Abstract
The conventional imaging geometry for small animal cone beam computed tomography (CBCT) is that a detector panel rotates around the head-to-tail axis of an imaged animal (“tubular” geometry). Another unusual but possible imaging geometry is that the detector panel rotates around the anterior-to-posterior axis of the animal (“pancake” geometry). The small animal radiation research platform (SARRP) developed at Johns Hopkins University employs the pancake geometry where a prone-positioned animal is rotated horizontally between an x-ray source and detector panel. This study is to assess the CBCT image quality in the pancake geometry and investigate potential methods for improvement. We compared CBCT images acquired in the pancake geometry with those acquired in the tubular geometry when the phantom/animal was placed upright simulating the conventional CBCT geometry. Results showed signal-to-noise and contrast-to-noise ratios in the pancake geometry were reduced in comparison to the tubular geometry at the same dose level. But the overall spatial resolution within the transverse plane of the imaged cylinder/animal was better in the pancake geometry. A modest exposure increase to two folds in the pancake geometry can improve image quality to a level close to the tubular geometry. Image quality can also be improved by inclining the animal, which reduces streak artifacts caused by bony structures. The major factor resulting in the inferior image quality in the pancake geometry is the elevated beam attenuation along the long axis of the phantom/animal and consequently increased scatter-to-primary ratio in that orientation. Notwithstanding, the image quality in the pancake-geometry CBCT is adequate to support image guided animal positioning, while providing unique advantages of non-coplanar and multiple mice irradiation. This study also provides useful knowledge about the image quality in the two very different imaging geometries, i.e., pancake and tubular geometry, respectively.
Keywords: Cone beam computed tomography, image quality, image guidance, small animal irradiation
INTRODUCTION
There has been increasing interest in image-guided radiotherapy in preclinical cancer research. Several systems mimicking human treatment methodologies have been developed to facilitate accurate focal irradiation for small target in laboratory animals (Wong et al., 2008, Clarkson et al., 2011, Graves et al., 2007). One such system is the small animal radiation research platform (SARRP) developed at Johns Hopkins University (Wong et al., 2008). The capabilities of the SARRP have been characterized and reported, including its mechanical calibration, beam modeling, and ability of irradiating a target to within 0.2 mm accuracy (Matinfar et al., 2009, Tryggestad et al., 2009, Armour et al., 2010). Since SARRP became available in 2008, it has been employed in many preclinical studies that take advantage of the precision focal irradiation (Duan et al., 2008, Ford et al., 2011, Karikari et al., 2007, Purger et al., 2009, Cao et al., 2011, Zhou et al., 2011).
In typical computed tomography (CT) guided small animal radiation systems, a gantry with an opposite x-ray source and image detector is usually rotated around the longitudinal axis of a stationary animal for image acquisition (Clarkson et al., 2011, Graves et al., 2007). However, SARRP employs an unusual imaging geometry for cone beam CT (CBCT) acquisition where the animal atop a stage is rotated between a horizontally opposite, but stationary, x-ray source and flat panel detector (Figure 1(a)). This imaging setup is denoted as “pancake” geometry, describing the horizontal positioning of x-ray beam during CBCT of a prone or supine animal subject. The intent of this geometry is to provide non-coplanar irradiation on the SARRP where the beam isocenter lies on the axis of the rotating stage (Figure 1(b)). However, the pancake geometry results in uneven x-ray transmission from one angle to another during CBCT projection acquisition, particularly between the beam paths along the length and width of the animal object, which can possibly degrade the image quality. This study is to evaluate the SARRP CBCT image quality and propose potential methods for improvement. We compared CBCT images acquired in the pancake geometry with those acquired when the phantom/animal was placed upright simulating the conventional CBCT geometry (denoted as “tubular” geometry). This study provides useful knowledge about the image quality in the two very different imaging geometries, i.e., pancake and tubular geometry.
METHODS
We compared CBCT images acquired in the pancake geometry and those acquired in the conventional, i.e., tubular geometry, while keeping all parameters for image acquisition, reconstruction and analysis identical for paired comparisons. The significance level is determined at p < 0.05 for student t test. All animal experiments were performed with the approval of the Johns Hopkins Animal Care and Use Committee.
Image Acquisition
The SARRP x-ray source is mounted on a gantry with a 35 cm source-to-isocenter distance and a 53 cm source-to-detector distance. On-board CBCT is acquired by rotating the imaged object in the horizontal plane between a stationary x-ray source and a 20 cm × 20 cm flat panel detector of 1024 × 1024 pixels. In the pancake geometry, CBCT was acquired with the phantom placed flat or animal placed in prone position. In the tubular geometry, the phantom or animal was placed upright mimicking conventional CT geometry. The two imaging geometries were demonstrated in Figure 2 (a) and (b). CBCT projections were acquired with 65 kVp, 1.6 mA, a 0.1 mm-thick brass filter and a 0.4 mm x-ray focal spot (nominal value per IEC336 standard) for all studies. No automatic tube current modulation was used in this study. We used filtered back projection, i.e., FDK algorithm with Hamming filter, for CBCT reconstruction (Feldkamp et al., 1984).
Signal-to-Noise Ratio and Contrast-to-Noise Ratio
To obtain signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), we used cylindrical polymethyl methacrylate (PMMA, 1.18 g/cm3, background material) phantoms similar to mouse dimension (3 cm wide and 7.5 cm long) embedded with four heterogeneities: teflon (2.20 g/cm3), delrin (1.42 g/cm3), acrylic (1.20 g/cm3) and polyethylene (0.93 g/cm3). The phantom compositions are illustrated in Figure 2 (c) and (d). CBCT of a single phantom, three phantoms placed side-by-side, or a euthanized mouse, was acquired in the pancake and tubular geometry, respectively, and reconstructed with 0.2 mm × 0.2 mm × 0.2 mm voxels. The triple phantom arrangement is shown in Figure 2 (a) and (b). To examine the improvement by increasing radiation exposure, the imaging doses varied from 1 fold (1x, 1.6 cGy) to 2 folds (2x, 3.2 cGy) and 4 folds (4x, 6.4 cGy), by increasing the number of projections from 360 to 720 and 1440, correspondingly. In addition, CBCT of the PMMA phantom or euthanized mouse inclined at 14° and 28° were acquired to examine the effect of inclining the imaged object on the SNR and CNR. Image acquisition or measurement was repeated three times for each setup.
An in-house program developed in MATLAB (MathWorks, Natick, MA, USA) was used for the SNR and CNR analysis. A volumetric region of interest (ROI) was drawn at the center of each inhomogeneity with a diameter of 20 pixels and a length of 70 pixels (approximately 22000 pixels in total). Comparisons of SNR and CNR for all ROIs were made between the two imaging geometries. The SNR is defined as SIx/σx, where SIx is the mean signal intensity (SI) of ROIx and σx is the corresponding standard deviation. The CNR is defined as (SIx − SIbkg)/σbkg, where SIbkg is the SI of a reference volumetric ROI located in the PMMA background with similar size, and σbkg is the standard deviation of the reference ROI. For data analysis on animal CBCT, the volumetric ROIs were drawn in the liver and brain, respectively. The liver ROIs had a diameter of 20 pixels and length of 20 pixels, and brain ROIs had a diameter of 10 pixels and length of 10 pixels. SNR and CNR Data were reported in the format of mean ± standard deviation.
Spatial Resolution
To estimate the spatial resolution, a cylindrical phantom of 30 mm diameter and 70 mm length was fabricated with a straight steel wire of 50 μm diameter stretched inside a cylindrical hole. The phantom was made of acrylic and embedded with teflon, delrin, and polyethylene rods. The CBCT transection of the phantom is shown in Figure 3 (a) where the thin steel wire is shown as a bright dot in the dark air space. The CBCT was acquired with 720 projections and reconstructed with 0.25 mm × 0.25 mm × 0.25 mm voxels. The spatial resolution was measured in the transverse plane of the cylinder phantom, along radial and tangent directions, respectively, with an over-sampling method as described by Kwan et al (Kwan et al., 2007). The radial and tangent directions were defined in the coordinates of the transverse plane of the phantom, as shown in Figure 3 (a). To obtain line spread functions with over-sampling method, the wire was slightly tilted to a 5–15° angle away from the radial or tangent direction, by inclining the phantom. The resolution was measured in both pancake and tubular geometry, with single and triple phantoms, respectively. For the triple phantom study, the resolution phantom was positioned at center or side as shown in Figure 3 (a), to examine whether the resolution is dependent on the location within the reconstruction volume. Another two homogeneous acrylic cylinder of 30 mm × 70 mm were placed along with the resolution phantom. Modulation transfer functions (MTFs) were derived and plotted, and the frequency at 10% amplitude (f0.1) was used for comparison.
Scatter-to-Primary Ratio
Scatter-to-primary ratio (SPR) was measured in the pancake and tubular geometry, along the long and short axis of imaged phantoms, respectively. The definitions of the long and short axis are shown in Figure 4 (a). In the single phantom pancake geometry, the long axis setup is defined as the beam central axis passes through the long axis of the cylinder, and the short axis setup defined as the beam central axis passes through the diameter of the cylinder. In the single phantom tubular geometry, the long axis and short axis setup are identical, both referring to that the beam central axis passes through the cylinder diameter. In the triple phantom studies, the long axis setup is defined as the beam central axis passes through three phantoms, and the short axis setup as the beam central axis passes through the center phantom only. A tungsten wire of 1 mm diameter and10 mm length was attached along the long axis of a homogeneous acrylic cylinder of 30 mm diameter and 70 mm length to block part of primary x-ray beam as shown in Figure 4 (b). The wire was positioned on the cylinder phantom at beam entrance and exit, respectively. A projection image of 1024 × 1024 pixels (0.2 mm pixel size) was acquired with the shadow of the tungsten wire shown at the image center. Dark current was subtracted from the image, but image lag correction was not performed because single image was acquired instead of an image sequence. The x-ray attenuation caused by the tungsten wire resulted in a dip in the measured signal profile as shown in Figure 4(c). The x-ray path through the diameter of the tungsten wire is equivalent to approximately 40 tenth value layers at the mean energy of the 65 kVp beam. Therefore, the x-ray transmission through the wire was negligible, and the signal intensity at the dip was considered contributed solely from scatter (SIscatter=SIbottom, Figure 4(c)). The signals measured on the shoulders of the dip (SIleft and SIright) can be averaged and approximated as the total signal (SItotal) at the dip otherwise without the wire in place. The total signal SItotal is the summation of the primary SIprimary and scatter SIscatter components. The SPR is calculated as
(1) |
Angular-Dependent Transmission
The average x-ray transmission, defined as the mean value of all the pixels within a large ROI that contoured the entire phantom or mouse on two-dimensional projection images, was measured in both tubular and pancake geometries at stage rotation angles of 0°, 90°, 180° and 270°. In the pancake geometry, rotation angle 0° or 180° is defined at the position where the long axis of the imaged object is perpendicular to the beam central axis. In the tubular geometry, rotation angle 0° or 180° is defined at the position where the beam central axis passes through all three phantoms in the triple phantom study, and at the position where the beam central axis passes laterally in the single mouse study. There is no angular difference in the single phantom tubular geometry.
RESULTS
SNR and CNR
Single Phantom Study
The SNR and CNR measurement results are shown in Table 1 and 2 respectively. The SNR and CNR in the pancake geometry were inferior to their counterparts in the tubular geometry at the same dose level. The SNRs in the pancake geometry were 17.3±2.7, 20.7±2.9, 23.1±4.1 and 26.0±5.6, for polyethylene, acrylic, delrin and teflon, respectively, versus 23.2±1.8, 26.3±0.5, 33.9±1.2, and 48.5±3.0 in the tubular geometry (p < 0.05 for all comparisons). The most dramatic discrepancy occurred in the high density heterogeneity teflon. At 2x exposure, the SNRs in the pancake geometry were superior or nearly equal to those in the tubular geometry at 1x exposure, for all heterogeneities except the teflon. The SNR for the teflon was recovered when the exposure was increased to 4x. Besides increasing exposure, inclining the phantom also enhanced the SNR. With a modest 14° inclination, the SNRs were improved by 16.5%, 17.3%, 22.8% and 55.1% for polyethylene, acrylic, delrin and teflon, respectively, compared to the pancake geometry without inclination. The most significant improvement by inclination occurred for the teflon.
Table 1.
ROI | Tubular 1x | Pancake 1x | Pancake 2x | Pancake 4x | Pancake 14°, 1x |
---|---|---|---|---|---|
Teflon | 48.5±3.0 | 26.0±5.6 | 36.7±3.6 | 48.7±3.6 | 40.4±2.2 |
Delrin | 33.9±1.2 | 23.1±4.1 | 33.7±1.0 | 45.9±2.0 | 28.3±1.0 |
Acrylic | 26.3±0.5 | 20.7±2.9 | 30.0±0.5 | 39.5±1.0 | 24.3±0.8 |
Polyethylene | 23.2±1.8 | 17.3±2.7 | 24.9±2.1 | 34.2±1.5 | 20.1±1.2 |
PMMA | 20.0±1.5 | 17.3±2.2 | 25.3±1.5 | 33.0±1.0 | 20.6±1.0 |
Table 2.
ROI | Tubular 1x | Pancake 1x | Pancake 2x | Pancake 4x | Pancake 14°, 1x |
---|---|---|---|---|---|
Teflon | 21.5±2.0 | 12.0±1.3 | 17.5±0.9 | 22.5±1.4 | 18.6±1.8 |
Delrin | 5.8±0.9 | 4.0±0.3 | 5.8±0.7 | 7.6±1.0 | 5.3±0.7 |
Acrylic | 1.0±0.3 | 0.7±0.3 | 1.1±0.6 | 1.4±0.8 | 0.7±0.3 |
Polyethylene | −3.7±0.3 | −2.6±0.3 | −3.8±0.2 | −4.8±0.2 | −3.1±0.3 |
The CNRs in the pancake geometry at 1x exposure were −2.6±0.3, 0.7±0.3, 4.0±0.3, and 12.0±1.3, for polyethylene, acrylic, delrin and teflon, respectively, versus −3.7±0.3, 1.0±0.3, 5.8±0.9, and 21.5±2.0 in the tubular geometry (p < 0.05 for all comparisons except the acrylic). Similar to the SNR results, the largest discrepancy occurred for the teflon. We increased the CNRs to −3.8±0.2, 1.1±0.6, 5.8±0.7 and 17.5±0.9, respectively, by doubling the exposure in the pancake geometry. The CNRs in the pancake geometry at 2x exposure were superior or nearly equal to those in the tubular geometry at 1x exposure, for all heterogeneities except the teflon. The CNR for the teflon was recovered when the exposure was increased to 4x. Again, inclining the phantom to 14° improved the CNRs in the pancake geometry for all the four heterogeneities.
Triple Phantom Study
The SNR and CNR measurement results for the triple phantom study are shown in Table 3 and 4, respectively. The SNR and CNR in the pancake geometry were inferior to those in the tubular geometry. The SNRs in the pancake geometry were 13.1±1.0, 12.9±1.4, 14.8±1.5, and 18.5±1.4, for polyethylene, acrylic, delrin and teflon, respectively, versus 17.1±1.3, 23.2±1.1, 27.5±0.4 and 39.5±2.5 in the tubular geometry (p < 0.05 for all comparisons). The CNRs in the pancake geometry were −1.8±0.2, 0.5±0.1, 2.9±0.2, and 8.7±0.8, versus −3.4±0.3, 1.6±0.1, 6.1±0.3 and 20.5±1.3 in the tubular geometry (p < 0.05 for all comparisons). Similar to the single phantom study, increasing exposure improved the SNR and CNR in the pancake geometry. The SNRs in pancake geometry were recovered at 4x exposure level for all heterogeneities except the teflon. The CNRs, however, required more than 4x exposure to achieve a level identical to the tubular geometry.
Table 3.
ROI | Tubular 1x | Pancake 1x | Pancake 2x | Pancake 4x |
---|---|---|---|---|
Teflon | 39.5±2.5 | 18.5±1.4 | 25.1±2.7 | 32.5±4.3 |
Delrin | 27.5±0.4 | 14.8±1.5 | 20.0±2.3 | 29.0±4.2 |
Acrylic | 23.2±1.1 | 12.9±1.4 | 17.5±2.4 | 23.9±2.5 |
Polyethylene | 17.1±1.3 | 13.1±1.0 | 17.6±1.8 | 23.9±3.7 |
Background | 20.0±1.5 | 11.9±1.3 | 16.1±2.1 | 21.9±3.6 |
Table 4.
ROI | Tubular 1x | Pancake 1x | Pancake 2x | Pancake 4x |
---|---|---|---|---|
Teflon | 20.5±1.3 | 8.7±0.8 | 11.7±1.4 | 15.7±2.3 |
Delrin | 6.1±0.3 | 2.9±0.2 | 3.9±0.4 | 5.3±0.8 |
Acrylic | 1.6±0.1 | 0.5±0.1 | 0.7±0.1 | 0.9±0.0 |
Polyethylene | −3.4±0.3 | −1.8±0.2 | −2.5±0.3 | −3.3±0.6 |
We also compared the SNR and CNR between the phantom at center and side in the pancake geometry at 1x exposure. The SNRs at side were better than those at center by 16.0%, 21.8%, 26.2%, and 14.1%, and the CNRs at side were better than those at center by 17.0%, 16.3%, 29.9%, and 21.8%, for teflon, delrin, acrylic and polyethylene, respectively. The non-uniformity, defined as the maximum percentage difference between the mean signal intensity of ROIs at center and side, was 1.5±0.1%.
The images from both pancake and tubular geometries at different exposures are shown in Figure 5. Although the SNRs and CNRs in the pancake geometry were inferior to those in the tubular geometry, they were adequate for visual differentiation of all four heterogeneities from the background.
Animal Imaging
Mouse CBCTs acquired at different imaging geometries are shown in Figure 6. Images acquired in the pancake geometry have noticeable streak artifacts in brain and thoracic regions, owing to increased x-ray attenuation by bony structures. The streak artifacts in brain was more severe than other anatomical sites due to abundance of bony structures in mouse brain. We reduced the streak artifacts by inclining the animal. The SNR analysis in brain and liver showed that either doubling the radiation exposure or inclining the animal could improve image quality in the pancake geometry, as shown in Table 5. The SNRs in the pancake geometry at 2x exposure approached to those in the tubular geometry at 1x exposure (25.6±1.0 versus 26.1±0.6 in brain and 23.6±1.0 versus 26.9±0.1 in liver, p < 0.05 for both comparisons). At 28° inclination, the SNR in the pancake geometry at 1x exposure was 22.4±1.3 in brain and 26.8±1.5 in liver, and improved by 42.7% and 21.8%, respectively, compared to the 14° inclination (p = 0.01 and 0.89, respectively). No obvious SNR improvement was observed at 14° inclination compared to the non-inclination.
Table 5.
ROI | Tubular 1x | Pancake 14°, 1x | Pancake 28°, 1x | Pancake 2x | Pancake 14°, 2x |
---|---|---|---|---|---|
Brain | 26.1±0.6 | 15.7±0.4 | 22.4±1.3 | 25.6±1.0 | 25.2±0.9 |
Liver | 26.9±0.1 | 21.8±1.4 | 26.8±1.5 | 23.6±1.0 | 23.5±1.4 |
Spatial Resolution
Single Phantom Study
The MTFs for single phantom study are shown in Figure 3 (b). The tangent resolution in the pancake geometry (f0.1 = 2.4 mm−1) is better than either resolution (f0.1 = 1.6 mm−1 for the tangent resolution and 1.9 mm−1 for the radial resolution) in the tubular geometry. The radial resolution in the pancake geometry (f0.1 = 1.8 mm−1) is similar to its counterpart in the tubular geometry.
Triple Phantom Study
The MTFs for triple phantom study are shown in Figure 3 (c). Similar to the single phantom study, the resolutions in the pancake geometry are either similar or better than those in the tubular geometry. In the pancake geometry, the tangent resolutions (f0.1 = 2.4 mm−1 at center and 2.5 mm−1 at side) are better than the radial resolutions (f0.1 = 1.8 mm−1 for both center and side). In the tubular geometry, the radial resolutions (f0.1 = 1.8 mm−1 for both center and side) are slightly better than the tangent ones (f0.1 = 1.6 mm−1 for both center and side). The location of the phantom, either at center or side, has minimal effect on the resolution. The resolution at side is not worse, at least, than that at center. The spatial resolutions in the triple phantom study were very similar to their counterparts in the single phantom study.
Angular Dependent Transmission
The angular dependent transmission data for phantom studies are shown in Figure 7 (a) and (b). In the single phantom pancake geometry, the amount of radiation transmitting through the single phantom was dependent on the x-ray projection angle. The transmission was smallest at the projection angles 90° and 270°, where x-ray beam central axis aligned with the long axis of the cylinder. In the single phantom tubular geometry, the transmission was similar at all projection angles owing to the axial symmetry of the phantom rotation. The reverse was observed when transmission was measured in the triple phantom study. The transmission was nearly independent of projection angles in the pancake geometry. In the tubular geometry, the transmission was smallest when the beam central axis passed through three phantoms, and was greatest when it only passed through one phantom.
The transmission data for the single mouse study are shown in Figure 7(c). In the pancake geometry, the transmission shows strong angle dependency at no inclination. The greatest transmission occurred when the beam central axis passed the animal body laterally, and smallest when the beam central axis passed along the craniocaudal direction. The transmission uniformity over projection angles was improved by animal inclination in the pancake geometry. The tubular geometry can be viewed as the extreme case of the pancake geometry at 90° inclination, which resulted in maximal transmission along the anterior-posterior direction and minimal transmission along the lateral direction in the single mouse study. Animal inclination in the pancake geometry improved transmission uniformity, approaching that in the tubular geometry with increasing inclination angles.
Scatter-to-Primary Ratio
The SPRs measured at x-ray beam entrance and exit were almost identical. Therefore the mean values were used and shown in Table 6. The primary and scatter x-ray passing through the imaged object at any project angle in the tubular geometry were similar to those in the pancake geometry when the beam central axis was perpendicular to the phantom long axis, resulting in similar SPRs (0.14 and 0.13 in the tubular geometry, versus 0.14 for the short axis setup in the pancake geometry). When the beam central axis penetrated through the long axis in the pancake geometry, the primary x-ray intensity deceased by 68.6%, and the SPR increased by 64.3% to 0.23. In the triple phantom tubular geometry, the SPR was greater when the beam central axis penetrated through three phantoms (0.30 in the long axis setup versus 0.16 in the short axis setup). In the pancake geometry, the SPR was 0.27 when the beam central axis penetrated through single phantom (the short axis setup), and 0.33 when it penetrated through three phantoms (the long axis setup). Increased scatter portion at certain projection angles is one of the major causes that worsen the image quality in the pancake geometry.
Table 6.
Geometry | Single Phantom | Triple phantom | ||
---|---|---|---|---|
| ||||
Tubular | Pancake | Tubular | Pancake | |
Short axis | 0.14 | 0.14 | 0.16 | 0.27 |
Long axis | 0.13 | 0.23 | 0.30 | 0.33 |
Note: the standard deviations are not presented because they are negligible.
DISCUSSION
This study investigated the image quality of the CBCT acquired in an unconventional geometry, i.e., pancake geometry and compared it with the conventional geometry, i.e., tubular geometry. At the same exposure level, the SNR and CNR in the pancake geometry were inferior to those in the tubular geometry. The degradation in the SNR and CNR is expected owing to appreciable non-uniformity in the x-ray transmission over projection angles when an elongated mouse-like object is rotated in the stationary x-ray beam, as demonstrated in the angular dependent transmission study. The transmission can be up to two times difference in one rotation in the single phantom pancake geometry. When the x-ray penetrates through the long axis, particularly of a high density material such as a teflon rod in a phantom or the spinal column in an animal, the large attenuation would significantly limit the transmission of the primary photons and result in considerable SPR increase. Such elongated objects are usually of substantial contrast and thus would not deter visualization. They would, however, impart streak artifacts caused by the increased scatter portion and beam hardening effect. To decouple the two effects warrants nontrivial effort and is beyond the scope of this study. Overall, CBCT acquired with the pancake geometry appears to be adequate to provide image guidance for small animal radiation therapy. Images of the contrast phantom shown in Figure 5 illustrate that 5 mm objects at a low 2% contrast, such as acrylic embedded in PMMA, can be detected. It should be noted that a SNR value of five was achieved for all the objects in the phantom imaging study at all exposure levels which satisfied the simple SNR Rose Criteria for visualization (Rose, 1973).
The SNR and CNR in the pancake geometry can be improved by either increasing the exposure or inclining the imaged object. Doubling the exposure can increase SNR and CNR to the level approaching the tubular geometry. The imaging dose is about 3.2 cGy at 2x exposure level, which is still lower than the 10 cGy threshold, below which neither deterministic nor stochastic effect due to radiation has ever been demonstrated (Mole, 1957, ICRP, 2007). Inclining the imaged objects can effectively reduce the streak artifact caused by elongated bony structures. This method is attractive because it does not incur additional dose to the animal, although more involved image manipulation utility, such as image tilting, is needed to allow visualization in customary orientation. On the other hand, while inclination of the imaged mouse can improve CBCT image quality, it may need extra efforts to ensure a reproducible animal setup for fractionated radiation treatment. A special animal holder with a properly tilted bed is needed to incline the mouse. Additional attention should be paid to ensure a steady fixation while taping the animal to the holder. Notwithstanding, with the CBCT guidance, the animal positon can always be adjusted to best reproduce the simulation or initial treatment setup. Another option is to hang up the animal vertically, which mimics the tubular geometry. Sharma et al has successfully delivered fractionated radiation treatment to rats by hanging up animals with a custom-made holder (Sharma et al., 2014). There are other potential methods for reducing the streak artifacts caused by bony structures, one of which is using higher energy photon beams which are considered less susceptible to bone attenuation. The general image quality can also be improved by implementing automatic current modulation technique to offset the angular dependent variation of x-ray transmission in the pancake geometry.
One concern of using pancake CBCT for image guidance is whether or not the non-conventional imaging geometry induces imaging artefacts that detract from targeting accuracy, particularly in multiple mice irradiation. Our group performed a study that investigated the capability of image guidance with the pancake geometry (Armour et al., 2010). The positioning error for repeated alignments with bony structures in the pancake CBCT was −0.18±0.05 mm for translation and −0.4±0.9° for rotation, better than that when mouse was setup without CBCT guidance, which was 0.8±0.49mm for translation and 0±2° for rotation. Therefore, we think the image quality of pancake CBCT is sufficient for image guidance purpose. In the triple phantom study, we found that the image quality, including SNR, CNR and spatial resolution, at side was not worse, if not better, than that at center, and the signal intensity non-uniformity was < 2% between center and side. In addition, the spatial resolution was not degraded by the increased scatter in the triple phantom study. Therefore, the targeting accuracy in multiple mice study should not be compromised by the location of the irradiated mouse.
We used a tungsten wire of 1 mm diameter to block the primary radiation beam in the SPR estimation in this study. A more typical approach is using small blocks of decreasing diameter, and then extrapolate the scatter contribution. However, one study by Niu et al (Niu et al., 2011) showed that scatter was estimated within 5% around the object center with single-width lead blockers of 2 mm thickness positioned at the beam entrance. Therefore our method should be sufficient to estimate and relatively compare the scatter contribution between the two imaging geometries.
It should be noted that the imaging quality in either pancake or tubular geometry is dependent on the shapes of the imaged objects. It is possible for pancake geometry to be more advantageous for certain object shapes/setups, although CBCT of cylindrically shaped objects can be aptly acquired with the conventional, i.e. tubular, geometry. One gain from the pancake geometry is the reduced cone angle that can fully cover the imaged animal, which may improve the vertical spatial resolution along the cone-opening direction. Previous studies have shown the spatial resolution along this direction deceases with increased cone angle (Mettivier et al., 2011). Based on our study, the spatial resolution along certain direction in the pancake geometry was actually better than that in the tubular geometry. Spatial resolution in CBCT varies depending on the direction along which it is measured (Mettivier et al., 2011). Usually it can be measured in three directions: the beam central axis direction, the cathode to anode direction, and the third direction perpendicular to the above two directions. In our study, the measurement directions were defined in the image coordinates, particularly in the transverse plane, the most common plane in image browsing, of the imaged phantom or animal. In the pancake geometry, the radial resolution was defined along the beam central axis direction, and the tangent resolution defined along the direction perpendicular to both the beam central axis and cathode to anode direction. In the tubular geometry, the radial resolution was defined along the beam central axis, and the tangent resolution defined along the cathode to anode direction. This probably explains why the radial resolutions were similar between the two imaging geometries. Comparing the single phantom study with triple phantom study, the corresponding resolutions were almost the same, indicating that the increased scatter in the triple phantom study does not degrade spatial resolution. We recognized, however, that the above conclusions drawn upon the high contrast resolution used in this study do not necessarily apply to low contrast resolution.
The pancake geometry on the SARRP is chosen primarily for its irradiation advantages. In image guided radiation therapy, with the axes of its translation and rotation stages readily aligned with the isocenter of the x-ray gantry, delivery of non-coplanar beams is simplified with conical beam tracks as shown in Figure 1(b). To mimic human radiotherapy situation, we need deliver focal radiation to the tumor target while sparing normal tissues. One example requiring precision focal irradiation is to model the radiation induced lung injury by mimicking clinical stereotactic radiotherapy often delivered with CyberKnife non-coplanar radiation beams (Bibault, et al., 2012, Gibbs, et al., 2010). Another example is to investigate the synergetic effect of combining radiotherapy and immunotherapy (Zeng et al., 2013, Burnette et al., 2013). It is highly desired to minimize the immune response induced by unnecessary irradiation of normal tissues. SARRP can deliver stereotactic radiation with non-coplanar beams in a simplified conical beam path, and has the ability of even better sparing normal tissues by adding several layers of conical beam paths. Another advantage of the pancake setup is its ability to support image guided multiple mice non-coplanar irradiation which is difficult, if not impossible, to achieve for the conventional geometry with a non-rotating animal stage. The pancake arrangement can accommodate non-coplanar irradiation of a target in each mouse, such as using a conical beam trajectory, provided that beam paths do not traverse the other mice. This is useful because efficient multiple mice irradiation is most desirable for preclinical radiation experiments that require a large number of animals.
COCLUSION
The SNR and CNR in the pancake geometry were inferior to those in the conventional, i.e., tubular geometry, for both single and triple phantom imaging. But the overall spatial resolution within the transverse plane of the imaged cylinder/animal in the pancake geometry was better. A modest exposure increase by a factor of two in the pancake geometry improved and produced SNR and CNR comparable to those in the tubular geometry. The image quality in the pancake geometry can also be enhanced by inclining the animal, which improves x-ray transmission and reduces streak artifacts caused by bony structures. The major factor contributing to the inferior image quality in the pancake geometry is the escalated primary beam attenuation along the long axis of the phantom/animal and consequently increased SPR. Notwithstanding, the CBCT acquired in the pancake geometry is adequate for image guided animal positioning, while this geometry provides advantages of non-coplanar and multiple mice irradiation.
Acknowledgments
This study was partially supported by a research agreement with Xstrahl, Inc., and by a grant from the National Cancer Institute (NIH/NCI; grant number R01CA158100-01). We also want to acknowledge Dr. Bin Zhang at Johns Hopkins University for his assistance during CBCT image acquisition.
Abbreviations
- CBCT
Cone Beam Computed Tomography
- CNR
Contrast to Noise Ratio
- MTF
Modulation Transfer Function
- PMMA
Polymethyl Methacrylate
- ROI
Regions of Interest
- SARRP
Small Animal Radiation Research Platform
- SI
Signal Intensity
- SNR
Signal to Noise Ratio
- SPR
Scatter to Primary Ratio
References
- ARMOUR M, FORD E, IORDACHITA I, WONG J. CT guidance is needed to achieve reproducible positioning of the mouse head for repeat precision cranial irradiation. Radiat Res. 2010;173:119–23. doi: 10.1667/RR1845.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- BIBAULT JE, PREVOST B, DANSIN E, MIRABEL X, LACORNERIE T, LARTIGAU E. Image-guided robotic stereotactic radiation therapy with fiducial-free tumor tracking for lung cancer. Radiat Oncol. 2012;7:102. doi: 10.1186/1748-717X-7-102. [DOI] [PMC free article] [PubMed] [Google Scholar]
- BURNETTE B, WEICHSELBAUM RR. Radiation as an Immune Modulator. Semin Radiat Oncol. 2013;23:273–80. doi: 10.1016/j.semradonc.2013.05.009. [DOI] [PubMed] [Google Scholar]
- CAO X, WU X, FRASSICA D, YU B, PANG L, XIAN L, WAN M, LEI W, ARMOUR M, TRYGGESTAD E, WONG J, WEN CY, LU WW, FRASSICA FJ. Irradiation induces bone injury by damaging bone marrow microenvironment for stem cells. Proc Natl Acad Sci U S A. 2011;108:1609–14. doi: 10.1073/pnas.1015350108. [DOI] [PMC free article] [PubMed] [Google Scholar]
- CLARKSON R, LINDSAY PE, ANSELL S, WILSON G, JELVEH S, HILL RP, JAFFRAY DA. Characterization of image quality and image-guidance performance of a preclinical microirradiator. Med Phys. 2011;38:845–56. doi: 10.1118/1.3533947. [DOI] [PMC free article] [PubMed] [Google Scholar]
- DUAN W, PENG Q, MASUDA N, FORD E, TRYGGESTAD E, LADENHEIM B, ZHAO M, CADET JL, WONG J, ROSS CA. Sertraline slows disease progression and increases neurogenesis in N171-82Q mouse model of Huntington’s disease. Neurobiol Dis. 2008;30:312–22. doi: 10.1016/j.nbd.2008.01.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- FELDKAMP LA, DAVIS LC, KRESS JW. Practical Cone-Beam Algorithm. J Opt Soc Am A. 1984;1:612–9. [Google Scholar]
- FORD EC, ACHANTA P, PURGER D, ARMOUR M, REYES J, FONG J, KLEINBERG L, REDMOND K, WONG J, JANG MH, JUN H, SONG HJ, QUINONES-HINOJOSA A. Localized CT-guided irradiation inhibits neurogenesis in specific regions of the adult mouse brain. Radiat Res. 2011;175:774–83. doi: 10.1667/RR2214.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- GIBBS IC, LOO BW., JR CyberKnife stereotactic ablative radiotherapy for lung tumors. Technol Cancer Res Treat. 2010;9:589–96. doi: 10.1177/153303461000900607. [DOI] [PubMed] [Google Scholar]
- GRAVES EE, ZHOU H, CHATTERJEE R, KEALL PJ, GAMBHIR SS, CONTAG CH, BOYER AL. Design and evaluation of a variable aperture collimator for conformal radiotherapy of small animals using a microCT scanner. Med Phys. 2007;34:4359–67. doi: 10.1118/1.2789498. [DOI] [PubMed] [Google Scholar]
- ICRP. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37:1–332. doi: 10.1016/j.icrp.2007.10.003. [DOI] [PubMed] [Google Scholar]
- KARIKARI CA, ROY I, TRYGGESTAD E, FELDMANN G, PINILLA C, WELSH K, REED JC, ARMOUR EP, WONG J, HERMAN J, RAKHEJA D, MAITRA A. Targeting the apoptotic machinery in pancreatic cancers using small-molecule antagonists of the X-linked inhibitor of apoptosis protein. Mol Cancer Ther. 2007;6:957–66. doi: 10.1158/1535-7163.MCT-06-0634. [DOI] [PMC free article] [PubMed] [Google Scholar]
- KWAN ALC, BOONE JM, YANG K, HUANG SY. Evaluation of the spatial resolution characteristics of a cone-beam breast CT scanner. Med Phys. 2007;34:275–81. doi: 10.1118/1.2400830. [DOI] [PubMed] [Google Scholar]
- MATINFAR M, FORD E, IORDACHITA I, WONG J, KAZANZIDES P. Image-guided small animal radiation research platform: calibration of treatment beam alignment. Phys Med Biol. 2009;54:891–905. doi: 10.1088/0031-9155/54/4/005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- METTIVIER G, RUSSO P. Measurement of the MTF of a cone-beam breast computed tomography laboratory scanner. IEEE Trans Nucl Sci. 2011;58:703–13. [Google Scholar]
- MOLE RH. Quantitative observations on recovery from whole body irradiation in mice. II. Recovery during and after daily irradiation. Br J Radiol. 1957;30:40–6. doi: 10.1259/0007-1285-30-349-40. [DOI] [PubMed] [Google Scholar]
- NIU TY, ZHU L. Scatter correction for full-fan volumetric CT using a stationary beam blocker in a single full scan. Med Phys. 2011;38:6027–38. doi: 10.1118/1.3651619. [DOI] [PMC free article] [PubMed] [Google Scholar]
- PURGER D, MCNUTT T, ACHANTA P, QUINONES-HINOJOSA A, WONG J, FORD E. A histology-based atlas of the C57BL/6J mouse brain deformably registered to in vivo MRI for localized radiation and surgical targeting. Phys Med Biol. 2009;54:7315–27. doi: 10.1088/0031-9155/54/24/005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- ROSE A. Vision: human and electronic. New York: Plenum Press; 1973. [Google Scholar]
- SHARMA S, MOROS EG, BOERMA M, SRIDHARAN V, HAN EY, CLARKSON R, HAUER-JENSEN M, CORRY PM. A Novel Technique for Image-Guided Local Heart Irradiation in the Rat. Technol Cancer Res T. 2014;13:593–603. doi: 10.7785/tcrtexpress.2013.600256. [DOI] [PMC free article] [PubMed] [Google Scholar]
- TRYGGESTAD E, ARMOUR M, IORDACHITA I, VERHAEGEN F, WONG JW. A comprehensive system for dosimetric commissioning and Monte Carlo validation for the small animal radiation research platform. Phys Med Biol. 2009;54:5341–57. doi: 10.1088/0031-9155/54/17/017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- WONG J, ARMOUR E, KAZANZIDES P, IORDACHITA I, TRYGGESTAD E, DENG H, MATINFAR M, KENNEDY C, LIU Z, CHAN T, GRAY O, VERHAEGEN F, MCNUTT T, FORD E, DEWEESE TL. High-resolution, small animal radiation research platform with x-ray tomographic guidance capabilities. Int J Radiat Oncol Biol Phys. 2008;71:1591–9. doi: 10.1016/j.ijrobp.2008.04.025. [DOI] [PMC free article] [PubMed] [Google Scholar]
- ZENG J, HARRIS TJ, LIM M, DRAKE CG, TRAN PT. Immune modulation and stereotactic radiation: improving local and abscopal responses. Biomed Res Int. 2013;2013:658126. doi: 10.1155/2013/658126. [DOI] [PMC free article] [PubMed] [Google Scholar]
- ZHOU J, TRYGGESTAD E, WEN Z, LAL B, ZHOU T, GROSSMAN R, WANG S, YAN K, FU DX, FORD E, TYLER B, BLAKELEY J, LATERRA J, VAN ZIJL PC. Differentiation between glioma and radiation necrosis using molecular magnetic resonance imaging of endogenous proteins and peptides. Nat Med. 2011;17:130–4. doi: 10.1038/nm.2268. [DOI] [PMC free article] [PubMed] [Google Scholar]