Abstract
The cardiovascular implantable electronic device (CIED) infection rate is rising disproportionately to the rate of device implantation. Identification of microorganisms that cause CIED infections is not always achieved using present laboratory techniques. We conducted a prospective study to determine whether device vortexing-sonication followed by culture of the resulting sonicate fluid would enhance microbial detection compared with traditional swab or pocket tissue cultures. Forty-two subjects with noninfected and 35 with infected CIEDs were prospectively enrolled over 12 months. One swab each from the device pocket and device surface, pocket tissue, and the CIED were collected from each patient. Swabs and tissues were cultured using routine methods. The CIED was processed in Ringer’s solution using vortexing-sonication and the resultant fluid semiquantitatively cultured. Tissue and swab growth was considered significant when colonies grew on ≥2 quadrants of the culture plate and device was considered significant when ≥20 colonies were isolated from 10 ml of sonicate fluid. In noninfected group, 5% of sonicate fluids yielded significant bacterial growth, compared with 5% of tissue cultures (p = 1.00) and 2% of both pocket and device swab cultures (p = 0.317 each). In infected group, significant bacterial growth was observed in 54% of sonicate fluids, significantly greater than the sensitivities of pocket swab (20%, p = 0.001), device swab (9%, p <0.001), or tissue (9%, p <0.001) culture. In conclusion, vortexing-sonication of CIEDs with semiquantitative culture of the resultant sonicate fluid results in a significant increase in the sensitivity of culture results, compared with swab or tissue cultures.
The cardiovascular implantable electronic device (CIED) implantation rate has markedly increased, largely because of expanding indications for device implantation.1 This has been accompanied by an increasing rate of device infections.2–4 Infection is a serious complication of CIED implantation that necessitates device removal through percutaneous or surgical approach that is associated with significant morbidity and mortality and considerable financial burden for the patient and the health care system.5–7 Current guidelines recommend generator pocket tissue Gram stain and culture and lead-tip culture for identification of the causative microorganism(s).8 However, Gram stain has been shown to have limited utility in the diagnosis of device-related infections,9 and cultures may be negative for a variety of reasons, including concentration of organisms in biofilms on the device surface and consequently not in the surrounding tissue and the presence of so-called “small colony variants” that may be more difficult to isolate by routine cultures.10–12 Vortexing-sonication of implants followed by culture of the resultant sonicate fluid is more sensitive and specific compared with conventional periprosthetic tissue culture for the diagnosis of prosthetic joint infection.13,14 Consequently, this technique has been adopted in clinical microbiology laboratories across the world for the diagnosis of prosthetic joint infection. On the basis of findings from these investigations, we hypothesized that vortexing-sonication, followed by culture of the resulting sonicate fluid, will enhance microbial detection compared with traditional swab or pocket tissue culture for the diagnosis of CIED infection.
Methods
The study was conducted at Mayo Clinic, Rochester, from November 2011 to November 2012. Potential subjects were identified using the institutional operating room census and by direct communication with the electrophysiology and cardiac surgical services. Written informed consent was obtained from all study subjects. For patients who consented for participation and underwent explantation of a CIED, the following samples were collected: (1) CIED; (2) device surface swab; (3) pocket tissue swab; and (4) pocket tissue (~1 cm3 in size). The Mayo Clinic Institutional Review Board approved the study protocol.
CIED infection was defined as the presence of inflammatory changes (erythema, warmth, fluctuation, or purulent discharge) at the generator pocket site, persistently positive blood cultures in the absence of any other defined focus of infection or pathologic analysis of pocket tissue demonstrating acute inflammation. CIED-related endocarditis was defined on the basis of modified Duke criteria.15,16 Device generator or leads that eroded through the pocket were also classified as infected. We concurrently enrolled subjects with no clinical or pathologic signs of infection, typically patients who underwent replacement of a generator for “end of battery life,” as noninfected controls.
The swabs and tissues were subjected to routine microbiologic culture involving inoculation onto aerobic blood and chocolate agars, and in cases of tissues, onto anaerobic blood agar and into thioglycollate broth (BD Diagnostic Systems, Sparks, Maryland) as well. Aerobic and anaerobic sheep blood agar plates (BD Diagnostic Systems) were incubated at 35°C to 37°C in 5% to 7% CO2 aerobically and anaerobically for 4 and 7 to 14 days, respectively. Cloudy thioglycollate broth was subcultured.
The cardiac device was sent to the laboratory in a sterile 1-liter, straight-sided, wide-mouthed polypropylene (Nalgene) jar. Four hundred milliliters of Ringer’s solution was added to the container. The container was vortexed for 30 seconds using a Vortex Genie (Scientific Industries Inc., Bohemia, New York) and then subjected to sonication (frequency 40 ± 2 kHz, power density 0.22 ± 0.04 W/cm2) in 5510 model ultrasound bath (Branson Ultrasonics Corp., Danbury, Connecticut) for 5 minutes, followed by vortexing for an additional 30 seconds (Figure 1). Fifty milliliters of sonicate fluid was placed into a conical tube and centrifuged at ×3,150g for 5 minutes. The supernatant was aspirated leaving 0.5 ml remaining in the tube, 0.1 ml which was plated onto an aerobic and another 0.1 ml onto an anaerobic sheep blood agar plate which were incubated at 35°C to 37°C in 5% to 7% CO2 aerobically for 4 days and anaerobically for 14 days. Microorganisms were enumerated and classified using routine microbiologic techniques (Figure 2).
Figure 1.

Apparatus for sonication. (A) Autoclaved container used for transportation and sonication of cardiac device; (B) vortexing apparatus; and (C) sonicator.
Figure 2.

Colony-forming units visualized on standard blood agar culture plate following (A) device (sonicate fluid) culture; (B) swab culture; and (C) tissue culture from 1 study subject.
Host demographics, comorbidities, clinical signs and symptoms, laboratory parameters, and microbiologic results were collected using a data collection instrument and entered into a REDCap version 4.13.17 (Vanderbilt University, Nashville, Tennessee) database.17
Sensitivity and specificity were calculated for device, swab, and tissue cultures, with infection status (gold standard) defined by clinical and pathologic criteria described previously. Device sonicate fluid cultures were classified as a positive versus negative result on the basis of a cut-off point of <20 colony-forming units/10 ml (CFU/10 ml) derived from a previous validation study of prosthetic joint infections at our institution.18 To test whether device sonicate fluid culture had greater sensitivity or specificity than each of the swab or tissue cultures for the diagnosis of CIED infection, McNemar test of paired proportions was used. A p value <0.05 was considered statistically significant. All analyses were performed using the SAS statistical software package version 9.3 (SAS Institute Inc., Cary, North Carolina).
Results
Ninety-three subjects were enrolled, of which 16 were excluded because of gross contamination of specimens during device extraction, specimen transportation delay of more than 12 hours, or failure to collect the requisite specimens. Of the 77 subjects included in the final analysis, 35 met the criteria for CIED infection and 42 were considered uninfected. There were no statistically significant differences in host, demographic, or clinical characteristics between these 2 groups (Table 1) although infected subjects had marginally higher rates of diabetes (34 vs 17%, p = 0.074), chronic kidney disease (stage 3 or greater/glomerular filtration rate <60 ml/min/1.73 m2; 29 vs 12%, p = 0.066), and previous history of CIED infection (14 vs 2%, p = 0.052).
Table 1.
Demographic and clinical characteristics of infected and non-infected cardiovascular implantable electronic device groups
| Demographic Characteristic |
Infected*
(n=35) |
Non-infected*
(n=42) |
p-value |
|---|---|---|---|
| Age (in years) | 63.7±17.3 | 67.4±15.4 | 0.319 |
| BMI (kg/m2) | 31.1±7.0 | 29.2±7.2 | 0.270 |
| Men | 24 (69%) | 25 (60%) | 0.411 |
| White | 33 (94%) | 41 (98%) | 0.588* |
| Hypertension | 21 (60%) | 24 (57%) | 0.800 |
| Hyperlipidemia | 23 (66%) | 30 (71%) | 0.590 |
| Diabetes Mellitus | 12 (34%) | 7 (17%) | 0.074 |
| Chronic Kidney Disease (GFR <60) |
10 (29%) | 5 (12%) | 0.066 |
| Immunosuppression | 4 (11%) | 1 (2%) | 0.109 |
| Coronary Artery Disease | 20 (57%) | 18 (43%) | 0.212 |
| Cerebrovascular Disease | 5 (14%) | 4 (10%) | 0.517 |
| Peripheral Vascular Disease | 2 (6%) | 1 (2%) | 0.588* |
| Chronic Obstructive Pulmonary Disease |
4 (11%) | 4 (10%) | 0.785 |
| Prosthetic Valve | 8 (23%) | 6 (14%) | 0.332 |
| Prior CIED Infection | 5 (14%) | 1 (2%) | 0.052 |
| Atrial Fibrillation | 9 (26%) | 17 (40%) | 0.173 |
| Heart Failure | 21 (60%) | 20 (48%) | 0.278 |
Continuous data summarized with mean±SD, and compared between groups via two-sample t test; categorical variables reported with counts and percentages, and compared using the Chi-square test unless specified otherwise.
BMI = Body Mass Index; CIED = Cardiovascular Implantable Electronic device; GFR = Glomerular Filtration Rate (in mL/min); HDL = High Density Lipoprotein (in mg/dL); LDL = Low Density Lipoprotein (in mg/dL), HDL = High Density Lipoprotein (in mg/dL), GFR = Glomerular Filtration Rate (in mL/min).
Definitions: Hypertension was defined as blood pressure greater than 140/90 mm Hg, Hyperlipidemia was defined as Low Density Lipoprotein >160 mg/dl or High Density Lipoprotein <40 mg/dl. Coronary artery disease was defined as previously documented evidence of angina pectoris, myocardial infarction, history of angioplasty for coronary artery obstruction or history of coronary artery bypass grafting.
p-value from Fisher’s exact test due to sparse data.
In the infected group, there were 12 cases of pocket cellulitis, 4 cases of device or lead erosion (without clinical signs of cellulitis around the pocket or positive blood cultures), 14 cases of bloodstream infection, presumably involving the intravascular portion of the lead(s), and 5 cases of CIED-related bloodstream infection simultaneously involving native or prosthetic heart valve(s) (Figure 3). Twenty-nine (83%) subjects (4 with pocket infection and 2 with device erosion) had received antibiotic therapy before cardiac device extraction and specimen collection.
Figure 3.

Clinical presentation of patients with CIED infection.
In the noninfected group, a number of culture plates yielded microbial growth, typically with low numbers of colonies (i.e., growth in the first quadrant/1 + growth/growth in broth only) of low-virulence organisms such as Bacillus sp., Corynebacterium sp., or Propionibacterium sp., which were likely representative of contamination. Moreover, comparing all specimens from each patient, none of the subjects had consistent growth of the same organism across all 4 specimens analyzed.
Because causative organisms (most commonly Staphylococcus aureus or Coagulase-negative Staphylococcus) were known in cases of lead or valve endocarditis on the basis of persistently positive blood cultures, we analyzed the level of discordance between blood culture results and cultures of the specimens removed at the time of device removal. Of 19 cases of lead or valve endocarditis, sonicate fluid culture results were discordant with blood culture results in 3 subjects. Tissue and/or swab culture results were also discordant with the results of the blood cultures in 3 subjects. Results of both sonicate fluid and tissue/swab culture results were different from those of blood culture in 2 subjects, whereas 1 patient had discordant results exclusively with the sonicate fluid and another exclusively with the tissue/swab cultures.
Comparative results of sensitivity and specificity across different culture types are summarized in Table 2. Although in noninfected subjects, 40% of sonicate fluids yielded any microbial growth on culture, compared with 33% of tissues (p = 0.366), 7% of pocket swabs (p <0.001), and 5% of device swabs (p <0.001), when disregarding minimal growth likely to represent contamination (i.e., <20 CFU/10 ml sonicate fluid and 1 + growth or growth in broth only for swab and tissue cultures, hereafter considered nonsignificant), 5% of sonicate fluids yielded a positive microbiologic result, compared with 5% of tissue cultures (p = 1.00) and 2% of both pocket and device swab (p = 0.317 each) cultures. Also based on significant positive growth, the overall sensitivity of sonicate fluid culture (54%) was greater than that of pocket swab (20%, p = 0.001), device swab (9%, p <0.001), and tissue (9%, p <0.001) culture.
Table 2.
Sensitivity and specificity of culture for cardiovascular implantable electronic device infection based on the specimen-types studied
| Test | Proportion | Sensitivity Rate, % (95% CI) |
P-value* | Proportion | Specificity Rate, % (95% CI) |
P-value* |
|---|---|---|---|---|---|---|
| Significant Growth | ||||||
| Device Swab | 3/35 | 9 (2-23%)† | <.001 | 41/42 | 98 (87-100%) | 0.317 |
| Pocket Swab | 7/35 | 20 (8-37%)† | 0.001 | 41/42 | 98 (87-100%) | 0.317 |
| Tissue | 3/35 | 9 (2-23%)† | <.001 | 40/42 | 95 (84-99%) | 1.000 |
| Sonicate Fluid | 19/35 | 54 (37-71%) | 40/42 | 95 (84-99%) | ||
| Any Growth | ||||||
| Device Swab | 10/35 | 29 (15-46%)† | <.001 | 40/42 | 95 (84-99%)† | <.001 |
| Pocket Swab | 15/35 | 43 (26-61%)† | 0.008 | 39/42 | 93 (81-99%)† | <.001 |
| Tissue | 16/35 | 46 (29-63%)† | 0.004 | 28/42 | 67 (50-80%) | 0.366 |
| Sonicate Fluid | 26/35 | 74 (57-88%) | 25/42 | 60 (43-74%) |
p-value testing for a difference in performance rate relative to sonicate fluid culture via McNemar’s test.
Value is significantly different from that by sonicate fluid culture (p <0.05 from McNemar’s test).
Subgroup analyses revealed evidence that the device sonicate fluid culture was generally more sensitive than the other tests in the infected subset with device erosion/pocket cellulitis and lead or valve endocarditis (Table 3). For the 16 cases of device erosion/pocket cellulitis, sensitivity of either swab or tissue culture (based on significant positive growth) was only 38% (95% confidence interval [CI], 15% to 65%) compared with 69% (95% CI 41% to 89%) for device sonicate fluid culture (p = 0.059). For the 19 cases of lead or valve endocarditis, sensitivity of either swab or tissue culture (based on significant positive growth) was only 11% (95% CI 1% to 33%) compared to 42% (20% to 67%) for device culture (p = 0.014).
Table 3.
Sensitivity of culture based on the specimen types studied among patients with cardiovascular implantable electronic device infection; stratified based on subgroups with a) device erosions and pocket infections and b) lead or valve associated endocarditis
| Test Subgroup |
Sensitivity | |||||
|---|---|---|---|---|---|---|
|
| ||||||
| Proportion | Rate, % (95% CI) | P-value* | Proportion | Rate, % (95% CI) | P-value* | |
| Device Erosions/Pocket Infections (n=16) | Significant Growth | Any Growth | ||||
| Device Swab | 2/16 | 13 (2-38%)† | 0.003 | 6/16 | 38 (15-65%)† | 0.035 |
| Pocket Swab | 5/16 | 31 (11-59%)† | 0.034 | 10/16 | 63 (35-85%) | 0.257 |
| Tissue | 2/16 | 13 (2-38%)† | 0.003 | 10/16 | 63 (35-85%) | 0.180 |
| Sonicate Fluid | 11/16 | 69 (41-89%) | 13/16 | 81 (54-96%) | ||
| Lead or Valve associated Endocarditis Infections (n=19) | Significant Growth | Any Growth | ||||
| Device Swab | 1/19 | 5 (0-26%)† | 0.008 | 4/19 | 21 (6-46%)† | 0.003 |
| Pocket Swab | 2/19 | 11 (1-33%)† | 0.014 | 5/19 | 26 (9-51%)† | 0.011 |
| Tissue | 1/19 | 5 (0-26%)† | 0.008 | 6/19 | 32 (13-57%)† | 0.008 |
| Sonicate Fluid | 8/19 | 42 (20-67%) | 13/19 | 68 (43-87%) | ||
p-value testing for a difference in performance rate relative to sonication via McNemar’s test.
Value is significantly different from that by Sonication test (p <.05 from McNemar’s test).
Discussion
The present investigation is the largest and provides the most comprehensive description of a vortexing-sonication technique in clinical practice to enhance microbial detection in cases of CIED infection. Recent guidelines issued by the American Heart Association8 recommend swab and tissue Gram stain and cultures to identify pathogens. However, several factors impact Gram stain and culture yield.
Vortexing-sonication of prosthetic orthopedic devices has been shown to be more sensitive than periprosthetic tissue culture for the diagnosis of prosthetic joint and implant infections.13,14,19,20 Our study findings suggest that sonication is superior to device or pocket swabs and pocket tissue cultures, techniques that are routinely used for establishing a microbiologic diagnosis, irrespective of the type of CIED infection.
Although 5% of sonicate fluid cultures in our study revealed significant bacterial growth from noninfected devices, there was some growth from 40% of noninfected devices overall. The significance of this finding, which appears to be a repetitive theme in similar studies,21–25 is unclear.26 In this investigation involving CIEDs, as with our previous prosthetic joint infection studies, the low virulence of isolated microorganisms and the small quantity of growth observed in most noninfected cases were considered consistent with contamination. Therefore, after adjudicating these cultures with small quantity of growth as possible contaminants, our estimated specificity of device sonicate fluid cultures was 95%, similar to that of swab (98%) and tissue (95%) cultures.
False-positive microbiologic results are also a theoretical concern in clinically infected devices. Contamination of CIEDs can occur at the time of collection, during transport, specimen processing, or during incubation. To address this possibility, we compared positive results from various specimens (swabs, tissues, and sonicate fluid) with those of blood cultures in cases of CIED lead or valve endocarditis. In the 19 cases with definite lead or valve endocarditis, results of device sonication cultures differed from blood cultures in only 3 cases. This was similar to a discrepancy between blood cultures and swab or tissue cultures. Therefore, the likelihood of a positive sonicate fluid culture representing a false-positive result does not seem to be greater than that of false-positive results from traditional swab or tissue cultures.
Despite the use of sonication technique, some cultures from clinically infected devices also resulted in bacterial growth in small numbers. Although we classified these results as representing contamination based on our predefined cut-off values, it is possible that in some cases, these results represent true pathogens as 83% of subjects with CIED infection had received antibiotics before device explantation and collection of specimens for cultures.
Usefulness of vortexing-sonication technique for the microbiologic diagnosis of CIED infections has been evaluated in earlier investigations (Table 4).22,24,25 However, these studies were limited by relatively small numbers of patients with infected devices, and there were no attempts to explain or eliminate the possibility of false-positive cultures results observed with either of the culture techniques. Because all positive culture results in these studies were considered to represent true infection, they may have overestimated the sensitivity of both sonication and traditional culture techniques. In contrast, we have tried to make a clear distinction between results that are likely to represent pathogens versus those that represent contamination.
Table 4.
Literature review of the microbiological yield of sonicate fluid and swab/tissue cultures
| Number of subjects with CIED infection | Sensitivity of sonicate fluid culture* | Sensitivity of swab/tissue culture* | |
|---|---|---|---|
| Present study* | 35 | 54% | 9–20% |
| Oliva et al.24 | 20 | 67% | 50% |
| Mason et al.25 | 16 | 94% | 75–81% |
| Rohacek et al.14 | 6 | 100% | 67% |
Values represent sensitivity of cultures for detecting significant growth.
Overall, our study findings suggest that vortexing-sonication of CIEDs with semiquantitative culture of the resultant sonicate fluid results in a significant increase in the sensitivity of culture results, compared with swab or tissue cultures, without compromising the specificity of positive results when only significant growth is deemed true positive. Results of this study should influence the future guidelines regarding diagnostic evaluation of CIED infection.
There are several limitations of present investigation, the most important being the small number of subjects. However, considering the generally low rates of CIED infections, this is the largest cohort of patients with infected CIED systems in which sonication technique has been used for microbial detection. The implication of a positive sonicate fluid culture in couple of subjects in the noninfected group remains unclear, and we are prospectively following these patients to see if they develop a clinically overt CIED infection. Another limitation in our study cohort was initiation of antimicrobial therapy before the collection of specimens for cultures. However, this should have affected the results of all culture techniques studied, regardless of specimen type. Moreover, this is a true representation of current practice of medicine where initiation of treatment cannot always be withheld before collection of specimens because of unstable clinical condition of subjects with device infection, especially those with systemic signs and symptoms or evidence of bloodstream infection, or because of infections at other sites. Finally, our study population consisted largely of Caucasian patients, reflective of the general demographic of the population served at our institution. However, considering that the focus of our investigation was isolation of microorganisms rather than studying host response, we do not anticipate that results will vary on the basis of inclusion of more ethnically diverse patient populations.
Acknowledgment
The authors thank Lisa F. Fanning, RCES, Peggy C. Kohner, BS, MT(ASCP), and Lisa M. Nyre, MA, MT(ASCP) for their administrative support during the conduct of this study and the technologists and laboratory assistants of the Mayo Clinic Division of Clinical Microbiology for processing and culturing the specimens.
This study was supported by the Ann and Leo Markin Endowed Professorship Fund, Small Grants Program of the Division of Infectious Diseases, and a Career Development Award to Dr. Sohail from Department of Medicine, Mayo Foundation for Medical Education and Research. The study was supported by Grant Number UL1 TR000135 from the National Center for Advancing Translational Sciences (NCATS).
Paul A. Friedman receives honoraria from Medtronic, Guidant, and Astra Zeneca; receives research sponsorship from Medtronic, Astra Zeneca via Beth Israel, Guidant, St. Jude, and Bard; and has intellectual property rights in Bard EP, Hewlett Packard, and Medical Positioning, Inc. David L. Hayes receives honoraria from Medtronic, Boston Scientific, St. Jude Medical, Sorin Medical, and Biotronik; is in the advisory board of St. Jude Medical and Medtronic; and is in the steering committee of Medtronic and St. Jude Medical. Larry M. Baddour receives royalty payments (authorship) from UpToDate Inc. (<$20,000) and Editor-in-Chief payments from Massachusetts Medical Society (Journal Watch Infectious Diseases; <$20,000). M. Rizwan Sohail received funding from TyRx Inc. for previous research unrelated to this study (Bloom et al. PACE 2011 34, 133–42), administered according to a sponsored research agreement that prospectively defined the scope of the research effort and corresponding budget, and receives honoraria from Medtronic and Spectranetics. Robin Patel has a US patent pending for a method and an apparatus for sonication (has forgone right to receive royalties in the event that the patent is licensed) and patents for an antibiofilm substance and Bordetella pertussis/parapertussis PCR; receives research support (grants, reagents, and equipment loan) from Pfizer, Tornier, Pradama, Astellas, Pocared, 3M, nanoMR, Bruker, Nanosphere, BioFire bioMérieux, Curetis is a consultant in Thermo Fisher Scientific, is supported by the National Institute of Arthritis and Musculoskeletal and Skin Diseases of the National Institutes of Health (Award Number R01 AR056647) and the National Institute of Allergy and Infectious Diseases (Award Number R01 AI091594). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.
Footnotes
Disclosures
The remaining authors have no disclosures.
References
- 1.Epstein AE, DiMarco JP, Ellenbogen KA, Estes NA, 3rd, Freedman RA, Gettes LS, Gillinov AM, Gregoratos G, Hammill SC, Hayes DL, Hlatky MA, Newby LK, Page RL, Schoenfeld MH, Silka MJ, Stevenson LW, Sweeney MO, Smith SC, Jr, Jacobs AK, Adams CD, Anderson JL, Buller CE, Creager MA, Ettinger SM, Faxon DP, Halperin JL, Hiratzka LF, Hunt SA, Krumholz HM, Kushner FG, Lytle BW, Nishimura RA, Ornato JP, Riegel B, Tarkington LG, Yancy CW. ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology/American heart Association Task Force on practice guidelines (Writing committee to Revise the ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation. 2008;117:e350–e408. doi: 10.1161/CIRCUALTIONAHA.108.189742. [DOI] [PubMed] [Google Scholar]
- 2.Voigt A, Shalaby A, Saba S. Continued rise in rates of cardiovascular implantable electronic device infections in the United States: temporal trends and causative insights. Pacing Clin Electrophysiol. 2010;33:414–419. doi: 10.1111/j.1540-8159.2009.02569.x. [DOI] [PubMed] [Google Scholar]
- 3.Cabell CH, Heidenreich PA, Chu VH, Moore CM, Stryjewski ME, Corey GR, Fowler VG., Jr Increasing rates of cardiac device infections among Medicare beneficiaries: 1990-1999. Am Heart J. 2004;147:582–586. doi: 10.1016/j.ahj.2003.06.005. [DOI] [PubMed] [Google Scholar]
- 4.Voigt A, Shalaby A, Saba S. Rising rates of cardiac rhythm management device infections in the United States: 1996 through 2003. J Am Coll Cardiol. 2006;48:590–591. doi: 10.1016/j.jacc.2006.05.016. [DOI] [PubMed] [Google Scholar]
- 5.Sohail MR, Henrikson CA, Braid-Forbes MJ, Forbes KF, Lerner DJ. Mortality and cost associated with cardiovascular implantable electronic device infections. Arch Intern Med. 2011;171:1821–1828. doi: 10.1001/archinternmed.2011.441. [DOI] [PubMed] [Google Scholar]
- 6.Cacoub P, Leprince P, Nataf P, Hausfater P, Dorent R, Wechsler B, Bors V, Pavie A, Piette JC, Gandjbakhch I. Pacemaker infective endocarditis. Am J Cardiol. 1998;82:480–484. doi: 10.1016/s0002-9149(98)00365-8. [DOI] [PubMed] [Google Scholar]
- 7.Darouiche RO. Treatment of infections associated with surgical implants. N Engl J Med. 2004;350:1422–1429. doi: 10.1056/NEJMra035415. [DOI] [PubMed] [Google Scholar]
- 8.Baddour LM, Epstein AE, Erickson CC, Knight BP, Levison ME, Lockhart PB, Masoudi FA, Okum EJ, Wilson WR, Beerman LB, Bolger AF, Estes NA, 3rd, Gewitz M, Newburger JW, Schron EB, Taubert KA. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation. 2010;121:458–477. doi: 10.1161/CIRCULATIONAHA.109.192665. [DOI] [PubMed] [Google Scholar]
- 9.Atkins BL, Athanasou N, Deeks JJ, Crook DW, Simpson H, Peto TE, McLardy-Smith P, Berendt AR, The OSIRIS Collaborative Study Group Prospective evaluation of criteria for microbiological diagnosis of prosthetic-joint infection at revision arthroplasty. J Clin Microbiol. 1998;36:2932–2939. doi: 10.1128/jcm.36.10.2932-2939.1998. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Baddour LM, Barker LP, Christensen GD, Parisi JT, Simpson WA. Phenotypic variation of Staphylococcus epidermidis in infection of transvenous endocardial pacemaker electrodes. J Clin Microbiol. 1990;28:676–679. doi: 10.1128/jcm.28.4.676-679.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Seifert H, Wisplinghoff H, Schnabel P, von Eiff C. Small colony variants of Staphylococcus aureus and pacemaker-related infection. Emerg Infect Dis. 2003;9:1316–1318. doi: 10.3201/eid0910.030200. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.von Eiff C. Staphylococcus aureus small colony variants: a challenge to microbiologists and clinicians. Int J Antimicrob Agents. 2008;31:507–510. doi: 10.1016/j.ijantimicag.2007.10.026. [DOI] [PubMed] [Google Scholar]
- 13.Piper KE, Jacobson MJ, Cofield RH, Sperling JW, Sanchez-Sotelo J, Osmon DR, McDowell A, Patrick S, Steckelberg JM, Mandrekar JN, Fernandez Sampedro M, Patel R. Microbiologic diagnosis of prosthetic shoulder infection by use of implant sonication. J Clin Microbiol. 2009;47:1878–1884. doi: 10.1128/JCM.01686-08. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Trampuz A, Piper KE, Jacobson MJ, Hanssen AD, Unni KK, Osmon DR, Mandrekar JN, Cockerill FR, Steckelberg JM, Greenleaf JF, Patel R. Sonication of removed hip and knee prostheses for diagnosis of infection. N Engl J Med. 2007;357:654–663. doi: 10.1056/NEJMoa061588. [DOI] [PubMed] [Google Scholar]
- 15.Greenspon AJ, Prutkin JM, Sohail MR, Vikram HR, Baddour LM, Danik SB, Peacock J, Falces C, Miro JM, Blank E, Naber C, Carrillo RG, Tseng CH, Uslan DZ. Timing of the most recent device procedure influences the clinical outcome of lead-associated endocarditis results of the MEDIC (Multicenter Electrophysiologic Device Infection Cohort) J Am Coll Cardiol. 2012;59:681–687. doi: 10.1016/j.jacc.2011.11.011. [DOI] [PubMed] [Google Scholar]
- 16.Li JS, Sexton DJ, Mick N, Nettles R, Fowler VG, Jr, Ryan T, Bashore T, Corey GR. Proposed modifications to the Duke criteria for the diagnosis of infective endocarditis. Clin Infect Dis. 2000;30:633–638. doi: 10.1086/313753. [DOI] [PubMed] [Google Scholar]
- 17.Harris PA, Taylor R, Thielke R, Payne J, Gonzalez N, Conde JG. Research electronic data capture (REDCap)ea metadata-driven methodology and workflow process for providing translational research informatics support. J Biomed Inform. 2009;42:377–381. doi: 10.1016/j.jbi.2008.08.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Piper KE, Jacobson MJ, Steckelberg JM, Patel R. Microbiological diagnosis of hip and knee prosthetic joint infection using explanted prostheses sonication followed by concentration of sonicate fluid 48th Interscience Conference on Antimicrobial Agents and Chemotherapy; Wash DC. 2008. [Google Scholar]
- 19.Vergidis P, Greenwood-Quaintance KE, Sanchez-Sotelo J, Morrey BF, Steinmann SP, Karau MJ, Osmon DR, Mandrekar JN, Steckelberg JM, Patel R. Implant sonication for the diagnosis of prosthetic elbow infection. J Shoulder Elbow Surg. 2011;20:1275–1281. doi: 10.1016/j.jse.2011.06.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Sampedro MF, Huddleston PM, Piper KE, Karau MJ, Dekutoski MB, Yaszemski MJ, Currier BL, Mandrekar JN, Osmon DR, McDowell A, Patrick S, Steckelberg JM, Patel R. A biofilm approach to detect bacteria on removed spinal implants. Spine (Phila Pa 1976) 2010;35:1218–1224. doi: 10.1097/BRS.0b013e3181c3b2f3. [DOI] [PubMed] [Google Scholar]
- 21.Pichlmaier M, Marwitz V, Kuhn C, Niehaus M, Klein G, Bara C, Haverich A, Abraham WR. High prevalence of asymptomatic bacterial colonization of rhythm management devices. Europace. 2008;10:1067–1072. doi: 10.1093/europace/eun191. [DOI] [PubMed] [Google Scholar]
- 22.Rohacek M, Weisser M, Kobza R, Schoenenberger AW, Pfyffer GE, Frei R, Erne P, Trampuz A. Bacterial colonization and infection of electrophysiological cardiac devices detected with sonication and swab culture. Circulation. 2010;121:1691–1697. doi: 10.1161/CIRCULATIONAHA.109.906461. [DOI] [PubMed] [Google Scholar]
- 23.Kleemann T, Becker T, Strauss M, Dyck N, Weisse U, Saggau W, Burkhardt U, Seidl K. Prevalence of bacterial colonization of generator pockets in implantable cardioverter defibrillator patients without signs of infection undergoing generator replacement or lead revision. Europace. 2010;12:58–63. doi: 10.1093/europace/eup334. [DOI] [PubMed] [Google Scholar]
- 24.Oliva A, Nguyen BL, Mascellino MT, D’Abramo A, Iannetta M, Ciccaglioni A, Vullo V, Mastroianni CM. Sonication of explanted cardiac implants improves microbial detection in cardiac device infections. J Clin Microbiol. 2013;51:496–502. doi: 10.1128/JCM.02230-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Mason PK, Dimarco JP, Ferguson JD, Mahapatra S, Mangrum JM, Bilchick KC, Moorman JR, Lake DE, Bergin JD. Sonication of explanted cardiac rhythm management devices for the diagnosis of pocket infections and asymptomatic bacterial colonization. Pacing Clin Electrophysiol. 2011;34:143–149. doi: 10.1111/j.1540-8159.2010.02820.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Baddour LM. Cardiac device infection—or not. Circulation. 2010;121:1686–1687. doi: 10.1161/CIR.0b013e3181de0334. [DOI] [PubMed] [Google Scholar]
