Abstract
The purpose of the study was to evaluate the efficacy and safety of wet chamber warming goggles (Blephasteam®) in patients with meibomian gland dysfunction (MGD) unresponsive to warm compress treatment. We consecutively enrolled 50 adult patients with low-delivery, non-cicatricial, MGD, and we instructed them to apply warm compresses twice a day for 10 min for 3 weeks and to use Blephasteam® (Laboratoires Thea, Clermont-Ferrand, France) twice a day for 10 min for the following 3 weeks. We considered “not-responders” to warm compress treatment the patients who showed no clinically significant Ocular Surface Disease Index (OSDI) improvement after the first 3 weeks. Clinical and in vivo confocal outcome measures were assessed in the worst eye (lower BUT) at baseline, after 3 weeks, and after 6 weeks. Eighteen/50 patients were not-responders to warm compress treatment. These patients, after 3 weeks of treatment with Blephasteam®, showed significant improvement of OSDI score (36.4 ± 15.8 vs 20.2 ± 12.4; P < 0.05, paired samples t test), increased BUT (3.4 ± 1.6 vs 7.6 ± 2.7; P < 0.05), and decreased acinar diameter and area (98.4 ± 18.6 vs 64.5 ± 14.4 and 8,037 ± 1,411 vs 5,532 ± 1,172, respectively; P < 0.05). Neither warm compresses nor Blephasteam® determined adverse responses. In conclusion, eyelid warming is the mainstay of the clinical treatment of MGD and its poor results may be often due to lack of compliance and standardization. Blephasteam® wet chamber warming goggles are a promising alternative to classical warm compress treatment, potentially able to improve the effectiveness of the “warming approach.”
Keywords: Meibomian gland, MGD, Confocal microscopy, Blephasteam, Dry eye, Ocular surface
Introduction
Meibomian gland dysfunction (MGD) is a common chronic condition, affecting the tear film and the ocular surface and causing symptoms of eye irritation [1, 2].
Meibomian gland obstruction, due to either terminal duct obstruction or altered secretion, is the most common form of MGD [2]. Eyelid warming, usually achieved with simple warm compresses, is regarded as the mainstay of the clinical treatment of this condition, but its efficacy is affected by lack of standardization, in terms of duration and maintenance of temperature, and by a scarce compliance [3]. In the last few years, different devices have been developed in order to try to improve the heat therapy efficacy [3–8].
In vivo laser scanning confocal microscopy (LSCM) is an emerging technology to study the ocular surface in several conditions, including dry eye and MGD [9–11]. At present, LSCM is showing promising clinical applications [9] and recent studies reported its helpfulness in detecting ocular surface response to treatment [12–14].
The aim of this research is to evaluate the safety and efficacy of Blephasteam® (Laboratoires Thea, Clermont-Ferrand, France) eyelid-warming device in the management of MGD unresponsive to warm compress treatment and to study treatment-related clinical and confocal changes.
Methods
We consecutively studied 50 adult patients with mild to moderate low-delivery, non-cicatricial, MGD. Written informed consent was obtained from all subjects before enrollment, and this study adhered to the tenets of the declaration of Helsinki. MGD classification and grading were performed according to the 2011 International Workshop on MGD [2, 15]. Briefly, we included patients with symptoms of ocular discomfort (Ocular Surface Disease Index—OSDI—score >12) [16], tear fluorescein break-up time (BUT) <5, mild to moderate meibum quality abnormality (score 11–20, according to Bron’ Scale) [17, 18], and mild to moderate expressibility reduction [15]. Exclusion criteria included blepharitis, ocular allergies, contact lens wear, hyposecretive dry eye, history of ocular trauma or surgery, cicatricial ocular surface diseases, and systemic or topical therapies (tear substitutes excepted) that would interfere with tear film and ocular surface.
All these patients were instructed to perform warm compress treatment twice a day for 10 min.
After 3 weeks of treatment, we defined as “not-responders to warm compress treatment” patients who did not show clinically significant OSDI improvement, based on the previously validated OSDI minimal clinically important difference [19].
Both “responder” and “not-responder” patients were then instructed to use Blephasteam® (Laboratoires Thea, Clermont-Ferrand, France) twice a day for 10 min, following the manufacturer instructions, for the following 3 weeks.
All the visits (screening and enrollment—V0, visit at day 21 ± 2—V1, and visit at day 42 ± 2—V2) included the same procedures, performed in the order suggested by the 2007 International Dry Eye Workshop [20]: OSDI questionnaire, fluorescein BUT, fluorescein corneal staining (quantified using the CLEK scheme) [21], Schirmer test without topical anesthesia, meibomian gland expression, and LSCM (HRT II Corneal Rostock Module, Heidelberg Engineering GmbH, Dossenheim, Germany) of meibomian glands. Confocal examination was performed at the lower eyelid margin, following a previously published procedure [22–24]. Meibomian acinar units were analyzed, quantifying their density, mean diameter, and area.
No changes in the concomitant medications, including artificial tears, were allowed during the study period.
The outcome measures were assessed in the worst eye, defined as the eye with the lower BUT.
Statistical analysis
The statistical analysis was conducted with commercial software (SPSS for Windows, ver. 12.0; SPSS, Chicago, IL, USA). The comparisons between consecutive visits were performed using the t test for repeated measures for parametric variables and with the Wilcoxon test for non-parametric variables. The comparisons between “responders” and “not-responders” to warm compress treatment were done using the t test for independent samples for parametric variables and the Mann–Whitney U test for non-parametric variables. The minimum criterion for tests of significance was P < 0.01.
Results
The 50 enrolled patients (31 females and 19 males) had a mean age of 64 ± 12 years.
After 3 weeks of warm compress treatment, 18 patients (36 %) were classified as “not-responders” and 32 (64 %) as “responders.”
No significant differences were found between the baseline characteristics of “responders” and “not-responders.”
Neither warm compresses nor Blephasteam® caused adverse events or problems of tolerability in our patients.
In the “not-responders” group, comparing V0 to V1, we found no significant treatment-related (warm compresses) improvement in clinical findings. Furthermore, in this same group, we observed significant improvement of OSDI score (Table 1), increase of BUT (Table 2), and decrease of both acinar mean diameter (Table 3) and area (Table 4) (Fig. 1) from V1 to V2.
Table 1.
V0 | V1 | V2 | *P (V0 vs V1) | *P (V1 vs V2) | |
---|---|---|---|---|---|
Responders (n = 32) | 36.3 ± 17.1 | 22.7 ± 13.1 | 20.5 ± 14.2 | <0.05 | n.s. |
Not-responders (n = 18) | 38.2 ± 15.5 | 36.4 ± 15.8 | 20.2 ± 12.4 | n.s. | <0.05 |
**P | n.s. | <0.05 | n.s. |
n.s. Not significant
* P by t test for repeated measures
** P by t test for independent samples
Table 2.
V0 | V1 | V2 | *P (V0 vs V1) | *P (V1 vs V2) | |
---|---|---|---|---|---|
Responders (n = 32) | 3.3 ± 1.8 | 6.7 ± 2.1 | 7.2 ± 2.2 | <0.05 | n.s. |
Not-responders (n = 18) | 3.8 ± 1.4 | 3.4 ± 1.6 | 7.6 ± 2.7 | n.s. | <0.05 |
**P | n.s. | <0.05 | n.s. |
n.s. Not significant
* P by t test for repeated measures
** P by t test for independent samples
Table 3.
V0 | V1 | V2 | *P (V0 vs V1) | *P (V1 vs V2) | |
---|---|---|---|---|---|
Responders (n = 32) | 108.3 ± 19.4 | 84.2 ± 17.6 | 77.5 ± 18.0 | <0.05 | n.s. |
Not-responders (n = 18) | 104.8 ± 15.1 | 98.4 ± 18.6 | 64.5 ± 14.4 | n.s. | <0.05 |
**P | n.s. | <0.05 | n.s. |
n.s. Not significant
* P by t test for repeated measures
** P by t test for independent samples
Table 4.
V0 | V1 | V2 | *P (V0 vs V1) | *P (V1 vs V2) | |
---|---|---|---|---|---|
Responders (n = 32) | 8,645 ± 1,980 | 6,026 ± 1,883 | 5,879 ± 1,820 | <0.05 | n.s. |
Not-responders (n = 18) | 8,276 ± 1,691 | 8,037 ± 1,411 | 5,532 ± 1,172 | n.s. | <0.05 |
**P | n.s. | <0.05 | n.s. |
n.s. Not significant
* P by t test for repeated measures
** P by t test for independent samples
Fluorescein staining, meibomian gland expressibility, and meibum quality showed no significant differences during the follow-up (Wilcoxon test).
Discussion
MGD, specifically in the low-delivery, non-cicatricial form, is an increasingly prevalent affliction/with potentially severe detriments to well-being [1]. Dedicated and reliable methods to monitor and effective approaches to manage the disease are still partially unmet needs. LSCM offers new opportunities to perform in vivo, non-invasive examinations of meibomian glands. This technology has proven to have the potential to diagnose MGD with high sensitivity and specificity [25] and to explore the different patterns of the disease, providing new information on the pathogenetic process and the acinar morphological changes [22–24]. Moreover, the previously hypothesized [12] suitability of LSCM to detect and quantify the MGD response to treatment may be confirmed by the present study. Our results showed agreement between clinical (symptoms and BUT) and confocal changes, in the absence of significant variations of traditional meibomian expression scores. These interesting data suggest the need for future studies to compare confocal and clinical examination of meibomian glands and to confirm the utility of incorporating LSCM analysis in the assessment of patient response to therapy.
Our study confirms the usefulness of the well-known and broadly accepted [3] eyelid-warming approach to MGD, but it also highlights that this treatment may be ineffective in 1/3 of the patients, although carefully selected. The good clinical and morphological response to Blephasteam® of subjects “not-responder” to warm compress treatment suggests that ineffectiveness may be due to poor standardization and compliance more than to poor rationale. Medical devices dedicated to eyelid warming try to bridge this gap in the management of MGD patients. Blephasteam® is an electrical pair of goggles that provides warmth and steam, with controlled treatment temperature and duration. Previous studies in healthy volunteers [7, 8] already showed that this device, compared to traditional warm compresses, provides longer warming of the eyelid margin without any adverse ocular response. Our results show its safety and efficacy in MGD patients and its potential superiority to warm compress treatment in ideal candidates to warming therapy.
In conclusion, new technologies as LSCM and eyelid-warming devices promise to play an important role in the management of MGD, improving the effectiveness of non-pharmacological treatment of this common condition.
Acknowledgments
The study was supported by Laboratoires Thea, Clermont-Ferrand, France which provided free of charge the devices used during the research.
References
- 1.Schaumberg DA, Nichols JJ, Papas EB, Tong L, Uchino M, Nichols KK. The international workshop on meibomian gland dysfunction: report of the subcommittee on the epidemiology of, and associated risk factors for MGD. Invest Ophthalmol Vis Sci. 2011;52:1994–2005. doi: 10.1167/iovs.10-6997e. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Nelson JD, Shimazaki J, Benitez-del-Castillo JM, et al. The international workshop on meibomian gland dysfunction: report of the definition and classification subcommittee. Invest Ophthalmol Vis Sci. 2011;52:1930–1937. doi: 10.1167/iovs.10-6997b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Geerling G, Tauber J, Baudouin C, et al. The international workshop on meibomian gland dysfunction: report of the subcommittee on management and treatment of meibomian gland dysfunction. Invest Ophthalmol Vis Sci. 2011;52:2050–2064. doi: 10.1167/iovs.10-6997g. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Ishida R, Matsumoto Y, Onguchi T, et al. Tear film with “Orgahexa EyeMasks” in patients with meibomian gland dysfunction. Optom Vis Sci. 2008;85:684–691. doi: 10.1097/OPX.0b013e318181ae83. [DOI] [PubMed] [Google Scholar]
- 5.Lane SS, DuBiner HB, Epstein RJ, et al. A new system, the LipiFlow, for the treatment of meibomian gland dysfunction. Cornea. 2012;31:396–404. doi: 10.1097/ICO.0b013e318239aaea. [DOI] [PubMed] [Google Scholar]
- 6.Friedland BR, Fleming CP, Blackie CA, Korb DR. A novel thermodynamic treatment for meibomian gland dysfunction. Curr Eye Res. 2011;36:79–87. doi: 10.3109/02713683.2010.509529. [DOI] [PubMed] [Google Scholar]
- 7.Purslow C. Evaluation of the ocular tolerance of a novel eyelid-warming device used for meibomian gland dysfunction. Contact Lens Anterior Eye. 2013;36:226–231. doi: 10.1016/j.clae.2013.02.009. [DOI] [PubMed] [Google Scholar]
- 8.Pult H, Riede-Pult BH, Purslow C. A comparison of an eyelid-warming device to traditional compress therapy. Optom Vis Sci. 2012;89:E1035–E1041. doi: 10.1097/OPX.0b013e31825c3479. [DOI] [PubMed] [Google Scholar]
- 9.Villani E, Magnani F, Viola F, et al. In vivo confocal evaluation of the ocular surface morpho-functional unit in dry eye. Optom Vis Sci. 2013;90:576–586. doi: 10.1097/OPX.0b013e318294c184. [DOI] [PubMed] [Google Scholar]
- 10.Villani E, Baudouin C, Efron N, et al. In vivo confocal microscopy of the ocular surface: from bench to bedside. Curr Eye Res. 2014;39:213–231. doi: 10.3109/02713683.2013.842592. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Villani E, Mantelli F, Nucci P. In-vivo confocal microscopy of the ocular surface: ocular allergy and dry eye. Curr Opin Allergy Clin Immunol. 2013;13:569–576. doi: 10.1097/ACI.0b013e328364ec92. [DOI] [PubMed] [Google Scholar]
- 12.Matsumoto Y, Shigeno Y, Sato EA, et al. The evaluation of the treatment response in obstructive meibomian gland disease by in vivo laser confocal microscopy. Graefes Arch Clin Exp Ophthalmol. 2009;247:821–829. doi: 10.1007/s00417-008-1017-y. [DOI] [PubMed] [Google Scholar]
- 13.Rao K, Leveque C, Pflugfelder SC. Corneal nerve regeneration in neurotrophic keratopathy following autologous plasma therapy. Br J Ophthalmol. 2010;94:584–591. doi: 10.1136/bjo.2009.164780. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Villani E, Galimberti D, Del Papa N, Nucci P, Ratiglia R. Inflammation in dry eye associated with rheumatoid arthritis: cytokine and in vivo confocal microscopy study. Innate Immun. 2013;19:420–427. doi: 10.1177/1753425912471692. [DOI] [PubMed] [Google Scholar]
- 15.Tomlinson A, Bron AJ, Korb DR, et al. The international workshop on meibomian gland dysfunction: report of the diagnosis subcommittee. Invest Ophthalmol Vis Sci. 2011;52:2006–2049. doi: 10.1167/iovs.10-6997f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Schiffman RM, Christianson MD, Jacobsen G, Hirsch JD, Reis BL. Reliability and validity of the Ocular Surface Disease Index. Arch Ophthalmol. 2000;118:615–621. doi: 10.1001/archopht.118.5.615. [DOI] [PubMed] [Google Scholar]
- 17.Bron AJ, Benjamin L, Snibson GR. Meibomian gland disease. Classification and grading of lid changes. Eye. 1991;5:395–411. doi: 10.1038/eye.1991.65. [DOI] [PubMed] [Google Scholar]
- 18.Shimazaki J, Sakata M, Tsubota K. Ocular surface changes and discomfort in patients with meibomian gland dysfunction. Arch Ophthalmol. 1995;113:1266–1270. doi: 10.1001/archopht.1995.01100100054027. [DOI] [PubMed] [Google Scholar]
- 19.Miller KL, Walt JG, Mink DR, et al. Minimal clinically important difference for the Ocular Surface Disease Index. Arch Ophthalmol. 2010;128:94–101. doi: 10.1001/archophthalmol.2009.356. [DOI] [PubMed] [Google Scholar]
- 20.Methodologies to diagnose and monitor dry eye disease: report of the Diagnostic Methodology Subcommittee of the International Dry Eye WorkShop (2007). Ocul Surf 5:108–52 [DOI] [PubMed]
- 21.McMahon TT, Szczotka-Flynn L, Barr JT, et al. CLEK Study Group. A new method for grading the severity of keratoconus: the keratoconus severity score (KSS) Cornea. 2006;25:794–800. doi: 10.1097/01.ico.0000226359.26678.d1. [DOI] [PubMed] [Google Scholar]
- 22.Villani E, Canton V, Magnani F, Viola F, Nucci P, Ratiglia R. The aging meibomian gland: an in vivo confocal study. Invest Ophthalmol Vis Sci. 2013;54:4735–4740. doi: 10.1167/iovs.13-11914. [DOI] [PubMed] [Google Scholar]
- 23.Villani E, Ceresara G, Beretta S, Magnani F, Viola F, Ratiglia R. In vivo confocal microscopy of meibomian glands in contact lens wearers. Invest Ophthalmol Vis Sci. 2011;52:5215–5219. doi: 10.1167/iovs.11-7427. [DOI] [PubMed] [Google Scholar]
- 24.Villani E, Beretta S, De Capitani M, Galimberti D, Viola F, Ratiglia R. In vivo confocal microscopy of meibomian glands in Sjögren’s syndrome. Invest Ophthalmol Vis Sci. 2011;52:933–939. doi: 10.1167/iovs.10-5995. [DOI] [PubMed] [Google Scholar]
- 25.Ibrahim OM, Matsumoto Y, Dogru M, et al. The efficacy, sensitivity, and specificity of in vivo laser confocal microscopy in the diagnosis of meibomian gland dysfunction. Ophthalmology. 2010;117:665–672. doi: 10.1016/j.ophtha.2009.12.029. [DOI] [PubMed] [Google Scholar]