar]
  • 17.Tucker-Schwartz J. M., Hong T., Colvin D. C., Xu Y., Skala M. C., “Dual-modality photothermal optical coherence tomography and magnetic-resonance imaging of carbon nanotubes,” Opt. Lett. 37, 872–874 (2012). 10.1364/OL.37.000872 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Subhash H. M., Xie H., Smith J. W., McCarty O. J. T., “Optical detection of indocyanine green encapsulated biocompatible poly (lactic-co-glycolic) acid nanoparticles with photothermal optical coherence tomography,” Opt. Lett. 37, 981–983 (2012). 10.1364/OL.37.000981 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Telenkov S. A., Dave D. P., Sethuraman S., Akkin T., Milner T. E., “Differential phase optical coherence probe for depth-resolved detection of photothermal response in tissue,” Phys. Med. Biol. 49, 111–119 (2004). 10.1088/0031-9155/49/1/008 [DOI] [PubMed] [Google Scholar]
  • 20.Guan G., Reif R., Huang Z., Wang R. K., “Depth profiling of photothermal compound concentrations using phase sensitive optical coherence tomography,” J. Biomed. Opt. 16, 126003 (2011). 10.1117/1.3659211 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Gemert M. J. C. v., Lucassen G. W., Welch A. J., “Time constants in thermal laser medicine: II. Distributions of time constants and thermal relaxation of tissue,” Phys. Med. Biol. 41, 1381 (1996). 10.1088/0031-9155/41/8/009 [DOI] [PubMed] [Google Scholar]
  • 22.Carslaw H. S., Conduction of Heat in Solids, 2nd ed. (Oxford University Press, 1986). [Google Scholar]
  • 23.Roider J., Birngruber R., “Solution of the heat conduction equation,” in Optical-Thermal Response of Laser-Irradiated Tissue, 1st ed. Welch A. J., Gemert M. J. C. v., eds. (Springer, 1995), pp. 385–409. 10.1007/978-1-4757-6092-7_12 [DOI] [Google Scholar]
  • 24.Bialkowski S., Photothermal Spectroscopy Methods for Chemical Analysis, (John Wiley & Sons, 1996). [Google Scholar]
  • 25.Song S., Huang Z., Wang R. K., “Tracking mechanical wave propagation within tissue using phase-sensitive optical coherence tomography: motion artifact and its compensation,” J. Biomed. Opt. 18, 121505 (2013). 10.1117/1.JBO.18.12.121505 [DOI] [PubMed] [Google Scholar]
  • 26.Kim J., Oh J., Milner T. E., “Measurement of optical path length change following pulsed laser irradiation using differential phase optical coherence tomography,” J. Biomed. Opt. 11, 041122 (2006). 10.1117/1.2236289 [DOI] [PubMed] [Google Scholar]
  • 27.Kupradze V. D., Three-Dimensional Problems of Elasticity and Thermoelasticity (Elsevier, 1979). [Google Scholar]
  • 28.Vakoc B. J., Tearney G. J., Bouma B. E., “Statistical Properties of Phase-Decorrelation in Phase-Resolved Doppler Optical Coherence Tomography,” IEEE Trans. Med. Imaging 28, 814–821 (2009). 10.1109/TMI.2009.2012891 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Makita S., Jaillon F., Jahan I., Yasuno Y., “Noise statistics of phase-resolved optical coherence tomography imaging: single-and dual-beam-scan Doppler optical coherence tomography,” Opt. Express 22, 4830–4848 (2014). 10.1364/OE.22.004830 [DOI] [PubMed] [Google Scholar]
  • 30.Park B. H., Pierce M. C., Cense B., Yun S.-H., Mujat M., Tearney G. J., Bouma B. E., Boer J. F. d., “Real-time fiber-based multi-functional spectral-domain optical coherence tomography at 1.3 μm,” Opt. Express 13, 3931–3944 (2005). 10.1364/OPEX.13.003931 [DOI] [PubMed] [Google Scholar]
  • 31.International Electrotechnical Commission , Safety of laser products – Part 1: Equipment classification and requirements, (International Electrotechnical Commission, 2007), 2nd ed. [Google Scholar]
  • 32.Walther J., Koch E., “Transverse motion as a source of noise and reduced correlation of the Doppler phase shift in spectral domain OCT,” Opt. Express 17, 19698–19713 (2009). 10.1364/OE.17.019698 [DOI] [PubMed] [Google Scholar]
  • 33.Boyer D., Tamarat P., Maali A., Lounis B., Orrit M., “Photothermal imaging of nanometer-sized metal particles among scatterers,” Science 297, 1160–1163 (2002). 10.1126/science.1073765 [DOI] [PubMed] [Google Scholar]
  • 34.Makita S., Fabritius T., Yasuno Y., “Full-range, high-speed, high-resolution 1-μm spectral-domain optical coherence tomography using BM-scan for volumetric imaging of the human posterior eye,” Opt. Express 16, 8406–8420 (2008). 10.1364/OE.16.008406 [DOI] [PubMed] [Google Scholar]
  • 35.Jain N. C., Essentials of Veterinary Hematology (Lea & Febiger, 1993). [Google Scholar]
  • 36.Hong Y.-J., Makita S., Jaillon F., Ju M. J., Min E. J., Lee B. H., Itoh M., Miura M., Yasuno Y., “High-penetration swept source Doppler optical coherence angiography by fully numerical phase stabilization,” Opt. Express 20, 2740–2760 (2012). 10.1364/OE.20.002740 [DOI] [PubMed] [Google Scholar]
  • 37.Kurokawa K., Sasaki K., Makita S., Hong Y.-J., Yasuno Y., “Three-dimensional retinal and choroidal capillary imaging by power Doppler optical coherence angiography with adaptive optics,” Opt. Express 20, 22796–22812 (2012). 10.1364/OE.20.022796 [DOI] [PubMed] [Google Scholar]
  • 38.Skobe M., Detmar M., “Structure, Function, and Molecular Control of the Skin Lymphatic System,” J. Investig. Dermatol. Symp. Proc. 5, 14–19 (2000). 10.1046/j.1087-0024.2000.00001.x [DOI] [PubMed] [Google Scholar]
  • 39.An L., Qin J., Wang R. K., “Ultrahigh sensitive optical microangiography for in vivo imaging of microcirculations within human skin tissue beds,” Opt. Express 18, 8220–8228 (2010). 10.1364/OE.18.008220 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Blatter C., Weingast J., Alex A., Grajciar B., Wieser W., Drexler W., Huber R., Leitgeb R. A., “In situ structural and microangiographic assessment of human skin lesions with high-speed OCT,” Biomed. Opt. Express 3, 2636–2646 (2012). 10.1364/BOE.3.002636 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Pache C., Bocchio N. L., Bouwens A., Villiger M., Berclaz C., Goulley J., Gibson M. I., Santschi C., Lasser T., “Fast three-dimensional imaging of gold nanoparticles in living cells with photothermal optical lock-in Optical Coherence Microscopy,” Opt. Express 20, 21385–21399 (2012). 10.1364/OE.20.021385 [DOI] [PubMed] [Google Scholar]
  • 42.Lee J., Wu W., Jiang J. Y., Zhu B., Boas D. A., “Dynamic light scattering optical coherence tomography,” Opt. Express 20, 22262–22277 (2012). 10.1364/OE.20.022262 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Weiss N., van Leeuwen T. G., Kalkman J., “Localized measurement of longitudinal and transverse flow velocities in colloidal suspensions using optical coherence tomography,” Phys. Rev. E 88, 042312 (2013). 10.1103/PhysRevE.88.042312 [DOI] [PubMed] [Google Scholar]
  • 44.Kernick D. P., Tooke J. E., Shore A. C., “The biological zero signal in laser doppler fluximetry – origins and practical implications,” Pflügers Arch. 437, 624–631 (1999). 10.1007/s004240050826 [DOI] [PubMed] [Google Scholar]
  • 45.Prahl S., “Optical Absorption of Hemoglobin,” (1999). http://omlc.org/spectra/hemoglobin/index.html.
  • 46.Bashkatov A. N., Genina E. A., Kochubey V. I., Tuchin V. V., “Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from 400 to 2000 nm,” J. Phys. D: Appl. Phys. 38, 2543 (2005). 10.1088/0022-3727/38/15/004 [DOI] [Google Scholar]

Articles from Biomedical Optics Express are provided here courtesy of Optica Publishing Group

RESOURCES