Abstract
Aims
To compare overnight closed‐loop and sensor‐augmented pump therapy in patients with type 1 diabetes by combining data collected during free‐living unsupervised randomized crossover home studies.
Methods
A total of 40 participants with type 1 diabetes, of whom 24 were adults [mean ± standard deviation (s.d.) age 43 ± 12 years and glycated haemoglobin (HbA1c) 8.0 ± 0.9%] and 16 were adolescents (mean ± s.d. age 15.6 ± 3.6 years and HbA1c 8.1 ± 0.8%), underwent two periods of sensor‐augmented pump therapy in the home setting, in combination with or without an overnight closed‐loop insulin delivery system that uses a model predictive control algorithm to direct insulin delivery. The order of the two interventions was random; each period lasted 4 weeks in adults and 3 weeks in adolescents. The primary outcome was time during which sensor glucose readings were in the target range of 3.9–8.0 mmol/l.
Results
The proportion of time when sensor glucose was in the target range (3.9–8.0 mmol/l) overnight (between 24:00 and 08:00 hours) was 18.5% greater during closed‐loop insulin delivery than during sensor‐augmented therapy (p < 0.001). Closed‐loop therapy significantly reduced mean overnight glucose levels by 0.9 mmol/l (p < 0.001), with no difference in glycaemic variability, as measured by the standard deviation of sensor glucose. Time spent above the target range was reduced (p = 0.001), as was time spent in hypoglycaemia (<3.9 mmol/l; p = 0.014) during closed‐loop therapy. Lower mean overnight glucose levels during closed‐loop therapy were brought about by increased overnight insulin delivery (p < 0.001) without changes to the total daily delivery (p = 0.84).
Conclusion
Overnight closed‐loop insulin therapy at home in adults and adolescents with type 1 diabetes is feasible, showing improvements in glucose control and reducing the risk of nocturnal hypoglycaemia.
Keywords: clinical trial, glycaemic control, closed‐loop insulin delivery, type 1 diabetes
Introduction
Intensive insulin therapy reduces the risks of macro‐ and microvascular complications in type 1 diabetes 1, 2 but is limited by hypoglycaemia, a key barrier to achieving tight glycaemic control 3. Nocturnal hypoglycaemia remains a significant concern because of its morbidity and link with sudden unexpected death in type 1 diabetes 4, 5. Achieving persistent euglycaemia safely overnight may help reduce the risk and burden to patients, carers and their families.
The closed‐loop system delivers insulin in a continually glucose‐responsive manner by combining subcutaneous continuous glucose monitoring and subcutaneous insulin pump delivery 6. This novel approach differs from conventional insulin pump therapy in its use of a control algorithm, which automatically directs insulin delivery based on real‐time sensor glucose levels. Randomized controlled trials in research facility settings in adults and children with type 1 diabetes have reported improvements in overnight glucose control whilst reducing the risk of nocturnal hypoglycaemia 7, 8, 9 and transitional studies, characterized by close monitoring in diabetes camp and hotel settings, have paved the way for closed‐loop system use in out‐of‐hospital settings 10, 11.
Unsupervised home studies constitute the ultimate ‘testbed’, as investigations under free‐living conditions provide unequivocal assessment of closed‐loop performance and its usability in the target environment. To increase our understanding of the benefits of unsupervised overnight closed‐loop therapy across a wide age range, we pooled data from two recent home studies in adolescents and adults with type 1 diabetes 12, 13. We were particularly interested to ascertain if closed‐loop therapy reduced time spent in hypoglycaemia, something which the two individual studies failed to detect because of their limited sample size and low background rates of nocturnal hypoglycaemia.
Participants and Methods
Participants and Experimental Protocol
Data were pooled from adult and adolescence overnight closed‐loop studies performed in unsupervised settings in the patient's home 12, 13. Between 1 July 2012 and 23 December 2013, adults aged ≥18 years were enrolled from diabetes clinics at Addenbrooke's Hospital, Cambridge, Sheffield Teaching Hospitals, Sheffield and Kings College Hospital, London, and adolescents aged 12–18 years were enrolled from Paediatric Diabetes Clinics at Addenbrooke's Hospital, Cambridge and University College Hospital, London. Inclusion criteria were: type 1 diabetes (World Health Organization criteria); >1 year from diagnosis or confirmed C‐peptide‐negative; insulin pump therapy for ≥3 months; four or more capillary glucose measurements per day; and glycated haemoglobin (HbA1c) level ≤10% (86 mmol/mol). Exclusion criteria included: established nephropathy; neuropathy or proliferative retinopathy; total daily insulin dose ≥2.0 U/kg; regular use of continuous glucose monitoring <1 month before enrolment; severe visual or hearing impairment; and pregnancy or breast feeding.
After a 2–4‐week run‐in phase, participants applied insulin pump therapy with real‐time continuous glucose monitoring at home on two periods with or without overnight closed‐loop therapy. Each period lasted 4 (adults) or 3 (adolescents) weeks. Identical insulin pump and real‐time continuous glucose monitoring devices were used during the two study periods, which were separated by a 3–4‐week washout period, during which participants used their own pump and discontinued continuous glucose monitoring.
Signed informed consent was obtained from all participants aged ≥16 years, and from parents or guardians of participants aged <16 years (assent was obtained from minors). Both study protocols were approved by the East of England Central Cambridge Ethics Committee. The trials were registered under Clinical Trials numbers NCT01440140 and NCT01221467.
Randomization and Blinding
The order of the two study periods was random and was determined after the run‐in phase using computer‐generated permuted block randomization. During the run‐in phase, the continuous glucose monitor receiver was modified to mask the recorded sensor glucose concentrations. Downloaded sensor glucose readings were used to optimize the study insulin pump therapy before randomization. Participants had access to sensor glucose readings after the end of the run‐in phase.
Study Procedures
Blood samples for plasma glucose and C‐peptide measurements were taken after enrolment. Participants were trained on the features of the study insulin pump (Dana R Diabecare; Sooil, Seoul, South Korea) and continuous glucose monitoring (FreeStyle Navigator; Abbott Diabetes Care, Alameda, CA, USA). After the run‐in period, compliance was demonstrated by the use the continuous glucose monitor for at least 14 (adults) and 8 days (adolescents). Participants used the rapid‐acting insulin analogue normally used in their usual clinical care and the built‐in bolus wizard of the insulin pump during both interventions to calculate insulin boluses at mealtimes and when administering correction boluses. Usual basal insulin delivery settings were also applied on the study pump.
The continuous glucose monitoring device was calibrated according to the manufacturer's instructions. The sensor glucose threshold alarm for hypoglycaemia was initially set at 3.5 mmol/l and could be modified by participants. During the closed‐loop period, overnight insulin delivery was automatically directed by a closed‐loop model predictive control algorithm 14.
The adult participants spent the first closed‐loop night in the local clinical research facility and received training in the use of the closed‐loop system. In the adolescence cohort, closed‐loop training was provided to the participants and parents at home during the first closed‐loop night. Training lasted for 60–90 min and covered initiation and discontinuation of the closed‐loop system and problem troubleshooting. Participants were instructed to start the system at home after their evening meal, and to discontinue it before breakfast the next morning. Participants were trained to check capillary glucose versus sensor glucose values (calibration check) before evening meals; if sensor glucose was greater than capillary glucose by >3 mmol/l, the continuous glucose monitor was recalibrated and the calibration check was repeated before starting the closed‐loop system. These instructions reduced the risk of sensor error and the calibration check approach, expected to occur infrequently once every 3–4 weeks, was effective when assessed by computer modelling 15.
No further supervision took place over the following nights and participants used the system fully unsupervised. Participants were not restricted in dietary intake or daily activities. A 24‐h telephone support service assisted participants in clinical or technical issues that arose during the study. All participants were given troubleshooting literature and user manuals for all study devices. Standard local hypoglycaemia and hyperglycaemia treatment guidelines were followed. The data from the study insulin pumps and continuous glucose monitoring devices were downloaded by the research team at the end of study periods. In addition, during closed‐loop therapy, encrypted data from the closed‐loop system was emailed by the participants to the study team once a week.
Closed‐Loop System
The Florence automated closed‐loop system comprised a model predictive control algorithm residing on a hand‐held computer linked by cable to the continuous glucose monitoring receiver 14. Every 12 min, the treat‐to‐target algorithm calculated a new insulin infusion rate, which was automatically set on the study pump via wireless communication.
At setup, the research team entered participants' weight and total daily insulin dose on the first night during which the closed‐loop system was used. Data for carbohydrate intake, as entered by participants into the insulin pump built‐in bolus wizard, were automatically downloaded to the hand‐held computer during the use of the closed‐loop system. Insulin delivery history, including manually instructed insulin boluses, was also automatically downloaded.
Laboratory Assays
Baseline random C‐peptide levels were measured using a chemiluminescence immunoassay (Liaison XL; DiaSorin Deutschland GmbH, Dietzenbach, Germany; interassay coefficients of variation 5.6% at 563 pmol/l, 4.5% at 2529 pmol/l and 5.8% at 5449 pmol/l). HbA1c was measured centrally with ion exchange high‐performance liquid chromatography (G8 HPLC Analyzer; Tosoh Bioscience, South San Francisco, CA, USA; interassay coefficients of variation 1.3% at 31.2 mmol/mol and 0.8% at 80.5 mmol/mol).
Outcomes
The primary efficacy outcome was the time spent in the target glucose range (3.9–8.0 mmol/l) between 24:00 and 08:00 hours, as recorded by continuous glucose monitoring. Secondary outcomes included mean glucose concentration, time spent at glucose levels <3.9 mmol/l (hypoglycaemia) and >8.0 mmol/l (hyperglycaemia), and insulin delivery. Overnight glucose variability was assessed by the standard deviation and the coefficient of variation of continuous glucose monitoring levels. We assessed hypoglycaemia burden by calculating the glucose sensor area under the curve <3.5 mmol/l. We calculated secondary outcomes from 24:00 to 08:00 hours and over 24 h.
Statistics
Analyses were carried out on an intention‐to‐treat basis. Normally distributed data were compared using the paired t‐test and non‐normally distributed data using the Wilcoxon signed‐rank test. Results are presented as mean [standard deviation (s.d.)] or median [interquartile range (IQR)], unless stated otherwise. p values <0.05 were taken to indicate statistical significance. Outcomes were calculated using gstat software (version 2.0, University of Cambridge) and statistical analyses were conducted using spss (version 21).
Results
Baseline Characteristics
A total of 24 adults and 16 adolescents completed both study periods. Baseline characteristics are shown in Table 1.
Table 1.
Baseline characteristics
| Adults (n = 24)* | Adolescents (n = 16)* | |
|---|---|---|
| Age, years | 43 ± 12 | 15.6 ± 2.1 |
| Gender: male/female | 13/11 | 10/6 |
| Body mass index, kg/m2 | 26.0 ± 3.5 | 22.4 ± 3.7 |
| HbA1c | ||
| % | 8.1 ± 0.8 | 8.0 ± 0.9 |
| mmol/mol | 64.9 ± 8.9 | 63.9 ± 9.4 |
| Duration of diabetes, years | 29 ± 11 | 7.2 ± 4.3 |
| Duration on pump, years | 6.3 ± 4.4 | 3.0 ± 2.3 |
| Total daily insulin, U/kg/day | 0.5 ± 0.1 | 0.8 ± 0.2 |
HbA1c, glycated haemoglobin. Data are mean ± standard deviation unless otherwise indicated.
All with C‐peptide levels lower than 33 pmol/l.
Primary and Secondary Outcomes Overnight
The primary endpoint, the proportion of time when overnight sensor glucose levels were in the target range, was significantly increased by a mean of 18.4% [95% confidence interval (CI) 13.5–23.4%; p < 0.001] during closed‐loop therapy (Table 2). Closed‐loop therapy significantly reduced mean overnight sensor glucose by 0.9 mmol/l (95% CI 0.4–1.3 mmol/l; p < 0.001), as well as the proportion of time when sensor glucose values were in hyperglycaemia (>8.0 mmol/l; mean 15.9%, 95% CI 10.7–21.0%; p < 0.001) and hypoglycaemia (<3.9 mmol/l; median 0.9%, 95% CI 0.2–2.2; p = 0.014; Table 2, Figure 1). There was no significant difference in the burden of hypoglycaemia as measured by the area under the curve when sensor glucose levels were <3.5 mmol/l (p = 0.15).
Table 2.
Intention‐to‐treat comparison of glucose outcomes from 24:00 to 08:00 hours during overnight closed‐loop therapy and the control period in the home setting over 3–4 weeks
| Closed‐loop period (n = 40) | Control period (n = 40) | p | |
|---|---|---|---|
| Mean ± s.d. glucose, mmol/l | 7.9 ± 0.9 | 8.7 ± 1.4 | <0.001 |
| Mean ± s.d. within‐night standard deviation of glucose, mmol/l | 2.0 ± 0.3 | 1.9 ± 0.3 | 0.47 |
| Mean ± s.d. within‐night coefficient of variation of glucose, % | 25.2 ± 3.5 | 22.8 ± 5.0 | 0.023 |
| Mean ± s.d. between‐nights coefficient of variation of glucose, % | 24.8 ± 6.9 | 29.1 ± 6.9 | 0.002 |
| Mean ± s.d. proportion of time spent at glucose level, % | |||
| 3.9–8.0 mmol/l* | 59.2 ± 11.5 | 40.7 ± 13.4 | <0.001 |
| 3.9–10.0 mmol/l | 77.4 ± 8.6 | 61.8 ± 13.3 | <0.001 |
| >8.0 mmol/l | 37.9 ± 12.4 | 53.8 ± 17.0 | 0.001 |
| Median (IQR) proportion of time spent at glucose level, % | |||
| >16.7 mmol/l | 0.66 (0.0, 2.3) | 1.3 (0.0, 2.8) | 0.15 |
| <3.9 mmol/l | 1.9 (0.7, 3.5) | 2.9 (1.0, 6.4) | 0.014 |
| <3.5 mmol/l | 0.7 (0.3, 2.0) | 1.1 (0.21, 3.9) | 0.098 |
| <2.8 mmol/l | 0.2 (0.0, 0.6) | 0.2 (0.0, 1.3) | 0.29 |
| Median (IQR) AUCday <3.5 mmol/l, mmol/l × min | 4.7 (1.2, 16.9) | 6.4 (0.4, 33.8) | 0.15 |
| Mean ± s.d. glucose at 24:00 hours, mmol/l | 9.1 ± 1.3 | 9.0 ± 1.8 | 0.39 |
| Mean ± s.d. glucose at 08:00 hours, mmol/l | 7.7 ± 0.92 | 9.0 ± 1.4 | <0.001 |
AUCday, Glucose area under curve below 3.5 mmol/l per day; IQR, interquartile range; s.d., standard deviation.
Primary endpoint.
Figure 1.

Individual values of mean glucose levels and proportion of time spent at low glucose levels (<3.9 mmol/l) in the whole cohort. Line plots show individual values of mean overnight glucose during the intention‐to‐treat analysis. The diameter of each circle is proportional to the percentage of time that each participant spent with a low glucose value <3.9 mmol/l.
Glucose variability overnight measured as standard deviation was similar during the two interventions. The mean ± s.d. coefficient of variation of overnight glucose within each night was higher during closed‐loop than during sensor‐augmented pump therapy (25.2 ± 3.5% vs. 22.8 ± 5.0%; p = 0.023; Table 2). Conversely, the between‐nights coefficient of variation of overnight glucose was significantly reduced by closed‐loop therapy (24.8 ± 6.9% vs. 29.1 ± 6.9%; p = 0.002).
Exploratory per‐protocol analysis, considering nights when the closed‐loop system was turned on for at least 6 h (closed‐loop period) or sensor data were available for at least 6 h (control period), showed that closed‐loop therapy increased the proportion of time overnight spent in target by 20.7% (95% CI 15.7–25.6%; p < 0.001; Table 3). This was achieved by reducing mean glucose levels by 1.0 mmol/l (95% CI 0.5–1.4 mmol/l; p < 0.001) as well as reducing time spent below target overnight by a median of 3.2% (95% CI 0.8–5.0%; p = 0.005).
Table 3.
Comparison of glucose outcomes from 24:00 to 08:00 hours, with at least 6 h of closed‐loop operation (intervention group) and 6 h of sensor data (control group)
| Closed‐loop group (n = 40) | Control group (n = 40) | p | |
|---|---|---|---|
| Number of nights | 786 | 881 | |
| Mean ± s.d. glucose, mmol/l | 7.8 ± 0.96 | 8.7 ± 1.4 | <0.001 |
| Mean ± s.d. proportion of time spent at glucose level 3.9–8.0 mmol/l, % | 61.3 ± 12.5 | 40.6 ± 13.4 | <0.001 |
| Median (IQR) proportion of time spent at glucose level <3.9 mmol/l, % | 1.1 (0.58, 2.9) | 2.9 (0.8, 5.8) | 0.005 |
| Median (IQR) AUCday <3.5 mmol/l, mmol/l × min | 2.5 (0.34, 10.1) | 6.2 (0.21, 33.8) | 0.09 |
| Mean ± s.d. lucose at 24:00 hours, mmol/l | 9.0 ± 1.3 | 9.0 ± 1.9 | 0.69 |
| Mean ± s.d. lucose at 08:00 hours, mmol/l | 7.7 ± 1.1 | 9.0 ± 1.3 | <0.001 |
AUCday, Glucose area under curve below 3.5 mmol/l per day; IQR, interquartile range; s.d., standard deviation.
Secondary Outcomes From Midnight to Midnight
Endpoints calculated over 24 h showed that the overnight closed‐loop system significantly increased the proportion of time spent within a wider target range of 3.9–10.0 mmol/l by a mean of 4.7% (95% CI 2.0–7.5%; p = 0.001; Table 4). Overnight closed‐loop therapy reduced the overall mean sensor glucose level by a mean of 0.5 mmol/l (95% CI 0.1–1.0 mmol/l; p = 0.016). This was achieved by significantly reducing both proportion of time spent over 24 h with glucose above the target by a mean of 7.0% (95% CI 3.6–10.3%; p < 0.001) and by reducing the proportion of time spent in hypoglycaemia by a median of 1.6% (95% CI 0.4–2.8%; p = 0.015).
Table 4.
Comparison of 24‐h glucose outcomes during overnight closed‐loop therapy and the control period in the home setting over 3–4 weeks
| Closed‐loop period (n = 40) | Control period (n = 40) | p | |
|---|---|---|---|
| Mean ± s.d. glucose, mmol/l | 8.6 ± 1.2 | 9.1 ± 1.4 | 0.016 |
| Mean ± s.d. proportion of time spent at glucose level, % | |||
| 3.9–10.0 mmol/l | 63.6 ± 9.9 | 58.8 ± 10.5 | 0.001 |
| >10.0 mmol/l | 29.5 ± 10.5 | 36.5 ± 13.3 | <0.001 |
| Median (IQR) proportion of time spent at glucose level <3.9 mmol/l, % | 1.8 (0.91, 3.2) | 2.7 (1.3, 5.7) | 0.015 |
| Median (IQR) AUCday < 3.5 mmol/l, mmol/l × min | 5.2 (2.0, 12.1) | 6.9 (2.4, 29.5) | 0.078 |
AUCday, Glucose area under curve below 3.5 mmol/l per day; IQR, interquartile range; s.d., standard deviation.
Insulin Requirements and Utility of the Closed‐Loop System
The closed‐loop system delivered 17% more insulin overnight (p < 0.001; Table 5). The total daily insulin delivered was similar during the two interventions (p = 0.84). Adult and adolescent participants used closed‐loop therapy of their own volition for 87% (856 nights) of the whole study duration (Table S1). Closed‐loop therapy was used for a median of 9 h at home each night; the median closed‐loop therapy start time was 22:37 hours and the stop time was 07:30 hours. The causes and frequency of unintentional closed‐loop interruptions are listed in Table S1.
Table 5.
Insulin delivery overnight (24:00–08:00 hours) and over 24 h
| Closed‐loop period, n = 40 | Control period, n = 40 | p | |
|---|---|---|---|
| Median (IQR) | Median (IQR) | ||
| Overnight insulin delivery, U | 7.0 (5.4, 9.3) | 6.0 (4.7, 7.4) | <0.001 |
| Total daily insulin delivery, U | 40.3 (32.9, 52.6) | 39.4 (32.8, 55.8) | 0.84 |
| s.d. of overnight insulin delivery, U | 0.6 (0.5, 0.8) | 0.1 (0.07, 0.2) | <0.001 |
IQR, interquartile range; s.d., standard deviation.
Adverse Events
There were two episodes of hyperglycaemia associated with ketosis (blood ketones >1.5 mmol/l) during the control period, and one occasion during the closed‐loop period. Two episodes of severe hypoglycaemia occurred during the closed‐loop period in two adults who each had a history of hypoglycaemia unawareness, and both events happened at a time when the closed‐loop system was not operational. Post hoc analysis showed that closed‐loop operation was interrupted 1 h before these two episodes occurred because of disrupted wireless connectivity with the insulin pump. Before this interruption, insulin delivery had already been suspended by the algorithm because of predicted low glucose concentrations. In 1 participant, the event was probably compounded by increased physical activity during the day. User error, that is, overdelivery of insulin whilst changing infusion set overnight, was deemed to be the likely contributory cause in the second participant. Both participants recovered fully with no clinical sequelae.
Discussion
This pooled analysis adds to the observations in two previous studies that adults and adolescents with type 1 diabetes are able to use closed‐loop therapy safely and effectively overnight at home over an extended period, without the need for remote monitoring or close supervision. The combined evaluation of the two aforementioned studies 12, 13 provides greater power and allows comprehensive assessment of closed‐loop benefits across a wide age range. A clinically significant reduction in overnight glucose was observed accompanied by reduced time spent in a hypoglycaemic state. Such combined effect has not been documented with any other means of intensified conventional insulin delivery in type 1 diabetes. Glucose variability between nights was reduced by closed‐loop therapy in the study, whilst increased variability within each night is explained by similarly elevated sensor glucose levels at midnight during both interventions, accompanied by consistent lower morning glucose at 08:00 hours during closed‐loop therapy.
Glucose concentrations remained lower over 24 h, suggesting there are extended benefits of improved overnight glucose control even when closed‐loop therapy is stopped. Furthermore, fasting morning glucose levels were 14% lower after overnight closed‐loop therapy. As a result, participants were able to administer fewer insulin boluses and corrections during the day. This may explain the similar total daily insulin dosage for both study periods, in spite of the 17% increase in overnight insulin delivery during closed‐loop.
Real‐time continuous glucose monitoring is associated with a mean HbA1c reduction of 0.5% (5.5 mmol/mol) 16, 17, 18 and an unchanged risk of nocturnal hypoglycaemia 19. Adolescents appear to benefit less, owing to a reduction in time when the glucose sensor was used 20. Frequent insulin dosing and infusion rate adjustments to match glycaemia levels and day‐to‐day variations in insulin sensitivity represent significant challenges and burden to subjects, their families and healthcare professionals. A more recent technological approach, the threshold‐suspend insulin pump, interrupts insulin delivery for up to 2 h when at a low glucose threshold. Its application reduces nocturnal frequency and duration of hypoglycaemia 21, 22, but is unable to step up insulin delivery during episodes of elevated glucose levels and does not alter overall glucose control 22.
Our closed‐loop algorithm accounts for variations in insulin requirements, finely modulating insulin delivery based on real‐time sensor glucose levels and maintaining more consistent glucose levels overnight. Other closed‐loop systems have been studied in outpatient and home settings, albeit for a shorter duration or under remote monitoring and supervision 11, 23, 24, 25. Compared with standard pump therapy, subjects on a bihormonal closed‐loop system in an outpatient setting for 5 days had lower mean glucose levels and less frequent hypoglycaemic episodes 26. Overnight closed‐loop therapy in children and adolescents at a diabetes camp over 5–6 days showed significant reductions in times spent at various hypoglycaemia glucose levels, but no significant difference in the intention‐to‐treat analysis was observed for median percentage of time in the glucose range of 3.9–8.3 mmol/l 26. A single night closed‐loop intervention at a diabetes camp showed a reduction in the frequency and duration of hypoglycaemia <3.5 and 3.3 mmol/l, with a reduction of overnight median glucose by 0.8 mmol/l 11. The same group reported significant reductions in time spent with sensor glucose levels <3.9 mmol/l and an increased percentage of time spent in the target range (3.9–7.8 mmol/l) during a 6‐week single‐centre overnight closed‐loop home study 27. In contrast to the present multicentre study, however, a remote monitoring system was used and thus the research team was immediately alerted of technical failures as well as hypo‐ and hyperglycaemia events during overnight closed‐loop therapy.
The strength of the present study is the rigour of the intention‐to‐treat analysis and duration of closed‐loop use. Its distinguishing feature is free‐living unsupervised use in home settings. Other studies performed in out‐of‐hospital settings were either shorter or were performed under remote monitoring, whereby investigators could intervene clinically during episodes of significant hypo‐ or hyperglycaemia, and could manage technical faults during the study on behalf of participants. Participants in the present study were not restricted in their activities or dietary habits, and used closed‐loop therapy at home of their own volition; thus, experimentally, these studies are the first to report on the utility and efficacy of closed‐loop therapy under real‐life day‐to‐day conditions without participants being directly monitored or supervised. The randomized crossover design of the present study allowed the participants to act as their own control. We acknowledge that the findings were limited by usability of the system. Closed‐loop therapy interruptions occurred on average once every five nights. More than 60% of interruptions were attributable to disruptions to the pump wireless connectivity or loss of sensor glucose availability. Participants' positive and negative experiences and quality of life may thus have been affected by various aspects of system utility 28. The experience and utilisation rate of 87% may be further enhanced by improvements in device connectivity.
In conclusion, unsupervised use of overnight closed‐loop therapy in adults and adolescents with type 1 diabetes is feasible. Benefits include improved overnight glucose control with reduction in time spent in a hypoglycaemic state. Longer‐term studies are warranted using devices with more reliable connectivity.
Conflict of Interest
R. H. has received speaker honoraria from Minimed Medtronic, Lifescan, Eli Lilly, BBrau, and Novo Nordisk, serves on an advisory panel for Animas, Minimed Medtronic and Eli Lilly, has received license fees from BBraun and Beckton Dickinson, and has served as a consultant to Beckton Dickinson, BBraun, Sanofi‐Aventis and Profil. S. R. H. has undertaken consultancy work for Novo Nordisk and Eli Lilly, for which his institution has received payment. He has spoken at meetings for which he has received payment from NovoNordisk, Eli Lilly and Beckton Dickinson. Medtronic has provided research support for some of his work. M. L. E. has received speaker honoraria from Eli Lilly, Animas and Abbott Diabetes Care and served on advisory panels for Medtronic, Roche, Sanofi‐Aventis and Cellnovo. P. C. declares speaker honoraria and travel support from Medtronic, Roche and Lifescan, and has undertaken consultancy for Novo Nordisk and Eli Lilly, for which his institution has received payment. He has spoken at meetings for which he has received payment from NovoNordisk, Eli Lilly and Beckton Dickinson. Medtronic has provided research support for some of his work. MEW reports receiving licensing fees from Beckton Dickinson. R. H., D. B. D. and M. E. W. report patent applications. H. T., D. E., L. L., J. M. A., A. L. S., M. S., E. W., J. M. A., A. I., K. D. B. and S. A. A. have no competing financial interests.
R. H. and H. T. had full access to all of the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis. R. H. coordinated the study. R. H., M. L. E., S. R. H., S. A. A., D. B. D., H. T., D. E., E. W., M. E. W. and K. D. B. co‐designed the studies. H. T., D. E., J. A., A. L.‐S., M. S., E. W., A. I. and P. C. were responsible for screening and enrolment of participants. H. T., D. E., L. L., J. A., A. L.‐S., M. S., E. W., A. I. and P. C. provided patient care and/or took samples and arranged informed consent from the participants. M. W. generated the random allocation sequence using a computer‐generated random code. H. T., L. L., and D. E. carried out or supported data analysis, including the statistical analyses. R. H. designed and implemented the glucose controller. H. T. drafted the manuscript. H. T. and R. H. edited the manuscript. All authors reviewed the manuscript and approved the final version.
Supporting information
Table S1. Utility and failure analysis of closed‐loop operation.
Acknowledgements
Funding for this study was received from the Juvenile Diabetes Research Foundation (#22‐2009‐802) and Diabetes UK (BDA07/0003549), with additional support for the Artificial Pancreas work by National Institute of Diabetes and Digestive and Kidney Diseases (1R01DK085621), Wellcome Strategic Award (100574/Z/12/Z) and National Institute for Health Research Cambridge Biomedical Research Centre. Abbott Diabetes Care supplied continuous glucose delivery devices and sensors and modified devices to facilitate real‐time connectivity. No funder had any role in the study design, data collection/analysis/interpretation or manuscript preparation.
We are grateful to study volunteers for their participation. Prof. Peter Hindmarsh (University College, London) helped in identifying potential recruits. We acknowledge support from the staff at the Addenbrooke's Wellcome Trust Clinical Research Facility, Sheffield's Centre for Biomedical Research Clinical Research Facility and the NIHR‐Welcome Trust King's Clinical Research Facility. John Lum (Jaeb Centre) and Jasdip Mangat (Cambridge University Hospitals NHS Foundation Trust) supported the development and validation of the closed‐loop system. Marianna Nodale (University of Cambridge) provided data and device management support. Josephine Hayes (Institute of Metabolic Science, University of Cambridge) provided administrative support. Andrew Pernet, Bula Wilson and Louisa Green provided clinical support at the research facility (Kings College London). Arti Gulati (University of Cambridge) provided data management support. Karen Whitehead (University of Cambridge) provided laboratory support. The Core Biochemical Assay Laboratory, University of Cambridge (Keith Burling) carried out biochemical analyses.
References
- 1. Nathan DM, Cleary PA, Backlund JY et al. Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes. N Engl J Med 2005; 353: 2643–2653. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. The effect of intensive treatment of diabetes on the development and progression of long‐term complications in insulin‐dependent diabetes mellitus. The Diabetes Control and Complications Trial Research Group. N Engl J Med 1993; 329: 977–986. [DOI] [PubMed] [Google Scholar]
- 3. Epidemiology of severe hypoglycemia in the diabetes control and complications trial. The DCCT Research Group. Am J Med 1991; 90: 450–459. [PubMed] [Google Scholar]
- 4. Frier BM, Schernthaner G, Heller SR. Hypoglycemia and cardiovascular risks. Diabetes Care 2011; 34(Suppl. 2): S132–137. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Frier BM. Morbidity of hypoglycemia in type 1 diabetes. Diabetes Res Clin Pract 2004; 65(Suppl. 1): S47–52. [DOI] [PubMed] [Google Scholar]
- 6. Hovorka R. Closed‐loop insulin delivery: from bench to clinical practice. Nat Rev Endocrinol 2011; 7: 385–395. [DOI] [PubMed] [Google Scholar]
- 7. Hovorka R, Kumareswaran K, Harris J et al. Overnight closed loop insulin delivery (artificial pancreas) in adults with type 1 diabetes: crossover randomised controlled studies. BMJ 2011; 342: d1855. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Luijf YM, DeVries JH, Zwinderman K et al. Day and night closed‐loop control in adults with type 1 diabetes: a comparison of two closed‐loop algorithms driving continuous subcutaneous insulin infusion versus patient self‐management. Diabetes Care 2013; 36: 3882–3887. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Breton M, Farret A, Bruttomesso D et al. Fully integrated artificial pancreas in type 1 diabetes: modular closed‐loop glucose control maintains near normoglycemia. Diabetes 2012; 61: 2230–2237. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Kovatchev BP, Renard E, Cobelli C et al. Feasibility of outpatient fully integrated closed‐loop control: first studies of wearable artificial pancreas. Diabetes Care 2013; 36: 1851–1858. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Phillip M, Battelino T, Atlas E et al. Nocturnal glucose control with an artificial pancreas at a diabetes camp. N Engl J Med 2013; 368: 824–833. [DOI] [PubMed] [Google Scholar]
- 12. Thabit H, Lubina‐Solomon A, Stadler M et al. Home use of closed‐loop insulin delivery for overnight glucose control in adults with type 1 diabetes: a 4‐week, multicentre, randomised crossover study. Lancet Diabetes Endocrinol 2014; 2: 701–709. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Hovorka R, Elleri D, Thabit H et al. Overnight closed‐loop insulin delivery in young people with type 1 diabetes: a free‐living, randomized clinical trial. Diabetes Care 2014; 37: 1204–1211. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Elleri D, Allen JM, Biagioni M et al. Evaluation of a portable ambulatory prototype for automated overnight closed‐loop insulin delivery in young people with type 1 diabetes. Pediatr Diabetes 2012; 13: 449–453. [DOI] [PubMed] [Google Scholar]
- 15. Wilinska ME, Chassin LJ, Acerini CL, Allen JM, Dunger DB, Hovorka R. Simulation environment to evaluate closed‐loop insulin delivery systems in type 1 diabetes. J Diabetes Sci Technol 2010; 4: 132–144. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Group JDRFCGMS . Effectiveness of continuous glucose monitoring in a clinical care environment: evidence from the Juvenile Diabetes Research Foundation continuous glucose monitoring (JDRF‐CGM) trial. Diabetes Care 2010; 33: 17–22. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Battelino T, Conget I, Olsen B et al. The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial. Diabetologia 2012; 55: 3155–3162. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Langendam M, Luijf YM, Hooft L, DeVries JH, Mudde AH, Scholten RJPM. Continuous glucose monitoring systems for type 1 diabetes mellitus. Cochrane Database of Systematic Reviews 2012: Issue 1. Art. No.: CD008101. DOI: 10.1002/14651858.CD008101.pub2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group , Fiallo‐Scharer R, Cheng J et al. Factors predictive of severe hypoglycemia in type 1 diabetes: analysis from the Juvenile Diabetes Research Foundation continuous glucose monitoring randomized control trial dataset. Diabetes Care 2011; 34: 586–590. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group , Tamborlane WV, Beck RW et al. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med 2008; 359: 1464–1476. [DOI] [PubMed] [Google Scholar]
- 21. Ly TT, Nicholas JA, Retterath A, Lim EM, Davis EA, Jones TW. Effect of sensor‐augmented insulin pump therapy and automated insulin suspension vs standard insulin pump therapy on hypoglycemia in patients with type 1 diabetes: a randomized clinical trial. JAMA 2013; 310: 1240–1247. [DOI] [PubMed] [Google Scholar]
- 22. Bergenstal RM, Klonoff DC, Garg SK et al. Threshold‐based insulin‐pump interruption for reduction of hypoglycemia. N Engl J Med 2013; 369: 224–232. [DOI] [PubMed] [Google Scholar]
- 23. Kovatchev BP, Renard E, Cobelli C et al. Safety of outpatient closed‐loop control: first randomized crossover trials of a wearable artificial pancreas. Diabetes Care 2014; 37: 1789–1796. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Leelarathna L, Dellweg S, Mader JK et al. Day and night home closed‐loop insulin delivery in adults with type 1 diabetes: three‐center randomized crossover study. Diabetes Care 2014; 37: 1931–1937. [DOI] [PubMed] [Google Scholar]
- 25. Nimri R, Muller I, Atlas E et al. Night glucose control with MD‐Logic artificial pancreas in home setting: a single blind, randomized crossover trial‐interim analysis. Pediatr Diabetes 2014; 15: 91–99. [DOI] [PubMed] [Google Scholar]
- 26. Russell SJ, El‐Khatib FH, Sinha M et al. Outpatient glycemic control with a bionic pancreas in type 1 diabetes. N Engl J Med 2014; 371: 313–325. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Nimri R, Muller I, Atlas E et al. MD‐logic overnight control for 6 weeks of home use in patients with type 1 diabetes: randomized crossover trial. Diabetes Care 2014; 37: 3025–3032. [DOI] [PubMed] [Google Scholar]
- 28. Barnard K, Wysocki T, Allen J et al. Closing the loop overnight at home setting: psychosocial impact for adolescents with type 1 diabetes and their parents. BMJ Open Diabetes Res Care 2014; 2: e000025. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Table S1. Utility and failure analysis of closed‐loop operation.