Abstract
We compared the performance of an in-house and a commercial malaria polymerase chain reaction (PCR) assay using freeze–thawed hemolytic blood samples.
A total of 116 freeze–thawed ethylenediamine tetraacetic acid (EDTA) blood samples of patients with suspicion of malaria were analyzed by an in-house as well as by a commercially available real-time PCR.
Concordant malaria negative PCR results were reported for 39 samples and malaria-positive PCR results for 67 samples. The in-house assay further detected one case of Plasmodium falciparum infection, which was negative in the commercial assay as well as five cases of P. falciparum malaria and three cases of Plasmodium vivax malaria, which showed sample inhibition in the commercial assay. The commercial malaria assay was positive in spite of a negative in-house PCR result in one case. In all concordant results, cycle threshold values of P. falciparum-positive samples were lower in the commercial PCR than in the in-house assay.
Although Ct values of the commercial PCR kit suggest higher sensitivity in case of concordant results, it is prone to inhibition if it is applied to hemolytic freeze–thawed blood samples. The number of misidentifications was, however, identical for both real-time PCR assays.
Keywords: malaria, PCR, test comparison, difficult sample matrix, freeze–thawed blood, molecular diagnostics, in-house, real-time
Introduction
Malaria, travel-associated diarrhea as well as skin and sexually transmitted diseases are major health concerns on deployments in subtropical and tropical endemic settings [1]. Rapid diagnosis in case of suspected malaria is mandatory. Traditional microscopy of Giemsa-stained thin and thick blood smears remains the diagnostic golden standard. However, this cheap and easy-to-perform procedure requires expert knowledge and substantial skills in parasite microscopy, which is not easy to maintain in non-endemic areas, where malaria is rare or constitutes a travel-associated disease.
In resource-limited deployment settings, alternative, less investigator-dependent diagnostic approaches are desirable. Rapid diagnostic tests (RDTs) are a frequently chosen option. Such tests are usually based on lateral-flow immunochromatography, and parasites are detected by respective antibodies. In detail, type 2 RDTs are based on anti-HRP-(histidine rich protein)-2-antibodies for Plasmodium falciparum and anti-aldolase-antibodies for all species; type 3 RDTs on anti-HRP-2-antibodies for P. falciparum and anti-pLDH-(plasmodium lactate de hydrogenase)-anti bodies for all species; and type 4 RDTs on anti-pLDH-antibodies for P. falciparum and anti-pLDH-antibodies for all species. In a recent Cochrane meta-analysis, average sensitivities and specificities were 78% and 99% for type 2 and 3 tests, and 89% and 98% for type 4 tests [2]. Thus, sensitivity of RDTs is still a matter of concern. Moreover, in the presence of heterophilic IgM antibodies or in cases of exceedingly high parasitemia, RDTs may even produce false negative results [3, 4].
Molecular DNA-based assays have been proven as alternative approaches, with high sensitivity and specificity for malaria diagnostics, which can be standardized and require less experienced skills than microscopy. If the respective infrastructural prerequisites are in place in a tropical field setting, polymerase chain reaction (PCR) might substitute for microscopy or RTDs. However, PCR can be limited by sample inhibition, e.g., due to hemolysis of blood samples [5]. This is a relevant concern, as adequate sample storage and transport cannot always be guaranteed under tropical deployment conditions. To assess the dimension of this problem, we evaluated a commercial and an in-house real-time PCR assay using repeatedly freeze–thawed, hemolytic blood of patients with suspected malaria.
Methods
Samples
A total of 116 freeze–thawed ethylenediamine tetraacetic acid (EDTA) blood samples of patients with suspicion of malaria were analyzed by traditional thick and thin blood smear microscopy by experienced technical assistants at the Bernhard Nocht Institute for Tropical Medicine Hamburg, the German National Reference Center for Tropical Diseases, which is also responsible for malaria slide reading proficiency testing of diagnostic German laboratories. Parasitemia was determined according to WHO standards and recorded as parasites per microliter.
Residual materials of EDTA blood were subjected to repeated freeze–thawing to ensure progressive hemolysis prior to further processing.
Nucleic acid extraction
From the hemolytic EDTA blood samples, DNA was extracted and enriched using the EZ1 DNA Blood 200 µl Kit (Qiagen, Hilden, Germany) on the automated EZ1 system (Qiagen) exactly as described by the manufacturer. EZ1-based extraction has recently been shown to be equally efficient than extraction by the column-based QIAamp DNA Blood Mini Kit (Qiagen) for such specimens [6].
PCR analyses
All samples were subjected to real-time in-house one-tube SybrGreen malaria PCR and commercial altona Diagnostics Realstar Malaria PCR, a Taqman approach. In detail, the in-house SybrGreen melting curve analysis-based approach targeting Plasmodium (P.) falciparum, P. malariae, P. ovale, and P. vivax was performed as previously described [7]. The Plasmodium spp.-specific real-time PCR RealStar Malaria PCR Kit 1.0 (altona Diagnostics, Hamburg, Germany) was applied exactly as described by the manufacturer.
Analysis
If concordant positive PCR results in both PCR approaches were observed, achieved cycle threshold (Ct) values were assessed including calculation of median and mean values as well as standard deviations (SDs). Commercial Real-Star Malaria PCR included an inhibition control, which allowed monitoring of PCR inhibition as well. Of note, negative inhibition control PCR in PCR-positive samples is caused by competitive amplification in this assay and not a sign of procedural failure.
Matched pairs of recorded Ct values were compared using Wilcoxon matched-pairs signed ranks testing with the software GraphPad InStat, version 3.06 (GraphPad Software Inc., La Jolla, CA, USA). Significance was accepted in case of a two-tailed P value ≤ 0.05. The nonparametric Spearman correlation coefficient was calculated (GraphPad Software Inc., La Jolla, CA, USA) to confirm effective pairing.
Results
Microscopic assessment
Of the 116 samples analyzed, 76 were microscopically positive for plasmodia. These comprised 61 cases of mono-infections with P. falciparum, seven cases with P. vivax, two cases with P. ovale, and one case with P. malariae, respectively. Parasite densities are shown in Table 1. Of note, parasitemia was not determined in one case of P. vivax infection and there were two cases in which microscopic discrimination between P. vivax and P. ovale was uncertain (1240/µl and 8720/µl). There was one case of P. vivax infection with a possible coinfection with P. malariae (100,000/µl), one case with a double infection of P. falciparum and P. malariae (500,000/µl), and one case in which microscopic discrimination of Pasmodium spp. beyond the genus level failed (480/µl). In 40 cases, malaria microscopy was negative.
Table 1.
Parasite density of mono-infections with African Plasmodium spp. as assessed by thick and thin smear microscopy
| P. falciparum | P. vivax | P. ovale | P. malariae | |
|---|---|---|---|---|
| Lowest observed density [/µl] | 0.2 | 2400 | 714 | 1400 |
| Highest observed density [/µl] | 350,000 | 8040 | 5200 | 1400 |
| Median [/µl] | 7000 | 3040 | 3117 | n.a. |
| Mean [/µl] | 43,753 | 4120 | 3117 | n.a. |
| Standard deviation [/µl] | 81,990 | 2167 | 3398 | n.a. |
n.a. = Not applicable
PCR results
In direct comparison of the two PCR assays regarding the overall detection of malaria, concordant results were recorded for 106 of the 116 samples of which 67 were parasite positive and 39 were parasite negative according to microscopy. The in-house assays further detected one case of a P. falciparum infection, which was negative by the commercial assay as well as five cases of P. falciparum and three cases of P. vivax infections, which showed sample inhibition in the commercial assay (Table 2). All of the latter discordant nine samples were positive by microscopy. In one case, the commercial malaria assay was positive with a relatively high Ct value of 31, but was negative with the in-house PCR. Of note, microscopy was negative in this sample (Table 2).
Table 2.
Characterization of samples with discordant PCR results
| Species in microscopy | Parasite density in microscopy [/µl] | Species in in-house PCR | Ct value in in-house PCR | Ct value in commercial PCR |
|---|---|---|---|---|
| P. falciparum | 0.4 | P. falciparum | 28 | Negative |
| P. falciparum | 800 | P. falciparum | 22 | Inhibited |
| P. falciparum | 96 | P. falciparum | 24 | Inhibited |
| P. falciparum | 0.2 | P. falciparum | 25 | Inhibited |
| P. falciparum | 37,600 | P. falciparum | 16 | Inhibited |
| P. falciparum | 50,000 | P. falciparum | 15 | Inhibited |
| P. vivax/ovale | 1240 | P. vivax | 20 | Inhibited |
| P. vivax | 2920 | P. vivax | 19 | Inhibited |
| P. vivax | 2920 | P. vivax | 17 | Inhibited |
| Negative | Negative | Negative | Negative | 31 |
Inhibition control PCR of the commercial PCR approach remained negative for 46 of 116 samples. Negative inhibition control PCR was observed in 29 PCR-positive samples, in nine PCR-negative samples in concordance with microscopic results, and in eight PCR-negative samples in discordance with microscopic results. In total, 17 out of 116 (14.7%) PCR results of the commercial assay were non-interpretable, reducing the number of unambiguously false negative results of this approach with inconspicuous inhibition control PCR to one.
Of note, in-house PCR recorded P. vivax only in two microscopically uncertain cases of P. vivax/P. ovale infections and P. falciparum in a case in which microscopic differentiation beyond the genus level of Plasmodium spp. failed. In addition, in-house PCR recorded P. falciparum in a case of microscopic finding of P. vivax infection with possible coinfection with P. malariae and, in another case, in-house PCR recorded P. falciparum mono-infection, but microscopy suggested double infection with both P. falciparum and P. malariae (Table 3).
Table 3.
Characterization of samples with discordant results in in-house PCR and microscopy
| Species in in-house PCR | Ct value in in-house PCR | Ct value in commercial PCR | Species in microscopy | Parasite density in microscopy [/µl] |
|---|---|---|---|---|
| P. vivax | 16 | 15 | P. vivax/ovale | 8720 |
| P. vivax | 20 | Inhibited | P. vivax/ovale | 1240 |
| P. falciparum | 16 | 12 | P. vivax with possible coinfection with P. malariae | 100,000 |
| P. falciparum | 19 | 16 | P. falciparum with coinfection with P. malariae | 500,000 |
| P. falciparum | 23 | 20 | Plasmodium spp. | 480 |
Cycle threshold comparison of the two assessed PCR assays
In cases of concordant results, cycle threshold values of P. falciparum-positive samples were significantly lower with the commercial PCR than with the in-house assay. However, there were no differences in Ct values between the two assays for samples containing P. vivax. The number of samples containing P. ovale or P. malariae was too low for meaningful statistical analyses (Table 4).
Table 4.
Comparison of Ct values of commercial PCR and the in-house approach
| P. falciparum | P. vivax | P. ovale | P. malariae | |
|---|---|---|---|---|
| Median Ct value in-house PCR | 18 | 16.5 | 22.5 | 26 |
| Median Ct value commercial PCR | 15 | 16 | 21 | 20 |
| Mean Ct value in-house PCR | 18.6 | 16.3 | 22.5 | 26 |
| Mean Ct value commercial PCR | 15.7 | 15.8 | 21 | 20 |
| Standard deviation in-house PCR | 3.7 | 1.4 | 3.5 | n.a. |
| Standard deviation commercial PCR | 4.0 | 2.4 | 2.8 | n.a. |
| P value in Wilcoxon matched-pairs signed ranks testing | P < 0.0001 | P = 0.4375 | n.a. | n.a. |
| P value of nonparametric Spearman correlation | P < 0.0001 | P = 0.0167 | n.a. | n.a. |
n.a. = Not applicable
Discussion
The role of PCR for the diagnosis of malaria is still under ongoing debate. As PCR assays have been shown to be more sensitive and specific than other approaches, some authors argue that this technology should be regarded as the new golden standard for the diagnosis of malaria at least in non-endemic countries. These authors justify their claim with the fact that a negative PCR finding shows the highest reliability in excluding malaria infections, a matter of high importance in returning travelers with a low risk of acquiring malaria [8]. Diagnostic PCRs for, e.g., P. falciparum, typically target genes such as SSUrRNA, pf155/resa, or cox1 [9]. Alternative diagnostic approaches such as flow cytometry and mass spectrometry that might allow for high-throughput screenings are still in experimental states [10].
More critical authors argue that several factors limit the use of PCR for the diagnosis of malaria at least in resource-limited high-endemicity settings. These factors comprise the time lag between sample collection, transportation and processing, and dissemination of results back to the physician in comparison to traditional microscopy next to limited financial resources, persistent subclinical parasitemia, and inadequate laboratory infrastructures in resource-limited, remote areas [11]. Such problems can be overcome in deployment situations, e.g., in medical field camps. Therefore, the developmental interest should be focused on potential technical limitations.
Indeed, the here described analysis showed good concordance between microscopic findings and PCR results in spite of a considerably high inhibition rate of the hemolytic samples as detected by the internal control approach of the commercial PCR. Regarding the overall detection of malaria, non-inhibited, truly discordant PCR results were observed in only two cases. Commercial PCR failed in one case without indicating inhibition. This sample contained a very low number of parasites. In contrast, commercial PCR was positive with a high cycle threshold value in a sample which was negative in in-house PCR as well as in microscopy. Of note, this sample was positive for hemozoin in matrix-assisted laser desorption–ionization time-of-flight mass spectrometry (MALDI-TOF-MS) [12, 13] (data not shown). This result makes a false-positive result, e.g., due to contamination, unlikely. Instead, it suggests either a case of extremely low parasitemia or a post-treatment case with parasite DNA and antigen remnants still circulating in the blood.
The high number of eight malaria cases that were missed due to inhibition of the commercial PCR is bothersome. Regarding this aspect, in-house PCR proved to be more inhibition-resistant than the commercial approach. In cases of concordant results between the two PCRs, however, cycle threshold values of the commercial PCR were lower, suggesting higher sensitivity. However, statistical significance of this phenomenon was observed for P. falciparum only, while detection of other Plasmodium ssp. was too rare for reliable statistical results.
Melting curve-based species identification of the in-house PCR approach allowed for the identification of one case of P. falciparum infection and two cases of P. vivax infections in which microscopy was not able to reliably discriminate the parasites on species level. This is in line with the widely accepted fact that PCR is more suitable for Plasmodium species identification than microscopy, in particular in case of low parasite counts, which make microscopic identification challenging [9].
However, discordant results of in-house PCR were observed for two cases of microscopically detected double infections. In case of a double infection with P. falciparum and P. malariae, only P. falciparum was identified by PCR. In the applied one-tube SybrGreen approach, the primers obviously reacted more readily with the quantitatively dominating DNA of P. falciparum, leading to a neglect of P. malariae coinfection. This is a known disadvantage of this procedure [7], although other molecular approaches were reported to be able to identify coinfections with different Plasmodium spp. even more reliable than microscopy [9]. In another instance, in-house PCR indicated P. falciparum although microscopy suggested P. vivax with a possible P. malariae coinfection. Obviously, schizonts of P. vivax and P. malariae had been observed in microscopy and the ring stages of P. falciparum had been attributed to one of these species. PCR, again, detected only DNA of quantitatively dominating P. falciparum in this sample which contained two or maybe even three different Plasmodium spp. The disadvantage of the low discriminatory potential of the one-tube SybrGreen in-house approach in case of infections with more than one Plasmodium spp. has to be balanced against the obvious advantage of high robustness even with hemolytic samples as observed in this assessment.
The development of molecular approaches for the diagnosis of malaria is ongoing. The use of microfluidic and chip-based approaches has been suggested several years ago [14]. More recently, a review on currently introduced loop-mediated isothermal amplification (LAMP) schemes as robust molecular approaches for the detection of Plasmodium spp. has been published [15]. On the intermediate and long term, the targeted detection of drug resistance-related mutations to support targeted antimalarial therapy might become an option for the diagnostic routine setting in well-equipped laboratories [9].
Conclusions
The analysis presented here suggests that malaria PCR leads to reliable results even in cases of poor sample quality due to hemolysis of blood. The one-tube SybrGreen in-house approach proved to be more robust than the commercial assay, in particular with regard to inhibition tolerance, although the cycle threshold values were lower in the commercial PCR. If sample inhibition in the commercial assay and associated negative Plasmodium spp. specific PCRs are observed, other diagnostic approaches have to be applied in order to rule out malaria infections.
Acknowledgements
Simone Priesnitz, Christine Wegner, and Iris Zielke are greatly acknowledged for skillful technical assistance.
Declaration of interest
The authors declare that there are no conflicts of interest.
References
- 1.Frickmann H, Schwarz NG, Holtherm HU, Maaßen W, Vorderwülbecke F, Erkens K, Fischer M, Morwinsky T, Hagen RM: Compliance with antimalarial chemoprophylaxis in German soldiers: a 6-year survey. Infection 41, 311–320 (2013) [DOI] [PubMed] [Google Scholar]
- 2.Abba K, Kirkham AJ, Olliaro PL, Deeks JJ, Donegan S, Garner P, Takwoingi Y: Rapid diagnostic tests for diagnosing uncomplicated non-falciparum or Plasmodium vivax malaria in endemic countries. Cochrane Database Syst Rev 12, CD011431 (2014) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Gillet P, Mori M, Van Esbroeck M, Van den Ende J, Jacobs J: Assessment of the prozone effect in malaria rapid diagnostic tests. Malar J 8, 271 (2009) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Gillet P, Scheirlinck A, Stokx J, De Weggheleire A, Chaúque HS, Canhanga OD, Tadeu BT, Mosse CD, Tiago A, Mabunda S, Bruggeman C, Bottieau E, Jacobs J: Prozone in malaria rapid diagnostics tests: how many cases are missed? Malar J 10, 166 (2011) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Mahajan B, Zheng H, Pham PT, Sedegah MY, Majam VF, Akolkar N, Rios M, Ankrah I, Madjitey P, Amoah G, Addison E, Quakyi IA, Kumar S: Polymerase chain reaction-based tests for pan-species and species-specific detection of human Plasmodium parasites. Transfusion 52, 1949–1956 (2012) [DOI] [PubMed] [Google Scholar]
- 6.Frickmann H, Hinz R, Hagen RM: Comparison of an automated nucleic acid extraction system with the column-based procedure. Eur J Microbiol Immunol (Bp) 5, 94–102 (2015) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Mangold KA, Manson RU, Koay ES, Stephens L, Regner M, Thomson RB, Jr, Peterson LR, Kaul KL: Real-time PCR for detection and identification of Plasmodium spp. J Clin Microbiol 43, 2435–2440 (2005) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Berry A, Benoit-Vical F, Fabre R, Cassaing S, Magnaval JF: PCR-based methods to the diagnosis of imported malaria. Parasite 15, 484–488 (2008) [DOI] [PubMed] [Google Scholar]
- 9.Berry A, Fabre R, Benoit-Vical F, Cassaing S, Magnaval JF: Contribution of PCR-based methods to diagnosis and management of imported malaria. Med Trop (Mars) 65, 176–183 (2005) [PubMed] [Google Scholar]
- 10.Hawkes M, Kain KC: Advances in malaria diagnosis. Expert Rev Anti Infect Ther 5, 485–495 (2007) [DOI] [PubMed] [Google Scholar]
- 11.Hänscheid T, Grobusch MP: How useful is PCR in the diagnosis of malaria? Trends Parasitol 18, 395–398 (2002) [DOI] [PubMed] [Google Scholar]
- 12.Scholl PF, Kongkasuriyachai D, Demirev PA, Feldman AB, Lin JS, Sullivan DJ, Jr, Kumar N: Rapid detection of malaria infection in vivo by laser desorption mass spectrometry. Am J Trop Med Hyg 71, 546–551 (2004) [PubMed] [Google Scholar]
- 13.Nyunt M, Pisciotta J, Feldman AB, Thuma P, Scholl PF, Demirev PA, Lin JS, Shi L, Kumar N, Sullivan DJ, Jr: Detection of Plasmodium falciparum in pregnancy by laser desorption mass spectrometry. Am J Trop Med Hyg 73, 485–490 (2005) [PubMed] [Google Scholar]
- 14.Gascoyne P, Satayavivad J, Ruchirawat M: Microfluidic approaches to malaria detection. Acta Trop 89, 357–369 (2004) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Oriero EC, Jacobs J, Van Geertruyden JP, Nwakanma D, D’Alessandro U: Molecular-based isothermal tests for field diagnosis of malaria and their potential contribution to malaria elimination. J Antimicrob Chemother 70, 2–13 (2015) [DOI] [PMC free article] [PubMed] [Google Scholar]