Graphical abstract
Method name: Construction of resistant optical-fiber cables for behavioral optogenetics
Keywords: Optogenetics, Optical fiber, Behavior
Abstract
In vivo optogenetic experiments commonly employ long lengths of optical fiber to connect the light source (commonly a laser) to the optical fiber implants in the brain. Commercially available patch cords are expensive and break easily. Researchers have developed methods to build these cables in house for invivo experiments with rodents [1], [2], [3], [4]. However, the half-life of those patch cords is greatly reduced when they are used with behaving rats, which are strong enough to break the delicate cable tip and to bite through the optical fiber and furcation tubing. Based on [3] we have strengthened the patch-cord tip that connects to the optical implant, and we have incorporated multiple layers of shielding to produce more robust and resistant cladding. Here, we illustrate how to build these patch cords with FC or M3 connectors. However, the design can be adapted for use with other common optical-fiber connectors. We have saved time and money by using this design in our optical self-stimulation experiments with rats, which are commonly several months long and last four to eleven hours per session. The main advantages are:
-
•
Long half-life.
-
•
Resistant to moderate rodent bites.
-
•
Suitable for long in vivo optogenetic experiments with large rodents.
Method details
The following steps illustrate how to build a robust optical patch-cord suitable for in vivo experiments with rats. This protocol complements other methods [1], [2], [3], [4] to build optical components for optogenetic experiments. Overall, we start with a 200 μm core multi-mode optical fiber of the desired length. We then attach the appropriate connector to be used with the behavioral testing setup (FC or M3 in this protocol). The length of the optical fiber is strengthened and protected by incorporating a layer of black shrink tubing and a stainless steel compression spring. We strengthen the opposite tip of the patch cord, which is attached to the optical implant on the animal’s head, by merging two stainless alloy ferrules and adding several layers of shrink tubing. This results in a strong and semi-rigid tip that serves as a holder and it is resistant to being dislodged when the animal moves around the setup.
Step 1: Before you begin, read this section completely and refer to Table 1 for a list of required materials. The assembling process takes sixty to ninety minutes. We prefer to use optical fibers with a smaller diameter core and/or smaller numerical aperture than that of the optical implant to which the patch-cord will be attached to. This is a preventive measure to avoid light loss. In principle, increasing the core diameter and/or the numerical aperture should retain light transmission in the face of small alignment errors. When two ferrules are joined by means of a ceramic sleeve, alignment may well vary somewhat as a function of the animal’s movements. Avoiding variation in light transmission is of particular importance in psychophysical experiments, which require tight, graded control of optical power. The following steps and inventory are appropriate to make a 200 μm core multi-mode 0.39 NA fiber patch-cord compatible with a 300 μm multimode implant, but they can be adapted for other core diameters, specially if experimenters are concerned about the larger tissue damage produced by using thicker fibers.
Table 1.
Material | Company | Catalog no. | Quantity |
---|---|---|---|
200 μm, 0.39 NA multimode optical fiber | Thorlabs Inc., Newton, New Jersey, USA | FT200EMT | Cut to desired length |
200 μm optical fiber stripper | Thorlabs Inc., Newton, New Jersey, USA | T12S21 | 1 |
Heat curable epoxy resin part A | Precision Fiber Products Inc., Silicon Valley, California, USA | PFP-353ND-16OZ-A | 1 ml |
Heat curable epoxy resin part B | Precision Fiber Products Inc., Silicon Valley, California, USA | PFP-353NC-16OZ-B | 0.1 ml |
M3-230 μm connectora | Doric Lenses Inc., QC, Canada | F210-0409 | 1 |
FC alloy connector drilled to 240 μmb | Fiber Instruments Sales Inc., New York, USA | 30126G2-240 | 1 |
FC crimp toolb | Thorlabs Inc., Newton, New Jersey, USA | CT042 | 1 |
Diamond wedge scribe | Fiber Instruments Sales Inc., New York, USA | FO9OW | 1 |
3/64′′ shrink tubing | Newark Element14. Palatine, IL, USA | 84N583 | ∼10% shorter than fiber |
1/4′′ shrink tubing | Newark Element14., Palatine, IL, USA | 84N588 | ∼2′′ |
0.125′′ ID adhesive shrink tubing | Newark Element14., Palatine, IL, USA | 48W4786 | ∼2′′ |
Stainless steel compression spring, 0.1′′ ED, 0.07′′ ID | Heilplex east end., Montreal, QC, Canada | Custom order | Cut to desired length |
Stainless alloy ferrule drilled to 240 μm | Fiber Instruments Sales Inc., New York, USA | F10061F240 | 2 |
Ceramic split sleeve | Fiber Instruments Sales Inc., New York, USA | F18300SSC25 | 2 |
5 μm silicon carbide polishing paper | Thorlabs Inc., Newton, New Jersey, USA | LFG5P | 1 |
3 μm aluminum oxide polishing paper | Thorlabs Inc., Newton, New Jersey, USA | LFG3P | 1 |
1 μm aluminum oxide polishing paper | Thorlabs Inc., Newton, New Jersey, USA | LFG1P | 1 |
0.3 μm aluminum oxide polishing paper | Thorlabs Inc., Newton, New Jersey, USA | LFG03P | 1 |
Polishing puck | Thorlabs Inc., Newton, New Jersey, USA | D50-FC | 1 |
Polishing pad and polishing plate | Thorlabs Inc., Newton, New Jersey, USA | NRS913A CTG913 |
1 |
200× fiber scope | Thorlabs Inc., Newton, New Jersey, USA | FS200 | 1 |
Digital power meter | Thorlabs Inc., Newton, New Jersey, USA | PM100D | 1 |
Photodiode power sensor | Thorlabs Inc., Newton, New Jersey, USA | S121C | 1 |
Threaded fiber adapter for sensor | Thorlabs Inc., Newton, New Jersey, USA | S120-FC | 1 |
List of required materials and tools.
The M3 connector is not necessary if working with the FC connector.
These materials are not necessary if working with the M3 connector.
Step 2: Use the optical fiber stripper to remove the optical fiber buffer and expose 35 mm of the optical-fiber cladding (Fig. 1A). Measure the optical fiber from the unstripped edge (Fig. 1B) and cut it to the desired length using scissors or wire cutters.
Step 3: Mix the heat curable epoxy resin according to the manufacturer’s instructions, and load a 3 ml syringe with the mixed resin. Attach a 21 gauge blunt tip needle (Braintree Scientific, Inc., Braintree, MA) to the syringe.
Place the FC (Fiber Instruments Sales Inc., New York, USA) or M3 (Doric lenses, Québec, Québec, Canada) connector (Fig. 1C and D) on the third hand with the small lumen facing downwards. Fill the internal volume of the connector with resin. Insert the fiber core all the way through the connector, leave in place, and cure the resin with the heat gun until it darkens. Be careful not to overheat the epoxy: some optical fibers may be damaged by excessive heat, resulting in lower light transmission. Any resin spills on the connector should be wiped off before heat-curing the resin.
When the resin has cured, the tip of the M3 or FC connector should have a small excess drop of resin on the 240 μm lumen (Fig. 1E), which will hold the optical fiber core in place when the end of the fiber is polished. If the excess resin is not present, a 26 gauge needle or a piece of optical-fiber buffer can be used to apply a small drop of resin to the area (Fig. 1F).
Step 4: Use the diamond scribe to make a small straight groove in the side of the optical fiber core, right at the edge of the hardened resin on the tip of the connector (Fig. 2A). Flick-off the protruding length of optical fiber core to leave a level surface. Avoid breaking the core with the diamond scribe.
If working with the M3 connector, insert the internally threaded nut (provided by the manufacturer) until it reaches the connector (Fig. 2B).
If working with an FC connector, pass the optical fiber through the metal boot (Fig. 2C) and crimp it in place using the special-purpose tool (not shown).
Step 5: Cut a piece of 3/64′′ shrink tubing (approximately 10% shorter than the length of the optical fiber), and pass the optical fiber through it until it is in contact with the connector (Fig. 2D). Secure the fiber to a vertical surface, and with the connector facing downwards, apply heat to evenly and tightly shrink the tubing around the fiber (Fig. 2E). Make sure that the shrink tubing cladding is even; avoid having bumps on the tubing.
If working with M3 connectors, pass the small-bore shrink tubing (provided by the manufacturer, Fig. 2F) over the fiber until it protrudes slightly beyond the metallic end of the M3 connector. Do not shrink this tubing until the fiber tip has been properly polished (see Step 11).
If working with the FC connector, slide the rubber protector provided by the manufacturer over the metallic boot (not shown).
Step 6: Pass the optical fiber through the stainless-steel compression steel spring (Fig. 3A), and strip 45–50 mm of the optical fiber beyond the connector (Fig. 3B). When stripping, avoid leaving a large gap of unstripped fiber between the shrink tubing and the exposed optical fiber cladding (a small gap, between 5 and 8 mm long, is acceptable, Fig. 3C).
Step 7: Place a stainless-steel ferrule in the third hand (the 240 μm drilled part should face downwards). Fill up the internal volume of the ferrule with heat-curable epoxy, and pass the long section of exposed fiber core all the way through until it cannot be pushed any further. It is very important to remove all traces of resin from the inserted optical-fiber core. This can be done using a delicate-task wipe and alcohol. Do not cure the resin inside the ferrule until certain that the fiber core is resin free. Apply heat to cure the resin and fix the ferrule in place, making sure to leave approximately 40 mm of optical fiber core protruding from the ferrule (Fig. 3D).
Step 8: Fasten a ceramic split sleeve to another ferrule, making sure the 240 μm lumen is facing away from the sleeve and that most of the ferrule remains exposed (i.e., about 1/4 of the length of the ferrule should enter the ceramic sleeve, Fig. 3E). Attaching this second ferrule to the patch cord provides a rigid area to grip when attaching and detaching the cable. This also prevents the cable from bending sharply as the animal makes sudden movements and reduces the likelihood that contact with the levers or the chamber walls will detach, break, or dislodge the patch cord. In our hands, cables with single ferrules at the tip have a shorter half-life than those built using two ferrules in tandem.
Place the attached ferrule on the third hand and load it with heat curable epoxy (Fig. 3F). Allow for some resin to reach the inside of the ceramic sleeve. Insert the optical fiber core through the ferrule, until both ferrules are in contact with each other within the ceramic sleeve (Fig. 3G). Remove excess resin and heat cure the remainder in place.
When the resin has cured, a small excess should remain at the tip of the ferrule. If not present, gently apply a very small amount of resin to the area (as in Fig. 1F). Again, use the diamond scribe to make a groove on the side of the optical fiber core, at the edge of the cured resin. Make sure not to break the fiber when making the groove. Flick to cut the fiber at the level of the groove.
Step 9: Push the spring over the fiber until it is in contact with the ferrule. Insert the end with the attached ferrules through a 2” piece of the adhesive shrink tubing (Fig. 4A). Push the ferrules through the adhesive shrink tubing until the ferrule glued to the tip of the cord protrudes completely and some shrink tubing covers a section of the spring; the attached ceramic sleeve should be almost completely covered by the tubing. Heat-shrink the tubing in place, making sure the spring is in close contact with the ferrule. Apply light pressure around the tubing while it is still hot to produce tight contact with the spring and the ceramic sleeve below it (Be careful: the tubing can be very hot!). Some adhesive from the inside of the tubing may spill out. Avoid getting it on the exposed ferrule. This step stiffens the tip of the patch cord and secures the spring in place (Fig. 4B).
Step 10: Strengthen the tip by inserting it into a piece of 1/4′′ shrink tubing (approximately 2′′ long) and heat-shrinking this tubing over the already stiffened patch-cord tip (Fig. 4C), taking care to leave the ferrule exposed. This step can be repeated until the desired rigidity or external diameter is achieved. We recommend using at least one layer of 1/4′′ shrink tubing.
Step 11: Fibers need to be adequately polished to achieve optimal performance; instructions can be downloaded for free from the manufacturer’s web site: https://www.thorlabs.com/thorproduct.cfm?partnumber=FN96A. Polish the tip of the patch cord on the 5 μm silicon carbide polishing paper by repeatedly tracing a figure eight pattern with the exposed ferrule tip. Use the polishing puck to keep the ferrule vertical while polishing (Fig. 4D). Applying too much pressure when polishing can damage the optical-fiber core. Once the cured resin at the tip of the ferrule has been polished away, clean the tip of the fiber using isopropyl alcohol and a delicate-task wipe. Replace the silicon carbide paper with the 3 μm polishing paper and repeat the polishing process. It is helpful to keep this fine polishing paper clean between polishes by using isopropyl alcohol and delicate-task wipes. Lubricating the patch-cord tip with a drop of distilled water makes the polishing process easier. Repeat the process with progressively finer polishing paper (i.e., 1 μm and 0.3 μm) until the polished optical fiber core appears circular, smooth, light gray, and without scratches when viewed through the 200× fiber scope. Repeat these steps to polish the FC connector.
If using an M3 connector, clamp a straight hemostat to the connector tip and keep it flat against the polishing paper [3]. Make sure the long axis of the tip of the connector is perpendicular to, and in touch with, the paper. Polish by drawing a series of figure eight on the silicon-carbide paper with the tip of the connector. The hemostat should help avoid uneven polishing. Continue the process on progressively finer paper, as you did to the other end of the patch cord. The optical fiber core should be circular, light gray, and glossy when viewed through the fiber scope. The small shrink tubing from the M3 connector should extend slightly over the metallic part of the connector so as to wrap around it and the optical fiber, while still allowing the threaded nut to rotate and move along the base of the connector (Fig. 4E).
Step 12: Test the patch cord in the behavioral apparatus. Attach the patch-cord connector to the optical rotary joint (Fig. 4F) or the corresponding light output of your setup. Carefully point the patch cord tip toward a vertical non-reflective surface and turn on the light source (Caution: do not use full laser power in this test, and wear appropriate protective goggles, except when you are certain that the beam is pointing away from you, and you wish to view the pattern it forms on the non-reflective surface). A properly polished patch cord will produce a pattern of closely spaced concentric rings (Fig. 5A). Confirm that the desired optical power is emitted by the patch cord by using the photodiode sensor and the optical-power meter. Adjust the laser power to determine the maximal light output from the cord. If the patch cord is properly built and polished, the light loss should not exceed 15% per connection (top and bottom). It is best to provide “headroom” by using patch cords that exceed the required output. Avoid using patch cords that cannot transmit at least 10% more power than required in the experiment.
Insert the metal ferrule at the patch-cord tip halfway into a ceramic sleeve (Fig. 5B). The metallic ferrule attached to the optical implant should fill the empty half of the ceramic sleeve so that both ferrules are in close contact inside the sleeve (Fig. 5B inset).
These patch cords have held up well for several months under heavy use in daily experimental sessions lasting from four to eleven hours. We have used them successfully with or without a counterbalancing arm (MED Associates Inc., St. Albans, VT. Product number: PHM-110-SAI). When cut to a length of 2.32 ft (71 cm) the patch cords weigh approximately 0.49 oz (14 g).
Additional information
Damage prevention and maintenance
Always handle the patch cord by the reinforced ferrule tip. This is particularly important when detaching it from the optical implant: pulling directly on the cord may damage it.
On rare occasions, the patch cord may detach from the rat’s head during an experimental session, and the rat may seize it and bite it. This can be prevented by routinely replacing the ceramic sleeve that connects the cord to the optical implant; these sleeves tend to loosen with continued use.
The metallic shielding and tubing layers tend to survive some nibbling by a rat. When this occurs, new layers of 1/4′′ shrink tubing can be used to re-strengthen the damaged part. In rare cases, a rat may become adept at detaching the patch cord, and may destroy it if given enough time. Thus, it is useful to apply a few drops of a quinine solution to the patch cord tip to dissuade the animal from biting.
After several months of usage, some loss of power may occur. This can be remedied by gently wiping the ferrule end with alcohol, or by re-polishing it if the problem persists.
Acknowledgments
Dr. Daniel Madularu took most of the photos. We thank David Munro for technical assistance and advice. ITP was supported by a MELS-PBEE scholarship from the Fonds de la recherche en santé de Québec (#140478) and a scholarship from Consejo Nacional de Ciencia y Tecnologia (CONACYT, #309126). This project was also supported by an NSERC grant to PS (#RGPIN308-11). Additional financial support was provided by two Facilities Optimization Program grants from Concordia University to PS and NC and the Concordia University Research Chair held by PS.
MethodsX thanks the reviewers of this article (David Christopher Klorig and a second reviewer who would like to remain anonymous) for taking the time to provide valuable feedback.
References
- 1.Aravanis A.M., Wang L.-P., Zhang F., Meltzer L.A., Mogri M.Z., Schneider M.B., Deisseroth K. An optical neural interface: in vivo control of rodent motor cortex with integrated fiberoptic and optogenetic technology. J. Neural Eng. 2007;4(3):S143–S156. doi: 10.1088/1741-2560/4/3/S02. [DOI] [PubMed] [Google Scholar]
- 2.Klorig D.C., Godwin D.W. A magnetic rotary optical fiber connector for optogenetic experiments in freely moving animals. J. Neurosci. Methods. 2014;227:132–139. doi: 10.1016/j.jneumeth.2014.02.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Sparta D.R., Stamatakis A.M., Phillips J.L., Hovelsø N., van Zessen R., Stuber G.D. Construction of implantable optical fibers for long-term optogenetic manipulation of neural circuits. Nat. Protoc. 2011;7(1):12–23. doi: 10.1038/nprot.2011.413. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Ung K., Arenkiel B.R. Fiber-optic implantation for chronic optogenetic stimulation of brain tissue. J. Vis. Exp. 2012;68:e50004. doi: 10.3791/50004. [DOI] [PMC free article] [PubMed] [Google Scholar]