Published in final edited form as: IEEE Trans Med Imaging. 2014 Dec;33(12):2213–2222. doi: 10.1109/TMI.2014.2333370

Fast T2 Mapping With Improved Accuracy Using Undersampled Spin-Echo MRI and Model-Based Reconstructions With a Generating Function

Tilman J Sumpf 1,, Andreas Petrovic 2, Martin Uecker 3, Florian Knoll 4, Jens Frahm 5
PMCID: PMC4469336  NIHMSID: NIHMS695520  PMID: 24988592

Abstract

A model-based reconstruction technique for accelerated T2 mapping with improved accuracy is proposed using under-sampled Cartesian spin-echo magnetic resonance imaging (MRI) data. The technique employs an advanced signal model for T2 relaxation that accounts for contributions from indirect echoes in a train of multiple spin echoes. An iterative solution of the nonlinear inverse reconstruction problem directly estimates spin-density and T2 maps from undersampled raw data. The algorithm is validated for simulated data as well as phantom and human brain MRI at 3T. The performance of the advanced model is compared to conventional pixel-based fitting of echo-time images from fully sampled data. The proposed method yields more accurate T2 values than the mono-exponential model and allows for retrospective under-sampling factors of at least 6. Although limitations are observed for very long T2 relaxation times, respective reconstruction problems may be overcome by a gradient dampening approach. The analytical gradient of the utilized cost function is included as Appendix. The source code is made available to the community.

Index Terms: Fast spin echo (FSE), indirect echoes, model-based reconstruction, relaxometry, T2 mapping

I. Introduction

Quantitative evaluations of T2 relaxation times are of importance for an increasing number of clinical medical resonance imaging (MRI) studies [1]. Conventional T2 mapping relies on the time-demanding acquisition of fully sampled multi-echo spin-echo (MSE) MRI datasets. Recent advances exploit a model-based reconstruction strategy which allows for accelerated T2 mapping from undersampled data [2]–[5]. So far, however, these approaches have been limited to a mono-exponential signal model, whereas experimental spin-echo trains are well known to deviate from the idealized behavior, for example, because of B1 inhomogeneities and nonrectangular slice profiles [6]–[8]. Under these circumstances, model-based reconstructions lead to T2 errors even for fully sampled conditions and additional deviations for different degrees of undersampling. As a consequence, it is common practice to discard the most prominently affected first echo of a MSE acquisition [4], [9], [10].

To overcome the aforementioned problem, an analytical formula has been proposed which models the MSE signal more accurately [11], [12]. This work demonstrates the application of this advanced model [12] for a model-based reconstruction. The new approach allows for highly accelerated as well as accurate T2 mapping from undersampled Cartesian MSE MRI data. Part of this work has been presented in [13]. A related approach was recently proposed in [14], where the extended phase graph (EPG) model is used in a dictionary-based linear reconstruction algorithm. First promising results for a nonlinear inversion of the computationally demanding EPG algorithm have been reported in [15] for small matrices (64×64 samples).

II. Theory

T2 mapping relies on a train of successively refocused spin echoes. In practical MRI, however, the assumption that these signals represent the true T2 relaxation decay is violated due to the formation of indirect echoes [16], [17]. The most notable consequence is a hypointense first spin echo which clearly disrupts the expected mono-exponential signal decay. A variety of attempts to better reproduce the echo amplitudes employ the EPG algorithm [18]–[21] which considers different magnetization pathways in a recursive description. In 2007 Lukzen et al. [11] obtained an explicit analytical expression for the problem by exploiting the generating function (GF) formalism [22], [23]. In contrast to the EPG algorithm, the model formula can be implemented very efficiently using the fast Fourier transform. An example of the improved fitting of the extended GF model [12], which includes nonideal slice profiles, against the mono-exponential fitting is shown in Fig. 1 for human brain MRI data.

Fig. 1.

Fig. 1

GF and single exponential (Exp) fitted to the magnitude signal (circles) of a multi-echo spin-echo MRI acquisition of the human brain (25 echoes, single pixel = arrow). Because the GF is only valid at exact echo times, the solid curve is an interpolation for display purposes.

A. Generating Function

The GF for the MSE signal amplitudes is given in the z-transform domain [11]

Gρ,α,k1,k2(z)=ρ2+ρ2(1+zk2)[1-z(k1+k2)cosα+z2k1k2](-1+zk2)[-1+z(k1-k2)cosα+z2k1k2] (1)

where ρ is the spin density, k1 = exp(−τ/T1) and k2 = exp(−τ/T2) are the T1 and T2 relaxation terms, α is the refocusing flip angle, and τ the echo spacing. z denotes a complex variable in the z-domain. Evaluation of (1) on the unit circle, i.e., for z = exp(), ω = 0…2π, yields a frequency-domain representation of the echo amplitudes at echo times TE.

The nonuniform flip-angle distribution of a nonideal slice profile can be accounted for by superimposing evaluations of (1) for different values of α [12]. The final formulation in frequency domain is given by

Sρ,k2(ω)=1Qq=1QGρ,αq,k1,k2(eiω) (2)

with αq finite number of Q supporting points characterizing the profile of the refocusing pulse in slice direction. The values for k1 and the absolute magnitudes of the angles αq have to be taken from experimentally determined T1 and B1 maps. The echo amplitudes in time domain can be recovered by application of a discrete Fourier transform (DFT) on the resulting frequency-domain samples.

Given a series of magnitude images from a MSE train, the GF approach can be used to determine quantitative T2 values at different spatial positions by pixel-wise fitting. The method has been demonstrated to yield more accurate T2 estimations than a mono-exponential fit [12]. As a potential limitation, (2) requires a valid T1 and B1 map prior to T2 reconstruction as well as an estimation of the pulse profile in slice direction. The influence of errors in these estimates on the reconstructed T2 maps has previously been elaborated for fully sampled data [12].

B. Reconstruction From Undersampled Data

In addition to the DFT Fω along the samples in frequency domain, (2) can be extended by a 2-D DFT Fxy to synthesize k-space samples from estimated parameter maps. Similar to the approaches described in [2], [4] the conformity of these (synthetic) data with the experimentally available samples sc from a MSE acquisition can be quantified with a cost function

Φ(x)=12cM(x)-sc22 (3)
M(x)=PFxyCcFωW(x) (4)
x=(ρk2) (5)
ρ=(ρ1,,ρNp) (6)
k2=(k2,1,,k2,Np)T (7)

where the diagonal matrices P and Cc contain the binary sampling pattern and the complex coil sensitivities of the coil elements c, respectively. W(x) represents the combined results from evaluating (2) at all Np pixel positions and Nω frequencies. Minimization of (3) with respect to the components of x allows for the direct reconstruction of T2 = −τ/ln(k2) and ρ parameter maps from undersampled data.

C. Column-Wise Reconstruction

When using Cartesian sampling schemes where undersampling is only performed in the phase-direction (y), the Fourier transform Fx in read-direction can be excluded from the cost function (3) and replaced by a respective inverse DFT of the data samples sc prior the iterative reconstruction. This approach not only reduces the computational costs for the evaluation of each cost function, but also allows for an independent and parallel reconstruction of image columns. This strategy splits the overall image-reconstruction into much smaller problems, which usually converge significantly faster than the respective global optimization approach. Another advantage is the possibility to remove noise columns from the reconstruction, e.g., by masking columns with an overall energy below a given threshold.

D. Oversampling on the z-Plane

As stated before, evaluation of the GF on the unit circle in z-domain allows for the calculation of MSE amplitudes by application of a DFT in z-direction. The range of echo times is inversely proportional to the frequency resolution, so that for Nω frequency samples the longest modeled echo time yields

TEmax=Nωτ. (8)

As a rule of thumb, TEmax should be at least six times the longest T2 within the measured object to ensure a proper coverage of the T2 signal decay. If this requirement is violated, the modeled echo amplitudes (i.e., the DFT of the GF) become distorted due to aliasing in time direction. In practice, typical T2 mapping protocols use 16 echoes with an echo spacing of 10 ms. To avoid reconstruction errors for T2 values longer than about 30 ms, it is therefore reasonable to evaluate (1) for a considerably higher number of frequency samples than the number NE of actually measured echoes. Unfortunately, this oversampling on the z-plane yields a substantial increase of the computational costs. The current implementation therefore applied only a moderate oversampling (Nω; = 128) to define T2 values of up to 213 ms (τ = 10 ms) which cover most tissues in brain [24] and other organ systems except for fluid compartments. In fact, the precise determination of these very long T2 values has only limited clinical relevance. For conventional pixel-wise fitting it is therefore reasonable to accept quantitative errors in regions with T2 values above this limit. For model-based reconstructions, however, pixels with an implausible signal behavior or long T2 cannot simply be excluded from the vector of unknowns. Respective artifacts, in particular for reconstructions from undersampled data, need to be treated by additional means (see below).

III. Materials and Methods

A. Optimization and Gradient Scaling

In accordance to [2] and [4] we used the CG-DESCENT algorithm [25] to minimize the nonlinear cost function (3). This method offers a guaranteed descent and an efficient line search but requires the gradient of the objective function to be balanced with respect to its partial derivatives. Therefore, the vectors k2 and ρ in (5) have been substituted by scaled variants

ρ=Lρ-1ρ (9)
k2=Lk-1k2 (10)

with Lp and Lk being diagonal scaling matrices. The dimensioning of the scaling involves several challenges. For example, we found large T2 values (T2 > TEmax/6) to provoke very high entries in the cost function gradient, so that even few discrete regions may lead to a global failure of the reconstruction process when using scalar values for the scaling. To counterbalance these effects, a dynamic validity mask has been implemented to detect potentially destructive pixels during reconstruction and to reduce the gradient scaling in respective regions. Because such regions are initially unknown for under-sampled data, the reconstruction was performed in three steps with a fixed number of 3×20 CG-iterations for each image column. After each of the three optimization blocks, the validity mask was updated and the gradient scaling was dampened by a factor of 10−5 in regions with a T2 of either more than TEmax/6 or less than 0 ms. With all data initially standardized by their overall L2 norm, the scaling was initialized with heuristically chosen scalar values of Lp = 1 and Lk = 40 . The same values were used for all reconstructions in this paper.

Pixel-wise fitting of echo-time images involved the Matlab nlinfit program (MathWorks, Natick, MA, USA), i.e., the Levenberg Marquard algorithm. In contrast to the CG-DESCENT optimization approach, the gradient was approximated using the internal finite-difference method of nlinfit rather than the explicit analytical expression given in the Appendix.

B. Regularization

Especially in the first optimization block, the model insufficiencies for large T2 values can lead to strong outliers in the reconstructed maps. The effect can be reduced by penalizing the L2 norm of the difference between estimates ρ and k2 and their initial guesses ρ0 and k2,0. The technique, known as Tikhonov regularization, introduces the regularization parameters λp and λk to the cost function

Φ(x)=12cM(x)-sc22+λρρ-ρ022+λkk2-k2,022. (11)

The parameters balance the data fitting term and the penalty terms and need to be tuned accordingly. Here, we chose an iteratively regularized approach with starting values of λp = 1·10−3 and λk = 3·10−3. After each of the three optimization blocks, both scaling factors are reduced by a factor of 10−3. This yields a moderate regularization at the beginning (far from the solution) and only minimal regularization at the end.

C. Numerical Simulations

To test the algorithm against different conditions, we used simulated data for a numerical phantom with multiple objects exhibiting equal spin density but different T2 relaxation times ranging from 50 to 800 ms. Simulated noiseless k-space samples for a 160×160 data matrix were derived from superimposed circles using the analytical Fourier space representation of an ellipse [26], [27]. The signal amplitudes in time domain were derived using (2) with subsequent DFT as a forward model. To avoid aliasing in time direction, the simulated data were derived from 4096 frequency-domain samples for every phantom compartment. Only the first 17 points were further used in time domain, corresponding to echo times of 10 to 170 ms. The simulated profile for the refocusing pulse was derived from a Gaussian using 16 support points. The T1 map was simulated to yield a constant value of T1 = 1000 ms throughout the FOV. The B1 field was simulated to be homogeneous and to correspond to an ideal flip angle of 180° at the center of the slice profile.

D. MRI Measurements

To experimentally validate the accuracy of the approach using multi-channel data, we used MRI data of a commercially available relaxation phantom (DiagnosticSonar, Eurospin II, gel nos 3, 4, 7, 10, 14, and 16). The phantom contains six compartments with T2 values ranging from 46 to 166 ms and T1 values ranging from 311 to 1408 ms (at 21 °C). For human brain MRI, a young healthy volunteer with no known abnormalities participated in this study and gave written informed consent before the examination.

All MRI experiments were conducted at 3T (Tim Trio, Siemens Healthcare, Erlangen, Germany). While radio-frequency excitation was accomplished with the use of a body coil, signal reception was performed by a 12-channel head coil in CP mode (circular polarized), thus yielding data from four virtual elements. The field map was acquired using the method by Bloch–Siegert [28] (Gaussian pulse, B1peak = 0.11 G, KBS = 21.3 rad G−2 ms−1, 8 s duration, fOR = 8 kHz, FOV = 250×250 mm2 matrix = 64 × 32, slice thickness = 8 mm, α = 60°/120°, scan time = 13 s). T1 values were obtained using a Turbo Inversion Recovery (TIR) sequence (TR = 7 s, TE = 7.8 ms, TI = 100 − 3100 ms, turbo factor = 5, FOV = 250×250 mm2, matrix = 192×192, slice thickness = 4 mm, scan time = 17 : 05 min) and a traditional magnitude fitting procedure. The data samples sc were acquired using a MSE sequence (TR = 4 s, echo spacing = 10 ms, 25 echoes, FOV = 250×250 mm2 matrix = 192×192, slice thickness = 4 mm, scan time = 12:54 min). For the relaxation phantom, 16 additional (single) spin-echo (SE) experiments were conducted with the same parameters and equally spaced echo times TE from 10 to 160 ms (scan time = 206 : 24 min).

E. Reconstruction and Undersampling

Apart from conventional mono-exponential fitting of echo-time images, simulated and measured SE and MSE data were analyzed using the proposed model-based reconstruction with the GF model (GF-MARTINI = Model-based Accelerated Relaxometry by Iterative Nonlinear Inversion). Undersampling for acceleration factors R of up to 12 was simulated using a “block” pattern, as depicted in Fig. 2. In contrast to the scheme used for mono-exponential MARTINI [4], the pattern was designed such that the first two echo times are sampled around the k-space center. In combination with the GF model, this strategy was found to minimize deviations in the T2 estimation at different undersampling factors.

Fig. 2.

Fig. 2

Cartesian encoding scheme with a blocked undersampling pattern. First two echoes are sampled in a block at the k-space center. For subsequent echoes, the block is circularly shifted by its size. The examples refer to acceleration factors of 3 (left) and 4 (right) and a k-space of 24 phase-encoded lines and nine echoes. Solid symbols represent acquired lines, while open symbols refer to lines not measured.

The initial spin-density map for GF-MARTINI was set to zero. The map for k2 was initialized with a value that corresponds to TEmax/6 for all pixels. T1 values were limited to 100 ms < T1 < 5000 ms. B1 values were limited to correspond to refocussing flip angles of at least 90° in the center of the assumed Gaussian slice profile. Coil sensitivities were estimated in a preprocessing step according to [29] using a fully sampled composite dataset derived from the mean samples of all echoes. The algorithm was constrained to shift all phase information into the complex coil sensitivities, while all parameter maps were assumed to be real. Finally, the sensitivity map of each receiver channel was standardized by the root sum square (RSS) of all channels.

Reference maps from fully sampled data were created by applying the nlinfit tool to magnitude images from the RSS of all receiver channels. For a constant number of echoes in time domain, GF-MARTINI was performed for different numbers of frequency-domain samples with and without the proposed validity mask. For mono-exponential fits the first echo was discarded.

IV. Results

A. Oversampling and Adaptive Masking

Fig. 3 shows spin-density and T2 maps of a noiseless numerical phantom reconstructed by pixel-wise GF fitting and the proposed GF-MARTINI method without and with the adaptive validity mask. The 17 time-domain data points (echoes) per pixel were derived from either 128 or 512 samples in the frequency-domain that define the degree of oversampling on the z-plane. As expected, even for the pixel-based reconstruction (Fig. 3, left), both the spin-density and T2 map yield errors for the compartment with the longest T2 of 800 ms when using only 128 samples (128, arrow). However, T2 values of the remaining compartments were accurately reconstructed with less than 2 ms deviation from the true value. The problem may largely be reduced by extensive oversampling with 512 frequency-domain samples (512) as revealed by a mean T2 value of 810 ms in the rightmost compartment, which is less than 2% deviation from the true value.

Fig. 3.

Fig. 3

Spin-density and color-coded T2 maps of a fully sampled numerical phantom with T2 compartments of 50, 100, and 800 ms (surrounding: 80 ms). Maps were obtained by (left) pixel-based GF fitting (128 and 512 frequency-domain samples) as well as GF-MARTINI reconstruction (middle) without and (right) with adaptive validity mask. All artifacts due to model violations can be avoided by the proposed method. For details see text.

Fig. 3 (center) shows the corresponding maps for GF-MARTINI when using 60 iterations per column and an optimized but constant gradient scaling for every pixel (no masking of invalid regions). The results in (128) not only demonstrate the expected deviations for the long-T2 compartment, but also artifacts (arrows) in the remote part of the columns that comprise the compartment. While the effect is again reduced for the higher number of frequency-domain samples (512), residual artifacts originating from model violations at the compartment borders persist.

Finally, Fig. 3 (right) demonstrates that all artifacts can be avoided by the proposed adaptive mask, even for moderate over-sampling of Nω = 128 samples. In this case the numerical results are in good agreement with the corresponding pixel-based GF fit (128) except for the masked-out right compartment. Here, the values for GF-MARTINI are mainly influenced by the applied L2 regularization, which is not included in the pixel-based fit.

B. Accuracy of the GF Model

Fig. 4 compares spin-density and T2 reconstructions obtained for a fully sampled SE and fully sampled MSE dataset. The results for a pixel-wise mono-exponential fitting of the SE magnitude (measurement time = 206 min) serve as a reference (ground truth). On the other hand, the corresponding fitting of the MSE images (first echo removed) yields the expected T2 overestimation in all compartments (arrows). In contrast, the GF fitting of the same data results in T2 values which are remarkably similar to the reference. Deviations in the surrounding water compartment, which are most notable in the spin-density map, are again an effect of the limited frequency domain oversampling (Nω = 128). The quantitative results from a ROI analysis in Table I confirm the visual impression. The relative error of the GF fit to the reference is 1% or less in all compartments. The mono-exponential fit, on the other hand, yields errors between 16% and 28%.

Fig. 4.

Fig. 4

Spin-density and color-coded T2 maps from fully sampled spin-echo (SE) and multi-echo spin-echo (MSE) data. In comparison to the SE reference, exponential fitting of the MSE data leads to an overestimation of T2 values (arrows), while GF fitting recovers the true values with high accuracy even when assuming a constant T1 = 1000 ms for all pixels.

TABLE I.

T2 Relaxation Times of a Phantom for Different Fitting Methods

Compartment 1 2 3 4 5 6
SE Reference 46 ± 2 81 ± 3 101 ± 2 132 ± 5 138 ± 4 166 ± 5
Mono-Exp 59 ± 4
28.3%
101 ± 5
24.7%
117 ± 4
15.8%
161 ± 5
22.0%
170 ± 4
23.2%
205 ± 7
23.5%
GF 46 ± 3
0.0%
81 ± 4
0.0%
100 ± 3
−1.0%
131 ± 4
−0.8%
137 ± 3
−0.7%
165 ± 5
−0.6%
GF, const T1 44 ± 3
−4.3%
79 ± 4
−2.5%
93 ± 3
−7.9%
130 ± 4
−1.5%
138 ± 3
0.0%
168 ± 6
1.2%

Absolute values represent a ROI analysis and are given in ms (mean ± SD).

Relative values for T2 estimates represent the deviation to the reference.

A practical limitation of the accurate GF fit is the necessity for additional B1 and T1 measurements. However, as has already been pointed out in [12] and [21], the influence of T1 on the GF result is relatively small. This finding is confirmed in Fig. 4 (right), where additional reconstructions were performed under the assumption of a constant T1 value of 1000 ms for all pixels. Although the relative error in T2 increases the stronger the chosen T1 deviates from the true value, the results are surprisingly accurate and even for worst cases, the largest error in the T2 map is still lower than the smallest error of the mono-exponential fit.

C. Undersampling

For the relaxation phantom, Fig. 5 demonstrates the results of the proposed GF-MARTINI method (128 frequency samples, validity mask, 3×20 CG iterations) for different degrees of un-dersampling. Again, effects of the limited oversampling in the frequency domain are most notable only in the spin-density map for the surrounding water compartment. However, this region was successfully identified as invalid after the first optimization block. Apart from that, all other regions are reconstructed with very high accuracy for all undersampling factors. Artifacts are barely notable except for some minor blurring around the compartment with the smallest T2 (red arrow). For undersampling factor greater than 6, some small vertical ghosts appear, as most notable in compartments 3 and 6 (blue arrows). For all cases a corresponding ROI analysis of T2 values is given in Table II. The resulting values are remarkably stable with deviations of less than ±2 ms from the fully sampled SE reference for all compartments and undersampling factors.

Fig. 5.

Fig. 5

Spin-density and color-coded T2 maps obtained by GF-MARTINI reconstructions (with validity mask) for undersampling factors of 1, 2, 6, 8, and 12. Corresponding measurement times were 12:54, 6:27, 2:09, 1:37, and 1:05 min. Highly undersampled results reveal minor blurring (red arrow) as well as occasional ghosts (blue arrows). Mean T2 values are accurate for all undersampling factors.

TABLE II.

T2 Relaxation Times of a Phantom Using GF-MARTINI With Different Undersampling Factors R

Compartment 1 2 3 4 5 6
R= 1 45 ± 3 81 ± 4 97 ± 3 132 ± 4 138 ± 3 166 ± 5
R = 2 46 ± 4 82 ± 4 97 ± 4 132 ± 4 138 ± 3 166 ± 5
R = 4 46 ± 4 82 ± 4 98 ± 4 133 ± 5 139 ± 3 168 ± 6
R = 6 46 ± 5 82 ± 5 98 ± 4 132 ± 5 139 ± 4 167 ± 6
R = 8 46 ± 4 82 ± 4 98 ± 4 132 ± 6 139 ±4 167 ± 6
R= 12 46 ± 4 82 ± 6 98 ± 7 133 ± 8 139 ± 5 168 ± 7

Values represent a ROI analysis and are given in ms (mean ± SD).

Corresponding results for a transverse section of the human brain are shown in Fig. 6. Whereas a pixel-wise exponential fit of the fully sampled MSE data (first echo removed) again leads to a systematic T2 overestimation, the GF-MARTINI method presents with good accuracy for all undersampling factors. The most notable distortions are small vertical ghosts near the hemispheric fissure, which increase for higher undersampling factors. For the highest acceleration factor of 12, also the spin-density map suffers from edge enhancement and blurring. A quantitative ROI analysis of T2 values in various brain tissues is summarized in Table III. The mean T2 values obtained by GF-MARTINI are in remarkably good agreement with the fully sampled reference and very stable for all undersampling factors. Moreover, Fig. 7 depicts reconstructions of the same data with the assumption of a constant T1 = 1000 ms for all pixels. It turns out that the GF-MARTINI reconstructions with constant T1 are almost indistinguishable from the results obtained with a full T1 map. The ROI analysis in Table III confirms this observation as all measurable deviations between the two GF-MARTINI versions remain within ±1 ms.

Fig. 6.

Fig. 6

(Top) Spin-density and (center) color-coded T2 maps of the human brain obtained by (left) exponential fitting of fully sampled data and (right) GF-MARTINI reconstructions with validity mask for undersampling factors of 1, 2, 6, and 12. Corresponding measurement times were 12:54, 6:27, 2:09, and 1:05 min. Reconstruction was performed with the use of a previously measured T1 map, requiring an additional measurement time of 17:05 min. (Bottom) T2 difference maps with respect to the fully sampled reference.

TABLE III.

T2 Relaxation Times of the Human Brain Using Pixel-Based Methods and GF-MARTINI for Different Undersampling Factors R

Anterior Cingulate Insular Cortex Thalamus Putamen Internal Capsule Frontal White Matter
Mono-Exp 101 ± 8 100 ± 10 84 ± 3 76 ± 5 96 ± 6 81 ± 3
GF 73 ± 6 76 ± 8 60 ± 2 56 ± 4 70 ± 4 60 ± 2
GF-MARTINI
R= 1 73 ± 6 76 ± 8 59 ± 2 55 ± 4 69 ± 4 59 ± 2
R = 2 71 ± 6 76 ± 8 59 ± 2 55 ± 4 69 ± 4 58 ± 3
R = 6 71 ± 6 76 ± 8 60 ± 3 56 ± 4 69 ± 4 59 ± 3
R= 12 71 ± 7 76 ± 8 60 ± 3 56 ± 5 70 ± 5 59 ± 3
GF-MARTINI, const T1
R= 1 73 ± 6 77 ± 8 59 ± 2 55 ± 4 69 ± 4 58 ± 2
R = 2 72 ± 6 76 ± 8 59 ± 2 56 ± 5 68 ± 4 58 ± 2
R = 6 72 ± 6 77 ± 9 60 ± 3 56 ± 5 69 ± 4 59 ± 3
R= 12 72 ± 7 77 ± 8 60 ± 3 56 ± 5 69 ± 5 59 ± 3

Values represent a ROI analysis and are given in ms (mean ± SD).

Fig. 7.

Fig. 7

Spin-density and color-coded T2 maps of the human brain obtained by GF-MARTINI reconstructions with validity mask and constant T1 = 1000 ms for undersampling factors of 1, 2, 6, and 12. The corresponding measurement times were 12:54, 6:27, 2:09, and 1:05 min.

D. Reconstruction Time

The computation time of the algorithm mainly depends on the matrix size, number of receive coils and number of frequency samples in the z-domain. In addition, the number of badly conditioned regions in the image can affect the reconstruction speed. Running on an Intel X5650 Xeon PC with 12 virtual CPUs@2.67 GHz, phantom and human brain MRI reconstructions in this study took between 2 and 3:45 min when using 128 frequency samples. Phantom calculations with Nω = 512 took up to 16 min. However, a major speedup is expected when performing the calculations on graphic processing units due to the highly parallelizable structure of the algorithm.

V. Discussion

This work demonstrates the adaptation of the GF approach, which properly describes the signal decay of a multiple spin-echo MRI acquisition, for a model-based reconstruction from highly undersampled MSE data in order to achieve both fast and accurate T2 mapping. For human brain MRI and in contrast to a mono-exponential model, the proposed GF-MARTINI method yields accurate T2 values for retrospective undersampling factors of at least 6, while reasonable T2 maps may even be obtained for retrospective undersampling factors of 12. Limitations for long T2 relaxation times of fluids are caused by echo-time aliasing due to the finite sampling of the spin-echo train in the frequency domain. However, for T2 values within a defined range of interest, the related errors can be avoided by moderate frequency-domain oversampling. The numerical instabilities caused by regions that strongly violate the GF model have further been shown to be controllable by adaptive masking and local dampening. Another strategy for a more tolerant performance for long T2 relaxation times might result from apodization techniques as proposed in [11]. So far, however, preliminary trials let to a notable loss of T2 accuracy.

In principle, the estimation of accurate T2 values by GF-MARTINI requires prior knowledge of T1 and B1 field distributions. This can be an obstacle in practice, especially if the maps are spatially inconsistent due to motion. However, while deviations in B1 indeed affect the T2 decay of MSE experiments, the T1 influence has been shown to be relatively small [12], [21]. Depending on the application and actual range of expected T1 values, the present findings support the practical feasibility of reliable reconstructions with the use of a single reasonable T1 estimate. B1 maps may be acquired at low resolution due to their smooth shape. It is also conceivable to treat the B1 map as an additional unknown parameter of the cost function. By jointly estimating B1 along with T2 and spin-density, additional measurements may be avoided. However, while preliminary trials yielded promising results for pixel-based fits, we encountered stability issues when reconstructing from undersampled data.

A potential limitation of the proposed method is its dependency on properly chosen scaling and regularization variables that balance the gradient components of the nonlinear cost function. In the current implementation, these variables are kept at a fixed starting value, while a relatively large number of 3×20 CG iterations per column is used to counter small deviations from an optimal choice. For a more efficient and general application of the method, an automatic mechanism for adjusting the scaling would be highly desirable. Such development, however, is outside the scope of this work.

Acknowledgments

The authors would like to thank T. Hilbert from the EPFL Lausanne for valuable ideas to improve the computational efficiency of the reconstruction algorithm.

Appendix

Derivation of the Cost Function Gradient

Optimization of the cost function Φ by means of the CG-DESCENT algorithm requires the implementation of the gradient of Φ with respect to the components in the scaled vector of real-valued unknowns . By combining all linear operations in (3) into a single system matrix Ac for every coil element , the cost function can be written as

Φ(x)=12cAcW(x)-sc22. (12)

The gradient of this term is given by

Φ=cR{DWH(x)AcH[AcW(x)-sc]} (13)

with DW the Jacobian matrix of W and (·)H referring to the adjoint. With (·̄) denoting complex conjugation, the adjoint system matrix AcH yields

AcH=FωHC¯cFxyHP (14)

where FxyH and FωH correspond to the inverse DFT in xy and frequency direction, respectively. Calculation of the Jacobian matrix DW requires the partial derivatives of the scaled model function

Gρ,α,k1,k2(z)=Lρρ2(1+λ1λ2λ3λ4) (15)
λ1=1+zLkk2 (16)
λ2=1-z(k1+Lkk2)cosα+z2k1Lkk2 (17)
λ3=-1+zLkk2 (18)
λ4=-1+z(k1-Lkk2)cosα+z2k1Lkk2 (19)

which are given by

ρGρ,α,k1,k2(z)=Lρ2(1+λ1λ2λ3λ4) (20)

and

k2Gρ,α,k1,k2(z)=Lρρ4Lk(λ2z+λ1λ5λ3λ4-zλ1λ2λ32λ4-λ1λ2λ5λ3λ42)(λ1λ2λ3λ4)-1/2 (21)
λ5=z2k1-zcosα (22)

for every pixel in image space.

The source code of the algorithm is available at: http://www.biomednmr.mpg.de/sw/gfmartini.html.

Footnotes

Color versions of one or more of the figures in this paper are available online at http://ieeexplore.ieee.org.

Contributor Information

Tilman J. Sumpf, Email: tsumpf@gwdg.de, Biomedizinische NMR Forschungs GmbH, Max-Planck-Institute for Biophysical Chemistry, 37077 Göttingen, Germany

Andreas Petrovic, Institute for Medical Engineering, Graz University of Technology, 8010 Graz, Austria and also with the Ludwig Boltzmann Institute for Clinical Forensic Imaging, 8010 Graz, Austria.

Martin Uecker, Department of Electrical Engineering and Computer Sciences, University of California, Berkeley, Berkeley, 94720 CA USA.

Florian Knoll, Institute for Medical Engineering, Graz University of Technology, 8010 Graz, Austria, and also with the Center for Biomedical Imaging, New York University School of Medicine, New York, NY 10016 USA.

Jens Frahm, Biomedizinische NMR Forschungs GmbH, Max-Planck-Institute for Biophysical Chemistry, 37077 Göttingen, Germany.

References

  • 1.Cheng HM, Stikov N, Ghugre NR, Wright GA. Practical medical applications of quantitative MR relaxometry. J Magn Reson Imag. 2012;36:805–824. doi: 10.1002/jmri.23718. [DOI] [PubMed] [Google Scholar]
  • 2.Block KT, Uecker M, Frahm J. Model-based iterative reconstruction for radial fast spin-echo MRI. IEEE Trans Med Imag. 2009 Nov;28(11):1759–1769. doi: 10.1109/TMI.2009.2023119. [DOI] [PubMed] [Google Scholar]
  • 3.Doneva M, Börnert P, Eggers H, Stehning C, Sénégas J, Mertins A. Compressed sensing reconstruction for magnetic resonance parameter mapping. Magn Reson Med. 2010;64:1114–1120. doi: 10.1002/mrm.22483. [DOI] [PubMed] [Google Scholar]
  • 4.Sumpf TJ, Uecker M, Boretius S, Frahm J. Model-based nonlinear inverse reconstruction for T2 mapping using highly undersampled spin-echo MRI. J Magn Reson Imag. 2011;34:420–428. doi: 10.1002/jmri.22634. [DOI] [PubMed] [Google Scholar]
  • 5.Huang C, Graff CG, Clarkson EW, Bilgin A, Altbach MI. T2 mapping from highly undersampled data by reconstruction of principal component coefficient maps using compressed sensing. Magn Reson Med. 2012;67:1355–1366. doi: 10.1002/mrm.23128. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Majumdar S, Orphanoudakis SC, Gmitro A, Odonnell M, Gore JC. Errors in the measurements of T2 using multiple-echo MRI techniques. 1. Effects of radiofrequency pulse imperfections. Magn Reson Med. 1986;3:397–417. doi: 10.1002/mrm.1910030305. [DOI] [PubMed] [Google Scholar]
  • 7.Majumdar S, Orphanoudakis SC, Gmitro A, Odonnell M, Gore JC. Errors in the measurements of T2 using multiple-echo MRI techniques. 2. Effects of static-field inhomogeneity. Magn Reson Med. 1986;3:562–574. doi: 10.1002/mrm.1910030410. [DOI] [PubMed] [Google Scholar]
  • 8.Crawley AP, Henkelman RM. Errors in T2 estimation using multislice multiple-echo imaging. Magn Reson Med. 1987;4:34–47. doi: 10.1002/mrm.1910040105. [DOI] [PubMed] [Google Scholar]
  • 9.Smith HE, Mosher TJ, Dardzinski BJ, Collins BG, Collins CM, Yang QX, Schmithorst VJ, Smith MB. Spatial variation in cartilage T2 of the knee. J Magn Reson Imag. 2001;14:50–55. doi: 10.1002/jmri.1150. [DOI] [PubMed] [Google Scholar]
  • 10.Mosher TJ, Liu Y, Torok CM. Functional cartilage MRI T2 mapping: Evaluating the effect of age and training on knee cartilage response to running. Osteo Cart. 2010;18:358–364. doi: 10.1016/j.joca.2009.11.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Lukzen NN, Savelov AA. Analytical derivation of multiple spin echo amplitudes with arbitrary refocusing angle. J Magn Reson. 2007;185:71–76. doi: 10.1016/j.jmr.2006.11.010. [DOI] [PubMed] [Google Scholar]
  • 12.Petrovic A, Scheurer E, Stollberger R. Closed-form solution for T2 mapping with nonideal refocusing of slice-selective CPMG sequences. Magn Reson Med. 2014 doi: 10.1002/mrm.25170. [DOI] [PubMed] [Google Scholar]
  • 13.Sumpf TJ, Knoll F, Frahm J, Stollberger R, Petrovic A. Nonlinear inverse reconstruction for T2 mapping using the generating function formalism on undersampled Cartesian data. Proc 20th Ann Meet ISMRM. 2012:2398. [Google Scholar]
  • 14.Huang C, Bilgin A, Barr T, Altbach MI. T2 relaxometry with indirect echo compensation from highly undersampled data. Magn Reson Med. 2013;70:1026–1037. doi: 10.1002/mrm.24540. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Lankford CL, Does MD. Fast, artifact-free T2 mapping with fast spin echo using the extended phase graph and joint parameter reconstruction. Proc. 21th Annu. Meet. ISMRM; Salt Lake City, UT. 2013; p. 2461. [Google Scholar]
  • 16.Hennig J. Multiecho imaging sequences with low refocusing flip angles. J Magn Reson. 1988;78:397–407. [Google Scholar]
  • 17.Hennig J. Echoes—How to generate, recognize, use or avoid them in MR-imaging sequences. Part I: Fundamental and not so fundamental properties of spin echoes. Conc Magn Reson. 2005;3:125–143. [Google Scholar]
  • 18.Woessner DE. Effects of diffusion in nuclear magnetic resonance spin-echo experiments. J Chem Phys. 1961;34:2057. [Google Scholar]
  • 19.Kiselev VG. Calculation of diffusion effect for arbitrary pulse sequences. J Magn Reson. 2003;164:205–211. doi: 10.1016/s1090-7807(03)00241-6. [DOI] [PubMed] [Google Scholar]
  • 20.Zur Y. An algorithm to calculate the NMR signal of a multi spin-echo sequence with relaxation and spin-diffusion. J Magn Reson. 2004;171:97–106. doi: 10.1016/j.jmr.2004.07.018. [DOI] [PubMed] [Google Scholar]
  • 21.Lebel RM, Wilman AH. Transverse relaxometry with stimulated echo compensation. Magn Reson Med. 2010;64:1005–1014. doi: 10.1002/mrm.22487. [DOI] [PubMed] [Google Scholar]
  • 22.Comtet L. Advanced Combinatorics: The art of Finite and Infinite Expansions. Boston, MA: Reidel; 1974. [Google Scholar]
  • 23.Wilf HS. Generating Functionology. Natick, MA: A. K. Peters; 1990. [Google Scholar]
  • 24.Poon CS, Henkelman RM. Practical T2 quantitation for clinical applications. J Mag Resonance Imag. 1992;2:541–553. doi: 10.1002/jmri.1880020512. [DOI] [PubMed] [Google Scholar]
  • 25.Hager WW, Zhang HA. A new conjugate gradient method with guaranteed descent and an efficient line search. SIAM J Optim. 2006;16:170–192. [Google Scholar]
  • 26.Slaney M, Kak A. Principles of Computerized Tomographic Imaging. Philadelphia, PA: SIAM; 1988. [Google Scholar]
  • 27.Van de Walle R, Barrett HH, Myers KJ, Altbach MI, Desplanques B, Gmitro AF, Cornelis J, Lemahieu I. Reconstruction of MR images from data acquired on a general nonregular grid by pseudoinverse calculation. IEEE Trans Med Imag. 2000 Dec;19(12):1160–1167. doi: 10.1109/42.897806. [DOI] [PubMed] [Google Scholar]
  • 28.Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B1 mapping by Bloch-Siegert shift. Magn Reson Med. 2010;63:1315–1322. doi: 10.1002/mrm.22357. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Uecker M, Hohage T, Block KT, Frahm J. Image reconstruction by regularized nonlinear inversion—Joint estimation of coil sensitivities and image content. Magn Reson Med. 2008;60:674–682. doi: 10.1002/mrm.21691. [DOI] [PubMed] [Google Scholar]

RESOURCES