Hybrid monopole/loop coil array for human head MR imaging at 7T

Xinqiang Yan 1,2,3, Long Wei 1,2, Rong Xue 3,4, Xiaoliang Zhang 5,6
PMCID: PMC4479412  NIHMSID: NIHMS669016  PMID: 26120252

Abstract

The monopole coil and loop coil have orthogonal radiofrequency (RF) fields and thus are intrinsically decoupled electromagnetically if they are laid out appropriately. In this study, we proposed a hybrid monopole/loop technique which could combine the advantages of both loop arrays and monopole arrays. To investigate this technique, a hybrid RF coil array containing 4 monopole channels and 4 loop channels was developed for human head MR imaging at 7T. In vivo MR imaging and g-factor results using monopole-only channels, loop-only channels and all channels of the hybrid array were acquired and evaluated. Compared with the monopole-only and loop-only channels, the proposed hybrid array has higher SNR and better parallel imaging performance. Sufficient electromagnetic decoupling and diverse RF magnetic field (B1) distributions of monopole channels and loop channels may contribute to this performance improvement. From experimental results, the hybrid monopole/loop array has low g-factor and excellent SNR at both periphery and center of the brain, which is valuable for human head imaging at ultrahigh fields.

Keywords: Hybrid, monopole, loop, decouple, RF coil, array, parallel imaging, ultrahigh field MRI

Introduction

Ultrahigh field (i.e., 7T and higher) MRI offers significant advantages of higher signal-to-noise ratio (SNR), better imaging contrast and higher spatial resolution [13]. However, radiofrequency (RF) design challenges associated with ultrahigh fields, such as enlarged electromagnetic coupling among the resonant elements and limited field diversity of array channel, are augmented. A variety of RF coils arrays for ultrahigh field MR applications, e.g., L/C loop array [48], microstrip transmission line (MTL) array [914], monopole array [15] and dipole array [16], have been proposed and applied for human head imaging at the ultrahigh field of 7T.

The previous work of hybrid monopole/loop dual-tuned coil designs has demonstrated that monopole and loop have two orthogonal and intrinsically decoupled RF fields [17] at two different frequencies. If this decoupling property still holds at the same frequency, it will be much useful to build a mono-nuclear array with both monopole channels and loop channels for proton MR imaging. Owing to the increased number of independent receive channels, the hybrid array should promise better coil performance over monopole-only and loop-only arrays in terms of SNR and parallel imaging capability [1821]. In designing parallel imaging arrays, it is desirable that each element has a unique B1 field. It has been demonstrated that the monopole and loop have different B1 profile owing to different coil structures and current distributions, which might further improve the g-factors and thus the parallel imaging performance of the hybrid monopole/loop array.

In this work, we proposed a hybrid monopole/loop array design technique for proton parallel MR imaging applications. This technique was validated by implementing a RF array containing 4 hybrid coil elements for human head imaging at 7T. Each hybrid coil element or block has a monopole channel and a loop channel, which are intrinsically decoupled. Bench tests and in vivo MR images of the hybrid monopole/loop array were obtained to demonstrate its feasibility and performance in ultrahigh field MRI. To investigate the benefits of the proposed technique, MR imaging, SNR and g-factor results of the hybrid array were also compared with those from monopole-only channels and loop-only channels.

Materials and methods

a) Construction of the 8-channel monopole/loop hybrid array

A volume-typed monopole/loop hybrid array containing 4 independent blocks was designed and constructed for human head imaging at 7T, as shown in Fig. 1. Each block consists of a monopole channel and a loop channel, and the total channel number of this volume array is 8. The monopole was positioned across the center of the loop to ensure that the two have orthogonal RF fields and thus are intrinsically decoupled. Mi and Li (i=1, 2, 3 and 4) in Fig. 1 represent the element numbering of monopoles and loops, respectively.

Fig. 1.

Fig. 1

Diagram of the monopole/loop array containing 4 hybrid elements. Monopoles were placed across the center of the loops, resulting in intrinsic decoupling performance between them. Mi and Li (i=1, 2, 3 and 4) represent the element numbering of monopoles and loops, respectively.

An acrylic tube with dimensions of 24 cm ID, 25 cm OD and 26 cm in length was used as the mechanical support for this coil array. Four monopoles (25 cm in length) and four loops (9 cm×9 cm in dimension) were equally spaced along the circumference of the acyclic tube. An acrylic plate (40 cm × 40 cm in dimension) with upper corners cut used as the former for the ground, which is perpendicular to the acrylic tube. As shown in Fig. 2, trimmer capacitors Ct_m, Cm_m, Ct_L and Cm_L (Voltronics Corp., Denville, NJ) were used for tuning and matching for the monopole channel and loop channel [22]. Each loop has three fixed capacitors (3.3 pF, American Technical Ceramics Corp., Huntington Station, NY) distributed along the conductors. The chosen dimensions of loops ensure the coupling between neighboring blocks can be ignored and meanwhile keep the imaging coverage acceptable. Monopoles and loops were made of 10-mm wide and 5-mm wide copper tapes, respectively. The ground was made from adhesive-backed copper foil (3M, St, Paul, MN).

Fig. 2.

Fig. 2

Equivalent circuit of one hybrid element which consists of a monopole and a loop. Cm_m and Ct_m were used for the matching and tuning of the monopole; while Cm_L and Ct_L were used for the matching and tuning of the loop. Three fixed capacitors (3.3 pF) were distributed along the loop conductor to avoid the value of tuning capacitor (Ct_L) is too small.

All coil elements or channels were matched to 50 and tuned to 297.2 MHz, which is the Larmor frequency of the utilized 7T MRI system. RF traps were placed between cables and loop channels to avoid possible “cable resonance” at the high frequency of 297.2 MHz. It is worth noting that RF traps are not necessary for monopoles due to their instinct unbalanced structure.

b) Bench measurement and MR imaging experiment

Scattering (S-) parameters including reflection coefficient (S11) and transmission coefficient (S21) of the hybrid coil array were measured with a calibrated network analyzer (5071C, Agilent Technologies Inc., CA, USA). The S-parameter results were recorded when the coils were loaded with human head. S11 measurement was also used to calculate the Q values.

In vivo human head MR images were obtained from a healthy human volunteer using the 8-channel hybrid coil array. The human MRI experimental protocol was approved by the local Institutional Review Board (IRB), and the volunteer signed an informed consent form before the experiment. In the experiment, scans were conducted two times using 4 loop channels and 4 monopole channels separately. In each scan, 4 channels (monopole or loop) were operated by using an RF interface box and all unused channels were terminated with 50-Ω loads. The RF interface box is composed of a 4-way power splitter, phase shifters, T/R switches and 50-Ω preamplifiers. The phase difference between adjacent coil elements was 90 degree. During the two scans, the positioning of the subject was carefully kept unchanged via the line markers made on the patient table.

The sequence used for MR imaging is gradient recalled echo (GRE). The imaging parameters were: flip angle (FA)=25 degree, TR=100 ms, TE=10 ms, FOV=200×200 mm2, matrix=256×256, slice thickness=2 mm, bandwidth=320 Hz/pixel, number of excitation (NEX)=1. All imaging data were acquired on a 7T whole-body MRI scanner (MAGNETOM, Siemens Healthcare, Erlangen, Germany). To demonstrate the parallel imaging capability, g-factor maps and average g-factors in the sagittal plane were measured and calculated with reduction factor (R) of 2, 3 and 4. The g-factor maps and average g-factors were calculated using an RF coil array design and the analysis software Musaik (Speag, Switzerland).

Results

a) Bench test results

Fig. 3 shows the S-parameter matrix of the 8-channel hybrid array loaded with human head. Reflection coefficient (S11) of each coil element was better than −25 dB, indicating excellent matching performance. Average isolation between a loop and a monopole from the same block is −20.1 dB, which verifies their intrinsic decoupling performance. Average isolation between a loop and a monopole from neighboring blocks is −15.2 dB; and average isolation of neighboring loops and neighboring monopoles are −13.7 dB and −13.4 dB, respectively. These results demonstrates that the isolation between adjacent blocks is still acceptable even no decoupling methods are employed. The average unloaded and loaded Q values of loop channels are 124 and 34, respectively; whereas the average unloaded and loaded Q values of the monopole channels are 9.4 and 3.1, respectively.

Fig. 3.

Fig. 3

S-parameter matrix of the 8-channel hybrid monopole/loop array loaded with a healthy human head. The worst-case isolation between any two elements is −13.1 dB, as shown in red arrow. The element numbering here is corresponds that shown in Fig. 1.

b) MR imaging results

Fig. 4 shows the combined images in different axial planes from all loop channels, all monopole channels and their combination. These images were reconstructed from raw data without using any image processing method, and were shown with the same signal intensity scale. The combined method is rooted-sum-of-squares (RSS). Axial images from loop channels have high signal intensities at the periphery, but these intensities decrease fast with depth. Thus loops face difficulty when imaging the area located deeply in human tissues, e.g. the center of brain. On the contrary, the intensities of images from monopole channels had higher signal intensity at central area, which is consistent with previous work [15]. These results indicated that combination of the loop channels and monopole channels, i.e., the hybrid coil array, could provide high signal sensitivities at both the peripheral and central areas. Local SNR at the four peripheral areas and the central area of human brain were calculated and marked in red color in Fig. 4. Compared with the loop-only array, the hybrid array had overall SNR gains of 24%~26% and central SNR gains of 89%~108% for different slices.

Fig. 4.

Fig. 4

Combined axial images from all loop channels (left column), all monopole channels (middle column) and their combination (right column). These images were combined with RSS method and shown in the same signal intensity scale. Local SNR at peripheral and central areas of the brain were marked in red color in these images. As expected, loop channels have high SNR at the periphery but quite low SNR at the center. With the help of monopole channels, the hybrid array has almost double SNR at the center of brain compared with the loop-only array.

G-factor maps with R of 2, 3, and 4 were calculated using the images from loop channels, monopole channels and all channels, as shown in Fig. 5. Average g-factors were also calculated and marked in white color in Fig. 5. At the high accelerate factor of 4, the average g-factor using monopole channels, loop channels and all channels were 1.52, 1.65 and 1.23, respectively. These results demonstrated that the parallel imaging performance could be largely improved when using the hybrid design, i.e., both the monopole channels and loop channels were utilized. Fig. 6 shows the original and rate-4 accelerated images of human head in the sagittal plane. The accelerated images were reconstructed with generalized auto calibrating partially parallel acquisitions (GRAPPA) using the PULSAR toolbox [23]. In the reconstruction, 32 auto-calibration signal (ACS) lines were used. From Fig. 6 we found that excellent quality images still could be obtained even reduction factor achieves as high as 4.

Fig. 5.

Fig. 5

G-factor maps (R=2, 3 and 4) calculated using the images from loop channels, monopole channels and all channels in the sagittal plane. Average g-factors are marked in white color in the g-factor maps. The average g-factors of loop channels, monopole channels and all channels with R of 4 are 1.52, 1.65, and 1.23, respectively. These results demonstrates that parallel imaging performance could be highly improved by using the hybrid monopole/loop array.

Fig. 6.

Fig. 6

GRE images of human head in the sagittal plane using the 8-channel hybrid array: (a) fully encoded image without acceleration; (b) accelerated image with rate-4 GRAPPA acquisition.

Discussions and conclusion

Both L/C loop and monopole techniques have been successfully applied for human head RF coil array designs at ultrahigh fields. When the monopole was placed across the center of the loop, the two coils have orthogonal RF fields and thus were instinct EM decoupled. Based on this feature, we proposed a hybrid monopole/loop array concept in the study. This concept was realized by designing and implementing a head coil array containing 4 hybrid monopole/loop blocks at 7T. As expected, the coupling between the monopole and loop in the same block was better than −19 dB at the frequency of 297.2 MHz, indicating their instinct decoupling performance.

Compared with the monopole-only and loop-only arrays, the hybrid array provides improved SNR gain and better parallel imaging performance. As demonstrated in previous work [15] and this study, the loop array has higher SNR at the peripheral area while the monopole array have higher SNR at the central area. Hence the hybrid array with them together promises complementary advantage to provide excellent SNR at the entire brain. With the hybrid technique, coil element numbers or channel numbers could be doubled without adding additional decoupling treatments, leading to reduced g-factors (Fig. 5) and thus superior imaging quality in fast imaging. Sufficient electromagnetic decoupling and different B1 distribution between monopole element and loop elements should be another important factor to contribute to the improvement of SNR and parallel imaging performance. Beside the benefits of SNR and parallel imaging, the hybrid monopole/loop technique is also possible for building phased array coils with quadrature capability, which could potentially increase the transmit efficiency by reducing the excitation power [24, 25]. Furthermore, if the proposed hybrid array technique is used for RF shimming and parallel transmit (pTx) applications, the hybrid arrays promise additional degrees of freedom for the RF shimming or pTx due to increased independent transmit elements [26, 27], which could be advantageous to provide more homogeneous transmit field.

Due to the limitation of our RF interface, monopole channels and loop channels of the hybrid array were driven separately for imaging acquisition in this study, but they can be used simultaneously given that the coupling among any two channels was low. Note that the transmit field using only monopole channels or loop channels is not the same as that using all channels simultaneously. Therefore images obtained in the study might be different from those acquired using all channels simultaneously. In addition, the proposed hybrid array could be operated in other ways that, for example, loop-only or monopole-only channels are used for transmit while all channels are used for receive. Although the hybrid monopole/loop array in this study only contains 4 blocks, the number of block numbers could be extended to 8 or even more if appropriate decoupling schemes, e.g., individual shielding [28] and magnetic wall decoupling [22, 2933], are applied to ensure sufficient decoupling between the blocks. It has been demonstrated that magnetic wall decoupling is capable of decoupling both loop and monopole arrays [22, 32], and thus it should a promising approach to addressing the coupling issue of the proposed hybrid arrays.

Acknowledgments

This work was supported in part by the Chinese National Major Scientific Equipment R&D Project (Grant No. ZDYZ2010-2), the Ministry of Science and Technology of China Grant (2012CB825500), the National Natural Science Foundation of China Grant (51228702), the Chinese Academy of Sciences Grants (XDB02010001, XDB02050001), and National Institutes of Health (NIH) R01EB008699.

References

  • 1.Yacoub E, Shmuel A, Pfeuffer J, Van De Moortele PF, Adriany G, Andersen P, Vaughan JT, Merkle H, Ugurbil K, Hu X. Magn Reson Med. 2001;45:588–594. doi: 10.1002/mrm.1080. [DOI] [PubMed] [Google Scholar]
  • 2.Vaughan JT, Garwood M, Collins CM, Liu W, DelaBarre L, Adriany G, Andersen P, Merkle H, Goebel R, Smith MB, Ugurbil K. Magn Reson Med. 2001;24:24–30. doi: 10.1002/mrm.1156. [DOI] [PubMed] [Google Scholar]
  • 3.Zhang X, Ugurbil K, Chen W. Magn Reson Med. 2001;46:443–450. doi: 10.1002/mrm.1212. [DOI] [PubMed] [Google Scholar]
  • 4.Pinkerton RG, Barberi EA, Menon RS. Magn Reson Med. 2005;54:499–503. doi: 10.1002/mrm.20583. [DOI] [PubMed] [Google Scholar]
  • 5.Wiggins GC, Potthast A, Triantafyllou C, Wiggins CJ, Wald LL. Magn Reson Med. 2005;54:235–240. doi: 10.1002/mrm.20547. [DOI] [PubMed] [Google Scholar]
  • 6.Alagappan V, Nistler J, Adalsteinsson E, Setsompop K, Fontius U, Zelinski A, Vester M, Wiggins GC, Hebrank F, Renz W, Schmitt F, Wald LL. Magn Reson Med. 2007;57:1148–1158. doi: 10.1002/mrm.21247. [DOI] [PubMed] [Google Scholar]
  • 7.Avdievich NI. Appl Magn Reson. 2011;41:483–506. doi: 10.1007/s00723-011-0280-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Yan X, Ma C, Shi L, Zhuo Y, Zhou XJ, Wei L, Xue R. Appl Magn Reson. 2014;45:437–449. [Google Scholar]
  • 9.Lee RF, Westgate CR, Weiss RG, Newman DC, Bottomley PA. Magn Reson Med. 2001;45:673–683. doi: 10.1002/mrm.1090. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Adriany G, Van de Moortele PF, Wiesinger F, Moeller S, Strupp JP, Andersen P, Snyder C, Zhang X, Chen W, Pruessmann KP, Boesiger P, Vaughan T, Ugurbil K. Magn Reson Med. 2005;53:434–445. doi: 10.1002/mrm.20321. [DOI] [PubMed] [Google Scholar]
  • 11.Wu B, Zhang X, Qu P, Shen GX. J Magn Reson. 2006;182:126–132. doi: 10.1016/j.jmr.2006.04.013. [DOI] [PubMed] [Google Scholar]
  • 12.Wu B, Zhang X, Qu P, Shen GX. Magn Reson Imaging. 2007;25:418–424. doi: 10.1016/j.mri.2006.09.031. [DOI] [PubMed] [Google Scholar]
  • 13.Adriany G, De Moortele PFV, Ritter J, Moeller S, Auerbach EJ, Akgun C, Snyder CJ, Vaughan T, Ugurbill K. Magn Reson Med. 2008;59:590–597. doi: 10.1002/mrm.21488. [DOI] [PubMed] [Google Scholar]
  • 14.Adriany G, Auerbach EJ, Snyder CJ, Gozubuyuk A, Moeller S, Ritter J, Van de Moortele PF, Vaughan T, Ugurbil K. Magn Reson Med. 2010;63:1478–1485. doi: 10.1002/mrm.22413. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Hong SM, Park JH, Woo MK, Kim YB, Cho ZH. Magn Reson Med. 2014;71:1944–1952. doi: 10.1002/mrm.24844. [DOI] [PubMed] [Google Scholar]
  • 16.Raaijmakers AJ, Ipek O, Klomp DW, Possanzini C, Harvey PR, Lagendijk JJ, van den Berg CA. Magn Reson Med. 2011;66:1488–1497. doi: 10.1002/mrm.22886. [DOI] [PubMed] [Google Scholar]
  • 17.Yan X, Xue R, Zhang X. Quant Imaging Mag Surg. 2014;4(4):225–231. doi: 10.3978/j.issn.2223-4292.2014.08.03. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Sodickson DK, Manning WJ. Magn Reson Med. 1997;38:591–603. doi: 10.1002/mrm.1910380414. [DOI] [PubMed] [Google Scholar]
  • 19.Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. Magn Reson Med. 1999;42:952–962. [PubMed] [Google Scholar]
  • 20.Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A. Magn Reson Med. 2002;47:1202–1210. doi: 10.1002/mrm.10171. [DOI] [PubMed] [Google Scholar]
  • 21.Sodickson DK, McKenzie CA, Ohliger MA, Yeh EN, Price MD. Magma. 2002;13:158–163. doi: 10.1007/BF02678591. [DOI] [PubMed] [Google Scholar]
  • 22.Yan X, Zhang X, Wei L, Xue R. Quant Imaging Mag Surg. 2014;4(2):79–86. doi: 10.3978/j.issn.2223-4292.2014.04.10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Ji JX, Son JB, Rane SD. Concept Magn Reson B. 2007;31B:24–36. [Google Scholar]
  • 24.Glover GH, Hayes CE, Pelc NJ, Edelstein WA, Mueller OM, Hart HR, Hardy CJ, Odonnell M, Barber WD. J Magn Reson. 1985;64:255–270. [Google Scholar]
  • 25.Li Y, Yu BY, Pang Y, Vigneron DB, Zhang XL. Plos One. 2013;8(11):e80428. doi: 10.1371/journal.pone.0080428. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Lakshmanan K, Cloos RLM, Sodickson D, Wiggins G. Proceedings of the 22nd annual meeting of ISMRM; Milan. 2014. p. 397. [Google Scholar]
  • 27.Paska1 J, Leussler C. Proceedings of the 16th annual meeting of ISMRM; Toronto. 2008. p. 149. [Google Scholar]
  • 28.Gilbert KM, Curtis AT, Gati JS, Klassen LM, Villemaire LE, Menon RS. Magnet Reson Med. 2010;64:1640–1651. doi: 10.1002/mrm.22574. [DOI] [PubMed] [Google Scholar]
  • 29.Xie Z, Zhang X. Proceedings of the 16th annual meeting of ISMRM; Toronto. 2008. p. 1068. [Google Scholar]
  • 30.Xie Z, Zhang X. Proceedings of the 16th annual meeting of ISMRM; Toronto. 2008. p. 2972. [Google Scholar]
  • 31.Li Y, Xie Z, Pang Y, Vigneron D, Zhang X. Med Phys. 2011;38:4086–4093. doi: 10.1118/1.3598112. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Yan X, Zhang X, Feng B, Ma C, Wei L, Xue R. IEEE Trans Med Imaging. 2014;33:1981–1987. doi: 10.1109/TMI.2014.2313879. [DOI] [PubMed] [Google Scholar]
  • 33.Yan X, Zhang X, Wei L, Xue R. Appl Magn Reson. 2014 doi: 10.1007/s00723-014-0612-9. [Epub ahead of print] [DOI] [Google Scholar]

RESOURCES