Myocardial T2 Mapping With Respiratory Navigator and Automatic Nonrigid Motion Correction

Shivraman Giri 1,2, Saurabh Shah 3, Hui Xue 4, Yiu-Cho Chung 5, Michael L Pennell 6, Jens Guehring 4, Sven Zuehlsdorff 3, Subha V Raman 2,7,8, Orlando P Simonetti 1,2,7,8,*
PMCID: PMC4512252  NIHMSID: NIHMS510630  PMID: 22851292

Abstract

Quantitative T2 mapping was recently shown to be superior to T2-weighted imaging in detecting T2 changes across myocardium. Pixel-wise T2 mapping is sensitive to misregistration between the images used to generate the parameter map. In this study, utility of two motion-compensation strategies—(i) navigator gating with prospective slice correction and (ii) nonrigid registration—was investigated for myocardial T2 mapping in short axis and horizontal long axis views. Navigator gating provides respiratory motion compensation, whereas registration corrects for residual cardiac and respiratory motion between images; thus, the two strategies provided complementary functions. When these were combined, respiratory-motion-induced T2 variability, as measured by both standard deviation and interquartile range, was comparable to that in breath-hold T2 maps. In normal subjects, this combined motion-compensation strategy increased the percentage of myocardium with T2 measured to be within normal range from 60.1% to 92.2% in short axis and 62.3% to 92.7% in horizontal long axis. The new motion-compensated T2 mapping technique, which combines navigator gating, prospective slice correction, and nonrigid registration to provide through-plane and in-plane motion correction, enables a method for fully automatic and robust free-breathing T2 mapping.

Keywords: T2 mapping, motion-compensation, nonrigid registration, respiratory navigator


Myocardial T2 is altered in certain pathologies such as acute ischemia (1), myocarditis (2), and heart transplant rejection (3). Previous studies have proposed the use of T2-weighted imaging or T2 quantification (46) to detect these changes. Recently, we described a technique for T2 mapping of the myocardium and demonstrated its ability to overcome some of the limitations of qualitative T2-weighted imaging (7). Techniques for T2 quantification generally require multiple T2-weighted images, each acquired at a different echo or T2-prep (T2P) time; thereafter, T2 values are computed by fitting an exponential decay curve across the multiple images at each pixel (7) or within desired regions of interest (ROIs) (8). This process of fitting a curve through points (pixel values or average ROI intensities) from different images renders T2-quantification sensitive to misregistration of the images caused by breathing motion, inconsistent cardiac rhythm, or gross subject motion. The T2 values of misregistered pixels and ROIs may be inaccurate, degrading the overall precision of myocardial T2 estimation. Thus, it is important to ensure proper alignment of multiple-echo images prior to T2 quantification.

Recently, a fast nonrigid registration algorithm (9) was successfully adapted to registering first-pass perfusion images (10). However, to the best of our knowledge, the use of registration in T2 quantification has not yet been demonstrated.

In this work, we propose an enhancement of our previous T2 mapping scheme: the new sequence includes a respiratory navigator to gate the acquisition based on diaphragm position and to prospectively correct for head-to-foot displacement; thereafter, any residual misregistration is corrected by a nonrigid registration algorithm optimized for myocardial T2 mapping. We hypothesize that this combination of motion-correction strategies will enable free-breathing T2 mapping with results comparable to breath-hold acquisitions. We tested this hypothesis in normal subjects by comparing three parameters: (i) pixel-wise standard deviation (SD) of myocardial T2T2) throughout the myocardium, (ii) interquartile range of T2 (IQR-T2), and (iii) percentage of myocardium with normal T2 (%T2Normal). These parameters were measured using three different acquisition schemes (breath-hold, free-breathing without navigator, and free-breathing with navigator) and two postprocessing modes (with and without nonrigid registration). The effect of the two motion compensation schemes on these parameters was first evaluated individually; thereafter, the proposed technique—free-breathing acquisition with navigator and registration—was compared with breath-hold acquisition. Results were evaluated for two slices in each subject: a mid-ventricular short axis (SAX) in which the heart tends to demonstrate in-plane motion and the horizontal long axis (HLA) view in which breathing can induce through-plane motion of the heart.

MATERIALS AND METHODS

The basic T2 mapping sequence has been previously described (7). Briefly, T2 maps were generated using a monoexponential, log-transformed linear least-squares fit to signal intensities at each pixel of three T2 Prepared Steady-State Free Precession (T2P-SSFP) images (11). The T2P times used were 0 (no prep), 24, and 55 ms, and the ensuing SSFP readout was single-shot acquisition with linear k-space reordering. The nonzero T2P times were achieved by a T2P pulse consisting of a 90° tip-down followed by four 180° composite pulses with Malcolm-Levitt (MLEV) phase cycling scheme and a composite 90° tip-up pulse; all the pulses were nonselective. Such a T2P scheme has been shown to be robust against amplitude of static (polarizing) field and amplitude of (excitation) radiofrequency field inhomogeneities at 1.5 T (12).

Two new features were incorporated into this sequence: respiratory navigator and nonrigid motion correction. Figure 1a shows the new acquisition scheme; note that a navigator echo was acquired prior to each image, and T2P-SSFP images were acquired every 3 RR intervals to allow sufficient T1 recovery. Navigator echo tracking of diaphragm position was used to compensate for respiratory motion using a combined strategy of gating and prospective correction of slice position; for gating, an acceptance window of 8 mm was used, empirically chosen as a good balance between imaging efficiency and motion sensitivity. An empirical correction factor of 0.6 (13) was used to prospectively correct for residual displacement of the slice in the head-to-foot direction. Other imaging parameters are listed in Table 1.

FIG. 1.

FIG. 1

Acquisition scheme and postprocessing work-flow. a: Acquisition scheme for navigator-gated T2 map using T2P-bSSFP. Images were acquired once in every three cardiac cycles to allow for sufficient T1 recovery between acquisitions. Acquired images were accepted only if the diaphragm position was within the acceptance window of 8 mm (“prospective correction of slice position” is not shown in the schematic). The sequence consists of navigator pulses, T2P pulse (4 MLEV weighted), linear ramp catalyzation (for bSSFP) and single-shot bSSFP readout. b: Block diagram of motion correction using nonrigid registration. *“Registration sufficient” indicates that either the predefined similarity measure has been satisfied or the maximum number of iterations has been reached.

Table 1.

Imaging Parameters

Parameter name Value
Number of T2P 3
T2P times 0 (no prep), 24, 55 ms
Acquisition window ~160 ms
Acquisition mode Single shot
Image matrix (human studies) 130 × 160
Field of view (human studies) 203–285 × 250–350 mm2
Image matrix (phantom study) 104 × 128
Field of view (phantom study) 178 × 220 mm2
Slice thickness 8 mm
Parallel acquisition/acceleration/reference lines GRAPPA/2/24
Repetition time 3 RR intervals
Total scan time (without navigator gating) 7 RR intervals
Flip angle 70°
Field strength 1.5 T

A fast variational nonrigid registration algorithm (9) was applied to register the T2P-SSFP images in a pair-wise manner, that is, the 24-ms T2P image was selected as the reference frame, and both 0- and 55-ms images were registered to it. The registration algorithm could be considered as an extension of the classic optical flow method (14). In this framework, a dense deformation field is estimated as the solution to a calculus of variation problem, which is solved by performing a compositional update step corresponding to a transport equation. The regularization is added by low-pass filtering the gradient images which are in turn used as a velocity field to drive the transport equation. To accelerate the convergence and to avoid local minima, a multiscale image pyramid is created. We selected the local cross correlation as the image similarity measure, as its explicit derivative can be efficiently calculated, and it is general enough to cope with intensity fluctuation and imaging noise between two adjacent T2P images. Figure 1b summarizes the process.

The nonrigid registration of three T2P-SSFP images and the subsequent generation of T2 maps were performed in a fully automatic manner and implemented as part of the image reconstruction software of the MR scanner (1.5 T MAGNETOM Avanto, Siemens Healthcare, Erlangen, Germany). The motion-corrected images and T2 maps were typically available within ~1 s after image acquisition.

Phantom Study

To ensure that the nonrigid registration algorithm did not bias the estimated T2 values, the T2 maps of a static two-compartment nickel-doped agar-phantom (True T1/T2: 1099/49 ms and 1242/58 ms) were acquired with and without the nonrigid registration algorithm. True T1 and T2 values of these phantoms were measured using spin echo sequences with and without inversion recovery preparation, respectively, with pulse repetition time = 10 s for each sequence (acquisition time per image was ~17 min).

Volunteer Study

Ten healthy subjects consented to participate in this Health Insurance Portability and Accountability Act (HIPAA)-compliant study approved by the institution’s Human Subjects Committee. For each subject, localizer scouts were first acquired to define mid-ventricular SAX and HLA slices. Thereafter, T2P-SSFP images were acquired under three conditions: (i) free-breathing without respiratory navigator, (ii) breath-hold (duration = 7 RR intervals), and (iii) free-breathing with respiratory navigator. In each case, the acquisition was gated to mid-diastole, and T2 maps were generated with and without nonrigid registration, yielding a total of six maps per slice.

Image Analysis

As the registration algorithm uses the center frame (T2P = 24 ms) as reference, this frame was unaffected by the algorithm and was the same in both unregistered and registered T2P image sets; this image was selected to draw a ROI covering the entire myocardium in SAX view. For the HLA view, the apex was excluded from the ROI due to limited resolution and also to avoid partial volume effects. The ROI was then transferred to other images and to the T2 maps for both unregistered and registered sets. These steps were performed using Matlab software (The Mathworks, Natick, MA).

First, mean T2 values before and after registration were compared for any T2 variation that may be caused as a side-effect of registration. Thereafter, the performance of motion-compensation strategies was evaluated with two performance metrics:

  1. Variability of myocardial T2, which was measured using SD (σT2) and interquartile range of T2 (IQR-T2) values. While σT2 is a standard measure of dispersion, it is sensitive to extreme outliers; therefore, we include IQR-T2, as it is less sensitive to outliers.

  2. Percentage of myocardial pixels with normal T2 (%T2Normal), where normal T2 was defined as the range of values within two SDs of mean T2 in the breath-hold acquisition.

Additionally, pixels with abnormal T2 values were investigated for connectivity, and a set of 10 or more connected abnormal pixels were considered to be a region of T2 abnormality (15); given that the volunteers were healthy, all such cases represented false-positive diagnosis. The different motion-compensation strategies (registration, navigator, and navigator + registration) were compared with (i) free-breathing without compensation to demonstrate the improvement offered by the strategy and (ii) breath-hold acquisition (considered the gold-standard) to demonstrate the feasibility of using the strategy in a clinical setting.

Statistical Analysis

Myocardial T2 values are expressed as mean ± SD. To test the hypothesis that nonrigid registration did not alter mean myocardial T2 values, Student’s t-test was used for the phantom and in vivo data.

Signal intensities of misregistered pixels may not follow an exponential decay curve, or even a monotonically decreasing pattern; as a result, the fitted T2 can vary over a wide range, including negative and very large positive values. The presence of such outlier pixels, along with other misregistered pixels, had two implications: (i) the high and low T2 pixels could balance each other out, leaving the mean T2 value unaltered; thus, we do not report the mean T2 values and do not perform tests of significance for the mean T2 values for T2 maps generated from misregistered images. (ii) The resulting distributions of σT2, IQR-T2, and %T2Normal values across subjects were nongaussian; to account for this, σT2, IQR-T2, and %T2Normal for different cases (combinations of acquisition and postprocessing) were compared using nonparametric Wilcoxon signed rank tests. A P < 0.05 was deemed statistically significant, and Bonferroni correction was used for multiple comparisons.

All analyses were performed using MINITAB statistical software package (Minitab Inc., State College PA).

RESULTS

Effect of Registration on T2 Values

Phantom results are shown in Table 2 and indicate that the image registration algorithm caused no change in T2 values. For breath-hold acquisitions in the 10 subjects, the pixelwise T2 values in SAX view were 52.3 ± 6.5 ms before and 53.4 ± 6.1 ms after registration, whereas for HLA view, the values were 54.5 ± 7.1 ms and 56 ± 6.8 ms (P value = non-significant (NS) for both). These results show that the use of nonrigid registration did not have a significant side-effect on the distribution of myocardial T2 values.

Table 2.

Phantom Results Demonstrating the Effect of Nonrigid Registration on T2 Values

Phantom 1 (True T1/T2 = 1099 ms/49 ms) Phantom 2 (True T1/T2 = 1242 ms/58 ms)
Without registration 55.0 ms ± 1.0 ms 63.3 ms ± 0.9 ms
Without registration 54.9 ms ± 1.0 ms 63.2 ms ± 0.9 ms

Results from individual and combined motion-compensation strategies are shown in Tables 3 and 4. In summary, registration alone reduced σT2 and IQR-T2 and increased %T2Normal in all cases, giving T2 maps that were comparable to those from breath-hold acquisitions. On the other hand, although navigator alone reduced σT2 and IQR-T2 and increased %T2Normal, the resulting T2 maps had higher variability of pixel-wise T2 than those from breath-hold acquisitions; this was corrected for by combining navigator with registration, with the latter compensating for residual motion. Figures 3 and 4 demonstrate the benefit of adding registration to navigator acquisitions.

Table 3.

Effect of Different Motion-Compensation Strategies on Myocardial T2 Variability, Expressed (in ms) Using Two Parameters: σT2 and IQR of T2 Values of Myocardial Pixels

T2 variability (expressed in ms) in SAX T2 variability(expressed in ms) in HLA


Parameter None BH REG NAV NAV + REG None BH REG NAV NAV + REG
σT2 51.2 6.4 6.5 12.2 6.1 79.3 6.9 7.1 11.6 6.6
IQR-T2 25.8 7.6 9.1 9.4 8.4 26.1 8.5 10.1 9.9 9.2

Values represent estimate of median (Hodges–Lehmann estimate, computed in Wilcoxon signed rank test) of 10 volunteers. BH, breath-hold; REG, registration, NAV, navigator.

Table 4.

Effect of Different Motion-Compensation Strategies on Percentage of Pixels With Normala T2 (%T2Normal)

Percentage of pixels with normal T2 in SAX Percentage of pixels with normal T2 in HLA


Volunteer None BH REG NAV NAV + REG None BH REG NAV NAV + REG
1 42.9 94.3 86.7 83.6 78.3 61.1 94.9 98.8 80.4 93.0
2 74.2 96.4 100.0 95.8 100.0 47.7 94.5 91.2 72.8 92.0
3 42.7 94.6 85.5 85.3 88.1 58.7 94.7 88.8 95.3 93.2
4 57.4 94.7 79.1 73.4 84.4 73.7 95.0 83.9 85.9 95.3
5 66.8 94.4 99.7 92.4 96.5 81.6 95.5 91.8 87.3 97.8
6 48.0 94.7 94.6 78.4 92.9 51.0 95.3 86.9 83.6 86.0
7 80.6 94.4 100.0 99.7 100.0 59.6 95.5 94.8 84.7 91.2
8 49.4 95.7 91.4 80.6 91.6 63.7 93.5 79.0 86.3 93.1
9 77.2 95.2 93.8 90.3 97.5 75.8 95.4 89.3 90.2 91.7
10 61.5 95.0 95.5 84.6 87.3 62.2 93.6 88.0 90.7 93.3
Medianb 60.1 94.8 93.2 86.5 92.2 62.3 94.9 89.3 86 92.7

BH, breath-hold; REG, registration; NAV, navigator.

a

Normal T2 was defined as the range of values within two SDs of mean T2 in the breath-hold acquisition

b

Values represent estimate of median (Hodges–Lehmann estimate), computed in Wilcoxon signed rank test.

FIG. 3.

FIG. 3

SAX (a) and HLA (b) T2 maps from a healthy subject for each combination of acquisition and postprocessing. For each view, column 2 is free-breathing without navigator, and column 3 is free-breathing with navigator; nonrigid registration was used for bottom row only. Although the navigator reduced motion-induced T2 variability (row 1, column 2 vs. column 3 for both views), residual T2 heterogeneity was evident across the myocardium; registration improved T2 map (row 2, column 3), yielding more homogeneous myocardial T2 values. Also note that the nonrigid registration alone showed robust performance, correcting for motion-induced heterogeneity in T2 values (column 2, row 1 vs. row 2). Values expressed as mean ± SD. Display window: 0–150. [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.]

FIG. 4.

FIG. 4

Pixels with normal (gray) and abnormal (black) T2 values for different cases in the same subject shown in Fig. 3. Normal range of T2 was defined in the breath-hold case (mean ± 2SD); for this case, the percentage of pixels with normal T2 (%T2Normal) is 95% by definition. %T2Normal for other cases is shown in the images. Note that after registration, %T2Normal increases, with most of the outlier pixels scattered along the edges and not forming a cluster of connected pixels.

Performance of Nonrigid Registration Only

T2 maps from free-breathing acquisitions compensated with only nonrigid registration were comparable to those from breath-hold acquisitions. Figure 2 demonstrates the effect of registration on T2 values. In all 10 subjects, the proposed nonrigid registration significantly improved (P < 0.05) T2 maps by reducing pixel-wise T2 variability as defined by (i) σT2 from 51.2 to 6.5 ms in SAX and 79.3 to 7.1 ms in HLA and (ii) IQR-T2 from 25.8 to 6.5 ms for SAX and 26.1 to 10.1 ms for HLA, and increasing %T2Normal from 60.1 to 93.2% in SAX and 62.3 to 89.3% in HLA.

FIG. 2.

FIG. 2

Effect of nonrigid registration on SAX (a) and HLA (b) T2 maps of a healthy subject. Acquisition was free-breathing without respiratory navigator. For each view, T2P-SSFP images and T2 maps preregistration (top row) and postregistration (bottom row) registration are shown. Contours were drawn on T2P = 24 ms image, the reference frame in registration. Myocardial T2 map in the top row of each view contains misregistered pixels, resulting in heterogeneous T2; after nonrigid registration, the myocardial T2 is homogenous. Values expressed as mean ± SD. Display window: 0–150.

After registration, T2 maps from free-breathing acquisition were comparable to those from breath-hold acquisition, with the median difference of σT2 = 0.1 ms/0.1 ms (P = NS for both), of IQR-T2 = 1.1 ms (P = 0.08)/1.6 ms (P = 0.01), and that of %T2Normal = 1.9% (P = 0.48)/5.7% (P = 0.01) for SAX/HLA. Further, the residual outlier pixels in registered T2 maps did not form a region of 10 or more connected pixels in any view (see column 2 in Figs. 3 and 4).

Performance of Respiratory Navigator Only

For the 20 acquisitions (two views in 10 subjects), navigator efficiency ranged from 15 to 100% (Median = 50%; IQR = 67%), resulting in a total scan time that ranged from a minimum of 7 to a maximum of 40 RR intervals. The use of respiratory navigator alone improved T2 maps in free-breathing acquisition; these maps were, however, inferior to those from breath-hold acquisitions.

Although navigator alone significantly improved T2 maps (P value < 0.05) by reducing pixel-wise T2 variability as defined by (i) σT2 from 51.2 to 12.2 ms in SAX and 79.3 to 11.6 ms in HLA and (ii) IQR-T2 from 25.8 to 9.4 ms for SAX and 26.1 to 9.9 ms for HLA and increasing %T2Normal from 60.1 to 86.5% in SAX and 62.3 to 86% in HLA, Figs. 3 and 4 show volunteer images where navigator alone did not provide adequate motion compensation. Note also in Fig. 4 that the outlier pixels were connected, mimicking a region with T2 abnormality; similar results were observed in 8/20 other views.

Navigator alone gave T2 maps that were inferior to those from breath-hold acquisition, with the median difference of σT2 = 6.8 ms (P = 0.05)/5.2 ms (P = 0.01), and that of %T2Normal = 8.5%/8.7% (P value < 0.05) for SAX/HLA. However, when expressed as IQR, the pixel-wise T2 variability was comparable to those from breath-hold acquisition; the median difference of IQR-T2 was 1.7 ms for SAX and 1.8 ms for HLA (P value = NS).

Performance of Navigator + Registration

The benefit of additional compensation by registration over navigator gating alone is seen in column 3, bottom row of Figs. 3 and 4. Similar results were noted in other cases where navigator alone was inadequate.

For free-breathing T2 mapping, the combined use of navigator and registration significantly improved T2 maps (P value < 0.05) by reducing pixel-wise T2 variability as defined by (i) σT2 from 51.2 to 6.1 ms in SAX and HLA 79.3 to 6.6 ms in HLA and (ii) IQR-T2 from 25.8 to 8.4 ms for SAX and 26.1 to 9.2 ms for HLA, and increasing %T2Normal from 60.1 to 92.2% for SAX and 62.3 to 92.7% for HLA.

These T2 maps were comparable to those from breath-hold acquisition, with median difference of σT2 = 0.3 ms/0.4 ms (P = NS), of IQR-T2 = 0.5 ms/0.9 ms (P = NS), and of %T2Normal = 2.8% (P = 0.2)/1.9 % (P = 0.04) for SAX/HLA. The residual outlier pixels did not form regions of ten or more connected pixels in SAX or HLA views in any of the volunteers.

DISCUSSION

This work has demonstrated that the combination of navigator gating and nonrigid motion compensation enables free-breathing myocardial T2 mapping with results comparable to data acquisition during breath-hold. Further, the nonrigid registration and T2 map calculation are fully automatic, and the entire process is accomplished inline in less than 1 s, enabling this method to be easily applied in the clinical setting.

Although our study showed that nonrigid registration alone can successfully motion-correct the images acquired during free-breathing, we do not generally recommend this approach, especially for HLA view. Respiratory-induced cardiac displacement is predominantly in the superior–inferior direction, with the displacement in the other two orthogonal directions being relatively less (13); this translates to a greater through-plane motion for HLA and a greater in-plane motion for SAX in the majority of patients. Focal pathologies that are limited in spatial extent may not be captured in every 2D frame due to through-plane motion; indeed, Kellman et al. (16) reported this effect in multiple free-breathing single-shot late gadolinium enhancement HLA images but did not observe it in the SAX view. For T2 mapping, if the different T2P frames do not capture the pathology, then the subsequent registration and T2 curve-fitting will lead to inaccuracies in T2 quantification; this was not a problem in this study in either SAX or HLA view because of the homogeneous myocardium found in healthy volunteers. We included the case of free-breathing acquisition with only nonrigid registration to demonstrate the robustness of the registration algorithm.

In clinical applications, either breath-hold or respiratory navigator gating should be used for respiratory compensation to ensure a reproducible position of the heart between images. Breath-hold acquisitions are prone to diaphragm drift (13), and our results showed imperfect motion compensation with navigator gating and prospective slice correction. These inadequacies can be addressed by the additional motion compensation provided by nonrigid registration, which has shown robust performance in this study. Further, nonrigid registration can mitigate the effects of in-plane misregistrations caused by inconsistent cardiac rhythm or patient movement. Such misregistrations could not be experimentally controlled in vivo, and hence we chose to acquire free-breathing images without navigator to demonstrate the effectiveness of our nonrigid registration under extreme motion. In this way, the two motion-compensation strategies— respiratory navigator and nonrigid registration—provide complementary motion compensation.

While a breath-hold of 7 RR is within the reach of most individuals, we frequently encounter cardiovascular patients with substantial cardiac motion possibly due to diaphragm drift, failed attempt at voluntary breath-hold, patient noncooperation, or beat-to-beat variability in cardiac rhythm. In such cases, although registration alone can correct for in-plane motion, navigator gating must be used to compensate for through-plane motion. Furthermore, certain applications requiring high resolution T2 maps may warrant a segmented rather than single- shot k-space acquisition. Only navigator gating can compensate for motion between k-space segments; registration algorithms that work in image space will not be effective in this situation.

Recently, a nonrigid registration algorithm was proposed for motion-corrected averaging of myocardial late gadolinium enhancement images acquired in free-breathing (17); the objective was to improve signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) by averaging the registered images. The algorithm used by that study exploited B-splines to parameterize the motion field; the deformation between control points was interpolated using B-spline kernels. It should be noted that all the input images had similar signal and contrast levels; accordingly, mean square error was used as the similarity measure. Our nonrigid registration algorithm, on the other hand, solves a partial differential equation to get the pixel-wise deformation, using local cross-correlation as the similarity measure. Although both approaches rely on gradient descent optimization, with their respective performances being largely determined by the similarity measure (mean square error vs. local cross-correlation), the partial differential equation-based method leads to so-called diffeomorphic deformation, which is a motion field free of folding or tearing; this diffeomorphic property is generally not a characteristic of the B-spline deformation, unless explicit constraints are added to avoid any possible foldings by limiting the magnitude of spatial deformation on the B-spline control points (18); this will lead to slower convergence as a side-effect. Given this, and our objective of registering multiple T2P images, each with a different signal and contrast level, we chose to use the partial differential equation-based nonrigid registration algorithm.

The pixel-wise σT2 found in this study is higher than previously reported values (7), possibly because contours, drawn to cover the entire myocardium, included border pixels with partial-volume effects. Additionally, the misregistered pixels varied over a very wide range. In typical clinical studies, such pixels are excluded from image analysis by appropriate contouring. Given that the purpose of this study is to quantitatively demonstrate the effect of different motion-compensation strategies on T2 maps, we did not exclude such pixels. We used two statistical measures of dispersion—SD and IQR—to demonstrate how motion correction and navigator gating effectively reduced motion-induced variability in myocardial T2 values. While SD is a valid measure of dispersion even for nongaussian spread, it is sensitive to extreme outliers; accordingly, we utilized IQR as an additional measure of motion-induced T2 variability that is less sensitive to the influence of outliers.

Several recent studies have demonstrated T2 mapping with respiratory compensation. Foltz et al. (19) used respiratory bellows to monitor respiratory phase; this approach necessitated repetition of the acquisitions that were farthest from the median respiratory position. The authors of that study speculated that motion uncertainty was a primary contributor to remaining σT2 in vivo, suggesting inadequacy of their motion-compensation strategy. Huang et al. (20) used navigator gating for respiratory motion compensation and relied on a subjective analysis to either adjust ROI position or to discard images with more than 4 pixel misregistration. Blume et al. (21) also used navigator gating but did not account for the in-plane misregistration that may arise due to patient movement or inconsistent cardiac rhythm. Compared to all previous approaches, the solution proposed in this paper is the only one to combine two techniques that provide complementary motion compensation (through-plane and in-plane). The nonrigid registration has shown robust performance even in the extreme case of free-breathing acquisition without navigator. Further, the navigator gating, nonrigid registration, and T2 map generation are fully automatic and integrated with the sequence, obviating the need for any user input.

Limitations

Given that the primary purpose of this study was to validate the nonrigid registration for T2 mapping, patients were not recruited as the myocardial T2 variability in patients would likely have confounded the investigation of variability due to misregistration. We did not investigate the effects of variable myocardial signal, as may be found in pathologies, on registration and subsequent T2 mapping. We have recently demonstrated the feasibility of using breath-hold T2 mapping with nonrigid motion correction (without navigator gating) in patients with acute myocardial infarction, myocarditis, and Tako-tsubo cardiomyopathy (6,22).

While the sequence overestimated phantom T2 values, which is consistent with and discussed in our earlier work (7), phantom results showed that registration did not affect T2 quantification. Similarly, while the in vivo T2 values postregistration did show a small systematic bias (~+3%), it was not statistically significant. %T2Normal in the proposed technique (navigator + registration) was less than the target 95% in many cases (Table 4); given that σT2 was similar to breath-hold acquisition, the reduced %T2Normal may be due to the aforementioned bias. In a quantitative approach, the resulting outlier pixels could potentially be flagged as pathologic tissue. A recent clinical study (15) defined a region as pathologic if 10 or more connected pixels had abnormal T2-weighted signal. With this as a reference, the proposed technique (navigator + registration) performed well, because most of the outlier pixels were scattered around the borders, without forming a connected region (maximum eight connected pixels in this study), thus precluding potential false-positive cases (Fig. 4, column 3).

CONCLUSIONS

We have demonstrated a fully automatic and reliable free-breathing myocardial T2 mapping technique based on single-shot, T2-prepared SSFP that provides two complementary motion-compensation schemes: respiratory navigator and nonrigid registration. Both schemes were shown to independently provide significant motion compensation by reducing myocardial T2 variability and increasing the percentage of myocardium with normal T2 in healthy volunteers. While the navigator technique was used to ensure reproducible slice position, our results indicated that it alone did not provide adequate motion compensation for reliable T2 mapping. A new nonrigid registration algorithm optimized for myocardial T2 mapping was able to correct for any residual misregistrations; further, this algorithm alone showed robust performance even during breathing for both SAX and the motion-sensitive HLA views. When combined, these two strategies provide complementary motion compensation and can be used to generate fully automatic and reliable free-breathing myocardial T2 maps.

ACKNOWLEDGMENTS

The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Heart, Lung, and Blood Institute or the National Institutes of Health.

Grant sponsor: National Heart, Lung, and Blood Institute; Grant numbers: R01HL102450, R01HL095563.

REFERENCES

  • 1.Abdel-Aty H, Cocker M, Meek C, Tyberg JV, Friedrich MG. Edema as a very early marker for acute myocardial ischemia: a cardiovascular magnetic resonance study. J Am Coll Cardiol. 2009;53:1194–1201. doi: 10.1016/j.jacc.2008.10.065. [DOI] [PubMed] [Google Scholar]
  • 2.Friedrich MG, Sechtem U, Schulz-Menger J, Holmvang G, Alakija P, Cooper LT, White JA, Abdel-Aty H, Gutberlet M, Prasad S, Aletras A, Laissy JP, Paterson I, Filipchuk NG, Kumar A, Pauschinger M, Liu P. Cardiovascular magnetic resonance in myocarditis: A JACC White Paper. J Am Coll Cardiol. 2009;53:1475–1487. doi: 10.1016/j.jacc.2009.02.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Marie PY, Angioi M, Carteaux JP, Escanye JM, Mattei S, Tzvetanov K, Claudon O, Hassan N, Danchin N, Karcher G, Bertrand A, Walker PM, Villemot JP. Detection and prediction of acute heart transplant rejection with the myocardial T2 determination provided by a blackblood magnetic resonance imaging sequence. J Am Coll Cardiol. 2001;37:825–831. doi: 10.1016/s0735-1097(00)01196-7. [DOI] [PubMed] [Google Scholar]
  • 4.Abdel-Aty H, Zagrosek A, Schulz-Menger J, Taylor AJ, Messroghli D, Kumar A, Gross M, Dietz R, Friedrich MG. Delayed enhancement and T2-weighted cardiovascular magnetic resonance imaging differentiate acute from chronic myocardial infarction. Circulation. 2004;109:2411–2416. doi: 10.1161/01.CIR.0000127428.10985.C6. [DOI] [PubMed] [Google Scholar]
  • 5.Aletras AH, Tilak GS, Natanzon A, Hsu LY, Gonzalez FM, Hoyt RF, Jr, Arai AE. Retrospective determination of the area at risk for reperfused acute myocardial infarction with T2-weighted cardiac magnetic resonance imaging: histopathological and displacement encoding with stimulated echoes (DENSE) functional validations. Circulation. 2006;113:1865–1870. doi: 10.1161/CIRCULATIONAHA.105.576025. [DOI] [PubMed] [Google Scholar]
  • 6.Verhaert D, Thavendiranathan P, Giri S, Mihai G, Rajagopalan S, Simonetti OP, Raman SV. Direct T2 quantification of myocardial edema in acute ischemic injury. JACC Cardiovasc Imaging. 2011;4:269–278. doi: 10.1016/j.jcmg.2010.09.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Giri S, Chung YC, Merchant A, Mihai G, Rajagopalan S, Raman SV, Simonetti OP. T2 quantification for improved detection of myocardial edema. J Cardiovasc Magn Reson. 2009;11:56. doi: 10.1186/1532-429X-11-56. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.He T, Gatehouse PD, Anderson LJ, Tanner M, Keegan J, Pennell DJ, Firmin DN. Development of a novel optimized breathhold technique for myocardial T2 measurement in thalassemia. J Magn Reson Imaging. 2006;24:580–585. doi: 10.1002/jmri.20681. [DOI] [PubMed] [Google Scholar]
  • 9.Chef d’hotel C, Hermosillo G, Faugeras O. Flows of diffeomorphisms for multimodal image registration. 2002 IEEE International Symposium on Biomedical Imaging, Proceedings. 2002:753–756. [Google Scholar]
  • 10.Xue H, Guehring J, Srinivasan L, Zuehlsdorff S, Saddi K, Chefdhotel C, Hajnal JV, Rueckert D. Evaluation of rigid and non-rigid motion compensation of cardiac perfusion MRI. Med Image Comput Comput Assist Interv Int Conf Med Image Comput Comput Assist Interv. 2008;11(pt 2):35–43. doi: 10.1007/978-3-540-85990-1_5. [DOI] [PubMed] [Google Scholar]
  • 11.Kellman P, Aletras AH, Mancini C, McVeigh ER, Arai AE. T2-prepared SSFP improves diagnostic confidence in edema imaging in acute myocardial infarction compared to turbo spin echo. Magn Reson Med. 2007;57:891–897. doi: 10.1002/mrm.21215. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Brittain JH, Hu BS, Wright GA, Meyer CH, Macovski A, Nishimura DG. Coronary angiography with magnetization-prepared T2 contrast. Magn Reson Med. 1995;33:689–696. doi: 10.1002/mrm.1910330515. [DOI] [PubMed] [Google Scholar]
  • 13.Scott AD, Keegan J, Firmin DN. Motion in cardiovascular MR imaging. Radiology. 2009;250:331–351. doi: 10.1148/radiol.2502071998. [DOI] [PubMed] [Google Scholar]
  • 14.Xue H, Zuehlsdorff S, Kellman P, Arai A, Nielles-Vallespin S, Chefdhotel C, Lorenz CH, Guehring J. Unsupervised inline analysis of cardiac perfusion MRI. Med Image Comput Comput Assist Interv. 2009;12(pt 2):741–749. doi: 10.1007/978-3-642-04271-3_90. [DOI] [PubMed] [Google Scholar]
  • 15.Friedrich MG, Abdel-Aty H, Taylor A, Schulz-Menger J, Messroghli D, Dietz R. The salvaged area at risk in reperfused acute myocardial infarction as visualized by cardiovascular magnetic resonance. J Am Coll Cardiol. 2008;51:1581–1587. doi: 10.1016/j.jacc.2008.01.019. [DOI] [PubMed] [Google Scholar]
  • 16.Kellman P, Larson AC, Hsu LY, Chung YC, Simonetti OP, McVeigh ER, Arai AE. Motion-corrected free-breathing delayed enhancement imaging of myocardial infarction. Magn Reson Med. 2005;53:194–200. doi: 10.1002/mrm.20333. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Ledesma-Carbayo MJ, Kellman P, Hsu LY, Arai AE, McVeigh ER. Motion corrected free-breathing delayed-enhancement imaging of myocardial infarction using nonrigid registration. J Magn Reson Imaging. 2007;26:184–190. doi: 10.1002/jmri.20957. [DOI] [PubMed] [Google Scholar]
  • 18.Rueckert D, Aljabar P, Heckemann RA, Hajnal JV, Hammers A. Diffeomorphic registration using B-splines. Med Image Comput Comput Assist Interv. 2006;9(pt 2):702–709. doi: 10.1007/11866763_86. [DOI] [PubMed] [Google Scholar]
  • 19.Foltz WD, Al-Kwifi O, Sussman MS, Stainsby JA, Wright GA. Optimized spiral imaging for measurement of myocardial T2 relaxation. Magn Reson Med. 2003;49:1089–1097. doi: 10.1002/mrm.10467. [DOI] [PubMed] [Google Scholar]
  • 20.Huang TY, Liu YJ, Stemmer A, Poncelet BP. T2 measurement of the human myocardium using a T2-prepared transient-state TrueFISP sequence. Magn Reson Med. 2007;57:960–966. doi: 10.1002/mrm.21208. [DOI] [PubMed] [Google Scholar]
  • 21.Blume U, Lockie T, Stehning C, Sinclair S, Uribe S, Razavi R, Schaeffter T. Interleaved T(1) and T(2) relaxation time mapping for cardiac applications. J Magn Reson Imaging. 2009;29:480–487. doi: 10.1002/jmri.21652. [DOI] [PubMed] [Google Scholar]
  • 22.Thavendiranathan P, Walls M, Giri S, Verhaert D, Rajagopalan S, Moore S, Simonetti OP, Raman SV. Improved detection of myocardial involvement in acute inflammatory cardiomyopathies using T2 mapping. Circ Cardiovasc Imaging. 2011 Oct 28; doi: 10.1161/CIRCIMAGING.111.967836. Epub ahead of print. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES