Abstract
Introduction
We recently reported the development of the [18F]fluorodiethylene glycol ester of rhodamine B as a potential positron emission tomography (PET) tracer for myocardial perfusion imaging (MPI). This compound was developed by optimizing the ester moiety on the rhodamine B core, and its pharmacokinetic properties were found to be superior to those of the prototype ethyl ester. The goal of the present study was to optimize the rhodamine core while retaining the fluorodiethyleneglycol ester prosthetic group.
Methods
A series of different rhodamine cores (rhodamine 6G, rhodamine 101, and tetramethylrhodamine) were labeled with 18F using the corresponding rhodamine lactones as the precursors and [18F]fluorodiethylene glycol ester as the prosthetic group. The compounds were purified by semipreparative HPLC, and their biodistribution was measured in rats. Additionally, the uptake of the compounds was evaluated in isolated rat cardiomyocytes.
Results
As was the case with the different prosthetic groups, we found that the rhodamine core has a significant effect on the in vitro and in vivo properties of this series of compounds. Of the rhodamines evaluated to date, the pharmacologic properties of the 18F-labeled diethylene glycol ester of rhodamine 6G are superior to those of the 18F-labeled diethylene glycol esters of rhodamine B, rhodamine 101, and tetramethylrhodamine. As with 18F-labeled rhodamine B, [18F]rhodamine 6G was observed to localize in the mitochondria of isolated rat cardiomyocytes.
Conclusions
Based on these results, the 18F-labeled diethylene glycol ester of rhodamine 6G is the most promising potential PET MPI radiopharmaceutical of those that have been evaluated to date, and we are now preparing to carry out first-in-human clinical studies with this compound.
Keywords: Fluorine-18, rhodamine, myocardial perfusion imaging, PET
1. Introduction
Cardiovascular disease is a major health problem worldwide. Myocardial perfusion imaging (MPI) using non-invasive modalities such as SPECT (single-photon emission computer tomography) and PET (positron emission tomography) are the most widely used techniques for diagnosis and treatment planning in this disease. Currently, the primary method for MPI is SPECT using the single-photon emitters 99mTc-sestamibi, 99mTc-tetrofosmin, or 201Tl. However, the limitations of SPECT, including the absence of a standardized attenuation correction method, the inability to perform quantitative measurements, the lower spatial resolution and sensitivity compared to PET as well as recurring shortages of 99mTc have increased interest in PET MPI [1, 2].
Despite the advantages of PET for MPI, its widespread clinical use is hampered by the practical limitations of the currently available PET perfusion agents: [13N]NH3, [15O]H2O, and [82Rb]RbCl. The tracers [13N]NH3 and [15O]H2O are readily produced by accelerator methods, but the short half-lives of 13N (10 min) and 15O (2 min) restrict their use to clinical centers with an on-site cyclotron. Generator-produced 82Rb (t1/2 = 76 s) can be used at clinics without access to cyclotrons, but the high cost of the 82Sr/82Rb generator requires high patient throughput for the generator to be cost effective. Other limiting factors of 82Rb include its less than optimal myocardial extraction and its high positron energy, which results in decreased spatial resolution [1].
The limitations of the existing PET MPI agents and the near ideal physical properties of 18F (β+, 0.635 MeV [97%]; t1/2 = 110 min) have increased interest in the development of 18F-labeled myocardial perfusion tracers. An additional advantage of 18F is that the 110 min half-life is sufficiently long enough to allow centralized production of 18F-labeled radiopharmaceuticals and distribution to clinical centers without an on-site cyclotron, while still being short enough to allow repeated MPI studies of a patient on the same day.
In recent years, several 18F-labeled compounds have been evaluated as possible myocardial perfusion agents including quaternary ammonium salts [3]; tetraphenylphosphonium compounds [4–7]; rotenone derivatives [8, 9]; and pyridazinone analogs, such as BMS-747158-02 (Flurpiridaz F18; Lantheus Medical Imaging, Inc.) [10–17]. Flurpiridaz is currently in Phase 3 clinical trials [18–22], and BFPET (Fluoropharma) has recently completed phase 1 clinical trials [7, 23].
Neither of these 18F-labeled agents has, however, yet been approved by the FDA, and it is not certain that either, or both, will ultimately be approved. Given the clinical importance of MPI, the added clinical value of PET MPI, and the limitations of the existing PET MPI agents, the successful development of an effective 18F-labeled MPI radiopharmaceutical is essential to the care of patients with cardiovascular disease.
We recently reported the development of 18F-labeled esters of rhodamine B as potential MPI radiopharmaceuticals [24, 25]. The in vivo stability and pharmacokinetics of the 18F-labeled rhodamine B ethyl ester were, however, less than optimal resulting in unfavorable liver uptake. In a study of different rhodamine B esters where we sought to reduce the liver uptake and increase the myocardial uptake of the tracer, we found that the 18F-labeled diethylene glycol ester of rhodamine B ([18F]2) was superior to several other rhodamine B esters in terms of in vivo stability and pharmacokinetics [26]. In that study we also demonstrated the ability of [18F]2 to delineate myocardial infarction in a rat [26]. The objective of the present study was to compare the pharmacokinetic properties of several different 18F-labeled rhodamines dyes (i.e., rhodamine 6G, tetramethylrhodamine, and rhodamine 101, Fig. 1) and determine if they provided improved properties compared to [18F]2.
Figure 1.

Rhodamine dyes discussed in this study.
2. Methods
2.1. General
Rhodamine 6G chloride was obtained from Acros (Fair Lawn, NJ). Tetramethylrhodamine lactone, rhodamine 101 lactone, and tetrabutylammonium fluoride (1 M in THF) were purchased from Sigma-Aldrich (St. Louis, MO). Diethyleneglycol bistosylate was purchased from TCI America (Philadelphia, PA). For the radiosynthesis, extra dry reagent grade acetonitrile (Thermo Scientific, Bellefonte, PA) and Kryptofix (K 2.2.2, 98%, Sigma-Aldrich) were used. Potassium carbonate (99.97%) was purchased from Alfa Aesar (Ward Hill, MA). Other solvents and reagents were of the highest grade commercially available and used without further purification. Thin-layer chromatography (TLC) was performed using silica gel IB-F coated plastic sheets from J. T. Baker (Phillipsburg, NJ). Nuclear magnetic resonance spectra were obtained using a 400 MHz Varian 400-MR system (Palo Alto, CA). Chemical shifts are given as parts per million (ppm) and are reported relative to tetramethylsilane. Coupling constants are reported in hertz (Hz). The multiplicity of the NMR signals is described as follows: s = singlet, d = duplet, t = triplet, q = quartet, m = multiplet. High-resolution mass spectra (ESI-MS mode) were obtained at the University of Illinois Mass Spectrometry Facility using a Micromass 70-VSE spectrometer. Fluorine-18 (as F− in water) was purchased from Cardinal Healthcare (Woburn, MA) and the Brigham and Women’s Hospital BICOR (Boston, MA). The purity of the nonradioactive (19F) reference compounds was ≥95% as determined by analytical HPLC and NMR.
2.2. Purification and quality control
Analytical HPLC was carried out using a Hitachi 7000 system including an L-7455 diode array detector, an L-7100 pump, and a D-7000 interface. The radiometric HPLC detector was comprised of Canberra nuclear instrumentation modules and was optimized for 511 keV photons. An LaChrom PuroSphere Star C18e column (4 mm × 30 mm, 3 μm) was used for analytical measurements. The mobile phase consisted of 0.1% trifluoroacetic acid (TFA) in water (solvent A) and 0.1% TFA in acetonitrile (solvent B) at a flow rate of 1 mL/min at room temperature. The solvent gradient was 0–15 min (30–70% B), 15–25 min (70% B). For semi-preparative HPLC, an ISCO system comprised of an ISCO V4 variable wavelength UV-visible detector (operated at 550 nm), ISCO 2300 HPLC pumps, a radiometric detector similar to that described above, and a Grace Apollo C18 column (10 mm × 250 mm, 5 μm) was used. Semi-preparative HPLC method A (isocratic): 40% solvent A (0.1% TFA in water), 60% solvent B (0.1% TFA in acetonitrile); flow rate, 5 mL/min; room temperature. Semi-preparative HPLC method B (gradient): 0–10 min (40% B); 10–30 min (40–50% B); 30–35 min (50–100% B); 35–40 min (100% B); flow rate, 5 mL/min; room temperature. Radiofluorination yields were determined by thin-layer chromatography using silica gel plates and chloroform/methanol (8:1 v/v) as the solvent. After development, the TLC strips were cut into 1 cm pieces and counted with a Packard Cobra γ counter.
2.3. Synthesis of non-radiolabeled reference compounds
2.3.1. 2-(2-Fluoroethoxy)ethyl tosylate (1)
Compound 1 was prepared using the previously described procedure [26]. Yield: 262 mg (83%). 1H NMR (CDCl3): δ 7.80 (2H, d, J = 8.26), 7.35 (2H, d, J = 8.20), 4.55 (m, 1H), 4.43(m, 1H), 4.18 (t, 2H, J = 4.80), 3.71 (m, 3H), 3.63 (m, 1H), 2.45 (s, 3H).
2.3.2. Lactone precursors
The rhodamine 6G lactone precursor was prepared as previously described [27] and characterized by 1H NMR and mass spectrometry.
2.3.3. Tetramethylrhodamine 2-(2-fluoroethoxy)ethyl ester, TFA salt (3)
Tetramethylrhodamine lactone (20 mg, 0.05 mmol) was dissolved in 10 mL of acetonitrile, and the solution was heated to 80 °C before adding 40 mg (0.15 mmol) of 1 dissolved in 2 mL of acetonitrile containing 0.1 mL (0.6 mmol) of DIPEA. The mixture was refluxed for 3 h and allowed to cool to room temperature. The condenser was replaced with a serum stopper equipped with a venting needle and the mixture was heated to 90 °C, allowing the reaction mixture to evaporate slowly to complete dryness (approx. 7 h). HPLC analysis of the reaction mixture revealed incomplete conversion (crude yield 40%). The crude product was dissolved in 3 ml acetonitrile and filtered through a C-18 Sep-Pak cartridge to remove the unreacted lactone, which is retained on the cartridge. In order to obtain high purity reference material, a small fraction of the crude product was further purified by semi-preparative HPLC using the same separation conditions as for the radioactive compound (HPLC method B). Yield (crude): 6 mg (20%). 1H NMR (CDCl3): δ 8.33 (dd, 1H, J = 1.21, 7.84), 7.82-7.73 (m, 2H), 7.33 (dd, 1H, J1=7.34, J2=0.96), 7.10 (d, 2H, J=9.39), 6.94-6.88 (m, 4H), 4.52 (dt, 2H, J1 = 47.77, J2 = 4.03), 3.62–3.45 (m, 4H), 3.33 (s, 12H). 19F NMR (CDCl3): −74.7 ppm, (s, 3F) −222.9 ppm, (m, 1F). HRMS m/z (%): calcd for C32H38FN2O4+ [M+] 477.2190, found 477.2183 (100%).
2.3.4. Rhodamine 6G 2-(2-fluoroethoxy)ethyl ester, chloride salt (4)
This compound was prepared using the same method as described for 3 starting with 138 mg (0.33 mmol) of rhodamine 6G lactone. The crude yield was 70% (by HPLC). The crude product was dissolved in 3 mL acetonitrile and filtered through a C-18 Sep-Pak cartridge to remove the unreacted lactone. A high-purity sample was obtained by purifying a small amount of the crude product semi-preparative HPLC using the Method B. Yield: 89 mg (50%). 1H NMR (CDCl3, 400 MHz): δ 8.35 (dd, 1H, J = 1.34, 7.93), 7.82–7.72 (m, 4H), 7.29 (dd, 1H, J = 0.95, 7.42), 6.72 (s, 2H), 6.63 (s, 2H), 4.50 (dt, 2H, J1 = 47.71, J2 = 4.08), 4.15 (t, 2H, J = 4.71), 3.60–3.47 (m, 8H), 2.30 (s, 6H), 1.40 (t, 6H, J = 7.19). 19F NMR (CDCl3): −222.8 ppm (m, 1F). HRMS m/z (%): calcd for C32H38FN2O4+ [M+] 505.2497, found 505.2503 (100%).
2.3.5. Rhodamine 101 2-(2-fluoroethoxy)ethyl ester, tosyl salt (5)
This compound was prepared from the rhodamine 101 lactone (215 mg, 0.44 mmol) and initially purified using a C-18 Sep-Pak cartridge to remove unreacted lactone as described above for 3. A high-purity sample was obtained by C-18 column chromatography (22 mm × 350 mm) using 50% acetonitrile/water with 0.5% acetic acid as mobile phase. Yield (crude): 90 mg (30%). 1H NMR (CDCl3, 400 MHz): δ 8.33 (dd, 1H, J = 1.18, 7.84), 7.85–7.70 (m, 4H), 7.27 (dd, 1H, J = 0.98, 7.28), 7.08 (d, 2H, J = 8.02), 6.52 (s, 2H), 4.55 (dt, 2H, J1 = 47.76, J2 = 4.10), 4.20 (t, 2H, J = 4.63), 3.75–3.46 (m, 12H), 3.04 (m, 4H), 2.66 (t, 4H, J = 6.06), 2.28 (s, 3H), 2.09 (m, 6H), 1.96 (m, 3H). 19F NMR (CDCl3): −222.9 ppm, (m, 1F). HRMS m/z (%): calcd for C32H38FN2O4+ [M+] 581.2816, found 581.2815 (100%).
2.4. Synthesis of 18F-labeled rhodamines
The diethylene glycol esters [18F]3, [18F]4, and [18F]5 were prepared using the one-pot two-step procedure as previously described for [18F]2 [24–26]. Briefly, an aqueous [18F]fluoride solution was dried azeotropically in a 3 mL Pierce Reacti-Vail in the presence of K 2.2.2, K2CO3, and acetonitrile. Diethylene glycol ditosylate in 0.5 mL acetonitrile was added to the dried residue and the mixture was heated at 90° C for 10 min. After rapid cooling to room temperature, a solution of the rhodamine lactone in 0.8 mL acetonitrile and DIPEA (3–5 drops) were added to the vial. Further heating at 160° C for 30 min using a 30G ventilation needle gave the crude 18F-labeled rhodamine, which was further purified by semi-preparative HPLC. For each compound, HPLC fractions containing the radioactive product were combined and dried under a stream of nitrogen at 65 °C. The radioactive product was then redissolved in either 10% EtOH/water for the partition coefficient measurements or 10% EtOH/saline for animal experiments. The integrity and purity of the final radioactive product was confirmed by analytical HPLC.
2.5. Partition coefficient (logD) measurements
The logD values for the compounds were determined as described previously [26]. Briefly, 3.7–5.6 MBq (100–150 μCi) of [18F]3, [18F]4, or [18F]5 in 20 μL 10% EtOH/water was added to a mixture of 480 μL phosphate-buffered saline (pH 7.4, PBS) and 500 μL octanol in 1.5 mL Eppendorf tubes. Samples were vortexed for 1 min and then centrifuged at 13,200 rpm for 5 min. From each sample, three 100-μL samples of both the PBS and the organic layer were transferred into plastic tubes, and the samples were counted with a Packard Cobra gamma counter. The experiments were performed in triplicate.
2.6. Small-animal PET imaging studies
All in vivo experiments were carried out in rats because of the previously observed instability of [18F[2 in mouse serum [24]. All animal studies were carried out under a protocol approved by the Boston Children’s Hospital Institutional Animal Care and Use Committee. Before injection, the solution was filtered through a sterilized centrifugal filter (0.2 μm). For PET imaging studies, animals were injected with 100 μL of [18F]3, [18F]4, or [18F]5 (3.7–5.6 MBq, 100–150 μCi) and anesthetized with isoflurane (2–4% in air). Data acquisition was initiated as quickly as possible after injection of the tracer. Imaging was performed using a Siemens Focus 120 MicroPET scanner. Data were acquired for 60 min in list mode and reconstructed into either a single 60 min image or six 10 min frames. Reconstruction was performed using unweighted OSEM2D generating an image with a volume of 128×128×95 voxels (0.866×0.866×0.796 mm3). Image analysis was performed using the ASIPro software package (Siemens Medical Solutions). Time-activity curves (TACs) were constructed by manually drawing regions of interest (ROI) within the left ventricular myocardium, the chest, and medially in the liver. All ROIs were then copied on each of the frames, and time-activity curves of the ROI mean values were generated.
2.7. Biodistribution studies
For each time point (15, 30, 60, 120 min), 3–5 animals were injected with 1.1–2.2 MBq (30–60 μCi) of the 18F-labeled rhodamine in 100 μL 10% EtOH/saline via the tail vein, sacrificed by CO2 asphyxia at the appropriate time, and weighed. Selected tissues were then excised, weighed and assayed for 18F. The percent injected dose per gram (% ID/g) for each tissue was calculated by comparison of the tissue counts to a standard sample prepared from the injectate.
As part of the biodistribution study, a urine sample was obtained at 60 min post-injection and analyzed by HPLC to determine the identity of the excreted material.
2.8. Data Analysis
Statistical analysis was performed with Prism software (GraphPad Software Inc., La Jolla, CA) using a 2-tailed t test.
2.9. Cardiomyocyte uptake and fluorescent microscopy studies
Cardiomyocyte uptake and fluorescent microscopy studies were performed as previously described [26]. Briefly, ventricular cardiomyocytes were isolated from 2-day-old Lewis rat pups using the Neonatal Cardiomyocyte Isolation System (LK003300, Worthington Biochemical Corporation, Lakewood, NJ). For the cellular uptake experiments, [18F]4 was dissolved in serum-free medium to give a concentration of ~37 kBq (1 μCi)/μL at the beginning of the experiment. Six time points were used for uptake experiments: 0, 1, 5, 10, 30, and 60 minutes. After the appropriate incubation time, the supernatant was transferred into a separate tube for subsequent assay; the cardiomyocytes were washed with PBS, which was added to the supernatant; and the cells were harvested by addition of 1 N NaOH. The vials containing the combined supernatant and PBS washing solution and the harvested cells were assayed in a gamma counter to determine percent tracer uptake over time. The cell uptake experiments were performed in triplicate with three different batches of [18F]4.
For fluorescent microscopy imaging, cells were stained with either MitoTracker Green FM (M-7514, Life Technologies, Grand Island, NY) in combination with 4′,6-diamidino-2-phenylindole (DAPI) (Life Technologies) according to the manufacturer’s directions after incubation with the 18F-labeled rhodamine compound, and then mounted on slides and imaged using an FSX100 microscope (Olympus, Waltham, MA).
3. Results and Discussion
We synthesized three new 18F-labeled rhodamine dyes (rhodamine 6G, tetramethylrhodamine, and rhodamine 101) and compared their pharmacokinetic properties to those of rhodamine B ([18F]2) to determine if, by changing the rhodamine core, we could further reduce the liver concentration and/or increase the rate of clearance from the liver while simultaneously maintaining or increasing tracer accumulation in the heart.
3.1. Synthesis of the non-radioactive (19F) compounds
We prepared the non-radioactive compounds for use as reference materials for the HPLC characterization of the no-carrier-added 18F compounds. These reference compounds were characterized by HRMS and proton and fluorine NMR. The crude 19F compounds were purified by semi-preparative HPLC using the same elution conditions used for the purification of the 18F compounds. The tosyldiethyleneglycol fluoride prosthetic group was prepared in good yield (80%) from the bistosylate and tetrabutylammonium fluoride as described previously.[26] Reaction of diethylenglycol fluorotosylate with the rhodamine lactones in acetonitrile solution in the presence of DIPEA gave the corresponding esters in approximately 25% isolated yield.
3.2. Synthesis of 18F-labeled rhodamine esters
The 18F-labeled diethyleglycol esters of 18F-tetramethylrhodamine ([18F]3), rhodamine 6G ([18F]4), rhodamine 101 ([18F]5) were prepared using a procedure similar to that previously described for the 18F-labeled diethyleneglycol ester of rhodamine B ([18F]2, Fig. 2) [26]. The identities of the 18F-labeled rhodamines were confirmed by analytical HPLC using the non-radioactive compounds as reference materials. The total synthesis time was 120 min and the products were obtained in >97% radiochemical purity with decay-corrected yields of 18±1%, 19±1%, and 22±3% for [18F]3, [18F]4, and [18F]5, respectively. The specific activities of [18F]3, [18F]4, and [18F]5 are all on the order of 310–320 MBq/μmol (8.4–8.6 mCi/μmol) (Table 1). The partition coefficients for the compounds were similar (1.90 to 1.97). These results are summarized in Table 1.
Figure 2.
Radiosyntheses of the rhodamine esters: a) K2CO3/K 2.2.2; MeCN, 90°C, 10 min. b) DIPEA, MeCN, 165°C, 30 min.
Table 1.
Specific activities and logD values for the three new rhodamine derivatives in comparison to the previously described rhodamine B ester [18F]2.
| Compound | Rhodamine | LogD | Specific Activity MBq/μmol (mCi/μmol) |
|---|---|---|---|
| [18F]2 | B | 1.90 ± 0.01 | 310 (8.4) |
| [18F]3 | Me4 | 1.90 ± 0.01 | 310 (8.4) |
| [18F]4 | 6G | 1.96 ± 0.05 | 320 (8.6) |
| [18F]5 | 101 | 1.97 ± 0.01 | 320 (8.6) |
3.3. Small-animal PET imaging
Representative 60-min (summed) small-animal PET images obtained with [18F]3, [18F]4, and [18F]5 are shown in Figure 3. These images clearly show high and persistent retention in the myocardium and high contrast between the myocardium and blood, liver, bones and lungs for [18F]4 while [18F]3 and [18F]5 show lower uptake in the myocardium and higher uptake in the liver. The heart is much more clearly defined with [18F]4 than with [18F]3 and [18F]5, and the liver uptake is much lower.
Figure 3.
Coronal (top), sagittal (middle), and transverse (bottom) microPET images of [18F]4 (A), [18F]3 (B), and [18F]5 (C). Animals were injected with (3.7–5.6 MBq, 100–150 μCi) of [18F]3, [18F]4, or [18F]5, anesthetized with isoflurane (2–4% in air), and data were acquired for 60 min.
The microPET images also reveal high accumulation of 18F in the neck, as was observed for [18F]2, possibly representing the parathyroid or salivary gland accumulation, and in the muscles of the forelimbs. The bone uptake in all cases is minimal, with the skeleton being barely visible, suggesting minimal in vivo defluorination.
Time-activity curves for [18F]4 are shown in Figure 4. These curves show that, for [18F]4, the initial difference (at 5 min p.i.) between the heart and the liver is much greater than was previously seen for [18F]2 and the liver clearance is faster. Similarly to [18F]2, the heart uptake of [18F]4 remains constant through the study period with no washout. Thus, for [18F]4, the heart concentration (69 ± 5 kBq/cm3) at 5 min p.i. is somewhat higher than that in the liver (59 ± 2 kBq/cm3) resulting in a heart-to-liver ratio of 1.2. Since the tracer rapidly washes out of the liver but not the heart, at 35 min p.i., the heart-to-liver ratio for [18F]4 increases to 4.2. At 55 min p.i., the liver concentration of [18F]4 decreases to 25% of the 5 min value and the heart-to-liver ratio increases to approximately 4.5.
Figure 4.
Time-activity curves for [18F]rhodamine 6G ([18F]4) derived from small-animal PET images.
3.4. Biodistribution Studies
Due to the lower myocardial uptake and high liver uptake of [18F]3 and [18F]5 in the microPET imaging study, a biodistribution study was only carried out for [18F]4. The results of this study are summarized in Table 2.
Table 2.
Biodistribution of radioactivity after injection of [18F]4 into ratsa
| Tissue | Compound
|
||||
|---|---|---|---|---|---|
| [18F]4
|
30 min | 60 min | 120 min | [18F]2 b
|
|
| 15 min | 60 min | ||||
| Blood | 0.13 ± 0.03 | 0.07 ± 0.01 | 0.07 ± 0.01 | 0.06 ± 0.01 | 0.25 ± 0.02 |
| Heart | 1.31 ± 0.46 | 1.32 ± 0.06 | 1.32 ± 0.29 | 1.77 ± 0.15 | 2.51 ± 0.16 |
| Lung | 1.08 ± 0.34 | 0.58 ± 0.14 | 0.58 ± 0.19 | 0.82 ± 0.19 | 2.06 ± 0.16 |
| Liver | 1.01 ± 0.14 | 0.15 ± 0.11 | 0.15 ± 0.02 | 0.11 ± 0.01 | 0.49 ± 0.04 |
| Spleen | 3.16 ± 1.30 | 2.01 ± 0.17 | 2.01 ± 0.61 | 1.86 ± 0.41 | 3.53 ± 0.57 |
| Kidney | 10.2 ± 3.7 | 7.90 ± 1.05 | 7.90 ± 0.99 | 8.21 ± 1.22 | 9.28 ± 0.66 |
| Gut | 3.83 ± 3.50 | 1.39 ± 0.78 | 1.39 ± 0.58 | 1.07 ± 0.28 | 1.69 ± 0.42 |
| Skin | 0.13 ± 0.04 | 0.15 ± 0.02 | 0.15 ± 0.04 | 0.16 ± 0.04 | 0.36 ± 0.04 |
| Muscle | 0.10 ± 0.02 | 0.12 ± 0.01 | 0.12 ± 0.03 | 0.14 ± 0.02 | 0.54 ± 0.07 |
| Bone | 0.30 ± 0.08 | 0.29 ± 0.03 | 0.29 ± 0.05 | 0.37 ± 0.05 | 0.77 ± 0.04 |
| Thyroid/Trachea | 1.23 ± 0.47 | 1.05 ± 0.34 | 1.05 ± 0.24 | 1.08 ± 0.12 | - |
| Mandibular gland | 1.74 ± 0.50 | 2.01 ± 0.26 | 2.01 ± 0.31 | 2.31 ± 0.54 | - |
| Heart/blood c | 10.1 | 18.9 | 18.9 | 29.5 | 10.0 |
| Heart/lung c | 1.2 | 2.3 | 2.3 | 2.2 | 1.2 |
| Heart/liver c | 1.3 | 8.8 | 12.0 | 16.1 | 5.1 |
| Heart/bone c | 4.4 | 4.6 | 4.6 | 4.8 | 3.3 |
As was seen in the small-animal PET images, the change from rhodamine B to rhodamine 6G leads to a significantly improved pharmacokinetic profile. The heart concentration of [18F]4 reaches a plateau of 1.3% ID/g after 15 min and remains essentially constant over the entire 120 min study period with no washout of the tracer over that time. The absolute uptake by the heart is, however, somewhat lower than that seen for [18F]2 (2.5% ID/g). On the other hand, the concentration of [18F]4 in non-target tissues is significantly lower than that of [18F]2. For example, at 60 min p.i. the liver uptake of [18F]4 is 0.15% ID/g compared to 0.49% ID/g for [18F]2. Similarly, at 60 min p.i. the 18F concentration in the blood for [18F]4 is 0.07% ID/g (vs. 0.25% ID/g for [18F]2) and the 18F concentration in the lungs is 0.58% ID/g (vs. 2.06% ID/g for [18F]2). Results for other tissues follow a similar pattern. These results translate into much improved heart-to-non-target-tissue ratios. For example, at 60 min p.i. the heart-to-liver ratio for [18F]4 is 12.0 compared to 5.1 for [18F]2, and this value steadily increased throughout the 2-h study period, from 1.3 at 15 min p.i. to 16.1 at 2 h p.i. The heart-to-blood ratio also increased throughout the study period, from 10.1 at 15 min p.i. to 29.5 at 2 h p.i.
The concentration of the tracer in the kidneys remained relatively high and constant throughout the study, ranging from 10.2% at 15 min p.i. to 8.2% ID/g at 2 h while the concentration in the gut varied from 3.8% ID/g at 15 min to 1.0% ID/g at 2 h. Together these results suggest primarily renal excretion, in agreement with the small-animal PET images that show significant tracer accumulation in the bladder.
Analysis of urine samples (n=3) obtained at 60 min post-injection showed 79±7% of the material to be intact [18F]4, which suggests that the compound is sufficiently stable for use in myocardial perfusion imaging.
There is significant accumulation of [18F]4 in both the thyroid/parathyroid/trachea and the mandibular salivary gland. The former is presumably due to accumulation of the compound in the parathyroid, as is seen with other lipophilic cationic compounds such as 99mTc-MIBI and 99mTc-tetrofosmin [28]. These tissues were collected because of the high uptake observed in the neck on the small-animal PET scans. As is seen with the cardiac uptake, the concentration of the tracer in these tissues remains fixed throughout the 2-h measurement period. If the uptake in the parathyroid is, in fact, [18F]4, this suggests that it might also be useful for parathyroid imaging, similarly to 99mTc-MIBI [28]. Further studies are necessary to confirm exactly which these tissue(s) in the region actually accumulate 18F and whether or not the accumulation reflects intact [18F]4.
There is also accumulation of tracer in the spleen, which decreases slowly throughout the study. The concentration is, however, about 40% less than was seen for [18F]2 (2.01±0.61% ID/g vs. 3.53±0.57% ID/g at 60 min p.i.). This result and the lung uptake, which is also lower than was observed for [18F]2 (0.58±0.19% ID/g vs. 2.06±0.16% ID/g), may both be due to aggregation of the compound because of the relatively low specific activity (Table 1): Rhodamines are known to aggregate even at submicromolar concentrations [29]. The low specific activity (310–320 MBq/μmol) is primarily the result of the small amount of 18F used in these syntheses (typically 10 mCi), and pilot studies with larger amounts of 18F show a proportionally higher value for the specific activity. However, as can be seen from the small-animal PET images, the low specific activity does not interfere with visualization of the heart. Finally, it is worth noting that there is minimal tracer uptake in the bone (~0.3% ID/g) which does not increase during the study suggesting that there is no appreciable loss of 18F from the compound.
The biodistribution of [18F]4 may be compared to that of other 18F-labeled myocardial perfusion agents currently under development. Compared to BMS-747158-02 (Flurpiridaz F18; Lantheus Medical Imaging, Inc.), which is currently in late-stage clinical trials, the heart concentration of [18F]4 at 60 min p.i. (1.77±0.15% ID/g) is similar to the 2.06±0.83% ID/g observed in a study in Wistar rats (p = 0.51) and somewhat lower than the 3.3±0.3% ID/g (p < 0.001) observed in Sprague-Dawley rats [11, 12, 30]. On the other hand, the liver concentration of BMS-747158-02 is somewhat higher and the clearance from the liver is somewhat slower than that of [18F]4. The liver concentration of BMS-747158-02 decreases from 1.27±0.10% ID/g to 0.63±0.16% ID/g from 30 min to 120 min p.i. whereas the liver concentration of [18F]4 decreases from 0.14±0.09% ID/g to 0.05±0.02% ID/g in the same time frame. This is also reflected by a comparison of the heart-to-liver ratios. From 30 min to 120 min p.i., the heart-to-liver ratio of [18F]4 increases from 4.1 to 13.4 in contrast to BMS-747158-02 where the corresponding ratio only increases from 2.7 to 5.4 [12]. However, the heart-to-liver ratio for [18F]4 at 60 min p.i. (6.0) is similar to that of BMS-747158-02 (5.4 in Wistar rats and 3.7 in Sprague-Dawley rats) [11, 30]. In contrast, the heart-to-blood ratio of 19.3 for [18F]4 at 30 mins p.i., while quite high, is somewhat lower than that of BMS-747158-02 (30.0) [12].
Another 18F-labeled myocardial perfusion tracer currently in clinical trials is BFPET (Fluoropharma), an 18F-labeled analog of the tetraphenylphosphonium cation [7, 23]. This compound exhibits heart uptake similar to that of [18F]4. For example, the heart uptake of BFPET in Sprague-Dawley rats is 1.51±0.04% ID/g at 30 min, slightly higher than for [18F]4 (1.32±0.06% ID/g, p = 0.0007), and 1.57±0.18% ID/g at 60 min p.i., equivalent to that result for [18F]4 (1.77±0.15% ID/g, p = 0.1189) [7]. The concentration of the two compounds in the liver is also very similar. At 30 min p.i. the liver concentration of BFPET is 0.18±0.05% ID/g versus 0.15±0.11% ID/g for [18F]4, and at 60 min p.i. the liver concentration of BFPET is 0.17±0.03% ID/g compared to 0.15±0.02% ID/g for [18F]4.
3.5. Cardiomyocyte Studies
A unique aspect of the rhodamine-based tracers is that rhodamines are inherently fluorescent making 18F-labeled rhodamines intrinsic dual-modality imaging agents [31]. This allows the measurement of the uptake of the compound by cells by assaying the 18F content while at the same time the intracellular localization of the tracer can be directly observed without the need for any additional interventions.
In order to determine if [18F]4 accumulates in the mitochondria in the same manner as does non-radiolabeled rhodamine 6G [32] and 18F-labeled rhodamine B diethylene glycol ester [18F]2 [26], fluorescence microscopy experiments and cellular uptake studies were performed in rat cardiomyocytes. Figure 5 shows a representative fluorescence microscopy image of cardiomyocytes incubated with [18F]4. Figure 5A shows the mitochondria stained in green (MitoTracker Green) while figure 5B shows the same cells stained in red with [18F]4. Figure 5C shows the nuclei stained in blue (DAPI). In agreement with the results obtained for [18F]2, [18F]4 localizes in the mitochondria of rat cardiomyocytes as shown in the overlaid image in Figure 5D (Pearson correlation coefficient 0.85).
Figure 5.
Fluorescence microscopy images of rat cardiomyocytes showing mitochondria stained by MitoTracker green dye (A), [18F]4 (B), nuclei stained by DAPI (C), and the overlay (D). The tracer [18F]4 clearly co-localizes in the mitochondria of rat cardiomyocytes (Pearson correlation coefficient 0.85).
The uptake of [18F]4 in rat cardiomyocytes were measured as previously described for [18F]2 [26] and found to be similar (Figure 6). The 18F-labeled rhodamine 6G diethylene glycol ester, [18F]4, internalizes rapidly into rat cardiomyocytes with 4.6±0.7% of the initial activity internalized within 1 min, similar to the results for the rhodamine B derivative, [18F]2 (3.9±0.8% at 1 min). At 60 min, 55.0±6.8% of the initial activity of [18F]4 is internalized, not significantly higher than that of the rhodamine B derivative [18F]2 (43.1±7.7%, p = 0.1153). Incubation times exceeding 60 min resulted in variable results due to the experimental stress on the isolated cardiomyocytes, so the study could not be continued until an uptake plateau was reached.
Figure 6.
Time-dependent uptake of [18F]4 into rat cardiomyocytes. Error bars indicate estimated standard deviation.
One additional consideration in the in vivo evaluation of radiolabeled rhodamine dyes is that rhodamine 6G is a substrate of Pgp, the protein implicated in MDR1 drug resistance in cancer [33]. While this is not a factor in their evaluation as myocardial perfusion agents, it is an important consideration in their future evaluation as tumor imaging agents. Preliminary studies suggest that [18F]4 is, in fact a substrate for Pgp [34], and studies are currently being carried out to more completely evaluate this possible application.
4. Conclusions
This comparison of several different 18F-labeled rhodamines as potential myocardial perfusion agents demonstrated that 18F-labeled rhodamine 6G ([18F]4) is superior to the first generation 18F-labeled rhodamine tracer, rhodamine B ([18F]4), primarily in terms of its superior contrast between the heart and the liver. As was the case with all of the 18F-labeled rhodamine dyes that have been evaluated to date, the uptake by the heart is rapid and does not decrease over time. In contrast to other 18F-labeled rhodamines, the liver uptake of 18F-labeled rhodamine 6G is both initially lower and decreases more rapidly. As was the case with 18F-labeled rhodamine B, 18F-labeled rhodamine 6G localizes in the mitochondria of isolated rat cardiac myocytes. In combination, these results strongly suggest that 18F-labeled rhodamine 6G is a promising candidate for more extensive evaluation as a PET MPI tracer.
Acknowledgments
These studies were supported by the Children’s Hospital Radiology Foundation and NIH grant # 1 R01 HL108107-03 (ABP).
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Bengel FM, Higuchi T, Javadi MS, Lautamaki R. Cardiac positron emission tomography. J Am Coll Cardiol. 2009;54:1–15. doi: 10.1016/j.jacc.2009.02.065. [DOI] [PubMed] [Google Scholar]
- 2.Machac J. Cardiac positron emission tomography imaging. Semin Nucl Med. 2005;35:17–36. doi: 10.1053/j.semnuclmed.2004.09.002. [DOI] [PubMed] [Google Scholar]
- 3.Studenov AR, Berridge MS. Synthesis and properties of 18F-labeled potential myocardial blood flow tracers. Nucl Med Biol. 2001;28:683–93. doi: 10.1016/s0969-8051(01)00233-5. [DOI] [PubMed] [Google Scholar]
- 4.Ravert HT, Madar I, Dannals RF. Radiosynthesis of 3-[18F]fluoropropyl and 4-[18F]fluorobenzyl triarylphosphonium ions. J. Labelled Compd. Radiopharm. 2004;47:469–476. [Google Scholar]
- 5.Madar I, Ravert HT, Du Y, Hilton J, Volokh L, Dannals RF, et al. Characterization of uptake of the new PET imaging compound 18F-fluorobenzyl triphenyl phosphonium in dog myocardium. J Nucl Med. 2006;47:1359–66. [PubMed] [Google Scholar]
- 6.Madar I, Ravert H, Dipaula A, Du Y, Dannals RF, Becker L. Assessment of severity of coronary artery stenosis in a canine model using the PET agent 18F-fluorobenzyl triphenyl phosphonium: comparison with 99mTc-tetrofosmin. J Nucl Med. 2007;48:1021–30. doi: 10.2967/jnumed.106.038778. [DOI] [PubMed] [Google Scholar]
- 7.Shoup TM, Elmaleh DR, Brownell AL, Zhu A, Guerrero JL, Fischman AJ. Evaluation of (4-[(18)F]Fluorophenyl)triphenylphosphonium Ion. A Potential Myocardial Blood Flow Agent for PET. Mol Imaging Biol. 2011;13:511–7. doi: 10.1007/s11307-010-0349-2. [DOI] [PubMed] [Google Scholar]
- 8.Martarello L, Greenamyre JT, Goodman MM. Synthesis and evaluation of a new fluorine-18 labeled rotenoid as a potential PET probe of mitochondrial complex I activity. J Labelled Compd Radiopharm. 1999;42:1039–1051. [Google Scholar]
- 9.Marshall RC, Powers-Risius P, Reutter BW, O’Neil JP, La Belle M, Huesman RH, et al. Kinetic analysis of 18F-fluorodihydrorotenone as a deposited myocardial flow tracer: comparison to 201Tl. J Nucl Med. 2004;45:1950–9. [PubMed] [Google Scholar]
- 10.Yalamanchili P, Wexler E, Hayes M, Yu M, Bozek J, Kagan M, et al. Mechanism of uptake and retention of F-18 BMS-747158-02 in cardiomyocytes: a novel PET myocardial imaging agent. J Nucl Cardiol. 2007;14:782–8. doi: 10.1016/j.nuclcard.2007.07.009. [DOI] [PubMed] [Google Scholar]
- 11.Huisman MC, Higuchi T, Reder S, Nekolla SG, Poethko T, Wester HJ, et al. Initial characterization of an 18F-labeled myocardial perfusion tracer. J Nucl Med. 2008;49:630–6. doi: 10.2967/jnumed.107.044727. [DOI] [PubMed] [Google Scholar]
- 12.Purohit A, Radeke H, Azure M, Hanson K, Benetti R, Su F, et al. Synthesis and Biological Evaluation of Pyridazinone Analogues as Potential Cardiac Positron Emission Tomography Tracers. J Med Chem. 2008;51:2954–2970. doi: 10.1021/jm701443n. [DOI] [PubMed] [Google Scholar]
- 13.Yu M, Guaraldi M, Kagan M, Mistry M, McDonald J, Bozek J, et al. Assessment of 18F-labeled mitochondrial complex I inhibitors as PET myocardial perfusion imaging agents in rats, rabbits, and primates. Eur J Nucl Med Mol Imaging. 2009;36:63–72. doi: 10.1007/s00259-008-0909-8. [DOI] [PubMed] [Google Scholar]
- 14.Yu M, Bozek J, Guaraldi M, Kagan M, Azure M, Robinson SP. Cardiac imaging and safety evaluation of BMS747158, a novel PET myocardial perfusion imaging agent, in chronic myocardial compromised rabbits. J Nucl Cardiol. 2010;17:631–6. doi: 10.1007/s12350-010-9221-7. [DOI] [PubMed] [Google Scholar]
- 15.Mou T, Jing H, Yang W, Fang W, Peng C, Guo F, et al. Preparation and biodistribution of [18F]FP2OP as myocardial perfusion imaging agent for positron emission tomography. Bioorg Med Chem. 2010;18:1312–20. doi: 10.1016/j.bmc.2009.12.022. [DOI] [PubMed] [Google Scholar]
- 16.Sherif HM, Nekolla SG, Saraste A, Reder S, Yu M, Robinson S, et al. Simplified quantification of myocardial flow reserve with flurpiridaz F 18: validation with microspheres in a pig model. J Nucl Med. 2011;52:617–24. doi: 10.2967/jnumed.110.083196. [DOI] [PubMed] [Google Scholar]
- 17.Woo J, Tamarappoo B, Dey D, Nakazato R, Le Meunier L, Ramesh A, et al. Automatic 3D registration of dynamic stress and rest (82)Rb and flurpiridaz F 18 myocardial perfusion PET data for patient motion detection and correction. Med Phys. 2011;38:6313–26. doi: 10.1118/1.3656951. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Yu M, Nekolla SG, Schwaiger M, Robinson SP. The next generation of cardiac positron emission tomography imaging agents: discovery of flurpiridaz F-18 for detection of coronary disease. Semin Nucl Med. 2011;41:305–13. doi: 10.1053/j.semnuclmed.2011.02.004. [DOI] [PubMed] [Google Scholar]
- 19.Rischpler C, Park MJ, Fung GS, Javadi M, Tsui BM, Higuchi T. Advances in PET myocardial perfusion imaging: F-18 labeled tracers. Ann Nucl Med. 2012;26:1–6. doi: 10.1007/s12149-011-0552-5. [DOI] [PubMed] [Google Scholar]
- 20.Maddahi J. Properties of an ideal PET perfusion tracer: new PET tracer cases and data. J Nucl Cardiol. 2012;19(Suppl 1):S30–7. doi: 10.1007/s12350-011-9491-8. [DOI] [PubMed] [Google Scholar]
- 21.Berman DS, Germano G, Slomka PJ. Improvement in PET myocardial perfusion image quality and quantification with flurpiridaz F 18. J Nucl Cardiol. 2012;19(Suppl 1):S38–45. doi: 10.1007/s12350-011-9487-4. [DOI] [PubMed] [Google Scholar]
- 22.Berman DS, Maddahi J, Tamarappoo BK, Czernin J, Taillefer R, Udelson JE, et al. Phase II safety and clinical comparison with single-photon emission computed tomography myocardial perfusion imaging for detection of coronary artery disease: flurpiridaz F 18 positron emission tomography. J Am Coll Cardiol. 2013;61:469–77. doi: 10.1016/j.jacc.2012.11.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Elmaleh D, Kardan A, Barrow S, Dragotakes S, Correia J, Weise S, et al. A phase I study evaluating dosimetry and myocardial pharmacokinetic behavior of BFPET, a new F-18 labeled tracer for myocardial perfusion imaging. J Nucl Med Meeting Abstracts. 2009;50:420. [Google Scholar]
- 24.Heinrich TK, Gottumukkala V, Snay E, Dunning P, Fahey FH, Ted Treves S, et al. Synthesis of fluorine-18 labeled rhodamine B: A potential PET myocardial perfusion imaging agent. Appl Radiat Isot. 2010;68:96–100. doi: 10.1016/j.apradiso.2009.08.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Gottumukkala V, Heinrich TK, Baker A, Dunning P, Fahey FH, Treves ST, et al. Biodistribution and stability studies of [18F]fluoroethylrhodamine B, a potential PET myocardial perfusion agent. Nucl Med Biol. 2010;37:365–70. doi: 10.1016/j.nucmedbio.2009.12.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Bartholoma MD, Gottumukkala V, Zhang S, Baker A, Dunning P, Fahey FH, et al. Effect of the prosthetic group on the pharmacologic properties of 18F-labeled rhodamine B, a potential myocardial perfusion agent for positron emission tomography (PET) J Med Chem. 2012;55:11004–12. doi: 10.1021/jm301453p. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Afonso CAM, Santhakumar V, Lough A, Batey RA. An Expedient Synthesis of Cationic Rhodamine Fluorescent Probes Suitable for Conjugation to Amino Acids and Peptides. Synthesis. 2003;2003:2647, 2654. [Google Scholar]
- 28.Palestro CJ, Tomas MB, Tronco GG. Radionuclide imaging of the parathyroid glands. Semin Nucl Med. 2005;35:266–76. doi: 10.1053/j.semnuclmed.2005.06.001. [DOI] [PubMed] [Google Scholar]
- 29.Lopez AI, Ruiz OP. Dimeric states of Rhodamine B. Chem Phys Lett. 1982;87:556–60. [Google Scholar]
- 30.Yu M, Guaraldi MT, Mistry M, Kagan M, McDonald JL, Drew K, et al. BMS-747158-02: a novel PET myocardial perfusion imaging agent. J Nucl Cardiol. 2007;14:789–98. doi: 10.1016/j.nuclcard.2007.07.008. [DOI] [PubMed] [Google Scholar]
- 31.Beija M, Afonso CA, Martinho JM. Synthesis and applications of Rhodamine derivatives as fluorescent probes. Chem Soc Rev. 2009;38:2410–33. doi: 10.1039/b901612k. [DOI] [PubMed] [Google Scholar]
- 32.Lacerda SH, Abraham B, Stringfellow TC, Indig GL. Photophysical photochemical, and tumor-selectivity properties of bromine derivatives of rhodamine-123. Photochem Photobiol. 2005;81:1430–8. doi: 10.1562/2005-08-05-RA-639. [DOI] [PubMed] [Google Scholar]
- 33.Matsumoto Y, Sasaoka N, Tsuchida T, Fujiwara T, Nagao S, Ohmoto T. Fluorescent dye rhodamine 6G as a molecular probe to study drug resistance of C6 rat glioma cells. Journal of Neuro Oncology. 1992;13:217–22. doi: 10.1007/BF00172473. [DOI] [PubMed] [Google Scholar]
- 34.Akurathi V, Zhang S, Dearling J, Treves ST, Packard A. Preliminary evaluation of 18F-rhodamine 6G as a tumor-imaging agent. J Nucl Med Meeting Abstracts. 2013;54:1148. [Google Scholar]





