Abstract
For over 30 years scientists in the NIST radioactivity group have been using their pressurized ionization chamber “A” (PIC “A”) to make measurements of radioactivity and radioactive half-lives. We now have evidence that some of those reported measurements were incorrect due to slippage of the source positioning ring over time. The temporal change in the holder caused an error in the source-height within the chamber, which was thought to be invariant. This unaccounted-for height change caused a change in the detector response and thus a relative error in measured activity on the order of 10−5 to 10−3 per year, depending on the radionuclide. The drifting detector response affected calibration factors and half-life determinations. After discovering the problem, we carried out historic research and new sensitivity tests. As a result, we have created a quantitative model of the effect and have used that model to estimate corrections to some of the past measurement results from PIC “A”. In this paper we report the details and results of that model. Meanwhile, we have fixed the positioning ring and are recalibrating the detector using primary measurement methods and enhanced quality control measures.
Keywords: Half-life, ionization chamber
1. Introduction
Since the 1960’s the NIST pressurized ionization chamber “A” (PIC “A”) has been used for two main purposes. First, it has been used to maintain radioactivity standards through calibration factors for over 70 γ-ray emitting radionuclides. Most of those calibration factors were determined in the late 1970’s and some are still used for radioactivity determinations. Second, PIC “A” has been used to measure half-lives for over 50 γ-ray emitting radionuclides.
We now have evidence that some of those reported activity and half-life values were incorrect due to slippage of the sample positioning ring over time The temporal change in the holder caused an error in the source-height within the chamber, which was thought to be invariant. This unaccounted-for height change caused a change in the detector response and thus a relative error in measured activity on the order of 10−5 to 10−3 per year, depending on the radionuclide. The drifting detector response affected calibration factors and half-life determinations. After discovering the problem, we carried out historic research and new sensitivity tests. As a result, we have created a quantitative model of the effect and have used that model to estimate corrections to some of the past measurement results from PIC “A”. Meanwhile, we have fixed the positioning ring and are recalibrating the detector using primary measurement methods and enhanced quality control measures.
The route from investigation to corrections went as follows. First, historic artifacts and drawings of the instrument were used to estimate the height of the positioning ring at various dates between 1968 and 2010. Second, the detector response was measured as a function of height in the chamber for various radionuclides. Third, those data were used to model the response change for any γ-ray-emitting radionuclide over any time-span. Various aspects of the model were then validated by measurements of other radionuclides at various heights, analysis of historic NIST equivalent-activity values in the International Reference System (SIR), analysis of NIST half-life fit residuals and comparison of corrected calibration factors with new calibration factors determined by primary standardization methods.
2. The Measurement System
The NIST PIC “A” is a type “TPA Mk. 2” re-entrant chamber with a “1 inch”(nominal 25 cm) inner diameter, a “1/32 inch” (nominal 0.8 mm) wall thickness containing “20 atm” (20.×105 Pa) of argon [1]. The detector was obtained from the Chalk River Laboratories of the National Research Council of Canada and was a forerunner to the Centronic1 IG-11.
The sources typically consist of aqueous solutions or gases contained within 5-mL glass ampoules that are placed inside a cylindrical sample-holder made of polymethyl methacrylate (acrylic). A brass positioning ring is fitted onto the holder a few cm from the top. The ring position on the holder is called h, as is shown in Fig. 1. When the holder is inserted into the chamber, this positioning ring rests on the shoulder of the chamber, and thereby sets the height of the sample within the chamber. The ring was thought to have been fixed by a set-screw in 1968. In January 2010, it was discovered that the set-screw did not drill into the holder, but only pushed against it, and was held in position by compression. The positioning ring has apparently been slipping an average of about 1 mm per year. Thus, the position of the sample in the chamber has moved deeper in the detector as a function of time. Since the detector response is sensitive to the height of the source, this slippage has caused a temporal change in the detector response for each radionuclide.
Fig. 1.
Sample holder, rotated 90 degrees counter-clockwise from its usual orientation. An ampoule sits in the bottom cup. The height, h, is shown.
The standard procedures for using PIC “A” are described by Calhoun [2]. The methodology is similar to that used by Rytz [3]. The measurand is the ratio, R, of the detector response for a given radioactive source to that of a NIST radium-226 reference source. The calibration factor, K, (we call it the “K-value”), for a source of known activity, A, is defined as,
(1) |
The use of the ratio to radium mitigates uncertainties caused by variations in the electrometer response over time. However, since the height sensitivity of the detector is dependent on the energy of the emitted γ-rays, the change in height will not necessarily affect the source of interest and the radium reference source by the same amount. That is, use of the ratio will not eliminate the effect of height-change on measured activity. Therefore, to quantify the effect of the positioning-ring slippage, in addition to gleaning h as a function of time, the variation of R as a function of h for each radionuclide needed to be determined.
3. The Model
3.1 Height Variation
The height of the source positioning ring over time, h(t), was determined by studying historical drawings of the instrument from 1981, 1986, 1992, 1993 and 2010. No drawing of the sample holder from 1968, when the positioning-ring was set, has been found. One assumption is that the holder was made to be at the same height as the pre-1968 glass holder, which still exists. A second assumption is that the holder was made to be at the height which minimizes the slope of R(h) for a medium-energy γ-ray emitter like 57Co. An average of these two assumptions was chosen for the model. This assumption was consistent with various checks, described in Sec 3.3. Note that most calibration factors and half-life data date from after 1976, so the height at earlier dates is not very important. These height assumptions are listed in Table 1 and the resulting h(t) and its uncertainty are shown in Fig. 2. As evident from the graph, the slope of h(t) is not constant. Presumably h(t) is monotonic, since during routine operation the positioning ring would only feel forces that would tend to either increase h or leave it constant. Since the functional form of h(t) was unknown for times between the tabulated values, the uncertainty in h was estimated from a rectangular distribution (equal ignorance over the interval) between those measured values. The standard uncertainties2 are shown by the dashed lines in Fig. 2 and correspond to those standard uncertainty intervals calculated from the equally-probable (uniform) intervals divided by .
Table 1.
Assumed values for the height of the positioning-ring on specific dates.
Date | h (cm) | uh (cm) |
---|---|---|
1/19/2010 | 24.29 | 0.05 |
7/28/1993 | 23.38 | 0.05 |
11/4/1992 | 23.01 | 0.05 |
11/16/1986 | 22.54 | 0.05 |
4/1/1981 | 22.16 | 0.05 |
1/1/1968 | 19.6 | 1.1 |
Fig. 2.
The solid line represents the assumed sample-holder height vs. time. The dashed lines represent the standard uncertainty, based on a uniform rectangular distribution. Two possible values for 1967 are described in the text, Sec. 3.1.
To use a historic K-value for an activity determination, the source-holder as it is presently used is now raised by spacers to the h corresponding to the t at which the K-value was originally determined. Predicted activity corrections for radionuclides for which we do not have sources on-hand are calculated, as described in the rest of Sec. 3.
3.2 Relative Response Function
The height dependence of the ionization current depends on the photon-emission spectrum for each radionuclide. In practice, the response is always measured as a ratio, R, to a 226Ra reference source. The height sensitivity of R is largest for radionuclides with γ-ray spectra dissimilar from 226Ra, particularly, emitters of low energy γ-rays and x-rays. The variation of R with h was initially measured experimentally using 125I, 241Am, 57Co, 133Ba, 137Cs and 60Co, as shown in Fig. 3. In order to use this data as a predictor of the effect for other radionuclides, we created a model of R as a function of both h and γ-ray energy. Using that model, R at h for a given radionuclide, nuc, which emits γ-rays of energy Ei with probability Pi detected with relative efficiency (current/photon), εi, would be given by,
(2) |
Fig. 3.
Graph of R versus height. The height corresponding to 1976 is shown by a dashed line. Quadratic fits for R(h) are shown by solid lines. The typical relative fit residual is 0.08 %.
The ε curve was taken from the report by Calhoun, [2]. The ratio of R for a particular nuclide at two times would be given by,
(3) |
which can be expressed as,
(4) |
This f is the correction factor that would be used to adjust historic calibration factors from t1 to t2.
In order to determine R(E, h) from the data for the above-mentioned radionuclides (125I, 241Am, etc.), an effective energy was derived for each nuclide such that R(nuc,ℎ)=R(Eeff,ℎ). This was done by iteration. The first approximation was to take the effective γ-ray energy for each nuclide as,
(5) |
The second iteration began by fitting the 2-dimensional function (Eeff,ℎ), using the estimated Eeff values from the first iteration for each radionuclide. Then f was calculated for each radionuclide using Eqn. (4) with t1 = April 1981 and t2 = February 2010. Trial and error were then used to solve Eqn. (4) for a single energy Eeff(nuc)=Ei. A repeat of the second iteration showed that the Eeff values had converged to the values listed in Table 2. Note that these effective energies are appropriate for the calculation of f, but are not the same as the effective energies that would be used to characterize the detector efficiency.
Table 2.
Effective energies, Eeff, for each given radionuclide, nuc.
nuc | Eeff (keV) |
---|---|
125I | 33.6 |
241Am | 59.6 |
57Co | 124.0 |
133Ba | 315.0 |
137Cs | 662. |
60Co | 1251. |
Once Eeff were fixed, a least-squares fit of R(E,h), was performed using a function that was quadratic in h with both exponential and power terms in E. The function (Fig. 4) contains 8 parameters and was fit to 82 data points. To avoid edge effects, the fit used data with domains of h and E that exceeded those that would be used for calculating activity correction factors. The root mean square (RMS) relative fit residual was 0.08 %, which was about the same as the standard deviation of repeated R measurements for a given height.
Fig. 4.
Parameterization of R(E,h) with RMS residual of 0.08 %.
The effect due to bremsstrahlung was parameterized using a Monte Carlo model of 32P and 90Y, and confirmed by 85Kr measurements. For a 85Kr source moved between h = 21.9 cm and h = 24.3 cm (corresponding to October 1979 and February 2010 in the model), we calculate f = (0.982 ± 0.031). The measured value for the two heights is f = (0.985 ± 0.015), in good agreement with the calculation. About half of the height effect on R for 85Kr is due to bremsstrahlung. For the other radionuclides considered here, bremsstrahlung was not significant.
3.3 Model Validation
The R(E,h) function was tested by making PIC “A” measurements with sources of 109Cd, 131I, 99Mo, 207Bi, 152Eu and 166mHo. Setting t1 as the calibration date (mostly the late 1970’s) and t2 as the measurement date in 2010, the correction factors, f, differed from unity by between 0.6 % and 2.1 %. Measured values agreed well with the model, as shown in Fig. 5. Note that this test did not gauge the accuracy of h(t), since we merely chose two heights and both measured and calculated f as the source was moved between them.
Fig. 5.
Test of the model of the correction due to moving the height of the source by 4.0 cm. The data uncertainty bars are standard deviations of repeated measurements, and the model uncertainties are based on the PIC “A” efficiency, ε, and the parameterization of R(E,h). Here “99Moeq” refers to 99Mo in equilibrium with 99mTc.
Another test of the model was accomplished by comparing activity values determined by new primary measurements to those determined using corrected PIC “A” K-values. The three most stringent tests were 67Ga, 99mTc and 57Co. After correcting the old K-values from c. 1977 to 2010, the old and new K-values differ on average by (0.4 ± 0.3 %), where the uncertainty is only the average uncertainty in the correction. This is a test of the complete model, h(t) and R(E,h).
Parts of the model were also compared with historic NIST/NBS submissions to the International Reference System (SIR). For various radionuclides, NIST had submitted numerous sources over the years to the SIR based on the same K-value. If we assume that the SIR ionization chamber did not suffer a similar problem to the one described here, then we can use the ratio of early and late NIST equivalent-activity values, Ae, (similar to our K-values) as an indication of how PIC “A” has changed with time. To do this, we inverted R(nuc,h) to convert Ae into h for each radionuclide and then compared the implied height changes with our h(t) model. The SIR results for 19 radionuclides are shown in Fig. 6. As the SIR data only indicate the slope of h(t), not an absolute value, the height offset for each dataset (2 to 5 submissions for each radionuclide) was chosen so that the data straddle the model line. The slopes of the various datasets tend to agree with the model. One radionuclide, 59Fe, was excluded from this plot because it disagreed with the other data by 4 cm, probably for some reason other than this sample-holder effect. The uncertainties on the slopes are larger for high-energy γ-ray emitters and lower for low-energy γ-ray emitters, which are more sensitive to height. The scatter of the data in Fig. 6 indicates the uncertainty in this slope comparison.
Fig. 6.
The h(t) model overlaid with SIR submissions by NBS/NIST for all radionuclides that have been submitted multiple times based on the same K-value, excluding 59Fe, which was very discrepant. The R(nuc,h) function was inverted to solve for the apparent change in height. Datasets for each nuclide were shifted vertically to overlay the model line. See text for uncertainty discussion.
Another test was performed using half-live residuals. Although we later estimated half-life corrections, as presented in Sec. 4, we first used the existing data to check our model. We first calculated new world-average half-life values without NIST in the cohort. The data were taken from the Decay Data Evaluation Project [4]. We then fit the historic NIST data with the evaluated half-life and recorded the residuals from the fit. We then used the R(nuc, h) model to convert those residuals for R into residuals in h. Plotting these residuals versus time (Fig. 7) roughly confirmed the h(t) assumption beyond the 1981 height measurement. Recall that we set the 1968 value based on an assumption about what was done then. That assumed height is between the value from a linear extrapolation of measured heights and that suggested by these half-life residuals. Due to the extreme uncertainty of this assumption, we propose that any new half-life fits begin with 1981 data, ignoring data from before that time.
Fig. 7.
Half-life residuals confirming h(t) assumption. A typical uncertainty is shown for a 137Cs point. The scatter in the data also reflects the uncertainty. The data for each nuclide were shifted vertically to overlay the model line.
3.4 Results and Uncertainties for Calibration Factors
The uncertainties of the f calculations are usually mostly due to the h(t) interpolation. The standard uncertainties in h(t) are shown by the dashed lines in Fig. 2. This source of uncertainty caused asymmetric uncertainty in f. For ease of reporting, a symmetrized uncertainty was calculated to be the average of the plus and minus uncertainties. Due to the many assumptions underlying this analysis, there exists a large uncertainty in the uncertainty for these corrections. Multiplying the standard uncertainty by 2 is not a good estimate of a 95 % confidence level on the correction factor.
Table 3 shows an example for the result and uncertainty budget for 155Eu. In this case the K-value determination date was 1 January 1978, and the K-value usage date was 1 January 2001. Results for 11 radionuclides that are routinely measured in PIC “A” are listed in Table 4. They are tabulated by the relative change in response, D = 100(1 − f). The t2 value used for the calculations in Table 4 was February 2010, after the ring was fixed.
Table 3.
Model results and uncertainty budget for 155Eu Activity correction factor, f, for height change from 1978 to 2001.
Component | ui | Comments | ||
---|---|---|---|---|
h(t) | 0.00655 | From rectangular distribution uncertainty for height as a function of date. | ||
r(E,h) γ | 0.00020 | From residuals of fit. | ||
r(E,h) γ | 0.00098 | From tests of fit. | ||
ε | 0.00117 | From uncertainty in relative shape of efficiency curve for γ’s. | ||
r(h) β | 0.00001 | From 20 % uncertainty in R vs. height for β’s. | ||
r(E) β | 0.00003 | From 60 % uncertainty in shape of efficiency curve for β’s. | ||
uc | 0.0067 | combined uncertainty of f | ||
Symmetrized, and corresponding asymmetric uncertainty component due to h(t) | ||||
f | ± | + | − | |
0.9836 | 0.0066 | 0.0033 | 0.0104 | |
Fractional contributions to f | ||||
Β | 0.0020 | |||
γ | 0.9980 |
Table 4.
Calculated K-value relative changes, Δ, for radionuclides commonly-measured in PIC “A”.
Nuclide | K-value date | Δ (%) | uΔ (%) |
---|---|---|---|
201Tl | 7/15/1976 | 2.5 | 0.7 |
123I | 4/28/1976 | 1.5 | 0.4 |
67Ga | 10/7/1977 | 1.5 | 0.5 |
133Xe | 7/27/1988 | 1.40 | 0.15 |
111In | 1/8/1977 | 1.4 | 0.4 |
99Mo | 3/1/1977 | 1.3 | 0.4 |
131I | 5/22/1978 | 1.0 | 0.4 |
133Ba | 8/1/1983 | 0.97 | 0.11 |
54Mn | 2/8/1979 | 0.45 | 0.20 |
88Y | 9/1/1980 | 0.18 | 0.13 |
99mTc | 3/4/2009 | 0.03 | 0.19 |
3.5 SIR Equivalence Values
The SIR equivalent activities, Ae, are essentially calibration factors for the SIR ionization chamber [3], based on submissions from participating laboratories. For many radionuclides, NIST has repeatedly submitted samples to the SIR based on PIC “A” measurements. In those cases, the PIC “A” K-values were typically determined in the late 1970’s. Specifically, the more-recently reported NIST values have been higher than the actual activity. Thus, the NIST activity reported in those cases would be affected by the sample-holder drift. In Table 5 we present corrected Ae values and uncertainties for those cases, following the same methodology as employed for correcting K-values. We also list uD as the uncertainty in Ae due only to the height-correction, while uAe is the total uncertainty in Ae. Omitted from Table 5 are nuclides that have been superseded by more recent primary standardizations as well as the two cases in which the time duration was only 2 years between primary and secondary submission. Most of the changes in Ae have a magnitude similar to their combined standard uncertainties. However, we consider the uncertainty estimate for f to be much less robust than the uncertainty estimate for an activity determined by a primary measurement method. Therefore, these corrected activity values are meant to demonstrate the magnitude of the effect, but are not to be considered of acceptable quality to be included in any average leading to reference values.
Table 5.
Reported NIST SIR equivalent activity, Ae, corrected value, Ae′, and relative difference, D. For each radionuclide, PIC “A” was calibrated at time t1 and used for the SIR submission at t2.
Nuclide | t1 | t2 | Ae (MBq) | Ae′ (MBq) | (%) | D (%) | uD (%) | |
---|---|---|---|---|---|---|---|---|
203Hg | 1976 | 1985 | 68.00 | 0.35 | 67.65 | 0.46 | 0.40 | 0.28 |
99Mo | 1978 | 1998 | 64.69 | 0.24 | 64.13 | 0.36 | 0.80 | 0.29 |
111In | 1977 | 1998 | 43.45 | 0.12 | 43.00 | 0.23 | 0.93 | 0.31 |
131I | 1979 | 1998 | 40.50 | 0.14 | 40.23 | 0.21 | 0.62 | 0.26 |
51Cr | 1981 | 1999 | 0.4893 | 0.00 | 0.4863 | 0.00 | 0.61 | 0.28 |
137Cs | 1982 | 2001 | 27.63 | 0.10 | 27.54 | 0.10 | 0.36 | 0.12 |
85Sr | 1977 | 2001 | 30.09 | 0.09 | 29.88 | 0.13 | 0.61 | 0.21 |
59Fe | 1979 | 2001 | 14.64 | 0.06 | 14.61 | 0.06 | 0.19 | 0.12 |
54Mn | 1979 | 2002 | 19.27 | 0.05 | 19.20 | 0.07 | 0.33 | 0.15 |
88Y | 1980 | 2002 | 6.913 | 0.02 | 6.904 | 0.02 | 0.13 | 0.11 |
133Xe | 1992 | 2001 | 896.10 | 3.70 | 889.59 | 4.49 | 0.73 | 0.28 |
139Ce | 1987 | 1997 | 0.1344 | 0.00 | 0.1337 | 0.00 | 0.54 | 0.20 |
153Gd | 1989 | 1998 | 369.4 | 2.50 | 366.5 | 2.77 | 0.79 | 0.32 |
There is no change to the NIST 99mTc link to the SIR. In that case, the PIC “A” had been calibrated in March 2009 and used in May 2009 with the SIR Transfer Instrument. The model predicts 0.008 % correction for that case, which is insignificant compared to the activity uncertainty of 0.43 %. This was verified by checking the results from measuring the same 57Co sample before the PIC “A” calibration in 2009 and again in 2010. The results were the same with an uncertainty of 0.08 %.
4. Half-lives
4.1 Correction Method
For over 30 years, NIST has been reporting half-lives, Tm, measured using PIC “A” (Unterweger, 2002). These half-lives were determined over various time spans from c. 1974 until 2001. The recently-discovered drift in the sample-positioning ring has affected these Tm values. These discrepancies had been noted by other national laboratories and data compilers.
In this section we attempt to estimate the effect. To estimate the corrected half-life, T, we assume a linear change in h over time, as shown by the dashed line in Fig. 8.
Fig. 8.
The solid line represents assumed model of sample-holder height vs. time. The dashed line represents the simplified, linear model used for these estimated half-life corrections and the dotted lines represent the uncertainty in the slope.
Note that Tm had been determined by Unterweger [5] assuming the usual decay law for a source producing signal R0 at time zero and signal R at time t:
(6) |
But in reality, due to the shifting positioning-ring, the measured decay is better approximated by,
(7) |
Where the slope ξ = dR/dt was determined using the model of R(nuc,h) for each radionuclide and the linear h(t) shown in Fig. 8. By equating Eqns. (6) and (7), one can solve for the relative change in apparent half-life for a 2-point half-life determination,
(8) |
For all cases reported here ξt < 0.01, so to good approximation,
(9) |
Thus, the measured half-lives can be corrected using z from Eqn. (9) as follows,
(10) |
We can also express T in terms of the change in half-life, Δ,
(11) |
(12) |
The uncertainty in T due to Tm and ξ is given by,
(13) |
where is the uncertainty reported by (Unterweger, 2002).
A relevant metric is the change in half-life relative to ,
(14) |
This derivation was based on a 2-point half-life determination. We tested the applicability of this approximation to the Tm data, which were determined from a least-squares fit to multiple points of a decaying-source. We tested this approximation for 137Cs, which has a large T, large j and was measured for t = 0.9T and also tested it for 155Eu, which has a large ξ medium j and t =4T. We first simulated 30 decay-data points for the ideal case, using Eqn. (6) over the time-span t. We included Gaussian noise such that the uncertainty from a least-squares fit to the simulated data was . We then simulated data using Eqn. (7) with the same noise as before, and calculated Δ from the respective fit parameters, Tm and T It is important to note that the χ2 value and fit uncertainty were identical with and without the linear effect and no trend was visible in any of the residuals, shown in Fig. 9. Therefore, this effect would not have been detected from the original residuals. For 137Cs, the value from [5] is Tm = (11018.3 ± 9.5 d). The simulation gave Δ = 109 d, whereas Eqn. (7) gave Δ = (104 ± 54) d. Likewise for 155Eu, Tm = (1739.1 ± 0.5 d), the simulation gave Δ = 8.4 d and Eqn. (7) gave Δ = (8.5 ± 3.5) d. Those uncertainties are due to ξ = dR/dt only. Based on these two tests, I am satisfied that the approximations of Eqns. (7) through (9) are justified.
Fig. 9.
Relative residuals from least squares fit to simulated, ideal 137Cs decay data from Eqn. (6) (open squares) and warped data from Eqn. (7) (filled circles). The same Gaussian noise, the average magnitude of which is represented by the uncertainty bars, was applied to both data sets and used in the weighting of the fit. The fit value for half-life in the two cases differed by Δ = 109 d. In both cases the uncertainty returned was 9.5 d, χ2 per degree of freedom was 1, and no trend was visible in the residuals.
4.2 Corrected Half-lives
The results for all of the reported half-lives of [5] except 3H, which had been determined by gas counting, are listed in Table 6. For each nuclide, we list Tm and from [5] the half-life shift Δ and corrected values T and uT calculated here, using Eqns. (11) and (13) respectively. Also tabulated are j, the change in half-life relative to and the ratio of the new to old uncertainty. Of the 63 half-life corrections reported here, 18 have j > 1 and are indicated in bold face in Table 6. Long-lived radionuclides, especially those that are low-energy photon emitters (85Kr, 155Eu), tend to have the largest relative corrections. In most cases the change in half-life and its uncertainty are small. That is, in most cases j < 1 and . The uncertainty component due to dR/dt results from, in most cases, the estimate of the height of the positioning ring versus time. For radionuclides that emit γ-rays of various energies, the calculation of the ionization-chamber efficiency as a function of energy can contribute significantly to the overall uncertainty. Example uncertainty budgets for 137Cs and 155Eu are shown in Table 7.
Table 6.
Measured half-lives from Unterweger [5] and corrected half-lives, T, from present work. For each radionuclide, the same unit is used for all non-ratio quantities. Radionuclides for which the correction is larger than the original uncertainty [5] are in bold face.
Nuclide | Tm | Unit | T | uT | Δ | |||
---|---|---|---|---|---|---|---|---|
18F | 1.8295 | 0.0003 | h | 1.8295 | 0.0003 | 2E-07 | 0.0004 | 1.0 |
22Na | 950.97 | 0.15 | d | 950.4 | 0.4 | 0.6 | 4 | 2.4 |
24Na | 14.951 | 0.003 | h | 14.951 | 0.003 | 1E-06 | 0.0003 | 1.0 |
32P | 14.263 | 0.003 | d | 14.262 | 0.003 | 0.001 | 0.3 | 1.0 |
46Sc | 83.83 | 0.07 | d | 83.828 | 0.066 | 0.003 | 0.05 | 1.0 |
51Cr | 27.7010 | 0.0012 | d | 27.6999 | 0.0013 | 0.0011 | 0.9 | 1.1 |
54Mn | 312.03 | 0.03 | d | 311.97 | 0.05 | 0.06 | 1.8 | 1.4 |
57Co | 272.11 | 0.26 | d | 271.95 | 0.27 | 0.16 | 0.6 | 1.0 |
58Co | 70.77 | 0.11 | d | 70.77 | 0.11 | 0.004 | 0.03 | 1.0 |
59Fe | 44.507 | 0.007 | d | 44.507 | 0.007 | 0.001 | 0.10 | 1.0 |
60Co | 1925.20 | 0.25 | d | 1924.0 | 0.9 | 1.2 | 5 | 3.7 |
62Cu | 9.673 | 0.008 | m | 9.672 | 0.008 | 7E-08 | 0.00001 | 1.0 |
65Zn | 244.16 | 0.10 | d | 244.14 | 0.10 | 0.02 | 0.2 | 1.0 |
67Ga | 3.2615 | 0.0005 | d | 3.2615 | 0.0005 | 2E-05 | 0.04 | 1.0 |
75Se | 119.81 | 0.07 | d | 119.78 | 0.07 | 0.03 | 0.4 | 1.0 |
85Kr | 3935.7 | 1.2 | d | 3905 | 19 | 31 | 26 | 16 |
85Sr | 64.853 | 0.008 | d | 64.848 | 0.008 | 0.005 | 0.6 | 1.0 |
88Y | 106.63 | 0.04 | d | 106.62 | 0.04 | 0.003 | 0.07 | 1.0 |
99Mo | 65.924 | 0.006 | h | 65.924 | 0.006 | 0.0002 | 0.03 | 1.1 |
99mTc | 6.0072 | 0.0009 | h | 6.0072 | 0.0009 | 3E-06 | 0.004 | 1.0 |
99ATc | 6.012 | 0.003 | h | 6.012 | 0.003 | 3E-06 | 0.001 | 1.0 |
103Ru | 39.31 | 0.04 | d | 39.31 | 0.04 | 0.0017 | 0.04 | 1.0 |
109Cd | 463.3 | 0.6 | d | 462.6 | 0.7 | 0.7 | 1.1 | 1.1 |
110mAg | 249.950 | 0.024 | d | 249.91 | 0.03 | 0.04 | 1.6 | 1.4 |
111In | 2.8048 | 0.0005 | d | 2.8048 | 0.0005 | 1E-05 | 0.03 | 1.0 |
113Sn | 115.08 | 0.08 | d | 115.06 | 0.08 | 0.018 | 0.22 | 1.0 |
117mSn | 14.00 | 0.05 | d | 14.00 | 0.05 | 0.0004 | 0.01 | 1.0 |
123I | 13.2235 | 0.0019 | h | 13.2235 | 0.0019 | 1E-05 | 0.01 | 1.0 |
125I | 59.49 | 0.13 | d | 59.47 | 0.13 | 0.02 | 0.2 | 1.0 |
125Sb | 1007.56 | 0.10 | d | 1006.5 | 0.6 | 1.1 | 11 | 5.6 |
127Xe | 36.345 | 0.003 | d | 36.342 | 0.003 | 0.002 | 0.8 | 1.1 |
131I | 8.0197 | 0.0022 | d | 8.0196 | 0.0022 | 9E-05 | 0.04 | 1.0 |
131mXe | 11.934 | 0.021 | d | 11.934 | 0.021 | 0.0004 | 0.02 | 1.0 |
133Ba | 3854.7 | 2.8 | d | 3832.3 | 10.8 | 22.4 | 8 | 3.8 |
133Xe | 5.2475 | 0.0005 | d | 5.2474 | 0.0005 | 1E-04 | 0.2 | 1.0 |
134Cs | 753.88 | 0.15 | d | 753.43 | 0.28 | 0.4 | 3 | 1.9 |
137Cs | 11018.3 | 9.50 | d | 10915 | 55 | 104 | 11 | 5.8 |
139Ce | 137.73 | 0.09 | d | 137.70 | 0.09 | 0.04 | 0.4 | 1.0 |
140Ba | 12.7527 | 0.0023 | d | 12.7525 | 0.0023 | 0.0002 | 0.08 | 1.0 |
140La | 40.293 | 0.012 | h | 40.293 | 0.012 | 2E-05 | 0.002 | 1.0 |
141Ce | 32.510 | 0.024 | d | 32.508 | 0.024 | 0.0022 | 0.09 | 1.0 |
144Ce | 284.53 | 0.03 | d | 284.35 | 0.08 | 0.18 | 5 | 2.4 |
152Eu | 4947.2 | 1.1 | d | 4929 | 10 | 18 | 16 | 8.8 |
153Gd | 239.47 | 0.07 | d | 239.29 | 0.10 | 0.18 | 2.6 | 1.5 |
153Sm | 46.2853 | 0.0014 | h | 46.2850 | 0.0030 | 0.0003 | 0.2 | 2.2 |
154Eu | 3145.2 | 1.1 | d | 3138 | 4 | 7 | 6 | 3.5 |
155Eu | 1739.1 | 0.5 | d | 1731 | 3 | 8 | 19 | 7.8 |
166Ho | 26.794 | 0.023 | h | 26.794 | 0.023 | 8E-05 | 0.004 | 1.0 |
169Yb | 32.015 | 0.009 | d | 32.011 | 0.009 | 0.004 | 0.4 | 1.0 |
177Lu | 6.640 | 0.010 | d | 6.640 | 0.010 | 0.0001 | 0.01 | 1.0 |
181W | 121.10 | 0.06 | d | 121.03 | 0.07 | 0.06 | 1.0 | 1.1 |
186Re | 89.25 | 0.07 | h | 89.25 | 0.07 | 0.0010 | 0.014 | 1.0 |
188Re | 17.001 | 0.022 | h | 17.001 | 0.022 | 2E-5 | 0.001 | 1.0 |
188W | 69.78 | 0.05 | d | 69.77 | 0.05 | 0.012 | 0.25 | 1.0 |
192Ir | 73.81 | 0.02 | d | 73.802 | 0.019 | 0.008 | 0.4 | 1.0 |
195Au | 186.10 | 0.05 | d | 186.01 | 0.06 | 0.09 | 1.9 | 1.3 |
198Au | 2.69517 | 0.00021 | d | 2.69516 | 0.00021 | 9E-06 | 0.04 | 1.0 |
201Tl | 3.0456 | 0.0015 | d | 3.0456 | 0.0015 | 3E-05 | 0.02 | 1.0 |
202Tl | 12.47 | 0.08 | d | 12.47 | 0.08 | 0.0003 | 0.004 | 1.0 |
203Hg | 46.619 | 0.027 | d | 46.615 | 0.027 | 0.004 | 0.14 | 1.0 |
203Pb | 51.92 | 0.04 | h | 51.92 | 0.04 | 0.0003 | 0.007 | 1.0 |
207Bi | 11523 | 15 | d | 11403 | 61 | 120 | 8 | 4.1 |
228Th | 698.6 | 0.4 | d | 698.4 | 0.4 | 0.2 | 0.7 | 1.1 |
Table 6.
Uncertainty budgets for corrected half-lives, T, for 137Cs and 155Eu. The designation “A or B” refers to the type of uncertainty evaluation, as described in Taylor and Kuyatt [6].
Source | Description | A or B | 137Cs ui (%) | 155Eu ui (%) |
---|---|---|---|---|
Tm | Uncertainty for Tm reported by [5] | B | .086 | 0.026 |
h(t) | Due to standard uncertainty in position change with time. | B | 0.46 | 0.19 |
R(E,h) | Due to standard uncertainty in R(E,h) for a single photon energy. | B | 0.17 | 0.037 |
ε | Due to standard uncertainty in detector efficiency as a function of energy. | B | 0 | 0.032 |
uc (%) | 0.50 | 0.20 |
The top graph in Fig. 10 shows the relative change in half-life, Δ/Tm, versus Tm. The bottom graph in Fig. 10 shows . versus half-life. Also shown in the top plot are the differences of Schrader [7] measurements from Unterweger [5], (again relative to ), for comparison. This effect may account for the historic difference of Unterweger and Schrader for some long-lived radionuclides (e.g., 85Kr).
Fig. 10.
Top: Relative half-life correction. Closed circles are Δ with uncertainties described in text. Open squares are differences of Schrader [7] from Unterweger [5]. Bottom: Half-life corrections relative to original uncertainties [5]. The uncertainty bars are due to dR/dt only.
I consider these half-life corrections and their uncertainties to be rough approximations for use in identifying which half-lives are significantly affected, and at approximately what magnitude. For short half-lives (less than about 10 d), the calculated corrections are usually much smaller than the originally-stated uncertainties. A more accurate and more precise correction could be obtained by re-fitting corrected R data, restricting the dataset to dates beyond 1981, when the first known measurements of h exist. That method would avoid using the linear h(t) assumption, with its large uncertainty, as was done here. Such an approach would require the start and stop time for each half-life measurement.
Acknowledgments
The author is grateful to many scientists and technicians working in the NIST radioactivity group for their efforts that made this work possible. Dr. Larry Lucas discovered the discrepancy between the sample holder and a historic drawing, led the historic research, and made many of the response vs. height measurements that were used to constrain the model. Mr. Daniel Golas and Ms. Rheannan Young provided further PIC “A” data. Dr. Michael Unterweger provided his half-life data. (And the author took some data too!) Mr. Jeffrey Cessna led the communications effort. This manuscript benefited considerably from the expert criticism of Dr. Ronald Collé.
Biography
About the author: Ryan Fitzgerald is a physicist in the Radioactivity Group of the Ionizing Radiation Division of the Physical Measurement Laboratory at NIST. He carries out research in radioactivity metrology to meet demands from a variety of sectors: basic science, health, security, environmental and commercial.
Footnotes
Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.
All uncertainties reported in this paper are combined standard uncertainties in the corresponding measurand, unless otherwise noted.
5. References
- [1].Sharpe J, Wade F. A Re-entrant thimble ionisation chamber. Atomics. 1953 Feb 4;:32. [Google Scholar]
- [2].Calhoun JM. Radioactivity Calibrations with the NBS “4π”γ Ionization Chamber, and Other NBS Radioactivity Calibration Capabilities, NBS SP 250-10. Gaithersburg, MD, USA: Aug, 1986. pp. 11–12. [Google Scholar]
- [3].Rytz, A A. International coherence of activity measurements. Environ Int. 1978;1:15–18. [Google Scholar]
- [4].Bé MM. Decay Data Evaluation Project (DDEP) Laboratoire National Henri Becquerel; http://www.nucleide.org/DDEP_WG/DDEPdata.htm [accessed 2010] [Google Scholar]
- [5].Unterweger MP. Half-life measurements at the National Institute of Standards and Technology. Appl Radiat Isot. 2002;56:125–130. doi: 10.1016/s0969-8043(01)00177-4. [DOI] [PubMed] [Google Scholar]
- [6].Taylor BN, Kuyatt CE. Guidelines For evaluating and Expressing the Uncertainty of NIST Measurement Results, NIST TN 1297. Gaithersburg, MD, USA: 1994. [Google Scholar]
- [7].Schrader H. Half-life measurements with ionization chambers - A study of systematic effects and results. Appl Radiat Isot. 2004;60:317–323. doi: 10.1016/j.apradiso.2003.11.039. [DOI] [PubMed] [Google Scholar]