Direct evaluation of MR-derived attenuation correction maps for PET/MR of the mouse myocardium

Eleanor Evans 1,, Guiso Buonincontri 1, Rob C Hawkes 1, Richard E Ansorge 2, T Adrian Carpenter 1, Stephen J Sawiak 1,3
PMCID: PMC4545627  PMID: 26501677

Attenuation correction (AC) must be applied to provide accurate measurements of PET tracer activity concentrations. Due to the limited space available in PET/MR scanners, MR-derived AC (MRAC) is used as a substitute for gold standard transmission source scans [1]. We compared MRAC to transmission scans to evaluate its performance in mouse myocardium studies.

PET SUV values derived for 10 mice [2] using whole body MRAC maps were compared to those attained using AC maps from a transmission source. 3D FISP was acquired using a 4.7T Bruker BioSpec before the mouse was transferred on a standard Bruker animal bed (with single loop surface coil) to the Cambridge split magnet PET/MR [3]. A 10 minute transmission scan (68Ge) was performed. Emission data was acquired for 45 minutes following ~25MBq 18F-FDG administration.

MRAC comparison Following co-registration using SPMMouse [4], MR data were forward projected into 3D PET sinograms and thresholded to create an AC map, defined as a single region of tissue with uniform attenuation co-efficient of 0.095cm–1. SUV values were calculated from summed PET images (last 20 minutes) and compared on a voxel by voxel basis between images without AC, with transmission source AC, and with MRAC.

A 22.6 ± 0.9% (mean ± SD) improvement in mouse myocardium SUV values (shown in Figures 1 and 2) was seen by applying transmission AC and a 18.5 ± 0.9% improvement using MRAC, compared to not applying AC. The global attenuation correction over the whole mouse body was 20.7 ± 0.7% using transmission AC and 16.5 ± 1.3% using MRAC. Differences of up to 40% (mean: 30.1 ± 4.4%, range: 27-40%) were seen adjacent to the RF coil (see Figure 3).

Figure 1.

Figure 1

Single subject line profile for each AC method.

Figure 2.

Figure 2

SUV maps (transverse view) for single subject. (A) No AC applied, (B) Transmission AC, (C) MRAC.

Figure 3.

Figure 3

PET emission (blue) and transmission (grey) fused images showing mouse, coil and bed attenuation, (A) Transverse, (B) Coronal, (C) Sagittal.

Conclusion

A simple, one region MRAC approach provided acceptable AC compared to transmission scanning for myocardial imaging in mice.

References

  • 1.Wagenknecht G, et al. MRI for attenuation correction in PET: methods and challenges. Magn Reson Mater Phy. 2013;26:99–113. doi: 10.1007/s10334-012-0353-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Buonincontri G, et al. PET/MRI assessment of the infarcted mouse heart. NIMA A. 2014;734:152–5. doi: 10.1016/j.nima.2013.08.066. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Lucas AJ, et al. Development of a combined microPET®-MR system. IEEE Nucl Sci Symp Record. 2006;4:2345–8. [Google Scholar]
  • 4.Sawiak SJ, et al. MRI reveals brain asymmetry following 6-OHDA lesions in the mouse brain. Proc. ISMRM. 2009;17:1077. [Google Scholar]

Articles from EJNMMI Physics are provided here courtesy of Springer-Verlag

RESOURCES