Abstract
Background
In recent years there has been an improvement of zirconia ceramic materials to replace posterior missing teeth. To date little in vitro studies has been carried out on the fracture resistance of zirconia veneered posterior fixed dental prostheses. This study investigated the fracture resistance and the failure mode of 3-unit zirconia-based posterior fixed dental prostheses fabricated with two CAD/CAM systems.
Material and Methods
Twenty posterior fixed dental prostheses were studied. Samples were randomly divided into two groups (n=10 each) according to the zirconia ceramic analyzed: Lava and Procera. Specimens were loaded until fracture under static load. Data were analyzed using Wilcoxon´s rank sum test and Wilcoxon´s signed-rank test (P<0.05).
Results
Partial fracture of the veneering porcelain occurred in 100% of the samples. Within each group, significant differences were shown between the veneering and the framework fracture resistance (P=0.002). The failure occurred in the connector cervical area in 80% of the cases.
Conclusions
All fracture load values of the zirconia frameworks could be considered clinically acceptable. The connector area is the weak point of the restorations.
Key words:Fixed dental prostheses, zirconium-dioxide, zirconia, fracture resistance, failure mode.
Introduction
The porcelain fused to metal represents the gold standard technique for posterior fixed dental prostheses (FDPs). However, with the increasing demand in esthetic restorations, and the introduction of computer-aided design/computer-aided manufacturing (CAD-CAM) technology in dentistry, in the last years the use of ceramic restorations has been increased (1). Therefore, during the last decade yttrium oxide partially stabilized zirconia (Y-TZP) ceramic is used in dentistry for heavily loaded restoration, showing in vitro and clinical studies promising results (2,3). Zirconia restorations have good fracture resistance but are highly opaque. Therefore, to obtain a natural-looking and improve esthetics the framework should be veneered with porcelain (1), more over, some manufacturers make provision for zirconia colored cores in order to enhance esthetic outcomes (2). It is possible to manufacture the zirconia core in two different ways: milling a fully sintered piece of zirconia, or milling a partially sintered zirconia block and completing the sintering thereafter. Zirconia frameworks sintered after milling has better mechanical properties than densely sintered zirconia (4), but a 20% shrinkage must be allowed to obtain an optimal fit of the frameworks.
Although the good mechanical properties of zirconia allow the construction of posterior FDPs, the core-veneer interface is one of the weakest aspect of these restorations (2,5) so that, delamination or chipping of the veneering porcelain has been described as the most frequent reason for the failure of zirconia FDPs (3,6). It is important to understand how the veneer and framework material interact in a multi-layer configuration, so the study of veneered structures can provide information about the fracture resistance of each component and the failure mode and origin (1).
The aim of the present study was to compare the fracture resistance (FR) of two zirconia CAD/CAM systems with their corresponding porcelains and the failure mode of 3-unit zirconia posterior fixed dental prostheses (FDPs) with an intermediate pontic. The null hypothesis was, that no differences would be found in the fracture resistance and the failure mode between the zirconia systems.
Material and Methods
Twenty standardized specimens with 2 abutments and screwed onto a platform (30 mm in length, 17 mm in width, and 4.5 mm in thickness) were prepared from stainless steel to receive posterior 3-unit FDPs with an intermediate pontic (spanning the first premolar to the first molar). The abutments were prepared with 5 mm in height, a 1 mm wide chamfer, and a 6 degrees angle of convergence of the axial walls. The specimens were used as working dies and randomly divided in 2 groups (n=10 each). Two zirconia materials and two veneering porcelain were used: Group 1 (L) Lava All-ceramic System and Lava Ceram (respectively) (3M ESPE, Seefeld, Germany) and Group 2 (P): Nobel Procera Zirconia and NobelRondo (respectively) (Nobel Biocare, Zurich, Switzerland). The size of the connector area was 3 mm x 3 mm and the frameworks thickness was 0.5 mm, with a space of 50 µm for the cement agent. The digitization of the specimens was performed using an optical scanner in the L group (Lava Scan, 3M ESPE, Seefeld, Germany) and a mechanical scanner in the P group (Procera Forte, Nobel Biocare, Zurich, Switzerland).
Samples fabricated with each zirconia system were luted onto the master dies at room temperature with glass ionomer cement (Ketac Cem Easymix, 3M ESPE, Seefeld, Germany). The cement was placed on the axial surfaces of the abutments and a standardized load of 10 N was applied for 10 minutes with a dynamometric key (USAG 820/70, SWK Utensilerie, Milano, Italy) to ensure the correct distribution of the cement and to seat the FDPs properly. The cemented FDPs were stored in water for 1 week at 37 °C.
All FDPs were subjected, according to the ISO 6872:2008, with a three-point bending test until fracture using a universal testing machine (ME 405/10, SERVOSIS SA, Pinto, Spain) at a crosshead speed of 0.5 mm/min. Axial compressive loads were applied at the central fossa of the FDPs´ pontic. Data of the veneering ceramic fracture (partial fracture) and total fracture of the FDPs were automatically recorded. The force was measured in Newton (N). After testing the location of the fracture was examined visually and with a stereomicroscope (x15).
The Wilcoxon rank sum test and the Wilcoxon signed-rank test were run for FR (N) comparisons. Statistical significance was set at P < 0.05. All statistical analyses were handle with SAS 9.2 (SAS Institute Inc, Cary, NC, USA).
Results
Table 1 shows the mean and standard deviation values of fracture resistance for the experimental groups. The L and the P groups recorded comparable total FR with no significant differences between them (P=0.07). Significant differences in the fracture resistance of the veneering ceramic were recorded between both groups (P=0.0023). The fracture resistance of the veneering ceramic and the total FR were significantly different within each experimental group (P=0.002). The L group showed the highest mean value.
Table 1. Fracture resistance (in N) of veneer ceramic and total fracture of Lava and Procera group.
Regardless of the zirconia system used, the failure mainly occurred at the cervical area of the connector (80%) (Fig. 1). In both groups, occurred partial fracture in 100% of the samples, in which the porcelain veneer failed before fracture of the framework material, resulting in delamination of the porcelain layer. Seven specimens of the L group exhibited cohesive fracture within the veneering porcelain, and 3 specimens exhibited adhesive failure. In the P group, 6 specimens failed adhesively and 4 specimens exhibited mixed failure (Fig. 2).
Discussion
The present study analyses the fracture resistance of two zirconia systems and no differences were observed in total fracture, but differences were shown for veneering ceramic between the groups thus the data support the partially acceptance of the null hypothesis.
Little in vitro studies have been carried out regarding the strength of frameworks and porcelain veneering on zirconia posterior FDPs. The study results indicate that in all the specimens and in both zirconia systems the veneer porcelain fractures at a lower load than the framework, with significant differences in fracture values between the veneer and the framework. These results are in accordance with previous studies in which failure of the porcelain layer before failure of the framework material was observed (1,7-11). The porcelain delamination observed is probably related to zirconia superior mechanical properties (1). The results are also in accordance with those of clinical studies in which the main problem identified in posterior zirconia FDPs, is the chipping of the veneer porcelain (3,12-15).
In the present study, the Procera group revealed predominantly adhesive failure between the zirconia core and the veneer ceramic in 6 specimens and adhesive failure mode in 4 specimens, but in the Lava group, the failure pattern was predominantly cohesive, in agreement with previous studies in which the most frequent type of fracture occurs within the porcelain veneer rather than at the porcelain-zirconia interface (8,9,11,16-20). However, as was reported in a previous study the stereomicroscope evaluation may not be sufficient to truly determine the mode of fracture (9).
Different factors may influence and/or cause the fracture of the veneering porcelain as differences in termal expansión coefficients between core and ceramic, flexural strength of the veneering ceramic, firing shrinkage of ceramic, porcelain thickness, surface treatament of the framework, flaws on veneering and pour wetting by veneering on core (2,3,9,12,15). Special ceramics are nowadays developed for zirconia in order to minimize this unfavorable aspect (2).
Despite the many disadvantages of in vitro studies, it is important to evaluate isolated mechanical properties under standardized conditions (21). In this in vitro study, the methods used were chosen to reflect the clinical situation as far as possible. The tooth preparation design and dimensions of both zirconia groups are identical, and the veneer porcelain was fired according to the manufacturer´s, recommendations, with appropriate dimensions and an identical layered build-up technique, thus it is possible to make comparisons between the two zirconia systems. One important factor to take into consideration for the fracture resistance of the restorations is the anatomy of the framework. It is important to avoid the occurrence of areas with too little or too great a veneer thickness, which might reduce the restorations´ resistance (1,20,22-24), thus in the present study the design of the framework was anatomically shaped. The present study was done under compressive testing and although does not reproduce conditions in the oral environment as cyclic studies, the results of this type of test provide valid information (11,20), and previous studies showed that cyclic loading conditions does not affect the resistance of the material (9,22).
The most common fracture pattern of tested zirconia-based FDPs was at the loading point and through one or both connectors, being the iniciation of fracture in the gingival embrasure (15,22,25-27). This was in accordance with the present results in which the fracture was initiated from the gingival surface of the connector and propagated toward the pontic. Thus the results support that the connector design appears to be crucial for the fracture resistance and longevity of zirconia FDPs and should be taken into account when designing zirconia-based FDPs as previously reported (6,15).
Further laboratory and clinical studies are needed to evaluate the veneering porcelain on the fracture process of zirconia FDPs.
Conclusions
Within the limitations of this in vitro study, it can be concluded that the fracture resistance of the veneering ceramic of zirconia structures depends on the zirconia system used. The Lava system exhibited the greatest resistance of the veneering ceramic. However, the zirconia system had no effect on the ultimate resistance of zirconia frameworks. All tested groups demonstrated clinically acceptable fracture load values.
Acknowledgments
This study was supported by research grants from the University Complutense and 3M España S.A. (No. 26-2005), and Nobel Biocare Ibérica S.A. (No. 322-2005), through the last author. The authors would like to thank the Dident S.A., and Prótesis S.A. dental laboratories for their technical advice and for manufacturing the FPDs, Dr. Guillermo Caruana, National Centre for Metallurgical Research, for his assistance with the loading of the FDPs and Dr. Carmen Bravo, Centre of Data Processing, Computing Service for Research Support, University Complutense of Madrid, for her assistance with the statistical analysis.
References
- 1.Borba M, de Araujo MD, de Lima E, Yoshimura HN, Cesar PF, Grigss JA. Flexural strength and failure modes of layered ceramic structures. Dent Mater. 2011;27:1259–66. doi: 10.1016/j.dental.2011.09.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Manicone PF, Rossi Iommetti P, Raffaelli L. An overview of zirconia ceramics: basic properties and clinical applications. J Dent. 2007;35:819–26. doi: 10.1016/j.jdent.2007.07.008. [DOI] [PubMed] [Google Scholar]
- 3.Pelaez J, G Cogolludo P, Serrano B, L Lozano JF, Suarez MJ. A four-year prospective clinical evaluation of zirconia and metal-ceramic posterior fixed dental prostheses. Int J Prosthodont. 2012;25:451–8. [PubMed] [Google Scholar]
- 4.Sundh A, Sjogren G. A comparison of fracture strength of yttrium oxide partially-stabilized zirconia ceramic crowns with varying core thickness, shapes and veneer ceramics. J Oral Rehabil. 2004;31:682–8. doi: 10.1111/j.1365-2842.2004.01284.x. [DOI] [PubMed] [Google Scholar]
- 5.Aboushelib MN, de Jager N, Kleverlaan CJ, Feilzer AJ. Effect of loading method on the fracture mechanics of two layered all-ceramic restorative systems. Dent Mater. 2007;23:952–9. doi: 10.1016/j.dental.2006.06.036. [DOI] [PubMed] [Google Scholar]
- 6.Komine F, Blatz MB, Matsumura H. Current status of zirconia-based fixed restorations. J Oral Sci. 2010;52:531–9. doi: 10.2334/josnusd.52.531. [DOI] [PubMed] [Google Scholar]
- 7.Guazzato M, Proos K, Sara G, Swain MV. Strength, reliability, and mode of fracture of bilayered porcelain/core ceramics. Int J Prosthodont. 2004;17:142–9. [PubMed] [Google Scholar]
- 8.Fischer J, Grohmann P, Stawarczyk B. Effect of zirconia surface treatments on the shear strength of zirconia/veneering ceramic composites. Dent Mat J. 2008;27:448–54. doi: 10.4012/dmj.27.448. [DOI] [PubMed] [Google Scholar]
- 9.Ashkanani HM, Raigrodski AJ, Flinn BD, Heindl H, Mancl LA. Flexural and shear strengths of ZrO2 and a high-noble alloy bonded to their corresponding porcelains. J Prosthet Dent. 2008;100:274–84. doi: 10.1016/S0022-3913(08)60206-1. [DOI] [PubMed] [Google Scholar]
- 10.Aboushelib MN, Kleverlaan CJ, Feilzer AJ. Effect of zirconia type on its bond strength wit different veneer ceramics. J Prosthodont. 2008;17:401–8. doi: 10.1111/j.1532-849X.2008.00306.x. [DOI] [PubMed] [Google Scholar]
- 11.Agustin-Panadero R, Fons-Font A, Roman-Rodriguez JL, Granell-Ruiz M, del Rio-HGighsmith J, Sola-Ruiz MF. Zirconia versus metal: a preliminary comparative analysis of ceramic veneer behavior. Int J Prosthodont. 2012;25:294–300. [PubMed] [Google Scholar]
- 12.Raigrodski AJ, Hillstead MB, Meng GK, Chung KH. Survival and complications of zirconia-based fixed dental prostheses: A systematic Review. J Prosthet Dent. 2012;107:170–7. doi: 10.1016/S0022-3913(12)60051-1. [DOI] [PubMed] [Google Scholar]
- 13.Vigolo P, Mutinelli S. Evaluation of zirconium-oxide-based ceramic single-unit posterior fixed dental prostheses (FDPs) generated with two CAD/CAM systems compared to porcelain-fused-to-metal single-unit posterior FDPs: a 5-year clinical prospective study. J Prosthodont. 2012;21:265–9. doi: 10.1111/j.1532-849X.2011.00825.x. [DOI] [PubMed] [Google Scholar]
- 14.Rinke S, Gersdorff N, Lange K, Roediger M. Prospective evaluation of zirconia posterior fixed partial dentures: 7-year clinical results. Int J Prosthodont. 2013;26:164–71. doi: 10.11607/ijp.3229. [DOI] [PubMed] [Google Scholar]
- 15.Agustin-Panadero R, Román-Rodríguez JL, Ferreiroa A, Solá-Ruíz MF, Fons-Font A. Zirconia in fixed prosthesis. A literature review. J Clin Exp Dent. 2014;6:e66–73. doi: 10.4317/jced.51304. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Zhang D, Lu C, Zhang X, Mao S, Arola D. Contact fracture of full ceramic crowns subjected to occlusal loads. J Biomech. 2008;4:2995–3001. doi: 10.1016/j.jbiomech.2008.07.019. [DOI] [PubMed] [Google Scholar]
- 17.Tsalouchou E, Cattell M, Knowles J, Pittayachawan P, McDonald A. Fatigue and fracture properties of yttria partially stabilized zirconia crown systems. Dent Mater. 2008;24:308–18. doi: 10.1016/j.dental.2007.05.011. [DOI] [PubMed] [Google Scholar]
- 18.Saito A, Komine F, Blatz M, Matsumura H. A comparison of bond strength of layered veneering porcelains to zirconia and metal. J Prosthet Dent. 2010;104:247–57. doi: 10.1016/S0022-3913(10)60133-3. [DOI] [PubMed] [Google Scholar]
- 19.Rosentritt M, Hmaidouch R, Behr M, Handel G, Schneider-Feyrer S. Fracture resistance of zirconia FPDs with adhesive bonding versus conventional cementation. Int J Prosthodont. 2011;24:168–71. [PubMed] [Google Scholar]
- 20.Farga-Niñoles I, Agustin-Panadero R, Román-Rodríguez JL, Solá-Ruíz MF, Granell-Ruíz M, Fons-Font A. Fractographic study of the behabiour of different ceramic veneers on full coverage crowns in relation to supporting core materials. J Clin Exp Dent. 2013;5:e260–6. doi: 10.4317/jced.51293. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Blatz MB, Bergler M, Ozer F, Holst S, Phark JH, Chiche GJ. Bond strength of different veneering ceramics to zirconia and their susceptibility to thermocycling. Am J Dent. 2010;23:213–6. [PubMed] [Google Scholar]
- 22.Sundh A, Molin M, Sjögren G. Fracture resistance of yttrium oxide partially-stabilized zirconia all-ceramic bridges after veneering and mechanical fatigue testing. Dent Mater. 2005;21:476–82. doi: 10.1016/j.dental.2004.07.013. [DOI] [PubMed] [Google Scholar]
- 23.Molin MK, Karlsson SL. Five-year clinical prospective evaluation of zirconia-based Denzir 3-unit FPDs. Int J Prosthodont. 2008;21:223–7. [PubMed] [Google Scholar]
- 24.Kokubo Y, Tsumita M, Kano T, Fukushima S. The influence of zirconia coping designs on the fracture load of all-ceramic molar crowns. Dent Mater J. 2011;30:281–5. doi: 10.4012/dmj.2010-130. [DOI] [PubMed] [Google Scholar]
- 25.Kou W, Kou S, Liu H, Sjögren G. Numerical modeling of the fracture process in a three-unit all-ceramic fixed partial denture. Dent Mater. 2007;23:1042–9. doi: 10.1016/j.dental.2006.06.039. [DOI] [PubMed] [Google Scholar]
- 26.Larsson C, Holm L, Lovgren N, Kokubo Y, Vult von Steyern P. Fracture strength of four-unit Y-TZP FPD cores designed with varying connector diameter. An in-vitro study. J Oral Rehabil. 2007;34:702–9. doi: 10.1111/j.1365-2842.2007.01770.x. [DOI] [PubMed] [Google Scholar]
- 27.Silva NR, Bonfante EA, Rafferty BT, Zavanelli RA, Rekow ED, Thompson VP. Modified Y-TZP core design improves all-ceramic crown reliability. J Dent Res. 2011;90:104–8. doi: 10.1177/0022034510384617. [DOI] [PMC free article] [PubMed] [Google Scholar]