Abstract
High—speed atomic force microscopy has proven to be a valuable tool for the study of biomolecular systems at the nanoscale. Expanding its application to larger biological specimens such as membranes or cells has, however, proven difficult, often requiring fundamental changes in the AFM instrument. Here we show a way to utilize conventional AFM instrumentation with minor alterations to perform high-speed AFM imaging with a large scan range. Using a two—actuator design with adapted control systems, a 130 × 130 × 5 μm scanner with nearly 100 kHz open—loop small-signal Z—bandwidth is implemented. This allows for high-speed imaging of biologically relevant samples as well as high-speed measurements of nanomechanical surface properties. We demonstrate the system performance by real-time imaging of the effect of charged polymer nanoparticles on the integrity of lipid membranes at high imaging speeds and peak force tapping measurements at 32 kHz peak force rate.
In recent years, high-speed AFM (HS-AFM) has been demonstrated to be very powerful for studying biological systems at the nanoscale1,2. These studies were made possible by a continuous improvement in HS-AFM instrumentation toward even higher scan speeds and better feedback performance3,4,5. Current HS-AFMs are highly specialized, complex instruments tailored to measuring the dynamics of nanoscale systems such as molecular motors6, membrane proteins7,8, or antibodies9 with temporal imaging rates of less than one second per image. These measurements have stirred high hopes for advancements in other bio- and nanotechnology fields as well. However, many applications of AFM in cellular biology10,11,12 and materials science13,14 require scan sizes of 10 μm to 100 μm rather than the hundreds of nanometres of current HS-AFMs. In order to make HS-AFM available to the broad nanotechnology community it is therefore essential to enable large scan ranges in all directions at high speeds15,16,17. In addition, AFM has evolved away from a tool for dedicated experts into a routine tool where many users only operate the instruments occasionally. The high degree of complexity of current HS-AFMs is therefore still prohibitive for broader adoption. A HS-AFM for the general nanotechnology community should not be more difficult to use than a conventional AFM. These two requirements put especially high demand onto the scanner and its control. Increasing the scan size at a fixed scan rate results in an increased tip surface speed. Therefore, the spatial frequency of sample topography translates to higher temporal frequencies when scanning a larger area. Imaging at a larger scan size thus not only puts more burden on the X-Y axes, but also requires a very high bandwidth of the feedback in general and the Z-scanner in particular. In this paper, we describe an approach combining modification of a conventional AFM scanner with model-based control in all three directions. This leads to a total scan range of 130 × 130 × 5 μm in X, Y, and Z respectively, while maintaining a flat small—signal frequency response up to 90 kHz in Z—direction.
Traditionally, the main limitation for scanners has been the excitation of mechanical resonances by the scanner drive signal (see Fig. 1). Piezoelectric tube scanners which are predominantly used in AFM exhibit a large number of different resonant modes, of which the first lateral and the first length extensional modes are of primarily influence on imaging performance (see Fig. 1b).
Lateral resonances are excited when the scan direction is reversed at the end of a line; the sharp turnaround makes the tube structure ring (oscillate) in the direction of the fast scan axis. This ringing is visible as horizontal waves in the image, caused by a coupling of the X-Y resonance into Z-motion due to sample tilt and the well-known scanner bow effect of tube scanners (Fig. 1b inset). Less obvious, but just as severe is that the lateral ringing will severely displace the scan position, making vertical strips in the image appear alternately stretched and compressed18 (see original edge displacement in Fig. 1a). The length extensional resonance (Z-ringing) is visible as horizontal ripples that follow sharp steps in the topography.
Control systems have been used extensively in the past to extend the speed of available scanners19,20. This approach has shown to be a straightforward way to improve the performance of existing mechanical designs with electronics, but is ultimately still limited by the decreasing transducer efficiency at higher frequencies past their resonance. To push the resonance frequencies to higher values, a number of mechanical designs for high—speed large—range scanners have been proposed. Most high—speed scanners are flexure based designs16,21,22 and often use control techniques to deal with resonances23. These scanners are fast, but are typically limited in range, as they trade scan range for speed by decreasing the size of the actuators. More recently, feedback using two different Z-actuators has been shown to offer a promising increase in speed24,25. The tip and the sample are moved with different actuators, where one actuator is lower range but faster. Corresponding control systems have been proposed26. This approach is fast, but no longer provides a monolithic scanner and requires the user to interface with two actuators.
We propose a solution for a large range, high—speed system through simple modifications of a commercial system to move the sample with two different actuators in mechanical series and extending the control loop with model-based control for this scanner.
Mechanical modifications
HS-AFM is primarily enabled on the detection side by the use of small cantilevers with high resonance frequencies27 and on the actuation side by scanners with high bandwidth1,3 together with appropriate controllers28. These technologies generally require a complete redesign of the instrument. Recently, we reported an open-source add-on to a commercial instrument (Bruker Multimode 8) that allows the use of very small cantilevers. The replacement AFM head has a detection bandwidth for resonance frequencies up to 20 MHz29,30 and thereby allows fast topography detection on a standard AFM system.
Here we extend this system by enabling high-speed actuation through a simple addition to the standard piezoelectric tube scanner. We place a secondary stack piezo actuator on top of the tube scanner (see Fig. 2b). The piezo stack has a significantly higher resonance frequency than the tube, due to its small size and higher rigidity. The tube scanner has a scan range of 130 × 130 μm and a Z range of 5 μm. The stack is mounted in a custom flexure housing that acts as a flexure guide for uniaxial Z-motion, fluid sealing of the stack piezo and integrated preload, providing a sample movement of 1.5 μm. A silicone O-ring (Bruker FCO-10) can be seated on the housing for fluid imaging in a closed fluid cell. The sample is mounted on the radial flexure in the top part (titanium grade 5) of the housing (see Fig. 2c).
Using a second actuator on top of the original tube influences the dynamic behaviour of the whole scanner. Instead of the simple harmonic oscillator (SHO) behaviour of a single tube, the system acts as a coupled two mass oscillator (see Fig. 2d) with two inputs and therefore two transfer functions. By frequency domain system modelling (see appendix), we find the tube actuator to behave like an SHO with added mass, while the stiff stack actuator on top can resonantly couple into the much softer tube (Fig. 2e). The lateral resonances of the scanner are only affected insofar as far as that there is additional mass, which slightly lowers the lateral resonance frequency.
Control systems
Suppression of SHO-like scanner resonances, both lateral as well as axial, has been studied extensively in the past. Using a first order notch filter, the resonant behaviour of the tube can be anticipated and counteracted before the signal is applied to the actuators. Lateral resonances can be nearly completely supressed, and this approach has been shown to offer significant tracking-speed improvements over uncompensated systems, as the scanner can be used to a bandwidth close to the mechanical resonance frequency31,32,33.
Since the combined scanner has two actuators in Z-direction, it can be considered as a multiple—input (two piezo drives) single—output (sample height) system, and can achieve better overall performance by using both inputs. By balancing control authority over two actuators we can achieve the combination of high bandwidth and large range simultaneously. Several schemes exist in literature to utilize two actuators to control for one distance34,35, of which several have been used for AFM in the past36,37. The most direct approach to dual actuation with reduced range on one fast actuator is nested PID feedback. A high—frequency primary PID to the fast actuator tracks topography while a secondary slower loop keeps the mean of the fast signal centred. While simple in its implementation, this approach is not robust and cannot deal with coupled dynamics and requires extensive user interaction38. Other implementations that use a model based approach for the whole Z-feedback loop can achieve excellent performance and guaranteed stability39, but do not leave room to tune the feedback manually to get the best performance for each sample. We therefore chose an approach that is transparent for the AFM controller and does not require any change from standard AFM operating procedures.
To split a signal conservatively across the frequency spectrum, audio systems have long since utilized frequency crossovers. We use the same design to split our signals into low— and high—frequency components. We use the high resonance frequency of the stack actuator to transduce high frequency components, which generally have smaller amplitudes. Low frequency movements are sent to the tube actuator with its longer range.
Models are a requirement for dynamic filtering of dynamics. We use the fact that the AFM itself is a fast, precise sensor that is already in place to measure the dynamics of our system without additional hardware modifications. We can record the transfer functions of both Z actuators by exciting the respective actuator and measuring the deflection of a cantilever in contact mode. Burns et al. have presented a method that enables to also measure lateral dynamics directly with the cantilever. By exciting the turnaround and measuring the ringdown in the X—Y to Z coupling, the SHO resonance and damping can be extracted without the use of additional sensors31.
The tube roll—off of the cross—over is set to start well below the resonance of the tube (400 Hz) and residual effects of the first order resonance of the tube can be filtered with a simple notch filter at about 8 kHz. The first resonance of the stack actuator can be cancelled in a similar way with a notch filter at ca. 80 kHz. However, the coupling dynamics of the stack actuator into the resonances of the tube is more involved. Using model inversion, we filter the stack signal in elementary blocks, each compensating a resonant coupling (coupling filters) in addition to the notch filter to suppress the stack resonance (Fig. 3a) (See supplementary information for the models used for fitting). All filters are implemented as a sequence of second order sections on field programmable gate arrays (FPGAs) (National Instruments, X-Filter: NI PXI-7851R, Z-Filters: NI 7954R/NI 5781). The measured small—signal transfer functions of the systems as well as their corresponding filters are shown in Fig. 3b. Due to the frequency crossover, the second tube resonance is already strongly attenuated, and does not require an additional notch filter. The resulting small—signal open—loop response of the whole system is flat within ±3 dB to 90 kHz and results in a nearly 20 times faster closed loop response than the unmodified tube scanner could achieve. With our current implementation the system delay, in large part caused by analog—to—digital conversions and digital—to—analog conversions, restricts the non-peaking closed loop operation (Bessel type) to about 35 kHz. For imaging purposes a faster operation can be achieved by allowing some frequency domain peaking.
Lateral compensation (X-Y ringing) requires only a notch filter per resonance; usually only the fast scan direction requires compensation. The method for indirect system identification described by Burns et al.31 is used to determine the model parameters. The resonance frequency and damping is extracted from the ring down in the X-Y to Z coupling after turnaround (compare vertical waves in Fig. 1a).
To illustrate the effect of the individual filters, we have imaged a scratched muscovite mica sample at a moderate 4 Hz line rate in contact mode AFM with the feedback gains set to just avoid feedback oscillation. At these conditions we increase the line rate to 166 Hz (see Fig. 4a,b). While continuously scanning at high rate, we first enable the X-compensation to supress the lateral resonances, removing positional artefacts and the waviness in height (Fig. 4c). Enabling the interoperating two actuators removes the Z-ringing (Fig. 4d). Suppressing the Z-resonances and bypassing their associated phase delays allows the use of much more aggressive PID settings. Increasing the feedback bandwidth by choosing higher PID gains allows us to use the full bandwidth of the scanner and restore the imaging output to nearly the same quality as the original slow imaging (Fig. 4e).
Results
High—speed imaging of lipid detachment by charged dendrimers
To validate the capabilities of our system, we chose to study the interaction of positively charged dendrimer nanoparticles with lipid bilayers. The interaction of nanoparticles with biological membrane is an interesting question, both from an environmental, toxicological perspective40,41,42 as well as from a potentially therapeutic perspective43,44,45,46. In both cases, how the nano-particles interact with the membrane is critical in determining their ultimate effect. The combination of supported lipid bilayers and AFM provide an ideal combination to directly observe the interactions of nano-particles with the membrane in a simplified, controllable setting47,48,49,50. The mixed DLPC-DPPC bilayer used in our system allows for direct observation of the influence of lipid phase on the interaction with the positively charged nanoparticles, since both DLPC and DPPC have the same chemical head group, but have different melting temperatures. At room temperature (where we performed our experiments), the DLPC is in the fluid phase and the DPPC is in the gel phase. To avoid excessive heating of the cantilever and sample we reduced the laser power of the 635 nm laser diode in our custom AFM head29,30 to just above the lasing threshold.
This experiment is also an ideal test system for HS-AFM, because any remaining, uncompensated dynamics would result in nanometre sized height distortions that would clearly exceed the height difference between DLPC and DPPC (ca. 0.12 nm). The delicate structure of the bilayer requires excellent force control, which in turn requires a high Z-bandwidth. In previous AFM experiment on this system, the long time between frames (2–5 min) has limited the understanding of the mechanism, as the whole interaction happens inside one image. HS-AFM has been sucessfully used to study dynamic effects on lipid bilayers in the past51,52 and allows us to dramatically reduce the time between images in this experiment (29 seconds, 512 × 512 pixels, 17.6 lines/s) while scanning a relatively large area (5 μm). For this experiment, feedback was limited by the bandwidth of the cantilever in fluid (ca. 30 kHz), rather than the scanner itself. The additional temporal resolution has lead to the discovery of two key mechanistic results. First, the fluid DLPC is easier to disrupt than the gel DPPC. Initial defects appear only in the fluid region (DLPC), and until ca. 400 seconds after nanoparticle addition, the amount of DLPC removed outpaces that of DPPC. The second result is that the DPPC removal occurs almost exclusively after the removal of the surrounding DLPC proceeding from the destabilized edge inward. This highlights that the dendrimers are more likely to disrupt already weakened areas due to a lower activation barrier. Both of these observations would have been extremely difficult to make with conventional AFM because a single image could easily have taken more than 500 seconds completely obscuring these features. Our initial results hold promise for more mechanistic studies of membrane-nanoparticle interactions based on the measurement of kinetics and disruption patterns (Fig. 5). Permutations of charges both on the polymers and lipids will elucidate the different preferences in interaction. By adding sterols and other membrane integrating components one can probe their stabilizing and destabilizing effects. The large scan range (130 μm in X and Y) also enables the imaging of rare structures in native membranes that one could otherwise never find with the small scan sizes of current HS-AFMs. To transitioning form supported bilayers to suspended bilayer structures careful force control will be even more important, to investigate the role of lipid curvature on stability.
High—rate peak force tapping
The need for mechanical property characterization at the nanoscale has accelerated the use of off-resonance AFM modes in recent years53,54,55,56 and offers very promising applications to cell biology57,58. In these modes, the cantilever is moved up and down relative to the sample in a controlled trajectory. One implementation of the off-resonance modes, peak force tapping (PFT), uses scanner motion to perform approach—retract curves at rapid rates. A sinusoidal motion is superimposed on the feedback signal (Fig. 6a). The peak interaction force is extracted from each curve and used as feedback signal for the PID controller. This approach has the benefit that it offers very good force control, as well as ease of use53,59,60. Since it does not depend on exciting the cantilever resonances it is well suited for automated image acquisition and gain setting. Additionally, on—line model fitting can be used to extract and display materials properties such as Young’s modulus or adhesion in real time.
With conventional scanners, the PFT rate is limited by the scanner resonance, as the frequency of the sinusoidal modulation has to stay well below the scanners resonance frequency to avoid distorted motion (see orange arrow in Fig. 6b). Most commercial systems run at 2 kHz PFT rate, high—end systems can reach up to 8 kHz rates, which makes PFT generally slower than contact mode or tapping modes. The imaging speed in PFT is directly dependent upon the PFT rate, as the number of taps per pixel defines the feedback performance (Fig. 6b).
Using the fast flat response of our two—actuator scanner, we were able to operate at a 31.5 kHz PFT rate, enabling much higher line rates (6c). Even at only one tap per pixel an image can still be aquired at reasonable quality. We are currently limited by the 500 kHz sampling rate of our controller, as the peak force extraction starts to become unreliable at roughly 16 points per curve. Due to the flat frequency response of our scanner we anticipate that the system could be operated at PFT rates up to 85 kHz once faster controllers become available.
One eventual limitation for high speed peak force tapping with this implementation is that the required current for the piezo increases drastically as the peak-force frequency increases. The absolute calculated peak currents that our system is capable of are around 2 A. However, during operation in QNM this will typically stay below 100 mA. Since the majority of the power is not dissipated into the piezo, but into the amplifier, we have not observed any problematic heating of the piezo at the currently implemented rates.
Discussion
The performance of the system both in terms of scan range and tracking bandwidth show great promise for the dynamic imaging of biological samples. In particular the ability to track the small, high spatial frequencies in a sample that often exist on the background of large amplitude, low spatial frequency topography enables the use of HS-AFM on a much broader range of applicable samples. Imaging in fluid is shown to enable experiments on delicate biologically relevant systems. This approach requires the recording and processing of system dynamics, as sample weight and size can vary, especially when imaging in liquid. However since the complete identification is done directly on the cantilever deflection, the creation of the filters can be completely automated and in real-time, as well as made completely transparent to the end—user. The sample surface in the current design is small and comparable with other high—speed systems (some mm2), but could potentially be extended by using different piezo housings. The design is limited in range at high speeds, as fast changes can only be effected with the ±750 nm range of the stack piezo, while the tube scanner is primarily used to compensate large background topography, drift and sample tilt. As the higher spatial frequencies of most samples are small in magnitude we have not observed problems with tracking due to the limited range. One significant drawback of using two different actuators is that conventional hysteresis compensation techniques are not equipped to deal with two very distinct hysteresis behaviours, such as a stack piezo actuator in comparison to a tube piezo. This is not as problematic for biological imaging, but can be a problem for nanometrology applications, where better than 1% accuracy is required. Finally, the current implementation does not yet take into account resonant couplings between Z and X—Y.
Future work on the design should include cross compensation between the filters, as Z to X—Y coupling can create artefacts at very large steps. The compensation could be implemented without the need for any additional system characterization, because the X—Y dynamics are already identified for lateral compensation. Existing hysteresis models could be adapted into the control system to correct for aberrations. Alternatively, a Z sensor could be used to map the real position of the sample.
Independent of further development on the design, this system is well equipped to be used in research on larger—scale microbiological systems, such as membrane dynamics, both supported and in vivo.
Methods
AFM setup
The microscope used was a modified Bruker Multimode 8 with a home built small cantilever compatible head. The scanner that was modified was a Bruker J tube scanner. Signals were accessed with a Bruker Signal Access Module III. Necessary signal level adjustment between filters and microscope were done with homebuilt wideband scaling amplifiers. Piezo tube signals were amplified with an externalized Nanoscope 5 high—voltage amplifier, stack signals with a wideband piezo amplifier (Techproject, Austria).
Filter design
Transfer functions were recorded with a digital lock-in amplifier (Anfatec E-204) directly by AFM in contact mode using a 10 nm excitation signal. The data was then fitted to the models and discretized in Matlab. The filter coefficients were loaded into a series of generic filters implemented as second order sections in Labview FPGA. The FPGA boards were a PXI-7851R for lateral suppression filters and a 7954R with a 5781 baseband transceiver for z filtering (all National Instruments). Stack and tube movement were digitally scaled to the same physical displacement per input signal.
Lipid disruption imaging
Small unilamelar vesicle mixtures of 1,2-dilauroyl-sn-glycero-3-phosphocholine (DLPC) and 1,2-dipalmitoyl-sn-glycero-3-phosphocholine (DPPC) were prepared via sonication (both Avanti Polar Lipids Incorporated). Lipid powders were mixed before vesicle formation at a nominal molar ratio of 1:2, DLPC:DPPC. Vesicle solutions (1 mg/ml) were formed by transferring an appropriate mass of lipid into glass vials and dissolved with chloroform. The chloroform was evaporated off with dry Nitrogen gas, leaving a thin film on the glass vial. The film was hydrated with Milli-Q water (Milipore, Billerica, MA, USA), generating large multi-laminar vesicles, (LMVs). The LMVs were then sonicated with a probe sonicator (BioLogics Inc) to generate small unilaminar vesicles (SUVs). The SUVs were centrifuged to remove metal particles left from the probe sonicator. 35 μl of the lipid preparation was warmed to 37 °C and deposited onto freshly cleaved mica surfaces, forming bilayers via vesicle fusion. Surfaces were allowed to incubate for at least a half hour in a humid environment at room temperature.
Imaging was done in liquid with a Bruker FastScan-C cantilever in amplitude modulation. The imaging data has been flattened and corrected for drift in post processing.
Additional Information
How to cite this article: Nievergelt, A. P. et al. Studying biological membranes with extended range high-speed atomic force microscopy. Sci. Rep. 5, 11987; doi: 10.1038/srep11987 (2015).
Supplementary Material
Acknowledgments
The authors thank Prof. Matteo Dal Peraro and Thomas Lemmin at EPFL for molecular dynamics simulation for the depiction of the DLPC-DPPC interface, the Atelier de l’Institut de microtechnique (ATPR) at EPFL for their on-going support and fabrication of research equipment as well as Chanmin Su at Bruker Nano for providing the 31.5 kHz peak force tapping software. This work was funded by the European Union’s Seventh Framework Programme FP7/2007-2011 under the grant agreement 286146, Eurostars E!8213-Tripple-S, the European Union’s Seventh Framework Programme FP7/2007-2013/ERC grant agreement 307338 as well as the Swiss National Science Foundation through grants 205321_134786 and 205320_152675.
Footnotes
Author Contributions A.P.N. and B.W.E. contributed equally to this work. A.P.N. designed and built electronics and control systems. B.W.E. implemented filtering and corresponding software and operated the microscope. N.H. contributed to control systems. J.D.A. designed and built secondary actuators and housing. A.P.N. and G.E.F. wrote the paper.
References
- Ando T. et al. A high-speed atomic force microscope for studying biological macromolecules. Proceedings of the National Academy of Sciences of the United States of America 98, 12468–12472 (2001). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Katan A. J. & Dekker C. High-speed AFM reveals the dynamics of single biomolecules at the nanometer scale. Cell 147, 979–982 (2011). [DOI] [PubMed] [Google Scholar]
- Fantner G. E. et al. Components for high speed atomic force microscopy. Ultramicroscopy 106, 881–887 (2006). [DOI] [PubMed] [Google Scholar]
- Ando T. High-speed atomic force microscopy coming of age. Nanotechnology 23, 062001 (2012). [DOI] [PubMed] [Google Scholar]
- Kodera N., Sakashita M. & Ando T. Dynamic proportional-integral-differential controller for high-speed atomic force microscopy. Review of Scientific Instruments 77, 083704 (2006). [Google Scholar]
- Kodera N., Yamamoto D., Ishikawa R. & Ando T. Video imaging of walking myosin V by high-speed atomic force microscopy. Nature 468, 72–76 (2010). [DOI] [PubMed] [Google Scholar]
- Colom A., Casuso I., Rico F. & Scheuring S. A hybrid high-speed atomic force-optical microscope for visualizing single membrane proteins on eukaryotic cells. Nature communications 4, 2155, 10.1038/ncomms3155 (2013). [DOI] [PubMed] [Google Scholar]
- Suzuki Y. et al. High-speed atomic force microscopy combined with inverted optical microscopy for studying cellular events. Scientific reports 3, 2131, 10.1038/srep02131 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Preiner J. et al. IgGs are made for walking on bacterial and viral surfaces. Nature communications 5, 4394, 10.1038/ncomms5394 (2014). [DOI] [PubMed] [Google Scholar]
- Iyer S., Gaikwad R. M., Subba-Rao V., Woodworth C. D. & Sokolov I. Atomic force microscopy detects differences in the surface brush of normal and cancerous cells. Nature nanotechnology 4, 389–393 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fantner G. E., Barbero R. J., Gray D. S. & Belcher A. M. Kinetics of antimicrobial peptide activity measured on individual bacterial cells using high-speed atomic force microscopy. Nature nanotechnology 5, 280–285 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lekka M. Atomic force microscopy: A tip for diagnosing cancer. Nature nanotechnology 7, 691–692 (2012). [DOI] [PubMed] [Google Scholar]
- Coffey D. C. & Ginger D. S. Time-resolved electrostatic force microscopy of polymer solar cells. Nature materials 5, 735–740 (2006). [DOI] [PubMed] [Google Scholar]
- Herruzo E. T., Perrino A. P. & Garcia R. Fast nanomechanical spectroscopy of soft matter. Nature communications 5, 3126, 10.1038/ncomms4126 (2014). [DOI] [PubMed] [Google Scholar]
- Zhao B. et al. Large scan area high-speed atomic force microscopy using a resonant scanner. The Review of scientific instruments 80, 093707 (2009). [DOI] [PubMed] [Google Scholar]
- Braunsmann C. & Schäffer T. E. High-speed atomic force microscopy for large scan sizes using small cantilevers. Nanotechnology 21, 225705 (2010). [DOI] [PubMed] [Google Scholar]
- Shibata M., Uchihashi T., Ando T. & Yasuda R. Long-tip high-speed atomic force microscopy for nanometer-scale imaging in live cells Scientific Reports 5, 8724 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schitter G. & Stemmer a. Identification and Open-Loop Tracking Control of a Piezoelectric Tube Scanner for High-Speed Scanning-Probe Microscopy. IEEE Transactions on Control Systems Technology 12, 449–454 (2004). [Google Scholar]
- Schitter G., Allgöwer F. & Stemmer a. A new control strategy for high-speed atomic force microscopy. Nanotechnology 15, 108–114 (2004). [Google Scholar]
- Butterworth J. A., Pao L. Y. & Abramovitch D. Y. A comparison of control architectures for atomic force microscopes. Asian Journal of Control 11, 175–181 (2009). [Google Scholar]
- Kindt J. H., Fantner G. E., Cutroni J. a. & Hansma P. K. Rigid design of fast scanning probe microscopes using finite element analysis. Ultramicroscopy 100, 259–265 (2004). [DOI] [PubMed] [Google Scholar]
- Watanabe H. et al. Wide-area scanner for high-speed atomic force microscopy. The Review of scientific instruments 84, 053702 (2013). [DOI] [PubMed] [Google Scholar]
- Yong Y. K., Bhikkaji B. & Reza Moheimani S. O. R. Design, Modeling, and FPAA-Based Control of a High-Speed Atomic Force Microscope Nanopositioner. IEEE/ASME Transactions on Mechatronics 18, 1060–1071 (2013). [Google Scholar]
- Jeong Y., Jayanth G. R. & Menq C.-H. Control of tip-to-sample distance in atomic force microscopy: a dual-actuator tip-motion control scheme. The Review of scientific instruments 78, 093706 (2007). [DOI] [PubMed] [Google Scholar]
- Schitter G., Rijkee W. F. & Phan N. Dual actuation for high-bandwidth nanopositioning. Paper presented at 47th IEEE Conference on Decision and Control, Cancun, Mexico, Decision and Control, 2008. CDC 2008. 47th: IEEE; 10.1109/CDC.2008.4738876 (2008). [DOI]
- Kuiper S. & Schitter G. Model-based feedback controller design for dual actuated atomic force microscopy. Mechatronics 22, 327–337 (2012). [Google Scholar]
- Viani M. B. et al. Fast imaging and fast force spectroscopy of single biopolymers with a new atomic force microscope designed for small cantilevers. Review of Scientific Instruments 70, 4300 (1999). [Google Scholar]
- Schitter G., Fantner G., Kindt J., Thurner P. & Hansma P. On Recent Developments for High-Speed Atomic Force Microscopy. Paper presented at International Conference on Advanced Intelligent Mechatronics, Monterey, California, USA. Proceedings of the 2005 IEEE/ASME International Conference on Advanced Intelligent Mechatronics: IEEE; 10.1109/AIM.2005.1501000 (2005). [DOI]
- Adams J. D. et al. High-speed imaging upgrade for a standard sample scanning atomic force microscope using small cantilevers. The Review of scientific instruments 85, 093702 (2014). [DOI] [PubMed] [Google Scholar]
- Nievergelt A. P., Adams J. D., Odermatt P. D. & Fantner G. E. High-frequency multimodal atomic force microscopy. Beilstein Journal of Nanotechnology 5, 2459–2467, 10.3762/bjnano.5.255 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Burns D. J., Youcef-Toumi K. & Fantner G. E. Indirect identification and compensation of lateral scanner resonances in atomic force microscopes. Nanotechnology 22, 315701 (2011). [DOI] [PubMed] [Google Scholar]
- Kodera N., Yamashita H. & Ando T. Active damping of the scanner for high-speed atomic force microscopy. Review of Scientific Instruments 76, 053708 (2005). [Google Scholar]
- Aphale S. S., Devasia S. & Reza Moheimani S. O. High-bandwidth control of a piezoelectric nanopositioning stage in the presence of plant uncertainties. Nanotechnology 19, 125503 (2008). [DOI] [PubMed] [Google Scholar]
- Ri D., Dqg Q., Khqj P. M. & Hgx V. Model predictive control of a dual-stage actuator system for fast setpoint tracking. Paper presented at 2013 IEEE 8th Conference on Industrial Electronics and Applications (ICIEA), Melbourne, VIC. Proceedings of the 8th Conference on Industrial Electronics and Applications: IEEE, 10.1109/ICIEA.2013.6566625 (2013). [DOI]
- Ryba L., Voda A. & Besan G. An LQG / LTR approach towards piezoactuator vibration reduction with observer-based hysteresis compensation. Paper presented at 19th IFAC World Congress, Le Cap, South Africa. Preprints of the 19th World Congress of The International Federation of Automatic Control: IFAC. (2014. August).
- Sulchek T. et al. Dual integrated actuators for extended range high speed atomic force microscopy. Applied Physics Letters 75, 1637 (1999). [Google Scholar]
- Fleming a. J., Kenton B. J. & Leang K. K. Ultra-fast dual-stage vertical positioning for high performance SPMs. Paper presented at the 2010 American Control Conference, Baltimore, MD, USA. Proceedings of the 2010 American Control Conference : AACC; 10.1109/ACC.2010.5530950 (2010). [DOI]
- Sulchek T. et al. High-speed tapping mode imaging with active Q control for atomic force microscopy. Applied Physics Letters 76, 1473 (2000). [Google Scholar]
- Schitter G., Menold P., Knapp H. F., Allgower F. & Stemmer A. High performance feedback for fast scanning atomic force microscopes. Review of Scientific Instruments 72, 3320 (2001). [Google Scholar]
- Oberdörster G., Oberdörster E. & Oberdörster J. Nanotoxicology: An Emerging Discipline Evolving from Studies of Ultrafine Particles. Environmental Health Perspectives 113, 823–839, 10.1289/ehp.7339 (2005). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tetley T. D. Health effects of nanomaterials. Biochemical Society transactions 35, 527–531 (2007). [DOI] [PubMed] [Google Scholar]
- Nemmar A. Passage of Inhaled Particles Into the Blood Circulation in Humans. Circulation 105, 411–414 (2002). [DOI] [PubMed] [Google Scholar]
- Peer D. et al. Nanocarriers as an emerging platform for cancer therapy. Nature nanotechnology 2, 751–760 (2007). [DOI] [PubMed] [Google Scholar]
- Davis M. E., Chen Z. G. & Shin D. M. Nanoparticle therapeutics: an emerging treatment modality for cancer. Nature reviews. Drug discovery 7, 771–782 (2008). [DOI] [PubMed] [Google Scholar]
- Kukowska-Latallo J. F. et al. Nanoparticle targeting of anticancer drug improves therapeutic response in animal model of human epithelial cancer. Cancer research 65, 5317–5324 (2005). [DOI] [PubMed] [Google Scholar]
- Duncan R. Polymer conjugates as anticancer nanomedicines. Nature reviews. Cancer 6, 688–701 (2006). [DOI] [PubMed] [Google Scholar]
- Hong S. et al. Interaction of polycationic polymers with supported lipid bilayers and cells: nanoscale hole formation and enhanced membrane permeability. Bioconjugate chemistry 17, 728–734 (2006). [DOI] [PubMed] [Google Scholar]
- Erickson B. et al. Interactions of Poly (amidoamine) Dendrimers with Survanta Lung Surfactant : The Importance of Lipid Domains 149, 11003–11008 (2008). [DOI] [PMC free article] [PubMed]
- Mecke A., Lee D.-k., Ramamoorthy A., Orr B. G. & Holl M. M. B. Synthetic and natural polycationic polymer nanoparticles interact selectively with fluid-phase domains of DMPC lipid bilayers. Langmuir : the ACS journal of surfaces and colloids 21, 8588–8590 (2005). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mecke A. et al. Lipid bilayer disruption by polycationic polymers: the roles of size and chemical functional group. Langmuir : the ACS journal of surfaces and colloids 21, 10348–10354 (2005). [DOI] [PubMed] [Google Scholar]
- Giocondi M.-C. et al. Surface topography of membrane domains. Biochimica et biophysica acta 1798, 703–718 (2010). [DOI] [PubMed] [Google Scholar]
- Picas L., Milhiet P.-E. & Hernández-Borrell J. Atomic force microscopy: a versatile tool to probe the physical and chemical properties of supported membranes at the nanoscale. Chemistry and physics of lipids 165, 845–860 (2012). [DOI] [PubMed] [Google Scholar]
- Rosa-Zeiser a., Weilandt E., Hild S. & Marti O. The simultaneous measurement of elastic, electrostatic and adhesive properties by scanning force microscopy: pulsed-force mode operation. Measurement Science and Technology 8, 1333–1338 (1997). [Google Scholar]
- Wu H. W., Kuhn T. & Moy V. T. Dynamic atomic force microscopy methods. Scanning 20, 389–397 (2006). [DOI] [PubMed] [Google Scholar]
- Wu H. W., Kuhn T. & Moy V. T. Mechanical Properties of L929 Cells Measured by Atomic Force Microscopy : Effects of Anticytoskeletal Drugs and Membrane Crosslinking. Scanning 20, 389–397 (2006). [DOI] [PubMed] [Google Scholar]
- Adamcik J., Berquand A. & Mezzenga R. Single-step direct measurement of amyloid fibrils stiffness by peak force quantitative nanomechanical atomic force microscopy. Applied Physics Letters 98, 193701 (2011). [Google Scholar]
- Dufrêne Y. F., Martnez-Martn D., Medalsy I., Alsteens D. & Müller D. J. Multiparametric imaging of biological systems by force-distance curve-based AFM. Nature methods 10, 847–854 (2013). [DOI] [PubMed] [Google Scholar]
- Alsteens D., Trabelsi H., Soumillion P. & Dufrêne Y. F. Multiparametric atomic force microscopy imaging of single bacteriophages extruding from living bacteria. Nature communications 4, 2926, 10.1038/ncomms3926 (2013). [DOI] [PubMed] [Google Scholar]
- de Pablo P. J., Colchero J., Gomez-Herrero J. & Baro a. M. Jumping mode scanning force microscopy. Applied Physics Letters 73, 3300 (1998). [Google Scholar]
- Pletikapic G., Berquand A., Radic T. M. & Svetlicic V. Quantitative Nanomechanical Mapping of Marine Diatom in Seawater Using Peak Force Tapping Atomic Force Microscopy1. Journal of Phycology 48, 174–185 (2012). [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.