Abstract
Osteoporosis involves degradation of bone’s trabecular architecture, cortical thinning, and enlargement of cortical pores. Increased cortical porosity is a major cause of the decreased strength of osteoporotic bone. The majority of cortical pores, however, are below the resolution limit of MRI.
Recent work has shown that porosity can be evaluated by MRI-based quantification of bone water. Bi-exponential T2* fitting and adiabatic inversion preparation are the two most common methods purported to distinguish bound and pore water in order to quantify matrix density and porosity.
To assess the viability of T2* bi-component analysis as a method for quantifying bound and pore water fractions, we have applied this method to human cortical bone at 1.5T, 3T, 7T, and 9.4T, and validated the resulting pool fractions against μCT-derived porosity and gravimetrically-determined bone densities. We also investigated alternative methods: 2D T1–T2* bi-component fitting by incorporating saturation-recovery, 1D and 2D fitting of CPMG echo amplitudes, and deuterium inversion recovery.
Short-T2* pool fraction was moderately correlated with porosity (R2 = 0.70) and matrix density (R2 = 0.63) at 1.5T, but the strengths of these associations were found to diminish rapidly as field strength increases, falling below R2 = 0.5 at 3T. Addition of the T1 dimension to bi-component analysis only slightly improved the strengths of these correlations. T2*-based bi-component analysis should therefore be used with caution.
Performance of deuterium inversion-recovery at 9.4T was also poor (R2 = 0.50 versus porosity and R2 = 0.46 versus matrix density). CPMG-derived short-T2 fraction at 9.4T, however, is highly correlated with porosity (R2 = 0.87) and matrix density (R2 = 0.88), confirming the utility of this method for independent validation of bone water pools.
Keywords: NMR, MRI, bi-component analysis, bone, bone water, relaxation, bound water, pore water
1. Introduction
Osteoporosis is a common bone disease which involves deterioration of trabecular bone architecture (1) and thinning and enlargement of pores in cortical bone (2). This increased cortical porosity is a major cause of the impaired strength of osteoporotic bone (3,4). Measurement of cortical porosity is, therefore, of great interest for assessment of bone health.
Micro-computed tomography (μCT), due to its exquisite spatial resolution on the order of microns, is a gold-standard method for direct visualization of pores and quantification of cortical porosity (5,6), but is applicable only ex vivo, and segmentation of pore spaces depends on selection of an appropriate intensity threshold. Dual-energy x-ray absorptiometry (DXA) is a standard in vivo screening tool that measures apparent areal bone mineral density, which varies inversely with porosity as long as mineralization density is constant. This density is ‘apparent’ in that it represents mineral per total bone area, including both matrix and pore spaces. DXA density is therefore affected both by changes in porosity (mesoscopic scale) and tissue mineralization (microscopic scale). Due to its two-dimensional nature, it is also affected by bone thickness. Quantitative computed tomography (QCT) adds a third dimension, but is still subject to the other limitations of DXA.
NMR and MRI methods have recently been used to study bone density and porosity. Unfortunately, the majority of cortical pores are smaller than the spatial resolution achievable by in vivo MRI, necessitating other approaches that do not rely on resolution of pore spaces. The NMR 1H signal in bone arises from three major pools. Water in bone is partitioned into two compartments: mobile water within the Haversian and lacunocanalicular pore system (7), and water which is hydrogen-bonded to bone matrix collagen (8–10). The third pool consists of protons in collagen and mineral (11).
As bone substance is lost and pore spaces expand, pore water concentration increases (9,10,12–17). Bound water concentration, which should parallel collagen density (8), is proportional to bone density and, therefore, is inversely proportional to porosity (9,10,12–14,18–22). Total water concentration, the sum of bound and pore water concentrations, is also weakly associated with porosity (9,13,16,23–26).
The three 1H signal pools can be separated based on their T2 relaxation times (8): pore water has T2 > 1 ms, bound water has T2 ~ 300–400 μs, and collagen has T2 < 60 μs. These components are shown in a schematic T2 spectrum (Figure 1). Due to its extremely short transverse relaxation time, the collagen and mineral 1H signal is beyond the reach of clinical MRI hardware, even with specialized solid-state pulse sequences.
There are two main NMR methods in routine use for distinguishing between bound and pore water: bi-exponential T2* fitting and T2-selective magnetization preparation. The former involves acquisition of signals at several TEs and fitting of a sum of two weighted exponential functions to the resulting decaying signals. The weights of each exponential have been hypothesized to represent bound and pore water signal amplitudes (12,22). The latter method uses an adiabatic RF pulse to selectively invert long-T2 pore water magnetization while saturating short-T2 bound water magnetization (13,18,21,27–29). After a specific inversion-recovery time (TI), pore water’s longitudinal magnetization (Mz) will be nulled while bound water magnetization will have recovered to Mz > 0.
T2* of pore water, however, is shortened due to strong internal magnetic field gradients arising from the difference in magnetic susceptibility between water and bone tissue (∆χ ~ 2.5 ppm SI) (30). The reduced separation between bound and pore water T2* values, illustrated in Figure 2, complicates separation via bi-component fitting, which, as a form of inverse Laplace transformation, is an ill-posed problem (31). Because the strength of the induced magnetic fields increases linearly with field strength, this effect becomes more severe at higher field strengths.
To assess the viability of T2* bi-component analysis as a method for quantifying bound and pore water fractions in humans, we have scanned a set of human cortical bone specimens at 1.5T, 3T, 7T, and 9.4T, and validated bi-exponential fitting of the resulting FIDs against μCT-derived porosity and gravimetrically-determined bone densities. These specimens are expected to vary widely in bone density and porosity. We also compared T2* bi-component analysis at these four field strengths to T2 bi-exponential fitting of CPMG echo amplitudes (8) and deuterium inversion-recovery NMR (9) at 9.4T.
2. Materials and Methods
2.1. Specimen Source and Preparation
The tissue examined consisted of 15 specimens of cortical bone taken from the tibial mid-shaft of previously frozen male and female human donors, aged 27–97 years (National Disease Research Interchange, NDRI). Donors with bone demineralizing disorders were excluded; only age-related structural bone loss is expected. A 4-mm slice was cut from each thawed tibia with a rotating blade at the region of maximum cortical bone thickness, 38% of the length of the tibia from the medial malleolus to the medial condyle. A rectangular beam was cut from the longest of the three faces of the roughly triangular bone slice, and trimmed to fit into a 5-mm NMR tube. The direction of the long axis of the bone was indicated on the end of each beam by cutting a notch parallel to the bone’s axis. Specimens were stored individually in phosphate-buffered saline.
2.2. 1H NMR Spectroscopy
1.5T, 3T, and 7T experiments were performed on whole-body human MRI scanners (Siemens, Erlangen, Germany) using custom-built 1H-free solenoidal radiofrequency (RF) coils (to eliminate signal contamination from the coil), 10 mm in diameter and 25 mm in length. Each coil was constructed of two parallel six-turn windings of copper wire on a polytetrafluoroethylene (PTFE) tube, mounted to a PTFE board, and tuned and matched capacitively. Coils were connected to the transmit/receive interface box (Stark Contrast, Erlangen, Germany) using PTFE-dielectric and -insulated coaxial cable. At 9.4T, experiments were performed on a vertical-bore NMR spectrometer and micro-imaging system (Avance III, Bruker, Billerica, MA) using a standard commercially-available 5-mm broadband inverse (BBI) probe with a 1-axis gradient.
Each bone was removed from its storage solution, gently blotted dry, placed quickly into a small, air-tight NMR tube with minimum interior air volume to prevent evaporation of free water, and scanned with a saturation-recovery (SR) pulse sequence, shown in Figure 3a. Longitudinal magnetization was saturated by a train of 90° pulses followed by spoiler gradients, and the partially-recovered magnetization was measured after each of 12 saturation recovery times (TSR) arrayed logarithmically from 3 ms to 6 s. 32 signal acquisitions were averaged, and scan time for this sequence was 6 min.
An equation consisting of the sum of two decaying exponentials plus a noise offset term,
[1] |
was fitted using non-linear least squares (NLLS) to the magnitude free induction decay (FID) data after the longest TSR. Short-T2* fraction is given by MS/(MS+ML), and the corresponding short T2* relaxation time by T2S*. Similarly, long-T2* fraction is given by ML/(MS+ML) and the long T2* relaxation time by T2L*.
Two-dimensional bi-component T1–T2* fitting, which has been shown to improve accuracy (32), was also performed by fitting a similar equation,
[2] |
to the set of saturation recovery-prepared magnitude FIDs. Here, the short-T2* fraction is given by MS/(MS+ML), the short T2* relaxation time by T2S*, and the short- T2* pool’s T1 relaxation time by T1S; fractions and relaxation times are analogous for the long- T2* pool. All reconstruction and fitting was performed in Matlab (Mathworks, Natick, MA).
Additionally, taking advantage of the lack of specific absorption rate (SAR) limitations and the availability of high-powered hardware at 9.4T, each bone was scanned using a SR-prepared Carr-Purcell-Meiboom-Gill (SR-CPMG) pulse sequence (33), shown in Figure 3b. TSRs were arrayed identically to the SR-FID sequence, and the number of refocusing pulses, N, was arrayed logarithmically from 0 to 5000 in 20 steps, and one signal acquisition was performed. Scan time for this sequence was 29 min. Analysis was performed similarly to the 1D and 2D bi-exponential fitting of FIDs, with FIDs simply substituted for arrays of echo amplitudes, using the following equations:
[3] |
and
[4] |
Finally, each bone was scanned with a single adiabatic inversion recovery-prepared FID (SIR-FID) pulse sequence, shown in Figure 3c, at all fields. Adiabatic inversion pulses may be designed with both long duration and broad bandwidth, and are thus T2-selective, rather than T2*-selective. Inversion time (TI) was stepped from 10 to 270 ms at TR = 300 ms. All other parameters were identical to SR-FID. The magnitude FID from each TI was processed by fitting a sum of two exponentials in a similar manner as 1D FID data, but with the relaxation times of the two pools instead set as fixed constants equal to the fitted values from 2D T1–T2* bi-component analysis. The TI at which the long-T2* pool fraction was minimized was selected for each bone as the optimal inversion-nulling time for pore water signal.
2.3. 2H NMR Spectroscopy
Labile protons, consisting predominantly of bound and pore water, were exchanged with 2H by immersion in deuterium oxide (D2O). Bones were blotted dry and placed in a 20-fold volume excess of 99.9% purity D2O-saline for six days, and were removed and transferred to a container with the same volume of fresh D2O-saline on days two and four to ensure full exchange.
Following full deuterium exchange, specimens were scanned using 2H inversion-recovery (IR) (9). A 2H spectrum of bone at 9.4T consists of a narrow central peak with long T1 flanked by a doublet with short T1. The narrow single peak corresponds to bone water residing in the pore system of bone and whose motion is unimpeded by interaction with bone collagen. The doublet peak with splitting of 4.8 kHz results from quadrupole interaction of the deuteron with the electric field gradient along the O-2H bond in bone water that is hydrogen-bonded to matrix collagen (34–36).
The pulse sequence and relevant parameters are shown in Figure 3d. TI was stepped in 10 ms increments in order to capture the null point of the narrow pore water peak, the post-acquisition delay was 1 s to ensure return to equilibrium longitudinal magnetization, and 48 signal acquisitions were averaged. Scan time for this sequence was 21 minutes. A fully relaxed spectrum was also acquired. The integral of the fully-relaxed spectrum represents total bone water, the integral of the spectrum with the narrow pore water signal nulled represents bound water only, and the difference between these two represents pore water only. Bound and pore water fractions were calculated by dividing the integral of the pore water-nulled spectrum or the difference spectrum, respectively, by the integral of the fully relaxed spectrum.
2.4. μCT Imaging
Bone specimens were scanned on a Scanco μCT35 scanner (Scanco, Brüttisellen, Switzerland) at 18.5-μm isotropic resolution. Bone exteriors were masked by 3D active snakes using the ITK-SNAP software package (37), and pores were segmented from this masked 3D image by thresholding. Porosity was calculated as pore (segmented) volume divided by total (masked) volume.
2.5. Gravimetry
Fully hydrated bone specimens were removed from liquid, gently blotted dry, and weighed to establish their initial mass. The bones were then placed in tared crucibles and dried at 105° C for 110 hr to remove all bound and pore bone water. Completion of drying was verified by no change in mass over a 24-hour period. Bones were again weighed and the dry mass was recorded. The bones were then incinerated at 600° C for 30 hr to burn off all organic matrix, and the residual ash was weighed again. Total water mass was then obtained as the difference between initial and dry mass, total matrix mass was the difference between dry and ash mass, and total mineral mass was equal to the ash mass. These masses, divided by total bone volume obtained from the μCT bone exterior mask, yield total water, matrix, and mineral densities.
3. Results
3.1. 1H NMR Spectroscopy
The measured NMR pool fractions by bi-exponential fitting, along with validation measurements by μCT, 2H IR NMR, and gravimetry, are shown in Table 1. Short- T2* fraction at 1.5T was 69.6±12.7% (37.7–82.6%) (mean ± standard deviation and range (min-max)). At 3T, the mean short- T2* fraction was similar to 1.5T at 68.1%, but the standard deviation and range were larger, at 21.9% and 17.0–98.6%, respectively. Short- T2* fractions at 7T and 9.4T were 82.3±14.0% (49.8–99.9%) and 55.1±28.7% (18.6–98.3%).
Table 1.
Bone | 27F | 30F | 37M | 49M | 53M | 53F | 65F | 69M | 74F | 75M | 82F | 83F | 83M | 93M | 97F | Mean | StDev | ||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
μCT Porosity | 3.70 | 3.06 | 4.77 | 4.08 | 5.56 | 3.78 | 5.57 | 5.18 | 20.38 | 4.51 | 17.31 | 33.53 | 4.94 | 5.18 | 12.92 | 8.96 | 8.61 | ||||
| |||||||||||||||||||||
2H IR Bound Fraction (%) | 72.1 | 68.8 | 61.7 | 77.9 | 60.7 | 72.0 | 75.6 | 68.9 | 50.4 | 62.1 | 56.2 | 48.7 | 58.8 | 56.0 | 49.4 | 62.62 | 9.56 | ||||
| |||||||||||||||||||||
Mineral Density (mg/cc) | 1197 | 1208 | 1163 | 1207 | 1159 | 1192 | 1161 | 1172 | 930.7 | 1175 | 1015 | 751.2 | 1159 | 1219 | 1058 | 1118 | 129.6 | ||||
Organic Density (mg/cc) | 525.2 | 520.7 | 494.9 | 514.3 | 513.1 | 527.5 | 513.0 | 511.2 | 472.4 | 510.3 | 479.0 | 437.0 | 520.1 | 515.2 | 501.2 | 503.7 | 24.3 | ||||
Water Density (mg/cc) | 287.4 | 292.3 | 316.2 | 281.4 | 294.6 | 294.8 | 297.3 | 302.7 | 405.0 | 298.8 | 394.5 | 435.6 | 317.2 | 318.7 | 356.2 | 326.2 | 48.4 | ||||
| |||||||||||||||||||||
9.4T Short T1–T2 Fraction (%) | 86.6 | 86.2 | 78.4 | 83.8 | 82.9 | 86.6 | 81.2 | 79.3 | 62.4 | 80.1 | 66.4 | 55.4 | 77.3 | 77.6 | 71.4 | 77.0 | 9.3 | ||||
9.4T Short T2 Fraction (%) | 87.4 | 86.7 | 78.9 | 83.9 | 83.4 | 87.4 | 82.3 | 79.6 | 64.7 | 80.6 | 68.1 | 59.0 | 77.8 | 78.5 | 73.1 | 78.1 | 8.5 | ||||
| |||||||||||||||||||||
1.5T Short T2* Fraction (%) | 82.6 | 82.4 | 72.2 | 77.1 | 75.0 | 80.7 | 75.5 | 72.9 | 50.5 | 75.2 | 74.7 | 37.7 | 58.6 | 68.9 | 59.5 | 69.6 | 12.7 | ||||
3T Short T2* Fraction (%) | 80.0 | 81.0 | 72.5 | 73.2 | 71.2 | 94.5 | 76.9 | 75.2 | 37.8 | 75.6 | 98.6 | 17.0 | 62.9 | 67.2 | 37.2 | 68.1 | 21.9 | ||||
7T Short T2* Fraction (%) | 87.3 | 91.7 | 86.4 | 87.3 | 83.6 | 88.5 | 84.9 | 89.1 | 58.1 | 94.0 | 90.8 | 64.6 | 99.9 | 78.2 | 49.8 | 82.3 | 14.0 | ||||
9.4T Short T2* Fraction (%) | 19.4 | 95.0 | 59.4 | 79.4 | 95.5 | 18.6 | 43.0 | 49.0 | 38.8 | 80.7 | 31.2 | 60.7 | 21.8 | 98.3 | 35.2 | 55.1 | 28.7 | ||||
| |||||||||||||||||||||
1.5T Short T1–T2* Fraction (%) | 82.6 | 81.9 | 71.9 | 76.8 | 75.2 | 81.0 | 76.2 | 73.5 | 51.1 | 75.1 | 64.4 | 41.5 | 64.0 | 69.5 | 62.3 | 69.8 | 11.6 | ||||
3T Short T1–T2* Fraction (%) | 80.3 | 80.5 | 73.7 | 73.2 | 72.0 | 57.1 | 76.9 | 75.9 | 49.1 | 75.2 | 44.9 | 37.3 | 64.5 | 69.0 | 55.1 | 65.6 | 13.7 | ||||
7T Short T1–T2* Fraction (%) | 84.3 | 85.4 | 83.4 | 85.9 | 82.9 | 85.4 | 83.4 | 86.0 | 58.0 | 87.6 | 81.6 | 52.4 | 33.1 | 75.6 | 61.5 | 75.1 | 16.2 | ||||
9.4T Short T1–T2* Fraction (%) | 14.3 | 91.4 | 62.7 | 79.1 | 72.6 | 16.1 | 49.5 | 53.8 | 39.0 | 79.6 | 33.8 | 32.3 | 33.2 | 34.4 | 45.9 | 49.2 | 23.6 | ||||
| |||||||||||||||||||||
1.5T Optimal TI (ms) | 70 | 70 | 80 | 80 | 70 | 60 | 90 | 100 | 110 | 100 | 130 | 120 | 90 | 90 | 110 | 91.3 | 20.3 | ||||
3T Optimal TI (ms) | 50 | 50 | 80 | 60 | 60 | 70 | 80 | 90 | 110 | 80 | 110 | 130 | 100 | 70 | 80 | 81.3 | 23.3 | ||||
7T Optimal TI (ms) | 140 | 180 | 180 | 160 | 140 | 130 | 160 | 200 | 160 | 220 | 190 | 170 | 240 | 160 | 180 | 174 | 30 |
Short- T2* pool relaxation times were 401±119 (301–762) μs, 389±116 (249–768) μs, 368±76 (302–613) μs, and 302±150 (125–615) μs at 1.5T, 3T, 7T, and 9.4T, respectively. Long- T2* relaxation times decreased more dramatically at higher field strength: 4110±1230 (1840–7170) μs, 4350±8570 (980–35300) μs, 1300±420 (380–1860) μs, and 886±525 (410–2160) μs at 1.5T, 3T, 7T, and 9.4T, respectively. Short- and long- T2* relaxation times and short- T2* pool sizes are plotted in Figure 4. If one outlier, 35300 μs, is excluded from this mean and standard deviation at 3T, the long- T2* relaxation time becomes 2150±540 (980–2940) μs. This outlier was a result of poor fitting due to oscillations in the magnitude FID; this phenomenon, which appeared in several specimens at multiple fields, will be discussed in detail in the discussion section.
Addition of the T1 dimension reduced the standard deviations of 2D short- T2* fractions relative to 1D, particularly at 3T, but average pool fractions and T2* relaxation times were unchanged. T1 relaxation times of the short- T2* fractions were 82.6±10.4 (62.2–97.3) ms, 145±25 (103–186) ms, 400±68 (206–496) ms, and 358±240 (93–565) ms at 1.5T, 3T, 7T, and 9.4T, respectively, and T1s of the long- T2* fractions were 651±273 (379–1210) ms, 880±281 (465–1470) ms, 1790±470 (898–2470) ms, and 1300±370 (751–1940) ms.
Short-T2 fractions by bi-component T2 fitting of CPMG echo amplitudes at 9.4T were generally larger than short- T2* fractions by FID fitting: 1D short-T2 fraction was 78.1±8.5% (59.0–87.4%), and 2D short-T2 fraction was 77.0±9.3% (55.4–86.6%). Relaxation times for short- and long-T2 fractions were 540±150 (430–960) μs and 77±53 (22–220) ms, respectively, for 1D fitting. Short-T2 relaxation time for 2D fitting was unchanged at 540±150 (430–980) μs, while long-T2 relaxation time was 55±38 (17–161) ms. The corresponding T1 relaxation times of short- and long-T2 fractions obtained by 2D fitting were 480±80 (320–560) ms and 1210±300 (880–1910) ms, respectively.
Average signal to noise ratios (SNRs), defined as the magnitude of the first data point of the FID signal divided by the standard deviation of the magnitude of the final 25 data points, were 4130, 3040, 15600, and 34300 at 1.5T, 3T, 7T, and 9.4T, respectively. The noise terms, n, in Equations 1 and 2 for 1D and 2D bi-component T2* fitting were on the order of 0.1% of total signal (MS+ML) or less at all field strengths, and were therefore inconsequential. In Equations 3 and 4 for 1D and 2D bi-component T2 fitting of CPMG data, the noise terms, n, were 2–3% of total signal. This larger value of n is a result of the overly simple assumption of two discrete pools; the presence of a small fraction of signal that has T2 much longer than the fitted long-T2 value appears to the bi-component fitting method as a non-zero noise level.
Three example 2D relaxation spectra of a bone specimen taken from a 37 year old male donor are shown in Figure 5: T2*–T2 and T1–T2 spectra at 9.4T, and a T1–T2* spectrum at 3T. Relaxation times given in the labels next to each pool represent the centroid of each pool. T2 values were distributed across three orders of magnitude (horizontal axis in Figure 5a and 5b), while T2* values spanned less than two (vertical axis in Figure 5a and horizontal axis in Figure 5c).
Optimal inversion times were 91.3±20.3 (60–130) ms at 1.5T, 81.3±23.3 (50–130) ms at 3T, and 174±30 (130–240) ms at 7T. These TIs resulted in short- T2* signal fractions of 99.3±0.6% (97.7–100%) at 1.5T, 99.2±0.6% (98.1–100%) at 3T, and 99.7±0.2% (99.3–100%) at 7T.
3.2. 2H NMR Spectroscopy
Deuterium-exchanged 2H IR bound water fraction was 62.6±9.6% (48.7–77.9%). Mean T1 of the bound water pool was 11.2±1.7 (9.8–15.9) ms, and mean T1 of the pore water pool was 197±42 (129–282) ms. The mean inversion-recovery time to achieve nulling of the central pore water peak was 142±31 (92–206) ms. An example plot in Figure 6 shows the fully relaxed, pore water-nulled, and subtracted pore water-only spectra taken from the specimen from a 27 year old female donor.
3.3. μCT and Gravimetry
Porosity measured from μCT image segmentation was 8.96±8.61% (3.06–33.53%). Gravimetric mineral density was 1118±130 (751–1219) mg/cc and organic matrix density was 503.7±24.3 (437.0–527.5) mg/cc, consistent with previous observations in porcine femoral cortical bone by Cao et al. (18,38). Total water density, including both bound and pore water, was 326.2±48.4 (281.4–435.6) mg/cc. A matrix of R2 values for inter-parameter correlations is given in Table 2. Porosity and matrix density were highly negatively correlated (R2 = 0.91), supporting the notion that any increase in porosity occurs at the expense of a loss of matrix volume.
Table 2.
μCT Porosity | Organic Density | 1.5T 1D T2* Short- T2* Fraction | 1.5T 2D T1–T2* Short- T2* Fraction | 3T 1D T2* Short- T2* Fraction | 3T 2D T1–T2* Short- T2* Fraction | 7T 1D T2* Short- T2* Fraction | 7T 2D T1–T2* Short- T2* Fraction | 9.4T 1D T2* Short- T2* Fraction | 9.4T 2D T1–T2* Short- T2* Fraction | 9.4T 1D T2 Short-T2 Fraction | 9.4T 2D T1–T2 Short-T2 Fraction | |
---|---|---|---|---|---|---|---|---|---|---|---|---|
9.4T 2H IR Bound Fraction | 0.50 | 0.46 | 0.61 | 0.68 | 0.46 | 0.51 | 0.40 | 0.40 | 0.00 | 0.05 | 0.67 | 0.66 |
9.4T 2D T1–T2 Short-T2 Fraction | 0.90 | 0.89 | 0.77 | 0.94 | 0.50 | 0.74 | 0.39 | 0.35 | 0.03 | 0.08 | 1.00 | |
9.4T 1D T2 Short-T2 Fraction | 0.87 | 0.88 | 0.77 | 0.94 | 0.49 | 0.72 | 0.37 | 0.35 | 0.02 | 0.07 | ||
9.4T 2D T1–T2* Short- T2* Fraction | 0.10 | 0.02 | 0.08 | 0.08 | 0.01 | 0.24 | 0.04 | 0.14 | 0.55 | |||
9.4T 1D T2* Short- T2* Fraction | 0.02 | 0.00 | 0.02 | 0.02 | 0.00 | 0.12 | 0.00 | 0.11 | ||||
7T 2D T1–T2* Short- T2* Fraction | 0.25 | 0.16 | 0.66 | 0.54 | 0.47 | 0.30 | 0.12 | |||||
7T 1D T2* Short- T2* Fraction | 0.40 | 0.31 | 0.42 | 0.41 | 0.62 | 0.31 | ||||||
3T 2D T1–T2* Short- T2* Fraction | 0.76 | 0.61 | 0.54 | 0.69 | 0.26 | |||||||
3T 1D T2* Short- T2* Fraction | 0.50 | 0.44 | 0.83 | 0.67 | ||||||||
1.5T 2D T1–T2* Short- T2* Fraction | 0.84 | 0.80 | 0.93 | |||||||||
1.5T 1D T2* Short- T2* Fraction | 0.70 | 0.63 | ||||||||||
Organic Density | 0.91 |
1D short- T2* pool fraction was moderately negatively correlated with porosity and positively with matrix density at 1.5T (R2 = 0.70 and 0.63, respectively), but the strengths of these associations diminished rapidly as field strength increased. In fact, at 9.4T, no statistically significant correlation was observed. Scatter plots displaying 1D short- T2* pool fractions versus matrix density are shown in Figure 7. In general, addition of the T1 dimension improved the strengths of these correlations (except at 7T), but this improvement still did not raise the correlations at 9.4T to the level of statistical significance.
Fitted pool fractions by bi-component T2 fitting of CPMG echo amplitudes at 9.4T were generally better correlated with porosity and matrix density than were those derived from T2* fitting of FIDs and 2H IR. Coefficients of determination (R2) of short-T2 fraction by 1D fitting of CPMG echo amplitudes to porosity and matrix density were 0.87 and 0.88, respectively; with the addition of the T1 dimension, these increased slightly to 0.90 and 0.89, respectively. Bound water fraction by 2H IR, however, was only moderately correlated with porosity (R2 = 0.50) and matrix density (R2 = 0.46), but correlations of 2H IR bound water fraction with short- T2* fraction by 1D FID and 2D SR-FID fitting showed the same trend of reduced association as field strength increases.
4. Discussion
Bi-exponential fitting is applicable as long as the as the time constants representative of the two pools are sufficiently separated from one another. Due to the ill-posed nature of the inverse Laplace transform, significant errors may arise in the fitted pool fractions and time constants once the two time constants become similar. Because T2* of pore water is substantially shortened by dephasing due to internal magnetic field gradients arising from the large susceptibility difference between water and bone tissue (∆χv ~ 2.5 ppm SI) (30), resulting in decreased separation of bound and pore water relaxation times, T2* bi-component fitting of FIDs is inferior to T2 fitting of CPMG echo amplitudes (8). As field strength increases, these internal magnetic field gradients increase proportionally, and pore water T2* further decreases, more severely impacting the ability of bi-component fitting of FIDs to distinguish bound and pore water. This is reflected in the reduced strength of the correlations of short- T2* fractions versus porosity and matrix density as field strength increases. While the results are promising at 1.5T, less than half the variance in matrix density is explained by 1D short- T2* fraction at 3T. The phase dispersion resulting from the static internal field inhomogeneity is refocused in the CPMG sequence, yielding greater separation in T2 and improved fitting performance.
The measured short- T2* component fractions of 69.6% and 68.1% at 1.5T and 3T, respectively, are nearly identical to the 68.5% and 69% measured by Li et al. (39) in bovine bone, but are slightly lower than the 74.4% and 75.9% measured in human bone at these same field strengths. The ages and pore volume fractions of the human bones studied by Li et al., however, are unknown. While Li et al. observed decreases in short T2* relaxation time from 450 μs to 320 μs, and in long T2* relaxation time from 7.17 ms to 3.02 ms, at 1.5T compared to 3T, the relaxation times presented here are relatively consistent at these two field strengths. Most importantly, however, no significant differences between 1.5T and 3T short- T2* fraction are observed in the present results or in either of Li’s measurements.
At 3T and higher fields, oscillations appear in the magnitude FIDs of several bone specimens, presumably resulting from non-water off-resonant spins. Similar oscillations have previously been observed in bone (13) and tendon (19). In many cases, the amplitude of these oscillations is sufficient to disrupt the monotonic decrease of the FID, a condition that is not able to be fitted by a sum of monotonically decaying exponentials. An example of such an oscillatory FID at 7T of a bone specimen from a 53-year-old female donor, with attempted mono-exponential and bi-exponential fits, is shown in Figure 8a. These oscillations appear most regularly in bone from old female donors, who have greater porosity. Although care was taken in the preparation of these samples to remove marrow fat, some may remain trapped in the enlarged pore spaces of osteoporotic cortical bone. A small amount of lipid also exists within cortical bone matrix at the cement line surrounding each osteon (40).
Note also that the oscillation in Figure 8 does not appear to have a fixed period; rather, its period varies. If this signal were purely from the methylene resonance of fatty acid triglycerides, a frequency of 1040 Hz would be expected at 7T. It is possible that this oscillation arises from fat that exists in a spatially heterogeneous magnetic field due to susceptibility effects, from other non-fat sources within the same environment, or some combination of these effects (all of which would be refocused by the 180° pulses in CPMG), but further experimentation would be necessary to support any of these speculative hypotheses.
If bone specimens exhibiting these problematic oscillations are retrospectively excluded from correlation analyses (four females ages 53–97 years, one male age 83 years), the R2 values describing the association of short- T2* fraction with μCT porosity and matrix density in the remaining 10 specimens are substantially improved. At 1.5T, 3T, and 7T, R2 = 0.87, 0.93, and 0.86, respectively, for correlations with μCT porosity, and 0.83, 0.78, and 0.61 for correlations with matrix density. At 9.4T, none of the samples yielded data judged suitable for fitting. It must be noted that the specimens removed from this analysis were predominantly from the population to which an MRI study of bone composition would be of greatest benefit, and it is not possible to know in advance of such a study whether the resulting data will be usable.
Short-T2 fraction by 1D bi-component fitting of CPMG echo amplitudes outperforms bi-component fitting of FIDs at all field strengths. This is likely due to the much greater separation of bound and pore water T2 relaxation times. Whereas the two pools are separated by only one order of magnitude in T2*, the difference in T2 is two orders of magnitude. This T2-based method, however, is not applicable to clinical scanners due to SAR limitations and the requirement of extremely short, high-amplitude 180° refocusing pulses. Nevertheless, the method is suitable for independent determination of bound and pore water fractions.
Addition of a second dimension generally improves the stability and accuracy of bi-component fitting (32). In the present case, addition of T1 to T2* by incorporation of saturation-recovery into the sequence improves the strengths of correlations between fitted short- T2* fraction and validation methods, except at 7T. Though it would be time-prohibitive to incorporate saturation-recovery into a spatially-resolved in vivo bi-component T2* scanning protocol, this result suggests that methods which take advantage of differences in both T1 and T2 (or T2*) may be superior to those relying solely on T2*. An example of such an approach is single adiabatic inversion recovery (13,14,28,29,41), which selectively inverts long-T2 pore water while saturating short-T2 bound water. As the longitudinal magnetization of pore water (which was also found to have longer T1 than bound water) passes through its null point, a solid-state imaging readout is performed to selectively image bound water.
The optimal inversion times calculated in this study from SIR-FID experiments are consistent with previous work at 3T by Li et al. (42), but calculation of these results utlizes the same 1D bi-component T2* fitting method that is the main subject of this work. These TI results, therefore, should be interpreted with the same caution as 1D bi-component T2* fitting for calculation of bound and pore water fractions in general. Specifically, the value at 1.5T is well supported by strong correlations between 1D short- T2* fraction and both μCT porosity and gravimetric density, while the value at 3T is somewhat less well supported. Also consistent with Li et al., the long- T2* fraction is very well-suppressed (to less than ~5%) in a range of approximately ±20 ms surrounding the optimal value for each bone, suggesting that the effectiveness of adiabatic inversion nulling of long- T2* signal is not significantly decreased by even moderate deviations from the optimal TI.
Also note that the optimal inversion time of 174 ms at 7T is greater than one half of the TR = 300 ms used in this experiment. The spin dynamics, as predicted by the Bloch equations, preclude the possibility of a signal pool being nulled by inversion in steady-state at a TI greater than half the TR. Primarily on the basis of this observation, and secondarily due to the weak correlations between 7T 1D short- T2* fraction and the two confirmatory measurements, we have no confidence in this calculated optimal TI at 7T.
2H IR was found to be less effective in separating the two water (i.e. deuterium oxide) pools than bi-component T2 fitting of CPMG data. Bound water fraction obtained by this 2H IR method were considerably less strongly correlated with μCT porosity and matrix density than was short-T2 fraction by fitting of CPMG amplitudes. Correlations of short- T2* fraction at 1.5T with these same validation methods were also stronger than those of 2H IR bound water fraction. This unexpectedly inferior performance may be due to a distribution of T1 values within pore D2O, thereby preventing complete nulling of the central pore water peak. These results suggest that bi-component fitting of CPMG data is a more reliable NMR-based method for quantification of bound and pore water fractions than 2H IR.
It is important to keep in mind that bi-component analysis, whether based on T2* or T2, and whether performed in one or two dimensions, involves one major, and incorrect, assumption: the existence of two pools with discrete relaxation times. T2 values are instead distributed continuously over several orders of magnitude. While deviations from this assumption may not have severe consequences at low field, where relaxation times of bound and pore water are well-separated and, thus, well-approximated by this two-pool model, the results are more severely impacted as the relaxation times converge. Non-negative least squares methods, such as the MERA software package (43), do not require an assumption of the number of pools, but demand careful regularization to generate a valid relaxation spectrum, which may have any number of peaks. As bound and pore water relaxation times converge, this method often fails to distinguish multiple pools, instead returning a single peak containing 100% of 1H signal.
The T1 values obtained for the long- T2* fraction by 2D T1–T2* fitting at 7T and 9.4T (1790 ms and 1300 ms, respectively) are greater than 1200 ms and, therefore, greater than TSR/5. If these values are accurate, this may cause the magnetization after TSR = 6 s to be slightly below the equilibrium magnetization. As a consequence, long- T2* longitudinal magnetization at 7T measured by 1D T2* fitting may be attenuated by 3.5%, and the resulting 1D short- T2* fraction may be higher than the actual fraction by approximately 1%. It is unlikely that this effect is responsible for the large (>10%) difference in 1D T2* pool sizes at 7T compared to other field strengths. Also, the more reliable T1-T2 results at 9.4T show a long-T2 fraction T1 value of 1210 ms, which is sufficient for assumption of full longitudinal relaxation (TSR = 5T1).
This study has several limitations. First, we focus only on cortical bone, whereas in vivo examinations will likely target both cortical and trabecular bone. Though we expect our results to be generalizable (the important difference between trabecular and cortical bone being the predominance of pore versus bound water, respectively), this may benefit from further investigation into the effects of marrow fat. Also, we examine bi-component T2 fitting of CPMG echo amplitudes at only one field strength, 9.4T. Although this is a widely available field strength for spectroscopic hardware, it would be instructive to examine whether short-T2 fraction is also sensitive to field strength. Finally, the source of the oscillations in magnitude FIDs remains unclear. Further investigation, possibly by chemical removal of fat or a complementary method to quantify lipid and protein content, would likely yield additional insight into this phenomenon.
5. Conclusion
T2* bi-component fitting for quantification of bound and pore water fractions performs moderately well at 1.5T, but becomes less reliable as field strength increases. It should therefore be used with caution, and other methods for distinguishing between bone water fractions, namely those based on adiabatic inversion-recovery or dual-band saturation should be considered for use in in vivo examinations. For validation of bound and pore water fractions, bi-component fitting of CPMG echo amplitudes is superior to 2H IR, and is on par with μCT as a method of investigating bone porosity within a single specimen.
Acknowledgments
The authors thank Dr. Wei-Ju Tseng for assistance with μCT imaging protocol refinement, and Dr. Steven Szewczyk for facilitating the use of high-temperature furnaces for gravimetric validation.
Sources of Support: This work was supported by NIH R01 AR50068, NIH F31 AG042289 (A.C.S.), and the University of Pennsylvania Institute for Translational Medicine and Therapeutics. This work was also supported in part by the National Center for Research Resources Grant UL1RR024134, which is now the National Center for Advancing Translational Sciences Grant UL1TR000003. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH.
Abbreviations
- μCT
micro-computed tomography
- CPMG
Carr-Purcell-Meiboom-Gill
- DXA
dual-energy x-ray absorptiometry
- QCT
quantitative computed tomography
- TI
inversion time
- Mz
longitudinal magnetization
- RF
radiofrequency
- PTFE
polytetrafluoroethylene
- BBI
broadband inverse
- SR
saturation-recovery
- TSR
saturation-recovery time
- NLLS
non-linear least-squares
- FID
free induction decay
- SAR
specific absorption rate
- SIR
single adiabatic inversion recovery
- TI
inversion time
- D2O
deuterium oxide
- IR
inversion-recovery
- χv
volume magnetic susceptibility
References
- 1.Lindsay R. Osteoporosis. In: Fauci AS, Braunwald E, Kasper DL, Hauser SL, Longo DL, Jameson JL, Loscalzo J, editors. Harrison’s Principles of Internal Medicine. 17. McGraw-Hill Professional; 2008. [Google Scholar]
- 2.Zebaze R, Ghasem-Zadeh A, Bohte A, Iuliano-Burns S, Mackie E, Seeman E. Age-related bone loss: The effect of neglecting intracortical porosity. Bone. 2009;44:S117–S118. doi: 10.1016/j.bone.2009.01.261. [DOI] [Google Scholar]
- 3.Seeman E, Delmas PD. Bone quality–the material and structural basis of bone strength and fragility. N Engl J Med. 2006;354:2250–2261. doi: 10.1056/NEJMra053077. [DOI] [PubMed] [Google Scholar]
- 4.Bousson V, Peyrin F, Bergot C, Hausard M, Sautet A, Laredo J-D. Cortical Bone in the Human Femoral Neck: Three-Dimensional Appearance and Porosity Using Synchrotron Radiation. J Bone Miner Res. 2004;19:794–801. doi: 10.1359/jbmr.040124. [DOI] [PubMed] [Google Scholar]
- 5.Borah B, Dufresne T, Nurre J, Phipps R, Chmielewski P, Wagner L, Lundy M, Bouxsein M, Zebaze R, Seeman E. Risedronate reduces intracortical porosity in women with osteoporosis. J Bone Miner Res. 2009;25:41–47. doi: 10.1359/jbmr.090711. [DOI] [PubMed] [Google Scholar]
- 6.Nuzzo S, Peyrin F, Cloetens P, Baruchel J, Boivin G. Quantification of the degree of mineralization of bone in three dimensions using synchrotron radiation microtomography. Med Phys. 2002;29:2672. doi: 10.1118/1.1513161. [DOI] [PubMed] [Google Scholar]
- 7.Wang XD, Ni QW. Determination of cortical bone porosity and pore size distribution using a low field pulsed NMR approach. Journal of Orthopaedic Research. 2003;21:312–319. doi: 10.1016/S0736-0266(02)00157-2. [DOI] [PubMed] [Google Scholar]
- 8.Horch RA, Nyman JS, Gochberg DF, Dortch RD, Does MD. Characterization of 1H NMR signal in human cortical bone for magnetic resonance imaging. Magn Reson Med. 2010;64:680–687. doi: 10.1002/mrm.22459. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Ong HH, Wright AC, Wehrli FW. Deuterium nuclear magnetic resonance unambiguously quantifies pore and collagen-bound water in cortical bone. J Bone Miner Res. 2012;27:2573–2581. doi: 10.1002/jbmr.1709. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Ni Q, Nyman JS, Wang X, Santos ADL, Nicolella DP. Assessment of water distribution changes in human cortical bone by nuclear magnetic resonance. Meas Sci Technol. 2007;18:715–723. doi: 10.1088/0957-0233/18/3/022. [DOI] [Google Scholar]
- 11.Wilson EE, Awonusi A, Morris MD, Kohn DH, Tecklenburg MMJ, Beck LW. Three Structural Roles for Water in Bone Observed by Solid-State NMR. Biophysical Journal. 2006;90:3722–3731. doi: 10.1529/biophysj.105.070243. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Biswas R, Bae W, Diaz E, Masuda K, Chung CB, Bydder GM, Du J. Ultrashort echo time (UTE) imaging with bi-component analysis: Bound and free water evaluation of bovine cortical bone subject to sequential drying. Bone. 2012;50:749–755. doi: 10.1016/j.bone.2011.11.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Horch RA, Gochberg DF, Nyman JS, Does MD. Clinically compatible MRI strategies for discriminating bound and pore water in cortical bone. Magn Reson Med. 2012;68:1774–1784. doi: 10.1002/mrm.24186. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Manhard MK, Horch RA, Harkins KD, Gochberg DF, Nyman JS, Does MD. Validation of quantitative bound- and pore-water imaging in cortical bone. Magn Reson Med. 2014;71:2166–2171. doi: 10.1002/mrm.24870. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Nyman JS, Ni Q, Nicolella DP, Wang X. Measurements of mobile and bound water by nuclear magnetic resonance correlate with mechanical properties of bone. Bone. 2008;42:193–199. doi: 10.1016/j.bone.2007.09.049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Li C, Seifert AC, Rad HS, Bhagat YA, Rajapakse CS, Sun W, Lam SCB, Wehrli FW. Cortical Bone Water Concentration: Dependence of MR Imaging Measures on Age and Pore Volume Fraction. RY. 2014;272:796–806. doi: 10.1148/radiol.14132585. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Bae WC, Patil S, Biswas R, Li S, Chang EY, Statum S, D’Lima DD, Chung CB, Du J. Magnetic resonance imaging assessed cortical porosity is highly correlated with μCT porosity. Bone. 2014;66:1–6. doi: 10.1016/j.bone.2014.06.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Cao H, Ackerman JL, Hrovat MI, Graham L, Glimcher MJ, Wu Y. Quantitative bone matrix density measurement by water- and fat-suppressed proton projection MRI (WASPI) with polymer calibration phantoms. Magn Reson Med. 2008;60:1433–1443. doi: 10.1002/mrm.21771. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Diaz E, Chung CB, Bae WC, Statum S, Znamirowski R, Bydder GM, Du J. Ultrashort echo time spectroscopic imaging (UTESI): an efficient method for quantifying bound and free water. NMR Biomed. 2011;25:161–168. doi: 10.1002/nbm.1728. [DOI] [PubMed] [Google Scholar]
- 20.Du J, Hermida JC, Diaz E, Corbeil J, Znamirowski R, D’Lima DD, Bydder GM. Assessment of cortical bone with clinical and ultrashort echo time sequences. Magn Reson Med. 2012 doi: 10.1002/mrm.24497. [DOI] [PubMed] [Google Scholar]
- 21.Wu Y, Ackerman JL, Chesler DA, Graham L, Wang Y, Glimcher MJ. Density of organic matrix of native mineralized bone measured by water- and fat-suppressed proton projection MRI. Magn Reson Med. 2003;50:59–68. doi: 10.1002/mrm.10512. [DOI] [PubMed] [Google Scholar]
- 22.Bae WC, Chen PC, Chung CB, Masuda K, D’Lima D, Du J. Quantitative ultrashort echo time (UTE) MRI of human cortical bone: Correlation with porosity and biomechanical properties. J Bone Miner Res. 2012;27:848–857. doi: 10.1002/jbmr.1535. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Fernández-Seara MA, Wehrli SL, Takahashi M, Wehrli FW. Water Content Measured by Proton-Deuteron Exchange NMR Predicts Bone Mineral Density and Mechanical Properties. J Bone Miner Res. 2003;19:289–296. doi: 10.1359/JBMR.0301227. [DOI] [PubMed] [Google Scholar]
- 24.Wehrli FW, Fernández-Seara MA. Nuclear Magnetic Resonance Studies of Bone Water. Ann Biomed Eng. 2005;33:79–86. doi: 10.1007/s10439-005-8965-8. [DOI] [PubMed] [Google Scholar]
- 25.Techawiboonwong A, Song HK, Leonard MB, Wehrli FW. Cortical bone water: in vivo quantification with ultrashort echo-time MR imaging. RY. 2008;248:824–833. doi: 10.1148/radiol.2482071995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Rad HS, Lam SCB, Magland JF, Ong H, Li C, Song HK, Love J, Wehrli FW. Quantifying cortical bone water in vivo by three-dimensional ultra-short echo-time MRI. NMR Biomed. 2011;24:855–864. doi: 10.1002/nbm.1631. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Li C, Magland JF, Rad HS, Song HK, Wehrli FW. Comparison of optimized soft-tissue suppression schemes for ultrashort echo time MRI. Magn Reson Med. 2011;68:680–689. doi: 10.1002/mrm.23267. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Larson PEZ, Conolly SM, Pauly JM, Nishimura DG. Using adiabatic inversion pulses for long-T2 suppression in ultrashort echo time (UTE) imaging. Magn Reson Med. 2007;58:952–961. doi: 10.1002/mrm.21341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Du J, Carl M, Bydder M, Takahashi A, Chung CB, Bydder GM. Journal of Magnetic Resonance. Journal of Magnetic Resonance. 2010;207:304–311. doi: 10.1016/j.jmr.2010.09.013. [DOI] [PubMed] [Google Scholar]
- 30.Hopkins JA, Wehrli FW. Magnetic susceptibility measurement of insoluble solids by NMR: magnetic susceptibility of bone. Magn Reson Med. 1997;37:494–500. doi: 10.1002/mrm.1910370404. [DOI] [PubMed] [Google Scholar]
- 31.Epstein CL, Schotland J. The Bad Truth about Laplace’s Transform. SIAM Rev. 2008;50:504–520. doi: 10.1137/060657273. [DOI] [Google Scholar]
- 32.Celik H, Bouhrara M, Reiter DA, Fishbein KW, Spencer RG. Stabilization of the inverse Laplace transform of multiexponential decay through introduction of a second dimension. J Magn Reson. 2013;236:134–139. doi: 10.1016/j.jmr.2013.07.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Snaar JEM, Van As H. Probing water compartments and membrane permeability in plant cells by 1H NMR relaxation measurements. Biophysical Journal. 1992;63:1654–1658. doi: 10.1016/S0006-3495(92)81741-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Navon G, Eliav U, Demco DE, Blümich B. Study of order and dynamic processes in tendon by NMR and MRI. J Magn Reson Imaging. 2007;25:362–380. doi: 10.1002/jmri.20856. [DOI] [PubMed] [Google Scholar]
- 35.Shinar H, Navon G. Multinuclear NMR and microscopic MRI studies of the articular cartilage nanostructure. NMR Biomed. 2006;19:877–893. doi: 10.1002/nbm.1068. [DOI] [PubMed] [Google Scholar]
- 36.Eliav U, Navon G. Multiple Quantum Filtered NMR Studies of the Interaction between Collagen and Water in the Tendon. J Am Chem Soc. 2002;124:3125–3132. doi: 10.1021/ja011791n. [DOI] [PubMed] [Google Scholar]
- 37.Yushkevich PA, Piven J, Hazlett HC, Smith RG, Ho S, Gee JC, Gerig G. User-guided 3D active contour segmentation of anatomical structures: Significantly improved efficiency and reliability. NeuroImage. 2006;31:1116–1128. doi: 10.1016/j.neuroimage.2006.01.015. [DOI] [PubMed] [Google Scholar]
- 38.Cao H, Nazarian A, Ackerman JL, Snyder BD, Rosenberg AE, Nazarian RM, Hrovat MI, Dai G, Mintzopoulos D, Wu Y. Quantitative 31P NMR spectroscopy and 1H MRI measurements of bone mineral and matrix density differentiate metabolic bone diseases in rat models. Bone. 2010;46:1582–1590. doi: 10.1016/j.bone.2010.02.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Li S, Chang EY, Bae WC, Chung CB, Gao S, Bao S, Bydder GM, Hua Y, Du J. Ultrashort echo time bi-component analysis of cortical bone-a field dependence study. Magn Reson Med. 2013 doi: 10.1002/mrm.24769. [DOI] [PubMed] [Google Scholar]
- 40.Cowin SC. Bone poroelasticity. J Biomech. 1999;32:217–238. doi: 10.1016/s0021-9290(98)00161-4. [DOI] [PubMed] [Google Scholar]
- 41.Reichert ILH, Robson MD, Gatehouse PD, He T, Chappell KE, Holmes J, Girgis S, Bydder GM. Magnetic resonance imaging of cortical bone with ultrashort TE pulse sequences. Magnetic Resonance Imaging. 2005;23:611–618. doi: 10.1016/j.mri.2005.02.017. [DOI] [PubMed] [Google Scholar]
- 42.Li S, Ma L, Chang EY, Shao H, Chen J, Chung CB, Bydder GM, Du J. Effects of inversion time on inversion recovery prepared ultrashort echo time (IR-UTE) imaging of bound and pore water in cortical bone. NMR Biomed. 2014 doi: 10.1002/nbm.3228. n/a–n/a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Does MD. Multi-Exponential Relaxation Analysis (MERA) Toolbox [Google Scholar]