Abstract
Background and Objectives
Multiple-breath washout (MBW) is an attractive test to assess ventilation inhomogeneity, a marker of peripheral lung disease. Standardization of MBW is hampered as little data exists on possible measurement bias. We aimed to identify potential sources of measurement bias based on MBW software settings.
Methods
We used unprocessed data from nitrogen (N2) MBW (Exhalyzer D, Eco Medics AG) applied in 30 children aged 5–18 years: 10 with CF, 10 formerly preterm, and 10 healthy controls. This setup calculates the tracer gas N2 mainly from measured O2 and CO2concentrations. The following software settings for MBW signal processing were changed by at least 5 units or >10% in both directions or completely switched off: (i) environmental conditions, (ii) apparatus dead space, (iii) O2 and CO2 signal correction, and (iv) signal alignment (delay time). Primary outcome was the change in lung clearance index (LCI) compared to LCI calculated with the settings as recommended. A change in LCI exceeding 10% was considered relevant.
Results
Changes in both environmental and dead space settings resulted in uniform but modest LCI changes and exceeded >10% in only two measurements. Changes in signal alignment and O2 signal correction had the most relevant impact on LCI. Decrease of O2 delay time by 40 ms (7%) lead to a mean LCI increase of 12%, with >10% LCI change in 60% of the children. Increase of O2 delay time by 40 ms resulted in mean LCI decrease of 9% with LCI changing >10% in 43% of the children.
Conclusions
Accurate LCI results depend crucially on signal processing settings in MBW software. Especially correct signal delay times are possible sources of incorrect LCI measurements. Algorithms of signal processing and signal alignment should thus be optimized to avoid susceptibility of MBW measurements to this significant measurement bias.
Introduction
Assessment of impaired ventilation distribution in the lungs by multiple-breath washout (MBW) measurement has been increasingly used over the past few years. In children with cystic fibrosis (CF) but also with primary ciliary dyskinesia MBW has been shown to be more sensitive for detection of early structural lung changes than standard lung function tests [1–3]. MBW is able to assess treatment effects even in mild CF lung disease [4–6]. In some centers, MBW is already part of the routine clinical surveillance in patients with CF [7–9].
The recently published ATS/ERS consensus aims to standardize MBW signal recording, processing, and analysis [10]. Despite guidelines as to what software packages should be able to perform, most of the technical recommendations rely on little data only [11;12]. In particular, uncertainty exists as to what extent the different software settings impact upon MBW outcomes. Even if the MBW post-hoc quality criteria are met [10], technical flaws during or after the measurement may strongly impact on MBW results. Random measurement bias may particularly influence longitudinal MBW data. Further, some MBW data require post hoc offline analysis to adjust for incorrect settings of online measurements. This may introduce additional non-systematic bias on test results.
To avoid MBW measurement bias as good as possible, the most important software settings and the respective impact upon results need to be known. The aim of our study was thus to assess the influence of different software settings on MBW results. We used nitrogen (N2) MBW raw data from 30 children and adolescents with and without lung disease. We systematically evaluated the impact of changing different software parameters in order to identify the most important source of measurement bias. Primary outcome was the change in lung clearance index (LCI) and functional residual capacity (FRC), secondary outcome was change in phase III slope indices (SIII).
Methods
Subjects
To cover a wide age and disease range we used unprocessed raw N2MBW data (A-files from the recording software Spiroware 3.1.6, Eco Medics AG, Duernten, Switzerland) from children aged 5 to 18 years, 10 with CF, 10 formerly preterm born and 10 healthy term born children. The study was approved by the Ethics Committee of the Canton of Bern, Switzerland. The children’s assent was obtained and parents or caregivers provided written informed consent.
Nitrogen multiple-breath washout
We applied a previously described N2MBW setup [13;14] (Exhalyzer D and Spiroware 3.1.6, Eco Medics AG) as recommended by the current consensus and the manufacturer to record raw data [10]. In this device flow was measured by a mainstream ultrasonic flowmeter which derives tidal volumes (Fig 1). Gas concentrations were measured by a side-stream laser O2 sensor and a main-stream infra-red carbon dioxide (CO2) sensor. The N2 fraction was measured indirectly from O2, CO2 and the (estimated) Argon fraction. Flow-gas delay times were based on default settings as recommended by the manufacturer and further individually adjusted based on visual control of the N2 signal. Equipment dead space volume was divided in pre and post sampling-point volumes, defining pre- and post-capillary dead space as represented in Fig 1. Depending on the child’s bodyweight appropriate apparatus dead space reducer was used, set 2 (9.5mL) for children <35kg and set 3 (22mL) for children >35kg, except for the use of set 3 in two preterm children and two children with CF with <35kg bodyweight for logistic reasons.
All children performed triplicate N2MBW according to current consensus [10]. During measurement children were sitting upright, wearing a nose clip and quietly breathing through a snorkel mouth piece. N2MBW was stopped after 3 breaths below 1/40th of N2 starting concentration. We included the first high quality measurement per child for analysis, i.e. one test without visible breathing irregularity or leak [10].
Software settings
All data were recorded, processed and analyzed using Spiroware 3.1.6 (Eco Medics AG). To generate a consistent baseline, original measurements were analyzed with averaged delay times per set (Table 1), averaged ambient temperature and pressure. This was the standard baseline analysis to which we compared MBW results after rerunning measurements with alternative software settings. We calculated LCI and FRC as primary outcomes, and parameters calculated from SIII (Scond and Sacin) as secondary outcomes. Outcomes were calculated as currently recommended [10], LCI from cumulative expired volume used to wash out to 1/40th of initial N2 concentration, divided by FRC.
Table 1. Baseline settings per patient group.
Healthy (n = 10) | Preterm (n = 10) | CF (n = 10) | |
---|---|---|---|
Age (years) | 6.1 ± 0.1 [6.0–6.4] | 7.9 ± 0.8 [6.6–9.2] | 12.3 ± 3.6 [5.1–18.1] |
Dead space reducer set | 2 | 3* | 3* |
Individual delay times | |||
O2 delay (ms) | 601 ± 4 (599–604) | 615 ± 8 (608–620) | 622 ± 5 (619–626) |
CO2 delay (ms) | 51 ± 3 (49–53) | 58 ± 3 (56–60) | 62 ± 2 (61–64) |
Averaged delay times for baseline | |||
O2 delay (ms) | 601 | 618 | 618 |
CO2 delay (ms) | 51 | 60 | 60 |
Individual signal change | |||
O2 delay (ms): ±20, 40, 60 ms | 581, 561, 541 and 621, 641, 661 | 598, 578, 558 and 638, 658, 678 | 598, 578, 558 and 638, 658, 678 |
CO2 delay (ms): ±10 ms | 41 and 61 | 50 and 70 | 50 and 70 |
Data are given as mean ± SD, [range] and (95% CI). Baseline settings per patient group. Individual delay times per file, averaged delay times per dead space reducer set used as baseline, and individual changes for O2 and CO2 delay.
*For logistic reasons at the time of measurement in two children of the preterm group and two children of the CF group we had to use set 3 instead of 2.
We based the magnitude of change in the software settings either on given setting properties or on observed changes in clinical measurements. Accordingly, we inactivated certain settings (e.g. signal correction) or changed baseline setting in both directions (increase and decrease) using realistic steps of at least 5 units (e.g. 5°C ambient temperature or 5 mL dead space) or a relative change of at least 10% (e.g. flow-O2 offset). We categorized our changes into four groups:
1. Environmental conditions
Default ambient temperature was 21°C and pressure 980 hPa. We changed temperature from 21°C to 16°C and 26°C (±5°C or ±24%), and pressure from 980 hPa to 960 hPa and 1000 hPa (±20 hPa or ± 2%). Secondly, BTPS (body temperature pressure saturation) correction was completely switched off.
2. Apparatus dead space
Pre-capillary dead space was 24 mL; post-capillary dead space 9.5 mL in set 2, and 22 mL in set 3. First we altered the default pre- and post-capillary dead spaces separately. Pre-capillary dead space was changed from 24 mL to 19 mL and 29 mL (±5 mL or ±21%); post-capillary dead space from 9.5 mL to 7.5 mL and 11.5 mL (±2 mL or ±21%) in set 2, and from 22 mL to 17 mL and 27 mL (±5 mL or ±23%) in set 3. Secondly, we simultaneously lowered and elevated both pre- and post-capillary dead spaces by ±5 mL, respectively ± 2ml for post-capillary dead space in set 2.
3. Signal processing and breath detection limits
We switched off one by one the default algorithms processing raw gas signals as recommended by the manufacturer and by previous work [13]: (i) We separately and individually deactivated: automated O2-drift correction, dynamic CO2-correction (adjusting the CO2 signal for high O2 fractions), O2 response-time correction (normally set to 30 ms), and the correction for re-inspired N2 volume. (ii) We decreased the sensitivity for breath detection by elevating the required minimum tidal volume from 25 mL to 100 mL. (iii) We increased the cut-offs determining limits for SIII calculation from 65%–95% to 50%–80% of expired volume. The latter limits may include phase II in paediatric tracer gas expirograms [10].
4. Signal delay times
In the current setup side-stream O2 and main-stream CO2 signals must be aligned in time together with the flow signal to allow calculation of the tracer gas N2 volumes as illustrated in Fig 1. Default flow-O2 offset was 601 ms in set 2 and 618 ms in set 3, default flow-CO2 offset was 51 ms in set 2 and 60 ms in set 3 (Table 1). We assessed the susceptibility of measurements towards changes in delay times by the following steps. (i) First, we changed the signal delays individually using realistic steps: For flow-O2 offset from 601 ms in set 2 and from 618 ms in set 3 to ±20 ms (±3%), ±40 ms (±7%) and ±60 ms (±10%); for flow-CO2 offset from 51 ms to ±10 ms (±25%) in set 2, and from 60 ms to ±10 ms (±17%) in set 3 (Table 1). (ii) Secondly, we changed both O2 and CO2 signal delays simultaneously in four separate analyses adding or subtracting the same times as before. (iii) Third, we maintained always a constant delay difference between the two signals in three separate analyses: we changed both O2 and CO2 signal delays identically by -40 ms, +40 ms and +80 ms, respectively.
Statistics
Primary outcome parameters were changes in LCI and FRC. Change in either outcome exceeding 10% was considered relevant as 10% approximately reflects between day-to-day variability of LCI using the same equipment [13;14]. We further assessed changes in parameters calculated from SIII: Scond reflecting convection-dependent ventilation inhomogeneity and Sacin reflecting diffusion-convection-dependent ventilation inhomogeneity. We report raw Scond and Sacin as well as corrected for tidal volume as currently recommended in children [15].
FRC, LCI and Scond and Sacin data were not normally distributed while differences between tests were normally distributed. Therefore data from paired tests were compared by the non-parametric Wilcoxon signed-rank tests. Agreement between settings (differences between paired tests) were assessed by Bland Altman plots [16]. P-values <0.05 were considered statistically significant. All analyses were done using Stata (Stata Statistical Software: Release 11. College Station, TX: StataCorp LP).
Results
We analyzed 30 raw N2MBW files from 10 children with CF (5 boys), 10 former preterm children (6 boys) and 10 healthy children (3 boys). Mean (range) age of children with CF was 12.3 (5.1–18.1) years, 8.0 (6.9–9.2) in former preterm and 6.1 (6.0–6.4) years in healthy children (Table 1).
Environmental conditions
Temperature and pressure settings had small indirect proportional effects on FRC and LCI. Lowering the temperature by 5°C resulted in higher LCI and lower FRC values. Vice versa, 5°C higher temperature resulted in the opposite effect (Table 2). Lowering the pressure by 20 hPa lead to both higher LCI and FRC, while higher pressure lead to lower LCI and FRC (Fig 2). Effect size (mean change) of LCI in all children was lower than 10% (Table 3). Completely inactivating BTPS correction resulted in the same direction of change as lowering temperature with >10% LCI change observed in two children (one CF, one Healthy).
Table 2. Change of N2MBW results after reanalysis with different environmental conditions.
Settings | LCI | FRC | Scond | Scond*VT | Sacin | Sacin*VT |
---|---|---|---|---|---|---|
Temp. lower | ↑↑ | ↓↓ | ↓ | ↓↓ | ↔ | ↔ |
Temp. higher | ↓↓ | ↑↑ | ↑↑ | ↑↑ | (↓) | ↔ |
Pressure lower | ↑↑ | ↑↑ | ↓↓ | ↑↑ | ↓↓ | ↓↓ |
Pressure higher | ↓ | ↓ | ↑ | ↓↓ | ↑↑ | ↑↑ |
BTPS off | ↑↑ | ↓↓ | ↔ | (↓) | ↑↑ | (↑) |
Change of N2MBW test results of 30 children (10 with CF, 10 former preterm and 10 healthy children) after simulation with changed environmental conditions (temperature, pressure and BTPS correction). N2MBW results of the standard test were compared to the test with changed setting parameter by Wilcoxon signed-rank test. We compared lung clearance index (LCI), Scond raw (Scond) and Scond corrected for tidal volume (Scond*VT), Sacin raw (Sacin) and Sacin corrected for tidal volume (Sacin*VT). The arrows reflect the statistical significance (Wilcoxon Signed Rank Test): ↓↓ difference of p < 0.001, ↓ p < 0.01, (↓) p < 0.05, ↔ no significant change.
Table 3. Effect size of the change in LCI after reanalysis with different environmental conditions.
Settings | Changed LCI | % change of standard LCI | # children with 5–10% change | # children with >10% change |
---|---|---|---|---|
Baseline | 8.55 ± 2.64 [6.29–14.94] | |||
Temp. lower | 8.57 ±2.64 [6.30–14.98] | 0.18 ± 0.08 | 0 | 0 |
Temp. higher | 8.54 ± 2.63 [6.27–14.91] | -0.17 ± 0.07 | 0 | 0 |
Pressure lower | 8.56 ± 2.64 [6.29–14.95] | 0.04 ± 0.01 | 0 | 0 |
Pressure higher | 8.59 ± 2.66 [6.28–14.94] | 0.42 ± 1.44 | 1 (3%) | 0 |
BTPS off | 8.77 ± 2.59 [6.49–15.01] | 2.87 ± 4.54 | 5 (16.5%) | 2 (6.5%) |
The table shows mean ± SD [range] LCI after reanalysis of the tests with changed environmental conditions, mean ± SD change in percentage of standard LCI, number of children with a change between 5–10% and children with >10% change of standard LCI.
Apparatus dead space
Dead space settings had small indirect proportional effects on FRC and LCI. Lowering pre- and post-capillary dead space settings separately and simultaneously resulted in higher LCI and FRC (Table 4). Vice versa increased dead space decreased LCI and FRC. Mean effect size on LCI was smaller than 2% and lower than 10% in all children (Tables 4 and 5). Interestingly, measurement bias due to dead space changes was strongly non-linearly associated with FRC, with smaller FRCs being most affected (Fig 3). LCI showed a trend for an analogous but much weaker relationship (data not shown).
Table 4. Change of N2MBW results after reanalysis with different apparatus dead space.
Settings | LCI | FRC | Scond | Scond*VT | Sacin | Sacin*VT |
---|---|---|---|---|---|---|
Precap DS lower | ↑↑ | ↑↑ | ↓ | ↓ | ↔ | ↔ |
Precap DS higher | ↓↓ | ↓↓ | ↑↑ | ↑↑ | (↓) | (↓) |
Postcap DS lower | ↑↑ | ↑↑ | ↓↓ | ↓ | ↑↑ | ↑ |
Postcap DS higher | ↓ | (↓) | ↔ | (↑) | ↔ | ↔ |
Pre- and postcap DS lower | ↑↑ | ↑↑ | ↓ | ↓ | ↑ | ↑ |
Pre- and postcap DS higher | ↓↓ | ↓↓ | ↔ | ↔ | ↔ | ↔ |
Change of N2MBW test results of 30 children after simulation with changed apparatus dead space (pre-capillary dead space (precap DS), post-capillary DS (postcap DS) and both). Results of standard N2MBW test were compared to the test with changed setting parameter by Wilcoxon signed-rank test and marked as follows: ↓↓ difference of p < 0.001, ↓ p < 0.01, (↓)p < 0.05, ↔ no significant change.
Table 5. Effect size of the change in LCI after reanalysis with different apparatus dead space.
Settings | Changed LCI | % change of standard LCI | # children with 5–10% change | # children with >10% change |
---|---|---|---|---|
Baseline | 8.55 ± 2.64 [6.29–14.94] | |||
Precap DS lower | 8.64 ± 2.63 [6.37–15.07] | 1.13 ± 0.57 | 0 | 0 |
Precap DS higher | 8.46 ± 2.64 [6.20–14.82] | -1.14 ± 0.58 | 0 | 0 |
Postcap DS lower | 8.61 ± 2.65 [6.36–15.16] | 0.70 ± 0.69 | 0 | 0 |
Postcap DS higher | 8.52 ± 2.64 [6.20–14.69] | -0.40 ± 1.48 | 0 | 0 |
Pre- and postcap DS lower | 8.70 ± 2.65 [6.44–15.28] | 1.82 ± 0.96 | 0 | 0 |
Pre- and postcap DS higher | 8.40 ± 2.62 [6.13–14.60] | -1.85 ± 0.98 | 0 | 0 |
The table shows mean ± SD [range] LCI after reanalysis of the tests with changed apparatus dead space, mean ± SD change in percentage of standard LCI, number of children with a change between 5–10% and children with >10% change of standard LCI.
Signal processing and detection limits
Inactivating drift correction of the O2 signal marginally increased low LCI and FRC values and decreased high LCI and FRC values which resulted in comparable mean values but showed a clear influence on single measurements (Tables 6 and 7). Inactivation of dynamic CO2 correction leads to an overall increase of LCI and FRC without clear systematic effect. For both inactivation of O2 drift and dynamic CO2 correction 5 children (4 CF, 1 Preterm; and 1CF, 2 Preterm, 2 Healthy) showed >10% LCI change. Inactivation of O2 response-time correction lead to a significant increase in LCI with >10% change in 11 children (2 CF, 6 Preterm, 3 Healthy). Thereby measurement bias was weakly linearly and inversely related to FRC magnitude. Inactivating re-inspired N2 and increasing the sensitivity of breath volume detection lead to heterogeneous changes of LCI and FRC. Changing limits for SIII calculation from 65–95% to 50–80% of expired volume altered Scond non-systematically, while Sacin significantly increased on average by the double. Sacin showed a linear trend for more increase the higher standard Sacin. This effect was more pronounced in raw Sacin than in volume-corrected Sacin (Fig 4).
Table 6. Change of N2MBW results after reanalysis with different signal processing and detection limits.
Settings | LCI | FRC | Scond | Scond*VT | Sacin | Sacin*VT |
---|---|---|---|---|---|---|
O2 drift correction off | ↔ | ↓ | ↔ | (↑) | ↔ | (↓) |
Dynamic CO2 correction off | ↑↑ | ↑↑ | ↓↓ | ↓ | (↑) | ↑ |
O2 response-time correction off | ↑↑ | ↓↓ | ↔ | ↔ | ↑↑ | ↑↑ |
Integrated reinspired N2 off | ↔ | ↔ | ↔ | ↔ | ↔ | (↓) |
Volume Sensitivity higher | ↔ | (↓) | (↓) | (↓) | (↑) | ↑ |
SIII 50–80% | ↔↔ | ↔↔ | ↔ | ↔ | ↑↑ | ↑↑ |
Change of N2MBW test results of 30 children after simulation with changed default algorithms for signal correction. Results of standard N2MBW test were compared to the test with changed setting parameter by Wilcoxon signed-rank test and marked as follows: ↓↓ difference of p < 0.001, ↓ p < 0.01, (↓)p < 0.05, ↔ no significant change, ↔↔ no change at all.
Table 7. Effect size of the change in LCI after reanalysis with different algorithms for signal processing and detection limits.
Settings | Changed LCI | % change of standard LCI | # children with 5–10% change | # children with >10% change |
---|---|---|---|---|
Baseline | 8.55 ± 2.64 [6.29–14.94] | |||
O2 drift correction off | 8.29 ± 2.18 [6.38–14.95] | -1.75 ± 7.95 | 5 (17%) | 5 (17%) |
Dynamic CO2 correction off | 9.07 ± 2.60 [6.65–15.99] | 6.67 ± 5.23 | 15 (50%) | 5 (17%) |
Integrated reinspired N2 off | 8.51 ± 2.57 [6.25–14.76] | -0.41 ± 1.65 | 0 | 0 |
O2 response-time correction off | 9.24 ± 2.71 [6.63–18.05] | 8.65 ± 6.65 | 14 (47%) | 11 (37%) |
Volume Sensitivity higher | 8.57 ± 2.63 [6.25–14.94] | 0.23 ± 0.64 | 0 | 0 |
The table shows mean ± SD [range] LCI after reanalysis of the tests with changed default algorithms for signal correction, mean ± SD change in percentage of standard LCI, number of children with a change between 5–10% and children with >10% change of standard LCI.
Signal delay times
Change of O2 signal delay time settings had a significant impact on LCI and FRC, starting from changes of ±40 ms upwards (Tables 8 and 9). (i) Decreasing flow O2 offset by 40 ms (-7%) lead to a mean LCI increase of 12% and a significant FRC decrease, thereby 60% of the children showed >10% LCI change. The N2 signal typically showed spikes at end-expiration over the last washout breaths (Fig 5A). Elevating the flow O2 offset by 40 ms (+7%) resulted in the opposite effect with a mean LCI decrease of 9% and 43% of the children with >10% LCI change. This resulted in divots at beginning of inspired N2 signal (Fig 5B). Changing flow-CO2 offset showed the opposite effect on LCI and FRC but not exceeding 10%, and without visible change of the N2 signal. (ii) Combining changes in flow-O2 and-CO2 offset resulted in the same trend of changes as for flow-O2 offset alterations alone. (iii) When maintaining a constant delay difference between the two signals, effect sizes of changes were lower, but still significant. As expected, no effect was visible on the N2 signal even in those children with relevant LCI changes. Overall Scond changed non-systematically while Sacin showed the same trend of changes as LCI.
Table 8. Change of N2MBW results after reanalysis with different signal delay times.
Settings | LCI | FRC | Scond | Scond*VT | Sacin | Sacin*VT |
---|---|---|---|---|---|---|
Individual signal changes | ||||||
O2 delay -20ms | ↑↑ | ↓↓ | ↔ | ↔ | ↑↑ | ↑↑ |
O2 delay -40ms | ↑↑ | ↓↓ | ↔ | ↔ | ↑ | ↑↑ |
O2 delay -60ms | ↑↑ | ↓↓ | ↔ | ↔ | ↑↑ | ↑↑ |
O2 delay +20ms | ↓↓ | ↑↑ | ↔ | ↔ | ↓ | ↓ |
O2 delay +40ms | ↓↓ | ↑↑ | (↓) | (↓) | ↓↓ | ↓↓ |
O2 delay +60ms | ↓↓ | ↑↑ | ↓ | ↓ | ↓↓ | ↓↓ |
CO2 delay -10ms | ↓↓ | ↑↑ | (↓) | (↓) | ↔ | ↔ |
CO2 delay +10ms | ↑↑ | ↓↓ | ↑ | ↑ | ↔ | ↔ |
Simultaneous signal changes using different delays | ||||||
O2 -100ms and CO2 -10ms | ↑↑ | ↓↓ | ↔ | ↔ | ↑↑ | ↑↑ |
O2 +100ms and CO2 +10ms | ↓↓ | ↑↑ | ↓ | ↓ | ↓↓ | ↓↓ |
O2 -100ms and CO2 +10ms | ↑↑ | ↓↓ | ↔ | ↔ | ↑↑ | ↑↑ |
O2 +100ms and CO2 -10ms | ↓↓ | ↑↑ | ↓↓ | ↓↓ | ↓↓ | ↓↓ |
Simultaneous signal changes using similar delays | ||||||
O2 -40ms and CO2 -40ms | ↑↑ | ↓↓ | ↔ | ↔ | ↑ | ↑↑ |
O2 +40ms and CO2 +40ms | ↓↓ | ↑↑ | ↔ | ↔ | ↓↓ | ↓↓ |
O2 +80ms and CO2 +80ms | ↓↓ | ↑↑ | ↔ | ↔ | ↓↓ | ↓↓ |
Change of N2MBW test results of 30 children after simulation with changed signal delay times. Results of standard N2MBW test were compared to the test with changed setting parameter by Wilcoxon signed-rank test and marked as follows: ↓↓ difference of p < 0.001, ↓ p < 0.01, (↓) p < 0.05, ↔ no significant change. Of note, for simultaneous signal changes using similar time steps, delays between O2 and CO2 sensor were kept constant.
Table 9. Effect size of the change in LCI after reanalysis with different signal delay times.
Settings | Changed LCI | % change of standard LCI | # children with 5–10% change | # children with>10% change |
---|---|---|---|---|
Baseline | 8.55 ± 2.64 [6.29–14.94] | |||
Individual signal changes | ||||
O2 delay -20ms | 9.03 ± 2.79 [6.55–16.82] | 5.68 ± 3.85 | 16 (53%) | 3 (10%) |
O2 delay -40ms | 9.53 ± 2.86 [6.69–19.20] | 11.98 ± 8.21 | 7 (23%) | 18 (60%) |
O2 delay -60ms | 10.33 ± 3.20 [6.83–21.30] | 21.37 ±12.93 | 4 (13%) | 25 (83%) |
O2 delay +20ms | 8.13 ± 2.46 [5.93–13.54] | -4.87 ± 2.33 | 13 (43%) | 1 (3%) |
O2 delay +40ms | 7.76 ± 2.35 [5.55–12.85] | -9.12 ± 3.46 | 15 (50%) | 13 (43%) |
O2 delay +60ms | 7.47 ± 2.22 [5.27–12.36] | -12.47 ± 4.71 | 7 (23%) | 21 (70%) |
CO2 delay -10ms | 8.48 ± 2.60 [6.23–14.50] | -0.86 ± 1.22 | 0 | 0 |
CO2 delay +10ms | 8.62 ± 2.69 [6.32–15.25] | 0.65 ± 1.06 | 0 | 0 |
Simultaneous signal changes using different delays | ||||
O2 -100ms, CO2 -10ms | 13.88 ± 6.84 [7.23–40.43] | 61.48 ±54.85 | 0 | 30 (100%) |
O2 +100ms, CO2 +10ms | 7.14 ± 2.20 [5.04–12.03] | -16.38 ± 6.21 | 6 (20%) | 24 (80%) |
O2 -100ms, CO2 +10ms | 14.96 ± 8.61 [7.30–46.88] | 73.89 ±79.56 | 0 | 30 (100%) |
O2 +100ms, CO2 -10ms | 7.04 ± 2.16 [4.98–11.81] | -17.57 ± 6.33 | 4 (13%) | 26 (87%) |
Simultaneous signal changes using similar delays | ||||
O2 -40ms, CO2 -40ms | 9.28 ± 2.81 [6.61–17.29] | 8.79 ± 6.74 | 9 (30%) | 14 (47%) |
O2 +40ms, CO2 +40ms | 7.95 ± 2.47 [5.72–13.49] | -7.12 ± 3.36 | 18 (60%) | 6 (20%) |
O2 +80ms, CO2 +80ms | 7.49 ± 2.36 [5.33–12.69] | -12.56 ± 4.68 | 8 (27%) | 21 (70%) |
The table shows mean ± SD [range] LCI after reanalysis of the tests with changed signal delay times, mean ± SD change in percentage of standard LCI, number of children with a change between 5–10% and children with >10% change of standard LCI. Of note, for simultaneous signal changes using similar time steps, delays between O2 and CO2 sensor were kept constant.
Taken together, signal delay times had the largest impact upon the primary outcomes LCI (Table 10), FRC, and secondary outcomes Scond and Sacin. Thereby changes in O2 delay were relatively more sensitive than changes in CO2 delay; while 7% (40 ms) change in O2 delay showed a significant impact, 17–25% (10 ms) change in CO2 delay had no significant effect. The largest effect on Sacin was caused by changed limits for SIII calculation doubling Sacin values on average. Except for inactivation of O2 drift or reinspired N2 leading to different changes among disease groups, changes of software factors lead to the same trend of LCI and FRC changes in all children independent of disease state. This uniformity often generated an overall statistically significant difference which was not necessarily clinically relevant (LCI change >10%) in all subjects (Table 10).
Table 10. Summary of the impact of different software settings on LCI of N2MBW.
LCI change | Statistical significance | Clinical relevance | Max. No. of children with>10% change |
---|---|---|---|
Environmental conditions | Yes | No | 2 (7%) |
Apparatus dead space | Yes | No | 0 |
Signal processing and detection limits | Only for algorithms concerning O2 and CO2 signal processing | Yes, for those mentioned | 11 (37%) |
Signal delay times | Yes, for O2 delay changes of ±40 ms upwards | Yes | 18 (60%) |
Overview of impact on LCI after change of software settings. High quality N2MBW tests of 30 children with different lung disease were reanalyzed with changed software settings. Several parameters were changed according to four sectors, environmental conditions (e.g. ambient temperature), setup parameter (apparatus dead space), software algorithm (e.g. O2 response-time correction), and flow-gas signal delay times. Because of the uniformity of change statistical significance might be reached over the study population without clinical relevance (defined as >10% change of baseline LCI for the individual measurement; and for the change in software setting if >10% of all children showed such a LCI change >10%).
Discussion
This is the first study that systematically examines the implication of different software settings on washout results in a commercially available N2MBW software. We find that incorrect software settings introduce significant measurement bias with different effect sizes of various settings. While some settings (e.g. environmental settings) introduce only small random measurement errors, other software settings show significant impact upon results. The most susceptible settings were signal alignments, realistic changes in flow-O2 signal delay beyond ±40 ms resulted in >10% LCI change in 60% of the children for -40 ms O2 change and in 43% of the children for +40 ms O2 change. Incorrect apparatus dead space estimates may especially bias smaller FRC measurements in young children. While some of the current findings are specific for this setup, principles are also valid for other MBW devices currently available. Especially, signal alignment is done in all MBW setups.
The current data draws attention to important software settings which have not been reported in detail before. Various efforts have been made for validation of available washout setups using a realistic lung model for FRC measurement [13;17;18] or by comparison to mass spectrometry [19–22] using optimized software settings. While the ERS/ATS consensus statement gives recommendations for the optimal hard- and software there is only little data at what point technical factors lead to relevant measurement errors [11;12].
There is preliminary evidence that change of settings over time in two different software releases of the same software lead to different FRC and LCI results as seen in infant MBW using sulfur hexafluoride (SF6) [23]. Two other studies simulated specifically flow gas misalignment in different washout setups, both using 10 ms steps from -50 ms to +50 ms [24;25]. While Horsley et al. simulated flow-gas misalignment in a lung model using SF6 washouts with an Innocor gas analyzer (Innovision, Odense, Denmark), Buess et al. analyzed raw data files from N2MBW (ndd Medical Technologies, Switzerland) [24;25]. Both found an almost linear increase of FRC between ±2.5% and ±7.5% over the -50 to +50 ms delay-time change and a higher FRC error with higher breathing frequency. While different setups and simulation conditions hamper direct translation of results, direction and effect size seem comparable to FRC errors in our study. In any case, all those studies point towards the importance of precise flow and gas signal alignment on lung volume calculations particularly in young children with faster ventilation. This has now been shown for all commercially available MBW setups.
One limitation of our study is that we simulated technical measurement errors rather than having the subjects themselves performing the test repetitively under changed conditions. Change of certain settings might have a different impact on real-time washout measurement compared to simulated tests, e.g. the change of post-capillary dead space would be associated with a change of signal delay times concurrently. On the other hand we based our simulations on the reload of raw, unprocessed storage files of the original measurement. Thus we are confident that results reflect real-life impact. To enable multiple reloads of the tests, we used averaged delay times per set. As we did not change the setup, individual delay times of the tests varied only minimally within a narrow range (Table 1). In addition we confirmed proper flow-gas alignment for each test by visual control of the N2 signal shape. Thus, we believe this approach does not impede validity of results. Another limitation is that results are specific for the device used in our study. However only certain findings do not apply to other pieces of equipment, like the change of software specific algorithms. General findings such as for signal alignement and BTPS correction are not specific to this apparatus. Sampling flow and gas concentration at different points will always result in delay between signals. Most of the times this delay is even flow-dependent [26]. This applies to all available and customized washout setups such as ultrasonic flowmeter based MBW using either N2 [14] or SF6 [27], Innocor gas analyser using SF6 [18] and mass spectrometer using SF6 [28]. As mentioned above this is underlined by comparable findings of incorrect flow-gas delays on MBW results also for other devices [18;24;25].
We only tested single relevant changes for each software factor within our heterogeneous study population. Thus we could not assess the effect over the complete range of the factor, or define clear relationships with outcome parameters. The Bland-Altman plots suggest nonlinearity for many parameters. However the primary aim of our study was to tease out the most important software setting by including measurement of good quality [10] in a wide age range of children with different lung disease. The majority of our findings were consistent throughout the study population resulting in the same trend of change for all children. Depending on the role of software settings within the algorithm for outcome calculation, the effects of changes of software settings affecting primarily volume measurements were independent of underlying LCI (e.g. BTPS, temperature) while the effect of changes in settings primarily affecting gas concentrations (e.g. reinspired N2) was of course related to underlying LCI (e.g. reinspired N2).
The most relevant finding was the significant impact of flow-O2 signal misalignment on FRC and LCI results. The same applied for Scond and Sacin, with >10% change of their baseline values. The O2 signal mainly determines the tracer gas signal (N2) in the current setup. In order to derive FRC from MBW measurements, expired gas volumes are calculated by integrating the gas signal with flow. The same applies for measurement of respiratory dead space [26;29]. This requires precise alignment of gas and flow signals [30]. In the N2MBW setup we encounter two additional challenges for this alignment. During N2 washout gas composition changes due to increasing O2 concentration and viscosity [30]. This leads to changing delay times even in the side-stream sampling of O2 over the period of the washout. Moreover mainstream flow is not constant because of breathing cycles. This will influence signal alignment furthermore. So far, flow gas delay was calculated as fixed correction factor for the whole measurement. The susceptibility of this flow gas synchronization and the clinical implication in case of misalignment as shown in this study suggests strongly that this algorithm is prone to errors. Whether a dynamic flow-gas delay correction is superior, especially at younger age, in which these errors are more relevant [24;25], needs to be examined in future studies. Simulation of changes in equipment dead space also showed more impact in young children, where dead space is large relative to lung volumes. This is in line with recently published clinical data showing higher LCI values with increasing equipment dead space especially in young children [31].
Technical quality control of washout tests gets particularly important for longitudinal studies with potentially changing software settings. In the current setup, different signal processing settings (such as O2-drift correction, dynamic CO2 correction or correction for re-inspired N2 volume) were implemented continuously over three years. Moreover the storage of unprocessed raw data makes it possible to rerun original measurement with changed settings. We now know that the most important technical factor to check for the operator is correct signal alignment throughout the entire washout while other factors such as small deviations from environmental settings or BTPS correction only have small and clinically negligible impact. Unfortunately, if the N2 signal shows relevant flow-gas misalignment, it gets clearly visible to the operator only over the last washout breaths. Prior to software updates the companies should provide data showing the new software’s impact (if any) on MBW outcomes. This is recommended by the current consensus [10] and would enable the user to account for possible measurement bias, which seems especially important for repeated MBW measurements in longitudinal studies.
To sum up, we found that software settings have clear impacts upon MBW results. The most important technical factor is flow gas signal alignment. Particularly inaccurate flow-O2 offset in N2MBW can lead to wrongly elevated or false normal LCI in a wide age range of children with different lung disease. Whether a flow-adapted new algorithm for signal synchronization will lead to more robust results needs to be examined in the future.
Supporting Information
Acknowledgments
The authors would like to thank all children and their families for their participation in the study. We express our gratitude to Carmen Casaulta, Sylvia Nyilas, Anne Schmidt, Monika Graf, Lina Beul-Béguin, Sandra Luescher, Gisela Wirz and Sharon Schmid (University Children’s Hospital Bern and Basel) for their valuable work and support.
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
This study was supported by the Swiss Society for Cystic Fibrosis (CFCH), the Gottfried und Julia Bangerter-Rhyner-Stiftung and the FondationBotnar.
References
- 1. Aurora P. Multiple-breath inert gas washout test and early cystic fibrosis lung disease. Thorax 2010;65: 373–374. 10.1136/thx.2009.132100 [DOI] [PubMed] [Google Scholar]
- 2. Lum S, Gustafsson P, Ljungberg H, Hulskamp G, Bush A, Carr SB, et al. Early detection of cystic fibrosis lung disease: multiple-breath washout versus raised volume tests. Thorax 2007;62: 341–347. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Boon M, Vermeulen FL, Gysemans W, Proesmans M, Jorissen M, De Boeck K. Lung structure-function correlation in patients with primary ciliary dyskinesia. Thorax 2015;70: 339–345. 10.1136/thoraxjnl-2014-206578 [DOI] [PubMed] [Google Scholar]
- 4. Amin R, Subbarao P, Jabar A, Balkovec S, Jensen R, Kerrigan S, et al. Hypertonic saline improves the LCI in paediatric patients with CF with normal lung function. Thorax 2010;65: 379–383. 10.1136/thx.2009.125831 [DOI] [PubMed] [Google Scholar]
- 5. Amin R, Subbarao P, Lou W, Jabar A, Balkovec S, Jensen R, et al. The effect of dornase alfa on ventilation Inhomogeneity in patients with Cystic Fibrosis. Eur Respir J 2011;37: 806–812. 10.1183/09031936.00072510 [DOI] [PubMed] [Google Scholar]
- 6. Subbarao P, Stanojevic S, Brown M, Jensen R, Rosenfeld M, Davis S, et al. Lung Clearance Index as an Outcome Measure for Clinical Trials in Young Children with Cystic Fibrosis: A Pilot Study using Inhaled Hypertonic Saline. Am J Respir Crit Care 2013;188: 456–460. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Gustafsson PM, De Jong PA, Tiddens HA, Lindblad A. Multiple-breath inert gas washout and spirometry versus structural lung disease in cystic fibrosis. Thorax 2008;63: 129–134. [DOI] [PubMed] [Google Scholar]
- 8. Vermeulen F, Proesmans M, Boon M, Havermans T, De BK. Lung clearance index predicts pulmonary exacerbations in young patients with cystic fibrosis. Thorax 2014;69: 39–45. 10.1136/thoraxjnl-2013-203807 [DOI] [PubMed] [Google Scholar]
- 9. Kent L, Reix P, Innes JA, Zielen S, Le BM, Braggion C, et al. Lung clearance index: evidence for use in clinical trials in cystic fibrosis. J Cyst Fibros 2014;13: 123–138. 10.1016/j.jcf.2013.09.005 [DOI] [PubMed] [Google Scholar]
- 10. Robinson PD, Latzin P, Verbanck S, Hall GL, Horsley A, Gappa M, et al. Consensus statement for inert gas washout measurement using multiple- and single- breath tests. Eur Respir J 2013;41: 507–522. 10.1183/09031936.00069712 [DOI] [PubMed] [Google Scholar]
- 11. Arieli R, Van Liew HD. Corrections for the response time and delay of mass spectrometers. J Appl Physiol 1981;51: 1417–1422. [DOI] [PubMed] [Google Scholar]
- 12. Farmery AD, Hahn CE. Response-time enhancement of a clinical gas analyzer facilitates measurement of breath-by-breath gas exchange. J Appl Physiol 2000;89: 581–589. [DOI] [PubMed] [Google Scholar]
- 13. Singer F, Houltz B, Latzin P, Robinson P, Gustafsson P. A realistic validation study of a new nitrogen multiple-breath washout system. PLoS ONE 2012;7: e36083 10.1371/journal.pone.0036083 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Singer F, Kieninger E, Abbas C, Yammine S, Fuchs O, Proietti E, et al. Practicability of nitrogen multiple-breath washout measurements in a pediatric cystic fibrosis outpatient setting. Pediatr Pulmonol 2013;48: 739–746. 10.1002/ppul.22651 [DOI] [PubMed] [Google Scholar]
- 15. Aurora P, Kozlowska W, Stocks J. Gas mixing efficiency from birth to adulthood measured by multiple-breath washout. Respir Physiol Neurobiol 2005;148: 125–139. [DOI] [PubMed] [Google Scholar]
- 16. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1: 307–310. [PubMed] [Google Scholar]
- 17.Schmidt A, Yammine S, Proietti E, Frey U, Latzin P, Riedel T, et al. Validation of multiple-breath washout equipment for infants and young children. Pediatr Pulmonol 2014 Feb 25. 10.1002/ppul.23010 [DOI] [PubMed]
- 18. Gonem S, Singer F, Corkill S, Singapuri A, Siddiqui S, Gustafsson P. Validation of a photoacoustic gas analyser for the measurement of functional residual capacity using multiple-breath inert gas washout. Respiration 2014;87: 462–468. 10.1159/000357786 [DOI] [PubMed] [Google Scholar]
- 19. Pillow JJ, Ljungberg H, Hulskamp G, Stocks J. Functional residual capacity measurements in healthy infants: ultrasonic flow meter versus a mass spectrometer. Eur Respir J 2004;23: 763–768. [DOI] [PubMed] [Google Scholar]
- 20. Fuchs SI, Buess C, Lum S, Kozlowska W, Stocks J, Gappa M. Multiple breath washout with a sidestream ultrasonic flow sensor and mass spectrometry: a comparative study. Pediatr Pulmonol 2006;41: 1218–1225. [DOI] [PubMed] [Google Scholar]
- 21. Fuchs SI, Sturz J, Junge S, Ballmann M, Gappa M. A novel sidestream ultrasonic flow sensor for multiple breath washout in children. Pediatr Pulmonol 2008;43:731–738. 10.1002/ppul.20825 [DOI] [PubMed] [Google Scholar]
- 22. Jensen R, Stanojevic S, Gibney K, Salazar JG, Gustafsson P, Subbarao P, et al. Multiple breath nitrogen washout: a feasible alternative to mass spectrometry. PLoS ONE 2013;8: e56868 10.1371/journal.pone.0056868 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Ebdon AM, Skoric B, Robertson C, Ranganthan S. Comparing analysis software versions in infant multiple breath washout. Eur Respir J 2013;42: 228s. [Google Scholar]
- 24. Horsley A, Macleod K, Gupta R, Goddard N, Bell N. Enhanced photoacoustic gas analyser response time and impact on accuracy at fast ventilation rates during multiple breath washout. PLoS ONE 2014;9: e98487 10.1371/journal.pone.0098487 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Buess C, Fuchs S, Gappa M. Influence of tracer gas delay-time on FRC and Lung clearance index from Multiple-breath nitrogen washout. Am J Respir Crit Care Med 2012;185: A1691. [Google Scholar]
- 26. Thamrin C, Latzin P, Sauteur L, Riedel T, Hall GL, Frey U. Deadspace estimation from CO2 versus molar mass measurements in infants. Pediatr Pulmonol 2007;42: 920–927. [DOI] [PubMed] [Google Scholar]
- 27. Fuchs SI, Eder J, Ellemunter H, Gappa M. Lung clearance index: normal values, repeatability, and reproducibility in healthy children and adolescents. Pediatr Pulmonol 2009;44: 1180–1185. 10.1002/ppul.21093 [DOI] [PubMed] [Google Scholar]
- 28. Gustafsson PM, Aurora P, Lindblad A. Evaluation of ventilation maldistribution as an early indicator of lung disease in children with cystic fibrosis. Eur Respir J 2003;22:972–979. [DOI] [PubMed] [Google Scholar]
- 29. Tang Y, Turner MJ, Baker AB. Effects of lung time constant, gas analyser delay and rise time on measurements of respiratory dead-space. Physiol Meas 2005;26: 1103–1114. [DOI] [PubMed] [Google Scholar]
- 30. Brunner JX, Wolff G, Cumming G, Langenstein H. Accurate measurement of N2 volumes during N2 washout requires dynamic adjustment of delay time. J Appl Physiol 1985;59: 1008–1012. [DOI] [PubMed] [Google Scholar]
- 31. Benseler A, Stanojevic S, Jensen R, Gustafsson P, Ratjen F. Effect of equipment dead space on multiple breath washout measures. Respirology 2015;20: 459–466. 10.1111/resp.12470 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data