Abstract
Purpose
In this work the feasibility of using an array of electric dipole antennas for RF transmission in spine MRI at high field is explored.
Method
A 2-channel transmit array based on an electric dipole design was quantitatively optimized for 7T spine imaging and integrated with a receive array combining 8 loop coils. Using B1+ mapping, the transmit efficiency of the dipole array was compared to a design using quadrature loop pairs. The radio-frequency (RF) energy deposition for each array was measured using a home-built dielectric phantom and MR thermometry. The performance of the proposed array was qualitatively demonstrated in human studies.
Results
The results indicate dramatically improved transmit efficiency for the dipole design as compared to the loop excitation. Up to 76% gain was achieved within the spinal region.
Conclusion
For imaging of the spine, electric-dipole based transmitters provided an attractive alternative to the traditional loop-based design. Easy integration with existing receive array technology facilitates practical use at high field.
Keywords: High field MRI, dipole antenna, spinal imaging, transmit array, receive array
INTRODUCTION
MRI of the spine at 3T has been found to offer improvement over 1.5T in terms of signal-to-noise ratio (SNR) and tissue delineation, but with new challenges related to B1 and B0 inhomogeneity (1). MRI of the spine at 7T is hampered by a lack of commercial radio-frequency (RF) receive coils and a body transmit coil. Development of 7T spine arrays therefore requires implementation of both transmit and receive functions in the same coil assembly, either via transceiver elements or via separate transmit and receive elements. In the past few years, several groups have proposed various designs (2-7) utilizing loop-based transmit structures. However, given the generally large field-of-view (FOV) required for spine imaging in comparison with the RF wavelength at 7T (e.g. ~13 cm in muscle), loop-based transmit elements generally require a rather complicated layout and cabling (2-5). Although it is possible to simplify RF excitation by exploiting the traveling wave phenomenon (e.g. by employing a patch antenna at magnet bore entrance (8,9)), this may result in reduced transmit efficiency when compared with transmitters placed close to the area of interest (10).
In recent studies, use of electric dipole antennas has been explored as an alternative mode of RF transmission at high field (11-18), mostly for small FOV imaging. For head or prostate imaging at 7T, the antennas were shortened to ~20 cm to avoid unnecessary RF deposition outside the region-of-interest (ROI). However, when the conventional “half-wave dipole” (19) is constructed only with conductor strips and without additional lumped elements such as inductors, its length is well-suited for the larger superior-inferior (S-I) FOV required for spine imaging. In one application, the electric dipole was used in a so-called “dark mode” (i.e. only generating longitudinal B1+ field, which is invisible in MRI) to modify the electric field distribution for a cervical spine array based on loop designs (20). But for spine imaging at 7T, the use of electrical dipoles for B1 transmission has not been reported. In this work, the feasibility of using electrical dipole antennas is explored by evaluating both transmit efficiency and energy deposition, relative to a conventional loop-based design (4).
METHODS
Transmit and receive elements of our spine array were divided between a top layer immediately under the patient surface housing the receive-only loop array and a bottom layer located 4 cm below the patient housing the transmit-only dipole array.
Design and Optimization of the Electric Dipole Array
Our transmit-only dipole array (Figure 1a) employed two transmit elements, each consisting of a conventional half-wave dipole. The dipoles were constructed with straight 25 mm wide copper tape and fed at their centers with RF. For each of the two segments of each dipole element, the copper strips mounted on a 5 mm thick acrylic board. The dipoles were positioned 4 cm below the patient surface to allow space for the receive array. As expected, each dipole was found to have a resonance with a wavelength corresponding to twice the total dipole length. The resonant frequency of the dipole shifted lower when it was placed near a highly dielectric and conductive load such as human body. For tuning, a phantom that presented loading characteristics similar to a human body at 297.2 MHz was placed 4 cm above the dipoles. The lengths of the dipole segments were adjusted to bring the resonant frequencies close to the NMR Larmor frequency (297.2 MHz at 7T). Each dipole was fed via a Lorenz (lattice) balun (21) (Figure 1b) matched to ~90 Ω (i.e. two 35 nH inductors and two 8 pF capacitors, consistent with theoretical prediction). Two diodes were added in series with the feeding points of each dipole so that it could be detuned during signal reception (Figure 1b). The RF power of the transmit port of the 7T scanner (Siemens Magnetom 7T, Erlangen, Germany) was split equally between each dipole using a home-built lumped-element quadrature hybrid (22), using a minimal amount of metal to avoid eddy currents in close proximity to the dipoles. Reflected power from the isolation port of the hybrid was terminated with a high-power 50-Ω load (Figure 1a).
A body-mimicking phantom was constructed using a plastic container of 425 mm width, 600 mm length, and 170 mm depth. It was used to optimize coil tuning and perform initial imaging experiments. To mimic the bulk electrical properties of muscle of the back at 7T (σ = 0.79 S/m, εr = 59) (23), the phantom was filled to a depth of 110 mm with a sugar-gel recipe developed earlier (24) (available at http://www.amri.ninds.nih.gov/phantomrecipe.html).
Two additional optimization steps were performed on the dipole array before the detuning circuits were added. First, with the elements connected in quadrature (90° inter-channel phase delay), the B1+ fields of the elements were mapped in the central axial plane using a modified Bloch-Siegert approach (25). The separation between the dipoles was empirically minimized until coupling between the elements significantly changed their B1+ distributions. In the next step, the transmit efficiency (defined as B1+ in μT per unit input voltage in V) was measured at the center axial plane by varying the phase delay between the hybrid and individual dipoles. Phase delays were adjusted by inserting additional cable of different lengths between the dipole feeding points and either hybrid output ports. The phase delay was varied in ~30° steps, and the average transmit efficiencies were fit using 6th order polynomials as a function of inter-channel phase delay. Beside phase delay, the offset in L-R direction between the center of the array and the center of the ROI could also affect the transmit efficiency. Thus similar to (4), the transmit efficiency within a 20 mm diameter circular ROI at 50 mm depth was then numerically maximized over the two variables. After this optimization, the dipole array was converted to a transmit-only configuration by implementing the detuning circuits as mentioned above.
Receive Array Design
Since the main goal of this work was to improve transmit efficiency, a conventional array of overlapping loops was used for reception. Starting with the array described in (4), two columns of 10cm square coils, four per column, were configured in receive-only mode. A third column of transmit-only coil as well as the transmit feature in the center column was removed. Each coil was tuned and matched using the loading phantom. For reception, high reflection, low noise GaAs-FET preamplifiers (type Siemens 10185702 + 10185751; Stark Contrast, Erlangen, Germany) were used, which included active detuning circuits on their circuit boards. For additional protection from RF transmit interference, passive detuning (including an inductor and a pair of cross-diodes) was added to one capacitor of each element. To prevent inter-channel coupling through the RF receive cables, floating cable traps (26) were added on each cable bundle that contains all four coaxial cables connecting the preamplifiers on receive coils to the coil system plug.
Use of the Loop-based Transmit-Receive Spine Array as Reference
In order to compare the transmit efficiency and the RF heating of the dipole-based transmit array to the loop-based design (4), the latter was slightly modified: 1) inline traps were added on the coax cables connecting the four receive-only coils; 2) the bundle traps were replaced with custom-built floating traps (26); and 3) the transmit cables were cut to the determined optimal lengths.
MR Experiments
Phantom measurement of RF transmit efficiency
The experimental setup is illustrated in Figure 1c. A ROI representing the expected location of the spine is marked as a white circle in the center axial plane of the phantom (marked by yellow/black dashed rectangle), centered 5 cm above the bottom of the phantom in anterior-posterior (A-P) direction. Unless specified otherwise, experiments were conducted with the center planes of the phantom (x’=0 plane and z’=0) coinciding with x=0 plane and z=0 plane, as indicated in Figure 1c. For simplicity, we define this position as the “center position”.
Transmit efficiency was measured using Bloch-Siegert B1+ mapping (25) as follows. First, for calibration, the RF envelope voltage required to generate an 11.7 μT (500 Hz) B1+ field was estimated at the center of the ROI (x’=0,y’=0,z’=0) at the center position. After transmitter calibration, axial B1+ maps were acquired. Given the length of the dipole and the increasing gradient nonlinearity in the S-I direction, mapping was performed in sequential axial scans by moving the patient table to cover sections of the FOV. Each section covered a 10 cm region along the long-axis (S-I) of the FOV. Five sections centered on the long-axis of the phantom (S-I) were selected and imaged. The acquired B1+ maps were converted to B1+ efficiency by dividing the measured B1+ amplitudes (in μT) by the RF envelop voltages (in V) used to acquire the data.
Phantom measurement of RF induced heating
To evaluate the RF energy deposition, MR thermometry based on the water proton resonant frequency (PRF) shift was performed for each array. Multi-echo gradient echo (GRE) images were acquired while depositing RF energy with a 32 kHz off-resonance pre-pulse that deposited substantial RF energy deposition without affecting MRI signal. In order to avoid impractically long experiments (due to the required cool-down time between the acquisitions), thermometry was only performed at a single section that covered the area of largest expected temperature increase, which was close to the feeding point (i.e. z’=0) based on the theory (19) and simulation. To minimize the effects of gradient non-linearities, the experiment was performed at the center position, as indicated in Figure 1c. Coronal rather than axial slices were chosen to maximize coverage parallel to the bottom part of the phantom. Duration of the heating was 10 minutes with an average power deposition of approximately 33 W, as measured using the power monitor on the scanner. Temperature changes could be derived from changes in image phase (induced by PRF shift) before and after the heating. As mentioned in (24), the multiple resonances of sucrose, whose resonance frequencies do not change notably with temperature, can confound PRF-based MR thermometry (similar to what is observed in water-lipid environments (27)). Therefore, a nonlinear least squares fitting approach was used to determine changes in water proton resonance frequency in the presence of the known sucrose signal (28), rather than a linear scaling of phase changes as is done in conventional PRF-based MR thermometry (29) .
Anatomical MRI
To demonstrate practical use of the array, anatomical images of the vertebral column and the spinal cord were acquired on two healthy volunteers; all imaging was approved by the Institutional Review Board. On the first volunteer, single slice images were acquired with breathholding and without fat saturation or any saturation band. First, an axial B1+ map with a matching anatomical GRE image was acquired at z=0 to confirm the B1+ efficiency in vivo. Then two additional single slice GRE images in axial and sagittal planes were acquired with imaging parameters similar to what was presented in the literature, e.g. (2,4,5). On the second volunteer, a high-resolution image of the spinal cord was acquired without breathhold using a 2D GRE sequence. Imaging artifacts caused by breathing were corrected based on the built-in navigator echoes, in combination with information about the TE dependence of the artifact, derived from three otherwise identical pre-scan repetitions without phase encoding.
RESULTS
Array Configurations
The prototype transmit dipole array, receive array, and transmit/receive (T/R) loop array are shown in Figure 2a, 2b, and 2c respectively. For the transmit array, a length of 370 mm was found to provide the best tuning for the 7T resonance frequency (297.2 MHz in this case), which is shorter than the half wavelength in vacuum (~505 mm). This shortening is attributed to the interaction of the dipole with the high dielectric constant of the body (or the phantom with similar properties). A minimum separation of 6 cm between the dipoles was required to prevent significant inter-element coupling. In this setup, the reflection coefficient S11 for each dipole was < −30 dB, the coupling coefficient S21 between the elements was ~ −12 dB, and the active detuning circuit typically provided better than 20 dB of detuning using an S21 measurement with a pick-up probe. The S-parameters for the loop-based arrays (receive-only and T/R arrays) were typically, S11 = −35 dB for matching, S21 = −24 dB between elements (measured without the present of other electronics), and the detuning circuit typically provided 33 dB of detuning. The unloaded/loaded Q-ratio for each channel was around 7.3. Finally, the preamplifiers typically provided about 17 dB of preamp decoupling on each receive-only element.
Optimal Positioning and Phasing of the Dipole Array
With the dipole elements connected in T/R configuration, the transmit efficiency of the dipole pair as a function of inter-element phase delay was measured using B1+ mapping. The maps of transmit efficiency (in μT/V) are shown in Figure 3a, with regions of low MRI signal masked out. The capability of “steering” the transmit field by incrementing the phase shift between the elements is clearly demonstrated. Figure 3b shows the dependence of transmit efficiency in the target ROI (at the expected depth of the spine) on the phase delay and on the ROI position relative to the center of the array (i.e. spatial offset in L-R direction). Given that the width of the phantom was much larger than the separation of dipole pair (425 mm v.s. 60 mm), the effect of the spatial offset could be simulated by moving the ROI in the data analysis, as opposed to repeating the B1+ mapping while physically shifting the phantom over the array. The optimum phase delay and spatial offset were found to be 142° and 9 mm, respectively. However, it was found that performance remained within 97% of this optimum value (Figure 3c) if the dipole pair were simply driven in quadrature (90º) and no L-R offset was used. Based on this, in vivo experiments were performed with quadrature drive and with center line of the array in S-I direction approximately aligned with the spinal cord. To further simplify the setup, the centerlines of the receive array (in L-R and S-I) were aligned with those of the transmit array. While this may not be optimal for receiving in the presence of twisting effects in the receive B1 profile (4), the deterioration in receive sensitivity was expected to be minimal.
Performance Comparison between Dipole and Loop Array
Estimated calibration voltages for the RF amplitude were 347 V for the dipole array versus 548 V for the T/R loop array, corresponding to transmit efficiencies as of 0.0338 μT/V and 0.0214 μT/V respectively. The latter is comparable with that reported in (4). Axial B1+ efficiency maps (in μT/V) calculated from B1+ mapping experiments for the two arrays are summarized in Figure 4. It is clear that the dipole-based transmit array has much higher transmit efficiency along the longitudinal centerline (S-I) as compared with the loop-based array. We attribute this to the difference in B1+ field distributions between the two arrays, given that the average B1+ efficiencies over the entire phantom were nearly identical (0.026 μT/V for both arrays).
Maps of temperature change after normalization for input power for the two arrays are shown in Figure 5. The dipole transmitter (Figure 5a) showed ~10% higher maximum RF induced heating as compared to the loop transmitter (Figure 5b) for the 10-minute heating test. Heating for the dipole array is more concentrated on the left side, toward the dipole with 90° phase delay, whereas the heating generated by the loop transmitters is more evenly distributed.
These results suggest that, for a given amount of input power, the dipole array provides a higher B1+ in the region of the spine than the loop-based array, but it also generates higher peak local SAR. Therefore, to properly compare transmit efficiency under SAR-limited conditions, we normalized the B1+ by the square-root of associated maximum local heating for a 10 minute session (from now on referred to as the heating-corrected transmit efficiency, in as temperature change in Kelvin). The square root is needed for the temperature change because B1+ is proportional to the input RF envelope voltage, whereas the local temperature change is proportional to the power (the square of voltage). Figure 6 shows line profiles through the center of the spinal ROI in three orthogonal directions: L-R, A-P, and S-I. These results confirm that the gain in B1+ efficiency by the dipole array is higher than the associated slightly increased peak RF heating. In fact, the gain in these three profiles within 10 cm around the center of the spinal ROI (x’=0,y’=0,z’=0) (highlighted in grey in Figure 6) can be as high as 76%.
Figure 7a shows an in vivo B1+ efficiency map overlaid on an anatomical image. For comparison, the B1+ efficiency map acquired from the same location on the phantom (center slice in Figure 4a) is shown in Figure 7b, with the ROI corresponding to the cord region marked with white square (about 4cm deep). The average B1+ efficiency within the ROI was 0.0405 μT/V for the in vivo data and 0.0399 μT/V for the phantom. Examples of in vivo images obtained with the dipole array are shown in Figure 7c-e and demonstrate the ability to obtain excellent sensitivity in the region of the spine at high resolution.
DISCUSSION
In this study, the feasibility of using electric dipole antennae as transmit elements in 7T MR spinal imaging was explored with careful positioning of the dipole elements. Phantom experiments showed that significant gains in transmit efficiency can be achieved without significantly increasing tissue heating. Given the excellent match between the length of the “half-wave dipole” and the expected FOV for spine imaging at 7T, use of electric dipole elements can simplify transmit array design for human spinal applications by reducing total number of channels required to cover a large area (in this example, two dipoles instead of eight loops). Minimizing the number of cables that are used can further reduce potential cable coupling and minimize the parts count for supporting components, such as RF traps and power splitters. One potential drawback with this approach is that the large size of the dipoles used here limits the maximum number of the dipoles that can be used simultaneously without detrimental inter-element coupling. For spine imaging at 7 T, 6 cm separation was found to be necessary to prevent excessive inter-element coupling. Using more half-wave dipoles in parallel with similar separations in L-R direction would place them too far away from the spine ROI. Other dipole configurations would be required to fully take advantage of multi-channel transmission, which usually comes with eight parallel transmit channels. For clinical MRI of the spine, it may be necessary to shorten dipoles (14,18), or combine them with loop transmitters.
The gain in the transmit efficiency at the spinal cord region for the dipole seems mostly due to a different distribution of the B1+ field in comparison with the loop-based design, given the average B1+ efficiency was almost identical. This is probably due to the removal of the return path that is always 180-degrees out-of-phase, i.e. less field cancellation for positions far from the coil. However, the actual mechanism needs further investigation.
In this work, array configurations were optimized experimentally with the help of a modified Bloch-Siegert B1+ mapping method (25). Because this is a laborious approach, only a relatively small portion of the parameter space was explored. Further optimization of the dipole transmitters may be possible by using RF simulations, which readily allow coverage of a larger parameter space. In addition, RF simulations would facilitate the optimization of the dipole configuration using heating-corrected transmit efficiency (or equivalent) rather than RF amplitude normalized transmit efficiency, as was done here, and this may result in further improvement in performance.
CONCLUSION
A coil that employed an efficient transmit array based on electric dipoles and designed for use with a loop-based receive array was constructed for 7T spine imaging. Steering of the B1+ field in the axial plane was accomplished by varying the phase delay between the dipoles, which allowed RF shimming and transmit efficiency optimization for a pre-selected ROI. The transmit efficiency of the dipole array was compared with a previously described design based on quadrature-loop pairs. The dipole antenna provides a much simpler and more efficient transmitter design, and it can be readily combined with standard receive arrays.
Acknowledgements
This research was supported by the Intramural Research Program of the National Institute of Neurological Disorders and Stroke.
REFERENCES
- 1.Shapiro MD. MR Imaging of the Spine at 3T. Magnetic Resonance Imaging Clinics of North America. 2006;14(1):97–108. doi: 10.1016/j.mric.2006.01.005. [DOI] [PubMed] [Google Scholar]
- 2.Kraff O, Bitz AK, Kruszona S, Orzada S, Schaefer LC, Theysohn JM, Maderwald S, Ladd ME, Quick HH. An eight-channel phased array RF coil for spine MR imaging at 7 T. Invest Radiol. 2009;44(11):734–740. doi: 10.1097/RLI.0b013e3181b24ab7. [DOI] [PubMed] [Google Scholar]
- 3.Wu B, Wang C, Krug R, Kelley DA, Xu D, Pang Y, Banerjee S, Vigneron DB, Nelson SJ, Majumdar S, Zhang X. 7T Human Spine Imaging Arrays With Adjustable Inductive Decoupling. Biomedical Engineering, IEEE Transactions on. 2010;57(2):397–403. doi: 10.1109/TBME.2009.2030170. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Duan Q, Sodickson DK, Lattanzi R, Zhang B, Wiggins G. Optimizing 7 T spine array design through offsetting of transmit and receive elements and quadrature excitation. International Society for Magnetic Resonance in Medicine; 18th Annual Meeting & Exhibition of ISMRM; Stockholm, Sweden. 2010. p. 324. [Google Scholar]
- 5.Vossen M, Teeuwisse W, Reijnierse M, Collins CM, Smith NB, Webb AG. A radiofrequency coil configuration for imaging the human vertebral column at 7 T. J Magn Reson. 2011;208(2):291–297. doi: 10.1016/j.jmr.2010.11.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Sigmund EE, Suero GA, Hu C, McGorty K, Sodickson DK, Wiggins GC, Helpern JA. High-resolution human cervical spinal cord imaging at 7 T. NMR Biomed. 2012;25(7):891–899. doi: 10.1002/nbm.1809. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Zhao W, Cohen-Adad J, Polimeni JR, Keil B, Guerin B, Setsompop K, Serano P, Mareyam A, Hoecht P, Wald LL. Nineteen-channel receive array and four-channel transmit array coil for cervical spinal cord imaging at 7T. Magn Reson Med. 2014;72(1):291–300. doi: 10.1002/mrm.24911. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Brunner DO, De Zanche N, Frohlich J, Paska J, Pruessmann KP. Travelling-wave nuclear magnetic resonance. Nature. 2009;457(7232):994–998. doi: 10.1038/nature07752. [DOI] [PubMed] [Google Scholar]
- 9.Andreychenko A, Voogt I, Kroeze H, Klomp D, Lagendijk J, Luijten P, van den Berg C. An Advantageous Combination of Travelling Wave and Local Receive for Spine MR Imaging at 7T: Local SAR Reduction and SENSE Reconstruction; 18th Annual Meeting of ISMRM; Stockholm, Sweden. 2010. p. 430. [Google Scholar]
- 10.Zhang B, Sodickson DK, Lattanzi R, Duan Q, Stoeckel B, Wiggins GC. Whole body traveling wave magnetic resonance imaging at high field strength: Homogeneity, efficiency, and energy deposition as compared with traditional excitation mechanisms. Magnetic Resonance in Medicine. 2012;67(4):1183–1193. doi: 10.1002/mrm.23107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Raaijmakers AJ, Ipek O, Klomp DW, Possanzini C, Harvey PR, Lagendijk JJ, van den Berg CA. Design of a radiative surface coil array element at 7 T: the single-side adapted dipole antenna. Magn Reson Med. 2011;66(5):1488–1497. doi: 10.1002/mrm.22886. [DOI] [PubMed] [Google Scholar]
- 12.Ipek O, Raaijmakers AJE, Klomp DWJ, Lagendijk JJW, Luijten PR, Berg CATvd Characterization of transceive surface element designs for 7 tesla magnetic resonance imaging of the prostate: radiative antenna and microstrip. PHYSICS IN MEDICINE AND BIOLOGY. 2012;57(2):343. doi: 10.1088/0031-9155/57/2/343. [DOI] [PubMed] [Google Scholar]
- 13.Ipek O, Raaijmakers A, Lagendijk J, Luijten P, van den Berg C. Optimization of the radiative antenna for 7-T magnetic resonance body imaging. Concepts in Magnetic Resonance Part B: Magnetic Resonance Engineering. 2013;43B(1):1–10. [Google Scholar]
- 14.Raaijmakers AJE, Italiaander M, Voogt IJ, Luijten PR, Hoogduin JM, Klomp DWJ, van den Berg CAT. The fractionated dipole antenna: A new antenna for body imaging at 7 Tesla. Magnetic Resonance in Medicine. 2015 doi: 10.1002/mrm.25596. 10.1002/mrm.25596. [DOI] [PubMed] [Google Scholar]
- 15.Lakshmanan K, Cloos M, Lattanzi R, Sodickson D, Novivkov D, Wiggins G. The Circular Dipole. International Society for Magnetic Resonance in Medicine; Joint Annual Meeting ISMRM-ESMRMB; Milan, Italy. 2014. p. 315. [Google Scholar]
- 16.Duan Q, Gudino N, Zwart JAd, Gelderen Pv, Murphy-Boesch J, Duyn JH, Merkle H. Array RF Transmitter for 7T MRI of the Spine Based on Dipole Antennas. International Society for Magnetic Resonance in Medicine; Joint Annual Meeting ISMRM-ESMRMB; Milan, Italy. 2014. p. 316. [Google Scholar]
- 17.Chen G, Cloos M, Lattanzi R, Sodickson DK, Wiggins G. Bent Electric Dipoles: A Novel Coil Design Inspired by the Ideal Current Pattern for Central SNR at 7 Tesla. International Society for Magnetic Resonance in Medicine; Joint Annual Meeting ISMRM-ESMRMB; Milan, Italy. 2014. p. 402. [Google Scholar]
- 18.Chen G, Cloos M, Sodickson DK, Wiggins G. A 7T 8 Channel Transmit-Receive Dipole Array for Head Imaging: Dipole Element and Coil Evaluation. International Society for Magnetic Resonance in Medicine; Joint Annual Meeting ISMRM-ESMRMB; Milan, Italy. 2014. p. 621. [Google Scholar]
- 19.Silver S. The "Half-wave Dipole". In: Silver S, editor. Microwave Antenna Theory and Design. McGraw-Hill Book Company, Inc.; New York: 1949. pp. 98–99. [Google Scholar]
- 20.Eryaman Y, Guerin B, Keil B, Mareyam A, Herraiz JL, Kosior RK, Martin A, Torrado-Carvajal A, Malpica N, Hernandez-Tamames JA, Schiavi E, Adalsteinsson E, Wald LL. SAR reduction in 7T C-spine imaging using a "dark modes" transmit array strategy. Magn Reson Med. 2014 doi: 10.1002/mrm.25246. DOI: 10.1002/mrm.25246. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Lorenz C. Schaltungsanordnung zum Übergang von einer symmetrischen, elektrischen Anordnung zu einer unsymmetrischen, insbesondere bei Hochfrequenzanwendungen. Germany patent. 1932 603816. [Google Scholar]
- 22.Mispelter J, Lupu M, Briguet A. Quadrature Driving. In: Wei TK, editor. NMR Probeheads for Biophysical and Biomedical Experiments. Mainland Press; Singapore: 2006. pp. 131–186. [Google Scholar]
- 23.Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. Phys Med Biol. 1996;41(11):2251–2269. doi: 10.1088/0031-9155/41/11/002. [DOI] [PubMed] [Google Scholar]
- 24.Duan Q, Duyn JH, Gudino N, de Zwart JA, van Gelderen P, Sodickson DK, Brown R. Characterization of a dielectric phantom for high-field magnetic resonance imaging applications. Medical Physics. 2014;41(10):102303. doi: 10.1118/1.4895823. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Duan Q, van Gelderen P, Duyn J. Improved Bloch-Siegert based B(1) mapping by reducing off-resonance shift. NMR Biomed. 2013;26(9):1070–1078. doi: 10.1002/nbm.2920. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Seeber DA, Jevtic J, Menon A. Floating shield current suppression trap. Concepts in Magnetic Resonance Part B: Magnetic Resonance Engineering. 2004;21B(1):26–31. [Google Scholar]
- 27.de Zwart JA, Vimeux FC, Delalande C, Canioni P, Moonen CT. Fast lipid-suppressed MR temperature mapping with echo-shifted gradient-echo imaging and spectral-spatial excitation. Magn Reson Med. 1999;42(1):53–59. doi: 10.1002/(sici)1522-2594(199907)42:1<53::aid-mrm9>3.0.co;2-s. [DOI] [PubMed] [Google Scholar]
- 28.Deniz CM, Brown R, de Zwart JA, Collins CM, Sodickson DK. ISMRM scientific workshop "Safety in MRI: Guidelines, Rationale & Challenges". International Society for Magnetic Resonance in Medicine; Washington, DC, USA: 2014. Multi-Channel Array Safety Using Least Squares Fitting based MR Thermometry. [Google Scholar]
- 29.Ishihara Y, Calderon A, Watanabe H, Okamoto K, Suzuki Y, Kuroda K, Suzuki Y. A precise and fast temperature mapping using water proton chemical shift. Magnetic Resonance in Medicine. 1995;34(6):814–823. doi: 10.1002/mrm.1910340606. [DOI] [PubMed] [Google Scholar]