Abstract
Over the past decade, the number of known eicosanoids has expanded immensely and we have now developed an ultra-performance liquid chromatography - electrospray ionization triple quadrupole mass spectrometric (UPLC-QTRAP/MS/MS) method to monitor and quantify numerous eicosanoids. The UPLC-QTRAP/MS/MS approach utilizes scheduled multiple reaction monitoring (MRM) to optimize sensitivity, number of metabolites that can be analyzed and the time requirement of the analysis. A total of 184 eicosanoids including 26 deuterated internal standards can be separated and monitored in a single 5 min UPLC run. To demonstrate a practical application, human plasma samples were analyzed following solid-phase extraction (SPE) and the recovery rate and matrix effects were determined for the 26 deuterated internal standards added to the plasma. The method was validated and shown to be sensitive with the limit of quantitation at pg levels for most compounds, accurate with recovery rates of 70-120%, and precise with a CV<30 for all compounds. Also, the method showed a linear response over a range spanning several orders of magnitude. In a QC human plasma sample, we identified and rigorously quantified over 120 eicosanoids.
Keywords: Eicosanoids, Ultra-performance liquid chromatography, Tandem mass spectrometry, Plasma, Lipidomics
1. Introduction
Eicosanoids comprise a class of bioactive lipids derived from a unique group of polyunsaturated essential fatty acids that mediate a wide variety of important physiological functions [1]. They exert complex control over many physiological processes, including inflammation [2]. Also, many eicosanoids are associated with chronic disease conditions including heart disease [3], cancer [4] and arthritis [5].
Arachidonic acid (AA) and related polyunsaturated fatty acids serve as the metabolic precursors for eicosanoid synthesis. Biologically, these molecules are generally stored in the sn-2 position of the glycerol backbone of membrane phospholipids. To be used for biosynthesis, the arachidonic acid must first be released from phospholipids via phospholipase A2 (PLA2) [6], and which is then acted on by enzymes of the cyclooxygenase pathway (COX) or the lipoxygenase pathway (LOX) to form prostaglandins (PG) and thromboxanes (TX) or leukotrienes (LT), respectively. The COX site incorporates molecular oxygen at the 11- and 15-carbons on arachidonic acid to form PGG2, followed by a peroxidase activity that reduces the peroxide to a hydroxyl to form PGH2. PGH2 is an intermediate for a number of different bioactive products through the action of PG synthases that can form a number of important signaling molecules, including PGI2, TXA, PGE2, PGD2, and PGF2a. Alternatively, the LOX pathway produces LTs including 5-HETE, 15-HETE, LTA4, LTB4, LTC4 etc, which constitute a family of biologically active molecules formed in response to immunological and non-immunological stimuli. For example, 5-LOX produces 5-hydroperoxy-eicosatetraenoic acid (5-HpETE) by incorporating one molecular oxygen at the C-5 position of arachidonic acid that can be reduced to 5-HETE, or undergo a catalytic rearrangement in the 5-LOX active site to form LTA4. An additional set of enzymes catalyzes the stereospecific rearrangement of LTA4 to produce LTB4, LTC4, and LTD4.
This complete enzymatic system produces hundreds of eicosanoids derived from AA and related polyunsaturated fatty acids with very similar structures, chemistries and physical properties [7], which makes the analysis of eicosanoids a challenging task, especially in biological samples. The concentration of eicosanoids in plasma or serum is the lowest among all endogenous lipid metabolites [8]. However, under certain conditions the plasma level of eicosanoids may change considerably and thus, eicosanoids may serve as a useful readout reflecting disease progression. As a result, current research is focused on developing fast, sensitive, and reliable methods that accurately profile and quantify eicosanoid biomarkers [9,10].
In the past, eicosanoids were mainly analyzed by enzyme-linked immunosorbent assays (EIA) [11, 12], gas chromatography-mass spectrometry (GC-MS) [13] and liquid chromatography –mass spectrometry (LC-MS) [14]. The drawback of EIA is a lack of specificity and the ability to determine multiple analytes in a single set of analyses. GC-MS provides greater sensitivity and selectivity for eicosanoid analysis, but requires chemical derivatization steps that limit its application. The rapid progress of liquid-chromatography-electrospray ionization tandem mass spectrometry (LC-ESI-MS/MS) has facilitated the use of this technology for accurate monitoring of eicosanoid metabolites in biological samples [7]. Previous reports include liquid-liquid extraction for the determination of PGE2 and LTB4 in plasma using LC-MS [15], the analysis of four kinds of PGs and LTs in cell culture media by LC-MS [16], an on-line two dimensional reverse-phase LC-MS for the simultaneous determine of PGE2, PGF2a and 13,4-dihydro-15-keto PGF2a [17], and a LC-MS method for the simultaneous determination of twenty-three eicosanoids [18], a UPLC-MS platform that enables profiling of 122 eicosanoids from human whole blood[19], a targeted HPLC-MS/MS analysis platform for 100 oxylipins and 36 oxylipins was detected from 250 uL human plasma in 26 min[20], a LC–MS/MS method for rapid and concomitant quantification of 26 PUFA metabolites from Caco-2 cells [21], a LC–MS/MS for the simultaneous analysis of arachidonic acid and 32 related metabolites in 1 mL human plasma [22], a online HPLC-MS/MS analyzed more than 20 different oxidized fatty acids and their precursors from 200 uL plasma or urine [23], a fast LC–MS/MS method including on-line SPE and LIT fragment confirmation for the profiling of 7 PUFAs and 94 oxidized metabolites from 200 uL plasma [24]. We previously developed a high-throughput lipidomic analysis of eicosanoids using LC-MS to monitor 141 unique species in a single 22 min analysis [25], and applied it to investigate the human plasma lipidome [26,27].
However, these methods may need much more biolgocial sample or longer analysis time. In order to address the need for large-scale high-throughput analysis of eicosanoids in small quantities of human plasma and tissues, we now report the development of a method utilizing an ultra-high performance liquid chromatography-QTRAP MS/MS (UPLC-MS/MS) for monitoring 184 eicosanoids in a 5 min run. The new methodology is validated by identification and quantitation of eicosanoids in 20 uL human plasma.
2. Experimental
2.1 Reagents
All eicosanoids and deuterated internal standards were purchased from Cayman Chemical. Optima LC-MS grade acetonitrile (ACN), methanol (MeOH), and water were obtained from Fisher Scientific. Isopropanol (IPA) was purchased from Sigma-Aldrich. Formic acid (FA) was obtained from EMD Technologies. Dulbecco's Phosphate Buffered Saline (DPBS) was obtained from Corning Life Science.
2.2 Sample preparation
2.2.1 Preparation of primary standard and internal standard solutions
For the preparation of calibration curves, stock solutions were prepared in ethanol that contained all eicosanoid standards, each at a concentration of 0.25 ng/uL. Working standard solutions for all eicosanoids were prepared by serial dilution of the stock solutions to create the necessary concentrations. A solution containing 26 internal (deuterated) eicosanoid standards was prepared at 0.01 ng/uL in ethanol. All solutions were stored at -80°C when not in use.
2.2.2 Extraction of metabolites from plasma and tissue
Aliquots of 20 ul control plasma (Human Source Plasma, Gemini Bio-Products) were diluted to 1 mL with phosphate salt buffer spiked with 100 uL of the internal standard solution. The eicosanoids were extracted using Strata-X reversed-phase SPE columns (8B-S100-UBJ, Phenomenex). Columns were washed with 3 mL MeOH and then equilibrated with 3 mL H2O. After loading the sample, the columns were washed with 10% MeOH to remove impurities, and the metabolites were then eluted with 1 mL of MeOH and stored at -80°C to prevent metabolite degradation. Prior to analysis, the eluant was dried under vacuum and redissolved in 50 uL of the UPLC solvent A (water/acetonitrile/acetic acid (60:40:0.02; v/v/v)) for UPLC/MS/MS analysis.
Took tissue sample, then weighed, and transferred to 1 mL PBS buffer containing 10% MeOH tube. The tissue was homogenized using Beadbug Microtube Homogenizer (Benchmark Scientific). The eicosanoids extracted from homogenates with SPE followed the same protocol as plasma sample.
2.3 UPLC-MS/MS
An Acquity UPLC system (Waters Corp) was used. Reversed-phase separation was performed on an Acquity UPLC BEH shield RP18 column (2.1×100 mm 1.7 μm; Waters). The mobile phase consisted of (A) ACN/water/acetic acid (60/40/0.02, v/v) and (B) ACN/IPA(50/50, v/v). Gradient elution without splitting was carried out for 5 min at a flow rate of 0.5 mL/min. Gradient conditions were as follows: 0-4.0 min, 0.1-55% B; 4.0-4.5 min, 55-99% B; 4.5-5.0 min, 99% B; A 10 μL aliquot of each sample was injected into the column. The column temperature was kept at 40 °C. All samples were incubated at 4 °C during analysis.
Mass spectrometry was performed on an ABI/Sciex 6500 QTRAP hybrid, triple quadrupole, linear ion trap mass spectrometer equipped with a Turbo V ion source. Eicosanoids were detected in negative electrospray ion mode. Curtain gas (CUR), nebulizer gas (GS1) and turbo-gas (GS2) were set at 10 psi, 30 psi and 30 psi, respectively. The electrospray voltage was -4.5 kV, and the turbo ion spray source temperature was 525 °C. Nitrogen was employed as collision gas and the collisionally-activated dissociation (CID) was set at a high level.
Eicosanoids were analyzed using scheduled multiple reaction monitoring (MRM). Mass spectrometer parameters including the declustering potentials and collision energies were optimized for each analyte. Data acquisitions were performed using Analyst 1.6.2 software (Applied Biosystems). Multiquant software (Applied Biosystems) was used to quantify all metabolites.
2.4 Analytical validation
2.4.1 Linearity and LOD and LOQ
A typical standard curve was prepared by adding 1 ng of each internal (deuterated) eicosanoid standard to the following amounts of eicosanoid (nondeuterated) primary standards: 0.005, 0.015, 0.025, 0.035, 0.05, 0.15, 0.25, 0.35, 0.5, 1.5, 2.5, 3.5, and 5.0 ng. Quantitation of eicosanoid levels was performed using linear regression of the response ratios (peak area analyte/peak area internal standard) obtained from the calibration curve to calculate the corresponding eicosanoid amount. The limit of detection (LOD) was defined as the concentration that resulted in a peak with a signal-to-noise ratio (S/N) greater than 3:1 (3 S/N) and the limit of quantitation (LOQ) was defined as (S/N) greater than 7:1 (7 S/N)
2.4.2 Recovery rate and matrix effect
Recovery from control plasma was determined by comparison of the analyte peak area of the internal standard spiked into a plasma sample compared with the corresponding peak area of standard solutions extracted in a similar manner. The determination was performed in triplicate. The matrix effect was calculated by dividing the peak area of the internal standard spiked into a plasma sample and extracted into mobile phase A by the area of the deuterated standard in the standard solution. This determination was also performed in triplicate.
2.4.3 Accuracy and precision
Accuracy and precision of the method was determined using quality control (QC) samples spiked at three levels of eicosanoids: low (LQC) 0.15 ng, medium (MQC) 1.5 ng, and high QC (HQC) 5 ng. QC samples were prepared by spiking blank plasma with three levels of eisosanoids and internal standards prior to plasma extraction as described. Intra-batch and inter-batch (three different batches) accuracy and precision were determined by analyzing five QC samples covering the calibration range. The precision of the quantitation was expressed as percent coefficient of variance (CV %), calculated by dividing the standard deviation by the mean and then multiplied by 100.
These QC amounts included the known fortified level added to the control plasma plus the endogenous amount of analytes. The endogenous amounts of analytes in plasma were determined in five replicate measurements. The accuracy of the analytic method was denoted by the relative error (RE %), calculated as percent of the mean deviation from the known amount plus endogenous amount, RE % = [(amount found − (known amount + endogenous amount) × 100/(known amount + endogenous amount)].
2.4.4 Stability
To determine the stability of the processed samples, they were kept at 4°C in the autosampler and injected three times at 0, 4 and 8 h with three levels of quality control samples, respectively. The peak area of the analytes at the initial point (0 h) was used as the reference to determine the relative stability at subsequent points.
3. Results and Discussion
3.1 Method development
SPE techniques were used to extract eicosanoid metabolites from plasma, which is more suitable for processing a large number of samples than a more efficient liquid/liquid extraction (LLE) technique [28]. The extraction efficiency of LLE is generally higher than SPE, but this method also extracts many endogenous impurities that can affect the separation and quantitation of target analytes. The ability of SPE to eliminate impurities is better than that of LLE, which improves the detection of eicosanoids in biological matrices, especially when present at low levels.
A crucial aspect of our method is the inclusion of 26 deuterated internal standards. All samples are spiked with a mixture of deuterated internal standards prior to lipid extraction. An internal standard is used to correct for run-to-run variation in extraction efficiency, for monitoring the chromatographic response, and for normalization purpose which allows for accurate quantification. Also, eicosanoid quantitation was performed by the stable isotope dilution method. For each eicosanoid to be quantify, an internal standard was selected that had a different precursor ion mass than the target analyte, but was chemically and structurally as similar to the target analyte as possible. This is ideally achieved by using a deuterated analog of the analyte. We employed these standards whenever they were commercially available. For example, (d4) PGE2 was employed as the internal standard for PGE2. In other cases, we employed a deuterated analog that was the closest to the desired analog in characteristics. For example, (d4) 15d PGJ2 was employed as the internal standard for PGJ2, 15d- PGJ2, and 15d-PGD2. The 26 deuterated internal standards, which presently used to quantify, assigned to each of the 158 eicosanoids analyzed are listed in Table 1.
Table 1. Internal Standard Assigned for Analytes.
| No. | Internal Standard | Analytes Assigned | No. | Internal Standard | Analytes Assigned |
|---|---|---|---|---|---|
| 1 | (d4) 6k PGF1α | 6 | 14 | (d6) 20-HETE | 10 |
| 2 | (d4) TXB2 | 5 | 15 | (d4) 9-HODE | 2 |
| 3 | (d4) PGF2α | 12 | 16 | (d4) 13-HODE | 4 |
| 4 | (d4) PGE2 | 10 | 17 | (d7) 5-oxoETE | 6 |
| 5 | (d4) PGD2 | 11 | 18 | (d4) Resolvin E1 | 6 |
| 6 | (d4) 15d PGJ2 | 6 | 19 | (d11) 8,9 EET | 1 |
| 7 | (d4) dhk PGF2α | 1 | 20 | (d11) 11,12 EET | 5 |
| 8 | (d4) dhk PGD2 | 2 | 21 | (d11) 14,15 EET | 1 |
| 9 | (d11) 5-iso PGF2αVI | 4 | 22 | (d4) 9,10 diHOME | 1 |
| 10 | (d4) LTB4 | 18 | 23 | (d4) 12,13 diHOME | 1 |
| 11 | (d8) 5-HETE | 13 | 24 | (d5) LTC4 | 1 |
| 12 | (d8) 12-HETE | 13 | 25 | (d5) LTE4 | 2 |
| 13 | (d8) 15-HETE | 9 | 26 | (d8) Arachidonic Acid | 8 |
A targeted approach was used to identify and quantify lipids using mass spectrometry (MS) coupled with ultra-high performance liquid chromatography (UPLC-MS). A targeted MRM approach provided for higher sensitivity than an unbiased full scan MS analysis. UPLC also provided enhanced chromatographic resolution, sensitivity, reproducibility and fast separation. Additionally, a QTRAP 6500 was used in the present study. The advantages of the QTRAP 6500 are scan speeds of up to 20,000 Da/second for optimized UPLC strategies, a twenty-fold increase in the detector dynamic range, and LOQ improvement of up to five fold. In summary, the UPLC-QTRAP 6500 system provides a sensitive, accurate and fast separation and monitoring platform.
All eicosanoids are detected in the negative ion mode to take advantage of their conserved terminal carboxyl moiety. Fig. 1 shows a representative extracted ion chromatogram for the separation of all 184 eicosanoids including the 26 internal standards, which are monitored in a single 5 min LC-MS/MS analytical run. The optimal declustering potential (DP) and the collision energy (CE) has been determined for each MRM pair (Table 2). These values were optimized by directly infusing commercial standards into the mass spectrometer.
Fig. 1. Extracted Ion Chromatograms of a mixture 187 Eicosanoid Standard (A) and a Magnified View of the Chromatograms of HETE Metabolites.

Table 2. Optimized MRM Pairs and Parameters for Eicosanoids.
| No. | Common name | Retention Time (min) | m/z | DP (V) | CE (V) | LOQ (pg) | |
|---|---|---|---|---|---|---|---|
|
| |||||||
| Parent | Daughter | ||||||
| 001 | (d4) 6k PGF1α | 0.76 | 373 | 167 | -60 | -34 | - |
| 002 | (d4) TXB2 | 0.98 | 373 | 173 | -50 | -22 | - |
| 003 | (d4) PGF2α | 1.14 | 357 | 197 | -50 | -35 | - |
| 004 | (d4) PGE2 | 1.29 | 355 | 275 | -50 | -23 | - |
| 005 | (d4) PGD2 | 1.47 | 355 | 275 | -50 | -23 | - |
| 006 | (d4) 15d PGJ2 | 3.54 | 319 | 203 | -30 | -20 | - |
| 007 | (d4) dhk PGF2α | 1.74 | 357 | 295 | -80 | -28 | - |
| 008 | (d4) dhk PGD2 | 2.13 | 355 | 297 | -40 | -26 | - |
| 009 | (d4) 8-iso PGF2α VI | 0.99 | 357 | 197 | -20 | -22 | - |
| 010 | (d4) LTB4 | 2.85 | 339 | 197 | -45 | -23 | - |
| 011 | (d8) 5-HETE | 4.17 | 327 | 116 | -40 | -20 | - |
| 012 | (d8) 12-HETE | 4.05 | 327 | 184 | -50 | -19 | - |
| 013 | (d8) 15-HETE | 3.92 | 327 | 226 | -40 | -19 | - |
| 014 | (d6) 20-HETE | 3.66 | 325 | 281 | -50 | -24 | - |
| 015 | (d4) 9-HODE | 3.89 | 299 | 172 | -60 | -23 | - |
| 016 | (d4) 13-HODE | 3.89 | 299 | 198 | -60 | -23 | - |
| 017 | (d7) 5-oxoETE | 4.31 | 323 | 279 | -60 | -22 | - |
| 018 | (d4) Resolvin E1 | 0.80 | 353 | 197 | -40 | -20 | - |
| 019 | (d11) 8,9 EET | 4.37 | 330 | 155 | -50 | -19 | - |
| 020 | (d11) 11,12 EET | 4.34 | 330 | 167 | -50 | -19 | - |
| 021 | (d11) 14,15 EET | 4.23 | 330 | 175 | -40 | -19 | - |
| 022 | (d4) 9,10 diHOME | 3.12 | 317 | 203 | -50 | -29 | - |
| 023 | (d4) 12,13 diHOME | 3.07 | 317 | 185 | -50 | -29 | - |
| 024 | (d5) LTC4 | 2.00 | 629 | 272 | -50 | -34 | - |
| 025 | (d5) LTE4 | 2.25 | 443 | 338 | -30 | -26 | - |
| 026 | (d8) Arachiconic Acid | 4.87 | 311 | 267 | -55 | -20 | - |
| 027 | 6k PGF1α | 0.76 | 369 | 245 | -60 | -34 | 5 |
| 028 | TxB2 | 0.98 | 369 | 169 | -50 | -22 | 3 |
| 029 | PGF2α | 1.12 | 353 | 197 | -50 | -35 | 70 |
| 030 | PGE2 | 1.27 | 351 | 271 | -50 | -23 | 3 |
| 031 | PGD2 | 1.45 | 351 | 27 | -50 | -23 | 3 |
| 032 | 11b PGF2α | 1.08 | 353 | 335 | -50 | -35 | 100 |
| 033 | TXB1 | 0.95 | 371 | 171 | -30 | -27 | 5 |
| 034 | PGF1α | 1.18 | 355 | 293 | -60 | -33 | 7 |
| 035 | PGE1 | 1.40 | 353 | 235 | -40 | -29 | 30 |
| 036 | PGD1 | 1.51 | 353 | 235 | -40 | -29 | 30 |
| 037 | d17 6k PGF1α | 0.66 | 367 | 163 | -60 | -34 | 7 |
| 038 | TXB3 | 0.81 | 367 | 169 | -40 | -27 | 7 |
| 039 | PGF3α | 0.94 | 351 | 193 | -50 | -30 | 50 |
| 040 | PGE3 | 1.03 | 349 | 269 | -30 | -24 | 3 |
| 041 | PGD3 | 1.14 | 349 | 269 | -30 | -24 | 7 |
| 042 | dihomo PGF2α | 1.85 | 381 | 337 | -40 | -37 | 3 |
| 043 | dihomo PGE2 | 2.20 | 379 | 299 | -40 | -37 | - |
| 044 | dihomo PGD2 | 2.30 | 379 | 299 | -40 | -37 | - |
| 045 | dihomo PGJ2 | 3.00 | 361 | 299 | -40 | -37 | - |
| 046 | dihomo 15d PGD2 | 3.30 | 361 | 299 | -40 | -37 | - |
| 047 | 6k PGE1 | 0.83 | 367 | 331 | -40 | -25 | 7 |
| 048 | 6,15 dk-,dh- PGF1α | 1.01 | 369 | 267 | -40 | -37 | 300 |
| 049 | 15k PGF1α | 1.59 | 353 | 221 | -50 | -38 | 30 |
| 050 | 15k PGF2α | 1.42 | 351 | 219 | -50 | -32 | 10 |
| 051 | 15k PGE2 | 1.56 | 349 | 235 | -30 | -26 | 7 |
| 052 | 15k PGD2 | 1.65 | 349 | 235 | -30 | -40 | 70 |
| 053 | dh PGF2α | 1.42 | 355 | 311 | -60 | -29 | 7 |
| 054 | dhk PGF2α | 1.71 | 353 | 291 | -80 | -28 | 5 |
| 055 | dhk PGE2 | 1.82 | 351 | 207 | -40 | -26 | 7 |
| 056 | dhk PGD2 | 2.10 | 351 | 207 | -40 | -26 | 3 |
| 057 | bicyclo PGE2 | 2.28 | 333 | 175 | -40 | -30 | 7 |
| 058 | 11b dhk PGF2α | 1.50 | 353 | 221 | -60 | -37 | 50 |
| 059 | 9oh PGF2α | 0.55 | 369 | 192 | -60 | -35 | 300 |
| 060 | 20oh PGF2α | 0.54 | 369 | 165 | -40 | -39 | 7 |
| 061 | 19oh PGE2 | 0.57 | 367 | 243 | -20 | -31 | 10 |
| 062 | 20oh PGE2 | 0.57 | 367 | 175 | -30 | -27 | 5 |
| 063 | 2,3 dinor 11b PGF2α | 0.83 | 325 | 227 | -30 | -22 | 5 |
| 064 | PGFM | 0.58 | 329 | 293 | -40 | -25 | 1 |
| 065 | PGEM | 0.60 | 327 | 291 | -30 | -23 | 7 |
| 066 | tetranor 12-HETE | 3.12 | 265 | 109 | -20 | -18 | 1 |
| 067 | 11b PGE2 | 1.37 | 351 | 271 | -40 | -23 | 3 |
| 068 | PGK2 | 2.02 | 349 | 249 | -40 | -31 | 10 |
| 069 | 12-HHTrE | 3.28 | 279 | 217 | -30 | -21 | 5 |
| 070 | 11-HETE | 4.00 | 319 | 167 | -40 | -23 | 1 |
| 071 | 11-HEPE | 3.58 | 317 | 215 | -40 | -20 | 3 |
| 072 | 13 HDoHE | 3.98 | 343 | 221 | -30 | -19 | 5 |
| 073 | PGA2 | 2.36 | 333 | 271 | -20 | -20 | 3 |
| 074 | PGB2 | 3.13 | 333 | 271 | -40 | -25 | 3 |
| 075 | 15d PGA2 | 3.64 | 315 | 255 | -40 | -20 | - |
| 076 | PGJ2 | 2.28 | 333 | 189 | -40 | -22 | 3 |
| 077 | 15d PGD2 | 2.83 | 333 | 271 | -30 | -22 | 3 |
| 078 | 15d PGJ2 | 3.52 | 315 | 203 | -30 | -20 | 3 |
| 079 | 5-iso PGF2αVI | 1.08 | 353 | 115 | -60 | -28 | 10 |
| 080 | 8-iso PGF2αIII | 0.96 | 353 | 193 | -40 | -33 | 7 |
| 081 | 9-HETE | 4.09 | 319 | 123 | -40 | -20 | 3 |
| 082 | 9-HEPE | 3.70 | 317 | 149 | -40 | -20 | 3 |
| 083 | 8 HDoHE | 4.11 | 343 | 109 | -40 | -20 | 5 |
| 084 | 16 HDoHE | 3.94 | 343 | 233 | -50 | -19 | 3 |
| 085 | 20 HDoHE | 3.86 | 343 | 241 | -30 | -18 | 3 |
| 086 | LTB4 | 2.82 | 335 | 195 | -45 | -23 | 3 |
| 087 | 20oh LTB4 | 0.83 | 351 | 195 | -40 | -23 | 5 |
| 088 | 20cooh LTB4 | 0.84 | 365 | 303 | -40 | -26 | 30 |
| 089 | 5,6-diHETE | 3.64 | 335 | 115 | -50 | -29 | 3 |
| 090 | 6t LTB4 | 2.90 | 335 | 195 | -45 | -22 | 3 |
| 091 | 12epi LTB4 | 2.83 | 335 | 195 | -45 | -22 | 3 |
| 092 | 6t,12epi LTB4 | 2.82 | 335 | 195 | -45 | -22 | 3 |
| 093 | 12oxo LTB4 | 2.84 | 335 | 253 | -50 | -22 | 3 |
| 094 | LTC4 | 1.80 | 624 | 272 | -50 | -33 | - |
| 095 | LTD4 | 1.47 | 495 | 177 | -50 | -29 | 5 |
| 096 | LTE4 | 2.18 | 438 | 333 | -30 | -25 | 3 |
| 097 | 11t LTC4 | 2.30 | 624 | 272 | -50 | -34 | - |
| 098 | 11t LTD4 | 1.77 | 495 | 177 | -50 | -29 | 5 |
| 099 | 11t LTE4 | 2.42 | 438 | 333 | -50 | -33 | 7 |
| 100 | 5-HETE | 4.16 | 319 | 115 | -40 | -20 | 3 |
| 101 | 5-HEPE | 3.77 | 317 | 115 | -30 | -22 | 3 |
| 102 | 7 HDoHE | 4.07 | 343 | 141 | -40 | -19 | 5 |
| 103 | 4 HDoHE | 4.28 | 343 | 101 | -60 | -18 | 3 |
| 104 | 9-HOTrE | 3.48 | 293 | 171 | -40 | -22 | 0.1 |
| 105 | 5-HETrE | 4.56 | 321 | 205 | -30 | -19 | 3 |
| 106 | 5,15-diHETE | 2.73 | 335 | 201 | -40 | -26 | 3 |
| 107 | 6R-LXA4 | 1.81 | 351 | 167 | -20 | -21 | - |
| 108 | 6S-LXA4 | 1.89 | 351 | 217 | -20 | -18 | - |
| 109 | 15R-LXA4 | 1.82 | 351 | 165 | -20 | -23 | 50 |
| 110 | LXA5 | 3.29 | 349 | 215 | -30 | -25 | - |
| 111 | LXB4 | 1.44 | 351 | 221 | -50 | -21 | 5 |
| 112 | Resolvin E1 | 0.80 | 349 | 195 | -40 | -20 | 70 |
| 113 | Resolvin D1 | 1.70 | 375 | 141 | -20 | -20 | 5 |
| 114 | Protectin D1 | 1.80 | 359 | 153 | -20 | -20 | 5 |
| 115 | 15t-Protectin D1 | 2.75 | 359 | 153 | -20 | -27 | 5 |
| 116 | 10S-Protectin D1 | 2.78 | 359 | 153 | -20 | -21 | 3 |
| 117 | 8,15-diHETE | 2.64 | 335 | 235 | -40 | -26 | 100 |
| 118 | 15-HETE | 3.91 | 319 | 175 | -40 | -19 | 3 |
| 119 | 15-HEPE | 3.61 | 317 | 219 | -40 | -18 | 3 |
| 120 | 17 HDoHE | 3.94 | 343 | 229 | -20 | -19 | 30 |
| 121 | 13-HODE | 3.89 | 295 | 195 | -60 | -23 | 3 |
| 122 | 13-HOTrE | 3.65 | 293 | 195 | -40 | -28 | 10 |
| 123 | 13-HOTrE(y) | 3.63 | 293 | 193 | -40 | -19 | 5 |
| 124 | 15-HETrE | 4.15 | 321 | 221 | -30 | -21 | 3 |
| 125 | 8-HETE | 4.05 | 319 | 155 | -40 | -19 | 5 |
| 126 | 8-HEPE | 3.74 | 317 | 155 | -50 | -29 | 7 |
| 127 | 10 HDoHE | 4.06 | 343 | 153 | -50 | -19 | 3 |
| 128 | 8-HETrE | 4.20 | 321 | 157 | -20 | -22 | 5 |
| 129 | 14,15 LTC4 | 3.29 | 624 | 272 | -30 | -32 | - |
| 130 | 14,15 LTD4 | 1.10 | 495 | 177 | -60 | -25 | 300 |
| 131 | 14,15 LTE4 | 1.61 | 438 | 333 | -40 | -22 | 7 |
| 132 | 12-HETE | 4.03 | 319 | 135 | -50 | -19 | 3 |
| 133 | 12-HEPE | 3.67 | 317 | 179 | -30 | -19 | 3 |
| 134 | 14 HDoHE | 4.00 | 343 | 205 | -30 | -18 | 5 |
| 135 | 11 HDoHE | 4.04 | 343 | 149 | -20 | -19 | 5 |
| 136 | 9-HODE | 3.88 | 295 | 171 | -60 | -23 | 3 |
| 137 | HXA3 | 3.48 | 335 | 195 | -60 | -26 | 70 |
| 138 | HXB3 | 3.46 | 335 | 183 | -40 | -21 | 5 |
| 139 | 5-oxoETE | 4.30 | 317 | 203 | -60 | -22 | 3 |
| 140 | 12-oxoETE | 4.06 | 317 | 153 | -50 | -23 | 5 |
| 141 | 15-oxoETE | 3.94 | 317 | 113 | -20 | -25 | 3 |
| 142 | 9-oxoODE | 3.97 | 293 | 185 | -50 | -28 | 3 |
| 143 | 13-oxoODE | 3.91 | 293 | 167 | -50 | -29 | 10 |
| 144 | 15 oxoEDE | 4.49 | 321 | 223 | -80 | -32 | 3 |
| 145 | 20-HETE | 3.64 | 319 | 245 | -50 | -24 | 3 |
| 146 | 19-HETE | 3.57 | 319 | 231 | -40 | -23 | 30 |
| 147 | 18-HETE | 3.67 | 319 | 261 | -60 | -20 | 3 |
| 148 | 17-HETE | 3.71 | 319 | 247 | -50 | -20 | 3 |
| 149 | 16-HETE | 3.73 | 319 | 189 | -30 | -21 | 3 |
| 150 | 18-HEPE | 3.52 | 317 | 215 | -50 | -20 | 3 |
| 151 | 5,6-EET | 4.42 | 319 | 191 | -30 | -17 | 3 |
| 152 | 8,9-EET | 4.37 | 319 | 155 | -30 | -18 | 7 |
| 153 | 11,12-EET | 4.34 | 319 | 167 | -30 | -17 | 3 |
| 154 | 14,15-EET | 4.23 | 319 | 175 | -30 | -17 | 3 |
| 155 | 14(15) EpETE | 3.99 | 317 | 207 | -30 | -19 | 3 |
| 156 | 17(18) EpETE | 3.85 | 317 | 259 | -40 | -18 | 10 |
| 157 | 16(17) EpDPE | 4.27 | 343 | 193 | -40 | -19 | 10 |
| 158 | 19(20) EpDPE | 4.16 | 343 | 241 | -50 | -18 | 5 |
| 159 | 19,20 DiHDPA | 3.20 | 361 | 229 | -40 | -22 | 5 |
| 160 | 9,10 EpOME | 4.25 | 295 | 171 | -60 | -21 | 3 |
| 161 | 12,13 EpOME | 4.23 | 295 | 195 | -50 | -23 | 3 |
| 162 | 5,6-diHETrE | 3.62 | 337 | 145 | -40 | -22 | 3 |
| 163 | 8,9-diHETrE | 3.44 | 337 | 127 | -30 | -27 | 3 |
| 164 | 11,12-diHETrE | 3.33 | 337 | 167 | -40 | -25 | 3 |
| 165 | 14,15-diHETrE | 3.19 | 337 | 207 | -30 | -24 | 3 |
| 166 | 9,10 diHOME | 3.09 | 313 | 201 | -50 | -29 | 3 |
| 167 | 12,13 diHOME | 3.04 | 313 | 183 | -50 | -29 | 3 |
| 168 | Arachidonic Acid | 4.85 | 303 | 259 | -55 | -20 | 0.1 |
| 169 | Adrenic Acid | 4.97 | 331 | 287 | -70 | -20 | 0.1 |
| 170 | EPA | 4.76 | 301 | 257 | -40 | -16 | 0.1 |
| 171 | DHA | 4.88 | 327 | 283 | -40 | -19 | 0.1 |
| 172 | 20cooh AA | 3.66 | 333 | 271 | -60 | -23 | 5 |
| 173 | 17k DPA | 4.21 | 343 | 247 | -40 | -23 | 3 |
| 174 | 2,3 dinor TXB2 | 0.84 | 341 | 137 | -20 | -31 | 50 |
| 175 | 11d-TXB2 | 1.38 | 367 | 305 | -20 | -26 | 7 |
| 176 | 2,3 dinor 8-iso PGF2a | 0.77 | 325 | 237 | -30 | -19 | 5 |
| 177 | 2,3 dinor-6k PGF1a | 4.32 | 363 | 281 | -30 | -23 | - |
| 178 | PGK1 | 2.02 | 351 | 251 | -40 | -26 | 5 |
| 179 | 8-iso PGF3a | 0.82 | 351 | 307 | -30 | -28 | 5 |
| 180 | 8-iso-15k PGF2b | 1.20 | 351 | 219 | -50 | -22 | 5 |
| 181 | 9-Nitrooleate | 4.74 | 326 | 168 | -40 | -20 | 3 |
| 182 | 10-Nitrooleate | 4.73 | 326 | 169 | -40 | -19 | 7 |
| 183 | tetranor-PGDM | 0.70 | 327 | 247 | -20 | -20 | 50 |
| 184 | 7(R) Maresin-1 | 2.75 | 359 | 177 | -30 | -22 | 30 |
Peak selectivity was demonstrated by comparison with individual MRM transitions and the retention time of each analyte, as shown in Table 2. Scheduled MRM is an improvement over traditional MRM allowing for better data collection and more analytes to be monitored in a single analysis. We found a 30 sec retention time window for each MRM pair sufficient to allow for potential small shifts in retention time. In addition to providing good sensitivity, MRM approaches are highly selective, reducing the need for extensive sample cleanup. As a proof of principle, control plasma was analyzed and we found no discernable distortions.
The MS/MS spectra of most eicosanoids show numerous product ions. Various product ions from the precursor were also listed In Table 2, and the product ions for MRM detection was marked with underline. The product ions employed here for the MRM detection were selected to yield the best discrimination from other eicosanoids that co-elute in the vicinity of the analyte and to yield the highest signal. By balancing LC retention time and product ion selection, we were able to successfully distinguish the large majority of the eicosanoids listed. For example, owing to the silimar structure, (d4) PGE2 and PGE2 have the same retention 7.1 min from LC, but the precursor and product of (d4) PGE2 and PGE2 was different, which was 355→275 and 351→271, respectively. In the other case, the precursor and product of PGE2 and PGD2 was the same, 351→271, but the retention time of PGE2 and PGD2 was different.
While a similar eicosanoid may have the same product ions, their relative intensities usually vary. The ratio of intensities of these product ions can be used to distinguish these species. In this case, multiple MRM transitions can then be analyzed, and the ratio of product ions found in the unknown can be compared with either an MS/MS library spectra or a pure standard run under the same conditions. This would aid in confirming the identity of a chromatographic peak.
3.2 Different precursor for eicosanoids
The key precursor polyunsaturated fatty acid (PUFA) for eicosanoids is arachidonic acid (AA, 20:4, n-6). Other eicosanoids and related compounds are formed from eicosapentaenoic acid (EPA, 20:5, n-3), docosahexaenoic acid (DHA, 22:6, n-3), and dihomo-γ-linolenic acid (DGLA, 20:3, n-6). Arachidonic acid is the main precursor for a wide spectrum of unique eicosanoids produced by COX, LOX, and CYP. Biochemical characterizations of EPA and DHA have generally suggested that these fatty acids are less prone to metabolism by eicosanoid pathway enzymes [29]. In the present studies, we probed for 88 eicosanoids derived from AA, 22 from DHA, 17 from EPA, 13 from DGLA, and 18 from other fatty acids, that were identified as outlined in Fig 2. The solid black lines around circles showed how many metabolites from the different PUFA. The comprehensive and simultaneous analysis of all eicosanoids is important because eicosanoids derived from different PUFA sources may have different physiological effects. For example, in neural trauma and neurodegenerative diseases, there is a dramatic rise in the levels of AA-derived eicosanoids and in contrast, DHA-derived metabolytes can prevent neuroinflammation [30].
Fig. 2. The Relationship of the Member Eicosanoid Analytes to PUFA Precursors.

3.3 Method validation
The assay was validated for analyte recovery rate, matrix effect, linearity, inter-batch and intra-batch precision and accuracy for the major analytes. This procedure was performed on five replicates and on three consecutive days. Analyte stability was performed on processed samples maintained at 4°C for 4h and 8h.
3.3.1 Recovery rate and matrix effect
(line 254)Matrix effect occurring between different matrics and adding to the complexity of a measurement and can lead to a bias. In the case of eicosanoid analysis, there is no true “blank” plasma available because of endogenous analytes present in all human plasma at various levels. Since a deuterated internal standard is either an analogous lipid metabolite or a molecule with similar chemical characteristics (chemical structure of the eicocanoids can be found in Ref. 2, 21 and 23), both lipid metabolites and internal standards will have similar ion suppression and extraction efficiencies. Thus, internal standards were used to evaluate the matrix effect and the recovery rate of the method.
The matrix effect was determined by spiking the internal standard mixture into a blank plasma sample after exaction. The results are shown in Table 3. Except for (d4) PGF2a, (d8) 5-HETE and (d8) arachidonic acid, the matrix effect of the other 23 internal standards are all (line 264) above 80 %, indicating that ion suppression affects during analysis are minimal for most analytes. For the three exceptions, it is possible that co-eluting matrix components may reduce the ion intensity of the analytes. Phospholipids, and in particular glycerophosphocholines and lysophosphatidylcholines, represent the major class of endogenous compounds causing significant matrix effects, which will suppress the signal of other co-eluting lipids. So the removal of phospholipids is preferable for analysis of other kinds of lipids in lipidomics study. Because Phospholipids are amphipathic compounds which generally consist of a phosphate-containing polar head group and one or two long hydrophobic fatty acid ester chains. So the absorption of phospholipids on the C18 column is much stronger than eicosanoids. If using C18 material, phospholipids generally can be eluted with isopropanol as mobile phase. In our experiment condition, phospholipids generally cannot be eluted from SPE column. So the interference from phospholipids may be neglected. However, lysophospholipids may be coeluted from SPE in our experiment conditions. So part of matrix effect may be caused from lysophospholipids. To minimize the matrix effect, a more efficient and selective extraction method or an efficient chromatographic separation of the analytes from matrix interferences could be developed.
Table 3. Recovery Rate and Matrix Effect of Internal Standards.
| No | Internal Standard | Recovery rate(%) | Matrix effect (%) |
|---|---|---|---|
| 1 | (d4) 6k PGF1a | 94.1 | 86.6 |
| 2 | (d4) TXB2 | 87.2 | 89.0 |
| 3 | (d4) PGF2a | 62.1 | 76.6 |
| 4 | (d4) PGE2 | 89.5 | 84.8 |
| 5 | (d4) PGD2 | 82.2 | 99.5 |
| 6 | (d4) 15d PGJ2 | 58.2 | 91.4 |
| 7 | (d4) dhk PGF2a | 94.7 | 97.3 |
| 8 | (d4) dhk PGD2 | 84.4 | 96.8 |
| 9 | (d11) 8-iso PGF2a III | 81.3 | 97.0 |
| 10 | (d4) LTB4 | 82.9 | 90.8 |
| 11 | (d8) 5-HETE | 67.9 | 68.1 |
| 12 | (d8) 12-HETE | 80.8 | 87.4 |
| 13 | (d8) 15-HETE | 82.5 | 88.4 |
| 14 | (d6) 20-HETE | 71.0 | 85.8 |
| 15 | (d4) 9-HODE | 76.2 | 82.1 |
| 16 | (d4) 13-HODE | 86.71 | 85.9 |
| 17 | (d7) 5-oxoETE | 92.8 | 85.2 |
| 18 | (d4) Resolvin E1 | 63.0 | 85.8 |
| 19 | (d11) 8,9 EET | 78.7 | 87.5 |
| 20 | (d11) 11,12 EET | 74.4 | 89.9 |
| 21 | (d11) 14,15 EET | 78.1 | 93.9 |
| 22 | (d4) 9,10 diHOME | 70.9 | 91.1 |
| 23 | (d4) 12,13 diHOME | 78.0 | 90.0 |
| 24 | (d5) LTC4 | 112.6 | 119.0 |
| 25 | (d5) LTE4 | 112.7 | 118.3 |
| 26 | (d8) Arachiconic Acid | 54.3 | 70.4 |
Absolute recovery was determined by spiking the internal standards into plasma samples before exaction, as shown in Table 3. Recovery of 21 of the internal standards was above 70%. The lower recovery for (d4) PGF2a, (d8) 5-HETE and (d8) arachidonic acid may be caused by the matrix effect noted above. The lower recovery for (d4) 14d PGJ2 and (d4) Resolvin E1 may be due to non-specific binding by the SPE absorbent. The error margin for the matrix effect and absolute recovery rate was ±10%.
3.3.2 Linear and lower limit of quantitation
The limit of quantitation for each analyte was defined as the lowest signal obtained with S/N ≥7. Based on these criteria, the LOQ of 86% of the analytes was less than 10 pg, as shown in Table 2. For target analytes, if the concentration of endogenous is much higher and separation interference is much lower, the LOQ is lower. Otherwise, the LOQ for target eicosanoids was much high.
Eicosanoid quantitation was performed by the stable isotope dilution method. For each eicosanoid to be quantified an internal standard was selected and a linear standard curve was generated where the ratio of analyte standard peak area to internal standard peak area was plotted versus the amount of analyte standard. The method offers good linearity for all analytes, the R2 value is above 0.97 and the typical standard eicosanoid curves are shown in Fig. 3.
Fig. 3. Linearity of Representative Standards.

3.3.3 Accuracy and Precision
The results of intra- and inter-day precision and accuracy testing using QC standard samples prepared in plasma as a blank matrix are summarized in Table 4. Human plasma contains a number of endogenous eicosanoids and we found that we could identify and quantify 60 endogenous metabolites in 20 ul of standard plasma.
For QC samples, the nominal amount of the QC sample was expressed as the sum of the endogenous amount and the spiked amount. In all, 147 eicosanoids from plasma can be monitored and 121 eicosanoids can be monitored and also quantified using QC samples (Supplementary Table S1, Supporting Information).
Precision (CV %) values for 95% of the analytes were within 20%, 100% of the analytes were within 30%. Accuracy (RE %) for 87% of the analytes was within 20%, 100% analytes were within 30%. These results indicate a good reproducibility for the determination of 121 analytes in human plasma using QC samples.
For the bioanalytical method validation and sample analysis, calibration standards and QC samples should ideally be prepared in the same matrix as the intended sample. This is especially true when electrospray ionization mass spectrometry is used because, as outlined above, components in the biological sample can lead to matrix effects [31]. Typically, suppression or enhancement of analyte response is accompanied by diminished precision and the accuracy of subsequent measurements [32]. Due to the endogenous presence of eicosanoids at various levels, it is difficult to prepare a calibration curve in plasma; thus, we had to use calibration curves prepared in ethanol that are devoid of interfering matrices and impurities. This may contribute to analytical difficulties in detecting and accurately quantitating some of the eicosanoid molecular species for plasma sample. For example, the matrix effect may contribute to the deviation of the measurements of arachidonic acid for spiked plasma samples. To analyze the free fatty acids themselves, we have found that GC-MS provides an optimal approach [33].
3.3.4 Stability
To test the stability of the processed samples, they were kept at 4°C in the loading tray for various times prior to injection. Post-preparative stability of resuspended extracts showed a precision (CV %) and accuracy (RE %) for 100% of the analytes were within 20% for 0, 4 and 8h of pre-loading. Therefore it is concluded that eicosanoids in resuspended plasma extracts are stable for at least 8h when stored at 4°C.
3.3.5 Application of method to other tissues
The method was also applied to detect eicosanoids in mouse and human tissues, including adipose tissues, liver tissues and muscle tissues. Complete eicosanoid profiles similar to these from plasma were obtained from sample sizes as small as 2 mg of tissue. Of course, homogenization of the tissue samples before SPE extraction may be needed. These results demonstrate that this method is broadly applicable and can be used to measure eicosanoids in various biological sources.
4. Conclusion
Eicosanoids have wide ranging physiological functions, thus it is important to be able to accurately quantify and profile them in biological materials. A targeted UPLC-MS/MS method to globally monitor and quantify eicosanoids from human plasma and other tissues was developed and validated. Based on this new separation and detection platform, 184 unique metabolites were simultaneous monitored in a single 5 min analytical run. This makes it possible to analyze a large number of samples on a routine basis, including clinical samples. Target eicosanoid profiling should help the identification of potential biomarkers and understanding their role in specific diseases. Moreover, due to the sensitivity of the method only small tissue samples (line 329) are needed (i.e., 20 uL of human plasma or 2 mg of tissue) to comprehensive analysis.
Supplementary Material
Acknowledgments
This work was supported by LIPID MAPS “Glue” Grant U54 GM069338 and R01 GM20,501, and the National Natural Science Foundation of China (21105064 and 21175092) and the program on the development of national key scientific instruments and equipment (2011YQ150072, 2011YQ15007204, 2011YQ15007207, 2011YQ15007210).
References
- 1.Funk CD. Science. 2001;294:1871. doi: 10.1126/science.294.5548.1871. [DOI] [PubMed] [Google Scholar]
- 2.Buczynski MW, Dumlao DS, Dennis EA. J Lipid Res. 2009;50:1015. doi: 10.1194/jlr.R900004-JLR200. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Jenkins CM, Cedars A, Gross RW. Cardiovascular Research. 2009;82:240. doi: 10.1093/cvr/cvn346. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Antona CR, Sundberg MI. Oncogene. 2006;25:1679. [Google Scholar]
- 5.Calder PC, Zurier RB. Curr Opin Clin Nutr Metab Care. 2001;4:115. doi: 10.1097/00075197-200103000-00006. [DOI] [PubMed] [Google Scholar]
- 6.Dennis EA, Cao J, YH, Magrioti V, Kokotos G. Chem Rev. 2011;111:6130–6185. doi: 10.1021/cr200085w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Deems R, Buczynski MW, Gentry RB, Harkewicz R, Dennis EA. Methods in Enzymology. 2007;432:59. doi: 10.1016/S0076-6879(07)32003-X. [DOI] [PubMed] [Google Scholar]
- 8.Kortz BS, Helmschrodt L, Thiery C, J, Ceglarek U. J Chromatogr B. 2012;883-884:68–75. doi: 10.1016/j.jchromb.2011.10.018. [DOI] [PubMed] [Google Scholar]
- 9.Ferreiro-Vera C, Mata-Granados JM, Priego-Capote F, Quesada-Gómez JM, Luque de Castro MD. Anal Bioanal Chem. 2011;399:1093. doi: 10.1007/s00216-010-4400-6. [DOI] [PubMed] [Google Scholar]
- 10.Hofseth LJ, Ying L. Biochim Biophys Acta. 2006;1765:74. doi: 10.1016/j.bbcan.2005.08.005. [DOI] [PubMed] [Google Scholar]
- 11.Reinke M. Am J Physiol. 1992;262:E658. doi: 10.1152/ajpendo.1992.262.5.E658. [DOI] [PubMed] [Google Scholar]
- 12.Shono F, Yokota K, Horie K, Yamamoto S, Yamashita K, Watanabe K, Miyazaki H. Anal Biochem. 1988;168:284. doi: 10.1016/0003-2697(88)90320-x. [DOI] [PubMed] [Google Scholar]
- 13.Baranowski R, Pacha K. Mini Rev Med Chem. 2002;2:135. doi: 10.2174/1389557024605465. [DOI] [PubMed] [Google Scholar]
- 14.Margalit A, Duffin KL, Isakson PC. Anal Biochem. 1996;235:73. doi: 10.1006/abio.1996.0093. [DOI] [PubMed] [Google Scholar]
- 15.Araujo P, Mengesha Z, Lucena E, Grung B. J Chromatogr A. 2013 doi: 10.1016/j.chroma.2013.10.082. online. [DOI] [PubMed] [Google Scholar]
- 16.Araujo P, Janagap S, Holen E. J Chromatogr A. 2012;1260:102. doi: 10.1016/j.chroma.2012.08.056. [DOI] [PubMed] [Google Scholar]
- 17.Komaba JJ, Matsuda D, Shibakawa K, Nakade S, Hashimoto Y, Miyataa Y, Ogawaa M. Biomed Chromatogr. 2009;23:315. doi: 10.1002/bmc.1117. [DOI] [PubMed] [Google Scholar]
- 18.Blewett AJ, Varma D, Gilles T, Libonati JR, Jansen Susan A. J Pharmaceut Biomed Anal. 2008;46:653. doi: 10.1016/j.jpba.2007.11.047. [DOI] [PubMed] [Google Scholar]
- 19.Song J, Liu XJ, Wu JJ, Meehan MJ, Blevitt JM, Dorrestein PC, Milla ME. Anal Biochem. 2013;433:181. doi: 10.1016/j.ab.2012.10.022. [DOI] [PubMed] [Google Scholar]
- 20.Strassburg K, Huijbrechts AML, Kortekaas KA, Lindeman JH, Pedersen TL, Dane A, Berger R, Brenkman A, Hankemeier T, van Duynhoven J, Kalkhoven E, Newman JW, Vreeken RJ. Anal Bioanal Chem. 2012;404:1413–1426. doi: 10.1007/s00216-012-6226-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Faouder PL, Baillif V, Spreadburyd I, Mottae JP, Roussete P, Chênec G, Guignéc C, Tercéa F, Vannerd S, Vergnollee N, Bertrand-Michel J, Dubourdeauc M, Cenace N. J Chromatogr B. 2013;932:123. doi: 10.1016/j.jchromb.2013.06.014. [DOI] [PubMed] [Google Scholar]
- 22.Shindea DD, Kima KB, Oha KS, Abdallaa N, Liub KH, Baec SK, Shona JH, Kima HS, Kima DH, Shina JG. J Chromatogr B. 2012;911:113–121. doi: 10.1016/j.jchromb.2012.11.004. [DOI] [PubMed] [Google Scholar]
- 23.Levison BS, Zhang RL, Wang ZN, Fu XM, Di Donato JA, Hazen SL. Free Radical Biology and Medicine. 2013;59:2–13. doi: 10.1016/j.freeradbiomed.2013.03.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Kortz L, Dorow JL, Becker S, Thiery J, Ceglareka U. J Chromatogr B. 2013;927:209–213. doi: 10.1016/j.jchromb.2013.03.012. [DOI] [PubMed] [Google Scholar]
- 25.Dumlao DS, Buczynski MW, Norris PC, Harkewicz R, Dennis EA. Biochim Biophy Acta. 2011;1811:7240. doi: 10.1016/j.bbalip.2011.06.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Quehenberger O, Armando AM, Brown AH, Milne SB, Myers DS, Merrill AH, Bandyopadhyay S, Jones KN, Kelly S, Shaner RL, Sullards CM, Wang E, Murphy RC, Barkley RM, Leiker TJ, Raetz CRH, Guan ZQ, Laird GM, Six DA, Russell DW, McDonald JG, Subramaniam S, Fahy E, Dennis EA. J Lip Res. 2010;51:3299. doi: 10.1194/jlr.M009449. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Quehenberger O, Dennis EA. New Engl J Med. 2013;365:1812. doi: 10.1056/NEJMra1104901. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Golovko MY, Murphy EJ. J Lipid Res. 2008;49:893. doi: 10.1194/jlr.D700030-JLR200. [DOI] [PubMed] [Google Scholar]
- 29.Wada M, De Long CJ, Hong YH, Rieke CJ, Song I, Sidhu RS, Yuan C, Warnock M, Schmaier AH, Yokoyama C, Smyth CEM, Wilson SJ, Fitz Gerald GA, Garavito RM, Sui DX, Regan JW, Smith WL. J Biol Chem. 2007;282:22254. doi: 10.1074/jbc.M703169200. [DOI] [PubMed] [Google Scholar]
- 30.Tassoni D, Kaur G, Weisinger RS, Sinclair AJ. Asia Pac J Clin Nutr. 2008;17:220. [PubMed] [Google Scholar]
- 31.Jian W, Edom R, Weng N, Zannikos P, Zhang Z, Wang ZH. J Chromatogr B. 2010;878:1687. doi: 10.1016/j.jchromb.2010.04.024. [DOI] [PubMed] [Google Scholar]
- 32.Dams R, Huestis MA, Lambert WE, Murphy CM. J Am Soc Mass Spectrom. 2003;14:1290. doi: 10.1016/S1044-0305(03)00574-9. [DOI] [PubMed] [Google Scholar]
- 33.Quehenberger O, Armando A, Dumlao D, Stephens DL, Dennis EA. Prostaglandins Leukot Essent Fatty Acids. 2008;79:123. doi: 10.1016/j.plefa.2008.09.021. [DOI] [PMC free article] [PubMed] [Google Scholar]